USER MANUAL FOR HIGH PRESSURE OXYGEN GENERATOR I r F-qg

Transcript of MANUAL FOR HIGH PRESSURE - Frank's Hospital Workshop...Net weisht (Ks) 65 118 125 Dimension (mm)...

Page 1: MANUAL FOR HIGH PRESSURE - Frank's Hospital Workshop...Net weisht (Ks) 65 118 125 Dimension (mm) 850.420.600 850.590.660 850.660.660 Low purity alarm When oxygen purity is > 84%, the

USER MANUAL FOR HIGH PRESSURE

OXYGEN GENERATOR

Ir F-qg

Page 2: MANUAL FOR HIGH PRESSURE - Frank's Hospital Workshop...Net weisht (Ks) 65 118 125 Dimension (mm) 850.420.600 850.590.660 850.660.660 Low purity alarm When oxygen purity is > 84%, the

Contents

Foreword. ..........3

Caution.... ........4

Safety lnstruction ............4

Attention.. ....4-5

Product lntroduction.............. ................6-7

Operating condition.. ........7

Scope of application........... ..................7

Technical parameter .......8

Name and function............ ........9-16

Operation steps........ ....16-17

Maintenance instruction.......... 17-20

Troubles and so1ution.............. .....21

Condition for transportation and stora9e............. .. .....22

Accompany spare parts........ ............22

Quality guarantee .........22

Appendix l: Schematic Diagram... ....... .....23

Appendix ll: Connection Schematic Diagram of Control board.. ...................24

Page 3: MANUAL FOR HIGH PRESSURE - Frank's Hospital Workshop...Net weisht (Ks) 65 118 125 Dimension (mm) 850.420.600 850.590.660 850.660.660 Low purity alarm When oxygen purity is > 84%, the

Foreword

Thank you for purchasing our products, hoping you will be satisfied with our products.

This operation manual contains function, operation steps, attention, basic trouble

solution and so on. To ensure your efficient use of the machine, please have a close

read of this operation manual before operating it.

\

Page 4: MANUAL FOR HIGH PRESSURE - Frank's Hospital Workshop...Net weisht (Ks) 65 118 125 Dimension (mm) 850.420.600 850.590.660 850.660.660 Low purity alarm When oxygen purity is > 84%, the

. CautionA

1. Please check the power (voltage and frequency),adjust needed oulet pressure

before use the machine.

2' When use oxygen generator, should pay special attention to reduce fire hazards, in

safety, all combustibles should be far away from oxygen generator

3. Oil, grease or some material under the pressure when meeting oxygen will easily fire

naturally, and these substances must stay away from oxygen generator, oxygen tube

and other oxygen parts. Please don't use any lubricant. unless Longfian Scitech Co.,

Ltd written recommendation,

4' ln order to achieve optimal performance of the oxygen generator, it is striclyprohibited to switch on or off the machine frequently in short time, which can reduce the

life of the machine.

5. Keep Oxygen generator away from the wall, furniture and similar objects at least 20

cm.

6. Don't touch oxygen generator with wet hand.

7. cut off the power plug, when not using the oxygen generator.

* Safety lnstruction A1. Oxygen generator should be kept far away from open fire and fire origin

2. When checking and maintenance, please pull off the power plug in order not to be

electric shock.

3. When using the machine, it was not allowed to open the panel freely. lf there is

quality problem or unusual alarm, customer should contact with dealer and

producer.But forbidden to open and maintain by themselves.

* Attention A1. Oxygen generator should be kept in well-ventilated space to avoid the pollution.

2. When using oxygen generator, the bottom should be guaranteed in good ventilation,

othenrvise it will cause machine overheating.

3. Oxygen generator may have the exhaust noise intervals.

4. From turn on to setting time less then 10 mins.

Page 5: MANUAL FOR HIGH PRESSURE - Frank's Hospital Workshop...Net weisht (Ks) 65 118 125 Dimension (mm) 850.420.600 850.590.660 850.660.660 Low purity alarm When oxygen purity is > 84%, the

5. When oxygen concentration display( OCSI)show unnormal, user should state dealer

for Maintenance.

6. What oxygen generator need to clean is the filter net and the filter which needs to

change regularly;

7. Clean requirement is as following chart:

7.1 JAY-10-4.0. JAY-20-4.0 Model

8.When concentration less than 84%+2%(VN), alarm LED is shown, turn off the

machine immediately, and check the fault.

9.The machine can't frequently start and stop: lnterval time between turning off and

next turning on should not be less than 5 minutes ( ln order to prolong the service life of

air compressor).

10.Before turning on the power flow regulating valve must be open.

11.This machine is strictly prohibited to touch oil, grease and lubricant.

1 2.Graphics, abbreviated description :

AC: power; VA: Watt fH, see usermanual; A:Warning @ :Oround connection

Part name Cleaning cycle Cleaning method Remark

First air filter Working 100 hours Clean water to flush Black mesh

Second air filter Working 600 hours replaced ( replaced for

the first piece every

time )

Air filter

Third air filter Working 3000 hours Replaced for filter Air filter

Fourth air filter Working 5000 hours Replaced for mid filter Bacterialfilter

7.2 JAY-15-4.0Model

Part name Cleaning cycle Cleaning method Remark

First air filter Working 100 hours Clean water to flush Black mesh

Second air filter Working 3000 hours replaced Paper material

Third air filter Working 5000 hours Replaced for mid filter Paper material

Fourth air filter Working 5000 hours Replaced for mid filter Bacterialfilter

* , =,".,ric

safety B type; d, ,rnufacturing date; * ,Manufacturing address;

Page 6: MANUAL FOR HIGH PRESSURE - Frank's Hospital Workshop...Net weisht (Ks) 65 118 125 Dimension (mm) 850.420.600 850.590.660 850.660.660 Low purity alarm When oxygen purity is > 84%, the

* Product lntroduction

1 ., General principle

The oxygen generator adopt Pressure Swing Adsorption(PSA)principle, which can

separate oxygen, nitrogen and other gas from the air, at constant temperature, as soon

as power is connected, the oxygen that meets medical standards can be separated

from air constantly. Oxygen is generated by pure physical method, with steady, safe

and reliable oxygen flow, low cost, and adjustable flowrate and outlet pressure. There

is no influence on indoor oxygen percentage during the concentrator operating.

2. General Compose

Oxygen generator is made up of main unit that include adopt oil-free air compressor

dynamic system. filtering, cooling and dehumidifiers system (JAY-10-4.0Model adopt

rotary automatic dehydrator;JAY-15-4.0. JAY-20-4.0 Model adopt cooling drying

device), Adsorption air separation system, the controlvalve (inlet valve, exhaust valve,

equalizing valve, emptying valve, drain valve) and the etectric control system and the

oxygen concentration monitoring system composition, and the outer plate surface

spray processing.

3. Working Principle

oil-free air compressor power system to produce pressure about 0.43 Mpa

compressed air, through condenser make the temperature lower then by rotary

automatic dehydrator or freeze drying machine removing the moist from compressed

air, through control valve supply oxygen gas to three adsorption separately to be

0.40Mpa concentration higher g0% oxygen according to control program requirements.

Page 7: MANUAL FOR HIGH PRESSURE - Frank's Hospital Workshop...Net weisht (Ks) 65 118 125 Dimension (mm) 850.420.600 850.590.660 850.660.660 Low purity alarm When oxygen purity is > 84%, the

flowmeter

*rr*. fi

EMPTY VALVE

AIR GAUGE

compressor

valve

check valve C

OXYGEN GAUGEANTI.FLOW VALVE

* operating condition (lnclude oxygen purity indicator ocsl)1. Environment temperature: 5C-40C

2. Relative humidity: 30o/o-75o/o

3. Atmospher€ pressure : 700hPa-1060hPa

4. The surrounding environment no corrosive gas and strong magnetic field.

* Scope of Application

1. Product is suitable for aerobic crowd, suitable for to chronic obstructive pulmonary

disease, cardiovascular disease and other oxygen therapy patients.

2. Product can supply oxygen to ventilator and anesthesia machine.

equalizing valve B

ourPUTA I ourpur a

equalizing valve A

Page 8: MANUAL FOR HIGH PRESSURE - Frank's Hospital Workshop...Net weisht (Ks) 65 118 125 Dimension (mm) 850.420.600 850.590.660 850.660.660 Low purity alarm When oxygen purity is > 84%, the

* Technica! parameter

Model JAY-10-4.0 JAY-15-4.0 JAY-20-4.0

Rated power (VA) 850 1 500 1 800Operation voltage

(Vl{z)AC220Vt22V

50Hz+1HzAC220Vr22V

50Hz+1 HzAC220Yt22V

50Hz+1HzOxygen Flow

( L/min )0- 10 0- 15 0-20

Oxygenconcentration (VA/) 93o/ot3o/o 93%t3% 93o/ot3oh

Outlet pressureMPa

0.14-0.40 0.14-0.40 0.14-0.40

Temperature alarm >80C alarm >80C alarm >80C alarm

Alarm(Optional)Low purity

High&low pressureLow purity

High&low pressureLow purity

High&low pressureSound level(dB) s55 <60 <60

Electrical category Class I Type B Class I Type B Class I Type BProduct category Classll Type a ClassllType a ClassllType a

DehumidifierAutomatic

DehumidifierRefrigerated air

DryerRefrigerated air

DryerNet weisht (Ks) 65 118 125Dimension (mm) 850.420.600 850.590.660 850.660.660

Low purityalarm

When oxygenpurity is > 84%,the blue lamp ison, when oxygen

purity is below84o/o ,red lamp is

on, indicating lowpurity.

(Accuracyx2%)

When oxygenpurity is > 84%,the blue lamp ison, when oxygen

purity is below84o/o ,red lamp ison, indicating low

purity.(Accuracyt2o/o)

When oxygenpurity is > 84%,the blue lamp is

on, when oxygenpurity is below

84o/o ,rad lamp is

on, indicating lowpurity.

(Accuracyx2%)

Relief valvepressure 0.5MPa 0.5MPa 0.5MPa

Page 9: MANUAL FOR HIGH PRESSURE - Frank's Hospital Workshop...Net weisht (Ks) 65 118 125 Dimension (mm) 850.420.600 850.590.660 850.660.660 Low purity alarm When oxygen purity is > 84%, the

* Name and function

JAY-1 0-4.0. JAY-1 5-4.0. JAY-20-4.0 model oxygen generator construction

1. General appearance of oxygen generator

JAY-10-4.0 JAY- 1 5-4. 0/JAY-20-4. 0

See as above picture, oxygen generator is mainly composed by main unit case, front

panel, front plate, back panel, head cover, left side door, right side door, trundles.2. Top view of Oxygen generator2.1 JAY-10-4.01) Output regulator valve2) Oxygen pressure gauge3) LCD display4) Oxygen tank5) Third air filter6) Second air filter7) Cooling fan ( compressor cooling )8) Reducing valve ( For oxygen sensor)9) Oxygen sensor10) Electric condenser1 1) A adsorption12) B adsorption13) C adsorption14) Fourth air filter15) Cooler16) Cooling fan (cooler cooling use)

JAY-10-4.0

I

W

Page 10: MANUAL FOR HIGH PRESSURE - Frank's Hospital Workshop...Net weisht (Ks) 65 118 125 Dimension (mm) 850.420.600 850.590.660 850.660.660 Low purity alarm When oxygen purity is > 84%, the

2.2 JAY-15-4.01 ) Output regulator valve2) Oxygen pressure gauge3) LCD display4) Oxygen tank5) Cooling fan (compressor cooling)6) Electric condenser7) Second air filter8) Refrigerated air Dryer9) A adsorption10) B adsorption11) C adsorption12) Cooler13) Cooling fan (cooler cooling use)14) Fourth air filter15) pressure gauge (compressed air pressure)

2.3 JAY-20-4.01 ) Output regulator valve2 ) Oxygen pressure gauge3) LCD display4) Oxygen tank5) Cooling fan (compressor cooling)6) Electric condenser7) Two position three way electromagnetic valve8) Second air filter9) Single way valve10) Fourth airfilter11) C adsorption12) B adsorption13 ) A adsorption14) Fan15) Refrigerated air Dryer

JAY-15-4.0

JAY-20-4.0

I

Page 11: MANUAL FOR HIGH PRESSURE - Frank's Hospital Workshop...Net weisht (Ks) 65 118 125 Dimension (mm) 850.420.600 850.590.660 850.660.660 Low purity alarm When oxygen purity is > 84%, the

3. Oxygen generator positive side view3.1 JAY-10-4.0

1) Flowmeter2) LCD display (display operation parameters)3) Power (Turn to left,power on)4) LED display(Machine works and alarms in different color LED lighton indicating relevant situation)Audio alarm' When machine works unnormally,relevant audio alarm will be heared .

6) Oxygen pressure gauge7) Pressure Regulator valve8) Sequence valve9) Main control board10) Power11 ) Line bank

3.2 JAY- 1 5-4.0/JAY-20-4.01 ) Flowmeter2) LCD display: (display operation parameters)3) Power: (Turn to left ,Power on)4) LED light indicator:(Machine works and alarms in different color LEDlight on indicating relevant situation)5) Audio alarm: When machine works abnormally,relevant audio alarm will be heard .

6) Oxygen pressure gauge7) Pressure Regulatorvalve8) Sequence valve9) Main control board10) Power11) Line bank12) Oxygen sensor13) Reducing valve (oxygen sensor use)14) First air filter

JAY-10-4.0

JAY- 1 5-4. 0/JAY-20-4.0

/l)l,13

{

,4/-

/cv

ul i

r. ,l

I i;.

7

Page 12: MANUAL FOR HIGH PRESSURE - Frank's Hospital Workshop...Net weisht (Ks) 65 118 125 Dimension (mm) 850.420.600 850.590.660 850.660.660 Low purity alarm When oxygen purity is > 84%, the

4. Right side view of oxygen generator4.1 JAY-10-4.01) Oxygen tank2 ) Equalizing electromagnetic valve and

holder3) Rubber tank (compressor)4) lnlet tube (compressor inlet)5) lntake Solenoid valve and holder6 ) Exhaust silencer7) Empty valve

4.2 JAY-15-4.01) Pressure gauge (compressed air)2 ) Empty valve3) Exhaust silencer

4) Drainage outlet

5) Drain valve6) Pressure gauge (oxygen)7) Pressure regulating valve8) Refrigerated air Dryer9) Oxygen sensor10) Reducing valve(for oxygen sensor)

4.3 JAY-20-4.01) Fan2 ) Exhaust silencer3) Drain valve

4) Drainage outlet

5) Refrigerated air Dryer6) Pressure gauge (oxygen)7) Pressure regulating valve8) Oxygen sensor9) Reducing valve(for oxygen sensor)

JAY-10-4.0

JAY-15-4.0

JAY-20-4.0

Page 13: MANUAL FOR HIGH PRESSURE - Frank's Hospital Workshop...Net weisht (Ks) 65 118 125 Dimension (mm) 850.420.600 850.590.660 850.660.660 Low purity alarm When oxygen purity is > 84%, the

5. Left side view of oxygen generator5.1 JAY-10-4.01) Axial flow air dryer (for radiator)2) Fourth air filter3) Radiator4) Adsorption5) Third air filter6) Equalizing electromagnetic valve and7) Rubber tank (compressor)B) Oxygen tank9) Pressure gauge (compressed air)10) lntake Solenoid valve and holder11) Exhaust Solenoid valve and holder12) Breaker

5.2 JAY-15-4.01) Adsorption2) Second air filter3) Oxygen tank4) Rubber tank (compressor)5) Equalizing electromagnetic valve and6) Radiator7) Axial flow air dryer (for radiator)8) lntake Solenoid valve and holder9) Breaker10) Exhaust Solenoid valve and holder

5.3 JAY-20-4.01) Adsorption2) Second air filter3) Third air filter4) Oxygen tank5) Rubber tank (compressor)6) Equalizing electromagnetic valve and holder7) lntake Solenoid valve and holder8) Breaker9) Exhaust Solenoid valve and holder10) Fourth air filter

JAY-10-4.0

JAY-15-4.0

JAY-20-4.0

h.,i ,

i/ tffi

Page 14: MANUAL FOR HIGH PRESSURE - Frank's Hospital Workshop...Net weisht (Ks) 65 118 125 Dimension (mm) 850.420.600 850.590.660 850.660.660 Low purity alarm When oxygen purity is > 84%, the

6. Back side view of oxygen generator6.1 JAY-10-4.01) Oxygen outlet2 ) First air filter3) Automatic dehydrator4) Drain valve5) Power cord

JAY-10-4.0

6.2 JAY- 1 5-4.0/JAY-20-4.01) Oxygen outlet

2) Waterspout

3) Fan4) Power cord

JAY- 1 5-4. o/JAY- 20 -4.07. LED indicating light

1) P.O.:Power On, Green LED on

2) H.OZ:High oxygen concentration light (284o/o, Blue LED on)

3) L.O2:Low oxygen concentration light (<84o/o, Red LED on)

4) L.P.: Low Pressure indicate light

(oxygen pressure<0.25MPa, yellow LED light on)

5) H.P.: High pressure light

(oxygen pressure)0.45MPa, Red LED light on)

6) H.T.:High temp light on ( Rubber tank temp>8Oc ,

Remark' JAY-10-4.0 and JAY-15-4.0 Model:

Red LED light on )

Page 15: MANUAL FOR HIGH PRESSURE - Frank's Hospital Workshop...Net weisht (Ks) 65 118 125 Dimension (mm) 850.420.600 850.590.660 850.660.660 Low purity alarm When oxygen purity is > 84%, the

H.O2:High oxygen concentration light (>84o/o, Blue LED on)

L.O2:Low oxygen concentration light (<g4o/o, Red LED on)

8. LCD SCreert:

8.1 After turning on the machine, LCD display on, follow information will be shown :

0.P. MPa: 0.0000. T. C: 00.00.Time:0000:00

Acc. Time: 0000000

O.P.: oxygen output pressure(normal scope: 0.35Mpa- 0.43Mpa).

O.T.: case temp of compressor (normal SCope : sgOC )"

O.Time: Operating time (hour: minute)

Acc.Time: Accumulative operating time ( hour)

8.2 Adjustment interface l, When pressing selected button once, following display will

be entered

T1: 0.00 <- T2:0.00

T3: 0.00 T4z 0.00

T5:0.00 T6:0.00

TTz 0.00 T8: 0.00

8.3 Adjustment of interface ll, When pressing

display will be entered

selected button nine times, follow

T9: 0.00 +

T10: 0.00

T11: 0.00

Tl2z 0.00 F: 000

Page 16: MANUAL FOR HIGH PRESSURE - Frank's Hospital Workshop...Net weisht (Ks) 65 118 125 Dimension (mm) 850.420.600 850.590.660 850.660.660 Low purity alarm When oxygen purity is > 84%, the

T1-T12: The setting time for inlet. exhaust. Equalizing solenoid valve which works

unit: second"

F: The setting interval time for drain valve starting ( Factory set to start for 10 seconds

in every 60 minutes; user should set it according to actual environment. ln high

humidity season this valve open in shorter interval properly; conversely, it will open in

long interval time), unit: minute

* Operation steps

1. Let oxygen generator power cord plugs access mains.

2. Turn on circuit breaker switch of the oxygen generator, and green LED indicator,

LCD screen, red LED indicator for low oxygen will be on.

3, Open the knob of flow meter.

4. Turn knob left to "ON" , oxygen generator began to work.

5' After the system is on and empty valve start for 10 seconds which guarantee the

compressor start (the valve for normally open valve)

6. Oxygen generator's outlet pressure after 60 seconds: JAY-10-4.0. JAy-15-4.0.

JAY-20-4.0 above 0. 35Mpa.

7. after 3 minutes of operation, flow meter should be adjusted to required flow. but not

greater than the standard flow value. Oxygen output pressure can be adjusted by the

pressure regulating valve, but not more than the rated pressure value(when>0.45 Mpa

will stop).

8. oxygen purity should be above 90% after operating 10 minutes.

9' lf not continuously be used, please shut off the power (disconnect oxygen generator

in circuit breaker switch) and unplugged.

10. Oxygen purity alarm function: after ten minutes it will reach the normal.

When oxygen purity < 84o/o, error+ 2o/o, "L.o2" red LED light, show low oxygen purity.

When oxygen purity 2 84o/o, error+ 2o/o, "H.O2', blue LED light.

11' lf starting fifteen minutes and the low oxygen concentration light "L.O2" still red,

the system operation is unnormal.

Now check if position of the pressure reducing valve for oxygen sensors is normal,

because too much flow in the oxygen sensor can make output purity of oxygen

Page 17: MANUAL FOR HIGH PRESSURE - Frank's Hospital Workshop...Net weisht (Ks) 65 118 125 Dimension (mm) 850.420.600 850.590.660 850.660.660 Low purity alarm When oxygen purity is > 84%, the

generator insufficient. By adjusting the pressure reducing valve (oxygen sensor with)

make its flow meet the requirements, or close concentration light of the valve also can't

make blue LED light,: the oxygen sensor should be checked for adjustment with

pressure reducing valve, with reference to the ((maintenance instructions) in item six .

12. Drain valve starts for 5 seconds according to interval setting time (note: discharge

of water should be properly handled, such as drainage or container collection).

13. when Turning off (rotary switch to "oFF"), axial fan, and cooling drying device

stop work. after 60 seconds

* Maintenance instructions'1 ' Please cut off the power supply, the external casing use clean soft wet towel touchsa few cleaner, wipe external casing everywhere, then dry towel cleaning, 1 - 2 times amonth.

2. clean primary filter is an important step of the dairy maintenance(see cleaning requirements table)Methods: the case should be cleaned with cleanser Thoroughly andthen risen with dry cloth twice every month.

3. Change second filter(see cleaning requirements table)3.1 JAY-10-4.0, JAY-20-4.0 Model (change filter cotton)Open the oxygen generator side door, open the secondaryfilter remove filter cotton, the most outside a piece of change(figure 2), on the inside, make originally the second piece in thefirst place,and rotation use.

3.2 JAY-15-4.0 model:Open the oxygen generator side door and checkif the secondary filter became black, if yes, changethe new filter.

4. Clean third inlet filter(see cleaning requirements table)4.1 JAY-10-4.0. JAY-20-4.0 Model (change the filter cotton),Open the oxygen generator side door, Unplug the silicone rubbertube on the top of third filter, turn on the third filter and change thefilter cotton. at the same time clean filter's cavity.

first filter

Second filter

Second filter

Page 18: MANUAL FOR HIGH PRESSURE - Frank's Hospital Workshop...Net weisht (Ks) 65 118 125 Dimension (mm) 850.420.600 850.590.660 850.660.660 Low purity alarm When oxygen purity is > 84%, the

Third air filter

4.2 JAY-15-4.0 model (change filter element),open the oxygen generator left side door, upload compressorlnternal case side board, unscrew the third filter fastening screw,open the black case, replace the internal filter core, and at thesame time clean filter cavity.

5. Replace fourth filter

Open the oxygen generator side door, ln the panel nearthe oxygen outlet position, check the fourth firter's color, if colorchanged,unscrew filter and change the element.

6, Use reducing valve check oxygen sensor

Reducing valve is used to check whether oxygen purity is normal.othenryise ,Oxygen sensor is supplied more oxygen, it willlead to loss of pressure of the oxygen ouflet, flow reducing,because too much oxygen to oxygen sensor will reducing thepurity and pressure of the outlet. The best way to check howmuch amount of oxygen to the oxygen sensor is to connect atube to the oxygen sensor output and put this tube to the water

(as shown in picture) and if one bubble will appear in the waterevery a second, it will be oK. lf more continuous bubbles appears,the reducing valve should be adjusted to its best condition7. CompressorCompressor positioning is made of a positioning screw, damping spring and statorcomposition, lt can show clear positioning screw and bottom friction sound whenhappened offset, external forces need to put to compressor in the right position.Compressor need maintain every year, blowing the dust on the surface and checkwiring.

Third air filter

Fourth air filter

,.fl*

Page 19: MANUAL FOR HIGH PRESSURE - Frank's Hospital Workshop...Net weisht (Ks) 65 118 125 Dimension (mm) 850.420.600 850.590.660 850.660.660 Low purity alarm When oxygen purity is > 84%, the

JAY-1 0-4.0. JAY_20_4.0

8. Control board

24 ',23 ',22 2t 2A 19 18 17 16 15 14

1) Axial & Refrigerated air dryer control; 2) Compressor control; 3) DC24V input; 4)horn connector; 5) LED lamp connector; 6) Down adjusting button 7) select button;B) Up adjusting button; g) LCD display connector; 10) DCSV input connector; 11)temperature sensor connector ; 12) pressure SerSot; 13 ) Turn on/off and power offalarm connector 14) drain valve connector; 15) atmospheric valve connector; 16)A tower intake valve connector ; 17 ) B tower intake valve connector; 18 ) C tower intakevalve connector; 19 ) A tower vent valve connector ; 20) B tower vent valve connector;21) C tower vent valve connector; 22) AB tower equalizing valve connector; 23) BCtower equalizing valve connector; 24) AC tower equalizing valve connector.

Every three months check the fastness for binding post, use screwdriver twist screwfor fixing line, should no loose, every year should complete maintenance, blowing thedust on the surface and check the wiring.

JAY-15-4.0

M @M#A

ffitr*it$IffiE

::,. wi,tl* T ,'iif :l rililo*'Lg,#,,;mffi*;m#

ffi

9i-r -

M;+-w,W,&WW

Page 20: MANUAL FOR HIGH PRESSURE - Frank's Hospital Workshop...Net weisht (Ks) 65 118 125 Dimension (mm) 850.420.600 850.590.660 850.660.660 Low purity alarm When oxygen purity is > 84%, the

9. Binding post1) locating plate;2) Axial flow fan, air dryerAC220V-N;3) axial flow fan, dryerAC220V-N;4) Axial flow fan, air dryerAC220V-N;5) axial flow fan, dryer AC22OV-L, connect relay part;6) compressor AC220V-L, connect relay part;7) AC/DC switch power connect electric supplyAC220V-L;8) Connect electric supply AC22OV-L;9) Compressor connect electric supply AC220V-N;10)AC/DCswitchpoWerconnectelectricsupplyAC220V-Nffi1 1 ) Connect electric supplyAC220V-N;

10' lf users need circuit diagram, components list, diagram note, correction rules andmaintenance material, manufacturers will provide necessary drawing material.

Page 21: MANUAL FOR HIGH PRESSURE - Frank's Hospital Workshop...Net weisht (Ks) 65 118 125 Dimension (mm) 850.420.600 850.590.660 850.660.660 Low purity alarm When oxygen purity is > 84%, the

* Troubles and solutionsNo. Trouble Causes Solution

1

\o operationrfter power:onnected

1.No connection between circuitof oxygen generator and power

2.Capacitor of compressorcroken3.Compressor broken4.Switch connection problem

l.Check out whether switch, plug,power line in good connection.

2. Replace start capacitor3.Change compressor4.Check switch connection line

2

No oxygenout or tinyouttake flow

l.Change of Relief valve settingpressure

2.Filter clogging, air inlet difficult3.Connection pipeline no screw,air leakage

4.Too much leakage rate ofreducing valve for oxygensensor

5.No working of solenoid valve

l.Replace and adjust relief valve2.Clean filter3.Check pipeline4.Check and adjust reducing valve forone bubble every second

5.Check solenoid valve and replace

3No exhaustsound

1. Air control valve can not work2. Electrical control board cannotwork

l.Change air control valve2.Change electrical control board

4l-oo noisy:xhaustion

1.The joint of exhaustion mufflerfallen off

2. Exhaustion muffler broken

l.Connect the joint well2. Have the muffler replaced

5Ioo noisyihake

1 .Compressor fixing positionrffset2.Compressor positioningcolumn deformation

1.Move compressor in the fixed position2. Reshaping positioning screw or

replacement

6Cxygen purityalarm light

l.Oxygen purity below 84% l.Check compressor air system2.Check oxygen sensor3.Check and adjust pressure reducingvalve if have no air outlet

3.Air leakage for oxygen generator

7

High/lowpressurealarm

l.Pressure regulating valve offse2.lncorrect flow meter outlet3.Pipeline air leakage

l.Check and adjust pressure regulatingvalve

2.Adjust and replace flow meter3.Check fastening pipeline

8No light forLCD backlight

l.Backlight damage2.Backlight burn out because ofthe fault alarm

1. Replace LCD2. Reconnect with main power supply

and restart the machine, backlight willlight

Page 22: MANUAL FOR HIGH PRESSURE - Frank's Hospital Workshop...Net weisht (Ks) 65 118 125 Dimension (mm) 850.420.600 850.590.660 850.660.660 Low purity alarm When oxygen purity is > 84%, the

* Condition for transportation and storage

Environment temperature scale: -4OC-S5C

Comparative humidity scale, <gs%

Air pressure scale, 700hpa-1060hpa

* Accompany spare parts

1, Air filter net 1 unit

2. Nylon tube 10*8*1 meter. 8*6*1 meter. 6*4*l meter

. Quality Warrant

The whole machine repair: one year guarantee.

After-sale service: Longfian Scitech Co., Ltd.

Add: 2F&3F, East Section, Building 12, Power valley pioneer park,

NO.369 Huiyang street, 071051,Baoding,Hebei province, China

Postcode: 071051

Service phone: 0086-31 2-5909505

Fax: 0086-312-5909515

Page 23: MANUAL FOR HIGH PRESSURE - Frank's Hospital Workshop...Net weisht (Ks) 65 118 125 Dimension (mm) 850.420.600 850.590.660 850.660.660 Low purity alarm When oxygen purity is > 84%, the

Appendix l: Oxygen generator connection diagram

ACZZOY N

ACZZOY L

breaker controlboard

breaker

comprBs$0r

Axial&Refrigeration

air dryer

Page 24: MANUAL FOR HIGH PRESSURE - Frank's Hospital Workshop...Net weisht (Ks) 65 118 125 Dimension (mm) 850.420.600 850.590.660 850.660.660 Low purity alarm When oxygen purity is > 84%, the

Appendix ll: oxygen generator control board connection diagram

A.P.AC

A.P.BO

A,P.AB

Exhaust valve C

Exhaustvalve B

Exhaustvalve A

lnputvalve C

lnputvalve B

lnputvalve A

Empty valve

l;rU

r[tt!rf-.1l'J

E

rl.lt!

t-J

l'l[]

option [.] UPEI L*l

Downf

rLCD screen connector lll

[lu[J

Temp sensor connector

HE]

Start&Power faliure alarm connector

Drainage

Electric control board PCB