Instrumental analysis in research

271
INSTRUMENTAL METHODS IN RESEARCH METHODOLOGY By Dr. M. Gopikrishna Reader, PG Dept. of Rasashastra SJG Ayurvedic Medical College, Koppal, Karnataka email: [email protected]

Transcript of Instrumental analysis in research

Page 1: Instrumental analysis in research

INSTRUMENTAL METHODS IN RESEARCH METHODOLOGY

ByDr. M. GopikrishnaReader, PG Dept. of RasashastraSJG Ayurvedic Medical College, Koppal, Karnatakaemail: [email protected]

Page 2: Instrumental analysis in research

TO IDENTIFIE NEW THINGS-

TO THROUGH MODERN LIGHT ON OLD FACTS-

Page 3: Instrumental analysis in research

TO UNDERSTAND ALL THESE WE NEED FEW PARAMETERS TO ASSES THEM HENCEFORTH FEW OF THE INSTRUMENTS ARE MENTIONED TO KNOW THE ANALYSIS OF THE DRUGS.EG;- ROOT PRESSURE IN VARIOUS SEASONS VARY AS BILVA IN GRISHMA RITU IS RICH IN TANINS OR COLOURING AGENT IS MORE,TO IDENTIFIE WE NEED INSTRUMENTS LIKE PHYTOCHEMICAL ANALYSIS AND MICROSCOPIC ANALYSIS.

Page 4: Instrumental analysis in research

PHYTOCHEMICAL ANALYSIS:-•

-TO ASSES AND ANALYSIS.

-WATER SOLUBLE MATERIALS WITH MANY COMPOUNDS ARE OBSERVED AND IDENTIFIED.

-SEPARATE CHEMICAL AND ISOLATE IT AND ASSES THE DIFFERENT BONDING WITH DIFFERENT ADVANCE TECHNIQUES.

Page 5: Instrumental analysis in research

-BIO-SYNTHETIC PATHWAY CAN BE ASSESED OF THE ISOLATED CHEMICAL.FOR THIS WE NEED TO-

* QUALITATIVE ANALYSIS.•

*QUANTITATIVE ANALYSIS.

SEMIQUALITATIVE AND SEMIQUANTITATIVEEG:-WATER EXTRACTS ETC

Page 6: Instrumental analysis in research

METHODS NEEDED FOR DOING ALL THESE-

1)INSTRUMENTAL–BY THIS WE CAN KNOW

PHYSICOCHEMICAL NATURE.

2)NON INSTRUMENTAL-AN ORGANOLEPTIC EVALUATION.

Page 7: Instrumental analysis in research

1)INSTRUMENTAL ANALYSIS:•

INSTRUMENTS PERFORM DIFFERENT FUNCTION EG;- REFRACTIVE INDEX OF OILS,OR SOLUBILITY OF ANY SUBSTANCE,TO KNOW ANY OF SUCH FACTORS WE NEED INSTRUMENTS TO CONFIRM THE FACTORS.

THE CONTACT OF THE PLANT TO A INSTRUMENT , ITS REACTION TO THE SAMPLE AND COMPARED TO DESIGN THIS GRASPED SIGNAL IN A MEASURABLE UNITS I.E TO CONVERT THE ORIGINAL SIGNALS TO A CONVENTIONAL SIGNAL.CONVERT THE VARIOUS ENERGIES TO A ELECTRIC ENERGY AND CAN BE ASSESED IN A GALVANOMETERETC.(IT’S A MEASURABLE FORM).SOMETIMES TRANSFORMING SIGNALS IS DIFFICULT THERE IN SUCH PLACES WE NEED AMPLIFIERS.

PRESENTATION OF DATA IN DIFFERENT WAYS IS TRANSFORMED,AMPLIFIED,AND GENERATED .I.E THESE OBSERVATIONS ARE DIRECTLY PRAPORTIONAL TO THE DATA.

Page 8: Instrumental analysis in research

ADVANTAGES:--TO ANALYSE A SAMPLE SMALL AMOUNT OF DRUG IS ENOUGH.-DETERMINATION OF QUALITATIVE AND QUANTITATIVE ANALYSIS IN SHORT TIME.-COMPLEX MIXTURES CAN ALSO BE ANALYSED WITH OR WITHOUT ISOLATION.-IT IS WITH SUFFICIENT RELIABILITY AND ACCURATE.

Page 9: Instrumental analysis in research

DISADVANTAGES:-

-TOO COSTLY.

-MAINTANANCE OF INSTRUMENT IS DIFFICULT.

-CHEMICAL CLEANLINESS IS ALSO COSTLY.

-SENSITIVITY DEPENDS ON ADVANCEMENT OF INSTRUMENT THEREFOR UPGRADE IT REGULARLY.

-SPECIALISED TEST FOR HANDLING OR USING IS NEEDED.

-FREQUENT NEED OF CHECKING DRUGS IS NECESSARY THEREFORE FREQUENTLY CHECK THE INSTRUMENT.

Page 10: Instrumental analysis in research

EXTRACTION OF DRUGS:-•

PHANTA,HIMA,SATWA,ARE EXTRACTS ONLY

-IN MODERN THE METHODS OF EXTACTIONS ARE.

*1)MACERATION•

*2)PERCOLATION.

*3)DECOCTION.

Page 11: Instrumental analysis in research

*1)MACERATION:-KEEPING THE POWDERED DRUG IN SUITABLE SOLVENT FOR FEW HOURS LIKE IN WATER,ETHER,METHENOL,WITHOUT APPLING HEAT TO DISSOLVE ITS SOLUBLE PORTION IN IT CALLED AS “ISOLATION MARC”

AND INSOLUBLE

PART CALLED AS “MARC”.SUB. NEEDED:-CONICAL GLASS,FILTER PAPER,EVOPARATING DISH,GLASS FUNNEL,HOT WATER BATH,ELECTRIC OVEN,ANALYTICAL WEIGHING BALANCE,PIPPETE,VOLUMETRIC FLASK,SOLVENTS,THE TEST DRUG.

Page 12: Instrumental analysis in research
Page 13: Instrumental analysis in research

PROCEDURE:TAKE 5GMS OF TEST DRUG POWDER I.E.AIR DRIED DRUG POWDER ,ADD 100ML OF WATER TO THIS AND PLACE IT IN A CONICAL FLASK SHAKE THE MIXTURE NOW AND THEN FOR 18HRS THEN FILTER IT.SEPERATE THE RESIDUE.FROM THE SOLVENT COLLECT 20ML AND PLACE IT IN A PORCELINE DISH AND DRY IT ON A HOT WATER BATH AND COLLECT THE RESIDUE AND WEIGH IT AND CALCULATE THE % OF THE EXTRACTIVE.

Page 14: Instrumental analysis in research
Page 15: Instrumental analysis in research
Page 16: Instrumental analysis in research

PERCOLATION:-IT IS THE METHOD OF EXTRACTION OF ALCOLOIDS ETC.BY PASSING A LIQUID THROUGH A COARS POWDER DRUG.THE LIQUID IS MADE TO PASS THE DRUG AND THE SOLVENT IS COLLECTED BY THE OPENING OF THE STOP CORK,ALSO THE SOLVENT IS MADE TO PASS THE COTTON PLUG AT THE NECK TO FILTER AND TO GAIN ONLY THE SOLLUBLE EXTRACT ONLY.

Page 17: Instrumental analysis in research

USE OF PERCOLATE: COLLECT THE PERCOLATE OF 5MIN INTERVAL PF 4-5 BATCHES OF SAME SOLVENT AND ASSES FOR THE RICH ALCOLOID SAMPLE.

EG;IF GOOD %IN FIRST SAMPLE OR 3RD

SAMPLE OR 5TH SAMPLE,THAT PARTICULAR SAMPLE CAN BE GIVEN TO THE PATIENT AND STANDARDISE THE SAMPLE.

Page 18: Instrumental analysis in research

DECOCTION:-•

IT IS A METHOD OF EXTRACTING OF EXTRACTIVES,HERE FIRE OR HEAT TREATMENT IS USED TO EXTRACT THE SOLVENTS,HERE THE COARSE POWDER IS TAKEN IN A APPARATUS AND A ARRANGEMENT IS MADE TO PASS THE HOT LIQUID THROUGH THE DRUG AND RECYCLE IT ,AS IT PASSES THROUGH THE DRUG IT IS FILTERED THROUGH A COTTON PLUG AND IS COLLECTED AT THE CHAMBER AT THE BASE WHICH IS HEATED AND THE VAPOUR IS MADE TO PASS THROUGH A PIPE INTO THE CONDENCER AREA AND MADE TO LIQUID FORM AND AGAIN PASS THROUGH THE DRUG.REPEAT THE PROCESS AS PER REQUIREMENT AND STANDARDISE IT..

Page 19: Instrumental analysis in research

SOXLET APPARATUS IS DESIGNED FOR THE SAME PURPOSE.

Page 20: Instrumental analysis in research
Page 21: Instrumental analysis in research

IF IT IS A COMPLEX COMPOUND SEPARATE THE EXTRACTS WITH ANY METHODS ,AS ONE METHOD IS-

BATCH EXTRACTION:-•

IT IS A LIQUID LIQUID EXTRACTION,OR EXTRACTION OF EXTRACTIVES FROM THE LIQUIDS.

TAKE A LIQUID SUBSTANCE IN A FUNNEL WITH CORKS ON EITHER SIDES,ALLOW IT FOR FEW HOURS AND LATER DISTINGUISH IT IN RESPECT TO ITS COLOUR OR CONSISTANCY,OF VISCOSITY,SEPARATE THEM AS PER REQUIREMENT INTO A BEAKER.

Page 22: Instrumental analysis in research

Separatory Funnel Extraction Procedure•

1. Inspect your separatory funnel.

The organic chem teaching labs have acquired several different types of sep funnels over the past years, with different types of stopcocks and stoppers, as illustrated in the photo below. The Teflon stopcocks work better than the ground glass stopcock; if you have a sep funnel with a ground glass stopcock, you may exchange for a Teflon style.

Page 23: Instrumental analysis in research

The sep funnel on the left is a 60 mL size, the others are 125 mL. The organic chem teaching labs are downscaling slowly to this smaller size.

Page 24: Instrumental analysis in research

There are two different styles of stopper, too. Some students swear by the plastic style, some by the ground glass style. The disadvantage of the ground glass style is that it can lodge permanently in the sep funnel if it is not removed and stored separately after use. Whichever style you have, make sure that the stopper fits snugly in the top of the flask.

Page 25: Instrumental analysis in research

2. Support the separatory funnel in a ring on a ringstand.The rings

are located on

the back shelves and they come in many sizes. Test to make sure that you haven’t chosen too large a ring before setting the funnel in it. You can add pieces of cut tygon

tubing

to the ring to cushion the funnel.

Make sure the stopcock of the separatory

funnel is closed!

Page 26: Instrumental analysis in research

3. Add the liquid to the separatory funnel.•

Place a stemmed funnel in the neck of the separatory funnel. Add the liquid to be extracted, then add the extraction solvent. The total volume in the separatory funnel should not be greater than three- quarters of the funnel volume. Insert the stopper in the neck of the separatory funnel.

Page 27: Instrumental analysis in research

pour in liquid to be extracted . . .

pour in the solvent . . .

add a stopper

Page 28: Instrumental analysis in research

4. Shake the separatory funnel.Pick up the separatory

funnel with the stopper in

place and the stopcock closed, and rock it once gently. Then, point the stem up and slowly open the stopcock to release excess pressure. Close the stopcock. Repeat this procedure until only a small amount of pressure is released when it is vented.

Page 29: Instrumental analysis in research

Now, shake the funnel vigorously for a few seconds. Release the pressure, then again shake vigorously. About 30 sec total vigorous shaking is usually sufficient to allow solutes to come to equilibrium between the two solvents.

Vent frequently to prevent pressure buildup, which can cause the stopcock and perhaps hazardous chemicals from blowing out. Take special care when washing acidic solutions with bicarbonate or carbonate since this produces a large volume of CO2

gas.

Page 30: Instrumental analysis in research

Let the funnel rest undisturbed until the layers are clearly separated.

While waiting, remove the stopper and place a beaker or flask under the sep funnel.

5. Separating the layers.

Page 31: Instrumental analysis in research

Carefully open the stopcock and allow the lower layer to drain into the flask. Drain just to the point that the upper liquid barely reaches the stopcock.

If the upper layer is to be removed from the funnel, remove it by pouring it out of the top of the funnel.

Page 32: Instrumental analysis in research

6. Perform multiple extractions as necessary.•

Often you will need to do repeat extractions with fresh solvent. You can leave the upper layer in the separatory funnel if this layer contains the compound of interest. If the compound of interest is in the lower layer, the upper layer must be removed from the separatory funnel and replaced with the drained-off lower layer, to which fresh solvent is then added.

Yes, it can be confusing! Plus, the beginning student often does not know in which layer resides the compound of interest. The best advice: Always save all layers until the experiment is completely finished!

Page 33: Instrumental analysis in research

7. Store your separatory funnel with the cap (stopper) separate from the funnel!

Please, remove the stopper before storage!!

Page 34: Instrumental analysis in research
Page 35: Instrumental analysis in research

DECANTATATION:-•

ALLOW THE SOLID PART TO SETTAL AND ONLY THE LIQUID PART TO BE REMOVED OR SEPERATED ITS DECANTATION,THIS CAN BE DONE EVEN WITH A CENTRIFUGAL MACHINE AND SEPERATED.

Page 36: Instrumental analysis in research

DISTILLATION:-

Page 37: Instrumental analysis in research

SEPERATION OF VOLATILE SUBSTANCESFROM A DRUG IN A DISTILATION APPARATUS.

Page 38: Instrumental analysis in research
Page 39: Instrumental analysis in research

CHROMATOGRAPHY:-IT IS A TECHNIQUE TO SEPARATEINDIVIDUAL COMPONENTS INA MIXTURE,IN A SPECIFIC MOBILE ANDSTATIONARY PHASE.

CHROMA= COLOUR.

GRAPHY= WRITING .

Page 40: Instrumental analysis in research

1ST INVENTED BY M.TSWETT IN 1906 BY A BOTONIST,HE USED CACO3 CRYSTALS IN A COLOUM AND POURED EXTRACTIVES IN COLOUM,ALSO POURED TEST SOUTION IN IT AND DEVELOPED COLOURED BANDS IN CACO3 .THEREFORE THIS SYSTEM WAS NAMED AS “SYSTEM OF COLOURED BANDS”.LATER NAMED AS CHROMATOGRAPHY WHICH TOOK TREMENDAROUS CHANGES TODAY TO SEPARATE ALMOST ANY SUBSTANCE IN A COMPLEX SOLVENT.

Page 41: Instrumental analysis in research

IN 1930-FEW VARITIES OF TLC AND ION EXCHANGE CROMATOGRAPHY WAS INTRODUCED.

IN 1941 PARTITION AND PAPER CHROMATOGRAPHY WAS DEVELOPED.

IN 1952 GAS CHROMATOGRAPHY WAS INTRODUCED.

TLC IS A TECHNIQUE PURELY BASED ON RATE OF MOUNT OF COMPONENT THROUGH A MEDIUM AND A STATIONARY PHASE,BINDING CAPACITY OF A MIXTURE IS SEPERATED .

Page 42: Instrumental analysis in research

2 PHASES IN CROMATOGAPHY.*STATIONARY PHASE -SOLID FORM OR A LIQUID IN SOLID FORM.*MOBILE PHASE- EITHER LIQUID OR GAS TECHNIQUEIN MOBILE PHASE PASSES THE STATIONARY PHASETRANSPORTS THE SEPARATE COMPONENT AT DIFFERENTSPEED AT DIRECTION OF FLOW OF MOBILE PHASE.

Page 43: Instrumental analysis in research
Page 44: Instrumental analysis in research

PRINCIPLE:•

SEPERATION OF SINGLE COMPONENT FROM A MIXTURE IN STABLE AND MOBILE PHASE.

2 PHASES NEEDED-

1)STATIONARY PHASE -USING SILICA GEL ALSO CELLULOS POWDER ETC.

-2)MOBILE PHASE- ETHANOL,BENZINE,CARBONTETRACHLORIDE,ETC . CAN BE USED.

Page 45: Instrumental analysis in research

APPLICATIONS OF TLC:-•

*IT IS MUCH BENIFICIAL WITH INDIVIDUAL COMPONENTS IN A MIXTURE.

*FOR CHECKING PURITY OF THE SAMPLE,ALSO FOR THE PURIFICATION PROCESS.

*HELPFUL IN CHEMISTRY LAB TO IDENTIFIE THE REACTION.

*FOR IDENTIFICATION OF INDIVIDUAL COMPONENTS.•

*STANDERD PARAMETER USEFUL FOR STANDERDISATION OF PHARMACEUTICAL PROCEDURE IN INDUSTRIES.

*ISOLATION OF MANY ORGANIC COMPOUNDS LIKE ALCOLOIDS,AMIDES,ACIDS ARE POSSIBLE.

*ALSO IN BIOCHEMICAL ANALYSIS IN METABOLITES LIKE PLASMA,SERUM ANALYSIS,URINE ANALYSIS.

Page 46: Instrumental analysis in research

ADVANTAGES:•

*SIMPLE AND EASY TEST IN STANDERDISATION.

*TIME IS 20-40MIN ,ITS VERY FAST.•

*SEPARATE INDIVIDUAL COMPONENTS FROM SMALL AMOUNT OF SAMPLE.

*ITS HIGHLY SEPERATION METHOD IN INDIVIDUAL COMPONENTS.

*LESS EXPENSIVE.•

*VERY EASY FOR DETECTION OF SAMPLE.

DIS ADVANTAGES:•

NOT POSSIBLE TO SEPARATE THE COMPONENTS IN LARGE SCALE.

Page 47: Instrumental analysis in research

Procedure for TLC1. Prepare the developing container.The developing container for TLC can be a specially designed chamber, a jar with a lid, or a beaker with a watch glass on the top:

In the teaching labs, we use a beaker with a watch glass on top.

Page 48: Instrumental analysis in research

Pour solvent into the beaker to a depth of just less than 0.5 cm.

Page 49: Instrumental analysis in research

To aid in the saturation of the TLC chamber with solvent vapors, line part of the inside of the beaker with filter paper.

Page 50: Instrumental analysis in research

Cover the beaker with a watch glass, swirl it gently, and allow it to stand while you prepare your TLC plate.

Page 51: Instrumental analysis in research

2. Prepare the TLC plate.TLC plates used in the organic chem. teaching labs are purchased as 5 cm x 20 cm sheets. Each large sheet is cut horizontally into plates which are 5 cm tall by various widths; the more samples you plan to run on a plate, the wider it needs to be.

Page 52: Instrumental analysis in research

Plates will usually be cut and ready for you when you come to lab.Handle the plates carefully so that you do not disturb the coating of adsorbent or get them dirty.

Page 53: Instrumental analysis in research

Measure 0.5 cm from the bottom of the plate. Take care not to press so hard with the pencil that you disturb the adsorbent.

Using a pencil, draw a line across the plate at the 0.5 cm mark. This is the origin: the line on which you will "spot" the plate.

Page 54: Instrumental analysis in research

Under the line, mark lightly the name of the samples you will spot on the plate, or mark numbers for time points. Leave enough space between the samples so that they do not run together, about 4 samples on a 5 cm wide plate is advised. Use a pencil and do not press down so hard that you disturb the surface of the plate. A close-up of a plate labeled "1 2 3" is shown to the right.

Page 55: Instrumental analysis in research

3. Spot the TLC plate•

The sample to be analyzed is added to the plate in a process called "spotting".

If the sample is not already in solution, dissolve about 1 mg in a few drops of a volatile solvent such as hexanes, ethyl acetate, or methylene chloride. As a rule of thumb, a concentration of "1%" or "1 gram in 100 mL" usually works well for TLC analysis. If the sample is too concentrated, it will run as a smear or streak; if it is not concentrated enough, you will see nothing on the plate. The "rule of thumb" above is usually a good estimate, however, sometimes only a process trial and error (as in, do it over) will result in well-sized, easy to read spots.

Page 56: Instrumental analysis in research

add a few drops of solvent . . .

. . . swirl until dissolved

The solution is applied to the TLC plate with a 1µL microcap.

Page 57: Instrumental analysis in research

Microcaps come in plastic vials inside red-and-white boxes. If you are opening a new vial, you will need to take off the silver cap, remove the white styrofoam plug, and put the silver cap back on. A small hole in the silver cap allows you to shake out one microcap at a time. Microcaps are very tiny; the arrow points to one, and it is hard to see in the photo.

Page 58: Instrumental analysis in research

Take a microcap and dip it into the solution of the sample to be spotted. Then, touch the end of the microcap gently to the adsorbent on the origin in the place which you have marked for the sample. Let all of the contents of the microcap run onto the plate. Be careful not to disturb the coating of adsorbent. dip the microcap into solution - the arrow points to the microcap, it is tiny and hard to see

Page 59: Instrumental analysis in research

make sure it is filled - hold it up to the light if necessary

touch the filled microcap to TLC plate to spot it - make sure you watch to see that all the liquid has drained from the microcap

Page 60: Instrumental analysis in research

rinse the microcap with clean solvent by first filling it . . .

. . . and then draining it by touching it to a paper towel

do this rinse process 3 times!

Page 61: Instrumental analysis in research

here's the TLC plate, spotted and ready to be developed

Page 62: Instrumental analysis in research

4. Develop the plate.•

Place the prepared TLC plate in the developing beaker, cover the beaker with the watch glass, and leave it undisturbed on your bench top. Run until the solvent is about half a centimeter below the top of the plate (see photos below).

Page 63: Instrumental analysis in research

place the TLC plate in the developing container - make sure the solvent is not too deep

Page 64: Instrumental analysis in research

The solvent will rise up the TLC plate by capillary action. In this photo, it is not quite halfway up the plate.

In this photo, it is about 3/4 of the way up the

plate.

The solvent front is about half a cm below the top of the plate - it is now ready to be removed

Page 65: Instrumental analysis in research

Remove the plate from the beaker.

quickly mark a line across the plate at the solvent front with a pencil

Allow the solvent to evaporate completely from the plate. If the spots are colored, simply mark them with a pencil.

Page 66: Instrumental analysis in research

5. Visualize the spots•

If your samples are colored, mark them before they fade by circling them lightly with a pencil.

Most samples are not colored and need to be visualized with a UV lamp. Hold a UV lamp over the plate and mark any spots which you see lightly with a pencil.

Beware! UV light is damaging both to your eyes and to your skin! Make sure you are wearing your goggles and do not look directly into the lamp. Protect your skin by wearing gloves.

If the TLC plate runs samples which are too concentrated, the spots will be streaked and/or run together. If this happens, you will have to start over with a more dilute sample to spot and run on a TLC plate.

Page 67: Instrumental analysis in research

this is a UV lamp

here are two proper sized spots, viewed under a UV lamp (you would circle these while viewing them)

Page 68: Instrumental analysis in research

The plate to the left shows three compounds run at three different concentrations. The middle and right plate show reasonable spots; the left plate is run too concentrated and the spots are running together, making it difficult to get a good and accurate Rf reading.

Page 69: Instrumental analysis in research

Here's what overloaded plates look like compared to well-spotted plates. The plate on the left has a large yellow smear; this smear

contains the same two compounds which are nicely resolved on the plate next to it. The plate to the far right is a UV visualization of

the same overloaded plate.

Page 70: Instrumental analysis in research

DETECTION OF THE ISOLATED COMPONENT:-

1)non specific method:-•

By colours-florensent phase

Iodine chamber.•

H2 so4 spray.

u.v chambers.•

Color spray reagents can be used.

Page 71: Instrumental analysis in research

2)specific method:-•

For different types of components-

i)phenolic compounds and tannins-ferric chloride spray is advised.

ii)for alcoloids-dragandroffs reagent.•

iii)for amino acids-ninhydrin in acetone.

iv)for cardiac glycosides-spray with 3,5dinitro benzoic acid.

Page 72: Instrumental analysis in research

FURTHER EVALUATION OF SEPERATED COMPONENTS:-

QUALITATIVE AND QUANTITATIVE-•

QUALITATIVE ANALYSIS:-

-By visual assessment by observing size,density,number of spots with different reagents of components can be identified.

-separately measure the spot in mm is directly proportional to substance present in that spot.

Page 73: Instrumental analysis in research

RETARDATION OR RETENTION FACTOR:‐Rf

• Measuring Rf

values

measurements are often taken from the plate in  order to help identify the compounds present. 

These measurements are the distance travelled

by  the solvent, and the distance travelled

by individual 

spots.

When the solvent front gets close to the top of the  plate, the plate is removed from the beaker and the  position of the solvent is marked with another line 

before it has a chance to evaporate.

Page 74: Instrumental analysis in research

The Rf

value for each dye is then worked out using the formula:

These measurements are then taken:

Page 75: Instrumental analysis in research

For example, if the red component travelled

1.7 cm from the base line while the solvent had travelled

5.0 cm, then the Rf

value for the red dye is:

Page 76: Instrumental analysis in research

If you could repeat this experiment under exactly the same conditions, then the Rf values for each dye would always be the same. For example, the Rf value for the red dye would always be 0.34. However, if anything changes (the temperature, the exact composition of the solvent, and so on), that is no longer true. You have to bear this in mind if you want to use this technique to identify a particular dye

Page 77: Instrumental analysis in research

QUANTITATIVE ANALYSIS:-•

Carried out in 2 ways-

Direct method:-•

a) on the plate i.e., after spray of different reagents.

b)by assessing the density of elute.•

c)measurement of spot area in mm is proportion to amount of quantity more in the sample.

d)densitometer:- method where intensity of color of substance is measured in chromatogram –in situ method.

E)densitometer-method where optical density of separated spots is measured.

f)spectrophotometer-instrument which gives qualitative and quantitative analysis. by wave length of maximum absorption of different spots and compared with standard spots.

Page 78: Instrumental analysis in research

densitometer

Page 79: Instrumental analysis in research

spectrophotometer-instrument

Page 80: Instrumental analysis in research

Indirect method:-•

Components scraped is assessed further with different test or different chromatograms etc. in different titrations.

Microanalysis can be performed by colorimeter ,electroporosis.etc.,

Page 81: Instrumental analysis in research

HPTLC:-HIGH PERFORMANCE THIN LAYER CROMATOGRAPHY-

In this a pre coated stationary phase is used and the chemicals used are extremely small sized particles so that adsorbent capacity is highly active.

Instead of maneuver samples the standard samples are available here for spotting and a new type of development chamber which requires less amount of solvent for development, more efficacy in separation and shorter analysis time because of advance type of densitometer scanner and improved data possessing capacity by which you will know the readings in computer.

Page 82: Instrumental analysis in research
Page 83: Instrumental analysis in research

PAPER CROMATOGRAPHY:-•

Technique in which analysis of unknown substance with the help of the mobile phase and a stationary phase is specially designed filter paper also known as whattman chromatography paper, all principals of chromatography holds good.

Page 84: Instrumental analysis in research

Adsorption ChromatographyAdsorption chromatography is probably one of the oldest types of chromatography around. It utilizes a mobile liquid or gaseous phase that is adsorbed onto the surface of a stationary solid phase. The equilibriation

between

the mobile and stationary phase accounts for the separation of different solutes.

Page 85: Instrumental analysis in research

Partition ChromatographyThis form of 

chromatography is based  on a thin film formed on the 

surface of a solid support by  a liquid stationary phase. 

Solute equilibriates between the mobile phase 

and the stationary liquid.

Page 86: Instrumental analysis in research

Types of Chromatography -

Radial chromatography

-

Ascending chromatography

-

Descending chromatography

Radial Chromatography In this type of chromatography, as the pigment separates, the different colours

move

outwards.

Page 87: Instrumental analysis in research

Ascending Chromatography The solvent moves upwards on the separating media

Page 88: Instrumental analysis in research

Descending Chromatography The solvent moves downwards on the separating media.

Page 89: Instrumental analysis in research

Ion Exchange  Chromatography

In this type of chromatography, the use of a resin (the stationary solid phase) is used to covalently attach anions or cations

onto it. Solute ions of the opposite charge in the mobile liquid phase are attracted to the resin by electrostatic forces.

Page 90: Instrumental analysis in research

Column Chromatography•

Procedure for Microscale Flash Column Chromatography

In microscale flash chromatography, the column does not need either a pinchclamp or a stopcock at the bottom of the column to control the flow, nor does it need air-pressure connections at the top of the column. Instead, the solvent flows very slowly through the column by gravity until you apply air pressure at the top of the column with an ordinary Pasteur pipet bulb.

Page 91: Instrumental analysis in research

(1) Prepare the column.The column is packed using a simple dry‐pack method. 

Plug a Pasteur pipet

with a small amount of cotton; use a wood applicator stick to tamp it down lightly. Take care that you do not use either too much cotton or pack it too tightly. You just need enough to prevent the adsorbent from leaking out.

Page 92: Instrumental analysis in research

Add dry silica gel adsorbent, 230-400 mesh --

usually the jar

is labeled "for flash chromatography." One way to fill the column is to invert it into the jar of silica gel and scoop it out . . .

Page 93: Instrumental analysis in research

. . . then tamp it down before scooping more out.

Another way to fill the column is to pour the gel into the column using a 10 mL beaker.

Page 94: Instrumental analysis in research

Whichever method you use to fill the column, you must tamp it down on the bench top to pack the silica gel. You can also use a pipet

bulb to force air into the

column and pack the silica gel.

When properly packed, the silica gel fills the column to just below the indent on the pipet. This leaves a space of 4–5 cm on top of the adsorbent for the addition of solvent. Clamp the filled column securely to a ring stand using a small 3-pronged clamp.

Page 95: Instrumental analysis in research

(2) Pre-elute the column.•

The procedure for the experiment that you are doing will probably specify which solvent to use to pre- elute the column. A non-polar solvent such as hexanes is a common choice.

Add hexanes (or other specified solvent) to the top  of the silica gel. The solvent flows 

slowly down the column; on the                           column above,it

has flowed down 

to the point marked by the arrow.

Page 96: Instrumental analysis in research

Monitor the solvent level, both as it flows through the silica gel and the level at the top. If you are not in a hurry (or busy doing something else), you can let the top level drop by gravity, but make sure it does not go below the top of the silica. Again, the arrow marks how far the solvent has flowed down the column.

Page 97: Instrumental analysis in research

Speed up the process by using a pipet

bulb to force the solvent

through the silica gel -

this puts the flash in microscale

flash

chromatography. Place the pipet

bulb on top of the column,

squeeze the bulb, and then remove the bulb while it is still squeezed. You must be careful not to allow the pipet

bulb to

expand before you remove it from the column, or you will draw solvent and silica gel into the bulb.

Page 98: Instrumental analysis in research

When the bottom solvent level is at the bottom of the column, the pre-elution process is completed and the column is ready to load.

Page 99: Instrumental analysis in research

If you are not ready to load your sample onto the column, it is okay to leave the column at this point. Just make sure that it does not go dry --

keep the top

solvent level above the top of the silica (as shown in the picture to the left) by adding solvent as necessary.

Page 100: Instrumental analysis in research

(3) Load the sample onto the silica gel column.Two different methods are used to load the column: the wet method and the dry method: wet and dry. Below are illustrations of both methods of loading a crude sample of ferrocene

onto

a column. In the wet method, the sample to be purified (or separated into components) is dissolved in a small amount of solvent, such as hexanes, acetone, or other solvent. This solution is loaded onto the column.

Page 101: Instrumental analysis in research

Wet loading method

The column is being loaded by the wet method.  Follow the thumbnails below to see close‐up details 

of the sample as it is allowed to sink into the  column. Once it's in the column, fresh eluting  solvent is added to the top and you are ready to 

begin the elution process (see step 4).

Page 102: Instrumental analysis in research

Sometimes the solvent of choice to load the sample onto the column is more polar than the eluting solvents. In this case, if you use the wet method of column loading, it is critical that you only use a few drops of solvent to load the sample. If you use too much solvent, the loading solvent will interfere with the elution and hence the purification or separation of the mixture. In such cases, the dry method of column loading is recommended.

Page 103: Instrumental analysis in research

Dry loading method

First dissolve the sample to be 

analyzed in the minimum 

amount of solvent and add 

about 100 mg of silica gel. Swirl 

the mixture until the solvent 

evaporates and only a dry 

powder remains. Place the dry 

powder on a folded piece of 

weighing paper and transfer it to 

the top of the prepared column. 

Add fresh eluting solvent to the 

top ‐‐

now you are ready to 

begin the elution process (see 

step 4). 

Page 104: Instrumental analysis in research

(4) Elute the column.•

Force the solvent through the column by pressing on the top of the Pasteur pipet

with a

pipet

bulb. Only force the solvent to the very top of the silica: do not let the silica go dry. Add fresh solvent as necessary.

Page 105: Instrumental analysis in research

The photo at the left shows the solvent being forced through the column with a pipet

bulb.

The series of 5 photos below show the colored compound as it moves through the column after successive applications of the pipet

bulb process.

The last two photos illustrate collection of the colored sample. Note that the collection beaker is changed as soon as the colored compound begins to elute.The process is complicated if the compound is not colored. In such experiments, equal sized fractions are collected sequentially and carefully labeled for later analysis.

Page 106: Instrumental analysis in research
Page 107: Instrumental analysis in research

(5) Elute the column with the second elution solvent.

If you are separating a mixture of one or more compounds, at this point you would change the eluting solvent to a more polar system, as previously determined by TLC. Elution would proceed as in step (4).

Page 108: Instrumental analysis in research

(6) Analyze the fractions.•

If the fractions are colored, you can simply combine like-colored fractions, although TLC before combination is usually advisable. If the fractions are not colored, they are analyzed by TLC (usually). Once the composition of each fraction is known, the fractions containing the desired compound(s) are combined.

Page 109: Instrumental analysis in research

Gas Chromatography: Procedure•

(1) Add the sample to be injected to the syringe.

A 25µL glass Hamilton syringe is used to inject the GC samples. Only 2-4 µL of sample is injected onto the column, which means that you fill only a small part of the barrel with sample. Examine the syringe carefully before you fill it. The divisions are marked "5 -

10 -

15 -

20 -

25".

Page 110: Instrumental analysis in research

This is a 25 µL glass Hamilton syringe. You only inject 2.5 µL, so it will NOT be filled to the top.

Page 111: Instrumental analysis in research

Place the tip of the needle in the liquid. Slowly draw up a small amount of liquid by raising the plunger, then press on the plunger to expel the liquid back into the liquid. This serves to “rinse”

the syringe with your sample, ensuring that what you will measure in the GC run is the

composition of your mixture. Repeat the rinse process one or two

times. Then, draw up the plunger slowly again while the needle is in the liquid and carefully fill the syringe with liquid about halfway to the “5”.

Page 112: Instrumental analysis in research

It is often hard to see the liquid in the syringe. If the syringe is clogged, the plunger will be in the correct position but the barrel of the syringe will be filled with only air, as in the bottom syringe in the photo to the left. The best thing to do is to carefully examing

the syringe after you think that you have filled it. Hold it up to the light to get a better view.Small air bubbles in the syringe will not affect the GC run (middle syringe in the photo to the left). As long as there is enough liquid in the syringe, the GC run will work fine. If you keep getting bubbles, just pull the plunger up a bit past the "halfway to the 5" mark to compensate. If you have a VERY large air bubble, you will not have enough liquid to show a reading on the GC (e.g., the bottom syringe in the photo).

Page 113: Instrumental analysis in research

(2) Inject the sample into the injector port.

You are need to do two things sequentially and quickly, so  make sure you know where the injection port is and where  the start button on the recorder is.

Push the needle of the syringe through the injection port  and immediately press the plunger to inject the sample, 

then immediately press the start button on the recorder.

You will feel a bit of resistance from the rubber septum in  the injection port; this is to be expected and you should be  prepared to apply some pressure to the syringe as you 

force the needle into the instrument all the way to the  base of the needle.

Page 114: Instrumental analysis in research

Push the needle of the filled syringe through the injector (as far as it will go) and quickly push the plunger.

Remove the syringe immediately . . .

Page 115: Instrumental analysis in research

. . . and quickly press the start button on the integrating recorder or the start recording button on the computer

Here's a close-up of the integrating recorder.

Page 116: Instrumental analysis in research

MECHANISM:-•

Gas is used as a mobile phase and solid or liquid

coated on a solid support is used as a stationary phase. The test mixture is converted to vapor and to this vapor the mobile phase is passed, hence the components through stationary phase with help of mobile phase and separated components which is less soluble in stationary phase travels faster and which is greater soluble capacity in stationary phase. Here with help of some gas , if you heat the test sample it travels and one with higher affinity will react here and one with less affinity will move far, we have detectors and amplifiers and lastly the recorder in different forms.

Page 117: Instrumental analysis in research

Important criteria in GC-•

1) Mobile phase as gas.

2) Test mixture heated and evaporated which should be inert to mobile phase.

3) To vaporize mixture if they posses volatile oil or volatile mixture are separated in technique.

4) Mobile should not mix with vapor in stationary phase.

5) Mixture sample should not change by heat i.e. thermostable, as we heat the mixture.

Page 118: Instrumental analysis in research
Page 119: Instrumental analysis in research

Here mobile gas as hydrogen,helium,nitrogen,and

argon are used,

among these helium is expensive.•

Result is in form of curves, individual peaks is of individual components and retention time is the time between injection of sample to the time to get maximum peak of components, retention factor of standard is time of maximum peak.i.e, rf.

Compare the test sample curve with standard curve asses it with considering the retention time even.

Page 120: Instrumental analysis in research

(3) Sit back and wait.•

Observe the recorder. Within several minutes, it should record several peaks.

(4) End the GC run.•

When you have seen all of the peaks which you suspect are in the mixture, or when the recorder has shown a flat baseline for a few minutes or so, press stop on the recorder.

When you press stop, the recorder will print out the peaks, the retention times, and the areas under the peaks. When it is done printing, you can press “enter”

a couple times to advance the

paper.•

Carefully tear the paper off the recorder. The paper is not perforated, so do not try to pull up and expect it to pop out of

the recorder. Instead, pull it down to start a tear from one edge, and then continue the tear until the paper is cut and free.

Page 121: Instrumental analysis in research
Page 122: Instrumental analysis in research

Uses:

mainly for detecting steroids,food

components    ,identification of various substances.

Page 123: Instrumental analysis in research

Study of colors

Page 124: Instrumental analysis in research

BEER LAMBERT LAW

Page 125: Instrumental analysis in research

MECHANISM

Page 126: Instrumental analysis in research

PROCESS INVOLVED

Page 127: Instrumental analysis in research

When a beam of light pass through a object , this object absorbs some amount of light and transmits the light to some extent.

I0 = I0 + IR + IT .•

Incidence of light = incidence of absorbance + incidence of refractence

+ incidence of transmitted.

Thickness of the object and concentration is much important for change in transmitted light.i.e., in various frequency depends on thickness.

Page 128: Instrumental analysis in research

LAMBERTS LAW:-•

When a beam of light is allowed to pass through a transparent media , the rate of decrease of intensity with the thickness of medium is directly proportional to the intensity of light.

If intensity is more. The frequency varies or •

If thickness is more. Changes.

Page 129: Instrumental analysis in research
Page 130: Instrumental analysis in research

Beers law:-Intensity of beam of light decreases with the increase in the concentration of absorbing substance.

Page 131: Instrumental analysis in research

1 cm

X                                                      y

x

2 cm

Y1       y

Page 132: Instrumental analysis in research

Similarly a concentration of solution with the light ray through them , the greater or lesser amount of concentration energy.

»»

Page 133: Instrumental analysis in research

Standard incidence energy = Is

Absorbance                            = As

Transmittance                         = Ts•

Suppose 

Is

= 230

As

= 4 of given standard.

Ts

= 6 of a given drug.

Then 

4     ‐

230

6     ‐

?

Calculate   6  X  230

4 =  345

Page 134: Instrumental analysis in research

colorimeter

Page 135: Instrumental analysis in research

••

Colorimeter Definition

A colorimeter is a device used in the practice of colorimetery, or the science of color. Colorimetry

has also been termed the

quantitative study of color perception. A number of instruments are available to determine the concentration of a solution. The simplest of those instruments is a colorimeter.

Page 136: Instrumental analysis in research

Function–

A colorimeter is a device used to measure the absorption of a specific wavelength of light by a solution, which can in turn be used to determine the concentration of a solute in a solution.

Components–

The critical parts of a colorimeter are a light source (commonly a low-voltage filament lamp), adjustable aperture, colored filters, solution cuvette, a transmitted light detector (such as a photoresistor) and an output display meter. Some colorimeters might also contain a regulator to prevent damage related to the machine from voltage fluctuations as well as a second light path to allow comparison between two solutions.

Page 137: Instrumental analysis in research

Filters–

A filter is used to select and measure the wavelength of light that the solution absorbs the most. The measurements are done in nanometers (nm). Typically, the wavelength used is between 400nm and 700nm.

Results–

Following the reading, the colorimeter will provide data in either an analog or digital formation, depending upon the machine. The data can be shown as transmittance on a linear percentage scale between 0 percent and 100 percent. It can also be shown as absorbance on a logarithmic scale between zero and infinity. Typically, the range of absorbance is from 0 to 2 with the ideal range being between 0 to 1.

Page 138: Instrumental analysis in research

History–

The colorimeter was invented by Jan Szczepanik

and 

applies the Beer‐Lambert law when determining the  concentration. The Beer‐Lambert law states that 

absorbance is proportional to the concentration of a  solute.

Different Parts of a Colorimeter

A colorimeter is a tool to measure the ability of a  solution to absorb a specified wavelength of light.  This absorption is used to determine the 

concentration of a substance in the solution.

Page 139: Instrumental analysis in research
Page 140: Instrumental analysis in research

Components–

The basic colorimeter is made up of a low‐voltage light 

source that shines through the solution held in a  cuvette. The cuvette

fits in a small receptacle and, as the 

light shines through it, a detector measures the light  that is passed through. The detector's readings are 

shown on an LCD display.

Options–

Colored filters are inserted in front of the cuvette

depending on which wavelength of light you need. This  is determined by knowing which wavelength of light the  solute you are measuring absorbs the best. Field and lab  manuals will have lists of these wavelengths for 

reference.

Page 141: Instrumental analysis in research

Function–

Once you select and insert the filter for the wavelength 

you need, you must analyze a blank and two to three  standards (minimum) using the colorimeter. The blank is 

typically pure water. Standards are solutions of various  concentrations of the solute you are measuring in your  sample. The readings from the blank and the standards  are used to set up a chart from which you determine 

your sample concentration after you analyze it.

Page 142: Instrumental analysis in research

Colorimeter Type

Colorimeters measure the perception of color. 

Colorimetry

is a technique to describe and quantify  the human perception of color by focusing on the 

physical aspects of color. A colorimeter measures  the amount of color from a given medium. Various 

different applications of colorimeters exist today to  quantify color, ranging from laboratories to the 

electronic industry.

Page 143: Instrumental analysis in research

Tristimulus

Colorimeter

Tristimulus

colorimeters are often used in the  application of digital imaging. The tristimulus

colorimeter measures color from light sources such  as lamps, monitors and screens. By taking multiple 

wideband spectral energy readings along the visible  spectrum, this colorimeter can profile and calibrate  specific output devices. The measured quantities 

can approximate tristimulus

values, which are the  three primary colors needed to match a test color

Page 144: Instrumental analysis in research

Densitometer–

A densitometer measures the density of light passing 

through a given frame. Density can be characterized as  the level of darkness in film or print. When an image is  printed, the ink pigments block light naturally when 

deposited by the printing process. Graphics industry  professionals use densitometers to help control color in 

the various steps of the printing process.

Page 145: Instrumental analysis in research

Spectroradiometer–

Spectroradiometers

quantify the spectral power 

distribution emitted from a given light source. In other  words, the spectroradiometer

measures the intensity of 

color.

Characteristically similar to spectrophotometers,  spectroradiometers

are used to evaluate lighting for 

sales within manufacturing and for quality control  purposes.

Other applications include confirming a 

customer's light source specifications and calibrating  liquid crystal displays for televisions and laptops.

Page 146: Instrumental analysis in research

Spectrophotometer–

A spectrophotometer is an analytical tool that measures 

the reflection and transmittance properties of a color  sample. Using functions of light wavelengths, the 

spectrophotometer passes a beam of light through the  sample to record both absorbance and transmittance. 

The instrument does not require human interpretation  and is much more complex than a standard colorimeter. 

Common applications for the spectrophotometer  include color formulation and industry research and 

development.

Page 147: Instrumental analysis in research

What Is Absorbance When Dealing With a Colorimeter?

Absorption is measured by color intensity. 

Absorbency is crucial in determining concentration of a  substance in a sample through colorimeter analysis. 

Colorimeters measure the intensity of color and light  transmittance by the sample to achieve the concentration.

Function

A colorimeter works by shining a white light through an  optical filter into a cuvette

that holds the sample. By 

adding a color reagent to the sample, the substance will  darken in color depending on the amount of 

concentration. The darkening (intensity) of the color is  measured by how much light is absorbed by the sample. 

The higher the intensity, the higher the concentration.

Page 148: Instrumental analysis in research

Significance

Absorbance can be defined as a logarithmic measurement of the 

amount of light at a particular wavelength taken in by sample. 

This measurement is known as the Beer‐Lambert Law or Beer's 

Law. The law states that absorbance is equal to the concentration 

multiplied by the path length of light and the molar extinction 

coefficient (how strongly the solution absorbs color).

Color Importance

The color of the sample depends on the transmittance of light, 

not absorption. For example, a sample is seen as red because it 

absorbs blue and purple wavelengths. Therefore in that example, 

the wavelength of light used to determine absorbance will be in 

the blue region of the visible color spectrum.

Page 149: Instrumental analysis in research

How to Use Colorimeters•

A colorimeter is any device that measures the color of a particular substance. It's a generic term that may refer to a range of devices such as a spectrophotometer, which is a specific type of colorimeter typically used to measure a solution's absorbance of a particular wavelength of light. This allows the concentration of a known solute in the solution to be

calculated with considerable accuracy. This type of colorimeter is a standard piece of equipment in analytical chemistry.

InstructionsThings You'll Need

Colorimeter •

Reference solution

Test solution

Page 150: Instrumental analysis in research

1 •

Determine the wavelength of light that the solution absorbs most strongly. A colorimeter has a set of changeable filters that can show which color of light to examine for the greatest accuracy. Set the colorimeter to this wavelength.–

2

Calibrate the colorimeter. Turn the unit on and wait 30 minutes for the colorimeter to warm up before taking any measurements. Remove the container (cuvette) from its chamber.–

3

Read the colorimeter. The meter is extremely sensitive, and you must use the correct measurement. Most models have a reflective surface under the needle and you should look at the needle so that you can't see its reflection in the mirror.

Page 151: Instrumental analysis in research

4 •

Fill the reference cuvette

with the test solution as

indicated by the colorimeter model you're using. Clean the reference cuvette

carefully to ensure it doesn't

have any smudges and place it in the chamber. Read the absorbance and adjust the colorimeter to show an absorbance reading of 0.–

5

Put the test solution in the specimen cuvette

and clean the outside of the cuvette. Place it in the chamber and read the absorbance of the test solution. You will typically record this value and use it in calculations to determine the solution's concentration.

Page 152: Instrumental analysis in research

Colorimeter Analysis•

Colorimeters plot concentration and absorbance on a line graph.

Using a colorimeter is one of the fastest and easiest ways to measure unknown concentrations of a sample substance. Measuring absorbency is crucial for colorimetry

analysis since it is in a linear relationship

with concentration.•

Function–

Substances in a sample absorb light, and color pigments also absorb light at different wavelengths. Colorimeters use color and light at different wavelengths to measure the intensity of color and how much light that color absorbs in a sample.

Page 153: Instrumental analysis in research

Beer-Lambert Law–

To obtain the concentration of a substance in a sample, colorimeters use the Beer-Lambert law to gain results. The mathematical formula states that concentration is equal to the absorbance divided by the path length of light and the molar extinction coefficient (how strong the solution absorbs color).

Considerations–

To get the curve for the Beer-Lambert law, standards of known concentration are measured for their absorbance of color and light and then plotted on a graph. This graph uses the y = mx

+ b formula to achieve a straight line with

concentration on the x-axis and absorbance on the y-axis. Unknown samples will then be measured using this curve. For example, if y = .301, then the absorbance (determined by the colorimeter) multiplied by .301 will equal the concentration.

Page 154: Instrumental analysis in research

Application of Colorimeter

Essentially, a colorimeter is a scientific instrument that  measures the amount of light passing through a solution 

relative to the amount that passes through a sample of  pure solvent. Colorimeters have many applications in the 

fields of biology and chemistry.

Beer‐Lambert Law

The Beer‐Lambert Law states that the concentration of a  dissolved substance, or solute, is proportional to the 

amount of light that it absorbs. A common application of a  colorimeter is therefore to determine the concentration of  a known solute in a given solution

Page 155: Instrumental analysis in research

Biological Culture–

In biology, a colorimeter can be used to monitor the 

growth of a bacterial or yeast culture. As the culture  grows, the medium in which it is growing becomes 

increasingly cloudy and absorbs more light.

Bird Plumage Coloration–

A colorimeter can also be used to eliminate subjectivity, 

or personal opinion, from the assessment of color in bird  plumage. Modern colorimeters provide highly accurate 

results, under standard, repeatable conditions and have  proved to be very reliable.

Page 156: Instrumental analysis in research

What is the Difference Between a Colorimeter and a  Spectrophotometer? 

Composed of many different wavelengths and colors,  perceived light can be deconstructed using colorimeters 

and spectrophotometers in different ways. 

Since the advent of the XYZ color system established by the  Commission Internationale

de l'Eclairage

(CIE) in 1931, 

many devices have been invented to measure and  interpret light ‐‐‐

a science known as spectroscopy. Two of 

these machines ‐‐‐

colorimeters and reflectance  spectrophotometers ‐‐‐

have made homes in laboratories 

and design studios, but their similar quantitative outputs  can be confusing despite their very different functionality  and design.

Page 157: Instrumental analysis in research

Color Measurement–

In 1931 the Commission Internationale

de l'Eclairage

attempted to quantify the light that humans perceive by matching the three primary colors that make up all colors with three values, called the tristimulus

values ---

x, y and z -

--

which approximately correspond to red, blue and green. Any visible color can be quantified using these three values, and this allowed for objective measuring and comparing of colors.

Colorimeters–

Relying on exactly that system laid out by the CIE, colorimeters measure the color of a sample compared to a white control surface and output data for x, y and z values. Used to color-match or color-mix, these relatively simple devices can often be found in the textile, paint and design industries. Colorimeters are generally inexpensive and rugged devices.

Page 158: Instrumental analysis in research

Colorimeters in the Lab–

Known concentrations of a sample can be tested with a 

colorimeter and plotted on a graph to create a  calibration curve. Using this data, a sample with an 

unknown concentration can be tested and compared  against the graph to ascertain its concentration.

Page 159: Instrumental analysis in research

Reflectance Spectrophotometers

Similar in the respect that reflectance spectrophotometers can 

also be used to determine a sample's color, these more complex 

devices provide a greater amount of data, and are otherwise 

completely different machines. Rather than just one broad 

wavelength analysis, spectrophotometers analyze the intensities 

of 16 or more narrow wavelengths of the sample light by 

diffracting the light into its component wavelengths. Unlike 

colorimeters, spectrophotometers typically include adjustments 

for many variables such as observer angle and illuminant, and are 

usually connected directly to a computer for data analysis. Some

spectrophotometers include a gloss trap to either include or 

exclude the specular

components of reflected light from either 

xenon or tungsten lamps.

Page 160: Instrumental analysis in research

Spectrophotometers in the Lab–

Reflectance spectrophotometers have a variety of 

applications in many laboratory disciplines. Similar to  colorimeters, spectrophotometers can more accurately 

determine the concentration of solute in a solution  against a graphed curve. But these more expensive  devices can also be used to, for example, measure the 

rate of bacterial growth in a sample or calculate the  color‐changing effect of a chemical reaction in real time.

Page 161: Instrumental analysis in research

Difference Between Colorimeter & Spectrophotometer

Colorimeters and spectrophotometers are used in the medical, research, and scientific industries to provide rapid results of unknown samples through the use of color. Although both the colorimeter and the spectrophotometer use color to analyze samples, they operate differently.

Light–

Colorimeters use a colored light beam to measure sample concentration. Spectrophotometers use a white light that is passed through a slit and filter to analyze samples.

Page 162: Instrumental analysis in research

Wavelengths

In colorimetry, colored light passes through an optical  filter to produce a single band of wavelengths. With 

spectrophotometers, the white light is passed through a  special filter that disperses the light into many bands of 

wavelengths.

Measurement–

Both of these techniques use the Beer‐Lambert Law to 

determine concentration. The difference is that  colorimeters measure the absorbency of light in a 

sample while spectrophotometers measure the amount  of light that passes through it.

Page 163: Instrumental analysis in research

Function–

The colorimeter uses psychophysical analysis, comparing color in the same manner as human eye-brain perception. The spectrophotometer uses only physical analysis, using different wavelengths of light to determine the reflection and transmission properties of color.

Page 164: Instrumental analysis in research

STUDY OF MULTICOLOR

Page 165: Instrumental analysis in research

SPECTROPHOTOMETER

Page 166: Instrumental analysis in research

The spectrophotometer

is an instrument which measures the amount of light of a specificed

wavelength which passes through a medium. According to Beer's law, the amount of light absorbed by a medium is proportional to the concentration of the absorbing material or solute present.

Page 167: Instrumental analysis in research

Thus the concentration of a colored solute in a solution may be determined in the lab by measuring the absorbency of light at a given wavelength.

Wavelength (often

abbreviated as lambda) is measured in nm.

The

spectrophotometer allows selection of a wavelength pass through the solution.

Usually,

the wavelength chosen which corresponds to the absorption maximum of the solute. Absorbency is indicated with a capital A.

Page 168: Instrumental analysis in research

To familiarize yourself with the spectrophotometer, illustrate and label the following features which are important to its proper use. You should know the function and/or significance of each of these features before you use the instrument.

Page 169: Instrumental analysis in research

At the spectrophotometer, you should have two cuvettes

in a plastic

rack.

Solutions which are to be read are poured into cuvettes

which are inserted into the machine. One should be marked "B"for

the blank and one "S" for your sample.

Page 170: Instrumental analysis in research

.

A wipette

should be available to polish them before insertion into the cuvette

chamber.

Cuvettes are carefully manufactured for their optical uniformity and are quite expensive.

They should be handled with

care so that they do not get scratched, and stored separate from standard test tubes.

Page 171: Instrumental analysis in research

Try not to touch them except at the top of the tube to prevent finger smudges which alter the reading.

For experiments in which minor

inprecision

is acceptible, clean, unscratched 13 x 100 mm test tubes may be used.

Page 172: Instrumental analysis in research
Page 173: Instrumental analysis in research

WARM-UP: 1. Plug in and turn on (left hand front dial, labeled ZERO in the illustration). Allow about 30 minutes for warm up.

Page 175: Instrumental analysis in research

Do not touch this knob again during the rest of the following procedure.

Page 180: Instrumental analysis in research

8.

Note that the scale for absorbance is the lower scale on the dial, and should be read from R to L .

Page 181: Instrumental analysis in research

For all readings of the dial, line up the reflection of the needle in the mirror behind the dial with the needle itself.

Otherwise,

parallax error will occur, giving an erroneous reading.

The illustration shows the correctly

aligned dial with a reading of 0.116.

Page 182: Instrumental analysis in research

PARALLAX ERROR: Here, the picture was taken of the identical solution as in the previous image, but with the point of view too far to the right.

Note that

the needle reflection is to the right of the needle.

The apparent reading is 0.120.

Page 183: Instrumental analysis in research

PARALLAX ERROR: Here, the picture was taken of the identical solution as in the previous image but with the point of view too far to the left.

Note

that the needle reflection is to the left of the needle.

The apparent reading is 0.113.

Page 184: Instrumental analysis in research

9. If you read additional specimens, you  should confirm that the machine is still 

zeroed and blanked out, as in steps 2, 4 and 5  for all readings.

Page 185: Instrumental analysis in research

CLEAN UP: 10. Remove cuvette

from machine, carefully

wash and store spectrophotometer cuvettes keeping them separate from regular test tubes.

Return spectophotometer

to its storage location.

Page 186: Instrumental analysis in research
Page 187: Instrumental analysis in research
Page 188: Instrumental analysis in research
Page 189: Instrumental analysis in research
Page 190: Instrumental analysis in research
Page 191: Instrumental analysis in research

INFRARED SPECTROSCOPY

INFRARED SPECTROSCOPY

Page 192: Instrumental analysis in research

Infrared spectroscopy exploits the fact that molecules absorb specific frequencies that are characteristic of their structure. These absorptions are resonant frequencies, i.e. the frequency of the absorbed radiation matches the frequency of the bond or group that vibrates. The energies are determined by the shape of the molecular potential energy surfaces, the masses of the atoms, and the associated vibronic

coupling.

Page 193: Instrumental analysis in research

In particular, in the Born–Oppenheimer

and harmonic approximations, i.e. when the molecular Hamiltonian

corresponding to the

electronic ground state

can be approximated by a harmonic oscillator

in the neighborhood of the

equilibrium molecular geometry, the resonant frequencies are determined by the normal modes

corresponding to the molecular

electronic ground state potential energy surface. Nevertheless, the resonant frequencies can be in a first approach related to the strength of the bond, and the mass of the atoms

at either

end of it. Thus, the frequency of the vibrations can be associated with a particular bond type.

Page 196: Instrumental analysis in research

Uses and applications•

Infrared spectroscopy is a simple and reliable technique widely used in both organic and inorganic chemistry, in research and industry. It is used in quality control, dynamic measurement, and monitoring applications such as the long-term unattended measurement of CO2

concentrations in greenhouses and growth chambers by infrared gas analyzers.

It is also used in forensic analysis

in both criminal and civil cases, for example in identifying polymer degradation. It can be used in detecting how much alcohol is in the blood of a suspected drink driver measured as 1/10,000 g/mL

= 100 μg/mL.

Page 197: Instrumental analysis in research

A useful way of analysing

solid samples without the need for cutting samples uses ATR or attenuated total reflectance

spectroscopy. Using this approach,

samples are pressed against the face of a single crystal. The infrared radiation passes through the crystal and only interacts with the sample at the interface between the two materials.

With increasing technology in computer filtering and manipulation of the results, samples in solution can now be measured accurately (water produces a broad absorbance across the range of interest, and thus renders the spectra unreadable without this computer treatment).

Page 198: Instrumental analysis in research

Infrared spectroscopy has also been successfully utilized in the field of semiconductor microelectronics,for

example,

infrared spectroscopy can be applied to semiconductors like silicon, gallium arsenide, gallium nitride, zinc selenide, amorphous silicon, silicon nitride, etc.

The instruments are now small, and can be transported, even for use in field trials.

Infrared spectroscopy is also useful in measuring the degree of polymerization in polymer

manufacture.

Page 199: Instrumental analysis in research
Page 203: Instrumental analysis in research
Page 204: Instrumental analysis in research

ULTRAVIOLET–VISIBLE SPECTROSCOPY

Page 205: Instrumental analysis in research

•Violet: 400 -

420 nm •Indigo: 420 -

440 nm •Blue: 440 -

490 nm •Green: 490 -

570 nm •Yellow: 570 -

585 nm •Orange: 585 -

620 nm Red: 620 -

780 nm

Page 206: Instrumental analysis in research
Page 207: Instrumental analysis in research

A diagram of the components of a typical  spectrometer are shown in the following diagram. 

The functioning of this instrument is relatively  straightforward. A beam of light from a visible  and/or UV light source (colored red) is separated 

into its component wavelengths by a prism or  diffraction grating. Each monochromatic (single 

wavelength) beam in turn is split into two equal  intensity beams by a half‐mirrored device. One 

beam, the sample beam (colored magenta), passes  through a small transparent container (cuvette) 

containing a solution of the compound being  studied in a transparent solvent.

Page 208: Instrumental analysis in research

The other beam, the reference (colored blue),  passes through an identical cuvette

containing only 

the solvent. The intensities of these light beams are  then measured by electronic detectors and 

compared. The intensity of the reference beam,  which should have suffered little or no light 

absorption, is defined as I0

. The intensity of the  sample beam is defined as I. Over a short period of 

time, the spectrometer automatically scans all the  component wavelengths in the manner described.  The ultraviolet (UV) region scanned is normally 

from 200 to 400 nm, and the visible portion is from  400 to 800 nm.

Page 209: Instrumental analysis in research

Applications•

UV/Vis

spectroscopy is routinely used in analytical chemistry

for

the quantitative

determination of different analytes, such as transition metal

ions, highly conjugated

organic compounds,

and biological macromolecules. Determination is usually carried out in solutions.

Solutions of transition metal ions can be colored (i.e., absorb visible light) because d electrons

within the metal atoms can be

excited from one electronic state to another. The colour

of metal ion solutions is strongly affected by the presence of other species, such as certain anions or ligands. For instance, the colour

of a dilute solution of copper sulfate

is a very light blue;

adding ammonia

intensifies the colour

and changes the wavelength of maximum absorption (λmax ).

Page 210: Instrumental analysis in research

Organic compounds, especially those with a high degree of conjugation, also absorb light in the UV or visible regions of the electromagnetic spectrum. The solvents for these determinations are often water for water soluble compounds, or ethanol

for organic-

soluble compounds. (Organic solvents may have significant UV absorption; not all solvents are suitable for use in UV spectroscopy. Ethanol absorbs very weakly at most wavelengths.) Solvent polarity and pH can affect the absorption spectrum of an organic compound. Tyrosine, for example, increases in absorption maxima and molar extinction coefficient when pH increases from 6 to 13 or when solvent polarity decreases.

Page 211: Instrumental analysis in research

While charge transfer complexes

also give rise to  colours, the colours

are often too intense to be 

used for quantitative measurement. 

Page 213: Instrumental analysis in research
Page 214: Instrumental analysis in research

Raman spectroscopy

Page 215: Instrumental analysis in research

named after C. V. Raman

is a spectroscopic technique used to study vibrational, rotational,

and other low-frequency modes in a system.[1]

It

relies on inelastic scattering, or Raman scattering, of monochromatic

light, usually from

a laser

in the visible, near infrared, or near ultraviolet

range. The laser light interacts with

molecular vibrations, phonons

or other excitations in the system, resulting in the energy of the laser photons being shifted up or down. The shift in energy gives information about the vibrational

modes in the system.

Infrared spectroscopy

yields similar, but complementary, information.

Page 216: Instrumental analysis in research

Typically, a sample is illuminated with a laser beam. Light from the illuminated spot is collected with a lens

and sent through a

monochromator. Wavelengths close to the laser line, due to elastic Rayleigh

scattering, are

filtered out while the rest of the collected light is dispersed onto a detector.

Spontaneous Raman scattering

is typically very weak, and as a result the main difficulty of Raman spectroscopy is separating the weak inelastically

scattered light from the intense

Rayleigh

scattered laser light.

Page 217: Instrumental analysis in research

Applications•

Raman spectroscopy is commonly used in chemistry, since vibrational

information is

specific to the chemical bonds

and symmetry of molecules. Therefore, it provides a fingerprint by which the molecule can be identified.

Another way that the technique is used to study changes in chemical bonding

Raman gas analyzers have many practical applications. For instance, they are used in medicine for real-time monitoring of anaesthetic

and respiratory gas mixtures during surgery.

Page 218: Instrumental analysis in research

In solid state physics, spontaneous Raman spectroscopy is used to, among other things, characterize materials, measure temperature, and find the crystallographic orientation of a sample

Raman spectroscopy is being investigated as a means to detect explosives

for airport security

Page 220: Instrumental analysis in research

Nuclear magnetic resonance (NMR) is a physical  phenomenon

in which magnetic nuclei in a 

magnetic field absorb and re‐emit electromagnetic  radiation. This energy is at a specific resonance

frequency which depends on the strength of the  magnetic field and the magnetic properties of the 

isotope

of the atoms. NMR allows the observation  of specific quantum mechanical

magnetic

properties of the atomic nucleus

Page 221: Instrumental analysis in research

Many scientific techniques exploit NMR phenomena to study molecular physics, crystals, and non-crystalline materials through NMR spectroscopy. NMR is also routinely used in advanced medical imaging

techniques, such

as in magnetic resonance imaging

(MRI).•

All isotopes that contain an odd number of protons

and/or of neutrons

(see Isotope) have

an intrinsic magnetic moment

and angular momentum, in other words a nonzero spin, while all nuclides

with even numbers of both

have a total spin of zero. The most commonly studied nuclei are 1 H

Page 222: Instrumental analysis in research

A key feature of NMR is that the resonance

frequency of a particular substance is directly proportional to the strength of the applied magnetic field. It is this feature that is exploited in imaging techniquesThe principle of NMR usually involves two sequential steps:

The alignment (polarization) of the magnetic nuclear spins in an applied, constant magnetic field

H0

. •

The perturbation of this alignment of the nuclear spins by employing an electro-magnetic, usually radio frequency (RF) pulse. The required perturbing frequency is dependent upon the static magnetic field (H0

) and the nuclei of observation.

Page 223: Instrumental analysis in research

History•

Nuclear magnetic resonance was first described and measured in molecular beams by Isidor

Rabi

in 1938,[1]

and in 1944, Rabi was

awarded the Nobel Prize in physics

for this work.[2]

In 1946, Felix Bloch

and Edward Mills

Purcell

expanded the technique for use on liquids and solids, for which they shared the Nobel Prize in Physics

in 1952

Page 224: Instrumental analysis in research

Theory of nuclear magnetic resonance•

Nuclear spin and magnets

All nucleons, that is neutrons

and protons, composing any atomic nucleus, have the intrinsic quantum property of spin. The overall spin of the nucleus is determined by the spin quantum number

S.

. It is this magnetic moment that allows the observation of NMR absorption spectra caused by transitions between nuclear spin levels. Most nuclides (with some rare exceptions) that have both even numbers of protons and even numbers of neutrons, also have zero nuclear magnetic moments, and they also have zero magnetic dipole and quadrupole

moments. Hence, such nuclides do not exhibit any NMR absorption spectra.

Page 225: Instrumental analysis in research

NMR spectroscopy is one of the principal techniques used to obtain physical, chemical, electronic and structural information about molecules

due to either the chemical shift,

Zeeman

effect, or the Knight shift

effect, or a combination of both, on the resonant frequencies of the nuclei present in the sample. It is a powerful technique that can provide detailed information on the topology, dynamics and three-dimensional structure of molecules in solution and the solid state. Additional structural and chemical information may be obtained by performing double-quantum NMR experiments for quadrupolar

nuclei

Page 226: Instrumental analysis in research

Applications•

Medical MRI

The application of nuclear magnetic resonance best known to the general public is magnetic resonance imaging

for medical diagnosis and magnetic

resonance microscopy

in research settings, however, it is also widely used in chemical studies, notably in NMR spectroscopy such as proton NMR, carbon-13 NMR, deuterium NMR and phosphorus-31 NMR. Biochemical information can also be obtained from living tissue (e.g. human brain

tumors) with the

technique known as in vivo magnetic resonance spectroscopy

or chemical shift NMR Microscopy.

Page 228: Instrumental analysis in research

Chemistry

By studying the peaks of nuclear magnetic  resonance spectra, chemists can determine the 

structure of many compounds. It can be a very  selective technique, distinguishing among many 

atoms within a molecule or collection of molecules  of the same type but which differ only in terms of 

their local chemical environment

Page 229: Instrumental analysis in research

ELECTROPORESIS

Page 230: Instrumental analysis in research

Electrophoresis is a molecular separation technique that involves the use of high voltage electric current for inducing the movement of charged molecules---proteins, DNA, nucleic acids---in a support medium.The movement of charged molecules is called mobility. The mobility of molecules is towards the opposite charge, for instance, a protein molecule with a negative charge moves toward the positive pole of the support medium. The medium may be a paper, a gel or a capillary tube.

Page 231: Instrumental analysis in research

HOW TO DEFINE ELECTROPHORESIS•

Electrophoresis is any process that uses electricity to separate particles in a fluid. It's a common method of identifying molecules according to some criteria such as molecular weight. Electrophoresis may also be used to prepare a sample for some other scientific technique. There are many different types of electrophoresis and this term doesn't refer to a specific process.

Page 232: Instrumental analysis in research

ELECTRICAL CHARGE•

Electrical current is placed at one end of the gel, with positive electrodes at the top of the gel --

closest to

the pits with the DNA inside --

and negative electrodes at the bottom. Due to the fact DNA is negatively charged, it will move through the gel toward the bottom and the positive electrodes. The gel will offer resistance for the DNA, so it will take larger strands a longer time to pass through the gel than shorter strands. The process is stopped at a predetermined time, and wherever the DNA is in the gel, that will give an indication of how large each of the strands were. This final image, after the electrophoresis, is the DNA "fingerprint."

Page 233: Instrumental analysis in research

Capillary electrophoresis is an analytical technique that separates ions based on their electrophoretic

mobility with the use of an applied voltage.

The electrophoretic

mobility is dependent upon the charge

of the molecule, the viscosity, and the atom's radius.

The rate at which the particle moves is directly

proportional to the applied electric field--the greater the field strength, the fast the mobility. Neutral species are not affected, only ions move with the electric field.

Page 234: Instrumental analysis in research

If two ions are the same size, the one with greater charge will move the fastest.

For ions

of the same charge, the smaller particle has less friction and overall faster migration rate. Capillary electrophoresis is used most predominately because it gives faster results and provides high resolution separation.

It is a

useful technique because there is a large range of detection methods Available.

Page 235: Instrumental analysis in research

MICROSCOPIC ELECTROPHORESIS•

A technique in which the electrophoresis of individual particles is observedwith the aid of a microscope or ultra-

microscope.

Page 236: Instrumental analysis in research

The moving boundary method was the first to be used by Tiselius

to demonstrate the efficacy

of the electrophortic

process. The apparatus consisted of a U tube the horizontal lower portion of the U tube being filled with a mixture of the substances under examination dispersed in a suitable buffer. The two vertical limbs were filled solely with buffer and the cathode and anode dipped into the buffer at the top of each limb respectively.

Page 237: Instrumental analysis in research

ELECTROPHORESIS IN FREE SOLUTIONMOVING BOUNDARY ELECTROPHORESIS: Arne Tiselius

(1937) developed the

moving boundary technique for the electrophoretic separation of substances,

for which, besides his work on adsorption analysis, hereceived the Nobel prize in 1948. The sample, a mixture

of proteinsfor example, is applied in a U-shaped cell filled with a

buffer solutionand at the end of which electrodes are immersed. Under

the influence

Page 238: Instrumental analysis in research

of the applied voltage, the compounds will migrate at different velocities towards the anode or the cathode depending on their charges. The changes in the refractive index at the boundary during migration can be detected at both ends on the solution using Schlieren

optics.

Page 239: Instrumental analysis in research

ZONE ELECTROPORESIS:-– a sophisticated instrument, which has a chamber ,

two separate containers for buffers, a stage and electric supply.

watt man paper is taken onto the stage and a small quantity of the test solution is placed over it and switch on the instrument wherein due to the electric supply and the capillary action of the substance, the migration of the test solution particles occur and separate bands are formed where in the electrically charged particles can be separated.

Page 240: Instrumental analysis in research

TYPES:‐2 types‐

a) vertical

b) horizontal.

Page 241: Instrumental analysis in research

Make up the gel which the DNA will be  put into

Presenter
Presentation Notes
Square tray 2-3 cm of agarose gel which is left to set, can also be made of starch or polyacrylamide, special comb put in so that there are small wells left in the gel
Page 242: Instrumental analysis in research
Page 243: Instrumental analysis in research
Presenter
Presentation Notes
A comb is put in the gel to create holes which we call wells
Page 244: Instrumental analysis in research
Page 245: Instrumental analysis in research

Dye added to the DNA

Presenter
Presentation Notes
Makes the sample visible when it is put into the agarose wells
Page 246: Instrumental analysis in research

Buffer solution added to the tank

Presenter
Presentation Notes
This ensures that the electric current goes through the whole tank and that maintains that ions can move in the solution
Page 247: Instrumental analysis in research

DNA samples loaded into wells 

Presenter
Presentation Notes
Glycerol also in the loading dye
Page 248: Instrumental analysis in research

Electrical current applied to the  chamber

Presenter
Presentation Notes
Safety cover is put over the top and the current is switched on The dye will migrate through the gel toward the positive electrode, as will the DNA Depending on how much voltage is applied and how warm the gel is and size and shape of molecules will depend on how fast the mols move through the gel Smaller fragments will move easier so they will be closer to the positive electrode Once the dye has moved through the gel to the buffer, the electrical current is switched off and gel is removed from the tray
Page 249: Instrumental analysis in research

DNA is stained using ethidium

bromide

Presenter
Presentation Notes
Gel is stained using ethidium bromide which binds to DNA it shows up as bands in UV light Draw attention to the fact that small mols are at the bottom of the gel and large ones stay nearest to the wells
Page 252: Instrumental analysis in research
Page 253: Instrumental analysis in research
Page 254: Instrumental analysis in research
Page 255: Instrumental analysis in research

POINTS TO CONSIDER:‐

Selection of buffer

Temperature

Electro osmosis

U.v

florescent chamber

Page 256: Instrumental analysis in research

FLORIMETRIC ANALYSIS:-•

a large number of substances absorb,transmits,reflects

light-during this

process there is a production of heat of varied quantum( less or more ) which form a new wavelength or radiation called as ELECTROMAGNETIC RADIATION .new light is different from absorbed light which emit longer electromagnetic radiation known as LUMINESCENCE.

IT IS OF TWO TYPES-FLUORASCENCE-PHOSPHOROSCENCE

Page 257: Instrumental analysis in research
Page 258: Instrumental analysis in research

FLUORESCENCE:‐

Means when a beam of light is incident on certain  substance they emit visible light or visible radiation 

and this phenomena is called as fluorescence  substance ,this phenomena is instantaneous and it 

starts immediately after absorption of light and  stops as soon as the incidence light cuts off    

Page 259: Instrumental analysis in research
Page 260: Instrumental analysis in research

PHOSPHOROSCENCE•

Means they emit light continuously even after the incident light is cut off.

in florescence the light emit 10-6

to 10-4 seconds

of absorption whereas in phosphorescence the radiation of light re emit in 10-4

to 20 seconds or

longer than that , therefore it takes much time to re emit radiation even after the incident light is cut off.

The florescent substance is proportional to the concentration of the incident of light , the devise to analyze such substance of intensity of florescence is florimetry.

Page 261: Instrumental analysis in research
Page 262: Instrumental analysis in research

Fluorimetry

is the instrument to measure the  fluorescence of substance nature , routinely used in 

lab to identifie

the drugs of different fluorescence   substance , hormones , certain protines

components of haemostasis

etc.

Types:‐

A) fluorimeter

( filters are used.)

B) spectrofluorimeter.( prisms and gratings are  used.)

Page 263: Instrumental analysis in research

SCHEMATIC DIAGRAM OF FLUORIMETER

Page 264: Instrumental analysis in research

APPLICATIONS:‐

Useful in clinical laboratory to identify specific  proteins and also to determine uranium in salts , 

henceforth used in nuclear research .

It is one of the sensitive analysis of many elements  like aluminum in alloys , boron in steel.vit

B1

and  B2,analysis of food products, qualitative and 

quantitative analysis of aromatic substance.

Page 265: Instrumental analysis in research

LOTS TO READ

Page 266: Instrumental analysis in research
Page 267: Instrumental analysis in research
Page 268: Instrumental analysis in research
Page 269: Instrumental analysis in research
Page 270: Instrumental analysis in research
Page 271: Instrumental analysis in research