CRONO 50: IDENTIFYING THE RESERVOIR COMPONENTS CANÈS.p.A. Medical Technology Via Cuorgnè, 42/a 10090 Rivoli Cascine Vica (TORINO) - Italy Tel. +39.011.9574872 Fax +39.011.9598880 www.canespa.it [email protected] MAN RS 01/EN/03 CRONO 50 07/09 The Crono 50 pump uses a 50 ml specific reservoir. Features of the reservoir: Sterile. Disposable. Apyrogenous. Do not use if the package is damaged or open. Reservoir body Piston Luer-Lock cap (detachable) Stem RESERVOIR Warning For safety purposes the use of original CRN ® Crono ® Reservoir syringes is recommended. The use of reservoirs of other types may damage the pump and harm its user. CANÈ S.p.A. is exempt from any responsibility should the device be used with other syringes than those recommended. Needle cover IDENTIFYING THE NEEDLE COMPONENTS PROFILL-CRN NEEDLE Use the PROFILL-CRN needle for: piercing the rubber membrane of the containers of the medication; draining the medication to fill the 50 ml reservoir. Features of the needle: 14G - 38 mm, sterile, disposable, apyrogenous, do not use if the package is damaged or open. The luer lock cap is connected to the stem of the reservoir from which it is detached by applying a little pressure. Functions of the luer lock cap: After filling the reservoir, it assists in unscrewing the stem from the piston; It assists in the correct connection between the pushing unit of the pump and the rubber piston of the syringe; It protects the medication inside the reservoir should it not be immediately used. LUER LOCK CAP Needle Luer Lock cap male Warning Do not use the needle for injections and/or samples. 1

Transcript of CRONO 50: IDENTIFYING THE RESERVOIR COMPONENTS

Page 1: CRONO 50: IDENTIFYING THE RESERVOIR COMPONENTS

CRONO 50: IDENTIFYING THE RESERVOIR COMPONENTS

CANÈ S.p.A. Medical Technology Via Cuorgnè, 42/a 10090 Rivoli Cascine Vica (TORINO) - Italy Tel. +39.011.9574872 Fax +39.011.9598880 www.canespa.it [email protected]

MA

NR

S01

/EN

/03

CR

ON

O50

07/0

9

The Crono 50 pump uses a 50 ml specific reservoir.

Features of the reservoir:• Sterile.• Disposable.• Apyrogenous.• Do not use if the package is damaged or open.

Reservoir body

Piston Luer-Lock cap(detachable)

Stem

RESERVOIR

Warning

• For safety purposes the use of original CRN® Crono® Reservoir syringesis recommended.

• The use of reservoirs of other types may damage the pump and harm itsuser.

• CANÈ S.p.A. is exempt from any responsibility should the device be usedwith other syringes than those recommended.

Needle cover

IDENTIFYING THE NEEDLE COMPONENTS

PROFILL-CRN NEEDLE

Use the PROFILL-CRN needle for:• piercing the rubber membrane of the containers of the medication;• draining the medication to fill the 50 ml reservoir.

Features of the needle: 14G - 38 mm, sterile, disposable, apyrogenous, do notuse if the package is damaged or open.

The luer lock cap is connected to the stem of thereservoir from which it is detached by applying alittle pressure. Functions of the luer lock cap:

• After filling the reservoir, it assists in unscrewing the stem from the piston;

• It assists in the correct connection betweenthe pushing unit of the pump and the rubberpiston of the syringe;

• It protects the medication inside the reservoir should it not be immediately used.

LUER LOCK CAP

Needle Luer Lock capmale

Warning

Do not use the needle for injections and/or samples.

1

Page 2: CRONO 50: IDENTIFYING THE RESERVOIR COMPONENTS

CRONO 50: PREPARING THE RESERVOIR

CANÈ S.p.A. Medical Technology Cuorgnè, 42/a 10090 Rivoli Cascine Vica (TORINO) - Italy Tel. +39.011.9574872 Fax +39.011.9598880 www.canespa.it [email protected]

The term “free flow” means possible uncon-trolled administration of a medication cau-sed by the effect of the force of gravity.To avoid this, the pushing unit is equippedwith a special profile that is attached to theinside of the rubber piston of the reservoirand binds them together.

The effectiveness of the safety system isguaranteed only by the correct attachmentof the piston to the pushing unit as indicatedin this figure:

ANTI FREE FLOW SAFETY SYSTEM

ATTACHING THE RESERVOIR TO THE DEVICE AttentionThe reservoir must be completely filled to allow the connection of the rubber piston tothe pushing unit.The rubber piston of the reservoir must be connected to the pushing unit of the pump;if this does not occur the following conditions may happen:• The infusion incurs a delay and is imprecise.• The connection between the pushing unit and the piston is incorrect and therefore

the safety of the anti free flow system is not guaranteed.

2

1. Unscrew the male luer lock cap from the PROFILL - CRN needle;2. Screw the needle on to the reservoir turning it clockwise, and remove the needle

cover;3. Fill the reservoir by drawing up the liquid slowly and checking that the amount of

drug does not exceed the 50 ml capacity or the partial volume set;4. Screw the luer lock cap on to the reservoir and unscrew the rod turning it

anticlockwise with a slightly rapid movement;

5. Insert the rubber piston onto the pusher while connecting it to the reservoir; a 90° rotation followed by a release will confirm the lockout.

6. Screw the anti-siphon valve between reservoir and infusion-set; the anti-siphon valve has to be used for intravenous and epidural infusions. The valve acts as a further safety device so as to avoid free flow should the syringe piston not be connected onto the pump’s pusher.

Anti-Siphon Valve

WARNINGThe piston must not be removed

from the syringe

1 2 3 4

NO YES1 2 3 4 5 6

Performance Specification

Cracking pressure 3.5 PSI +/- 1

Back Pressure > 45 PSI

Flow rate @ 5 PSI 203 ml/min

Static position Closed

Applications Liquids and air