LEO-5 - Instrumart · Thank you for your purchase of the LEO 5 high resolution digital pressure...

7
LEO-5 High-Resolution Digital Pressure Gauge Quickstart Guide Keller America Inc. - Newport News, Virginia, U.S.A Revision 1, October 2014

Transcript of LEO-5 - Instrumart · Thank you for your purchase of the LEO 5 high resolution digital pressure...

Page 1: LEO-5 - Instrumart · Thank you for your purchase of the LEO 5 high resolution digital pressure gauge. The LEO 5 combines the notable fea-tures of Keller’s successful LEO-Record

LEO-5

High-Resolution Digital Pressure Gauge

Quickstart Guide

Keller America Inc. - Newport News, Virginia, U.S.ARevision 1, October 2014

Page 2: LEO-5 - Instrumart · Thank you for your purchase of the LEO 5 high resolution digital pressure gauge. The LEO 5 combines the notable fea-tures of Keller’s successful LEO-Record

Keller America Inc LEO 5 Quickstart Guide Rev. 10/14

Table of Contents

I. CompanyInformation.…………………………………………………………………………………2 II. ContactInformation……………………………………………………………………………………..2 III. ProductIntroduction…………………………………………………………………………………….2 IV. Diagram...............………………………………………………………………………………………….3 V. GettingStarted....………………………………………………………………………………………….3 VI. MenuOptions……....................………………………………………………………………………4 VII. ExplanationofFunctions……………………………………………………………………………….5 Electronic copies of this user’s guide can be found at: http://www.kelleramerica.com/manuals-and-software/

Page 3: LEO-5 - Instrumart · Thank you for your purchase of the LEO 5 high resolution digital pressure gauge. The LEO 5 combines the notable fea-tures of Keller’s successful LEO-Record

I. Company Information

KellerAmericaIncistheNorthAmericansubsidiaryofKellerAGfürDruckmesstechnik,whichcontinuesastheleaderinlevelandpressuresensing.Fromhumblebeginnings,Kellernowprovidesitsproductsandexpertisethroughaworldwidenetwork of factory-trained professionals.

Withtheadventofinexpensive,miniaturizedmicroprocessors,Kellercontinuestoadvancethestate-of-the-artinfullypackagedpressuretransmitters,withTotalErrorBandperformancenotpossiblejustafewshortyearsago.Annualsen-sorproductionnowexceeds1millionpieces.

DoingbusinesswithKelleriseasyandmostinquiriesareansweredwithin24hours.AtKeller,providingcustomerswithsuperior value is a way of life.

II. Contact Information

Keller America, Inc.351BellKingRdNewportNews,Virginia,23606P:877-253-5537F:757-596-6659Web: www.kelleramerica.comEmail: [email protected]

III. Product Introduction:

ThankyouforyourpurchaseoftheLEO5highresolutiondigitalpressuregauge.TheLEO5combinesthenotablefea-turesofKeller’ssuccessfulLEO-RecordandLEX1digitalpressuregauges,featuringanIP67-ratedstainlesssteelenclo-sure.Thisrobusthousingcombinescontemporarymicrocontroller-basedelectronicsandcapacitive-touchcontrols,operated through the environmentally-sealed safety glass front panel. The large backlit LCD display ensures readability in anylightingconditions.

Twoselectablemeasurementmodes,standardandpeak,areincludedtoensuremaximumversatility.Instandardmode,high-resolutionpressuremeasurementsaretakentwicepersecond.Whenoperatinginpeakmode,samplingfrequencyincreasesto5kHzwith16bitresolution.

TheLEO5interfaceswithaPCviaUSBconnectivity,allowingaccesstoactivemeasurementsandrecordeddatainclud-ingpressure,peakpressure,temperature,andmeasurementinterval.PCconnectionalsoallowsdeviceconfiguration,firmwareupdates,andrechargingoftheintegralaccumulatorpowersupply.

TheuniquemodulardesignoftheLEO5makespossibleadaptationtocustomer-specificmeasuringrequirements.

Page2

Keller America Inc LEO 5 Quickstart Guide Rev. 10/14

Page 4: LEO-5 - Instrumart · Thank you for your purchase of the LEO 5 high resolution digital pressure gauge. The LEO 5 combines the notable fea-tures of Keller’s successful LEO-Record

IV. Diagram:

V. Getting Started:

1. Toturnonthegauge,pressthe“Select”keyandthenthe“Enter”button.2. Thegaugewillpowerupanddisplayandthepressureandtemperaturereadingswillautomaticallyafterafew secondsofstartuptimeandthegaugeisnowon.3. Toturnoffthegaugepressthe“Select”keytwiceand“OFF”shouldappearandthenpressthe“Enter”button.

VI. Menu Options: 1. Pressthe“Select”key3timesoruntil“MENU”appearsonthescreen.2. Pressthe“Enter”keywhenyouseethe“MENU”optionandyouhaveenteredintothemenuoptions.3. Onceinthemenuoptions,youmaykeepselectingthe“Select”buttontoscrollthroughtheoptions: A. RecordingOnorOff B. PeakPressure/TempStartorStop C. BluetoothOnorOff(additionaloption) D. ZeroReferencePointSetting E. ResettingtheGaugeToFactoryDisplaySettings F. UnitofMeasure G. Continuousfunction H. Display(Max/Min) I. BacklightOnorOff J. ResolutionSetting K. Accumulator/BatteryInfo4. Onceyouhavefoundyourmenuitemyoumayselectthe“Enter”buttontoselectthatmenuitemtoeitherturn onorofforstartorstop.

Caution! Please allow a second when selecting any items on the gauge. Selecting either button multiple times within a fraction of a second will throw the capacitive logic off and may freeze any functions for that moment.

Page3

Keller America Inc LEO 5 Quickstart Guide Rev. 10/14

Pressure Connection

Select Button

Enter Button

Current Pressure Display

Media TemperatureDisplay

Unit of Measure

Battery ChargeIndicator

Page 5: LEO-5 - Instrumart · Thank you for your purchase of the LEO 5 high resolution digital pressure gauge. The LEO 5 combines the notable fea-tures of Keller’s successful LEO-Record

VII. Explanation of Functions:

RecordingOn/Off:

ThisfunctionallowstheendusertostartorstopalogmanuallyinlieuofusingaPCtodoso. » TheloggingintervalsareconfiguredbyKeller’sLoggersoftwareandacommunicationcable. » Softwarecanbefoundhere:http://www.keller-druck.com/home_e/paprod_e/software_e.asp

PeakOn/Off:

Thisfunctionallowstheusertoautomaticallyrecordthepeakpressure/temperatureonthedisplayonly. » Thepeakvalueswillnotberecordedintheinternalmemory,unlesstheunitisin“Record”mode. » Toresetseeoption“G”,Continuousmeasurement.

BluetoothOn/Off(ifequipped):

ThisfunctionallowstheusertowirelesslyconnecttheLEO5toaBluetoothenabledWindows™tabletora PCtotransferdataorlivetimeviewing.

» Dependingonthereceivingdevice,transmissiondistancesarelimitedtoapproximately30feet. » The user can program the device wirelessly.

Zeroreferencepointsetting:

Thezeroreferencepointsettingallowstheusertosetthereferencezeropoint.Thisfunctionisusefulfor users that need a higher than normal zero reference point, i.e., elevated tank level monitoring or to zeroing the gaugeatthatmomentfortesting.Thesecond“Zero”functionzerosthedisplayfromthecurrentpeakmeasure ment, and/or max/min measuring.

Resettingthegauge:

Thisallowstheusertoresetallofthegaugedisplayvariablesbacktofactorysettings. » Pressurewillrevertto“bar”units » Temperature will revert to °C units

Changing the unit of measure:

Tochangetheunitofmeasureyoumustaccessintothemenuoptionsuntilyouseethe“UNIT”.Onceyousee “UNIT”youmustselectthe“Enter”buttononceandthenusethe“SELECT”buttontoscrollthroughtheunitof measure:bar,mbar,kpa,mPa,PSI,KP/cm2,cmH20,mh20,inH20,ftH20,mmHG,inHg,Liter,Hliter,°F,°C.Once youfindtheunitofmeasurethatbestfitsyourapplicationyoucanselectthe“Enter”button.

Caution! Please change the temperature unit of measure before changing the pressure unit of measure.

Page 4

Keller America Inc LEO 5 Quickstart Guide Rev. 10/14

Page 6: LEO-5 - Instrumart · Thank you for your purchase of the LEO 5 high resolution digital pressure gauge. The LEO 5 combines the notable fea-tures of Keller’s successful LEO-Record

ContinuousFunction:

ContON-Thegaugewillstayoncontinuouslyuntiltheuserturnsoffthegauge. ContOFF-Activatestheautomaticofffunction.Thegaugewillturnoffapproximately15minsafterthelastkeyis pressed.

DisplayFunction: Thisfunctionallowstheusertoturnonoroffthelivetimemaximumandminimumpressuremonitoring. » Toturnonoroffthisfunctionyoumustgothroughthemenuuntil“DISPL”appears.Thenyoumust selectthe“Enter”buttonandthenusethe“Select”buttontotogglebetweenstandardmeasuringand min/maxmeasuring.Oncethedesiredfunctionisdisplayed,youmayselectthe“Enter”button.

Backlight:

Therearetwobacklightoptions: » Backlightonfor10secondsafteryoutoucheitherthe“Select”or“Enter”button. » Backlightonconstantuntilyouturnitoffmanually.

Resolution:

Thisfunctionallowstheusertosetthepressureresolutionaccordingly.

Accumulator/BatteryInfo:

Thisfunctionallowstheusertoviewthebatteryvoltageinformation. » 4.3Visafullychargedbattery » 3.0Vorlessthebatterywillneedchargingforreliableservice

Page 5

Keller America Inc LEO 5 Quickstart Guide Rev. 10/14

Page 7: LEO-5 - Instrumart · Thank you for your purchase of the LEO 5 high resolution digital pressure gauge. The LEO 5 combines the notable fea-tures of Keller’s successful LEO-Record

Electronic copies of this user’s guide can be found at: http://www.kelleramerica.com/manuals-and-software/

KELLER AMERICA INC351 BELL KING ROAD NEWPORT NEWS, VIRGINIA 23606

PHONE: 877-253-5537 FAX: 757-596-6659WWW.KELLERAMERICA.COM