DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

40
0197 IMPORTANT This manual provides operating instructions for the CLESTA. The instructions contained in this booklet should be thoroughly read and understood before operating the unit. After the installation has been completed, keep this manual in a safe place and refer to it for future maintenance. If you have any questions about this Manual or this product, please contact us. If manual becomes unreadable or is lost, please request a new manual by contacting your dealer. Installation should be conducted by authorized personnel only. Follow instructions on installation manual. DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

Transcript of DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

Page 1: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

0197

IMPORTANT

This manual provides operating instructions for the CLESTA. The instructions contained in this booklet should be thoroughly read and

understood before operating the unit. After the installation has been completed, keep this manual in a safe place and refer to it for future maintenance.

If you have any questions about this Manual or this product, please contact us. If manual becomes unreadable or is lost, please request a new manual by contacting your dealer.

Installation should be conducted by authorized personnel only. Follow instructions on installation manual.

DENTAL UNIT

INSTALLATION andOPERATING INSTRUCTIONS

Page 2: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS
Page 3: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

Page

SAFETYPRECAUTIONS------------------------------------------------- 1

1.GENERALINTRODUCTIONS---------------------------------------- 5

2.INSTALLATIONINSTRUCTIONS----------------------------------- 8 3.OPERATINGINSTRUCTIONS--------------------------------------- 19

4.DAILYMAINTANANCE----------------------------------------------- 23

5.MAINTANANCEANDINSPECTION5-1.GUIDEFORDAILYMAINTENANCEANDINSPECTION---- 27(Maintenanceandinspectionbyuser)5-2.DUIDEFORPERIODICALCHECK-UP------------------------ 29

6.ELECTROMAGNETICCOMPATIBILITY(EMC)----------------- 30

7.LISTOFCOMPATIBLEHANDPIECES----------------------------- 33

TABLE OF CONTENTS

Page 4: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

Intended Use of the ProductThisproductisanactivetherapeuticdeviceintendedfortheexclusiveusefordiagnoses,treatmentsandrelativeproceduresofdentistry.The product must be operated or handled by the qualified dentists or by dental staffs under thesupervisionofthedentist.Suchdentistsordentalstaffsshouldinstructand/orassistthepatientstoapproachtoandleavefromtheproduct.Patientsshouldnotbeallowedtooperateorhandletheproductunlesshe/sheissoinstructed.Theproductissuppliedtogetherwiththehandpieceslikeelectricmicromotor,airturbineand/ormotor,scalerandsoon.

Compatibility of HandpiecesUsethecompatiblehandpiecesasshownontheattachedlistforthisunit.(Listofcompatiblehandpieces).

Important Notes In case of the troubles, please contact Takara Belmont offices or your dealers. Donotdisassembleorattempttorepair. Disassembly, repair or modifications should only be done by a qualified repair technician. Attempts at disassembly, repair or modifications may lead to abnormal operation and accidents.

Disposal of residue material

Pleaserequestaspecialcontractorwhenyoudisposeamalgam.

In case of disposal of equipment Whendisposingtheunit,appropriatelydisposecomplyingwithallcurrentapplicableregulationsandlocalcodes.InEUarea,EUdirective2002/96/EConwasteelectricalandelectronicequipment(WEEE)isappliedonthisproduct.Inthisdirective,environmentconsciousrecycling/abandonmentisobligated.

Page 5: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

W A

EC REP

SN

Protective earth (ground)

ON (power) OFF (power)

Chair last position Chair auto return Chair preset 1 Chair preset 2

To raise the chair To recline the backrest To lower the chair

To raise the bacrest

Chairmanual control

Fiber optic handpiece light on/off

Micro motor rotation speed

Scaler power control

Micro Motor Forward/Reverseselect

Syringe Saliva ejector Bowl flush

Cupfiller Service outlet (water) Water Air

Water heater Serial number

Authorizedrepresentative in the European community

Manufacturer Date of manufacture

Non-ionizingradiaion

Separatecollection for electrical and electronicequipment

Type Bapplied parts

alternating current

CautionIt means “caution, warnings, or possibility to danger”.

Refer to instructionmanual/booklet

Inthismanual,onthelabelsoronthecontrolpanelofCLESTA,followingsymbolsareused.Confirm the meaning of each symbol.

SYMBOLS

Page 6: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

■Beforeuse,readthe“Safetyprecautions”carefullytoensureproperuse.

■Thefollowinginformationisdesignedtoensuresafeuseofthisproductandtopreventinjuryanddamage

to you and others. The precautions contained here are classified depending on the severity and degree of

imminenceofpossibleinjuryordamageresultingfromimproperuse.Besuretofollowallthe

information,whichisimportantforsafety.

SAFETY PRECAUTIONS

Thissymbolindicatesthat“itisrecommendedtofollowtheseprecautionsforsafety.”

Classification of precautions Severityanddegreeofimminenceofpossibleinjuryordamage

WARNING

CAUTION

NOTICE

Thissymbolindicatesthat“ignoranceoftheseprecautionsmayleadtosevereinjuryorevendeathasaresultofimproperuse.”

Thissymbolindicatesthat“ignoranceoftheseprecautionsmayleadtomildormoderatephysicalinjuryordamagetopropertyasaresultofimproperuse.”

 

WARNING1.Besuretoturnoffbreakersforequipmentintheclinicwhenthisproductwillnotbeusedforalongperiodoftime

Besuretoturnoffbreakersforequipmentintheclinicwhenthisproductwillnotbeusedforalongperiodoftime(followingthecompletionofwork,duringthesuspensionofbusiness,etc.).Insulationdegradationmaycause electrical fire.

2.BesuretoturnoffthemainswitchuponcompletionofworkorduringworkbreaksBesure to turnoff themainswitchuponcompletionofworkorduringworkbreaks.Thisprevents incorrectoperationduetoaccidentalcontactandassociatedhazards.

4.DonotplaceanundueloadonthearmDonotgetonorplaceanundueloadonthearmofthisunitordentalchairarmrest.Thiscouldcausetheunittotoppleorotheraccidents.

5.BesuretoestablishagroundingconnectionBesure toestablishapropergroundingconnection.(Refer toavendorforgroundingconnection.)Failureorelectricleakagemayleadtoelectricshock.

6.Neverdisassemble,repairormodifythisproductIndividuals other than certified repair technicians should not disassemble or attempt to repair and modify this product. This could lead to an accident, failure, electric shock or fire.

3.DonotsitonotherthanseatWhenthebackrestisattheforwardposition.donotsitonorplaceanundueloadontheheadrestorlegreatofdentalchair.Thiscouldcausetheunittotoppleorcoulddamagetheunit.

7.UsewithcautioninthepresenceofelectromagneticinterferencewavesDonotplace thisproductaroundequipmentgeneratingelectromagneticwaves(includingcommunicationsequipment,elevators,etc.)asincorrectoperationofthisproductmayoccurinthepresenceofelectromagneticinterferencewaves.Donotuseequipmentgeneratingelectromagneticwaves,suchasmobilephones,aroundthisproduct.

-1-

Page 7: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

SAFETY PRECAUTIONS

 

WARNING8.Besuretoturnoffthemainswitchwhenelectrocauteryisinuse

Besuretoturnoffthemainswitchwhenelectrocauteryisinuse,becausenoisemaycauseincorrectoperationofthisproduct.

10.Donotplaceobjectsweighing3kgormoreontheDoctor'stable

•Failuretomaintainthisproductmayleadtophysicalinjuryorpropertydamage.•Refertothesectionofmaintenance.

11.Besuretousethemirrorcover•Besuretousethemirrorcoverofthedentallightwhenthelightisturnedon.Directcontactwithlampsmaycauseburns.•SeetheInstructionManualofthedentallightforfurtherinformation.

12.Besuretoturnoffthepowerwhenreplacinglamps

9.Ensurethemaintenanceofthisproduct

DonotplaceobjectsweighingthreekilogramsormoreontheDoctor'stable.ThiscouldcausedamagetotheDoctor'stable,defectivefunctionoraccidents.

•Besuretoturnoffthepowerwhenreplacingthedentallight.Thiscouldresultinelectricshock.•Useonlydedicatedhalogenlamps.•Immediatelyafterahalogenlamphasburntout,thelampandthelampholderarestillhot.Replacethelampaftertheycooldown.•Donottouchhalogenlampswithbarehands.•SeetheInstructionManualofthedentallightforfurtherinformation.

13. Immediately wipe off any water spills or leakage on the floorImmediately wipe off any water spills or leakage on the floor. Decreased strength of the floor may lead tophysicalinjuryincludingfall,orpropertydamage.

14.UsewithcautiononpatientswithacardiacpacemakerUsethisproductwithextremecautiononpatientswithacardiacpacemaker.Inthecaseofanyabnormalitiesinpatientsduringuse,immediatelyturnoffthisproductanddiscontinueuse.

CAUTION1.Onlyexperiencedpersonnelshouldusethisproduct

Onlydentistsorotherdentalprofessionalsshouldusethisproduct.

2. Confirm safety before use.Before use, confirm that the parts are correctly and safely operating and that there are no obstacles around this product.

4.Discontinueuseifyoufeelthat“somethingiswrong”

Keepyoureyesonpatients(especially,children)sothatmischieforinadvertentoperationofequipmentwillnotleadtounexpectedaccidents.

3.Payattentiontopatientsandchildren

Alwaysbecarefultoinspectthisproductforlooseness,rattling,tilting,wobbling,sounds,temperature,odors,etc. Immediately discontinue use at the first feeling that “something is wrong.”

-2-

Page 8: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

SAFETY PRECAUTIONS

 

CAUTION5.keepyoureyesonthepatientduringoperation.

6.PayattentionduringmovementoftheDoctor'stable•PayattentiontosurroundingswhenyoumovetheDoctor'stable.Injurybythetipsofhandpieces,etc.,mayoccur.•BesuretomovetheDoctor'stablebyholdingthehandleoftheunit.•Besuretomoveup&downtheDoctor'stablebyreleasingthebalancearmbrakeoftheunit.

7.DonotplaceanythinghotontheDoctor'stable

12.PrecautionsforcleaningtheresincoverForcleaning,donotusecleaningagentscontainingsolventorabrasives,thinnersoroil-basedalcohol(butanolandisopropylalcohol),whichmaycausecracks.

DonotplaceanythinghotontheDoctor'stable.Thiscouldcausedeformationordiscoloration.

10.DonotsmackorrubthisproductDonotsmackorrubthisproductforcefully.Thiscouldcausedamagetocoversordefectivefunction.

8.Cautionswhenadjustingtheheightofthecart-typeinstrumenttable(Carttypeonly)Afteradjustingtheheight,besuretofastentheloosenedlockscrew.Failuretoheedthiswarningmaycausethetabletodrop,resultinginanaccident.

9.Observethecarthose(Carttypeonly)Donottreadonthecarthose.Failuretoheedthiswarningmayresultindamagetothehose,aswellaspeopletrippingover.

11.Precautionsforcleaningaspittoonbowl

•Confirm that the patient is seated in the proper position. Keep your eyes on the patient during the operation.•Payspecialattentiontosurroundingsatautomaticoperationofthedentaltreatmenttable.Damagetothebackrest,stoolorDoctor'stablemayoccur.

Thespittoonbowlismadeofglass.Handlewithcare.Donotwashitwithhotwater.Otherwise,itmaybreak.

13.Immediatelywipeoffdrugsolutionwhenitcomesintocontactwiththisunit

Shoulddrugsolutionorwatercomesintocontactwiththisunit,immediatelywipeitoffwithadrysofttowel,etc.Thiscouldresultindefectivefunctionorelectricleakageaswellasspottingorrusting.

14.ClosethewatermainvalveandturnoffthemainswitchuponcompletionofworkBesuretoclosethewatermainvalveandthemainswitchattheendofeachworkdaytopreventwaterleakagefromoccurring.

16.Payattentionduringtheheadrestoperation

Besuretooperateswitcheswithyourhands,except thefootcontroller,whichisoperatedwithyourfoot.Operationwithbodypartsotherthanhandsmaycausedamageorincorrectoperation.

17.BecarefulnottoheatanemptywaterheaterExercise caution as heating of an empty water heater may result in burning of the heater, leading to fire.

Do not allow hands, fingers, or hair to become entangled in the moving parts of the headrest during operationg

15.Besuretooperateswitcheswithyourhands

-3-

Page 9: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

SAFETY PRECAUTIONS

 

CAUTION18.Precautionsforcleaningtheoperationpanel(membraneswitches)

The water tank is intended only for use with purified water, distilled water and pure water. Do not use mouthwash orelectrolyzedwater,suchasConCoolorpovidoneiodine,astheymaycausecloggedtubingoraffectinternalvalvesandequipment.

22. Precaautions for insert the bowlflush nozzle and the cupfiller nozzle to the spittoon bowl

Penetrationofdropletsofsanitizingsprayintothebackoftheoperationpanelmaybeassociatedwithswitchfailure.Useapapertowelsoakedwithsanitizingsolutiontocleanthesurfaceoftheoperationpanel.

20.Setthepressureofthewatertankat200kPaorlessAdjust theairsupplypressurefor thewater tankto200kPaorless.Anexcessivelyhighpressuremaycausedamagetothewatertank.

19.Precautionsforcleaning

21. Do not use water other than purified water, distilled water or pure water for the water tank

 

NOTICE1.Troubleshootingandcontactinformation

Inthecaseofanyproblems,discontinueuse,turnoffthemainswitchandcontactthedealerorourcompany.

2.CheckoperationofthecompressorWithnoairsupplied,thisproductdoesnotoperateevenafterturningonthemainswitch.Turnonthepowerofthecompressorbeforeoperatingthisproduct.

4.Usetheturbinewithawatercheckvalve

Thewaterunitisintendedforusewithtapwater.Cautionshouldbeexercisedastheuseofwaterotherthantapwater(waterthroughasterilizerofwatersystems,etc.)mayresultinfailureofequipment.

5.HandlingofequipmentinthecaseofapowerfailurePutthehandpieceintheholderandturnoffthemainswitchifequipmentstopsworkingduringuseduetoapowerfailureorotherreasons.

3.Precautionswhenusingwaterotherthantapwater

Usetheturbinewithawatercheckvalve.Contactthedealerorourcompanywhenaturbinewithoutawatercheckvalvewillbeused.

•Neverusesandpaper,metalscrubbrushesandabrasivecleaningagentstocleantheunit.•Donotusestronglyacidiccleaningagentsoralkalinepipecleaningagentstoavoidcorrosionofmetals,etc.

23.ReadthedocumentsaccompanyingthevariouspiecesofequipmentBeforeuse,besure tocarefullyreadthepackageinsertsandInstructionManualsaccompanyingthevariouspiecesofequipment(includingoptionalarticles)toensureproperuse.

Aftercleaningofthespottoonsection,besuretoinsertthenozzlesstraightlytospittoonbowl.Ifnot,itmaybedamagetheo-ringanditmayleakthewater.

-4-

Page 10: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

-5-

Bowl Flush Nozzle

Cuspidor Top

Assistant Holder Bar

Saliva Ejector

Assistant Holder

Light Pole

Cupfiller Nozzle

Syringe

High Volume Evacuator

Foot Control

Dental Light

Horizontal Arm

Light Post

Cuspidor Unit

Balance Arm

Dental Size Film Viewer

Doctor Table TrayControl Panel

Handle

Umbilical Hose

Junction Box

Syringe

Waste Receptacle

Handpiece Holder

Handpiece Tubing

Handpieces(Air Turbine, Air Motor, Scaler and etc.)Manufacturers recommend to use the handpieces with CE markings.

[1] GENERAL INTRODUCTIONS

(1)UNITTYPEVARIATIONS&MAJORPARTSIDENTIFICATIONS1.UNITMOUNTTYPE

Fig.1-1Overview

Page 11: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

-6-

Dental Size Film Viewer

Doctor Table Tray

Caster

Cart Base

Cart Hose

Junction Box

Foot Control

Handle

Umbilical Hose

2.SPLITCARTHOLDERTYPEWITHCUSPIDOR

Fig.1-2Overview

Page 12: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

188

171 287 364

294493

76

529

98

98250

610

52

860

783 65

0

423

521

562

905

330°

14°

38°

180°

330°

90°90°

(2)DIMENSIONS

Fig.1-3Dimensions

Ratedpowersupply-------------------------------------------------AC230V50/60HZ0.7AAirmainpressure---------------------------------------------------0.45to0.5MPaWatermainpressure------------------------------------------------0.1to0.2MPaNetweight-----------------------------------------------------------80.5kg(withoutdentallight)Dentallight----------------------------------------------------------AL-501Workingenvironment-----------------------------------------------Temperature:(Operating)+5℃~+40℃

(Storage)-10℃~+50℃

Humidity:10~80%Atmosphericpressure:600hPa~1060hPaServiceLife-----------------------------------------------------------10Years

Classification of foot controller ------------------------------------IPX1(applicablestandardIEC60529)Protectionclassagainstelectricshock----------------------------ClassIequipmentAppliedparts---------------------------------------------------------TypeBappliedparts:Handpieceforunit(Listofcompatiblehandpieces)Equipment that is not suitable for use in air, flammable anesthetic gas, oxygen or nitrous oxide

(3)SPECIFICATIONS

-7-

-mm-

Page 13: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

-8-

220

130

527

85 121

135

11055

7070

311

243

175

CM/OP MIN.1000 SC MIN.1200

MIN.800

MIN

.600

CENTER OF CLESTA CHAIR

VACUUMDRAIN

VACUUMOPERATINGPOWER

SOURCE

AIR

WATER

[2] INSTALLATION INSTRUCTIONS(1)PREPARINGFORINSTALLATIONTherecommendedlocation,plumbingandlayoutofjunctionboxareshowninFug.2-1,2-2and2-3respectively.1.LOCATIONANDPLUMBNGLAYOUTOFJUNCTIONBOX

Fig.2-1LocationandPlumbingLayout

StandardType

Page 14: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

-9-

220

130

527

80 121

140

110

7070

311

243

175

CM/OP MIN.1000 SC MIN.1200

MIN.800M

IN.6

00

VACUUM

DRAIN

VACUUMOPERATING

POWER SOURCE

AIR

WATER

CENTER OF CLESTA CHAIR

BelmontType

NOTE:CM/OP.....CHAIRMOUNTOVERTHEPATIENTTYPE

SC............SPLITCARTTYPE

Fig.2-2LocationandPlumbingLayout

Page 15: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

-10-

40

20 30

80

F

A

D EG

B

C

VACUUM DRAIN POWER SOURCE &VACUUM OPERATING

AIR & WATER

2.PLUMBINGHEIGHTOFJUNCTUONBOX

Fig.2-3PlumbingHeight

A:NIPPLEB:SOCKETJOINTC:AIR

WATER

D:DRAIN

E:POWERSOURCE

SUPPLYCABLEEARTHWIREF:VACUUM

G:VACUUMOPERATING

;PT1/2-1/2(Brass);UnplasticiziedPolyvinylChloridePipeFitting;UnplasticiziedPolyvinylChloridePipe(ShockResistingType)'HI13'18mmO.D.-13mmI.D.;UnplasticiziedPolyvinylChloridePipeforWaterWorksService'VP13'18mmO.D.-13mmI.D.;UnplasticiziedPolyvinylChloridePipe'VP40'48mmO.D.-40mmI.D.;UnplasticiziedPolyvinylChlorideConduit'VE16'22mmO.D.-16mmI.D.;PolyvinylChlorideInsulatedandSheathedCable'VVF2.0-2C';PolyvinylChlorideInsulatedWire'VE1.6';UnplasticiziedPolyvinylChloridePipe'VP20'26mmO.D.-20mmI.D.;UnplasticiziedPolyvinylChlorideConduit'VE16'22mmO.D.-16mmI.D.

Page 16: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

-11-

1.CHECKTHEPACKINGCheckeachquantityofthepartsforinstallationaslistedbelow. *Standard Specification1)SHUTOFFVALVESFORAIR&WATER 1SET2)WOODSCREW 4PCS.3)FLEXIBLEPIPE 2PCS.4)DRAINELBOW 1PCE.5)VACUUMELBOW 1PCE.6)DRAINCAP 1PCE.7)BASKETSTRAINER 1PCE.8)CUPFILLERNOZZLE 1PCE.9)BOWLFLUSHNOZZLE 1PCE.10)SALIVAEJECTORTIP 1PCE.11)VACUUMTIPS(BENT&STRAIGHT) 1SET12)SILICONTIP 10PCS.13)LEVELADJUSTMENTSCREW 4PCS.14)WASTERECEPTACLE 1PCE.15)LIGHTPOLE 1SET16)SILICONMATFORDR.TABLE 1PCE.17)WIRECONNECTOR 3PCS.18)SEAL 4PCS.19)PATIENTCUP 1PCE.20)VIALSWITHSTAND 1SET21)COTTONCONTAINERWITHSTAND 1SET21)MOUNTINGBRACKETCOVER 1PCE.

2.ATTACHUNITTOCHAIR2-1.MountingbracketAType(Fig.2-4)1)Raisetheupperstructureofdentalchairtothehighestposition.2)InstallthecuspidorunitonMountingBracketATypewith4pcs.ofM10-50FlatHeadScrews,SpringWasherandNuts.3)Adjustthelevelofcuspidorunitwith4pcs.ofM10-25LevelAdjustmentScrews.4)SticktheSealsovertheFlatHeadScrews.5)Removethestopperscrewfromthemountingbracket.6)Attachthemountingbracketcoveronthe mounting bracket, and fix with the stopperscrew.

[2] INSTALLATION PROCEDURES

Fig.2-4AttachUnittoChair

Flat Head Screw

MountingBracket A

Level Adjustment Screw

Cuspidor Unit

NutSpring Washer

Mounting Bracket

Seal

StopperScrew

Mounting BracketCover

StopperScrew

Page 17: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

-12-

Dental LightPlastic Wahser

Light Pole

Set Screw

Cover

Cable from Dental Light

Cable from Cuspidor Unit

2-2.MountingBracketBType(Option)(Fig.2-5)1)Raisetheupperstructureofthedentalchairtothehighestposition.2)AttachtheAdaptortoMountingBracketBTypewith4pcs.oftheM10-40SocketHeadCapScrewsandSpringWashers.3)InstallthecuspidorunitonMountingBracketBTypewith4pcs.ofM10-50FlatHeadScrews,SpringWasherandNuts.4)Adjustthelevelofcuspidorunitwith4pcs.ofM10-25LevelAdjustmentScrews.5)SticktheSealsovertheFlatHeadScrews.6)Removethestopperscrewfromthemountingbracket.7)Attachthemountingbracketcoveronthe mounting bracket, and fix with the stopperscrew.

Fig.2-5AttachUnittoChair

Fig.2-6InstallDentalLight

3.INSTALLDENTALLIGHT(Fig.2-6)1)RemoveCoverfromthebottomoftheunit2)PutPlasticWasherontheLightPole.3)RuntheCablethroughLightPoleandattachtheDentalLighttotheLightPole.4)RuntheCablethroughLightPostandinstallthePole,thentighten2pcs.oftheSetScrews.5)ConnectthewiresfromtheDentalLightandCuspidorUnitinaccordancewiththe identification color.6)Re-attachtheCover.

Flat Head Screw

Socket Head ScrewSpringWasher

Mounting Bracket BLevel Adjustment Screw

Nut

Adaptor

Cuspidor Unit

Spring Washer

Seal

Mounting Bracket

Stopper

Mounting BracketCover

Stopper

Page 18: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

1 2 3 4 4 5 5 6 7 8

1 2 3 4 4 5 5 6 7 8

4 5

4 5

L

C C

L

Umbilical Hose

Umbilical Hose Clamp

Brown 1/8”Blue 5/16”

Yellow 1/4”

Light Green 1/4”Wirings for Water Heater & Dental Light

White Black

White Black

Brown 1/8”Blue 1/16”

Yellow 1/4”

Light Green 1/4”

Wirings for Water Heater

Wirings for Dental Light

Blue Brown

BlueBrown

24V Line Voltage

Small thin Screw Driver

Electric Terminal

Wiring

-13-

4.CONNECTUMBILICALHOSEWITHJUNCTIONBOX(Fig.2-7,2-8&2-9)ThissectionisonlyforSPLITCARTTYPE.Forothertypeofunit,proceedto5.1)OpenthecoverofJunctionBoxbylooseningtwoscrews.2)LoosenthescrewofUmbilicalHoseClamp.3)Insertallhoses,tubingsandwiringsofUmbilicalHosefromCuspidorunitintoJunctionBoxthroughtheUmbilicalHoseClamp.4)CorrectthetwistoftheUmbilicalHoseastheJunctionBoxisplacedattheplannedlocation.5)SettheendofUmbilicalHoseattheUmbilicalHoseClampthenre-tightenthescrewfortheUmbilicalHoseClamp.6)Connecttubingswithnylonsleevestothe fittings as shown in Fig.2-8 or 2-9.7)ConnectwiringstoElectricTerminalasshowninFig.2-8or2-9.

*HowtoconnectwiringtotheElectricTerminal(Fig.2-10)DuringpressingtheslotoftheElectricTerminalbytheSmallThinScrewDriver,insertthewiringtotheElectricTerminal.

Fig.2-7ConnectUmbilicalHose

Fig.2-8 Standard Specification

Fig.2-9 Low Voltage SpecificationFig.2-10HowtoConnectWiring

Page 19: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

-14-

Standard Type Belmont Type

Center of Clesta Chair Center of Clesta Chair

Water Shut Off Valve

Air Shut Off Valve

Water Shut Off Valve

Air Shut Off Valve

ElectricalReceptacle

Shut Off Valve

Nipple

ElectricalReceptacle

Wood ScrewVacuum Drain Hose

Drain Hose

Drain Elbow

Vacuum Elbow

Vacuum Drain Hose

Vacuum Elbow Drain HoseDrain Elbow

Vacuum Pipe Drain Pipe

Drain Bush (Option)

Drain Joint (Option)

Drain Bush (Option)

5.CONNECTUTILITYSECTION(Fig.2-11&2-12)1)AttachShutOffValvestotheairandwatersupplylines,pointingtoproperdirectionsforfurtherconnections.2)AttachElectricalReceptacletoMainPowerSupplyCable.(TheElectricalReceptaclemustbepreparedlocally.)3)OpeneachShutOffValvetoremoveanychipsordirtintheairandwatersupplyline,thenclosetheValves.4)OpenthecoverofJunctionBoxbylooseningtwoscrews,andplaceJunction Box at the planned location, then fix it on the floor with 4pcs. of Wood Screws.5)CuttheDrainHoseandtheVacuumDrainHoseattheproperlength,andconnectandgluethemtoeachelbow.6)ConnecttheVacuumElbowtotheSuctionPipe.7)InserttheDrainElbowintotheDrainPipe.

*AttachingDrainBushing(Option)totheDrainPipeisdesirable,andDrainElbow should be bonded firmly into the Drain Pipe.* For Air Vacuum specification, prepare DrainJoint(Option)asshowninFig.2-13,sothatthedrainlineandthevacuumdrainlinemaybeinsertedtogetherintotheDrainPipe.

Fig.2-11ConnectSupplyLines

Fig.2-12ConnectDrainLines(1)

Fig.2-13ConnectDrainLines(2)

Page 20: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

-15-

Shut Off ValvePacking

Filter

Filter

Standard Type Belmont Type

Air Filter Air FilterWater Filter Water Filter

1 2 3 4 4 5 5 6 7 8

1 2 3 4 4 5 5 6 7 8

Cable for Chair Control (9P)

Cable for Chair Safety Device (4P)

Cable from Chair

Main Power Supply Cable

VacuumOperatingCable (9P)

Plug

VacuumOperating Cable

Cable for Chair Control (9P)

Cable for Chair Safety Device (4P)

Cable from Chair

Main Power Supply Cable

VacuumOperatingCable (9P)

Plug

VacuumOperating Cable

SeeFig.2-14&2-15.8)ConnectWaterShutOffValveandWaterFilterwithStainlessFlexiblePipeandPacking.ConnectAirShutOffValveandAirFilterinthesameway.

SeeFig.2-16or2-17.9)PreparetheElectricalPlug,andconnectittoMainPowerSupplyCable.(ElectricalPlugmustbepreparedlocally.)10)ConnectVacuumOperatingCabletotheElectricTerminalasshowninFig.2-16or2-17. (For Central Vacuum Specification only)11)Runthethreecablesfromchair(PowerSupplyCablewithplug,Cablewith9PconnectorandCablewith4Pconnector),throughtheholeoftheJunctionBoxwallabovetheUmbilicalHoseintotheJunctionBox.12)ConnectthePowerSupplyCableofthechairtothereceptacle,andconnectothertwocablestothecorrespondingconnectors.

Fig.2-14ConnectFilters(1)

Fig.2-15ConnectFilters(2)

Fig.2-16ConnectElectricalcable(1)

Fig.2-17ConnectElectricalcable(2)

Page 21: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

-16-

6.ATTACHFILMVIEWER(Fig.2-18&2-19)1)RemovetheDoctorTableTraybylooseningeightFlatHeadScrews.2)RunthecablefromFilmViewerintoDoctorTable.3)AttachFilmViewertothebracketforDoctorTablewithtwoPhilipsScrews.4)ConnectthecablefromFilmViewertotheElectricalTerminal.

7)SETUPCUSIPDORTOP(Fig.2-20)1) Insert Cupfiller Nozzle and Bowl Flush Nozzle,andtighteneachnut.2)PutBasketStrainerandDrainCapintothecuspidorbowl.

Flat Head Screw

Philips Screw

DoctorTableTary

ElectricalTerminal

FilmViewer

Electrical Terminal

Cupfiller Nozzle

Bowl Flush NozzleDrain Cap

Basket Strainer

Fig.2-18AttachFilmViewer

Fig.2-19ConnectWiringsforFilmViewer

Fig.2-20CuspidorTop

Page 22: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

-17-

(3)ADJUSTMENTBEFOREUSE1.AIR&WATERSUPPLYLINE(Fig.2-21&2-22)1)ChecktheAirPressureGaugeonthesideofJunctionBox.ThePressureshouldbeapproximately0.45Mpato0.5Mpaforair.Foradjustmentthepressure,pullupandturntheknobofAirRegulatorwhiletheair is flowing (use 3-way Syringe for example).Toincreasethepressure,turnclockwise.Todecrease,turncounter-clockwise. After adjustment, push down the knob to fix it.

2)CheckiftheWaterPressureGaugeindicatesthepressureataround0.1Mpato0.2Mpa.AdjustmentcanbedonebyWaterRegulator.Toadjustthepressure,liftandturnthewater regulator knob until the water starts flowing. (use3-waySyringeforexample).Toincreasethepressure,turnclockwise.Todecrease,turncounter-clockwise.Afteradjustment,lowertheknobbackinplace.

3)Astheunitisreadytorun,turnontheMasterSwitchlocatedunderneaththedoctortable(seeFig.2-23.),thenrunthesalivaejectorfor a while to flow the water through the unit andmakesurenoleakagetakesplaceattheutilitysection.SeeFig.2-23&26.4)TurnallCoolantWaterControlKnobscounter-clockwisefullyandpickupallhandpiecetubingfromtheholder,thendepressDrivePedalofthefootcontrol(SpraySwitch-ON)foramoment to flush out the water from the tubing into thecuspidorbowl.

Pull Up

Push DownKnob

Increase decrease

Air Regulator

Water RegulatorIncrease

decrease

Knob

Pull UpPush Down

Water Air

Air Pressure Gauge

Water Pressure Gauge

Master Switch

ONOFF

Coolant WaterControl Knob

Increase decrease

Fig.2-21InsideofJunctionBox

Fig.2-22SideViewofJunctionBox

Fig.2-23BottomViewofDoctorTable

Page 23: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

-18-

2.DriveAir&SprayforHandpieces(Fig.2-24,2-25&2-26)1)Connecthandpiecestoeachtubing,andmakeadjustmentonebyone.2)Toadjustthedriveairpressureforeachhandpiece,turntheappropriateAdjustScrewforDriveAironetheAuto-SelectValveinthedoctortable.Turnthescrewclockwisetoclosethelinefully.ThenduringdepressingDrivePedalofFootControlfully,graduallyincrease the sir flow by turning the screw counter-clockwiseuptoHandpiecePressureGaugereadsthefullrecommendedpressure.3)Toadjustsprayofhandpiece,useAdjustScrewforCoolantAirontheAuto-SelectValveandCoolantWaterControlKnob.

Flat Head Screw

DoctorTableTray

Auto-SelectValve

PressureGauge

Adjust Screw for Drive Air

Adjust Screw for Coolant Air

Drive Pedal

Fig.2-24InsideofDoctorTable

Fig.2-25Auto-SelectValve

Fig.2-26FootControl

Page 24: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

-19-

(1)DOCTORUNITSECTION1.MASTERSWITCH(Fig.3-1&3-2)1)TurnontheMasterSwitchandthePowerIndicatorislitingreen.2)Aftertheoperation,turnofftheMasterSwitch,thenthesuppliesofair,waterandelectricityareallshutdownattheutilitysection.

2.TABLEPOSITION(Fig.3-1)Forstandardmodelofthechairmountoverpatienttype,theheightofthedoctortableisalways fixed by Air Brake. The height canbeeasilychangedbypressingBrakeReleaseButton.

3.CHAIRCONTROL(Fig.3-3)ThechairmovementcanbecontrolledbypressingChairControlonMembraneSwitchA.Pleaserefertooperatinginstructionsforchair.

4.CUPLILLER&BOWLFLUSHCONTROL(ByAIRTIMERSYSTEM)(Fig.3-3)Bypressingeachmembraneswitch,waterissuppliedforacertaintime.For cupfiller, filling time is set for 4 seconds atfactory.For bowl flush flushing time is set for 6 seconds atfactory.

5.3-WAYSYRINGE(Fig.3-2&3-4)1)Bypressingeitherorbothbuttons,Syringeoffersair,waterandspray.2)SyringeTipcanberotatedfreely.3)Toadjustair/water/spray,useFlowControlKnobslocatedunderneaththedoctortable.

[3] OPERATION INSTRUCTIONSFilm ViewerPower Indicator

Membrane Switch A

Brake Release Button

Membrane Switch B

Handpiece Holder

Master Switch ONOFFAir

Water

Flow Control Knobfor 3-Way Syringe

Coolant Water Control Knob

2

LP0

1

Cupfiller Switch

Chair Manual Control

Chair Automatic Control

Bowl Flush Switch

w A

Air Button

Water Button

Syringe Nozzle

Fig.3-1DoctorUnit

Fig.3-2BottomView

Fig.3-3MembraneSwitchA

Fig.3-43-WaySyringe

Page 25: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

-20-

6.OPERATINGHANDPIECE(Fig.3-5)1)Pickupahandpiecefromthepandpieceholder,anddepressDrivePedalofFootControl.2)Forthedry-cutting,turnoffSpraySwitchofFootControl.3)WhiledepressingChipPedal,onlythechipblowcomesoutfromthehandpiece.

7.FILMVIEWER(Fig.3-6)FilmViewerSwitchislocatedontherightside of film viewer. Press the switch once toilluminatetheviewer,pressitagaintoturnitoff.

8.FIBEROPTICHANDPIECE(OPTION)(Fig.3-7)FiberOpticHandpieceSwitchislocatedontheMembraneSwitchB.Pickupaoptichandpiecefromthehandpieceholderandthehandpiceislit.Returnthehandpiecetotheholderanditisputout.

9.ELECTRICSCALER(OPTION)(Fig.3-7)PowerControlforElectricScalerislocatedundertheMembraneSwitchB.

ON

OFFCoolant Water Switch

Drive Air Pedal

Chip Blower Button

Film Viewer Switch

Fig.3-5FootControl

Fig.3-6FilmViewer

Fig.3-7MembraneSwitchB

1

0

2 3

4

5

Fiber Optic Handpiece Switch

Power Control for Electric Scaler

HandpiecesRefertohandpiecemanufacturersoperatinginstructions.

CAUTION

Page 26: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

-21-

(2)ASSISTANTUNITSECTION1.CUPFILLER&BOWLFLUSHCONTROL(ByAIRTIMERSYSTEM)(Fig.3-8)Bypressingeachmembraneswitch,waterissuppliedforacertaintime.For cupfiller, filling time is set for 4 seconds afactory.For bowl flush. Flushing time is set for 5 secondsatfactory.

2.3-WAYSYRINGE(Fig.3-4&3-9)1)Bypressingeitherorbothbuttons,Syringeoffersair,waterandspray.2)SyringeTipcanberotatedfreely.3)Toadjustair/water/spray,useFlowControlKnobslocatedinthecuspidorbody.

3.MASTERVALVEBLOCK(Fig.3-9)Openthesidepanelofthecuspidorbody.Flow Control knobs for Cupfiller and BowlFlushareaccessible.

4.CUSPIDORBODYROTATION&TELESCOPICASSISTANTHOLDERBAR(Fig.3-10)1)CuspidorBodycanrotateby90degreesforassistant’sconvenience.2)Fortwo-handstreatment,AssistantHolderBarcanbeextendedby50cm.

5.ASSISTANTINSTRUMENTHOLDER(Fig.3-10)Pickuptheinstrument(SalivaEjectororHighVolumeEvacuator)fromtheholder,anditstartsworking.

Bowl Flush Switch

Cupfiller Switch

Cupfiller Flow Control Knob

Bowl Flush Flow Control Knob

Air Flow Control Knob for 3-Way Syringe

Water Flow Control Knob for 3-Way Syringe

Air Vacuum Flow Control Knob

Increase decrease

TelescopicAssistantHolder bar

AssistantHolder

Fig.3-8MembraneSwitchforAssistant

Fig.3-9MasterValveBlock

Fig.3-10CuspidorUnit

Page 27: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

-22-

Water Heater Switch

Bowl Flush Switch

Service Water OutletFlow Control Knob for Service Water Outlet

- +Cup Base

MANUAL SENSOR

H

M LH

Touchless Switch

6.WATERHEATERSWITCH(Fig.3-11)TurnontheWaterHeaterSwitchwhenthewarmwaterforthepatientcupisrequired.

7.SERVICEWATEROUTLETWITHFLOWCONTROL(Fig.3-11)WaterOutletwithFlowControlislocatedonthepanelwhichadaptstothequickconnector.

8.BOWLFLUSHSWITCH(OPTION)(Fig.3-11)TurnontheBowlFlushSwitchandthewaterisflushed to the bowl continuously.

9.AUTOMATICCUPFILLER&BOWLFLUSH(OPTION)(Fig.3-12)1)Placethepatientcupweighting45to80gonCupBaseandwaterissuppliedfora certain quantity to the patient cup and flushedforacertaintimetothebowl.2)QuantityofwaterinpatientcupcanbeadjustedbyturningCupBase.

(3)LIGHTSECTION(MODELIO5000)1.MODESELETIONSWITCH(Fig.3-13)TurntheswitchtoSENSORside,andON/OFFofthelightcanbemadebyapproachinghandofthesensor.(TouchlessSwitch).

2.INTENSITYSWITCH(Fig.3-14)ThenintensityoflightcanbeselectedL(Low),M(Medium)orH(High).RefertothemanualofIO5000.

Fig.3-11CuspidorControlPanel

Fig.3-12 Automatic Cupfiller

Fig.3-13ModeSelectionSwitch

Fig.3-14IntensitySwitch

Page 28: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

-23-

Drain Cap

BasketStrainer

Solid Collector

Solid Collector Filter

(1)DRAIN&SUCTIONLINEOFUNIT(Fig.4-1)1.BasketStrainer.RemovetheBasketStrainerCap,andcleantheBasketStrainer.2.SolidCollectorFilterOpenthesidepanelofthecuspidorbodyandtakeoutSolidCollectorFilterandcleanit.PullingontheSolidCollectorCapwillbringtheSolidCollectorFilterwithit.

(2)AIRSUPPLYLINE(Fig.4-2)1.AirFilterDrainValve 1) Release the water in the air filter by opening DrainValveofJunctionBox.2)Turnoffthemainswitchofthecompressor.Releasecompressedairanddrainaccordingtotheinstructionsofthecompressor. Confirm the air Pressure Gauge indicates0Mpa.

[4] DAILY MAINTENANCE AND CARE

Drain Valve

Air Pressure Gauge

Fig.4-1BasketStrainer&SolidCollector

Fig.4-2JunctionBox

Fig.4-3VacuumHndpiece&SalivaEjectorHandpiece

Vacuum Tip

Saliva EjectorHandpiece

VacuumHandpiece

Vacuum

Vacuum Cap

Saliva EjectorHose

Body

Body

Hose

Slide Knob Slide Knob

SalivaEjectorCap

Hose Connector

Filter

(3)HANDPIECE1-1.VacuumHandpieceandSalivaEjectorHandpieceForeffectivesterilization,washingforremovingcontaminationandimmersionbyacleaningagentarerequired.Then,rinsebywaterinordertoremoveresidualcleaningagentonmedicaldevice.Takefollowingproceduresfromcleaningtosterilization.*Usethedisposablesalivaejectortip.

Disassembly (Fig.4-3)Disassemblethehandpieceforthepreparationof cleaning as the following figures show. Pullthehoseconnectortodisconnectthevacuumhose.

Page 29: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

CAUTION

Sterilization pouch

Fig.4-7VacuumHandpieceandSalivaEjectorHandpiece

-24-

Cleaning by hand A.Wipeoffthesurfacecontaminationbyaclothwhilerinsingthesurfacebyrunningcleanwarmwaterat40±5degrees.(Fig.4-4) Scrub the intubation or hole, slide groove and filter by acleaningbrushorbyatoothbrushwithrunningcleanwarmwaterat40±5degrees.(Fig.4-5)Wipeoffbyaclothfortheareawhichbrushisunabletoreach.(Fig.4-6)

B.Rinsethoroughlybydistilledwateratordinarytemperatureorbycleanwaterformorethan1minute.C.Checkwhethercontaminationisremovedornotaftercleaning.Continuethecleaningifcontaminationisremained.D.Immersedwithanalkalinedetergentfor5minutes.(WerecommendtouseID212madebyDURR)E.Rinsethoroughlybydistilledwateratordinarytemperatureorbycleanwaterformorethan1minute.

Sterilization (Fig.4-7)VacuumTip/VacuumCap/VacuumHandpieceBody/SalivaEjectorHandpieceBodycanbeautoclave.Vacuumhandpiecebodyandsalivaejectorbodyhavetoassemblebeforeautoclave.A.Insertthehandpieceinasterilizationpouchandsealit. B. Autoclave for 20 min. at 121℃ or 4 min. at 135℃. Sterilizationwithautoclaveispermittedupto250times.

Cleaning must be done within 1 hour after use.Throw out and do not take autoclave handpiece forfollowing cases. • Any waste material can not be removed by clogged hole. • Contamination and solid material attached to handpiece can not be removed.

StorageAftercleaningthevacuumtip,keepitinthecleanplace.

• Sterilization by class B cycles. • Sterilization temperature is 135 degrees or less. • The cap, filter and body are made of resin. They may become deteriorated if they are sterilized in an autoclave many times. • After autoclave sterilization, the cap, filter, body and valve are subject to discoloration, which does not have a negative effect on performance. • The slide knob can be autoclave 100 times and is expendable supplies. • Skip the drying process if the temperature is to exceed 135℃. • If damage occurs to the sterilization pouch, discard and sterilize again using a new pouch.

CAUTION

Cloth

Fitting cleaning brushor Tooth brush

Slide groove

Cloth

Fig.4-4Surfacecleaning

Fig.4-5Cleaningofholeandslidingpart

Fig.4-6Cleaningthearea(brushisunabletoreach)

Page 30: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

-25-

2.VacuumHose/SalivaEjectorHose(Fig.4-9)Itispracticaltoremovethehosesinordertocleanthem.Note:Attheendoftheday,cleanthevacuumandsalivaejectorhandpiecesbysuckingintwocups’worthofwater.

3.AIRMOTOR/TURBINE/3-WAYSYRINGESterilizethehandpieceaccordingtomanufacturer’soperatingmanual.

VacuumHose

SalivaEjectorHose

Fig.4-9VacuumHose&SalivaEjectorHose

CAUTION

4.Belmont77Syringe(Fig.4-10)DisassemblyRemovethenozzlefromsyringebyturningitindirectionA.

Cleaning by hand A.Wipeoffthesurfacecontaminationbyaclothwhilerinsingthesurfacebyrunningcleanwarmwaterat40±5degrees. Scrubthetipandjointpartofnozzlebyacleaningbrushorbyatoothbrushwithrunningcleanwarmwaterat40±5degrees.(Fig.4-11)B.Rinsethoroughlybydistilledwateratordinarytemperatureorbycleanwaterformorethan1minute.C.Checkwhethercontaminationisremovedornotaftercleaning.Continuethecleaningifcontaminationisremained.D.Immersedwithanalkalinedetergentfor5minutes.(WerecommendtouseID212madebyDURR)E.Rinsethoroughlybydistilledwateratordinarytemperatureorbycleanwaterformorethan1minute.

Cleaning must be done within 1 hour after use.Throw out and do not take autoclave handpiece for following cases. • Contamination and solid material attached to handpiece can not be removed.

1-2.Cleaningvacuumandsalivaejectorlines(Fig.4-8)Thesuckingunitcomesintocontactwithsecretions,spitandbloodthatcontainbacteriaeveryday.Besuretocleanandsterilizeitattheendofeachworkday.Recommendedcleaner:OrotolPlusmadebyDURR.

Donotusestronglyacidiccleaningagentsoralkalinepipecleaningagents,whichmaycausecorrosionofmetals,etc/

CAUTION Fig.4-8CleaningVacuum/SalivaElectorLines

Fig.4-11CleaningBelmont77SyringeNozzle

A

B

w A

Nozzle

Nut

Syringe body

Fig.4-10Belmont77Syringe

Cleaning brushor Tooth brush

Page 31: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

(8)FILTERREPLACEMENT(Fig.4-13)The water filter in the junction box needs to bereplacedatleastonceayear.The air filter in the junction box needs to bereplacedatleastonceeverythreeyears.Contactyourlocalservicerepresentativeforreplacement.

-26-

Water Filter

Air Filter

Fig.4-13Filter

(6)LUBRICATINGHANDPIECESRefertothemanualofthehandpieces.

(4)HOSESANDTUBINGCleanhosesandtubingusingDURRFD333.

(5)CLEANINGUNITAllsurfacescanbecleanedwithDURRFD333cleaner.Spraythecleaner(FD333)onclothandwipethesurfaceswiththecloth.

CAUTION

Donotdrenchtheunit.Wipeallsurfacesdryaftercleaning.

StorageAftercleaningthenozzle,keepitinthecleanplace.

• Sterilization by class B cycles. • Sterilization temperature is 135 degrees or less. • Skip the drying process if the temperature is to exceed 135℃. • If damage occurs to the sterilization pouch, discard and sterilize again using a new pouch. • Before use, make sure that the nut is firmly tightened.

Sterilization (Fig.4-12)Thenozzlecanbesterilizedwithautoclave.A.Insertthehandpieceinasterilizationpouchandsealit. B. Autoclave for 20 min. at 121℃ or 4 min. at 135℃.Sterilizationwithautoclaveispermittedupto250times.ApplyVaselinethinlyandevenlytothetwoO-ringsaftersterilization. Tighten the nut firmly in direction B to undo the nozzle.

CAUTIONFig.4-12SterilizationBelmont77SyringeNozzle

Sterilization pouch

(7)MASTERSWITCHAfterthedailyoperationandmaintenance,turnofftheMasterSwitch.

Page 32: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

[5] MAINTANANCE AND INSPECTION5-1. Guide for daily maintenance and inspection (Maintenance and inspection by user)•Managementofmaintenanceandinspectionofmedicalequipmentshouldbeimplementedbytheuser(medicalinstitution).Incasetheuserdoesnotimplementsuchmanagement,itispermitted that such management is outsourced to a qualified entity such as a medical equipment repaircompany. • For safe use of this product, it is necessary that inspection should be conducted in the specified frequencyonthetemsdescribedbelow.

Air turbine revolution, water flow, air flow and so forth shall be free of abnormality.

Micromotor revolution, water flow and so forth shall be free of abnormality.

Scaler vibration, water flow and so forth shall be free of abnormality.

No. Item Frequency Inspection method and diagnosis

Influence if inspection not conducted

Maintenance required in case of nonconformity

Check of safety functions

Before start

Before start

Before start

Before start

For each patient

For each patient

For each patient

After closing

Unexpected personal injury and troubles may arise due to motion of the chair during medical treatment and due to pinching between doctor section and chair.

Contact your dealer or our office if any abnormality arises.

Check for leakage of water and air

Leakage of water and air shall not be observed around the product.

The product will not normally work, and troubles may arise.

Contact your dealer or our office if any abnormality arises.

Cupfiller When a paper cup is placed on the cupfiller, the cup shall be detected and cupfilling shall be executed.* Malfunction may arise if the cup is of another material grade (such as stainless steel and plastics) or if the paper cup is of dark color or pattern.

Cupfilling may not be executed.

Execute re-inspection in accor-dance with "Method for operation" described in the instruction manual. Contact your dealer or our office if recovery is not achieved as a result of re-inspection.

Check of motions of equipment

Troubles such as injury in patient's oral cavity and equipment failure may arise.

Control the water flow in accor-dance with "Control of compo-nents" described in the instruction manual. If any other abnormality arises, refer to the instruction manual attached to individual equipment. Contact your dealer or our office if recovery is not achieved.

Check of air turbine bar

Appropriate bar shall be positively mounted.Make sure to refer to the instruction manual attached to individual equipment.

The bar will not normally work and troubles may arise.

If abnormality such as flaw and deformation is found on the bar, replace the bar in accordance with the instruction manual attached to individual equipment.

Check of scaler tip

Appropriate tip shall be positively mounted and be correctly used.Make sure to refer to the instruction manual attached to the scaler.

The tip will not normally work and troubles may arise.

If the tip was worn or deformed, replace the tip in accordance with the instruction manual attached to the scaler.Contact your dealer or our office if any other trouble arises.

Check of tightness of syringe nut

The nut for fixing the nut of Type 77, 3-way syringe shall be positively tightened.

Troubles may arise if the nut comes off.

Turn and positively retighten the nut that fixes the nozzle.

Matters attached to micromotor

Excessive handpiece oil or the like shall not be attached to the motor section.

The motor section will not work normally and troubles may arise.

Execute care in accordance with the instruction manual attached to individual micromotor.

Make sure the chair movement stops by any of the following actions.

when foot controller pedal is depressed.

During chair auto movement, depress of any operation switch.

While setup is in progress with function switch on the doctor membrane switch panel.

When the spittoon bowl is turned to patient side (Pedestal type)

-27-

Page 33: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

No. Item Frequency Inspection method and diagnosis

Influence if inspection not conducted

Maintenance required in case of nonconformity

After closing

CareVacuum and saliva ejector handpiece

Flush the suction line, and then clean the filter of the vacuum or saliva ejector handpiece.

Faulty suction may arise.

Clean the suction line and filter in accordance with "Method for care" described in the instruction manual.

9

CareCuspidor section

Clean the cuspidor and dust filter.

After closing

Faulty water drainage may arise.

Execute dust removal and cleaning in accordance with "Method for care" described in the instruction manual.

CareSolid collector

Clean the filter of the solid collector.

Vacuum suction will become weak.

Clean the filter in accordance with "Method for care" described in the instruction manual.

After closing

CareExterior

Chemical, filthy water and so forth shall not be found (attached or remaining) on the product exterior.

Discoloration and deterioration to the exterior, and corrosion and rusting to metallic components may arise.

Execute wiping in accordance with "Method for care" described in the instruction manual.

Check of main switch and main valves

The product main switch shall be off, and water/air main valves shall be closed.

Product failure and troubles may arise.

Contact your dealer or our office if the main switch cannot be turned off or if the main valve cannot be closed.

Product's moving parts

No abnormal noise or the like shall be produced from product's moving parts when the product is operated.

The product will not normally work and troubles may arise.

Contact your dealer or our office if any abnormality arises.

CareDrain valve

Drain the water from the air filter drain valve.

Water may enter the air line, and equipment failure may arise.

Drain the water from the air filter in accordance with "Method for care" described in the instruction manual.

Check of water pressure and air pressure

Check the water pressure and air pressure by reading pressure gauges in the U-Box section.Basic set pressure : Water: 0.1 to 0.2 MPaAir: 0.45 to 0.5 MPa

The product will not normally work, and troubles may arise.

Contact your dealer or our office if the set pressure is abnormally high or low.

Check of conditions of table section

The table shall be free of inclination, and water shall not flow on the table.

Injury caused by falling of goods located on the table and other troubles may arise.

Contact your dealer or our office if any abnormality arises.

Care Oil mist separator

The oil level in the oil mist separator shall be lower than the red line.

Normal output will not be produced due to inferior handpiece exhaust.

Discharge the oil in accordance with "Method for care" described in the instruction manual.

After closing

After closing

10

11

12

Once every week

13

14

15

16

17

18

Once every week

Once every month

Once every month

Once every month

-28-

Page 34: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

5-2. Guide for Periodical Check-up

Somepartsandcomponentsoftheproductsaredegradedordeteriorateddependingonthe

frequencyofuse.Annualcheck-upandmaintenance,aswellasreplacementofconsumableparts,

arerequired.

Therequiredparts(includingconsumableparts)arelistedbelow.Itmaybedifferentfromthe

followinglistdependingontheoptionoftheunit.

Forcheck-upandrepair,callatechnicianofourauthorizeddealer.

Parts and components that require periodical check-up

Parts Description

Vacuum handpiece body

Saliva ejector handpiece body

Foot controller

Water supply hose

Drain hose

Air supply hose

Electric wiring of moving part

Standard LifetimeNo.

3 years 1

2 3 years

3 5 years

3 years

3

4

4

5 3 years

Consumable parts

No. Parts Description

1

2

Parts Description Standard LifetimeNo.

3 years 8

9 3 years

10 5 years

5 years11

12 3 years

Regulator

Valves

Switches

Film viewer body part

Pressure gauge

Arm section of moving part

Control PCBs.

6 3 years 13 7 years

7 5 years 14 5 years

Valve for vacum handpiece body

Vacuum tip

Handpiece tubings

Vacuum hose

Saliva ejector hose

8

No. Parts Description

6

7

Filter for oil mist separator

Filter (Air & Water)

O-ring, Packing, Diaphragm

5

Executethemaintenanceinaccordancewiththisinstractionmanualandoperatingmanualattachedtoeachindividualequipment(Dentallight,Handpiece,etc..).Failuretomaintainthisproductmayleadtophysicalinjuryorpropertydamage.

WARNING

-29-

Page 35: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

[6] ELECTROMAGNETIC COMPATIBILITY(EMC)MedicalelectricalequipmentneedsspecialprecautionsregardingEMCandneedstobeinstalledandputintoserviceaccordingtotheEMCinformationprovidedinthismanual.PortableandmobileRFcommunicationsequipmentcanaffectmedicalelectricalequipment.Theequipmentorsystemshouldnotbeusedadjacenttoorstackedwithotherequipment.Ifadjacentorstackeduseisnecessary, theequipmentorsystemshouldbeobservedtoverifynormaloperationintheconfiguration in which it will be used.

Guidance and manufacture’s declaration – electromagnetic emissionsThe CLESTA Unit is intended for use in the electromagnetic environment specified below. The customer or the useroftheCLESTAUnitshouldassurethatitisusedinsuchanenvironment.

Emissions test Compliance Electromagnetic environment - guidanceRFemissionsCISPR11 Group1

TheCLESTAUnitusesRFenergyonlyforitsinternalfunction.Therefore,itsRFemissionsareverylowandarenotlikelytocauseanyinterferenceinnearbyelectronicequipment.

RFemissionsCISPR11 ClassB TheCLESTAUnitissuitableforuseinallestablishments,

includingdomesticestablishmentsandthosedirectlyconnectedtothepubliclow-voltagepowersupplynetworkthatsuppliesbuildingsusedfordomesticpurposes.

HarmonicemissionsIEC61000-3-2 ClassA

Voltage fluctuations/FlickeremissionsIEC61000-3-3

Complies

Guidance and manufacture’s declaration – electromagnetic immunityThe CLESTA Unit is intended for use in the electromagnetic environment specified below. The customer or the useroftheCLESTAUnitshouldassurethatitisusedinsuchanenvironment.

Immunity test IEC 60601test level Compliance level Electromagnetic environment-

guidanceElectrostaticdischarge(ESD)IEC61000-4-2

±6kVcontact±8kVair

±6kVcontact±8kVair

Floorsshouldbewood,concreteorceramic file. If floors are covered withsyntheticmaterial,therelativehumidityshouldbeatleast30%.

Electricalfasttransient/burstIEC61000-4-4

±2kVforpowersupplylines±1kVforinput/outputlines

±2kVforpowersupplylines±1kVforinput/outputlines

Mainspowerqualityshouldbethatofatypicalcommercialorhospitalenvironment.

SurgeIEC61000-4-5

±1kVdifferentialmode±2kVcommonmode

±1kVdifferentialmode±2kVcommonmode

Mainspowerqualityshouldbethatofatypicalcommercialorhospitalenvironment.

Voltagedips,shortinterruptionsandvoltagevariationsonpowersupplyinputlinesIEC61000-4-11

<5%UT

(>95%dipinUT)for0.5cycle40%UT

(60%dipinUT)for5cycle70%UT

(30%dipinUT)for25cycle<5%UT

(>95%dipinUT)for5s

<5%UT

(>95%dipinUT)for0.5cycle40%UT

(60%dipinUT)for5cycle70%UT

(30%dipinUT)for25cycle<5%UT

(>95%dipinUT)for5s

Mainspowerqualityshouldbethatofatypicalcommercialorhospitalenvironment.IftheuseroftheCLESTAUnitrequirescontinuedoperationduringpowermainsinterruptions,itisrecommendedthattheCLESTAUnitbepoweredfromanuninterruptiblepowersupplyorabattery.

Powerfrequency(50/60Hz)magnetic fieldIEC61000-4-8

3A/m 3A/m Power frequency magnetic fields shouldbeatlevelscharacteristicofatypicallocationinatypicalcommercialorhospitalenvironment.

NOTEUTisthea.c.mainsvoltagepriortoapplicationsofthetestlevel.

-30-

Page 36: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

Guidance and manufacture’s declaration – electromagnetic immunityThe CLESTA Unit is intended for use in the electromagnetic environment specified below. The customer or the useroftheCLESTAUnitshouldassurethatitisusedinsuchanenvironment.

Immunity test IEC 60601 test level Compliance level Electromagnetic environment -guidance

PortableandmobileRFcommunicationsequipmentshouldbeusednoclosertoanypartoftheCLESTAUnit,includingcables,thantherecommendedseparationdistancecalculatedfromtheequationapplicationstotheFrequencyofthetransmitter.

Recommended separation distance

ConductedRFIEC61000-4-6

3Vrms150kHzto80MHzoutsideISMbandsa

3Vrms d = 1.2√P

RadiatedRFIEC61000-4-3

3V/m80MHzto2.5GHz

3V/m d = 1.2√P 80MHzto800MHzd = 2.3√P 800MHzto2.5GHz

WherePisthemaximumoutputpowerratingofthetransmitterinwatts(W)accordingtothetransmittermanufactureranddistherecommendedseparationdistanceinmetres(m).

Field strengths from fixed RF transmitters, as determinedbyanelectromagneticsitesurvey,ashouldbelessthanthecompliancelevelineachfrequencyrange.b

Interferencemayoccurinthevicinityofequipmentmarkedwiththefollowingsymbol:

NOTE1At80MHzand800MHz,thehigherfrequencyrangeapplies.NOTE2Theseguidelinesmaynotapplyinallsituations.Electromagneticpropagationisaffectedby adsorption and reflection from structures, objects and people.

a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobileradios,amateurradio,AMandFMradiobroadcastandTVbroadcastcannotbepredictedtheoreticallywith accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the CLESTA Unit isusedexceedstheapplicableRFcompliancelevelabove,theCLESTAUnitshouldbeobservedtoverifynormaloperation.Ifabnormalperformanceisobserved,additionalmeasuresmaybenecessary,suchasreorientingorrelocatingtheCLESTAUnit.

b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.

Essential performance (purpose of IMMUNITY testing)Unless operated by the switches for chair control, the chair connected to the CLESTA does not make any movements, except for sounding a buzzer and switching on/off the indicator.

-31-

Page 37: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

Recommended separation distances betweenPortable and mobile RF communications equipment and the CLESTA Unit

TheCLESTAUnitisintendedforuseinanelectromagneticenvironmentinwhichradiatedRFdisturbancesarecontrolled.ThecustomerortheuseroftheCLESTAUnitcanhelppreventelectromagneticinterferencebymaintainingaminimumdistancebetweenportableandmobileRFcommunicationsequipment(transmitters)andtheCLESTAUnitasrecommendedbelow,accordingtothemaximumoutputpowerofthecommunicationsequipment.

Rated maximum output power of transmitter

W

Separation distance according to frequency of transmitterm

150 kHz to 80 MHzd = 1.2√P

80 MHz to 800 MHzd = 1.2√P

800 MHz to 2.5 GHzd = 2.3√P

0.01 0.12 0.12 0.230.1 0.38 0.38 0.731 1.2 1.2 2.3

10 3.8 3.8 7.3100 12 12 23

Fortransmittersratedatamaximumoutputpowernotlistedabove,therecommendedseparationdistancedinmetres(m)canbeestimatedusingtheequationapplicabletothefrequencyofthetransmitter,wherePisthemaximumoutputpowerratingofthetransmitterinwatts(W)accordingtothetransmittermanufacturer.NOTE1At80MHzand800MHz,theseparationdistanceforthehigherfrequencyrangeapplies.NOTE2Theseguidelinesmaynotapplyinallsituations.Electromagneticpropagationisaffectedbyadsorption and reflection from structures, objects and people.

-32-

Page 38: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

[7] LIST OF COMPATIBLE HANDPIECES

MODEL

SYRINGELUZZANI(3-way) Minilight w/Light

DCI(3-way)

TURBINEBIEN AIR BORA S36L / UNIFIX with LIGHT

NSK Ti-Max X

AIR MOTORBIEN AIR Aquilon 830 / UNIFIX with LIGHT/PM1132

NSK EX-203/EX-6

SCALER SATELEC SP4055 w/Light

-33-

Page 39: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS
Page 40: DENTAL UNIT INSTALLATION and OPERATING INSTRUCTIONS

2-1-1,Higashishinsaibashi,Chuo-ku,Osaka,JapanTEL : (06) 6213-5945 FAX : (06) 6212-3680

TAKARA BELMONT CORPORATION

NOTE

Takara Belmont (UK) Ltd.Belmont HouseOne St.Andrews Way,Bow,London E3 3PA U.K.Tel: (44)20 7515 0333Fax:(44)20 7987 3596

EC REP

Book No.FEFA25H0Printed in JAPAN 2012-04