DNVGL-CP-0485 5 ppm bilge alarms · the sample being supplied to the 5 ppm bilge alarm and the ppm...

26
The electronic pdf version of this document, available free of charge from http://www.dnvgl.com, is the officially binding version. DNV GL AS CLASS PROGRAMME Type approval DNVGL-CP-0485 Edition March 2016 5 ppm bilge alarms

Transcript of DNVGL-CP-0485 5 ppm bilge alarms · the sample being supplied to the 5 ppm bilge alarm and the ppm...

Page 1: DNVGL-CP-0485 5 ppm bilge alarms · the sample being supplied to the 5 ppm bilge alarm and the ppm display showing the correct response, shall not exceed 5 seconds. 1.2.6 The 5 ppm

The electronic pdf version of this document, available free of chargefrom http://www.dnvgl.com, is the officially binding version.

DNV GL AS

CLASS PROGRAMME

Type approval

DNVGL-CP-0485 Edition March 2016

5 ppm bilge alarms

Page 2: DNVGL-CP-0485 5 ppm bilge alarms · the sample being supplied to the 5 ppm bilge alarm and the ppm display showing the correct response, shall not exceed 5 seconds. 1.2.6 The 5 ppm

FOREWORD

DNV GL class programmes contain procedural and technical requirements including acceptancecriteria for obtaining and retaining certificates for objects and organisations related toclassification.

© DNV GL AS March 2016

Any comments may be sent by e-mail to [email protected]

This service document has been prepared based on available knowledge, technology and/or information at the time of issuance of thisdocument. The use of this document by others than DNV GL is at the user's sole risk. DNV GL does not accept any liability or responsibilityfor loss or damages resulting from any use of this document.

Page 3: DNVGL-CP-0485 5 ppm bilge alarms · the sample being supplied to the 5 ppm bilge alarm and the ppm display showing the correct response, shall not exceed 5 seconds. 1.2.6 The 5 ppm

Cha

nges

- c

urre

nt

Class programme — DNVGL-CP-0485. Edition March 2016 Page 35 ppm bilge alarms

DNV GL AS

CHANGES – CURRENT

This is a new document.

Page 4: DNVGL-CP-0485 5 ppm bilge alarms · the sample being supplied to the 5 ppm bilge alarm and the ppm display showing the correct response, shall not exceed 5 seconds. 1.2.6 The 5 ppm

Con

tent

s

Class programme — DNVGL-CP-0485. Edition March 2016 Page 45 ppm bilge alarms

DNV GL AS

CONTENTS

Changes – current.................................................................................................. 3

Section 1 General....................................................................................................51 Introduction.........................................................................................52 Documentation.....................................................................................6

Section 2 Product requirements..............................................................................81 Design requirements............................................................................82 Test requirements................................................................................93 Limitations in the certificate................................................................94 Product marking.................................................................................. 95 Installation requirements to the 5 ppm bilge arrangement..................9

Appendix A Specifications for performance test................................................... 121 General.............................................................................................. 122 Test specifications............................................................................. 12

Appendix B Specifications for environmental testing............................................ 171 General.............................................................................................. 172 Test specifications............................................................................. 17

Appendix C Method for determination of oil content.............................................191 General.............................................................................................. 19

Appendix D Documentation of test data and results............................................. 201 General.............................................................................................. 202 Oil-content analysis........................................................................... 203 Test fluids data..................................................................................204 Test rig and sampling arrangement...................................................215 Display calibration and activation-accuracy tests.............................. 216 Contaminant and colour test............................................................. 217 Sample pressure test.........................................................................228 Shut-off test...................................................................................... 239 Utilities supply variation test.............................................................2310 Calibration and zero-drift test......................................................... 2411 Response time test.......................................................................... 24

Changes – historic................................................................................................25

Page 5: DNVGL-CP-0485 5 ppm bilge alarms · the sample being supplied to the 5 ppm bilge alarm and the ppm display showing the correct response, shall not exceed 5 seconds. 1.2.6 The 5 ppm

Sec

tion

1

Class programme — DNVGL-CP-0485. Edition March 2016 Page 55 ppm bilge alarms

DNV GL AS

SECTION 1 GENERAL

1 Introduction

1.1 ObjectiveThe objective of this class programme (CP) is to give a description of the type approval (TA) of 5 ppm bilgealarms.For a description of the DNV GL type approval scheme in general and further information on generalconditions and procedures for obtaining DNV GL TA certificate, see class programme CP 0338 – DNV GL TypeApproval scheme.The procedures and requirements described in this CP are applicable for obtaining DNV GL TA certificatebased on requirements in:

— DNV GL rules RU SHIP Pt.4 Ch.6— DNV GL rules RU SHIP Pt.6 Ch.7— Resolution MEPC.107(49) “Revised Guidelines and Specifications for Pollution Prevention Equipment for

Machinery Space Bilges of Ships”Note:This class programme is not applicable for obtaining EU Marine Equipment Directive (MED) certificates. Visit www.dnvgl.com forinformation on MED certification.

---e-n-d---o-f---n-o-t-e---

1.2 ScopeThis CP gives a description of the procedures and requirements related to documentation, design and typetesting applicable for TA of 5 ppm bilge alarms.TA is based on compliance with design requirements given in the rules and/or other regulations andstandards. The CP describes the applicable design requirements and how to document compliance with therequirements in order to obtain a TA certificate for the equipment. This includes, where relevant, technicalrequirements for how the type tests shall be performed.

1.3 ApplicationDNV GL rules RU SHIP Pt.6 Ch.7 require that 5 ppm bilge alarms are DNV GL type approved, and TA inaccordance with this CP is thus mandatory for equipment to be installed on DNV GL classed vessels of Clean(Design) notation.A TA certificate in accordance with this CP will confirm compliance with the requirements in the rules asspecified in [1.1]. The TA certificate will not confirm compliance with requirements in other parts of the rules.In case additional requirements in other parts of the rules shall be covered by the TA certificate, this shall bespecified in the application for TA and will be stated in the TA certificate.

1.4 DefinitionsTable 1 Definitions

5 ppm Bilge (water) separator Any combination of a separator, filter, coalesces or other means, and also a single unitdesigned to produce an effluent with oil content not exceeding 5 ppm.

5 ppm Bilge alarm An oil content meter with associated alarm which is capable of detecting and measuring5 ppm or less of oil in a ship’s bilge water overboard discharge.

Page 6: DNVGL-CP-0485 5 ppm bilge alarms · the sample being supplied to the 5 ppm bilge alarm and the ppm display showing the correct response, shall not exceed 5 seconds. 1.2.6 The 5 ppm

Sec

tion

1

Class programme — DNVGL-CP-0485. Edition March 2016 Page 65 ppm bilge alarms

DNV GL AS

Accredited laboratory

— DNV GL's laboratories at Høvik, Norway.— A european laboratory accredited for all the required tests by an Accreditation Body

being member of European Accreditation, EA.— A non-european laboratory accredited by an organisation who has signed a

multilateral agreement (MLA) with EA.— A laboratory having the quality system audited by DNV GL. A quality audit by DNV

GL means that a competent person has gone through the quality system of thelaboratory in accordance with ISO/IEC 17025 (and EN45001) and that a "Statementof Recognition" has been issued.

— A laboratory recognised/certified by the marine administration of his country marineadministration provided acceptance by the responsible approval center in DNV GLbefore testing.

— Any laboratory when testing is witnessed by a qualified surveyor.— The manufacturer's premises providing the test set-up is approved and the testing

is witnessed by a qualified surveyor. Such arrangement shall be approved by theresponsible approval center in DNV GL before testing.

Automatic stopping device

A device used to automatically stop any discharge overboard of oily mixture whenthe oil content of the effluent exceeds 5 ppm. The automatic stopping device shallconsist of a valve arrangement installed in the effluent outlet line of the 5 ppm bilgewater separator which automatically diverts the effluent mixture from being dischargedoverboard back to the ship's bilges or bilge tank when the oil content of the effluentexceeds 5 ppm.

Bilge waterWater which may be contaminated by oil resulting from incidents such as leakage ormaintenance work in machinery spaces. Any liquid entering the bilge system, bilgepiping, tank top or bilge holding tanks is considered bilge water.

ppm Parts of oil per million parts of water by volume

ppm display A numerical scale display of the 5 ppm bilge alarm.

2 DocumentationFor TA of 5 ppm Bilge alarms the following documentation shall be submitted:

— DNV GL application form for TA— Assembly drawings showing main dimensions and equipment arrangement— Mechanical drawings with detailed drawings of all essential mechanical components— Electrical system and instrumentation layout and diagrams and list of components— Alarm list (list of all events that activate the alarm).— Instruction manuals including operation, installation, maintenance and permitted operating parameters— Information regarding the manufacturer’s type designation marking (or other marking) of the product for

unambiguous identification of the product with reference to technical specifications— Specification of materials applied— Performance test plan according to App.A with oil-content analysis according to App.C, including

laboratory accreditation certificate if test and/or sample analysis is not witnessed by DNV GL, andincluding model designation of DNV type-approved 5 ppm bilge alarm to be used during testing, and usingtest forms outlined in App.D

— Environmental test plan according to App.B, including laboratory accreditation certificate— Hardware, firmware and software revision information, as applicable, necessary to identify the equipment

under test

Page 7: DNVGL-CP-0485 5 ppm bilge alarms · the sample being supplied to the 5 ppm bilge alarm and the ppm display showing the correct response, shall not exceed 5 seconds. 1.2.6 The 5 ppm

Sec

tion

1

Class programme — DNVGL-CP-0485. Edition March 2016 Page 75 ppm bilge alarms

DNV GL AS

— Reports with results from performance test App.A with oil-content analysis according to App.C, includingmodel designation of DNV type-approved 5 ppm bilge alarm used during testing, and using test formsoutlined in App.D.

— Reports with results from environmental test (App.B)— Proposed field of application and operational limitations— Drawings of permanent plates giving any operational or installation limits, if any— Hazardous-area (EX) certification documentation as applicable if the equipment is to be certified for

hazardous area— Functional description of all essential parts of the equipment— Asbestos-free declaration— Report from initial assessment

The test plans shall be approved prior to testing.The manufacturer shall provide documentation on all the requirements in this document, not limited to thedocument list above. The documentation submitted shall be clearly labelled and divided in such a way that allelements below can be easily identified. All drawings and descriptions shall be marked with drawing number,item name, revision no. and/or issue date, etc. The documentation shall be generic drawings suitable for atype approval.

Page 8: DNVGL-CP-0485 5 ppm bilge alarms · the sample being supplied to the 5 ppm bilge alarm and the ppm display showing the correct response, shall not exceed 5 seconds. 1.2.6 The 5 ppm

Sec

tion

2

Class programme — DNVGL-CP-0485. Edition March 2016 Page 85 ppm bilge alarms

DNV GL AS

SECTION 2 PRODUCT REQUIREMENTS

1 Design requirements

1.1 GeneralThe production model of the 5 ppm bilge alarm, for which the approval will apply, shall be identical to the 5ppm bilge alarm type tested and environmental tested in accordance with [2].The 5 ppm bilge alarm shall always be activated when the 5 ppm level is not maintained by the bilgeseparator, and the automatic stopping device shall then automatically stop the overboard discharge.The routine maintenance permitted operating parameters and procedures for the 5 ppm bilge alarm shall beclearly defined by the manufacturer in the associated operation, installation and maintenance manual.

1.2 Specific requirements to 5 ppm bilge alarm

1.2.1 The 5 ppm bilge alarm shall be robust and practical instrument suitable for shipboard use. It shallresist corrosion in conditions of the marine environment and shall be compatible with the liquids that it willmonitor.

1.2.2 The 5 ppm bilge alarm shall, if intended to be fitted in locations where flammable atmosphere maybe present, comply with the relevant safety regulations for such spaces. Any electrical equipment which ispart of the 5 ppm bilge alarm shall be placed in a non-hazardous area, or shall be certified as safe for use ina hazardous atmosphere. Any moving parts which are fitted in hazardous areas shall be arranged so as toavoid the formation of static electricity.

1.2.3 The 5 ppm bilge alarm shall not contain or use any substance of a dangerous nature, unless adequatearrangements are provided to eliminate any hazards introduced thereby.

1.2.4 A ppm display shall be provided. The ppm display shall not be affected by emulsions and/or the type ofoil given that the test fluid detailed in App.A [2.1.4] is deemed to represent a mixture that may be expectedin the machinery space bilges of a ship. It shall not be necessary to calibrate the 5 ppm bilge alarm on boardship, but on board testing according to the manufacturer’s instructions shall be permitted. The accuracy ofthe meter shall be such that the reading will represent within +/-2 ppm the actual oil content of the samplebeing tested.

1.2.5 The response time of the 5 ppm bilge alarm, that is, the time which elapses between an alteration inthe sample being supplied to the 5 ppm bilge alarm and the ppm display showing the correct response, shallnot exceed 5 seconds.

1.2.6 The 5 ppm bilge alarm shall be fitted with an electrical/electronic device which shall be pre-set by themanufacturer to activate when the effluent exceeds 5 ppm and shall do so with an accuracy of +/- 2 ppm.Activation of alarm shall also operate automatically if at any time the 5 ppm bilge alarm shall fail to function,require a warm-up period or otherwise be de-energized. The accuracy of a 5 ppm bilge alarm shall remainwithin the above limits despite the presence of contaminants other than oil, and the power supply varying by10% from the design value – i.e. in respect of electricity, compressed air, etc.

1.2.7 It is required that a simple means be provided aboard ship to check on instrument drift, repeatabilityof the instrument reading, and the ability to re-zero the instrument.

1.2.8 The 5 ppm bilge alarm shall record date, time and alarm status, and operating status of the BilgeSeparator. The recording device shall also store data for at least 18 months and shall be able to display orprint a protocol for official inspections as required. In the event the 5 ppm bilge alarm is replaced, meansshall be provided to ensure the data recorded remains available on board for 18 months.

Page 9: DNVGL-CP-0485 5 ppm bilge alarms · the sample being supplied to the 5 ppm bilge alarm and the ppm display showing the correct response, shall not exceed 5 seconds. 1.2.6 The 5 ppm

Sec

tion

2

Class programme — DNVGL-CP-0485. Edition March 2016 Page 95 ppm bilge alarms

DNV GL AS

1.2.9 To avoid wilful manipulation of 5 ppm bilge alarms, the fail-safe arrangements listed in [5] e) shall beincluded.

1.2.10 The accuracy of the 5 ppm bilge alarms shall be checked according to manufacturer recommendation,but not less than at least every 2.5 years at IOPP or class certificate intermediate and renewal surveys.Alternatively the unit may be replaced by a calibrated type approved 5 ppm bilge alarm. The calibrationcertificate for the 5 ppm bilge alarm, certifying date of last calibration check, shall be retained on board forinspection purposes. The accuracy checks can only be done by the manufacturer or persons authorized bythe manufacturer.

1.2.11 The 5 ppm bilge alarm must be designed to operate in each plane that forms an angle of 22.5° withthe plane of its normal operating position, and this shall be verified through the test requirements in App.B[2.2.4].

1.2.12 The 5 ppm bilge alarm shall be tested according to and comply with test requirements of [2].

2 Test requirementsThe 5 ppm bilge alarm shall undergo the following prototype tests:

— Performance test according to App.A with oil-content analysis according to App.C and using the forms asoutlined in App.D

— Environmental test according to App.B.

All test reports shall include the type designation or model identification of the equipment tested, capacity,serial number, as well as hardware, firmware and software revision information, as applicable, necessary toidentify the equipment under test.It is the manufacturer’s responsibility to ensure that the testing is performed according to the requirementsin this type approval programme.

3 Limitations in the certificate

3.1 If the bilge alarm is not EX type approved, the following limitation will be included in the certificate:“Installation of the equipment in spaces subject to explosion hazard is not permitted.”

3.2 If the bilge alarm is not equipped with no-flow sensor or does not activate whenever clean is used forcleaning or zeroing purposes, this should be listed as a limitation in the certificate.

4 Product markingThe final product shall be provided with visible marking, which completely identifies the product and itscomponents according to drawings/equipment specification.The marking shall give at least the following information:

— Identification of manufacturer— Equipment type designation or model identification— Serial number— Revision information, as applicable, for all firmware or software modules installed per hardware unit,

necessary to identify the equipment, permanently marked.

Page 10: DNVGL-CP-0485 5 ppm bilge alarms · the sample being supplied to the 5 ppm bilge alarm and the ppm display showing the correct response, shall not exceed 5 seconds. 1.2.6 The 5 ppm

Sec

tion

2

Class programme — DNVGL-CP-0485. Edition March 2016 Page 105 ppm bilge alarms

DNV GL AS

5 Installation requirements to the 5 ppm bilge arrangement

5.1 For future inspection purposes on board ship, a sampling point shall be provided in a vertical section of thewater effluent piping as close as is practicable to the 5 ppm bilge separator outlet. Re-circulating facilitiesshall be provided, after and adjacent to the overboard outlet of the stopping device to enable the 5 ppm bilgeseparator system, including the 5 ppm bilge alarm and the automatic stopping device, to be tested with theoverboard discharge closed (see Figure 1).The re-circulating facility shall be so configured as to prevent under all operating conditions any by-pass ofthe 5 ppm bilge separator.

5.2 The layout of the installation shall be arranged so that the overall response time (including the response timeof the 5 ppm bilge alarm) between an effluent discharge from the 5 ppm bilge separator exceeding 5 ppm,and the operation of the automatic stopping device preventing overboard discharge, shall be as short aspossible and in any case not more than 20 s.

5.3 The arrangement on board ship for the extraction of samples from the 5 ppm bilge separator discharge lineto the 5 ppm bilge alarm shall give a truly representative sample of the effluent.

5.4 A vessel fitted with a 5 ppm bilge alarm shall, at all times, have on board a copy of the operation, installationand maintenance manual.

5.5 The onboard arrangement of the 5 bilge alarm, 5 ppm bilge separator and automatic stopping device shallfunction automatically. However, fail-safe arrangements to avoid any discharge in case of malfunction shall beprovided. The requirements to automatics and fail safe are:

1) The alarm is always activated whenever clean water is used for cleaning or zeroing purposes.2) The alarm is always activated whenever no flow of sample through the alarm is detected by the flow

sensor.3) Any alarm will activate the automatic stopping device and lead to recirculation.4) The overall response time (including the response time of the alarm) between the effluent discharge

exceeding 5 ppm oil and to the automatic stopping device preventing the overboard discharge is lessthan 20 s.

5) By-passing the alarm during normal operation shall by no means be possible.6) Every access of the alarm (beyond check on instrument drift, repeatability of the instrument reading,

and the ability to re-zero the instrument) requires the breaking of a seal.

Page 11: DNVGL-CP-0485 5 ppm bilge alarms · the sample being supplied to the 5 ppm bilge alarm and the ppm display showing the correct response, shall not exceed 5 seconds. 1.2.6 The 5 ppm

Sec

tion

2

Class programme — DNVGL-CP-0485. Edition March 2016 Page 115 ppm bilge alarms

DNV GL AS

Figure 1 Typical on board arrangement of 5 ppm bilge separator, 5 ppm bilge alarm and automaticstopping device

Page 12: DNVGL-CP-0485 5 ppm bilge alarms · the sample being supplied to the 5 ppm bilge alarm and the ppm display showing the correct response, shall not exceed 5 seconds. 1.2.6 The 5 ppm

App

endi

x A

Class programme — DNVGL-CP-0485. Edition March 2016 Page 125 ppm bilge alarms

DNV GL AS

APPENDIX A SPECIFICATIONS FOR PERFORMANCE TEST

1 GeneralThese performance test specifications relate to 5 ppm bilge alarms. The oil content of the samples shall bedetermined in accordance with App.C. Result forms according to App.D should be used.The 5 ppm bilge alarm being tested shall comply with all the relevant requirements of the technical specificrequirements contained in Sec.2 [1.2].The test reports shall include the type designation or model identification of the equipment tested, serialnumber, as well as hardware, firmware and software revision information, as applicable, necessary to identifythe equipment under test.

2 Test specifications[2.1] gives the requirements to the testing arrangements, i.e. to the test facility, test fluids and grabsampling. The requirements in [2.1] apply to the performance tests. The technical details for theperformance tests are specified in [2.2].

2.1 Specifications of testing arrangements2.1.1 Test facilityA diagrammatic arrangement of a test facility for evaluating the performance of the 5 ppm bilge alarmis given in Figure 1. The accuracy of the 5 ppm bilge alarm will be determined by comparing its readingsagainst a known flow of test fluid injected into a known flow of water. The grab samples taken shall beanalyzed in the accredited laboratory by the methods specified in App.C. The results of the laboratoryanalysis shall be used for correction and to indicate sampling and test equipment variability. The water flowrate shall be adjusted so that the entire test fluid-water flow passes through the 5 ppm bilge alarm, exceptthe intermittent grab sample stream. Special care shall be given to keep, continuously, a constant test fluidcontent in the water that flows into the 5 ppm bilge alarm. The metering pumps shall be adjusted to delivera nearly continuous quantity of test fluid. If test fluid injection becomes intermittent at low concentrations,the test fluid may be pre-mixed with water to provide continuous flow. The test fluid injection point shall beimmediately up-stream of the 5 ppm bilge alarm inlet to minimize time lags.A diagram of the test facility used for the performance testing in [2.2] shall be submitted. (See App.D [3]).

Page 13: DNVGL-CP-0485 5 ppm bilge alarms · the sample being supplied to the 5 ppm bilge alarm and the ppm display showing the correct response, shall not exceed 5 seconds. 1.2.6 The 5 ppm

App

endi

x A

Class programme — DNVGL-CP-0485. Edition March 2016 Page 135 ppm bilge alarms

DNV GL AS

Figure 1 Diagram of test facility

2.1.2 Sampling arrangementThe sampling arrangement for the test rig shall be such that a representative homogeneous sample isobtained under all conditions of operation and under all operational proportions of oil content. The sampleshall be obtained from the full flow through the 5 ppm bilge alarm, but when this is impracticable thesampling arrangements shown in Figure 2 shall be used. Special care shall be given to this stage of theprocess and the validity of the resultant findings.A diagram of the sampling arrangement used for the performance testing in [2.2] should be submitted. (SeeApp.D [3]).

Page 14: DNVGL-CP-0485 5 ppm bilge alarms · the sample being supplied to the 5 ppm bilge alarm and the ppm display showing the correct response, shall not exceed 5 seconds. 1.2.6 The 5 ppm

App

endi

x A

Class programme — DNVGL-CP-0485. Edition March 2016 Page 145 ppm bilge alarms

DNV GL AS

Distance A: not greater than 400 mm

Distance B: sufficient to insert sampling bottle

Dimension C: straight length shall not be less than 60 mm

Dimension D: pipe thickness shall not be greater than 2 mm

Detail E: chisel-edged chamfer (30°C)

Figure 2 Diagram of sampling arrangement

2.1.3 Determination of oil contentThe method for determination of the oil content of the grab samples shall be as specified in App.C. Thesamples are to be extracted on the same day of collection, and they are to be sealed and labelled in thepresence of the accredited laboratory or a qualified surveyor. Arrangements shall be made for analysis assoon as possible, and in any case within seven days, provided the samples are being kept between 2°C and6°C at the accredited laboratories.Form according to App.D [1] should be submitted.

2.1.4 The test fluidsTests shall be performed using three grades of test fluids prepared in full accordance with MEPC.107(49)Annex Part 1 item [1.2.4]. The three test fluids are:

— Test fluid “A” which is a marine residual fuel oil in accordance with ISO 8217, type RMG 35 (density at15°C not less than 980 kg/m3).

— Test fluid “B” which is a marine distillate fuel oil in accordance with ISO 8217, type DMA (density at 15°Cnot less than 830 kg/m3).

— Test fluid “C” which is a mixture of an oil-in-fresh water emulsion in the ratio whereby 1 kg of the mixtureconsists of:

— 947.8 g of fresh water— 25.0 g of test fluid “A”

Page 15: DNVGL-CP-0485 5 ppm bilge alarms · the sample being supplied to the 5 ppm bilge alarm and the ppm display showing the correct response, shall not exceed 5 seconds. 1.2.6 The 5 ppm

App

endi

x A

Class programme — DNVGL-CP-0485. Edition March 2016 Page 155 ppm bilge alarms

DNV GL AS

— 25.0 g of test fluid “B”— 0.5 g surfactant (sodium salt of dodecylbenzene sulfonic acid) in the dry form— 1.7 g “iron oxides” (The term “iron oxide” is used to describe black ferrosoferric oxide (Fe3O4) with

a particle size distribution of which 90% is less than 10 microns, the remainder having a maximumparticle size of 100 microns).

The following data for the test fluids shall be reported:

1) Properties of test fluids A and B (form filled in according to App.D [2]):

— Density at 15°C— Kinematic viscosity (centistokes at 100°C/40°C)— Flashpoint— Ash— Water content.

2) Properties of test fluid C (form filled in according to App.D [2]):

— Type of surfactant— Particle size percentage of the non-soluble suspended solids; and— Surfactant and iron oxide quality verification.

3) Properties of the water in the water tank (form according to App.D [2] should be filled in):

— Density of water at 20°C— Details of any solid matter present.

2.2 Specification of performance tests2.2.1 Calibration and activation accuracy testThe 5 ppm bilge alarm will be calibrated and zeroed as per the manufacturer's instructions. It will then betested with the three test fluids “A”, “B” and “C”, as specified in [2.1.4], at the following oil concentrations inparts per million: 0, 2, 4 and 6 ppm.The following apply for each concentration test on all test fluids: Each concentration test will last for 15 min.Meter display is noted. It shall be noted if the alarm activates or not. Grab samples shall be taken and sentto analysis to determine the actual oil content. After each concentration test, the 5 ppm alarm will be run onclean water for 15 min and the display reading noted.If it proves necessary to re-zero or re-calibrate the 5 ppm bilge alarm during this test, this fact will be noted.Form according to App.D [4] should be filled in. The meter accurancy in ppm is calculated from the differencebetween the meter display reading and the measured oil content in grab samples. The meter accuracy shallbe according to Sec.2 [1.2] d). The activation accuracy shall be according to Sec.2 [1.2] f).

2.2.2 Contaminant and colour testThe 5 ppm bilge alarm shall undergo contaminant and colour tests as follows:

1) The 5 ppm bilge alarm shall be run on a mixture of clean water and 4 ppm test fluid “B”. The alarmreading is noted.

2) The water supply shall be changed from 4 ppm test fluid “B” and clean water to 4 ppm test fluid “B” andwater contaminated with iron oxide in a concentration of 4 ppm.

3) Any shift in the 5-ppm-bilge-alarm reading shall be noted. The reading shall be within the accuracy limitsspecified in Sec.2 [1.2] f).

4) The procedure specified in 2) and 3) above shall be repeated with iron oxide concentrations of 50 ppmand 100 ppm respectively.

5) The 5 ppm bilge alarm shall be run on a mixture of clean water and 4 ppm test fluid “B” and its readingnoted.

Page 16: DNVGL-CP-0485 5 ppm bilge alarms · the sample being supplied to the 5 ppm bilge alarm and the ppm display showing the correct response, shall not exceed 5 seconds. 1.2.6 The 5 ppm

App

endi

x A

Class programme — DNVGL-CP-0485. Edition March 2016 Page 165 ppm bilge alarms

DNV GL AS

6) The water supply shall be changed from clean water to very salt water (a solution of 6% common saltwith clean water).

7) Any shift in the 5 ppm bilge alarm reading shall be noted. The reading shall be within the accuracy limitsspecified in Sec.2 [1.2] f); and

8) Sufficient water shall be available in the mixing tank to ensure an effective test of not less than 15 min.

Form according to App.D [5] should be filled in.

2.2.3 Sample pressure or flow testThe 5 ppm bilge alarm shall be run on 4 ppm test fluid “B” sample. The water pressure or flow rate of themixture shall be adjusted from one half normal, normal and twice normal. Any effect of these changes (zeroshift or calibration shift) on the 5 ppm bilge alarm ppm display reading shall be noted and recorded on thecertificate. This test may require modification for 5 ppm bilge alarms with flow or pressure regulators or 5ppm bilge alarms designed to discharge into an ambient pressure sump.Any zero shift and calibration shift shall be within the accuracy limits specified in Sec.2 [1.2] f).Form according to App.D [6] should be filled in.

2.2.4 Shut off testsThe 5 ppm bilge alarm shall be run on clean water before and after the testing. The 5 ppm bilge alarm shallbe run on 4 ppm test fluid “B” sample. The water and test fluid injection pumps shall be shut off. The 5ppm bilge alarm will be left turned on with no other changes made. After 8 hours, the water and test fluidinjection pump shall be turned on and set to provide the mixture of 4 ppm. The 5 ppm bilge alarm ppmdisplay readings before and after each test, and any damage to the 5 ppm alarm shall be noted and recordedon the certificate.Any zero shift and calibration shift shall be within the accuracy limits specified in Sec.2 [1.2] f). Formaccording to App.D [7] should be filled in.

2.2.5 Utilities supply variation testIf the 5 ppm bilge alarm requires any utilities besides electricity, it shall be tested with these utilities at110% and 90% of the design figures. The 5 ppm bilge alarm shall be run on clean water before and after thetesting. The 5 ppm bilge alarm shall be run on 4 ppm test fluid “B” sample. Any zero shift or calibration shiftto the 5 ppm bilge alarm ppm display reading shall be noted and recorded on the certificate.Any zero shift and calibration shift shall be within the accuracy limits specified in Sec.2 [1.2] f). Formaccording to App.D [8] should be filled in.

2.2.6 Calibration and zero drift testThe 5 ppm bilge alarm shall be calibrated and zeroed. The meter display reading on oil-free water is noted.4 ppm test fluid “B” sample shall run through the 5 ppm bilge alarm for eight hours and any calibration driftnoted. During this test, grab samples shall be taken 0, 2, 4, 6, and 8 hours into the test schedule to verifythe actual oil content and to use for determining the calibration shift and calibration drift. After the test, the 5ppm bilge alarm shall run on oil-free water for 15 minutes, and any zero drift shall be noted and recorded onthe certificate.Any zero drift and calibration shift/drift shall be within the accuracy limits specified in Sec.2 [1.2] f). Formaccording to App.D [9] should be filled in.

2.2.7 Response-time testThe response time shall be taken for the 5 ppm bilge alarm to give an alarm at 5 ppm oil concentration afterthe supply to the 5 ppm bilge alarm is changed from clean water to oily water having a concentration ofmore than 5 ppm oil. Test shall be carried out with test fluids “A”, “B” and “C”. Between each test fluid, thealarm shall run on clean water for 15 minutes. The response-time test is performed in order to verify thecompliance with Sec.2 [1.2] e). The response time, i.e. the time from the flow having more than 5 ppm oilconcentration to the meter display showing the correct response, is accurately measured and noted.Form according to App.D [10] should be filled in.

Page 17: DNVGL-CP-0485 5 ppm bilge alarms · the sample being supplied to the 5 ppm bilge alarm and the ppm display showing the correct response, shall not exceed 5 seconds. 1.2.6 The 5 ppm

App

endi

x B

Class programme — DNVGL-CP-0485. Edition March 2016 Page 175 ppm bilge alarms

DNV GL AS

APPENDIX B SPECIFICATIONS FOR ENVIRONMENTAL TESTING

1 GeneralThe specifications for environmental testing for type approval relate to the electrical and electronic sectionsof 5 ppm bilge alarm.All electrical and electronic components of the 5 ppm separator shall be tested in accordance with thisappendix.Satisfactory compliance with the environmental tests laid down in these guidelines and specifications, whereapplicable, shall be shown on the environmental test protocol issued by the accredited testing laboratory.The protocol shall include at least a statement of the tests conducted on the equipment, including the resultsthereof. The protocol shall be signed and dated by the person in charge of the laboratory.The test report shall include the type designation or model identification of the equipment tested, serialnumber, as well as hardware, firmware and software revision information, as applicable, necessary to identifythe equipment under test.

2 Test specifications

2.1 Testing requirementsThe electrical and electronic sections of the equipment in the standard production configuration shall besubjected to the programme of environmental tests set out in this specification at the accredited laboratory.A copy of the environmental test document, in a format similar to that specified in [1], shall be submitted tothe class society by the manufacturer, together with the application for type approval.

2.2 Test specification detailsEquipment shall operate satisfactorily on completion of each of the following environmental tests:

2.2.1 Vibration tests:

1) A search shall be made for resonance over the following range of frequency and amplitude ofacceleration:

a) 2 to 13.2 Hz with an amplitude of ± 1mmb) 13.2 to 80 Hz with an acceleration of ± 0.7 g. This search shall be made in each of the three planes

at a rate sufficiently low to permit detection of resonance.

2) The equipment shall be vibrated in the planes at each major resonant frequency for a period of 2 hours.

3) If there is no resonant frequency, the equipment shall be vibrated in each of the planes at 30 Hz with anacceleration of ± 0.7 g for a period of 2 hours.

4) After completion of the tests specified in 2) or 3) a search shall again be made for resonance and thereshall be no significant change in the vibration pattern.

2.2.2 Temperature tests (procedure according to IEC60068-2-2 Tests Bb, Bd)

1) Equipment that may be installed in an enclosed space that is environmentally controlled, including anengine-room, shall be subjected, for a period of not less than 2 h, to:

a) a low temperature test at 0°Cb) a high temperature test at 55°C.

2) At the end of each of the tests referred to, the equipment shall be switched on and it shall functionnormally under the test conditions.

Page 18: DNVGL-CP-0485 5 ppm bilge alarms · the sample being supplied to the 5 ppm bilge alarm and the ppm display showing the correct response, shall not exceed 5 seconds. 1.2.6 The 5 ppm

App

endi

x B

Class programme — DNVGL-CP-0485. Edition March 2016 Page 185 ppm bilge alarms

DNV GL AS

2.2.3 Humidity testsEquipment shall be left switched off for a period of 2 h at a temperature of 55°C in an atmosphere with arelative humidity of 90%. At the end of this period the equipment shall be switched on and shall operatesatisfactorily for 1 hour (First the temperature test of the electronic part shall be operated under the testcondition mentioned under [2.2.2], followed by the humidity test.).

2.2.4 Inclination testEquipment shall operate satisfactorily at angles of inclination up to 22.5° in any plane from the normaloperating position.

Page 19: DNVGL-CP-0485 5 ppm bilge alarms · the sample being supplied to the 5 ppm bilge alarm and the ppm display showing the correct response, shall not exceed 5 seconds. 1.2.6 The 5 ppm

App

endi

x C

Class programme — DNVGL-CP-0485. Edition March 2016 Page 195 ppm bilge alarms

DNV GL AS

APPENDIX C METHOD FOR DETERMINATION OF OIL CONTENT

1 GeneralThe international standard ISO 9377-2:2000 “Water quality – Determination of hydrocarbon oil index –Part 2: Method using solvent extraction and gas chromatography” shall be used for the determination of oilcontent requirements outlined in this programme.

Page 20: DNVGL-CP-0485 5 ppm bilge alarms · the sample being supplied to the 5 ppm bilge alarm and the ppm display showing the correct response, shall not exceed 5 seconds. 1.2.6 The 5 ppm

App

endi

x D

Class programme — DNVGL-CP-0485. Edition March 2016 Page 205 ppm bilge alarms

DNV GL AS

APPENDIX D DOCUMENTATION OF TEST DATA AND RESULTS

1 GeneralTest data and results of tests conducted on a 5 ppm bilge alarm in accordance with App.A shall bedocumented on a format as shown below.

2 Oil-content analysisManufacturer of 5 ppm bilge alarm

Test location

Date of test

Samples sealed and labelled by

Method of sample analysis

Samples analysed by

3 Test fluids dataTest fluid “A”

Density at 15°C

Viscosity Centistokes at 100°C

Flashpoint °C

Ash content %

Water content at start of test %

Test fluid “B”

Density at 15°C

Viscosity Centistokes at 40°C

Flashpoint °C

Ash content %

Water content at start of test %

Test fluid “C”

Surfactant – documentary evidence*

Page 21: DNVGL-CP-0485 5 ppm bilge alarms · the sample being supplied to the 5 ppm bilge alarm and the ppm display showing the correct response, shall not exceed 5 seconds. 1.2.6 The 5 ppm

App

endi

x D

Class programme — DNVGL-CP-0485. Edition March 2016 Page 215 ppm bilge alarms

DNV GL AS

Test fluid “C”

Iron oxides – documentary evidence*

* Certificate or laboratory analysis.

Test water

Density at 20°C

Solid matter present

Test temperatures

Ambient °C

Test fluid “A” °C

Test fluid “B” °C

Test fluid “C” °C

Test water °C

4 Test rig and sampling arrangementDiagram of test rig submitted Yes/No

Diagram of sampling arrangement submitted Yes/No

5 Display calibration and activation-accuracy testsTable example is given for test fluid “A”. Identical tables shall be made also for test fluids “B” and “C”.

Test Fluid “A” TimeExpected

oil content(ppm)

Meter displayreading(ppm)

Measured oilcontent in

grab sample(ppm)

Meter displayaccuracy(ppm)

Activation?(Yes/No)

Clean water

2 ppm

Clean water

4 ppm

Clean water

6 ppm

Water temperature [°C]

Re-zero Yes/No and if yes: when

Re-calibrate Yes/No and if yes: when

Page 22: DNVGL-CP-0485 5 ppm bilge alarms · the sample being supplied to the 5 ppm bilge alarm and the ppm display showing the correct response, shall not exceed 5 seconds. 1.2.6 The 5 ppm

App

endi

x D

Class programme — DNVGL-CP-0485. Edition March 2016 Page 225 ppm bilge alarms

DNV GL AS

6 Contaminant and colour test

Fluid TimeExpected oil content

(ppm)Bilge-alarm reading

[ppm]Calibration shift

(ppm)

Clean water

Clean water and 4 ppmtest fluid “B”

4 ppm test fluid “B” andiron oxide 4 ppm

4 ppm test fluid “B” andiron oxide 50 ppm

4 ppm test fluid “B” andiron oxide 100 ppm

4 ppm test fluid “B” andclean water

4 ppm test fluid “B” andvery salt water (6%)

Total testing time: _____________ min

7 Sample pressure test

Fluid Pressure [bar]or flow [l/min]

Expectedoil content

(ppm)

alarm displayreading[ppm]

Zero shift(ppm)

Calibrationshift

(ppm)

Clean water, normalpressure/flow

4 ppm test fluid “B”, half ofnormal pressure/flow:

4 ppm test fluid “B”, normalpressure/flow:

4 ppm test fluid “B”, twice ofnormal pressure/flow

Clean water, normalpressure/flow

Deviations and othercomments:

Page 23: DNVGL-CP-0485 5 ppm bilge alarms · the sample being supplied to the 5 ppm bilge alarm and the ppm display showing the correct response, shall not exceed 5 seconds. 1.2.6 The 5 ppm

App

endi

x D

Class programme — DNVGL-CP-0485. Edition March 2016 Page 235 ppm bilge alarms

DNV GL AS

8 Shut-off test

Fluid TimeExpected

oil content(ppm)

Alarm displayreading(ppm)

Zero shift(ppm)

Calibration shift(ppm)

Clean water

4 ppm test fluid “B”, before shutoff

Shut off, minimum 8 hours

4 ppm test fluid “B”, minimum 8hours after shut off

Clean water

Deviations or damage to bilgealarm:

9 Utilities supply variation test

Utility supplyExpected oil content

(ppm)

Alarm displayreading(ppm)

Zero shift(ppm)

Calibration shift(ppm)

Clean water

110% voltage

90% voltage

110% air pressure

90% air pressure

110% hydraulic pressure

90% hydraulic pressure

Clean water

Deviations and other comments:

Page 24: DNVGL-CP-0485 5 ppm bilge alarms · the sample being supplied to the 5 ppm bilge alarm and the ppm display showing the correct response, shall not exceed 5 seconds. 1.2.6 The 5 ppm

App

endi

x D

Class programme — DNVGL-CP-0485. Edition March 2016 Page 245 ppm bilge alarms

DNV GL AS

10 Calibration and zero-drift test

Fluid TimeExpected

oil content(ppm)

Alarmdisplayreading[ppm]

Measured oilcontent (grab

samples)(ppm)

Calibrationshift

(ppm)

Calibrationdrift

(ppm)

Zero drift(ppm)

Clean water

4 ppm test fluid “B”after 0 hours

4 ppm test fluid “B”after 2 hours

4 ppm test fluid “B”after 4 hours

4 ppm test fluid “B”after 6 hours

4 ppm test fluid “B”after 8 hours

Clean water

11 Response time test

Medium TimeExpected oil content

(ppm)Meter display reading

(ppm)Response time

[s]

Clean water

Test fluid “A”

Clean water

Test fluid “B”

Clean water

Test fluid “C”

Page 25: DNVGL-CP-0485 5 ppm bilge alarms · the sample being supplied to the 5 ppm bilge alarm and the ppm display showing the correct response, shall not exceed 5 seconds. 1.2.6 The 5 ppm

Cha

nges

– h

isto

ric

Class programme — DNVGL-CP-0485. Edition March 2016 Page 255 ppm bilge alarms

DNV GL AS

CHANGES – HISTORICThere are currently no historical changes for this document.

Page 26: DNVGL-CP-0485 5 ppm bilge alarms · the sample being supplied to the 5 ppm bilge alarm and the ppm display showing the correct response, shall not exceed 5 seconds. 1.2.6 The 5 ppm

DNV GLDriven by our purpose of safeguarding life, property and the environment, DNV GL enablesorganizations to advance the safety and sustainability of their business. We provide classification andtechnical assurance along with software and independent expert advisory services to the maritime,oil and gas, and energy industries. We also provide certification services to customers across a widerange of industries. Operating in more than 100 countries, our 16 000 professionals are dedicated tohelping our customers make the world safer, smarter and greener.

SAFER, SMARTER, GREENER