STATEMENT: This test may not be duplicated or used in part without prior written consent from Bay Area Compliance Laboratories Corp. (Shenzhen). The test data was only valid for the test sample(s). This report must not be used by the customer to claim product certification, approval, or endorsement by NVLAP, NIST, or any agency of the U.S. Federal Government. This report is valid only with a valid digital signature. The digital signature may be available only under the Adobe software above version 7.0. IES LM-79-08 MEASUREMENT AND TEST REPORT For ATG Electronics Corp 9020 Rancho Park Court Rancho Cucamonga, CA 91730 Test Model: CPPG40HU400000 Report Type: Electrical and Photometric tests including: Luminous Flux, Color, Luminous Intensity Distribution, THD, Off-state Power Test Engineer: Daniel Duan Report Number: RSZ130611501-10 Test Date: 2014-04-02 to 2014-04-03 Report Date: 2014-04-11 Reviewed By: Jeanne Han/Safety Manager Prepared By: Bay Area Compliance Laboratories Corp. (Shenzhen) 6/F, the 3rd Phase of WanLi Industrial Building, ShiHua Road, FuTian Free Trade Zone Shenzhen, Guangdong, China Tel: +86-755-33320018 Fax: +86-755-33320008 Test Facility: Test facility was located at Pu Long Cun 69, Puxinghu Industrial Area, Tangxia Town, Dongguan, Guangdong, P.R.China. Accreditation: The NVLAP Lab Code is 200707-0.

Transcript of IES LM-79-08 - LED light Fixtures Manufacturer -ATG …€¦ ·  · 2015-12-18IESNA LM-79-08: ......

STATEMENT: This test may not be duplicated or used in part without prior written consent from Bay Area Compliance Laboratories Corp.

(Shenzhen). The test data was only valid for the test sample(s). This report must not be used by the customer to claim product certification,

approval, or endorsement by NVLAP, NIST, or any agency of the U.S. Federal Government.

This report is valid only with a valid digital signature. The digital signature may be available only under the Adobe software above version 7.0.

IES LM-79-08

MEASUREMENT AND TEST REPORT For

ATG Electronics Corp 9020 Rancho Park Court Rancho Cucamonga, CA 91730

Test Model: CPPG40HU400000

Report Type:

Electrical and Photometric tests including: Luminous Flux,

Color, Luminous Intensity Distribution, THD, Off-state

Power

Test Engineer: Daniel Duan

Report Number: RSZ130611501-10

Test Date: 2014-04-02 to 2014-04-03

Report Date: 2014-04-11

Reviewed By: Jeanne Han/Safety Manager

Prepared By:

Bay Area Compliance Laboratories Corp. (Shenzhen)

6/F, the 3rd Phase of WanLi Industrial Building,

ShiHua Road, FuTian Free Trade Zone

Shenzhen, Guangdong, China

Tel: +86-755-33320018

Fax: +86-755-33320008

Test Facility: Test facility was located at Pu Long Cun 69, Puxinghu Industrial

Area, Tangxia Town, Dongguan, Guangdong, P.R.China.

Accreditation: The NVLAP Lab Code is 200707-0.

Bay Area Compliance Laboratories Corp. (Shenzhen) 6/F, the 3rd Phase of WanLi Industrial Building, ShiHua Road, FuTian

Free Trade Zone Shenzhen, Guangdong, China

Report No. RSZ130611501-10 Page 2 of 12 The NVLAP Lab Code is 200707-0.

1. Product Description

General Information:

Sample No.: RSZ130611501-S01 Model: CPPG40HU400000 One sample was received on 2013-06-11 and used for testing.

Model Tested: CPPG40HU400000

Manufacturer: ATG Electronics Corp

Brand Name: ATG

Product Designation: LED CANOPY

Burning Time Before Test: 0 hour(For New Products)

Rated Values:

Rated Voltage/Frequency: 100-277 VAC 50/60Hz

Rated Power: 40W

Nominal CCT: 4000K

2. Standards Used

IESNA LM-79-08: Approved Method: Electrical & Photometric Measurement of Solid-state Lighting Products

ANSI C82.77-2002: Harmonic Emission Limits – Related Power Quality Requirements for Lighting

3. Description of Test Equipment

Device Manufacture Model No Serial No Test Range Calibration date

Integrating Sphere SENSING SPR-600 S09008 25~50℃ 2014-03-16

Spectral photometer SENSING SPR3000 90902027 350nm~800nm 2014-03-16

Power Meter YOKOGAWA WT-210 91j926132 15/30/60/150/300/60

0 V 2014-03-12

AC Power Supply ALL Power APW-105N 970613 0V-300V 50-400Hz 2014-03-12

Standard Light Source SENSING N/A LSD090808 N/A 2013-05-13

Thermal Meter SENSING N/A T-08-EE006-1 25℃ 2014-03-16

DC Power Supply ITECH IT6154 0061 0417

6471 0010 19 0~60V 2014-03-12

AC Power Supply EVERFINE VPS1060 PWM 1101006 0-150V, 0-300V 2014-03-12

DC Power Supply EVERFINE WY12010 1009009 30V/5A 2014-03-12

Power Meter YOKOGAWA WT-210 91KB35700 N/A 2014-03-12

Goniophotometer EVERFINE GO-R5000 YG108492N1

0120001 1600mm,3000W/10A 2014-03-04

Thermal Meter Victor VC230 EE091 0~40℃0~90% 2013-04-01

Standard Light Source EVERFINE D908 1012001 N/A 2013-05-28

Digital Power Meter EVERFINE PF9811 507047 0~35V DC 2013-11-12

AC POWER SUPPLY SZHPC HPA 1103 0003394 3KVA 2014-03-12

Bay Area Compliance Laboratories Corp. (Shenzhen) 6/F, the 3rd Phase of WanLi Industrial Building, ShiHua Road, FuTian

Free Trade Zone Shenzhen, Guangdong, China

Report No. RSZ130611501-10 Page 3 of 12 The NVLAP Lab Code is 200707-0.

Statement of Traceability: Bay Area Compliance Laboratories Corp. (Shenzhen) attested that all calibration has been

performed using suitable standards traceable to National Primary Standards and International System of Units (SI).

4. Test Method

Product was tested with no seasoning. All stabilization and measurements were made in compliance with IES LM-79-08. The product was operated at rated voltage or at voltage required by manufacturer. The ambient temperature of the sample was maintained at 25°C±1°C during measurement. And relative humidity is less than 65%.

Integrating Sphere System

The system includes AC power source, digital power meter, DC power supply, spectrophotometer, and integrating sphere. The integrating sphere system is calibrated by standard light source before measurement. The system and standard light source has been calibrated regularly and traceable to the National Primary Standards.

4π geometry was used during measurement. The product was operated in its intended orientation in application and was recorded in this report.

The uncertainty of the light output (luminous flux) measurements is U=1.64% (K=2), at the 95% confidence level. The

uncertainty of the correlated color temperature measurements is U=21K (K=2), at the 95% confidence level. The

uncertainty of the CRI is U=1.6 (K=2) , at the 95% confidence level.

Goniophotometer System

The goniophotometer system is calibrated by standard light source before measurement. The standard light source has been calibrated regularly and traceable to the National Primary Standards.

Type C goniophotometer was used for measuring total luminous flux, luminous intensity distribution, and color spatial uniformity. The product was operated in its intended orientation in application and was recorded in this report.

The uncertainty of the luminous intensity is U=2.82% (K=2) , at the 95% confidence level.

Additional Test

The Additional Test item may not be covered by IESNA LM-79-2008. Additional test including power factor, off-state power and THD, was measured by Digital Power Meter after stabilized at 25°C±1°C.Test voltage for THD and power factor test would be equal to rated voltage or, in case of a voltage range, maximum value of that range.

The uncertainty of power meter AC current U=0.19 % of rdg, AC Voltage U=0.15% of rdg, Power U=0.20%) (K=2), at

the 95% confidence level.

Bay Area Compliance Laboratories Corp. (Shenzhen) 6/F, the 3rd Phase of WanLi Industrial Building, ShiHua Road, FuTian

Free Trade Zone Shenzhen, Guangdong, China

Report No. RSZ130611501-10 Page 4 of 12 The NVLAP Lab Code is 200707-0.

5. Test Result

[Integrating Sphere System]

Total operating time for integrating sphere test: 1.0 hours

Test orientation: Downward

Electrical Measurement

Voltage (V) Frequency (Hz) Current (A) Power (W) Power Factor

120.05 60.0 0.3125 37.34 0.995

Photometric Measurement

Luminous Flux (lm) Radiant Flux (W) Efficacy (lm/W) CCT (K) Duv

3302.839 7.860 88.453 4110 -1.40E-03

Chromaticity Coordinate

x y u v u' v'

0.3748 0.3701 0.2240 0.3318 0.2240 0.4978

Color Rendering Index

Ra

84.3

R1 R2 R3 R4

83 89 93 84

R5 R6 R7 R8

83 84 88 71

R9 R10 R11 R12

23 73 83 62

R13 R14 R15

85 96 80

Bay Area Compliance Laboratories Corp. (Shenzhen) 6/F, the 3rd Phase of WanLi Industrial Building, ShiHua Road, FuTian

Free Trade Zone Shenzhen, Guangdong, China

Report No. RSZ130611501-10 Page 5 of 12 The NVLAP Lab Code is 200707-0.

Relative Spectral Power Distribution

nm mW

nm mW

nm mW

nm mW

nm mW

380 8.941E-03 465 1.103E-01 550 1.744E-01 635 1.517E-01 720 2.126E-02

385 5.257E-03 470 8.085E-02 555 1.789E-01 640 1.420E-01 725 1.823E-02

390 3.664E-03 475 6.302E-02 560 1.845E-01 645 1.315E-01 730 1.597E-02

395 2.784E-03 480 6.070E-02 565 1.895E-01 650 1.208E-01 735 1.361E-02

400 2.279E-03 485 6.346E-02 570 1.932E-01 655 1.109E-01 740 1.206E-02

405 2.002E-03 490 7.212E-02 575 1.972E-01 660 1.013E-01 745 1.024E-02

410 2.716E-03 495 8.407E-02 580 1.988E-01 665 9.157E-02 750 9.019E-03

415 4.525E-03 500 9.864E-02 585 1.999E-01 670 8.155E-02 755 8.077E-03

420 9.388E-03 505 1.122E-01 590 2.027E-01 675 7.318E-02 760 6.946E-03

425 1.905E-02 510 1.247E-01 595 2.034E-01 680 6.383E-02 765 6.233E-03

430 3.671E-02 515 1.346E-01 600 2.012E-01 685 5.645E-02 770 5.363E-03

435 6.751E-02 520 1.420E-01 605 2.005E-01 690 4.999E-02 775 4.827E-03

440 1.262E-01 525 1.477E-01 610 1.969E-01 695 4.282E-02 780 4.486E-03

445 2.214E-01 530 1.537E-01 615 1.909E-01 700 3.745E-02 785 4.093E-03

450 2.645E-01 535 1.581E-01 620 1.816E-01 705 3.221E-02 790 3.505E-03

455 1.957E-01 540 1.629E-01 625 1.717E-01 710 2.810E-02 795 3.522E-03

460 1.407E-01 545 1.684E-01 630 1.624E-01 715 2.483E-02 800 3.068E-03

0.0

0.2

0.4

0.6

0.8

1.0

1.2

380 440 500 560 620 680 740 800

Bay Area Compliance Laboratories Corp. (Shenzhen) 6/F, the 3rd Phase of WanLi Industrial Building, ShiHua Road, FuTian

Free Trade Zone Shenzhen, Guangdong, China

Report No. RSZ130611501-10 Page 6 of 12 The NVLAP Lab Code is 200707-0.

CIE 1931 x y Chromaticity Diagram

7-Step & 4-Step Chromaticity Quadrangles

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8

0.30

0.35

0.40

0.45

0.30 0.35 0.40 0.45 0.50

sample

7-STEP 2700K

7-STEP 3000K

7-STEP 3500K

7-STEP 4000K

7-STEP 4500K

7-STEP 5000K

7-STEP 5700K

7-STEP 6500K

4-STEP 2700K

4-STEP 3000K

4-STEP 3500K

4-STEP 4000K

4-STEP 5000K

4-STEP 6500K

Bay Area Compliance Laboratories Corp. (Shenzhen) 6/F, the 3rd Phase of WanLi Industrial Building, ShiHua Road, FuTian

Free Trade Zone Shenzhen, Guangdong, China

Report No. RSZ130611501-10 Page 7 of 12 The NVLAP Lab Code is 200707-0.

[Goniophotometer System]

Total operating time for luminous intensity distribution: 1.5 hours

Test orientation: Downward

Electrical Measurement

Input Voltage (V) Frequency (Hz) Input Current (A) Power (W) Power Factor

120.2 60.0 0.3111 37.17 0.9942

Photometric Measurement

Luminous Flux (lm) Efficacy (lm/W) CBCP (cd) S/MH (C0/180) S/MH (C90/270)

3337.3 89.78 989 1.30 1.32

Luminous Intensity Distribution

C0/180 C45/225 C90/270 C135/315 AVG.

Beam Angle (50% Imax): 115.3 119.9 118.6 119.0 118.2

Field Angle (10% Imax): 166.0 171.7 166.8 172.2 169.2

Bay Area Compliance Laboratories Corp. (Shenzhen) 6/F, the 3rd Phase of WanLi Industrial Building, ShiHua Road, FuTian

Free Trade Zone Shenzhen, Guangdong, China

Report No. RSZ130611501-10 Page 8 of 12 The NVLAP Lab Code is 200707-0.

Luminous Intensity (cd) Distribution Data

C

0° 22.5° 45° 67.5° 90° 112.5° 135° 157.5° γ

0.0° 989 989 989 989 989 989 989 989

5.0° 987 983 995 1006 1013 1003 986 976

10.0° 959 986 999 986 997 982 998 969

15.0° 943 964 977 985 989 981 961 947

20.0° 917 945 952 979 967 968 935 913

25.0° 889 914 932 922 919 921 904 890

30.0° 834 873 885 873 883 857 864 852

35.0° 795 848 823 835 820 822 806 803

40.0° 765 797 794 787 780 776 766 771

45.0° 693 727 725 716 733 709 707 693

50.0° 618 649 653 666 663 651 633 617

55.0° 547 574 583 578 586 563 555 540

60.0° 470 483 498 489 482 476 476 460

65.0° 372 397 411 404 399 394 395 374

70.0° 284 306 322 315 302 311 313 289

75.0° 205 223 235 224 212 224 233 215

80.0° 135 153 162 154 139 156 164 150

85.0° 79 97 107 99 82 101 109 98

90.0° 67 79 86 79 67 80 86 79

95.0° 4 8 59 70 63 70 55 8

100.0° 67 75 22 31 39 30 22 73

105.0° 65 70 61 22 13 17 55 67

110.0° 62 64 56 47 53 42 46 59

115.0° 57 56 49 45 62 44 37 47

120.0° 48 45 40 42 55 41 29 34

125.0° 5 30 29 37 47 37 20 21

130.0° 17 14 19 33 40 33 17 7

135.0° 10 5 15 29 34 29 14 0

140.0° 2 0 11 24 28 24 10 0

145.0° 0 0 6 17 21 17 6 0

150.0° 0 0 1 10 13 10 1 0

155.0° 0 0 0 4 6 4 0 0

160.0° 0 0 0 0 0 0 0 0

165.0° 0 0 0 0 0 0 0 0

170.0° 0 0 0 0 0 0 0 0

175.0° 0 0 0 0 0 0 0 0

180.0° 0 0 0 0 0 0 0 0

Bay Area Compliance Laboratories Corp. (Shenzhen) 6/F, the 3rd Phase of WanLi Industrial Building, ShiHua Road, FuTian

Free Trade Zone Shenzhen, Guangdong, China

Report No. RSZ130611501-10 Page 9 of 12 The NVLAP Lab Code is 200707-0.

Luminous Intensity (cd) Distribution Data (cont.)

C

180° 202.5° 225° 247.5° 270° 292.5° 315° 337.5° γ

0.0° 989 989 989 989 989 989 989 989

5.0° 973 974 980 995 991 999 986 973

10.0° 960 970 981 992 1001 987 981 974

15.0° 930 951 974 998 1012 1004 983 955

20.0° 900 914 966 977 981 989 959 923

25.0° 866 903 922 972 960 967 941 909

30.0° 835 867 898 888 910 890 905 870

35.0° 783 830 833 831 860 839 841 847

40.0° 744 793 784 802 846 810 796 803

45.0° 683 711 739 749 750 757 739 721

50.0° 613 634 661 684 659 691 667 657

55.0° 533 556 590 611 569 622 593 569

60.0° 448 471 510 523 513 531 515 479

65.0° 359 389 421 424 416 435 424 395

70.0° 272 298 330 332 339 341 332 303

75.0° 198 221 241 237 229 241 243 221

80.0° 132 154 171 165 151 168 170 153

85.0° 79 99 109 102 86 103 109 97

90.0° 65 77 85 79 68 81 86 79

95.0° 4 20 46 66 63 68 51 10

100.0° 65 74 29 24 34 25 27 75

105.0° 63 68 60 26 19 28 64 70

110.0° 60 60 50 50 64 53 56 63

115.0° 52 48 41 47 67 46 49 55

120.0° 38 35 33 43 58 42 40 44

125.0° 12 22 25 38 49 37 30 28

130.0° 10 8 19 35 42 34 21 13

135.0° 3 1 16 30 36 30 15 6

140.0° 0 0 11 25 30 25 11 0

145.0° 0 0 5 17 23 19 7 0

150.0° 0 0 0 9 14 12 3 0

155.0° 0 0 0 0 5 4 0 0

160.0° 0 0 0 0 0 0 0 0

165.0° 0 0 0 0 0 0 0 0

170.0° 0 0 0 0 0 0 0 0

175.0° 0 0 0 0 0 0 0 0

180.0° 0 0 0 0 0 0 0 0

Bay Area Compliance Laboratories Corp. (Shenzhen) 6/F, the 3rd Phase of WanLi Industrial Building, ShiHua Road, FuTian

Free Trade Zone Shenzhen, Guangdong, China

Report No. RSZ130611501-10 Page 10 of 12 The NVLAP Lab Code is 200707-0.

Average Area Illumination Figure

Height (m) Diameter

(cm) Eavg(lx) Emax(lx)

0.5 167.1 1010.0 4003.0

1.0 334.2 252.4 1001.0

1.5 501.3 112.2 444.8

2.0 668.4 63.1 250.2

2.5 835.4 40.4 160.1

3.0 1002.5 28.1 111.2

3.5 1169.6 20.6 81.7

4.0 1336.7 15.8 62.6

4.5 1503.8 12.5 49.4

5.0 1670.9 10.1 40.0

Angle: 118.2°. Flux out: 2379.0 lm.

Bay Area Compliance Laboratories Corp. (Shenzhen) 6/F, the 3rd Phase of WanLi Industrial Building, ShiHua Road, FuTian

Free Trade Zone Shenzhen, Guangdong, China

Report No. RSZ130611501-10 Page 11 of 12 The NVLAP Lab Code is 200707-0.

Zonal Lumen Density Measurement

Deg Flux (lm) %

Deg Flux (lm) %

0-5 23.5 0.70 0-5 23.5 0.70

5-10 70.6 2.12 0-10 94.1 2.82

10-15 116.1 3.48 0-15 210.2 6.30

15-20 158.6 4.75 0-20 368.8 11.05

20-25 196.0 5.88 0-25 564.9 16.93

25-30 226.2 6.77 0-30 791.0 23.70

30-35 252.2 7.56 0-35 1043.2 31.26

35-40 269.8 8.08 0-40 1313.0 39.34

40-45 279.2 8.37 0-45 1592.2 47.71

45-50 277.8 8.32 0-50 1870.0 56.03

50-55 264.8 7.94 0-55 2134.7 63.97

55-60 244.2 7.31 0-60 2379.0 71.28

60-65 215.6 6.46 0-65 2594.6 77.74

65-70 180.2 5.40 0-70 2774.8 83.14

70-75 139.6 4.19 0-75 2914.4 87.33

75-80 101.1 3.03 0-80 3015.5 90.36

80-85 67.8 2.03 0-85 3083.3 92.39

85-90 46.3 1.39 0-90 3129.6 93.78

90-95 34.6 1.03 0-95 3164.1 94.81

95-100 24.6 0.74 0-100 3188.7 95.55

100-105 23.8 0.71 0-105 3212.5 96.26

105-110 27.3 0.82 0-110 3239.8 97.08

110-115 27.2 0.81 0-115 3266.9 97.89

115-120 22.4 0.67 0-120 3289.3 98.56

120-125 16.8 0.51 0-125 3306.1 99.07

125-130 11.3 0.33 0-130 3317.4 99.40

130-135 8.0 0.24 0-135 3325.4 99.64

135-140 5.5 0.17 0-140 3330.8 99.81

140-145 3.6 0.10 0-145 3334.4 99.91

145-150 2.0 0.06 0-150 3336.4 99.97

150-155 0.8 0.02 0-155 3337.1 99.99

155-160 0.1 0.01 0-160 3337.3 100.00

160-165 0.0 0.00 0-165 3337.3 100.00

165-170 0.0 0.00 0-170 3337.3 100.00

170-175 0.0 0.00 0-175 3337.3 100.00

175-180 0.0 0.00 0-180 3337.3 100.00

Bay Area Compliance Laboratories Corp. (Shenzhen) 6/F, the 3rd Phase of WanLi Industrial Building, ShiHua Road, FuTian

Free Trade Zone Shenzhen, Guangdong, China

Report No. RSZ130611501-10 Page 12 of 12 The NVLAP Lab Code is 200707-0.

[Additional Test]

Test Item Test Voltage (V) Frequency (Hz) Test Result

Power Factor: 277 60 0.922

Total Harmonic Distortion: 120 60 5.9%

Total Harmonic Distortion: 277 60 15.8%

Off State Power (W): 120 60 0.0

6. Product Photo

***********END OF REPORT***********