Lecture HPLC Student

download Lecture HPLC Student

of 243

Transcript of Lecture HPLC Student

  • 7/28/2019 Lecture HPLC Student

    1/243

    1

    TOPIC

  • 7/28/2019 Lecture HPLC Student

    2/243

    2

    Able to :

    1. Explain the fundamental concepts & theories ofseparation techniques in HPLC.

    2. Sketch & label the schematic diagrams in HPLC.3. Identify the strength & limitations of each type of HPLC

    technique.

    4. Suggest & justify the most suitable & efficient separationtechnique to be employed for an analysis.

    Course Learning Outcomes

  • 7/28/2019 Lecture HPLC Student

    3/243

    3

    Introduction Mobile phase is liquid.

    Analytes are first dissolved in a solvent / mobile

    phase, then passed through column under highpressure.

    The separation dependent upon interaction ofthe analytes with mobile & stationary phases.

  • 7/28/2019 Lecture HPLC Student

    4/243

    HPLC applications4

    Analysis ofmixtures of non-volatile or

    thermally unstable compounds (thosethat cannot be analyzed by GC)

  • 7/28/2019 Lecture HPLC Student

    5/243

    5

    To hold the

    liquid mobilephase

    Topressurizethe liquidmobile

    phase

    To allowinjection ofthe samplemixture

    Columncontainingthe bed ofstationaryphase

    To detect thepresence ofcomponents

    as they exitthe column

    To recordthe detectorsignal

  • 7/28/2019 Lecture HPLC Student

    6/243

    6

    HPLC

    setup

    http://upload.wikimedia.org/wikipedia/en/4/4b/Agilent1200HPLC.jpg
  • 7/28/2019 Lecture HPLC Student

    7/243

    Mobile phase delivery7

    Consists of pump to provide the high pressure required.

    Equipped with one/more glass reservoirs. Require very clean & pure HPLC grade solvents to prevent

    column degradation with impurities & to minimize detectorbackground signals from contaminants (usually UVtransparent).

    Solvents must be filtered (to remove dust) through 0.2/0.45m filters & degassed.

  • 7/28/2019 Lecture HPLC Student

    8/243

    8

    Membrane Filter Holder Assembly

  • 7/28/2019 Lecture HPLC Student

    9/243

    9

    Degassing (removal of dissolved N2

    , O2

    , CO2

    ..) by :

    9 Vacuum devices (vacuum applied to headspaceabove solvent).

    9 Ultra sonication (high frequency vibration drives

    gasses out of solvent).9 Heating (decreases solubility of gases).

    9 He sparging. Sparging is a process in whichdissolved gases are swept out of a solvent.

  • 7/28/2019 Lecture HPLC Student

    10/243

    10

    If solvents are not degassed, gases that are solubleat high pressure (ie. 4,000 psi) become insoluble atlow pressure (ie. at the end of the column or

    detector).

    Gas bubbles create difficulties with pumps

    (destabilize pressure), columns (bad separation) &detectors.

    Dissolved O2 absorb UV radiation in the 250-200 nmwavelength range & interfering with UV detection.

  • 7/28/2019 Lecture HPLC Student

    11/243

    Sample problem 111

    Why is high pressure is needed in HPLC?

  • 7/28/2019 Lecture HPLC Student

    12/243

    Elution12

    The mobile phase may comprise of a single solvent or,

    Various solvents can be mixed to give mixture with asuitable polarity.

    Isocratic elution

    The composition of the mobile phase remainsconstant during the separation

  • 7/28/2019 Lecture HPLC Student

    13/243

    13

    Binary gradient 2 solvents (very common).

    Tertiary gradient 3 solvents (quite common).

    Quaternary gradient 4 solvents (very rarely).

    Gradient elution

    Solvent composition of the mobile phase is

    changed with time during the separation

  • 7/28/2019 Lecture HPLC Student

    14/243

    14

    E.g. of a gradient :Starting at 10% methanol & ending at 90% methanolafter 20 minutes.

    The two components of the mobile phase are

    typically termed "A" & "B";9 A is the "weak" solvent which allows the solute to

    elute only slowly.9 B is the "strong" solvent which rapidly elutes the

    solutes from the column.

  • 7/28/2019 Lecture HPLC Student

    15/243

    15

    Solvent A is often water, while B is an organicsolvent miscible with water, such as acetonitrile,methanol, THF or isopropanol.

    Organic solvent of the mobile phase is to reducethe retarding strength of the aqueous component.

  • 7/28/2019 Lecture HPLC Student

    16/243

    16

    Different compounds are eluted by increasing thestrength of the organic solvent.

    The sample is injected while a weaker mobile phase isbeing applied to the system.

    The strength of the mobile phase is later increased in

    increments by raising the organic solvent fraction, whichsubsequently results in elution of retained components.

  • 7/28/2019 Lecture HPLC Student

    17/243

    17

    Advantages :

    9 Decreases the retention of the later-elutingcomponents so that they elute faster, giving

    narrower (& taller) peaks for most components.

    9 Improves the peak shape for tailed peaks.

  • 7/28/2019 Lecture HPLC Student

    18/243

    Pumping system18

    Pumping system pumps (force) the mobile phase

    through the manual injector, column, detector & finallyto the waste container.

    The quality of a pump for HPLC is measured by howsteady & reproducible a flow it can produce.

    A fluctuating flow rate can create detector noise thatobscures weak signals.

  • 7/28/2019 Lecture HPLC Student

    19/243

    19

    Requirements for HPLC :9 Pressures up to 6000 psi.9

    Pulse free, prevents remixing of solutes.9 Control flow rate from 0.1 to 10 mL/min.

  • 7/28/2019 Lecture HPLC Student

    20/243

    20

    Reciprocating pump

    9 The most used.

    9

    Mechanical high pressure liquid pumpingachieved by piston, check valves.

    9 Pressure surges can be somewhat overcomewith pulse dampeners.

    9 Advantages : Provide constant flow rates &independent of solvent viscosity or columnbackpressure.

  • 7/28/2019 Lecture HPLC Student

    21/243

    21

    Reciprocating pump

  • 7/28/2019 Lecture HPLC Student

    22/243

    22

    Displacement pump

    9 Continuous pressure applied to large volume(syringe pump).

    9 Advantage is lack of pressure surges.

    9 Disadvantage is price.

  • 7/28/2019 Lecture HPLC Student

    23/243

    Sample injection system23

    Injection port :

    9 Device for injection of sample into system.

    9 Consists of a stainless steel ring with 6 differentports, 1 to the column.

    9 Has interchangeable sample loops, which holda fix volume.

  • 7/28/2019 Lecture HPLC Student

    24/243

    24

    9 Sample volume is no more than volume

    indicated on a loop.9 Samples size ranges from 2-1000 L.

    9 Samples introduced manually into the valve witha syringe to fill the sample loop.

    9 Sample injection only with flat end needle (toprevent damage to the injection port).

    X

  • 7/28/2019 Lecture HPLC Student

    25/243

    25

    Manual injection port

    Sample loop

  • 7/28/2019 Lecture HPLC Student

    26/243

    26

    Injection valve

  • 7/28/2019 Lecture HPLC Student

    27/243

    27

    Injection procedure :

    1. Place in load position& load sample into loop.

    2. Rotate to inject position the content of thesample loop is injected into the column at highpressure.

  • 7/28/2019 Lecture HPLC Student

    28/243

    28

    Can inject a large number ofsamples automatically.

    Inject continuously variablevolume 1 L1 mL.

    Small vials with a septum placedin a tray.

    Needle penetrates septum &

    withdraws the sample.

    Valve introduces the sample into

    the column.

    Autosamplers

  • 7/28/2019 Lecture HPLC Student

    29/243

    Columns29

    1. Guard column

    2. Analytical column

  • 7/28/2019 Lecture HPLC Student

    30/243

    30

    Guard column (precolumn)

    310 cm long, placed between the injector & the

    analytical column.

    Contains the same packing as the analytical column.

  • 7/28/2019 Lecture HPLC Student

    31/243

    31

    Function : To protect & increase the lifetime of the

    analytical column by :

    9 Retain debris (eg. pump seal fragments), particulatematter & contaminants that would otherwise get onanalytical column & foul/clog it & changing columnefficiency.

    9 Retains highly sorbed compounds that would be

    caught on & not be eluted from the analytical column. Must be replaced or regenerated periodically.

  • 7/28/2019 Lecture HPLC Student

    32/243

    32

    Sample problem 2For the analysis of biological sample using HPLC,a guard column is normally required.

  • 7/28/2019 Lecture HPLC Student

    33/243

    33

    Analytical column Stainless steel.

    10-30 cm long (10,15 & 25 cm).

    4-10 mm internal diameter (3.9or 4.6 mm).

    3, 5 or 10 m particle size (dp).

    40,000-60,000 plates/m.

    Eg. 5 m ODS column (250 x 4.5 mm)

  • 7/28/2019 Lecture HPLC Student

    34/243

    34

    van Deemter equation,

    H = A + B/v + Cv

    Which of the 3 components is the largestcontribution to H? Consider the following :

    9 B/v effects diffusion is 100x less in liquidsthan in the gas & can be neglected.

    Band Broadening in LC

  • 7/28/2019 Lecture HPLC Student

    35/243

    35

    9 Cv effects we will assume that mass transport(M.T.) effects are the largest contribution.

    The mobile phase mass transfer coefficient :

    CMv = (fM(k)dp2/DM)v

    Smaller dp increases the surface area/volume

    ratio & thus decreases M.T. in the m.p.

  • 7/28/2019 Lecture HPLC Student

    36/243

    36

    Reduction of dp from 45 to 6 m results in a tenfoldor more decrease in H

  • 7/28/2019 Lecture HPLC Student

    37/243

    37

    Sample problem 3Discuss the effect of the particle size on the separation

    performance in packed column liquid chromatography.

  • 7/28/2019 Lecture HPLC Student

    38/243

    38

    Stationary phase Normally, columns are filled with silica gel because its :

    9 particle shape (spherical, can be packed morehomogeneously),

    9 surface properties,

    9 pore size (60-100 range)

    structure help to get a good separation.

  • 7/28/2019 Lecture HPLC Student

    39/243

    39

    Silica structures

  • 7/28/2019 Lecture HPLC Student

    40/243

    40

    Silica :

    9 Is wetted by nearly every potential mobile phase.

    9 Is inert to most compounds.

    9 Has a high surface activity which can bemodified easily with water & other agents.

    9 Can be used to separate a wide variety ofchemical compounds.

  • 7/28/2019 Lecture HPLC Student

    41/243

    41

    Limitation of silica stationary phase :

    X Operating pH range 2-8.

    X Dissolves above pH 8. Crosslinked polymericparticles (eg., polystyrene / polymethacrylates) areused for separation of bases.

    X Siloxane bond (Si-O-Si) to the stationary phasehydrolyses below pH 2.

    These defects will cause changes in tR & loss ofresolution.

  • 7/28/2019 Lecture HPLC Student

    42/243

    Detector types42

    UV absorbance

    9 Variable Wavelength UV-Vis detector.9 Diode Array Detector (DAD) : measure absorbance

    across a broad spectrum of wavelengths

    simultaneously. Fluorescence

    Electrochemical

    Refractive Index

    Mass spectrometric

  • 7/28/2019 Lecture HPLC Student

    43/243

    43

    HPLC detectors will depend on the nature of thesample.

    The most widely used detectors for HPLC are

    based on the absorption of ultraviolet or visibleradiation.

    UV region : 100-400 nm

    Vis region : 400-780 nm

  • 7/28/2019 Lecture HPLC Student

    44/243

    44

    The detectors often make use of the 254 nm &280 nm lines from a mercury source because

    many organic functional groups absorb in theregion.

    Usually set the UV at 254 nm. This will detect any to * transitions. So, it would detect anything

    with the benzene ring.

  • 7/28/2019 Lecture HPLC Student

    45/243

    45

    Based on electronic transitions withinmolecules.

    Most common type of detector for LC.

    Fixed wavelength, Hg lamp 254 nm ( => *).

    UV-Vis absorbance detector

  • 7/28/2019 Lecture HPLC Student

    46/243

    46

    A diagram of the UV detector

  • 7/28/2019 Lecture HPLC Student

    47/243

    47

    The detectorsensorconsists of a small cylindrical

    cell (normally 2.0 to 10 l in volume) through which,the eluent from the column flows.

    UV light from a UV lamp (usually a Hg dischargelamp) emitting UV light mostly at 254 nm.

  • 7/28/2019 Lecture HPLC Student

    48/243

    48

    Variable wavelength, selectable for specificwavelengths in the visible region (eg.deuterium lamp for use down to 190-400 nm).

    Still limited to single wavelengths.

  • 7/28/2019 Lecture HPLC Student

    49/243

    49

    Analyte must absorb in the visible region (haschromophore).

    You might wonder why the solventsused don't absorb UV light. They do!

    But different compounds absorb moststrongly in different parts of the UVspectrum.

  • 7/28/2019 Lecture HPLC Student

    50/243

    50

    Methanol, for eg., absorbs at wavelengths below205 nm & water below 190 nm.

    If you were using a methanol-water mixture as

    the solvent, you would therefore have to use awavelength greater than 205 nm to avoid falsereadings from the solvent.

  • 7/28/2019 Lecture HPLC Student

    51/243

    51

    Measure absorbance across a broad spectrum ofwavelengths simultaneously.

    Allows for the recording of the entire spectrum of eachsolute as it passed through the diode array detector.

    Utilizes a deuterium or xenon lamp that emits light overthe UV spectrum range.

    UV-Vis absorbance detectorDIODE ARRAY (DAD)

  • 7/28/2019 Lecture HPLC Student

    52/243

    52

    Sample problem 4

    A sample mixture containing acetophenone, phenol, 1-hydroxyphenol & methyl benzene was analyzed usingHPLC with a mobile phase system of a ACN/H2O 80/20 &

    a UV detector set at wavelength 420 nm.Quantitative analysis was done using external standard

    method. Discuss the reason for each of the followingproblems encountered & suggest an approach toovercome the problem.

  • 7/28/2019 Lecture HPLC Student

    53/243

    53

    i. Very low detection signals were obtained.

  • 7/28/2019 Lecture HPLC Student

    54/243

    54

    ii. After doing 5 replicate measurements, the peakareas obtained vary considerably.

  • 7/28/2019 Lecture HPLC Student

    55/243

    55

    Based on the n, index of refraction. How much the

    light is bent in a substance is unique for all liquids &relative to their density.

    Advantages :9 universal, insensitive to flow, economical, robust.

    Disadvantages :X not sensitive, highly temperature dependent

    Refractive Index detector

  • 7/28/2019 Lecture HPLC Student

    56/243

  • 7/28/2019 Lecture HPLC Student

    57/243

    57

    Fluorescence :

    The most sensitive, 10-1000 times higher than thatof UV detector for strong UV absorbing materials.

    Electrochemical :Based on the measurements of the current resulting

    from oxidation/reduction reaction of the analyte at asuitable electrode.

    Other detectors

  • 7/28/2019 Lecture HPLC Student

    58/243

    58

    Conductivity :

    The conductivity of the column effluent iscontinuously measured & the appearance of the

    analyte in the cell is indicated by a change inconductivity.

  • 7/28/2019 Lecture HPLC Student

    59/243

    59

    Mass spectrometry (MS):

    Examining a compound (in a gas phase), that will providethe structural information of each compounds match fromthe library of mass spectra of known compounds storedon the computer & thus can identify the compound.

    Measures the mass-to-charge-ratio (m/z) of charged

    analytes.

  • 7/28/2019 Lecture HPLC Student

    60/243

    60

    Eg. mass spectrum for benzene

  • 7/28/2019 Lecture HPLC Student

    61/243

    61

    It is not a simple matter to connect the MS to LC,because

    9 LC operates in the liquid phase whereas MS is agas phase method.

    9 Analytes are non volatile & may be thermally labile

    but must be presented in gaseous form.

    In order to allow separated components of a mixture to

    be passed sequentially from the LC into the MS, aninterface is needed.

  • 7/28/2019 Lecture HPLC Student

    62/243

    62

    Sample problem 5Identification of compound is more confirmative by using

    mass spectrometer as a detector compare to UV detector.

  • 7/28/2019 Lecture HPLC Student

    63/243

    Sample problem 663

    Explain the advantage of using MS detector.

  • 7/28/2019 Lecture HPLC Student

    64/243

    Performance of HPLC detectors64

    HPLC detector LOD

    Absorbance 10 pgFluorescence 10 fg

    Electrochemical 100 pgRefractive index 1 ng

    Conductivity 100 pg 1 ngMass spectrometry < 1 pg

  • 7/28/2019 Lecture HPLC Student

    65/243

    65

    Chemical derivatization of a sample in HPLC isundertaken to convert a non-UV/fluorescenceresponding analyte into one that is.

    The derivatization step may enhance the separation.

    Sample derivatization

  • 7/28/2019 Lecture HPLC Student

    66/243

    Types of liquid chromatography66

    1. Partition chromatography

    (separation based on polarity)

    2. Adsorption chromatography

    3. Size exclusion chromatography(separation based on molecular size)

    4. Ion exchange chromatography(separation based on charge)

  • 7/28/2019 Lecture HPLC Student

    67/243

    67

    Partition chromatography Is a chromatographic technique in which solute are

    separated based on their partition between a liquid

    mobile phase & a liquid stationary phase coated ona solid support.

  • 7/28/2019 Lecture HPLC Student

    68/243

    68

  • 7/28/2019 Lecture HPLC Student

    69/243

    69

    The most widely used type of HPLC.

    Subdivided into :

    1. Liquid-liquid partition chromatography.2. Liquid-bonded phase chromatography.

  • 7/28/2019 Lecture HPLC Student

    70/243

    70

    1. Liquid-liquid partition chromatography

    9 The stationary phase is a solvent that is held inplace by adsorption on the surface of packing

    particles.

    9 The liquid coat might itself be soluble in the mobile

    phase which will cause the column to bleed.

    9 To prevent loss of stationary phase, liquid-

    bonded-phase partition chromatography is used.

  • 7/28/2019 Lecture HPLC Student

    71/243

    71

    2. Liquid-bonded-phase partition chromatography

    9 The stationary phase is an organic species thatis attached to the surface of the packing particles

    by chemical bonds.

  • 7/28/2019 Lecture HPLC Student

    72/243

    72

    Most bonded phase packings are prepared by

    reaction of an organochlorosilane with the OHgroups formed on the surface of silica particles byhydrolysis in hot, dilute hydrochloric acid.

    The product is an organosiloxane.

    Bonded phase packings

  • 7/28/2019 Lecture HPLC Student

    73/243

    73

    Bondedstationaryphase

    Alkyl /substituted

    alkyl group

  • 7/28/2019 Lecture HPLC Student

    74/243

    74

    The reaction for one such SiOH to produce C18 :

  • 7/28/2019 Lecture HPLC Student

    75/243

    75

    R Structure Polarity

    OctadecylOctyl

    Phenyl

    (CH2)17CH3 (C18)(CH2)7CH3 (C8)

    (CH2)3C6H5

    NON

    POLAR

    AminoCyano

    Diol

    (CH2)3NH2(CH2)3CN

    (CH2)2OCH2CH(OH)CH2(OH)

    POLAR

    Many different polarities forthe bonded stationary phaseare available

  • 7/28/2019 Lecture HPLC Student

    76/243

    76

    REVERSED PHASE

    NORMAL PHASE

  • 7/28/2019 Lecture HPLC Student

    77/243

    77

    It is not possible to bond all thesurface silanol (Si-OH) groups.

    Unreacted silanols are capableofadsorbing polar molecules.

  • 7/28/2019 Lecture HPLC Student

    78/243

    78

    Can be reduced by

    end-capping process :

    Silanols are reacted with asilylating agent e.g.trimethylchlorosilane.

  • 7/28/2019 Lecture HPLC Student

    79/243

    79

    2 types of partition chromatography aredistinguishable based on the relative polarities

    of the mobile & stationary phases.

    1. Normal Phase

    2. Reversed Phase

  • 7/28/2019 Lecture HPLC Student

    80/243

    80

    The stationary phase is a bonded siloxane with a polarfunctional group (e.g. cyano < diol < amino ELUTED FIRST.

    Polar molecules- retained by stationary phase--> ELUTED LATER.

  • 7/28/2019 Lecture HPLC Student

    82/243

    82

    Example :

  • 7/28/2019 Lecture HPLC Student

    83/243

    83

    i. Discuss the relative polarity of the compounds in the

    above analysis.

    Sample problem 7

  • 7/28/2019 Lecture HPLC Student

    84/243

    84

    Partition HPLC : Reversed phase (RP) The polarity of stationary phase is reversed (non-polar;

    C18 > C8 > phenyl groups).

    The mobile phase is polar(H2

    O, methanol, acetonitrile,THF orusually is a mixture of H2O with one of theorganic solvents).

    The organic solvent is called the modifier.

  • 7/28/2019 Lecture HPLC Student

    85/243

    85

    Metanol is used for acidic compounds. Acetonitrile for basic compounds.

    Tetrahydrofuran for those with large dipoles.

    Eluent strength is increased by adding a lesspolar solvent.

    A less polar solvent has a higher eluent strength

  • 7/28/2019 Lecture HPLC Student

    86/243

    86

    Hydrophobic groups (C18, C8, phenyl group) haveto be chemically bonded on to matrix.

    Octadecylsilane (ODS or C18H37 or C18 chain)widely used.

    Gives non-polar (hydrophobic) interaction with

    sample molecules & eluents.

  • 7/28/2019 Lecture HPLC Student

    87/243

    87

    CH3 CH2COOCH3

    CH3 CH2COOCH3

    Silica-C18 (ODS)

    Hydrophobic interaction

  • 7/28/2019 Lecture HPLC Student

    88/243

    88

    In RP chromatography,

    Polar sample molecules - interact least with

    stationary phase --> ELUTED FIRST.

    Non-polar- retained by stationary phase --> ELUTED LATER.

    These phases retain non polar comps inpreference to polar comps

  • 7/28/2019 Lecture HPLC Student

    89/243

    89

  • 7/28/2019 Lecture HPLC Student

    90/243

    90

    9 The most polar component is eluted first; increasingthe polarity of the mobile phase increases the elutiontime.

    9 The affinity of the hydrophobic analyte for thehydrophobic stationary phase stronger relative to the

    now more hydrophilic (polar) mobile phase.

    9 Can decrease tR by adding more organic solvent tothe eluent.

  • 7/28/2019 Lecture HPLC Student

    91/243

    91

    A

    B

    C

    0 0 05 5 5 10 (min)

    A

    AB

    B

    C

    C

    ACN / H2O(70/30) (60/40) (50/50)

    HighLow Polarity of Mobile phase

    Column : Finepak SIL C18

    Retention behavior in RP HPLC

    < <

  • 7/28/2019 Lecture HPLC Student

    92/243

    92

    In normal phase chromatography :

    9 The least polar component is eluted first ;increasing the polarity of the mobile phase

    decreases the elution time.

    Solvents

  • 7/28/2019 Lecture HPLC Student

    93/243

    Solvents polarity93

    water > methanol > acetonitrile > ethanol >

    propanol = ethyl acetate > tetrahydrofuran > butanol> chloroform > dichloromethane > n-hexane >

    cyclohexane

    Polar solvents

    Non-polar Solvents

    Polarities of organic functional groups

  • 7/28/2019 Lecture HPLC Student

    94/243

    Polarities of organic functional groups

    94

    Hydrocarbons < olefins < aromatic hydrocarbons