For Review Only Determination of Aroclor 1260 in Soil Samples by GC/MS with Solid Phase Microextraction Journal: Journal of Separation Science Manuscript ID: jssc.201400102 Wiley - Manuscript type: Original Paper Date Submitted by the Author: 28-Jan-2014 Complete List of Authors: Zhang, Mengliang; Ohio University, Chemistry and Biochemistry Jackson, Glen; West Virginia University, Department of Chemistry Kruse, Natalie; Ohio University, Voinovich School of Leadership and Public Affairs Bowman, Jennifer; Ohio University, Voinovich School of Leadership and Public Affairs Harrington, Peter; Ohio University, Chemistry and Biochemistry Keywords: Environmental samples, GC/MS, Polychlorinated biphenyls, Soil , Solid phase microextraction Wiley-VCH Journal of Separation Science

Transcript of For Review Only · For Review Only Determin ation of Aroclor 1260 in Soil Samples by GC/MS with...

Page 1: For Review Only · For Review Only Determin ation of Aroclor 1260 in Soil Samples by GC/MS with Solid Phase Microextraction Journal: Journal of Separation Science Manuscript ID: jssc.201400102

For Review O

nly

Determination of Aroclor 1260 in Soil Samples by GC/MS

with Solid Phase Microextraction

Journal: Journal of Separation Science

Manuscript ID: jssc.201400102

Wiley - Manuscript type: Original Paper

Date Submitted by the Author: 28-Jan-2014

Complete List of Authors: Zhang, Mengliang; Ohio University, Chemistry and Biochemistry Jackson, Glen; West Virginia University, Department of Chemistry Kruse, Natalie; Ohio University, Voinovich School of Leadership and Public Affairs Bowman, Jennifer; Ohio University, Voinovich School of Leadership and Public Affairs

Harrington, Peter; Ohio University, Chemistry and Biochemistry

Keywords: Environmental samples, GC/MS, Polychlorinated biphenyls, Soil , Solid phase microextraction

Wiley-VCH

Journal of Separation Science

Page 2: For Review Only · For Review Only Determin ation of Aroclor 1260 in Soil Samples by GC/MS with Solid Phase Microextraction Journal: Journal of Separation Science Manuscript ID: jssc.201400102

For Review O

nly

1 / 20

Determination of Aroclor 1260 in Soil Samples by GC/MS with Solid 1

Phase Microextraction 2

Mengliang Zhang1, Glen P. Jackson2, Natalie A. Kruse3, Jennifer R. Bowman3 3

and Peter de B. Harrington1,∗ 4

1Center for Intelligent Chemical Instrumentation, Clippinger Laboratories, 5

Department of Chemistry and Biochemistry, Ohio University, Athens, Ohio 6

45701-2979, USA. 7

2Forensic and Investigative Science Program & C. Eugene Bennett 8

Department of Chemistry, West Virginia University, 308 Oglebay Hall, 9

Morgantown, WV 26506-6121 USA. 10

3Voinovich School of Leadership and Public Affairs, Ohio University, Athens, 11

OH 45701-2979 USA. 12

Abbreviations: ECD, electron capture detector; EI, electron ionization; 13

EICs, extracted ion chromatograms; EPA, the US Environmental Protection 14

Agency; PCB, polychlorinated biphenyls; PDMS, polydimethylsiloxane; RE, 15

relative error; RSD, relative standard deviation; SPME, solid phase 16

microextraction; TCMX, tetrachloro-m-xylene; TIC, total ion current. 17

Keywords: Environmental samples/ GC/MS/ Polychlorinated biphenyls/ Soil 18

/Solid phase microextraction 19

20

∗ Corresponding author: Tel: +01 740-994-0265/fax: +01 740-593-0148.

E-mail address: [email protected]

Page 1 of 25

Wiley-VCH

Journal of Separation Science

123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960

Page 3: For Review Only · For Review Only Determin ation of Aroclor 1260 in Soil Samples by GC/MS with Solid Phase Microextraction Journal: Journal of Separation Science Manuscript ID: jssc.201400102

For Review O

nly

2 / 20

Abstract 21

A novel fast screening method was developed for the determination of 22

polychlorinated biphenyls (PCBs) commercial mixture, Aroclor 1260, in soil 23

matrices by GC/MS combined with solid phase microextraction (SPME). The 24

non-equilibrium headspace SPME with a 100 µm polydimethylsiloxane fiber 25

was used to extract PCBs from 0.5 g soil matrices. The use of 2 mL of 26

saturated potassium dichromate in 6 M sulfuric acid solution improved the 27

reproducibility of the extractions and the mass transfer of PCBs from the soil 28

matrix to the headspace and SPME fiber for extraction times of 30 min at 29

100 °C. The percent recoveries for the SPME method, which were evaluated 30

using an Aroclor 1260 standard liquid injection calibration data set, were 31

within the range of 54.9-65.7%. Extracted ion chromatogram data were 32

used to construct the calibration curves. The accuracy was less than ±15% 33

expressed as relative error and the precision was less than 15% expressed 34

as relative standard deviation. The method was validated with certified soil 35

samples and the predicted concentrations for Aroclor 1260 agreed with the 36

certified values. The method was demonstrated to be linear from 10 ng/g to 37

1000 ng/g for Aroclor 1260 in dry soil. 38

1 Introduction 39

Polychlorinated biphenyls (PCBs) represent a class of organic pollutants that 40

are characterized by biphenyl with a number of chlorine substituents that 41

may range from 1 to 10 chlorine atoms per molecule. There are 209 42

Page 2 of 25

Wiley-VCH

Journal of Separation Science

123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960

Page 4: For Review Only · For Review Only Determin ation of Aroclor 1260 in Soil Samples by GC/MS with Solid Phase Microextraction Journal: Journal of Separation Science Manuscript ID: jssc.201400102

For Review O

nly

3 / 20

possible congeners. As persistent organic pollutants, PCBs are a major 43

environmental concern due to their ubiquity, toxicity, and persistence. In 44

North America, commercial PCBs were produced under the trade name 45

Aroclor by the Monsanto Company and were banned for use in 1977 [1, 2]. 46

PCBs have been found in different matrices including soil, surface water, 47

sediments, and air; so many approaches have been developed for the 48

determination of PCBs and Aroclors depending on the PCB concentration in 49

the matrices and regulatory requirements [3]. Although several 50

bioanalytical screening methods such as multianalyte enzyme-linked 51

immunosorbent assay [4], chemically activated luciferase gene expression 52

assay [5] have been developed, GC coupled with an electron capture 53

detector (ECD) or mass spectrometry (MS) are still the most widely accepted 54

and reliable techniques for the quantification of PCBs because of their high 55

sensitivity, good selectivity, and reproducibility [6, 7]. Traditional chemical 56

analyses are usually very time-consuming and expensive due to the 57

requirement of extensive extraction and cleanup procedures that are coupled 58

to lengthy high-resolution gas chromatographic programs [8]. For soil 59

samples, the US Environmental Protection Agency (EPA) recommends a 60

variety of different extraction methods, including Soxhlet extraction, 61

automated Soxhlet extraction, pressurized fluid extraction, microwave 62

extraction, ultrasonic extraction, and supercritical fluid extraction [7]. These 63

methods all require large volumes of organic solvent (15 ~ 200 mL) and 64

Page 3 of 25

Wiley-VCH

Journal of Separation Science

123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960

Page 5: For Review Only · For Review Only Determin ation of Aroclor 1260 in Soil Samples by GC/MS with Solid Phase Microextraction Journal: Journal of Separation Science Manuscript ID: jssc.201400102

For Review O

nly

4 / 20

long extraction times (1.5 ~ 20 h) [9]. Some recent improvements to 65

reduce the extraction time and the use of extraction solvent have been 66

reported, such as low-pressure microwave assisted extraction [3], on-line or 67

selective pressurized fluid extraction [10-12], and miniaturized ultrasonic 68

solvent extraction [13]. Extraction methods with higher efficiency, shorter 69

times, and lower costs have also been developed, including vortex assisted 70

liquid-liquid microextraction [14], dispersive liquid–liquid microextraction 71

[15], hollow-fiber liquid-phase microextraction [16, 17], and ultrasound 72

assisted emulsification-microextraction [18]. However, these techniques are 73

only applicable to liquid samples, are difficult to automate, unstable, and/or 74

prone to cause irreversible damage to the analytes [14]. 75

The technique of solid phase microextraction (SPME) was first reported by 76

Pawliszyn and co-workers in 1989 [19], and after more than 20 years of 77

development, has a very wide range of applications following 78

commercialization through Supelco [20, 21]. Compared with traditional 79

sample preparation techniques such as liquid-liquid extraction and solid 80

phase extraction, SPME has the following advantages. First, the method is 81

fast, simple, sensitive, and is solvent free. Second, sampling, extraction, 82

and concentration are integrated into one step. Third, injection can be easily 83

coupled with major separation systems such as GC, HPLC, and CE, which 84

makes it easy to implement and automate [20]. Fourth, fibers with different 85

selectivities [e.g., polydimethylsiloxane (PDMS), polyacrylate, carboxen, 86

Page 4 of 25

Wiley-VCH

Journal of Separation Science

123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960

Page 6: For Review Only · For Review Only Determin ation of Aroclor 1260 in Soil Samples by GC/MS with Solid Phase Microextraction Journal: Journal of Separation Science Manuscript ID: jssc.201400102

For Review O

nly

5 / 20

carbowax and divinylbenzene] are commercially available. SPME has been 87

applied to the determination of PCBs in different matrices such as water [22], 88

soil [23], ash [24], and tissues [25]. All of these applications focused on the 89

quantitative analysis of selected PCB congeners rather than modeling the 90

Aroclors directly. 91

In this study, a fast method for the determination of Aroclor 1260 in soil 92

matrices using headspace SPME-GC/MS was developed. The optimization of 93

headspace SPME is discussed. The sample analysis was accomplished within 94

35 min by staggering the sample preparation and GC/MS analysis. The total 95

peak areas of tetra-chlorinated biphenyls (tetra-CB, m/z 292), penta-96

chlorinated biphenyls (penta-CB, m/z 326), hexa-chlorinated biphenyls 97

(hexa-CB, m/z 360), hepta-chlorinated biphenyls (hepta-CB, m/z 394), and 98

octa-chlorinated biphenyls (octa-CB, m/z 430) were used to construct the 99

calibration models for the quantification of Aroclor 1260. The method was 100

then validated with certified soil samples. Although we do not show data for 101

other Aroclors in this particular study, we have tested the method using 102

different Aroclors with similar success. 103

2 Experimental 104

2.1 Reagents 105

An Aroclor 1260 stock solution at a concentration of 100 µg/mL in methanol 106

was obtained from AccuStandard, Inc. (New Haven, CT). Standard solutions 107

of Aroclor 1260 with concentrations of 0.1, 0.3, 1, 3, and 10 µg/mL were 108

Page 5 of 25

Wiley-VCH

Journal of Separation Science

123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960

Page 7: For Review Only · For Review Only Determin ation of Aroclor 1260 in Soil Samples by GC/MS with Solid Phase Microextraction Journal: Journal of Separation Science Manuscript ID: jssc.201400102

For Review O

nly

6 / 20

prepared by dilutions of aliquots of the stock solution with methanol. A 109

mixture in hexane containing 1 mg/mL of decachlorobiphenyl (deca-CB) and 110

of tetrachloro-m-xylene (TCMX) was also obtained from AccuStandard, Inc. 111

(New Haven, CT). Potassium permanganate, potassium dichromate, the 112

SPME fibers coated with polydimethylsiloxane (PDMS, 7 µm or 100 µm film 113

thickness), 20-mL headspace glass vials, and crimp seals with PTFE/silicone 114

septa were purchased from Sigma-Aldrich Co. LLC. (St. Louis, MO). The 115

clean soil and certified Aroclor 1260 soil samples were purchased from RT 116

Corp (Laramie, WY). 117

The standard soil samples were prepared by thoroughly mixing 50 µL of 118

standard solutions with 0.5 g clean soil and completely evaporating the 119

solvent in a hood at room temperature. The internal standard solution 120

containing 10 µg/mL of deca-CB and TCMX was prepared by dilution with 121

hexane from the 1 mg/mL stock solution, but only deca-CB was used as an 122

internal standard for the Aroclor quantification. A saturated potassium 123

dichromate solution was prepared by dissolving an excess of potassium 124

dichromate in 6.0 M sulfuric acid. 125

2.2 Instruments 126

All the experimental data were collected on a Thermo Finnigan PolarisQ 127

quadrupole ion trap mass spectrometer/Trace GC system with a Triplus 128

AS2000 autosampler (San Francisco, CA, USA). The GC/MS system was 129

controlled by the XCalibur software version 2.0.7 provided by Thermo. The 130

Page 6 of 25

Wiley-VCH

Journal of Separation Science

123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960

Page 8: For Review Only · For Review Only Determin ation of Aroclor 1260 in Soil Samples by GC/MS with Solid Phase Microextraction Journal: Journal of Separation Science Manuscript ID: jssc.201400102

For Review O

nly

7 / 20

GC separation was accomplished on a SHRXI-5MS capillary column (5% 131

diphenyl/95% dimethylpolysiloxane cross-linked, 30 m×0.25 mm i.d., 0.1 132

µm film thickness) from Shimadzu Scientific Instruments Inc. (Columbia, 133

MD). MATLAB R2012b (MathWorks Inc., Natick, MA) was used to process 134

the data. 135

The RAW files of the two-way (retention time × mass-to-charge ratio) 136

GC/MS data were converted to the network common document format (CDF) 137

with the ‘File Converter Tool’ in the XCalibur Software. The CDF files were 138

read directly into MATLAB using the netcdf functions. 139

2.3 Sample preparation 140

Soil samples of 0.5 g were added to the 20-mL SPME vial and spiked with 141

20 µL of internal standard solution. The samples were left in the hood at 142

room temperature to evaporate the solvent. Then 2 mL of saturated 143

potassium dichromate solution was added to the vial and the vial was sealed 144

with a PTFE/silicone septum using a crimp seal. After 30 s of vortexing, the 145

mixtures were placed in the autosampler tray for analysis. The sample vial 146

was incubated at 100 °C for 0.5 min. A PDMS fiber was then exposed to the 147

headspace for 30 min. The agitation was sequentially pulsed on for 10 s and 148

then off for 10 s for the 30 min exposure. 149

The fiber was thermally desorbed in the GC injector at 280 °C for 5 min 150

to prevent carryover. The analytes were separated using the following oven 151

temperature program at a constant helium flow of 1 mL/min: 50 °C, hold for 152

Page 7 of 25

Wiley-VCH

Journal of Separation Science

123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960

Page 9: For Review Only · For Review Only Determin ation of Aroclor 1260 in Soil Samples by GC/MS with Solid Phase Microextraction Journal: Journal of Separation Science Manuscript ID: jssc.201400102

For Review O

nly

8 / 20

1 min, ramp at 20 °C/min to 280 °C, hold for 10 min. The transfer line and 153

ion source temperatures were both maintained at 280 °C. The mass 154

spectrometer was operated in positive ion electron ionization (EI) mode at 155

40 eV and mass spectra were collected after a 4 min solvent delay. Full 156

scan mode was selected for the mass spectrometer and the scan range was 157

from mass-to-charge ratio (m/z) 140 to 550. 158

Five blank soil samples without any Aroclor and internal standard were 159

treated in the same way. The blank soil sample data were used for 160

correcting the baselines of the Aroclor soil samples. 161

3 Results and discussion 162

Aroclor 1260 soil samples at the concentration of 30 ng/g (ppb) were used 163

to evaluate the optimization of SPME and instrument conditions. The peak 164

areas of hexa-chlorinated biphenyls (hexa-CBs) were selected as references 165

to compare the effects of different conditions because hexa-CBs are one of 166

the major PCBs in Aroclor 1260 (46.9 weight %) [26]. Extracted ion 167

monitoring at m/z 360 was used to quantify the hexa-CBs by integrating the 168

peak areas of the extracted ion chromatogram from 12.50 to 13.78 min. 169

3.1 Optimization of SPME conditions 170

Directly exposing a PDMS fiber to the headspace of a vial containing 0.5 g of 171

30 ppb Aroclor 1260 soil sample demonstrated that PCBs were unable to 172

transfer to the headspace efficiently (Supporting information Fig. S1). The 173

low efficiency was attributed to the low boiling point and lipophilicity of PCBs, 174

Page 8 of 25

Wiley-VCH

Journal of Separation Science

123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960

Page 10: For Review Only · For Review Only Determin ation of Aroclor 1260 in Soil Samples by GC/MS with Solid Phase Microextraction Journal: Journal of Separation Science Manuscript ID: jssc.201400102

For Review O

nly

9 / 20

which caused sorption of the PCBs to the surfaces of the soil particles. 175

Elemental sulfur (S6 and S8) is another common interference in the soil 176

matrix which could significantly decrease the extraction efficiency of the 177

PCBs [23]. Montes et al. have demonstrated that the employment of strong 178

oxidative conditions such as the addition of potassium permanganate 179

solution (0.1 M in 6 M sulfuric acid) to the soil assists in the release of the 180

PCBs from the soil and the removal of organic matter and sulfur 181

interferences [23]. In this study, an additional two strong oxidants, 182

potassium dichromate and chromium trioxide in 6 M sulfuric acid, were 183

compared with potassium permanganate in 6 M sulfuric acid. 184

All the parameters in these initial studies were the same as section 2.3., 185

except that the EI energy was set to 70 eV instead of 40 eV, and internal 186

standards were not added to the samples. Three extraction solution systems 187

were compared: 1) 2 mL KMnO4 (0.1 M), 0.5 mL H2SO4 (6 M); 2) 2 mL 0.2 188

M CrO3 in 6 M H2SO4; 3) 2 mL 0.2 M K2Cr2O7 in 6 M H2SO4. The extraction 189

efficiency of the KMnO4 system was significantly higher (average about 2.5 190

times higher) than the other two extraction systems, but the repeatability 191

was significantly worse based on four replicate extractions (Supporting 192

information Fig. Error! Reference source not found.S2A). These 193

preliminary SPME experiments were accomplished with 10-mL SPME vials 194

and it was found that the fiber was fouled by the oxidative conditions, which 195

may have accounted for the poor repeatability. In an attempt to mitigate the 196

Page 9 of 25

Wiley-VCH

Journal of Separation Science

123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960

Page 11: For Review Only · For Review Only Determin ation of Aroclor 1260 in Soil Samples by GC/MS with Solid Phase Microextraction Journal: Journal of Separation Science Manuscript ID: jssc.201400102

For Review O

nly

10 / 20

oxidative fouling problem, 20-mL vials were used to create a larger volume 197

for headspace for the extractions. However, after approximately 20 198

analyses, the PDMS fiber still turned black, which indicated that even the 199

larger headspace volume was not able to prevent the fiber from being 200

contaminated by the KMnO4 (Supporting information Fig. S2B). 201

The use of CrO3 and K2Cr2O7 offered stable extraction efficiencies and less 202

degradation of the SPME fibers. Other treatments such as with a strong 203

basic solution (10 M NaOH) or a strong acidic solution (10 M H2SO4) or single 204

addition of water were investigated, but failed to effectively release the PCBs 205

from the soil to the headspace (Supporting information Fig. S1). 206

The effect of CrO3 versus K2Cr2O7, the effect of concentration of K2Cr2O7 207

in 6 M H2SO4, and the effect of solution volume on absolute recoveries were 208

also studied. The responses obtained from different solutions with a 0.5 g 209

soil sample are given in Supporting information Fig. S3. There was no 210

significant difference between different extraction systems (p = 0.15 by one-211

way analysis of variance). Therefore, K2Cr2O7 was chosen as the extraction 212

solution because of its availability. The concentrations and amounts of 213

K2Cr2O7 solution added to the sample had no significant effect on the 214

extraction efficiency. Saturated K2Cr2O7 in 6 M H2SO4 was chosen to oxidize 215

organic matter to the largest extent and the amount of solution was set to 2 216

mL instead of 4 mL to create more headspace and prevent SPME fiber 217

degradation by the extraction solution. 218

Page 10 of 25

Wiley-VCH

Journal of Separation Science

123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960

Page 12: For Review Only · For Review Only Determin ation of Aroclor 1260 in Soil Samples by GC/MS with Solid Phase Microextraction Journal: Journal of Separation Science Manuscript ID: jssc.201400102

For Review O

nly

11 / 20

The coatings of SPME fiber were selected between 7 µm PDMS and 100 219

µm PDMS because PDMS had a higher affinity for PCBs than the other 220

coatings in previous studies [22, 23, 27, 28]. To evaluate the influence of 221

the fiber thickness, both fibers were exposed to the headspace at 100 °C for 222

30 min, and the 100 µm PDMS fiber was chosen for further study because 223

the signals were approximately three times better (Supporting information 224

Fig. S4). 225

After SPME extraction, the desorption of the fiber was accomplished in the 226

GC injection port at 280 °C (the maximum operation temperature for 100 227

µm PDMS fiber) for 5 min to avoid the carryover effect. The absence of 228

carryover was also validated by a system blank injection after each sample 229

analysis. 230

The effects of extraction temperature and extraction time on the hexa-231

CBs extracted by headspace SPME with 100 µm PDMS fiber were 232

investigated. Soil samples of 0.5 g were extracted by 2 mL saturated 233

K2Cr2O7 in 6 M H2SO4 for 30 min at 25, 60, and 100 °C and responses 234

obtained plotted with respect to temperature as given in Fig. 1A. The 235

mobility of the PCBs through liquid and gas phases was significantly 236

improved with the increase in extraction temperature, so the responses 237

obtained at 100 °C were much larger than the responses at the other two 238

lower temperatures. Finally, 100 °C was selected as the extraction 239

temperature. Equilibrium was not achieved, so even higher temperatures 240

Page 11 of 25

Wiley-VCH

Journal of Separation Science

123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960

Page 13: For Review Only · For Review Only Determin ation of Aroclor 1260 in Soil Samples by GC/MS with Solid Phase Microextraction Journal: Journal of Separation Science Manuscript ID: jssc.201400102

For Review O

nly

12 / 20

may increase mass transport, but could exceed the pressure safety limits of 241

the SPME vial. The extraction time profiles for 5, 15, 30, and 60 min at 100 242

°C are given in Fig. 1B. The adsorption of PCBs to the fiber had not 243

equilibrated after a 30 min extraction. To keep the analysis within a 244

reasonable time, the extraction time was fixed at 30 min. 245

3.2 GC/MS analysis 246

To develop an efficient method, a full separation of all 209 possible PCB 247

congeners was not attempted. A 22 min GC temperature program was used 248

in this study which was reported earlier [29]. About 40 total ion current 249

(TIC) chromatographic peaks can be separated for Aroclor 1260 (Fig. 2A). 250

Each chromatographic peak may contain multiple co-eluted PCBs. 251

Full scan mode was used for MS and the mass scan range was from m/z 252

140 to 550 because most of the ions for the PCB mass spectra are larger 253

than m/z 145. The effects of EI energy on signal response were evaluated 254

at electron energies of 15, 40, and 70 eV. Very low responses were 255

obtained by using an EI energy of 15 eV, and an EI energy at 40 eV gave a 256

factor of two times better response than using an EI energy of 70 eV (Fig. 257

1C). Therefore, the EI energy was set at 40 eV for further work in this 258

study. 259

3.3 Analytical method performance 260

The data sets were pretreated by orthogonal baseline correction (using 261

bases of 10 components) for which the GC/MS baseline/background was 262

Page 12 of 25

Wiley-VCH

Journal of Separation Science

123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960

Page 14: For Review Only · For Review Only Determin ation of Aroclor 1260 in Soil Samples by GC/MS with Solid Phase Microextraction Journal: Journal of Separation Science Manuscript ID: jssc.201400102

For Review O

nly

13 / 20

reconstructed from a best fitting orthonormal bases constructed from one of 263

the blank SPME runs. The full details of baseline correction are described in 264

a previous study [29]. Using this approach, the artifact peaks (e.g., PDMS 265

peaks) and baseline were thereby significantly reduced in the TIC 266

chromatograms (Fig. 2B). The correction method was less effective for 267

extracted ion chromatograms (EICs), because the EICs are relatively 268

indifferent to PDMS fragment ions from column bleed and the SPME fiber. 269

Although the internal standard solution contained both TCMX at 9.5 min 270

and deca-CB at 19.5 min, only deca-CB was used as internal standard 271

because of its closer structure and chemistry to the PCBs of interest, and 272

because TCMX eluted early and overlapped with some of the matrix peaks. 273

The molecular ion of deca-CB (m/z 498) was extracted from TIC and the 274

peak area was integrated at retention time window between 19.3 to 19.6 275

min. Each chromatogram was normalized to the peak area of deca-CB 276

respectively. 277

EICs at retention time windows between 11.0 to 16.0 min were used to 278

construct smaller two-way GC/MS data sets which were selective for the 279

PCBs. The molecular ions of tetra-chlorinated biphenyls (tetra-CB, m/z 280

292), penta-chlorinated biphenyls (penta-CB, m/z 326), hexa-CB (m/z 360), 281

hepta-chlorinated biphenyls (hepta-CB, m/z 394), and octa-chlorinated 282

biphenyls (octa-CB, m/z 430) were used to create EIC two-way data (Fig. 283

Page 13 of 25

Wiley-VCH

Journal of Separation Science

123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960

Page 15: For Review Only · For Review Only Determin ation of Aroclor 1260 in Soil Samples by GC/MS with Solid Phase Microextraction Journal: Journal of Separation Science Manuscript ID: jssc.201400102

For Review O

nly

14 / 20

2C). All five PCBs mentioned above represent more than 99% of the PCBs in 284

Aroclor 1260 [26]. 285

The proposed method resulted in a linear dynamic range of 10-1000 ng/g 286

of Aroclor 1260 with a regression equation y = 73.38x + 0.94 (R2 = 287

0.9992). The accuracy and precision of the method were evaluated by the 288

prepared soil samples at three different concentrations with three replicates 289

at each concentration. As reported in Table 1, the accuracy is in the range 290

of 0 to 0.9% expressed by the relative error (RE, %) and the precision is in 291

the range of 4.6 to 12.6%, expressed by relative standard deviation (RSD). 292

In this study, the limit of detection (LOD) was calculated from three times 293

the standard deviation of the blank signal [30]. Five blank soil samples with 294

internal standard were treated the same as described in section 2.3. The 295

predicted concentrations for blank samples were calculated. Then three 296

times the standard deviation of predicted concentrations was taken as the 297

LOD, which yielded a value of 5.2 ng/g. 298

3.4 Recovery evaluation of SPME-GC/MS method 299

To evaluate the recovery of the SPME method, another calibration data set 300

using standard liquid injection was collected. All the instrumental 301

parameters were the same as those described in section 3.2., except that 302

the injection mode was changed from SPME mode to splitless liquid injection 303

mode. A set of 0.5 g Aroclor 1260 soil samples at the concentrations of 10, 304

30, 100, and 300 ng/g in five replicates were collected using the SPME-305

Page 14 of 25

Wiley-VCH

Journal of Separation Science

123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960

Page 16: For Review Only · For Review Only Determin ation of Aroclor 1260 in Soil Samples by GC/MS with Solid Phase Microextraction Journal: Journal of Separation Science Manuscript ID: jssc.201400102

For Review O

nly

15 / 20

GC/MS method. The calibration mode was constructed using EIC data and 306

the mass of Aroclor 1260 extracted by SPME was determined. The SPME 307

peak areas were compared to those of the liquid injection calibration curve 308

to assess the mass loading on the column. The percent recovery was 309

calculated using the calculated mass on-column of the SPME-extracted 310

sample relative to the absolute mass contained within the vial before SPME 311

extraction. The results are listed in Table 2. The recoveries ranged between 312

55-66% for the SPME samples at the four concentrations studied. The 313

results are not surprising if one considers that the adsorption equilibrium 314

between PCBs and the SPME fiber was not properly established within 30 315

min (Fig. 1B). However, the low recoveries did not affect the accuracy of 316

the method because they were reproducible. 317

3.5 Validation of method by certified soil samples 318

Certified soil samples originally contaminated with Aroclor 1260 at 1.50 µg/g 319

(prediction interval 0.65-2.34 µg/g) were measured using the previously 320

optimized conditions. The certified soil samples were diluted with certified 321

clean soils to the concentrations at 50 ng/g and 500 ng/g. Each soil sample 322

was analyzed by SPME-GC/MS in four replicate trials. As given in Table 3, 323

the estimated concentrations are inside of the certified prediction intervals. 324

4 Conclusions 325

A fast Aroclor-based quantitative method for PCBs in soil samples by SPME-326

GC/MS has been proposed in this study. The combination of potassium 327

Page 15 of 25

Wiley-VCH

Journal of Separation Science

123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960

Page 17: For Review Only · For Review Only Determin ation of Aroclor 1260 in Soil Samples by GC/MS with Solid Phase Microextraction Journal: Journal of Separation Science Manuscript ID: jssc.201400102

For Review O

nly

16 / 20

dichromate and sulfuric acid solution was used to extract PCB from soil for 328

the first time, and the parameters for the non-equilibrium headspace SPME 329

were optimized. The extracted ion two-way (EIC) data sets were used to 330

construct calibration curves and the method has been validated by 331

commercial certified soil samples. The predicted concentrations of Aroclor 332

1260 were all in the prediction intervals for the certified soil samples. The 333

proposed method has the advantage of the high sample throughput, with a 334

soil sample being prepared and analyzed staggeringly about every 35 min. 335

The headspace SPME method is easy to perform and has the potential to be 336

adapted for onsite analysis. Other preliminary studies have demonstrated 337

its application to the field study combined with a portable GC/MS instrument 338

[31]. The method required a low sample amount (0.5 g) which can benefit 339

applications for which sample availability is a limiting factor. 340

Acknowledgments 341

The research is funded by a grant from US Department of Energy, 342

Office of Environmental Management, Portsmouth/Paducah Project Office#. 343

The Center for Intelligent Chemical Instrumentation and Department of 344

Chemistry and Biochemistry at Ohio University are acknowledged for the 345

financial support. 346

#The project was supported by US Department of Energy. Any opinions, 347

findings, and conclusions or recommendations expressed in this material are 348

those of the author(s) and do not necessarily reflect the views of the US 349

Page 16 of 25

Wiley-VCH

Journal of Separation Science

123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960

Page 18: For Review Only · For Review Only Determin ation of Aroclor 1260 in Soil Samples by GC/MS with Solid Phase Microextraction Journal: Journal of Separation Science Manuscript ID: jssc.201400102

For Review O

nly

17 / 20

Department of Energy Office of Environmental Management 350

Portsmouth/Paducah Project Office, or of the Voinovich School of Leadership 351

and Public Affairs at Ohio University. 352

The authors have declared no conflict of interest. 353

5 References 354

[1] A.L. Prignano, C.T. Narquis, Hyatt, J. E., WM2008 Conference 2008, 24-355

28. 356

[2] Safe, S., Crit. Rev. Toxicol. 1984, 13, 319-395. 357

[3] Bruzzoniti, M. C., Maina, R., Tumiatti, V., Sarzanini, C., et al., J. 358

Chromatogr. A. 2012, 1265, 31-38. 359

[4] Bronshtein, A., Chuang, J. C., Van Emon, J. M., Altstein, M., J. Agr. Food 360

Chem. 2012, 60, 4235-4242. 361

[5] Hasegawa, J., Guruge, K. S., Seike, N., Shirai, Y., et al., Chemosphere 362

2007, 69, 1188-1194. 363

[6] Mydlova-Memersheimerova, J., Tienpont, B., David, F., Krupcik, J., 364

Sandra, P., J. Chromatogr. A. 2009, 1216, 6043-6062. 365

[7] Wise, S. A., Barcelo, D., Garrigues, P., Turle, R., Anal. Bioanal. Chem. 366

2006, 386, 765-767. 367

[8] Castro-Jimenez, J., Gonzalez, C., J. Environ. Monit. 2011, 13, 894-900. 368

[9] EPA, EPA method 8082a 2000. 369

[10] Zhang, Z., Ohiozebau, E., Rhind, S. M., J. Chromatogr. A. 2011, 1218, 370

1203-1209. 371

Page 17 of 25

Wiley-VCH

Journal of Separation Science

123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960

Page 19: For Review Only · For Review Only Determin ation of Aroclor 1260 in Soil Samples by GC/MS with Solid Phase Microextraction Journal: Journal of Separation Science Manuscript ID: jssc.201400102

For Review O

nly

18 / 20

[11] Gomez-Ariza, J. L., Bujalance, M., Giraldez, I., Velasco, A., Morales, E., 372

J. Chromatogr. A. 2002, 946, 209-219. 373

[12] Sporring, S., Von Holst, C., Björklund, E., Chromatographia 2006, 64, 374

553-557. 375

[13] Aydin, M. E., Tor, A., Ozcan, S., Anal. Chim. Acta 2006, 577, 232-237. 376

[14] Ozcan, S., J Sep Sci 2011, 34, 574-584. 377

[15] Liu, X., Zhao, A., Zhang, A., Liu, H., et al., J. Sep. Sci. 2011, 34, 1-7. 378

[16] Li, G., Zhang, L., Zhang, Z., J. Chromatogr. A. 2008, 1204, 119-122. 379

[17] Basheer, C., Lee, H. K., Obbard, J. P., J. Chromatogr. A. 2004, 1022, 380

161-169. 381

[18] Ozcan, S., Tor, A., Aydin, M. E., Anal. Chim. Acta 2009, 647, 182-188. 382

[19] Arthur, C. L., Pawliszyn, J., Anal. Chem. 1990, 62, 2145-2148. 383

[20] Vas, G., Vekey, K., J. Mass Spectrom. 2004, 39, 233-254. 384

[21] Duan, C. F., Shen, Z., Wu, D. P., Guan, Y. F., TrAC Trends Anal. Chem. 385

2011, 30, 1568-1574. 386

[22] Hawthorne, S. B., Grabanski, C. B., Miller, D. J., Anal. Chem. 2009, 81, 387

6936-6943. 388

[23] Montes, R., Ramil, M., Rodriguez, I., Rubi, E., Cela, R., J. Chromatogr. 389

A. 2006, 1124, 43-50. 390

[24] Criado, M. R., Pereiro, I. R., Torrijos, R. C., Talanta 2004, 63, 533-540. 391

[25] Poli, D., Caglieri, A., Goldoni, M., Castoldi, A. F., et al., J. Chromatogr. 392

B. 2009, 877, 773-783. 393

Page 18 of 25

Wiley-VCH

Journal of Separation Science

123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960

Page 20: For Review Only · For Review Only Determin ation of Aroclor 1260 in Soil Samples by GC/MS with Solid Phase Microextraction Journal: Journal of Separation Science Manuscript ID: jssc.201400102

For Review O

nly

19 / 20

[26] Frame, G. M., Wagner, R. E., Carnahan, J. C., Brown, J. F., et al., 394

Chemosphere 1996, 33, 603-623. 395

[27] Llompart, M., Li, K., Fingas, M., J. Microcolumn. Sep. 1999, 11, 397-396

402. 397

[28] Derouiche, A., Driss, M. R., Morizur, J. P., Taphanel, M. H., J. 398

Chromatogr. A. 2007, 1138, 231-243. 399

[29] Zhang, M., Harrington, P. d. B., Talanta 2013, 117, 483-491. 400

[30] Long, G. L., Winefordner, J. D., Anal. Chem. 1983, 55, 712A-724A. 401

[31] Zhang, M., Harrington, P. d. B., Kruse, N. A., Bowman, J. R., et al., 402

61st ASMS Conference on Mass Spectrometry and Allied Topics, Minneapolis, 403

MN 2013. 404

405

406

407

408

409

410

411

412

413

414

415

Page 19 of 25

Wiley-VCH

Journal of Separation Science

123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960

Page 21: For Review Only · For Review Only Determin ation of Aroclor 1260 in Soil Samples by GC/MS with Solid Phase Microextraction Journal: Journal of Separation Science Manuscript ID: jssc.201400102

For Review O

nly

20 / 20

Legends of figures: 416

Figure 1. The effect of extraction temperature (A), extraction time (B), and 417

electron ionization energy (C) on the extraction efficiency of PCBs from soil 418

samples spiked with Aroclor 1260. (n = 3) 419

Figure 2. GC/MS TIC chromatograms of Aroclor 1260 before (A) and after (B) 420

baseline/background correction for 30 ng/g soil sample after headspace 421

SPME extraction. On the right (C) are EICs for tetra-, penta-, hexa-, hepta-, 422

and octa-CBs at zoom-in retention time window. 423

424

425

426

427

428

429

430

431

432

433

434

Page 20 of 25

Wiley-VCH

Journal of Separation Science

123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960

Page 22: For Review Only · For Review Only Determin ation of Aroclor 1260 in Soil Samples by GC/MS with Solid Phase Microextraction Journal: Journal of Separation Science Manuscript ID: jssc.201400102

For Review O

nly

Figure 1. The effect of extraction temperature (A), extraction time (B), and electron ionization energy (C) on the extraction efficiency of PCBs from soil samples spiked with Aroclor 1260. (n = 3)

273x464mm (300 x 300 DPI)

Page 21 of 25

Wiley-VCH

Journal of Separation Science

123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960

Page 23: For Review Only · For Review Only Determin ation of Aroclor 1260 in Soil Samples by GC/MS with Solid Phase Microextraction Journal: Journal of Separation Science Manuscript ID: jssc.201400102

For Review O

nly

Figure 2. GC/MS TIC chromatograms of Aroclor 1260 before (A) and after (B) baseline/background correction for 30 ng/g soil sample after headspace SPME extraction. On the right (C) are EICs for tetra-,

penta-, hexa-, hepta-, and octa-CBs at zoom-in retention time window. 304x170mm (300 x 300 DPI)

Page 22 of 25

Wiley-VCH

Journal of Separation Science

123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960

Page 24: For Review Only · For Review Only Determin ation of Aroclor 1260 in Soil Samples by GC/MS with Solid Phase Microextraction Journal: Journal of Separation Science Manuscript ID: jssc.201400102

For Review O

nly

1 / 3

Table 1. Accuracy and precision of developed method

Aroclor 1260 concentration

(ppb)

Measured concentration

(ppb)

Mean concentration

(ppb)

Accuracy

(RE, %)

Precision

(RSD, %)

30 26.8

30 0 11.2 30 29.7

30 33.5

300 309

301 0.3 4.6 300 285

300 309

1000 1078

1009 0.9 12.6 1000 1085

1000 864

1

2

3

4

5

6

Page 23 of 25

Wiley-VCH

Journal of Separation Science

123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960

Page 25: For Review Only · For Review Only Determin ation of Aroclor 1260 in Soil Samples by GC/MS with Solid Phase Microextraction Journal: Journal of Separation Science Manuscript ID: jssc.201400102

For Review O

nly

2 / 3

Table 2. The percentage recoveries of Aroclor 1260 by SPME-GC/MS.

Aroclor1260 concentration

(ppb)

Aroclor1260 in the vial

(ng)

Aroclor1260 on column

(ng)

Mean recovery

(%)

RSD

(%)

10 5 2.9

61.4 11.9

10 5 2.6

10 5 2.9

10 5 2.8

10 5 4.1

30 15 10.3

65.7 6.1

30 15 9.8

30 15 9.4

30 15 8.6

30 15 11.1

100 50 33.8

64.5 5.0

100 50 31.2

100 50 31.3

100 50 29.2

100 50 35.7

300 150 82.9

54.9 0.8

300 150 82.8

300 150 82.5

300 150 80.2

300 150 83.4

7

8

9

Page 24 of 25

Wiley-VCH

Journal of Separation Science

123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960

Page 26: For Review Only · For Review Only Determin ation of Aroclor 1260 in Soil Samples by GC/MS with Solid Phase Microextraction Journal: Journal of Separation Science Manuscript ID: jssc.201400102

For Review O

nly

3 / 3

Table 3. Application of the method to certified soil samples (n = 4).

Prediction interval (certified reference value) (ppb)

Concentration found (ppb)

217-780 (500) 545±91

22-78 (50) 64±15

10

11