Hardware Design and Implementation of an Electrocardiogram System
-
Upload
sarah-alharthey -
Category
Documents
-
view
217 -
download
0
description
Transcript of Hardware Design and Implementation of an Electrocardiogram System
By: Karima Khandaker, Israa Al-Qassas, Reemaz Hetaimish, Haneen Mohammed, Danah Nour, Lina Babsail, Sarah Al- Harthey Course: Introduction to Electronic (Integrated Circuits) - ECE210L Supervisor: Dr. Nema Salem Department: Electrical and Computer Engineering
HARDWARE DESIGN AND IMPLEMENTAT ION OF AN ELECTROCARDIOGRAM SYSTEM
• What is an ECG system ? According to the Encyclopaedia of the 20th Century Technology, "the
electrocardiogram (ECG or EKG) is a graphic measurement of the electrical activity of the heart produced by an electrocardiograph, or ECG machine." The ECG is used to monitor the electrical activity of the heart which is extracted from the human body in the form of analog signals.
• Why to build an ECG system ? to represent efficient design and implementation of a complete ECG
system starting from the signal acquisition from the human body till displaying both of the ECG signals and the rate of heart beats as well.
• What does the ECG do ?
The Electrocardiogram (ECG or EKG) is used to measure the electrical activity of the heart over a certain period of time depending on the purpose of taking the reading. It can be used to either monitor over longer periods of time or diagnostic purposes to analyze the status of the heart at a certain time.
ELECTROCARDIOGRAM “ECG”
Fig. 1.0 Standard Setup of ECG System Used for Monitoring
• With each heartbeat, an electrical signal is generated in a group of cells in the upper right chamber or right atrium of the heart. This signal passes through the two upper chambers (right and left atria) causing them to contract which in turn causes the blood to move in the lower chambers of the heart called ventricles. This electrical signal passing through the atria is shown as the P Wave on the ECG. As the signal passes from the atria to the ventricles it slows down to allow the ventricles to fill up with blood and this is shown as the flat region between the P and Q waves. The signal then distributes across the ventricles instigating them to contract and pump blood to the various parts of the body. This section causes the QRS waves on the ECG. Once the ventricles return to their normal state after this, it can be seen on the ECG as the T wave in Fig 2.0. The ECG allows the medical staff to deduce different conditions based on the deflection from the general expected ECG waveform. It may be based on the amplitude, shape, interval lengths or even the absence of expected waves. Table 2.0 shows some examples of comparisons between normal (expected) data and abnormal data and the conclusions that can be drawn in each case.
• 3 lead wire ECG System have three colored wires that connect to three electrodes forming the Einthoven’s triangle as shown in Fig. 3.0. This system provides three views of the heart.
HEART BEATS : PQRST WAVE ON THE ECG
Fig. 2.0 PQRST Wave
Fig. 3.0 Position Of 3 Leads Forming Einthoven’s Triangle
HARDWAR IMPLEMENTATION
Signal Extraction
Amplification System
Noise Filter System
Peak Detector System
Displays and Outputs
The overall system of the ECG can be divided into the shown 5 subsystems: Signal Extraction
• Electrodes • Patient Protection • Multiplexer
Further Amplification Noise Filter
• Amplification • Instrumentation Amplifier
Peak Detector
• To calculate heart rate
Output and Displays • From Peak Detector Circuit - Seven Segment Displays • ECG display – Oscilloscope
Fig. 2.0 Overall ECG System Block Diagram
Signal Extraction
Amplification System
Noise Filter System
Peak Detector System
Displays and Outputs
Signals are extracted from lead wire and connected to
the multiplexer that is used to select the lead to be
processed is a Dual 4-Line To 1-Line Data Selector/
Multiplexer.
The extracted signal from the human body cannot be
fed to the circuit because it is very weak (around 0.5 mV).
For this reason an amplifier should be used to increase
the amplitude of the signal
Filter the noise from the amplified signal , the band pass filters were selected to cancel the noise that surrounds that signal and any value that is close to zero, which will be considered a DC signal. Furthermore, the filter will select a specific range in a signal, which starts from 0.5Hz up to a maximum value of 175 Hz . The filter circuit has been designed based on specific values of bandwidths as shown below.
Count the most significant peak using a comparator which compares a signal voltage applied at one input of an op-amp with a known reference voltage at the other
input . The next step would be to count the number of square pulses by giving the square waveform as an input
to the 4-bit synchronous counter . The output from the 4-bit synchronous counter which represented the number of heart pulses in 10 seconds
The output viewed through oscilloscope and 3 digits of 7
segment display where it's known that the maximum
heart rate might exceed 200 which is 3 digits numbers .
[1] National Heart, Lung and Blood Institute. (2012). Understanding the Heart’s Electrical System and EKG Results. Retrieved from
http://www.nhlbi.nih.gov/health/health-topics/topics/hb/understanding.html
[2] STMicroelectronics. (2001). General Purpose Single Operational Amplifier Datasheet. Retrieved from http://www.datasheetcatalog.org/datasheet/stmicroelectronics/5304.pdf
[3] LG Semicon Co. Ltd. (2013). Dual 4-line to 1-line Data Selectors/Multiplexers. Retrieved from
http://www.datasheetarchive.com/GD74LS153-datasheet.html
[4] Texas Instruments. (2011). ADS1258 24 Bit, Fast Channel Cycling Delta-Sigma ADC. Retrieved from
http://www.ti.com/product/ads1258
[5] Nursecom Educational Technologies. (2004). An ECG Primer.
[6] Donald, N. (2010). Microelectronics: Circuit Analysis and Design (4th ed.). New York: McGraw-Hill.
[7] Paulus, D., Meier, T. (2009). ECG Amplifier. Retrieved from
http://www.mayr.informatik.tu-uenchen.de/konferenzen/MB-Jass09/courses/1/Paulus.pdf
[8] Texas Instruments Incorporated. (2005). Getting The Most out of Your Instrumentation Amplifier Design. Retrieved from http://www.ti.com/lit/an/slyt226/slyt226.pdf
[9] Texas Instruments (1998). LM741 Operational Amplifier. Retrieved from
http://www.ti.com/lit/ds/symlink/lm741.pdf
[10] Townsend, N. (2001). Medical Electronics. Retrieved from
www.robots.ox.ac.uk/~neil/teaching/med_elecwww.robots.ox.ac.uk/~neil/teaching/med_eleca
[11] Clayton, G., & Winder, S. (2003). Operational Amplifiers. (5th ed.). England: Newens.
[12] Priemer, R. (1991). Introductory Signal Processing. World Scientific.
[13] Roy, C. D., & Jain, S. (2003). Comparators and Waveforms Generators. In Linear integrated circuits. New Delhi: New Age International.
[14] Educational Publishing for Design & Technology. (2013). 4511 BCD to 7-segment Decoder. Doctronics. Retrieved from http://www.doctronics.co.uk/4511.htm
[15] Computer-Related Circuit. (2013). 8-bit Binary to 3-digit Decimal Display Decoder For 8-bit Microprocessor Systems With 256 by 4 PROM three 7-segment Decoder Drivers With 9374 Input Latches and Two Gates. Retrieved from
http://www.seekic.com/circuit_diagram/Computer-Related_Circuit/8_bit_binary_to_3_digit_decimal_display_decoder_for_8_bit_microprocessor_systems_with_256_by_4_PROM_three_7_segment_decoder_drivers_with_9374_input_latches_and_two_gates.html
[16] Tech. Philips Semiconductors. (1995). BCD to 7-segment Latch/decoder/driver. Retrieved from
http://www.uib.es/depart/dfs/GTE/staff/jfont/InstrETT/hef4511b.pdf
[17] Vishay Semiconductors. (2013). Standard 7-Segment Display 13 Mm. Tech. no. 83126. Retrieved from http://www.vishay.com/docs/83126/83126.pdf
REFERENCES