ICRM Gamma Spectrometry Workshop Paris, France 23-24 February 2009 COINCIDENCE-SUMMING CORRECTION FACTORS OF ICRM INTERCOMPARISON EXERCISE USING GEANT4 USING GEANT4 S.Hurtado Universidad de Sevilla SPAIN 1

Transcript of COINCIDENCE-SUMMING CORRECTION FACTORS · PDF fileICRM Gamma Spectrometry Workshop Paris,...

Page 1: COINCIDENCE-SUMMING CORRECTION FACTORS · PDF fileICRM Gamma Spectrometry Workshop Paris, France 23-24 February 2009 COINCIDENCE-SUMMING CORRECTION FACTORS OF ICRM INTERCOMPARISON

ICRM Gamma Spectrometry Workshop

Paris, France

23-24 February 2009

COINCIDENCE-SUMMING CORRECTION FACTORS

OF ICRM INTERCOMPARISON EXERCISE

USING GEANT4USING GEANT4

S.Hurtado

Universidad de Sevilla

SPAIN

1

Page 2: COINCIDENCE-SUMMING CORRECTION FACTORS · PDF fileICRM Gamma Spectrometry Workshop Paris, France 23-24 February 2009 COINCIDENCE-SUMMING CORRECTION FACTORS OF ICRM INTERCOMPARISON

COINCIDENCE-SUMMING CORRECTION FACTORS OF ICRM INTERCOMPARISON EXERCISE USING GEANT4

COMPUTATIONAL ENVIRONMENT

GEANT4 v. 9.1

Linux Ubuntu 8.10

ICRM GAMMA SPECTROMETRY WORKSHOP Paris, France 23-24 February 2009

Linux Ubuntu 8.10

AMD Athlon 64 3700+

gcc-4.3.2

ROOT 5.14

2

Page 3: COINCIDENCE-SUMMING CORRECTION FACTORS · PDF fileICRM Gamma Spectrometry Workshop Paris, France 23-24 February 2009 COINCIDENCE-SUMMING CORRECTION FACTORS OF ICRM INTERCOMPARISON

COINCIDENCE-SUMMING CORRECTION FACTORS OF ICRM INTERCOMPARISON EXERCISE USING GEANT4

PHYSICS Standard EM Gammas :

photo-electric effect

Compton scattering

Pair production

electron, muon pair production

Cuts:

0.001 mm

Cuts:

0.001 mm

(Ge: 250 eV gamma –2.7 keV electron)

ICRM GAMMA SPECTROMETRY WORKSHOP Paris, France 23-24 February 2009

electron, muon pair production

Standard EM Electrons/Positrons :

Ionization and delta ray production

Bremsstrahlung

e+e- annihilation

synchrotron radiation

Nuclear Database:

NUCLÉIDE (152Eu and 134Cs)

3

Page 4: COINCIDENCE-SUMMING CORRECTION FACTORS · PDF fileICRM Gamma Spectrometry Workshop Paris, France 23-24 February 2009 COINCIDENCE-SUMMING CORRECTION FACTORS OF ICRM INTERCOMPARISON

COINCIDENCE-SUMMING CORRECTION FACTORS OF ICRM INTERCOMPARISON EXERCISE USING GEANT4

PARAMETER X-ray

Crystal radius R 24.75 mm

Crystal height L 47.81 mm

DETECTOR SETUP

ICRM GAMMA SPECTROMETRY WORKSHOP Paris, France 23-24 February 2009

Crystal height L 47.81 mm

Distance to window D 5.0 mm

Hole inner radius r 4.4 mm

Hole height d 39.5 mm

Inner dead layer th 1.0 mm

Front dead layer tf 0.3 m4

Page 5: COINCIDENCE-SUMMING CORRECTION FACTORS · PDF fileICRM Gamma Spectrometry Workshop Paris, France 23-24 February 2009 COINCIDENCE-SUMMING CORRECTION FACTORS OF ICRM INTERCOMPARISON

COINCIDENCE-SUMMING CORRECTION FACTORS OF ICRM INTERCOMPARISON EXERCISE USING GEANT4

500

1000

1500

2000

F.E.P. EFFICIENCY Experimental vs MC (10 cm)

ICRM GAMMA SPECTROMETRY WORKSHOP Paris, France 23-24 February 2009

-7.0 -6.0 -5.0 -4.0 -3.0 -2.0 -1.0 0.0 1.0

100

200

300

100 200 300 500 1000 1500 2000

D (%) -6.6 -4.2 -3.2 -3.7 -3.1 -2.7 -4.8MC 1.08E+00 7.51E-01 4.98E-01 2.92E-01 1.53E-01 1.05E-01 8.03E-02Experimental 1.02E+00 7.21E-01 4.83E-01 2.81E-01 1.48E-01 1.02E-01 7.66E-02

5

Page 6: COINCIDENCE-SUMMING CORRECTION FACTORS · PDF fileICRM Gamma Spectrometry Workshop Paris, France 23-24 February 2009 COINCIDENCE-SUMMING CORRECTION FACTORS OF ICRM INTERCOMPARISON

COINCIDENCE-SUMMING CORRECTION FACTORS OF ICRM INTERCOMPARISON EXERCISE USING GEANT4

SIMULATION OF DECAY SCHEME

• Sampling decay mode: β- decay or EC / β+ decay

• Nuclear transition:

ICRM GAMMA SPECTROMETRY WORKSHOP Paris, France 23-24 February 2009

• Nuclear transition: γ emission or IC

• Atomic relaxation: X-rays and Auger e- by IC or EC

6

Page 7: COINCIDENCE-SUMMING CORRECTION FACTORS · PDF fileICRM Gamma Spectrometry Workshop Paris, France 23-24 February 2009 COINCIDENCE-SUMMING CORRECTION FACTORS OF ICRM INTERCOMPARISON

COINCIDENCE-SUMMING CORRECTION FACTORS OF ICRM INTERCOMPARISON EXERCISE USING GEANT4

RESULTS – 2 cm152Eu 134Cs

Energy/keV C error Energy/keV C error121.8 1.100 0.001 242.8 1.234 0.036244.7 1.167 0.002 326.5 1.334 0.057344.3 1.058 0.002 475.3 1.168 0.006411.1 1.148 0.004 563.2 1.162 0.003444.0 1.148 0.004 569.3 1.166 0.003

ICRM GAMMA SPECTROMETRY WORKSHOP Paris, France 23-24 February 2009

444.0 1.148 0.004 569.3 1.166 0.003564.0 1.140 0.011 604.7 1.094 0.002688.6 1.142 0.009 795.8 1.100 0.002778.9 1.094 0.003 801.9 1.153 0.004867.4 1.211 0.005 1038.6 1.030 0.010964.1 1.123 0.003 1167.9 0.913 0.007

1085.8 0.977 0.004 1365.2 0.855 0.0061089.7 1.101 0.0081112.1 1.076 0.0041212.9 1.230 0.0101299.1 1.107 0.0091408.0 1.091 0.004 7

Page 8: COINCIDENCE-SUMMING CORRECTION FACTORS · PDF fileICRM Gamma Spectrometry Workshop Paris, France 23-24 February 2009 COINCIDENCE-SUMMING CORRECTION FACTORS OF ICRM INTERCOMPARISON

COINCIDENCE-SUMMING CORRECTION FACTORS OF ICRM INTERCOMPARISON EXERCISE USING GEANT4

RESULTS – 2 cm

1.000

1.200

1.400

1.600

NUCLEIDE

ICRM GAMMA SPECTROMETRY WORKSHOP Paris, France 23-24 February 2009

0.000

0.200

0.400

0.600

0.800

121.

8

244.

7

344.

3

411.

1

444.

0

564.

0

688.

6

778.

9

867.

4

964.

1

1085

.8

1089

.7

1112

.1

1212

.9

1299

.1

1408

.0

242.

8

326.

5

475.

3

563.

2

569.

3

604.

7

795.

8

801.

9

1038

.6

1167

.9

1365

.2

8

Page 9: COINCIDENCE-SUMMING CORRECTION FACTORS · PDF fileICRM Gamma Spectrometry Workshop Paris, France 23-24 February 2009 COINCIDENCE-SUMMING CORRECTION FACTORS OF ICRM INTERCOMPARISON

COINCIDENCE-SUMMING CORRECTION FACTORS OF ICRM INTERCOMPARISON EXERCISE USING GEANT4

RESULTS – 5 cm152Eu 134Cs

Energy/keV C error Energy/keV C error121.8 1.030 0.001 242.8 1.080 0.039244.7 1.054 0.003 326.5 1.091 0.057344.3 1.017 0.003 475.3 1.069 0.007411.1 1.048 0.005 563.2 1.051 0.004444.0 1.042 0.005 569.3 1.058 0.004

ICRM GAMMA SPECTROMETRY WORKSHOP Paris, France 23-24 February 2009

444.0 1.042 0.005 569.3 1.058 0.004564.0 1.040 0.013 604.7 1.037 0.003688.6 1.065 0.011 795.8 1.036 0.004778.9 1.032 0.005 801.9 1.065 0.005867.4 1.069 0.007 1038.6 1.041 0.012964.1 1.047 0.005 1167.9 0.966 0.009

1085.8 1.004 0.006 1365.2 0.941 0.0081089.7 1.034 0.0101112.1 1.023 0.0061212.9 1.075 0.0111299.1 1.056 0.0111408.0 1.045 0.006 9

Page 10: COINCIDENCE-SUMMING CORRECTION FACTORS · PDF fileICRM Gamma Spectrometry Workshop Paris, France 23-24 February 2009 COINCIDENCE-SUMMING CORRECTION FACTORS OF ICRM INTERCOMPARISON

COINCIDENCE-SUMMING CORRECTION FACTORS OF ICRM INTERCOMPARISON EXERCISE USING GEANT4

RESULTS – 5 cm

0.800

1.000

1.200

1.400

NUCLEIDE

ICRM GAMMA SPECTROMETRY WORKSHOP Paris, France 23-24 February 2009

0.000

0.200

0.400

0.600

0.800

121.

8

244.

7

344.

3

411.

1

444.

0

564.

0

688.

6

778.

9

867.

4

964.

1

1085

.8

1089

.7

1112

.1

1212

.9

1299

.1

1408

.0

242.

8

326.

5

475.

3

563.

2

569.

3

604.

7

795.

8

801.

9

1038

.6

1167

.9

1365

.2

10

Page 11: COINCIDENCE-SUMMING CORRECTION FACTORS · PDF fileICRM Gamma Spectrometry Workshop Paris, France 23-24 February 2009 COINCIDENCE-SUMMING CORRECTION FACTORS OF ICRM INTERCOMPARISON

COINCIDENCE-SUMMING CORRECTION FACTORS OF ICRM INTERCOMPARISON EXERCISE USING GEANT4

RESULTS – 10 cm152Eu 134Cs

Energy/keV C error Energy/keV C error121.8 1.005 0.003 242.8 1.127 0.044244.7 1.021 0.004 326.5 0.928 0.051344.3 1.011 0.005 475.3 1.024 0.008411.1 1.019 0.007 563.2 1.022 0.006444.0 1.018 0.007 569.3 1.019 0.006

ICRM GAMMA SPECTROMETRY WORKSHOP Paris, France 23-24 February 2009

444.0 1.018 0.007 569.3 1.019 0.006564.0 1.011 0.014 604.7 1.023 0.006688.6 1.010 0.012 795.8 1.013 0.007778.9 1.027 0.007 801.9 1.014 0.008867.4 1.032 0.009 1038.6 1.018 0.014964.1 1.039 0.008 1167.9 0.985 0.012

1085.8 1.017 0.009 1365.2 1.000 0.0121089.7 0.986 0.0121112.1 0.991 0.0091212.9 1.027 0.0141299.1 1.035 0.0131408.0 1.004 0.009 11

Page 12: COINCIDENCE-SUMMING CORRECTION FACTORS · PDF fileICRM Gamma Spectrometry Workshop Paris, France 23-24 February 2009 COINCIDENCE-SUMMING CORRECTION FACTORS OF ICRM INTERCOMPARISON

COINCIDENCE-SUMMING CORRECTION FACTORS OF ICRM INTERCOMPARISON EXERCISE USING GEANT4

RESULTS – 10 cm

0.800

1.000

1.200

1.400

NUCLEIDE

ICRM GAMMA SPECTROMETRY WORKSHOP Paris, France 23-24 February 2009

0.000

0.200

0.400

0.600

0.800

121.

8

244.

7

344.

3

411.

1

444.

0

564.

0

688.

6

778.

9

867.

4

964.

1

1085

.8

1089

.7

1112

.1

1212

.9

1299

.1

1408

.0

242.

8

326.

5

475.

3

563.

2

569.

3

604.

7

795.

8

801.

9

1038

.6

1167

.9

1365

.2

12