DENTAL CHAIR
OPERATING INSTRUCTIONS
IMPORTANT
This manual provides operating instructions for the CLESTA II. The instructions contained in this booklet should be thoroughly read and understood before operating the chair. After the installation is completed, keep this manual in a safe place and referto it for future maintenance.

Page
SAFETYPRECAUTIONS---------------------------------------------------------- 11.OVERVIEWANDMAJORCOMPONENTS--------------------------------- 4 2.DIMENSIONSANDSPECIFICATIONS-------------------------------------- 4 3.OPERATINGINSTRUCTIONS 3-1.MAINSWITCH-------------------------------------------------------------- 5 3-2.CONTROLS------------------------------------------------------------------ 5 3-3.SAFETYLOCKDEVICE-------------------------------------------------- 6 3-4.HEADRESTOPERATIONS----------------------------------------------- 6 4.AUTOMODEPOSITIONADJUSTMENT----------------------------------- 6 5.CAREANDMAINTENANCE-------------------------------------------------- 7
6.MAINTANANCEANDINSPECTION6-1.GUIDEFORDAILYMAINTENANCEANDINSPECTION-------- 8(Maintenanceandinspectionbyuser)6-2.GUIDEFORPERIODICALCHECK-UP-------------------------------- 9
7.BEFOREASKINGFORREPAIRS--------------------------------------------- 9
8.ELECTROMAGNETICCOMPATIBILITY(EMC)--------------------------10
TABLE OF CONTENTS
Chair last position Chair auto return Chair preset1 Chair preset2
To raise the chair To Recline the backrest
Non-ionizingradiation
Date of manufacture
Authorized representative in the European community
Separate collection for electrical and electronic equipment
Manufacturer
Type B Applied Parts
CautionIt means “caution, warnings, or possibility to danger”.
To lower the chairTo raise the backrest
Chair manual control
Refer to instructionmanual/booklet
Alternating current Protective earth (ground)
Functional earth (ground) SN Serial number
SYMBOLS
Following symbols are used in this manual of CLESTA II CHAIR. Confirm the meaning of eachsymbol.

Intended Use of the ProductThisproductis intendedfortheexclusiveusefordiagnoses,treatmentsandrelativeproceduresof dentistry, and must be operated or handled by the qualified dentists or by dental staffs under thesupervisionofthedentist.Suchdentistsordentalstaffsshouldinstructand/orassistthepatientstoapproachtoandleavefromtheproduct.Patientsshouldnotbeallowedtooperateorhandletheproductunlesshe/sheissoinstructed.
Environmental RequirementsAmbientTemperatureOperating+5 -+40 Storage-10 -+50Humidity10%-80%AtmosphericalPressure600hPa-1060hPa
℃℃ ℃ ℃
Important NotesIn case of the troubles, please contat Takara Belmont offices or your dealers.
Donotdisassembleorattempttorepair.
Disassembly, repair or modifications shoud only be done by a qualified repair technician.
Attempts at disassembly, repair or modifications may lead to abnormal operation and accidents.
In case of disposal of equipmentWhendisposingthechair,appropriatelydisposecomplyingwithallcurrentapplicableregulationsand
localcodes.
InEUarea,EUdirective2002/96/EConwasteelectricalandelectronicequipment(WEEE)isapplied
onthisproduct.Inthisdirective,environmentconsciousrecycling/abandonmentisobligated.

■Beforeuse,readthe“Safetyprecautions”carefullytoensureproperuse.■Thefollowinginformationisdesignedtoensuresafeuseofthisproductandtopreventinjuryanddamage to you and others. The precautions contained here are classified depending on the severity and degree ofimminenceofpossibleinjuryordamageresultingfromimproperuse.Besuretofollowalltheinformation,whichisimportantforsafety.
SAFETYPRECAUTIONS
Thissymbolindicatesthat“itisrecommendedtofollowtheseprecautionsforsafety.”
Classification of precautions Severityanddegreeofimminenceofpossibleinjuryordamage
WARNING
CAUTION
NOTICE
Thissymbolindicatesthat“ignoranceoftheseprecautionsmayleadtosevereinjuryorevendeathasaresultofimproperuse.”
Thissymbolindicatesthat“ignoranceoftheseprecautionsmayleadtomildormoderatephysicalinjuryordamagetopropertyasaresultofimproperuse.”
WARNING1.Besuretoturnoffbreakersforequipmentintheclinicwhenthisproductwillnotbeusedforalongperiodoftime
Besuretoturnoffbreakersforequipmentintheclinicwhenthisproductwillnotbeusedforalongperiodoftime(followingthecompletionofwork,duringthesuspensionofbusiness,etc.).Insulationdegradationmaycause electrical fire.
2.BesuretoturnoffthemainswitchuponcompletionofworkorduringworkbreaksBesure to turnoff themainswitchuponcompletionofworkorduringworkbreaks.Thisprevents incorrectoperationduetoaccidentalcontactandassociatedhazards.
4.DonotplaceanundueloadonthearmDonotgetonorplaceanundueloadonthedentalchairarmrest.Thiscouldcausetheunittotoppleorotheraccidents.
7.Neverdisassemble,repairormodifythisproductIndividuals other than certified repair technicians should not disassemble or attempt to repair and modify this product. This could lead to an accident, failure, electric shock or fire.
3.DonotsitonotherthanseatWhenthebackrestisattheforwardposition.donotsitonorplaceanundueloadontheheadrestorlegreatofdentalchair.Thiscouldcausetheunittotoppleorcoulddamagetheunit.
8.UsewithcautioninthepresenceofelectromagneticinterferencewavesDonotplace thisproductaroundequipmentgeneratingelectromagneticwaves(includingcommunicationsequipment,elevators,etc.)asincorrectoperationofthisproductmayoccurinthepresenceofelectromagneticinterferencewaves.Donotuseequipmentgeneratingelectromagneticwaves,suchasmobilephones,aroundthisproduct.
5.ConnecttoanappropriatepowersourceThisequipmentmustbeconnectedtoanappropriatepowersource.
6.BesuretoestablishagroundingconnectionBesure toestablishapropergroundingconnection.(Refer toavendorforgroundingconnection.)Failureorelectricleakagemayleadtoelectricshock.
-1-

WARNING9.Besuretoturnoffthemainswitchwhenelectrocauteryisinuse
Besuretoturnoffthemainswitchwhenelectrocauteryisinuse,becausenoisemaycauseincorrectoperationofthisproduct.
•Failuretomaintainthisproductmayleadtophysicalinjuryorpropertydamage.•Refertothesectionofmaintenance.
10.Ensurethemaintenanceofthisproduct
11. Immediately wipe off any water spills or leakage on the floor
Immediately wipe off any water spills or leakage on the floor. Decreased strength of the floor may lead tophysicalinjuryincludingfall,orpropertydamage.
12.UsewithcautiononpatientswithacardiacpacemakerUsethisproductwithextremecautiononpatientswithacardiacpacemaker.Inthecaseofanyabnormalitiesinpatientsduringuse,immediatelyturnoffthisproductanddiscontinueuse.
CAUTION1.Onlyexperiencedpersonnelshouldusethisproduct
Onlydentistsorotherdentalprofessionalsshouldusethisproduct.
2. Confirm safety before use.Before use, confirm that the parts are correctly and safely operating and that there are no obstacles around this product.
4.Discontinueuseifyoufeelthat“somethingiswrong”
Keepyoureyesonpatients(especially,children)sothatmischieforinadvertentoperationofequipmentwillnotleadtounexpectedaccidents.
5.keepyoureyesonthepatientduringoperation.
3.Payattentiontopatientsandchildren
Alwaysbecarefultoinspectthisproductforlooseness,rattling,tilting,wobbling,sounds,temperature,odors,etc. Immediately discontinue use at the first feeling that “something is wrong.”
7.PrecautionsforcleaningtheresincoverForcleaning,donotusecleaningagentscontainingsolventorabrasives,thinnersoroil-basedalcohol(butanolandisopropylalcohol),whichmaycausecracks.
6.DonotsmackorrubthisproductDonotsmackorrubthisproductforcefully.Thiscouldcausedamagetocoversordefectivefunction.
•Confirm that the patient is seated in the proper position. Keep your eyes on the patient during the operation.•Payspecialattentiontosurroundingsatautomaticoperationofthedentaltreatmenttable.Damagetothebackrest,stoolorDoctor'stablemayoccur.
SAFETYPRECAUTIONS
-2-

CAUTION
9.Payattentionduringtheheadrestoperation
Besuretooperateswitcheswithyourhands,except thefootcontroller,whichisoperatedwithyourfoot.Operationwithbodypartsotherthanhandsmaycausedamageorincorrectoperation.
11.ReadthedocumentsaccompanyingthevariouspiecesofequipmentBeforeuse,besure tocarefullyreadthepackageinsertsandInstructionManualsaccompanyingthevariouspiecesofequipment(includingoptionalarticles)toensureproperuse.
10.Precautionsforcleaning•Neverusesandpaper,metalscrubbrushesandabrasivecleaningagentstocleantheunit.•Donotusestronglyacidiccleaningagentsoralkalinepipecleaningagentstoavoidcorrosionofmetals,etc.
Do not allow hands, fingers, or hair to become entangled in the moving parts of the headrest during operationg
8.Besuretooperateswitcheswithyourhands
SAFETYPRECAUTIONS
NOTICE1.Troubleshootingandcontactinformation
Inthecaseofanyproblems,discontinueuse,turnoffthemainswitchandcontactthedealerorourcompany.
2.HandlingofequipmentinthecaseofapowerfailurePutthehandpieceintheholderandturnoffthemainswitchifequipmentstopsworkingduringuseduetoapowerfailureorotherreasons.
-3-

-4-
1. OVERVIEW AND MAJOR COMPONENTS
Fig.1-1OverallView
(1)Headrest(2)Backrest(3)BackSupport(4)Armrest(Left)(5)Seat(6)Flange(7)LowerFlangeCover(8)FlangeHoseCover(9)Base(10)PumpCover(11)HeadrestRearCover(12)BackrestRearCover(13)BackSupportCover(14)UpperFlangeCover(15)MainSwitchPanel(16)FootControl
2
4
3
1
567
8916
10
11
12
13
14
7
1615
2. DIMENSIONS AND SPECIFICATIONS
2-1.DIMENSIONS-mm-
Fig.2-1Dimensions2-2.SPECIFICATIONS
SeatInitialHeight----------------------------------------------------- 410mmSeatLiftingStroke----------------------------------------------------- 380mmBackrestMovement--------------------------------------------------- 0˚ ~ 80˚ above HorizontalTiltingMechanism----------------------------------------------------- Backrest Synchronized Tilting (5 ̊~ 11 ̊above Horizontal)AutoMovements ----------------------------------------------------- 2-Preset,1-LastPositionMemoryand1-AutoReturnControlVoltage ----------------------------------------------------- DC.12VPowerConsumption--------------------------------------------------- AC230V,50Hz2.0AFuse--------------------------------------------------------------------- 5A/250V(CurrentRating:50Aat250VAC)Fast-blowWeight------------------------------------------------------------------- 115KgMaximumLoad-------------------------------------------------------- 135kgClassofFootSwirch-------------------------------------------------- IPX1(applicablestandars,IEC60529)AppliedParts----------------------------------------------------------- TypeBappliedpart:SeatforchairModeofOperation---------------------------------------------------- Non-ContinuousOperation/ONtime:3min,OFFtime:15minServiceLife------------------------------------------------------------- 10years
50˚ 8˚
80˚
1105
410
178
11˚5˚
565 1188 601813
6025
5
600630
510225
610
170

CAUTION
-5-
Fig.3-1MainSwitch
Fig.3-2FootControl
1 2
0 LP
Auto ReturnLast Position Memory
Preset-1 Preset-2Foot Switch Disc
Main Switch
ON
OFFMAINS
FUSE
I
O
3. OPERATING INSTRUCTIONS3-1.MAINSWITCH(Fig.1-1&Fig.3-1)Turnonthemainswitchlocatedontheleftsideofthepumpcover.Agreenlampinthemainswitchwillilluminate.
Operatethemainswitchbyhandonly.Turnoffthemainswitchafterdailyoperation.
3-2.CONTROLS(Fig.3-2) Before operating the chair, confirm that it is safe for the patientandtheoperator.
Allchairelectricalmovementscanbecontrolledbythefootswitch.
(1)ManualModeControlA.SeatLifting Keepdepressing()sideofthefootcontroldiscuntiltheseatislifteduptothedesiredposition.B.SeatLowering Keepdepressing()sideofthefootcontroldiscuntiltheseatisloweredtothedesiredposition.C.BackrestRecliningKeepdepressing()sideofthefootcontroldiscuntilthebackrestisreclinedtothedesiredposition.D.BackrestRaising Keepdepressing()sideofthefootcontroldiscuntilthebackrestisraseduptothedesiredposition.
(2)AutoModeControlE.PresetControl CLESTAIIchairhastwopresetpositions.(Preset-1andPreset-2) Momentarilydepress(1)buttononthefootcontrol,thechairwillmovetothepreset-1positionautomatically.(Preset-2isoperatedby(2)button.)F.AutoReturn Momentarilydepress(0)buttononthefootcontrol,thechairwillreturntotheinitialposition. (Theseatisfullyloweredandthebackrestisintheuprightposition.),G.LastPositionMemory Momentarilydepress(LP)buttonatthereclinedbackrestposition(treatmentposition),thebackrest willraisetothemouthrinsingpositionautomatically. Momentarilydepress(LP)buttonagain,thebackrestwillreclinetotheprevioustreatment positionautomatically.Note:Seatheightcannotbechangedby(LP)button.H.EmergencyStopDuringautomaticprocedure(Preset,AutoreturnandLastpositionmemory),depressofanysideofthediscorbuttonsonthefootcontrolwillcanceltheautomaticmovementimmediately.Note:Donotdepressautomodebutton(1)(2)(0)(LP)forover3seconds,becausethe memorizedpositioninautomodemaybechanged.
CAUTION

Fig.3-2FootControl
3-3.SAFETYLOCKDEVICE(Fig.3-3) Allchairmovementscanbestoppedautomatically bythesafetylockdevicewhenpressureisdetected betweenthebaseandthesublinkcover. Ifthesafetydevicehasbeenactivated,simplyoperatethebaseupbuttonandremovetheobjectcausingthesafetydevicetoactivatefromthisarea. Note:Seatliftingandbackrestrasingcanbe operatedwhenthesafetylockdeviceis activated.
3-4.HEADREST(Fig.3-4)(1)HeightAdjustment Pressdownorpulluptheheadresttothedesiredheight.(2)AngleAdjustment Pushtheheadrestforwardasrequired. Lifttheheadrestlevertorotatebackwardandreleasetheleveratthedesiredangle.
4. AUTO MODE POSITION ADJUSTMENT (1)PresetpositionAdjustment(Fig.4-1) Twopresetpositionscanbeset. A.Settheseatandthebackresttothedesiredpresetpositionbymanualcontrolswitch. B.Keepdepressingpreset1switch(1)untilbuzzersounds(inabout3seconds),thenreleaseit. C.ThepositionismemorizedforPreset-1. D.Preset-2canbememorizedbydepressingpreset2switch(2),asfollowingAtoC.
Fig.3-3SafetyLockDevice
Fig.3-4Headrest
Fig.4-1FootControl
Safety LockDevice
Sub Link Cover
Base
(2)MouthRinsingPositionAdjustment(Fig.4-1) Mouthrinsingpositioninlastpositionmemorymovementcanbeadjusted. A.Setthebackresttothedesiredmouthrinsingpositionbymanualcontrolswitch. B.Keepdepressinglastpositionmemoryswitch(LP)untilbuzzersounds(inabout3seconds) andreleaseit. C.Thisbackrestpositionisthenmemorizedasthemouthrinsingposition.
HeightAdjustment
Headrest Lever
AngleAdjustment
3-5.DOUBLEARTICULATINGHEADREST(Optional)(Fig.3-5)(1)HeightAdjustment Pressdownorpulluptheheadresttothedesiredheight.(2)AngleAdjustment Angleofheadrestcanbechangedbygraspingtheheadrestreleaseleveronheadrestmechanism.
Fig.3-5DoubleArticulatingHeadrest
Headrest Release�Lever
1 2
0 LP
Auto ReturnLast Position Memory
Preset-1 Preset-2
Manual Control Switch
-6-

CAUTION
TurnOFFthemainswitchatthelowestseatpositionafterdailyoperationandforalongterminterval.
CAUTION
5. CARE AND MAINTENANCE
Chair CleaningThesurfaceofthechair’sseatingareaismadeofsyntheticleather.Applydrywipingorwipethesurfacewith
clothmoistenedwitheitherwaterordilutedneutraldetergentforthecare.
Ifthecolorofclothingorbeltremainedonthesyntheticleather,wipeitoffwithclothmoistenedwithdiluted
neutraldetergentassoonaspossible,toavoiditspenetrationcausedbyplasticizer.
Incasethesyntheticleatheriswipedwithawetcloth,fullywipeoffthemoisture.Ifitremains,hydrolytic
degradationmaybeaccelerated.Donotusesolventorbleach.
Applydrywipingusingadryandsoftclothtometallicareas.Ifanymetallicareaiswetted,wipeoffthemoisture
assoonaspossible.Itwillrustotherwise.
Wipetheresincoverwithawetandsoftcloth.
Chair SterilizationUseFD333madebyDurrsprayedtoasoftclothorpapertowelforcleaningandsterilizationoftheproductexterior.Donotoperatetheproductuntiltheliquidusedforsterilizationhasfullydriedup.
Donotplaceanyhardandheavyarticleoranyarticlehavingasharptiporedgeonthechair’sseatingarea.Thesyntheticleathermaybedamagedotherwise.Forcleaningtheresincover,donotusesolventordetergentcontainingabrasiveagent.Inaddition,donot use any chemical that is other than specified for this unit. The resin cover may crack otherwise.Useasoftclothmoistenedwitheitherwaterordilutedneutraldetergent.Neverusedetergentofstrongacidityoralkalipipedetergent,ormaterialsincludingmetalmaybecorroded.
-7-

6. MAINTANANCE AND INSPECTION
6-1. Guide for daily maintenance and inspection (Maintenance and inspection by user)
Managementofmaintenanceand inspectionofmedicalequipment shouldbe implementedby theuser (medicalinstitution).Incasetheuserdoesnotimplementsuchmanagement,itispermittedthatsuchmanagementisoutsourced to a qualified entity such as a medical equipment repair company. For safe use of this product, it is necessary that inspection should be conducted in the specified frequency on the itemsdescribedbelow.
No. Item Frequency Inspection method and diagnosis
Influence if inspection not conducted
Maintenance required in case of nonconformity
Check of safety functions
Before start
After closing
Unexpected personal injury and troubles may arise due to motion of the chair during medical treatment and due to pinching between doctor section and chair.
Contact your dealer or our office if any abnormality arises.
Call to a technician of ourauthorized dealer.
During pre-set movement orauto-return movement, press any key for chair operation and the chair should automatically stop.
Power supplycable
Before start
Power supply cable should not be abnormally bent, pintched or damaged, and the plug should be firmly inserted to the poweroutlet and not covered with dust.
Accident or defect of the equipment
Clean and rearrange thecable. For replacement, contact your dealer or our office.
Function of eachswitch
Before start
Each switch for chair operationshould work correctly.
Defect of movement may cause an accident.
Contact your dealer or our office if any abnormality arises.
Moving parts
Before start
Make sure that the chair does not make any abnor-mal noise during automatic and manual movement
Accidend or defect of the equipment Contact your dealer or our
office if any abnormality arises.
Oil leakage
Before start
Make sure that there is no hydraulic oil on the floor or chair base leaking from the cylinder.
Defect of the equipment Contact your dealer or our office if any abnormality arises.
Wiping off water
After closing
Make sure that no water is left that were spilt during treatment.
Water will cause rust Wipe off water with a softand dry cloth.
Stain removal ofexternal parts
After closing
Visual check the attach-ment of dust, chemicals or drug and grasp the total product condition
Discoloration, transfor-mation or breakage of resin parts may arise.
Clean up according to the “CARE AND MAINTANANCE “ in operating instructions.
Main powerswitch
Make sure that the main power switch is off
Accident or defect of the equipment
In the case the main power cannot be turned off, contact your dealer or our office.
-8-

Phenomenon Check point and result Action to be taken
The product does not work at all.
Equipment circuit breaker in the clinic cabinet panel is not on.
Main switch is not on. Turn on the main switch.
Turn on the equipment circuit breaker.
The chair does not work.
Locking device for motion stop is activated. Unlock the device. See section for SAFETY LOCK DEVICE.
7. BEFORE ASKING FOR REPAIRSIfanyofphenomenadescribedbelowhasoccurred,makethefollowingchecksbeforeaskingforrepairs.
If theunitdoesnotnormallyworkeven ifactionswere takenuponcheckupstatedabove, thenstopusing the unit, turn off the main switch and contact your dealer or our office.
6-2. Guide for Periodical Check-up
Somepartsandcomponentsoftheproductsaredegradedordeteriorateddependingonthefrequencyof
use.Annualcheck-upandmaintenance,aswellasreplacementofconsumableparts,arerequired.
Therequiredparts(includingconsumableparts)arelistedbelow.Itmaybedifferentfromthefollowing
listdependingontheoptionoftheunit.
Forcheck-upandrepair,callatechnicianofourauthorizeddealer.
Executethemaintenanceinaccordancewiththisinstractionmanualandoperatingmanualattachedtoeachindividualequipment(Dentallight,Handpiece,etc..).Failuretomaintainthisproductmayleadtophysicalinjuryorpropertydamage.
WARNING
Parts and components that require periodical check-up
Parts Description Standard LifetimeNo.
7 years 1
2 5 years
3 5 years
5 years4
Consumable parts
No. Parts Description
1 O-ring, Packing, Diaphragm
Moving part
Electric wiring of moving parts
Switches
Control PCBs
-9-

Medicalelectricalequipmentneedsspecialprecautions regardingEMCandneeds tobe installedandput intoserviceaccordingtotheEMCinformationprovidedinthismanual.PortableandmobileRFcommunicationsequipmentcanaffectmedicalelectricalequipment.Theequipmentorsystemshouldnotbeusedadjacent toorstackedwithotherequipment. Ifadjacentorstackeduse isnecessary, the equipment or system should be observed to verify normal operation in the configuration in which it will be used.
Guidance and manufacture’s declaration - electromagnetic emissionsTheCLESTAII(CHAIR)is intended for use in the electromagnetic environment specified below. The customer or the useroftheCLESTAII(CHAIR)shouldassurethatitisusedinsuchanenvironment.
Emissions test Compliance Electromagnetic environment -guidanceRFemissionsCISPR11
Group1
TheCLESTAII(CHAIR)usesRFenergyonlyforitsinternalfunction.Therefore,itsRFemissionsareverylowandarenotlikelytocauseanyinterferenceinnearbyelectronicequipment.
RFemissionsCISPR11 ClassB
TheCLESTAII(CHAIR)issuitableforuseinallestablishments,includingdomesticestablishmentsandthosedirectlyconnectedtothepubliclow-voltagepowersupplynetworkthatsuppliesbuildingsusedfordomesticpurposes.
HarmonicemissionsIEC61000-3-2 ClassAVoltage fluctuations/FlickeremissionsIEC61000-3-3
Complies
Guidance and manufacture’s declaration - electromagnetic immunityThe CLESTA II (CHAIR) is intended for use in the electromagnetic environment specified below. The customer or the useroftheCLESTAII(CHAIR)shouldassurethatitisusedinsuchanenvironment.
Immunity test IEC 60601test level Compliance level Electromagnetic environment -
guidanceElectrostaticdischarge(ESD)IEC61000-4-2
±6kVcontact±8kVair
±6kVcontact±8kVair
Floorsshouldbewood,concreteorceramic file. If floors are covered withsyntheticmaterial,therelativehumidityshouldbeatleast30%.
Electricalfasttransient/burstIEC61000-4-4
±2kVforpowersupplylines±1kVforinput/outputlines
±2kVforpowersupplylines±1kVforinput/outputlines
Mainspowerqualityshouldbethatofatypicalcommercialorhospitalenvironment.
SurgeIEC61000-4-5
±1kVdifferentialmode±2kVcommonmode
±1kVdifferentialmode±2kVcommonmode
Mainspowerqualityshouldbethatofatypicalcommercialorhospitalenvironment.
Voltagedips,shortinterruptionsandvoltagevariationsonpowersupplyinputlinesIEC61000-4-11
<5%UT
(>95%dipinUT)for0.5cycle40%UT
(60%dipinUT)for5cycle70%UT
(30%dipinUT)for25cycle<5%UT
(>95%dipinUT)for5s
<5%UT
(>95%dipinUT)for0.5cycle40%UT
(60%dipinUT)for5cycle70%UT
(30%dipinUT)for25cycle<5%UT
(>95%dipinUT)for5s
Mainspowerqualityshouldbethatofatypicalcommercialorhospitalenvironment.IftheuseroftheCLESTAII(CHAIR)requirescontinuedoperationduringpowermainsinterruptions,itisrecommendedthattheCLESTAII(CHAIR)bepoweredfromanuninterruptiblepowersupplyorabattery.
Powerfrequency(50/60Hz)magnetic fieldIEC61000-4-8
3A/m 3A/m Power frequency magnetic fields shouldbeatlevelscharacteristicofatypicallocationinatypicalcommercialorhospitalenvironment.
NOTEUTisthea.c.mainsvoltagepriortoapplicationsofthetestlevel.
8. ELECTROMAGNETIC COMPATIBILITY
-10-

Guidance and manufacture’s declaration - electromagnetic immunityTheCLESTAII(CHAIR)is intended for use in the electromagnetic environment specified below. The customer or the useroftheCLESTAII(CHAIR)shouldassurethatitisusedinsuchanenvironment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment -guidance
PortableandmobileRFcommunicationsequipmentshouldbeusednoclosertoanypartoftheCLESTAII(CHAIR),includingcables,thantherecommendedseparationdistancecalculatedfromtheequationap-plicationstotheFrequencyofthetransmitter.
Recommended separation distanceConductedRFIEC61000-4-6
3Vrms150kHzto80MHzoutsideISMbandsa
3Vrms d = 1.2√P
RadiatedRFIEC61000-4-3
3V/m80MHzto2.5GHz
3V/m d = 1.2√P 80MHzto800MHzd = 2.3√P 800MHzto2.5GHz
WherePisthemaximumoutputpowerratingofthetransmitterinwatts(W)accordingtothetransmittermanufactureranddistherecommendedseparationdistanceinmetres(m).
Field strengths from fixed RF transmitters, as determinedbyanelectromagneticsitesurvey,ashouldbelessthanthecompliancelevelineachfrequencyrange.b
Interferencemayoccurinthevicinityofequipmentmarkedwiththefollowingsymbol:
NOTE1At80MHzand800MHz,thehigherfrequencyrangeapplies.NOTE2Theseguidelinesmaynotapplyinallsituations.Electromagneticpropagationisaffectedbyadsorptionand reflection from structures, objects and people.a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobileradios,amateurradio,AMandFMradiobroadcastandTVbroadcastcannotbepredictedtheoreticallywith accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the CLESTA II (CHAIR)isusedexceedstheapplicableRFcompliancelevelabove,theCLESTAII(CHAIR)shouldbeobservedtoverifynormaloperation.Ifabnormalperformanceisobserved,additionalmeasuresmaybenecessary,suchasreorientingorrelocatingtheCLESTAII(CHAIR).
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.
-11-
Essential performance (purpose of IMMUNITY testing)Unless operated by the switches for chair control, the CLESTA II (CHAIR) does not make any movements, except for sounding a buzzer and switching on/off the indicator.

Recommended separation distances betweenPortable and mobile RF communications equipment and the CLESTA II (CHAIR)
TheCLESTAII(CHAIR)isintendedforuseinanelectromagneticenvironmentinwhichradiatedRFdisturbancesarecontrolled.ThecustomerortheuseroftheCLESTAII(CHAIR)canhelppreventelectromagneticinterferencebymaintainingaminimumdistancebetweenportableandmobileRFcommunicationsequipment(transmitters)andtheCLESTAII(CHAIR)asrecommendedbelow,accordingtothemaximumoutputpowerofthecommunicationsequipment.
Rated maximum output power of transmitter
W
Separation distance according to frequency of transmitterm
150 kHz to 80 MHzd = 1.2√P
80 MHz to 800 MHzd = 1.2√P
800 MHz to 2.5 GHzd = 2.3√P
0.01 0.12 0.12 0.230.1 0.38 0.38 0.731 1.2 1.2 2.310 3.8 3.8 7.3100 12 12 23
Fortransmittersratedatamaximumoutputpowernotlistedabove,therecommendedseparationdistancedinmetres(m)canbeestimatedusingtheequationapplicabletothefrequencyofthetransmitter,wherePisthemaximumoutputpowerratingofthetransmitterinwatts(W)accordingtothetransmittermanufacturer.NOTE1At80MHzand800MHz,theseparationdistanceforthehigherfrequencyrangeapplies.NOTE2Theseguidelinesmaynotapplyinallsituations.Electromagneticpropagationisaffectedbyadsorptionandreflection from structures, objects and people.
-12-

2-1-1, Higashishinsaibashi,Chuo-ku,Osaka, 542-0083, JapanTEL : 81-6-6213-5945 FAX : 81-6-6212-3680
TAKARA BELMONT CORPORATION
NOTE
Printed in Japan 2012-04
Takara Belmont (UK) Ltd.Bemont HouseOne St.Andrews Way, Bow,London E3 3PA U.K.Tel : (44) 20-7515-0333Fax : (44)20-7987-3596 BOOK NO. AEFT02G0