Operating Manual SAP No. FDK-521H1200 SFIDK.PS.029.D1.02 INFOLOG Hand-held terminal type PSION for SITRANS F US ultrasonic flowmeter type SONOCELL and SONO 3301 [ ] *085R9464* s

Transcript of Hand-held terminal type PSION - Siemens AG Siemens data transfer system consists of 6 parts. ... The...

Page 1: Hand-held terminal type PSION - Siemens AG Siemens data transfer system consists of 6 parts. ... The hand-held terminal type PSION ... nn--dd/mm/yyyy this will appear in the display.

Operating Manual

SAP No. FDK-521H1200

SFIDK.PS.029.D1.02

INFOLOGHand-held terminal type PSION

for SITRANS F US ultrasonic flowmetertype SONOCELL and SONO 3301

[ ]

*085R9464*

s

Page 2: Hand-held terminal type PSION - Siemens AG Siemens data transfer system consists of 6 parts. ... The hand-held terminal type PSION ... nn--dd/mm/yyyy this will appear in the display.

INFOLOG

2 SFIDK.PS.029.D1.02

INFOLOG

3SFIDK.PS.029.D1.02

Contents

The Siemens data transfer system consists of 6 parts. The INFOLOG unit, an infrared optical head, solid state disk (flash) containing the software program for INFOLOG, a CD with soft-ware program for data transfer from INFOLOG to PC, a 9 pol. 0-modem cable and a copy of the Operating Manual.The hand-held terminal type PSION (Workabout MX) ("INFOLOG"), is used to read-out the actual values, account values or monthly data of Siemens flow measuring units of the SITRANS F US ultrasonic flowmeter types SONOCELL and SONO 3301 ("SITRANS F US 105") and transfer data into e.g. an Excel spread sheet for further data processing.

The read-out can be made by use of infrared optical head (order No. FDK-087H0108).

In addition to read-outs, the INFOLOG can be used for changing customer numbers, SONOCELLand SONO 3301 account date and read-out/reset of error codes.

1.1 System requirements • Windows 98, Windows 2000 and Windows XP• 5 MB after installationAs a standard the SONOCELL Data Transfer Programme (SDTP) will be distributed on CD with install shield based installation package and the software for INFOLOG will be distributed on static flash.

1. Introduction

1. Introduction ....................................................................................................................... 31.1 System requirements.......................................................................................................... 3

2. Quick start ........................................................................................................................ 42.1 Collect data ........................................................................................................................ 42.2 View data ........................................................................................................................... 42.3 Transfer data to PC ........................................................................................................... 52.4 Set-up mode....................................................................................................................... 6

3. Service menu ................................................................................................................... 73.1 Error status ......................................................................................................................... 73.2 Customer number............................................................................................................... 83.3 Account day........................................................................................................................ 83.4 SITRANS F US SONOFLO type SONOCELL and SONO 3301 time ............................... 83.5 SITRANS F US SONOFLO type SONOCELL and SONO 3301 date ............................... 8

4. Installing the data transfer software on your PC ......... ...................................................... 94.1 Running the data transfer software program. ..................................................................... 94.2 Storage of data ................................................................................................................... 94.3 Update software ................................................................................................................ 94.4 Uninstalling ......................................................................................................................... 11

5. Trouble shooting ............................................................................................................... 115.1 No communication between PC and INFOLOG ................................................................ 115.2 Program re-load/re-start ..................................................................................................... 115.3 Data.csv file in Microsoft Excel .......................................................................................... 11

6. Read-out examples of SITRANS F US SONOFLO type SONOCELL and SONO 3301 .. 116.1 SITRANS F US SONOFLO type SONOCELL and SONO 3301 actual data (from Excel) ...... 11

1. Introduction

Navigate arrows

Collect data

Transfer data

See data

Set-up mode

Viev available menu

Page 3: Hand-held terminal type PSION - Siemens AG Siemens data transfer system consists of 6 parts. ... The hand-held terminal type PSION ... nn--dd/mm/yyyy this will appear in the display.

INFOLOG

2 SFIDK.PS.029.D1.02

INFOLOG

3SFIDK.PS.029.D1.02

Connect the optical eye to SONOCELL or SONO 3301. Pressing starts data collection - and a bar code indicates the status.

A small beep indicates the INFOLOG has received data successfully.Costumer number and total accumulated volume are then displayed - indicating that INFOLOG has stored data successfully.

The INFOLOG can store up to 1500 readings. View data (F2) may be slow due to the high data capacity when approx. 400 readings have been made. Note Users in F1 mode are only allowed to read accumulated values as preset under the menu F4. As default the INFOLOG is set to data collect and actual data.

2. Quick start When you receive a complete set from Siemens the unit is almost ready for use. In order to protect batteries from leaking into the INFOLOG at storage, the 2 batteries and the lithium cell are added separately.

1. Open the slot by pressing the tab2. Insert the 2 x 1.5 V penlight batteries in the INFOLOG and insert the back-up lithium cell3. Insert the flash solid state disk in "drive A"

The software program on the flash now has to be copied to the INFOLOG.

1. Switch on the INFOLOG by pressing

2.

The program will start installing the software on your terminal.When asked:

3. "Stop link?" Press and wait for the program to continue4. "Create M\OVD" Press and wait for the program to start 5. The program is now installed on your INFOLOG and the display shows

SHIFT DEL On/Esc+ +

To get a general understanding of the INFOLOG functionality review the INFOLOG menu structure at Appendix A.

2.1 Collect data F1

F1

2.2 View data is used to view data.

By using the following arrows it is possible to scroll through the collected data. As the values are relatively small in the display, it is possible to enlarge the digits by pressing whenever the digits are highlighted.

F2

F2

2. Quick start

YES

YES

ENTER

Page 4: Hand-held terminal type PSION - Siemens AG Siemens data transfer system consists of 6 parts. ... The hand-held terminal type PSION ... nn--dd/mm/yyyy this will appear in the display.

INFOLOG

4 SFIDK.PS.029.D1.02

INFOLOG

5SFIDK.PS.029.D1.02

2.3 Transfer data to PC F3

To export the data from the INFOLOG to the PC the software programme (SDTP.exe) should be installed.

Install the software program as described in chapter 4.

The cable supplied with INFOLOG is a 0-modem cable.

Start the program and ensure the INFOLOG is in "F3" mode in order to establish communication to the PC.

Activate "Download data from INFOLOG" and data will now be transferred. (If this was the first time to use the program you may have to activate the same function again).

A red wire between the INFOLOG and the PC icon indicates that connection is established. Data will normally be stored on your PC under folder: (Installation directory) doc\data.csv.

NoteTo save disk space on INFOLOG, users should delete the old data file after transferring it to PC. A new file will be created for new data. Before starting the deleted operation, users are asked to enter the INFOLOG password. This will limit the use of this operation to permitted users.

2. Quick start

Page 5: Hand-held terminal type PSION - Siemens AG Siemens data transfer system consists of 6 parts. ... The hand-held terminal type PSION ... nn--dd/mm/yyyy this will appear in the display.

INFOLOG

4 SFIDK.PS.029.D1.02

INFOLOG

5SFIDK.PS.029.D1.02

ApplicationSelect This menu offers the following choices:• DataCollect Mode - the INFOLOG is used only as a datalogger, without any possibilities for set-up of data.• ServiceMode - opens for change/views of: 1. Read-out of actual values, account values, monthly values (previous 24 months). 2. Error code, customer number, account date and SONOCELL and SONO 3301 (SITRANS F US 105) date and time adjust. See chapter 3 for further information regarding choices in service mode.

DataSelect This menu offers the following choices:• Actual data: This data includes: Customer number, serial no., total volume, hour counter, peak flow in the actual month, SONOCELL and SONO 3301 - date and time, error code. Date/time of INFOLOG. • Account A data: The same types of data as above but this data represents the accumulated values at preset read off date.

CommunicationThe menu contains 3 choices:• Data reading port: Is set by default "A".• Communication rate: Is set by default to 300. SONOCELL and SONO 3301 supports only 300 baud whenever communication is via the optical eye.• Data transfer to PC port: Must be set to "A" (If used with dokingsstation the set port to "C").

ChangePasswordThis provides dialog to change the password of INFOLOG.

Time date adjustThis provides the functionality to adjust date and time of the INFOLOG.

SW VersionDisplays the software version in the INFOLOG.

2. Quick start

2.4 Set-up modeF4

F4

This mode is password protected.

gives you the following possibilities:

Page 6: Hand-held terminal type PSION - Siemens AG Siemens data transfer system consists of 6 parts. ... The hand-held terminal type PSION ... nn--dd/mm/yyyy this will appear in the display.

INFOLOG

6 SFIDK.PS.029.D1.02

INFOLOG

7SFIDK.PS.029.D1.02

To set the INFOLOG into the service mode follow the following steps: (refer to appendix A). 1. Press .

2. A password dialog will be displayed. Enter INFOLOG password then press .

3. Press and a menu-bar will be displayed as shown in "Set-up" paragraph.

4. Highlight the "Application Set-up" and press .

5. In "Application Select" dialog select "Server mode" by using the arrow keys and press . Now the INFOLOG is enabled to be set in service mode.

6. Press to enter the service mode (Shown in the next 2 figures)

3. Service Menu

F4

Menu

In general readings and changing settings will take a 30 seconds.

In the bottom of the display a status line indicates the actual commands when using the service menu.

Use arrows and "del" and keypads to enter new settings.

NoteTo erase digits under the item write, move the curser using the arrows so that the curser is flashing behind the digit. Use del to change or erase digits.

3.1 Error status EnterSelect menu

READ reads the actual error codeRESET resets all error codesESC escape from error status

F: 00000 represents no failure.

The failure code is represented with F3 and F5 depending on failure type.F3: Internal faultF5: Volume flow exceeds the preset Qs value.

3. Service menu

Enter

Enter

Enter

F1

In service mode the INFOLOG is able to read different type of data. Use arrows to select required data type:• Actual readings• Account A• Month-1, months-2, months-3……..months-24.

Select to enter into service menu Use the arrows to scroll up/down in menu.

Menu

Page 7: Hand-held terminal type PSION - Siemens AG Siemens data transfer system consists of 6 parts. ... The hand-held terminal type PSION ... nn--dd/mm/yyyy this will appear in the display.

INFOLOG

6 SFIDK.PS.029.D1.02

INFOLOG

7SFIDK.PS.029.D1.02

3.2 Customer number EnterSelect menu

READ reads the customer numberWRITE type new customer number ESC escape from customer number dialog

SONOCELL and SONO 3301: Password is default 1000.

The customer number as default is as v, ww, y, ssss, where "v" represents the software version number, "ww" represents the week of production, "y" represents the production year and "ssss" represents the 4 digits in the serial number.

Example: Customer number: 125300021 = software version25 = production week3 = year 20030002 = represents serial number

Maximum customer number length: 8 digits

Whenever changing the SITRANS F US 105 customer number you are prompted to insert the password.

3.3 Account day EnterSelect menu

READ reads the account dateWRITE enter new day/months (alteration of account date requires SONOCELL and SONO 3301 passwords) ESC escape from account day dialog

3.4 SITRANS F US 105time

EnterSelect menu/SITRANS F US 105 time

READ reads the actual time in hours and minutes WRITE type new hour: minutes (hh:nn) ESC escape from SITRANS F US 105 time dialog

NoteSONOCELL and SONO 3301 uses the actual time and date to store any failures. It is recommended that the real time clock is set correctly.If INFOLOG is set correctly with hh::nn--dd/mm/yyyy this will appear in the display. (See chapter 2.4).

3.5 SITRANS F US 105date

EnterSelect menu/SITRANS F US 105 date READ reads the actual day/month/year WRITE type new date (day/month/year) (dd.mm.yy)ESC escape from SONOCELL and SONO 3301 date

NoteSONOCELL and SONO 3301 uses the actual time/date to store any failures. It is recommended the real time clock is set correctly.If INFOLOG is set correctly with time/minutes/day/months/year this will appear in the display. (See chapter 2.4).

3. Service menu

Page 8: Hand-held terminal type PSION - Siemens AG Siemens data transfer system consists of 6 parts. ... The hand-held terminal type PSION ... nn--dd/mm/yyyy this will appear in the display.

INFOLOG

8 SFIDK.PS.029.D1.02

INFOLOG

9SFIDK.PS.029.D1.02

4. Installing the data transfer software on your PC

• Insert CD in your computer and this will auto-start the installation program • Follow the instruction on your screen

This will install "SDTP" program on your computer and all necessary subdirectories. It will also create 2 short-cuts: one on your program files and one on the desktop. Double click this short-cut to run the SONOCELL data transfer program. The created directories are:

• "Data" directory: Holds the data transfer from INFOLOG• "Doc" directory: License.txt and infolist.txt files• "HTSW" directory: Includes the upgrade of INFOLOG software

4.1 Running the data transfer software program

The easiest way to start the program is to double click the short-cut made by the installation pro-gram on your desktop or on the program menu (start\programs\INFOLOG\SDTP). A connection indicator will appear on the right corner on your screen (Tool menu) trying to connect. Connection is made when this indicator shows a sine wave.To exit the program press quit. The sine wave will be a horizontal line indicating that the program is ready to reconnect when it is reactivated. Right click the icon indicator to disconnect, reconnect or terminate the link between the hand terminal and PC.

Press the "Download Data from INFOLOG" button to copy data from INFOLOG to PC. If the connection is not already established, it will be established and the data file will be transferred. In case of a connection error the function will terminate and the button should be repressed when the connection is established. If the file exists you can choose to rename the file, overwrite the old copy or cancel the operation.

4.2 Storage of data Data will normally be stored under the (installation)folder: doc\data.csv.

4.3 Update software It is possible to download the latest new SONOCELL data transfer program via our homepage www.siemens.com/flow

First you have downloaded the SONOCELL data transfer program on your PC. It is possible then to update/install the INFOLOG software program.

Install the software on your INFOLOG from the SONOCELL data transfer program: Press "download program for INFOLOG". The program will give you the following message to guide you through the software update:

4. Installing the data transfer software on your PC (SDTP)

Page 9: Hand-held terminal type PSION - Siemens AG Siemens data transfer system consists of 6 parts. ... The hand-held terminal type PSION ... nn--dd/mm/yyyy this will appear in the display.

INFOLOG

8 SFIDK.PS.029.D1.02

INFOLOG

9SFIDK.PS.029.D1.02

1. Update software: Be sure that INFOLOG is in F3 mode and continue the installation by pressning on .

2. First time installation (no software program in the INFOLOG (Virgin unit)). This means that is not active, and the INFOLOG has to be set-up into a "Link"

mode where it is possible to communicate with the PC.

Set-up of INFOLOG in "Link" mode:

1. Push to start PSION device

2. Press and select command processor and press

3. M>\ appears in the display

4. Press and use to step to the menu "Special"

5. Select remote link followed by

6. By use of select "On" and "port A" followed by

7. INFOLOG is now ready to receive the software program from the PC.

4.3 Update software (continued)

4. Installing the data transfer software on your PC

Step 2 of 4

Step 3 of 4

Select download software for INFOLOG

Step 4 of 4

Ok

F3

On/Esc

Menu

Menu

Enter

Enter

Enter

Step 1 of 4

Page 10: Hand-held terminal type PSION - Siemens AG Siemens data transfer system consists of 6 parts. ... The hand-held terminal type PSION ... nn--dd/mm/yyyy this will appear in the display.

INFOLOG

10 SFIDK.PS.029.D1.02

INFOLOG

11SFIDK.PS.029.D1.02

5. Trouble shooting

5.1 No communication between PC and INFOLOG

• Ensure that the data transfer cable is connected between the PC and INFOLOG• Ensure that INFOLOG is in F3 mode• Ensure that the cable is a 0-modem cable• The connection between INFOLOG and PC is indicated by the redline between their icons • Set link to the connected port A (in F4 menu)

Following the instruction above complete the update of the INFOLOG software and the following picture will be present:

• Uninstalling the program ♦ Closing the program and related programs ♦ From control panel\removeSW\uninstall the program ♦ Answer yes to remove the unused dll’s, and check the box to avoid the same message

from re-appearing ♦ It may be necessary to manually remove some of the copied files

4.4 Uninstalling

5.2 Program re-load/re-start

The software in the INFOLOG is suitable to all kind of input. If you need to re-start the program carry out the following to preserve data:• Software reset (shift+del+on), this will re-load the program and run the exe file• Re-install the program (see chapter 4.3)

5.3 Data.csv file in Microsoft Excel

5. Trouble shooting

4.3 Update software (continued)

4. Installing the data transfer software on your PC

Note: Data format is the same as above, date/time of SONOCELL and SONO 3301 is the time of the preset account date.

6. Readout examples of SONOCELL and SONO 3301

6.1 SONOCELL and SONO 3301 actual data (from Excel)

Cust. No. Serial No. Total volume

Unit Peak0 Hour counter

Unit INFOCAL date

INFOCAL time

ERROR_Status

IN-FOLOG_

Date

IN-FOLOG_

Time

1 0312 233.45 MUSG 345.45 213 h 3.03.2003 12:14 F:00000 25.01.01 12:21

6. Readout examples of SONOCELL and SONO 3301

If the format of the data in the Excel sheet is not displayed correctly, try the following in Excel:• Mark the first column and press text to columns in data menu• Ensure delimited is marked in the convert text to columns wizard step 1 of 3 and press • Insure "Comma" check box is marked in step 2 and press Mark column "General\Total\Volume" and press on Set decimal separator to "." and thousands separator to "," and press Press To align all data to the left do the following:• Mark all the data by pressing on the left upper corner• Click the mouse right button and chose format cells from pop up menu• Chose in the chose left in the horizontal box

Note: The data is transferred in ASCII format with "." as decimal separator.

Next

OK

Advanced

Next

Alignment Text alignment

Finish

Page 11: Hand-held terminal type PSION - Siemens AG Siemens data transfer system consists of 6 parts. ... The hand-held terminal type PSION ... nn--dd/mm/yyyy this will appear in the display.

INFOLOG

10 SFIDK.PS.029.D1.02

INFOLOG

11SFIDK.PS.029.D1.02

Appendix A: INFOLOG Menu structure

Page 12: Hand-held terminal type PSION - Siemens AG Siemens data transfer system consists of 6 parts. ... The hand-held terminal type PSION ... nn--dd/mm/yyyy this will appear in the display.

We have checked the contents of this manual for agreement with the hardware and software described. Since deviations cannot be precluded entirely, we cannot guaran-tee full agreement. However, the data in this manual are reviewed regularly and any necessary corrections included in subsequent editions. Suggestions for improvement are always welcomed.

Technical data subject to change without prior notice.

The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model or design, are reserved.

Copyright © Siemens AG 01.2005 All Rights Reserved