HAMMERTECH - Fusion Orthopedics

14
Surgical Technique HAMMERTECH Fixation System

Transcript of HAMMERTECH - Fusion Orthopedics

Page 1: HAMMERTECH - Fusion Orthopedics

Surgical Technique

HAMMERTECHFixation System

Page 2: HAMMERTECH - Fusion Orthopedics

HammerTech Fixation System

DESCRIPTION OF THE MEDICAL DEVICE The HammerTech Fixation System implant is a PEEK (Polyetheretherketone (ASTM F2026)) threaded bone implant intended for fixation of the proximal interphalangeal (PIP) joint of the lesser toes. The device is offered as a straight cannulated implant in five different sizes. The specialized instruments are made primarily of surgical grade stainless steel (ASTM F899) and polyphenylsulfone (ASTM D6394). The implant and associated instrumentation is supplied sterile.

1

HammerTech Fixation System Surgical Technique

INDICATIONS FOR USE The HammerTech device is indicated for the fixation of osteotomies and reconstruction of the lesser toes following correction procedures for hammertoe, claw toe and mallet toe.

CONTRAINDICATIONS •Localorsystemic,acuteorchronicinflammation; •Activeinfectionorinflammation; •Growingpatientswithopenepiphyses.

Page 3: HAMMERTECH - Fusion Orthopedics

DIRECTIONS FOR USE To implant the HammerTech device, use only the specialized HammerTech instrumentation. Do not use implants or instruments from any other system or manufacturer. Before using the HammerTech System, the surgeon should be thoroughly familiar with the HammerTech System Surgical Technique Manual. Pre-Operative planning by the surgeon should determine the size of implant required and an adequate supply of the implant sizes should be available prior to surgery, including larger and smaller sizes than those expected to be used.

HammerTech Fixation System

2

HammerTech Fixation System Surgical Technique

Size

PetiteExtra SmallSmallMediumLarge

BlackGoldPurpleGreenBlue

2.4mm2.4mm2.75mm2.95mm3.25mm

Implant Kit Color Drill

K-Wire

1.0mm1.0mm1.4mm1.4mm1.6mm

3.5

5.2

6.0

Petite

3.7

6.5

6.5

Extra Small

4.1

6.7

6.8

Small

4.3

7.1

7.4

Medium

7.2

7.8

Large

4.6

Page 4: HAMMERTECH - Fusion Orthopedics

System Basics

3

HammerTech Fixation System

One Simple Sterile System

Implant5 Sizes Available

Cannulated DriverPrecision Fitment

K-Wires 1.1mm, 1.4mm and 1.6 mm Double-Trocar

Handle Ergonomic Design

Cannulated Drill Bit2.2mm, 2.5mm, 2.7mm & 3.0 Hard Stop

AccessoriesPackaged Separately

AwlDesigned for joint

preparation

Removal ToolFor easy and

efficient dislodge of implant

Sizer WheelTo determine

optimal Implantfitting

Implant SinglesAdditional

implants packaged separately

HammerTech Fixation System Surgical Technique

Page 5: HAMMERTECH - Fusion Orthopedics

After creating an incision over the dorsal aspect of the PIP joint, perform soft tissue releases as necessary. Resect the required amount of the distal aspect of the proximal phalanx.

A template may be used intra-operatively to determine the optimal implant size. The length and width of the template correspond to the length and major diameter of the proximal barbed end of the implant. The implant should be sized to the proximal phalanx for proper fitment.

1

2

HammerTech Fixation System Cannulated Straight Technique

4

HammerTech Fixation System Surgical Technique

Page 6: HAMMERTECH - Fusion Orthopedics

Denude the articular cartilage of the middle phalanx. The HammerTech Awl may be used to denude if desired. During joint preparation, take care to avoid excessive resection, which may result in a shortened digit.

Place the K-wire into the proximal phalanx along its central axis, approximately 10mm. Verify proper positioning with AP and lateral fluoroscopicviews.

3

4

HammerTech Fixation System Cannulated Straight Technique

5

HammerTech Fixation System Surgical Technique

Use of the K-wire for placement and fixation may be according to surgeon preference.

Page 7: HAMMERTECH - Fusion Orthopedics

Drill over the K-wire and advance until the positive stop on the cannulated drill is reached. This will create space for the barbed end of the implant. Drilling by hand is recommended although power may be used if desired.

Insert the K-wire into the central axis of the middle phalanx. After verifyingproperpositioningwithAPandlateralfluoroscopicviews, continue to drive the K-wire distally through both the middle and distal phalanx until it exits the toe. Retrograde the K-wire until the laser mark is burried within the middle phalanx and no longer visible, leaving 3mm to guide the implant.

5

6

Use caution and avoid overdrilling if power is utilized.

HammerTech Fixation System Cannulated Straight Technique

6

HammerTech Fixation System Surgical Technique

Page 8: HAMMERTECH - Fusion Orthopedics

Advance the cannulated drill over the K-wire to prepare the middle phalanx for the threaded side of the implant. Advance until the positive stop on the drill bit is reached.

Place the barbed segment of the implant into the driver and ensure that the driver is in the dorsal position. Screw the distal threaded portion into the middle phalanx, sliding the implant over the K-wire as a guide. Advance the implant until the threaded portion of the implant is buried within the middle phalanx leaving the unthreaded neck of the implant at the bone-to-bone interface.

7

8

Advance the Implant only if the entire barbed segment of the implant isFULLYSEATEDinsidethe driver. (Never advance the implant partially seated utilizing the end of the implant wings for torque.)

HammerTech Fixation System Cannulated Straight Technique

7

HammerTech Fixation System Surgical Technique

Page 9: HAMMERTECH - Fusion Orthopedics

Withdraw the K-wire distally until the black laser mark on the wire is no longer visible. Insert the barbed segment of the implant into the pre-drilled hole in the proximal phalanx. Firmly compress the joint until the resected surfaces of the joint are fully opposed. The distal end of the Handle can be used over the K-wire as a pusher to help reduce the joint.

Confirmunderfluoroscopyforadequatejointreductionandfinal placement. The K-wire is not indicated for permanent implantation (according to surgeon preference, not to exceed 30 days). Postoperative protocol is the responsibility of the medical professional.

9

10

It is recommended to advance the K-wire through the barbed segment of the implant, as this will enhance implant stability in the proximal phalanx. The K-wire can then be advanced through the metatarsophalangeal joint for fixation during the initial recovery period to prevent joint subluxation.

HammerTech Fixation System Cannulated Straight Technique

8

HammerTech Fixation System Surgical Technique

Page 10: HAMMERTECH - Fusion Orthopedics

Using a small power saw, cut the implant in half at the osteotomy site.

Attach the removal tool to a quick connect AO driver handle.

1

2

HammerTech Fixation System Removal Technique

9

HammerTech Fixation System Surgical Technique

Page 11: HAMMERTECH - Fusion Orthopedics

Advance the removal tool into the distal portion of the implant using a counter clockwise rotation to engage the reverse thread with the PEEK implant. After the threads have engaged with the implant, continue rotating counter-clockwise until the implant is removed.

The proximal portion of the implant can be removed with the same technique. Continue rotating the implant counter clockwise until the implant is removed.

3

4

HammerTech Fixation System Removal Technique

10

HammerTech Fixation System Surgical Technique

Page 12: HAMMERTECH - Fusion Orthopedics

HammerTech Fixation System Implants & Instruments

11

HammerTech Fixation System Surgical Technique

3.5

5.2

6.0

Petite

3.7

6.5

6.5

Extra Small

4.1

6.7

6.8

Small

4.3

7.1

7.4

Medium

7.2

7.8

Large

4.6

Description Sterile

Petite ImplantExtra Small ImplantSmall ImplantMedium ImplantLargeImplant

HT-10-1110-SHT-10-1120-SHT-10-1130-SHT-10-1140-SHT-10-1150-S

Implants

HT-10-1110HT-10-1120HT-10-1130HT-10-1140HT-10-1150

Non-Sterile

Description Sterile

Removal Tool HTAwl HTHT Sizer Wheel K-Wire Double Trocar 1.0mmK-Wire Double Trocar 1.4mmK-Wire Double Trocar 1.6mm HT Handle Reusable Driver HT Petite Driver HT Extra Small Driver HT Small Driver HT Medium DriverHTLarge Drill Cannulated HT 2.4mmDrill Cannulated HT 2.7mmDrill Cannulated HT 2.9mmDrill Cannulated HT 3.2mm

CD-09-6400-SCD-09-6200-STN-05-7010-S

-------------

Instruments

CD-09-6400CD-09-6200TN-05-7010HT-KW-4010 HT-KW-4014HT-KW-4016TN-TM-7000 TN-03-1110 TN-03-1120 TN-03-1130 TN-03-1140 TN-03-1150 CD-01-6024CD-01-6027CD-01-6029CD-01-6032

Non-Sterile

Page 13: HAMMERTECH - Fusion Orthopedics

HammerTech Fixation System Instrument Kits

12

HammerTech Fixation System Surgical Technique

K-Wire Double Trocar 1.0mm HT-KW-4010-S Handle HT Disposable CD-06-7000 Driver HT Petite TN-03-1 1 1 0 -S Drill Cannulated HT 2.4mm CD-01-6024-S

HT-20-1110-SImplant InstrumentsPetite

K-Wire Double Trocar 1.0mm HT-KW-4010-S Handle HT Disposable CD-06-7000 Driver HT Extra Small TN-03-1 1 2 0 -S Drill Cannulated HT 2.4mm CD-01-6024-S

HT-20-1120-SImplant InstrumentsExtra Small

K-Wire Double Trocar 1.4mm HT-KW-4014-S Handle HT Disposable CD-06-7000 Driver HT Small TN-03-1 1 3 0 -S Drill Cannulated HT 2.7mm CD-01-6027-S

HT-20-1130-SImplant InstrumentsSmall

K-Wire Double Trocar 1.4mm HT-KW-4014-S Handle HT Disposable CD-06-7000 Driver HT Medium TN-03-1 1 4 0 -S Drill Cannulated HT 2.9mm CD-01-6029-S

HT-20-1140-SImplant InstrumentsMedium

K-Wire Double Trocar 1.6mm HT-KW-4016-S Handle HT Disposable CD-06-7000 DriverHTLarge TN-03-1 1 5 0 -S Drill Cannulated HT 3.2mm CD-01-6032-S

HT-20-1150-SImplant InstrumentsLarge

Page 14: HAMMERTECH - Fusion Orthopedics

Call to schedule a Surgery:

800-403-6876

fusionorthopedics.com

4135 South Power Road, #110

Mesa, Arizona 85212, U.S.A.

800-403-6876

[email protected]

fusionorthopedics.com

D-4018-STG-1

9-25-18 Rev A

A surgeon must always rely on his or her own professional clinical judgment when deciding whether to use a particular product when treating a particular patient.  Fusion Orthopedics does not dispense medical advice and recommends that surgeons be trained in the use of any particular product before using it in surgery.  The information presented is intended to demonstrate the breadth of Fusion Orthopedics’ product offerings. A surgeon must always refer to the package insert, product label, and instructions for use before using any Fusion Orthopedics product.  Products may not be available in all markets because product availability is subject to the regulatory and/or medical practices in individual markets.  Please contact your representative if you have questions about the availability of products in your area.

FDA Cleared All Rights Reserved. HammerTech is a Trademark of Fusion Orthopedics, LLC.