    EPOCH XT B-Scan Software Option Available on EPOCH XT instruments beginning in April 2008.

    Instrument software revision 1.1.1.1 or later

    EPOCH XT must be equipped with the Hardware I/O Port (optional)in order to allow encoded scans.

  • 7/27/2019 EPOCH XT B-Scan Option Presentation

    3/29

    EPOCH XT B-Scan Key Features Scan Rate has been increased to up to 1000 measurements / second.

    With Floating Gate active the maximum scan rate is 500 meas. / sec. With A-Scan Storage active the maximum scan rate is 60 meas. / sec.

    The limit on the EPOCH 4 B-Scan is 180 meas. / sec.

    Three B-Scan Modes: Bi-Directional encoded, Uni-Directional encoded,and Manual non-encoded.

    Display Modes B-Scan only (split or full screen) A-Scan only (split or full screen) B-Scan and A-Scan simultaneously

    Store up to 150,000 points per B-scan and up to 100 scans at maximumsize.

    Store A-Scans during B-Scan collection based on: Alarm

    Minimum Depth

    Store All

    Limits scan rate to 60 measurements / second

  • 7/27/2019 EPOCH XT B-Scan Option Presentation

    4/29

    EPOCH XT B-Scan Key Features B-Scan is compatible with Echo-to-Echo for coated surfaces or

    when there is a varying water path.

    Gate 1 width must be set wide enough to completely cover Gate 2 to avoid lossof signal.

    Full compatibility with GageView Pro 3.5

    See the GageView Pro 3.5 presentation for the EPOCH XT. This is located in

    the HELP menu in GageView Pro.

    New color palettes for B-Scan

    Scan can change from Green to Red on alarm

    Grayscale gradient and Color gradient available

  • 7/27/2019 EPOCH XT B-Scan Option Presentation

    5/29

    EPOCH XT B-Scan InstrumentConfiguration and Parts

  • 7/27/2019 EPOCH XT B-Scan Option Presentation

    6/29

    Hardware I/O Explanation EPOCH XTs must be equipped with the Hardware I/O Port to accept

    the encoder cable used for B-Scans.

    If you are not sure if your EPOCH XT has this port, look at the topof the instrument and compare to the pictures below.

    Installation of this port requires that the EPOCH XT be returned toan Olympus NDT Service Center.

    OK NOT OK

  • 7/27/2019 EPOCH XT B-Scan Option Presentation

    7/29

    Parts Required for B-Scan EPOCH XT with Hardware I/O Port

    EPOCH XT B-Scan Software Option PN: EPXT-BSCAN

    EPOCH XT Encoder Cable PN: EPXT-ENC-CBL-XX Choose length (10, 25, 50, or 100 feet).

    Encoder Car PN: EP4/ENC

    Transducer Typically a D790-SM

    Transducer Cable BNC: BCMD-316-XXB (Choose length to match encoder cable)

    LEMO: L1CMD-316-XXB (Choose length to match encoder cable)

  • 7/27/2019 EPOCH XT B-Scan Option Presentation

    8/29

    B-Scan Kits B-Scan Kits Most customers choose to purchase B-Scan kits

    that contain the Instrument Software Option, B-Scan EncoderCar, and B-Scan Encoder Cable.

    EPXT-BSCAN-KIT-10: (10 foot encoder cable)

    EPXT-BSCAN-KIT-25: (10 foot encoder cable)

    EPXT-BSCAN-KIT-50: (10 foot encoder cable)

    EPXT-BSCAN-KIT-100: (10 foot encoder cable)

    The transducer and transducer cable must be purchasedseparately.

    GageView Pro must be purchased separately.

  • 7/27/2019 EPOCH XT B-Scan Option Presentation

    9/29

    B-Scan Encoder & Accessories

    D790-SMTransducer Cable

    Extension Pole BracketOption: EP4/EPB

    Magnetic Wheels Option:EP4/MW (1 Pair)

    Encoder CableInput

    Probe Holder:EP4/PH

    Water Input

  • 7/27/2019 EPOCH XT B-Scan Option Presentation

    10/29

    B-Scan Option Installation The EPOCH XT must be updated to Operating Software version

    1.1.1.1 or higher.

    If the B-Scan Option is NOT activated in your instrument, go tothe INSTR SETUP Menu STATUS Tab as shown below.

    Provide Olympus NDT with the last 4 characters of yourinstruments serial number.

    Olympus NDT will provide an Option Key number to activate the

    software option.

  • 7/27/2019 EPOCH XT B-Scan Option Presentation

    11/29

    EPOCH XT B-Scan Setup & Activation

  • 7/27/2019 EPOCH XT B-Scan Option Presentation

    12/29

    B-Scan Option Initial Setup The EPOCH XT must be properly setup and calibrated before the

    B-Scan option is turned ON.

    Pulser / Receiver Settings

    Calibrate for Zero Offset and Velocity

    Gate 1 should be setup to cover the minimum and maximum expectedthickness values for the inspection.

    The screenshots below show a complete setup ready for B-Scan

    to be turned on.

  • 7/27/2019 EPOCH XT B-Scan Option Presentation

    13/29

    B-Scan Option Initial Setup A Minimum Depth Alarm may be used to alert the operator to

    areas where the part thickness has dropped below a userdefined threshold.

    The Floating Gate option may be used to set the height of theEPOCH XTs gates. This option allows the gates to be set at alevel from -1dB to -14dB (1 dB increment) from the height of thegated echo.

    Avoid echo drop out when scanning on heavily corroded parts.

    Maximize measurement consistency by measuring at the same point onthe edge of the echo envelope.

  • 7/27/2019 EPOCH XT B-Scan Option Presentation

    14/29

    B-Scan Option Activation The B-Scan Option is activated in the MEAS SETUP Menu

    OPTIONS Tab B-SCAN Sub Tab.

    There are 3 Encoder Modes available: Manual The encoder is not active. B-Scan updates according to a time rate

    that is set with the PRF control.

    Uni-Directional Encoder is active. Scan moves forward when encodermoves forward or backward.

    Bi-Directional Encoder is active. Scan moves in the same direction asencoder.

  • 7/27/2019 EPOCH XT B-Scan Option Presentation

    15/29

    EPOCH XT B-Scan Bi-Directional Example

  • 7/27/2019 EPOCH XT B-Scan Option Presentation

    16/29

    Bi-Directional B-Scan Example Turn the B-Scan Feature ON

    Choose Encoder Mode = Bi-Directional

    Encoder Pulses: This is an encoder

    specification. The Remote Scanencoder that Olympus NDT sells outputs50 pulses / inch.

    Take Reading Every: This is theresolution of the B-Scan

    Start Coordinate: Set to zero or todesired value.

    Save Ascan Modes: No Storage: No A-Scans saved during scan

    Store on MinDepth: Store A-scans thatcorrespond to the minimum thickness in thescan.

    Store on Alarm: Store A-scans when theinstrument alarm is triggered.

    Store All: Store all A-Scans during the B-Scan.

    Press MEAS to return to Live Screen.

    NOTE: Saving A-Scans during a B-Scancurrently limits the scan rate to 60Hz.

  • 7/27/2019 EPOCH XT B-Scan Option Presentation

    17/29

    Bi-Directional B-Scan Example B-Scan View

    SR = Scan Rate

    Min Depth Alarm

    Bb = B-ScanBi-Directional

    Start New Scan (Old Scan

    is erased)

    Change to dual A & BScan View

    DT = Distance Travelled

    Thickness Limits defined

    by Gate 1 Start and Width

    Start = Start Scan

    Change to A-Scan View

  • 7/27/2019 EPOCH XT B-Scan Option Presentation

    18/29

    Bi-Directional B-Scan Example A-Scan View

    SR = Scan Rate

    Bb = B-ScanBi-Directional

    Start New Scan (Old Scan

    is erased)

    Change to dual A & BScan View

    DT = Distance Travelled

    Minimum Depth Alarm

    Stop = Stop Scan

    Change to B-Scan View

  • 7/27/2019 EPOCH XT B-Scan Option Presentation

    19/29

    Bi-Directional B-Scan Example A & B-Scan View

    SR = Scan Rate

    (only active in encoded

    scans)

    Bb = B-ScanBi-Directional

    Min Depth Alarm

    Start New Scan (Old Scan

    is erased)

    Change to B-Scan View

    DT = Distance Travelled

    (only active in encoded

    scans)

    Min Depth Alarm

    Thickness Limits

    defined by Gate 1 Start

    and Width

    Stop = Stop Scan

  • 7/27/2019 EPOCH XT B-Scan Option Presentation

    20/29

    Bi-Directional B-Scan Example Review

    Min Depth Alarm Triggered

    To review the stored B-Scan you must first stop the scan using the F1 key.

    As the Review Cursor is moved along the B-Scan the EPOCH XT will displaythe Distance Travelled (DT) for the point as well as any saved A-Scans.

    Review Cursor Move

    Left and Right with

    arrow keys to review

    B-Scan thickness data.

    This view shows the A-

    Scan that has been

    stored at this point in

    the B-Scan.

  • 7/27/2019 EPOCH XT B-Scan Option Presentation

    21/29

    EPOCH XT B-Scan Manual Example

  • 7/27/2019 EPOCH XT B-Scan Option Presentation

    22/29

    Manual B-Scan Example Turn the B-Scan Feature ON

    Choose Encoder Mode = Manual

    Direction: The EPOCH XT will allow

    the Manual Scan to run from Left toRight or from Right to Left.

    Stop on Loss Of Signal (LOS): Thiswill cause the scan to stop if there isno signal in the gate. When a signal

    reappears the instrument inserts ablank line in the scan.

    Save Ascan Modes: No Storage: No A-Scans saved during scan

    Store on MinDepth: Store A-scans thatcorrespond to the minimum thickness in thescan.

    Store on Alarm: Store A-scans when theinstrument alarm is triggered.

    Store All: Store all A-Scans during the B-

    Scan. Press MEAS to return to Live Screen.

    NOTE: Saving A-Scans during a B-Scancurrently limits the scan rate to 60Hz.

  • 7/27/2019 EPOCH XT B-Scan Option Presentation

    23/29

    Manual B-Scan Example B-Scan ViewSR = Scan Rate

    (only active in encoded

    scans)

    Min Depth Alarm

    Bm = B-Scan Manual

    Start New Scan (Old Scan

    is erased)

    Change to dual A & B

    Scan View

    DT = Distance Travelled

    (only active in encoded

    scans)

    Thickness Limits defined

    by Gate 1 Start and Width

    Start = Start Scan

    Change to A-Scan View

  • 7/27/2019 EPOCH XT B-Scan Option Presentation

    24/29

    Manual B-Scan Example A-Scan ViewSR = Scan Rate

    (only active in encoded

    scans)

    Bm = B-Scan Manual

    Start New Scan (Old Scan

    is erased)

    Change to dual A & B

    Scan View

    DT = Distance Travelled

    (only active in encoded

    scans)

    Minimum Depth Alarm

    Start = Start Scan

    Change to B-Scan View

  • 7/27/2019 EPOCH XT B-Scan Option Presentation

    25/29

    Manual B-Scan Example A & B-Scan ViewSR = Scan Rate

    (only active in encoded

    scans)

    Bm = B-Scan Manual

    Min Depth Alarm

    Start New Scan (Old Scan

    is erased)

    Change B Scan View

    DT = Distance Travelled

    (only active in encoded

    scans)

    Thickness Limits

    defined by Gate 1 Start

    and Width

    Start = Start Scan

  • 7/27/2019 EPOCH XT B-Scan Option Presentation

    26/29

    Manual B-Scan Example A & B-Scan View

    Min Depth Alarm TriggeredLoss of Signal

    When running a Manual B-Scan the scan rate is controlled with the PRFcontrol.

    Distance Traveled (DT) and Scan Rate (SR) are not active in Manual Scan

    mode. You may choose different measurements in Manual MeasurementSelection mode.

    Review Cursor Move

    Left and Right with

    arrow keys to review

    B-Scan thickness data.

  • 7/27/2019 EPOCH XT B-Scan Option Presentation

    27/29

    EPOCH XT B-Scan Data Logging

  • 7/27/2019 EPOCH XT B-Scan Option Presentation

    28/29

    EPOCH XT B-Scan Saving Files B-Scan files may be stored in any file type in the EPOCH XT. The most

    common are: CAL Files: Used to store B-Scan setups

    INCREMENTAL Files: Used for most inspection data storage

    When the SAVE key is pressed the operator has two choices: SAVE SCREEN: This saves a screen

    shot of the B-Scan along with the

    instrument setup. This can be recalled

    as a setup if desired.

    SAVE FULL: This saves the entire

    B-Scan along with stored A-Scans andthe instrument setup. This can be

    recalled as a setup if desired.

  • 7/27/2019 EPOCH XT B-Scan Option Presentation

    29/29

    EPOCH XT B-Scan File Review When a Full B-Scan has been saved the instrument will display a

    compressed B-Scan view in the Review Screen. To view the data in itsuncompressed format you must do one of the following: Transfer the file to GageView Pro

    Recall the file and review using the review cursor.

    When a B-Scan screenshot has been saved the instrument will display

    an uncompressed view of the B-Scan in the Review Screen.