Ion Exchange Chromatography
description
Transcript of Ion Exchange Chromatography
Ion Exchange ChromatographyIon exchange chromatography is the purification
technique, which involves the separation of the proteins based on the ions exchange property between the
proteins and the column Related LOs: Column packing material, Retention and Elution time > Prior Viewing – IDD-6. Extraction of serum protein, IDD-41. Liquid chromatography - Gel filtration, IDD-42. Liquid chromatography - affinity chromatography > Future Viewing – IDD-36. Isobaric tag for relative and absolute quantitation (iTRAQ), IDD-37. Isotope-coded affinity tags (ICAT), IDD-39. LC-MSMS data analysis
Course Name: Ion exchange Chromatography Level(UG/PG): UG Author(s): Dinesh Raghu, Vinayak Pachapur Mentor: Dr. Sanjeeva Srivastava
*The contents in this ppt are licensed under Creative Commons Attribution-NonCommercial-ShareAlike 2.5 India license
Learning objectivesAfter interacting with this learning object, the learner will
be able to:
1. Define buffer preparation2. Prepare ion exchange column3. Analyze the elution/purification of the sample4. Assess the troubleshooting steps involved in the
experiments.
5
3
2
4
1
Master Layout
5
3
2
4
1Sample preparation (Slide: 5-11)
Resin/buffer preparation (Slide:12-14)
Sample addition (Slide: 15-17)
Elution (Slide: 18-19)
UV-Visible spectrometry (Slide 20-26)
Analysis (Slide:27-31)
Please animate the layout, by taking figures from each of the steps
Definitions and Keywords
5
3
2
4
1 1. Stationary Phase: The resins inside the column which interacts with the protein of different charge densities and enable it to separate .
2. Mobile phase: The buffers/ salt solution that is used to elute the protein by interacting with the resin and thereby free the proteins from the column.
3. DEAE-Sepharose: Diethyl amino ethyl sepharose is an anion exchanger which interacts more with cationic proteins.
4. CM-Cellulose: Carboxy methyl cellulose is an cation exchanger which interacts more with anionic proteins.
Audio Narration Description of the action
5
2
1
4
3The animator should draw graduated measuring cylinder as shown in slide with graduation 100ml, 250 ml,500ml,1000ml.
The user should click on the appropriate cylinder for usage
100100
250250
500500
10001000
Step 1: T1:Reagents for Ion exchange chromatography
Measuring cylinder are used to make up the required final volume.
Step 1:
Audio Narration (if any)Description of the action
5
2
1
4
3Show a measuring balance, with display, ON, OFF and TARE/0 buttons on it. let user ON it, display reading as 0.000g, let user picks up the paper from the rack, makes 1/10 of folding on the sides and places it on the balance. Now the display reading changes to 0.003g. Instruct user to TARE the reading. And animate to click the tare button. Once user clicks it, reading must show ”0”
T1:Reagents for Ion exchange chromatography
When measuring with
paper, the weight of the paper need to be tarred from
actual reading.
Video file: Balancing
Step 1:
Audio Narration Description of the action
5
2
1
4
3 Show the bottle labeled as “DEAE sepharose and CM –cellulose, deionized water”. Let user pick up the bottle labeled as DEAE-sepharose. Instruct user to click on the bottle to open the lid of the bottle and click on the spoon to take in hand and add the DEAE from the bottle to the paper show like gradual increase in the reading till it weighs about 1gm. Follow the same procedure for weighing CM –Cellulose. Instruct the user to set the pipette to 1000ul and open the lid of deionized water bottle to pipette out 1000ul and add to a empty DEAE containing tube. Repeat it for one more time to make 2ml and repeat the same for CM cellulose in another tube. Let user label the tube accordingly. User must click on the hands for the events to happen.
DEAEDeionized Water
CM-cellulose
Prepare DEAE-sepharose and CM-cellulose resin. DEAE is used to prepare the anion exchange column while CM cellulose for cation exchange column.
T1:Reagents for Ion exchange chromatography
Step 2:
Audio Narration Description of the action
5
2
1
4
3 Instruct user to prepare 10mM phosphate buffer. Let user pick up bottle ” Monosodium phosphate” and “disodium phosphate”, spatula, measuring cylinder from the rack and keeps it on the table next to balance. let user tare the balance, user should click on the monosodium bottle, uncap it, with help of spatula weigh the required amount on a paper over the balance. Instruct user to weigh 0.04g of Monosodium phosphate and 0.18 g disodium phosphate”. Display a gradual increase in reading with quantity addition. if the gram exceeds user should remove some quantity or if it less add the quantity to get the exact required amount.
Monosodium phosphate
Prepare 10mM phosphate buffer of pH7.2
T1:Reagents for Ion exchange chromatography
Disodium phosphate
Step 2:
Audio Narration Description of the action
5
2
1
4
3 Instruct the user to click on the water bottle to pour in the measuring cylinder till it reaches 80ml. add to the weighed phosphate bottle from previous slide and then ask the user to take the bottle for mixing on the magnetic stirrer (see slide:10 for proper animation instruction). Show the clear solution, now instruct user to pour the solution from bottle to the beaker. Let user label the beaker accordingly as 10mM phosphate buffer.
Prepare 10mM phosphate buffer of pH7.2 , the prepared buffer need to checked for pH to set it at 7.2.
T1:Reagents for Ion exchange chromatography
5
2
1
4
3Beaker
Magnetic bead
Audio Narration (if any)
Description of the action
Show magnetic stirrer instrument. Let user place the beaker on it. Display the beaker containing powder at bottom, liquid layer on top and a magnetic bead at the bottom. Instruct user to ON the instrument, let user control the speed knob and regulate it accordingly to control the mixing speed in the beaker. Animate powder getting into the solution.Show a turbid solution turning colorless
Video file: Magnetic stirrer
The magnetic stirrer is used for evenly distribution of solute into the solvent.
Step 3:
Audio Narration Description of the action
5
2
1
4
3 Show the bottles labeled as “Sodium Chloride, deionized water”. Let user pick up the bottle labeled as Sodium chloride. Instruct the user to click on the bottle to open the lid of the bottle and click on the spoon to take in hand to add sodium chloride from the bottle on to the paper. show like gradual increase in the reading till it weighs about 0.292gm. Follow the same procedure for weighing1.421gm, 2.922gm of sodium chloride and animate like putting them in separate beaker. Instruct user to click on the phosphate buffer (earlier prepared) to pour in the measuring cylinder till it reaches 80ml and then ask the user to place the bottle for magnetic stirrer (see slide:10 for proper animation instruction). Show the clear solution after sometimes, now instruct the user to pour the solution to the measuring cylinder and show the volume till 90ml and instruct user to add 10ml of phosphate buffer to make the final volume to 100ml.
NACLDeionized Water
Prepare 50mM,250mM, 500mM of High salt buffer solution.
T1:Reagents for Ion exchange chromatography
Step 4:
Audio Narration Description of the action
5
2
1
4
3
Before the pH reading, pH instrument need to be calibrated with standards. Once with STD-1 at pH 4 and with STD-2 at pH 9.
STD 1STD 1STD 2STD 2
T2: pH meter standarization
Display standard pH bottles and pH instrument and deionized water, discard placed on a table. Instruct user to calibrate the instrument. Let user ON the instrument. Initially for the pH rod is dipped in water, when user clicks on read button, display must show a reading “7”. Now show like taking out the rod and washing it with deionized-water, let user cleans the rod with tissue. Now pick the STD-1, uncap it, dip the cleaned rod into the solution, user must click read button with display showing “4”. now clean the rod and repeat the step to note down the reading for STD-2 and now the display should show “9”
Video file: pH meter
Step 5:
Audio Narration Description of the action
5
2
1
4
3 Instruct user to set the pH for phosphate buffer pH at 7.2. Now take the bottle labeled as “Phosphate buffer“, uncap it, dip the cleaned pH rod into the solution. User need to click on read button. Initially display must show a reading 6. now instruct user to add NaOH to adjust the pH. Now allow the user to click on NaOH bottle so that drops of NaOH should be added with filler, user need to mix the solution with glass rod, click on read button and the reading should show anywhere near 6.1- 6.3. let user keeps adding the NaOH drop till the pH display shows 7.2.
Prepare Phosphate buffer of pH 7.2 , buffers play a very important role during equilibration step.
NaOHNaOHHClHCl
T2: pH measurement
Step 6:
Audio Narration Description of the action
5
2
1
4
3 Animate like the user transferring the beaker solution to 100ml measuring cylinder and reading should show 92ml.Instruct the user to click on the water and animate like the user pouring it to makeup the volume to 100ml. All action should happen when the user clicks the hand image.
NaOHNaOH HClHCl
T2: pH measurement
Prepare Phosphate buffer of pH 7.2
Step 7:
Audio NarrationDescription of the action
T3: Column Preparation
5
2
1
4
3Animate like the user tightening the knob at the bottom in the column. The user must click on the beaker labeled as DEAE and animate like the user pouring the solution inside the tube. Show like the particles are settling in the tube with water at the top. Repeat the same for CM cellulose. Animate as shown in the figureLet user click on the hand for the things to happen Show the blue ring with positive charge in column 1 and negative charge in column 2
DEAE
CM
Pack the columns by DEAE and CM cellulose resins. The packing is the main step which determines the effective separation for the samples.
Column 1Column 2
Step 8:
Audio NarrationDescription of the action
5
2
1
4
3Instruct user to “Equilibrate the column”. Animate like the user tightening the knob at the bottom in the column. The user must click on the beaker labeled as phosphate buffer and animate like the user pouring the solution inside the tube. Now instruct the user to click on hands to place a beakers at the bottom of the columns and open the stopper. Animate like the liquid comes out of the tube in drop to the beaker and show like closing the stopper and show a liquid layer at the top of the column
Buffer
Buffer
Equilibrate the column using Phosphate buffer of pH 7.2, which helps to attain a same condition through out the column.
stopper
Column
Buffer
T3: Column Preparation
Step 9:
Audio NarrationDescription of the action
5
2
1
4
3Instruct user to take out sample stored at -80’C. animate user taking out the sample, thawing the tube, animate frozen state to liquid state in tube, now the user should take the pipette set to 200ul, pipette out the sample and add to the column as shown in figure.Events must happen as and when the user clicks on the pipette animate a clock for 10 minutes
Load the sample to both anion exchange and cation exchange columns to carry out separation.
stopper
Column
Buffer
sample
T3: Sample Loading
Step 10: T4: Mechanism of separation
5
2
1
4
3
Column 1Column 2
Step 10: Audio Narration Description of the action
5
2
1
4
3Animate rings of different color with each ring possesing positive or negative charges passing through the column and bind to the resin.
Negative charge rings in the sample should bind to resin in column1 and positive charge to resin in column 2.
Show like positive charge ring unbinding to column 1 similarly negative charge ring unbinding to column 2.
Cationic protein bind in CM-cellulose resin while anionic protein bind to DEAE resin based on the strength of charge interaction
T4: Mechanism of separation
Step 11:
5
2
1
4
3
MOBILE PHASE FUNNEL
50mM
Nacl
T5: Elution
Step 12: Audio Narration Description of the action
5
2
1
4
3
Draw a mobile phase funnel as shown in previous slide. Instruct the user to take the funnel and place on the top of the column. Now show three beakers labeled as 50mM NaCl, 250mM Nacl and 500mM NaCl.
Now instruct the user to click on 50 mM NaCl solution beaker to pour into the funnel.
Pour the low ionic strength mobile phase through the column for elution step.
T5: Elution
Step 13:
5
2
1
4
3
T5: Elution
Step 13: Audio Narration Description of the action
5
2
1
4
3
Animate like the rings with the charges moving down as in previous slide figure .
Animate like negative rings binding to the resin in column 1 while the positive rings flow through the column.Similarly Animate like positive rings binding to the resin in column 2 while the negative ions rings flow through the column.
T5: Elution
The elution step helps for the collection of the sample at different fractions.
Step 14:
5
2
1
4
3
T5: Elution
Step 15: )
5
2
1
4
3
T5: Elution
collection tube
Step 14,15:
Audio Narration Description of the action
5
2
1
4
3
Now instruct the user to click on the knob at side to open and show like the sample flowing out into collection tube as shown in previous slide
Animate similarly from 20-26 with beaker labeled as 250mM (in slide 20)
Once this is done again show the same animation now with the beaker should be labeled as 500mM NaCl (in slide 20).So the animation must be for 3 times starting with 500mM, 250mM and last with 500mM NaCl.
When adding the 250mM and 500mM solution show like the movement of bound ion as in 22. Animate both column 1 and column 2 separately as per the instruction from 20-26 and follow the instruction carefully in slide 23 and animate the ions flowing accordingly
Collect the eluted sample at different fractions. Proteins with low ionic interaction will be eluted first and when the ionic strength is increased proteins with high ionic interaction will be eluted out later.
T5: Elution
Step 16:
5
2
1
4
3 Cuvette
T6: UV-visible spectrometry analysis
Video file: : UV spectrometry
Step 16: Audio Narration
Description of the action
5
2
1
4
3
Show a instrument labeled as “UV –visible spectrometry” and the samples in the stand as shown in figure. Animate buttons like “start, auto zero, absorbance, stop” on the instrument
Now instruct the user to switch on the instrument, set the wavelength to 595nm by pressing on numbers-open the lid of the instrument and take a cuvette as in figure and click on phosphate buffer to take it into cuvette and animate like keeping it inside the instrument to press “auto zero”. Display a value on the system as “0.000” animate like the user opening the lid to take the cuvette and discarding the solution, now animate like the user taking the sample-1 (collection tube-1) and adding it to the cuvette, keeping it inside, closing the lid and press absorbance. show the values as in next slide for each sample follow the same for (collection tube 2-15)
Detect the presence of protein of using the UV –visible spectrometry. The high absorbance reading indicate the presence of protein.
T6: UV-visible spectrometry analysis
Step 16:
5
2
1
4
3
T6: UV-visible spectrometry analysis
Sample Volume of sample Absorbance
1 2ml 0.12
2 2ml 0.229
3 2ml 0.303
4 2ml 0.457
5 2ml 0.533
6 2ml 0.681
7 2ml 0.71
8 2ml 0.62
9 2ml 0.65
10 2ml 0.8
11 2ml 0.98
12 2ml 0.6
13 2ml 0.44
14 2ml 0.23
15 2ml 0.1
Step 17:
5
2
1
4
3
T6: UV-visible spectrometry analysis
Step 17: Audio Narration Description of the action
5
2
1
4
3
Animate like the user drawing the graph as shown in previous slide with absorbance at Y axis and volume of eluent in x axis
Draw the chromatrogram graph with absorbance at Y axis and volume of eluent in x axis to determine the protein using the literature or to know the presence of protein. The sample collected can be directly injected into the LC-MS analysis. For further information please go through future viewing IDD.
T6: UV-visible spectrometry analysis
Animation areaInteraction 1: slide-19: provide user sample mixture
of both cation and anion proteins, ask for user input.
Instruction: user can start with any one column at first place, later by going through the absorbance reading user must make out the with high protein concentration the sample must undergo another step of column separation.
Instructions/ Working area
Credits
Name of the section/stage Interactivity
area
Tab 02 Tab 03 Tab 04 Tab 05 Tab 06 Tab 07
Button 01
Button 02
Button 03
Tab 01
Slide 5-11
Slide 12-14
Slide 15-17 Slide
18,19Slide 20-26 Slide
27-31
Questionnaire:APPENDIX
1
Question 1:The cation exchange resin is a)CM celluloseb)DEAEc)Sepharosed)Cellulose
Question 2
The anion exchange resin isa)DEAE-sepharoseb)CM-cellulosec)Cellulosed)Sepharose
Question 3
The stationary phase of the ion exchange chromatography isa)Naclb)Phosphate bufferc)Resind)Proteins
Questionnaire:APPENDIX
1
Question 4:
The mobile Phase in ion exchange chromatography isa)High salt phosphate buffer containing Naclb)Phosphate bufferc)Resind)Proteins
Question 5
What is the absorption wavelength of the proteins useda)595nmb)500nmc)600nmd)800nm
Links for further reading
Reference websites:1.http://oscar.iitb.ac.in
2.http://www.mnstate.edu/provost/ionexchangeprotocol.pdf
3.http://sbio.uct.ac.za/Sbio/documentation/Ion_exchange_chromatography.pdf
APPENDIX 2
SummaryAPPENDIX
3
The experiment mostly involves the sample preparation, column preparation, buffer preparation, elution buffer, followed by sample addition to the column, separation and mobile phase addition to the column in increasing concentration followed by elution, collection of sample for analysis and spectrometric analysis and interpretation.