EVALUATION OF THE
INTERLABORATORY COMPARISON
TEST
Volatile halogenated hydrocarbons
(VOC), BTEX and MTBE - CB01
Sample dispatch on 28th June 2016
Address: Umweltbundesamt GmbH
Spittelauer Lände 5
1090 Vienna/Austria
Contact: Dr. Sandra Kulcsar
Telephone: +43 (0) 1 31304 4334
E-mail: [email protected]
Website: http://www.umweltbundesamt.at/en/interlaboratory_comparison/
www.ifatest.eu
Management:
Dipl.-Ing. Monika Denner
2
Table of contents
1 Interlaboratory comparison test: Volatile halogenated hydrocarbons (VOC),
BTEX and MTBE – CB01 ............................................................................................................4
1.1 Participants and time schedule ...................................................................... 4
1.2 Sampling, sample material and distribution ................................................... 4
1.3 Control testing ............................................................................................... 5
2 Evaluation ................................................................................................................................5
3 Representation and interpretation of measurement results .........................................6
4 Explanatory notes ..................................................................................................................6
5 Annotations on tables and charts .......................................................................................7
6 Summary of results, after removal of outliers ..................................................... 11
7 Parameter oriented report .................................................................................. 13
8 Laboratory oriented report ................................................................................ 160
3
1 Interlaboratory comparison test: Volatile halogenated
hydrocarbons (VOC), BTEX and MTBE – CB01
1.1 Participants and time schedule
Number of registrations: 25
Number of submitted data records: 25
Dispatch of samples: 28th June 2016
Closing date for submission of data: 26th July 2016
For the interlaboratory comparison test CB01 the participants could participate in C-CB01 (VOC) and/or B-CB01 (BTEX and MTBE). To anonymize results, each laboratory was given a laboratory code on a random
basis.
1.2 Sampling, sample material and distribution
The following samples were made available:
2 Samples ground water (CB01 A – VOC, CB01 A – BTEX / MTBE)
2 Sample municipal waste water (CB01 B – VOC, CB01 B – BTEX / MTBE)
The sampling of the ground- and waste water was carried out on 27th June 2016.
Both samples were stored at < 4 °C until further processing.
The samples were partly spiked with specific substances and were filled into bottles
under continuous stirring to achieve homogeneous samples. The samples were
dispatched on 28th June 2016.
Each participant received (according to the order) :
2 samples (each 600 ml), each filled in 600 ml Aluminium bottles or
4 samples (each 600 ml), each filled in 600 ml Aluminium bottles
4
1.3 Control testing
During filling the bottles, aliquots of each sample were collected randomly for control
testing. Testing was performed close to the time of sample dispatch.
In the parameter-oriented evaluation, the results of the control testing are given in the
form of arithmetic means of the detected concentrations as check value ± U.
2 Evaluation
The analytical results had to be made available to the organiser not later than 26th
July 2016. Any values received at a later date were not considered. A statistical
evaluation of interlaboratory comparison data was only carried out if at least 6 valid
results per parameter were available.
To evaluate the data, outliers were detected first by using the outlier test method
according to Hampel. Values identified as conspicuous by this test method are
marked specifically in the parameter-oriented evaluation. Further evaluation was
performed in accordance with DIN ISO 5725-2. Results < LOQ or < LOD are not
taken into account for calculation.
The adjusted average value (after removal of outliers) for all submitted results was
used as a basis for the calculation of recovery rates and z-scores.
z-Score
z-Scores were calculated on the basis of the following formula:
SD
Xxscorez i
In this context,
ix is the measurement value of the participating laboratory.
X is the adjusted average value (i.e. after removal of outliers) of the participants‘ results.
SD is the reproducibility standard deviation, calculated from the participants‘ results (after removal of outliers) in the relevant test round.
5
Interpretation of z-Scores in the parameter-oriented evaluation:
|z| < 2 result: good
2 < |z| < 3 result: questionable
|z| > 3 result: not satisfactory
3 Representation and interpretation of measurement results
The parameter oriented report shows the measurement values including uncertainty,
recovery rate, calculated z-Score and the outliers in tabular form. The results listed in
the table are also represented graphically.
The laboratory oriented report shows the results of the individual laboratories,
including the recovery rates and z-Scores.
An annotation of the tables and graphics is given in section 5.
4 Explanatory notes
As explained in section 2, the z-Score is calculated using the reproducibility standard
deviation, calculated from the participants’ results (after removal of outliers) in the
relevant test round. As a consequence it might occur that the z-Score between -2 and
2 covers an extraordinary range, due to a high variance of the results.
The recovery rate is calculated for the individual result based on the target value.
Therefore, in the case of a high variance of the results, participants should also
consider recovery rates as an indicator for the necessity of internal quality assurance
measures.
Cf. Tetrachloromethane sample CB01 B - VOC (n=23)
Cf. Bromodichloromethane sample CB01 B - VOC (n=20)
Sample CB01 A: For the parameters 1,1,1-Trichloroethane, 1,1-Dichloroethene and
Tribromomethane no target value was calculated because of the low analyte content.
6
5 Annotations on tables and charts
5.1 Information and abbreviations in tables
Parameter Analyte identifier
Sample Sample identifier
Unit Given unit for result and uncertainty (e.g. µg/l)
Mean Mean of the participants results, without outliers (3
significant digits)
CI (99 %) 99% confidence interval (3 significant digits)
Minimum Minimum of all submitted results, after removal of outliers
(3 significant digits)
Maximum Maximum of all submitted results, after removal of outliers
(3 significant digits)
SD Reproducibility standard deviation, calculated from the
participants results, after removal of outliers (3 significant
digits)
RSD % Reproducibility standard deviation, calculated from the
participants results relative to the target value, given in %,
after removal of outliers (2 significant digits)
Check value ± U Mean of check value ± measurement uncertainty (3
significant digits)
Labcode Laboratory identifier (anonymized)
Result Result as indicated by participant (max. 5 decimal places)
± U Results uncertainty as indicated by participant (max. 5
decimal places)
LOQ Limit of quantification
LOD Limit of detection
Recovery Recovery rate in % based on target value (3 significant
digits, max. one decimal place given)
z-Score Deviation of result based on target value depicted as a
multiple of the criteria (3 significant digits, max. 2 decimal
places given)
- No data available
Comments Comment on the respective result (e.g. H, FN, FP)
H Outlier according to Hampel-Test
FN False negative – for a result < LOQ or result < LOD: The
absolute value of the LOQ or LOD fulfils the condition of
an outlier according to the Hampel test.
FP False positive – for parameters where no target value is
available because of a too low analyte content (n < 6):
7
Result that exceeds the median of the absolute values of
the transmitted LOQs or LODs by more than 100 %.
Standard deviation Reproducibility standard deviation, calculated from the
participants results (3 significant digits)
Rel. standard deviation Reproducibility standard deviation, calculated from the
participants results relative to the target value, given in %,
(3 significant digits)
n Number of results
Target value Mean of the participants results, without outliers (3
significant digits)
Criteria Criteria for z-Score calculation. The given value matches
the reproducibility standard deviation, calculated from the
participants’ results, after removal of outliers (3 significant
digits).
8
5.2 Graphical presentation of results
Example chart: Results
|z-Score|=2
���� ������������ �
|z-Score|=3
|z-Score|=2
|z-Score|=3
Target value (100 %)
Recovery rate of laboratory
Outliner according to the
Hampel test ± uncertainty
Result ± uncertainty
Confidence interval (95 %) of
target value
Result outside displayed area
Mean of participants
results, without outliners =
target value
Result < LOQ or < LOD, the basis of the
triangle is equal to the absolute value
of the LOQ or LOD
Example chart: Recovery
Result outside displayed area
Confidence interval (95 %) of
target value
9
|z-Score|=2
z-score of laboratory
Target value (z-score =0)
|z-Score|=2
|z-Score|>2
z-score of laboratory
Result outside displayed area
Target value (z-score =0)
Example chart: z-score
Example chart: z-score - laboratory oriented report
Sample
Sample A Sample B
Parameter 1
Parameter 2
Parameter 3
Parameter 4
Parameter 5
Parameter 6
Parameter 7
Parameter 8
Parameter 9
Parameter 10
Parameter 11
Parameter 12
10
Summary of results, after removal of outliers: Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
6 Summary of results, after removal of outliers
SampleParameter Unit Number of results for calculation
Number of outliers
Mean ± Cl (99%) Minimum Maximum SD RSD %
CB01 A - BTEX / MTBEBenzene µg/l 16 0.58 ± 0.0519 0.47 0.73 0.0692 121CB01 B - BTEX / MTBE µg/l 18 2.07 ± 0.241 1.6 2.89 0.341 160CB01 A - BTEX / MTBEEthylbenzene µg/l 16 0.442 ± 0.0383 0.32 0.52 0.051 121CB01 B - BTEX / MTBE µg/l 17 1.45 ± 0.201 0.87 2 0.276 191CB01 A - BTEX / MTBEo-Xylene µg/l 16 4.41 ± 0.513 2.88 5.7 0.684 162CB01 B - BTEX / MTBE µg/l 16 3.03 ± 0.363 2.2 4.02 0.484 162CB01 A - BTEX / MTBESum of m-Xylene and p-Xylene µg/l 16 5.69 ± 0.719 3.48 7.36 0.959 172CB01 B - BTEX / MTBE µg/l 16 8.55 ± 1.04 5.33 10.7 1.38 162CB01 A - BTEX / MTBEToluene µg/l 18 0.85 ± 0.0889 0.61 1.09 0.126 150CB01 B - BTEX / MTBE µg/l 17 5.33 ± 0.512 4.21 6.8 0.704 131CB01 A - BTEX / MTBEMethyl-tert-butyl-ether µg/l 13 3.25 ± 0.361 2.5 4.1 0.434 130CB01 B - BTEX / MTBE µg/l 13 0.882 ± 0.103 0.72 1.1 0.124 140CB01 A - VOC1,1,1-Trichloroethane µg/l 0 - ± - - - - -0CB01 B - VOC µg/l 22 7.27 ± 0.671 5.62 9.6 1.05 141CB01 A - VOC1,1-Dichloroethene µg/l 0 - ± - - - - -0CB01 B - VOC µg/l 19 9.06 ± 1.3 4.75 12.7 1.9 212CB01 A - VOC1,2-Dichloroethane µg/l 20 12.7 ± 0.888 10.5 15.4 1.32 101CB01 B - VOC µg/l 18 3.11 ± 0.227 2.51 3.7 0.322 103CB01 A - VOCBromodichloromethane µg/l 22 2.26 ± 0.141 1.92 2.82 0.221 9.71CB01 B - VOC µg/l 20 7.33 ± 0.346 6.51 8.68 0.515 73CB01 A - VOCcis-1,2-Dichloroethene µg/l 21 0.833 ± 0.063 0.68 1.01 0.0962 120CB01 B - VOC µg/l 19 4.68 ± 0.313 4.02 6.07 0.455 9.72CB01 A - VOCDibromochloromethane µg/l 23 1.1 ± 0.0847 0.87 1.4 0.135 120CB01 B - VOC µg/l 23 8.15 ± 0.698 6.51 10.8 1.12 140CB01 A - VOCDichloromethane µg/l 19 1.89 ± 0.147 1.66 2.4 0.213 111CB01 B - VOC µg/l 17 8.04 ± 0.611 5.92 9.39 0.84 104CB01 A - VOCTetrachloroethene µg/l 22 3.94 ± 0.494 2.35 5.72 0.773 201
This report was created with PROLab, a software by QuoData: http://www.quodata.de/
11
Summary of results, after removal of outliers: Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
SampleParameter Unit Number of results for calculation
Number of outliers
Mean ± Cl (99%) Minimum Maximum SD RSD %
CB01 B - VOCTetrachloroethene µg/l 23 11.3 ± 1.38 8.18 15.7 2.21 200CB01 A - VOCTetrachloromethane µg/l 23 1.19 ± 0.127 0.75 1.63 0.204 170CB01 B - VOC µg/l 23 16.1 ± 2.37 7.28 24.2 3.78 240CB01 A - VOCtrans-1,2-Dichloroethene µg/l 21 0.852 ± 0.112 0.514 1.2 0.171 200CB01 B - VOC µg/l 20 3.72 ± 0.446 2.99 5.34 0.665 181CB01 A - VOCTribromomethane µg/l 1 - ± - 0.03 0.03 - -0CB01 B - VOC µg/l 21 3.43 ± 0.245 2.83 4 0.374 111CB01 A - VOCTrichloroethene µg/l 22 0.652 ± 0.0506 0.5 0.82 0.0791 121CB01 B - VOC µg/l 22 3.9 ± 0.324 3.08 5.19 0.507 131CB01 A - VOCTrichloromethane µg/l 22 9.95 ± 0.664 8.32 12.5 1.04 101CB01 B - VOC µg/l 21 1.04 ± 0.072 0.84 1.26 0.11 112
12
7 Parameter oriented report
Benzene ..........................................................................................................................14
Ethylbenzene ..................................................................................................................22
o-Xylene ..........................................................................................................................30
Sum of m-Xylene and p-Xylene .......................................................................................38
Toluene ...........................................................................................................................46
Methyl-tertiary-butyl ether ................................................................................................54
1,1,1-Trichloroethane ......................................................................................................62
1,1-Dichloroethene ..........................................................................................................68
1,2-Dichloroethane ..........................................................................................................74
Bromdichloromethane .....................................................................................................82
cis-1,2-Dichloroethene .....................................................................................................90
Dibromochloromethane ...................................................................................................98
Dichloromethane .............................................................................................................106
Tetrachloroethene ...........................................................................................................114
Tetrachloromethane ........................................................................................................122
trans-1,2-Dichloroethene .................................................................................................130
Tribromomethane ............................................................................................................138
Trichloroethene ...............................................................................................................144
Trichloromethane ............................................................................................................152
13
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ABTX, Parameter: Benzene
Benzene
CB01 A - BTEX / MTBE
µg/l
Unit
Mean ± Cl (99%)
Control test value ± U
Parameter oriented report
Minimum - Maximum
µg/l
0.58 ± 0.05190.47 - 0.730.560 ± 0.0141
Labcode Result ± U Recovery [%] z-score Comments
LC0001 0.5 0.05 86.2 -1.16LC0004 0.57 0.11 98.2 -0.15LC0005 0.6 0.2 103 0.29LC0006 - - - -LC0007 < 0.5 (LOQ) - - -LC0009 0.5 0.125 86.2 -1.16LC0010 0.62 0.117 107 0.57LC0011 0.73 0.15 126 2.17LC0012 0.55 0.1 94.8 -0.44LC0014 0.647 0.07 112 0.96LC0015 0.56 0.129 96.5 -0.29LC0016 0.53 0.1 91.3 -0.73LC0017 0.59 0.178 102 0.14LC0018 0.693 0.208 119 1.63LC0019 0.47 0.14 81 -1.59LC0020 0.55 0.05 94.8 -0.44LC0022 0.594 0.08316 102 0.2LC0024 0.58 0.17 100 0.00LC0025 0.927 - 160 5.01 H
all results ithout outliers Unit
Mean ± Cl (99%) µg/l
µg/l
µg/l
µg/l
%
-n
rel. Standard deviation
Standard deviation
Maximum
Minimum
Characteristics of parameter
0.601 ± 0.0782 0.58 ± 0.0519
0.470.47
0.927 0.73
0.06920.108
17.9 11.9
1617
w
This report was created with PROLab, a software by QuoData: http://www.quodata.de/
14
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ABTX, Parameter: Benzene
Laboratory
LC00
01
LC00
04
LC00
05
LC00
07
LC00
09
LC00
10
LC00
11
LC00
12
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
22
LC00
24
LC00
25
µg/l
1.0
0.9
0.8
0.7
0.6
0.5
0.4
0.3
0.2
H
Graphical presentation of resultsResults
15
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ABTX, Parameter: Benzene
Laboratory
LC00
01
LC00
04
LC00
05
LC00
09
LC00
10
LC00
11
LC00
12
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
22
LC00
24
LC00
25
Rec
over
y [%
]
150
140
130
120
110
100
90
80
70
60
50
Recovery rate
16
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ABTX, Parameter: Benzene
Laboratory
LC00
01
LC00
04
LC00
05
LC00
09
LC00
10
LC00
11
LC00
12
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
22
LC00
24
LC00
25
z-sc
ore
3
2
1
0
-1
-2
-3
Z-score
17
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BBTX, Parameter: Benzene
Benzene
CB01 B - BTEX / MTBE
µg/l
Unit
Mean ± Cl (99%)
Control test value ± U
Parameter oriented report
Minimum - Maximum
µg/l
2.07 ± 0.2411.6 - 2.891.85 ± 0.135
Labcode Result ± U Recovery [%] z-score Comments
LC0001 1.6 0.25 77.4 -1.37LC0004 2.05 0.41 99.2 -0.05LC0005 2.3 0.5 111 0.68LC0006 - - - -LC0007 2.06 0.37 99.7 -0.02LC0009 1.78 0.445 86.1 -0.84LC0010 2.14 0.412 104 0.21LC0011 2.49 0.5 120 1.24LC0012 2.5 0.3 121 1.27LC0014 2.07 0.13 100 0.01LC0015 1.76 0.405 85.1 -0.9LC0016 1.84 0.34 89 -0.67LC0017 1.96 0.586 94.8 -0.31LC0018 2.44 0.78 118 1.09LC0019 1.63 0.49 78.9 -1.28LC0020 1.91 0.2 92.4 -0.46LC0022 1.94 0.2716 93.9 -0.37LC0024 1.85 0.56 89.5 -0.64LC0025 2.886 - 140 2.4
all results ithout outliers Unit
Mean ± Cl (99%) µg/l
µg/l
µg/l
µg/l
%
-n
rel. Standard deviation
Standard deviation
Maximum
Minimum
Characteristics of parameter
2.07 ± 0.241 2.07 ± 0.241
1.61.6
2.89 2.89
0.3410.341
16.5 16.5
1818
w
18
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BBTX, Parameter: Benzene
Laboratory
LC00
01
LC00
04
LC00
05
LC00
07
LC00
09
LC00
10
LC00
11
LC00
12
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
22
LC00
24
LC00
25
µg/l
3.50
3.25
3.00
2.75
2.50
2.25
2.00
1.75
1.50
1.25
1.00
0.75
0.50
Graphical presentation of resultsResults
19
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BBTX, Parameter: Benzene
Laboratory
LC00
01
LC00
04
LC00
05
LC00
07
LC00
09
LC00
10
LC00
11
LC00
12
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
22
LC00
24
LC00
25
Rec
over
y [%
]
160
140
120
100
80
60
40
Recovery rate
20
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BBTX, Parameter: Benzene
Laboratory
Z-scoreLC
0001
LC00
04
LC00
05
LC00
07
LC00
09
LC00
10
LC00
11
LC00
12
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
22
LC00
24
LC00
25
Z-S
core
3
2
1
0
-1
-2
-3
21
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ABTX, Parameter: Ethylbenzene
Ethylbenzene
CB01 A - BTEX / MTBE
µg/l
Unit
Mean ± Cl (99%)
Control test value ± U
Parameter oriented report
Minimum - Maximum
µg/l
0.442 ± 0.03830.32 - 0.520.424 ± 0.0415
Labcode Result ± U Recovery [%] z-score Comments
LC0001 0.39 0.05 88.3 -1.01LC0004 0.49 0.1 111 0.95LC0005 0.45 0.1 102 0.17LC0006 0.32 0.06 72.5 -2.38LC0007 < 0.5 (LOQ) - - -LC0009 0.22 0.055 49.8 -4.34 HLC0010 0.49 0.09 111 0.95LC0011 0.46 0.09 104 0.36LC0012 0.45 0.1 102 0.17LC0014 0.401 0.04 90.8 -0.8LC0015 0.404 0.093 91.5 -0.74LC0016 0.41 0.08 92.9 -0.62LC0017 0.52 0.155 118 1.54LC0018 0.505 0.187 114 1.24LC0019 0.44 0.13 99.6 -0.03LC0020 0.41 0.04 92.9 -0.62LC0022 0.475 0.04275 108 0.66LC0024 0.45 0.14 102 0.17LC0025 - - - -
all results ithout outliers Unit
Mean ± Cl (99%) µg/l
µg/l
µg/l
µg/l
%
-n
rel. Standard deviation
Standard deviation
Maximum
Minimum
Characteristics of parameter
0.429 ± 0.0531 0.442 ± 0.0383
0.320.22
0.52 0.52
0.0510.073
17 11.6
1617
w
22
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ABTX, Parameter: Ethylbenzene
Laboratory
LC00
01
LC00
04
LC00
05
LC00
06
LC00
07
LC00
09
LC00
10
LC00
11
LC00
12
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
22
LC00
24
µg/l
0.7
0.6
0.5
0.4
0.3
0.2
H
Graphical presentation of resultsResults
23
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ABTX, Parameter: Ethylbenzene
Laboratory
LC00
01
LC00
04
LC00
05
LC00
06
LC00
09
LC00
10
LC00
11
LC00
12
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
22
LC00
24
Rec
over
y [%
]
150
140
130
120
110
100
90
80
70
60
50
Recovery rate
24
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ABTX, Parameter: Ethylbenzene
Laboratory
LC00
01
LC00
04
LC00
05
LC00
06
LC00
09
LC00
10
LC00
11
LC00
12
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
22
LC00
24
z-sc
ore
3
2
1
0
-1
-2
-3
Z-score
25
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BBTX, Parameter: Ethylbenzene
Ethylbenzene
CB01 B - BTEX / MTBE
µg/l
Unit
Mean ± Cl (99%)
Control test value ± U
Parameter oriented report
Minimum - Maximum
µg/l
1.45 ± 0.2010.87 - 21.37 ± 0.228
Labcode Result ± U Recovery [%] z-score Comments
LC0001 1.1 0.08 75.8 -1.27LC0004 1.64 0.33 113 0.69LC0005 1.8 0.4 124 1.27LC0006 0.55 0.06 37.9 -3.26 HLC0007 1.374 0.26 94.7 -0.28LC0009 0.87 0.218 60 -2.1LC0010 1.62 0.298 112 0.61LC0011 1.41 0.28 97.2 -0.15LC0012 1.31 0.1 90.3 -0.51LC0014 1.47 0.15 101 0.07LC0015 1.4 0.322 96.5 -0.18LC0016 1.56 0.29 108 0.4LC0017 1.72 0.515 119 0.98LC0018 2 0.62 138 1.99LC0019 1.08 0.32 74.5 -1.34LC0020 1.33 0.1 91.7 -0.44LC0022 1.51 0.1359 104 0.22LC0024 1.46 0.44 101 0.04LC0025 - - - -
all results ithout outliers Unit
Mean ± Cl (99%) µg/l
µg/l
µg/l
µg/l
%
-n
rel. Standard deviation
Standard deviation
Maximum
Minimum
Characteristics of parameter
1.4 ± 0.242 1.45 ± 0.201
0.870.55
2 2
0.2760.342
24.4 19
1718
w
26
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BBTX, Parameter: Ethylbenzene
Laboratory
LC00
01
LC00
04
LC00
05
LC00
06
LC00
07
LC00
09
LC00
10
LC00
11
LC00
12
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
22
LC00
24
µg/l
2.75
2.50
2.25
2.00
1.75
1.50
1.25
1.00
0.75
0.50
0.25
H
Graphical presentation of resultsResults
27
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BBTX, Parameter: Ethylbenzene
Laboratory
LC00
01
LC00
04
LC00
05
LC00
06
LC00
07
LC00
09
LC00
10
LC00
11
LC00
12
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
22
LC00
24
Rec
over
y [%
]
180
160
140
120
100
80
60
40
20
H
Recovery rate
28
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BBTX, Parameter: Ethylbenzene
Laboratory
LC00
01
LC00
04
LC00
05
LC00
06
LC00
07
LC00
09
LC00
10
LC00
11
LC00
12
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
22
LC00
24
z-sc
ore
3
2
1
0
-1
-2
-3
Z-score
29
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ABTX, Parameter: o-Xylene
o-Xylene
CB01 A - BTEX / MTBE
µg/l
Unit
Mean ± Cl (99%)
Control test value ± U
Parameter oriented report
Minimum - Maximum
µg/l
4.41 ± 0.5132.88 - 5.74.40 ± 0.0906
Labcode Result ± U Recovery [%] z-score Comments
LC0001 4.1 0.19 93 -0.45LC0004 4.82 0.96 109 0.6LC0005 5.4 1.1 122 1.45LC0006 2.14 0.2 48.5 -3.32 HLC0007 4.374 0.89 99.2 -0.05LC0009 2.88 0.72 65.3 -2.24LC0010 4.44 0.813 101 0.04LC0011 4.58 0.92 104 0.25LC0012 7.87 0.8 178 5.06 HLC0014 4.34 0.67 98.4 -0.1LC0015 4.43 1.063 100 0.03LC0016 4.6 0.83 104 0.28LC0017 4.79 1.438 109 0.55LC0018 5.7 1.68 129 1.88LC0019 3.7 1.11 83.9 -1.04LC0020 3.42 0.3 77.5 -1.45LC0022 4.34 0.3906 98.4 -0.1LC0024 4.66 1.4 106 0.36LC0025 - - - -
all results ithout outliers Unit
Mean ± Cl (99%) µg/l
µg/l
µg/l
µg/l
%
-n
rel. Standard deviation
Standard deviation
Maximum
Minimum
Characteristics of parameter
4.48 ± 0.841 4.41 ± 0.513
2.882.14
7.87 5.7
0.6841.19
26.6 15.5
1618
w
30
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ABTX, Parameter: o-Xylene
Laboratory
LC00
01
LC00
04
LC00
05
LC00
06
LC00
07
LC00
09
LC00
10
LC00
11
LC00
12
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
22
LC00
24
µg/l
9
8
7
6
5
4
3
2
1
H
H
Graphical presentation of resultsResults
31
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ABTX, Parameter: o-Xylene
Laboratory
LC00
01
LC00
04
LC00
05
LC00
06
LC00
07
LC00
09
LC00
10
LC00
11
LC00
12
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
22
LC00
24
Rec
over
y [%
]
160
150
140
130
120
110
100
90
80
70
60
50
40
H
Recovery rate
32
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ABTX, Parameter: o-Xylene
Laboratory
Z-scoreLC
0001
LC00
04
LC00
05
LC00
06
LC00
07
LC00
09
LC00
10
LC00
11
LC00
12
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
22
LC00
24
Z-S
core
3
2
1
0
-1
-2
-3
33
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BBTX, Parameter: o-Xylene
o-Xylene
CB01 B - BTEX / MTBE
µg/l
Unit
Mean ± Cl (99%)
Control test value ± U
Parameter oriented report
Minimum - Maximum
µg/l
3.03 ± 0.3632.2 - 4.022.96 ± 0.166
Labcode Result ± U Recovery [%] z-score Comments
LC0001 2.5 0.14 82.4 -1.1LC0004 3.28 0.66 108 0.51LC0005 3.4 0.7 112 0.76LC0006 0.94 0.2 31 -4.33 HLC0007 3.247 0.66 107 0.44LC0009 2.2 0.55 72.5 -1.72LC0010 3.16 0.588 104 0.26LC0011 3.02 0.6 99.6 -0.03LC0012 6.95 0.7 229 8.1 HLC0014 3.03 0.11 99.9 -0.01LC0015 2.98 0.715 98.2 -0.11LC0016 3.22 0.58 106 0.39LC0017 3.67 1.1 121 1.32LC0018 4.02 1.18 133 2.04LC0019 2.34 0.7 77.1 -1.43LC0020 2.43 0.2 80.1 -1.25LC0022 2.95 0.2655 97.2 -0.17LC0024 3.09 0.93 102 0.12LC0025 - - - -
all results ithout outliers Unit
Mean ± Cl (99%) µg/l
µg/l
µg/l
µg/l
%
-n
rel. Standard deviation
Standard deviation
Maximum
Minimum
Characteristics of parameter
3.13 ± 0.823 3.03 ± 0.363
2.20.94
6.95 4.02
0.4841.16
37.1 15.9
1618
w
34
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BBTX, Parameter: o-Xylene
Laboratory
LC00
01
LC00
04
LC00
05
LC00
06
LC00
07
LC00
09
LC00
10
LC00
11
LC00
12
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
22
LC00
24
µg/l
5
4
3
2
1
H
Graphical presentation of resultsResults
35
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BBTX, Parameter: o-Xylene
Laboratory
LC00
01
LC00
04
LC00
05
LC00
06
LC00
07
LC00
09
LC00
10
LC00
11
LC00
12
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
22
LC00
24
Rec
over
y [%
]
160
150
140
130
120
110
100
90
80
70
60
50
40
Recovery rate
36
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BBTX, Parameter: o-Xylene
Laboratory
Z-scoreLC
0001
LC00
04
LC00
05
LC00
06
LC00
07
LC00
09
LC00
10
LC00
11
LC00
12
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
22
LC00
24
Z-S
core
3
2
1
0
-1
-2
-3
37
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ABTX, Parameter: Sum of m-Xylene and p-Xylene
Sum of m-Xylene and p-Xylene
CB01 A - BTEX / MTBE
µg/l
Unit
Mean ± Cl (99%)
Control test value ± U
Parameter oriented report
Minimum - Maximum
µg/l
5.69 ± 0.7193.48 - 7.365.42 ± 0.0167
Labcode Result ± U Recovery [%] z-score Comments
LC0001 5 0.39 87.8 -0.72LC0004 6.53 1.31 115 0.87LC0005 6.5 1.3 114 0.84LC0006 3.48 0.4 61.1 -2.31LC0007 5.714 1.37 100 0.02LC0009 2.95 0.738 51.8 -2.86 HLC0010 5.67 1.05 99.6 -0.02LC0011 6.05 1.21 106 0.37LC0012 7.31 0.7 128 1.69LC0014 5.39 0.96 94.7 -0.32LC0015 5.4 1.674 94.8 -0.31LC0016 5.49 0.99 96.4 -0.21LC0017 5.25 1.576 92.2 -0.46LC0018 7.36 2.41 129 1.74LC0019 2.34 0.7 41.1 -3.5 HLC0020 4.65 0.5 81.7 -1.09LC0022 5.5 0.44 96.6 -0.2LC0024 5.8 1.74 102 0.11LC0025 - - - -
all results ithout outliers Unit
Mean ± Cl (99%) µg/l
µg/l
µg/l
µg/l
%
-n
rel. Standard deviation
Standard deviation
Maximum
Minimum
Characteristics of parameter
5.35 ± 0.947 5.69 ± 0.719
3.482.34
7.36 7.36
0.9591.34
25 16.8
1618
w
38
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ABTX, Parameter: Sum of m-Xylene and p-Xylene
Laboratory
LC00
01
LC00
04
LC00
05
LC00
06
LC00
07
LC00
09
LC00
10
LC00
11
LC00
12
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
22
LC00
24
µg/l
10
9
8
7
6
5
4
3
2
1
H
H
Graphical presentation of resultsResults
39
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ABTX, Parameter: Sum of m-Xylene and p-Xylene
Laboratory
LC00
01
LC00
04
LC00
05
LC00
06
LC00
07
LC00
09
LC00
10
LC00
11
LC00
12
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
22
LC00
24
Rec
over
y [%
]
160
140
120
100
80
60
40
H H
Recovery rate
40
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ABTX, Parameter: Sum of m-Xylene and p-Xylene
Laboratory
Z-scoreLC
0001
LC00
04
LC00
05
LC00
06
LC00
07
LC00
09
LC00
10
LC00
11
LC00
12
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
22
LC00
24
Z-S
core
3
2
1
0
-1
-2
-3
H
41
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BBTX, Parameter: Sum of m-Xylene and p-Xylene
Sum of m-Xylene and p-Xylene
CB01 B - BTEX / MTBE
µg/l
Unit
Mean ± Cl (99%)
Control test value ± U
Parameter oriented report
Minimum - Maximum
µg/l
8.55 ± 1.045.33 - 10.77.89 ± 0.315
Labcode Result ± U Recovery [%] z-score Comments
LC0001 6.9 0.45 80.7 -1.19LC0004 9.7 1.94 113 0.83LC0005 10.7 2.2 125 1.56LC0006 3.44 0.4 40.2 -3.7 HLC0007 9.016 2.16 105 0.34LC0009 5.33 1.333 62.4 -2.33LC0010 9.15 1.629 107 0.44LC0011 8.74 1.75 102 0.14LC0012 9.7 0.9 113 0.83LC0014 8.2 0.78 95.9 -0.25LC0015 7.66 2.375 89.6 -0.64LC0016 8.92 1.61 104 0.27LC0017 8.12 2.44 95 -0.31LC0018 10.3 3.08 121 1.27LC0019 3.59 1.08 42 -3.59 HLC0020 6.75 0.7 79 -1.3LC0022 8.89 0.7112 104 0.25LC0024 8.67 2.6 101 0.09LC0025 - - - -
all results ithout outliers Unit
Mean ± Cl (99%) µg/l
µg/l
µg/l
µg/l
%
-n
rel. Standard deviation
Standard deviation
Maximum
Minimum
Characteristics of parameter
7.99 ± 1.47 8.55 ± 1.04
5.333.44
10.7 10.7
1.382.08
26.1 16.2
1618
w
42
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BBTX, Parameter: Sum of m-Xylene and p-Xylene
Laboratory
LC00
01
LC00
04
LC00
05
LC00
06
LC00
07
LC00
09
LC00
10
LC00
11
LC00
12
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
22
LC00
24
µg/l
14
13
12
11
10
9
8
7
6
5
4
3
2
H
H
Graphical presentation of resultsResults
43
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BBTX, Parameter: Sum of m-Xylene and p-Xylene
Laboratory
LC00
01
LC00
04
LC00
05
LC00
06
LC00
07
LC00
09
LC00
10
LC00
11
LC00
12
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
22
LC00
24
Rec
over
y [%
]
170
160
150
140
130
120
110
100
90
80
70
60
50
40
30
H H
Recovery rate
44
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BBTX, Parameter: Sum of m-Xylene and p-Xylene
Laboratory
LC00
01
LC00
04
LC00
05
LC00
06
LC00
07
LC00
09
LC00
10
LC00
11
LC00
12
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
22
LC00
24
z-sc
ore
3
2
1
0
-1
-2
-3
Z-score
45
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ABTX, Parameter: Toluene
Toluene
CB01 A - BTEX / MTBE
µg/l
Unit
Mean ± Cl (99%)
Control test value ± U
Parameter oriented report
Minimum - Maximum
µg/l
0.85 ± 0.08890.61 - 1.090.818 ± 0.0167
Labcode Result ± U Recovery [%] z-score Comments
LC0001 0.72 0.14 84.7 -1.03LC0004 0.9 0.18 106 0.4LC0005 0.9 0.2 106 0.4LC0006 0.69 0.12 81.2 -1.27LC0007 0.982 0.2 116 1.05LC0009 0.61 0.153 71.8 -1.91LC0010 0.94 0.179 111 0.72LC0011 1 0.2 118 1.19LC0012 0.77 0.1 90.6 -0.63LC0014 0.85 0.07 100 0.00LC0015 0.87 0.252 102 0.16LC0016 0.81 0.15 95.3 -0.32LC0017 0.97 0.292 114 0.96LC0018 1.09 0.38 128 1.91LC0019 0.67 0.2 78.8 -1.43LC0020 0.8 0.08 94.1 -0.4LC0022 0.854 0.07686 100 0.03LC0024 0.87 0.26 102 0.16LC0025 - - - -
all results ithout outliers Unit
Mean ± Cl (99%) µg/l
µg/l
µg/l
µg/l
%
-n
rel. Standard deviation
Standard deviation
Maximum
Minimum
Characteristics of parameter
0.85 ± 0.0889 0.85 ± 0.0889
0.610.61
1.09 1.09
0.1260.126
14.8 14.8
1818
w
46
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ABTX, Parameter: Toluene
Laboratory
LC00
01
LC00
04
LC00
05
LC00
06
LC00
07
LC00
09
LC00
10
LC00
11
LC00
12
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
22
LC00
24
µg/l
1.5
1.4
1.3
1.2
1.1
1.0
0.9
0.8
0.7
0.6
0.5
0.4
0.3
Graphical presentation of resultsResults
47
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ABTX, Parameter: Toluene
Laboratory
LC00
01
LC00
04
LC00
05
LC00
06
LC00
07
LC00
09
LC00
10
LC00
11
LC00
12
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
22
LC00
24
Rec
over
y [%
]
160
150
140
130
120
110
100
90
80
70
60
50
40
Recovery rate
48
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ABTX, Parameter: Toluene
Laboratory
LC00
01
LC00
04
LC00
05
LC00
06
LC00
07
LC00
09
LC00
10
LC00
11
LC00
12
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
22
LC00
24
z-sc
ore
3
2
1
0
-1
-2
-3
Z-score
49
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BBTX, Parameter: Toluene
Toluene
CB01 B - BTEX / MTBE
µg/l
Unit
Mean ± Cl (99%)
Control test value ± U
Parameter oriented report
Minimum - Maximum
µg/l
5.33 ± 0.5124.21 - 6.84.91 ± 0.369
Labcode Result ± U Recovery [%] z-score Comments
LC0001 4.6 0.34 86.3 -1.03LC0004 5.66 1.13 106 0.47LC0005 6.4 1.3 120 1.52LC0006 2.71 0.12 50.9 -3.72 HLC0007 5.599 1.15 105 0.38LC0009 4.21 1.053 79 -1.59LC0010 6.24 1.11 117 1.29LC0011 5.38 1.08 101 0.07LC0012 5.1 0.5 95.7 -0.33LC0014 5.26 0.49 98.7 -0.1LC0015 4.8 1.392 90.1 -0.75LC0016 5.42 0.98 102 0.13LC0017 5.83 1.75 109 0.71LC0018 6.8 2.06 128 2.09LC0019 4.51 1.35 84.6 -1.16LC0020 4.73 0.5 88.8 -0.85LC0022 4.94 0.4446 92.7 -0.55LC0024 5.11 1.53 95.9 -0.31LC0025 - - - -
all results ithout outliers Unit
Mean ± Cl (99%) µg/l
µg/l
µg/l
µg/l
%
-n
rel. Standard deviation
Standard deviation
Maximum
Minimum
Characteristics of parameter
5.18 ± 0.651 5.33 ± 0.512
4.212.71
6.8 6.8
0.7040.921
17.8 13.2
1718
w
50
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BBTX, Parameter: Toluene
Laboratory
LC00
01
LC00
04
LC00
05
LC00
06
LC00
07
LC00
09
LC00
10
LC00
11
LC00
12
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
22
LC00
24
µg/l
9
8
7
6
5
4
3
2
H
Graphical presentation of resultsResults
51
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BBTX, Parameter: Toluene
Laboratory
LC00
01
LC00
04
LC00
05
LC00
06
LC00
07
LC00
09
LC00
10
LC00
11
LC00
12
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
22
LC00
24
Rec
over
y [%
]
160
150
140
130
120
110
100
90
80
70
60
50
40
H
Recovery rate
52
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BBTX, Parameter: Toluene
Laboratory
LC00
01
LC00
04
LC00
05
LC00
06
LC00
07
LC00
09
LC00
10
LC00
11
LC00
12
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
22
LC00
24
z-sc
ore
3
2
1
0
-1
-2
-3
Z-score
53
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ABTX, Parameter: Methyl-tert-butyl-ether
Methyl-tert-butyl-ether
CB01 A - BTEX / MTBE
µg/l
Unit
Mean ± Cl (99%)
Control test value ± U
Parameter oriented report
Minimum - Maximum
µg/l
3.25 ± 0.3612.5 - 4.13.16 ± 0.0805
Labcode Result ± U Recovery [%] z-score Comments
LC0001 - - - -LC0004 3.64 0.73 112 0.89LC0005 4.1 0.9 126 1.95LC0006 - - - -LC0007 3.37 0.71 104 0.27LC0009 2.5 0.625 76.8 -1.74LC0010 3.69 0.708 113 1LC0011 2.74 0.55 84.2 -1.19LC0012 - - - -LC0014 - - - -LC0015 3.46 0.657 106 0.47LC0016 3.1 0.56 95.2 -0.36LC0017 3.25 0.974 99.9 -0.01LC0018 3.49 1.05 107 0.54LC0019 2.96 0.89 90.9 -0.68LC0020 2.9 0.3 89.1 -0.82LC0022 3.11 0.5287 95.6 -0.33LC0024 - - - -LC0025 - - - -
all results ithout outliers Unit
Mean ± Cl (99%) µg/l
µg/l
µg/l
µg/l
%
-n
rel. Standard deviation
Standard deviation
Maximum
Minimum
Characteristics of parameter
3.25 ± 0.361 3.25 ± 0.361
2.52.5
4.1 4.1
0.4340.434
13.3 13.3
1313
w
54
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ABTX, Parameter: Methyl-tert-butyl-ether
Laboratory
LC00
04
LC00
05
LC00
07
LC00
09
LC00
10
LC00
11
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
22
µg/l
5.0
4.5
4.0
3.5
3.0
2.5
2.0
1.5
Graphical presentation of resultsResults
55
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ABTX, Parameter: Methyl-tert-butyl-ether
Laboratory
LC00
04
LC00
05
LC00
07
LC00
09
LC00
10
LC00
11
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
22
Rec
over
y [%
]
160
150
140
130
120
110
100
90
80
70
60
50
40
Recovery rate
56
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ABTX, Parameter: Methyl-tert-butyl-ether
Laboratory
LC00
04
LC00
05
LC00
07
LC00
09
LC00
10
LC00
11
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
22
z-sc
ore
3
2
1
0
-1
-2
-3
Z-score
57
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BBTX, Parameter: Methyl-tert-butyl-ether
Methyl-tert-butyl-ether
CB01 B - BTEX / MTBE
µg/l
Unit
Mean ± Cl (99%)
Control test value ± U
Parameter oriented report
Minimum - Maximum
µg/l
0.882 ± 0.1030.72 - 1.10.880 ± 0.0825
Labcode Result ± U Recovery [%] z-score Comments
LC0001 - - - -LC0004 0.97 0.19 110 0.71LC0005 1.1 0.3 125 1.76LC0006 - - - -LC0007 0.951 0.2 108 0.55LC0009 0.72 0.18 81.6 -1.31LC0010 0.95 0.177 108 0.55LC0011 0.81 0.16 91.8 -0.58LC0012 - - - -LC0014 - - - -LC0015 0.91 0.173 103 0.22LC0016 0.83 0.15 94.1 -0.42LC0017 0.72 0.215 81.6 -1.31LC0018 1.05 0.32 119 1.35LC0019 0.83 0.25 94.1 -0.42LC0020 0.72 0.07 81.6 -1.31LC0022 0.911 0.15487 103 0.23LC0024 - - - -LC0025 - - - -
all results ithout outliers Unit
Mean ± Cl (99%) µg/l
µg/l
µg/l
µg/l
%
-n
rel. Standard deviation
Standard deviation
Maximum
Minimum
Characteristics of parameter
0.882 ± 0.103 0.882 ± 0.103
0.720.72
1.1 1.1
0.1240.124
14 14
1313
w
58
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BBTX, Parameter: Methyl-tert-butyl-ether
Laboratory
LC00
04
LC00
05
LC00
07
LC00
09
LC00
10
LC00
11
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
22
µg/l
1.4
1.3
1.2
1.1
1.0
0.9
0.8
0.7
0.6
0.5
0.4
Graphical presentation of resultsResults
59
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BBTX, Parameter: Methyl-tert-butyl-ether
Laboratory
LC00
04
LC00
05
LC00
07
LC00
09
LC00
10
LC00
11
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
22
Rec
over
y [%
]
160
150
140
130
120
110
100
90
80
70
60
50
40
Recovery rate
60
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BBTX, Parameter: Methyl-tert-butyl-ether
Laboratory
LC00
04
LC00
05
LC00
07
LC00
09
LC00
10
LC00
11
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
22
z-sc
ore
3
2
1
0
-1
-2
-3
Z-score
61
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: 1,1,1-Trichloroethane
1,1,1-Trichloroethane
CB01 A - VOC
µg/l
Unit
Mean ± Cl (99%)
Control test value ± U
Parameter oriented report
Minimum - Maximum
µg/l
-- < 0.0668 (LOD)
Labcode Result ± U Recovery [%] z-score Comments
LC0002 <0.1 (LOD) - - -LC0003 < 0.01 (LOQ) - - -LC0004 <0.06 (LOD) - - -LC0005 < 0.1 (LOQ) - - -LC0006 < 0.1 (LOQ) - - -LC0007 < 0.1 (LOQ) - - -LC0008 <0.004 (LOD) - - -LC0009 - - - -LC0010 < 0.1 (LOQ) - - -LC0011 < 0.05 (LOQ) - - -LC0012 < 0.2 (LOQ) - - -LC0013 < 0.1 (LOQ) - - -LC0014 < 0.1 (LOQ) - - -LC0015 < 0.05 (LOQ) - - -LC0016 < 0.05 (LOQ) - - -LC0017 < 0.2 (LOQ) - - -LC0018 < 0.1 (LOQ) - - -LC0019 < 0.1 (LOQ) - - -LC0020 < 0.08 (LOQ) - - -LC0021 < 0.1 (LOQ) - - -LC0022 < 0.05 (LOQ) - - -LC0023 < 0.02 (LOQ) - - -LC0024 < 0.04 (LOQ) - - -
all results ithout outliers Unit
Mean ± Cl (99%) µg/l
µg/l
µg/l
µg/l
%
-n
rel. Standard deviation
Standard deviation
Maximum
Minimum
Characteristics of parameter
- -
--
- -
--
- -
00
w
62
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: 1,1,1-Trichloroethane
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
µg/l
0.2
0.15
0.1
0.05
0
Graphical presentation of resultsResults
63
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: 1,1,1-Trichloroethane
1,1,1-Trichloroethane
CB01 B - VOC
µg/l
Unit
Mean ± Cl (99%)
Control test value ± U
Parameter oriented report
Minimum - Maximum
µg/l
7.27 ± 0.6715.62 - 9.67.27 ± 0.567
Labcode Result ± U Recovery [%] z-score Comments
LC0002 10.6 - 146 3.17 HLC0003 7.089 0.851 97.5 -0.17LC0004 7.72 1.54 106 0.43LC0005 9.6 2 132 2.22LC0006 5.81 0.7 79.9 -1.39LC0007 6.84 1.231 94.1 -0.41LC0008 7.06 0.447 97.1 -0.2LC0009 5.89 1.473 81 -1.32LC0010 7.55 1.3 104 0.27LC0011 8.37 1.67 115 1.05LC0012 5.62 0.6 77.3 -1.57LC0013 8.42 1.25 116 1.09LC0014 7.24 0.23 99.6 -0.03LC0015 6.06 1.333 83.3 -1.15LC0016 6.82 1.23 93.8 -0.43LC0017 7.72 2.32 106 0.43LC0018 9.3 2.73 128 1.93LC0019 6.56 1.97 90.2 -0.68LC0020 7.17 0.7 98.6 -0.1LC0021 6.51 0.98 89.5 -0.73LC0022 6.99 1.1883 96.1 -0.27LC0023 8.04 1.61 111 0.73LC0024 7.59 2.28 104 0.3
all results ithout outliers Unit
Mean ± Cl (99%) µg/l
µg/l
µg/l
µg/l
%
-n
rel. Standard deviation
Standard deviation
Maximum
Minimum
Characteristics of parameter
7.42 ± 0.774 7.27 ± 0.671
5.625.62
10.6 9.6
1.051.24
16.7 14.4
2223
w
64
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: 1,1,1-Trichloroethane
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
µg/l
12
11
10
9
8
7
6
5
4
3
H
Graphical presentation of resultsResults
65
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: 1,1,1-Trichloroethane
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
Rec
over
y [%
]
160
150
140
130
120
110
100
90
80
70
60
50
40
H
Recovery rate
66
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: 1,1,1-Trichloroethane
Laboratory
Z-scoreLC
0002
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
Z-S
core
3
2
1
0
-1
-2
-3
67
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: 1,1-Dichloroethene
1,1-Dichloroethene
CB01 A - VOC
µg/l
Unit
Mean ± Cl (99%)
Control test value ± U
Parameter oriented report
Minimum - Maximum
µg/l
-- < 0.061 (LOD)
Labcode Result ± U Recovery [%] z-score Comments
LC0002 - - - -LC0003 < 0.1 (LOQ) - - -LC0004 <0.04 (LOD) - - -LC0005 < 0.1 (LOQ) - - -LC0006 - - - -LC0007 < 0.2 (LOQ) - - -LC0008 < 0.05 (LOQ) - - -LC0009 - - - -LC0010 < 0.1 (LOQ) - - -LC0011 < 0.05 (LOQ) - - -LC0012 < 0.2 (LOQ) - - -LC0013 < 0.1 (LOQ) - - -LC0014 < 0.1 (LOQ) - - -LC0015 < 0.08 (LOQ) - - -LC0016 < 0.05 (LOQ) - - -LC0017 < 2 (LOQ) - - -LC0018 < 0.1 (LOQ) - - -LC0019 < 0.1 (LOQ) - - -LC0020 < 0.1 (LOQ) - - -LC0021 < 0.2 (LOQ) - - -LC0022 < 0.05 (LOQ) - - -LC0023 < 0.03 (LOQ) - - -LC0024 < 0.04 (LOQ) - - -
all results ithout outliers Unit
Mean ± Cl (99%) µg/l
µg/l
µg/l
µg/l
%
-n
rel. Standard deviation
Standard deviation
Maximum
Minimum
Characteristics of parameter
- -
--
- -
--
- -
00
w
68
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: 1,1-Dichloroethene
Laboratory
LC00
03
LC00
04
LC00
05
LC00
07
LC00
08
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
µg/l
2.00
1.50
1.00
0.50
0.00
Graphical presentation of resultsResults
69
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: 1,1-Dichloroethene
1,1-Dichloroethene
CB01 B - VOC
µg/l
Unit
Mean ± Cl (99%)
Control test value ± U
Parameter oriented report
Minimum - Maximum
µg/l
9.06 ± 1.34.75 - 12.79.32 ± 0.644
Labcode Result ± U Recovery [%] z-score Comments
LC0002 - - - -LC0003 8.864 1.064 97.8 -0.1LC0004 9.43 1.89 104 0.19LC0005 12.7 2.6 140 1.92LC0006 - - - -LC0007 11.5 3.33 127 1.29LC0008 9.19 0.565 101 0.07LC0009 7.7 1.925 85 -0.72LC0010 10.18 1.68 112 0.59LC0011 15.37 3.07 170 3.33 HLC0012 6.54 0.6 72.2 -1.33LC0013 4.75 0.71 52.4 -2.28LC0014 8.38 0.6 92.5 -0.36LC0015 7.54 1.206 83.2 -0.8LC0016 8.15 1.47 89.9 -0.48LC0017 11.5 3.44 127 1.29LC0018 14.9 4.7 164 3.08 HLC0019 8.84 2.65 97.5 -0.12LC0020 8.88 0.9 98 -0.1LC0021 8.47 1.27 93.5 -0.31LC0022 8.46 0.6768 93.3 -0.32LC0023 9.47 1.89 104 0.21LC0024 11.66 3.5 129 1.37
all results ithout outliers Unit
Mean ± Cl (99%) µg/l
µg/l
µg/l
µg/l
%
-n
rel. Standard deviation
Standard deviation
Maximum
Minimum
Characteristics of parameter
9.64 ± 1.68 9.06 ± 1.3
4.754.75
15.4 12.7
1.92.56
26.6 20.9
1921
w
70
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: 1,1-Dichloroethene
Laboratory
LC00
03
LC00
04
LC00
05
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
µg/l
20.0
17.5
15.0
12.5
10.0
7.5
5.0
2.5
0.0
HH
Graphical presentation of resultsResults
71
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: 1,1-Dichloroethene
Laboratory
LC00
03
LC00
04
LC00
05
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
Rec
over
y [%
]
180
160
140
120
100
80
60
40
20
HH
Recovery rate
72
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: 1,1-Dichloroethene
Laboratory
Z-scoreLC
0003
LC00
04
LC00
05
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
Z-S
core
3
2
1
0
-1
-2
-3
73
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: 1,2-Dichloroethane
1,2-Dichloroethane
CB01 A - VOC
µg/l
Unit
Mean ± Cl (99%)
Control test value ± U
Parameter oriented report
Minimum - Maximum
µg/l
12.7 ± 0.88810.5 - 15.411.9 ± 0.161
Labcode Result ± U Recovery [%] z-score Comments
LC0002 - - - -LC0003 13.65 1.64 108 0.74LC0004 11.85 2.37 93.5 -0.62LC0005 15.2 3.1 120 1.91LC0006 - - - -LC0007 11.568 1.69 91.3 -0.83LC0008 12 0.71 94.7 -0.51LC0009 13.02 3.255 103 0.26LC0010 12.31 1.97 97.1 -0.27LC0011 14.47 2.89 114 1.36LC0012 10.47 1.05 82.6 -1.66LC0013 11.56 1.73 91.2 -0.84LC0014 13.9 1.1 110 0.93LC0015 12.98 4.413 102 0.23LC0016 12.8 2.3 101 0.1LC0017 11.8 3.54 93.1 -0.66LC0018 17.9 3 141 3.95 HLC0019 12.14 3.64 95.8 -0.4LC0020 11.3 1 89.2 -1.04LC0021 11.27 1.69 88.9 -1.06LC0022 15.4 2.464 122 2.06LC0023 12.8 2.5 101 0.1LC0024 12.96 3.89 102 0.22
all results ithout outliers Unit
Mean ± Cl (99%) µg/l
µg/l
µg/l
µg/l
%
-n
rel. Standard deviation
Standard deviation
Maximum
Minimum
Characteristics of parameter
12.9 ± 1.13 12.7 ± 0.888
10.510.5
17.9 15.4
1.321.72
13.3 10.4
2021
w
74
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: 1,2-Dichloroethane
Laboratory
LC00
03
LC00
04
LC00
05
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
µg/l
22
20
18
16
14
12
10
8
6
H
Graphical presentation of resultsResults
75
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: 1,2-Dichloroethane
Laboratory
LC00
03
LC00
04
LC00
05
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
Rec
over
y [%
]
150
140
130
120
110
100
90
80
70
60
H
Recovery rate
76
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: 1,2-Dichloroethane
Laboratory
Z-scoreLC
0003
LC00
04
LC00
05
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
Z-S
core
3
2
1
0
-1
-2
-3
77
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: 1,2-Dichloroethane
1,2-Dichloroethane
CB01 B - VOC
µg/l
Unit
Mean ± Cl (99%)
Control test value ± U
Parameter oriented report
Minimum - Maximum
µg/l
3.11 ± 0.2272.51 - 3.72.85 ± 0.0955
Labcode Result ± U Recovery [%] z-score Comments
LC0002 - - - -LC0003 3.208 0.385 103 0.31LC0004 3.19 0.64 103 0.26LC0005 3.7 0.8 119 1.84LC0006 - - - -LC0007 3.029 0.44 97.5 -0.24LC0008 2.51 0.067 80.8 -1.86LC0009 2.75 0.688 88.5 -1.11LC0010 2.98 0.542 95.9 -0.4LC0011 3.67 0.73 118 1.75LC0012 2.1 0.2 67.6 -3.13 HLC0013 3.42 0.51 110 0.97LC0014 3.21 0.15 103 0.32LC0015 3.17 1.078 102 0.19LC0016 2.85 0.52 91.7 -0.8LC0017 5.05 1.52 163 6.04 HLC0018 4.96 1.61 160 5.76 HLC0019 2.85 0.85 91.7 -0.8LC0020 3.02 0.3 97.2 -0.27LC0021 2.99 0.45 96.2 -0.37LC0022 3.56 0.5696 115 1.41LC0023 3.02 0.6 97.2 -0.27LC0024 2.81 0.84 90.4 -0.93
all results ithout outliers Unit
Mean ± Cl (99%) µg/l
µg/l
µg/l
µg/l
%
-n
rel. Standard deviation
Standard deviation
Maximum
Minimum
Characteristics of parameter
3.24 ± 0.454 3.11 ± 0.227
2.512.1
5.05 3.7
0.3220.693
21.4 10.4
1821
w
78
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: 1,2-Dichloroethane
Laboratory
LC00
03
LC00
04
LC00
05
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
µg/l
4.5
4.0
3.5
3.0
2.5
2.0
1.5
H
Graphical presentation of resultsResults
79
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: 1,2-Dichloroethane
Laboratory
LC00
03
LC00
04
LC00
05
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
Rec
over
y [%
]
140
130
120
110
100
90
80
70
60
H
Recovery rate
80
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: 1,2-Dichloroethane
Laboratory
Z-scoreLC
0003
LC00
04
LC00
05
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
Z-S
core
3
2
1
0
-1
-2
-3
81
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: Bromodichloromethane
Bromodichloromethane
CB01 A - VOC
µg/l
Unit
Mean ± Cl (99%)
Control test value ± U
Parameter oriented report
Minimum - Maximum
µg/l
2.26 ± 0.1411.92 - 2.822.28 ± 0.0986
Labcode Result ± U Recovery [%] z-score Comments
LC0002 2.4 - 106 0.61LC0003 2.245 0.269 99.1 -0.09LC0004 2.35 0.47 104 0.39LC0005 3 0.6 132 3.33 HLC0006 2.12 0.08 93.6 -0.66LC0007 2.393 0.34 106 0.58LC0008 2.24 0.047 98.9 -0.11LC0009 2.29 0.573 101 0.11LC0010 1.99 0.308 87.9 -1.25LC0011 2.55 0.51 113 1.29LC0012 2.24 0.2 98.9 -0.11LC0013 2.66 0.4 117 1.79LC0014 2.44 0.3 108 0.79LC0015 2.12 0.53 93.6 -0.66LC0016 2.05 0.37 90.5 -0.97LC0017 1.92 0.575 84.8 -1.56LC0018 2.82 0.56 125 2.51LC0019 2.04 0.61 90.1 -1.02LC0020 2.08 0.2 91.8 -0.84LC0021 2.18 0.33 96.3 -0.39LC0022 2.2 0.264 97.1 -0.29LC0023 2.36 0.47 104 0.43LC0024 2.14 0.64 94.5 -0.57
all results ithout outliers Unit
Mean ± Cl (99%) µg/l
µg/l
µg/l
µg/l
%
-n
rel. Standard deviation
Standard deviation
Maximum
Minimum
Characteristics of parameter
2.3 ± 0.166 2.26 ± 0.141
1.921.92
3 2.82
0.2210.265
11.5 9.75
2223
w
82
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: Bromodichloromethane
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
µg/l
3.50
3.25
3.00
2.75
2.50
2.25
2.00
1.75
1.50
1.25
1.00
H
Graphical presentation of resultsResults
83
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: Bromodichloromethane
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
Rec
over
y [%
]
140
130
120
110
100
90
80
70
60
H
Recovery rate
84
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: Bromodichloromethane
Laboratory
Z-scoreLC
0002
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
Z-S
core
3
2
1
0
-1
-2
-3
85
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: Bromodichloromethane
Bromodichloromethane
CB01 B - VOC
µg/l
Unit
Mean ± Cl (99%)
Control test value ± U
Parameter oriented report
Minimum - Maximum
µg/l
7.33 ± 0.3466.51 - 8.687.70 ± 0.411
Labcode Result ± U Recovery [%] z-score Comments
LC0002 7.5 - 102 0.32LC0003 7.111 0.853 97 -0.43LC0004 8.07 1.61 110 1.43LC0005 9.1 2 124 3.43 HLC0006 6.75 0.62 92 -1.13LC0007 7.65 1.1 104 0.61LC0008 7.14 0.815 97.3 -0.38LC0009 7.14 1.785 97.3 -0.38LC0010 6.87 1.32 93.7 -0.9LC0011 8.68 1.74 118 2.61LC0012 7.26 0.7 99 -0.14LC0013 10.24 1.5 140 5.64 HLC0014 7.63 0.38 104 0.57LC0015 7.22 1.805 98.4 -0.22LC0016 7.11 1.28 96.9 -0.44LC0017 6.51 1.95 88.8 -1.6LC0018 9.87 2.97 135 4.92 HLC0019 7 2.1 95.4 -0.65LC0020 7.5 0.7 102 0.32LC0021 6.69 1 91.2 -1.25LC0022 7.71 0.9252 105 0.73LC0023 7.92 1.58 108 1.14LC0024 7.23 2.17 98.6 -0.2
all results ithout outliers Unit
Mean ± Cl (99%) µg/l
µg/l
µg/l
µg/l
%
-n
rel. Standard deviation
Standard deviation
Maximum
Minimum
Characteristics of parameter
7.65 ± 0.608 7.33 ± 0.346
6.516.51
10.2 8.68
0.5150.972
12.7 7.03
2023
w
86
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: Bromodichloromethane
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
µg/l
10
9
8
7
6
5
H
H
H
Graphical presentation of resultsResults
87
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: Bromodichloromethane
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
Rec
over
y [%
]
130
120
110
100
90
80
70
H
Recovery rate
88
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: Bromodichloromethane
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
z-sc
ore
3
2
1
0
-1
-2
-3
Z-score
89
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: cis-1,2-Dichloroethene
cis-1,2-Dichloroethene
CB01 A - VOC
µg/l
Unit
Mean ± Cl (99%)
Control test value ± U
Parameter oriented report
Minimum - Maximum
µg/l
0.833 ± 0.0630.68 - 1.010.870 ± 0.04
Labcode Result ± U Recovery [%] z-score Comments
LC0002 - - - -LC0003 0.868 0.104 104 0.36LC0004 0.86 0.17 103 0.28LC0005 0.85 0.17 102 0.17LC0006 - - - -LC0007 0.857 0.22 103 0.24LC0008 0.887 0.059 106 0.56LC0009 0.86 0.215 103 0.28LC0010 0.75 0.143 90 -0.87LC0011 1.01 0.2 121 1.84LC0012 0.68 0.1 81.6 -1.6LC0013 0.933 0.14 112 1.04LC0014 0.804 0.08 96.5 -0.31LC0015 0.68 0.156 81.6 -1.6LC0016 0.72 0.13 86.4 -1.18LC0017 0.81 0.241 97.2 -0.24LC0018 1.01 0.26 121 1.84LC0019 0.71 0.21 85.2 -1.28LC0020 0.82 0.08 98.4 -0.14LC0021 0.73 0.11 87.6 -1.08LC0022 0.843 0.11802 101 0.1LC0023 0.9 0.18 108 0.69LC0024 0.92 0.28 110 0.9
all results ithout outliers Unit
Mean ± Cl (99%) µg/l
µg/l
µg/l
µg/l
%
-n
rel. Standard deviation
Standard deviation
Maximum
Minimum
Characteristics of parameter
0.833 ± 0.063 0.833 ± 0.063
0.680.68
1.01 1.01
0.09620.0962
11.5 11.5
2121
w
90
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: cis-1,2-Dichloroethene
Laboratory
LC00
03
LC00
04
LC00
05
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
µg/l
1.3
1.2
1.1
1.0
0.9
0.8
0.7
0.6
0.5
0.4
Graphical presentation of resultsResults
91
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: cis-1,2-Dichloroethene
Laboratory
LC00
03
LC00
04
LC00
05
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
Rec
over
y [%
]
150
140
130
120
110
100
90
80
70
60
50
Recovery rate
92
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: cis-1,2-Dichloroethene
Laboratory
LC00
03
LC00
04
LC00
05
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
z-sc
ore
3
2
1
0
-1
-2
-3
Z-score
93
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: cis-1,2-Dichloroethene
cis-1,2-Dichloroethene
CB01 B - VOC
µg/l
Unit
Mean ± Cl (99%)
Control test value ± U
Parameter oriented report
Minimum - Maximum
µg/l
4.68 ± 0.3134.02 - 6.074.91 ± 0.414
Labcode Result ± U Recovery [%] z-score Comments
LC0002 - - - -LC0003 4.871 0.585 104 0.41LC0004 4.67 0.93 99.7 -0.03LC0005 5.2 1.1 111 1.14LC0006 - - - -LC0007 4.352 1.1 92.9 -0.73LC0008 4.55 0.054 97.2 -0.29LC0009 4.3 1.075 91.8 -0.84LC0010 4.77 0.754 102 0.19LC0011 6.07 1.21 130 3.05LC0012 4.56 0.5 97.4 -0.27LC0013 6.27 0.94 134 3.49 HLC0014 4.33 0.23 92.5 -0.78LC0015 4.3 0.989 91.8 -0.84LC0016 4.35 0.79 92.9 -0.73LC0017 5.09 1.53 109 0.9LC0018 6.74 1.75 144 4.52 HLC0019 4.02 1.21 85.9 -1.46LC0020 5.03 0.5 107 0.77LC0021 4.67 0.7 99.7 -0.03LC0022 4.54 0.6356 97 -0.31LC0023 4.83 0.96 103 0.33LC0024 4.46 1.34 95.3 -0.49
all results ithout outliers Unit
Mean ± Cl (99%) µg/l
µg/l
µg/l
µg/l
%
-n
rel. Standard deviation
Standard deviation
Maximum
Minimum
Characteristics of parameter
4.86 ± 0.459 4.68 ± 0.313
4.024.02
6.74 6.07
0.4550.702
14.4 9.71
1921
w
94
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: cis-1,2-Dichloroethene
Laboratory
LC00
03
LC00
04
LC00
05
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
µg/l
7
6
5
4
3
2
H
H
Graphical presentation of resultsResults
95
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: cis-1,2-Dichloroethene
Laboratory
LC00
03
LC00
04
LC00
05
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
Rec
over
y [%
]
140
130
120
110
100
90
80
70
60
H
Recovery rate
96
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: cis-1,2-Dichloroethene
Laboratory
LC00
03
LC00
04
LC00
05
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
z-sc
ore
3
2
1
0
-1
-2
-3
Z-score
97
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: Dibromochloromethane
Dibromochloromethane
CB01 A - VOC
µg/l
Unit
Mean ± Cl (99%)
Control test value ± U
Parameter oriented report
Minimum - Maximum
µg/l
1.1 ± 0.08470.87 - 1.41.03 ± 0.0976
Labcode Result ± U Recovery [%] z-score Comments
LC0002 1.4 - 127 2.2LC0003 1.09 0.131 98.9 -0.09LC0004 1.19 0.24 108 0.65LC0005 1.4 0.3 127 2.2LC0006 1.08 0.05 98 -0.16LC0007 1.182 0.2 107 0.59LC0008 1.19 0.041 108 0.65LC0009 1.13 0.283 103 0.2LC0010 0.87 0.148 78.9 -1.71LC0011 1.14 0.23 103 0.28LC0012 0.98 0.1 88.9 -0.9LC0013 1.2 0.18 109 0.72LC0014 1.06 0.24 96.2 -0.31LC0015 1.1 0.286 99.8 -0.02LC0016 0.98 0.18 88.9 -0.9LC0017 1.1 0.329 99.8 -0.02LC0018 1.11 0.2 101 0.06LC0019 0.92 0.27 83.5 -1.35LC0020 1.08 0.1 98 -0.16LC0021 0.98 0.15 88.9 -0.9LC0022 1.07 0.107 97.1 -0.24LC0023 1.21 0.24 110 0.8LC0024 0.89 0.27 80.7 -1.57
all results ithout outliers Unit
Mean ± Cl (99%) µg/l
µg/l
µg/l
µg/l
%
-n
rel. Standard deviation
Standard deviation
Maximum
Minimum
Characteristics of parameter
1.1 ± 0.0847 1.1 ± 0.0847
0.870.87
1.4 1.4
0.1350.135
12.3 12.3
2323
w
98
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: Dibromochloromethane
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
µg/l
1.7
1.6
1.5
1.4
1.3
1.2
1.1
1.0
0.9
0.8
0.7
0.6
0.5
Graphical presentation of resultsResults
99
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: Dibromochloromethane
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
Rec
over
y [%
]
150
140
130
120
110
100
90
80
70
60
50
Recovery rate
100
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: Dibromochloromethane
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
z-sc
ore
3
2
1
0
-1
-2
-3
Z-score
101
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: Dibromochloromethane
Dibromochloromethane
CB01 B - VOC
µg/l
Unit
Mean ± Cl (99%)
Control test value ± U
Parameter oriented report
Minimum - Maximum
µg/l
8.15 ± 0.6986.51 - 10.88.07 ± 0.33
Labcode Result ± U Recovery [%] z-score Comments
LC0002 9.9 - 121 1.57LC0003 7.821 0.939 96 -0.29LC0004 8.89 1.78 109 0.66LC0005 9.6 2 118 1.3LC0006 7.46 0.54 91.5 -0.62LC0007 8.87 1.48 109 0.65LC0008 7.52 0.778 92.3 -0.56LC0009 7.91 1.978 97 -0.22LC0010 7.21 1.15 88.5 -0.84LC0011 8.49 1.7 104 0.3LC0012 8 0.8 98.2 -0.14LC0013 10.82 1.62 133 2.39LC0014 7.25 1.03 89 -0.81LC0015 8.54 2.22 105 0.35LC0016 7.64 1.38 93.7 -0.46LC0017 7.19 2.16 88.2 -0.86LC0018 9.91 3.02 122 1.58LC0019 7.09 2.13 87 -0.95LC0020 8.14 0.8 99.9 -0.01LC0021 7.01 1.05 86 -1.02LC0022 6.51 0.651 79.9 -1.47LC0023 8.69 1.74 107 0.48LC0024 7 2.1 85.9 -1.03
all results ithout outliers Unit
Mean ± Cl (99%) µg/l
µg/l
µg/l
µg/l
%
-n
rel. Standard deviation
Standard deviation
Maximum
Minimum
Characteristics of parameter
8.15 ± 0.698 8.15 ± 0.698
6.516.51
10.8 10.8
1.121.12
13.7 13.7
2323
w
102
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: Dibromochloromethane
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
µg/l
13
12
11
10
9
8
7
6
5
4
3
Graphical presentation of resultsResults
103
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: Dibromochloromethane
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
Rec
over
y [%
]
160
150
140
130
120
110
100
90
80
70
60
50
40
Recovery rate
104
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: Dibromochloromethane
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
z-sc
ore
3
2
1
0
-1
-2
-3
Z-score
105
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: Dichloromethane
Dichloromethane
CB01 A - VOC
µg/l
Unit
Mean ± Cl (99%)
Control test value ± U
Parameter oriented report
Minimum - Maximum
µg/l
1.89 ± 0.1471.66 - 2.41.82 ± 0.169
Labcode Result ± U Recovery [%] z-score Comments
LC0002 - - - -LC0003 2.033 0.244 108 0.67LC0004 1.69 0.34 89.4 -0.94LC0005 2.4 0.5 127 2.39LC0006 - - - -LC0007 1.684 0.3 89.1 -0.97LC0008 1.67 0.08 88.4 -1.03LC0009 1.83 0.458 96.8 -0.28LC0010 1.99 0.311 105 0.47LC0011 0.97 0.19 51.3 -4.31 HLC0012 < 1 (LOQ) - - - FNLC0013 1.88 0.28 99.5 -0.05LC0014 2.04 0.15 108 0.7LC0015 1.823 0.529 96.5 -0.31LC0016 1.67 0.31 88.4 -1.03LC0017 1.69 0.505 89.4 -0.94LC0018 2.25 0.51 119 1.69LC0019 1.76 0.6 93.1 -0.61LC0020 1.66 0.2 87.8 -1.08LC0021 2.04 0.31 108 0.7LC0022 2.09 0.418 111 0.94LC0023 1.77 0.35 93.7 -0.56LC0024 1.94 0.58 103 0.23
all results ithout outliers Unit
Mean ± Cl (99%) µg/l
µg/l
µg/l
µg/l
%
-n
rel. Standard deviation
Standard deviation
Maximum
Minimum
Characteristics of parameter
1.84 ± 0.196 1.89 ± 0.147
1.660.97
2.4 2.4
0.2130.292
15.9 11.3
1920
w
106
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: Dichloromethane
Laboratory
LC00
03
LC00
04
LC00
05
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
µg/l
3.00
2.75
2.50
2.25
2.00
1.75
1.50
1.25
1.00
0.75
H
Graphical presentation of resultsResults
107
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: Dichloromethane
Laboratory
LC00
03
LC00
04
LC00
05
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
Rec
over
y [%
]
150
140
130
120
110
100
90
80
70
60
50
Recovery rate
108
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: Dichloromethane
Laboratory
LC00
03
LC00
04
LC00
05
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
z-sc
ore
3
2
1
0
-1
-2
-3
Z-score
109
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: Dichloromethane
Dichloromethane
CB01 B - VOC
µg/l
Unit
Mean ± Cl (99%)
Control test value ± U
Parameter oriented report
Minimum - Maximum
µg/l
8.04 ± 0.6115.92 - 9.398.07 ± 0.758
Labcode Result ± U Recovery [%] z-score Comments
LC0002 - - - -LC0003 8.697 1.044 108 0.79LC0004 8.04 1.61 100 0.00LC0005 11.9 2.4 148 4.6 HLC0006 - - - -LC0007 7.148 1.45 88.9 -1.06LC0008 7.13 0.307 88.7 -1.08LC0009 7.91 1.978 98.4 -0.15LC0010 8.72 1.69 108 0.81LC0011 4.58 0.92 57 -4.12 HLC0012 5.92 0.6 73.7 -2.52LC0013 11.29 1.7 140 3.87 HLC0014 8.29 0.54 103 0.3LC0015 8 2.32 99.5 -0.05LC0016 7.23 1.31 89.9 -0.96LC0017 9.39 2.82 117 1.61LC0018 11.9 3.6 148 4.6 HLC0019 8.03 2.41 99.9 -0.01LC0020 8.7 0.9 108 0.79LC0021 7.83 1.17 97.4 -0.25LC0022 8.3 1.66 103 0.31LC0023 8.2 1.64 102 0.19LC0024 9.11 2.73 113 1.28
all results ithout outliers Unit
Mean ± Cl (99%) µg/l
µg/l
µg/l
µg/l
%
-n
rel. Standard deviation
Standard deviation
Maximum
Minimum
Characteristics of parameter
8.4 ± 1.14 8.04 ± 0.611
5.924.58
11.9 9.39
0.841.75
20.8 10.5
1721
w
110
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: Dichloromethane
Laboratory
LC00
03
LC00
04
LC00
05
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
µg/l
13
12
11
10
9
8
7
6
5
4
H
H
H
H
Graphical presentation of resultsResults
111
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: Dichloromethane
Laboratory
LC00
03
LC00
04
LC00
05
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
Rec
over
y [%
]
150
140
130
120
110
100
90
80
70
60
50
H
Recovery rate
112
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: Dichloromethane
Laboratory
LC00
03
LC00
04
LC00
05
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
z-sc
ore
3
2
1
0
-1
-2
-3
Z-score
113
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: Tetrachloroethene
Tetrachloroethene
CB01 A - VOC
µg/l
Unit
Mean ± Cl (99%)
Control test value ± U
Parameter oriented report
Minimum - Maximum
µg/l
3.94 ± 0.4942.35 - 5.724.12 ± 0.25
Labcode Result ± U Recovery [%] z-score Comments
LC0002 6.9 - 175 3.83 HLC0003 3.999 0.48 101 0.07LC0004 4.52 0.9 115 0.75LC0005 5.4 1.1 137 1.89LC0006 2.35 0.26 59.6 -2.06LC0007 3.895 0.72 98.8 -0.06LC0008 3.8 0.059 96.4 -0.18LC0009 3.49 0.873 88.5 -0.58LC0010 4.32 0.652 110 0.49LC0011 4.73 0.95 120 1.02LC0012 3.34 0.3 84.7 -0.78LC0013 5.72 0.86 145 2.3LC0014 3.65 0.63 92.6 -0.38LC0015 3.38 1.115 85.8 -0.73LC0016 3.6 0.65 91.3 -0.44LC0017 3.9 1.17 98.9 -0.05LC0018 4.95 1.73 126 1.3LC0019 3.38 1.01 85.8 -0.73LC0020 3.25 0.3 82.5 -0.9LC0021 3.56 0.53 90.3 -0.49LC0022 3.24 0.972 82.2 -0.91LC0023 4.17 0.83 106 0.3LC0024 4.07 1.22 103 0.17
all results ithout outliers Unit
Mean ± Cl (99%) µg/l
µg/l
µg/l
µg/l
%
-n
rel. Standard deviation
Standard deviation
Maximum
Minimum
Characteristics of parameter
4.07 ± 0.61 3.94 ± 0.494
2.352.35
6.9 5.72
0.7730.975
24 19.6
2223
w
114
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: Tetrachloroethene
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
µg/l
7
6
5
4
3
2
1
H
Graphical presentation of resultsResults
115
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: Tetrachloroethene
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
Rec
over
y [%
]
180
160
140
120
100
80
60
40
20
H
Recovery rate
116
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: Tetrachloroethene
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
z-sc
ore
3
2
1
0
-1
-2
-3
Z-score
117
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: Tetrachloroethene
Tetrachloroethene
CB01 B - VOC
µg/l
Unit
Mean ± Cl (99%)
Control test value ± U
Parameter oriented report
Minimum - Maximum
µg/l
11.3 ± 1.388.18 - 15.711.78 ± 0.489
Labcode Result ± U Recovery [%] z-score Comments
LC0002 15.4 - 136 1.85LC0003 10.45 1.254 92.4 -0.39LC0004 12.34 2.47 109 0.46LC0005 15.1 3.1 133 1.71LC0006 10.74 3.42 94.9 -0.26LC0007 11.148 2.06 98.5 -0.08LC0008 11.8 0.327 104 0.22LC0009 9.07 2.268 80.2 -1.02LC0010 11.38 1.81 101 0.03LC0011 13.84 2.77 122 1.14LC0012 8.18 0.8 72.3 -1.42LC0013 15.74 2.36 139 2LC0014 9.36 0.89 82.7 -0.89LC0015 8.46 2.792 74.8 -1.29LC0016 10.4 1.9 91.9 -0.41LC0017 12.6 3.78 111 0.58LC0018 13.6 4.9 120 1.03LC0019 9.68 2.9 85.6 -0.74LC0020 10.1 1 89.3 -0.55LC0021 9.32 1.4 82.4 -0.9LC0022 9.02 2.706 79.7 -1.04LC0023 11.4 2.28 101 0.04LC0024 11.11 3.33 98.2 -0.09
all results ithout outliers Unit
Mean ± Cl (99%) µg/l
µg/l
µg/l
µg/l
%
-n
rel. Standard deviation
Standard deviation
Maximum
Minimum
Characteristics of parameter
11.3 ± 1.38 11.3 ± 1.38
8.188.18
15.7 15.7
2.212.21
19.5 19.5
2323
w
118
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: Tetrachloroethene
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
µg/l
20
18
16
14
12
10
8
6
4
2
Graphical presentation of resultsResults
119
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: Tetrachloroethene
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
Rec
over
y [%
]
180
160
140
120
100
80
60
40
20
Recovery rate
120
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: Tetrachloroethene
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
z-sc
ore
3
2
1
0
-1
-2
-3
Z-score
121
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: Tetrachloromethane
Tetrachloromethane
CB01 A - VOC
µg/l
Unit
Mean ± Cl (99%)
Control test value ± U
Parameter oriented report
Minimum - Maximum
µg/l
1.19 ± 0.1270.75 - 1.631.33 ± 0.0853
Labcode Result ± U Recovery [%] z-score Comments
LC0002 1.5 - 126 1.52LC0003 1.284 0.154 108 0.46LC0004 1.51 0.3 127 1.57LC0005 1.5 0.3 126 1.52LC0006 0.75 0.04 63 -2.16LC0007 1.095 0.19 92 -0.47LC0008 1.14 0.025 95.8 -0.25LC0009 1.06 0.265 89.1 -0.64LC0010 1.31 0.237 110 0.59LC0011 1.2 0.24 101 0.05LC0012 0.95 0.1 79.8 -1.18LC0013 1.155 0.17 97 -0.17LC0014 1.263 0.09 106 0.36LC0015 0.97 0.184 81.5 -1.08LC0016 1.03 0.19 86.5 -0.79LC0017 1.13 0.34 94.9 -0.3LC0018 1.63 0.44 137 2.16LC0019 1.12 0.34 94.1 -0.34LC0020 1.25 0.1 105 0.29LC0021 1.03 0.15 86.5 -0.79LC0022 1.13 0.2034 94.9 -0.3LC0023 1.17 0.23 98.3 -0.1LC0024 1.2 0.36 101 0.05
all results ithout outliers Unit
Mean ± Cl (99%) µg/l
µg/l
µg/l
µg/l
%
-n
rel. Standard deviation
Standard deviation
Maximum
Minimum
Characteristics of parameter
1.19 ± 0.127 1.19 ± 0.127
0.750.75
1.63 1.63
0.2040.204
17.1 17.1
2323
w
122
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: Tetrachloromethane
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
µg/l
2.2
2.0
1.8
1.6
1.4
1.2
1.0
0.8
0.6
0.4
0.2
Graphical presentation of resultsResults
123
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: Tetrachloromethane
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
Rec
over
y [%
]
180
160
140
120
100
80
60
40
20
Recovery rate
124
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: Tetrachloromethane
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
z-sc
ore
3
2
1
0
-1
-2
-3
Z-score
125
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: Tetrachloromethane
Tetrachloromethane
CB01 B - VOC
µg/l
Unit
Mean ± Cl (99%)
Control test value ± U
Parameter oriented report
Minimum - Maximum
µg/l
16.1 ± 2.377.28 - 24.218.3 ± 1.01
Labcode Result ± U Recovery [%] z-score Comments
LC0002 14.9 - 92.7 -0.31LC0003 17.22 2.066 107 0.3LC0004 19.43 3.89 121 0.89LC0005 24.2 4.9 151 2.15LC0006 13.82 1.34 86 -0.59LC0007 16.45 2.83 102 0.1LC0008 18.3 0.4 114 0.59LC0009 12.96 3.24 80.7 -0.82LC0010 7.28 1.22 45.3 -2.32LC0011 14.75 2.95 91.8 -0.35LC0012 11.24 1.1 70 -1.28LC0013 20.7 3.1053 129 1.22LC0014 18.3 1.1 114 0.59LC0015 10.9 2.071 67.8 -1.37LC0016 15.9 2.9 99 -0.04LC0017 16.6 4.97 103 0.14LC0018 22 6.6 137 1.57LC0019 13.04 3.91 81.2 -0.8LC0020 16.3 2 101 0.06LC0021 13.25 1.99 82.5 -0.74LC0022 16.4 2.952 102 0.09LC0023 17.1 3.4 106 0.27LC0024 18.46 5.54 115 0.63
all results ithout outliers Unit
Mean ± Cl (99%) µg/l
µg/l
µg/l
µg/l
%
-n
rel. Standard deviation
Standard deviation
Maximum
Minimum
Characteristics of parameter
16.1 ± 2.37 16.1 ± 2.37
7.287.28
24.2 24.2
3.783.78
23.6 23.6
2323
w
126
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: Tetrachloromethane
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
µg/l
35
30
25
20
15
10
5
0
Graphical presentation of resultsResults
127
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: Tetrachloromethane
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
Rec
over
y [%
]
200
180
160
140
120
100
80
60
40
20
0
Recovery rate
128
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: Tetrachloromethane
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
z-sc
ore
3
2
1
0
-1
-2
-3
Z-score
129
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: trans-1,2-Dichloroethene
trans-1,2-Dichloroethene
CB01 A - VOC
µg/l
Unit
Mean ± Cl (99%)
Control test value ± U
Parameter oriented report
Minimum - Maximum
µg/l
0.852 ± 0.1120.514 - 1.20.794 ± 0.0482
Labcode Result ± U Recovery [%] z-score Comments
LC0002 - - - -LC0003 0.932 0.112 109 0.47LC0004 0.81 0.16 95.1 -0.25LC0005 1.2 0.3 141 2.04LC0006 - - - -LC0007 0.855 0.25 100 0.02LC0008 0.514 0.071 60.3 -1.98LC0009 0.73 0.183 85.7 -0.71LC0010 0.72 0.131 84.5 -0.77LC0011 1.18 0.24 138 1.92LC0012 0.72 0.1 84.5 -0.77LC0013 1.04 0.16 122 1.1LC0014 0.816 0.07 95.8 -0.21LC0015 0.795 0.207 93.3 -0.33LC0016 0.65 0.12 76.3 -1.18LC0017 0.9 0.269 106 0.28LC0018 1.07 0.34 126 1.28LC0019 0.72 0.22 84.5 -0.77LC0020 0.83 0.08 97.4 -0.13LC0021 0.79 0.12 92.7 -0.36LC0022 0.78 0.0624 91.5 -0.42LC0023 0.83 0.17 97.4 -0.13LC0024 1.01 0.3 119 0.93
all results ithout outliers Unit
Mean ± Cl (99%) µg/l
µg/l
µg/l
µg/l
%
-n
rel. Standard deviation
Standard deviation
Maximum
Minimum
Characteristics of parameter
0.852 ± 0.112 0.852 ± 0.112
0.5140.514
1.2 1.2
0.1710.171
20.1 20.1
2121
w
130
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: trans-1,2-Dichloroethene
Laboratory
LC00
03
LC00
04
LC00
05
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
µg/l
1.6
1.4
1.2
1.0
0.8
0.6
0.4
0.2
Graphical presentation of resultsResults
131
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: trans-1,2-Dichloroethene
Laboratory
LC00
03
LC00
04
LC00
05
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
Rec
over
y [%
]
180
160
140
120
100
80
60
40
20
Recovery rate
132
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: trans-1,2-Dichloroethene
Laboratory
LC00
03
LC00
04
LC00
05
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
z-sc
ore
3
2
1
0
-1
-2
-3
Z-score
133
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: trans-1,2-Dichloroethene
trans-1,2-Dichloroethene
CB01 B - VOC
µg/l
Unit
Mean ± Cl (99%)
Control test value ± U
Parameter oriented report
Minimum - Maximum
µg/l
3.72 ± 0.4462.99 - 5.343.55 ± 0.264
Labcode Result ± U Recovery [%] z-score Comments
LC0002 - - - -LC0003 4.169 0.5 112 0.68LC0004 3.5 0.7 94.2 -0.33LC0005 4.8 1 129 1.63LC0006 - - - -LC0007 3.418 1.01 92 -0.45LC0008 3.13 0.068 84.2 -0.88LC0009 2.99 0.748 80.5 -1.09LC0010 3.12 0.583 84 -0.9LC0011 5.64 1.13 152 2.89 HLC0012 3.25 0.3 87.5 -0.7LC0013 4.75 0.71 128 1.56LC0014 3.39 0.29 91.2 -0.49LC0015 3.04 0.79 81.8 -1.02LC0016 3.13 0.57 84.2 -0.88LC0017 3.91 1.17 105 0.29LC0018 5.34 1.6 144 2.44LC0019 3.1 0.93 83.4 -0.93LC0020 3.93 0.4 106 0.32LC0021 3.94 0.59 106 0.34LC0022 3.36 0.2688 90.4 -0.54LC0023 4.04 0.81 109 0.49LC0024 4.01 1.2 108 0.44
all results ithout outliers Unit
Mean ± Cl (99%) µg/l
µg/l
µg/l
µg/l
%
-n
rel. Standard deviation
Standard deviation
Maximum
Minimum
Characteristics of parameter
3.81 ± 0.506 3.72 ± 0.446
2.992.99
5.64 5.34
0.6650.772
20.3 17.9
2021
w
134
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: trans-1,2-Dichloroethene
Laboratory
LC00
03
LC00
04
LC00
05
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
µg/l
7
6
5
4
3
2
1
H
Graphical presentation of resultsResults
135
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: trans-1,2-Dichloroethene
Laboratory
LC00
03
LC00
04
LC00
05
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
Rec
over
y [%
]
180
160
140
120
100
80
60
40
20
H
Recovery rate
136
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: trans-1,2-Dichloroethene
Laboratory
LC00
03
LC00
04
LC00
05
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
z-sc
ore
3
2
1
0
-1
-2
-3
H
Z-score
137
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: Tribromomethane
Tribromomethane
CB01 A - VOC
µg/l
Unit
Mean ± Cl (99%)
Control test value ± U
Parameter oriented report
Minimum - Maximum
µg/l
-0.03 - 0.03< 0.14 (LOD)
Labcode Result ± U Recovery [%] z-score Comments
LC0002 <0.2 (LOD) - - -LC0003 < 0.1 (LOQ) - - -LC0004 < 0.07 (LOQ) - - -LC0005 < 0.1 (LOQ) - - -LC0006 - - - -LC0007 < 0.1 (LOQ) - - -LC0008 < 0.04 (LOQ) - - -LC0009 - - - -LC0010 < 0.1 (LOQ) - - -LC0011 < 0.05 (LOQ) - - -LC0012 < 0.2 (LOQ) - - -LC0013 < 0.1 (LOQ) - - -LC0014 < 0.1 (LOQ) - - -LC0015 < 0.05 (LOQ) - - -LC0016 < 0.05 (LOQ) - - -LC0017 < 0.5 (LOQ) - - -LC0018 < 0.1 (LOQ) - - -LC0019 < 0.1 (LOQ) - - -LC0020 < 0.08 (LOQ) - - -LC0021 < 0.1 (LOQ) - - -LC0022 < 0.05 (LOQ) - - -LC0023 0.03 0.01 - -LC0024 < 0.04 (LOQ) - - -
all results ithout outliers Unit
Mean ± Cl (99%) µg/l
µg/l
µg/l
µg/l
%
-n
rel. Standard deviation
Standard deviation
Maximum
Minimum
Characteristics of parameter
0.03 -
0.030.03
0.03 0.03
--
- -
11
w
138
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: Tribromomethane
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
07
LC00
08
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
µg/l
0.5
0.4
0.3
0.2
0.1
0
Graphical presentation of resultsResults
139
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: Tribromomethane
Tribromomethane
CB01 B - VOC
µg/l
Unit
Mean ± Cl (99%)
Control test value ± U
Parameter oriented report
Minimum - Maximum
µg/l
3.43 ± 0.2452.83 - 43.24 ± 0.0735
Labcode Result ± U Recovery [%] z-score Comments
LC0002 3.7 - 108 0.73LC0003 3.236 0.388 94.4 -0.51LC0004 3.99 0.8 116 1.51LC0005 4 0.8 117 1.53LC0006 - - - -LC0007 3.53 0.65 103 0.28LC0008 3.95 0.056 115 1.4LC0009 3.35 0.838 97.8 -0.21LC0010 3.2 0.522 93.4 -0.61LC0011 3.51 0.7 102 0.22LC0012 3.48 0.4 102 0.14LC0013 4.93 0.74 144 4.02 HLC0014 3.34 0.13 97.5 -0.23LC0015 3.68 1.104 107 0.68LC0016 3.67 0.67 107 0.65LC0017 2.83 0.85 82.6 -1.6LC0018 3.4 1.04 99.2 -0.07LC0019 2.91 0.87 84.9 -1.38LC0020 3.32 0.3 96.9 -0.29LC0021 3.07 0.46 89.6 -0.96LC0022 3 0.24 87.5 -1.14LC0023 3.95 0.79 115 1.4LC0024 2.85 0.86 83.2 -1.54
all results ithout outliers Unit
Mean ± Cl (99%) µg/l
µg/l
µg/l
µg/l
%
-n
rel. Standard deviation
Standard deviation
Maximum
Minimum
Characteristics of parameter
3.5 ± 0.31 3.43 ± 0.245
2.832.83
4.93 4
0.3740.485
13.9 10.9
2122
w
140
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: Tribromomethane
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
µg/l
5.5
5.0
4.5
4.0
3.5
3.0
2.5
2.0
1.5
H
Graphical presentation of resultsResults
141
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: Tribromomethane
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
Rec
over
y [%
]
150
140
130
120
110
100
90
80
70
60
50
Recovery rate
142
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: Tribromomethane
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
z-sc
ore
3
2
1
0
-1
-2
-3
Z-score
143
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: Trichloroethene
Trichloroethene
CB01 A - VOC
µg/l
Unit
Mean ± Cl (99%)
Control test value ± U
Parameter oriented report
Minimum - Maximum
µg/l
0.652 ± 0.05060.5 - 0.820.69 ± 0.0167
Labcode Result ± U Recovery [%] z-score Comments
LC0002 0.8 - 123 1.88LC0003 0.654 0.079 100 0.03LC0004 0.73 0.15 112 0.99LC0005 0.6 0.12 92.1 -0.65LC0006 0.5 0.02 76.7 -1.92LC0007 0.636 0.12 97.6 -0.2LC0008 0.631 0.135 96.8 -0.26LC0009 0.59 0.148 90.6 -0.78LC0010 0.7 0.118 107 0.61LC0011 0.69 0.14 106 0.49LC0012 0.54 0.1 82.9 -1.41LC0013 0.82 0.12 126 2.13LC0014 0.591 0.05 90.7 -0.77LC0015 0.63 0.208 96.7 -0.27LC0016 0.58 0.11 89 -0.91LC0017 0.67 0.2 103 0.23LC0018 0.965 0.292 148 3.96 HLC0019 0.58 0.17 89 -0.91LC0020 0.66 0.07 101 0.11LC0021 0.63 0.09 96.7 -0.27LC0022 0.642 0.1605 98.5 -0.12LC0023 0.75 0.15 115 1.24LC0024 0.71 0.21 109 0.74
all results ithout outliers Unit
Mean ± Cl (99%) µg/l
µg/l
µg/l
µg/l
%
-n
rel. Standard deviation
Standard deviation
Maximum
Minimum
Characteristics of parameter
0.665 ± 0.0633 0.652 ± 0.0506
0.50.5
0.965 0.82
0.07910.101
15.2 12.1
2223
w
144
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: Trichloroethene
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
µg/l
1.0
0.9
0.8
0.7
0.6
0.5
0.4
0.3
H
Graphical presentation of resultsResults
145
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: Trichloroethene
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
Rec
over
y [%
]
150
140
130
120
110
100
90
80
70
60
50
H
Recovery rate
146
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: Trichloroethene
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
z-sc
ore
3
2
1
0
-1
-2
-3
Z-score
147
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: Trichloroethene
Trichloroethene
CB01 B - VOC
µg/l
Unit
Mean ± Cl (99%)
Control test value ± U
Parameter oriented report
Minimum - Maximum
µg/l
3.9 ± 0.3243.08 - 5.194.10 ± 0.112
Labcode Result ± U Recovery [%] z-score Comments
LC0002 5.6 - 144 3.35 HLC0003 3.91 0.469 100 0.02LC0004 4.29 0.86 110 0.77LC0005 3.3 0.7 84.6 -1.18LC0006 3.65 0.56 93.6 -0.49LC0007 3.478 0.66 89.2 -0.83LC0008 3.68 0.133 94.3 -0.43LC0009 3.08 0.77 79 -1.62LC0010 4.01 0.674 103 0.21LC0011 4.21 0.84 108 0.61LC0012 3.56 0.4 91.3 -0.67LC0013 5.19 0.77 133 2.54LC0014 3.53 0.24 90.5 -0.73LC0015 3.72 1.228 95.4 -0.36LC0016 3.81 0.69 97.7 -0.18LC0017 4.49 1.35 115 1.16LC0018 4.86 1.38 125 1.89LC0019 3.38 1.01 86.6 -1.03LC0020 4.24 0.4 109 0.67LC0021 3.68 0.55 94.3 -0.43LC0022 3.67 0.9175 94.1 -0.46LC0023 4.22 0.84 108 0.63LC0024 3.86 1.16 99 -0.08
all results ithout outliers Unit
Mean ± Cl (99%) µg/l
µg/l
µg/l
µg/l
%
-n
rel. Standard deviation
Standard deviation
Maximum
Minimum
Characteristics of parameter
3.97 ± 0.381 3.9 ± 0.324
3.083.08
5.6 5.19
0.5070.609
15.3 13
2223
w
148
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: Trichloroethene
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
µg/l
6.5
6.0
5.5
5.0
4.5
4.0
3.5
3.0
2.5
2.0
H
Graphical presentation of resultsResults
149
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: Trichloroethene
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
Rec
over
y [%
]
150
140
130
120
110
100
90
80
70
60
50
H
Recovery rate
150
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: Trichloroethene
Laboratory
Z-scoreLC
0002
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
Z-S
core
3
2
1
0
-1
-2
-3
151
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: Trichloromethane
Trichloromethane
CB01 A - VOC
µg/l
Unit
Mean ± Cl (99%)
Control test value ± U
Parameter oriented report
Minimum - Maximum
µg/l
9.95 ± 0.6648.32 - 12.59.81 ± 0.128
Labcode Result ± U Recovery [%] z-score Comments
LC0002 11.4 - 115 1.4LC0003 9.653 1.158 97 -0.28LC0004 10.01 2 101 0.06LC0005 11.5 2.3 116 1.49LC0006 9.27 0.5 93.2 -0.65LC0007 10.2 1.68 103 0.24LC0008 8.32 0.198 83.6 -1.57LC0009 9.86 2.465 99.1 -0.09LC0010 9.83 1.71 98.8 -0.12LC0011 11.07 2.21 111 1.08LC0012 8.6 0.8 86.4 -1.3LC0013 10.11 1.5 102 0.15LC0014 12.5 1.4 126 2.46LC0015 8.7 2.349 87.4 -1.2LC0016 9.69 1.75 97.4 -0.25LC0017 8.89 2.67 89.4 -1.02LC0018 13.1 3.2 132 3.03 HLC0019 9.53 2.86 95.8 -0.4LC0020 9.35 1 94 -0.58LC0021 8.92 1.34 89.7 -0.99LC0022 10.5 1.575 106 0.53LC0023 10.4 2.1 105 0.43LC0024 10.58 3.18 106 0.61
all results ithout outliers Unit
Mean ± Cl (99%) µg/l
µg/l
µg/l
µg/l
%
-n
rel. Standard deviation
Standard deviation
Maximum
Minimum
Characteristics of parameter
10.1 ± 0.756 9.95 ± 0.664
8.328.32
13.1 12.5
1.041.21
12 10.4
2223
w
152
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: Trichloromethane
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
µg/l
15
14
13
12
11
10
9
8
7
6
5
H
Graphical presentation of resultsResults
153
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: Trichloromethane
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
Rec
over
y [%
]
140
130
120
110
100
90
80
70
60
H
Recovery rate
154
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01ACKW, Parameter: Trichloromethane
Laboratory
Z-scoreLC
0002
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
Z-S
core
3
2
1
0
-1
-2
-3
155
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: Trichloromethane
Trichloromethane
CB01 B - VOC
µg/l
Unit
Mean ± Cl (99%)
Control test value ± U
Parameter oriented report
Minimum - Maximum
µg/l
1.04 ± 0.0720.84 - 1.261.20 ± 0.105
Labcode Result ± U Recovery [%] z-score Comments
LC0002 1.1 - 106 0.58LC0003 0.997 0.12 96.2 -0.35LC0004 1.09 0.22 105 0.49LC0005 1.4 0.3 135 3.31 HLC0006 0.97 0.1 93.6 -0.6LC0007 1.09 0.18 105 0.49LC0008 0.847 0.066 81.8 -1.72LC0009 1.1 0.275 106 0.58LC0010 1.95 0.311 188 8.32 HLC0011 1.26 0.25 122 2.04LC0012 0.86 0.1 83 -1.6LC0013 1.16 0.17 112 1.13LC0014 1.04 0.09 100 0.04LC0015 1.02 0.275 98.5 -0.14LC0016 0.84 0.16 81.1 -1.78LC0017 1.02 0.307 98.5 -0.14LC0018 1.16 0.35 112 1.13LC0019 1.03 0.31 99.4 -0.05LC0020 1.14 0.1 110 0.95LC0021 0.92 0.14 88.8 -1.05LC0022 1.01 0.1515 97.5 -0.24LC0023 1.12 0.22 108 0.77LC0024 0.98 0.3 94.6 -0.51
all results ithout outliers Unit
Mean ± Cl (99%) µg/l
µg/l
µg/l
µg/l
%
-n
rel. Standard deviation
Standard deviation
Maximum
Minimum
Characteristics of parameter
1.09 ± 0.142 1.04 ± 0.072
0.840.84
1.95 1.26
0.110.228
20.8 10.6
2123
w
156
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: Trichloromethane
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
µg/l
1.6
1.5
1.4
1.3
1.2
1.1
1.0
0.9
0.8
0.7
0.6
0.5
H
Graphical presentation of resultsResults
157
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: Trichloromethane
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
Rec
over
y [%
]
150
140
130
120
110
100
90
80
70
60
50
H
Recovery rate
158
Parameter oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Sample: CB01BCKW, Parameter: Trichloromethane
Laboratory
LC00
02
LC00
03
LC00
04
LC00
05
LC00
06
LC00
07
LC00
08
LC00
09
LC00
10
LC00
11
LC00
12
LC00
13
LC00
14
LC00
15
LC00
16
LC00
17
LC00
18
LC00
19
LC00
20
LC00
21
LC00
22
LC00
23
LC00
24
z-sc
ore
3
2
1
0
-1
-2
-3
Z-score
159
_________________________________________________________________________________
8 Laboratory oriented report
The laboratory oriented report is sorted by laboratory code.
160
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0001
The following results were achieved:
CB01ABTXParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
Benzene µg/l 0.5 86.20.050.0519 0.06920.58 ± -1.16
Ethylbenzene µg/l 0.39 88.30.050.0383 0.0510.442 ± -1.01
o-Xylene µg/l 4.1 930.190.513 0.6844.41 ± -0.45
Sum of m-Xylene and p-Xylene µg/l 5 87.80.390.719 0.9595.69 ± -0.72
Toluene µg/l 0.72 84.70.140.0889 0.1260.85 ± -1.03
Methyl-tert-butyl-ether µg/l - --0.361 0.4343.25 ± -
CB01BBTXParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
Benzene µg/l 1.6 77.40.250.241 0.3412.07 ± -1.37
Ethylbenzene µg/l 1.1 75.80.080.201 0.2761.45 ± -1.27
o-Xylene µg/l 2.5 82.40.140.363 0.4843.03 ± -1.1
Sum of m-Xylene and p-Xylene µg/l 6.9 80.70.451.04 1.388.55 ± -1.19
Toluene µg/l 4.6 86.30.340.512 0.7045.33 ± -1.03
Methyl-tert-butyl-ether µg/l - --0.103 0.1240.882 ± -
This report was created with PROLab, a software by QuoData: http://www.quodata.de/
161
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0001
z-score-2 0 2 -2 0 2
Sample
CB01 A - BTEX / MTBE CB01 B - BTEX / MTBE
Mea
sura
nd
Benzene
Ethylbenzene
o-Xylene
Sum of m-Xylene and p-Xylene
Toluene
162
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0002
The following results were achieved:
CB01ACKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l <0.1 (LOD) --- -- ± -
1,1-Dichloroethene µg/l - --- -- ± -
1,2-Dichloroethane µg/l - --0.888 1.3212.7 ± -
Bromodichloromethane µg/l 2.4 106-0.141 0.2212.26 ± 0.61
cis-1,2-Dichloroethene µg/l - --0.063 0.09620.833 ± -
Dibromochloromethane µg/l 1.4 127-0.0847 0.1351.1 ± 2.2
Dichloromethane µg/l - --0.147 0.2131.89 ± -
Tetrachloroethene µg/l 6.9 175-0.494 0.7733.94 ± 3.83
Tetrachloromethane µg/l 1.5 126-0.127 0.2041.19 ± 1.52
trans-1,2-Dichloroethene µg/l - --0.112 0.1710.852 ± -
Tribromomethane µg/l <0.2 (LOD) --- -- ± -
Trichloroethene µg/l 0.8 123-0.0506 0.07910.652 ± 1.88
Trichloromethane µg/l 11.4 115-0.664 1.049.95 ± 1.4
CB01BCKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l 10.6 146-0.671 1.057.27 ± 3.17
1,1-Dichloroethene µg/l - --1.3 1.99.06 ± -
1,2-Dichloroethane µg/l - --0.227 0.3223.11 ± -
Bromodichloromethane µg/l 7.5 102-0.346 0.5157.33 ± 0.32
cis-1,2-Dichloroethene µg/l - --0.313 0.4554.68 ± -
Dibromochloromethane µg/l 9.9 121-0.698 1.128.15 ± 1.57
Dichloromethane µg/l - --0.611 0.848.04 ± -
Tetrachloroethene µg/l 15.4 136-1.38 2.2111.3 ± 1.85
Tetrachloromethane µg/l 14.9 92.7-2.37 3.7816.1 ± -0.31
trans-1,2-Dichloroethene µg/l - --0.446 0.6653.72 ± -
Tribromomethane µg/l 3.7 108-0.245 0.3743.43 ± 0.73
Trichloroethene µg/l 5.6 144-0.324 0.5073.9 ± 3.35
Trichloromethane µg/l 1.1 106-0.072 0.111.04 ± 0.58
This report was created with PROLab, a software by QuoData: http://www.quodata.de/
163
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0002
z-score-2 0 2 -2 0 2
Sample
CB01 A - VOC CB01 B - VOC
Mea
sura
nd
1,1,1-Trichloroethane
Bromodichloromethane
Dibromochloromethane
Tetrachloroethene
Tetrachloromethane
Tribromomethane
Trichloroethene
Trichloromethane
2.2
3.8
3.2
3.4
164
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0003
The following results were achieved:
CB01ACKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l <0.01 (LOQ) --- -- ± -
1,1-Dichloroethene µg/l <0.1 (LOQ) --- -- ± -
1,2-Dichloroethane µg/l 13.65 1081.640.888 1.3212.7 ± 0.74
Bromodichloromethane µg/l 2.245 99.10.2690.141 0.2212.26 ± -0.09
cis-1,2-Dichloroethene µg/l 0.868 1040.1040.063 0.09620.833 ± 0.36
Dibromochloromethane µg/l 1.09 98.90.1310.0847 0.1351.1 ± -0.09
Dichloromethane µg/l 2.033 1080.2440.147 0.2131.89 ± 0.67
Tetrachloroethene µg/l 3.999 1010.480.494 0.7733.94 ± 0.07
Tetrachloromethane µg/l 1.284 1080.1540.127 0.2041.19 ± 0.46
trans-1,2-Dichloroethene µg/l 0.932 1090.1120.112 0.1710.852 ± 0.47
Tribromomethane µg/l <0.1 (LOQ) --- -- ± -
Trichloroethene µg/l 0.654 1000.0790.0506 0.07910.652 ± 0.03
Trichloromethane µg/l 9.653 971.1580.664 1.049.95 ± -0.28
CB01BCKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l 7.089 97.50.8510.671 1.057.27 ± -0.17
1,1-Dichloroethene µg/l 8.864 97.81.0641.3 1.99.06 ± -0.1
1,2-Dichloroethane µg/l 3.208 1030.3850.227 0.3223.11 ± 0.31
Bromodichloromethane µg/l 7.111 970.8530.346 0.5157.33 ± -0.43
cis-1,2-Dichloroethene µg/l 4.871 1040.5850.313 0.4554.68 ± 0.41
Dibromochloromethane µg/l 7.821 960.9390.698 1.128.15 ± -0.29
Dichloromethane µg/l 8.697 1081.0440.611 0.848.04 ± 0.79
Tetrachloroethene µg/l 10.45 92.41.2541.38 2.2111.3 ± -0.39
Tetrachloromethane µg/l 17.22 1072.0662.37 3.7816.1 ± 0.3
trans-1,2-Dichloroethene µg/l 4.169 1120.50.446 0.6653.72 ± 0.68
Tribromomethane µg/l 3.236 94.40.3880.245 0.3743.43 ± -0.51
Trichloroethene µg/l 3.91 1000.4690.324 0.5073.9 ± 0.02
Trichloromethane µg/l 0.997 96.20.120.072 0.111.04 ± -0.35
This report was created with PROLab, a software by QuoData: http://www.quodata.de/
165
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0003
z-score-2 0 2 -2 0 2
Sample
CB01 A - VOC CB01 B - VOC
Mea
sura
nd
1,1,1-Trichloroethane
1,1-Dichloroethene
1,2-Dichloroethane
Bromodichloromethane
cis-1,2-Dichloroethene
Dibromochloromethane
Dichloromethane
Tetrachloroethene
Tetrachloromethane
trans-1,2-Dichloroethene
Tribromomethane
Trichloroethene
Trichloromethane
166
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0004
The following results were achieved:
CB01ABTXParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
Benzene µg/l 0.57 98.20.110.0519 0.06920.58 ± -0.15
Ethylbenzene µg/l 0.49 1110.10.0383 0.0510.442 ± 0.95
o-Xylene µg/l 4.82 1090.960.513 0.6844.41 ± 0.6
Sum of m-Xylene and p-Xylene µg/l 6.53 1151.310.719 0.9595.69 ± 0.87
Toluene µg/l 0.9 1060.180.0889 0.1260.85 ± 0.4
Methyl-tert-butyl-ether µg/l 3.64 1120.730.361 0.4343.25 ± 0.89
CB01ACKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l <0.06 (LOD) --- -- ± -
1,1-Dichloroethene µg/l <0.04 (LOD) --- -- ± -
1,2-Dichloroethane µg/l 11.85 93.52.370.888 1.3212.7 ± -0.62
Bromodichloromethane µg/l 2.35 1040.470.141 0.2212.26 ± 0.39
cis-1,2-Dichloroethene µg/l 0.86 1030.170.063 0.09620.833 ± 0.28
Dibromochloromethane µg/l 1.19 1080.240.0847 0.1351.1 ± 0.65
Dichloromethane µg/l 1.69 89.40.340.147 0.2131.89 ± -0.94
Tetrachloroethene µg/l 4.52 1150.90.494 0.7733.94 ± 0.75
Tetrachloromethane µg/l 1.51 1270.30.127 0.2041.19 ± 1.57
trans-1,2-Dichloroethene µg/l 0.81 95.10.160.112 0.1710.852 ± -0.25
Tribromomethane µg/l <0.07 (LOQ) --- -- ± -
Trichloroethene µg/l 0.73 1120.150.0506 0.07910.652 ± 0.99
Trichloromethane µg/l 10.01 10120.664 1.049.95 ± 0.06
CB01BBTXParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
Benzene µg/l 2.05 99.20.410.241 0.3412.07 ± -0.05
Ethylbenzene µg/l 1.64 1130.330.201 0.2761.45 ± 0.69
o-Xylene µg/l 3.28 1080.660.363 0.4843.03 ± 0.51
Sum of m-Xylene and p-Xylene µg/l 9.7 1131.941.04 1.388.55 ± 0.83
Toluene µg/l 5.66 1061.130.512 0.7045.33 ± 0.47
Methyl-tert-butyl-ether µg/l 0.97 1100.190.103 0.1240.882 ± 0.71
CB01BCKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l 7.72 1061.540.671 1.057.27 ± 0.43
This report was created with PROLab, a software by QuoData: http://www.quodata.de/
167
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0004
Parameter Unit Target ± Cl(99%) Result ± U Criteria Recovery z-score
1,1-Dichloroethene µg/l 9.43 1041.891.3 1.99.06 ± 0.19
1,2-Dichloroethane µg/l 3.19 1030.640.227 0.3223.11 ± 0.26
Bromodichloromethane µg/l 8.07 1101.610.346 0.5157.33 ± 1.43
cis-1,2-Dichloroethene µg/l 4.67 99.70.930.313 0.4554.68 ± -0.03
Dibromochloromethane µg/l 8.89 1091.780.698 1.128.15 ± 0.66
Dichloromethane µg/l 8.04 1001.610.611 0.848.04 ± 0.00
Tetrachloroethene µg/l 12.34 1092.471.38 2.2111.3 ± 0.46
Tetrachloromethane µg/l 19.43 1213.892.37 3.7816.1 ± 0.89
trans-1,2-Dichloroethene µg/l 3.5 94.20.70.446 0.6653.72 ± -0.33
Tribromomethane µg/l 3.99 1160.80.245 0.3743.43 ± 1.51
Trichloroethene µg/l 4.29 1100.860.324 0.5073.9 ± 0.77
Trichloromethane µg/l 1.09 1050.220.072 0.111.04 ± 0.49
168
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0004
z-score-2 0 2 -2 0 2 -2 0 2 -2 0 2
Sample
CB01 A - BTEX / MTBE VOC VOC
Mea
sura
nd
Benzene
Ethylbenzene
o-Xylene
Sum of m-Xylene and p-Xylene
Toluene
Methyl-tert-butyl-ether
1,1,1-Trichloroethane
1,1-Dichloroethene
1,2-Dichloroethane
Bromodichloromethane
cis-1,2-Dichloroethene
Dibromochloromethane
Dichloromethane
Tetrachloroethene
Tetrachloromethane
trans-1,2-Dichloroethene
Tribromomethane
Trichloroethene
Trichloromethane
CB01 B - BTEX / MTBE
169
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0005
The following results were achieved:
CB01ABTXParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
Benzene µg/l 0.6 1030.20.0519 0.06920.58 ± 0.29
Ethylbenzene µg/l 0.45 1020.10.0383 0.0510.442 ± 0.17
o-Xylene µg/l 5.4 1221.10.513 0.6844.41 ± 1.45
Sum of m-Xylene and p-Xylene µg/l 6.5 1141.30.719 0.9595.69 ± 0.84
Toluene µg/l 0.9 1060.20.0889 0.1260.85 ± 0.4
Methyl-tert-butyl-ether µg/l 4.1 1260.90.361 0.4343.25 ± 1.95
CB01ACKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l <0.1 (LOQ) --- -- ± -
1,1-Dichloroethene µg/l <0.1 (LOQ) --- -- ± -
1,2-Dichloroethane µg/l 15.2 1203.10.888 1.3212.7 ± 1.91
Bromodichloromethane µg/l 3 1320.60.141 0.2212.26 ± 3.33
cis-1,2-Dichloroethene µg/l 0.85 1020.170.063 0.09620.833 ± 0.17
Dibromochloromethane µg/l 1.4 1270.30.0847 0.1351.1 ± 2.2
Dichloromethane µg/l 2.4 1270.50.147 0.2131.89 ± 2.39
Tetrachloroethene µg/l 5.4 1371.10.494 0.7733.94 ± 1.89
Tetrachloromethane µg/l 1.5 1260.30.127 0.2041.19 ± 1.52
trans-1,2-Dichloroethene µg/l 1.2 1410.30.112 0.1710.852 ± 2.04
Tribromomethane µg/l <0.1 (LOQ) --- -- ± -
Trichloroethene µg/l 0.6 92.10.120.0506 0.07910.652 ± -0.65
Trichloromethane µg/l 11.5 1162.30.664 1.049.95 ± 1.49
CB01BBTXParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
Benzene µg/l 2.3 1110.50.241 0.3412.07 ± 0.68
Ethylbenzene µg/l 1.8 1240.40.201 0.2761.45 ± 1.27
o-Xylene µg/l 3.4 1120.70.363 0.4843.03 ± 0.76
Sum of m-Xylene and p-Xylene µg/l 10.7 1252.21.04 1.388.55 ± 1.56
Toluene µg/l 6.4 1201.30.512 0.7045.33 ± 1.52
Methyl-tert-butyl-ether µg/l 1.1 1250.30.103 0.1240.882 ± 1.76
CB01BCKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l 9.6 13220.671 1.057.27 ± 2.22
This report was created with PROLab, a software by QuoData: http://www.quodata.de/
170
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0005
Parameter Unit Target ± Cl(99%) Result ± U Criteria Recovery z-score
1,1-Dichloroethene µg/l 12.7 1402.61.3 1.99.06 ± 1.92
1,2-Dichloroethane µg/l 3.7 1190.80.227 0.3223.11 ± 1.84
Bromodichloromethane µg/l 9.1 12420.346 0.5157.33 ± 3.43
cis-1,2-Dichloroethene µg/l 5.2 1111.10.313 0.4554.68 ± 1.14
Dibromochloromethane µg/l 9.6 11820.698 1.128.15 ± 1.3
Dichloromethane µg/l 11.9 1482.40.611 0.848.04 ± 4.6
Tetrachloroethene µg/l 15.1 1333.11.38 2.2111.3 ± 1.71
Tetrachloromethane µg/l 24.2 1514.92.37 3.7816.1 ± 2.15
trans-1,2-Dichloroethene µg/l 4.8 12910.446 0.6653.72 ± 1.63
Tribromomethane µg/l 4 1170.80.245 0.3743.43 ± 1.53
Trichloroethene µg/l 3.3 84.60.70.324 0.5073.9 ± -1.18
Trichloromethane µg/l 1.4 1350.30.072 0.111.04 ± 3.31
171
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0005
z-score-2 0 2 -2 0 2 -2 0 2 -2 0 2
Sample
CB01 A - BTEX / MTBE
Mea
sura
nd
Benzene
Ethylbenzene
o-Xylene
Sum of m-Xylene and p-Xylene
Toluene
Methyl-tert-butyl-ether
1,1,1-Trichloroethane
1,1-Dichloroethene
1,2-Dichloroethane
Bromodichloromethane
cis-1,2-Dichloroethene
Dibromochloromethane
Dichloromethane
Tetrachloroethene
Tetrachloromethane
trans-1,2-Dichloroethene
Tribromomethane
Trichloroethene
Trichloromethane
3.3
2.2
2.4
2.0
2.2
3.4
4.6
2.2
3.3
VOC VOCCB01 B - BTEX / MTBE
172
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0006
The following results were achieved:
CB01ABTXParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
Benzene µg/l - --0.0519 0.06920.58 ± -
Ethylbenzene µg/l 0.32 72.50.060.0383 0.0510.442 ± -2.38
o-Xylene µg/l 2.14 48.50.20.513 0.6844.41 ± -3.32
Sum of m-Xylene and p-Xylene µg/l 3.48 61.10.40.719 0.9595.69 ± -2.31
Toluene µg/l 0.69 81.20.120.0889 0.1260.85 ± -1.27
Methyl-tert-butyl-ether µg/l - --0.361 0.4343.25 ± -
CB01ACKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l <0.1 (LOQ) --- -- ± -
1,1-Dichloroethene µg/l - --- -- ± -
1,2-Dichloroethane µg/l - --0.888 1.3212.7 ± -
Bromodichloromethane µg/l 2.12 93.60.080.141 0.2212.26 ± -0.66
cis-1,2-Dichloroethene µg/l - --0.063 0.09620.833 ± -
Dibromochloromethane µg/l 1.08 980.050.0847 0.1351.1 ± -0.16
Dichloromethane µg/l - --0.147 0.2131.89 ± -
Tetrachloroethene µg/l 2.35 59.60.260.494 0.7733.94 ± -2.06
Tetrachloromethane µg/l 0.75 630.040.127 0.2041.19 ± -2.16
trans-1,2-Dichloroethene µg/l - --0.112 0.1710.852 ± -
Tribromomethane µg/l - --- -- ± -
Trichloroethene µg/l 0.5 76.70.020.0506 0.07910.652 ± -1.92
Trichloromethane µg/l 9.27 93.20.50.664 1.049.95 ± -0.65
CB01BBTXParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
Benzene µg/l - --0.241 0.3412.07 ± -
Ethylbenzene µg/l 0.55 37.90.060.201 0.2761.45 ± -3.26
o-Xylene µg/l 0.94 310.20.363 0.4843.03 ± -4.33
Sum of m-Xylene and p-Xylene µg/l 3.44 40.20.41.04 1.388.55 ± -3.7
Toluene µg/l 2.71 50.90.120.512 0.7045.33 ± -3.72
Methyl-tert-butyl-ether µg/l - --0.103 0.1240.882 ± -
CB01BCKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l 5.81 79.90.70.671 1.057.27 ± -1.39
This report was created with PROLab, a software by QuoData: http://www.quodata.de/
173
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0006
Parameter Unit Target ± Cl(99%) Result ± U Criteria Recovery z-score
1,1-Dichloroethene µg/l - --1.3 1.99.06 ± -
1,2-Dichloroethane µg/l - --0.227 0.3223.11 ± -
Bromodichloromethane µg/l 6.75 920.620.346 0.5157.33 ± -1.13
cis-1,2-Dichloroethene µg/l - --0.313 0.4554.68 ± -
Dibromochloromethane µg/l 7.46 91.50.540.698 1.128.15 ± -0.62
Dichloromethane µg/l - --0.611 0.848.04 ± -
Tetrachloroethene µg/l 10.74 94.93.421.38 2.2111.3 ± -0.26
Tetrachloromethane µg/l 13.82 861.342.37 3.7816.1 ± -0.59
trans-1,2-Dichloroethene µg/l - --0.446 0.6653.72 ± -
Tribromomethane µg/l - --0.245 0.3743.43 ± -
Trichloroethene µg/l 3.65 93.60.560.324 0.5073.9 ± -0.49
Trichloromethane µg/l 0.97 93.60.10.072 0.111.04 ± -0.6
174
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0006
z-score-2 0 2 -2 0 2 -2 0 2 -2 0 2
Sample
CB01 A - BTEX / MTBE
Mea
sura
nd
Ethylbenzene
o-Xylene
Sum of m-Xylene and p-Xylene
Toluene
1,1,1-Trichloroethane
Bromodichloromethane
Dibromochloromethane
Tetrachloroethene
Tetrachloromethane
Trichloroethene
Trichloromethane
-2.4
-3.3
-2.3
-2.1
-2.2
-3.3
-4.3
-3.7
-3.7
VOC VOCCB01 B - BTEX / MTBE
175
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0007
The following results were achieved:
CB01ABTXParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
Benzene µg/l <0.5 (LOQ) --0.0519 0.06920.58 ± -
Ethylbenzene µg/l <0.5 (LOQ) --0.0383 0.0510.442 ± -
o-Xylene µg/l 4.374 99.20.890.513 0.6844.41 ± -0.05
Sum of m-Xylene and p-Xylene µg/l 5.714 1001.370.719 0.9595.69 ± 0.02
Toluene µg/l 0.982 1160.20.0889 0.1260.85 ± 1.05
Methyl-tert-butyl-ether µg/l 3.37 1040.710.361 0.4343.25 ± 0.27
CB01ACKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l <0.1 (LOQ) --- -- ± -
1,1-Dichloroethene µg/l <0.2 (LOQ) --- -- ± -
1,2-Dichloroethane µg/l 11.568 91.31.690.888 1.3212.7 ± -0.83
Bromodichloromethane µg/l 2.393 1060.340.141 0.2212.26 ± 0.58
cis-1,2-Dichloroethene µg/l 0.857 1030.220.063 0.09620.833 ± 0.24
Dibromochloromethane µg/l 1.182 1070.20.0847 0.1351.1 ± 0.59
Dichloromethane µg/l 1.684 89.10.30.147 0.2131.89 ± -0.97
Tetrachloroethene µg/l 3.895 98.80.720.494 0.7733.94 ± -0.06
Tetrachloromethane µg/l 1.095 920.190.127 0.2041.19 ± -0.47
trans-1,2-Dichloroethene µg/l 0.855 1000.250.112 0.1710.852 ± 0.02
Tribromomethane µg/l <0.1 (LOQ) --- -- ± -
Trichloroethene µg/l 0.636 97.60.120.0506 0.07910.652 ± -0.2
Trichloromethane µg/l 10.2 1031.680.664 1.049.95 ± 0.24
CB01BBTXParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
Benzene µg/l 2.06 99.70.370.241 0.3412.07 ± -0.02
Ethylbenzene µg/l 1.374 94.70.260.201 0.2761.45 ± -0.28
o-Xylene µg/l 3.247 1070.660.363 0.4843.03 ± 0.44
Sum of m-Xylene and p-Xylene µg/l 9.016 1052.161.04 1.388.55 ± 0.34
Toluene µg/l 5.599 1051.150.512 0.7045.33 ± 0.38
Methyl-tert-butyl-ether µg/l 0.951 1080.20.103 0.1240.882 ± 0.55
CB01BCKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l 6.84 94.11.2310.671 1.057.27 ± -0.41
This report was created with PROLab, a software by QuoData: http://www.quodata.de/
176
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0007
Parameter Unit Target ± Cl(99%) Result ± U Criteria Recovery z-score
1,1-Dichloroethene µg/l 11.5 1273.331.3 1.99.06 ± 1.29
1,2-Dichloroethane µg/l 3.029 97.50.440.227 0.3223.11 ± -0.24
Bromodichloromethane µg/l 7.65 1041.10.346 0.5157.33 ± 0.61
cis-1,2-Dichloroethene µg/l 4.352 92.91.10.313 0.4554.68 ± -0.73
Dibromochloromethane µg/l 8.87 1091.480.698 1.128.15 ± 0.65
Dichloromethane µg/l 7.148 88.91.450.611 0.848.04 ± -1.06
Tetrachloroethene µg/l 11.148 98.52.061.38 2.2111.3 ± -0.08
Tetrachloromethane µg/l 16.45 1022.832.37 3.7816.1 ± 0.1
trans-1,2-Dichloroethene µg/l 3.418 921.010.446 0.6653.72 ± -0.45
Tribromomethane µg/l 3.53 1030.650.245 0.3743.43 ± 0.28
Trichloroethene µg/l 3.478 89.20.660.324 0.5073.9 ± -0.83
Trichloromethane µg/l 1.09 1050.180.072 0.111.04 ± 0.49
177
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0007
z-score-2 0 2 -2 0 2 -2 0 2 -2 0 2
Sample
CB01 A - BTEX / MTBE
Mea
sura
nd
Benzene
Ethylbenzene
o-Xylene
Sum of m-Xylene and p-Xylene
Toluene
Methyl-tert-butyl-ether
1,1,1-Trichloroethane
1,1-Dichloroethene
1,2-Dichloroethane
Bromodichloromethane
cis-1,2-Dichloroethene
Dibromochloromethane
Dichloromethane
Tetrachloroethene
Tetrachloromethane
trans-1,2-Dichloroethene
Tribromomethane
Trichloroethene
Trichloromethane
VOC VOCCB01 B - BTEX / MTBE
178
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0008
The following results were achieved:
CB01ACKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l <0.004 (LOD) --- -- ± -
1,1-Dichloroethene µg/l <0.05 (LOQ) --- -- ± -
1,2-Dichloroethane µg/l 12 94.70.710.888 1.3212.7 ± -0.51
Bromodichloromethane µg/l 2.24 98.90.0470.141 0.2212.26 ± -0.11
cis-1,2-Dichloroethene µg/l 0.887 1060.0590.063 0.09620.833 ± 0.56
Dibromochloromethane µg/l 1.19 1080.0410.0847 0.1351.1 ± 0.65
Dichloromethane µg/l 1.67 88.40.080.147 0.2131.89 ± -1.03
Tetrachloroethene µg/l 3.8 96.40.0590.494 0.7733.94 ± -0.18
Tetrachloromethane µg/l 1.14 95.80.0250.127 0.2041.19 ± -0.25
trans-1,2-Dichloroethene µg/l 0.514 60.30.0710.112 0.1710.852 ± -1.98
Tribromomethane µg/l <0.04 (LOQ) --- -- ± -
Trichloroethene µg/l 0.631 96.80.1350.0506 0.07910.652 ± -0.26
Trichloromethane µg/l 8.32 83.60.1980.664 1.049.95 ± -1.57
CB01BCKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l 7.06 97.10.4470.671 1.057.27 ± -0.2
1,1-Dichloroethene µg/l 9.19 1010.5651.3 1.99.06 ± 0.07
1,2-Dichloroethane µg/l 2.51 80.80.0670.227 0.3223.11 ± -1.86
Bromodichloromethane µg/l 7.14 97.30.8150.346 0.5157.33 ± -0.38
cis-1,2-Dichloroethene µg/l 4.55 97.20.0540.313 0.4554.68 ± -0.29
Dibromochloromethane µg/l 7.52 92.30.7780.698 1.128.15 ± -0.56
Dichloromethane µg/l 7.13 88.70.3070.611 0.848.04 ± -1.08
Tetrachloroethene µg/l 11.8 1040.3271.38 2.2111.3 ± 0.22
Tetrachloromethane µg/l 18.3 1140.42.37 3.7816.1 ± 0.59
trans-1,2-Dichloroethene µg/l 3.13 84.20.0680.446 0.6653.72 ± -0.88
Tribromomethane µg/l 3.95 1150.0560.245 0.3743.43 ± 1.4
Trichloroethene µg/l 3.68 94.30.1330.324 0.5073.9 ± -0.43
Trichloromethane µg/l 0.847 81.80.0660.072 0.111.04 ± -1.72
This report was created with PROLab, a software by QuoData: http://www.quodata.de/
179
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0008
z-score-2 0 2 -2 0 2
Sample
CB01 A - VOC CB01 B - VOC
Mea
sura
nd
1,1,1-Trichloroethane
1,1-Dichloroethene
1,2-Dichloroethane
Bromodichloromethane
cis-1,2-Dichloroethene
Dibromochloromethane
Dichloromethane
Tetrachloroethene
Tetrachloromethane
trans-1,2-Dichloroethene
Tribromomethane
Trichloroethene
Trichloromethane
180
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0009
The following results were achieved:
CB01ABTXParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
Benzene µg/l 0.5 86.20.1250.0519 0.06920.58 ± -1.16
Ethylbenzene µg/l 0.22 49.80.0550.0383 0.0510.442 ± -4.34
o-Xylene µg/l 2.88 65.30.720.513 0.6844.41 ± -2.24
Sum of m-Xylene and p-Xylene µg/l 2.95 51.80.7380.719 0.9595.69 ± -2.86
Toluene µg/l 0.61 71.80.1530.0889 0.1260.85 ± -1.91
Methyl-tert-butyl-ether µg/l 2.5 76.80.6250.361 0.4343.25 ± -1.74
CB01ACKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l - --- -- ± -
1,1-Dichloroethene µg/l - --- -- ± -
1,2-Dichloroethane µg/l 13.02 1033.2550.888 1.3212.7 ± 0.26
Bromodichloromethane µg/l 2.29 1010.5730.141 0.2212.26 ± 0.11
cis-1,2-Dichloroethene µg/l 0.86 1030.2150.063 0.09620.833 ± 0.28
Dibromochloromethane µg/l 1.13 1030.2830.0847 0.1351.1 ± 0.2
Dichloromethane µg/l 1.83 96.80.4580.147 0.2131.89 ± -0.28
Tetrachloroethene µg/l 3.49 88.50.8730.494 0.7733.94 ± -0.58
Tetrachloromethane µg/l 1.06 89.10.2650.127 0.2041.19 ± -0.64
trans-1,2-Dichloroethene µg/l 0.73 85.70.1830.112 0.1710.852 ± -0.71
Tribromomethane µg/l - --- -- ± -
Trichloroethene µg/l 0.59 90.60.1480.0506 0.07910.652 ± -0.78
Trichloromethane µg/l 9.86 99.12.4650.664 1.049.95 ± -0.09
CB01BBTXParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
Benzene µg/l 1.78 86.10.4450.241 0.3412.07 ± -0.84
Ethylbenzene µg/l 0.87 600.2180.201 0.2761.45 ± -2.1
o-Xylene µg/l 2.2 72.50.550.363 0.4843.03 ± -1.72
Sum of m-Xylene and p-Xylene µg/l 5.33 62.41.3331.04 1.388.55 ± -2.33
Toluene µg/l 4.21 791.0530.512 0.7045.33 ± -1.59
Methyl-tert-butyl-ether µg/l 0.72 81.60.180.103 0.1240.882 ± -1.31
CB01BCKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l 5.89 811.4730.671 1.057.27 ± -1.32
This report was created with PROLab, a software by QuoData: http://www.quodata.de/
181
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0009
Parameter Unit Target ± Cl(99%) Result ± U Criteria Recovery z-score
1,1-Dichloroethene µg/l 7.7 851.9251.3 1.99.06 ± -0.72
1,2-Dichloroethane µg/l 2.75 88.50.6880.227 0.3223.11 ± -1.11
Bromodichloromethane µg/l 7.14 97.31.7850.346 0.5157.33 ± -0.38
cis-1,2-Dichloroethene µg/l 4.3 91.81.0750.313 0.4554.68 ± -0.84
Dibromochloromethane µg/l 7.91 971.9780.698 1.128.15 ± -0.22
Dichloromethane µg/l 7.91 98.41.9780.611 0.848.04 ± -0.15
Tetrachloroethene µg/l 9.07 80.22.2681.38 2.2111.3 ± -1.02
Tetrachloromethane µg/l 12.96 80.73.242.37 3.7816.1 ± -0.82
trans-1,2-Dichloroethene µg/l 2.99 80.50.7480.446 0.6653.72 ± -1.09
Tribromomethane µg/l 3.35 97.80.8380.245 0.3743.43 ± -0.21
Trichloroethene µg/l 3.08 790.770.324 0.5073.9 ± -1.62
Trichloromethane µg/l 1.1 1060.2750.072 0.111.04 ± 0.58
182
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0009
z-score-2 0 2 -2 0 2 -2 0 2 -2 0 2
Sample
CB01 A - BTEX / MTBE
Mea
sura
nd
Benzene
Ethylbenzene
o-Xylene
Sum of m-Xylene and p-Xylene
Toluene
Methyl-tert-butyl-ether
1,1,1-Trichloroethane
1,1-Dichloroethene
1,2-Dichloroethane
Bromodichloromethane
cis-1,2-Dichloroethene
Dibromochloromethane
Dichloromethane
Tetrachloroethene
Tetrachloromethane
trans-1,2-Dichloroethene
Tribromomethane
Trichloroethene
Trichloromethane
-4.3
-2.2
-2.9
-2.1
-2.3
VOC VOCCB01 B - BTEX / MTBE
183
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0010
The following results were achieved:
CB01ABTXParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
Benzene µg/l 0.62 1070.1170.0519 0.06920.58 ± 0.57
Ethylbenzene µg/l 0.49 1110.090.0383 0.0510.442 ± 0.95
o-Xylene µg/l 4.44 1010.8130.513 0.6844.41 ± 0.04
Sum of m-Xylene and p-Xylene µg/l 5.67 99.61.050.719 0.9595.69 ± -0.02
Toluene µg/l 0.94 1110.1790.0889 0.1260.85 ± 0.72
Methyl-tert-butyl-ether µg/l 3.69 1130.7080.361 0.4343.25 ± 1
CB01ACKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l <0.1 (LOQ) --- -- ± -
1,1-Dichloroethene µg/l <0.1 (LOQ) --- -- ± -
1,2-Dichloroethane µg/l 12.31 97.11.970.888 1.3212.7 ± -0.27
Bromodichloromethane µg/l 1.99 87.90.3080.141 0.2212.26 ± -1.25
cis-1,2-Dichloroethene µg/l 0.75 900.1430.063 0.09620.833 ± -0.87
Dibromochloromethane µg/l 0.87 78.90.1480.0847 0.1351.1 ± -1.71
Dichloromethane µg/l 1.99 1050.3110.147 0.2131.89 ± 0.47
Tetrachloroethene µg/l 4.32 1100.6520.494 0.7733.94 ± 0.49
Tetrachloromethane µg/l 1.31 1100.2370.127 0.2041.19 ± 0.59
trans-1,2-Dichloroethene µg/l 0.72 84.50.1310.112 0.1710.852 ± -0.77
Tribromomethane µg/l <0.1 (LOQ) --- -- ± -
Trichloroethene µg/l 0.7 1070.1180.0506 0.07910.652 ± 0.61
Trichloromethane µg/l 9.83 98.81.710.664 1.049.95 ± -0.12
CB01BBTXParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
Benzene µg/l 2.14 1040.4120.241 0.3412.07 ± 0.21
Ethylbenzene µg/l 1.62 1120.2980.201 0.2761.45 ± 0.61
o-Xylene µg/l 3.16 1040.5880.363 0.4843.03 ± 0.26
Sum of m-Xylene and p-Xylene µg/l 9.15 1071.6291.04 1.388.55 ± 0.44
Toluene µg/l 6.24 1171.110.512 0.7045.33 ± 1.29
Methyl-tert-butyl-ether µg/l 0.95 1080.1770.103 0.1240.882 ± 0.55
CB01BCKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l 7.55 1041.30.671 1.057.27 ± 0.27
This report was created with PROLab, a software by QuoData: http://www.quodata.de/
184
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0010
Parameter Unit Target ± Cl(99%) Result ± U Criteria Recovery z-score
1,1-Dichloroethene µg/l 10.18 1121.681.3 1.99.06 ± 0.59
1,2-Dichloroethane µg/l 2.98 95.90.5420.227 0.3223.11 ± -0.4
Bromodichloromethane µg/l 6.87 93.71.320.346 0.5157.33 ± -0.9
cis-1,2-Dichloroethene µg/l 4.77 1020.7540.313 0.4554.68 ± 0.19
Dibromochloromethane µg/l 7.21 88.51.150.698 1.128.15 ± -0.84
Dichloromethane µg/l 8.72 1081.690.611 0.848.04 ± 0.81
Tetrachloroethene µg/l 11.38 1011.811.38 2.2111.3 ± 0.03
Tetrachloromethane µg/l 7.28 45.31.222.37 3.7816.1 ± -2.32
trans-1,2-Dichloroethene µg/l 3.12 840.5830.446 0.6653.72 ± -0.9
Tribromomethane µg/l 3.2 93.40.5220.245 0.3743.43 ± -0.61
Trichloroethene µg/l 4.01 1030.6740.324 0.5073.9 ± 0.21
Trichloromethane µg/l 1.95 1880.3110.072 0.111.04 ± 8.32
185
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0010
z-score-2 0 2 -2 0 2 -2 0 2 -2 0 2
Sample
CB01 A - BTEX / MTBE
Mea
sura
nd
Benzene
Ethylbenzene
o-Xylene
Sum of m-Xylene and p-Xylene
Toluene
Methyl-tert-butyl-ether
1,1,1-Trichloroethane
1,1-Dichloroethene
1,2-Dichloroethane
Bromodichloromethane
cis-1,2-Dichloroethene
Dibromochloromethane
Dichloromethane
Tetrachloroethene
Tetrachloromethane
trans-1,2-Dichloroethene
Tribromomethane
Trichloroethene
Trichloromethane 8.3
-2.3
VOC VOCCB01 B - BTEX / MTBE
186
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0011
The following results were achieved:
CB01ABTXParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
Benzene µg/l 0.73 1260.150.0519 0.06920.58 ± 2.17
Ethylbenzene µg/l 0.46 1040.090.0383 0.0510.442 ± 0.36
o-Xylene µg/l 4.58 1040.920.513 0.6844.41 ± 0.25
Sum of m-Xylene and p-Xylene µg/l 6.05 1061.210.719 0.9595.69 ± 0.37
Toluene µg/l 1 1180.20.0889 0.1260.85 ± 1.19
Methyl-tert-butyl-ether µg/l 2.74 84.20.550.361 0.4343.25 ± -1.19
CB01ACKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l <0.05 (LOQ) --- -- ± -
1,1-Dichloroethene µg/l <0.05 (LOQ) --- -- ± -
1,2-Dichloroethane µg/l 14.47 1142.890.888 1.3212.7 ± 1.36
Bromodichloromethane µg/l 2.55 1130.510.141 0.2212.26 ± 1.29
cis-1,2-Dichloroethene µg/l 1.01 1210.20.063 0.09620.833 ± 1.84
Dibromochloromethane µg/l 1.14 1030.230.0847 0.1351.1 ± 0.28
Dichloromethane µg/l 0.97 51.30.190.147 0.2131.89 ± -4.31
Tetrachloroethene µg/l 4.73 1200.950.494 0.7733.94 ± 1.02
Tetrachloromethane µg/l 1.2 1010.240.127 0.2041.19 ± 0.05
trans-1,2-Dichloroethene µg/l 1.18 1380.240.112 0.1710.852 ± 1.92
Tribromomethane µg/l <0.05 (LOQ) --- -- ± -
Trichloroethene µg/l 0.69 1060.140.0506 0.07910.652 ± 0.49
Trichloromethane µg/l 11.07 1112.210.664 1.049.95 ± 1.08
CB01BBTXParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
Benzene µg/l 2.49 1200.50.241 0.3412.07 ± 1.24
Ethylbenzene µg/l 1.41 97.20.280.201 0.2761.45 ± -0.15
o-Xylene µg/l 3.02 99.60.60.363 0.4843.03 ± -0.03
Sum of m-Xylene and p-Xylene µg/l 8.74 1021.751.04 1.388.55 ± 0.14
Toluene µg/l 5.38 1011.080.512 0.7045.33 ± 0.07
Methyl-tert-butyl-ether µg/l 0.81 91.80.160.103 0.1240.882 ± -0.58
CB01BCKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l 8.37 1151.670.671 1.057.27 ± 1.05
This report was created with PROLab, a software by QuoData: http://www.quodata.de/
187
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0011
Parameter Unit Target ± Cl(99%) Result ± U Criteria Recovery z-score
1,1-Dichloroethene µg/l 15.37 1703.071.3 1.99.06 ± 3.33
1,2-Dichloroethane µg/l 3.67 1180.730.227 0.3223.11 ± 1.75
Bromodichloromethane µg/l 8.68 1181.740.346 0.5157.33 ± 2.61
cis-1,2-Dichloroethene µg/l 6.07 1301.210.313 0.4554.68 ± 3.05
Dibromochloromethane µg/l 8.49 1041.70.698 1.128.15 ± 0.3
Dichloromethane µg/l 4.58 570.920.611 0.848.04 ± -4.12
Tetrachloroethene µg/l 13.84 1222.771.38 2.2111.3 ± 1.14
Tetrachloromethane µg/l 14.75 91.82.952.37 3.7816.1 ± -0.35
trans-1,2-Dichloroethene µg/l 5.64 1521.130.446 0.6653.72 ± 2.89
Tribromomethane µg/l 3.51 1020.70.245 0.3743.43 ± 0.22
Trichloroethene µg/l 4.21 1080.840.324 0.5073.9 ± 0.61
Trichloromethane µg/l 1.26 1220.250.072 0.111.04 ± 2.04
188
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0011
z-score-2 0 2 -2 0 2 -2 0 2 -2 0 2
Sample
CB01 A - BTEX / MTBE
Mea
sura
nd
Benzene
Ethylbenzene
o-Xylene
Sum of m-Xylene and p-Xylene
Toluene
Methyl-tert-butyl-ether
1,1,1-Trichloroethane
1,1-Dichloroethene
1,2-Dichloroethane
Bromodichloromethane
cis-1,2-Dichloroethene
Dibromochloromethane
Dichloromethane
Tetrachloroethene
Tetrachloromethane
trans-1,2-Dichloroethene
Tribromomethane
Trichloroethene
Trichloromethane
2.2
3.3
2.6
3.1
2.9
2.0
-4.3 -4.1
VOC VOCCB01 B - BTEX / MTBE
189
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0012
The following results were achieved:
CB01ABTXParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
Benzene µg/l 0.55 94.80.10.0519 0.06920.58 ± -0.44
Ethylbenzene µg/l 0.45 1020.10.0383 0.0510.442 ± 0.17
o-Xylene µg/l 7.87 1780.80.513 0.6844.41 ± 5.06
Sum of m-Xylene and p-Xylene µg/l 7.31 1280.70.719 0.9595.69 ± 1.69
Toluene µg/l 0.77 90.60.10.0889 0.1260.85 ± -0.63
Methyl-tert-butyl-ether µg/l - --0.361 0.4343.25 ± -
CB01ACKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l <0.2 (LOQ) --- -- ± -
1,1-Dichloroethene µg/l <0.2 (LOQ) --- -- ± -
1,2-Dichloroethane µg/l 10.47 82.61.050.888 1.3212.7 ± -1.66
Bromodichloromethane µg/l 2.24 98.90.20.141 0.2212.26 ± -0.11
cis-1,2-Dichloroethene µg/l 0.68 81.60.10.063 0.09620.833 ± -1.6
Dibromochloromethane µg/l 0.98 88.90.10.0847 0.1351.1 ± -0.9
Dichloromethane µg/l <1 (LOQ) --0.147 0.2131.89 ± -
Tetrachloroethene µg/l 3.34 84.70.30.494 0.7733.94 ± -0.78
Tetrachloromethane µg/l 0.95 79.80.10.127 0.2041.19 ± -1.18
trans-1,2-Dichloroethene µg/l 0.72 84.50.10.112 0.1710.852 ± -0.77
Tribromomethane µg/l <0.2 (LOQ) --- -- ± -
Trichloroethene µg/l 0.54 82.90.10.0506 0.07910.652 ± -1.41
Trichloromethane µg/l 8.6 86.40.80.664 1.049.95 ± -1.3
CB01BBTXParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
Benzene µg/l 2.5 1210.30.241 0.3412.07 ± 1.27
Ethylbenzene µg/l 1.31 90.30.10.201 0.2761.45 ± -0.51
o-Xylene µg/l 6.95 2290.70.363 0.4843.03 ± 8.1
Sum of m-Xylene and p-Xylene µg/l 9.7 1130.91.04 1.388.55 ± 0.83
Toluene µg/l 5.1 95.70.50.512 0.7045.33 ± -0.33
Methyl-tert-butyl-ether µg/l - --0.103 0.1240.882 ± -
CB01BCKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l 5.62 77.30.60.671 1.057.27 ± -1.57
This report was created with PROLab, a software by QuoData: http://www.quodata.de/
190
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0012
Parameter Unit Target ± Cl(99%) Result ± U Criteria Recovery z-score
1,1-Dichloroethene µg/l 6.54 72.20.61.3 1.99.06 ± -1.33
1,2-Dichloroethane µg/l 2.1 67.60.20.227 0.3223.11 ± -3.13
Bromodichloromethane µg/l 7.26 990.70.346 0.5157.33 ± -0.14
cis-1,2-Dichloroethene µg/l 4.56 97.40.50.313 0.4554.68 ± -0.27
Dibromochloromethane µg/l 8 98.20.80.698 1.128.15 ± -0.14
Dichloromethane µg/l 5.92 73.70.60.611 0.848.04 ± -2.52
Tetrachloroethene µg/l 8.18 72.30.81.38 2.2111.3 ± -1.42
Tetrachloromethane µg/l 11.24 701.12.37 3.7816.1 ± -1.28
trans-1,2-Dichloroethene µg/l 3.25 87.50.30.446 0.6653.72 ± -0.7
Tribromomethane µg/l 3.48 1020.40.245 0.3743.43 ± 0.14
Trichloroethene µg/l 3.56 91.30.40.324 0.5073.9 ± -0.67
Trichloromethane µg/l 0.86 830.10.072 0.111.04 ± -1.6
191
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0012
z-score-2 0 2 -2 0 2 -2 0 2 -2 0 2
Sample
CB01 A - BTEX / MTBE
Mea
sura
nd
Benzene
Ethylbenzene
o-Xylene
Sum of m-Xylene and p-Xylene
Toluene
1,1,1-Trichloroethane
1,1-Dichloroethene
1,2-Dichloroethane
Bromodichloromethane
cis-1,2-Dichloroethene
Dibromochloromethane
Dichloromethane
Tetrachloroethene
Tetrachloromethane
trans-1,2-Dichloroethene
Tribromomethane
Trichloroethene
Trichloromethane
5.1 8.1
-3.1
-2.5
VOC VOCCB01 B - BTEX / MTBE
192
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0013
The following results were achieved:
CB01ACKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l <0.1 (LOQ) --- -- ± -
1,1-Dichloroethene µg/l <0.1 (LOQ) --- -- ± -
1,2-Dichloroethane µg/l 11.56 91.21.730.888 1.3212.7 ± -0.84
Bromodichloromethane µg/l 2.66 1170.40.141 0.2212.26 ± 1.79
cis-1,2-Dichloroethene µg/l 0.933 1120.140.063 0.09620.833 ± 1.04
Dibromochloromethane µg/l 1.2 1090.180.0847 0.1351.1 ± 0.72
Dichloromethane µg/l 1.88 99.50.280.147 0.2131.89 ± -0.05
Tetrachloroethene µg/l 5.72 1450.860.494 0.7733.94 ± 2.3
Tetrachloromethane µg/l 1.155 970.170.127 0.2041.19 ± -0.17
trans-1,2-Dichloroethene µg/l 1.04 1220.160.112 0.1710.852 ± 1.1
Tribromomethane µg/l <0.1 (LOQ) --- -- ± -
Trichloroethene µg/l 0.82 1260.120.0506 0.07910.652 ± 2.13
Trichloromethane µg/l 10.11 1021.50.664 1.049.95 ± 0.15
CB01BCKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l 8.42 1161.250.671 1.057.27 ± 1.09
1,1-Dichloroethene µg/l 4.75 52.40.711.3 1.99.06 ± -2.28
1,2-Dichloroethane µg/l 3.42 1100.510.227 0.3223.11 ± 0.97
Bromodichloromethane µg/l 10.24 1401.50.346 0.5157.33 ± 5.64
cis-1,2-Dichloroethene µg/l 6.27 1340.940.313 0.4554.68 ± 3.49
Dibromochloromethane µg/l 10.82 1331.620.698 1.128.15 ± 2.39
Dichloromethane µg/l 11.29 1401.70.611 0.848.04 ± 3.87
Tetrachloroethene µg/l 15.74 1392.361.38 2.2111.3 ± 2
Tetrachloromethane µg/l 20.7 1293.10532.37 3.7816.1 ± 1.22
trans-1,2-Dichloroethene µg/l 4.75 1280.710.446 0.6653.72 ± 1.56
Tribromomethane µg/l 4.93 1440.740.245 0.3743.43 ± 4.02
Trichloroethene µg/l 5.19 1330.770.324 0.5073.9 ± 2.54
Trichloromethane µg/l 1.16 1120.170.072 0.111.04 ± 1.13
This report was created with PROLab, a software by QuoData: http://www.quodata.de/
193
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0013
z-score-2 0 2 -2 0 2
Sample
CB01 A - VOC CB01 B - VOC
Mea
sura
nd
1,1,1-Trichloroethane
1,1-Dichloroethene
1,2-Dichloroethane
Bromodichloromethane
cis-1,2-Dichloroethene
Dibromochloromethane
Dichloromethane
Tetrachloroethene
Tetrachloromethane
trans-1,2-Dichloroethene
Tribromomethane
Trichloroethene
Trichloromethane
2.3
2.1
5.6
3.5
2.4
3.9
2.0
4.0
2.5
-2.3
194
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0014
The following results were achieved:
CB01ABTXParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
Benzene µg/l 0.647 1120.070.0519 0.06920.58 ± 0.96
Ethylbenzene µg/l 0.401 90.80.040.0383 0.0510.442 ± -0.8
o-Xylene µg/l 4.34 98.40.670.513 0.6844.41 ± -0.1
Sum of m-Xylene and p-Xylene µg/l 5.39 94.70.960.719 0.9595.69 ± -0.32
Toluene µg/l 0.85 1000.070.0889 0.1260.85 ± 0.00
Methyl-tert-butyl-ether µg/l - --0.361 0.4343.25 ± -
CB01ACKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l <0.1 (LOQ) --- -- ± -
1,1-Dichloroethene µg/l <0.1 (LOQ) --- -- ± -
1,2-Dichloroethane µg/l 13.9 1101.10.888 1.3212.7 ± 0.93
Bromodichloromethane µg/l 2.44 1080.30.141 0.2212.26 ± 0.79
cis-1,2-Dichloroethene µg/l 0.804 96.50.080.063 0.09620.833 ± -0.31
Dibromochloromethane µg/l 1.06 96.20.240.0847 0.1351.1 ± -0.31
Dichloromethane µg/l 2.04 1080.150.147 0.2131.89 ± 0.7
Tetrachloroethene µg/l 3.65 92.60.630.494 0.7733.94 ± -0.38
Tetrachloromethane µg/l 1.263 1060.090.127 0.2041.19 ± 0.36
trans-1,2-Dichloroethene µg/l 0.816 95.80.070.112 0.1710.852 ± -0.21
Tribromomethane µg/l <0.1 (LOQ) --- -- ± -
Trichloroethene µg/l 0.591 90.70.050.0506 0.07910.652 ± -0.77
Trichloromethane µg/l 12.5 1261.40.664 1.049.95 ± 2.46
CB01BBTXParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
Benzene µg/l 2.07 1000.130.241 0.3412.07 ± 0.01
Ethylbenzene µg/l 1.47 1010.150.201 0.2761.45 ± 0.07
o-Xylene µg/l 3.03 99.90.110.363 0.4843.03 ± -0.01
Sum of m-Xylene and p-Xylene µg/l 8.2 95.90.781.04 1.388.55 ± -0.25
Toluene µg/l 5.26 98.70.490.512 0.7045.33 ± -0.1
Methyl-tert-butyl-ether µg/l - --0.103 0.1240.882 ± -
CB01BCKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l 7.24 99.60.230.671 1.057.27 ± -0.03
This report was created with PROLab, a software by QuoData: http://www.quodata.de/
195
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0014
Parameter Unit Target ± Cl(99%) Result ± U Criteria Recovery z-score
1,1-Dichloroethene µg/l 8.38 92.50.61.3 1.99.06 ± -0.36
1,2-Dichloroethane µg/l 3.21 1030.150.227 0.3223.11 ± 0.32
Bromodichloromethane µg/l 7.63 1040.380.346 0.5157.33 ± 0.57
cis-1,2-Dichloroethene µg/l 4.33 92.50.230.313 0.4554.68 ± -0.78
Dibromochloromethane µg/l 7.25 891.030.698 1.128.15 ± -0.81
Dichloromethane µg/l 8.29 1030.540.611 0.848.04 ± 0.3
Tetrachloroethene µg/l 9.36 82.70.891.38 2.2111.3 ± -0.89
Tetrachloromethane µg/l 18.3 1141.12.37 3.7816.1 ± 0.59
trans-1,2-Dichloroethene µg/l 3.39 91.20.290.446 0.6653.72 ± -0.49
Tribromomethane µg/l 3.34 97.50.130.245 0.3743.43 ± -0.23
Trichloroethene µg/l 3.53 90.50.240.324 0.5073.9 ± -0.73
Trichloromethane µg/l 1.04 1000.090.072 0.111.04 ± 0.04
196
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0014
z-score-2 0 2 -2 0 2 -2 0 2 -2 0 2
Sample
CB01 A - BTEX / MTBE
Mea
sura
nd
Benzene
Ethylbenzene
o-Xylene
Sum of m-Xylene and p-Xylene
Toluene
1,1,1-Trichloroethane
1,1-Dichloroethene
1,2-Dichloroethane
Bromodichloromethane
cis-1,2-Dichloroethene
Dibromochloromethane
Dichloromethane
Tetrachloroethene
Tetrachloromethane
trans-1,2-Dichloroethene
Tribromomethane
Trichloroethene
Trichloromethane 2.5
VOC VOCCB01 B - BTEX / MTBE
197
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0015
The following results were achieved:
CB01ABTXParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
Benzene µg/l 0.56 96.50.1290.0519 0.06920.58 ± -0.29
Ethylbenzene µg/l 0.404 91.50.0930.0383 0.0510.442 ± -0.74
o-Xylene µg/l 4.43 1001.0630.513 0.6844.41 ± 0.03
Sum of m-Xylene and p-Xylene µg/l 5.4 94.81.6740.719 0.9595.69 ± -0.31
Toluene µg/l 0.87 1020.2520.0889 0.1260.85 ± 0.16
Methyl-tert-butyl-ether µg/l 3.46 1060.6570.361 0.4343.25 ± 0.47
CB01ACKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l <0.05 (LOQ) --- -- ± -
1,1-Dichloroethene µg/l <0.08 (LOQ) --- -- ± -
1,2-Dichloroethane µg/l 12.98 1024.4130.888 1.3212.7 ± 0.23
Bromodichloromethane µg/l 2.12 93.60.530.141 0.2212.26 ± -0.66
cis-1,2-Dichloroethene µg/l 0.68 81.60.1560.063 0.09620.833 ± -1.6
Dibromochloromethane µg/l 1.1 99.80.2860.0847 0.1351.1 ± -0.02
Dichloromethane µg/l 1.823 96.50.5290.147 0.2131.89 ± -0.31
Tetrachloroethene µg/l 3.38 85.81.1150.494 0.7733.94 ± -0.73
Tetrachloromethane µg/l 0.97 81.50.1840.127 0.2041.19 ± -1.08
trans-1,2-Dichloroethene µg/l 0.795 93.30.2070.112 0.1710.852 ± -0.33
Tribromomethane µg/l <0.05 (LOQ) --- -- ± -
Trichloroethene µg/l 0.63 96.70.2080.0506 0.07910.652 ± -0.27
Trichloromethane µg/l 8.7 87.42.3490.664 1.049.95 ± -1.2
CB01BBTXParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
Benzene µg/l 1.76 85.10.4050.241 0.3412.07 ± -0.9
Ethylbenzene µg/l 1.4 96.50.3220.201 0.2761.45 ± -0.18
o-Xylene µg/l 2.98 98.20.7150.363 0.4843.03 ± -0.11
Sum of m-Xylene and p-Xylene µg/l 7.66 89.62.3751.04 1.388.55 ± -0.64
Toluene µg/l 4.8 90.11.3920.512 0.7045.33 ± -0.75
Methyl-tert-butyl-ether µg/l 0.91 1030.1730.103 0.1240.882 ± 0.22
CB01BCKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l 6.06 83.31.3330.671 1.057.27 ± -1.15
This report was created with PROLab, a software by QuoData: http://www.quodata.de/
198
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0015
Parameter Unit Target ± Cl(99%) Result ± U Criteria Recovery z-score
1,1-Dichloroethene µg/l 7.54 83.21.2061.3 1.99.06 ± -0.8
1,2-Dichloroethane µg/l 3.17 1021.0780.227 0.3223.11 ± 0.19
Bromodichloromethane µg/l 7.22 98.41.8050.346 0.5157.33 ± -0.22
cis-1,2-Dichloroethene µg/l 4.3 91.80.9890.313 0.4554.68 ± -0.84
Dibromochloromethane µg/l 8.54 1052.220.698 1.128.15 ± 0.35
Dichloromethane µg/l 8 99.52.320.611 0.848.04 ± -0.05
Tetrachloroethene µg/l 8.46 74.82.7921.38 2.2111.3 ± -1.29
Tetrachloromethane µg/l 10.9 67.82.0712.37 3.7816.1 ± -1.37
trans-1,2-Dichloroethene µg/l 3.04 81.80.790.446 0.6653.72 ± -1.02
Tribromomethane µg/l 3.68 1071.1040.245 0.3743.43 ± 0.68
Trichloroethene µg/l 3.72 95.41.2280.324 0.5073.9 ± -0.36
Trichloromethane µg/l 1.02 98.50.2750.072 0.111.04 ± -0.14
199
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0015
z-score-2 0 2 -2 0 2 -2 0 2 -2 0 2
Sample
CB01 A - BTEX / MTBE
Mea
sura
nd
Benzene
Ethylbenzene
o-Xylene
Sum of m-Xylene and p-Xylene
Toluene
Methyl-tert-butyl-ether
1,1,1-Trichloroethane
1,1-Dichloroethene
1,2-Dichloroethane
Bromodichloromethane
cis-1,2-Dichloroethene
Dibromochloromethane
Dichloromethane
Tetrachloroethene
Tetrachloromethane
trans-1,2-Dichloroethene
Tribromomethane
Trichloroethene
Trichloromethane
VOC VOCCB01 B - BTEX / MTBE
200
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0016
The following results were achieved:
CB01ABTXParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
Benzene µg/l 0.53 91.30.10.0519 0.06920.58 ± -0.73
Ethylbenzene µg/l 0.41 92.90.080.0383 0.0510.442 ± -0.62
o-Xylene µg/l 4.6 1040.830.513 0.6844.41 ± 0.28
Sum of m-Xylene and p-Xylene µg/l 5.49 96.40.990.719 0.9595.69 ± -0.21
Toluene µg/l 0.81 95.30.150.0889 0.1260.85 ± -0.32
Methyl-tert-butyl-ether µg/l 3.1 95.20.560.361 0.4343.25 ± -0.36
CB01ACKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l <0.05 (LOQ) --- -- ± -
1,1-Dichloroethene µg/l <0.05 (LOQ) --- -- ± -
1,2-Dichloroethane µg/l 12.8 1012.30.888 1.3212.7 ± 0.1
Bromodichloromethane µg/l 2.05 90.50.370.141 0.2212.26 ± -0.97
cis-1,2-Dichloroethene µg/l 0.72 86.40.130.063 0.09620.833 ± -1.18
Dibromochloromethane µg/l 0.98 88.90.180.0847 0.1351.1 ± -0.9
Dichloromethane µg/l 1.67 88.40.310.147 0.2131.89 ± -1.03
Tetrachloroethene µg/l 3.6 91.30.650.494 0.7733.94 ± -0.44
Tetrachloromethane µg/l 1.03 86.50.190.127 0.2041.19 ± -0.79
trans-1,2-Dichloroethene µg/l 0.65 76.30.120.112 0.1710.852 ± -1.18
Tribromomethane µg/l <0.05 (LOQ) --- -- ± -
Trichloroethene µg/l 0.58 890.110.0506 0.07910.652 ± -0.91
Trichloromethane µg/l 9.69 97.41.750.664 1.049.95 ± -0.25
CB01BBTXParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
Benzene µg/l 1.84 890.340.241 0.3412.07 ± -0.67
Ethylbenzene µg/l 1.56 1080.290.201 0.2761.45 ± 0.4
o-Xylene µg/l 3.22 1060.580.363 0.4843.03 ± 0.39
Sum of m-Xylene and p-Xylene µg/l 8.92 1041.611.04 1.388.55 ± 0.27
Toluene µg/l 5.42 1020.980.512 0.7045.33 ± 0.13
Methyl-tert-butyl-ether µg/l 0.83 94.10.150.103 0.1240.882 ± -0.42
CB01BCKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l 6.82 93.81.230.671 1.057.27 ± -0.43
This report was created with PROLab, a software by QuoData: http://www.quodata.de/
201
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0016
Parameter Unit Target ± Cl(99%) Result ± U Criteria Recovery z-score
1,1-Dichloroethene µg/l 8.15 89.91.471.3 1.99.06 ± -0.48
1,2-Dichloroethane µg/l 2.85 91.70.520.227 0.3223.11 ± -0.8
Bromodichloromethane µg/l 7.11 96.91.280.346 0.5157.33 ± -0.44
cis-1,2-Dichloroethene µg/l 4.35 92.90.790.313 0.4554.68 ± -0.73
Dibromochloromethane µg/l 7.64 93.71.380.698 1.128.15 ± -0.46
Dichloromethane µg/l 7.23 89.91.310.611 0.848.04 ± -0.96
Tetrachloroethene µg/l 10.4 91.91.91.38 2.2111.3 ± -0.41
Tetrachloromethane µg/l 15.9 992.92.37 3.7816.1 ± -0.04
trans-1,2-Dichloroethene µg/l 3.13 84.20.570.446 0.6653.72 ± -0.88
Tribromomethane µg/l 3.67 1070.670.245 0.3743.43 ± 0.65
Trichloroethene µg/l 3.81 97.70.690.324 0.5073.9 ± -0.18
Trichloromethane µg/l 0.84 81.10.160.072 0.111.04 ± -1.78
202
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0016
z-score-2 0 2 -2 0 2 -2 0 2 -2 0 2
Sample
CB01 A - BTEX / MTBE
Mea
sura
nd
Benzene
Ethylbenzene
o-Xylene
Sum of m-Xylene and p-Xylene
Toluene
Methyl-tert-butyl-ether
1,1,1-Trichloroethane
1,1-Dichloroethene
1,2-Dichloroethane
Bromodichloromethane
cis-1,2-Dichloroethene
Dibromochloromethane
Dichloromethane
Tetrachloroethene
Tetrachloromethane
trans-1,2-Dichloroethene
Tribromomethane
Trichloroethene
Trichloromethane
VOC VOCCB01 B - BTEX / MTBE
203
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0017
The following results were achieved:
CB01ABTXParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
Benzene µg/l 0.59 1020.1780.0519 0.06920.58 ± 0.14
Ethylbenzene µg/l 0.52 1180.1550.0383 0.0510.442 ± 1.54
o-Xylene µg/l 4.79 1091.4380.513 0.6844.41 ± 0.55
Sum of m-Xylene and p-Xylene µg/l 5.25 92.21.5760.719 0.9595.69 ± -0.46
Toluene µg/l 0.97 1140.2920.0889 0.1260.85 ± 0.96
Methyl-tert-butyl-ether µg/l 3.25 99.90.9740.361 0.4343.25 ± -0.01
CB01ACKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l <0.2 (LOQ) --- -- ± -
1,1-Dichloroethene µg/l <2 (LOQ) --- -- ± -
1,2-Dichloroethane µg/l 11.8 93.13.540.888 1.3212.7 ± -0.66
Bromodichloromethane µg/l 1.92 84.80.5750.141 0.2212.26 ± -1.56
cis-1,2-Dichloroethene µg/l 0.81 97.20.2410.063 0.09620.833 ± -0.24
Dibromochloromethane µg/l 1.1 99.80.3290.0847 0.1351.1 ± -0.02
Dichloromethane µg/l 1.69 89.40.5050.147 0.2131.89 ± -0.94
Tetrachloroethene µg/l 3.9 98.91.170.494 0.7733.94 ± -0.05
Tetrachloromethane µg/l 1.13 94.90.340.127 0.2041.19 ± -0.3
trans-1,2-Dichloroethene µg/l 0.9 1060.2690.112 0.1710.852 ± 0.28
Tribromomethane µg/l <0.5 (LOQ) --- -- ± -
Trichloroethene µg/l 0.67 1030.20.0506 0.07910.652 ± 0.23
Trichloromethane µg/l 8.89 89.42.670.664 1.049.95 ± -1.02
CB01BBTXParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
Benzene µg/l 1.96 94.80.5860.241 0.3412.07 ± -0.31
Ethylbenzene µg/l 1.72 1190.5150.201 0.2761.45 ± 0.98
o-Xylene µg/l 3.67 1211.10.363 0.4843.03 ± 1.32
Sum of m-Xylene and p-Xylene µg/l 8.12 952.441.04 1.388.55 ± -0.31
Toluene µg/l 5.83 1091.750.512 0.7045.33 ± 0.71
Methyl-tert-butyl-ether µg/l 0.72 81.60.2150.103 0.1240.882 ± -1.31
CB01BCKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l 7.72 1062.320.671 1.057.27 ± 0.43
This report was created with PROLab, a software by QuoData: http://www.quodata.de/
204
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0017
Parameter Unit Target ± Cl(99%) Result ± U Criteria Recovery z-score
1,1-Dichloroethene µg/l 11.5 1273.441.3 1.99.06 ± 1.29
1,2-Dichloroethane µg/l 5.05 1631.520.227 0.3223.11 ± 6.04
Bromodichloromethane µg/l 6.51 88.81.950.346 0.5157.33 ± -1.6
cis-1,2-Dichloroethene µg/l 5.09 1091.530.313 0.4554.68 ± 0.9
Dibromochloromethane µg/l 7.19 88.22.160.698 1.128.15 ± -0.86
Dichloromethane µg/l 9.39 1172.820.611 0.848.04 ± 1.61
Tetrachloroethene µg/l 12.6 1113.781.38 2.2111.3 ± 0.58
Tetrachloromethane µg/l 16.6 1034.972.37 3.7816.1 ± 0.14
trans-1,2-Dichloroethene µg/l 3.91 1051.170.446 0.6653.72 ± 0.29
Tribromomethane µg/l 2.83 82.60.850.245 0.3743.43 ± -1.6
Trichloroethene µg/l 4.49 1151.350.324 0.5073.9 ± 1.16
Trichloromethane µg/l 1.02 98.50.3070.072 0.111.04 ± -0.14
205
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0017
z-score-2 0 2 -2 0 2 -2 0 2 -2 0 2
Sample
CB01 A - BTEX / MTBE
Mea
sura
nd
Benzene
Ethylbenzene
o-Xylene
Sum of m-Xylene and p-Xylene
Toluene
Methyl-tert-butyl-ether
1,1,1-Trichloroethane
1,1-Dichloroethene
1,2-Dichloroethane
Bromodichloromethane
cis-1,2-Dichloroethene
Dibromochloromethane
Dichloromethane
Tetrachloroethene
Tetrachloromethane
trans-1,2-Dichloroethene
Tribromomethane
Trichloroethene
Trichloromethane
6.0
VOC VOCCB01 B - BTEX / MTBE
206
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0018
The following results were achieved:
CB01ABTXParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
Benzene µg/l 0.693 1190.2080.0519 0.06920.58 ± 1.63
Ethylbenzene µg/l 0.505 1140.1870.0383 0.0510.442 ± 1.24
o-Xylene µg/l 5.7 1291.680.513 0.6844.41 ± 1.88
Sum of m-Xylene and p-Xylene µg/l 7.36 1292.410.719 0.9595.69 ± 1.74
Toluene µg/l 1.09 1280.380.0889 0.1260.85 ± 1.91
Methyl-tert-butyl-ether µg/l 3.49 1071.050.361 0.4343.25 ± 0.54
CB01ACKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l <0.1 (LOQ) --- -- ± -
1,1-Dichloroethene µg/l <0.1 (LOQ) --- -- ± -
1,2-Dichloroethane µg/l 17.9 14130.888 1.3212.7 ± 3.95
Bromodichloromethane µg/l 2.82 1250.560.141 0.2212.26 ± 2.51
cis-1,2-Dichloroethene µg/l 1.01 1210.260.063 0.09620.833 ± 1.84
Dibromochloromethane µg/l 1.11 1010.20.0847 0.1351.1 ± 0.06
Dichloromethane µg/l 2.25 1190.510.147 0.2131.89 ± 1.69
Tetrachloroethene µg/l 4.95 1261.730.494 0.7733.94 ± 1.3
Tetrachloromethane µg/l 1.63 1370.440.127 0.2041.19 ± 2.16
trans-1,2-Dichloroethene µg/l 1.07 1260.340.112 0.1710.852 ± 1.28
Tribromomethane µg/l <0.1 (LOQ) --- -- ± -
Trichloroethene µg/l 0.965 1480.2920.0506 0.07910.652 ± 3.96
Trichloromethane µg/l 13.1 1323.20.664 1.049.95 ± 3.03
CB01BBTXParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
Benzene µg/l 2.44 1180.780.241 0.3412.07 ± 1.09
Ethylbenzene µg/l 2 1380.620.201 0.2761.45 ± 1.99
o-Xylene µg/l 4.02 1331.180.363 0.4843.03 ± 2.04
Sum of m-Xylene and p-Xylene µg/l 10.3 1213.081.04 1.388.55 ± 1.27
Toluene µg/l 6.8 1282.060.512 0.7045.33 ± 2.09
Methyl-tert-butyl-ether µg/l 1.05 1190.320.103 0.1240.882 ± 1.35
CB01BCKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l 9.3 1282.730.671 1.057.27 ± 1.93
This report was created with PROLab, a software by QuoData: http://www.quodata.de/
207
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0018
Parameter Unit Target ± Cl(99%) Result ± U Criteria Recovery z-score
1,1-Dichloroethene µg/l 14.9 1644.71.3 1.99.06 ± 3.08
1,2-Dichloroethane µg/l 4.96 1601.610.227 0.3223.11 ± 5.76
Bromodichloromethane µg/l 9.87 1352.970.346 0.5157.33 ± 4.92
cis-1,2-Dichloroethene µg/l 6.74 1441.750.313 0.4554.68 ± 4.52
Dibromochloromethane µg/l 9.91 1223.020.698 1.128.15 ± 1.58
Dichloromethane µg/l 11.9 1483.60.611 0.848.04 ± 4.6
Tetrachloroethene µg/l 13.6 1204.91.38 2.2111.3 ± 1.03
Tetrachloromethane µg/l 22 1376.62.37 3.7816.1 ± 1.57
trans-1,2-Dichloroethene µg/l 5.34 1441.60.446 0.6653.72 ± 2.44
Tribromomethane µg/l 3.4 99.21.040.245 0.3743.43 ± -0.07
Trichloroethene µg/l 4.86 1251.380.324 0.5073.9 ± 1.89
Trichloromethane µg/l 1.16 1120.350.072 0.111.04 ± 1.13
208
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0018
z-score-2 0 2 -2 0 2 -2 0 2 -2 0 2
Sample
CB01 A - BTEX / MTBE
Mea
sura
nd
Benzene
Ethylbenzene
o-Xylene
Sum of m-Xylene and p-Xylene
Toluene
Methyl-tert-butyl-ether
1,1,1-Trichloroethane
1,1-Dichloroethene
1,2-Dichloroethane
Bromodichloromethane
cis-1,2-Dichloroethene
Dibromochloromethane
Dichloromethane
Tetrachloroethene
Tetrachloromethane
trans-1,2-Dichloroethene
Tribromomethane
Trichloroethene
Trichloromethane
3.9
2.5
2.2
4.0
3.0
2.0
2.1
3.1
5.8
4.9
4.5
4.6
2.4
VOC VOCCB01 B - BTEX / MTBE
209
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0019
The following results were achieved:
CB01ABTXParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
Benzene µg/l 0.47 810.140.0519 0.06920.58 ± -1.59
Ethylbenzene µg/l 0.44 99.60.130.0383 0.0510.442 ± -0.03
o-Xylene µg/l 3.7 83.91.110.513 0.6844.41 ± -1.04
Sum of m-Xylene and p-Xylene µg/l 2.34 41.10.70.719 0.9595.69 ± -3.5
Toluene µg/l 0.67 78.80.20.0889 0.1260.85 ± -1.43
Methyl-tert-butyl-ether µg/l 2.96 90.90.890.361 0.4343.25 ± -0.68
CB01ACKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l <0.1 (LOQ) --- -- ± -
1,1-Dichloroethene µg/l <0.1 (LOQ) --- -- ± -
1,2-Dichloroethane µg/l 12.14 95.83.640.888 1.3212.7 ± -0.4
Bromodichloromethane µg/l 2.04 90.10.610.141 0.2212.26 ± -1.02
cis-1,2-Dichloroethene µg/l 0.71 85.20.210.063 0.09620.833 ± -1.28
Dibromochloromethane µg/l 0.92 83.50.270.0847 0.1351.1 ± -1.35
Dichloromethane µg/l 1.76 93.10.60.147 0.2131.89 ± -0.61
Tetrachloroethene µg/l 3.38 85.81.010.494 0.7733.94 ± -0.73
Tetrachloromethane µg/l 1.12 94.10.340.127 0.2041.19 ± -0.34
trans-1,2-Dichloroethene µg/l 0.72 84.50.220.112 0.1710.852 ± -0.77
Tribromomethane µg/l <0.1 (LOQ) --- -- ± -
Trichloroethene µg/l 0.58 890.170.0506 0.07910.652 ± -0.91
Trichloromethane µg/l 9.53 95.82.860.664 1.049.95 ± -0.4
CB01BBTXParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
Benzene µg/l 1.63 78.90.490.241 0.3412.07 ± -1.28
Ethylbenzene µg/l 1.08 74.50.320.201 0.2761.45 ± -1.34
o-Xylene µg/l 2.34 77.10.70.363 0.4843.03 ± -1.43
Sum of m-Xylene and p-Xylene µg/l 3.59 421.081.04 1.388.55 ± -3.59
Toluene µg/l 4.51 84.61.350.512 0.7045.33 ± -1.16
Methyl-tert-butyl-ether µg/l 0.83 94.10.250.103 0.1240.882 ± -0.42
CB01BCKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l 6.56 90.21.970.671 1.057.27 ± -0.68
This report was created with PROLab, a software by QuoData: http://www.quodata.de/
210
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0019
Parameter Unit Target ± Cl(99%) Result ± U Criteria Recovery z-score
1,1-Dichloroethene µg/l 8.84 97.52.651.3 1.99.06 ± -0.12
1,2-Dichloroethane µg/l 2.85 91.70.850.227 0.3223.11 ± -0.8
Bromodichloromethane µg/l 7 95.42.10.346 0.5157.33 ± -0.65
cis-1,2-Dichloroethene µg/l 4.02 85.91.210.313 0.4554.68 ± -1.46
Dibromochloromethane µg/l 7.09 872.130.698 1.128.15 ± -0.95
Dichloromethane µg/l 8.03 99.92.410.611 0.848.04 ± -0.01
Tetrachloroethene µg/l 9.68 85.62.91.38 2.2111.3 ± -0.74
Tetrachloromethane µg/l 13.04 81.23.912.37 3.7816.1 ± -0.8
trans-1,2-Dichloroethene µg/l 3.1 83.40.930.446 0.6653.72 ± -0.93
Tribromomethane µg/l 2.91 84.90.870.245 0.3743.43 ± -1.38
Trichloroethene µg/l 3.38 86.61.010.324 0.5073.9 ± -1.03
Trichloromethane µg/l 1.03 99.40.310.072 0.111.04 ± -0.05
211
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0019
z-score-2 0 2 -2 0 2 -2 0 2 -2 0 2
Sample
CB01 A - BTEX / MTBE
Mea
sura
nd
Benzene
Ethylbenzene
o-Xylene
Sum of m-Xylene and p-Xylene
Toluene
Methyl-tert-butyl-ether
1,1,1-Trichloroethane
1,1-Dichloroethene
1,2-Dichloroethane
Bromodichloromethane
cis-1,2-Dichloroethene
Dibromochloromethane
Dichloromethane
Tetrachloroethene
Tetrachloromethane
trans-1,2-Dichloroethene
Tribromomethane
Trichloroethene
Trichloromethane
-3.5 -3.6
VOC VOCCB01 B - BTEX / MTBE
212
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0020
The following results were achieved:
CB01ABTXParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
Benzene µg/l 0.55 94.80.050.0519 0.06920.58 ± -0.44
Ethylbenzene µg/l 0.41 92.90.040.0383 0.0510.442 ± -0.62
o-Xylene µg/l 3.42 77.50.30.513 0.6844.41 ± -1.45
Sum of m-Xylene and p-Xylene µg/l 4.65 81.70.50.719 0.9595.69 ± -1.09
Toluene µg/l 0.8 94.10.080.0889 0.1260.85 ± -0.4
Methyl-tert-butyl-ether µg/l 2.9 89.10.30.361 0.4343.25 ± -0.82
CB01ACKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l <0.08 (LOQ) --- -- ± -
1,1-Dichloroethene µg/l <0.1 (LOQ) --- -- ± -
1,2-Dichloroethane µg/l 11.3 89.210.888 1.3212.7 ± -1.04
Bromodichloromethane µg/l 2.08 91.80.20.141 0.2212.26 ± -0.84
cis-1,2-Dichloroethene µg/l 0.82 98.40.080.063 0.09620.833 ± -0.14
Dibromochloromethane µg/l 1.08 980.10.0847 0.1351.1 ± -0.16
Dichloromethane µg/l 1.66 87.80.20.147 0.2131.89 ± -1.08
Tetrachloroethene µg/l 3.25 82.50.30.494 0.7733.94 ± -0.9
Tetrachloromethane µg/l 1.25 1050.10.127 0.2041.19 ± 0.29
trans-1,2-Dichloroethene µg/l 0.83 97.40.080.112 0.1710.852 ± -0.13
Tribromomethane µg/l <0.08 (LOQ) --- -- ± -
Trichloroethene µg/l 0.66 1010.070.0506 0.07910.652 ± 0.11
Trichloromethane µg/l 9.35 9410.664 1.049.95 ± -0.58
CB01BBTXParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
Benzene µg/l 1.91 92.40.20.241 0.3412.07 ± -0.46
Ethylbenzene µg/l 1.33 91.70.10.201 0.2761.45 ± -0.44
o-Xylene µg/l 2.43 80.10.20.363 0.4843.03 ± -1.25
Sum of m-Xylene and p-Xylene µg/l 6.75 790.71.04 1.388.55 ± -1.3
Toluene µg/l 4.73 88.80.50.512 0.7045.33 ± -0.85
Methyl-tert-butyl-ether µg/l 0.72 81.60.070.103 0.1240.882 ± -1.31
CB01BCKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l 7.17 98.60.70.671 1.057.27 ± -0.1
This report was created with PROLab, a software by QuoData: http://www.quodata.de/
213
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0020
Parameter Unit Target ± Cl(99%) Result ± U Criteria Recovery z-score
1,1-Dichloroethene µg/l 8.88 980.91.3 1.99.06 ± -0.1
1,2-Dichloroethane µg/l 3.02 97.20.30.227 0.3223.11 ± -0.27
Bromodichloromethane µg/l 7.5 1020.70.346 0.5157.33 ± 0.32
cis-1,2-Dichloroethene µg/l 5.03 1070.50.313 0.4554.68 ± 0.77
Dibromochloromethane µg/l 8.14 99.90.80.698 1.128.15 ± -0.01
Dichloromethane µg/l 8.7 1080.90.611 0.848.04 ± 0.79
Tetrachloroethene µg/l 10.1 89.311.38 2.2111.3 ± -0.55
Tetrachloromethane µg/l 16.3 10122.37 3.7816.1 ± 0.06
trans-1,2-Dichloroethene µg/l 3.93 1060.40.446 0.6653.72 ± 0.32
Tribromomethane µg/l 3.32 96.90.30.245 0.3743.43 ± -0.29
Trichloroethene µg/l 4.24 1090.40.324 0.5073.9 ± 0.67
Trichloromethane µg/l 1.14 1100.10.072 0.111.04 ± 0.95
214
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0020
z-score-2 0 2 -2 0 2 -2 0 2 -2 0 2
Sample
CB01 A - BTEX / MTBE
Mea
sura
nd
Benzene
Ethylbenzene
o-Xylene
Sum of m-Xylene and p-Xylene
Toluene
Methyl-tert-butyl-ether
1,1,1-Trichloroethane
1,1-Dichloroethene
1,2-Dichloroethane
Bromodichloromethane
cis-1,2-Dichloroethene
Dibromochloromethane
Dichloromethane
Tetrachloroethene
Tetrachloromethane
trans-1,2-Dichloroethene
Tribromomethane
Trichloroethene
Trichloromethane
VOC VOCCB01 B - BTEX / MTBE
215
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0021
The following results were achieved:
CB01ACKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l <0.1 (LOQ) --- -- ± -
1,1-Dichloroethene µg/l <0.2 (LOQ) --- -- ± -
1,2-Dichloroethane µg/l 11.27 88.91.690.888 1.3212.7 ± -1.06
Bromodichloromethane µg/l 2.18 96.30.330.141 0.2212.26 ± -0.39
cis-1,2-Dichloroethene µg/l 0.73 87.60.110.063 0.09620.833 ± -1.08
Dibromochloromethane µg/l 0.98 88.90.150.0847 0.1351.1 ± -0.9
Dichloromethane µg/l 2.04 1080.310.147 0.2131.89 ± 0.7
Tetrachloroethene µg/l 3.56 90.30.530.494 0.7733.94 ± -0.49
Tetrachloromethane µg/l 1.03 86.50.150.127 0.2041.19 ± -0.79
trans-1,2-Dichloroethene µg/l 0.79 92.70.120.112 0.1710.852 ± -0.36
Tribromomethane µg/l <0.1 (LOQ) --- -- ± -
Trichloroethene µg/l 0.63 96.70.090.0506 0.07910.652 ± -0.27
Trichloromethane µg/l 8.92 89.71.340.664 1.049.95 ± -0.99
CB01BCKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l 6.51 89.50.980.671 1.057.27 ± -0.73
1,1-Dichloroethene µg/l 8.47 93.51.271.3 1.99.06 ± -0.31
1,2-Dichloroethane µg/l 2.99 96.20.450.227 0.3223.11 ± -0.37
Bromodichloromethane µg/l 6.69 91.210.346 0.5157.33 ± -1.25
cis-1,2-Dichloroethene µg/l 4.67 99.70.70.313 0.4554.68 ± -0.03
Dibromochloromethane µg/l 7.01 861.050.698 1.128.15 ± -1.02
Dichloromethane µg/l 7.83 97.41.170.611 0.848.04 ± -0.25
Tetrachloroethene µg/l 9.32 82.41.41.38 2.2111.3 ± -0.9
Tetrachloromethane µg/l 13.25 82.51.992.37 3.7816.1 ± -0.74
trans-1,2-Dichloroethene µg/l 3.94 1060.590.446 0.6653.72 ± 0.34
Tribromomethane µg/l 3.07 89.60.460.245 0.3743.43 ± -0.96
Trichloroethene µg/l 3.68 94.30.550.324 0.5073.9 ± -0.43
Trichloromethane µg/l 0.92 88.80.140.072 0.111.04 ± -1.05
This report was created with PROLab, a software by QuoData: http://www.quodata.de/
216
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0021
z-score-2 0 2 -2 0 2
Sample
CB01 A - VOC CB01 B - VOC
Mea
sura
nd
1,1,1-Trichloroethane
1,1-Dichloroethene
1,2-Dichloroethane
Bromodichloromethane
cis-1,2-Dichloroethene
Dibromochloromethane
Dichloromethane
Tetrachloroethene
Tetrachloromethane
trans-1,2-Dichloroethene
Tribromomethane
Trichloroethene
Trichloromethane
217
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0022
The following results were achieved:
CB01ABTXParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
Benzene µg/l 0.594 1020.083160.0519 0.06920.58 ± 0.2
Ethylbenzene µg/l 0.475 1080.042750.0383 0.0510.442 ± 0.66
o-Xylene µg/l 4.34 98.40.39060.513 0.6844.41 ± -0.1
Sum of m-Xylene and p-Xylene µg/l 5.5 96.60.440.719 0.9595.69 ± -0.2
Toluene µg/l 0.854 1000.076860.0889 0.1260.85 ± 0.03
Methyl-tert-butyl-ether µg/l 3.11 95.60.52870.361 0.4343.25 ± -0.33
CB01ACKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l <0.05 (LOQ) --- -- ± -
1,1-Dichloroethene µg/l <0.05 (LOQ) --- -- ± -
1,2-Dichloroethane µg/l 15.4 1222.4640.888 1.3212.7 ± 2.06
Bromodichloromethane µg/l 2.2 97.10.2640.141 0.2212.26 ± -0.29
cis-1,2-Dichloroethene µg/l 0.843 1010.118020.063 0.09620.833 ± 0.1
Dibromochloromethane µg/l 1.07 97.10.1070.0847 0.1351.1 ± -0.24
Dichloromethane µg/l 2.09 1110.4180.147 0.2131.89 ± 0.94
Tetrachloroethene µg/l 3.24 82.20.9720.494 0.7733.94 ± -0.91
Tetrachloromethane µg/l 1.13 94.90.20340.127 0.2041.19 ± -0.3
trans-1,2-Dichloroethene µg/l 0.78 91.50.06240.112 0.1710.852 ± -0.42
Tribromomethane µg/l <0.05 (LOQ) --- -- ± -
Trichloroethene µg/l 0.642 98.50.16050.0506 0.07910.652 ± -0.12
Trichloromethane µg/l 10.5 1061.5750.664 1.049.95 ± 0.53
CB01BBTXParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
Benzene µg/l 1.94 93.90.27160.241 0.3412.07 ± -0.37
Ethylbenzene µg/l 1.51 1040.13590.201 0.2761.45 ± 0.22
o-Xylene µg/l 2.95 97.20.26550.363 0.4843.03 ± -0.17
Sum of m-Xylene and p-Xylene µg/l 8.89 1040.71121.04 1.388.55 ± 0.25
Toluene µg/l 4.94 92.70.44460.512 0.7045.33 ± -0.55
Methyl-tert-butyl-ether µg/l 0.911 1030.154870.103 0.1240.882 ± 0.23
CB01BCKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l 6.99 96.11.18830.671 1.057.27 ± -0.27
This report was created with PROLab, a software by QuoData: http://www.quodata.de/
218
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0022
Parameter Unit Target ± Cl(99%) Result ± U Criteria Recovery z-score
1,1-Dichloroethene µg/l 8.46 93.30.67681.3 1.99.06 ± -0.32
1,2-Dichloroethane µg/l 3.56 1150.56960.227 0.3223.11 ± 1.41
Bromodichloromethane µg/l 7.71 1050.92520.346 0.5157.33 ± 0.73
cis-1,2-Dichloroethene µg/l 4.54 970.63560.313 0.4554.68 ± -0.31
Dibromochloromethane µg/l 6.51 79.90.6510.698 1.128.15 ± -1.47
Dichloromethane µg/l 8.3 1031.660.611 0.848.04 ± 0.31
Tetrachloroethene µg/l 9.02 79.72.7061.38 2.2111.3 ± -1.04
Tetrachloromethane µg/l 16.4 1022.9522.37 3.7816.1 ± 0.09
trans-1,2-Dichloroethene µg/l 3.36 90.40.26880.446 0.6653.72 ± -0.54
Tribromomethane µg/l 3 87.50.240.245 0.3743.43 ± -1.14
Trichloroethene µg/l 3.67 94.10.91750.324 0.5073.9 ± -0.46
Trichloromethane µg/l 1.01 97.50.15150.072 0.111.04 ± -0.24
219
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0022
z-score-2 0 2 -2 0 2 -2 0 2 -2 0 2
Sample
CB01 A - BTEX / MTBE
Mea
sura
nd
Benzene
Ethylbenzene
o-Xylene
Sum of m-Xylene and p-Xylene
Toluene
Methyl-tert-butyl-ether
1,1,1-Trichloroethane
1,1-Dichloroethene
1,2-Dichloroethane
Bromodichloromethane
cis-1,2-Dichloroethene
Dibromochloromethane
Dichloromethane
Tetrachloroethene
Tetrachloromethane
trans-1,2-Dichloroethene
Tribromomethane
Trichloroethene
Trichloromethane
2.1
VOC VOCCB01 B - BTEX / MTBE
220
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0023
The following results were achieved:
CB01ACKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l <0.02 (LOQ) --- -- ± -
1,1-Dichloroethene µg/l <0.03 (LOQ) --- -- ± -
1,2-Dichloroethane µg/l 12.8 1012.50.888 1.3212.7 ± 0.1
Bromodichloromethane µg/l 2.36 1040.470.141 0.2212.26 ± 0.43
cis-1,2-Dichloroethene µg/l 0.9 1080.180.063 0.09620.833 ± 0.69
Dibromochloromethane µg/l 1.21 1100.240.0847 0.1351.1 ± 0.8
Dichloromethane µg/l 1.77 93.70.350.147 0.2131.89 ± -0.56
Tetrachloroethene µg/l 4.17 1060.830.494 0.7733.94 ± 0.3
Tetrachloromethane µg/l 1.17 98.30.230.127 0.2041.19 ± -0.1
trans-1,2-Dichloroethene µg/l 0.83 97.40.170.112 0.1710.852 ± -0.13
Tribromomethane µg/l 0.03 -0.01- -- ± -
Trichloroethene µg/l 0.75 1150.150.0506 0.07910.652 ± 1.24
Trichloromethane µg/l 10.4 1052.10.664 1.049.95 ± 0.43
CB01BCKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l 8.04 1111.610.671 1.057.27 ± 0.73
1,1-Dichloroethene µg/l 9.47 1041.891.3 1.99.06 ± 0.21
1,2-Dichloroethane µg/l 3.02 97.20.60.227 0.3223.11 ± -0.27
Bromodichloromethane µg/l 7.92 1081.580.346 0.5157.33 ± 1.14
cis-1,2-Dichloroethene µg/l 4.83 1030.960.313 0.4554.68 ± 0.33
Dibromochloromethane µg/l 8.69 1071.740.698 1.128.15 ± 0.48
Dichloromethane µg/l 8.2 1021.640.611 0.848.04 ± 0.19
Tetrachloroethene µg/l 11.4 1012.281.38 2.2111.3 ± 0.04
Tetrachloromethane µg/l 17.1 1063.42.37 3.7816.1 ± 0.27
trans-1,2-Dichloroethene µg/l 4.04 1090.810.446 0.6653.72 ± 0.49
Tribromomethane µg/l 3.95 1150.790.245 0.3743.43 ± 1.4
Trichloroethene µg/l 4.22 1080.840.324 0.5073.9 ± 0.63
Trichloromethane µg/l 1.12 1080.220.072 0.111.04 ± 0.77
This report was created with PROLab, a software by QuoData: http://www.quodata.de/
221
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0023
z-score-2 0 2 -2 0 2
Sample
CB01 A - VOC CB01 B - VOC
Mea
sura
nd
1,1,1-Trichloroethane
1,1-Dichloroethene
1,2-Dichloroethane
Bromodichloromethane
cis-1,2-Dichloroethene
Dibromochloromethane
Dichloromethane
Tetrachloroethene
Tetrachloromethane
trans-1,2-Dichloroethene
Tribromomethane
Trichloroethene
Trichloromethane
222
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0024
The following results were achieved:
CB01ABTXParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
Benzene µg/l 0.58 1000.170.0519 0.06920.58 ± 0.00
Ethylbenzene µg/l 0.45 1020.140.0383 0.0510.442 ± 0.17
o-Xylene µg/l 4.66 1061.40.513 0.6844.41 ± 0.36
Sum of m-Xylene and p-Xylene µg/l 5.8 1021.740.719 0.9595.69 ± 0.11
Toluene µg/l 0.87 1020.260.0889 0.1260.85 ± 0.16
Methyl-tert-butyl-ether µg/l - --0.361 0.4343.25 ± -
CB01ACKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l <0.04 (LOQ) --- -- ± -
1,1-Dichloroethene µg/l <0.04 (LOQ) --- -- ± -
1,2-Dichloroethane µg/l 12.96 1023.890.888 1.3212.7 ± 0.22
Bromodichloromethane µg/l 2.14 94.50.640.141 0.2212.26 ± -0.57
cis-1,2-Dichloroethene µg/l 0.92 1100.280.063 0.09620.833 ± 0.9
Dibromochloromethane µg/l 0.89 80.70.270.0847 0.1351.1 ± -1.57
Dichloromethane µg/l 1.94 1030.580.147 0.2131.89 ± 0.23
Tetrachloroethene µg/l 4.07 1031.220.494 0.7733.94 ± 0.17
Tetrachloromethane µg/l 1.2 1010.360.127 0.2041.19 ± 0.05
trans-1,2-Dichloroethene µg/l 1.01 1190.30.112 0.1710.852 ± 0.93
Tribromomethane µg/l <0.04 (LOQ) --- -- ± -
Trichloroethene µg/l 0.71 1090.210.0506 0.07910.652 ± 0.74
Trichloromethane µg/l 10.58 1063.180.664 1.049.95 ± 0.61
CB01BBTXParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
Benzene µg/l 1.85 89.50.560.241 0.3412.07 ± -0.64
Ethylbenzene µg/l 1.46 1010.440.201 0.2761.45 ± 0.04
o-Xylene µg/l 3.09 1020.930.363 0.4843.03 ± 0.12
Sum of m-Xylene and p-Xylene µg/l 8.67 1012.61.04 1.388.55 ± 0.09
Toluene µg/l 5.11 95.91.530.512 0.7045.33 ± -0.31
Methyl-tert-butyl-ether µg/l - --0.103 0.1240.882 ± -
CB01BCKWParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
1,1,1-Trichloroethane µg/l 7.59 1042.280.671 1.057.27 ± 0.3
This report was created with PROLab, a software by QuoData: http://www.quodata.de/
223
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0024
Parameter Unit Target ± Cl(99%) Result ± U Criteria Recovery z-score
1,1-Dichloroethene µg/l 11.66 1293.51.3 1.99.06 ± 1.37
1,2-Dichloroethane µg/l 2.81 90.40.840.227 0.3223.11 ± -0.93
Bromodichloromethane µg/l 7.23 98.62.170.346 0.5157.33 ± -0.2
cis-1,2-Dichloroethene µg/l 4.46 95.31.340.313 0.4554.68 ± -0.49
Dibromochloromethane µg/l 7 85.92.10.698 1.128.15 ± -1.03
Dichloromethane µg/l 9.11 1132.730.611 0.848.04 ± 1.28
Tetrachloroethene µg/l 11.11 98.23.331.38 2.2111.3 ± -0.09
Tetrachloromethane µg/l 18.46 1155.542.37 3.7816.1 ± 0.63
trans-1,2-Dichloroethene µg/l 4.01 1081.20.446 0.6653.72 ± 0.44
Tribromomethane µg/l 2.85 83.20.860.245 0.3743.43 ± -1.54
Trichloroethene µg/l 3.86 991.160.324 0.5073.9 ± -0.08
Trichloromethane µg/l 0.98 94.60.30.072 0.111.04 ± -0.51
224
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0024
z-score-2 0 2 -2 0 2 -2 0 2 -2 0 2
Sample
CB01 A - BTEX / MTBE
Mea
sura
nd
Benzene
Ethylbenzene
o-Xylene
Sum of m-Xylene and p-Xylene
Toluene
1,1,1-Trichloroethane
1,1-Dichloroethene
1,2-Dichloroethane
Bromodichloromethane
cis-1,2-Dichloroethene
Dibromochloromethane
Dichloromethane
Tetrachloroethene
Tetrachloromethane
trans-1,2-Dichloroethene
Tribromomethane
Trichloroethene
Trichloromethane
VOC VOCCB01 B - BTEX / MTBE
225
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0025
The following results were achieved:
CB01ABTXParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
Benzene µg/l 0.927 160-0.0519 0.06920.58 ± 5.01
Ethylbenzene µg/l - --0.0383 0.0510.442 ± -
o-Xylene µg/l - --0.513 0.6844.41 ± -
Sum of m-Xylene and p-Xylene µg/l - --0.719 0.9595.69 ± -
Toluene µg/l - --0.0889 0.1260.85 ± -
Methyl-tert-butyl-ether µg/l - --0.361 0.4343.25 ± -
CB01BBTXParameter Target ± Cl(99%) Result ± UUnit Criteria Recovery z-score
Sample:
Benzene µg/l 2.886 140-0.241 0.3412.07 ± 2.4
Ethylbenzene µg/l - --0.201 0.2761.45 ± -
o-Xylene µg/l - --0.363 0.4843.03 ± -
Sum of m-Xylene and p-Xylene µg/l - --1.04 1.388.55 ± -
Toluene µg/l - --0.512 0.7045.33 ± -
Methyl-tert-butyl-ether µg/l - --0.103 0.1240.882 ± -
This report was created with PROLab, a software by QuoData: http://www.quodata.de/
226
Laboratory oriented report Volatile halogenated hydrocarbons (VOC), BTEX and MTBE - CB01
Laboratory: LC0025
z-score-2 0 2 -2 0 2
Sample
CB01 A - BTEX / MTBE CB01 B - BTEX / MTBE
Mea
sura
nd
Benzene 5.0 2.4
227