Gladiator Conductivity Datasheet

download Gladiator Conductivity Datasheet

of 12

Transcript of Gladiator Conductivity Datasheet

  • 7/29/2019 Gladiator Conductivity Datasheet

    1/12

    Data Sheet

    The Gladiator Conductivity Switch is a third generation, state-

    of-the-art level probe, designed to operate in tough industrial

    environments.

    Principle of OperationsA low voltage AC signal is applied between the probe elec-

    trode and the tank wall or reference electrode in the case of a

    non-metallic tank. When the liquid comes into contact with

    the electrode tip, a conductive path is established between the

    sense electrode and the metallic tank wall/reference electrode.

    Current ow due to the conductive path is sensed, amplifed

    and used to switch a relay for indication or control purposes.

    Typical UsesFail-safe high-level/low-level alarm

    High-level alarmLow-level alarm

    High and low-level

    Interface Detection

    Pump control

    FunctionPoint level switch for conductive liquids. Metallic tanks may

    use either single or dual probe versions whilst non metallic

    tanks must used a dual probe version.

    Primary Areas of Application- Brewing- Chemical

    - Dairy

    - Edible Oil

    - Fertilizer

    - Food & Beverage

    - Glass

    - Mining & Metals

    - Oil & Gas

    - Packaging

    A higher level of performance

    Features:No moving parts - low maintenance

    Low voltage on probe for operational safetySimple 1-minute setup

    Remote sensor or Smart all in one types

    Relays outputs: Smart probe (1) Remote (2)

    Remote test function

    Adjustable ON and OFF delays (0-20 sec)

    Smart communication options: GosHawk,

    Modbus, HART, Probus DP, DeviceNet

    Remote GSM Connection option

    Remote amplier to probe separation up to

    500 m (1640 ft)

    Bright visual status indication on probe

    Independent housing alignment after

    mounting thread locked

    Conductivity Switch Series- A level switch for liquids and slurrys -

    - Paint

    - Paper

    - Pharmaceutical

    - Power Generation

    - Refning

    - Semiconductor

    - Sugar

    - Textile

    - Water & Wastewater

  • 7/29/2019 Gladiator Conductivity Datasheet

    2/12Gladiator - Conductivity Switch Series

    Typical Applications

    Non-metallic tank

    Metallic tank

    Sump pump control

    High level switch in water tank* mounting must be electrically

    connected to tank wall.

  • 7/29/2019 Gladiator Conductivity Datasheet

    3/12Gladiator - Conductivity Switch Series

    Single Probe Two Probes

    single switch point

    Dimensions

    Remote Amplier Enclosure

    Bareprobe

    50mm(

    2)

    85mm (3.3)

    probelength=L1

    M20 cable gland

    or 3/4 NPT adaptoron smart type.

    Window forviewing status

    LEDs

    Mounting

    thread

    90mm(

    3.5)

    50mm(

    2)

    5mm (0.2)

    85mm (3.3)Window for

    viewing statusLEDs

    Optional

    Flange

    90mm(

    3.5)

    50mm(

    2)

    2 x M20

    cable glandsor 3/4 NPT adaptors

    on remote type.

    8mm (0.3)

    teflonisolatedprobe

    forbuild-upprotec

    tion

    Bareprobe

    50mm(

    2)

    probelength=L1=L2

    teflonisolatedprob

    e

    forbuild-upprotection

    85mm (3.3)Window for

    viewing statusLEDs

    90mm(

    3.5)

    50mm(

    2)

    probelen

    gthL2

    probelengthL1

    10mm(

    0.3)

    teflonisolatedpro

    be

    forbuild-upprotection

    Mountingthread, 1.5 NPT

    or BSP only

    Mounting

    thread, 1.5 NPT

    or BSP only

    Two Probes

    dual switch points

    Standard probe lengths (L1 or L2):

    30 cm (11.8)

    50 cm (19.7)

    100 cm (39.4)

    Optional Flange Dimensions - 50mm ()

    ANSI (Class150)

    A

    120.7 4.75

    B

    152.4 6

    C

    19.1 0.75

    DIN (PN40) 125 4.9 165 6.5 18 0.7

    JIS (10K) 120 4.7 155 6 .1 19 0.75

    A BC

    131.5mm(

    5.2

    )

    7.5

    m

    m(

    0.3

    )

    192.5 mm (7.6)

    141.5mm(

    5.6

    )

    190mm(

    7.5

    )

    182.5 mm (7.2)

    147 mm (5.8)

    167.5mm(

    6.6

    )

    147 mm (5.8)

    30.7m

    m(

    1.2

    )

    158 mm (6.2)

    14 mm (0.6)

    74 mm (2.9)

    78 mm (3.1)

    107mm(

    4.2

    )

    111.5 mm (4.4)

    4 mm (0.2)

    50 mm (2)

    108mm(

    4.3

    )

    190mm(

    7.5

    )

    174 mm (6.9)192.5 mm (7.6)

    182.5 mm (7.2)

  • 7/29/2019 Gladiator Conductivity Datasheet

    4/12Gladiator - Conductivity Switch Series

    Laptop or PC Communications

    or PLC / DCS with

    MODBUS RTU Port

    GosHawk Software for

    inventory monitoring on PC

    GSM Network

    or

    CDMA Network

    Flotation Cells

    Sultan AcousticWave Transmitter

    Slurries

    Gladiator

    Admittance

    Switch

    Sultan Acoustic Wave TransmitterSilo, bin levels, coal, plastic powder,

    woodchip, sawdust, cement,clinker, iron ore, lime etc.

    Orca Sonar InterfaceThickener, CCD

    GSM or CDMA Network Typically up to 31 transmitters or switches per string.

    Maximum 250 transmitters or switches.

    Using GSM/CDMA network, transmitters and switches can bemonitored, calibrated remotely.

    Alarm status, diagnostics can be monitored.

    Support from factory engineering for customer application problems.

    Sultan Acoustic Wave TransmitterStockpiles, Stackers,

    Reclaimers

    Sultan AcousticWave SwitchBlocked Chute Detection

    Sultan, Gladiator & Guided RadarFarm Tanks, Grain Terminals

    Orca Sonar InterfaceClarier

    SULTAN 234

    SULTAN 234

    Gladiator

    Admittance

    Switch

    Gladiator

    Admittance

    Switch

    Gladiator

    Admittance

    Switch

    GLadiator

    Microwave

    Low Level

    Gladiator

    Microwave

    Low Level

    Sultan Master/Slave Positioning System

    Gladiator

    Conductivity

    Switch

    Gladiator

    Conductivity

    Switch

    (Limited Modbus query rate for Switches only)

    Sultan AcousticWave Transmitter

    Multidrop Connections

    Communication Network Overview

  • 7/29/2019 Gladiator Conductivity Datasheet

    5/12Gladiator - Conductivity Switch Series5

    Wiring

    Smart Probe Wiring

    RELAY

    1.

    NC

    2.

    COM

    3.

    NO

    COMMS DC-IN AC-IN

    4.

    Test

    7-30Vdc 80-265 Vac

    7.

    8.

    N

    9.

    L1

    10.

    1 2 3 4 5 6 7 8 9 10

    RS 485

    5.

    B

    6.

    A

    SENSI

    TIV

    ITY

    DELAY

    HI FSH TESTCAL

    -

    GLADIATOR SMART PROBE TERMINAL LAYOUT

    Hole for securing ofoptional identification tag

    M4 grounding screw

    Ground the housing tovessel, if vessel is metallic.Ground the housing toplant ground, if vessel isnon-metallic.

    The AC earth/ground

    cable must be connected

    to the ground screw

    inside the housing when

    using AC power.

    If only one cable is used for both

    power and output signal, then the

    second entry port must be

    plugged or blinded. Every Smart

    unit is supplied with two M20

    glands (or 3/4NPT adaptors)

    mounted on the unit and one

    blind plug loose.

    Remove Plug-In

    terminal block for

    easier wiring.

    !

  • 7/29/2019 Gladiator Conductivity Datasheet

    6/12Gladiator - Conductivity Switch Series

    Remote Amplier Wiring Diagram

    1 2 3 4 5 6 7 8 9 10

    GLADIATOR REMOTE AMPLIFIER TERMINAL LAYOUT

    GLADIATOR REMOTE PROBE TERMINAL LAYOUT

    M4 grounding screw

    7.1. 2. 3. 4. 5. 6. 8. 9. 10.

    RED

    W

    HITE

    BLUE

    BLACK

    Cable type between Amplifier and Probe

    4 conductor shielded twisted pair instrument cable.

    Conductor size dependent on cable length.

    BELDEN 3084A, DEKORON or equivalent.

    Max: BELDEN 3084A = 500m (1640 ft)Max: DEKORON IED183AA002 = 350m (1150 ft)

    Hole for securing of

    optional identification tag

    Ground the housing tovessel, if vessel is metallic.Ground the housing to

    plant ground, if vessel isnon-metallic.

    Remove Plug-Interminal block for

    easier wiring.

    Terminals 1, 2, 3, 4, 9, 10 not used.

    AC-IN

    A L1+

    DC-INCOMMSSENSOR

    NB

    RELAY 2(FAIL SAFE)

    NC

    COM

    NO

    RED

    BLACK

    BLUE

    WHITE

    24 VDC 80-265 VAC

    RELAY 1

    NC

    COM

    NO

    Testin

    NOTUSED

    CURRENT

    + Is

    NOT USED

    Wiring

  • 7/29/2019 Gladiator Conductivity Datasheet

    7/12Gladiator - Conductivity Switch Series

    FailSafe Low

    FSL

    1 2 3

    1 2 3

    1 2 3

    1 2 3

    FailSafe High

    FSH (default)

    1 2 3

    1 2 3

    1 2 3

    1 2 3

    RISING LEVEL

    NC NO

    NC NO

    NC NO

    NC NO

    NC NO

    NC NO

    NC NO

    COM COM

    COM

    COM COM

    COM COM

    POWER FAILURE

    NC NO NC NOCOM COMINTERNAL FAILURE

    POWER FAILURE

    OR

    SYSTEM OPERATING

    NORMALLY

    NC NOCOM NC NOCOM

    Relay Status

    LED Status

    State

    1

    State

    2

    State

    1

    Single Switch Point Dual Switch Points

    FALLING LEVEL

    OR

    FALLING LEVELCONTACT LEVEL CONTACT LEVEL

    RISING LEVEL

    Smart Probe

    terminal numbers

    Remote Amplifier

    terminal functionlabels

    Relay Action

    Relay Functions

    Wiring

    Fail-Safe Switch Contact Action

    Relay - Remote version only.

    For Smart Probes the Test terminal

    can act as a solid state output with a

    similar function.

    Level Switch

    Contact Action

    Relay - for Smart Probe Version

    (Set Relay Action selection

    switch)

    Relay 1 - for Remote Version

    (Set Relay Action parameter)

  • 7/29/2019 Gladiator Conductivity Datasheet

    8/12Gladiator - Conductivity Switch Series

    Correct

    Correct (non-preferred

    low level mounting)

    Incorrect1

    2

    Correct Correct

    If necessary, mount a

    protection plate to

    prevent direct impact.

    1

    Housing can be rotated

    within 200 after the

    mounting thread is

    tightened, to allow cable

    entries to face downwards

    or allow optimal cable

    clearance.

    30-45

    Possible

    product

    build-up

    Possible conductive

    build-up

    Correct

    Correct

    Incorrect

    Incorrect

    Incorrect 2

    Mounting Examples

    MountingProbes can be mounted from above or from the side.

    Use a protection plate for side mounting where the probe may

    be subject to impact damage.

    Install the Probe far enough away from the vessel wall to

    prevent the probe from coming into contact with the wall, and

    prevent conductive build-up from bridging the probe to the wall

    over time.

  • 7/29/2019 Gladiator Conductivity Datasheet

    9/12Gladiator - Conductivity Switch Series

    Advanced Remote Communication

    GSM/CDMA CommunicationHawkLink GSM/CDMA communication device allows anyauthorized computer with a standard modem and GosHawksoftware to dial in and calibrate, test or check on the perform-

    ance of the connected Hawk product. The HawkLink devicecan be wired to the standard communication terminals of theHawk products.

    Remote technical support and complete commissioning ofequipment in applications via our GSM/CDMA module allowsmonitoring and adjustments of settings no matter what cornerof the world.

    Remote connection worldwide!

  • 7/29/2019 Gladiator Conductivity Datasheet

    10/12Gladiator - Conductivity Switch Series 10

    Part Numbering

    Smart Probe Version

    CS100 SMART Gladiator Conductivity Switch

    Power Supply

    B 24 VDC standard (7-30VDC)U Universal AC power supply (80-260 VAC input) and 7-30VDC

    Output Options

    S Switch. 1 level relay, with Modbus

    Z Special Request

    Housing

    S Standard Powder Coated, Diecast with glass lidC Corrosion Resistant, Stainless Steel Housing

    Type of Electrodes

    S Single Point 1 Electrode Type for Metallic tanks

    D Dual Point 2 Electrode Type for Metallic tanks

    N Single Point 2 Electrode Type for Non-Metallic tanks (L1 must = L2)

    MountingTN10 1 NPT Thread - Size only available for 1 electrode type

    TB10 1 BSP Thread - Size only available for 1 electrode typeTN15 1.5 NPT Thread

    TB15 1.5 BSP Thread

    TZ Special Thread RequestFA2 2 Flange ANSI (Class 150)

    FD2 2 Flange DN 50 PN 40

    FJ2 2 Flange JIS (10K)

    FZ Special Flange Request

    Approval Standard

    X Standard CE ApprovedA20 ATEX 20 (DC power only)

    A22 ATEX 22 (DC power only)

    L1. Probe Length (Active Probe)

    P0 30 cm (11.8) rigid probeP50 50 cm (18.7) rigid probe

    P100 100 cm (39.4) rigid probeZ Special length

    L. Probe Length (Reference Probe - equal to or longer than L1)

    X Not required for 1 electrode type

    P30 30 cm (11.8) rigid probe

    P50 50 cm (19.7) rigid probeP100 100 cm (39.3) rigid probe

    Z Special length

    CS100 B S S S TN10 X P0 X Single Point Switch - Metallic Tank

    CS100 B S S D TN15 X P0 P50 Dual Point Switch - Metallic Tank

    CS100 B S S N TN15 X P0 P0 Single Point Switch - Non-Metallic Tank

  • 7/29/2019 Gladiator Conductivity Datasheet

    11/12Gladiator - Conductivity Switch Series11

    Part Numbering

    Remote Amplier

    GSA Gladiator Amplier - Remote (compatible with all Gladiator products) Housing

    S Standard polycarbonate electronics housing

    Power Supply

    B 24 VDC standard (12-30VDC)

    C 48 VDCU Universal AC power supply (80-260 VAC input) and 7-30VDC

    Output Options

    S Switch. 1 level relay, 1 failsafe relay, with Modbus

    I HART Isolated. 1 level relay, 1 failsafe relay

    D Devicenet. 1 level relay, 1 failsafe relayP Probus DP. 1 level relay, 1 failsafe relay

    Z Special Request

    GSA S C S

    Remote ProbeCS3200 Remote Gladiator Conductivity Probe (must be connected to GSA transmitter)

    Housing

    S Standard Powder Coated, Diecast with glass lid

    C Corrosion Resistant, Stainless Steel Housing

    Type of Electrodes

    S Single Point 1 Electrode Type for Metallic tanksD Dual Point 2 Electrode Type for Metallic tanks

    N Single Point 2 Electrode Type for Non-Metallic tanks (L1 must = L2)

    Mounting

    TN10 1 NPT Thread - Size only available with 1 electrode type for metallic tanks

    TB10 1 BSP Thread - Size only available with 1 electrode type for metallic tanks

    TN15 1.5 NPT ThreadTB15 1.5 BSP Thread

    TZ Special Thread Request

    FA2 2 Flange ANSI (Class 150)

    FD2 2 Flange DN 50 PN 40

    FJ2 2 Flange JIS (10K)FZ Special Flange Request

    Approval Standard

    X Standard CE Approved

    A20 ATEX 20 (DC power only)

    A22 ATEX 22 (DC power only)

    L1. Probe Length (Active Probe)P30 30 cm (11.8) rigid probe

    P50 50 cm ( 19.7) rigid probe

    P100 100 cm (39.4) rigid probe

    Z Special length

    L. Probe Length (Reference Probe - equal to or longer than L1)

    X Not required for 1 electrode typeP30 30 cm (11.8) rigid probe

    P50 50 cm (19.7) rigid probe

    P100 100 cm (39.3) rigid probe

    Z Special length

    CS00 S S TN10 X P0 X Single Point Switch - Metallic Tank

    CS00 S D TN15 X P0 P50 Dual Point Switch - Metallic Tank

    CS00 S N TN15 X P0 P0 Single Point Switch - Non-Metallic Tank

    Remote Version

  • 7/29/2019 Gladiator Conductivity Datasheet

    12/12

    Specications

    Hawk Measurement Systems (Head Ofce)

    15-17 Maurice Court

    Nunawading VIC 3131

    Australia

    Phone: +61 3 9873 4750

    Fax: +61 3 9873 4538

    [email protected]

    Contact

    Hawk Measurement

    7 River Street

    Middleton, MA 01949

    USA

    Phone: +1 888 HAWKLEVEL (1-888-429-5538)

    Phone: +1 978 304 3000

    Fax: +1 978 304 [email protected]

    Represented by:

    Additional product warranty and applic ation guarantees upon request . Technical data subject to change without notice.

    Operating Voltage

    7 - 30VDC (residual ripple no greater than 100mV) 80 - 260VAC 50/60Hz

    Power Consumption