Development Status of a Stable BPM System for the SPring-8...

1
Readout Electronics • We have two candidates for the BPM electronics: Original design based on MTCA.4 [6] and the new generation of Libera Brilliance+ [7] • Most of the modules for the original design are same as the new LLRF system [8]. • Calibration tone will be used for the calibration of the original BPM electronics. • Single-pass and COD resolutions of Libera Brilliance+ were confirmed to be sufficient. • MADOCA [9] support is needed for the new generation of Libera Brilliance+. • Button diameter: 7 mm (Gap: 0.5 mm ) • Button material: Molybdenum - Non-magnetic material - High conductivity for small heat generation due to ohmic losses of trapped modes. • Machining accuracy: a few 10 µm → Error on the electric center with respect to the reference plane < 50 µm rms • Heat input: 1.1 W maximum (100 mA) for the whole BPM head [5] Prototype • 24 prototype electrodes were produced. All the electrodes passed a heat cycle test (150 ˚C) without any vacuum leaks. Some prototype BPM blocks by using prototype electrodes were also produced. Button electrodes were bonded to the block by electron beam welding (EBW). - Electrode was pulled out by a few 10 µm after EBW due to the shrinkage of welded part. - Height of the electrode was adjusted at the machining stage by considering the shrinkage. - Height of the electrode surface was controlled within 50 µm. • We decided not to plate copper in the beam duct. - To reduce the dimension error. - Resistive wall impedance is sufficiently small thanks to the compactness. » Other vacuum chambers has inner copper plating. SPring-8 Upgrade Project [1] Lace: 5-bend Achromat (5BA) Natural Emittance: 140 pm rad (bare lattice) 100 pm rad (with radiation damping of IDs) • X-ray Brilliance: more than 1 orders of magnitude higher than the present SPring-8 Breakthroughs of X-ray optical systems, such as Direct nano-focusing scheme [2]. DEVELOPMENT STATUS OF A STABLE BPM SYSTEM FOR THE SPRING-8 UPGRADE H. Maesaka 1 , H. Dewa 2 , T. Fujita 2 , M. Masaki 2 , S. Takano 1, 2 1 RIKEN SPring-8 Center 2 Japan Synchrotron Radiation Research Institute (JASRI) Requirements for the BPM System BPM Head [5] • Direct nano-focusing scheme demands that the photon axis stability should be well within the intrinsic photon divergence [3]. → Electron beam stability: sub-µm for position and sub-µrad for direction • COD resolution is sufficient for this demand. - COD stability requirement of 5 µm is challenging, although it is not sufficient for the direct nano-focusing beamline. - Present BPM system of SPring-8 shows some drifts of BPM data more than 10 µm [4]. • Beam position is required to be adjusted to the field center of multipole magnets within a few 10 µm. Precise beam-based alignment (BBA) with 10 µm accuracy is needed. • Single-pass resolution of 100 µm rms (100 pC) is required for first-turn steering. • BPM electric center must be aligned to an adjacent quadrupole magnets with 100 µm rms (±200 µm maximum) accuracy to reduce an unwanted kick. Summary • We are developing a stable and precise BPM system for the low-emittance upgrade of SPring-8 . • Requirements for the BPM are 5 µm long-term stability and 0.1 µm resolution for COD and high single-pass resolution and accuracy of 100 µm rms. • Precise BPM electrode and block were designed with the error on the electric center < 50 µm . • Some prototypes of the BPM head with suffi cient machining accuracy were produced. • Candidates for the readout electronics are a MTCA.4-based system and new Libera Brilliance+ . • Radiation-hard cable and radiation shield are under study to reduce humidity-dependent drift . • Beam-based alignment technique has been tested for precise orbit tuning within 10 µm rms . • Beam test at the present SPring-8 will be performed. COD Resolution 0.1 µm rms (100 mA, 1 kHz BW) COD Accuracy after BBA 10 µm rms COD Stability 5 µm max. (1 month) Single-pass Resolution 100 µm rms (100 pC single-bunch) Single-pass Accuracy before BBA 100 µm rms (±200 µm max.) References [1] SPring-8-II Conceptual Design Report, Nov. 2014, http://rsc.riken.jp/pdf/SPring-8-II.pdf [2] C. T. Schroer and G. H. Falkenberg, Journal of Synchrotron Radiaon, vol. 21, Pt. 5, pp. 996-1005, 2014. [3] H. Maesaka, “Comparison of Beam Diagnoscs for 3rd and 4th Generaon Ring-based Light Sources”, Proc. IPAC’15, THYC1, pp. 3657-3661. [4] T. Fujita, et al., “Long-term Stability of the Beam Position Monitors at SPring-8”, Proc. IBIC’15, TUPB020, pp. 359-363. [5] M. Masaki, et al., “Design Opmizaon of Buon-type BPM Electrode for the SPring-8 Upgrade”, Proc. IBIC’16, TUPG18, this conference. [6] PICMG MicroTCA open standard, https://www.picmg.org/openstandards/microtca/ [7] Instrumentaon Technologies, http://www.i-tech.si [8] H. Ego, et al., “RF System of the SPring-8 Upgrade Project”, Proc. IPAC’16, MOPMW009, pp. 414-416. [9] R. Tanaka, et al., “Control System of the SPring-8 Storage Ring”, Proc. ICALEPCS’95, W2B-c, p. 201. TUPG06 Mo φ7mm Ceramic Stainless Steel Reverse- Polarity SMA Jack BPM Signals Digitizer Module RF Frontend ... ... ... ... ... ADC ADC ADC ADC Cal. Tone FPGA BPF Level Adj. Balun 16 12 20 Cooling water channel Prototype of button electrode Prototype of BPM head Signal Cable • One of the most significant causes of the drift of the present BPM system : Radiation damage to signal cables Humidity-dependent drift • Balance error is defined to estimate the drift. - Beam position can be calculated from 3 electrodes out of 4. - 4 beam position values are obtained from 4 combinations of 3 electrodes. - Balance error is defined as the maximum difference among the 4 values. • Humidity dependence of the charactoristic impedance was confirmed by chemical analyses. • Replacement of the damaged cables Humidity-dependent drift disappeared • Radiation-resistant coaxial cables and radiation shields are under study. Support and Alignment • BPM head is supported from the same girder as magnets and vacuum chambers. - Relative position between the BPM and magnets is not changed after a realignment of the girder. • BPM head position is measured by a laser tracker survay within 50 µm rms accuracy. • Total error on the BPM electric center: 70 µm rms ( 50 2 +50 2 ) - Remained margin in the alignment errror is reserved for the calibration error of the electronics etc. Acknowledgment The authors acknowledge Hiroshi Sumitomo from SPring-8 Service Co., Ltd. for his great efforts in the development of the BBA tool and the data analysis. 0.0e+00 5.0e-05 1.0e-04 1.5e-04 2.0e-04 2.5e-04 3.0e-04 3.5e-04 -0.2 0 0.2 0.4 0.6 0.8 Horizontal 0.1mm bpm_05_3(bpm_27) offset 0.180560 [mm] Variance of COD [mm 2 ] (rms 2 ) Horizontal position at BPM_05_3 [mm] dk=+1.28[%] dk=-1.28[%] 1.04e-03 (x -0.180560) 2 +3.0e-06 Beam Test • Beam test setup of the new BPM system in the present SPring-8 has just installed. • Cables and readout electronics for the up-graded SPring-8 are also evaluated in this test. Signal intensity, waveform, temperature rise, long-term stability will be evaluated. Beam-based Alignment • Beam position was shifted by generating a local bump orbit. • Field strength of an adjacent quadrupole magnet was changed by 1% • Reproducibility of the result was 10 µm rms. • We will confirm the measurement accuracy and speed up the measurement procedure. 0 20 40 60 80 100 2015 04/09 2015 04/23 2015 05/07 2015 05/21 2015 06/04 2015 06/18 2015 07/02 -10 -5 0 5 10 A E D C C H H A E H balance error (μm) relative temperature x 10, dewpoint-10 (°C) C08-1 (#043) Balance Error Elec. Dew Point Elec. Temp. Tunnel Dew Point 0 20 40 60 80 100 2015 09/10 2015 09/24 2015 10/08 2015 10/22 2015 11/05 2015 11/19 2015 12/03 -10 -5 0 5 10 tuning H C F C A F A C balance error (μm) relative temperature x 10, dewpoint-10 (°C) C08-1 (#043) Balance Error Elec. Dew Point Elec. Temp. Tunnel Dew Point Before cable replacement After cable replacement

Transcript of Development Status of a Stable BPM System for the SPring-8...

Page 1: Development Status of a Stable BPM System for the SPring-8 ...accelconf.web.cern.ch/AccelConf/ibic2016/posters/tupg06_poster.pdf · • Button electrodes were bonded to the block

Readout Electronics• We have two candidates for the BPM electronics:

Original design based on MTCA.4 [6] and the new generation of Libera Brilliance+ [7]• Most of the modules for the original design are same as the new LLRF system [8].• Calibration tone will be used for the calibration of the original BPM electronics.• Single-pass and COD resolutions of Libera Brilliance+ were confirmed to be sufficient.• MADOCA [9] support is needed for the new generation of Libera Brilliance+.

• Button diameter: 7 mm (Gap: 0.5 mm)• Button material: Molybdenum

- Non-magnetic material - High conductivity for small heat generation due to ohmic losses of trapped modes.

• Machining accuracy: a few 10 µm → Error on the electric center with respect to the reference plane < 50 µm rms

• Heat input: 1.1 W maximum (100 mA) for the whole BPM head [5]Prototype

• 24 prototype electrodes were produced.• All the electrodes passed a heat cycle test (150 ˚C) without any vacuum leaks.• Some prototype BPM blocks by using prototype electrodes were also produced.• Button electrodes were bonded to the block by electron beam welding (EBW).

- Electrode was pulled out by a few 10 µm after EBW due to the shrinkage of welded part. - Height of the electrode was adjusted at the machining stage by considering the shrinkage. - Height of the electrode surface was controlled within 50 µm.

• We decided not to plate copper in the beam duct. - To reduce the dimension error. - Resistive wall impedance is sufficiently small thanks to the compactness.

»Other vacuum chambers has inner copper plating.

SPring-8 Upgrade Project [1]• Lattice: 5-bend Achromat (5BA)• Natural Emittance: 140 pm rad (bare lattice)

100 pm rad (with radiation damping of IDs)• X-ray Brilliance: more than 1 orders of magnitude higher than the present SPring-8• Breakthroughs of X-ray optical systems, such as Direct nano-focusing scheme [2].

DEVELOPMENT STATUS OF A STABLE BPM SYSTEM FOR THE SPRING-8 UPGRADE

H. Maesaka 1, H. Dewa 2, T. Fujita 2, M. Masaki 2, S. Takano 1, 2 1 RIKEN SPring-8 Center

2 Japan Synchrotron Radiation Research Institute (JASRI)

Requirements for the BPM System

BPM Head [5]

• Direct nano-focusing scheme demands that the photon axis stability should be well within the intrinsic photon divergence [3]. → Electron beam stability: sub-µm for position and sub-µrad for direction

• COD resolution is sufficient for this demand. - COD stability requirement of 5 µm is challenging, although it is not sufficient for the direct nano-focusing beamline.

- Present BPM system of SPring-8 shows some drifts of BPM data more than 10 µm [4].• Beam position is required to be adjusted to the field center of multipole magnets

within a few 10 µm. → Precise beam-based alignment (BBA) with 10 µm accuracy is needed.

• Single-pass resolution of 100 µm rms (100 pC) is required for first-turn steering.• BPM electric center must be aligned to an adjacent quadrupole magnets with

100 µm rms (±200 µm maximum) accuracy to reduce an unwanted kick.

Summary• We are developing a stable and precise BPM system for the low-emittance upgrade of SPring-8.• Requirements for the BPM are 5 µm long-term stability and 0.1 µm resolution for COD

and high single-pass resolution and accuracy of 100 µm rms.• Precise BPM electrode and block were designed with the error on the electric center < 50 µm.• Some prototypes of the BPM head with suffi cient machining accuracy were produced.• Candidates for the readout electronics are a MTCA.4-based system and new Libera Brilliance+.• Radiation-hard cable and radiation shield are under study to reduce humidity-dependent drift.• Beam-based alignment technique has been tested for precise orbit tuning within 10 µm rms.• Beam test at the present SPring-8 will be performed.

COD Resolution 0.1 µm rms (100 mA, 1 kHz BW)COD Accuracy after BBA 10 µm rmsCOD Stability 5 µm max. (1 month)Single-pass Resolution 100 µm rms (100 pC single-bunch)Single-pass Accuracy before BBA 100 µm rms (±200 µm max.)

References[1] SPring-8-II Conceptual Design Report, Nov. 2014, http://rsc.riken.jp/pdf/SPring-8-II.pdf[2] C. T. Schroer and G. H. Falkenberg, Journal of Synchrotron Radiation, vol. 21, Pt. 5, pp. 996-1005, 2014.[3] H. Maesaka, “Comparison of Beam Diagnostics for 3rd and 4th Generation Ring-based Light Sources”, Proc. IPAC’15, THYC1, pp. 3657-3661.[4] T. Fujita, et al., “Long-term Stability of the Beam Position Monitors at SPring-8”, Proc. IBIC’15, TUPB020, pp. 359-363.[5] M. Masaki, et al., “Design Optimization of Button-type BPM Electrode for the SPring-8 Upgrade”, Proc. IBIC’16, TUPG18, this conference.[6] PICMG MicroTCA open standard, https://www.picmg.org/openstandards/microtca/[7] Instrumentation Technologies, http://www.i-tech.si[8] H. Ego, et al., “RF System of the SPring-8 Upgrade Project”, Proc. IPAC’16, MOPMW009, pp. 414-416.[9] R. Tanaka, et al., “Control System of the SPring-8 Storage Ring”, Proc. ICALEPCS’95, W2B-c, p. 201.

TUPG06

Mo φ7mm

Ceramic

StainlessSteel

Reverse-PolaritySMA Jack

BPM

Sig

nals

Digitizer ModuleRF Frontend

...

...

......

...ADCADCADC

ADCCal. Tone

FPGABPF Level Adj. Balun

16

12

20

Cooling water channel

Prototype of button electrode Prototype of BPM head

Signal Cable• One of the most significant causes of the drift of the present BPM system:

Radiation damage to signal cables → Humidity-dependent drift• Balance error is defined to estimate the drift.

- Beam position can be calculated from 3 electrodes out of 4. - 4 beam position values are obtained from 4 combinations of 3 electrodes. - Balance error is defined as the maximum difference among the 4 values.

• Humidity dependence of the charactoristic impedance was confirmed by chemical analyses.• Replacement of the damaged cables → Humidity-dependent drift disappeared• Radiation-resistant coaxial cables and radiation shields are under study.

Support and Alignment• BPM head is supported from the same girder as magnets and vacuum chambers.

- Relative position between the BPM and magnets is not changed after a realignment of the girder.• BPM head position is measured by a laser tracker survay within 50 µm rms accuracy.• Total error on the BPM electric center: 70 µm rms (√502+502)

- Remained margin in the alignment errror is reserved for the calibration error of the electronics etc.

Acknowledgment• The authors acknowledge Hiroshi Sumitomo from SPring-8 Service Co., Ltd. for his

great efforts in the development of the BBA tool and the data analysis.

0.0e+00

5.0e-05

1.0e-04

1.5e-04

2.0e-04

2.5e-04

3.0e-04

3.5e-04

-0.2 0 0.2 0.4 0.6 0.8

Horizontal

0.1mm

bpm_05_3(bpm_27)offset 0.180560 [mm]

Var

ianc

e of

CO

D [m

m2 ] (rm

s2 )

Horizontal position at BPM_05_3 [mm]

dk=+1.28[%]dk=-1.28[%]

1.04e-03 (x -0.180560)2+3.0e-06

Beam Test• Beam test setup of the new BPM system in the present SPring-8 has just installed.• Cables and readout electronics for the up-graded SPring-8 are also evaluated in this test.• Signal intensity, waveform, temperature rise, long-term stability will be evaluated.

Beam-based Alignment• Beam position was shifted by generating a local bump orbit.• Field strength of an adjacent quadrupole magnet was changed by 1%• Reproducibility of the result was 10 µm rms.• We will confirm the measurement accuracy and speed up the measurement procedure.

0

20

40

60

80

100

201504/09

201504/23

201505/07

201505/21

201506/04

201506/18

201507/02

-10

-5

0

5

10A E D C C H H A E H

bala

nce

erro

r (μm

)

rela

tive

tem

pera

ture

x 1

0,de

wpo

int-1

0 (°

C)

C08-1 (#043)

Balance Error

Elec. Dew Point

Elec. Temp.

Tunnel Dew Point

0

20

40

60

80

100

201509/10

201509/24

201510/08

201510/22

201511/05

201511/19

201512/03

-10

-5

0

5

10tuning H C F C A F A C

bala

nce

erro

r (μm

)

rela

tive

tem

pera

ture

x 1

0,de

wpo

int-1

0 (°

C)

C08-1 (#043)

Balance Error

Elec. Dew Point

Elec. Temp.

Tunnel Dew Point

Before cable replacement After cable replacement