06/09/2004 Sebastien Couet 1
Desy Summerstudent Program 2004
Calibration of a 3D Hall probe for magnetic field investigations in
undalators and wigglers.
Outline:
•Motivation
•Experimental procedure
•Magnetic field characterisation
•Hall probe calibration
•Conclusion
06/09/2004 Sebastien Couet 2
Schematic wiggler
Electron remains in the ring for a long time.
They have to stay on track!
Motivation
We have to take great care of the influence of the magnetic
field in the wiggler.
Front view
Bx
Bz
06/09/2004 Sebastien Couet 3
Motivation
Use of a Hall probe:
• Gives a voltage function of the magnetic field.
• Point like measurement.
• Fast response.
Problem:
•We need a precision of 10-6 T.
Apparition of non linear effect.
06/09/2004 Sebastien Couet 4
23
3210
3210
xzxzz
xxx
VVcVcBccV
BaBaaV
Given Hall potentials:
Procedure:
• Characterisation of the field
- coil(horizontal)
- dipole magnet(vertical)
• Determination of c1
• Determination of a1, a2, c2
• Determination of cross sensitive parameter c3
Procedure
06/09/2004 Sebastien Couet 5
Magnetic field characterisation
Use of a NMR probe:
Advantages:
- Gives a very precise absolute value of the field.
Disadvantages:
- No direction information.
- Need homogenous field(not point like).
06/09/2004 Sebastien Couet 6
In the coil:
ImTBz *2185.40212.0)(
Magnetic field characterisation
(horizontal field)
06/09/2004 Sebastien Couet 7
Magnetic Field Characterisation
In the dipole magnet:
Bx
Bz
BNMR22zxNMR BBB
22 )( baIBx
(Vertical + horizontal)
06/09/2004 Sebastien Couet 8
Magnetic field characterisation
Finally, we get the evolution of the slope coefficient:
Conclusion.
Field determination is possible but:
- use of bigger coil
- need of stronger horizontal field
To get more accurate results.
06/09/2004 Sebastien Couet 9
Hall probe calibration
Offset measurement:
Use of a zero gauss chamber.
Temperature dependence:
Dependence: 10-4 V/ºC
06/09/2004 Sebastien Couet 10
Hall probe calibration
Horizontal component:zz BccV 10
Need a very precise alignment.
C1=5.10134 0.005 V/T
06/09/2004 Sebastien Couet 11
Conclusion
Determination of a two component magnetic field is possible with NMR.
Need: - Better homogeneity
- bigger horizontal field
Bigger coil.
Calibration of a 3D hall probe.
- need to improve the precision of the measurement.