LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND...

54
LEADLESS AND NON TRANS - VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD

Transcript of LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND...

Page 1: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral

LEADLESS AND NON TRANS-VENOUS TECHNOLOGY:

CURRENT AND FUTURE DIRECTIONS

JONATHAN LOWY MD

Page 2: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral
Page 3: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral

• 58 year old woman with severe rheumatic mitral stenosis, Tricuspid Insufficiency, Aortic Stenosis factor 5 Leiden, and Type II DM had prior AVR replacement in Ukraine 10 years

• Complete Heart block after surgery sternal infection, and she developed bilateral venous thrombosis of her upper extremity venous return.

• Epicardial post-surgical pacing wires began to fail after 7days with intermittent capture• Temporary femoral Trans-venous pacing was required • Surgeons did not want to place an epicardial lead and EP was uncomfortable with

anything permanent in the venous system due to risk of thrombosis. • She remained in the ICU for 36 days with temporary trans-venous pacing before being

transferred to our institution

Page 4: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral
Page 5: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral
Page 6: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral

Tong, F Reddy,V Circulation. 2017;135:1458–1470. DOI: 10.1161/CIRCULATIONAHA.116.025037

Page 7: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral
Page 8: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral

Why leadless pacemakers?• 1,000,000 pacemakers implanted worldwide

• 200,000 implanted in the US every year

Page 9: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral
Page 10: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral
Page 11: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral
Page 12: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral
Page 13: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral
Page 14: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral
Page 15: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral
Page 16: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral

Leadless pacing

Page 17: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral

Requirements for a leadless pacemakerDeliverableStable and SafeBattery life 8-12 yearsRate responseMRI compatibleTelemetry Remote Monitoring

Retrievable/ExtractableModular

Page 18: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral
Page 19: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral

CURRENTLY THE MICRA SINGLE CHAMBER VVIR PACEMAKER IS

THE ONLY FDA APPROVED LEADLESS CARDIAC PACEMAKER

Page 20: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral

NANOSTIM: ST. JUDE MEDICAL(NOT FDA APPROVED)

N Engl J Med 2015; 373:1125-113

Page 21: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral

Nanostim Implants Suspended in 2016 prior to FDA approvalBattery failure in a small number of implants resulted in voluntary recall in 2016Delivery button malfunction in 2017

Per Abbott they are working on a dual chamber version of Nanostim with updated battery and design. Plans for a clinical IDE trial in the “near future” The plans for the dual chamber system have not been publically disclosed.

Page 22: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral

MICRA MEDTRONIC-FDA APPROVED

Page 23: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral

MICRA LEADLESS PACEMAKEREasy to Implant Battery life 8-12 yearsMRI compatibleRate response with 3 vector accelerometerRetrievable/Extractable +/-Reliable In animal models up to 3 devices implanted in the RV

Page 24: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral
Page 25: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral
Page 26: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral
Page 27: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral

Updated performance of the Micra transcatheter pacemaker in the real-world setting: A comparison to the investigational study and a transvenous historical control

Page 28: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral

Boston Scientific “EMPOWR”(Investigational Device)

32.1 mm x 6.0 mm0.8 cc

EMPOWER Delivery and Retrieval(Investigational Device & System)

Dedicated delivery and retrieval catheters

tether loop/ retrieval groove

20.7 Fr Preloaded delivery catheter withextendable inner catheter

Dedicated retrieval system with single & tri-loop snares

Active fixationtalons

Page 29: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral

FDA PMA FOR MICRA:

Page 30: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral
Page 31: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral
Page 32: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral
Page 33: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral

Data for 40k patients show 20% TV-lead malfunction at 10 years

1. Koneru, JN, Jones, PW, Hammill, EF, Wold, N, and Ellenbogen, KA, Risk Factors and Temporal Trends of Complications Associated With Transvenous Implantable Cardiac Defibrillator Leads. J Am Heart Assoc, 2018. 7(10).

50%

55%

60%

65%

70%

75%

80%

85%

90%

95%

100%

1 2 3 4 5 6 7 8 9 10

Free

dom

from

Mec

hani

cal C

ompl

icat

ions

Mechanical Failure of Transvenous LeadsTrends from 10 Year Optum Data for > 40,000 Patients

Page 34: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral

WHY DO WE NEED AN ALTERNATIVE TO TRANSVENOUS ICDS?

• Patients are living longer requiring lead durability that may not last a patient’s lifespan• Sicker patients with more systemic disease increasing the risk of infection• A vast majority of patients do not need pacing or ATP• Extraction of Transvenous systems is not without significant risks

Page 35: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral
Page 36: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral

Evolution of S-ICD therapy

2å0022009

2015

2016

1st Generation

2nd Generation

3rd Generation

• EMBLEM Electrode (Model 3501)

• Electrode Delivery System (EDS)

• 2 Incision Technique Labeling

2017

• SMART Pass1

• MRI compatible1**

• AF MonitorTM 1

• Smaller, thinner,

longer-lasting device*

• LATITUDETM

compatible1

2012 Automated Screening Tool

Page 37: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral
Page 38: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral
Page 39: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral
Page 40: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral

DETECTION

Page 41: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral
Page 42: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral

CONDITIONAL ZONE DISCRIMINATORS

Page 43: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral

ONGOING TRIALS:• UNTOUCHED:• Understanding Outcomes With the EMBLEM™ S-ICD in Primary Prevention Patients With Low Ejection Fraction

(UNTOUCHED)

• LATE BREAKING UPDATE AT HRS 2019

• PRAETORIAN TRIAL:• Prospective, RAndomizEd comparison of subcuTaneOus and tRansvenous ImplANtable cardioverter-defibrillator therapy.

• ATLAS S-ICD• AVOID TRANSVENOUS LEADS IN APPROPRIATE SUBJECTS

• RANDOMIZE S-ICD VS SINGLE CHAMBER ICD

• PRAETORIAN-DFT• RANDOMIZING DFT VS NO DFT WITH XRAY TO CONFIRM APPROPRIATE DEVICE LOCATION

Page 44: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral
Page 45: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral
Page 46: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral

MEDTRONIC EV SUB-STERNAL ICD-EXPERIMENTAL-

• PILOT TRIAL: SPACE• 9 PATIENTS IMPLANTED IN AUSTRALIA AND

NEW ZEALAND

• HAS ABILITY TO DELIVER ATP AND POST SHOCK PACING

• WORLD WIDE STUDY INCLUDING US SITES NOW UNDWERWAY

IMAGE COURTESY MEDTRONIC

Page 47: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral
Page 48: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral
Page 49: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral

Courtesy of Brandon Fellows BSCI

Page 50: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral
Page 51: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral
Page 52: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral

WISE-CRT/SELECT LVEXPERIMENTAL NOT FDA APPROVED

Page 53: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral

THE FUTURE:

Page 54: LEADLESS AND NON TRANS-VENOUS TECHNOLOGY: CURRENT …€¦ · TRANS-VENOUS TECHNOLOGY: CURRENT AND FUTURE DIRECTIONS JONATHAN LOWY MD • 58 year old woman with severe rheumatic mitral