Determination of semiconductor resistance under a contact
Transcript of Determination of semiconductor resistance under a contact
Determination of semiconductor resistance under a contact
M. Ahmad a,b,*, A.P. Shah c, D.K. Sharma b, N.R. Roy a, B.M. Arora c
a Department of Physics, Ranchi University, Ranchi 834008, Indiab Department of Electrical Engineering, B-61, IIT, Powai, Mumbai 400076, India
c Tata Institute of Fundamental Research, Colaba, Bombay 400005, India
Received 28 February 2001; received in revised form 28 March 2001; accepted 20 August 2001
Abstract
In this work the expression for end resistance has been modified for the case of finite metal resistance. It has been
shown that using this approach the contact resistivity for alloyed contacts can be determined even in case of small
transfer length and large contact length. This is contrary to the case where metal resistance is assumed to be zero.
The changes required for different probe positions have also been considered. � 2002 Elsevier Science Ltd. All rights
reserved.
1. Introduction
Contacts to thin semiconductor layers on non-con-
ducting substrates are generally evaluated by using
either a transmission line method (TLM) or Kelvin re-
sistor method. Despite the advantages of the Kelvin
resistor method, the TLM [1] is still widely used [2] for
evaluating contacts to III–V compound semiconductors.
This is because it is difficult to fabricate a passivation
layer on III–V compound semiconductors, which is re-
quired in the Kelvin resistor method. In comparison,
passivation is a common part of Si technology. Apart
from the classical TLM method, another technique
called trilayer TLM has been reported in the literature
[3] for separately determining contact resistivities of the
metal-alloyed layer interface and the alloyed layer-
semiconductor interface. Although the trilayer TLM
provides this important information about the inter-
faces, the overall contact resistivity as seen by the device
is still given by the standard TLM. In short, the TLM
remains a popular method for quick evaluation of the
contact resistivity [4–6]. In the analysis of measurements
made by using TLM, some assumptions are usually
made: (i) metal resistance Rsm is zero, (ii) the sheet re-
sistance of the semiconductor Rsc under the contact
is equal to the semiconductor sheet resistance Rss be-tween the contacts. Neither of these assumptions is
correct in general. It is known that Rsc can be estimatedexperimentally by measuring the end resistance Re asshown schematically in Fig. 1 by passing a constant
current i0 between two contact pads and measuring thevoltage Ve between one of these pads and a third adja-cent pad, Re ¼ Ve=i0 [7]. In this paper we provide an
expression for the end resistance, which is an improve-
ment compared to that of Refs. [7–9], taking into ac-
count the finite value of the metal resistance. Since the
end resistance and the contact resistance measurements
can depend on the placement of the electrodes, partic-
ularly in the case when the metal film has significant
resistance, we provide expressions for the effect of shift
in the position of the current probe and the voltage
probe on these measurements. Note that the effect of
shift of the voltage probe position alone has been con-
sidered earlier in Refs. [10,11]. Next, we show that the
semiconductor sheet resistance Rsc can be evaluated byan alternative method in which the contact resistance
measurements are made on rows of TLM patterns
of varying width and length. Finally we give a few
experimental example cases to illustrate the use of end
resistance measurement in determination of contact
resistivity.
Solid-State Electronics 46 (2002) 505–512
*Corresponding author. Present address: Department of
Electrical Engineering, B-61, IIT, Powai, Mumbai 400076,
India.
E-mail address: [email protected] (M. Ahmad).
0038-1101/02/$ - see front matter � 2002 Elsevier Science Ltd. All rights reserved.
PII: S0038-1101 (01 )00286-6
2. Theory
2.1. End resistance method
2.1.1. Case of zero metal resistance (Rsm ¼ 0)
In the standard TLM, a number of contact pads of
rectangular geometry are fabricated on the semicon-
ductor with variable spacing L between the pads shown
schematically in Fig. 1(a). In this case the total resistance
R measured between any two contact pads (Fig. 2(a)) is
twice the contact resistance Rc, plus the resistance of thesemiconductor layer [1].
R ¼ 2Rc þ RssLW
ð1aÞ
Under the assumption of zero metal resistance, the
contact resistance Rc is given by
Rc ¼RscLtW
cothdLt
ð1bÞ
where Lt is the transfer length defined as Lt ¼ ðqc=RscÞ1=2,
qc is the contact resistivity, Rsc and Rss are the semi-conductor sheet resistance values below the contact and
between the pads respectively, W and d are the width
Nomenclature
qc contact resistivity
W , d contact width and contact length, respec-
tively
Rc contact resistance
Re end resistance
Lt transfer length
a modified transfer length
Rsm sheet resistance of the contact metal
Rsc sheet resistance of the semiconductor under
the contact
Rss sheet resistance of the semiconductor out-
side the contact
Dx shift of the current probe from the contact
edge
Dx1 shift of the voltage probe from the current
probe
d1 contact length minus Dxq0c apparent contact resistivity obtained with-
out end resistance correction
Fig. 1. (a) Experimental arrangement for end resistance measurement; (b) double transmission line model for rectangular geometry,
where x ¼ 0 is the current probe position. The lower bank of resistors relate to the sheet resistance of the semiconductor under the
metal ððRsc dxÞ=W Þ and away from the metal ððRss=W ÞLÞ. The top bank of resistors relate to the sheet resistance of the metal
ððRsm dxÞ=W Þ. The connecting resistors qc= ðW dxÞ denote the interface resistance between the metal and the semiconductor. End re-sistance is measured by the probe arrangement shown.
506 M. Ahmad et al. / Solid-State Electronics 46 (2002) 505–512
and the length of the contact pad and L is the spacing
between the contact pads. In Eq. (1a) R and the geo-
metrical factors are measured quantities and Rss is aknown material parameter. Therefore Rc can be foundby curve fitting the data of R vs. contact spacing L. The
transfer length Lt can be found from Eq. (1b) provided
Rsc is known, which in turn leads to the determination ofthe specific contact resistivity qc from
qc ¼ RscL2t ð2Þ
Since Rsc is generally not known, an extra measure-ment of the end resistance Re can be used to separatelyfind Lt and Rsc. For Rsm ¼ 0, the expression for the end
resistance [7] is given by
Re ¼LtRsc
W sinh dLt
¼ Rccosh d
Lt
ð3Þ
Various parameters in this expression have been de-
fined above. We can see that the value of Re is measur-able reliably only if the contact length is of the order of
the transfer length. For example, consider the case of
transmission line measurement in which the contact
length d is equal to 3Lt. In this case the value of endresistance will be about 10% of the contact resistance.
2.1.2. Case of non-zero metal resistance (Rsm 6¼ 0)
The case of the contact resistance for Rsm 6¼ 0 has
been treated earlier in Refs. [12,13]. Following the
method of Marlow and Das [12] the double transmission
line equivalent circuit for the contacts with non-zero Rsmis shown in Fig. 1(b). Using the notations given in Sec-
tion 2.1.1, the expression for the contact resistance Rc forcontacts with length d P 4a is given in a symmetrical
form as follows:
Rc ¼RscRsmd þ ðR2sc þ R2smÞa
ðRsc þ RsmÞWð4aÞ
where,
a ¼ffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi
qcRsc þ Rsm
rð4bÞ
a is the modified transfer length and reduces to Lt forRsm¼ 0. The above relation has also been derived in Ref.[13] without the above restriction on the contact length
in which case one obtains
Rc ¼1
ðRsc þ RsmÞWðR2sc
"þ R2smÞa coth
da
þ RscRsm d
(þ 2asinh d
a
)#ð5Þ
The corresponding end resistance is given by
Re ¼aRsc
W sinh da
þ aRsmW
cothda
ð6Þ
This expression can be obtained from Eq. (8), derived by
us later in Section 2.2. In this case the current i2 reducesto i0ðRsc þ eRsmÞ=ðeðRsc þ RsmÞÞ at x ¼ a for d � a as
compared to i0=e at the transfer length for Rsm¼ 0.Now, one can have an idea of the effect of the metal
sheet resistance Rsm on the values of Re. (i) According toEq. (3) with Rsm¼ 0 the end resistance is extremely smallfor large d. It decreases to zero exponentially with in-creasing d. (ii) According to Eq. (6) however, with Rsm6¼ 0 we see that a significant difference occurs. A finite
value of end resistance, larger by orders of magnitude, is
found even in the case of (a) small transfer lengths, (b)
small Rsm and (c) large d. For example, consider a typicalcase of d, W ¼ 50 lm, Rsc¼ 22 X, qc ¼ 1:5� 106
X cm2. Re ffi 2� 109 X for Rsm¼ 0; while Re ffi 0:5 mXfor Rsm ¼ 0.01 X, and 5 mX for Rsm¼ 0.1 X. Corre-sponding Rc ffi 1 X in all the three cases. (iii) According
to Eq. (6) the end resistance decreases exponentially with
increasing value of d, but it reaches a finite asymptoticvalue of Rsma=W for large d. From the ratio of Eqs. (4a)
and (6), for small Rsm and large d, Re=Rc ¼ Rsm=Rsc.
2.2. Probe position corrections
In Eqs. (4a) and (5) the current and voltage probes
are assumed to be at the extreme ends as shown in Fig.
2(a). Correction due to shift in the voltage probe posi-
tion from the pad end has been reported earlier [10] for
d P 4a contact geometry, but there is no mention of theeffect of shift in the current probe position. In practice,
neither of the probes are positioned at exact edges of the
pads as shown in Fig. 2(a) or 3(a). Probe position related
corrections to the above equations for Rc and Re becomeparticularly important in the case of finite metal resis-
tance. These are considered next.
Fig. 2. Probe positions in contact resistance measurement.
M. Ahmad et al. / Solid-State Electronics 46 (2002) 505–512 507
2.2.1. Current and voltage probes shifted from the contact
edge
Consider that both current and voltage probes are
shifted from the pad edge by the same amount Dx asshown by the dashed line in Fig. 2(b). For this case, we
have derived the expressions for Rc and Re using theboundary conditions i1ðd1Þ ¼ 0, i2ðd1Þ ¼ i0, i2ð0Þ ¼ i02ð0Þ,and
d
dxi2ðxÞ ¼
d
dxi02ðxÞ at x ¼ 0
as follows (the details being given in Appendix A):
Rc ¼1
ðRsc þ RsmÞW
�
RscRsm d1 þ2a
sinh d1a þ cosh d1
a tanhDxa
( )þ
ðR2sc þ R2smÞa cosh d1a þ R2sca sinh
d1a tanh
Dxa
sinh d1a þ cosh d1
a tanhDxa
266666664
377777775
ð7Þ
where d1 ¼ d Dx. This equation is similar to that ob-tained by Suzuki et al. [14,15] for the case of source
drain contact resistance of silicided thin film SOI
MOSFET’s. Corresponding expression for end resis-
tance is given by
Re ¼
RscRsm
þ coshd1a
sinhd1aþ cosh
d1atanh
Dxa
aRsmW
� 1
"
Rsc 1 exp Dxa
� �2ðRsc þ RsmÞ 1þ exp 2 Dx
a
� �#
ð8Þ
The above equation reduces to Eq. (6) for Dx ¼ 0. Now,
using Eqs. (5) or (7) and Eqs. (6) or (8) for d, d16 4a, asimultaneous solution for a and Rsc can be obtained,
which in turn can be used to get qc using Eq. (4a)
2.2.2. Voltage probe position between 0 and d1If the voltage probe is shifted to the right of the
current probe on pad 1 by an amount Dx1, and to the lefton pad 2 by the same amount, as shown by the solid line
in Fig. 2(b), there is an additional correction term for Rcby an amount DRc, to be subtracted from the RHS of
Eq. (7), as discussed in Appendix A. It is given by
DRc ¼ 0 for Dx1 ¼ 0; and,
DRc ¼RscRsmDx1
ðRsc þ RsmÞW
þ R2smaðRsc þ RsmÞW
1 cosh Dx1a
�sinh d
a
RscRsm
�"þ exp
� d
a
��
þ 1 exp
� Dx1
a
�#for Dx ¼ 0 ð9bÞ
For d1 P 4a, this is simplified to
DRc ¼RscRsm
ðRsc þ RsmÞWDx1
�þ a
RsmRsc
1
� exp
� Dx1
a
���ð9cÞ
Similarly, in the case of end resistance measurement,
when the voltage probe is shifted by a distance Dx1 to theright of the current probe on pad 2, as shown in Fig. 3(c),
the term DRe to be subtracted from the value of Re givenin Eq. (8) is the same as given in Eqs. (9a)–(9c), i.e.,
DRe ¼ DRc ð9dÞ
We may note that the position of voltage probes on
pad 1 is not important as no current is flowing in this
pad. Similarly the position of current probe on pad 2 is
not important since the voltage is measured only be-
tween the other two pads.
2.2.3. Voltage probe position between 0 and DxNow, if the voltage probe is placed to the left of the
current probe on pad 1 and to the right on pad 2, the
term to be subtracted from the RHS of Eq. (7), for
the shift of the voltage probe by Dx1 from the current
probe, i.e., the x ¼ 0 position, can be shown to be as
follows (detail given in Appendix A)
DRc ¼R2sma
ðRsc þ RsmÞW
RscRsm
þ cosh d1a
sinh d1a þ cosh d1
a tanhDxa
�1 exp Dx1
a
� �þ exp 2Dx
a
�1 exp Dx1
a
� �1þ exp 2Dx
a
�ð10aÞ
The term DRc reduces to zero for Dx1 ¼ 0. In the case of
Re measurement also the correction term to be sub-
tracted for a similar shift of the voltage probe by Dx1 asshown in Fig. 3(d) is again given by Eq. (10a), i.e.
DRc ¼RscRsmDx1
ðRsc þ RsmÞW
þ R2smaðRsc þ RsmÞW
1 cosh Dx1a sinh Dx1
a tanh Dxa
�þ cosh d1
a exp Dx1a
�sinh d1
a exp d1a
�cosh Dx1
a
sinh d1a þ cosh d1
a tanhDxa
" #ð9aÞ
508 M. Ahmad et al. / Solid-State Electronics 46 (2002) 505–512
DRe ¼ DRc ð10bÞ
2.3. Variable contact size method
The variation of contact length is not of much use in
end resistance measurement, since for small transfer
length the values of Rsc and the contact resistivity be-come geometry dependent and stabilise only for d of
about 50 lm [3]. We may note that, in case of d P 4a,the variation of d can also be used to estimate the
modified resultant semiconductor resistance under a
contact and hence the correct value of contact resistivity.
This is discussed in the following:
Let the contact resistance determined for a particular
row of transmission line model with contact pads of
length d 0 and varying contact spacing be Rc1 as foundfrom Eq. (1a) and Rc2 for another row of TLM pattern
of contact length d 00. Then using Eq. (5) for the case of
finite metal resistance one obtains
DRcjW ¼ Rc1 Rc2 ¼RscRsm
W ðRsc þ RsmÞðd 0 d 00Þ ð11Þ
Thus we see that the value of Rsc can be found fromEq. (9a) as other parameters can be measured experi-
mentally. DRc can be optimised using a suitable range oftest structures with variation in d and W. The accuracy
of Rsc determination is dependent on the measurement ofDRc and Rsm. Therefore, the value of Rsc has to be ob-tained from a set of different d values to improve the
accuracy. The proposed method works better when the
metal sheet resistance is high and comparable to Rsc.
3. Experimental detail
In order to explore the effect of metal sheet resistance
on end resistance and to find the contact resistance re-
lated parameters a and Rsc, the following samples wereused. Two samples are p type GaAs (i) A (p � 4:8� 1017
cm3) and B (p � 1:6� 1017 cm3), with a 2 lm thick
epitaxial layer grown on semi-insulating (SI) GaAs with
contacts of AuZn. Sheet resistance of the contact metal
was deliberately enhanced by using about 200 A thick
Au–Zn film. The other three samples are n type GaAs
(n � 8� 1016 cm3) of 1 lm thick epitaxial layer grown
on SI GaAs with different contacts of about (i) 2000 �AAthick AuGeNi/Au (sample C), (ii) 2000 �AA AuGeNi /Pd/
Au, (sample D) and (iii) 1000 �AA AuGeNi (sample E).
Samples C and D were aged at 450 �C for 3.5 h and
sample E was aged at 400 �C for about 1.5 h before
measurements. All samples were prepared by organic
cleaning followed by defining a mesa photolithographi-
cally, etching the epilayer in orthophosphoric acid:hy-
drogen peroxide:DI water::3:1:50 and rinsing in DI
water. TLM geometry contacts were formed by lift-off
technique, patterning with positive photoresist followed
by vacuum evaporation of the contact metal. The TLM
pattern typically has a pad length d of 170 lm and a
spacing between the pads varying from 25 to 125 lm. Toevaluate metal sheet resistance, contact materials were
also deposited on reference SI GaAs. Subsequent to
deposition, contacts were formed by rapid thermal an-
nealing at 430 �C in flowing nitrogen on a graphite boat
for 30 s for p-GaAs and at 445 �C for 2 min on a tan-
talum boat heater for n-GaAs. The reference samples
were also subjected to the same heat treatment as the
samples under test.
The contact resistance was measured using a Karl
Suss prober AP 4. A constant current of 0.1 mA was
passed through two contact pads using a Keithley 220
programmable current source and voltage was measured
between the same pads using a Keithley 2000 digital
multimeter. Four probes were used, two for passing
current and two for measuring voltage. The current and
voltage probes were kept collinear so as to minimise the
errors which could be introduced by the two dimen-
sional flow of current. The end resistance was measured
by passing current between two pads and measuring
voltage between one of these and a third adjacent probe
as shown in Figs. 1 and 3. The sheet resistance of the
metal was measured by four probe technique on the
blanket metallised and alloyed SI GaAs sample for
AuZn contact. For AuGeNi contacts the resistance Rsmwas measured on a line contact in the SI region of GaAs
Fig. 3. Probe positions in end resistance measurement.
M. Ahmad et al. / Solid-State Electronics 46 (2002) 505–512 509
by passing current between a length and measuring
voltage using two separate probes.
4. Results and discussion
Next we consider the experimental results of end re-
sistance and contact resistance measurements. The val-
ues of Rc found from the R–L plots are given in Table 1together with the measured end resistance values for
different contacts. Using these values and the known
contact parameters d, W and metal sheet resistance Rsm,the values of Rsc, a and contact resistivity are obtainedby solving Eqs. (5) and (6) or Eqs. (7) and (8) simulta-
neously. The corrections due to shift in the probe posi-
tions from the pad end have also been accounted in the
calculations presented in Table 1.
We first discuss the results of AuZn contacts to p-
GaAs, i.e. the samples A and B. The values of Re arehigh as expected because of the high values of Rsm of
these samples. If one uses Eqs. (3) and (5), no solution
for Rsc can be found. The obvious reason being that thevalue of Re expected from Eq. (3) is smaller than the
experimental value by about more than two orders of
magnitude. When the expression including the effect of
metal resistance, i.e., Eq. (8) is used, reasonable values
of a and Rsc are found. Eq. (4b) is then used to find thecontact resistivity qc. Thus we see for large values of Rsmthat the results of end resistance experiment cannot be
used in the analysis of contact resistivity unless the
modified expression including the metal sheet resistance
is used.
We find results similar to the above for the aged
contacts on n type samples. In samples C, D and E, the
metal sheet resistance is relatively smaller. Still, using
measured Rc and Re one cannot find a solution for theparameters a and Rsc if one uses the expressions whichneglect Rsm. On the other hand these parameters are
readily obtained using the expression which includes
Rsm. The values of qc obtained for these contacts werehigh because the contacts were aged as specified above.
We also note here that the Rsc values obtained by ex-periments are considerably different from the original
semiconductor sheet resistance Rss. If one had used
Rsc ¼ Rss, one would get apparent contact resistivity
values q0c which are also listed in Table 1 and are quite
low but very misleading.
The experiments conducted above are somewhat
contrived to enhance the end resistance and demonstrate
its effectiveness in the determination of realistic qc. Evenin the cases where the actual contact resistivity and Rsmare small, it is important to measure the end resistance
Re which could be in the milliohm range in realistic cases
where qc could be as small as 106 X cm2 and Rsm ffi 0:1–
0.01 X/sq. This is considerably larger than the nano-ohmvalue calculated if Rsm is neglected altogether.
Finally for non-zero Rsm, the correction required forthe displacement of probes from the exact edges is
considerable, specially for Rsm values of a few ohm or
more in the above cases, i.e. for sample A, B and E. To
avoid errors and more complex formula it is better to
keep the probes as near to the edges of the pads as
possible. In case Rsm is made nearly zero by using a thickoverlayer or plating then the formula for Rc becomesmuch simpler and DRc, DRe tends to be negligible. Onlythe expression for Re as given by Eq. (6) for Dx ¼ 0 or
Eq. (8) Dx 6¼ 0, remains significantly different from the
case given by Eq. (3), when Rsm is exactly zero.
5. Conclusion
In this work a modified expression for end resistance
has been presented for the case of finite metal resistance.
It has been shown that the end resistance does not de-
crease exponentially with increasing contact length but
has a finite asymptotic value, once the metal resistance,
however small it may be, is taken into account. Further,
it is shown that the end resistance is easily measurable
even in case of small transfer length, large contact length
and small metal resistance. Several cases are presented
which underline the importance of using the modified
formula to include the end resistance correction for es-
timating the actual value of contact resistivity. In the
case of alloyed contacts, the correction can be fairly
large because of the modification of the semiconductor
resistance Rsc from the nominal value of Rss. It has beenshown that the experimental data on contacts with finite
Rsm and large contact length can be explained only whenthe resistive effect of metal resistance is included in the
Table 1
Calculation of contact resistivity qc from the measured values of the end resistance, Re and contact resistance Rc
Sample Rss (X/sq) Rsm (X/sq) d, W (lm ) Rc (X) Re (X) Rsc (X/sq) a (lm) qc (X cm2) q0
c (X cm2)
A 258 39 170, 73 45 18 38 46 1:6� 103 9� 106
B 775 39 170, 73 80 28 70 57 3:7� 103 1:0� 104
C 445 2.2 127, 1050 2.5 1.1 28 81 1:9� 103 1:3� 104
D 445 2.8 130, 1050 3.5 1.7 34 90 3� 103 2:7� 104
E 142 5.5 170, 670 1.25 0.25 14 40 3:2� 104 9:6� 106
510 M. Ahmad et al. / Solid-State Electronics 46 (2002) 505–512
end resistance. An alternative method has also been
proposed to evaluate the resultant modified semicon-
ductor resistance under the contact which uses rou-
tine Rc measurements with a mask set of different
contact lengths or by measuring Rc for different probepositions.
Acknowledgements
The author would like to thank Prof. R. Lal, Prof. J.
Wasi and Mrs. Rani of IIT, Bombay, for their labora-
tory support in electrical measurements.
Appendix A
Following the method used by Marlow and Das a
one dimensional double transmission line model for
rectangular geometry is being used as shown in Fig. 1(b).
In this analysis we have assumed that voltage drops
along the y-axis are negligible compared to drops
along the current flow path which is x-axis. In this
condition one dimensional approximation is appropri-
ate and the formula derived allows us to evaluate the
contact resistivity with minimal error. In practice this
happens when the voltage and current probes are col-
linear.
The distributed resistance is divided into infinitesimal
elements whose values are given in Fig. 1(b). Let us
assume that the current enters or leaves the contact
at a point away from the edge by Dx as shown in
Fig. 1(b). Let the total current entering and leaving the
contact be i0, it is divided into two components i1 andi2 in the region between x ¼ 0 to d1, i1 flowing throughthe metal and i2 through the semiconductor. In the
second region, i.e. between 0 and Dx let the cur-
rent flowing in the semiconductor and the metal be i02and i01 respectively. The resulting potential drops in themetal and semiconductor are V1ðxÞ and V2ðxÞ respec-tively. The following Eqs. (A.1)–(A.3) are written re-
lating the voltage, current and parameter qc, Rsc andRsm.
V2ðxÞ V1ðxÞ ¼ di2qc
W dxðA:1Þ
d
dxV1ðxÞ ¼ i1ðxÞ
RsmW
ðA:2Þ
d
dxV2ðxÞ ¼ i2ðxÞ
RscW
ðA:3Þ
Now, in the region between x ¼ 0 and d1
i1 ¼ i0 i2 ðA:4Þ
Differentiation of Eq. (A.1) and substituting for the
voltages from Eqs. (A.2) and (A.3) results in a second
order inhomogeneous differential equation
d2
dx2i2ðxÞ ðRsc þ RsmÞi2ðxÞ þ
Rsmi0qc
¼ 0 ðA:5Þ
The solution is sum of a general solution obtained
by equating the term without i2ðxÞ to zero, plus a par-ticular solution. Since the third term is a constant,
the particular solution is obtained by assuming i2ðxÞ aconstant.
i2ðxÞ ¼ C1 þ C2 expxa
� �þ C3 exp
� xa
�ðA:6Þ
where,
C1 ¼Rsmi0
Rsc þ RsmðA:7Þ
is the particular solution.
Using Eq. (A.4) we get
i1ðxÞ ¼ C1RscRsm
C2 expxa
� � C3 exp
� xa
�ðA:8Þ
Now, in the region between x ¼ 0 and Dx
i01 ¼ i02 ðA:9Þ
Using Eqs. (A.1)–(A.3) with i2 replaced by i02 and Eq.(9a), the following differential equation is obtained
d2
dx2i02ðxÞ
ðRsc þ RsmÞi02ðxÞqc
¼ 0 ðA:10Þ
The solution for which can be written as
i02ðxÞ ¼ C4 expxa
� �þ C5 exp
� xa
�Using the boundary condition: i02ðxÞ ¼ 0 at x ¼ Dx
the equation reduces to
i02ðxÞ ¼ C4 expxa
� �� exp
� 2Dx
a
�exp
� xa
��ðA:11Þ
The constants C2, C3, C4 are found by applying thefollowing boundary conditions,
i2ð0Þ ¼ i02ð0Þ ðA:12aÞ
d
dxi2ðxÞ ¼
d
dxi02ðxÞ
����at x ¼ 0 ðA:12bÞ
and
i1ðd1Þ ¼ C1RscRsm
C2 expd1a
� � C3 exp
� d1
a
�¼ 0
ðA:12cÞ
M. Ahmad et al. / Solid-State Electronics 46 (2002) 505–512 511
It can be shown that
C2 ¼ C1RscRsm
1þ tanh Dxa
�þ exp d1
a
�2 sinh d1
a þ 2 cosh d1a tanh
Dxa
" #ðA:13aÞ
C3 ¼ C1RscRsm
1 tanh Dxa
�þ exp d1
a
�2 sinh d1
a þ 2 cosh d1a tanh
Dxa
" #ðA:13bÞ
C4 ¼C1
1þ exp 2Dxa
� RscRsm
þ cosh d1a
sinh d1a þ cosh d1
a tanhDxa
" #
ðA:13cÞ
According to the definition of end resistance Re givenabove in Section 2.1.2, it is V2ðxÞ V1ðxÞ at the end of thecontact divided by i0, but since the current probe is as-sumed to be shifted from the edge by Dx; the potentialdifference measured by the voltage probes is now ¼ VAC.It has been assumed that one voltage probe is at the
same position as the current probe and the other one is
at the adjacent pad; since no current is flowing in this
pad the voltage measured is same as that at A.
Following the path ABC,
VAC ¼ RscW
Z Dx
0
i02 dxþqcW
d
dxi02
����x¼0
ðA:14aÞ
Substituting i02 from Eq. (A.11), we obtain
Re ¼VACi0
¼ aC4Wi0
" Rsc 1
� exp
� Dx
a
��2
þ qca2
1
�þ exp
� 2Dx
a
��#ðA:14bÞ
The expression for end resistance Re is found by sub-stituting for the constants. The result has been given in
Eq. (4a) of Section 2.1.2.
According to its definition the contact resistance is
the voltage drop at the front edge of the contact divided
by the total current flowing through the contact. Thus, it
is given as V2ðxÞ V1ðxÞ at x ¼ d1, plusR 0d1dV1 divided by
i0, this is ¼ ðVEF þ VFCÞ=i0, assuming that the voltageprobe is also at Dx. Using Eqs. (A.1) and (A.2), we get
V2ðxÞ V1ðxÞjd1 ¼qcWa
C2 expd1a
� �� C3 exp
� d1
a
��ðA:15aÞ
Z d1
0
dV1 ¼RsmW
C1xRscRsm
� C2a exp
xa
� �
þ C3a exp� xa
��d1
0
ðA:15bÞ
On substituting for the constants, adding and simplify-
ing and diving the sum by i0, the expression for Rc isobtained and given in Eq. (4a) of Section 2.1.2.
Now, when the voltage probe position is shifted to a
place between 0 and d1 from current probe, i.e., x ¼ 0
position, by Dx1 then a termR Dx10
dV1 ¼ ðRsm=W ÞR Dx10
i1 dxis to be subtracted from the expression for Rc. Now, whenthe voltage probe position is shifted to a place between 0
andDx from current probe by Dx1 on pad 1 then a termRDx10
dV1 ¼ ðRsm=W ÞRDx10
i01 dx is to be subtracted fromthe expression for Rc. The same result holds for the caseof end resistance.
References
[1] Berger HH. Solid State Electron 1972;15:145–58.
[2] Shen TC, Gao GB, Morkoc H. J Vac Sci Technol B
1992;10(5):2113–31.
[3] Reeves GK, Harrison HB. IEEE Trans Electron Dev
1995;EDL-42(8):1536–47.
[4] Santhrama L, Schumacher H, Leblanc HP, Esagui R, Bhat
R, Koza M. Electron Lett 1990;26:1127–9.
[5] Vogt A, Simon A, Hartnagel HL, et al. J Appl Phys
1998;83(12):7715–9.
[6] Wakita AS, Mott N, Rosner SJ, Fischer-Collbrie A. J Vac
Sci Technol B 1995;5:2092–9.
[7] Reeves GK, Harrison HB. IEEE Trans Electron Dev Lett
1982;EDL-3(5):111–3.
[8] Scorzoni A, Finetti M. Mater Sci Rep 1988;3:79–138.
[9] Heiblum M, Nathan MI, Chang CA. Solid State Electron
1982;25(3):185–95.
[10] Porges M, Lalinsky T. Solid State Electron 1992;35(2):
157–8.
[11] Ahmad M, Ganguli T, Patil S, Major S, Roy NR, Arora
BM. Solid State Electron 1998;42(4):473–6.
[12] MarlowGS, DasMB. Solid State Electron 1982;25(2):91–4.
[13] Reeves GK, Harrison HB. Electron Lett 1982;18(25/
26):1083–5.
[14] Suzuki K, Tanaka T, Tosaka Y, Sugii T, Andoh S. IEEE
Trans Electron Dev 1994;ED 41(6):1007–12.
[15] Ahmad M, Arora BM. Solid State Electron 1992;
35(10):1441–5.
512 M. Ahmad et al. / Solid-State Electronics 46 (2002) 505–512