![Page 1: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/1.jpg)
DEFIBRILLATOR MACHINE BAsic TECHNICAL
TRAINING
By: Armando Darino Ngojo Senior Biomedical Engr.
![Page 2: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/2.jpg)
ContentsAims & ObjectivesIntroductionPrincipleAnatomy and PhysiologyApplicationTypesOperationSafetyMaintenanceTroubleshootingPPM NGOJO
![Page 3: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/3.jpg)
Aims
To provide basic understanding about the Defibrillator Machine
To perform and understand the basic technical troubleshooting steps and PPM tasks
NGOJO
![Page 4: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/4.jpg)
Objectives
- as a result of completing this module, the biomedical technician / engineer should be able to:
understand the concept of the Defibrillator including its applications
perform and identify basic problems, errors and basic troubleshooting solutions.
Perform PPM tasks
NGOJO
![Page 5: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/5.jpg)
History1899 Prevost and Batelli first introduced the concept of electrical
fibrillation after noticing that large voltages applied across the animal's heart could convert ventricular fibrillation into a sinus rhythm.
1933
1950s
1956
Hooker, Kouwenhoven and Langworthy published an account of successful alternating current (AC) internal animal defibrillation
Kouwenhoven was able to defibrillate dogs by applying the electrodes to the chest wall, that was the external electric defibrillator.
Zoll defibrillated a human subject in the same manner .
NGOJO
![Page 6: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/6.jpg)
History
1960s
1970s
Edmark and Lown et al found that direct current (DC) or pulse defibrillators were more effective and produced fewer side effects than AC defibrillator. The DC pulse waveform was further improved.
Experimental internal and external devices were designed to automatically detect ventricular fibrillation.
1980s
Present times
The first automatic internal defibrillator was implanted in human
A lot of improvements were introduced to the defibrillator with the aim of improving the survival rate of the cardiac arrested patient
NGOJO
![Page 7: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/7.jpg)
What is Defibrillator
- A device that reverses the Fibrillation of the heart. Fibrillation causes the heart to stop
pumping blood, leading to brain damage.
NGOJO
![Page 8: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/8.jpg)
Defibrillation
Defibrillation is a process in which an electronic device sends an electric shock to the heart to stop an extremely rapid, irregular heartbeat, and restore the normal heart rhythm.
It is a common treatment for life threatening cardiac dysrhythmia, ventricular fibrillation, and pulse less ventricular tachycardia.
NGOJO
![Page 9: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/9.jpg)
Principle
A high voltage electric current is applied to the Heart muscle either directly (Internal Defibrillator) through the open chest or indirectly (External Defibrillator) through the chest wall to terminate Ventricular Fibrillation.
NGOJO
![Page 10: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/10.jpg)
Physiology of Defibrillation
If enough current is delivered to the heart then a majority of Ventricular cells will be depolarized
A critical mass of cells should be simultaneously depolarized to achieve defibrillation
NGOJO
![Page 11: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/11.jpg)
Bio Electric Signal
Polarization Potassium ions pumped in Sodium ions pumped outDepolarization Potassium ions pumped out Sodium ions pumped in Repolarization Potassium ions pumped in Sodium ions pumped out
NGOJO
![Page 12: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/12.jpg)
Anatomy - HeartLocated between the lungs in the middle of chest, behind and slightly to the left of breastbone (sternum)
Size of ones fist and shaped like a cone.An average heart pumps 2.4 ounces (70 milliliters) per heartbeat. An average heartbeat is 72 beats per minute. Therefore an average heart pumps 1.3 gallons (5 Liters) per minute.
NGOJO
![Page 13: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/13.jpg)
Cont…
Heart is divided into four chambers
Right AtriumLeft AtriumRight VentricleLeft Ventricle
A wall of muscle called as a 'Septum', separates all chambers of heart.
NGOJO
![Page 14: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/14.jpg)
ECG tracingElectrical pattern
Normal Cardiac Conduction
NGOJO
![Page 15: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/15.jpg)
Normal ECG tracing
NGOJO
![Page 16: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/16.jpg)
NGOJO
whenABNORMALITIEShappen….
![Page 17: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/17.jpg)
1. Ventricular Fibrillation
Ventricular Fibrillation is a very fast, irregular heart rhythm in the lower heart chambers (ventricles). During VF the heart quivers and pumps little or no blood to the body. Consciousness is lost in seconds. If not treated immediately, VF will cause sudden cardiac arrest
NGOJO
![Page 18: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/18.jpg)
Occlusion of the coronary artery leads to ischemia.
Ischemia leads to infarct which causes interruption of normal cardiac conduction
Infarct = VF/VT
Cardiac Arrest
NGOJO
![Page 19: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/19.jpg)
Ventricular Fibrillation Ventricular Tachycardia
SAMPLE SHOCKABLE RHYTHMS
NGOJO
![Page 20: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/20.jpg)
2. Atrial Fibrillation
Atrial Fibrillation is a very fast, irregular heart rhythm in the upper heart chambers. During AF, the waveform is almost similar to normal waveform but for the loss of 'P' waveform.
NGOJO
![Page 21: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/21.jpg)
AF treatment:
For Atrial fibrillation, the shock should be avoided to be delivered in the T period otherwise it will lead to Ventricular Fibrillation
This is achieved by Synchronous mode
NGOJO
![Page 22: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/22.jpg)
ApplicationEmergency department
Anesthesiology
Cardiology
Operation theatre
Intensive care areas
Ambulance services NGOJO
![Page 23: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/23.jpg)
Types:
NGOJO
![Page 24: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/24.jpg)
1. Manual
Clinical expertise is needed to interpret the heart rhythm and decide whether to charge the defibrillator and deliver the shock to patient. Energy selection and delivery is given to the patient manually.
NGOJO
![Page 25: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/25.jpg)
2. Automatic
These defibrillators are small, safe, simple and lightweight with two pads that can be applied to the patient. The defibrillator guides the operator step-by-step through a programmed protocol. It records and analyses the rhythm and instructs the user to deliver the shock using clear voice prompts, reinforced by displayed messages.
NGOJO
![Page 26: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/26.jpg)
3. External
External Defibrillator is the device which delivers the high energy shock to patients Heart externally on patient's chest by using a Defibrillator Paddle. The maximum energy deliver to the patient is about 360 Joules in Monophasic & 200 Joules in Biphasic Defibrillator.
NGOJO
![Page 27: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/27.jpg)
4. Internal
Internal defibrillator consist of sterilized internal Handle/Paddle through which shock is delivered directly to the heart.
NGOJO
![Page 28: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/28.jpg)
5. AED Automatic External DefibrillatorAED can be classified as either fully automatic or semiautomatic. In fully automatic models disposable paddles are kept connected to the patient whilst the AED analyzes the ECG rhythm , decides and determines whether a defibrillation counter-shock is needed. Then the device automatically charges and discharges.
Semi Automatic AED analyze the patient's ECG and notify the operator when defibrillation is indicated. The operator then activates defibrillator and discharge.
NGOJO
![Page 29: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/29.jpg)
AED
Semi Automatic Fully Automatic
NGOJO
![Page 30: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/30.jpg)
6. ICD
Implantable Cardioverter Defibrillator
If it detects an abnormally fast heart rhythm, it either electrically paces the heart very fast or delivers a small electrical shock to the heart to convert the heart rhythm back to normal. The rapid pacing is not felt by the patient but the electrical shock, if used, is felt as a strong jolt in the chest. The ICD is used to treat life-threatening heart rhythms that lead to sudden death.
NGOJO
![Page 31: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/31.jpg)
ICD
NGOJO
![Page 32: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/32.jpg)
7. Pacemaker
A pacemaker is an electronic stimulator that produces periodic electric stimulation to the heart It is classified into two types: Internal Pacemaker External Pacemaker
NGOJO
![Page 33: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/33.jpg)
Internal Pacemaker
Battery operated device that is implanted inside the patient’s body to pace the Heart
NGOJO
![Page 34: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/34.jpg)
External Pacemaker
This is used to pace the heart temporally. It may use leads or electrodes for the stimulation
NGOJO
![Page 35: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/35.jpg)
What is Joule?
It is the unit of energy delivered by the Defibrillator
It means - “The energy released in one second by a current of one ampere through a resistance of one ohm”
Also called as watt-second
NGOJO
![Page 36: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/36.jpg)
– intrinsic characteristics of patient – patient’s disease – duration of arrhythmia – patient’s age – type of arrhythmia (more energy required for
VF)– type of the machine used
- The delivered energy is in the range of 50-360 joules and depends on:
NGOJO
![Page 37: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/37.jpg)
Operating Principle
Monophasic Defibrillator
Bi- Phasic Defibrillator
NGOJO
![Page 38: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/38.jpg)
Monophasic DefibrillatorDelivers its current in one forward direction (positive)
Requires higher escalating energy levels (200-300J) to convert VF/ pulse-less VT
NGOJO
![Page 39: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/39.jpg)
Bi Phasic Defibrillator
The Biphasic waveform type defibrillator delivers the current in one direction during the first phase and in opposite direction during the second phase.
Biphasic waveform shocks of 200 J are safe, equivalent or higher efficiency than damped sinusoidal waveform shocks of 360 J.
NGOJO
![Page 40: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/40.jpg)
Generation of bi-phasic waveform
NGOJO
![Page 41: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/41.jpg)
Waveform Comparison
NGOJO
![Page 42: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/42.jpg)
Operation
NGOJO
Defibrillator: Normal and Synchronous Mode Monitor
Recorder
Pacemaker
NIBP
SPO2
CO2
![Page 43: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/43.jpg)
1. Manual Defibrillation
Switch 'ON' the Machine
Wait for initialisation and self test
Make sure it is not in SYNC Mode
Apply gel to the paddles
Place them properly on the chestNGOJO
![Page 44: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/44.jpg)
Cont…
Select 'ENERGY' to be delivered( energy in Joules)
Press 'CHARGE' button
Wait for Charging to complete. This is usually denoted by a continuous /long beep sound.
Apply pressure to the paddles ( 12 Kg of pressure)
NGOJO
![Page 45: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/45.jpg)
Cont…
Make sure all the personnel are away from the patientPress both 'DISCHARGE' button simultaneouslyObserve patient and monitor ECGIf required, defibrillate againWhen finished, turn off and clean the paddles
NGOJO
![Page 46: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/46.jpg)
2. Synchronization Mode
Wait for initialization and self test.
Connect ECG leads to get a tall R wave
Select 'SYNC / CARDIOVERSION' mode..
Check for sync marker on the QRS waveform.
If possible sedate the patient and maintain airway
NGOJO
![Page 47: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/47.jpg)
Cont…
Apply gel on the paddles and place it properly on chestSelect 'ENERGY' to be delivered( energy in Joules)Press 'CHARGE' buttonWait for the Charge to be completed. This is usually denoted by a continuous /long beep sound.Press both 'DISCHARGE' button simultaneously and hold till energy is delivered.
NGOJO
![Page 48: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/48.jpg)
Cont…
Check patient condition and Heart rhythm.
If required, defibrillate again.
Monitor the patient for few hours.
Switch off and clean the paddles
NGOJO
![Page 49: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/49.jpg)
3. Internal Defibrillation
Connect the Internal Paddles to the machine
Place one paddle over the apex of the left ventricles and the other over the base of the right ventricles
Switch on the machine Select Energy Charge and Discharge
NGOJO
![Page 50: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/50.jpg)
What are the types of Paddles?
NGOJO
![Page 51: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/51.jpg)
Paddle Placement
There are two notable methods of paddle placement recommended by AHA
Anterior- Anterior
Anterior-Posterior
NGOJO
![Page 52: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/52.jpg)
Paddle PlacementAnterior- Anterior Place one paddle near
the second or the third right sternal border and the other on the cardiac apex.
NGOJO
![Page 53: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/53.jpg)
Paddle Placement
Anterior-Posterior - One paddle on sternum and the other on the left infra-scapular region
NGOJO
![Page 54: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/54.jpg)
Important Factors: 1.TimeEarly defibrillation allows more success or the longer period of VF, the less success of defibrillation.
Early initiation of CPR improves the success rate> 8 mins neurological damage sets in> 10 mins survival probability becomes very low
NGOJO
![Page 55: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/55.jpg)
2. Energy Level
AHA Recommendation for Adults
First shock 200 jSecond shock 200 j to 300 j Third and above shocks 360 j
AHA Recommendation For PaedsFirst shock 2 joules per KgSubsequent shock double the energy
NGOJO
![Page 56: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/56.jpg)
3. Paddle Size
Adult paddles should be 8 to 13 cm in diameterChild paddles should be 4.5 cm in diameterInfants use Anterior Posterior position
NGOJO
![Page 57: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/57.jpg)
4. Skin To Paddle Interface
Use the right gel
Too little gel increases possibility of burn
Too much gel causes electric current to arc from one electrode to another
If disposable paddles are used check the expire date
NGOJO
![Page 58: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/58.jpg)
Block Diagram of a typical Defibrillator machine
NGOJO
![Page 59: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/59.jpg)
Power Supply
Step up transformers used to convert 240 VAC to 5000 VAC
This is converted to DC by rectifierIn battery mode the DC is converted to AC by inverter
This AC is amplified and then again rectified to DC.
NGOJO
![Page 60: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/60.jpg)
Capacitor Most important part of Defibrillator
Stores large energy which can be discharged in a short burst
Unit of measurement is Farad
NGOJO
![Page 61: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/61.jpg)
Inductor
Coils of wire that produce a magnetic field when current flow through them
Used to prolong the duration of current flow Unit of measurement is Henry
NGOJO
![Page 62: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/62.jpg)
Safety
NGOJO
![Page 63: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/63.jpg)
Safety - General
The Defibrillator generates High voltage. It must be operated by trained, professional and qualified personnel only.
Never use defibrillator with improper grounding or electrical leak socket.
Keep away the Defibrillator from any x-ray, Ultrasonic or other electronic instruments.
NGOJO
![Page 64: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/64.jpg)
Cont…
Check the patient lead wire, cable and paddles for any damage or mishandling, otherwise replace the same immediately.
Recommend using proper size and placement of recording paper.
Clean the print head regularly for clear printout
NGOJO
![Page 65: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/65.jpg)
Safety - Monitoring
Use only the specified patient cable as recommended by manufacturer.
Place the patient cable in proper winding position or hang on to patient cable arm attached with Defibrillator trolley
Use good ECG electrodes to monitor ECG waveform.
NGOJO
![Page 66: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/66.jpg)
Cont…
Don't use damaged patient cable.
Confirm there is no ECG waveform because of electrical interference or defective patient cable. This may misinterpreted as QRS in synchronize mode.
NGOJO
![Page 67: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/67.jpg)
Safety - Defibrillation
Excessive Gels can cause arcing of the current along the chest wall
Malfunction of permanent pacemakers can result from placing defibrillator pads or paddles near the pacemaker
Defibrillation in the absence of an ECG rhythm (ie, 'blind defibrillation') to be avoided
NGOJO
![Page 68: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/68.jpg)
Cont…
A shock can be accidentally delivered to other rescuersThe initial three shocks should be delivered in sequence, without interruption for CPR, medication administration, or pulse checks.If transthoracic impedance is high, a low energy shock (< 100 J) may fail to generate enough current to achieve successful defibrillation.
NGOJO
![Page 69: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/69.jpg)
Cont…
Alcohol should never be used as conducting material for paddles because serious burns can result.
Never discharge the Defibrillator in Air to check its performance
Never discharge with paddles shorted
Always clean the paddles after useNGOJO
![Page 70: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/70.jpg)
Maintenance - Cleaning
For Exterior surface of defibrillator and battery /battery well or compartment Isopropyl alcohol(70% in water) Mild soap and water Chlorine bleach (3% in water) While cleaning take care that solution should not go inside the defibrillator. Use soft cloth for cleaning display, to prevent scratching.
NGOJO
![Page 71: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/71.jpg)
Maintenance - Cleaning
For Paddles, pads, Electrodes & cables
Disposable pads & monitoring Electrodes do not require cleaning.The paddles, paddle cable and patient cable, may cleaned withAlcohol free hand soap Chlorine bleach solution 10%in water Isopropyl Alcohol
NGOJO
![Page 72: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/72.jpg)
Maintenance - Cleaning
Internal Paddles / Cables
Internal Paddle Handle/ Cable / Electrodes can be cleaned ultrasonically, autoclave or steam sterilized but follow the manufacturer's instruction for sterilization procedure
NGOJO
![Page 73: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/73.jpg)
The paddles used in the procedure should not be placed:-
• on a woman's breasts • over an internal pacemaker patients.
Before the paddle is used, a gel must be applied to the patient's skin
PRECAUTIONS
NGOJO
![Page 74: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/74.jpg)
• Skin burns from the defibrillator paddles are the most common complication of defibrillation.
• Other risks include injury to the heart muscle,
abnormal heart rhythms, and blood clots.
RISKS IN DEFIBRILLATION
NGOJO
![Page 75: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/75.jpg)
• Attach the external and internal paddles if the monitor reads, "No paddles."
• Check to ensure that the leads are securely attached if the monitor reads, "No leads.“
• Connect the unit to AC power if the message reads, "Low battery."
• Verify that the Energy Select control settings are correct if the defibrillator does not charge.
TROUBLESHOOTING
NGOJO
![Page 76: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/76.jpg)
• Change the electrodes and make sure that the electrodes adapter cable is properly connected if you receive a message of "PACER FAILURE." Restart the pacer.
• Close the recorder door and the paper roll if the monitor message reads, "Check recorder”.
TROUBLESHOOTING
NGOJO
![Page 77: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/77.jpg)
PPM (Planned Preventive Maintenance)
Qualitative Task
PPM Task
Quantitative Task
Electrical Safety Test
NGOJO
![Page 78: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/78.jpg)
Test tool: Defibrillator Analyzer
NGOJO
![Page 79: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/79.jpg)
Defibrillator Analyzer
Basic Functions
Measures output energy Measures cardioversion delay time Measures output of pacemakers Simulates range of ECG waveforms Provides clinical training
NGOJO
![Page 80: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/80.jpg)
Practical Training and
Assessment
NGOJO
![Page 81: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/81.jpg)
Questions!!!
NGOJO
![Page 82: DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING By: Armando Darino Ngojo Senior Biomedical Engr.](https://reader038.fdocuments.in/reader038/viewer/2022103005/56649d8c5503460f94a732bf/html5/thumbnails/82.jpg)
THANK
YOU NGOJO