Mut.59.1.11.Satb.b (User Manual Insite Free)

download Mut.59.1.11.Satb.b (User Manual Insite Free)

of 32

Transcript of Mut.59.1.11.Satb.b (User Manual Insite Free)

  • 7/27/2019 Mut.59.1.11.Satb.b (User Manual Insite Free)

    1/32

  • 7/27/2019 Mut.59.1.11.Satb.b (User Manual Insite Free)

    2/32

  • 7/27/2019 Mut.59.1.11.Satb.b (User Manual Insite Free)

    3/32

    MUT.59.1.11.SATB.B

    Table of Content

    1

    GETTING STARTED 4

    1.1 Install the software 41.2 Define hardware configuration 51.3 Define software options 71.4 Define frequency bands 81.5 Define measurement parameters 111.6 Define channels and carriers 13

    1.6.1 Define channels 131.6.2 Define carriers 14

    1.7 Define channel width 141.8 Choose the attenuation mode 14

    2 STARTING A MEASUREMENT 152.1 Launch a measurement 152.2 Visualize data 15

    3 ANALYSIS/REMEASUREMENT 173.1 Analysis in peaks mode 183.2 Analysis in BCCH mode 203.3 Re-measure in UMTS mode 213.4 Additional functions for re-measurements 24

    4 INSITE Frees LEXICON 26

  • 7/27/2019 Mut.59.1.11.Satb.b (User Manual Insite Free)

    4/32

    INSITE Free User manual

    4/32

    The INSITE Free software has been developped to perform selective RF

    measurements and to analyse measurements results with the adapted

    method, depending on the nature of the signal.

    1 GETTING STARTED

    1.1 Install the software

    Before installing the software, check first the PC answers the followinghardware requirements

    Insert the software CD-ROM into the CD drive. Go to the folder andlaunch the application named setup.exe. Then, follow the instructions

    displayed on the screen. The INSITE Free software is ready to start

    When opening the software, the following main window appears:

    HARDWARE REQUIREMENTS

    PC Pentium 500 MHzCable link USB Port

    Operating system XP / WIN7

    Memory 1GB RAM

    Free space 200 MB free space on hard disc

    Spectrum analyser Serial Port, Ethernet or GPIB

    Application Microsoft Excel / OpenOffice Calc

  • 7/27/2019 Mut.59.1.11.Satb.b (User Manual Insite Free)

    5/32

  • 7/27/2019 Mut.59.1.11.Satb.b (User Manual Insite Free)

    6/32

    INSITE Free User manual

    6/32

    STEP 2: Choose UMTS scanner

    Choose the UMTS scanner on the list Choose the port communication Click on check to verify that the communication is working

    STEP 3: choose GPS

    Choose the GPS device on the list Choose the port communication Click on check to verify that the communication is working

    Antennessa supplies the probe, cable and switch and their corresponding

    calibration files for the INSITE Free software. Calibration files are ASCII files

    containing rows of data separated by tabulations.

    Example:

    0.1 0.002 0.002 0.003

    0.15 0.003 0.002 0.003

    0.2 0.003 0.004 0.003

    ...

    The first column contains the frequency in MHz.

    The next three columns contain the antenna factor of the probe or loss in dB

    of the cable for the selected way (the gain of each device is calculated using

    an affine function between the measured points).

    The switch has a somehow particular file, because it contains 7 columns for

    the lowest frequencies (below 30 MHz). The column 2, 3 and 4 correspond to

    the loss without amplifier and the three last ones to the gain of the amplifier.

    STEP 4: Choose the probe

    Click on select probe and browse to select the calibration file (.probe)STEP 5: Choose cable

    Click on select cable and browse to select the calibration file (. cable)STEP 6: Choose switch box

    Click on select switch and browse to select the calibration file (switch)

  • 7/27/2019 Mut.59.1.11.Satb.b (User Manual Insite Free)

    7/32

    INSITE Free User manual

    7/32

    Click on check to verify that the communication is working1.3 Define software options

    In the main window, go to edit and select software configuration, the

    following window will appear:

    Figure 3: Software configuration window

    The measurement results can be exported in an Excel or OpenOfficeCalc file. These results are presented compared to an exposure limit. To

    choose the applying recommandation, click in the corresponding box.

    The Custom standard recommendation permits to apply your own threshold recommandation, from a text file under the Standard

  • 7/27/2019 Mut.59.1.11.Satb.b (User Manual Insite Free)

    8/32

    INSITE Free User manual

    8/32

    directory. It contains two columns. The first one corresponds to the

    frequency in MHz. The second one corresponds to the standard level in

    V/m. The two columns are separated by one tabulation. The extension

    file is.std.

    Choose the maximum number of scan to be made per band, the type ofscan for the first measure as well as the automatic peaks selection.

    Restrict users rights Click on save and exit to validate

    Note 1: If one and only one report application (Excel / OpenOffice Cac) is

    installed on your PC, the Application Selection control is not displayed.

    1.4 Define frequency bands

    In the main window, go to the edit menu and select frequency bands. This

    window is made of three parts:

    Default band: for pre-defined band list.

    Measurement scenario: to select the bands to be measured.

    Select band information, to choose measurement parameters for each

    band.

    STEP 1: load default band list

    Select load default bands list, a pop up window will appear with aselection of files. A default list of frequency bands are available, like for

    instance the france.def file.

    Note 2: The default files are XML files. The files are usually named with the

    country name (e.g. france.def or uk.def).

    Select one .def file in the scenarii subdirectory, the following windowwill appear:

  • 7/27/2019 Mut.59.1.11.Satb.b (User Manual Insite Free)

    9/32

    INSITE Free User manual

    9/32

    Figure 4: Frequency band configuration window

    Note 3: The file path will be automatically saved and do not need to be

    specified again.

    STEP 2: Select frequency band(s) to be measured

    In the default window are listed the available frequency bands.

    Click on the wanted band(s), the following sign will appear at thebottom of the window: >>

    To select all the bands of the list, click on select all at the bottom ofthe window

    Click on the >> sign, the selected band(s) will appear in themeasurement scenario list

    STEP 3: Remove frequency band(s) to be measured

    To remove a band, select it in the measurement scenario list, thefollowing sign will appear at the bottom of the window:

  • 7/27/2019 Mut.59.1.11.Satb.b (User Manual Insite Free)

    10/32

    INSITE Free User manual

    10/32

    Click on the

  • 7/27/2019 Mut.59.1.11.Satb.b (User Manual Insite Free)

    11/32

    INSITE Free User manual

    11/32

    1.5 Define measurement parameters

    The measurement parameters can be defined in the selected band

    information window. This window becomes available, after clicking on a band

    in the measurement scenario list.

    Figure 5: Selected band information

    Three types of post measurement analysis of the acquired data are available:

    Standard (peaks) analysis: allows to select peaks and calculate thetotal field by adding the selected peaks

    BCCH Analysis: select some of the channels (this is done for a GSM orDCS band) that are BCCH (Broadcast Control Channels), re-measure

    them (if necessary) and associate them with a Trx. For each BCCH the

    field is then extrapolated by multiplying the field level by the square

    root of the Trx Number. The total field is processed by adding theselected extrapolated BCCH values

  • 7/27/2019 Mut.59.1.11.Satb.b (User Manual Insite Free)

    12/32

    INSITE Free User manual

    12/32

    UMTS Analysis: when using an UMTS scanner, the scanner will dosome measurements on some specified carriers, extract the scrambling

    codes and then extrapolate the maximum level

    Note 5: The Level for automatic peaks selection (V/m) can be modified

    only if the Standard (peaks) analysis is selected. By default the level

    corresponds to the selected recommendation in the software configuration.

    This default level can be adjusted to a level just above the noise level for an

    automatic selection of the significant peaks.

    The Detection Mode and Measurement Mode can be selected. If the Max

    Hold, Min Hold or Average measurement modes are selected, the

    Temporal Measurement check box is displayed.

    Note 6: If Temporal Measurement check box is checked, duration in

    seconds must be filled. In this case, the three axes are scanned

    successively, during the indicated time. Then, the max, min or average field is

    calculated according to the selected measurement mode.

    Note 7: If the Temporal Measurement check box is not checked, in case ofMax Hold or Min Hold measurement mode selection, the duration in seconds

    must be filled. This duration corresponds here to the duration of the scan for

    each axis. IfAverage measurement mode is selected, the number of samples

    must be filled. In the same way, this one is applied for each axis.

    Note 8: The Level for automatic peaks selection (V/m) can be modified

    only if the Standard (peaks) analysis is selected. By default the level

    corresponds to the selected recommendation in the software configuration.

    This default level can be adjusted to a level just above the noise level for an

    automatic selection of the significant peaks.

    Note 9: By default the most appropriate RBw/VBw available is chosen. Click

    in the Select RBw/VBw manually box, to choose another one.

  • 7/27/2019 Mut.59.1.11.Satb.b (User Manual Insite Free)

    13/32

    INSITE Free User manual

    13/32

    1.6 Define channels and carriers

    The channels function is available for the Standard (Peaks) and BCCH analysis

    modes.

    The carriers function is available for the Standard (Peaks) and UMTS analysis

    modes.

    1.6.1 Define channels

    In the frequency bands configuration window, go to the edit menu and

    select define channels, the following window will appear:

    Figure 6: Channels definition window

    First channel freq: to select the centre frequency of the firstchannel in the band

    First channel number: this number will be associated with the firstchannel; the next ones will be numbered by adding 1, 2, and 3 to

    this number

    Channel width : this is the bandwidth that will separate twoconsecutive channels. The chosen channel width will pre-define

    automatically the filter span used for peaks calculation.

    Number of channels: total number of channels in the band: acontrol will be done to check that no channel is outside the bands

    limits.

  • 7/27/2019 Mut.59.1.11.Satb.b (User Manual Insite Free)

    14/32

    INSITE Free User manual

    14/32

    1.6.2 Define carriers

    In the frequency bands configuration window, go to the edit menu and

    select define carriers, the following window will appear:

    Figure 7: Define carriers window

    Up to 20 carriers can be defined. In the re-measure screen, the carriers

    names will be displayed on the X axis instead of frequencies. In the peaks

    analysis mode, the peaks will be selected within the chosen carrier(s), if the

    level is above the defined threshold level.

    1.7 Define channel width

    The channel width represents the bandwidth occupied by every sender in the

    band.

    This information is used to determine the appropriate RBw and VBw and to

    check the quality of the measure. A minimum knowledge on the band is

    necessary to define its width.

    For example, it will be of 200 KHz in GSM but may rise to 8 MHz in TV.

    1.8 Choose the attenuation mode

    The following mode of attenuation are available:

    Maximum found on full scan: a first scan will be done for every session,

    the attenuation will be then defined depending on the maximum level found.

    This mode is recommended when little information on the scanned zone are

    available.

  • 7/27/2019 Mut.59.1.11.Satb.b (User Manual Insite Free)

    15/32

    INSITE Free User manual

    15/32

    Attenuation checked on selected bands: a first scan will be done on

    every band defined in the scenario; this mode is recommended when a few

    bands only are scanned and the measurements are made close to a

    transmitter (frequency and types already known).

    Constant attenuation: the attenuation is constant and selected manually.

    This mode is recommended to perform many measurements in a restricted

    area, far from any transmitter.

    2 STARTING A MEASUREMENT

    2.1 Launch a measurement

    In the main window, go to measurement and select launch measurement

    scenario.

    The software will then perform the measurement of all the bands defined in

    the scenario, starting from the lowest one.

    If the attenuation is not constant, there will be a first scan to determine the

    good value for it.

    When some bands are in the amplified zone, there will be another scan to

    check if the amplification can be used without any damage of the acquired

    data.

    2.2 Visualize data

    To visualize data, go to the file menu in the main software window and

    select open a session.

  • 7/27/2019 Mut.59.1.11.Satb.b (User Manual Insite Free)

    16/32

    INSITE Free User manual

    16/32

    Figure 8: Data visualization window for all chosen frequency

    bands

    Standard Limit visualization

    The View Standard Limit permits the visualization of a black line

    corresponding to the selectected threshold recommendation.

    Zoom in/out

    Zoom in: drag an drop rightwards Zoom out: drag and drop leftwards

    Figure 9: Zoom in Change display unit

  • 7/27/2019 Mut.59.1.11.Satb.b (User Manual Insite Free)

    17/32

    INSITE Free User manual

    17/32

    The menu at the bottom left of the window allows to change the display unit.

    The available choices are:

    - dBV/m- V/m- A/m- W/m2- mW/cm2

    Change visualization mode

    Right from the display unit bottom, a menu allows to change the type of scale

    for the frequency axis. The available choices are:

    - Linear scale: make high bands visualization easier.

    - Logarithmic scale: make low bands visualization easier.

    - Distributed: all bands occupy the same width.

    Viewing elements properties

    Keep the mouse over a band areafor a few second to view the name ofthe band

    Right click on the band cursor to view the element information andmodify the analysis mode or the curve's colour.

    Auto peaks selection: if you have checked the Automatic peaks

    selection case in Software configuration, the Auto peaks selection button

    is visible. Click on it to select the peaks automatically according to the leveldefined for the band.

    Save a session: go to the file menu and select save session as. The

    session will be automatically saved in the session directory of the program.

    The sessions are saved as XML files (.ins file). XML is a standard of file syntax

    that is inspired by HTML and permits to save structured ASCII data.

    3 ANALYSIS/REMEASUREMENT

    The analysis mode has been chosen, when defining the frequency band to be

    measured. To access the analysis window, left click on the selected band in

    the main data visualization window for all chosen frequency bands.

  • 7/27/2019 Mut.59.1.11.Satb.b (User Manual Insite Free)

    18/32

    INSITE Free User manual

    18/32

    3.1 Analysis in peaks mode

    STEP 1: Define peaks (without channels or carriers)

    A peak can be defined in relation with a defined threshold and filter span:

    The level on this point is higher than the threshold And there is no higher point in the (filter span)-wide band centred on

    the point

    Figure 10: Peaks definition

    Click on threshold definition and left click on the graph: the thresholdwill be redefined to the Y-axis value of this point. The peaks will be

    represented with blue rectangles.

    The filter span is the width of the chosen channel. By default the filterspan corresponds to the chosen bandwidth. In this window, the filter

    span can be entered manually.There is a pre-control on the number of

    points used to determine if a point is a peak. The following equation is

    used for this assessment:

    FilterSpan*NumberOfPointAnalyser/ (Fmax-Fmin)< 1.3=> No peak selection.

    If the entered filter span does not answer to this equation, a window will pop

    up, where you can enter manually a new filter span.

  • 7/27/2019 Mut.59.1.11.Satb.b (User Manual Insite Free)

    19/32

    INSITE Free User manual

    19/32

    Note 10:If you have clicked on the auto peaks selection button in the main

    panel, when the panel Peak display and selection is displayed for the first

    time, some peaks are selected automatically in relation with the level defined

    in Frequency bands configuration. They do not take into account of

    threshold line.

    STEP 2: Re-measure one peak

    In the selected peaks list or directly on the graph, left click on theselected peak

    Right click to open the re-measurement window.

    STEP 3: Re-measure all peaks

    Select all the peaks to be re-measured in the selected peakslist. Go to measurement in the menu and select remeasure all selected

    peaks.

    Figure 11: Re-measure one peak

  • 7/27/2019 Mut.59.1.11.Satb.b (User Manual Insite Free)

    20/32

    INSITE Free User manual

    20/32

    In the selected peaks list, an indication is given on wether or not the

    RBw/VBw is proper. The mention OK, "usable" or Not OK appears.

    RBW/BandWith < 0.25 => Not OK

    RBW/BandWith < 0.5 => usable

    RBW/BandWith > 0.5 => OK

    3.2 Analysis in BCCH mode

    STEP 1: select channel

    The BCCH window displays the measured peak for each channel. Tocalculate the total field, click right on the selected channel: it will be

    displayed in the all BCCH channel list.

    Figure 12: Select channel for BCCH analysis mode

    STEP 2: Re-measure one channel

    Click on the chosen column with a left click Launch the re-measurement with a right click

  • 7/27/2019 Mut.59.1.11.Satb.b (User Manual Insite Free)

    21/32

    INSITE Free User manual

    21/32

    Figure 13: Selection of channel for BCCH re-measurement mode

    To extrapolate the electric field to the maximum, you can enter the number of

    TRX for each BCCH signal.

    STEP 3: Re-measure all chosen channels

    Go to the measurement menu and select remeasure all selectedBCCH)

    3.3 Re-measure in UMTS mode

    STEP 1: Select Carrier

    Click right on the curve and validate the selection. The chosencarrier will appear in the carriers list.

  • 7/27/2019 Mut.59.1.11.Satb.b (User Manual Insite Free)

    22/32

    INSITE Free User manual

    22/32

    Figure 14: Window for re-measurement in UMTS mode

    STEP 2: re-measure one carrier

    Click left on the chosen carrier and launch the measurement with aright click.

    STEP 3: re-measure all the carriers (CPICH value)

    Go to the measurement menu and select remeasure all selectedcarriers

    STEP 4: view the scrambling code

    To view the information about the detected scrambling code, double click

    on the carrier. The scrambling code values are expressed in the unit

    selected in the UMTS processing panel.

  • 7/27/2019 Mut.59.1.11.Satb.b (User Manual Insite Free)

    23/32

    INSITE Free User manual

    23/32

    Figure 15: Scrambling code information window

    STEP 5: check the case Extrapolation FactorIf you want to extrapolate the result, check the Extrapolation Factor case

    and indicate the extrapolation factor that you want to use.

    Extrapolated Level = SQRT (Extrapolation Factor) * Detected Level.

  • 7/27/2019 Mut.59.1.11.Satb.b (User Manual Insite Free)

    24/32

    INSITE Free User manual

    24/32

    Figure 16: Extrapolated Level

    3.4 Additional functions for re-measurements

    Free scan: a free scan is available directly from the BCCH, peak and UMTS

    measurement window. Select the zone of the curve to be scanned and click on

    the free scan button. The following window will appears:

    Figure 17: Pop up window for free scan re-measurement

    The RBW/VBW can be changed at this stage manually or not. Validate to

    launch the scan.

  • 7/27/2019 Mut.59.1.11.Satb.b (User Manual Insite Free)

    25/32

  • 7/27/2019 Mut.59.1.11.Satb.b (User Manual Insite Free)

    26/32

    INSITE Free User manual

    26/32

    4 INSITE Frees LEXICONSession

    A session is a measurement on a fixed point that contains all the measured

    bands. It includes the initial measurement as well as the re-measurement(s).

    Band

    A band is composed of one measure and all the measurements for one

    frequency band (between two limits). It is associated with an analysis mode

    that will lead to a re-measurement screen.

    RBw

    The RBw (Resolution Bandwidth) represents the bandwidth of a filter that will

    be applied to the input signal. With a broader RBw, the signal will be higher,

    as it is taken on a wider bandwidth). If the RBw is to low compared to the

    channel width, the whole power in a channel may not be measured. This will

    lead to an underestimation of the field level.

    VBw

    The VBw (Video Bandwidth) is linked to the RBw. It's the bandwidth of a filter

    that may be seen as the inverse of an integration time for each point.

    The lower it is, the longer the scanning takes. The bigger it is, the more

    important are the variations.

    It is generally advised to respect this relation:

    VBw = RBw*3s

    Attenuation

    Attenuation is placed between RF Input of the analyser and the mixer to avoiddamage on the mixer. It helps to measure high power signals and prevents

    the analyser to suffer damages (even if a band is not scanned). The

    attenuation should be constant for all the measurements.

    BCCH

    A BCCH (Broadcast Control Channel) is a fixed emitting channel with a

    constant level.

  • 7/27/2019 Mut.59.1.11.Satb.b (User Manual Insite Free)

    27/32

    INSITE Free User manual

    27/32

    In GSM and DCS, it is a Downlink channel (from the base station to the

    mobile) that enables the base station to control the power emitted on traffic

    channels.

    Determining the maximum electro magnetic field level is difficult due to the

    very versatile emission levels of all the signals.

    The internationally accepted method is to identify the BCCH channels (the

    shape of their spectrum is specific, and their level is constant), get the power

    level in these channels and then extrapolate the BCCH level by multiplying the

    level by the square root of the number of Trx for this transmitter.

    The Trx number is the maximum number traffic channels that could be

    activated when the traffic is at its maximum state. It is an operator-owned

    data, which depends mainly on the population density of the area. When this

    information is unknown, the following table can be used:

    GSM DCS

    High Density area (big station, fair) 6 8

    Large city (> 400k inh.) 4 8

    Small city (> 100k inh.) 3 6

    Countryside or small town (

  • 7/27/2019 Mut.59.1.11.Satb.b (User Manual Insite Free)

    28/32

  • 7/27/2019 Mut.59.1.11.Satb.b (User Manual Insite Free)

    29/32

  • 7/27/2019 Mut.59.1.11.Satb.b (User Manual Insite Free)

    30/32

  • 7/27/2019 Mut.59.1.11.Satb.b (User Manual Insite Free)

    31/32

  • 7/27/2019 Mut.59.1.11.Satb.b (User Manual Insite Free)

    32/32