MiniCapt Mobile Microbial Microbial Air Sampler and ... · The purpose of this installation and...
Transcript of MiniCapt Mobile Microbial Microbial Air Sampler and ... · The purpose of this installation and...
Document 1000021474
Revision B
Date May 2015 Page Cover
MiniCapt® Mobile Microbial Microbial
Air Sampler and Accessories Measuring what matters™ — simply
Installation Qualification (IQ) and
Operational Qualification (OQ)
HEADQUARTERS 5475 Airport Blvd Boulder, Colorado 80301 USA T: +1 303 443 7100, +1 800 238 1801 Instrument Service & Support E: +1 800 557 6363 Customer Response Center T: +1 877 475 3317 E: [email protected] GLOBAL OFFICES BRAZIL T: +55 11 5188 8227 E: [email protected]
CHINA T: +86 21 6113 3600 E: [email protected] FRANCE T: +33 (0)6 82 99 17 98 E: [email protected] GERMANY T: +49 6151 6671 632 E: [email protected] ITALY T: +39 06 9053 0130 E: [email protected]
JAPAN T: +81 3 5298 8175 E: [email protected] KOREA T: +82 31 286 5790 E: [email protected] MEXICO T: +52 55 2271 5106 E: [email protected] NORDIC T: +45 707 028 55 E: [email protected]
PUERTO RICO T: +1 787 718 9096 E: [email protected] SINGAPORE T: +65 6496 0330 E: [email protected] TAIWAN T: 886-3-5525300 Ext: 301 E: [email protected] UNITED KINGDOM T: +44 1684 581 000 E: [email protected]
Document 1000021474 Revision B Date May 2015 Page i of ii
MiniCapt® Microbial Air Sampler and Accessories IQ/OQ Manual
HEADQUARTERS 5475 Airport Blvd Boulder, Colorado 80301 USA T: +1 303 443 7100, +1 800 238 1801 Instrument Service & Support E: +1 800 557 6363 Customer Response Center T: +1 877 475 3317 E: [email protected] GLOBAL OFFICES BRAZIL T: +55 11 5188 8227 E: [email protected]
CHINA T: +86 21 6113 3600 E: [email protected] FRANCE T: +33 (0)6 82 99 17 98 E: [email protected] GERMANY T: +49 6151 6671 632 E: [email protected] ITALY T: +39 06 9053 0130 E: [email protected]
JAPAN T: +81 3 5298 8175 E: [email protected] KOREA T: +82 31 286 5790 E: [email protected] MEXICO T: +52 55 2271 5106 E: [email protected] NORDIC T: +45 707 028 55 E: [email protected]
PUERTO RICO T: +1 787 718 9096 E: [email protected] SINGAPORE T: +65 6496 0330 E: [email protected] TAIWAN T: 886-3-5525300 Ext: 301 E: [email protected] UNITED KINGDOM T: +44 1684 581 000 E: [email protected]
Document 1000021474 Revision B Date May 2015 Page ii of ii
DO NOT REPRODUCE OR DISTRIBUTE MiniCapt® Microbial Air Sampler and Accessories Installation Qualification (IQ) and Operational Qualification (OQ) Manual Part number (P/N) 1000021474 Rev B
© 2015 Particle Measuring Systems, Inc. All rights reserved.
BioCapt® and MiniCapt® are registered trademarks of Particle Measuring Systems, Inc.
All trademarks appearing in this manual are the property of their respective owners. The absence of any trademark or service mark from this manual does not constitute a waiver of Particle Measuring Systems trademark or other intellectual property rights with respect to that trademark or service mark.
Confidential document This confidential document contains proprietary information, which is protected by copyright. All rights are reserved. No part of this document may be reproduced, distributed, or transmitted in any form without the prior written consent of Particle Measuring Systems. The information contained in this document is subject to change without notice.
Quality statement The Quality Policy of Particle Measuring Systems is to strive to meet or exceed the needs and expectations of our customers and to align the activities of all employees with the common focus of customer satisfaction through continuous improvement in the quality of our products and services.
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 1 of 39
Installation Qualification and Operational
Qualification Protocol
for
MiniCapt® Mobile Microbial Air Sampler
P/N: 790045 Serial Number:
Client:
PMS Protocol
PARTICLE MEASURING SYSTEMS (PMS)
Document Prepared By:
Name:
Signature:
Date:
Global Products Manager, Microbiology /
Sterility Assurance
Validation Team Leader
Global Products Specialist - Aerosol
L. Ceresa
F. Lalla
D. Pandolfi
May, 2015
May, 2015
May, 2015
CUSTOMER
Document Approved By:
Name:
Signature:
Date:
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 2 of 39
Table of Contents
1 PURPOSE .......................................................................................................................................................3
2 DESCRIPTION ..............................................................................................................................................3
2.1 GENERAL CHARACTERISTIC ...........................................................................................................................3 2.2 SAMPLING ......................................................................................................................................................3 2.3 WARNINGS AND ALARMS ...............................................................................................................................4
3 RESPONSIBILITY ........................................................................................................................................4
3.1 CLIENT........................................................................................................................................................4 3.2 PARTICLE MEASURING SYSTEMS ..................................................................................................................4
4 SAFETY RELATED ISSUES .......................................................................................................................5
5 DOCUMENTATION PROCEDURES ............................................................................................................5
5.1 DATA PREPARATION ...............................................................................................................................5 5.2 ACCEPTANCE CRITERIA .........................................................................................................................6 5.3 DEVIATIONS – CORRECTIVE ACTION .................................................................................................6 5.4 PERSONNEL ...............................................................................................................................................6 5.5 SIGNATURES OF FINAL ACCEPTANCE ................................................................................................6
6 LIST OF VERIFICATIONS FOR INSTALLATION QUALIFICATIONS .............................................7
6.1 PRE-REQUISITES OF THE INSTALLATION QUALIFICATION ...........................................................8 6.2 VERIFICATION OF MANUFACTURER’S DOCUMENTATION .........................................................15
6.3 VERIFICATION OF THE INSTALLATION OF THE SYSTEM’S MAIN COMPONENTS ..................12
6.4 VERIFICATION OF LIGHT DISPLAYS ........................................................................................... … 15
6.5 VERIFICATION OF SOFTWARE’S VERSION ......................................................................................15
7 LIST OF VERIFICATIONS FOR OPERATIONAL QUALIFICATIONS ............................................18
7.1 PRE-REQUISITES OF THE OPERATIONAL QUALIFICATION ..........................................................19 7.2 FUNCTIONALITY TEST .........................................................................................................................21 7.3 ALARM TESTING ....................................................................................................................................25 7.4 POWER LOSS TESTING ..........................................................................................................................27 7.5 COMPUTER CONNECTIVITY TEST .....................................................................................................29 7.6 DATA DOWNLOAD .................................................................................................................................31
8 FINAL REVIEW OF PROTOCOL - PMS ................................................................................................35
9 REVIEW AND FINAL APPROVAL OF PROTOCOL - CLIENT .........................................................35
APPENDIX NO 1: LIST OF PERSONNEL INVOLVED IN THE QUALIFICATION .........................36
APPENDIX NO 2: DATA COLLECTION SHEET ...................................................................................37
APPENDIX NO 3: DEVIATIONS/CORRECTIVE ACTIONS ................................................................38
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 3 of 39
1 PURPOSE
The purpose of this installation and operational qualification protocol (IOQ) is to provide
documented evidence that the MiniCapt® Mobile Microbial Air Sampler is supplied as per
requested and will be properly installed and operate in a consistent manner.
2 DESCRIPTION
2.1 General Characteristic
The MiniCapt Mobile is a portable instrument that uses the BioCapt® Impactor sampler head
with unique design which enables efficient biological sample collection in sterile environments.
Airflow into the BioCapt Impactor is controlled by a mass flow controller, thus ensuring an
operating flow-rate of either 25L/min, 50 L/min or 100 L/min as per model specifications. This
mass flow controller ensures the fluid dynamic conditions are maintained throughout the sample,
and allows for tubing and altitude corrections.
2.2 Sampling
The MiniCapt Mobile requires programming of the following operating parameters for sampling:
Choice of sampling mode (time or volume)
o Sampling duration (time mode)
25L/min: from 1 to 400 minutes
50L/min: from 1 to 200 minutes
100L/min: from 1 to 100 minutes
o Sampling volume in liters (volume mode) from 1 to 9999 L
Start delay: time delay before sample is started 0 to 139 minutes
Number of sampling intervals: from 1 to 1000 intervals
Waiting (pause) between an interval and the next, from 0 to 139 minutes
The MiniCapt Mobile makes it possible to pre-set sampling programs that can be activated from
the touchscreen or via Modbus.
The data saved from previous samplings are stored in memory, which can then be downloaded via
the USB connection, or printed using the USB connect for a printer.
When there is not enough system memory to store data from a new sampling session the oldest
sample record is deleted and automatically overwritten.
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 4 of 39
From a software point of view, the MiniCapt Mobile is equipped with non-modifiable firmware
that provides the basic functionality.
2.3 WARNINGS AND ALARMS
The MiniCapt Mobile has operating alarms that automatically manage anomalous events that do
not compromise the sampling (warnings), could compromise the sampling quality (alarms). Both alarms and warnings are signaled if the event takes place for more than 10 seconds. They
are stored in memory by the system when triggered.
3 RESPONSIBILITY
The verification and documentation process are as follows:
3.1 CLIENT
Approve the protocol before tests are started.
Provide technical-operative assistance during the execution of the qualification protocol.
Approve the possible deviations arisen during the execution of this protocol.
Approve the corrective actions proposed and approve the resolution of the deviations.
Proceed to the review of the results and to the final approval (Section 8) at the end of the
activities.
Develop and approve the Final Qualification Report.
Archive all documentation.
3.2 PARTICLE MEASURING SYSTEMS (PMS)
Prepare the Qualification Protocol including all recommendations suggested by the Client
and proceed to its initial approval.
Submit the protocol to the Client for initial approval.
Start verification action and proceed with the protocol only after the official approval by
the Client.
Ensure that the verifications and / or tests are carried out by qualified personnel.
Ensure that the results of the tests are accurately recorded in an acceptable format in the
worksheets.
Ensure that all original data are signed and dated and that they are included in the
document.
Ensure that all instruments are calibrated beforehand with traceable references to primary
standard samples, and that a copy of the calibration certificate is included in the protocol.
Communicate any possible deviations to the client in a timely manner.
Propose corrective actions for the deviations.
Carry out the corrective actions after they have been approved by the Client and present
the results for the resolution of the deviations.
Review the results.
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 5 of 39
4 SAFETY RELATED ISSUES
All persons involved in the qualification activities must know and observe the safety-related
norms established by the Client.
Possible specific precautions, the required personal protection equipment, and the measures to be
observed in the activities related to the execution of the present protocol must be indicated in
advance by the Client’s authorized personnel.
Each operation of use, maintenance, cleaning, etc. of the instrumentation or of the system must be
carried out by or under the supervision of the Client’s specifically trained personnel.
5 DOCUMENTATION PROCEDURES
5.1 DATA PREPARATION
Each test must be carried out by qualified personnel, following the instructions detailed in the
appropriate section of the protocol.
Use the worksheets from #6.1 to #7.6 to collect the results.
Use APPENDIX No 2: Data Collection Sheet for additional comments.
All chapters of this protocol must be assembled in proper order, including the original data
collected and all other relevant documents.
This protocol, with the original data and all documents collected during the qualification phase,
must be kept in an access controlled place following the norms established by the Client.
NOTE:
The modification “after the fact” of the validation data can seriously compromise the
approval by the Regulatory Authorities.
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 6 of 39
5.2 ACCEPTANCE CRITERIA
The acceptance criteria are listed in the sections corresponding to each test. At the end of each
verification and/or test, there will be a check and an indication as to whether the results comply
with the acceptance criteria.
5.3 DEVIATIONS – CORRECTIVE ACTION
During the execution of the tests each possible deviation found, together with the corrective
action taken and the corresponding assessment of acceptability, is documented in APPENDIX
No 3: Deviations / Corrective Actions.
5.4 PERSONNEL
Use APPENDIX No 1: List of Personnel Involved in the Qualification to record the personnel
involved in this execution of this protocol.
5.5 SIGNATURES OF FINAL ACCEPTANCE
The final acceptance should be documented in Section 8 “final review of protocol – PMS” and
Section 9 “review and final approval of protocol – client”
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 7 of 39
6 LIST OF VERIFICATIONS FOR INSTALLATION QUALIFICATIONS
TITLE SECTION
Pre-requisites of the installation qualification 6.1
Verification of manufacturer’s documentation 6.2
Verification of installation of the system’s main components 6.3
Verification of the light displays 6.4
Verification of the software version 6.5
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 8 of 39
6.1 PRE-REQUISITES OF THE INSTALLATION QUALIFICATION
Objective
To provide the list of pre-requisites that can slow down the execution of this protocol if they are
not met. If the answer to some of the following questions is “NO”, ask the Client’s appropriate
personnel for authorization before proceeding.
Method of Verification
To verify the compliance of the list of pre-requisites.
Acceptance Criteria
The list of pre-requisites are satisfied.
Collection of Results
Use the worksheet # 6.1
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 9 of 39
6.1 PRE-REQUISITES OF THE INSTALLATION QUALIFICATION WORKSHEET
IQ TEST: Equipment Pre-requisites
Form #: 6.1 Equipment #: Page of
Title of Test: Equipment Pre-requisites
Purpose:
To verify the compliance of list of pre-requisites of the MiniCapt® Mobile Microbial Air
Sampler.
Method:
Verify that the protocol is approved.
Verify the availability of the manufacturer’s instrument documents and the equipment
manual.
Acceptance Criteria:
The list of pre-requisites are satisfied.
Results:
Document
Result Verified by Date
Has the present protocol of the MiniCapt
Mobile Microbial Air Sampler been approved?
Are the manufacturer’s documentation and the
equipment’s manuals available?
Rational for the Qualification:
Comments:
Executed by: Date:
Authorized by (Client): Date:
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 10 of 39
6.2 VERIFICATION OF MANUFACTURER’S DOCUMENTATION
Objective
To determine if the documentation provided by the manufacturer is proper for the instrument
and for the its use.
Verification Method
Provide a list of operating manual, maintenance manuals (if applicable), compliance certificates,
and any other type of documentation provided by the equipment manufacturer.
If the provider’s documentation is not sufficient, it must be complemented by documents prepared
by the equipment user.
Acceptance Criteria
The documentation must be adequate to describe the critical functionalities, to identify and to
characterize the main components of the equipment.
Collection of Results
Use the worksheet # 6.2
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 11 of 39
6.2: VERIFICATION OF MANUFACTURER’S DOCUMENTATION WORKSHEET
IQ TEST: Equipment Documentation
Form #: 6.2 Equipment #: Page of
Title of Test: Equipment Documentation
Purpose:
To verify the availability and location of all appropriate documentation associated with the
MiniCapt Mobile Microbial Air Sampler and its installation. Typical documents include
calibration certificate, operator manual and eventually the packing list and may include, but are
not limited to, drawings, schematics, purchase orders, warranty or guarantee documents
maintenance contracts, maintenance or service reports are discretional.
Method:
Ensure that all “vendor supplied documents” listed below have been issued to site and are
physically available.
Record the revision number and the revision date (where applicable), the location of the file
(its physical whereabouts) and sign and date the entry for each required document.
Any additional documents available or relevant should be recorded in the space available or
in the comments field.
Acceptance Criteria:
The specified “vendor supplied documents” for the unit are available
Results:
Document
Revision Date Location Verified by Date
User manual of the MiniCapt Mobile
Microbial Air Sampler P/N:
Calibration certificate: attach
certificate to this document
Comments:
Executed by: Date:
Reviewed by (Client): Date:
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 12 of 39
6.3 VERIFICATION OF THE INSTALLATION OF THE SYSTEM’S MAIN
COMPONENTS
Objective
The objective of this document is to record the inspection of the different components of the
installation.
Method of Verification
Visually inspect the equipment and its main components listed into the following check list:
Item Quantity
MiniCapt Mobile Microbial Air Sampler 1
BioCapt Impactor Sampler Head 1
Lithium Ion Battery Pack 1
Power Supply 1
Visually inspect the printer and its main components listed into the following check list (if
applicable):
Item Quantity
Thermal Printer Kit 1
Printer Datamax Microflash 1
Cable Power Supply Microflash 1
Connection Cable USB/ RJ45 to the Printer
Microflash
1
Acceptance Criteria
The equipment and its components correspond to what is recorded in the reference
documentation.
Collection of Results
Use the worksheet # 6.3
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 13 of 39
6.4 VERIFICATION OF THE INSTALLATION OF THE SYSTEM’S MAIN
COMPONENTS (CONTINUED)
IQ TEST: Equipment Inventory
Form #: 6.3 Equipment #: Page of
Title of Test: Equipment Inventory
Purpose:
This section is intended to document the inspection of the different components and the
completed installation of the MiniCapt Mobile Microbial Air Sampler and to demonstrate that
the unit complies with the Component Data Sheets and Specifications.
Method:
Enter the product code, model, the serial number, type and flow rate in the spaces below of
the MiniCapt Mobile.
Enter the details of the options also supplied (if these options have not been purchased
record ‘N/A’.
Acceptance Criteria:
The equipment has not been damaged during shipping and the specifications stated below in
the column “Results” have been met for the following different parts of the system.
Results:
Document
Result
Verified by
Date
Record the product code and model of the
MiniCapt Mobile Microbial Air Sampler
Record the serial number of the MiniCapt Mobile
(found on bottom of instrument)
Record the flow rate of the MiniCapt Mobile
Sampler type
Record the availability of the external AC to DC
power supply of the MiniCapt Mobile
Record the serial number of the #1 battery supplied
Record the serial number of the thermal printer and
the availability of external AC to DC power supply
and connecting cable
Record the serial number of the head impactor
Record the serial number of the additional head
impactor (if applicable)
Comments:
Executed by: Date:
Reviewed by (Client): Date:
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 14 of 39
6.4 VERIFICATION OF LIGHT DISPLAYS
IQ TEST: Equipment Inventory
Form #: 6.4 Equipment #: Page of
Title of Test: Equipment Start-Up Test
Purpose:
To check that the MiniCapt Mobile powers up and displays the current status.
Method:
Ensure that main adaptor and battery(s) supplied are available.
Remove the cover of the battery chamber in the bottom pf the sampler and install the
battery supplied into the battery chamber of the MiniCapt Mobile (see user manual for
details).
Insert the main adaptor cable into the main socket (see user manual for details).
Turn on the equipment power for the MiniCapt Mobile.
Acceptance Criteria:
The MiniCapt Mobile powers up. The home screen is visible. The MiniCapt Mobile and
powers down suitably.
Results:
Document
Description
Result (YES/NO)
Verified by
Date
POWER BUTTON On /off
STATUS (green
LED)
Signals the
instrument’s operating
status: on/off
SAMPLING (green
LED)
Signals the instrument
is sampling: on / off
STATUS (red LED)
of the printer
Signals the printer
status: on / recharging
STATUS (green
LED) of the printer
Signals the printer
status: on/charged
Comments:
Deviation # Description Date
Executed by: Date:
Reviewed by (Client): Date:
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 15 of 39
6.5 VERIFICATION OF SOFTWARE’S VERSION
Objective
To check that the MiniCapt Mobile Microbial Air Sampler, after power up, can display the
current status: screen displays information about the software version currently installed.
Verification Method
Execute the test procedure specified in the “method” section of form 6.5 “equipment start-up”.
Record the results as required.
Acceptance Criteria
Turn on the instrument and observe the touch screen displaying identification number of the
version of software installed.
Data Collection
Use worksheet # 6.5
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 16 of 39
6.5 VERIFICATION OF SOFTWARE’S VERSION WORKSHEET
IQ TEST: Equipment Start-up
Form #: 6.5 Equipment #: Page of
Title of Test: Equipment Start-up Test
Purpose:
To check that the MiniCapt Mobile displays the current status
Method:
Please, refer to Operational Manual section “About screen”.
Turn on the MiniCapt Mobile.
Enter the submenu followed by pressing the icon to enter into the Submenu and
then 2 times the icon.
Record the equipment and version of the instruments software.
Press to exit.
Shut down the MiniCapt Mobile
Acceptance Criteria:
The MiniCapt Mobile powers up and the home touch screen is visible.
The navigation to the about icon is accessible and the instrument information is available
to view.
It is possible to identify the software version.
The MiniCapt Mobile exits and powers down suitably.
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 17 of 39
6.5 VERIFICATION OF SOFTWARE’S VERSION (CONTINUED)
IQ TEST: Equipment Start-up
Form #: 6.5 Equipment #: Page of
Results:
Test
Result
Pass/Fail
Verified by
Date
MiniCapt Mobile
powers up. The home touch
screen is activated
The about screen can be
accessed via the touch
screen and suitable icons
The software version is
displayed on the screen
The instrument exits the
status screen
The instrument shuts down
Comments:
Executed by: Date:
Reviewed by (Client): Date:
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 18 of 39
7 LIST OF VERIFICATIONS FOR OPERATIONAL QUALIFICATIONS
PRE-REQUISITES OF THE OPERATIONAL QUALIFICATION 7.1
Functionality test 7.2
Alarm testing 7.3
Power loss testing 7.4
Local Area Network (LAN) connectivity test 7.5
Data download 7.6
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 19 of 39
7.1 PRE-REQUISITES OF THE OPERATIONAL QUALIFICATION
Objective
The objective of this section is to provide the list of pre-requisites that can slow down the
execution of this protocol if they are not met.
Acceptance Criteria
If the answer to some of the following questions is “NO”, ask the Client’s appropriate personnel
for authorization before proceeding
Data Collection
Use worksheet #7.1
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 20 of 39
7.1 PRE-REQUISITES OF OPERATIONAL QUALIFICATION WORKSHEET
OQ TEST: Pre-requisites Test
Form #: 7.1 Equipment #: Page of
ITEM Yes No Not Applicable
Has the present protocol been approved?
Has the Installation Qualification been
completed and have its results been reviewed and
approved?
Have the possible deficiencies detected during
the Installation Qualification been corrected in
such a way that they do not negate the validity of
the tests of the Operational Qualification?
Comments:
Executed by: Date:
Reviewed by (Client): Date:
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 21 of 39
7.2 FUNCTIONALITY TEST
Objective
To check that the MiniCapt Mobile Microbial Air Sampler is capable of being configured to
operate in different modes.
Verification Method
Execute the test procedure specified in the “Method” section of form 7.2, functionality test.
Record the result as required.
Acceptance Criteria
The date, time, units, sample volume, start delay, interval and pauses are all independently
configurable for the counter.
Data Collection
Use worksheet #7.2
How to Run the Sampler for One Cycle to Ensure the Date is Associated with the Sample:
Please refer to Operator Manual Table 3.2.
Touch the clock and input sampling time: 1 minute as “01:00” by touching on the value.
Delete the value ( ) and then to enter a number of minutes and seconds (mm:ss).
Press to exit.
Touch the test tube and input 25 liter volume by touching on the value.
Delete the value ( ) and then enter a numbers.
Press to exit.
To run the test alwys press:
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 22 of 39
7.2 FUNCTIONALITY TEST WORKSHEET
OQ TEST: Functionality Test
Form #: 7.2 Equipment #: Page of
Title of Test: Functionality Test
Purpose:
To check that the MiniCapt Mobile Microbial Air Sampler is capable of being configured to
operate in different modes.
Method:
DATE:
Insert the current date by change the date on the counter. Press and then .
Select the TIME AND DATE icon using the touch screen navigation change the
date to one week advanced. Save. Press to exit .
Run the sampler for one cycle to ensure the date is associated with the sample.
TIME:
Insert the current time by change the time on the counter. Press and then
Select the TIME AND DATE icon using the touch screen navigation change the
time to one hour advanced. Save.
Run the sampler for one cycle to ensure the time is associated with the sample. Press
to exit.
UNITS:
Press then and then
Change L to m3
Run the sampler for one cycle to ensure the units are associated with the sample. Return to
the units setup screen
Change m3 to L. Press to exit.
Run the sampler for one cycle to ensure the units are associated with the sample. Press
to exit
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 23 of 39
7.2 FUNCTIONALITY TEST (CONTINUED)
OQ TEST: Functionality Test
Form #: 7.2 Equipment #: Page of
START DELAY:
Sampling test with delay is possible by touching the screen where indicated the value of
“start delay”.
Change the time to 30 seconds as: “00:30” and select “done”.
Run the sampler for one cycle to ensure the sample begins after the start delay.
INTERVAL AND PAUSES:
Ensure the sample time is at least 1 minute (01:00).
Change the interval by touching the screen on the values of “Interval” where indicated by
the letter “a”.
Change the interval to 4 and select “done”.
Change the pauses by touching the screen on the values of “Pause” where indicated by the
letter “b”.
Change the pauses to 10 and select “done”.
Run the sampler for one cycle to ensure the pauses and interval function is working.
Acceptance Criteria:
The date, time, units, sample volume, start delay, interval and pauses are all independently
configurable for the sampler.
a b
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 24 of 39
7.2 FUNCTIONALITY TEST WORKSHEET (CONTINUED)
OQ TEST: Functionality Test
Form #: 7.2 Equipment #: Page of
Results:
Test
Result
Pass/Fail
Verified by
Date
Setting Time and Volume
Current date
Date changed to
MiniCapt Mobile operates after
date change
Current time
Time changed to
MiniCapt Mobile operates after
time change
Current unit:
Unit change to m3 units
Unit change back to L
MiniCapt Mobile operates after unit
change
MiniCapt Mobile operates after
requested start delay
MiniCapt Mobile operates the
number of intervals requested
MiniCapt Mobile operates pauses
as a requested time between
intervals
Comments:
Executed by: Date:
Reviewed by (Client): Date:
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 25 of 39
7.3 ALARM TESTING
Objective
To verify that the MiniCapt Mobile Microbial Air Sampler will display flow warnings and alarms
and aborts the sample in the event of a flow alarm.
Verification Method
Obstruct the flow of the instrument to create a flow warning and a flow alarm.
Acceptance Criteria
The instrument will display a flow warning when the flow is partially obstructed. The instrument
will display a flow alarm and abort the sample when the flow is completely blocked.
Data Collection
Use worksheet # 7.3
7.3 ALARM TESTING WORKSHEET
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 26 of 39
OQ TEST: Alarm Testing
Form #: 7.3 Equipment #: Page of
Title of Test: Alarm Test
Purpose:
To verify that the MiniCapt Mobile Microbial Air Sampler will display flow warnings and
alarms and aborts the sample in the event of a flow alarm.
Method:
Start a sample.
Partially obstruct the flow to create a flow warning.
After flow warning stop the sample.
Start a new sample.
Completely obstruct the flow to create a flow alarm.
Sampler should show flow alarm and stop the sample.
Acceptance criteria:
The MiniCapt Mobile screen displays an orange flow warning message and emits a beep
warning when the instrument flow is partially obstructed.
The MiniCapt Mobile screen displays a red flow alarm message, emits a beep alarm and
aborts the sample when the instrument flow is completely obstructed.
Results:
Test
Result
Pass/Fail
Verified by
Date
Screen displays warning after flow is
partially obstructed
Sound system response after flow is
partially obstructed
Screen display alarm after flow is
completely obstructed
Sound system response after flow is
completely obstructed
Comments:
Executed by: Date:
Reviewed by (Client): Date:
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 27 of 39
7.4 POWER LOSS TESTING
Objective
To verify that each MiniCapt Mobile Microbial Air Sampler shuts down during a catastrophic
power failure and after the power is restored, the instrument starts up and functions normally
without any error. Typically the instrument would be powered down and not have power
disconnected.
Verification Method
Power down the system, after a few minutes power the system back up. Record the system
behavior during a power loss.
Acceptance Criteria
Upon power restoration, the system starts up and resumes normal operation
Data Collection
Use worksheet #7.4
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 28 of 39
7.4 POWER LOSS TESTING WORKSHEET
OQ TEST: Power Loss Testing
Form #: 7.4 Equipment #: Page of
Title of Test: Power Loss Testing TEST
Purpose:
To verify that each MiniCapt Mobile Microbial Air Sampler shuts down during a catastrophic
power failure and after the power is restored, the instrument starts up and functions normally
without any errors. Typically the instrument would be powered down and not have power
disconnected.
Method:
Ensure the battery supplied with the MiniCapt Mobile is removed, connect the MiniCapt
Mobile to the main power supply using the adaptor provided, it is not required that both
power supplies be connected.
Power on the MiniCapt Mobile.
Remove the power connector from the MiniCapt Mobile.
Return the power connector to the socket on the MiniCapt Mobile.
Power on the instrument.
Acceptance criteria:
The MiniCapt Mobile screen is powered up on initial startup.
The MiniCapt Mobile screen is blank when the power cable is removed.
The MiniCapt Mobile restarts once the power is reconnected.
The system operates normally post repowering.
Results:
Test
Result
Pass/Fail
Verified by
Date
System response after power is
removed
Was the system integrity affected
due to power loss?
Comments:
Executed by: Date:
Reviewed by (Client): Date:
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 29 of 39
7.5 COMPUTER CONNECTIVITY TEST
Objective
To verify that the MiniCapt Mobile Microbial Air Sampler can be connected to the Local Area
Network (LAN) via Modbus TCP.
Verification Method
Power up the instruments, connect the MiniCapt Mobile to the computer using Modbus TCP and
establishes the correct settings.
Acceptance Criteria
The MiniCapt Mobile can be connected to the local area network via Modbus TCP and the
connection is effectively displayed on the PC screen.
Data Collection
Use worksheet #7.5
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 30 of 39
7.5 COMPUTER CONNECTIVITY TEST WORKSHEET
OQ TEST: Computer Connectivity Test
Form #: 7.5 Equipment #: Page of
Title of Test: Computer Connectivity Test TEST
Purpose:
To verify that the MiniCapt Mobile can be connected to the Local Area Network (LAN) using
Modbus TCP.
CAUTION
This test is only to prove that a connection can be made using a direct link via Modbus TCP.
Various network configurations can be established according to site IT policies, if this is the
case then the setting below will need to be altered in compliance with site policy.
Method:
Power on the MiniCapt Mobile
Touch the network button on the setup submenu screen.
Set the network parameters:
Connect MiniCapt Mobile to a computer using the ethernet port and ping the instrument to
verify the network configurations
Results:
Test
Result
Pass/Fail
Verified by
Date
MiniCapt IP Address
MiniCapt Subnet Mask Address
MiniCapt Gateway Address
Executed by: Date:
Reviewed by (Client): Date:
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 31 of 39
7.6 DATA DOWNLOAD
Objective
To verify that the data can be downloaded or printed via the USB port from the MiniCapt Mobile
Microbial Air Sampler
Verification Method
Attach a USB key and download the data and/or attach a printer and print the data.
Acceptance Criteria
The MiniCapt Mobile downloads the data to the USB key.
The MiniCapt Mobile prints the data.
The date, time, units, sample volume, start delay, interval and pauses are all independently
configurable for the sampler’s counter and/or the data is printed on the designated printer.
Data Collection
Use worksheet #7.6
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 32 of 39
7.6 DATA DOWNLOAD WORKSHEET
OQ TEST: Data Download
Form #: 7.6 Equipment #: Page of
Title of test: Data Download
Purpose:
To verify that the data can be downloaded and/or printed via the USB port from the MiniCapt
Mobile
Method for download of saved data to USB key:
Power on the MiniCapt Mobile
Select “data”
Select “report records”
Insert a USB key
Select the desired report record
Select the export key
Select desired location to save to, name the file and press the green export key
Attach USB to a computer and verify the data has been correctly downloaded.
Method for download of new sample
Power on the MiniCapt Mobile
Run a sample
At the conclusion of the sample the report record will be displayed
Insert a USB key
Select the export key
Select desired location to save to, name the file and press the green export key
Attach USB to a computer and verify the data has been correctly downloaded.
Method for printing saved data
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 33 of 39
Power on the MiniCapt Mobile
Select “data”
Plug the printer into the USB slot and ensure it is on.
Select a report record
Select the export key
Select desired location to save to, name the file and press the green export key
Report will print
Method for printing new data
Power on the MiniCapt Mobile
Run a sample
At the conclusion of the sample the report record will be displayed
Plug the printer into the USB slot and ensure it is on
Select the export key
Select desired location to save to, name the file and press the green export key
Report will print
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 34 of 39
7.6 DATA DOWNLOAD WORKSHEET (CONTINUED)
OQ TEST: Data Download
Form #: 7.6 Equipment #: Page of
Results: Test Result Pass/Fail Verified by Date
Report downloaded
via USB key saved
as a csv file
Report printed
Report printed
correspond to the
saved data
Attach report
Comments:
Executed by: Date:
Reviewed by (Client): Date:
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 35 of 39
8 FINAL REVIEW AND DELIVERY OF PROTOCOL - PMS
Name of the Executor: Title: Signature: Date:
__________ ____
9 REVIEW AND FINAL APPROVAL OF PROTOCOL - CLIENT
FINAL REVISION OF PROTOCOL
Name of the Reviewer (Client):
Title:
Signature:
Date:
_______ ________ ______ ____
FINAL APPROVAL
Protocol Approved by (Client):
Title:
Signature:
Date:
__________________________
Quality
Assurance ______
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 36 of 39
APPENDIX NO 1: List of Personnel Involved in the Qualification
LIST OF PERSONNEL INVOLVED IN THE QUALIFICATION
Name Title Signature Initials
Executed by: Date:
Approved by (Client): Date:
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 37 of 39
APPENDIX No 2: Data Collection Sheet
DATA COLLECTION SHEET
Page______ of ______
Section Name/No:
Name of Test:
Executed by: Date:
Reviewed by (Client): Date:
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 38 of 39
APPENDIX NO 3: DEVIATIONS/CORRECTIVE ACTIONS
Reference Section____________________ Dev. N° ____ of ____
DEVIATION REPORT DEV. N° ____ OF ____
Reference Section: ______________________________Page_______of_______
Description:
Compiled by: Date:
Received by: Date:
CRITICALITY OF DEVIATION
Minor □
Major □
Compiled by: Date:
Approved by: Date:
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 39 of 39
APPENDIX NO 3: DEVIATIONS/CORRECTIVE ACTIONS (CONTINUED)
CORRECTIVE ACTION
Description:
Executed by: Date:
Compiled by: Date:
Approved by: Date:
Result:
Compiled by: Date:
Approved by: Date:
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Mobile Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 1 of 18
Installation Qualification and Operational
Qualification Protocol for
MiniCapt® Mobile Microbial Air Sampler Compressed
Gas Kit
P/N: 790045 Serial Number:
Client:
PMS Protocol
PARTICLE MEASURING SYSTEMS (PMS)
Document Prepared By:
Name:
Signature:
Date:
Global Products Manager, Microbiology /
Sterility Assurance
Validation Team Leader
Global Products Specialist - Aerosol
L. Ceresa
F. Lalla
D. Pandolfi
May 2015
May 2015
May 2015
CUSTOMER
Document Approved By:
Name:
Signature:
Date:
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Mobile Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 2 of 18
Table of Contents
1 PURPOSE .......................................................................................................................................................3
2 DESCRIPTION ..............................................................................................................................................3
2.1 GENERAL CHARACTERISTIC…………………………………………………………………………...3
3 LIST OF VERIFICATIONS FOR INSTALLATION QUALIFICATIONS .............................................3
3.1 PRE-REQUISITES OF THE INSTALLATION QUALIFICATION ...........................................................4 3.2 VERIFICATION OF MANUFACTURER’S DOCUMENTATION ...........................................................6 3.3 VERIFICATION OF THE INSTALLATION OF THE SYSTEM’S MAIN COMPONENTS ....................8
4 LIST OF VERIFICATIONS FOR OPERATIONAL QUALIFICATIONS ............................................10
4.1 PRE-REQUISITES OF THE OPERATIONAL QUALIFICATION ..........................................................11
4.2 COMPRESSED GAS KIT FUNCTIONALITY TEST ..............................................................................12
5 FINAL REVIEW OF PROTOCOL - PMS ................................................................................................15
6 REVIEW AND FINAL APPROVAL OF PROTOCOL - CLIENT .........................................................15
APPENDIX NO 1: LIST OF PERSONNEL INVOLVED IN THE QUALIFICATION .........................16
APPENDIX NO 2: DATA COLLECTION SHEET ...................................................................................17
APPENDIX NO 3: DEVIATIONS/CORRECTIVE ACTIONS ................................................................18
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Mobile Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 3 of 18
1 PURPOSE
The purpose of this appendix of installation and operational qualification protocol (IOQ) is to
provide documented evidence that the MiniCapt Mobile Microbial Air Sampler Compressed Gas
Kit Accessory is supplied as per requested and will be properly installed and operated in a
consistent manner.
For all the other sections, please refer to the installation and operational qualification protocol
(IOQ) of the MiniCapt Mobile Microbial Air Sampler.
2 DESCRIPTION
2.1 General Characteristic
The Compressed Gas Kit Accessory allows for the MiniCapt to be connected to a compressed gas
line. It uses a critical orifice to decompress the incoming gas, this incoming gas (air or nitrogen)
then expands into a sampling plenum before being sampled through the same BioCapt® impactor
slit design, as found in the environmental version.
3 LIST OF VERIFICATIONS FOR INSTALLATION QUALIFICATIONS
TITLE SECTION
Pre-requisites of the installation qualification of the kit 3.1
Verification of manufacturer’s kit documentation 3.2
Verification of installation of the kit’s main components 3.3
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Mobile Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 4 of 18
3.1 PRE-REQUISITES OF THE INSTALLATION QUALIFICATION
Objective
To provide the list of pre-requisites that can slow down the execution of this protocol if they are
not met. If the answer to some of the following questions is “NO”, ask the Client’s appropriate
personnel for authorization before proceeding.
Method of Verification
To verify the compliance of the list of pre-requisites.
Acceptance Criteria
The list of pre-requisites are satisfied.
Collection of Results
Use the worksheet # 3.1
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Mobile Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 5 of 18
3.1 PRE-REQUISITES OF THE INSTALLATION QUALIFICATION WORKSHEET
IQ TEST: Equipment Pre-Requisites
Form #: 3.1 Equipment #: Page of
Title of Test: Equipment Kit Pre-Requisites
Purpose:
To verify the compliance of list of pre-requisites of the MiniCapt® Mobile Microbial Air
Sampler.
Method:
Verify that the protocol is approved .
Verify the availability of the manufacturer’s instrument documents and the equipment
manual.
Acceptance Criteria:
The list of pre-requisites are satisfied.
Results:
Document
Result Verified by Date
Has the present protocol of the MiniCapt
Mobile Microbial Air Sampler Kit been
approved?
Are the manufacturer’s documentation and the
equipment’s Kit manuals available?
Rational of the Qualification:
Comments:
Executed by: Date:
Authorized by (Client): Date:
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Mobile Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 6 of 18
3.2 VERIFICATION OF MANUFACTURER’S DOCUMENTATION
Objective
To determine if the documentation provided by the manufacturer of MiniCapt® Mobile Air
Sampler is proper for the instrument’s gas kit and for the its use.
Verification Method
Provide a list of operating manual or instruction for use (if applicable) and any other type of
documentation provided by the kit manufacturer.
If the provider’s documentation is not sufficient, it must be complemented by documents prepared
by the equipment user.
Acceptance Criteria
The documentation must be adequate to describe the critical functionalities, to identify and to
characterize the main components of the equipment.
Collection of results
Use the worksheet # 3.2
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Mobile Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 7 of 18
3.2 VERIFICATION OF MANUFACTURER’S DOCUMENTATION WORKSHEET
IQ TEST: Equipment Kit Documentation
Form #: 3.2 Equipment #: Page of
Title of Test: Equipment Documentation
Purpose:
To verify the availability and location of all appropriate documentation associated with the
MiniCapt® Mobile Microbial Air Sampler Compresses Gas Kit installation. Typical
documents include calibration certificate, operator manual and eventually the packing list and
may include, but are not limited to, drawings, schematics, purchase orders, warranty or
guarantee documents maintenance contracts, maintenance or service reports are discretional.
Method:
Ensure that all “vendor supplied documents” listed below have been issued to site and are
physically available.
Record the revision number and the revision date (where applicable), the location of the
file (its physical whereabouts) and sign and date the entry for each required document.
Any additional documents available or relevant should be recorded in the space available
or in the comments field.
Acceptance criteria:
The specified “vendor supplied documents” for the unit are available
Results:
Document
Revision Date Location Verified by Date
User manual section of the
MiniCapt Mobile Microbial Air
including the compressed gas
sampling accessory
Comments:
Executed by: Date:
Reviewed by (Client): Date:
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Mobile Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 8 of 18
3.3 VERIFICATION OF THE INSTALLATION OF THE KITS’ MAIN COMPONENTS
Objective
To document the inspection of the different kit components of the installation.
Method of Verification
Consult the available certificates and all catalogued documentation then visually the equipment
and its main components.
Acceptance Criteria
The equipment and its kit components correspond to what is recorded in the reference
documentation.
Collection of Results
Use the worksheet # 3.3
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Mobile Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 9 of 18
3.3 VERIFICATION OF INSTALLATION OF THE SYSTEM’S MAIN COMPONENTS
WORKSHEET
IQ TEST: Equipment Kit Inventory
Form #: 3.3 Equipment #: Page of
Title of Test: Equipment Kit Inventory
Purpose:
This section is intended to document the inspection of the different KIT components of the
completed installation and to demonstrate that the unit complies with the Component Data
Sheets and Specifications. Please refer to the Operational Manual for the description of the
various accessories parts.
Method:
Enter the model of the compresses gas kit.
Enter the serial number of the compresses gas kit.
Switch on the unit and ensure the status lights illuminate.
Acceptance Criteria:
The equipment kit has not been damaged during shipping and the specifications stated below
in the column “Results” have been met for the following different parts of the system.
Requirement
Result
Verified by
Date
Record the model of
compressed gas sampling
accessory
Record the serial number of the
compressed gas sampling
impactor
Comments:
Executed by: Date:
Reviewed by (Client): Date:
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Mobile Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 10 of 18
4 LIST OF VERIFICATIONS FOR OPERATIONAL QUALIFICATIONS
Pre-requisites of the operational qualification 4.1
Compress gas kit funtionality test 4.2
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Mobile Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 11 of 18
4.1 PRE-REQUISITES OF OPERATIONAL QUALIFICATION WORKSHEET
OQ TEST: Pre-Requisites Test
Form #: 4.1 Equipment #: Page of
ITEM Yes No Not Applicable
Has the present protocol been approved?
Has the Installation Qualification been
completed and have its results been reviewed and
approved?
Have the possible deficiencies detected during
the Installation Qualification been corrected in
such a way that they do not negate the validity of
the tests of the Operational Qualification?
Comments:
Executed by: Date:
Reviewed by (Client): Date:
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Mobile Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 12 of 18
4.2 COMPRESSES GAS KIT FUNCTIONALITY TEST
Objective
To verify that the MiniCapt Mobile Microbial Air Sampler can be connected to a suitable
compress air system using the Compressed Gas Sampling accessory and take a sample from the
system.
Verification Method
Attach the compressed gas sampling accessory to the MiniCapt Mobile following the installation
instructions. Power up the instrument, access the main screen and check the compressed gas
mode box. The instrument will then look to ensure the compressed gas sampling accessory is
attached and when running will require the gas regulator to be set to a suitable reading.
Acceptance Criteria
The MiniCapt Mobile can be connected to a compressed air line and take a sample.
Data Collection
Use worksheet # 4.2
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Mobile Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 13 of 18
4.2 COMPRESSED GAS SAMPLE TEST WORKSHEET
OQ TEST: Compressed Gas Sample Test
Form #: 4.2 Equipment #: Page of
Title of Test: Compressed Gas Sample Test
Purpose:
To verify that the MiniCapt® Mobile Microbial Air Sampler can be connected to a
compressed air line using the compressed gas sampling accessory and take a sample.
CAUTION
Ensure all safety measures are taken when connecting the MiniCapt and accessory to a
compressed air system.
Method:
Switch on the MiniCapt.
Install an agar plate onto the MiniCapt.
Install the dual connector plate onto the MiniCapt (this is the component part with the
BioCapt® impactor slits).
Install the diffusion chamber onto the dual connector plate.
Connect the solenoid valve tubing to the tube connector (behind the flexible cover).
Connect the solenoid cable to the electrical connection.
Connect the pressure tubing to the compressed air feed.
Select “Pressure Settings” from the Option Menu.
During this period the reading of the pressure in the diffusion chamber will be seen on the
front screen of the MiniCapt. Set the value to between 50 ÷ 200 for 25 Lpm Model and
100 ÷ 400 for 50 or 100 Lpm Model.
Exit the configuration when the pressure is within the above range.
Install an agar plate onto the MiniCapt.
Select a sample time of five minutes.
Select the compress gas accessory check box and ensure selection is active.
Press “START” to begin sampling.
Run the sample until completion.
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Mobile Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 14 of 18
4.2 COMPRESSED GAS SAMPLE TEST WORKSHEET (CONTINUED)
Results: Test Result Pass/Fail Verified by Date
MiniCapt Powers up
Compressed Gas accessory installs
correctly onto the MiniCapt
The compressed gas is connected to an
infeed of compressed air
The instrument is configured to take a 5
minute sample
The sample runs as required and finished
once the sample cycle is completed
Comments:
Executed by: Date:
Reviewed by (Client): Date:
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Mobile Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 15 of 18
5 FINAL REVIEW AND DELIVERY OF PROTOCOL - PMS
Name of the Executor: Title: Signature: Date:
__________ ____
6 REVIEW AND FINAL APPROVAL OF PROTOCOL - CLIENT
FINAL REVISION OF PROTOCOL
Name of the Reviewer (Client):
Title:
Signature:
Date:
_____ ______ ______ ____
FINAL APPROVAL
Protocol approved by (Client):
Title:
Signature:
Date:
_______________________ Quality
Assurance
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Mobile Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 16 of 18
APPENDIX NO 1: LIST OF PERSONNEL INVOLVED IN THE QUALIFICATION
LIST OF PERSONNEL INVOLVED IN THE QUALIFICATION
Name: Title: Signature: Initials:
Executed by: Date:
Approved by (Client): Date:
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Mobile Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 17 of 18
APPENDIX No 2: DATA COLLECTION SHEET
DATA COLLECTION SHEET
Page______ of _______
Section Name/No:
Name of Test:
Executed by: Date:
Reviewed by (Client): Date:
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Mobile Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 18 of 18
APPENDIX NO 3: DEVIATIONS/CORRECTIVE ACTIONS
Reference Section____________________ Dev. N° ____ of ____
DEVIATION REPORT DEV. N° ____ OF ____
Reference Section: ______________________________Page_______of_______
Description:
Compiled by: Date:
Received by: Date:
CRITICALITY OF DEVIATION
Minor □
Major □
Compiled by: Date:
Approved by: Date:
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Mobile Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 19 of 18
APPENDIX NO 3: DEVIATIONS/CORRECTIVE ACTIONS (CONTINUED)
CORRECTIVE ACTION
Description:
Executed by: Date:
Compiled by: Date:
Approved by: Date:
Result:
Compiled by: Date:
Approved by: Date:
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 1 of 19
Installation Qualification and Operational
Qualification Protocol for
MiniCapt® Mobile Microbial Air Sampler
Accessory Kits
P/N: 790045 Serial Number:
Client:
PMS Protocol
PARTICLE MEASURING SYSTEMS (PMS)
Document Prepared By:
Name:
Signature:
Date:
Global Products Manager, Microbiology /
Sterility Assurance
Validation Team Leader
Global Products Specialist - Aerosol
L. Ceresa
F. Lalla
D. Pandolfi
May, 2015
May, 2015
May, 2015
CUSTOMER
Document Approved By:
Name:
Signature:
Date:
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 2 of 19
Table of Contents
1 PURPOSE .......................................................................................................................................................3
2 DESCRIPTION ..............................................................................................................................................3
2.1 GENERAL CHARACTERISTIC……………………………………………………………………………..3
3 LIST OF VERIFICATIONS FOR INSTALLATION QUALIFICATIONS .............................................3
3.1 PRE-REQUISITES OF THE INSTALLATION QUALIFICATION ...........................................................4 3.2 VERIFICATION OF MANUFACTURER’S DOCUMENTATION ...........................................................6 3.3 VERIFICATION OF THE INSTALLATION OF THE SYSTEM’S MAIN COMPONENTS ....................8
4 LIST OF VERIFICATIONS FOR OPERATIONAL QUALIFICATIONS ............................................10
4.1 PRE-REQUISITES OF THE OPERATIONAL QUALIFICATION ..........................................................11
4.2 ACCESSORY KITS FOR REMOTE SAMPLING ...................................................................................12
5 FINAL REVIEW OF PROTOCOL - PMS ................................................................................................15
6 REVIEW AND FINAL APPROVAL OF PROTOCOL - CLIENT .........................................................15
APPENDIX NO 1: LIST OF PERSONNEL INVOLVED IN THE QUALIFICATION .........................16
APPENDIX NO 2: DATA COLLECTION SHEET ...................................................................................17
APPENDIX NO 3: DEVIATIONS/CORRECTIVE ACTIONS ................................................................18
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 3 of 19
1 PURPOSE
The purpose of this appendix of installation and operational qualification protocol (IOQ) is to
provide documented evidence that the MiniCapt® Mobile Microbial Air Sampler Accessory Kits
for remote sampling are supplied as per requested and will be properly installed and operate in a
consistent manner.
For all the other sections, please refer to the installation and operational qualification protocol
(IOQ) of the MiniCapt Mobile Microbial Air Sampler.
2 DESCRIPTION
2.1 General Characteristic
The MiniCapt Mobile Microbial Air Sampler is a portable instrument that uses the BioCapt®
Impactor sampler head with unique design which enables efficient biological sample collection in
sterile environments.
The Isolator Monitoring Kit, Remote ISP Sampling Kit and Remote Connection to the BioCapt®
Stainless Steel (SS) and BioCapt® Single-Use are add-on adapter kits that enable the MiniCapt
Mobile Microbial Air Sampler to remotely sample inside aseptic environments where placement
of the MiniCapt Mobile Microbial Air Sampler is impractical due to space constraints or other
limitations.
3 LIST OF VERIFICATIONS FOR INSTALLATION QUALIFICATIONS
TITLE SECTION
Pre-requisites of the installation qualification of the kit 3.1
Verification of manufacturer’s Kit documentation 3.2
Verification of installation of the KIT’s main components 3.3
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 4 of 19
3.1 PRE-REQUISITES OF THE INSTALLATION QUALIFICATION
Objective
To provide the list of pre-requisites that can slow down the execution of this protocol if they are
not met. If the answer to some of the following questions is “NO”, ask the Client’s appropriate
personnel for authorization before proceeding.
Method of Verification
To verify the compliance of the list of pre-requisites.
Acceptance Criteria
The list of pre-requisites are satisfied.
Collection of results
Use the worksheet # 3.1
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 5 of 19
3.1 PRE-REQUISITES OF THE INSTALLATION QUALIFICATION WORKSHEET
IQ TEST: Equipment Pre-Requisites
Form #: 6.1 Page of
Title of Test: Equipment Kit Pre-Requisites
Purpose:
To verify the compliance of list of pre-requisites of the MiniCapt® Mobile Microbial Air
Sampler kits.
Method:
Verify that the protocol is approved
Verify the availability of the manufacturer’s instrument documents and the equipment
manual
Acceptance Criteria:
The list of pre-requisites are satisfied.
Results:
Document
Result Verified by Date
Has the present protocol of the MiniCapt
Mobile Microbial Air Sampler KIT been
approved?
Are the manufacturer’s documentation and the
equipment’s KIT manuals/assembly
description available?
Rational of the Qualification:
Comments:
Executed by: Date:
Authorized by (Client): Date:
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 6 of 19
3.2 VERIFICATION OF MANUFACTURER’S DOCUMENTATION
Objective
To determine if the documentation provided by the manufacturer of MiniCapt® Mobile
Microbial Air Sampler is proper for the instrument’s kit and for the its use.
Verification Method
Provide a list of operating manual or instruction for use (if applicable) and any other type of
documentation provided by the kit manufacturer.
If the provider’s documentation is not sufficient, it must be complemented by documents prepared
by the equipment user.
Acceptance Criteria
The documentation must be adequate to describe the critical functionalities, to identify and to
characterize the main components of the equipment.
Collection of Results
Use the worksheet # 3.2
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 7 of 19
3.2 Verification of Manufacturer’s Documentation Worksheet
IQ TEST: Equipment Kit Documentation
Form #: 3.2 Equipment #: Page of
Title of Test: Equipment Documentation
Purpose:
To verify the availability and location of all appropriate documentation associated with the
MiniCapt® Mobile Microbial Air Sampler Kit installation. Typical documents include
calibration certificate, operator manual and eventually the packing list and may include, but
are not limited to, drawings, schematics, purchase orders, warranty or guarantee documents
maintenance contracts, maintenance or service reports are discretional.
Method:
Ensure that all “vendor supplied documents” listed below have been issued to site and are
physically available.
Record the revision number and the revision date (where applicable), the location of the
file (its physical whereabouts) and sign and date the entry for each required document.
Any additional documents available or relevant should be recorded in the space available
or in the comments field.
Acceptance Criteria:
The specified “vendor supplied documents” for the unit are available
Results:
Document
Revision Date Location Verified by Date
User manual section of the
MiniCapt Mobile Microbial Air
including the remote sampling
accessory
Comments:
Executed by: Date:
Reviewed by (Client): Date:
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 8 of 19
3.3 VERIFICATION OF THE INSTALLATION OF THE SYSTEM’S MAIN
COMPONENTS
Objective
To document the inspection of the different kit components of the installation.
Method of Verification
Consult the available certificates and all catalogued documentation then visually the equipment
and its main components.
Acceptance Criteria
The equipment and its kit components correspond to what is recorded in the reference
documentation.
Collection of Results
Use the worksheet # 3.3
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 9 of 19
3.4 VERIFICATION OF INSTALLATION OF THE SYSTEM’S MAIN COMPONENTS
WORKSHEET
IQ TEST: Equipment Kit Inventory
Form #: 3.3 Equipment #: Page of
Title of Test: Equipment Kit Inventory
Purpose:
This section is intended to document the inspection of the different kit components of the
completed installation and to demonstrate that the unit complies with the Component Data
Sheets and Specifications. Please refer to the Operational Manual for the description of the
various accessories parts.
Method:
Enter the model of the kit components.
Enter the serial number of the kit components.
Switch on the unit and ensure the status lights illuminate.
Acceptance Criteria:
The equipment kit has not been damaged during shipping and the specifications stated below
in the column “Results” have been met for the following different parts of the system.
Requirement
Result
Verified by
Date
Record the codes/model of
sampling accessory KIT
Comments:
Executed by: Date:
Reviewed by (Client): Date:
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 10 of 19
4 LIST OF VERIFICATIONS FOR OPERATIONAL QUALIFICATIONS
Pre-requisites of the operational qualification 4.1
Accessory kits for remote sampling 4.2
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 11 of 19
4.1 PRE-REQUISITES OF OPERATIONAL QUALIFICATION WORKSHEET
OQ TEST: Pre-Requisites Test
Form #: 4.1 Equipment #: Page of
Item Yes No Not Applicable
Has the present protocol been approved?
Has the Installation Qualification been
completed and have its results been reviewed and
approved?
Have the possible deficiencies detected during
the Installation Qualification been corrected in
such a way that they do not negate the validity of
the tests of the Operational Qualification?
Comments:
Executed by: Date:
Reviewed by (Client): Date:
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 12 of 19
4.2 ACCESSORY KITS FOR REMOTE SAMPLING TEST
Objective
To verify that the MiniCapt® Mobile Microbial Air Sampler and be connected to a suitable
accessory kit and take a sample.
Verification Method
Attach the accessory kit to the MiniCapt Mobile Microbial Air Sampler following the installation
instructions or list. Power up the instrument, access the main screen and take a sample.
Acceptance Criteria
The MiniCapt Mobile Microbial Air Sampler can be connected to an accessory kit and take a
sample.
Data Collection
Use worksheet #4.2
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 13 of 19
4.2 ACCESSORY KITS TEST WORKSHEET
OQ TEST: Accessory Kits
From # : 4.2 Equipment #: Page of
Title of Test: Accessory Kits
Purpose:
To verify that the MiniCapt® Mobile Microbial Air Sampler and be connected to a suitable
accessory kit and take a sample.
Method for the Isolator Monitoring Kit:
Switch on the MiniCapt Mobile Microbial Air Sampler.
Install the remote sampling top to the MiniCapt Mobile Microbial Air Sampler.
Attach one end of the tubing to the remote sampling top.
Attach the other end of the tubing to the isolator monitoring kit.
Install an agar plate onto the isolator monitoring kit.
Select a sample time of five minutes.
Press “START” to being sampling.
Run the sample until completion.
Method for the Remote ISP Sampling Kit
Switch on the MiniCapt Mobile Microbial Air Sampler.
Install an agar plate onto the MiniCapt Mobile Microbial Air Sampler.
Install the dual connector plate onto the MiniCapt Mobile Microbial Air Sampler (this is
the component part with the BioCapt® Impactor slits).
Install the remote sampling top to the dual connector plate.
Attach one end of the tubing to the remote sampling top.
Attach the other end of the tubing to the remote ISP.
Select a sample time of five minutes.
Press “START” to being sampling.
Run the sample until completion.
Method for the Remote Connection to the BioCapt SS Microbial Impactor
Switch on the MiniCapt Mobile Microbial Air Sampler.
Install the remote sampling top to the MiniCapt Mobile Microbial Air Sampler.
Attach one end of the tubing to the remote sampling top.
Attach the other end of the tubing to the BioCapt microbial impactor.
Install an agar plate onto the BioCapt microbial impactor.
Select a sample time of five minutes.
Press “START” to being sampling.
Run the sample until completion.
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 14 of 19
4.2 ACCESSORY KITS TEST WORKSHEET (CONTINUED)
Method for the Remote Connection to the BioCapt Single-Use Microbial Impactor
Switch on the MiniCapt Mobile Microbial Air Sampler.
Install the remote sampling top to the MiniCapt Mobile Microbial Air Sampler.
Attach one end of the tubing to the remote sampling top.
Attach the other end of the tubing to the BioCapt single use microbial impactor.
Install an agar plate onto the BioCapt single use microbial impactor.
Select a sample time of five minutes.
Press “START” to being sampling.
Run the sample until completion.
NOTE: BioCapt Single-Use Microbial Impactor can be used with 25 and 50 LPM MiniCapt
Mobile Microbial Air Sampler.
Results:
Test
Result Pass/Fail Verified by Date
MiniCapt Mobile
Microbial Air Sampler
powers up
Isolator monitoring kit
installs correctly
Remote ISP sampling kit
installs correctly
Remote connection to
BioCapt microbial
impactor installs
correctly
Remote connection to
BioCapt single use
microbial impactor
installs correctly
The instrument is
configured to take a 5
minute sample
The sample runs as
required and finished
once the sample cycle is
completed
Executed by: Date:
Reviewed by (Client): Date:
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 15 of 19
5 FINAL REVIEW AND DELIVERY OF PROTOCOL - PMS
Name of the Executor: Title: Signature: Date:
__________ ____
6 REVIEW AND FINAL APPROVAL OF PROTOCOL - CLIENT
FINAL REVISION OF PROTOCOL
Name of the Reviewer (Client):
Title:
Signature:
Date:
_____ ______ ______ ____
FINAL APPROVAL
Protocol Approved by (Client):
Title:
Signature:
Date:
___________________ Quality
Assurance
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 16 of 19
APPENDIX NO 1: LIST OF PERSONNEL INVOLVED IN THE QUALIFICATION
LIST OF PERSONNEL INVOLVED IN THE QUALIFICATION
Name: Title: Signature: Initials:
Executed by: Date:
Approved by (Client): Date:
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 17 of 19
APPENDIX NO 2: DATA COLLECTION SHEET
DATA COLLECTION SHEET
Page______ of _______
Section Name/No:
Name of Test:
Executed by: Date:
Reviewed by (Client): Date:
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 18 of 19
APPENDIX NO 3: DEVIATIONS / CORRECTIVE ACTIONS
Reference Section:____________________ Dev. N° ____ of ____
DEVIATION REPORT DEV. N° ____ OF ____
Reference Section: ______________________________Page_______of_______
Description:
Compiled by: Date:
Received by: Date:
CRITICALITY OF DEVIATION
Minor □
Major □
Compiled by: Date:
Approved by: Date:
IQ/OQ Installation Qualification and Operational Qualification MiniCapt® Microbial Air Sampler
Document: Revision: Date: Page:
1000021474 B May 2015 19 of 19
APPENDIX NO 3: DEVIATIONS / CORRECTIVE ACTIONS (CONTINUED)
CORRECTIVE ACTION
Description:
Executed by: Date:
Compiled by: Date:
Approved by: Date:
Result:
Compiled by: Date:
Approved by: Date: