GE Healthcare Life Sciences IN Cell Analyzer 2000 Instrument Version 1.0, Build 3.0.0.43 Software Upgrade Instructions March 30, 2012

Transcript of IN Cell Analyzer 2000 Instrument - GE...

Page 1: IN Cell Analyzer 2000 Instrument - GE Healthcareincelldownload.gehealthcare.com/bin/download_data... · 2012-04-06 · IN Cell Analyzer 2000 instrument will automatically re-initialize.

GE Healthcare Life Sciences

IN Cell Analyzer 2000 Instrument Version 1.0, Build 3.0.0.43

Software Upgrade Instructions March 30, 2012

Page 2: IN Cell Analyzer 2000 Instrument - GE Healthcareincelldownload.gehealthcare.com/bin/download_data... · 2012-04-06 · IN Cell Analyzer 2000 instrument will automatically re-initialize.

Release Notes

2

Upgrade instructions for SW Build 3.0.0.43* * For updating from 3.0.0.5 or higher.

Customers with software builds prior to 3.0.0.5 must contact Service for the upgrade.

These instructions apply to upgrading to build 3.0.0.43 on Windows XP workstations ONLY.

Prior to performing the software upgrade Several objective lens parameter settings need to be copied from your current software version to version 3.0.0.43. Follow the steps below to copy the configuration file containing these parameters:

1. Right click on the Windows Start button and select ‘Explore’

2. Navigate to C:\Program Files\GE Healthcare\IN Cell Analyzer 2000\config

3. In this folder is a file named config.xml

4. Right-click on the file and select Rename and change the file name to IN Cell Analyzer 2000.xml

Installing IN Cell Analyzer 2000 SW Build 3.0.0.43

Access to the software upgrade will be either on a CD or through a web link provided by GE Healthcare.

1. Locate the file install_incell_analyzer_2000_3.0.0.43.exe.

2. Save or download the file to the instrument workstation Desktop.

3. Double-click on the installation file icon. The installation wizard will begin.

4. The Introduction window will display the steps to occur during installation. Click Next to

continue.

Page 3: IN Cell Analyzer 2000 Instrument - GE Healthcareincelldownload.gehealthcare.com/bin/download_data... · 2012-04-06 · IN Cell Analyzer 2000 instrument will automatically re-initialize.

Release Notes

3

5. The Uninstall Previous Version pop up window will open. Click Continue Anyway to continue.

6. The License Agreement window displays. Review the agreement and mark the I accept the

terms of the License Agreement radio button. Click Next to continue.

Selecting I do NOT accept the terms of the License Agreement will abort the installation process.

7. On the Choose Install Folder window accept the default location. Click Next to continue.

Page 4: IN Cell Analyzer 2000 Instrument - GE Healthcareincelldownload.gehealthcare.com/bin/download_data... · 2012-04-06 · IN Cell Analyzer 2000 instrument will automatically re-initialize.

Release Notes

4

8. The Pre-Installation Summary window displays the parameters for the install. Click Install to continue.

A progess bar displays during the installation process. An Overwrite Existing File pop up window will display. Click No to continue.

9. On successful installation, the Install Complete window will display. Click Done to continue.

The IN Cell Analyzer 2000 acquisition software Build 3.0.0.43 is now installed.

Page 5: IN Cell Analyzer 2000 Instrument - GE Healthcareincelldownload.gehealthcare.com/bin/download_data... · 2012-04-06 · IN Cell Analyzer 2000 instrument will automatically re-initialize.

Release Notes

5

Installing the prerequisite applications

The upgrade will now continue with the installation of a number of prerequisite applications.

• Microsoft Visual Studio Redistributable.

• QPI SDK v2 Redistributable

• IN Cell Analyzer 2000 Server Installer

The first application to be installed is the Microsoft Visual Studio Redistributable

1. The Visual Studio Redistributable installation wizard begins.

2. At the Welcome window, mark the I have read and accept the license terms checkobx. Click

Install to continue.

Page 6: IN Cell Analyzer 2000 Instrument - GE Healthcareincelldownload.gehealthcare.com/bin/download_data... · 2012-04-06 · IN Cell Analyzer 2000 instrument will automatically re-initialize.

Release Notes

6

3. A progess bar displays during the installation process. On successful installation, the Installation is Complete window will display. Click Finish to continue.

Next, the upgrade will now progress to the installation of QPI SDK v2 Redistributable.

1. At the Welcome window, click Next to continue.

2. At the Program Maintenace window, confirm that the Repair radio button is marked. Click

Next to continue

Page 7: IN Cell Analyzer 2000 Instrument - GE Healthcareincelldownload.gehealthcare.com/bin/download_data... · 2012-04-06 · IN Cell Analyzer 2000 instrument will automatically re-initialize.

Release Notes

7

3. On the Ready to Repair the Program window. click Install to continue.

4. When the update is complete, the Completion window displays. Click Finish to continue

Finally, the upgrade will progress to the IN Cell Analyzer 2000 Server Installer.

1. On the Connect to Scanner window, confirm that the IP address = 159.159.159.1. Click Connect…

Note If you have trouble connecting to the scanner, refer to the Troubleshooting section below.

Page 8: IN Cell Analyzer 2000 Instrument - GE Healthcareincelldownload.gehealthcare.com/bin/download_data... · 2012-04-06 · IN Cell Analyzer 2000 instrument will automatically re-initialize.

Release Notes

8

2. A successful connection message will display, as seen below. Click Next to continue.

3. On the Installing… window, click Install to begin the server installation.

The following messages will disply current status during the install process.

Page 9: IN Cell Analyzer 2000 Instrument - GE Healthcareincelldownload.gehealthcare.com/bin/download_data... · 2012-04-06 · IN Cell Analyzer 2000 instrument will automatically re-initialize.

Release Notes

9

4. On the Install Complete window, click Done to close wizard.

System Initialization After completing the software installation, the IN Cell Analyzer 2000 instrument will automatically re-initialize. During the initialization, the IN Cell Analyzer 2000 runs through diagnostic checks to insure that it is functioning properly. You will notice the sample door open and close and may hear the stage move from one position to another. The green and yellow indicator lights on the front of the instrument will be lit. Once the yellow light goes out, the instrument is ready for use.

Continue to eLicensing setup to license your software.

Troubleshooting Problems connecting to scanner If, after clicking Connect… the software freezes or fails to connect, check the following:

The instrument is connected to the workstation via the provided Ethernet cable.

The Ethernet cable from the instrument to the workstation is connected to the lower Ethernet card on the workstation (the connector closest to the monitor adapter).

Turn off Windows Firewall and retry.

VERY IMPORTANT If the workstation is plugged into a network, it must be unplugged before shutting off the firewall! Unplug the Ethernet cable from the port located closest to the top of the instrument. After installation has completed, turn Windows Firewall back on and plug the network cable back in.

Check to see if any anti-virus software is blocking communication between the instrument and workstation; check with your local IT contact who may have installed such software.

Page 10: IN Cell Analyzer 2000 Instrument - GE Healthcareincelldownload.gehealthcare.com/bin/download_data... · 2012-04-06 · IN Cell Analyzer 2000 instrument will automatically re-initialize.

Release Notes

10

Customer/Technical Support Asia Pacific Tel: +852 2811 8693 Fax: +852 2811 5251 Australasia Tel: + 61 2 9899 0999 Fax: +61 2 9899 7511 Austria Tel: 01/57606-1619 Fax: 01/57606-1627 Belgium Tel: 0800 73 888 Fax: 03 272 1637 Canada Tel: 1 800 463 5800 Fax: 1 800 567 1008 Central, East, & South East Europe Tel: +43 1 982 3826 Fax: +43 1 985 8327 Denmark Tel: 45 16 2400 Fax: 45 16 2424 Finland & Baltics Tel: +358-(0)9-512 39 40 Fax: +358 (0)9 512 39 439 France Tel: 01 6935 6700 Fax: 01 6941 9677 Germany Tel 089 / 96281 660 Fax 089 / 96281 620 Italy Tel: 02 27322 1 Fax: 02 27302 212 Japan Tel: +81 3 5331 9336 Fax: +81 3 5331 9370

Latin America Tel: +55 11 3933 7300 Fax: +55 11 3933 7306 Middle East & Africa Tel: +30 210 9600 687 Fax: +30 210 9600 693 Netherlands Tel: 0165 580 410 Fax: 0165 580 401 Norway Tel: 815 65 555 Fax: 815 65 666 Portugal Tel: 21 417 7035 Fax: 21 417 3184 Russia & other C.I.S. & N.I.S Tel: +7 (095) 232 0250 +7 (095) 956 1137 Fax: +7 (095) 230 6377 South East Asia Tel: 60 3 8024 2080 Fax: 60 3 8024 2090 Spain Tel: 93 594 49 50 Fax: 93 594 49 55 Sweden Tel: 018 612 1900 Fax: 018 612 1910 Switzerland Tel: 0848 8028 12 Fax: 0848 8028 13 UK Tel: 0800 616928 Fax: 0800 616927 USA Tel: +1 800 526 3593 Fax: +1 877 295 8102