Mod13 Fluoro 20130829 v2-Printable

download Mod13 Fluoro 20130829 v2-Printable

of 54

Transcript of Mod13 Fluoro 20130829 v2-Printable

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    1/54

    FLUOROSCOPY & INTERVENTIONAL IMAGINGBUSHBERG CHAPTER 9

    RSNA & AAPM PHYSICS CURRICULUM: MODULE 13

    Rene (Dickinson) Butler, MS, DABRMedical Physicist

    University of Washington Medical Center

    Department of Radiology

    Diagnostic Physics Section

    [email protected]

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    2/54

    General Fluoroscopy:Modes of Operation & System Components

    UW and Rene Butler, MS, DABR

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    3/54

    Fluoroscopy vs Projection RadiographyPARAMETER G.I. FLUOROSCOPY RADIOGRAPHY

    kVps 60 - 120 50 - 130

    mA values 0 - 5 (continuous) 200 - 8000 - 100 (pulsed)

    X-ray Tube Focal Spot Size 0.3 - 0.6 mm 1.0 - 1.2 mm

    Exposure Duration 0.5 - 15 minutes 0.01 - 0.3 seconds

    Image Receptor Input Radiation Dose per image 0.01 - 0.15 mGy / image 4 - 10 mGy/image

    Patient skin Dose Rates 10 - 60 mGy / min 0.2 - 10 mGy / image

    Source-to- Skin Distance (SSD) 30 - 50 cm 60 - 145 cm

    Source-to-Image Receptor Distance (SID) 80 - 120 cm 100 or 182 cm

    Typical Spatial Resolution1 - 2.0 LP/mm (Image Int.)

    3 - 10 LP/mm2.5 - 3.0 LP/mm(Flat Panel)

    Image Quantum Mottle High Low

    Staff Exposure to Scattered Radiation Yes No

    c.f. AAPM/RSNA Web Module: Fluoroscopy systems. Section IV. Table1. UW and Rene Butler, MS, DABR

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    4/54

    Low(er) Exposure Rates general fluoroscopy

    Continuous typically display is 30 fps or greater

    Maximum dose rate = 10 R per min [87.3 mGy/min] (10 CFR Part 20) Pulsed fluoroscopy

    Maximum dose rate = 10 R per min [87.3 mGy/min] (10 CFR Part 20)

    High(er) Exposure Rates leads to rapid dose accumulation

    High dose rate Specially activiated fluoroscopy (enable buttons orseparate pedal) Maximum dose rate = 20 R/min [175 mGy/min] with audible signal (10 CFR Part 20)

    Not found on modern fluoroscopy suite systems, common on c-arms Road Mapping & Digital Subtraction Angiography (DSA): real-time

    subtraction of pre- and post-contrast injection images to improve theperception of low-contrast vessels

    3D rotational angiography or CT acquisitions

    FluoroscopyModes of Operation

    UW and Rene Butler, MS, DABR

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    5/54

    Typical entrance exposure rates to the patient:

    For thin patients/body parts: dose rate is roughly 1-2 R per min [8.7 to

    17 mGy per min] for thin body partsSkin injury threshold in normal mode can be reached in

    approximately 118 to 230 min

    For average patient sizes: dose rate is roughly 3-5 R per min [26 to

    44 mGy/minSkin injury threshold in normal mode can be reached in

    approximately 45 to 77 min

    For heavy patient sizes: dose rate is roughly 8-10 R per min [70 to

    87.3 mGy/minSkin injury threshold in normal mode can be reached in

    approximately 23 to 29 min

    FluoroscopyModes of Operation

    UW and Rene Butler, MS, DABR

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    6/54

    If temporal resolution is not needed (e.g. guiding a catheter from the

    femoral artery to aortic arch), use pulsed fluoro to reduce patient (and

    personnel!) dose Also reduces motion

    30 fps

    15 fps

    display

    display

    c.f. AAPM/RSNA Physics Tutorials for Residents: MS Van Lysel. Fluoroscopy: OpticalCoupling and the Video System. Radiographics. Nov 2000; 20: 1769-1786.

    FluoroscopyModes of Operation

    UW and Rene Butler, MS, DABR

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    7/54

    Continuous vs. pulsed

    Continuous 30 fps w/ 33 msec per frame @ 2 mA (0.066 mAs)

    Pulsed 30 fps w/ 10 msec per frame @ 6.6 mA (0.066 mAs)** Same exposure to patient, but less motion artifact **

    Pulsed 15 fps w 33 msec per frame @ 2 mA (0.033 mAs)

    1 second exposureContinuous Fluoro30 fps, 33 msec

    per frame

    Pulsed Fluoro15 fps, 33 msec

    per frame

    Pulsed Fluoro

    30 fps, 10 msecper frame

    2 mA

    6.6 mA

    2 mA

    FluoroscopyModes of Operation

    UW and Rene Butler, MS, DABR

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    8/54

    8.76 mGy per R

    14.7 R per min = 128.8 mGy per min

    34.6 R per min = 303.1 mGy per min

    63.1 R per min = 552.8 mGy per min

    FluoroscopyModes of Operation

    UW and Rene Butler, MS, DABR

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    9/54

    Digital Subtraction Angiography

    (DSA): real-time subtraction of

    pre- and post-contrast injectionimages to improve the

    perception of low-contrast

    vessels

    Removal of background anatomy

    and tissue

    Increased image noise

    Clinically used for diagnostic and

    therapeutic applications of vessel

    visualization throughout the entirebody

    c.f. AAPM/RSNA Physics Tutorial for Residents: Digital Fluoroscopy. Radiographics. Vol 21. March 2001.

    DSA cerebral arteriogram. a. Unsubtracted digital fluoro.b-d. Subtracted DSA images obtained at 3 progressivetime points during contrast injection.

    FluoroscopyImaging Techniques

    UW and Rene Butler, MS, DABR

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    10/54

    Road mapping

    A DSA sequence is performed and the frame w/ maximum vessel

    opacification is identified The road map mask is subtracted from subsequent live fluoro images to

    produce real-time subtracted fluoro images

    e.g.: a wire is steered by using the road map for cues on maneuvering through

    vasculature

    3D rotational angiography

    Philips Xper CT

    c.f. AAPM/RSNA Physics Tutorial for Residents: Digital Fluoroscopy. Radiographics. Vol 21. March 2001.

    FluoroscopyImaging Techniques

    UW and Rene Butler, MS, DABR

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    11/54

    What type of configuration is preferred for a urology suite?

    Why?

    Kidneys & bladder are closer to the image receptor, therefore, reducing

    focal spot blur.

    What is the negative of this room set-up?

    Because the tube is above the patient, the scatter radiation is projected

    back into the procedure room; whereas for GI fluoroscopy rooms, the

    entrance point is below the table and lead shield curtains attenuate the

    scatter radiation.

    UW and Rene Butler, MS, DABR

    Web ModulesClinical Applications

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    12/54

    Under table tube

    Gastrointestinal room

    Remote rooms

    C-arm (mobile)

    Operating rooms or floor clinics

    Orthopedic joint replacement, endoscopy, colonoscopy Positioning flexibility

    Angiography, IR rooms

    Over table tube

    Urology

    c.f. AAPM/RSNA Web Module: Fluoroscopy systems. Section III.A-E. UW and Rene Butler, MS, DABR

    Web ModulesClinical Applications

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    13/54

    Fluoroscopy System ComponentsImage Intensifier (II) vs Flat-Panel Detectors (FPD)

    Flat-panel

    detector

    c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2nd ed., p. 232. UW and Rene Butler, MS, DABR

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    14/54

    Image intensifier system components:

    II vacuum bottle (housing); input screen (x-ray

    to e-); electronic lenses; output phosphor (e-s to

    visible light) Lenses and aperature

    Optical coupling w/ accessory port

    Viewing electronic output image video or more

    commonly charged-coupled device (CCD)

    detectors

    c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2nd ed., p. 232, 239.

    Fluoroscopy System ComponentsImage Intensifier (II) vs Flat-Panel Detectors (FPD)

    UW and Rene Butler, MS, DABR

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    15/54

    Review QuestionsFor the image intensifier shown, match the following: (answers may

    be used more than once)

    A. Light photons.

    B. X-ray photons.

    C. Microwaves.

    D. Electrons.

    E. Infrared photons.

    I represents

    II represents

    III represents

    IV represents

    V represents

    B.

    B.

    A.

    D.

    A.

    UW and Rene Butler, MS, DABR

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    16/54

    In vacuum electrons (e-s) are influence by environment

    Input screen converts x-rays to e-s

    1 mm aluminum window creates vacuum

    Support layer

    Input phosphor Cesium Iodine (CsI) crystal; Converts x-rays to light

    Photocathode Layer of antimony and alkali metal; emits electrons when struck

    by light; 10-20% conversion efficiency

    Electronic lenses 25 kV 35 kV electric field between input and output

    (electronic gain)

    c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2nd ed., p. 233. UW and Rene Butler, MS, DABR

    Fluoroscopy System ComponentsImage Intensifier (II)

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    17/54

    Output phosphor converts e-s to visible light

    Zinc cadmium sulfide (ZnCdS)

    Anode thin coating of aluminum on the vacuum side of output phosphor

    Each e- interacts in the phosphor creating ~1000 light photons

    Some fraction of the output light emitted by ZnCdS phosphor is reflected at the

    glass window; known as veiling glare, which reduces image contrast

    c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2nd ed., p. 233. UW and Rene Butler, MS, DABR

    Fluoroscopy System ComponentsImage Intensifier (II)

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    18/54

    Solid-state devices

    Thin, carbon fiber protects CsI phosphor and photodiode array

    INDIRECT digital detector Input phosphor CsI converts x-rays to light photons

    X-rays interact in CsI and create ionizations; some deposited energy is

    emitted as light

    Photodiode array An array of detector elements each

    element is 200 microns (0.2 mm)

    Absorbs light and converts energy into free

    electron charge that is stored in each cellof the array

    Charge stored is proportional to the incident

    light, which is proportional to the # incident

    (absorbed) x-rayc.f. Granfors & Albagli. Scintillator-based flat-panel x-ray imaging detectors. Journal of the Society for Information Display. June 2009.

    UW and Rene Butler, MS, DABR

    Fluoroscopy System ComponentsFlat-Panel Detectors (FPD)

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    19/54

    Patient & Personnel Safety in Fluoroscopy

    UW and Rene Butler, MS, DABR

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    20/54

    Identify the technique factors and appropriate system

    features to use to optimize image quality while minimizing

    patient dose.

    1. FPD (or II) close to patient

    2. Grid in

    3. Collimate!

    4. Increase SSD (source-to-skin distance) to

    decrease ESD (entrance skin dose)

    *Note: geometry limitations withshorter personnel

    AAPM/RSNA Resident Physics Curriculum: Module 13: Fluoroscopy & Interventional Imaging

    Clinical Application

    UW and Rene Butler, MS, DABR

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    21/54

    Describe the geometric factors that affect operator dose

    during an IR procedure.

    1. Lead apron

    2. Thyroid shield

    3. Protective eyewear4. Radiation badge

    5. Adjust dose settings when possible (pulse) to

    reduce scatter

    6. Distance, when possible7. Shielding, when possible

    AAPM/RSNA Resident Physics Curriculum: Module 13: Fluoroscopy & Interventional Imaging

    Clinical Application

    UW and Rene Butler, MS, DABR

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    22/54

    Describe the geometric factors that affect operator dose during an IR procedure

    scatter geometry for the frontal and lateral tubes in a NIR suite.

    c.f. ZH Anastasian et.al. Radiation Exposure of the Anesthesiologist in theNeurointerventional Suite.Anesthesiology 2011; 114: 512-20.

    General set-up for angio/IR suites:

    Frontal tube: positioned w/ tubebelow patient

    Scatter to personnel minimizedby lead drapes

    Lateral tube: positioned so radiologistis on the same side as the FPD or II

    Scatter is projected from theskin back toward the x-ray tube Higher scatter for personnel on

    tube side of lateral tube. Usemoveable shields!

    AAPM/RSNA Resident Physics Curriculum: Module 13: Fluoroscopy & Interventional Imaging

    Clinical Application

    UW and Rene Butler, MS, DABR

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    23/54

    General set-up for angio/IR suites:

    Frontal tube: positioned w/ tubebelow patient

    Scatter to personnel minimizedby lead drapes

    Lateral tube: positioned so radiologistis on the same side as the FPD or II

    Scatter is projected from theskin back toward the x-ray tube

    Higher scatter for personnl ontube side of lateral tube. Usemoveable shields!

    AAPM/RSNA Resident Physics Curriculum: Module 13: Fluoroscopy & Interventional Imaging

    Clinical Application

    c.f. ZH Anastasian et.al. Radiation Exposure of the Anesthesiologist in theNeurointerventional Suite.Anesthesiology 2011; 114: 512-20. UW and Rene Butler, MS, DABR

    Describe the geometric factors that affect operator dose during an IR procedure

    scatter geometry for the frontal and lateral tubes in a NIR suite.

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    24/54

    c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2nd ed., p791.

    Bushberg Table 23-18. Nuclear Regulatory Commission (NRC) Regulatory Requirements:

    Maximum Permissible Dose Equivalent Limitsa

    Maximum Possible Annual Dose Limit

    Limits mSv rem

    Occupational Limits

    Total effective dose equivalent (ED) 50 5

    Total dose equivalent to any individual organ

    (except lens of eye)500 50

    Dose equivalent to the lens of the eye 150 15

    Dose equivalent to the skin or any extremity 500 50

    Minor (< 18 years old) 10% of adult limit 10% of adult limit

    Dose to an embryo/fetusb 5 in 9 months 0.5 in 9 months

    Non-occupational (Public) Limits

    Individual members of the public 1.0 per yr 0.1 per yr

    Unrestricted area 0.02 in any 1 hr c 0.002 in any 1 hrc

    a These limits are exclusive of natural background and any dose the individual has received for medical purposes; inclusive of internal committed doseequivalent & external effective dose equivalent (i.e., total effective dose equivalent).bApplies only to conceptus of a worker who declares her pregnancy. If the limit exceeds 4.5 mSv (450 mrem) at declaration, conceptus dose forremainder of gestation is not to exceed 0.5 mSv (50 mrem).c This means the dose to an area (irrespective of occupancy) shall not exceed 0.02 mSv (2 mrem) in any 1 hour. This is not a restriction ofinstantaneous dose rate to 0.02 mSv per hour (2 mrem per hour).

    Radiation Protection and Fluoroscopy/IR

    UW and Rene Butler, MS, DABR

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    25/54

    Radiation Protection and Fluoroscopy/IR Badging OUTSIDE the lead near the neck/collar Generally, no double badging (except pregnant personnel)

    Monitors total exposure, eye dose, extremity

    Total annual limit for occupation radiation worker is 50 mSv/year

    Total eye lens dose limit is 150 mSv (or 20 mSv; ICRP) per year

    Lead aprons Lead equivalent = 0.25 mm: absorbs > 90% of scatter

    Lead equivalent = 0.35 - 0.50 mm: absorbs 95 - 99% of scatter (but

    heavier, so is it feasible to wear??) Use a lead thyroid shield at all times

    Protective gloves of 0.5 mm lead of greater should be worn ifhands are going to be near but outside the primary beam Toprotect hands during fluoroscopy, it is recommended:

    Keep hands out of and away from the x-ray field when the beam is onunless physician control of invasive devices is required for patient careduring fluoroscopy

    Work on the exit-beam side of the patient whenever possible

    Monitor hand dose

    UW and Rene Butler, MS, DABR

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    26/54

    Radiation Protection and Fluoroscopy/IR Leaded glasses:

    Recommended that all full-time radiology interventionists andanesthesiologists should wear leaded eye protection

    Measured lens entrance dose is ~2.5 +/- 2.0 Sv (radiologist);therefore ~30,000 procedures to reach current 150 mSv/year limit(based on current NRC Dose Limits)

    NEW!! The ICRP recently released a statement stating lower dosethresholds for cataracts were appropriate.

    The previous ICRP threshold (and current NCRP or US threshold) of 4Gy (acute exposure) and 8 Gy (chronic exposure)

    Reduced to 0.5 Gy for acute and chronic exposures, based on recentstudies of patients and occupational workers.

    Note, even though the USA has yet to adopt this ICRP threshold, it isanticipated change.

    Because of this lower cataract threshold, the ICRP slashed theoccupational dose limit for the lens of the eye to 20 mSv in a year,averaged over a defined period of five years.

    The cumulative lens dose should not exceed 50 mSv in any single

    year.

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    27/54

    Fluoroscopy the Ten Commandments1. As patient size increases As image quality decreases, patient dose increases, personnel

    dose increases

    2. Exposure time Total fluoro time directly affects patient dose, but also distributing

    dose over the skin (can you rotate/move tube to a different

    position??)

    3. Use appropriate dose and dose-rate settings Pulsed vs continuous

    Standard FOV vs mag modes

    UW and Rene Butler, MS, DABR

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    28/54

    Fluoroscopy the Ten Commandments4. X-ray tube position Raise/lower patient away from x-ray tube to decrease ESD

    Lateral and oblique tube positions general have higher ESDs

    c.f. LK Wagner & BR Archer. Minimizing Risks from Fluoroscopic XRays: Bioeffects, Instrumentation, and Examination. 2004: 4th Edition. UW and Rene Butler, MS, DABR

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    29/54

    c.f. LK Wagner & BR Archer. Minimizing Risks from Fluoroscopic X Rays:Bioeffects, Instrumentation, and Examination. 2004: 4th Edition.

    Fluoroscopy the Ten Commandments5. Proximity of II or FPD to Patient improves image quality anddecreases radiation dose

    UW and Rene Butler, MS, DABR

    SID = 110 cm, varied SOD

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    30/54

    Fluoroscopy the Ten Commandments6. Inverse Square Law

    Reduce the dose through the use of distance if and when you can.

    Given: Exposure rate at 2 ft is 90 mR/hr

    UW and Rene Butler, MS, DABR

    9 ft24 ft21 ft2

    2 ft

    4 ft

    6 ft

    Exposure Rate at 4 ft = (90 mR/hr)(2ft/4ft) = 22.5 mR/hr2

    Exposure Rate at 6 ft = (90 mR/hr)(2ft/6ft) = 10 mR/hr2

    1 ft

    1 ft

    2 ft

    2 ft

    3 ft

    3 ft

    2

    2

    1

    12

    D

    DEE

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    31/54

    Fluoroscopy the Ten Commandments Magnification

    Electronic mag generally higher doses when using mag modes (though

    may not always be

    Geometric mag increase distance between patient and II; typically

    increases dose by the square of the magnification

    Grid remove grid for thin patients or if the image contrast is not

    affected by the scatter

    Collimation!!

    Personnel Safety use time, distance and shielding to your

    advantage whenever possible; always wear lead aprons, use badges

    to monitor individual dose

    UW and Rene Butler, MS, DABR

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    32/54

    Skin Injury Case Reports & Radiation Dose

    UW and Rene Butler, MS, DABR

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    33/54

    Skin Injuries Case Reports

    This is a 49-yr old woman with 8-yr history of refractory supraventricular tachycardia. The patient was exposed to about

    20 min of fluoroscopy with her elbow about 20-25 cm from the x-ray source. The circular port of the x-ray systemdefined the sharply demarcated border of the injury.

    Fig.8b show sharply demarcated erythema above right elbow at 3 weeks after RF cardiac cath ablation.

    Fig. 8c shows tissue necrosis 5 months after procedure.

    Fig. 8d shows injury with humerus visible about 6.5 months after procedure.

    Fig. 8e shows the surgical flap about 10 months after fluoroscopy procedure.

    Fig 8bFig 8c

    Fig 8dFig 8e

    UW and Rene Butler, MS, DABR

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    34/54

    Skin Injuries Case Reports

    ref: ICRP Publication 85, Case 1 (photographs courtesy of T. Shope).

    (a) The patients back 68 weeks after multiple coronary angiography and angioplasty procedures.

    (b) The injury approximately 1621 weeks after the procedures. A small, ulcerated area is present.

    (c) The injury approximately 1821 months after the procedures. Tissue necrosis is evident.

    (d) Close-up photograph of the lesion shown in (c).

    UW and Rene Butler, MS, DABR

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    35/54

    Skin Injuries Case Reports

    Radiation injury in a 60-year-old woman subsequent to

    successful neurointerventional procedure for the treatment

    of acute stroke. Estimated fluoroscopy time was more than

    70 minutes; 43 imaging series were performed during

    course of the procedure. The head was not shaved. Note

    focal epilation on scalp and skin injury on neck but not on

    scalp. No dose estimates were available for this case.

    ref: Balter et al. Radiology. Feb 2010. Vol 254:2

    UW and Rene Butler, MS, DABR

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    36/54

    UW and Rene Dickinson, MS

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    37/54

    FDA Advisory In the late 1980-early 1990s, the US FDA documented

    reports of at least 40 cases of radiation-induced burns to

    patients from fluoroscopically guided procedures.

    Fluoroscopic radiation is a carcinogen. While the risk of

    cancer from fluoroscopy is usually very small, it is

    essential that the radiation be properly controlled tominimize this risk to patients, to operators and to

    personnel.

    On September 9th, 1994, the FDA issued an advisory for

    facilities that use fluoroscopy for invasive procedures.

    UW and Rene Butler, MS, DABR

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    38/54

    FDA AdvisoryRecommendations

    Appropriate credentials and training for physicians

    performing fluoroscopy Operators be trained and understand system operation,

    and implications of radiation exposure for each mode of

    operation

    Physicians be educated in assessing risks and benefits

    on a case-by-case basis for patients

    Patients be counseled regarding the symptoms and

    risks of large radiation exposures Physicians justify and limit use of high dose rate modes

    of operation

    UW and Rene Butler, MS, DABR

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    39/54

    Washington State Law

    WAC 246-225-020

    Operators shall be adequately instructed in safe operatingprocedures and shall be able to demonstrate competence

    A medical x-ray machine operator shall be licensed,certified or registered by the department as either: a licensed health care practitioner

    a certified diagnostic or therapeutic RT

    a registered x-ray technician

    Nurses or PAs need training if asked to operate x-rayequipment

    Physician is ultimately responsible for assuring that the x-rays are safely and properly applied and that appropriateradiation protection measures are followed

    UW and Rene Butler, MS, DABR

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    40/54

    Image Quality (Dose) in Fluoroscopy

    UW and Rene Butler, MS, DABR

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    41/54

    Review QuestionsA 9-in. multi-mode image intensifier (II) is switched to the 6-in.

    mode. As a result, the image will be ________ , and the

    automatic brightness control system (ABC) will _________ the

    exposure to the II and the patient.

    A. magnified, decrease

    B. magnified, increase

    C. minified, increase

    D. magnified, not change

    E. minified, decrease

    UW and Rene Butler, MS, DABR

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    42/54

    Small FS general fluoroscopy (minimize blurring)

    Large FS digital spots only (tube loading)

    In general, the spatial resolution of the I.I. alone is 3.5-6.0

    LP/mm

    Smaller structures are minified less (spread over a larger portion ofthe output phosphor), this enlargement of the displayed image

    improves the limiting resolution of the imaging system

    The typical spatial resolution of most current FPD image

    receptors is about 2.5 3.0 LP/mm for all FOVs.

    UW and Rene Butler, MS, DABR

    Web ModulesClinical Applications Spatial Resolution

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    43/54

    With regard to a flat panel, what is binning?

    What is the advantage?

    What is the disadvantage?

    Example:

    8 x 8 matrix; 10 photons per pixel

    4 x 4 matrix; 160 photons per pixel

    10 10 10 10 10 10 10 10

    10 10 10 10 10 10 10 10

    10 10 10 10 10 10 10 10

    10 10 10 10 10 10 10 10

    10 10 10 10 10 10 10 10

    10 10 10 10 10 10 10 10

    10 10 10 10 10 10 10 10

    10 10 10 10 10 10 10 10

    160 160

    160 160

    Less Quantum Mottle, the patient radiation dose can be

    significantly reduced while maintaining the same image noise.

    However, binning does reduce the spatial resolution of the image.

    Binning is especially useful for a large FOV where there would be

    too many pixels in the image

    = sqrt(10) = 3.2SNR = N/ = sqrt(N) = 3.2

    = sqrt(160) = 12.6

    SNR = 12.6

    UW and Rene Butler, MS, DABR

    Web ModulesClinical Applications Spatial Resolution (and noise)

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    44/54

    What percent of dose is a single fluoroscopy image relative

    to a general radiography image?

    About 1% (typically 450 to 1800 images per minute of fluoroscopy)

    0.01-0.15 mGy per image (fluoro)4-10 mGy per image (radiography)

    UW and Rene Butler, MS, DABR

    Web ModulesClinical Applications Radiation Dose

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    45/54

    Spatial resolution

    System detector limitations FOV, matrix, DELs, video capabilities,

    binning

    e.g.: FPD detector element sizes

    GI studies @ 2.5-3 lp per mm

    Using mag, the resolution improves to @ 3.5-6 lp per mm

    Television systems limit resolution to about 1-2 lp per mm for GI systems and 2-4 lp per

    mm for angio systems

    Focal spot size and geometry keep patient adjacent to detector!! This

    reduces focal spot blur

    Motion, temporal factors affecting image blur

    In general, pulsed fluoro reduces motion blur improve resolution

    UW and Rene Butler, MS, DABR

    FluoroscopyFactors Affecting Image Quality

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    46/54

    Contrast

    Scattered x-rays (grid), veiling glare (II only)

    kVp and filtration if the average (effective) energy of the x-rays is

    increased, then contrast decreases

    Collimation decreases scatter contribution

    Radiation dose and noise increasing the mA, decreases the noise and

    therefore improved contrast

    Image processing smoothing algorithms and frame averaging reducesimage noise, which improves contrast; edge enhancement algorithms

    increase image noise, therefore contrast degrades

    Contrast media (iodine, barium, or air) enhances contrast of anatomical

    structures

    UW and Rene Butler, MS, DABR

    FluoroscopyFactors Affecting Image Quality

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    47/54

    Image Quality: II vs FPDIMAGE QUALITY II Systems FPD Systems

    If FOV is reduced (mag mode) improved spatial resolutionImproves or stays the same (no

    geometric minification for FPD)

    Binning n/areduces spatial resolution,improved SNR by decreasing

    mottle

    Focal spot size use small FS to decrease focal spot blur, except for digital spots

    Limiting resolutionII resolution (3.5-6 LP/mm), butsystem resolution is limited by

    television system (1-4 LP/mm)

    limited by DEL (2.5-3 LP/mm)

    Dynamic range limited generally, no limitation

    Contrastincreasing the kVp and/or filtration increases the average energyof the x-rays and the number of Compton scatter events, therefore

    image contrast is degraded

    Collimationlimits number of scattered photons and extraneous irradiation of

    tissue

    c.f. AAPM/RSNA Web Module: Fluoroscopy systems. Section VIII: A-F. UW and Rene Butler, MS, DABR

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    48/54

    Goal is to keep the # photons constant maintains SNR regardless of

    patient thickness (or attenuation)

    c.f. AAPM/RSNA Web Module: Fluoroscopy systems. Section X.A-7.

    Programming the generator:

    If generator responds byincrease kV for thicker (moreattenuating) regions, thencontrast is compromised butdose is lower

    For procedures where contrastis critical (e.g. angiography) thegenerator programmed toincrease mA first. Preserves

    contrast, but at the cost ofincreased dose to patient

    UW and Rene Butler, MS, DABR

    FluoroscopyFactors affecting Radiation Dose (and Image Quality)

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    49/54

    ABC = automatic brightness control

    higher kVps on the x-ray tube

    more x-ray tube current (higher mA) opening the aperture to increase the brightness

    longer x-ray pulse width

    less x-ray beam filtration

    or some combination of these

    factors

    UW and Rene Butler, MS, DABR

    FluoroscopyFactors affecting Radiation Dose (and Image Quality)

    c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2nd ed., p. 247.

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    50/54

    ABC systems X-ray beam kVp and filtration

    Aperture smaller aperture blocks more light from output phosphor,

    and decreases dose rate (the aperture is used to balance an acceptable

    amount of noise w/ an acceptable level of patient dose rate) Pulse vs continuous fluoro

    Conversion gain as an II ages, the amount of light produced in the

    input phosphor decreases and the conversion gain decreases this

    results in GREATER RADIATION DOSE because the II is less efficient

    Geometry

    Decreased SID dose

    Image receptor close to patient

    FOV selection

    II system: smaller (mag) FOVs increase the

    dose

    FPD: may not increase x-ray output

    c.f. AAPM/RSNA Web Module: Fluoroscopy systems. Section IX.B. UW and Rene Butler, MS, DABR

    FluoroscopyFactors affecting Radiation Dose (and Image Quality)

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    51/54

    Radiation Dose: II vs FPD

    Radiation Dose II Systems FPD Systems

    If FOV is reduced (mag mode) increases the radiation dose not necessary to increasethe dose

    Increasing the SID increases the radiation dose

    Adding filtration (e.g. Cu)preferentially removes low energy x-rays, therefore decreasing

    the dose

    Smaller aperature (large F)

    increases the radiation dose(aperature balances acceptable

    amount of image noise with an

    acceptable level of patient dose)

    n/a

    Increasing kVp reduction in radiation dose (more x-rays penetrate the patient)

    Reducing the pulse rate reduction in radiation dose

    Decreasing conversion gainincreases the radiation dose (in

    aging IIs)n/a

    c.f. AAPM/RSNA Web Module: Fluoroscopy systems. Section IX: A-J. UW and Rene Butler, MS, DABR

    R i Q ti

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    52/54

    Review QuestionsAn interventional radiologist performed 5000 fluoroscopically

    guided procedures last year. His annual occupational exposure

    was reported as background. What is the most likely

    explanation?

    A. All procedures were performed remotely from the x-ray control

    room.

    B. His badge was worn under a 0.5mm lead apron.

    C. The radiologist never wore his radiation badge while working.

    D. The departments control badges were stored in the

    interventional control room.

    E. There was a persistent failure at the radiation badge company.

    UW and Rene Butler, MS, DABR

    R i Q ti

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    53/54

    Review QuestionsWhich one of the following scenarios will result in the highest skin

    dose to the patient?

    A.B.

    C.

    D.

    Short SSD plus large SID results in higher patient dose. Use of the

    grid also requires higher patient dose.

    UW and Rene Butler, MS, DABR

  • 7/21/2019 Mod13 Fluoro 20130829 v2-Printable

    54/54

    Questions?

    UW and Rene Butler, MS, DABR