LB-CAL-013 · 2021. 5. 29. · Mass/ Electronic Balances Use of mass standards, acc. to EURAMET cg...
Transcript of LB-CAL-013 · 2021. 5. 29. · Mass/ Electronic Balances Use of mass standards, acc. to EURAMET cg...
LB-CAL-013
Emirates International Accreditation Centre
has accredited
Gulf Calibration & Technical Services
Ghala Industrial Area, Muscat- Sultanate of Oman
In accordance with the requirements of
ISO/IEC 17025:2017
General requirements for the competence of testing and calibration laboratories
to undertake the calibration in the attached accreditation scope
This Accreditation is invalid without the attached accreditation scope and shall remain in force within the validity
period printed below, subject to continuing compliance with the requirements of the accreditation criteria.
Validity: 30-03-2021 to 10-09-2022
Initial Accreditation Date: 11-09-2013
Accreditation ScopeLB-CAL-013
Date: 30-03-2021
Scope Issue No. Date
Temperature 5
Balance 4
Force 5
Mass 4
Dimensional 4
Pressure 5
Electrical 2
Volume 2
Humidity 2
Temperature 4
Balance 3
Force 4
Mass 3
Dimensional 3
Pressure 4
Gulf Calibration & Technical Services
Ghala Industrial Area, Muscat- Sultanate of Oman
Valid to: 10-09-2022
Extension in scope and first issuance under the name of
EIAC (which was formerly known as DAC)
First issuance under the name of EIAC (which was
formerly known as DAC)
3/30/2021
Re-issued due to: transition to the new version of ISO/IEC
17025:2017 and comply with the new accreditation
number format
Re-issued due to: transition to the new version of ISO/IEC
17025:2017 and comply with the new accreditation
number format
Re-issued due to: structural amendments made to
methods only, transition to the new version of ISO/IEC
17025:2017 and comply with the new accreditation
number format
Re-issued due to: transition to the new version of ISO/IEC
17025:2017 and comply with the new accreditation
number format
Re-issued due to: structural amendments made to
methods only, transition to the new version of ISO/IEC
17025:2017 and comply with the new accreditation
number format
Re-issued due to: structural amendments made to
methods only, transition to the new version of ISO/IEC
17025:2017 and comply with the new accreditation
number format
Accreditation History
Details
12/5/2019
One Central-Offices 2- Level 3, Suit No.: 302, 303 308.The Trade Centre District, Dubai-United Arab Emirates
Office: +971 4 872 26661 of 2
Accreditation ScopeLB-CAL-013
Date: 30-03-2021
Scope Issue No. Date
Gulf Calibration & Technical Services
Ghala Industrial Area, Muscat- Sultanate of Oman
Valid to: 10-09-2022
3/30/2021Re-issued due to: transition to the new version of ISO/IEC
Accreditation History
Details
Electrical 1
Volume1
Humidity1
Granted accreditation from EIAC 12/5/2019
One Central-Offices 2- Level 3, Suit No.: 302, 303 308.The Trade Centre District, Dubai-United Arab Emirates
Office: +971 4 872 26662 of 2
Accreditation Scope Temperature Calibration
LB-CAL-013
Issue no.: 05
Calibration Field/
Measuring QualityCalibration Method Range and Specification
Calibration Measurement
Capability
(CMC)*
Location
0 °C 0.02 °C
-20 °C to 150 °C 0.05 °C
>150 °C to 250 °C 0.07 °C
30 °C to 250 °C 0.3 °C
>250 °C to 400 °C 0.6 °C
>400 °C to 600°C 1.2 °C
0 °C 0.05 °C
-20 °C to 150 °C 0.1 °C
>150 °C to 250 °C 0.4 °C
>250 °C to 400 °C 0.6 °C
>400 °C to 600 °C 0.8 °C
>600 °C to 1100 °C 2.0 °C
-20 °C to 250 °C 0.25 °C
>250 °C to 400 °C 0.5 °C
>400 °C to 600 °C 1.0 °C
Liquid-in-glass
thermometers
GCTS-SOP-10 -20 °C to 250 °C 0.25 °C Laboratory
Gulf Calibration & Technical Services
Ghala Industrial Area, Muscat- Sultanate of Oman
Date: 30-03-2021 Valid to: 10-09-2022
Digital Thermometers
With PRT sensor
GCTS-SOP-08
In liquid bath
Laboratory
Digital Thermometers
With PRT sensor
GCTS-SOP-08
In dry block
Laboratory
Digital Thermometers
With TC sensor
GCTS-SOP-08
In liquid bath
Laboratory
Digital Thermometers
With TC sensor
GCTS-SOP-08
In dry block
Laboratory
Dial thermometers GCTS-SOP-09 Laboratory/
Customers
Premises
* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing
more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded
uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration
performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
One Central-Offices 2- Level 3, Suit No.: 302, 303 308.The Trade Centre District, Dubai-United Arab Emirates
Office: +971 4 872 26661 of 2
Accreditation Scope Temperature Calibration
LB-CAL-013
Issue no.: 05
Calibration Field/
Measuring QualityCalibration Method Range and Specification
Calibration Measurement
Capability
(CMC)*
Location
Gulf Calibration & Technical Services
Ghala Industrial Area, Muscat- Sultanate of Oman
Date: 30-03-2021 Valid to: 10-09-2022
Digital Thermometers GCTS-SOP-08 Laboratory50 °C to 300 °C 2.0 °C
>300 °C to 500 °C 3.0 °C
Environmental chambers,
Ovens
GCTS-SOP-11 20 °C to 300 °C 0.7 °C
Incubators
(5 sensors)
GCTS-SOP-13 Ambient to 100 °C 0.3 °C
Autoclaves (temperature GCTS-SOP-38 100 °C to 140 °C 0.3 °C Laboratory/
Customers
Premises20 °C to 95 °C 0.25 °C
>95 °C to 200 °C 0.4 °C
Ambient to 300 °C 1 °C
300 °C to 600 °C 3 °C
600 °C to 800 °C 6 °C
800 °C to 1200 °C 10 °C
* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing
more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded
uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration
performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
Liquid bath
(5 sensors)
GCTS-SOP-12 Laboratory/
Customers
Premises
Muffle Furnace
(1 sensor)
GCTS-SOP-16
IR thermometers GCTS-SOP-40 Laboratory
Chillers, Freezers
(9 sensors)
GCTS-SOP-11 .-20 °C to 20 °C 0.5 °C Laboratory/
Customers
Premises
One Central-Offices 2- Level 3, Suit No.: 302, 303 308.The Trade Centre District, Dubai-United Arab Emirates
Office: +971 4 872 26662 of 2
Accreditation ScopeBalance Calibration
LB-CAL-013
Issue no.: 04
Calibration Field/
Measuring QualityCalibration Method Range and Specification
Calibration Measurement
Capability
(CMC)*
Location
Up to 5 kg 5x 10-6 of FS
>5 kg up to 30 kg 1x 10-5 of FS
>30 kg up to 1000 kg 5x10-5 of FS
* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing
more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded
uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration
performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
Mass/ Electronic Balances Use of mass standards, acc.
to EURAMET cg 18:2015;
ASTM E898-20
GCTS-SOP-01
Customers
Premises
Gulf Calibration & Technical Services
Ghala Industrial Area, Muscat- Sultanate of Oman
Date: 30-03-2021 Valid to: 10-09-2022
One Central-Offices 2- Level 3, Suit No.: 302, 303 308.The Trade Centre District, Dubai-United Arab Emirates
Office: +971 4 872 26661 of 1
Accreditation ScopeDimensional Calibration
LB-CAL-013
Issue no.: 04
Calibration Field/
Measuring QualityCalibration Method Range and Specification
Calibration Measurement
Capability
(CMC)*
Location
All types of Calipers
Vernier
Dial
Electronic
GCTS -SOP 05
based on BS EN 887:2008
Up to 300 mm Overall Performance
10+ (15 x *L) μm
Laboratory
External Micrometer GCTS-SOP 06
based on BS EN 870:2008
Up to 100 mm 2 μm Laboratory
Dial Gauges GCTS -SOP 07
based on BS 907:2008
Up to 25 mm 4 μm Laboratory
Below 4 mm 4 µm Laboratory
Above 4 mm 20 µm Laboratory/
Customer
Premises
Performance Testing for
Laboratory Test Sieves
ASTM E2427 & UKAS LAB
22
Below 4 mm 1% Laboratory/
Customer
Premises
Feeler Gauges GCTS-SOP-39 based on BS
957:2008
0.03 mm – 1.00 mm 10 µm Laboratory
GCTS-SOP-14 based on
ASTM E11-17 / ISO 3310-
2:2013(E) / ISO 3310-
1:2016(E)
Laboratory Test Sieves
Gulf Calibration & Technical Services
Ghala Industrial Area, Muscat- Sultanate of Oman
Date: 30-03-2021 Valid to: 10-09-2022
* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing
more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded
uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration
performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
One Central-Offices 2- Level 3, Suit No.: 302, 303 308.The Trade Centre District, Dubai-United Arab Emirates
Office: +971 4 872 26661 of 2
Accreditation ScopeDimensional Calibration
LB-CAL-013
Issue no.: 04
Calibration Field/
Measuring QualityCalibration Method Range and Specification
Calibration Measurement
Capability
(CMC)*
Location
Gulf Calibration & Technical Services
Ghala Industrial Area, Muscat- Sultanate of Oman
Date: 30-03-2021 Valid to: 10-09-2022
All types of Depth
Gauges.(Dial/ Vernier
/Electronic)
GCTS-SOP-43 based on BS
EN ISO 13385-01:2011
Up to 100 mm 10 µm Laboratory
Depth Micrometers GCTS-SOP-44 based on BS
6468:2008
Up to 100 mm 10 µm Laboratory
Profile Projectors GCTS-SOP-45 based on JIS
B 7184 : 1999
Up to 300 mm ( (x-y axis) 2 µm (Linear), 10 min of
ARC (Angular) & 0.5%
(Magnification)
Laboratory/
Customer
Premises
Bore Gauge GCTS-SOP-47 Up to 2 mm travel 3 µm Laboratory
Extensometers: GCTS-SOP-48 based on
ASTM-E-83 & ISO 9513
Up to 50 mm (Extension) &
Gauge Length- up to 600
mm
5 µm Customer
Premises
* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing
more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded
uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration
performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
One Central-Offices 2- Level 3, Suit No.: 302, 303 308.The Trade Centre District, Dubai-United Arab Emirates
Office: +971 4 872 26662 of 2
Accreditation ScopeForce Calibration
LB-CAL-013
Issue no.: 05
Calibration Field/
Measuring QualityCalibration Method Range and Specification
Calibration Measurement
Capability
(CMC)*
Location
Force/ Material Testing
Machines in Compression
Verification/ Calibration of
the force measuring system
by force proving
instruments in
compression, acc. ISO 7500-
1: 2018
Alignment of upper
machine platen acc. EN
12390-4:2019
GCTS-SOP-02
300 kN up to 3000 Kn 0.27 % of reading Customer
Premises
Force/ Flexural Strength
Material Testing Machines
Verification/Calibration of
the force measuring system
by force proving
instruments in
compression, acc. ISO 7500-
1: 2018
GCTS-SOP-02
10 kN up to 250 kN 0.27 % of reading Customer
Premises
Force/ Proving Ring, Load
cell with indicator for CBR,
Marshall testing machine
Verification/ Calibration of
the force measuring system
by force proving
instruments in
compression, acc. ISO 7500-
1: 2018
GCTS-SOP-30
5 kN to 50 kN 1% of reading Laboratory/
Customer
Premises
Gulf Calibration & Technical Services
Ghala Industrial Area, Muscat- Sultanate of Oman
Date: 30-03-2021 Valid to: 10-09-2022
* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing
more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded
uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration
performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
One Central-Offices 2- Level 3, Suit No.: 302, 303 308.The Trade Centre District, Dubai-United Arab Emirates
Office: +971 4 872 26661 of 1
Accreditation ScopePressure Calibration
LB-CAL-013
Issue no.: 05
Calibration Field/
Measuring QualityCalibration Method Range and Specification
Calibration Measurement
Capability
(CMC)*
Location
Gas Pressure
(Negative Gauge)/ Digital
& Analogue Indicating
Devices
Comparison method, acc.
DKD R-6.1:2014
GCTS-SOP-04
100 kPa to 0 kPa 0.05 kPa Laboratory
Gas Pressure (gauge)/
Digital & Analogue
Indicating Devices
Comparison method, acc.
DKD R-6.1:2014
GCTS-SOP-03
0.1 to 3.5 MPa 1 kPa Laboratory
Hydraulic Pressure
(gauge)/ Digital &
Analogue Indicating Devices
Pressure generated by dead
weight tester, acc. DKD R-
6.1:2014 and EURAMET cg
17 v4: 2019
GCTS-SOP-31
0.1 MPa to 140 MPa 8 x 10 -5 pe + 50 Pa Laboratory
Hydraulic Pressure (gauge)
/ Digital & Analogue
Indicating Devices
Comparison method, acc.
DKD R-6.1:2014
GCTS-SOP-03
0.1 MPa to 300 MPa 0.25% of F.S Laboratory/
Customers
Premises
Gas Pressure (gauge)/
Digital & Analogue
Indicating Devices
Pressure generated by dead
weight tester, acc. DKD R-
6.1:2014 and EURAMET cg
17 v4: 2019
GCTS-SOP-031
0.1 MPa to 12 MPa 0.01 % of FS Laboratory
* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing
more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded
uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration
performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
Gulf Calibration & Technical Services
Ghala Industrial Area, Muscat- Sultanate of Oman
Date: 30-03-2021 Valid to: 10-09-2022
One Central-Offices 2- Level 3, Suit No.: 302, 303 308.The Trade Centre District, Dubai-United Arab Emirates
Office: +971 4 872 26661 of 1
Accreditation ScopeMass Calibration
LB-CAL-013
Issue no.: 04
Calibration Field/
Measuring QualityCalibration Method Range and Specification
Calibration Measurement
Capability
(CMC)*
Location
1 mg up to 5 kgOIML
class F2
1/3 of MPE Class F2 Laboratory
10 kg to 20 kg OIML
class M1
1/3 of MPE Class M1 Laboratory
* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing
more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded
uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration
performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
Gulf Calibration & Technical Services
Ghala Industrial Area, Muscat- Sultanate of Oman
Valid to: 10-09-2022 Date: 30-03-2021
Mass
Conventional Mass
Substitution Weighing
with Air buoyancy Corr.
OIML R 111
www.eiac.gov.aeOne Central-Offices 2- Level 3, Suit No.: 302, 303 308.The Trade Centre District, Dubai-United Arab Emirates
Office: +971 4 872 2666
1 of 1
Accreditation ScopeVolume Calibration
LB-CAL-013
Issue no.: 02
Calibration Field/
Measuring QualityCalibration Method Range and Specification
Calibration Measurement
Capability
(CMC)*
Location
1 to 100 µl 0.19 μl
100 to 500 µl 0.22 μl
500 to 1000 µl 0.30 μl
1 to 10 ml 3.3 µl
1 to 50 ml 0.05 ml
> 50 ml to 500 ml 0.41 ml
˃500 to 2000 ml 1.3 ml
1 to 10 ml 0.06 ml
5 to 25 ml 0.3 ml
10 to 100 ml 0.73 ml
25 to 250 ml 0.85 ml
50 to 500 ml 3.2 ml
100 to 1000 ml 6.7 ml
200 to 2000 ml 10 ml
0.1 to 1 ml 0.012 ml
>1 ml to 10 ml 0.02 ml
˃10 ml to 50 ml 0.04 ml
Volumetric Glass Wares
(Measuring cylinders)
GCTS-SOP-25 based on
ISO 4787 & EURAMET cg19
Gulf Calibration & Technical Services
Ghala Industrial Area, Muscat- Sultanate of Oman
Date: 30-03-2021 Valid to: 10-09-2022
Volumetric Glass Wares
(Graduated Pipettes.)
GCTS-SOP-25 based on
ISO 4787 & EURAMET cg19
* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing
more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded
uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration
performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
Laboratory
Laboratory
Laboratory
Laboratory
Piston Operated Volumetric
Apparatus (Micro-Pipettes
& Pipettes)
GCTS-SOP-51 based on
ISO 8655-6, ISO/TR 20461
& ISO/TR 20461
Corr1:2008
Volumetric Glass Wares
(Measuring flasks)
GCTS-SOP-25 based on
ISO 4787 & EURAMET cg19
One Central-Offices 2- Level 3, Suit No.: 302, 303 308.The Trade Centre District, Dubai-United Arab Emirates
Office: +971 4 872 26661 of 2
Accreditation ScopeVolume Calibration
LB-CAL-013
Issue no.: 02
Calibration Field/
Measuring QualityCalibration Method Range and Specification
Calibration Measurement
Capability
(CMC)*
Location
Gulf Calibration & Technical Services
Ghala Industrial Area, Muscat- Sultanate of Oman
Date: 30-03-2021 Valid to: 10-09-2022
LaboratoryPiston Operated Volumetric GCTS-SOP-51 based on Volumetric Glass Wares
(Bulb pipettes-Single
Volume)
GCTS-SOP-25 based on
ISO 4787 & EURAMET cg19
1 ml to 50 ml 0.03 ml Laboratory
Volumetric Glass Wares
(Graduated Burettes.)
GCTS-SOP-25 based on
ISO 4787 & EURAMET cg19
0.1 ml 1 ml to 50 ml 0.03 ml Laboratory
* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing
more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded
uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration
performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
One Central-Offices 2- Level 3, Suit No.: 302, 303 308.The Trade Centre District, Dubai-United Arab Emirates
Office: +971 4 872 26662 of 2
Accreditation ScopeHumidity Calibration
LB-CAL-013
Issue no.: 02
Calibration Field/
Measuring QualityCalibration Method Range and Specification
Calibration Measurement
Capability
(CMC)*
Location
10- 50 °C 0.4 °C
35- 80 % rH at 20°C 3 % rH
* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing
more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded
uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration
performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
Direct Reading Thermo /
Hygrometers including
Data loggers
GCTS-SOP-27 Laboratory
Gulf Calibration & Technical Services
Ghala Industrial Area, Muscat- Sultanate of Oman
Date: 30-03-2021 Valid to: 10-09-2022
One Central-Offices 2- Level 3, Suit No.: 302, 303 308.The Trade Centre District, Dubai-United Arab Emirates
Office: +971 4 872 26661 of 1
Accreditation Scope
Electrical Calibration
LB-CAL-013
Issue no.: 02
Calibration Field/
Measuring QualityCalibration Method Range and Specification
Calibration Measurement
Capability
(CMC)*
Location
0 mV to 600 mV 0.24 mV
>0.60 V to 6 V 8.9 mV
˃6.0 V to 60 V 89 mV
˃60.0 V to 600 V 0.89 V
˃600 V to 1000 V 4.12 V
0 mV to 100 mV 0.022 mV
˃ 0.10 V to 1 V 0.13 mV
˃ 1.0 V to 10 V 1.3 mV
˃ 10.0 V to 100 V 90 mV
˃ 100.0 V to 1000 V 0.13 V
DC High Voltage Direct Method using Fluke
179 DMM & High Voltage
Probe
U = Measured voltage value
1 kV to 30 kV 20 x 10-3 U Laboratory/
Customers
Premises
0 mV to 600 mV
@45 Hz to 1 kHz
0.46 mV
˃ 0.6 V to 6 V
@45 Hz to 1 kHz
0.073 V
Gulf Calibration & Technical Services
Ghala Industrial Area, Muscat- Sultanate of Oman
Date: 30-03-2021 Valid to: 10-09-2022
DC Volatage
Direct Method using Fluke
179 DMM
DC Voltage Laboratory/
Customers
Premises
Calibration of calibrators
* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing
more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded
uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration
performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
Laboratory/
Customers
Premises
Direct Method using Fluke
179 DMM
AC Voltage
Customers
Premises
Direct Method using
Transmille 1000 A Portable
Multi-Function Calibrator
www.eiac.gov.ae
One Central-Offices 2- Level 3, Suit No.: 302, 303 308.
The Trade Centre District, Dubai-United Arab Emirates
Office: +971 4 872 2666
1 of 17
Accreditation Scope
Electrical Calibration
LB-CAL-013
Issue no.: 02
Calibration Field/
Measuring QualityCalibration Method Range and Specification
Calibration Measurement
Capability
(CMC)*
Location
Gulf Calibration & Technical Services
Ghala Industrial Area, Muscat- Sultanate of Oman
Date: 30-03-2021 Valid to: 10-09-2022
Calibration of calibrators
AC Voltage Direct Method using Fluke
179 DMM
˃ 6.0 V to 60 V
@45 Hz to 1 kHz
0.73 V Laboratory/
Customers
˃ 60.0 V to 600 V
@45 Hz to 1 kHz
7.3 V
˃600 V to 1000 V
@45 Hz to 1 kHz
15.59 V
AC High Voltage Direct Method using Fluke
179 DMM & High Voltage
Probe
U = Measured voltage value
1 kV to 28 kV 20 x 10-3 U Laboratory/
Customers
Premises
1 Hz to 99.99 Hz 0.13 Hz
100.0 Hz to 999.9 Hz 1.3 Hz
1 kHz to 9.99 kHz 12.9 Hz
10.00 kHz to 99.99 kHz 130.0 Hz
0 Ω to 600 Ω 6.4 Ω
˃ 600 Ω to 6 kΩ 64 Ω
˃ 6.0 k Ω to 60 kΩ 0.64 kΩ
˃ 60.0 k Ω to 600 kΩ 6.4 kΩ
Laboratory/
Customers
Premises
Direct Method using Fluke
179 DMM
Resistance
* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing
more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded
uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration
performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
Laboratory/
Customers
Premises
Direct Method using Fluke
179 DMM
Frequency
Calibration of calibrators
www.eiac.gov.ae
One Central-Offices 2- Level 3, Suit No.: 302, 303 308.
The Trade Centre District, Dubai-United Arab Emirates
Office: +971 4 872 2666
2 of 17
Accreditation Scope
Electrical Calibration
LB-CAL-013
Issue no.: 02
Calibration Field/
Measuring QualityCalibration Method Range and Specification
Calibration Measurement
Capability
(CMC)*
Location
Gulf Calibration & Technical Services
Ghala Industrial Area, Muscat- Sultanate of Oman
Date: 30-03-2021 Valid to: 10-09-2022
Calibration of calibrators
˃ 600 k Ω to 6 MΩ 64 kΩ
˃ 6.0 M Ω to 50 MΩ 0.54 MΩ
0 mA to 60 mA 0.73 mA
˃ 60.0 mA to 400 mA 5.0 mA
˃ 0.4 A to 6 A 73 mA
˃ 6.0 A to 10 A 0.15 A
0 mA to 60 mA
@45 Hz to 1 kHz
1.1 mA
˃ 60.0mA to 400 mA
@45 Hz to 1 kHz
7.3 mA
˃ 0.40 A to 6 A
@45 Hz to 1 kHz
0.11 A
˃ 6.0 A to 10 A
@45 Hz to 1 kHz
0.21 A
RTD
PT 100
Direct Method using Beamex
MC6 Multi-Function
Calibrator
-200 °C to 850 °C 0.5 °C Laboratory/
Customers
Premises
Thermocouple
K Type Calibration of
calibrators
Direct Method using
BEAMEX MC6 Multi-
Function Calibrator
-200 °C to 1370 °C 0.6 °C Laboratory/
Customers
Premises
Direct Method using Fluke
179 DMM
DC Current Laboratory/
Customers
Premises
Laboratory/
Customers
Premises
Direct Method using Fluke
179 DMM
AC Current
Calibration of calibrators
* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing
more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded
uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration
performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
Direct Method using Fluke
179 DMM
Resistance Laboratory/
Customers
Premises
www.eiac.gov.ae
One Central-Offices 2- Level 3, Suit No.: 302, 303 308.
The Trade Centre District, Dubai-United Arab Emirates
Office: +971 4 872 2666
3 of 17
Accreditation Scope
Electrical Calibration
LB-CAL-013
Issue no.: 02
Calibration Field/
Measuring QualityCalibration Method Range and Specification
Calibration Measurement
Capability
(CMC)*
Location
Gulf Calibration & Technical Services
Ghala Industrial Area, Muscat- Sultanate of Oman
Date: 30-03-2021 Valid to: 10-09-2022
Calibration of calibrators
Thermocouple
J Type
Direct Method using Beamex
MC6 Multi-Function
Calibrator
-210 °C to 1200 °C 0.6 °C Laboratory/
Customers
Premises
Thermocouple
T Type
Direct Method using Beamex
MC6 Multi-Function
Calibrator
-250 °C to 400 °C 0.6 °C Laboratory/
Customers
Premises
Thermocouple
R Type Calibration of
calibrators
Direct Method using Beamex
MC6 Multi-Function
Calibrator
0 °C to 1760 °C 0.6 °C Laboratory/
Customers
Premises
Thermocouple
S Type
Direct Method using Beamex
MC6 Multi-Function
Calibrator
0 °C to 1760 °C 0.6 °C Laboratory/
Customers
Premises
Thermocouple
N Type
Direct Method using Beamex
MC6 Multi-Function
Calibrator
-200 °C to 1300 °C 0.6 °C Laboratory/
Customers
Premises
Thermocouple
B Type
Direct Method using Beamex
MC6 Multi-Function
Calibrator
600 °C to 1820 °C 0.6 °C Laboratory/
Customers
Premises
Calibration of calibrators
* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing
more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded
uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration
performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
www.eiac.gov.ae
One Central-Offices 2- Level 3, Suit No.: 302, 303 308.
The Trade Centre District, Dubai-United Arab Emirates
Office: +971 4 872 2666
4 of 17
Accreditation Scope
Electrical Calibration
LB-CAL-013
Issue no.: 02
Calibration Field/
Measuring QualityCalibration Method Range and Specification
Calibration Measurement
Capability
(CMC)*
Location
Gulf Calibration & Technical Services
Ghala Industrial Area, Muscat- Sultanate of Oman
Date: 30-03-2021 Valid to: 10-09-2022
Calibration of calibrators
Thermocouple
E Type
Direct Method using
Beamex MC6 Multi-Function
Calibrator
-250 °C to 1000 °C 0.6 °C Laboratory/
Customers
Premises
Thermocouple
L Type
Direct Method using Beamex
MC6 Multi-Function
Calibrator
-200 °C to 900 °C 0.6 °C Laboratory/
Customers
Premises
Thermocouple
U Type
Direct Method using
Beamex MC6 Multi-Function
Calibrator
-200 °C to 600 °C 0.6 °C Laboratory/
Customers
Premises
Thermocouple
C Type
Direct Method using
Beamex MC6 Multi-Function
Calibrator
0 °C to 2300 °C 0.6 °C Laboratory/
Customers
Premises
0 mV to 100 mV 0.2 mV
˃ 0.10 V to 1 V 0.77 mV
˃ 1.0 V to 10 V 14 mV
˃ 10.0 V to 100 V 0.17 V
˃ 100.0 V to 1000 V 2.4 V
Direct Method using
Transmille 1000A Portable
Multi-Function Calibrator
AC Voltage Customers
Premises
Calibration of calibrators
* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing
more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded
uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration
performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
Calibration of instruments
www.eiac.gov.ae
One Central-Offices 2- Level 3, Suit No.: 302, 303 308.
The Trade Centre District, Dubai-United Arab Emirates
Office: +971 4 872 2666
5 of 17
Accreditation Scope
Electrical Calibration
LB-CAL-013
Issue no.: 02
Calibration Field/
Measuring QualityCalibration Method Range and Specification
Calibration Measurement
Capability
(CMC)*
Location
Gulf Calibration & Technical Services
Ghala Industrial Area, Muscat- Sultanate of Oman
Date: 30-03-2021 Valid to: 10-09-2022
Calibration of calibrators
0.0 mV to 200 mV
@ 45 Hz to 1 kHz
0.14 mV
˃0.20 V to 2 V
@ 45 Hz to 1 kHz
1.2 mV
˃ 2.0 V to 20 V
@ 45 Hz to 1 kHz
12 mV
˃ 20.0 V 200 V
@ 45 Hz to 1 kHz
0.14V
˃ 200.0 V to 1000 V
@ 45 Hz to 1 kHz
0.72 V
0.0 mV to 200 mV 0.016 mV
˃0.20 V to 2 V 0.2 mV
˃ 2.0 V to 20 V 1.5 mV
˃ 20.0 V 200 V 15 mV
˃ 200.0 V to 1000 V 81 mV
1 Hz to 100 Hz 0.0023 Hz
>100Hz to 1 kHz 0.02 Hz
>1 kHz to 10 kHz 0.23 Hz
>10 kHz to 20 kHz 0.46 Hz
>20 kHz to 50 kHz 1.2 Hz
>50 kHz to 100 kHz 2.3 Hz
Direct Method using
Transmille 3050A/ 3041A
Multi-Function Calibrator
AC Voltage Laboratory
* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing
more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded
uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration
performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
Direct Method using
Transmille 1000A Portable
Multi-Function Calibrator
Frequency Customers
Premises
Calibration of instruments
Direct Method using
Transmille 3050A/ 3041A
Multi-Function Calibrator
DC Voltage Laboratory
www.eiac.gov.ae
One Central-Offices 2- Level 3, Suit No.: 302, 303 308.
The Trade Centre District, Dubai-United Arab Emirates
Office: +971 4 872 2666
6 of 17
Accreditation Scope
Electrical Calibration
LB-CAL-013
Issue no.: 02
Calibration Field/
Measuring QualityCalibration Method Range and Specification
Calibration Measurement
Capability
(CMC)*
Location
Gulf Calibration & Technical Services
Ghala Industrial Area, Muscat- Sultanate of Oman
Date: 30-03-2021 Valid to: 10-09-2022
Calibration of calibrators
1 Hz to 100 Hz 0.0023 Hz
˃10 Hz to 1 kHz 0.023 Hz
˃ 1.0 kHz to 10.0 kHz 0.23 Hz
˃ 10.0 kHz to 100 kHz 2.3 Hz
˃ 100.0 kHz to 1.0 MHz 20 Hz
˃ 1.0 MHz to 10 MHz 0.23 kHz
0 µA to 100 µA 0.07 μA
˃ 0.10 mA to 1 mA 0.46 μA
˃ 1.0 mA to 10 mA 4.6 μA
˃ 10.0 mA to 100 mA 46 µA
˃ 0.10 A to 1 A 0.52 mA
˃ 1.0 A to 10 A 8.3 mA
DC Current Direct Method using
Portable Multi-Function
Calibrator
(Transmille 1000A) & 2, 10
& 50 turn current coil
I = Measured current value
10 A to1000 A 10 x 10-3 I Customers
Premises
Frequency Laboratory
Customers
Premises
Direct Method using
Transmille 1000A Portable
Multi-Function Calibrator
DC Current
* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing
more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded
uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration
performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
Calibration of instruments
Direct Method using
Transmille 3050A/ 3041A
Multi-Function Calibrator
www.eiac.gov.ae
One Central-Offices 2- Level 3, Suit No.: 302, 303 308.
The Trade Centre District, Dubai-United Arab Emirates
Office: +971 4 872 2666
7 of 17
Accreditation Scope
Electrical Calibration
LB-CAL-013
Issue no.: 02
Calibration Field/
Measuring QualityCalibration Method Range and Specification
Calibration Measurement
Capability
(CMC)*
Location
Gulf Calibration & Technical Services
Ghala Industrial Area, Muscat- Sultanate of Oman
Date: 30-03-2021 Valid to: 10-09-2022
Calibration of calibrators
0.0 µA to 200 µA 0.05 μA
˃ 0.20 mA to 2 mA 0.3 μA
˃ 2.0 mA to 20 mA 3.3 μA
˃ 20.0 mA to 200 mA 40 µA
˃ 0.20 A to 2 A 1.3 mA
˃ 2.0 A to 30 A 13 mA
DC Current Direct Method using
Portable Multi-Function
Calibrator
(TRANSMILLE
3050A/3041A) & 2, 10 & 50
turn current coil
I = Measured current value
10 A to 1500 A 10 x 10-3 I Laboratory
0 µA to 100 µA @45 Hz to 1
kHz
0.6 μA
˃ 0.10 mA to 1 mA
@45 Hz to 1 kHz
2.6 μA
Direct Method using
Transmille 1000A Portable
Multi-Function Calibrator
AC Current Customers
Premises
Calibration of instruments
Direct Method using
Transmille 3050A/ 3041A
Multi-Function Calibrator
LaboratoryDC Current
* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing
more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded
uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration
performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
www.eiac.gov.ae
One Central-Offices 2- Level 3, Suit No.: 302, 303 308.
The Trade Centre District, Dubai-United Arab Emirates
Office: +971 4 872 2666
8 of 17
Accreditation Scope
Electrical Calibration
LB-CAL-013
Issue no.: 02
Calibration Field/
Measuring QualityCalibration Method Range and Specification
Calibration Measurement
Capability
(CMC)*
Location
Gulf Calibration & Technical Services
Ghala Industrial Area, Muscat- Sultanate of Oman
Date: 30-03-2021 Valid to: 10-09-2022
Calibration of calibrators
˃ 1.0 mA to 10 mA
@45 Hz to 1 kHz
22 µA
˃ 10.0 mA to 100 mA
@45 Hz to 1 kHz
0.22 mA
˃ 0.10 A to 1 A
@45 Hz to 1 kHz
2.2 mA
˃ 1.0 A to 10 A
@45 Hz to 1 kHz
0.03 A
AC Current Direct Method using
Portable Multi-Function
Calibrator
Transmille 1000A) & 2, 10 &
50 turn current coil
I = Measured current value
10 A to1000 A
@45 Hz to 200 Hz
10 x 10-3 I Customers
Premises
AC Current Direct Method using
Portable Multi-Function
Calibrator
Transmille 3050A/3041A) &
2, 10 & 50 turn current coil
I = Measured current value
10 A to 1500 A
@45 Hz to 200 Hz
10 x 10-3 I Laboratory
Calibration of instruments
Direct Method using
Transmille 1000A Portable
Multi-Function Calibrator
AC Current Customers
Premises
* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing
more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded
uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration
performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
www.eiac.gov.ae
One Central-Offices 2- Level 3, Suit No.: 302, 303 308.
The Trade Centre District, Dubai-United Arab Emirates
Office: +971 4 872 2666
9 of 17
Accreditation Scope
Electrical Calibration
LB-CAL-013
Issue no.: 02
Calibration Field/
Measuring QualityCalibration Method Range and Specification
Calibration Measurement
Capability
(CMC)*
Location
Gulf Calibration & Technical Services
Ghala Industrial Area, Muscat- Sultanate of Oman
Date: 30-03-2021 Valid to: 10-09-2022
Calibration of calibrators
0.0 µA to 200 µA
@45 Hz to 1kHz
0.52 μA
˃ 0.20 mA to 2 mA
@45 Hz to 1kHz
2.2 μA
˃ 2.0 mA to 20 mA
@45 Hz to 1kHz
21 µA
˃ 20.0 mA to 200 mA @45
Hz to 1kHz
0.21 mA
˃ 0.20 A to 2 A
@45 Hz to 1kHz
2.7 mA
˃ 2.0 A to 30 A
@45 Hz to 1kHz
56 mA
10 nF 0.1 nF
100 nF 1.0 nF
1 µF 20 nF
10 nF 0.052 nF
20 nF 0.11 nF
50 nF 0.24 nF
100 nF 0.48 nF
1 µF 4.7 nF
Customers
Premises
Direct Method using
Transmille 1000A Portable
Multi-Function Calibrator
Capacetance
Direct Method using
Transmille 3050A/ 3041A
Multi-Function Calibrator
(Fixed Values only)
Capacitance Laboratory
Calibration of instruments
Direct Method using
TRANSMILLE 3050A/
3041A Multi-Function
Calibrator
AC Current
45 Hz to 1kHz
Laboratory
* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing
more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded
uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration
performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
www.eiac.gov.ae
One Central-Offices 2- Level 3, Suit No.: 302, 303 308.
The Trade Centre District, Dubai-United Arab Emirates
Office: +971 4 872 2666
10 of 17
Accreditation Scope
Electrical Calibration
LB-CAL-013
Issue no.: 02
Calibration Field/
Measuring QualityCalibration Method Range and Specification
Calibration Measurement
Capability
(CMC)*
Location
Gulf Calibration & Technical Services
Ghala Industrial Area, Muscat- Sultanate of Oman
Date: 30-03-2021 Valid to: 10-09-2022
Calibration of calibrators
0 Ω to 10 Ω 0.07 Ω
˃ 10 Ω to 100 Ω 0.08 Ω
˃ 100.0 Ω to 1 kΩ 0.29 Ω
˃1.0 kΩ to 10 kΩ 2.4 Ω
>10.0 KΩ to 100 KΩ 0.024 kΩ
>100.1 KΩ to 1 MΩ 0.25 kΩ
˃1 MΩ to 10 MΩ 4.1 kΩ
10 Ω 13 mΩ
100 Ω 16 mΩ
1 kΩ 0.12 Ω
10 kΩ 1.2 Ω
100 kΩ 12 Ω
1 MΩ 0.22 kΩ
10 MΩ 7.4 kΩ
100 MΩ 0.75 MΩ
* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing
more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded
uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration
performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
Direct Method using
Transmille 3050A/ 3041A
Multi-Function Calibrator
Resistance
Fixed values only-2 WIRE
Laboratory
Calibration of instruments
Customers
Premises
Simulation Method using
Transmille 1000A
Portable Multi-Function
Calibrator
Resistance
Resistance
Fixed values only-2 WIRE
Direct Method using
Transmille 3050A/ 3041A
Multi-Function Calibrator
Laboratory
www.eiac.gov.ae
One Central-Offices 2- Level 3, Suit No.: 302, 303 308.
The Trade Centre District, Dubai-United Arab Emirates
Office: +971 4 872 2666
11 of 17
Accreditation Scope
Electrical Calibration
LB-CAL-013
Issue no.: 02
Calibration Field/
Measuring QualityCalibration Method Range and Specification
Calibration Measurement
Capability
(CMC)*
Location
Gulf Calibration & Technical Services
Ghala Industrial Area, Muscat- Sultanate of Oman
Date: 30-03-2021 Valid to: 10-09-2022
Calibration of calibrators
0.01 MΩ to 5 MΩ 10 x 10-3 R
˃ 5 MΩ to 2 GΩ 39 x 10-3 R
100mΩ to 900 mΩ 40 mΩ
1 Ω to 2 Ω 50 mΩ
˃2 Ω to 900 Ω 20 x 10-3 R
1 kΩ to 50 kΩ 20 x 10-3 R
10 mA to 100 mA 2.0 mA
>100 mA to 300 mA 5.0 mA
AC Voltage Out Direct Method using 3200B
Electrical Test Calibrator
U = Measured voltage value
100 V to 400 V@50 Hz 5.0 x 10-3U Laboratory
Insulation Resistance Voltage
Measure
Direct Method using 3200B
Electrical Test Calibrator
U = Measured voltage value
50 V to 1000 V 30 x 10-3 U Laboratory
Calibration of instruments
LaboratoryDirect Method using 3200B
Electrical Test Calibrator
R = Measured resistance
value
Continuity Resistance
Direct Method using 3200B
Electrical Test Calibrator
Continuity Current Laboratory
Direct Method using 3200B
Electrical Test Calibrator
R = Measured resistance
value
Insulation Resistance Laboratory
* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing
more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded
uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration
performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
www.eiac.gov.ae
One Central-Offices 2- Level 3, Suit No.: 302, 303 308.
The Trade Centre District, Dubai-United Arab Emirates
Office: +971 4 872 2666
12 of 17
Accreditation Scope
Electrical Calibration
LB-CAL-013
Issue no.: 02
Calibration Field/
Measuring QualityCalibration Method Range and Specification
Calibration Measurement
Capability
(CMC)*
Location
Gulf Calibration & Technical Services
Ghala Industrial Area, Muscat- Sultanate of Oman
Date: 30-03-2021 Valid to: 10-09-2022
Calibration of calibrators
Insulation Resistance DC
Current Measurement
Direct Method using 3200B
Electrical Test Calibrator
0.5 mA & 1 mA 20 µA Laboratory
0.05 Ω to 10.0 Ω 0.5 Ω
100 Ω 6.0 Ω
900.0 Ω 10 Ω
2 mA to 10 mA 0.2 mA
10.01 mA to 30 mA 0.5 mA
30.01 mA to 100 mA 1.6 mA
100.01 mA to 300 mA 4.5 mA
300.01 mA to 1000 mA 15 mA
1000.01 mA to 3000 mA 30 mA
20 ms to 390 ms 1.3 ms
900 ms 20 ms
PAT Insulation Resistance Direct Method using 3200B
Electrical Test Calibrator
R = Measured resistance
value
1 MΩ to 10 MΩ 10 x 10-3 R Laboratory
LaboratoryDirect Method using 3200B
Electrical Test CalibratorR
RCD Trip Time
* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing
more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded
uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration
performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
Calibration of instruments
Direct Method using 3200B
Electrical Test Calibrator
LaboratoryLoop Impedance Resistance
Measurement
Direct Method using 3200B
Electrical Test Calibrator
RCD Current Laboratory
www.eiac.gov.ae
One Central-Offices 2- Level 3, Suit No.: 302, 303 308.
The Trade Centre District, Dubai-United Arab Emirates
Office: +971 4 872 2666
13 of 17
Accreditation Scope
Electrical Calibration
LB-CAL-013
Issue no.: 02
Calibration Field/
Measuring QualityCalibration Method Range and Specification
Calibration Measurement
Capability
(CMC)*
Location
Gulf Calibration & Technical Services
Ghala Industrial Area, Muscat- Sultanate of Oman
Date: 30-03-2021 Valid to: 10-09-2022
Calibration of calibrators
0.05 Ω to 1.0 Ω 20 mΩ
˃1 Ω to 900.0 Ω 20 x 10-3 R
0 mA to 500 mA 16 mA
1 A to 10 A 0.2 A
11 A to 30 A 0.6 A
PAT Leakage Current Direct Method using 3200B
Electrical Test Calibrator
2 mA to 8 mA
@ 240 V
20 µA Laboratory
RTD
(PT 100 Source Mode)
Simulation Method using
Beamex MC6 Multi-Function
Calibrator
200 °C to 850 °C 0.5 °C Laboratory/
Customers
Premises
Thermocouple
K Type
Simulation Method using
Transmille 3050A Multi-
Function Calibrator &
Thermocouple Simulator/
Beamex MC6 Multi-Function
Calibrator
-200 °C to 1370 °C 0.6 °C Laboratory/
Customers
Premises
* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing
more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded
uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration
performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
Calibration of instruments
Direct Method using
Electrical Test Calibrator
R = Measured resistance
value
PAT Earth Bond Resistance Laboratory
Direct Method using
Electrical Test Calibrator
Earth Bond Current
Measurement
Laboratory
www.eiac.gov.ae
One Central-Offices 2- Level 3, Suit No.: 302, 303 308.
The Trade Centre District, Dubai-United Arab Emirates
Office: +971 4 872 2666
14 of 17
Accreditation Scope
Electrical Calibration
LB-CAL-013
Issue no.: 02
Calibration Field/
Measuring QualityCalibration Method Range and Specification
Calibration Measurement
Capability
(CMC)*
Location
Gulf Calibration & Technical Services
Ghala Industrial Area, Muscat- Sultanate of Oman
Date: 30-03-2021 Valid to: 10-09-2022
Calibration of calibrators
Thermocouple
J Type
Simulation Method using
Transmille 3050A Multi-
Function Calibrator &
Thermocouple Simulator/
Beamex MC6 Multi-Function
Calibrator
-210 °C to 1200 °C 0.6 °C Laboratory/
Customers
Premises
Thermocouple
T Type
Simulation Method using
Transmille 3050A Multi-
Function Calibrator &
Thermocouple Simulator/
Beamex MC6 Multi-Function
Calibrator
-250 °C to 400 °C 0.6 °C Laboratory/
Customers
Premises
Thermocouple
R Type
Simulation Method using
Transmille 3050A Multi-
Function Calibrator &
Thermocouple Simulator/
Beamex MC6 Multi-Function
Calibrator
0 °C to 1760 °C 0.6 °C Laboratory/
Customers
Premises
Thermocouple
S Type
Simulation Method using
Transmille 3050A Multi-
Function Calibrator &
Thermocouple Simulator/
Beamex MC6 Multi-Function
Calibrator
0 °C to 1760 °C 0.6 °C Laboratory/
Customers
Premises
Calibration of instruments
* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing
more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded
uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration
performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
www.eiac.gov.ae
One Central-Offices 2- Level 3, Suit No.: 302, 303 308.
The Trade Centre District, Dubai-United Arab Emirates
Office: +971 4 872 2666
15 of 17
Accreditation Scope
Electrical Calibration
LB-CAL-013
Issue no.: 02
Calibration Field/
Measuring QualityCalibration Method Range and Specification
Calibration Measurement
Capability
(CMC)*
Location
Gulf Calibration & Technical Services
Ghala Industrial Area, Muscat- Sultanate of Oman
Date: 30-03-2021 Valid to: 10-09-2022
Calibration of calibrators
Thermocouple
N Type
Simulation Method using
Transmille 3050A Multi-
Function Calibrator &
Thermocouple Simulator/
Beamex MC6 Multi-Function
Calibrator
-200 °C to 1300 °C 0.6 °C Laboratory/
Customers
Premises
Thermocouple
B Type
Simulation Method using
Transmille 3050A Multi-
Function Calibrator &
Thermocouple Simulator/
Beamex MC6 Multi-Function
Calibrator
600 °C to 1820 °C 0.6 °C Laboratory/
Customers
Premises
Thermocouple
E Type
Simulation Method using
Transmille 3050A Multi-
Function Calibrator &
Thermocouple Simulator/
Beamex MC6 Multi-Function
Calibrator
-250 °C to 1000 °C 0.6 °C Laboratory/
Customers
Premises
Thermocouple
L Type
Simulation Method using
Transmille 3050A Multi-
Function Calibrator &
Thermocouple Simulator/
Beamex MC6 Multi-Function
Calibrator
-200 °C to 900 °C 0.6 °C Laboratory/
Customers
Premises
* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing
more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded
uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration
performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
Calibration of instruments
www.eiac.gov.ae
One Central-Offices 2- Level 3, Suit No.: 302, 303 308.
The Trade Centre District, Dubai-United Arab Emirates
Office: +971 4 872 2666
16 of 17
Accreditation Scope
Electrical Calibration
LB-CAL-013
Issue no.: 02
Calibration Field/
Measuring QualityCalibration Method Range and Specification
Calibration Measurement
Capability
(CMC)*
Location
Gulf Calibration & Technical Services
Ghala Industrial Area, Muscat- Sultanate of Oman
Date: 30-03-2021 Valid to: 10-09-2022
Calibration of calibrators
Thermocouple
U Type
Simulation Method using
Transmille 3050A Multi-
Function Calibrator &
Thermocouple Simulator/
Beamex MC6 Multi-Function
Calibrator
-200 °C to 600 °C 0.6 °C Laboratory/
Customers
Premises
Thermocouple
C Type
Simulation Method using
Transmille 3050A Multi-
Function Calibrator &
Thermocouple Simulator/
Beamex MC6 Multi-Function
Calibrator
0 °C to 2300 °C 0.6 °C Laboratory/
Customers
Premises
RPM
Tachometers
Non-Contact
GCTS-SOP-54 using
Transmille EA015 Multi-
Function Workstation
240 rpm to 20,000 rpm 2 rpm Laboratory
* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing
more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded
uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration
performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
Calibration of instruments
www.eiac.gov.ae
One Central-Offices 2- Level 3, Suit No.: 302, 303 308.
The Trade Centre District, Dubai-United Arab Emirates
Office: +971 4 872 2666
17 of 17