Manufacturing Diving & Hyperbaric Chambers Manufacturing...

16
Monoplace Hyperbaric Oxygen Chamber Chambers O2mono Manufacturing Diving & Hyperbaric Chambers

Transcript of Manufacturing Diving & Hyperbaric Chambers Manufacturing...

  • Monoplace Hyperbaric Oxygen Chamber

    Manufacturing Diving & Hyperbaric Chambers

    O2mono

    Manufacturing Diving & Hyperbaric Chambers

  • 1

  • 2

    O2mono CHAMBER

    The Baroks O2mono Model Monoplace Hyperbaric Treatment System consists of one

    compartment, is designed to safely provide hyperbaric oxygen therapy to patients at pressure

    up to 2 bar( 3ATA). Monoplace chamber is designed according to European Standarts and is

    tested according to EN14931 requirements.

    Only quality-certified “Special Steel” with test reports is used in our Monoplace Chamber

    production. In addition; before production of the chamber, ultrasonic lamination tests and after

    welding process; x-ray, magnetic particle and ultrasonic tests (Non-destructive tests) have

    been done by professional engineering staff. All these process is controlled and reported by

    certification company. After chamber production, hydrostatic tests is done by the certification

    authority surveyors

    The chamber has cylindrical shape and one entrance door

  • 3

    We are manufacturing two type monoplace chamer : Half Acrylic (HA) and Full Acrylic (FA).

    General view of two types

    Half Acrylic (HA) O2mono Full Acrylic (FA) O2mono

  • 4

    Monoplace HBOT Chamber; are designed to treat a single person pressurized with 100% oxygen.

    Figure General O2mono Body

    Chamber consists of below mentioned parts

    Table 1: Chamber Parts

    No Part Name

    1 Cylindrical Body

    2 Torispherical Head

    3 Door

    4 Body Wall

    5 Viewport

    6 Nozulles

  • 5

    Cylindrical Shell (Body)

    The chamber length is 1000 mm and inner diameter is 900 mm.

    The square door size of the chamber is 1050 mm. X 1050 mm and it provides entrance and

    exit easily.

    Door

    Door size is 1050 mm X 1050 mm.

    And thickness is 40 mm. It has also

    locking mechanism for safety.

    Figure Door Parts

    The below table shows technical specifications about monoplace chamber system.

    TECHNICAL SPECIFICATIONS

    Working Pressure 3 ATA (2 bar)

    Hydrostatic Test Pressure 4 ATA (3 bar)

    Working Temperature 0C and 45C

    Chamber Capacity 1 Person

    Inner Diameter 900 mm.

    Chamber Length (Inside) 2,221 mm.

    Total Length 2494 mm.

    Material Pressure Steel

    Door Number 1

    Door Opening 900 mm.

    Weigth ≈ 1.5 ton

    Electrical Power Requirement 3 KW

    Electrical Connection Voltage 220 Volt, 50 Hz

    Approved By TUV

  • 6

    Lock Mechanism

    The door of the system can be opened or closed by using Lock Arm. The system cannot be

    pressurized if both lock key and door key are not closed. Camshaft provides maximum sticking

    on the door Automatic Locking Cylinder locks the door when the pressure reaches 0,1 bar.

    Viewport

    Viewports are manufactured from PMMA:( Poly Methl Meth Acrylate ) material where

    chemical and physical properties are given in ASME PVHO standard. .

  • 7

    CHAMBER EQUIPMENTS This chapter consists of complementary /supportive equipments inside/on the chambe

    Figure Chamber Equipments

    Kod Parça Adı

    1 Painting

    2 Breathing System

    3 Safety Valves

    4 Stretcher

    5 Entertaintment Monitor

    6 Temperature and Humidity Sensor

    7 Pressure Sensor

    8 Antistatic Wrıst Strap

  • 8

    Painting

    The chambers are covered with flame retardent non-toxic epoxy paint on both exterior

    and interior surfaces. The paint, due to its flame retardency property starts burning at

    high temperatures, the burning only ensuses at the centre of the heat source and is

    extinguished by itself when it is removed. Also the the fumation that ensues during

    burning is not toxic for humans. This feature also prevents any sickness that may ensue

    because of the natural fumation spread by the paint because of nominal heat.

    Breathing System

    Breathing system provides required breathing air/oxygen to patients and evacuates

    exhaust air via exhaust system

    Drager X-Plore 3300 BIBS mask is a breathing mask that is used to deliver the treatment

    gas to the patient.

    Avox Inlet Regulator Avox Outlet Regulator

  • 9

    CONTROL PANEL

    Figure O2mono Control Panel

    No Part Name

    1 Counter

    2 Door Locking Button

    3 Temperature Indicator

    4 Humidity Indicator

    5 CO2 Indicator

    6 Chamber Pressure

    7 O2 Source Pressure Manometer

    8 Air Source Pressure Manometer

    9 Chamber Pressure

    10 Flowmeter

    11 O2 Analyzer

    12 Patient Warning Reset Button On/Off

    13 Patient Warning Alarm

    14 Door Locked Light

    15 Pressure Inside Light

    16 Set Pressure

    17 TV On/Off

    18 Emergency Stop Button

    19 Intercom

    20 System On/Off Button

    21 BIBS Mask Air or O2 Selection

    22 Auxiliary Exhaust

    23 Decompression

    24 Compression

    25 Chamber Selection Air or O2

    9 7

    4

    3

    10

    12

    19

    5

    11

    2

    21 8

    14

    22 23 24

    6

    1

    13

    15

    20

    18

    25

    17

    16

  • 10

    O2mono Control panel consist of the following equipment’s which enables the operator take

    control of the chamber safely and reliable. All main operations and functions of the chamber

    can be done with this panel .The control panel of the hyperbaric chamber is located on the

    chamber and provides following operations.

    • Treatments control

    • Maintain and adapt the interior environment of the chamber

    • Communication and lighting

    • Supply, evacuation and exchange of the medical working gases

    • Monitoring of important information such as humidity, temperature and oxygen

    concentration and patients.

  • 11

    ENTERTAINTMENT UNIT

    During the treatment session, an LCD entertainment monitor is mounted on the chamber to

    spend time during the treatment. It provides opportunity to watch video and TV to patient.

    POWER SUPPLY UNIT

    O2mono electric system generally;

    Power Stage

    Communication

    Digital Indicators

    Analyzers

    The electrical system is supplied with 220V AC. It is also reduced 24V

    DC using electronic transformers.

    Communication System consists of the operator and operator warning

    system.

    Digital indicators show pressure, temperature, humidity and CO2 value

    shows.

    Oxygen and carbon dioxide analyzer is available.

    The input voltage system 220 V AC ± 10%, frequency 50 Hz ± 5%. There is power supply in

    order to support the system at least 30 minutes in cases where input voltage is not present,

    cut or go beyond the limit value.

    LCD Monitor

  • 12

    PATIENT TRANSFER UNIT

    Stretcher is used for patient transferring and it consists of two parts as chamber internal

    stretcher and transfer stretcher. Internal stretcher is transported with transfer stretcher while

    the patient is transported into the chamber by pushing The stretcher is designed for working in

    hyperbaric conditions as mechanically.

    Figure Stretcher

  • 13

  • 14

  • 15

    Adress: Aydıntepe Mah. Dinçel Sk. No:1 Lider Plaza Kat:4 34947 TUZLA – ISTANBUL, TURKEY

    www.baroks.eu [email protected]

    http://www.baroks.eu%[email protected]://www.baroks.eu%[email protected]