Improved eddy flux measurements by open-path gas analyzer ... of … · Improved eddy flux...

1
Improved eddy flux measurements by open-path gas analyzer and sonic anemometer co-location Ivan Bogoev Campbell Scientific, Inc., Logan, Utah, USA Conclusions 1. Co-locating the open-path gas-analyzer and sonic measurement volumes preserves the true covariance between all variables associated with the WPL terms and eliminates biases in the eddy-flux estimates. The correction factors accounting for the loss of correlation due to spatial separation in the individual WPL terms (Massman, 2004) are 6.5% and 13.8% for w´ρ v ´ and w´ρ c ´ respectively. 2. IRGASON temperature agrees with the ambient thermistor probe and CSAT3 sonic temperatures to within 1.1% and 2.4% respectively, which in- dicates that the housing surfaces adjacent to the open-path sensing vol- ume are not appreciably warmer or cooler than the ambient air. When cor- rected for humidity, IRGASON sonic temperature is accurate and reliable for calculating CO 2 mixing ratios. It has sufficient frequency response, and it is not affected by solar radiation. 3. Compared to the CSAT3, the IRGASON underestimates hourly and cumula- tive sensible heat flux by 5.7% and 0.7% respectively. 4. Calculating CO 2 flux using point-by-point conversion to mixing ratio is fea- sible for an open-path gas analyzer and a co-located sonic anemometer/ thermometer. The air density WPL terms can be implicitly accounted for with this approach. Differences between CO 2 flux calculated using point-by- point conversion to mixing ratio and flux computed following the traditional WPL methodology are less than 0.3%. The pressure term of the density correc- tions (Zhang et al., 2011) is small for this site and does not explain the differ- ence between WPL and molar-ratio-based fluxes. No apparent CO 2 uptake was observed during off-season and cold periods over snow-covered surfaces, which also suggests negligible instrument induced heat flux in the sensing path of the gas analyzer. Future work Validate the mixing-ratio method with flux measurements by a closed-path eddy-covariance system. Literature cited Grelle, A., Burba, G. (2007) Fine-wire thermometer to correct CO 2 fluxes by open- path analyzers for artificial density fluctuations. Ag. For. Meteorol 147, 48-57. Massman, W. (2004). Concerning the measurement of atmospheric trace gas flux- es with open- and closed-path eddy covariance system. In: Lee, X., Massman, W., Law, B. (eds). Handbook of micrometeorology. Dordrecht: Kluwer Academic Pub., 67-99. Webb E., G. Pearman, and R. Leuning (1980). Correction of flux measurements for density effects due to heat and water vapour transfer. Quart. J. Roy. Meteorol. Soc 106: 85-100. Zhang J., X. Lee, G. Songa, S. Hana (2011) Pressure correction to the long-term mea- surement of carbon dioxide flux. Agricultural and Forest Meteorology 151, 1, 70–77. Acknowledgements We thank Campbell Scientific for the support of this study, Dylan Finlayson for the site maintenance and data collection, Larry Jacobsen, Mark Blonquist, Steve Sargent, and Dave Meek for the review of the poster. We are grateful to Bert Tanner for the in- spiration and encouragement. Further information More details and specifications of the IRGASON instrument can be found at: www.campbellsci.com/irgason. Correspondence: +1 435 227 9702, [email protected] Introduction Eddy flux is systematically underestimated because of: Spatial separation between measurements of w´ (vertical wind) and ρ´ (gas density) Temporal asynchronicity between measurements of w´, T´, and ρ´ Open-path gas analyzers introduce biases in the flux estimates attributed to: Variations of air density with temperature T´ and water vapor ρ v ´ (Webb et al., 1980), (Massman, 2004) Instrument-induced surface-heat exchange (Grelle et al., 2007) The IRGASON addresses these problems with the following features: Simultaneously measures w´, T´, ρ v ´, and ρ c ´ in the same volume of air Reduces instrument self-heating and solar radiation loading due to low power consumption and small- diameter, aerodynamic housing Implicitly accounts for air density effects with the ability to compute CO 2 flux using point-by-point conversion to mixing ratio Research objectives This study was conducted to: 1. Examine the effect of anemometer and gas-analyzer separation on sensible (Hs), latent (Le), and CO 2 (Fc) fluxes 2. Compare the IRGASON and CSAT3 sonic temperatures 3. Evaluate the influence of instrument induced heat on ambient sensible heat flux measurements 4. Test the concept of calculating fluxes measured by an open-path analyzer using instantaneous point-by-point conversion to CO 2 mixing ratio Fig 1. Test setup at a pasture near Logan, Utah Measurement height: IRGASON: 1.65 m CSAT3: 2 m Spatial separation: Horizontal: 0.35 m Vertical: 0.2 m Sampling rate: 20 Hz Temperature probe CSAT3 IRGASON Materials and methods Operate the IRGASON and CSAT3 in the field over different environmental conditions.  Calculate flux from the IRGASON using instantaneous CO 2 mixing-ratio (MR) based on the provided w´, T´, ρ v ´, and ρ c ´ measurements and the following steps: a. Correct IRGASON sonic temperature for humidity on-line using the co-located water vapor density: + + + + = v v RairTair Patm v Ts Rair Patm v Ts Tair ρ ρ ρ ρ 51 . 0 1 1 51 . 0 1 1 b. Compute water-vapor pressure and instantaneous CO 2 mixing ratio using: v v M RT e c c c M e P RT ) ( c. Calculate CO 2 flux using the instantaneous CO 2 mixing ratio: ' ' c c MR c w T R m e P F Compare the results with CO 2 fluxes computed with the tradi- tional WPL approach: T T w m m w m m w F c d v v d v d c v d c WPL c ' ' 1 ' ' ' ' where the effect of humidity on sonic temperature is cor- rected with: ' ' 51 . 0 ' ' ' ' w T T w T w s Results 1. Effect of sensor spatial separation on eddy fluxes Eddy flux is computed when co-located measurements of w´ and T´ from the IRGASON are replaced with equivalent measurements from the CSAT3. 1A. Effect of spatial separation on Hs A 14.3% loss in cumulative Hs between co-located w´ and T´ and dis- placed (w´ CSAT3 , T´ IRGASON ) measurements was observed. The loss in- creases to 25.3% when w´ is underneath the T´ (w´ IRGASON , T´ CSAT3 ). 01/16 03/07 04/26 06/15 08/04 09/23 11/12 01/01 -50 0 50 100 150 200 250 300 350 Cumulative H s flux Date Cumulative H s flux [MJm -2 ] H s w’ IRGASON T’ IRGASON H s w’ CSAT3 T’ CSAT3 H s w’ CSAT3 T’ IRGASON H s w’ IRGASON T’ CSAT3 w’ and T’ co-located (IRGASON) w’ and T’ co-located (CSAT3) w’ and T’ separated (w’ below T’) w’ and T’ separated (w’ above T’) 1B. Effect of spatial separation on raw Le The cumulative uncorrected water vapor flux w´ρ v ´ from the IRGASON (w´ and ρ v ´ co-located) is 6.5% higher than the same flux computed using ρ v ´ from the IRGASON and w´ from the adjacent CSAT3. 01/16 03/07 04/26 06/15 08/04 09/23 11/12 01/01 0 100 200 300 400 500 600 700 Cumulative w’ ρv Date Cumulative w’ ρv [MJ m -2 ] w’ IRGASON ρv IRGASON w’ CSAT3 ρv IRGASON w’ separated from ρv w’ and ρv co-located 6.5% 1C. Effect of spatial separation on raw Fc The magnitude of the cumulative uncorrected CO 2 flux w´ρ c ´ from the IRGASON is 13.8% larger than the cumulative flux from the spatially displaced measurements: ρ c ´ IRGASON and w´ CSAT3. 01/16 03/07 04/26 06/15 08/04 09/23 11/12 01/01 -2000 -1500 -1000 -500 0 Cumulative w’ ρc Date Cumulative w’ ρc [g CO 2 m -2 ] w’ IRGASON ρc IRGASON w’ CSAT3 ρc IRGASON w’ separated from ρc w’ and ρc co-located 13.8% 1D. Effect of spatial separation on WPL corrected Fc Flux is underestimated 41% when w´ and T´ measurements are sep- arated from the ρ v ´ and ρ c ´. The error is reduced to 27% when T´ is co-located with ρ v ´ and ρ c ´. 01/16 03/07 04/26 06/15 08/04 09/23 11/12 01/01 -700 -600 -500 -400 -300 -200 -100 0 100 Cumulative CO 2 flux Date Cumulative CO 2 flux [g CO 2 m -2 ] IRGASON approach: w’,T’,ρ v ’,ρ c ’ all co-located 27% 41% Grelle 2007 approach: T’, ρ v ’,ρ c ’ co-located but w’ is separated Traditional approach: w’,T’ measured by CSAT3 are separated from ρ v ’,ρ c ’ measured by IRGA 2. Comparison of sonic temperature IRGASON and CSAT3 sonic temperatures X-Y slopes agree with the thermistor probe within 1.1% and 3.6% respectively. The CSAT3 overestimated the slope by 2.4% compared to the IRGASON. The CSAT3 has 0.49 °C offset compared to the IRGASON and the air-temperature probe. -30 -20 -10 0 10 20 30 40 -30 -20 -10 0 10 20 30 40 Mean air temperature Temperature Probe With Radiation Shield [C] Humidity corrected sonic temperature [C] CSAT3A slope = 1.0361 offset = 0.48709 R2 = 0.99731 IRGASON slope = 1.0114 offset = -0.0099898 R2 = 0.99654 CSAT3A fit CSAT3A IRGASON fit IRGASON -30 -20 -10 0 10 20 30 40 -30 -20 -10 0 10 20 30 40 Sonic temperature CSAT3A [C] IRGASON [C] slope = 0.97646 offset = -0.49277 R2 = 0.99965 1:1 3. Hourly sensible heat flux comparison Compared to the CSAT3, the IRGASON underestimates the sensible heat flux by 5.7%. Part of this error is attributed to the 2.4% gain error in the sonic temperature of the CSAT3. Cumulative fluxes agree within 0.7%. -100 -50 0 50 100 150 200 250 300 350 -100 -50 0 50 100 150 200 250 300 350 400 Sensible heat flux CSAT3A [W m -2 ] IRGASON [W m -2 ] slope = 1.0569 R2 = 0.94706 1:1 4. Comparison of CO 2 flux computed by the mixing-ratio method to the traditional WPL density-based approach Both methods yield identical results to within 0.25%. The pressure term (Zhang et al., 2011) is negligible for this site and does not explain the small difference between the two approaches. No apparent CO 2 uptake was observed during off-season and over snow-covered surfaces with either method. (No instrument heating corrections were applied.) 01/16 03/07 04/26 06/15 08/04 09/23 11/12 01/01 -700 -600 -500 -400 -300 -200 -100 0 100 IRGASON molar ratio and WPL cumulative CO 2 fluxes Date Cumulative CO 2 flux [g CO 2 m -2 ] F c Molar ratio F c WPL SNOW COVER 01/16 03/07 04/26 06/15 08/04 09/23 11/12 01/01 -1.5 -1 -0.5 0 0.5 Cumulative CO 2 flux error and WPL pressure term Date Flux error [g CO 2 m -2 ] F c Molar Ratio - F c WPL (wP) Measurements w´, T´, ρ v ´, ρ c ´ Gas analyzer Sonic anemometer

Transcript of Improved eddy flux measurements by open-path gas analyzer ... of … · Improved eddy flux...

Page 1: Improved eddy flux measurements by open-path gas analyzer ... of … · Improved eddy flux measurements by open-path gas analyzer and sonic anemometer co-location Ivan Bogoev Campbell

Improved eddy flux measurements by open-path gas analyzer and sonic anemometer co-locationIvan Bogoev

Campbell Scientific, Inc., Logan, Utah, USAConclusions1. Co-locating the open-path gas-analyzer and sonic measurement volumes

preserves the true covariance between all variables associated with the WPL terms and eliminates biases in the eddy-flux estimates. The correction factors accounting for the loss of correlation due to spatial separation in the individual WPL terms (Massman, 2004) are 6.5% and 13.8% for w´ρv´ and w´ρc´ respectively.

2. IRGASON temperature agrees with the ambient thermistor probe and CSAT3 sonic temperatures to within 1.1% and 2.4% respectively, which in-dicates that the housing surfaces adjacent to the open-path sensing vol-ume are not appreciably warmer or cooler than the ambient air. When cor-rected for humidity, IRGASON sonic temperature is accurate and reliable for calculating CO2 mixing ratios. It has sufficient frequency response, and it is not affected by solar radiation.

3. Compared to the CSAT3, the IRGASON underestimates hourly and cumula-tive sensible heat flux by 5.7% and 0.7% respectively.

4. Calculating CO2 flux using point-by-point conversion to mixing ratio is fea-sible for an open-path gas analyzer and a co-located sonic anemometer/thermometer. The air density WPL terms can be implicitly accounted for with this approach. Differences between CO2 flux calculated using point-by-point conversion to mixing ratio and flux computed following the traditional WPL methodology are less than 0.3%. The pressure term of the density correc-tions (Zhang et al., 2011) is small for this site and does not explain the differ-ence between WPL and molar-ratio-based fluxes.

No apparent CO2 uptake was observed during off-season and cold periods over snow-covered surfaces, which also suggests negligible instrument induced heat flux in the sensing path of the gas analyzer.

Future workValidate the mixing-ratio method with flux measurements by a closed-path eddy-covariance system.

Literature citedGrelle, A., Burba, G. (2007) Fine-wire thermometer to correct CO2 fluxes by open-

path analyzers for artificial density fluctuations. Ag. For. Meteorol 147, 48-57.Massman, W. (2004). Concerning the measurement of atmospheric trace gas flux-

es with open- and closed-path eddy covariance system. In: Lee, X., Massman, W., Law, B. (eds). Handbook of micrometeorology. Dordrecht: Kluwer Academic Pub., 67-99.

Webb E., G. Pearman, and R. Leuning (1980). Correction of flux measurements for density effects due to heat and water vapour transfer. Quart. J. Roy. Meteorol. Soc 106: 85-100.

Zhang J., X. Lee, G. Songa, S. Hana (2011) Pressure correction to the long-term mea-surement of carbon dioxide flux. Agricultural and Forest Meteorology 151, 1, 70–77.

AcknowledgementsWe thank Campbell Scientific for the support of this study, Dylan Finlayson for the site maintenance and data collection, Larry Jacobsen, Mark Blonquist, Steve Sargent, and Dave Meek for the review of the poster. We are grateful to Bert Tanner for the in-spiration and encouragement.

Further informationMore details and specifications of the IRGASON instrument can be found at: www.campbellsci.com/irgason.

Correspondence: +1 435 227 9702, [email protected]

IntroductionEddy flux is systematically underestimated because of:

• Spatial separation between measurements of w´ (vertical wind) and ρ´ (gas density)

• Temporal asynchronicity between measurements of w´, T´, and ρ´

Open-path gas analyzers introduce biases in the flux estimates attributed to:• Variations of air density with temperature T´ and water vapor ρv´

(Webb et al., 1980), (Massman, 2004)

• Instrument-induced surface-heat exchange (Grelle et al., 2007)

The IRGASON addresses these problems with the following features:• Simultaneously measures w´, T´, ρv´, and ρc´ in the same

volume of air

• Reduces instrument self-heating and solar radiation loading due to low power consumption and small- diameter, aerodynamic housing

• Implicitly accounts for air density effects with the ability to compute CO2 flux using point-by-point conversion to mixing ratio

Research objectivesThis study was conducted to:

1. Examine the effect of anemometer and gas-analyzer separation on sensible (Hs), latent (Le), and CO2 (Fc) fluxes

2. Compare the IRGASON and CSAT3 sonic temperatures

3. Evaluate the influence of instrument induced heat on ambient sensible heat flux measurements

4. Test the concept of calculating fluxes measured by an open-path analyzer using instantaneous point-by-point conversion to CO2 mixing ratio

Fig 1. Test setup at a pasture near Logan, Utah

Measurement height: IRGASON: 1.65 m CSAT3: 2 m

Spatial separation: Horizontal: 0.35 m Vertical: 0.2 m

Sampling rate: 20 Hz

Temperature probe CSAT3

IRGASON

Materials and methodsOperate the IRGASON and CSAT3 in the field over different environmental conditions. Calculate flux from the IRGASON using instantaneous CO2 mixing-ratio (MR) based on the provided w´, T´, ρv´, and ρc´ measurements and the following steps:

a. Correct IRGASON sonic temperature for humidity on-line using the co-located water vapor density:

+

++

+

=

v

vRairTair

Patmv

TsRair

Patmv

TsTair

ρ

ρ

ρ

ρ

51.01

1

51.01

1

b. Compute water-vapor pressure and instantaneous CO2 mixing ratio using:

'' c

cMRc w

TRmeP

F

vvdsv

sd

v

s

RPTT

RP

TT

))//((51.011

51.01

1

''51.0'''' wTTwTw s

v

v

MRT

e

c

cc MeP

RT)(

TTw

mmw

mmwF c

d

v

v

dv

d

c

v

dc

WPLc

''1''''

'' c

cMRc w

TRmeP

F

vvdsv

sd

v

s

RPTT

RP

TT

))//((51.011

51.01

1

''51.0'''' wTTwTw s

v

v

MRT

e

c

cc MeP

RT)(

TTw

mmw

mmwF c

d

v

v

dv

d

c

v

dc

WPLc

''1''''

c. Calculate CO2 flux using the instantaneous CO2 mixing ratio:

'' c

cMRc w

TRmeP

F

vvdsv

sd

v

s

RPTT

RP

TT

))//((51.011

51.01

1

''51.0'''' wTTwTw s

v

v

MRT

e

c

cc MeP

RT)(

TTw

mm

wmmwF c

d

v

v

dv

d

c

v

dc

WPLc

''1''''

Compare the results with CO2 fluxes computed with the tradi-tional WPL approach:

'' c

cMRc w

TRmeP

F

vvdsv

sd

v

s

RPTT

RP

TT

))//((51.011

51.01

1

''51.0'''' wTTwTw s

v

v

MRT

e

c

cc MeP

RT)(

TTw

mmw

mmwF c

d

v

v

dv

d

c

v

dc

WPLc

''1''''

where the effect of humidity on sonic temperature is cor-rected with:

'' c

cMRc w

TRmeP

F

vvdsv

sd

v

s

RPTT

RP

TT

))//((51.011

51.01

1

''51.0'''' wTTwTw s

v

v

MRT

e

c

cc MeP

RT)(

TTw

mmw

mm

wF cd

v

v

dv

d

c

v

dc

WPLc

''1''''

Results1. Effect of sensor spatial separation on eddy fluxes Eddy flux is computed when co-located measurements of w´ and T´ from the IRGASON are replaced with equivalent measurements from the CSAT3.

1A. Effect of spatial separation on Hs A 14.3% loss in cumulative Hs between co-located w´ and T´ and dis-placed (w´CSAT3, T´IRGASON) measurements was observed. The loss in-creases to 25.3% when w´ is underneath the T´ (w´IRGASON, T´CSAT3).

01/16 03/07 04/26 06/15 08/04 09/23 11/12 01/01−50

0

50

100

150

200

250

300

350

Cumulative Hs flux

Date

Cum

ulat

ive

Hs fl

ux [M

Jm −

2 ]

∫Hs w’

IRGASON T’

IRGASON

∫Hs w’

CSAT3 T’

CSAT3

∫Hs w’

CSAT3 T’

IRGASON

∫Hs w’

IRGASON T’

CSAT3

w’ and T’ co−located (IRGASON)

w’ and T’ co−located (CSAT3)

w’ and T’ separated(w’ below T’)

w’ and T’ separated(w’ above T’)

1B. Effect of spatial separation on raw LeThe cumulative uncorrected water vapor flux w´ρv´ from the IRGASON (w´ and ρv´ co-located) is 6.5% higher than the same flux computed using ρv´ from the IRGASON and w´ from the adjacent CSAT3.

01/16 03/07 04/26 06/15 08/04 09/23 11/12 01/01

0

100

200

300

400

500

600

700

Cumulative w’ρ’v

Date

Cum

ulat

ive

w’ρ

’ v [MJ

m−

2 ]

∫w’IRGASON

ρ’v IRGASON

∫w’CSAT3

ρ’v IRGASON

w’ separated from ρ’v

w’ and ρ’v co−located

6.5%

1C. Effect of spatial separation on raw FcThe magnitude of the cumulative uncorrected CO2 flux w´ρc´ from the IRGASON is 13.8% larger than the cumulative flux from the spatially displaced measurements: ρc´ IRGASON and w´ CSAT3.

01/16 03/07 04/26 06/15 08/04 09/23 11/12 01/01

−2000

−1500

−1000

−500

0

Cumulative w’ρ’c

Date

Cum

ulat

ive

w’ρ

’ c [g C

O2 m

−2 ]

∫w’IRGASON

ρ’c IRGASON

∫w’CSAT3

ρ’c IRGASON

w’ separated from ρ’c

w’ and ρ’c co−located

13.8%

1D. Effect of spatial separation on WPL corrected FcFlux is underestimated 41% when w´ and T´ measurements are sep-arated from the ρv´ and ρc´. The error is reduced to 27% when T´ is co-located with ρv´ and ρc´.

01/16 03/07 04/26 06/15 08/04 09/23 11/12 01/01

−700

−600

−500

−400

−300

−200

−100

0

100

Cumulative CO2 flux

Date

Cum

ulat

ive

CO

2 flux

[g C

O2 m

−2 ]

IRGASON approach: w’,T’,ρ

v’,ρ

c’ all co−located 27%41%

Grelle 2007 approach:T’, ρ

v’,ρ

c’ co−located

but w’ is separated

Traditional approach: w’,T’ measured by CSAT3are separated from ρ

v’,ρ

c’ measured by IRGA

2. Comparison of sonic temperatureIRGASON and CSAT3 sonic temperatures X-Y slopes agree with the thermistor probe within 1.1% and 3.6% respectively. The CSAT3 overestimated the slope by 2.4% compared to the IRGASON. The CSAT3 has 0.49 °C offset compared to the IRGASON and the air-temperature probe.

−30 −20 −10 0 10 20 30 40−30

−20

−10

0

10

20

30

40Mean air temperature

Temperature Probe With Radiation Shield [C]

Hum

idity

cor

rect

ed

soni

c te

mpe

ratu

re [C

]

CSAT3A slope = 1.0361 offset = 0.48709 R2 = 0.99731

IRGASON slope = 1.0114 offset = −0.0099898 R2 = 0.99654

CSAT3Afit CSAT3AIRGASONfit IRGASON

−30 −20 −10 0 10 20 30 40−30

−20

−10

0

10

20

30

40Sonic temperature

CSAT3A [C]

IRG

AS

ON

[C]

slope = 0.97646 offset = −0.49277 R2 = 0.99965

1:1

3. Hourly sensible heat flux comparisonCompared to the CSAT3, the IRGASON underestimates the sensible heat flux by 5.7%. Part of this error is attributed to the 2.4% gain error in the sonic temperature of the CSAT3. Cumulative fluxes agree within 0.7%.

−100 −50 0 50 100 150 200 250 300 350−100

−50

0

50

100

150

200

250

300

350

400Sensible heat flux

CS

AT

3A [W

m−

2 ]

IRGASON [W m−2]

slope = 1.0569 R2 = 0.94706

1:1

4. Comparison of CO2 flux computed by the mixing-ratio method to the traditional WPL density-based approach Both methods yield identical results to within 0.25%. The pressure term (Zhang et al., 2011) is negligible for this site and does not explain the small difference between the two approaches. No apparent CO2 uptake was observed during off-season and over snow-covered surfaces with either method. (No instrument heating corrections were applied.)

01/16 03/07 04/26 06/15 08/04 09/23 11/12 01/01

−700

−600

−500

−400

−300

−200

−100

0

100

IRGASON molar ratio and WPL cumulative CO2 fluxes

Date

Cum

ulat

ive

CO

2 flux

[g C

O2 m

−2 ]

∫Fc Molar ratio

∫Fc WPLSNOW

COVER

01/16 03/07 04/26 06/15 08/04 09/23 11/12 01/01

−1.5

−1

−0.5

0

0.5

Cumulative CO2 flux error and WPL pressure term

Date

Flu

x er

ror

[g C

O2 m

−2 ]

∫Fc Molar Ratio − ∫F

c WPL

∫(wP)

Measurementsw´, T´, ρv´, ρc´

Gas analyzer

Sonic anemometer