INSTRUCTION MANUAL Model 109 Temperature Probe Revision: 4/18 Copyright © 1983-2018 Campbell Scientific, Inc.

Transcript of INSTRUCTION - Campbell Sci › documents › us › manuals › 109.pdfdetermine which Campbell...

Page 1: INSTRUCTION - Campbell Sci › documents › us › manuals › 109.pdfdetermine which Campbell Scientific company serves your country. To obtain a Returned Materials Authorization

INST

RU

CT

ION

MA

NU

AL

Model 109 Temperature Probe Revision: 4/18

C o p y r i g h t © 1 9 8 3 - 2 0 1 8 C a m p b e l l S c i e n t i f i c , I n c .

Page 2: INSTRUCTION - Campbell Sci › documents › us › manuals › 109.pdfdetermine which Campbell Scientific company serves your country. To obtain a Returned Materials Authorization
Page 3: INSTRUCTION - Campbell Sci › documents › us › manuals › 109.pdfdetermine which Campbell Scientific company serves your country. To obtain a Returned Materials Authorization

Limited Warranty “Products manufactured by CSI are warranted by CSI to be free from defects in materials and workmanship under normal use and service for twelve months from the date of shipment unless otherwise specified in the corresponding product manual. (Product manuals are available for review online at www.campbellsci.com.) Products not manufactured by CSI, but that are resold by CSI, are warranted only to the limits extended by the original manufacturer. Batteries, fine-wire thermocouples, desiccant, and other consumables have no warranty. CSI’s obligation under this warranty is limited to repairing or replacing (at CSI’s option) defective Products, which shall be the sole and exclusive remedy under this warranty. The Customer assumes all costs of removing, reinstalling, and shipping defective Products to CSI. CSI will return such Products by surface carrier prepaid within the continental United States of America. To all other locations, CSI will return such Products best way CIP (port of entry) per Incoterms ® 2010. This warranty shall not apply to any Products which have been subjected to modification, misuse, neglect, improper service, accidents of nature, or shipping damage. This warranty is in lieu of all other warranties, expressed or implied. The warranty for installation services performed by CSI such as programming to customer specifications, electrical connections to Products manufactured by CSI, and Product specific training, is part of CSI's product warranty. CSI EXPRESSLY DISCLAIMS AND EXCLUDES ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. CSI hereby disclaims, to the fullest extent allowed by applicable law, any and all warranties and conditions with respect to the Products, whether express, implied or statutory, other than those expressly provided herein.”

Page 4: INSTRUCTION - Campbell Sci › documents › us › manuals › 109.pdfdetermine which Campbell Scientific company serves your country. To obtain a Returned Materials Authorization

Assistance Products may not be returned without prior authorization. The following contact information is for US and international customers residing in countries served by Campbell Scientific, Inc. directly. Affiliate companies handle repairs for customers within their territories. Please visit www.campbellsci.com to determine which Campbell Scientific company serves your country.

To obtain a Returned Materials Authorization (RMA) number, contact CAMPBELL SCIENTIFIC, INC., phone (435) 227-9000. Please write the issued RMA number clearly on the outside of the shipping container. Campbell Scientific’s shipping address is:

CAMPBELL SCIENTIFIC, INC. RMA#_____ 815 West 1800 North Logan, Utah 84321-1784

For all returns, the customer must fill out a “Statement of Product Cleanliness and Decontamination” form and comply with the requirements specified in it. The form is available from our website at www.campbellsci.com/repair. A completed form must be either emailed to [email protected] or faxed to (435) 227-9106. Campbell Scientific is unable to process any returns until we receive this form. If the form is not received within three days of product receipt or is incomplete, the product will be returned to the customer at the customer’s expense. Campbell Scientific reserves the right to refuse service on products that were exposed to contaminants that may cause health or safety concerns for our employees.

Page 5: INSTRUCTION - Campbell Sci › documents › us › manuals › 109.pdfdetermine which Campbell Scientific company serves your country. To obtain a Returned Materials Authorization

Safety DANGER — MANY HAZARDS ARE ASSOCIATED WITH INSTALLING, USING, MAINTAINING, AND WORKING ON OR AROUND TRIPODS, TOWERS, AND ANY ATTACHMENTS TO TRIPODS AND TOWERS SUCH AS SENSORS, CROSSARMS, ENCLOSURES, ANTENNAS, ETC. FAILURE TO PROPERLY AND COMPLETELY ASSEMBLE, INSTALL, OPERATE, USE, AND MAINTAIN TRIPODS, TOWERS, AND ATTACHMENTS, AND FAILURE TO HEED WARNINGS, INCREASES THE RISK OF DEATH, ACCIDENT, SERIOUS INJURY, PROPERTY DAMAGE, AND PRODUCT FAILURE. TAKE ALL REASONABLE PRECAUTIONS TO AVOID THESE HAZARDS. CHECK WITH YOUR ORGANIZATION'S SAFETY COORDINATOR (OR POLICY) FOR PROCEDURES AND REQUIRED PROTECTIVE EQUIPMENT PRIOR TO PERFORMING ANY WORK.

Use tripods, towers, and attachments to tripods and towers only for purposes for which they are designed. Do not exceed design limits. Be familiar and comply with all instructions provided in product manuals. Manuals are available at www.campbellsci.com or by telephoning (435) 227-9000 (USA). You are responsible for conformance with governing codes and regulations, including safety regulations, and the integrity and location of structures or land to which towers, tripods, and any attachments are attached. Installation sites should be evaluated and approved by a qualified engineer. If questions or concerns arise regarding installation, use, or maintenance of tripods, towers, attachments, or electrical connections, consult with a licensed and qualified engineer or electrician.

General • Prior to performing site or installation work, obtain required approvals and permits. Comply

with all governing structure-height regulations, such as those of the FAA in the USA. • Use only qualified personnel for installation, use, and maintenance of tripods and towers, and

any attachments to tripods and towers. The use of licensed and qualified contractors is highly recommended.

• Read all applicable instructions carefully and understand procedures thoroughly before beginning work.

• Wear a hardhat and eye protection, and take other appropriate safety precautions while working on or around tripods and towers.

• Do not climb tripods or towers at any time, and prohibit climbing by other persons. Take reasonable precautions to secure tripod and tower sites from trespassers.

• Use only manufacturer recommended parts, materials, and tools.

Utility and Electrical • You can be killed or sustain serious bodily injury if the tripod, tower, or attachments you are

installing, constructing, using, or maintaining, or a tool, stake, or anchor, come in contact with overhead or underground utility lines.

• Maintain a distance of at least one-and-one-half times structure height, 20 feet, or the distance required by applicable law, whichever is greater, between overhead utility lines and the structure (tripod, tower, attachments, or tools).

• Prior to performing site or installation work, inform all utility companies and have all underground utilities marked.

• Comply with all electrical codes. Electrical equipment and related grounding devices should be installed by a licensed and qualified electrician.

Elevated Work and Weather • Exercise extreme caution when performing elevated work. • Use appropriate equipment and safety practices. • During installation and maintenance, keep tower and tripod sites clear of un-trained or non-

essential personnel. Take precautions to prevent elevated tools and objects from dropping. • Do not perform any work in inclement weather, including wind, rain, snow, lightning, etc.

Maintenance • Periodically (at least yearly) check for wear and damage, including corrosion, stress cracks,

frayed cables, loose cable clamps, cable tightness, etc. and take necessary corrective actions. • Periodically (at least yearly) check electrical ground connections.

WHILE EVERY ATTEMPT IS MADE TO EMBODY THE HIGHEST DEGREE OF SAFETY IN ALL CAMPBELL SCIENTIFIC PRODUCTS, THE CUSTOMER ASSUMES ALL RISK FROM ANY INJURY RESULTING FROM IMPROPER INSTALLATION, USE, OR MAINTENANCE OF TRIPODS, TOWERS, OR ATTACHMENTS TO TRIPODS AND TOWERS SUCH AS SENSORS, CROSSARMS, ENCLOSURES, ANTENNAS, ETC.

Page 6: INSTRUCTION - Campbell Sci › documents › us › manuals › 109.pdfdetermine which Campbell Scientific company serves your country. To obtain a Returned Materials Authorization
Page 7: INSTRUCTION - Campbell Sci › documents › us › manuals › 109.pdfdetermine which Campbell Scientific company serves your country. To obtain a Returned Materials Authorization

i

Table of Contents PDF viewers: These page numbers refer to the printed version of this document. Use the PDF reader bookmarks tab for links to specific sections.

1. Introduction ................................................................ 1

2. Precautions ................................................................ 1

3. Initial Inspection ......................................................... 1

4. QuickStart ................................................................... 1

5. Overview ..................................................................... 3

6. Specifications ............................................................. 4

7. Installation .................................................................. 4

7.1 Wiring to Datalogger ........................................................................... 5 7.2 Datalogger Programming ..................................................................... 5

7.2.1 Therm109() Instruction ................................................................. 5 7.3 Air Temperature Installation ................................................................ 6 7.4 Water Temperature Installation ............................................................ 8 7.5 Soil Temperature Installation ............................................................... 8

8. Operation .................................................................... 8

8.1 Sensor Schematic ................................................................................. 8 8.2 Measurement and Output Linearization ............................................... 8 8.3 Electrically Noisy Environments ......................................................... 9 8.4 Long Cable Lengths ............................................................................. 9

9. Troubleshooting and Maintenance ......................... 10

9.1 Troubleshooting ................................................................................. 10 9.2 Maintenance ....................................................................................... 10 9.3 Calibration.......................................................................................... 11

10. Attributions and References ................................... 11

Appendices

A. Importing Short Cut Code Into CRBasic Editor ... A-1

B. Example Programs ................................................. B-1

C. Thermistor Resistance and Temperature ............. C-1

Page 8: INSTRUCTION - Campbell Sci › documents › us › manuals › 109.pdfdetermine which Campbell Scientific company serves your country. To obtain a Returned Materials Authorization

Table of Contents

ii

Figures 7-1. Installing a 109 and 41303-5A Radiation Shield on a

CM200-Series Crossarm .................................................................. 7 7-2. 109 and 41303-5A Radiation Shield on a tripod mast (left). 109

and RAD06 Radiation Shield on a CM200-Series Crossarm (right). .............................................................................................. 7

8-1. 109 Thermistor Probe Schematic ........................................................ 8

Tables 7-1. Wire Color, Function, and Datalogger Connection ............................. 5 C-1. 109 Thermistor Resistance and Temperature .................................. C-1

CRBasic Examples B-1. 109 Program Example for CR300-Series, CR800-Series,

CR1000, CR1000X-Series, CR3000, and CR5000 Dataloggers .................................................................................. B-1

B-2. 109 Program Example for CR6 Dataloggers ................................... B-1 B-3. 109 Program Example for CR200(X)-Series Dataloggers .............. B-2

Page 9: INSTRUCTION - Campbell Sci › documents › us › manuals › 109.pdfdetermine which Campbell Scientific company serves your country. To obtain a Returned Materials Authorization

1

Model 109 Temperature Probe 1. Introduction

The 109 Temperature Probe uses a thermistor to measure temperature in air, soil, and water. It is compatible with all CRBasic dataloggers except the CR9000(X). See Section 6, Specifications (p. 4), for a list of compatible CRBasic dataloggers.

For Edlog datalogger support, check the availability of an older manual at www.campbellsci.com/old-manuals, or contact Campbell Scientific for assistance.

2. Precautions READ AND UNDERSTAND the Safety section at the front of this manual.

Santoprene® rubber, which composes the black outer jacket of the 109 cable, will support combustion in air. It is used because of its resistance to temperature extremes, moisture, and UV degradation. It is rated as slow burning when tested according to U.L. 94 H.B. and passes FMVSS302. However, local fire codes may preclude its use inside buildings.

3. Initial Inspection Check the packaging and contents of the shipment. If damage occurred during transport, immediately file a claim with the carrier. Contact Campbell Scientific to facilitate repair or replacement.

Check model information against the shipping documents to ensure the expected products and the correct lengths of cable are received. Model numbers are found on each product. On cables and cabled items, the model number is usually found at the connection end of the cable. Report any shortages immediately to Campbell Scientific.

4. QuickStart A video that describes programming using Short Cut is available at: www.campbellsci.com/videos/cr1000x-datalogger-getting-started-program-part-3. Short Cut is an easy way to program your datalogger to measure the 109 probe and assign datalogger wiring terminals. Short Cut is available as a download on www.campbellsci.com. It is included in installations of LoggerNet, PC200W, PC400, or RTDAQ.

The following procedure also describes programming with Short Cut.

1. Open Short Cut and click Create New Program.

2. Double-click the datalogger model.

Page 10: INSTRUCTION - Campbell Sci › documents › us › manuals › 109.pdfdetermine which Campbell Scientific company serves your country. To obtain a Returned Materials Authorization

Model 109 Temperature Probe

2

3. In the search box under the Available Sensors and Devices heading, type in 109 or find the 109 in the Sensors | Temperature folder. Double-click the 109 Temperature Probe. Data defaults to degree Celsius. This can be changed by clicking the Deg C box and selecting Deg F, for degrees Fahrenheit, or K, for Kelvin. After entering the Properties, click on the Wiring tab to see how the sensor is to be wired to the datalogger.

4. Select any other sensors you have, and then finish the remaining Short Cut steps to complete the program.

Page 11: INSTRUCTION - Campbell Sci › documents › us › manuals › 109.pdfdetermine which Campbell Scientific company serves your country. To obtain a Returned Materials Authorization

Model 109 Temperature Probe

3

5. In Output Setup, enter the scan rate and Data Output Storage Interval.

6. Select the output option.

7. If LoggerNet, PC400, RTDAQ, or PC200W is running on your computer, and the computer to datalogger connection is active, you can click Finish in Short Cut and you will be prompted to send the program just created to the datalogger.

8. If the sensor is connected to the datalogger, as shown in the wiring diagram, check the output of the sensor in the datalogger support software data display to make sure it is making reasonable measurements.

5. Overview The 109 is a rugged probe that accurately measures air, soil, or water temperature in a variety of applications. The sensor consists of a thermistor encapsulated in an epoxy-filled aluminum housing. This design allows the probe to be buried or submerged in water to 15 m (50 ft) or 21 psi. When measuring air temperature, a 41303-5A radiation shield is normally used to

Page 12: INSTRUCTION - Campbell Sci › documents › us › manuals › 109.pdfdetermine which Campbell Scientific company serves your country. To obtain a Returned Materials Authorization

Model 109 Temperature Probe

4

mount the 109 and limit solar radiation loading. See Specifications for a complete list of compatible dataloggers.

6. Specifications Features:

• Measures air, soil, or water temperature • Compatible with AM16/32-series multiplexers • Easy to install or remove • Durable • Compatible with the following CRBasic dataloggers:

CR200(X) series, CR300 series, CR6 series, CR800 series, CR1000, CR1000X series, CR3000, and CR5000

Sensor Element: Measurement Specialties™ 10K3A1iA thermistor

Survival Range: –50 to 100 °C Measurement Range: –50 to 70 °C

Time Constant in Air: 30 to 60 s in a wind speed of 5 m/s

Maximum Cable Length: 1000 ft

Accuracy1 Worst case: ±0.60 °C (–50 to 70 °C) ±0.25 °C (–10 to 70 °C)

Interchangeability Error: ±0.10 °C (0 to 70 °C) ±0.13 °C at –10 °C ±0.15 °C at –20 °C ±0.18 °C at –30 °C ±0.20 °C at –40 °C ±0.50 °C at –50 °C

Steinhart-Hart Linearization Error: ≤ 0.03 °C (–50 to 70 °C)

Probe Weight and Dimensions Weight with 10 ft cable: 136 g (5 oz) Length: 10.4 cm (4.1 in) Diameter: 0.762 cm (0.3 in)

1Overall probe accuracy is a combination of thermistor interchangeability, bridge-resistor accuracy, and error of the Steinhart-Hart equation. Interchangeability is the principle component error. If needed, an estimate of the interchangeability error for 0 to 50 °C, that can be used as the Offset parameter of the Therm109() instruction, can be determined with a 1-point or 2-point calibration.

7. Installation If you are programming your datalogger with Short Cut, skip Section 7.1, Wiring to Datalogger (p. 5), and Section 7.2, Datalogger Programming (p. 5). Short Cut does this work for you. See Section 4, QuickStart (p. 1), for a Short Cut tutorial.

Page 13: INSTRUCTION - Campbell Sci › documents › us › manuals › 109.pdfdetermine which Campbell Scientific company serves your country. To obtain a Returned Materials Authorization

Model 109 Temperature Probe

5

7.1 Wiring to Datalogger TABLE 7-1. Wire Color, Function, and Datalogger Connection

Wire Color Wire Function Datalogger Connection Terminal

Black Voltage-excitation input U configured for voltage

excitation1, EX, VX (voltage excitation)

Red Analog-voltage output

U configured for single-ended analog input1, SE

(single-ended, analog-voltage input)

Purple Bridge-resistor lead (analog ground)

Clear EMF shield (analog ground)

1U channels are automatically configured by the measurement instruction.

7.2 Datalogger Programming Short Cut is the best source for up-to-date datalogger programming code.

If your data acquisition requirements are simple, you can probably create and maintain a datalogger program exclusively with Short Cut. If your data acquisition needs are more complex, the files that Short Cut creates are a great source for programming code to start a new program or add to an existing custom program.

Short Cut cannot edit programs after they are imported and edited in CRBasic Editor.

A Short Cut tutorial is available in Section 4, QuickStart (p. 1). If you wish to import Short Cut code into CRBasic Editor to create or add to a customized program, follow the procedure in Appendix A, Importing Short Cut Code Into CRBasic Editor (p. A-1). Programming basics are provided in the following section. A complete program example can be found in Appendix B, Example Programs (p. B-1).

If the 109 probe is to be used with long cable lengths or in electrically noisy environments, consider employing the measurement programming techniques outlined in Section 8.3, Electrically Noisy Environments (p. 9), and Section 8.4, Long Cable Lengths (p. 9).

Details of 109 probe measurement and linearization of the thermistor output are provided in Section 8.2, Measurement and Output Linearization (p. 8).

7.2.1 Therm109() Instruction The Therm109() measurement instruction programs CRBasic dataloggers to measure the 109 probe. It applies a precise excitation voltage, makes a half-bridge resistance measurement, and converts the result to temperature using the

NOTE

Page 14: INSTRUCTION - Campbell Sci › documents › us › manuals › 109.pdfdetermine which Campbell Scientific company serves your country. To obtain a Returned Materials Authorization

Model 109 Temperature Probe

6

Steinhart-Hart equation. See Section 8.2, Measurement and Output Linearization (p. 8), for more information. Therm109() instruction and parameters are as follows:

Therm109(Dest, Reps, SEChan, VxChan, SettlingTime, Integ/Fnotch, Mult, Offset)

The instruction for CR200(X)-series dataloggers excludes the SettlingTime and Integ parameters.

Variations:

• Temperature reported as °C — set Mult to 1 and Offset to 0 • Temperature reported as °F — set Mult to 1.8 and Offset to 32 • AC mains noise filtering — set Integ/Fnotch to the 60 Hz or 50 Hz

option (see Section 8.3, Electrically Noisy Environments (p. 9)) • Compensate for long cable lengths — Set SettlingTime to 20000 (see

Section 8.4, Long Cable Lengths (p. 9))

7.3 Air Temperature Installation For air temperature measurements, locate probes over an open, level area at least 9 m (EPA; Section 10, Attributions and References (p. 11)) in diameter. The surface should be covered by short grass or the natural earth surface where grass does not grow. Probes should be located at a distance of at least four times the height of any nearby obstruction, and at least 30 m (EPA) from large paved areas. Sensors should be protected from thermal radiation, and adequately ventilated.

Standard air temperature measurement heights:

• 1.25 to 2.0 m (WMO; Section 10, Attributions and References (p. 11)) • 2.0 m (EPA; Section 10, Attributions and References (p. 11)) • 2.0 m and 10.0 m for temperature difference (EPA)

When exposed to sunlight, the 109 should be housed in a six-plate solar radiation shield. Six-plate shields offered by Campbell Scientific are models 41303-5A, 41303-5B, or RAD06.

The white color of these shields reflects solar radiation, and the louvered construction allows air to pass freely through, thereby keeping the probe at or near ambient temperature. The RAD06 uses a double-louvered design that offers improved sensor protection from insect intrusion and driving rain and snow. In addition, the RAD06 shield has lower self-heating in bright sunlight combined with higher temperatures (> 24 °C (75 °F)) and low wind speeds (< 2 m/s (4.5 mph)), giving a better measurement.

The 41303-5A and RAD06 attach to a crossarm, mast, or user-supplied pipe with a 2.5 to 5.3 cm (1.0 to 2.1 in) outer diameter. The 41303-5B attaches to a CM500-series pole or a user-supplied pole with a 5.1 cm (2.4 in) outer diameter.

Page 15: INSTRUCTION - Campbell Sci › documents › us › manuals › 109.pdfdetermine which Campbell Scientific company serves your country. To obtain a Returned Materials Authorization

Model 109 Temperature Probe

7

Tools required for installing a radiation shield to a tripod or tower include:

• 1/2-inch open end wrench • UV resistant cable ties • adjustable wrench with a minimum 1-1/2 inch jaw size

FIGURE 7-1. Installing a 109 and 41303-5A Radiation Shield on a CM200-Series Crossarm

FIGURE 7-2. 109 and 41303-5A Radiation Shield on a tripod mast (left). 109 and RAD06 Radiation Shield on a CM200-Series Crossarm (right).

The 109 is held in the 41303-5A radiation shield by a mounting clamp at the bottom (FIGURE 7-1 and FIGURE 7-2 left). Loosen the mounting clamp screws, and insert the probe through the clamp. Tighten the screws to secure the sensor, and route the sensor cable to the instrument enclosure.

Page 16: INSTRUCTION - Campbell Sci › documents › us › manuals › 109.pdfdetermine which Campbell Scientific company serves your country. To obtain a Returned Materials Authorization

Model 109 Temperature Probe

8

The 109 is held in the RAD06 radiation shield by inserting the sensor through the sensor gland at the bottom of the shield (FIGURE 7-2 right). Loosen the nut on the gland, and insert the probe into the shield. Tighten the nut on the sensor gland using an adjustable wrench until the sensor is securely held in place. Route the sensor cable to the instrument enclosure.

Secure the cable to the tripod or tower using cable ties.

7.4 Water Temperature Installation 109 probes can be submerged to 15 m (50 ft) or 21 psi. The 109 is not weighted, so a weighting system should be added, or the probe secured to a submerged object such as a piling.

7.5 Soil Temperature Installation The 109 tends to measure the average temperature over its length, so burying the probe such that the measurement tip is horizontal to the soil surface at the desired depth is usually preferred. The maximum burial depth for soil that could become saturated with water is dictated by the maximum water pressure allowed for the sensor, which is 21 psi.

One or two coils of cable should also be buried in a shallow installation. Burial of some cable mitigates the effect of solar heating of the above ground cable on the temperature measurement.

Placement of the cable inside a rugged conduit may be necessary for long cable runs, especially in locations subject to digging, mowing, traffic, use of power tools, or lightning strikes.

8. Operation 8.1 Sensor Schematic

FIGURE 8-1. 109 Thermistor Probe Schematic

8.2 Measurement and Output Linearization CRBasic instruction Therm109() measures the 109 probe thermistor and automatically converts the result to temperature. With reference to the previous

Black

Red

Purple

Clear

Vx

Vs

— Analog Ground

Thermistor

24.9 kΩ, 0.1% 10 ppm

Page 17: INSTRUCTION - Campbell Sci › documents › us › manuals › 109.pdfdetermine which Campbell Scientific company serves your country. To obtain a Returned Materials Authorization

Model 109 Temperature Probe

9

FIGURE 8-1, 109 Thermistor Probe Schematic, Therm109() applies 2500 mV excitation at the Vx line and measures the voltage drop across the 24.9 kΩ resistor at the Vs line.

The ratio of measured voltage (Vs) to excitation voltage (Vx) is related to thermistor resistance (Rs) and the 24.9 kΩ bridge resistor as described in the following equations:

Vs/Vx = 24900 Ω / (Rs + 24900 Ω)

Solving for Rs:

Rs + 24900 Ω = 24900 Ω • (Vx/Vs)

Rs = 24900 Ω • ((Vx/Vs) – 1)

The relationship of Rs to temperature is tabulated in Appendix C, Thermistor Resistance and Temperature (p. C-1), but is calculated by Therm109() using the Steinhart-Hart equation, described as follows:

Tc = (1 / (A + B • ln (Rs) + C • (ln (Rs))3)) – 273.15

where:

Tc = temperature in degrees Celsius (°C)

A1 = 1.129241E–3

B1 = 2.341077E–4

C1 = 8.775468E–8 1Coefficients provided by Measurement Specialties™.

8.3 Electrically Noisy Environments EMF noise emanating from the ac mains power grid can be a significant source of measurement error. 60 Hz noise is common in the United States. 50 Hz noise is common in Europe and other regions. This noise can usually be filtered out.

The following code snip examples filter 60 Hz noise.

CR6-series datalogger example:

Therm109(T109_C,1,U1,U10,20000,60,1.0,0.0)

CR800-series and CR3000 dataloggers example:

Therm109(T109_C,1,1,1,20000,_60Hz,1.0,0.0)

An integration parameter is not available for CR200(X)-series dataloggers.

8.4 Long Cable Lengths Long cable lengths (>50 ft) may require longer than normal analog measurement settling times. Settling times are increased by adding a measurement delay to a datalogger program.

The 60 Hz and 50 Hz integration options include a 3 ms settling time; longer settling times can be entered into the Settling Time parameter. Campbell

Page 18: INSTRUCTION - Campbell Sci › documents › us › manuals › 109.pdfdetermine which Campbell Scientific company serves your country. To obtain a Returned Materials Authorization

Model 109 Temperature Probe

10

Scientific suggests doubling the settling time every 50 ft. The following code snip examples increase settling time by 20000 µs by placing 20000 as the argument in the SettlingTime parameter:

CR6-series datalogger example:

Therm109(T109_C,1,U1,U10,20000,60,1.0,0.0)

CR800-series and CR3000 dataloggers example:

Therm109(T109_C,1,1,1,20000,_60Hz,1.0,0.0)

A setting time parameter is not available for CR200(X)-series dataloggers.

9. Troubleshooting and Maintenance All factory repairs and recalibrations require a returned material authorization (RMA) and completion of the “Declaration of Hazardous Material and Decontamination” form. Refer to the Assistance page at the beginning of this manual for more information.

9.1 Troubleshooting Symptom: Temperature is reported as NAN, –INF, or incorrect temperature.

Verify wire leads are connected to the terminals specified in the Therm109() instruction: red to single-ended analog input (SE or U), black to switched excitation (VX/EX or U), and purple to ground ().

Symptom: Incorrect temperature is reported.

Verify the Mult and Offset arguments in Therm109() are correct for the desired units (Section 7.2, Datalogger Programming (p. 5)). Check the cable for signs of damage and possible moisture intrusion.

Symptom: Unstable temperature is reported.

Probably a result of electromagnetic interference. Try using the 50 Hz or 60 Hz options for the Fnotch parameter, and/or increasing the settling time as described in Section 8.3, Electrically Noisy Environments (p. 9), and Section 8.4, Long Cable Lengths (p. 9). Ensure the clear wire is connected to datalogger ground, and the datalogger is properly grounded.

9.2 Maintenance The 109 probe requires minimal maintenance. For air temperature measurements, check the radiation shield monthly to make sure it is clean and free from debris. Periodically check cabling for signs of damage and possible moisture intrusion.

NOTE

Page 19: INSTRUCTION - Campbell Sci › documents › us › manuals › 109.pdfdetermine which Campbell Scientific company serves your country. To obtain a Returned Materials Authorization

Model 109 Temperature Probe

11

9.3 Calibration If needed, an estimate of the interchangeability error for 0 to 50 °C, that can be used as the Offset parameter of the Therm109() instruction, can be determined with a 1-point or 2-point calibration. Calibration of the 109 probe is not necessary unless the accuracy needed in the sensor data requires correction of the thermistor interchangeability offset described in Section 6, Specifications (p. 4).

10. Attributions and References Santoprene® is a registered trademark of Exxon Mobile Corporation.

Measurement Specialties™ is a trademarked global designer and manufacturer of sensors and sensor-based systems.

EPA installation standard: Quality Assurance Handbook for Air Pollution Measurement Systems – Volume IV: Meteorological Measurements Version 2.0

WMO standard: WMO No. 8, Seventh edition, 6 Aug 2008 Guide to Meteorological Instruments and Methods of Observation

Page 20: INSTRUCTION - Campbell Sci › documents › us › manuals › 109.pdfdetermine which Campbell Scientific company serves your country. To obtain a Returned Materials Authorization
Page 21: INSTRUCTION - Campbell Sci › documents › us › manuals › 109.pdfdetermine which Campbell Scientific company serves your country. To obtain a Returned Materials Authorization

A-1

Appendix A. Importing Short Cut Code Into CRBasic Editor

This tutorial shows:

• How to import a Short Cut program into a program editor for additional refinement

• How to import a wiring diagram from Short Cut into the comments of a custom program

Short Cut creates files, which can be imported into CRBasic Editor. Assuming defaults were used when Short Cut was installed, these files reside in the C:\campbellsci\SCWin folder:

• .DEF (wiring and memory usage information) • .CR2 (CR200(X)-series datalogger code) • .CR300 (CR300-series datalogger code) • .CR6 (CR6-series datalogger code) • .CR8 (CR800-series datalogger code) • .CR1 (CR1000 datalogger code) • .CR1X (CR1000X-series datalogger code) • .CR3 (CR3000 datalogger code) • .CR5 (CR5000 datalogger code)

Use the following procedure to import Short Cut code and wiring diagram into CRBasic Editor.

1. Create the Short Cut program following the procedure in Section 4, QuickStart (p. 1). Finish the program. On the Advanced tab, click the CRBasic Editor button. The program opens in CRBasic with the name noname.CR_. Now save the program with your desired name in any folder.

Once the file is edited with CRBasic Editor, Short Cut can no longer be used to edit the datalogger program. Change the name of the program file or move it, or Short Cut may overwrite it next time it is used.

2. The program can now be edited, saved, and sent to the datalogger.

3. Import wiring information to the program by opening the associated .DEF file. By default, it will be in the c:\campbellsci\SCWin folder. Copy and paste the section beginning with heading “–Wiring for CRXXX–” into the CRBasic program, usually at the head of the file. After pasting, edit the information such that an apostrophe (') begins each line. This character instructs the datalogger compiler to ignore the line when compiling. You can highlight several lines of CRBasic code then right-click and select Comment Block. (This feature is demonstrated at about 5:10 in the CRBasic | Features video.)

NOTE

Page 22: INSTRUCTION - Campbell Sci › documents › us › manuals › 109.pdfdetermine which Campbell Scientific company serves your country. To obtain a Returned Materials Authorization
Page 23: INSTRUCTION - Campbell Sci › documents › us › manuals › 109.pdfdetermine which Campbell Scientific company serves your country. To obtain a Returned Materials Authorization

B-1

Appendix B. Example Programs This following example can be used directly with most of our CRBasic dataloggers.

CRBasic Example B-1. 109 Program Example for CR300-Series, CR800-Series, CR1000, CR1000X-Series, CR3000, and CR5000 Dataloggers

'Program measures one 109 temperature probe once a second and 'stores the average temperature every 60 minutes. 'Wiring Diagram '============== ' 109 ' Probe ' Lead Datalogger ' Color Function Terminal ' ----- -------- ------ ' Black Voltage-excitation input VX1 or EX1 ' Red Analog-voltage output SE1 ' Purple Bridge-resistor ground Ground Symbol ' Clear Shield Ground Symbol 'Declare the variables for the temperature measurement Public T109_C 'Define a data table for 60 minute averages: DataTable(Hourly,True,-1) DataInterval(0,60,Min,0) Average(1,T109_C,IEEE4,0) EndTable BeginProg Scan(1,Sec,1,0) 'Measure the temperature Therm109(T109_C,1,1,Vx1,0,_60Hz,1.0,0.0) 'Call Data Table CallTable(Hourly) NextScan EndProg

The following example can be used directly with CR6-series dataloggers.

CRBasic Example B-2. 109 Program Example for CR6 Dataloggers

'Program measures one 109 temperature probe once a second and 'stores the average temperature every 60 minutes. 'Wiring Diagram '============== ' 109 ' Probe ' Lead CR6 ' Color Function Terminal ' ----- -------- ------ ' Black Voltage-excitation input U10 ' Red Analog-voltage output U1 ' Purple Bridge-resistor ground Ground Symbol ' Clear Shield Ground Symbol 'Declare the variables for the temperature measurement Public T109_C

Page 24: INSTRUCTION - Campbell Sci › documents › us › manuals › 109.pdfdetermine which Campbell Scientific company serves your country. To obtain a Returned Materials Authorization

Appendix B. Example Programs

B-2

'Define a data table for 60 minute averages: DataTable(Hourly,True,-1) DataInterval(0,60,Min,0) Average(1,T109_C,IEEE4,0) EndTable BeginProg Scan(1,Sec,1,0) 'Measure the temperature Therm109(T109_C,1,U1,U10,0,60,1.0,0.0) 'Call Data Table CallTable(Hourly) NextScan EndProg

The following example can be used directly with CR200(X)-series dataloggers.

CRBasic Example B-3. 109 Program Example for CR200(X)-Series Dataloggers

'Program measures one 109 temperature probe once a second and 'stores the average temperature every 60 minutes. 'Wiring Diagram '============== ' 109 ' Probe ' Lead CR200(X) ' Color Function Terminal ' ----- -------- ------ ' Black Voltage-excitation input VX1/EX1 ' Red Analog-voltage output SE1 ' Purple Bridge-resistor ground Ground Symbol ' Clear Shield Ground Symbol 'Declare the variable for the temperature measurement Public T109_C 'Define a data table for 60 minute averages DataTable(Hourly,True,-1) DataInterval(0,60,min) Average(1,T109_C,False) EndTable BeginProg Scan(1,sec) 'Measure the temperature Therm109(T109_C,1,1,Ex1,1.0,0) 'Call Data Table CallTable Hourly NextScan EndProg

Page 25: INSTRUCTION - Campbell Sci › documents › us › manuals › 109.pdfdetermine which Campbell Scientific company serves your country. To obtain a Returned Materials Authorization

Appendix C. Thermistor Resistance and Temperature

TABLE C-1. 109 Thermistor Resistance and Temperature1

Actual Temperature (°C)

10K3A1iA Thermistor

Resistance (Ω)

CRBasic Therm109() Output (°C)

-40 336098 -40.00 -39 314553 -39.00 -38 294524 -38.00 -37 275897 -37.00 -36 258563 -36.00 -35 242427 -35.00 -34 227398 -34.00 -33 213394 -33.00 -32 200339 -32.00 -31 188163 -31.00 -30 176803 -30.00 -29 166198 -29.00 -28 156294 -28.00 -27 147042 -27.00 -26 138393 -26.00 -25 130306 -25.00 -24 122741 -24.00 -23 115661 -23.00 -22 109032 -22.00 -21 102824 -21.00 -20 97006 -20.00 -19 91553 -19.00 -18 86439 -18.00 -17 81641 -17.00 -16 77138 -16.00 -15 72911 -15.00 -14 68940 -14.00 -13 65209 -13.00 -12 61703 -12.00 -11 58405 -11.00 -10 55304 -10.00 -9 52385 -9.00 -8 49638 -8.00 -7 47050 -7.00 -6 44613 -6.00 -5 42317 -5.00 -4 40151 -4.00 -3 38110 -3.00 -2 36184 -2.00 -1 34366 -1.00 0 32651 0.00 1 31031 1.00 2 29500 2.00 3 28054 3.00

Page 26: INSTRUCTION - Campbell Sci › documents › us › manuals › 109.pdfdetermine which Campbell Scientific company serves your country. To obtain a Returned Materials Authorization

Appendix B. Example Programs

C-2

4 26687 4.00 5 25395 5.00 6 24172 6.00 7 23016 7.00 8 21921 8.00 9 20885 9.00

10 19903 10.00 11 18973 11.00 12 18092 12.00 13 17257 13.00 14 16465 14.00 15 15714 15.00 16 15001 16.00 17 14324 17.00 18 13682 18.00 19 13073 19.00 20 12493 20.00 21 11943 21.00 22 11420 22.00 23 10923 23.00 24 10450 24.00 25 10000 25.00 26 9572 26.00 27 9165 27.00 28 8777 28.00 29 8408 29.00 30 8056 30.00 31 7721 31.00 32 7402 32.00 33 7097 33.00 34 6807 34.00 35 6530 35.00 36 6266 36.00 37 6014 37.00 38 5774 38.00 39 5544 39.00 40 5325 40.00 41 5116 41.00 42 4916 42.00 43 4724 43.00 44 4542 44.00 45 4367 45.00 46 4200 46.00 47 4040 47.00 48 3887 48.00 49 3741 49.00 50 3601 50.00 51 3467 51.00 52 3339 52.00 53 3216 53.00 54 3098 54.00 55 2985 55.00 56 2877 56.00 57 2773 57.00 58 2674 58.00 59 2579 59.00

Page 27: INSTRUCTION - Campbell Sci › documents › us › manuals › 109.pdfdetermine which Campbell Scientific company serves your country. To obtain a Returned Materials Authorization

Appendix C. Conversion of Thermistor Resistance or Voltage Ratio to Temperature

C-3

60 2487 60.00 61 2399 61.00 62 2315 62.00 63 2234 63.01 64 2157 64.00 65 2082 65.00 66 2011 66.00 67 1942 67.00 68 1876 68.00 69 1813 68.99 70 1752 69.99 71 1693 71.00 72 1637 71.99 73 1582 73.01 74 1530 74.00 75 1480 75.00

1Data from Measurement Specialties™

Page 28: INSTRUCTION - Campbell Sci › documents › us › manuals › 109.pdfdetermine which Campbell Scientific company serves your country. To obtain a Returned Materials Authorization
Page 29: INSTRUCTION - Campbell Sci › documents › us › manuals › 109.pdfdetermine which Campbell Scientific company serves your country. To obtain a Returned Materials Authorization
Page 30: INSTRUCTION - Campbell Sci › documents › us › manuals › 109.pdfdetermine which Campbell Scientific company serves your country. To obtain a Returned Materials Authorization

Campbell Scientific Worldwide Offices

Australia Location: Garbutt, QLD Australia Email: [email protected]

Website: www.campbellsci.com.au

Germany Location: Bremen, Germany Email: [email protected]

Website: www.campbellsci.de

Brazil Location: São Paulo, SP Brazil

Email: [email protected] Website: www.campbellsci.com.br

South Africa Location: Stellenbosch, South Africa

Email: [email protected] Website: www.campbellscientific.co.za

Canada Location: Edmonton, AB Canada

Email: [email protected] Website: www.campbellsci.ca

Southeast Asia Location: Bangkok, Thailand Email: [email protected]

Website: www.campbellsci.asia

China Location: Beijing, P. R. China

Email: [email protected] Website: www.campbellsci.com.cn

Spain Location: Barcelona, Spain Email: [email protected]

Website: www.campbellsci.es

Costa Rica Location: San José, Costa Rica

Email: [email protected] Website: www.campbellsci.cc

UK Location: Shepshed, Loughborough, UK

Email: [email protected] Website: www.campbellsci.co.uk

France Location: Antony, France

Email: [email protected] Website: www.campbellsci.fr

USA Location: Logan, UT USA

Email: [email protected] Website: www.campbellsci.com

Please visit www.campbellsci.com/contact to obtain contact information for your local US or international representative.