Lecture HPLC Student
-
Upload
nur-hariani -
Category
Documents
-
view
217 -
download
0
Transcript of Lecture HPLC Student
-
7/28/2019 Lecture HPLC Student
1/243
1
TOPIC
-
7/28/2019 Lecture HPLC Student
2/243
2
Able to :
1. Explain the fundamental concepts & theories ofseparation techniques in HPLC.
2. Sketch & label the schematic diagrams in HPLC.3. Identify the strength & limitations of each type of HPLC
technique.
4. Suggest & justify the most suitable & efficient separationtechnique to be employed for an analysis.
Course Learning Outcomes
-
7/28/2019 Lecture HPLC Student
3/243
3
Introduction Mobile phase is liquid.
Analytes are first dissolved in a solvent / mobile
phase, then passed through column under highpressure.
The separation dependent upon interaction ofthe analytes with mobile & stationary phases.
-
7/28/2019 Lecture HPLC Student
4/243
HPLC applications4
Analysis ofmixtures of non-volatile or
thermally unstable compounds (thosethat cannot be analyzed by GC)
-
7/28/2019 Lecture HPLC Student
5/243
5
To hold the
liquid mobilephase
Topressurizethe liquidmobile
phase
To allowinjection ofthe samplemixture
Columncontainingthe bed ofstationaryphase
To detect thepresence ofcomponents
as they exitthe column
To recordthe detectorsignal
-
7/28/2019 Lecture HPLC Student
6/243
6
HPLC
setup
http://upload.wikimedia.org/wikipedia/en/4/4b/Agilent1200HPLC.jpg -
7/28/2019 Lecture HPLC Student
7/243
Mobile phase delivery7
Consists of pump to provide the high pressure required.
Equipped with one/more glass reservoirs. Require very clean & pure HPLC grade solvents to prevent
column degradation with impurities & to minimize detectorbackground signals from contaminants (usually UVtransparent).
Solvents must be filtered (to remove dust) through 0.2/0.45m filters & degassed.
-
7/28/2019 Lecture HPLC Student
8/243
8
Membrane Filter Holder Assembly
-
7/28/2019 Lecture HPLC Student
9/243
9
Degassing (removal of dissolved N2
, O2
, CO2
..) by :
9 Vacuum devices (vacuum applied to headspaceabove solvent).
9 Ultra sonication (high frequency vibration drives
gasses out of solvent).9 Heating (decreases solubility of gases).
9 He sparging. Sparging is a process in whichdissolved gases are swept out of a solvent.
-
7/28/2019 Lecture HPLC Student
10/243
10
If solvents are not degassed, gases that are solubleat high pressure (ie. 4,000 psi) become insoluble atlow pressure (ie. at the end of the column or
detector).
Gas bubbles create difficulties with pumps
(destabilize pressure), columns (bad separation) &detectors.
Dissolved O2 absorb UV radiation in the 250-200 nmwavelength range & interfering with UV detection.
-
7/28/2019 Lecture HPLC Student
11/243
Sample problem 111
Why is high pressure is needed in HPLC?
-
7/28/2019 Lecture HPLC Student
12/243
Elution12
The mobile phase may comprise of a single solvent or,
Various solvents can be mixed to give mixture with asuitable polarity.
Isocratic elution
The composition of the mobile phase remainsconstant during the separation
-
7/28/2019 Lecture HPLC Student
13/243
13
Binary gradient 2 solvents (very common).
Tertiary gradient 3 solvents (quite common).
Quaternary gradient 4 solvents (very rarely).
Gradient elution
Solvent composition of the mobile phase is
changed with time during the separation
-
7/28/2019 Lecture HPLC Student
14/243
14
E.g. of a gradient :Starting at 10% methanol & ending at 90% methanolafter 20 minutes.
The two components of the mobile phase are
typically termed "A" & "B";9 A is the "weak" solvent which allows the solute to
elute only slowly.9 B is the "strong" solvent which rapidly elutes the
solutes from the column.
-
7/28/2019 Lecture HPLC Student
15/243
15
Solvent A is often water, while B is an organicsolvent miscible with water, such as acetonitrile,methanol, THF or isopropanol.
Organic solvent of the mobile phase is to reducethe retarding strength of the aqueous component.
-
7/28/2019 Lecture HPLC Student
16/243
16
Different compounds are eluted by increasing thestrength of the organic solvent.
The sample is injected while a weaker mobile phase isbeing applied to the system.
The strength of the mobile phase is later increased in
increments by raising the organic solvent fraction, whichsubsequently results in elution of retained components.
-
7/28/2019 Lecture HPLC Student
17/243
17
Advantages :
9 Decreases the retention of the later-elutingcomponents so that they elute faster, giving
narrower (& taller) peaks for most components.
9 Improves the peak shape for tailed peaks.
-
7/28/2019 Lecture HPLC Student
18/243
Pumping system18
Pumping system pumps (force) the mobile phase
through the manual injector, column, detector & finallyto the waste container.
The quality of a pump for HPLC is measured by howsteady & reproducible a flow it can produce.
A fluctuating flow rate can create detector noise thatobscures weak signals.
-
7/28/2019 Lecture HPLC Student
19/243
19
Requirements for HPLC :9 Pressures up to 6000 psi.9
Pulse free, prevents remixing of solutes.9 Control flow rate from 0.1 to 10 mL/min.
-
7/28/2019 Lecture HPLC Student
20/243
20
Reciprocating pump
9 The most used.
9
Mechanical high pressure liquid pumpingachieved by piston, check valves.
9 Pressure surges can be somewhat overcomewith pulse dampeners.
9 Advantages : Provide constant flow rates &independent of solvent viscosity or columnbackpressure.
-
7/28/2019 Lecture HPLC Student
21/243
21
Reciprocating pump
-
7/28/2019 Lecture HPLC Student
22/243
22
Displacement pump
9 Continuous pressure applied to large volume(syringe pump).
9 Advantage is lack of pressure surges.
9 Disadvantage is price.
-
7/28/2019 Lecture HPLC Student
23/243
Sample injection system23
Injection port :
9 Device for injection of sample into system.
9 Consists of a stainless steel ring with 6 differentports, 1 to the column.
9 Has interchangeable sample loops, which holda fix volume.
-
7/28/2019 Lecture HPLC Student
24/243
24
9 Sample volume is no more than volume
indicated on a loop.9 Samples size ranges from 2-1000 L.
9 Samples introduced manually into the valve witha syringe to fill the sample loop.
9 Sample injection only with flat end needle (toprevent damage to the injection port).
X
-
7/28/2019 Lecture HPLC Student
25/243
25
Manual injection port
Sample loop
-
7/28/2019 Lecture HPLC Student
26/243
26
Injection valve
-
7/28/2019 Lecture HPLC Student
27/243
27
Injection procedure :
1. Place in load position& load sample into loop.
2. Rotate to inject position the content of thesample loop is injected into the column at highpressure.
-
7/28/2019 Lecture HPLC Student
28/243
28
Can inject a large number ofsamples automatically.
Inject continuously variablevolume 1 L1 mL.
Small vials with a septum placedin a tray.
Needle penetrates septum &
withdraws the sample.
Valve introduces the sample into
the column.
Autosamplers
-
7/28/2019 Lecture HPLC Student
29/243
Columns29
1. Guard column
2. Analytical column
-
7/28/2019 Lecture HPLC Student
30/243
30
Guard column (precolumn)
310 cm long, placed between the injector & the
analytical column.
Contains the same packing as the analytical column.
-
7/28/2019 Lecture HPLC Student
31/243
31
Function : To protect & increase the lifetime of the
analytical column by :
9 Retain debris (eg. pump seal fragments), particulatematter & contaminants that would otherwise get onanalytical column & foul/clog it & changing columnefficiency.
9 Retains highly sorbed compounds that would be
caught on & not be eluted from the analytical column. Must be replaced or regenerated periodically.
-
7/28/2019 Lecture HPLC Student
32/243
32
Sample problem 2For the analysis of biological sample using HPLC,a guard column is normally required.
-
7/28/2019 Lecture HPLC Student
33/243
33
Analytical column Stainless steel.
10-30 cm long (10,15 & 25 cm).
4-10 mm internal diameter (3.9or 4.6 mm).
3, 5 or 10 m particle size (dp).
40,000-60,000 plates/m.
Eg. 5 m ODS column (250 x 4.5 mm)
-
7/28/2019 Lecture HPLC Student
34/243
34
van Deemter equation,
H = A + B/v + Cv
Which of the 3 components is the largestcontribution to H? Consider the following :
9 B/v effects diffusion is 100x less in liquidsthan in the gas & can be neglected.
Band Broadening in LC
-
7/28/2019 Lecture HPLC Student
35/243
35
9 Cv effects we will assume that mass transport(M.T.) effects are the largest contribution.
The mobile phase mass transfer coefficient :
CMv = (fM(k)dp2/DM)v
Smaller dp increases the surface area/volume
ratio & thus decreases M.T. in the m.p.
-
7/28/2019 Lecture HPLC Student
36/243
36
Reduction of dp from 45 to 6 m results in a tenfoldor more decrease in H
-
7/28/2019 Lecture HPLC Student
37/243
37
Sample problem 3Discuss the effect of the particle size on the separation
performance in packed column liquid chromatography.
-
7/28/2019 Lecture HPLC Student
38/243
38
Stationary phase Normally, columns are filled with silica gel because its :
9 particle shape (spherical, can be packed morehomogeneously),
9 surface properties,
9 pore size (60-100 range)
structure help to get a good separation.
-
7/28/2019 Lecture HPLC Student
39/243
39
Silica structures
-
7/28/2019 Lecture HPLC Student
40/243
40
Silica :
9 Is wetted by nearly every potential mobile phase.
9 Is inert to most compounds.
9 Has a high surface activity which can bemodified easily with water & other agents.
9 Can be used to separate a wide variety ofchemical compounds.
-
7/28/2019 Lecture HPLC Student
41/243
41
Limitation of silica stationary phase :
X Operating pH range 2-8.
X Dissolves above pH 8. Crosslinked polymericparticles (eg., polystyrene / polymethacrylates) areused for separation of bases.
X Siloxane bond (Si-O-Si) to the stationary phasehydrolyses below pH 2.
These defects will cause changes in tR & loss ofresolution.
-
7/28/2019 Lecture HPLC Student
42/243
Detector types42
UV absorbance
9 Variable Wavelength UV-Vis detector.9 Diode Array Detector (DAD) : measure absorbance
across a broad spectrum of wavelengths
simultaneously. Fluorescence
Electrochemical
Refractive Index
Mass spectrometric
-
7/28/2019 Lecture HPLC Student
43/243
43
HPLC detectors will depend on the nature of thesample.
The most widely used detectors for HPLC are
based on the absorption of ultraviolet or visibleradiation.
UV region : 100-400 nm
Vis region : 400-780 nm
-
7/28/2019 Lecture HPLC Student
44/243
44
The detectors often make use of the 254 nm &280 nm lines from a mercury source because
many organic functional groups absorb in theregion.
Usually set the UV at 254 nm. This will detect any to * transitions. So, it would detect anything
with the benzene ring.
-
7/28/2019 Lecture HPLC Student
45/243
45
Based on electronic transitions withinmolecules.
Most common type of detector for LC.
Fixed wavelength, Hg lamp 254 nm ( => *).
UV-Vis absorbance detector
-
7/28/2019 Lecture HPLC Student
46/243
46
A diagram of the UV detector
-
7/28/2019 Lecture HPLC Student
47/243
47
The detectorsensorconsists of a small cylindrical
cell (normally 2.0 to 10 l in volume) through which,the eluent from the column flows.
UV light from a UV lamp (usually a Hg dischargelamp) emitting UV light mostly at 254 nm.
-
7/28/2019 Lecture HPLC Student
48/243
48
Variable wavelength, selectable for specificwavelengths in the visible region (eg.deuterium lamp for use down to 190-400 nm).
Still limited to single wavelengths.
-
7/28/2019 Lecture HPLC Student
49/243
49
Analyte must absorb in the visible region (haschromophore).
You might wonder why the solventsused don't absorb UV light. They do!
But different compounds absorb moststrongly in different parts of the UVspectrum.
-
7/28/2019 Lecture HPLC Student
50/243
50
Methanol, for eg., absorbs at wavelengths below205 nm & water below 190 nm.
If you were using a methanol-water mixture as
the solvent, you would therefore have to use awavelength greater than 205 nm to avoid falsereadings from the solvent.
-
7/28/2019 Lecture HPLC Student
51/243
51
Measure absorbance across a broad spectrum ofwavelengths simultaneously.
Allows for the recording of the entire spectrum of eachsolute as it passed through the diode array detector.
Utilizes a deuterium or xenon lamp that emits light overthe UV spectrum range.
UV-Vis absorbance detectorDIODE ARRAY (DAD)
-
7/28/2019 Lecture HPLC Student
52/243
52
Sample problem 4
A sample mixture containing acetophenone, phenol, 1-hydroxyphenol & methyl benzene was analyzed usingHPLC with a mobile phase system of a ACN/H2O 80/20 &
a UV detector set at wavelength 420 nm.Quantitative analysis was done using external standard
method. Discuss the reason for each of the followingproblems encountered & suggest an approach toovercome the problem.
-
7/28/2019 Lecture HPLC Student
53/243
53
i. Very low detection signals were obtained.
-
7/28/2019 Lecture HPLC Student
54/243
54
ii. After doing 5 replicate measurements, the peakareas obtained vary considerably.
-
7/28/2019 Lecture HPLC Student
55/243
55
Based on the n, index of refraction. How much the
light is bent in a substance is unique for all liquids &relative to their density.
Advantages :9 universal, insensitive to flow, economical, robust.
Disadvantages :X not sensitive, highly temperature dependent
Refractive Index detector
-
7/28/2019 Lecture HPLC Student
56/243
-
7/28/2019 Lecture HPLC Student
57/243
57
Fluorescence :
The most sensitive, 10-1000 times higher than thatof UV detector for strong UV absorbing materials.
Electrochemical :Based on the measurements of the current resulting
from oxidation/reduction reaction of the analyte at asuitable electrode.
Other detectors
-
7/28/2019 Lecture HPLC Student
58/243
58
Conductivity :
The conductivity of the column effluent iscontinuously measured & the appearance of the
analyte in the cell is indicated by a change inconductivity.
-
7/28/2019 Lecture HPLC Student
59/243
59
Mass spectrometry (MS):
Examining a compound (in a gas phase), that will providethe structural information of each compounds match fromthe library of mass spectra of known compounds storedon the computer & thus can identify the compound.
Measures the mass-to-charge-ratio (m/z) of charged
analytes.
-
7/28/2019 Lecture HPLC Student
60/243
60
Eg. mass spectrum for benzene
-
7/28/2019 Lecture HPLC Student
61/243
61
It is not a simple matter to connect the MS to LC,because
9 LC operates in the liquid phase whereas MS is agas phase method.
9 Analytes are non volatile & may be thermally labile
but must be presented in gaseous form.
In order to allow separated components of a mixture to
be passed sequentially from the LC into the MS, aninterface is needed.
-
7/28/2019 Lecture HPLC Student
62/243
62
Sample problem 5Identification of compound is more confirmative by using
mass spectrometer as a detector compare to UV detector.
-
7/28/2019 Lecture HPLC Student
63/243
Sample problem 663
Explain the advantage of using MS detector.
-
7/28/2019 Lecture HPLC Student
64/243
Performance of HPLC detectors64
HPLC detector LOD
Absorbance 10 pgFluorescence 10 fg
Electrochemical 100 pgRefractive index 1 ng
Conductivity 100 pg 1 ngMass spectrometry < 1 pg
-
7/28/2019 Lecture HPLC Student
65/243
65
Chemical derivatization of a sample in HPLC isundertaken to convert a non-UV/fluorescenceresponding analyte into one that is.
The derivatization step may enhance the separation.
Sample derivatization
-
7/28/2019 Lecture HPLC Student
66/243
Types of liquid chromatography66
1. Partition chromatography
(separation based on polarity)
2. Adsorption chromatography
3. Size exclusion chromatography(separation based on molecular size)
4. Ion exchange chromatography(separation based on charge)
-
7/28/2019 Lecture HPLC Student
67/243
67
Partition chromatography Is a chromatographic technique in which solute are
separated based on their partition between a liquid
mobile phase & a liquid stationary phase coated ona solid support.
-
7/28/2019 Lecture HPLC Student
68/243
68
-
7/28/2019 Lecture HPLC Student
69/243
69
The most widely used type of HPLC.
Subdivided into :
1. Liquid-liquid partition chromatography.2. Liquid-bonded phase chromatography.
-
7/28/2019 Lecture HPLC Student
70/243
70
1. Liquid-liquid partition chromatography
9 The stationary phase is a solvent that is held inplace by adsorption on the surface of packing
particles.
9 The liquid coat might itself be soluble in the mobile
phase which will cause the column to bleed.
9 To prevent loss of stationary phase, liquid-
bonded-phase partition chromatography is used.
-
7/28/2019 Lecture HPLC Student
71/243
71
2. Liquid-bonded-phase partition chromatography
9 The stationary phase is an organic species thatis attached to the surface of the packing particles
by chemical bonds.
-
7/28/2019 Lecture HPLC Student
72/243
72
Most bonded phase packings are prepared by
reaction of an organochlorosilane with the OHgroups formed on the surface of silica particles byhydrolysis in hot, dilute hydrochloric acid.
The product is an organosiloxane.
Bonded phase packings
-
7/28/2019 Lecture HPLC Student
73/243
73
Bondedstationaryphase
Alkyl /substituted
alkyl group
-
7/28/2019 Lecture HPLC Student
74/243
74
The reaction for one such SiOH to produce C18 :
-
7/28/2019 Lecture HPLC Student
75/243
75
R Structure Polarity
OctadecylOctyl
Phenyl
(CH2)17CH3 (C18)(CH2)7CH3 (C8)
(CH2)3C6H5
NON
POLAR
AminoCyano
Diol
(CH2)3NH2(CH2)3CN
(CH2)2OCH2CH(OH)CH2(OH)
POLAR
Many different polarities forthe bonded stationary phaseare available
-
7/28/2019 Lecture HPLC Student
76/243
76
REVERSED PHASE
NORMAL PHASE
-
7/28/2019 Lecture HPLC Student
77/243
77
It is not possible to bond all thesurface silanol (Si-OH) groups.
Unreacted silanols are capableofadsorbing polar molecules.
-
7/28/2019 Lecture HPLC Student
78/243
78
Can be reduced by
end-capping process :
Silanols are reacted with asilylating agent e.g.trimethylchlorosilane.
-
7/28/2019 Lecture HPLC Student
79/243
79
2 types of partition chromatography aredistinguishable based on the relative polarities
of the mobile & stationary phases.
1. Normal Phase
2. Reversed Phase
-
7/28/2019 Lecture HPLC Student
80/243
80
The stationary phase is a bonded siloxane with a polarfunctional group (e.g. cyano < diol < amino ELUTED FIRST.
Polar molecules- retained by stationary phase--> ELUTED LATER.
-
7/28/2019 Lecture HPLC Student
82/243
82
Example :
-
7/28/2019 Lecture HPLC Student
83/243
83
i. Discuss the relative polarity of the compounds in the
above analysis.
Sample problem 7
-
7/28/2019 Lecture HPLC Student
84/243
84
Partition HPLC : Reversed phase (RP) The polarity of stationary phase is reversed (non-polar;
C18 > C8 > phenyl groups).
The mobile phase is polar(H2
O, methanol, acetonitrile,THF orusually is a mixture of H2O with one of theorganic solvents).
The organic solvent is called the modifier.
-
7/28/2019 Lecture HPLC Student
85/243
85
Metanol is used for acidic compounds. Acetonitrile for basic compounds.
Tetrahydrofuran for those with large dipoles.
Eluent strength is increased by adding a lesspolar solvent.
A less polar solvent has a higher eluent strength
-
7/28/2019 Lecture HPLC Student
86/243
86
Hydrophobic groups (C18, C8, phenyl group) haveto be chemically bonded on to matrix.
Octadecylsilane (ODS or C18H37 or C18 chain)widely used.
Gives non-polar (hydrophobic) interaction with
sample molecules & eluents.
-
7/28/2019 Lecture HPLC Student
87/243
87
CH3 CH2COOCH3
CH3 CH2COOCH3
Silica-C18 (ODS)
Hydrophobic interaction
-
7/28/2019 Lecture HPLC Student
88/243
88
In RP chromatography,
Polar sample molecules - interact least with
stationary phase --> ELUTED FIRST.
Non-polar- retained by stationary phase --> ELUTED LATER.
These phases retain non polar comps inpreference to polar comps
-
7/28/2019 Lecture HPLC Student
89/243
89
-
7/28/2019 Lecture HPLC Student
90/243
90
9 The most polar component is eluted first; increasingthe polarity of the mobile phase increases the elutiontime.
9 The affinity of the hydrophobic analyte for thehydrophobic stationary phase stronger relative to the
now more hydrophilic (polar) mobile phase.
9 Can decrease tR by adding more organic solvent tothe eluent.
-
7/28/2019 Lecture HPLC Student
91/243
91
A
B
C
0 0 05 5 5 10 (min)
A
AB
B
C
C
ACN / H2O(70/30) (60/40) (50/50)
HighLow Polarity of Mobile phase
Column : Finepak SIL C18
Retention behavior in RP HPLC
< <
-
7/28/2019 Lecture HPLC Student
92/243
92
In normal phase chromatography :
9 The least polar component is eluted first ;increasing the polarity of the mobile phase
decreases the elution time.
Solvents
-
7/28/2019 Lecture HPLC Student
93/243
Solvents polarity93
water > methanol > acetonitrile > ethanol >
propanol = ethyl acetate > tetrahydrofuran > butanol> chloroform > dichloromethane > n-hexane >
cyclohexane
Polar solvents
Non-polar Solvents
Polarities of organic functional groups
-
7/28/2019 Lecture HPLC Student
94/243
Polarities of organic functional groups
94
Hydrocarbons < olefins < aromatic hydrocarbons