How to Get the Best Performance out of your CyTOF

Carina Torres

CyTOF User Group Meeting

CYTO 2013

23 May 2013

Page 2: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

Outline

• History• CyTOF Workflow

• Instrument Startup• Optimization• Running Samples• Instrument Shutdown• Troubleshooting

• Progress Report…1 ½ years later

Page 3: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

CyTOF at Lilly

2010 2011 2012 2013

A group of Lilly Senior Research Scientists went to a Garry Nolan talk

CyTOF installed, placed in Flow Core

SeptemberPurchased CyTOF (#9)

Panel ValidationPhenotyping Data analysis

Bendall, S.C., et al. Single-cell mass cytometry of differential immune and drug responses across a human hematopoietic continuum. Science 332 (6030): 687-696, 2011.

Experiments began

Page 4: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

Preparing your Lab

Page 5: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

Make sure Argon is on▪ 100 psi at wall▪ 50 psi on instrument

Check lights on instrument Turn Heater onInstrument setup Card Cage Tab Click Heater on 200oC

Start Syringe pump (HPLC)Start Plasma

Daily Workflow: Instrument Startup

Prepare Nebulizer and Syringe Pump

Page 6: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

Instrument Startup

TP1 TP2

VG1 VG2

BP

Argon

Ex Vent

CLOSE

Page 7: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

Make sure Argon is on▪ 100 psi at wall▪ 50 psi on instrument

Check lights on instrument Turn Heater onInstrument setup Card Cage Tab Click Heater on 200oC

Start Syringe pump (HPLC)Start Plasma

Instrument Startup

Prepare Nebulizer and Syringe Pump

Page 8: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

Instrument Startup

20 min

Page 9: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

Instrument Startup

Make sure Argon is on▪ 100 psi at wall▪ 50 psi on instrument

Check lights on instrument Turn Heater onInstrument setup Card Cage Tab Click Heater on 200oC

Prepare Nebulizer and Syringe Pump

Start PlasmaStart Syringe pump (HPLC)

Page 10: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

Instrument Startup

DI H20

Page 11: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

Instrument Startup

Make sure Argon is on▪ 100 psi at wall▪ 50 psi on instrument

Check lights on instrument Turn Heater onInstrument setup Card Cage Tab Click Heater on 200oC

Start PlasmaStart Syringe pump (HPLC)

Prepare Nebulizer and Syringe Pump

Page 12: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

Instrument Startup

• Chiller turns on• Start-up sequence• Plasma ignites• Chiller turns on• Start-up sequence• Plasma ignites

Page 13: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

Instrument Startup

Page 14: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

Instrument Startup

Make sure Argon is on▪ 100 psi at wall▪ 50 psi on instrument

Check lights on instrument Turn Heater onInstrument setup Card Cage Tab Click Heater on 200oC

Start PlasmaStart Syringe pump (HPLC)

Prepare Nebulizer and Syringe Pump

Page 15: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

Instrument Startup

Page 16: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

Optimization: Performance Check and Tuning

►To get the best resolution between detection channels

Mass Calibration ►Makes sure TOF values for CS and IR are <5% of peak height, calibration doesn’t typically need to be changed

►Optimize to ensure efficient detection of ions

►The dual count coefficient is calculated to generate the dual signal scale that is used to quantify ion signals

►Measure sensitivity by doing a dual counts tests to ensure 159Tb dual counts are >100,000

►Check Oxide ratio as an indication of plasma robustness. The ratio of 155Gd dual count /159Tb dual count should be less than 3%

Page 17: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

Optimization: Performance Check and Tuning

Tuning Solution

Page 18: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

=6

Peak height: x= 10425 (TOF) y= 525,000 (intensity)

½ max signal intensity

Example: 10425/6(2)= 868

Optimization: Performance Check and Tuning

Mass Resolutio

n

Tuning Solution

>400

Tb

Page 19: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

<5% of peak height <5% of peak height

Optimization: Performance Check and Tuning

Mass Resolutio

n

Mass Calibration

Tuning Solution

9537 11457

Cs Ir114579537

Page 20: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

Optimization: Performance Check and Tuning

Mass Resolutio

n

Mass Calibration

Detector gain

Tuning Solution

Max signal intensity Tb pulse counts

629242136495

= 4.61 (4-5)

TbAll

Page 21: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

All

Mass Resolutio

n

Mass Calibration

Detector gain

Tuning Solution

Optimization: Performance Check and Tuning

Dual CountCalibration

CYTOF is Ready!

Dual Count testOxidation

=<3%

=1.26%

155Gd dual count 159Tb dual count

2754218626

159Tb dual count >100,000

Page 22: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.
Page 23: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

Automated QC at Lilly

QC Table Pulse Dual Mass Mass PPM Counts Counts Intensity Error Error

Page 24: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

Automated QC at Lilly

QC Table Pulse Dual Mass Mass PPM Counts Counts Intensity Error Error

QC Table Pulse Dual Mass Mass PPM Counts Counts Intensity Error Error

Page 25: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

Automated QC at Lilly

Dual count calibration

Dual counts plot

Page 26: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

When Specs Are Not Met

Mass Calibration

Parameter Observation Possible Solution

<400 Call for service

159Tb ratio is<4 or >5 optimize detector voltage

ToF values for Cs and Ir at beginning of peak are >5% of peak height

Verify ToF values and enter new values in the mass calibration tab

Check X-Y alignment

Optimize gas flows

Optimize current

Swap nebulizer

Optimize gas flows

159Tb dual count <100,000

155Gd dual count159Tb dual count

>3%

Page 27: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

Mass Calibration

Parameter Observation Possible Solution

<400 Call for service

159Tb ratio is<4 or >5 optimize detector voltage

ToF values for Cs and Ir at beginning of peak are >5% of peak height

Verify ToF values and enter new values in the mass calibration tab

Check X-Y alignment

Optimize gas flows

Optimize current

Swap nebulizer

Optimize gas flows

159Tb dual count <100,000

155Gd dual count159Tb dual count

>3%

When Specs Are Not Met

Page 28: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

Mass Calibration

Parameter Observation Possible Solution

<400 Call for service

159Tb ratio is<4 or >5 optimize detector voltage

ToF values for Cs and Ir at beginning of peak are >5% of peak height

Verify ToF values and enter new values in the mass calibration tab

159Tb dual count <100,000

Check X-Y alignment

Optimize gas flows

Optimize current

Swap nebulizer

Optimize gas flows155Gd dual count159Tb dual count

>3%

When Specs Are Not Met

Page 29: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

IrCs

9533 11453

<5% of peak height <5% of peak height

Verify Mass Calibration

9537 11457

Page 30: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

Mass Calibration

Parameter Observation Possible Solution

<400 Call for service

159Tb ratio is<4 or >5 optimize detector voltage

ToF values for Cs and Ir at beginning of peak are >5% of peak height

Verify ToF values and enter new values in the mass calibration tab

Check X-Y alignment

Optimize gas flows

Optimize current

Swap nebulizer

Optimize gas flows

159Tb dual count <100,000

155Gd dual count159Tb dual count

>3%

When Specs Are Not Met

Page 31: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

.

. . . . . . . . .. . . . .

Optimize Detector Voltage

Page 32: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

Mass Calibration

Parameter Observation Possible Solution

<400 Call for service

159Tb ratio is<4 or >5 optimize detector voltage

ToF values for Cs and Ir at beginning of peak are >5% of peak height

Verify ToF values and enter new values in the mass calibration tab

Check X-Y alignment

Optimize gas flows

Optimize current

Swap nebulizer

Optimize gas flows

159Tb dual count <100,000

155Gd dual count159Tb dual count

>3%

When Specs Are Not Met

Page 33: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

Mass Calibration

Parameter Observation Possible Solution

<400 Call for service

159Tb ratio is<4 or >5 optimize detector voltage

ToF values for Cs and Ir at beginning of peak are >3% of peak height

Verify ToF values and enter new values in the mass calibration tab

Check X-Y alignment

Optimize gas flows

Optimize current

Swap nebulizer

Optimize gas flows

159Tb dual count <100,000

155Gd dual count159Tb dual count

>3%

When Specs Are Not Met

Page 34: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

X-Y Alignment of the Interface

All

Page 35: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

Optimize Gas Flows

All

Page 36: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

Optimize Current

All

Page 37: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

Re-Check QC

Mass Calibration

Dual Counts Plot

Page 38: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

• Make sure sample loop is clean: run H20, check Mass Cal screen to verify

Running Samples: User Guidelines

Page 39: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

Running Samples: User Guidelines

6003010

Page 40: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

• Filter samples and count cells. Resuspend cells to achieve collection rate

of ideally 300-500 events/sec

Running Samples: User Guidelines

• Clean with H20 between samples to minimize carryover• Clean instrument when done running samples (Wash solution and H20)

Page 41: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

Instrument Shutdown

Page 42: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

Instrument Shutdown

Page 43: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

Instrument Shutdown

• Turn off HPLC• Remove nebulizer and store in 3% Citranox

Page 44: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

Troubleshooting: Common Issues

1.

Cause SolutionProblem

Plasma won’t ignite

Argon off Turn it on

Argon not pure enough

Use Ultra High Purity Argon

Incorrect Gas? Double check gas

Air Leak Localize source

Plasma ignites

Plasma does not ignite

Look for air leak in the area of sample introduction

Check the torch, injector, coilfor condensation

Page 45: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

Troubleshooting: Common Issues

2.

ProblemNo/low signal

Cause

Sample line capillary pinched

Sample line capillary inserted improperly

Inspect capillary line

Reinsert capillary

Clogged nebulizer Check spray/swapnebulizer

Insufficient gas flows

Check gas flows

Solution

Page 46: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

• Pankaj came out and found the leak. Replaced torch and replaced RFG coil

Troubleshooting: Unusual Circumstances

• Leak in one of the argon lines

1.

Cause SolutionProblem

• Clean/Replace coil

• Arcing because

the coil not aligned

• Oxidation on the coils

Page 47: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

Troubleshooting: Unusual Circumstances

Poor contact of skimmer/reducer with the contact ring on the interface. This may be due to oxidation on the contact ring or on the reducer itself.

Vigorously cleaned the contact ring on the interface with methanol. Eventually, Daniel came out to replace the interface

2.

Cause SolutionProblem

Page 48: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

Troubleshooting: Unusual Circumstances

Originally thought it was a faulty syringe pump. Turned out to be a bad valve.

3. Replaced the entire valve assembly

Cause SolutionProblem

Page 49: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

Clean/swap nebulizer

Troubleshooting: Unusual Circumstances

Cause SolutionProblem

4. Partially clogged/dirty nebulizer

Sample capillarynot positioned properly

Reinsertcapillary

Bad syringe Replace syringe

Page 50: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

180-190

Tungsten contamination; but from where?

Troubleshooting: Unusual Circumstances

Cleaned ignitor pin with sandpaper

Cause SolutionProblem

5.

180-190

Page 51: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

Progress Report: 1 ½ years later

What have we done so far?

>100 expts: 60% validations & optimizations 40% phenotyping

Validated core panel: 16 parameters

*Replaced Rhodium with Cisplatin for

viability

Joined Cytobank; Transitioning from FlowJo to Spade for data analysis

ICP-MS characterization

*Reference: Feinberg, H.G., et al. A platinum-based covalent viability reagent for single-cell mass cytometry. Cytometry A 81 (6): 467-475, 2012

Lilly CyTOF Projects:

• Multi-parametric, high content immunophenotyping

• LRAP with the Sanford Burnham Institute to investigate signaling pathways

Page 52: How to Get the Best Performance out of your CyTOF Carina Torres CyTOF User Group Meeting CYTO 2013 23 May 2013.

Acknowledgements

Elaine ConnerMarian MastrangeloHan He

Andrew Glasebrook DVS SciencesPankaj ChaudhariLeslie FungTed YoungTad GeorgeRon CohnDaniel Tong

Mark DanielsJohn Fitchett