Curve Ceiling Mounted Image Guided Surgery Technical Specifications

6
p. 1 of 6 PLATFORM 15700 | CURVE CEILING-MOUNTED DUAL DISPLAY Ceiling-mounted high performance Image-Guided-Surgery navigation station combining the latest computer technology and superior OR efficiency. Direct mounting of the displays and camera, featuring advanced optical wireless passive marker tracking technology, to the OR ceiling enables greatly improved setup flexibility and ergonomics while optimizing OR space usage. Zero OR footprint'' concept provided through ceiling-mounted cameras and touch displays Sterile surgical control accessible via two large 26" widescreen touch displays Brilliant display quality with resolutions beyond full HD (1920x1200pixels per display). Image quality entirely preserved, no visualization limitations from the touch interface (surface acoustic wave technology) Motorized joints for remote-controlled camera alignment yields new degrees of freedom in terms of setup flexibility Infrared tracking cameras with extended detection volume and laser pointer for quick and intuitive positioning Central "Home Button" ensures intuitive system control Interface Box including for plug&play connectivity e.g. with surgical microscopes, fluoroscopes, endoscopes, ultrasound etc via state-of-the-art digital and analog video inputs supporting up to full HD resolution: 1x SDI 1080i (In), 1x S-Video (In), 1x Composite (In), 1x USB(B) for microscope, 1x DVI-I Fast simultaneous access to e.g. PACS/hospital network and integration with e.g. C-Arms via 2x high-speed network connection (up to 1 Gbit/s each) High-performance medical computer unit (Intel Xeon 3530 2.8 GHz Central Processor Unit, 6 GB RAM memory and 320 GB storage) Computer Unit and Connection Unit can be placed in and outside the OR (within 50 m connected via optical fiber cable) Direct patient data transfer from/to 4x USB and CD/DVD ± RW Easy cleansing and hygienic disinfection VESA 100 display interface for customized adaptation to existing ceiling booms Defined camera interface for customized adaptation to existing ceiling booms Brainlab Session Sharing: Viewing and control of Curve from Digital OR package or Digital Lightbox Customer site pre-requisite for Brainlab Session Sharing: systems connected to local network infrastructure (minimum 100 Mbit/s, recommended 1 Gbit/s, effective network speed 40 Mbit/s, recommended maximum latency 2 ms) Intuitive Brainlab operating system for patient-centric data handling, application and display management Central ''Home Button'' ensures intuitive system control Intuitive content management of available displays via "drag & drop" functionality Streamlined patient-centric access to DICOM and Brainlab xBrain image data from multiple sources (PACS, CD/DVD, USB, Network) - DICOM functionality requires 30038 UNIVERSAL DICOM TRANSFER Unified search and load of patient image data from all available sources with optimized usability, including intelligent pre-fetching and buffering of patient image data for increased performance Ability to merge different patient data sets Export of treatment documentation (e.g. screenshots) and plans to network storage or USB HIPAA-compliant feature set including authentication, accountability log and automatic log-off

description

Learn more: https://www.brainlab.com/curve Ceiling-mounted high performance Image-Guided-Surgery navigation station combining the latest computer technology and superior O.R. efficiency. Direct mounting of the displays and camera, featuring advanced optical wireless passive marker tracking technology, to the O.R. ceiling enables greatly improved setup flexibility and ergonomics while optimizing O.R. space usage.

Transcript of Curve Ceiling Mounted Image Guided Surgery Technical Specifications

Page 1: Curve Ceiling Mounted Image Guided Surgery Technical Specifications

p. 1 of 6

PLATFORM15700 | CURVE CEILING-MOUNTED DUAL DISPLAYCeiling-mounted high performance Image-Guided-Surgery navigation stationcombining the latest computer technology and superior OR efficiency. Directmounting of the displays and camera, featuring advanced optical wirelesspassive marker tracking technology, to the OR ceiling enables greatly improvedsetup flexibility and ergonomics while optimizing OR space usage.� Zero OR footprint'' concept provided through ceiling-mounted cameras and

touch displays� Sterile surgical control accessible via two large 26" widescreen touch

displays� Brilliant display quality with resolutions beyond full HD (1920x1200pixels per

display). Image quality entirely preserved, no visualization limitations fromthe touch interface (surface acoustic wave technology)

� Motorized joints for remote-controlled camera alignment yields new degreesof freedom in terms of setup flexibility

� Infrared tracking cameras with extended detection volume and laser pointerfor quick and intuitive positioning

� Central "Home Button" ensures intuitive system control� Interface Box including for plug&play connectivity e.g. with surgical

microscopes, fluoroscopes, endoscopes, ultrasound etc via state-of-the-artdigital and analog video inputs supporting up to full HD resolution: 1x SDI1080i (In), 1x S-Video (In), 1x Composite (In), 1x USB(B) for microscope, 1xDVI-I

� Fast simultaneous access to e.g. PACS/hospital network and integrationwith e.g. C-Arms via 2x high-speed network connection (up to 1 Gbit/s each)

� High-performance medical computer unit (Intel Xeon 3530 2.8 GHz CentralProcessor Unit, 6 GB RAM memory and 320 GB storage)

� Computer Unit and Connection Unit can be placed in and outside the OR(within 50 m connected via optical fiber cable)

� Direct patient data transfer from/to 4x USB and CD/DVD ± RW� Easy cleansing and hygienic disinfection� VESA 100 display interface for customized adaptation to existing ceiling

booms� Defined camera interface for customized adaptation to existing ceiling

booms� Brainlab Session Sharing: Viewing and control of Curve from Digital OR

package or Digital Lightbox� Customer site pre-requisite for Brainlab Session Sharing: systems

connected to local network infrastructure (minimum 100 Mbit/s,recommended 1 Gbit/s, effective network speed 40 Mbit/s, recommendedmaximum latency 2 ms)

� Intuitive Brainlab operating system for patient-centric data handling,application and display management

� Central ''Home Button'' ensures intuitive system control� Intuitive content management of available displays via "drag & drop"

functionality� Streamlined patient-centric access to DICOM and Brainlab xBrain image

data from multiple sources (PACS, CD/DVD, USB, Network) - DICOMfunctionality requires 30038 UNIVERSAL DICOM TRANSFER Unified searchand load of patient image data from all available sources with optimizedusability, including intelligent pre-fetching and buffering of patient imagedata for increased performance

� Ability to merge different patient data sets� Export of treatment documentation (e.g. screenshots) and plans to network

storage or USB� HIPAA-compliant feature set including authentication, accountability log and

automatic log-off

Page 2: Curve Ceiling Mounted Image Guided Surgery Technical Specifications

p. 2 of 6

IMPORTANT INFORMATIONThe Curve Ceiling-Mounted ("Curve CM") requires a ceiling mounted armsystem in order to operate within its intended use. Brainlab does NOT offer orprovide such ceiling mounted arm system.It is the customer's sole responsibility� to identify and provide, at his own cost, a ceiling mounted arm system from

a third party supplier that allows a safe combination with Curve CM;� to ensure that the ceiling mounted arm system meets all specifications and

requirements as stated in the CURVE CEILING MOUNTED SITE PLANNINGMANUAL provided by Brainlab to the Customer; and

� to ensure that the ceiling mounted arm system is installed and ready for theCurve CM installation by Brainlab at the mutually agreed installation date.

The CURVE CM SITE PLANNING MANUAL will be provided by Brainlabupon order confirmation. If required, a copy can be provided prior to thisdate upon request. Please contact your Brainlab sales representative orconsultant

Computer Unit

Power / Voltage Max 650 W

Frequency 100V � 240 VAC+-10% at 50 or 60 Hz

Processor / Memory Intel, Xeon W3530 2.8 GHz. 6 GB RAM

Operating System Windows 7 64-bit (Embedded Standard)

I/O Support FRONT SIDE:

· 1x mains power hard switch

· 2x USB 2.0 (inside - for service use only)

BACK SIDE:

· 1x LAN 1Gbit

· 4x USB 2.0 (for 3rd party integr.)

· 2x DVI-D (for ext.routing systems)

· 3x fiber optics to Connection Unit

· 1x fiber optics for each touch display

· 1x mains power plug C14

· 1x potential equalization plug

Certificates and Classification Electrical Standard Approval

by EN 60601-1 (IEC 60601-1)

by IEC 529

Dimensions Width 44 cm

Height 17 cm

Depth 50 cm

Weight 21kg

Heat Emmission Max 450 W

Connection Unit

Power / Voltage Max 650 W

Frequency 100V � 240 VAC+-10% at 50or 60 Hz

Operating System Windows 7 64-bit (Embedded Standard)

I/O Support FRONT SIDE:

· 4x USB 2.0

· 1x intraoperative LAN 100 Mbit

· 1x SDI 1080i (In)

· 1x S-Video (In)

· 1x CVBS (Composite) (In)

· USB(B) for microscope

· DVI-I

· IEEE 802.11n, b, g (supporting 2.4 and 5 GHz upto 150Mbit/s)

BACK SIDE:

· power connectors for displays (2x) andcamera(1x)

Page 3: Curve Ceiling Mounted Image Guided Surgery Technical Specifications

p. 3 of 6

· 4x fiber optic connectors:

- 1 for the camera

- 3 to Computer Unit

· 1x mains power plug C14

· 1x potential equalization plug

· 1x mains power hard switch

Dimensions Width 38 cm

Height 10 cm

Depth 38 cm

Weight 12 kg

Heat Emmission Max 250 W

Touch Display

Panel Size 26��

Resolution 1920x1080 px

Dimensions Width 61.5 cm

Height 40.5 cm

Depth 8.5 cm

Weight 8.5 kg

Scan Frequency Horizontal 30-80 kHz

Vertical 50-85 Hz

Display Technology surface acoustic wave technology

Heat Emmission max. 90 W

Camera System

Physical Characteristics Width 61.3 cm

Heigth 8.6 cm

Depth 15.5 cm

Weight 3 kg

Tracking Distance 300 cm

Heat Emmission max 20 W. If motorized: max: 75 W

Enviromental Requirements

Temperature (Operation) 10° C to +35° C

Temperature (Storage) -10° C to +50° C

Humidity (Operation) 30% to 75% non-condensing

Humidity (Storage) 10% to 90% non-condensing

Pressure (Operation) 700 hPA to 1060 hPa

Pressure (Storage) 500 hPA to 1060 hPa

15705 | CURVE CEILING-MOUNTED SINGLE DISPLAYCeiling-mounted high performance Image-Guided-Surgery navigation stationcombining the latest computer technology and superior OR efficiency. Directmounting of the display and cameras, featuring advanced optical wirelesspassive marker tracking technology, to the OR ceiling enables greatly improvedsetup flexibility and ergonomics while optimizing OR space usage.� Zero OR footprint'' concept provided through ceiling-mounted cameras and

touch display� Sterile surgical control accessible via large 26" widescreen touch display� Brilliant display quality with resolutions beyond full HD (1920x1200pixels per

display). Image quality entirely preserved, no visualization limitations fromthe touch interface (surface acoustic wave technology)

� Infrared tracking cameras with extended detection volume and laser pointerfor quick and intuitive positioning

� Central "Home Button" ensures intuitive system control� Connection Unit including for plug&play connectivity e.g. with surgical

microscopes, fluoroscopes, endoscopes, ultrasound etc via state-of-the-artdigital and analog video inputs supporting up to full HD resolution: 1x SDI1080i (In), 1x S-Video (In), 1x Composite (In), 1x USB(B) for microscope, 1xDVI-I

Page 4: Curve Ceiling Mounted Image Guided Surgery Technical Specifications

p. 4 of 6

� Fast simultaneous access to e.g. PACS/hospital network and integrationwith e.g. C-Arms via 2x high-speed network connection (up to 1 Gbit/s each)

� High-performance medical computer unit (Intel Xeon 3530 2.8 GHz CentralProcessor Unit, 6 GB RAM memory and 320 GB storage)

� Computer Unit and Connection Unit can be placed in and outside the OR(within 50 m connected via optical fiber cable)

� Direct patient data transfer from/to 4x USB and CD/DVD ± RW.� Easy cleansing and hygienic disinfection� VESA 100 display interface for customized adaptation to existing ceiling

booms� Defined camera interface for customized adaptation to existing ceiling

booms� Brainlab Session Sharing: Viewing and control of Curve from Digital OR

package or Digital Lightbox� Customer site pre-requisite for Brainlab Session Sharing: systems

connected to local network infrastructure (minimum 100 Mbit/s,recommended 1 Gbit/s, effective network speed 40 Mbit/s, recommendedmaximum latency 2 ms)

� Intuitive Brainlab operating system for patient-centric data handling,application and display management

� Central ''Home Button'' ensures intuitive system control� Intuitive content management of available displays via "drag & drop"

functionality� Streamlined patient-centric access to DICOM and Brainlab xBrain image

data from multiple sources (PACS, CD/DVD, USB, Network) - DICOMfunctionality requires 30038 UNIVERSAL DICOM TRANSFER

� Unified search and load of patient image data from all available sources withoptimized usability, including intelligent pre-fetching and buffering of patientimage data for increased performance

� Ability to merge different patient data sets� Export of treatment documentation (e.g. screenshots) and plans to network

storage or USB� HIPAA-compliant feature set including authentication, accountability log and

automatic log-off

IMPORTANT INFORMATIONThe Curve Ceiling-Mounted ("Curve CM") requires a ceiling mounted armsystem in order to operate within its intended use. Brainlab does NOT offer orprovide such ceiling mounted arm system.It is the customer's sole responsibility� to identify and provide, at his own cost, a ceiling mounted arm system from

a third party supplier that allows a safe combination with Curve CM;� to ensure that the ceiling mounted arm system meets all specifications and

requirements as stated in the CURVE CEILING MOUNTED SITE PLANNINGMANUAL provided by Brainlab to the Customer; and

� to ensure that the ceiling mounted arm system is installed and ready for theCurve CM installation by Brainlab at the mutually agreed installation date.

The CURVE CM SITE PLANNING MANUAL will be provided by Brainlabupon order confirmation. If required, a copy can be provided prior to thisdate upon request. Please contact your Brainlab sales representative orconsultant.

Computer Unit

Power / Voltage Max 650 W

Frequency 100V � 240 VAC+-10% at 50 or 60 Hz

Processor / Memory Intel, Xeon W3530 2.8 GHz. 6 GB RAM

Operating System Windows 7 64-bit (Embedded Standard)

I/O Support FRONT SIDE:

· 1x mains power hard switch

Page 5: Curve Ceiling Mounted Image Guided Surgery Technical Specifications

p. 5 of 6

· 2x USB 2.0 (inside - for service use only)

BACK SIDE:

· 1x LAN 1Gbit

· 4x USB 2.0 (for 3rd party integr.)

· 2x DVI-D (for ext.routing systems)

· 3x fiber optics to Connection Unit

· 1x fiber optics for each touch display

· 1x mains power plug C14

· 1x potential equalization plug

Certificates and Classification Electrical Standard Approval

by EN 60601-1 (IEC 60601-1)

by IEC 529

Dimensions Width 44 cm

Height 17 cm

Depth 50 cm

Weight 21kg

Heat Emmission Max 450 W

Connection Unit

Power / Voltage Max 650 W

Frequency 100V � 240 VAC+-10% at 50or 60 Hz

Operating System Windows 7 64-bit (Embedded Standard)

I/O Support FRONT SIDE:

· 4x USB 2.0

· 1x intraoperative LAN 100 Mbit

· 1x SDI 1080i (In)

· 1x S-Video (In)

· 1x CVBS (Composite) (In)

· USB(B) for microscope

· DVI-I

· IEEE 802.11n, b, g (supporting 2.4 and 5 GHz upto 150Mbit/s)

BACK SIDE:

· power connectors for displays (2x) andcamera(1x)· 4x fiber optic connectors:

- 1 for the camera

- 3 to Computer Unit

· 1x mains power plug C14

· 1x potential equalization plug

· 1x mains power hard switch

Dimensions Width 38 cm

Height 10 cm

Depth 38 cm

Weight 12 kg

Heat Emmission Max 250 W

Touch Display

Panel Size 26��

Resolution 1920x1080 px

Dimensions Width 61.5 cm

Height 40.5 cm

Depth 8.5 cm

Weight 8.5 kg

Scan Frequency Horizontal 30-80 kHz

Vertical 50-85 Hz

Display Technology surface acoustic wave technology

Heat Emmission max. 90 W

Camera System

Page 6: Curve Ceiling Mounted Image Guided Surgery Technical Specifications

p. 6 of 6

Physical Characteristics Width 61.3 cm

Heigth 8.6 cm

Depth 15.5 cm

Weight 3 kg

Tracking Distance 300 cm

Heat Emmission max 20 W. If motorized: max: 75 W

Enviromental Requirements

Temperature (Operation) 10° C to +35° C

Temperature (Storage) -10° C to +50° C

Humidity (Operation) 30% to 75% non-condensing

Humidity (Storage) 10% to 90% non-condensing

Pressure (Operation) 700 hPA to 1060 hPa

Pressure (Storage) 500 hPA to 1060 hPa