CM42 pH ORP Operating Instruction

download CM42 pH ORP Operating Instruction

of 44

Transcript of CM42 pH ORP Operating Instruction

  • 7/30/2019 CM42 pH ORP Operating Instruction

    1/44

    CM42-MAA301EAE00

    BA00381C/07/EN/14.12

    Valid as of:

    Software version 10.07.00

    Operating Instructions

    Liquiline M CM42Two-wire transmitter for pH/ORP measurement withMemosens glass sensors

    Part 1: Commissioning

  • 7/30/2019 CM42 pH ORP Operating Instruction

    2/44

    Documentation information

    Warnings

    The structure, signal words and color coding of the warnings follow the specifications of ANSI Z535.6

    ("Product safety information in product manuals, instructions and other collateral materials").

    Symbols used

    Structure of the safety

    symbol

    Meaning

    DANGER!

    Causes (/consequences)

    Possible consequences if ignored

    Preventive measures

    This symbol alerts you to a dangerous situation.

    Failure to avoid this situationwill result in serious or fatal injury.

    WARNING!

    Causes (/consequences)

    Possible consequences if ignored

    Preventive measures

    This symbol alerts you to a dangerous situation.Failure to avoid this situation can result in serious or fatal injury.

    CAUTION!

    Causes (/consequences)

    Possible consequences if ignored

    Preventive measures

    This symbol alerts you to a dangerous situation.

    Failure to avoid this situation can result in medium or minor

    injury.

    NOTICE

    Cause/situation

    Possible consequences if ignored

    Action/note

    This symbol alerts you to situations that can cause damage to

    equipment or property.

    Additional information, tips

    Permitted or recommended

    Forbidden or not recommended

  • 7/30/2019 CM42 pH ORP Operating Instruction

    3/44

    Documents supplied

    Operating Instructions

    CD-ROM

    Contents:

    Measuring technology fundamentals

    Communication fundamentals

    Overview of analytical measuring technology

    Accessories

    Safety

    Ordering information

    In addition

    1 manufacturer's certificate

    Adhesive labels:

    Interior nameplate

    Wiring adhesive label

    Attach these adhesive labels to the inside of the cover.

    The Operating Instructions are split into two parts:

    Commissioning Description of your device

    All steps that you have to carry outonce onlyduring initial commissioning

    Software history

    Technical data

    Operation

    Software description

    All steps that you have to carry out during the routine operation of your device, depending on

    your measuring task

    Maintenance and accessories Diagnostics and troubleshooting

  • 7/30/2019 CM42 pH ORP Operating Instruction

    4/44

    CM42-MAA301EAE00

    4

    Table of contents

    1 Basic safety instructions . . . . . . 51.1 Requirements for personnel . . . . . . . . . . . . 5

    1.2 Designated use . . . . . . . . . . . . . . . . . . . . . . 5

    1.3 Occupational safety . . . . . . . . . . . . . . . . . . 5

    1.4 Operational safety . . . . . . . . . . . . . . . . . . . . 5

    1.5 Product safety . . . . . . . . . . . . . . . . . . . . . . 6

    2 Identification. . . . . . . . . . . . . . . 72.1 Device designation . . . . . . . . . . . . . . . . . . . 7

    2.2 Scope of delivery . . . . . . . . . . . . . . . . . . . . 8

    2.3 Reordering documentation . . . . . . . . . . . . . 8

    2.4 Certificates and approvals . . . . . . . . . . . . . . 9

    3 Device description . . . . . . . . . 103.1 Housing closed . . . . . . . . . . . . . . . . . . . . . 10

    3.2 Housing open . . . . . . . . . . . . . . . . . . . . . . 10

    4 Installation . . . . . . . . . . . . . . . 114.1 System setup . . . . . . . . . . . . . . . . . . . . . . 11

    4.2 Installation conditions . . . . . . . . . . . . . . . 12

    4.3 Installation instructions . . . . . . . . . . . . . . 14

    4.4 Post-installation check . . . . . . . . . . . . . . . 16

    5 Wiring . . . . . . . . . . . . . . . . . . . 175.1 Electrical connection . . . . . . . . . . . . . . . . 17

    5.2 Post-connection check . . . . . . . . . . . . . . . 23

    6 Commissioning . . . . . . . . . . . . 246.1 Function check . . . . . . . . . . . . . . . . . . . . 24

    6.2 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

    6.3 Operation concept . . . . . . . . . . . . . . . . . . 26

    6.4 Position of the DIL switches . . . . . . . . . . . 266.5 Communication . . . . . . . . . . . . . . . . . . . . 27

    6.6 Quick Setup (local operation) . . . . . . . . . . 29

    6.7 Measuring display (MEAS) . . . . . . . . . . . . 30

    7 Calibration and adjustment. . . 317.1 Definitions . . . . . . . . . . . . . . . . . . . . . . . . 31

    7.2 Zero point and slope . . . . . . . . . . . . . . . . . 32

    7.3 Notes on the calibration . . . . . . . . . . . . . . 32

    7.4 Calibration intervals . . . . . . . . . . . . . . . . . 33

    7.5 Types of calibration . . . . . . . . . . . . . . . . . 34

    7.6 Two-point calibration . . . . . . . . . . . . . . . . 34

    7.7 Single-point calibration . . . . . . . . . . . . . . 35

    7.8 Temperature adjustment . . . . . . . . . . . . . 36

    7.9 Calibrate . . . . . . . . . . . . . . . . . . . . . . . . . 36

    8 Software history. . . . . . . . . . . .38

    9 Technical data . . . . . . . . . . . . .409.1 Input . . . . . . . . . . . . . . . . . . . . . . . . . . . 40

    9.2 Output . . . . . . . . . . . . . . . . . . . . . . . . . . 40

    9.3 Electrical connection . . . . . . . . . . . . . . . . 40

    9.4 Performance characteristics . . . . . . . . . . . 40

    9.5 Environment . . . . . . . . . . . . . . . . . . . . . . 41

    9.6 Mechanical construction . . . . . . . . . . . . . 41

    Index . . . . . . . . . . . . . . . . . . . .42

    http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/
  • 7/30/2019 CM42 pH ORP Operating Instruction

    5/44

    CM42-MAA301EAE00 Basic safety instructions

    Endress+Hauser 5

    1 Basic safety instructions

    1.1 Requirements for personnel

    Installation, commissioning, operation and maintenance of the measuring system must only be

    carried out by trained technical personnel. The technical personnel must be authorized to perform the tasks by the owner-operator.

    The electrical connection may only be established by an electrical technician.

    The technical personnel must have read and understood these Operating Instructions and must

    follow the instructions they contain.

    Faults at the measuring point may only be rectified by authorized and properly trained personnel.

    Repairs not described in the enclosed Operating Instructions may only be carried out directly at

    the manufacturer's or by the Service Organization.

    1.2 Designated use

    Liquiline M CM42 is a two-wire transmitter for liquid analysis in all areas of process engineering.

    The primary applications comprise:

    Chemical processes

    Pharmaceutical industry

    Food technology

    The use of the transmitter depends strongly on the sensor deployed. For this reason, please pay

    attention to the information on "designated use" provided in the sensor manual.

    The transmitter is suitable for pollution degree 3.

    If the device is used for any purpose other than that described, this poses a threat to the safety of people

    and the entire measuring system and is thus not permitted.

    The manufacturer does not accept liability for damage caused by improper or non-designated use.

    1.3 Occupational safety

    As the user, you are responsible for observing the following safety regulations:

    Installation guidelines

    Local standards and regulations

    1.4 Operational safety

    Prior to commissioning the entire measuring point, check that all connections are correct. Make

    sure that electric cables and hose connections are not damaged.

    Do not commission damaged products. Protect them against unintentional startup. Label and

    identify the damaged product as defective.

    If the faults cannot be eliminated, take the products out of service and protect them against

    unintentional startup.

  • 7/30/2019 CM42 pH ORP Operating Instruction

    6/44

    Basic safety instructions CM42-MAA301EAE00

    6 Endress+Hauser

    1.5 Product safety

    The product is designed to meet state-of-the-art safety requirements, has been tested and left the factory

    in a condition in which it is safe to operate.

    The applicable regulations and European standards have been taken into account.

  • 7/30/2019 CM42 pH ORP Operating Instruction

    7/44

    CM42-MAA301EAE00 Identification

    Endress+Hauser 7

    2 Identification

    2.1 Device designation

    2.1.1 Nameplate

    Nameplates can be found:

    On the outside of the housing (lasered)

    On the packaging (adhesive label, portrait format)

    On the adhesive label sheet provided (landscape format, for your own use)

    Compare the data on the nameplate with your order.

    The nameplate provides you with the following information on your device:

    Order code

    Serial number

    Device version (software package) Measuring range(s)

    Input, output values

    Approvals (depending on the version ordered)

    Protection class

    Ambient temperature range

    Approval symbols

    Safety notices, warnings

    2.1.2 Serial number and order codeThe order code and serial number of your device can be found in the following locations:

    On the nameplate

    On the front page of the Operating Instructions supplied

    In the delivery papers

    The documentation only contains information that was valid at the time of delivery. If you have

    any queries, contact your local sales center.

  • 7/30/2019 CM42 pH ORP Operating Instruction

    8/44

    Identification CM42-MAA301EAE00

    8 Endress+Hauser

    2.1.3 Interpreting the order code

    You can create a valid and complete order code on the Internet using the Configurator tool.

    Enter the following address in the browser to launch the product page:

    www.products.endress.com/cm42

    1. You can choose from the following options on the right-hand side of the Product page:

    2. Click "Configure this product".

    3. The Configurator opens in a new window.Using the selection buttons configure the order code on the nameplate of your device.

    4. You can export the order code and the product details as a PDF or Excel file.

    To do so, click the specific button at the top of the page.

    2.2 Scope of delivery

    1 transmitter in the version ordered

    1 mounting plate incl. 4 flat head screws

    1 sheet of adhesive labels (nameplate, terminal connection diagrams)

    Operating Instructions Part 1 and 2, BA381C "Commissioning" and BA382C "Operation" in the

    language ordered

    1 CD-ROM with additional documentation

    1 manufacturer's certificate

    If you have any queries, please contact your supplier or local sales center.

    2.3 Reordering documentation

    What to order?

    Reordering lost documentation or documentation of which several copies are required:You must quote the serial number of your device.

    You receive the same documentation as received with the delivery.

    First time ordering appropriate documentation for devices whose function you have modified yourself

    by installing a module:

    You have to indicate the order code the device corresponds to after installing the module. Specify a

    serial number also.

    You receive the documentation to suit your newly modified device.

    How to order?

    On the Internet: www.endress.com/liquiline-documentationYou receive PDF files of both parts of the Operating Instructions.

  • 7/30/2019 CM42 pH ORP Operating Instruction

    9/44

    CM42-MAA301EAE00 Identification

    Endress+Hauser 9

    Through your sales center:

    Indicate you want the documentation as hard copy.

    The order code of the device described, and its serial number, are printed on the front page of the

    original brochures. Furthermore, it is possible to specify a tag name.

    The serial number and tag name are notpart of the PDF files which you can download over the

    Internet or through W@M access.

    2.4 Certificates and approvals

    With this declaration, the manufacturer guarantees that the product conforms to the regulations of

    European Directive 2004/108/EC and Low Voltage Directive 2006/95/EC. This is proven by

    observing the standards listed in the Declaration of Conformity.

  • 7/30/2019 CM42 pH ORP Operating Instruction

    10/44

    Device description CM42-MAA301EAE00

    10 Endress+Hauser

    3 Device description

    3.1 Housing closed

    3.2 Housing open

    a0001067

    Fig. 1: Liquiline with closed display cover

    12345-8

    Wall securing plateAlarm LEDDisplayNavigatorSoft keys,assignment depends on menu

    a0010278

    Fig. 2: With open display cover (without wiring)

    1

    23

    Slot for DAT module

    CPU moduleSensor module

    4

    56

    DIN rail

    Threaded jointsCDI socket (service interface)

  • 7/30/2019 CM42 pH ORP Operating Instruction

    11/44

    CM42-MAA301EAE00 Installation

    Endress+Hauser 11

    4 Installation

    4.1 System setup

    A complete system consists of:

    Liquiline M CM42 An immersion assembly, e.g. CPA140 or flow assembly, e.g. CPA240 or retractable assembly, e.g.

    CPA475; each with or without a potential matching pin (PML)

    A pH/ORP sensor with Memosens technology: e.g. CPS11D

    A measuring cable: CYK10

    A programmable logic controller (PLC) with FOUNDATION Fieldbus interface incl. PC and operating

    software and terminating resistor

    a0002655

    Fig. 3: Example of a measuring system

    123

    PC with operating programPLCLiquiline M CM42

    456

    Special measuring cableSensor and assemblyTerminating resistor

  • 7/30/2019 CM42 pH ORP Operating Instruction

    12/44

    Installation CM42-MAA301EAE00

    12 Endress+Hauser

    4.2 Installation conditions

    4.2.1 Dimensions

    a0011834

    Fig. 4: Dimensions

    164

    (6

    .46)

    148

    (5

    .83)

    110

    (4

    .33)

    220 (8.66)

    177

    (6

    .97)

    2

    (0

    .0

    8) 130 (5.12)

    138 (5.43)

    144 (5.67)

    144

    (5

    .67)

    22

    (0

    .87)

    95

    (3

    .74)

    mm (inch)

  • 7/30/2019 CM42 pH ORP Operating Instruction

    13/44

    CM42-MAA301EAE00 Installation

    Endress+Hauser 13

    4.2.2 Mounting plate

    4.2.3 Weather protection cover

    NOTICE

    Effect of climatic conditions (rain, snow, direct sun etc.)

    Impaired operation to complete transmitter failure

    When installing outside, always use the weather protection cover (accessory).

    a0010363

    Fig. 5: Mounting plate

    a0001671

    Fig. 6: Weather protection cover

    220 (8.66)

    85 (3.35)

    70 (2.76)

    141 (5.55)

    196 (7.72)

    70

    (2

    .76)

    85 (3.35)

    95

    (3

    .74)

    mm (inch)

    122

    2 x R8 2 x R3

    R3

    2 x R8

    2

    (0

    .08

    )

    86 (3.39)

    112 (4.41)

    161 (6.34)

    4 x

    6(0.24)

    2

    x 11

    (0.43)

    4x

    7(0.28)

    80

    (3

    .15)

    2 x R70

  • 7/30/2019 CM42 pH ORP Operating Instruction

    14/44

    Installation CM42-MAA301EAE00

    14 Endress+Hauser

    4.3 Installation instructions

    4.3.1 Wall or field mounting

    There are many ways of mounting the device:

    Wall mounting

    Mounting on a vertical pipe or post (round or square)

    Mounting on railing or on a horizontal pipe (round or square)

    Diameter of a pipe, post or railing suitable for mounting:

    30 to 61 mm (1.18 to 2.40").

    You require the post retainer when mounting. This is an accessory and is not included in the scope

    of delivery (see "Accessories").

    a0010373

    Fig. 7: Mounting on a vertical pipe/post

    1 Liquiline 2 Mounting plate (part of scope of delivery for Liquiline)3 Pipe or post (circular/square)4 Mounting plate (post retainer, --> "Accessories")5 Threaded rods with spring washer, washer and nut (part of scope of delivery for post retainer)

    1 2 3

    45

  • 7/30/2019 CM42 pH ORP Operating Instruction

    15/44

    CM42-MAA301EAE00 Installation

    Endress+Hauser 15

    a0010371

    Fig. 8: Mounting on a horizontal pipe/railing

    1 Liquiline 2 Mounting plate (part of scope of delivery for Liquiline)3 Pipe or railing 4 Mounting plate (post retainer, --> "Accessories")5 Threaded rods with spring washer, washer and nut (part of scope of delivery for post retainer)

    a0010382

    Fig. 9: Wall mounting

  • 7/30/2019 CM42 pH ORP Operating Instruction

    16/44

    Installation CM42-MAA301EAE00

    16 Endress+Hauser

    4.3.2 Panel mounting

    For panel mounting, you need the installation kit consisting of tensioning screws and a front seal. This

    is an accessory (see "Accessories") and is not included in the scope of delivery.

    If installing the units above one another, you must observe a minimum distance for the cable glands

    of the upper device.

    If installing the units beside one another, you must observe a minimum distance for opening the

    front of the housing.

    If arrangingin a square, you must take into account the lengths of the mounting plates at the rear

    of the device or the cable glands for minimum spacing distance.

    4.4 Post-installation check

    After installation, check the transmitter for damage.

    Check whether the transmitter is protected against precipitation and direct sunlight (e.g. by the

    weather protection cover).

    a0002402

    Fig. 10: Panel mounting: left side view, right front view

  • 7/30/2019 CM42 pH ORP Operating Instruction

    17/44

    CM42-MAA301EAE00 Wiring

    Endress+Hauser 17

    5 Wiring

    5.1 Electrical connection

    WARNING!

    The device is live!

    Incorrect wiring can result in injury or fatality

    The electrical connection may only be established by an electrical technician.

    The electrical technician must have read and understood these Operating Instructions and must fol-

    low the instructions they contain.

    Prior to beginningany wiring work, make sure voltage is not applied to any of the cables.

    5.1.1 Opening the housing

    1. Loosen the four screws on the front with a Phillips screwdriver:

    2. Open the housing:

    NOTICE

    Pointed or sharp tools

    Damaged housing seal and scratched housing Do not use any sharp or pointed tools, such as screwdrivers or knives, to open the housing.

  • 7/30/2019 CM42 pH ORP Operating Instruction

    18/44

    Wiring CM42-MAA301EAE00

    18 Endress+Hauser

    5.1.2 Housing grounding

    WARNING!

    Electric voltage at ungrounded cable mounting rail

    No shock protection

    Connect the cable mounting rail to the foundation ground with a separate 2.5 mm2

    (14 AWG)functional ground.

    5.1.3 Cable grounding in housing

    If possible, only use terminated genuine cables.

    Ground the sensor cable in the housing of the transmitter as shown in the following diagrams

    (sample cable, does not necessarily correspond to the original cable).

    a0003617

    Fig. 11: Grounding the housing12

    Cable mounting rail2.5 mm2(14 AWG) functional ground

    a0002679

    Fig. 12: Terminated cable (example)

    1 Outer shield (exposed) 2 Cable cores with ferrules 3 Cable sheath (insulation)

    a0002680

    Fig. 13: Inserting the cable

    4 Grounding clip

    a0002681

    Fig. 14: Tightening the screw, the cableshield is grounded by the

    grounding clip

  • 7/30/2019 CM42 pH ORP Operating Instruction

    19/44

    CM42-MAA301EAE00 Wiring

    Endress+Hauser 19

    5.1.4 Cable terminals

    Single-wire, multi-wire and fine-wire cables can be used for the connection, with and without

    ferrules. Only one wire is permitted per terminal.

    5.1.5 Signal output / power supply

    NOTICE

    Incorrect supply voltage can cause damage or a malfunction

    Comply with the specifications on the permitted and minimum supply voltage (see "Electrical con-

    nection" under Technical data).

    Use a fieldbus cable grounded at both ends (device and PCS).

    There are various ways of connecting the unit:

    1. Two-wire cabling shielded on both sides, "hard grounding"

    (generally to be given priority over "capacitive connection to ground")

    2. Shielded two-wire cabling, "capacitive connection to ground" (shield grounded on device side via

    capacitor, "C-module" accessory required)

    If there is a risk of large potential equalization currents.

    3. Use the fieldbus connection socket (accessories)

    "Hard grounding"

    Connect the cable shield to the cable mounting rail.

    a0003602

    Fig. 15: Opening the terminal

    1. Insert a suitable screwdriver into the opening of the

    terminal spring (square opening) until the stop.

    a0003603

    Fig. 16: Inserting the cable

    2. Insert the terminated cable end into the terminal

    opening (round opening).

    a0003604

    Fig. 17: Releasing the terminal spring

    3. Remove the screwdriver. Make sure the cable is

    securely positioned in the terminal.

  • 7/30/2019 CM42 pH ORP Operating Instruction

    20/44

    Wiring CM42-MAA301EAE00

    20 Endress+Hauser

    Connect the cable cores as per the assignment ( 18).

    "Capacitance ground connection"

    Pull back the braided shield, put the stranded extension wire of the C-module (item 1) onto the

    exposed shield and tighten the clip:

    Connect the extension wire to the cable mounting rail.

    Connect the cable cores as per the assignment ( 20).

    a0004061

    Fig. 18: View in device (CPU module)a0002367

    Fig. 19: Wiring diagram

  • 7/30/2019 CM42 pH ORP Operating Instruction

    21/44

    CM42-MAA301EAE00 Wiring

    Endress+Hauser 21

    "Fieldbus connection socket"

    Screw the fieldbus connection socket into the housing bushing.

    Trim the connection cores of the socket to approx. 15 cm.

    Connect the cable cores as per the assignment. In doing so, you must connect the cable shield

    (GN/YE) to the cable mounting rail ( 22).

    a0004072

    Fig. 20: View in device (CPU module)a0004073

    Fig. 21: Wiring diagram

    a0002364

    Fig. 22: View in device (CPU module)a0004062

    Fig. 23: Wiring diagram

  • 7/30/2019 CM42 pH ORP Operating Instruction

    22/44

    Wiring CM42-MAA301EAE00

    22 Endress+Hauser

    5.1.6 Sensor connection

    pH sensors suitable for connection

    Glass electrodes: CPS11D, CPS41D, CPS71D, CPF81D, CPS91D

    Pfaudler electrodes: CPS341D

    ORP sensors suitable for connection CPS12D, CPS42D, CPS72D, CPF82D, CPS92D

    pH/ORP combined sensors suitable for connection

    CPS16D, CPS76D, CPS96D

    The list of Endress+Hauser sensors provides examples only and does not claim to be exhaustive. For

    more information on the technical data and the suitability of particular sensors for your measuring task,

    please visit our online portal: www.products.endress.com

    NOTICE

    No shielding against electrical and magnetic interference

    Interference can give erroneous measurement results

    You must connect shield connections or terminals to the functional ground ( ) (there is no

    protective ground on the plastic housing ( )).

    Explanation of abbreviations in the following diagrams:

    Abbreviation Meaning

    U+Power supply of the digital sensor

    U

    Com ACommunication signals of the digital sensor

    Com B

    )

  • 7/30/2019 CM42 pH ORP Operating Instruction

    23/44

    CM42-MAA301EAE00 Wiring

    Endress+Hauser 23

    5.2 Post-connection check

    WARNING!

    Wiring errorsIncorrect wiring puts the safety of people and the measuring point at risk. The manufacturer does not

    accept any responsibility for errors that result from failure to comply with the instructions in this

    manual.

    Only put the transmitter into operation if you can answeryes to all of the following questions.

    Device state and specifications

    1. Are the transmitter and cables free from damage on the outside?

    Electrical connection

    2. Are the mounted cables strain relieved?3. Are the cables run without loops and cross-overs?

    a0001087

    Fig. 24: View in device (sensor module)a0001078

    Fig. 25: Wiring diagram

  • 7/30/2019 CM42 pH ORP Operating Instruction

    24/44

    Commissioning CM42-MAA301EAE00

    24 Endress+Hauser

    4. Are the signal lines correctly connected in accordance with the wiring diagram?

    5. Have you connected unused connection wires to the cable mounting rail?

    6. Are all the connection wires securely positioned in the cable terminals?

    7. Are all the cable entries installed, tightened and sealed?

    6 Commissioning

    6.1 Function check

    WARNING!

    Incorrect connection, incorrect supply voltage

    Safety risks for staff and incorrect operation of the device

    Check that all connections have been established correctly in accordance with the wiring diagram. Make sure that the supply voltage matches the voltage indicated on the nameplate.

    6.2 Display

    6.2.1 Overview

    a0010019

    Fig. 26: Local display (measuring mode)

    123

    Status lineDisplay and setting fieldAssignment of the soft keys

  • 7/30/2019 CM42 pH ORP Operating Instruction

    25/44

    CM42-MAA301EAE00 Commissioning

    Endress+Hauser 25

    6.2.2 Status display

    Symbol for the measuring menu (MEAS)

    Device state is OK. No alarms or warnings are present.

    Service warning. Maintenance is recommended.The device is operational; the reliability of the measured value may be restricted.

    Alarm that points to an error.The device is no longer operational or the measured value is no longer reliable.

    Measured value display in the status line if a menu other than MEAS is launched.The main value and the secondary value are displayed.

    Symbol is displayed while the device is actively communicating via the fieldbus.

    Symbol is displayed if user administration is switched on.

  • 7/30/2019 CM42 pH ORP Operating Instruction

    26/44

    Commissioning CM42-MAA301EAE00

    26 Endress+Hauser

    6.3 Operation concept

    6.4 Position of the DIL switches

    You can set the simulation mode or the write protection directly in the device.To do so, change the switch positions of the DIL switch on the CPU module.

    a0010029

    Fig. 27: Pressing the soft key: selecting the menu directly

    a0010035-en

    Fig. 28: Turning the navigator: moving the cursor in the menu

    a0010036-en

    Fig. 29: Press navigator: select a functiona0010037-en

    Fig. 30: Turn navigator: change value

    a0010038-en

    Fig. 31: Press navigator: accept new value

    Operation concept

    1. Select a menu directly by pressing the

    appropriate soft key.

    2. Move the cursor in the menu by turning thenavigator.

    3. Select the desired function by pressing the

    navigator.

    4. Change the value by turning the navigator.

    5. Accept the new value by pressing the

    navigator.

  • 7/30/2019 CM42 pH ORP Operating Instruction

    27/44

    CM42-MAA301EAE00 Commissioning

    Endress+Hauser 27

    The factory setting of the switches is always "Off".

    6.5 Communication

    6.5.1 Device identification

    6.5.2 Sequence during commissioning

    Perform the steps of the commissioning procedure in the following sequence:

    1. Supply the device with power (--> "Wiring")

    2. Install the device description files3. Assign the PD_TAG

    a0003106

    Fig. 32: DIL switch on CPU module (without connectorstrip)

    1 Simulation 2 Write protection

    DIL switch assignment:

    Simulation (switch 1):

    (Only if switch 8 = "on")

    Liquiline simulates measured values and outputs

    these via the fieldbus interface. Write protection (switch 2):

    (Only if switch 8 = "on")

    You cannot make any device settings via the fieldbus

    interface when the write protection is switched

    "On".

    All other switches (3-8) are not assigned functions at

    the factory.

    Manufacturer name: Endress+Hauser

    Model name: Liquiline M CM42 pH/ORP

    Manufacturer ID (hex): 452B48

    Device Type (hex): 108FDevice Revision (hex): 1

    Device Class (hex): Link Master

    ITK Version: 4.6

    Function and other Blocks: 1xRB, 6xAI, 4xTB

    1

    2

    3

    4

    5

    6

    7

    8

    O

    FF

    O

    N

  • 7/30/2019 CM42 pH ORP Operating Instruction

    28/44

    Commissioning CM42-MAA301EAE00

    28 Endress+Hauser

    4. Configure the Resource Block

    5. Configure the Transducer Blocks

    6. Configure the Analog Input Blocks

    7. Connect the function blocks in the editor

    8. Download the configuration (Configure menu)

    9. If necessary, check the configuration (SIMULATE function).

    6.5.3 Required software and files

    FF configuration program, e.g. NI-FBUS Configurator

    Cff file (Common File Format: *.cff)

    DD (Device Description: *.sym, *.ffo)

    You must load these files into the intended directory before the configuration.

    Make sure that you always use the correct file versions.

    You can use the DEV_REV and DD_REV parameters in the Resource Block to read out therequired versions.

    You can download the device descriptions from our homepage.

    6.5.4 Configuring a device

    1. Start the configuration program, e.g. NI-FBUS Configurator.

    The network is then displayed as an extendable tree diagram.

    If the device is connected correctly, it is detected automatically:

    E+H_XXXXXXXXXXX

    2. Double-clickon the device name.

    The device data is then displayed:

    PD_TAG (physical name of the device)

    DEVICE_ID (unique device identification, consisting of:

    452B48 --> Vendor-ID for Endress+Hauser

    108F --> Dev_Type

    XXXXXXXXXXX --> Serial number of the device, cf. nameplate

    NODE_ADDRESS (device address in the FF network, is generally assigned automatically)

    3. Click on the device name using the right mouse button.

    This offers you access to the menu, in which you can change the PD_TAG.

    4. Click on the device name using the left mouse button.This opens the tree diagram and displays the function blocks:

    E+H_XXXXXXXXXX

    RESOURCE_XXXXXXXXXX (RB2)

    TRD_PHORP_XXXXXXXXXX (TBUL)

    TRD_TEMP_XXXXXXXXXX (TBUL)

    TRD_FSDG1_XXXXXXXXXX (TBUL)

    TRD_SERVICE_XXXXXXXXXX (TBUL)

    AI_1_XXXXXXXXXX (AI)

    AI_2_XXXXXXXXXX (AI)

    AI_3_XXXXXXXXXX (AI)

  • 7/30/2019 CM42 pH ORP Operating Instruction

    29/44

    CM42-MAA301EAE00 Commissioning

    Endress+Hauser 29

    AI_4_XXXXXXXXXX (AI)

    AI_5_XXXXXXXXXX (AI)

    6.5.5 Resource Block

    Click on Resource Block and:

    1. check the position of the hardware switch in WRITE_LOCK.

    If "locked" is displayed, you have to change the position of the DIL switch on the CPU module.

    2. Click on the Block-Tag using the right mouse button if you want to change it.

    3. Set MODE_BLK_TARGET to "Out-of-Service".

    4. Reset the device to the factory settings using the RESTART function.

    You can also access this function by clicking on the device name with the right mouse button.

    5. If necessary, assign a new tag description (TAG_DESC).

    6. Set MODE_BLK_TARGET to "AUTO".

    6.5.6 Transducer BlocksClick on the Transducer Blocks one after the other and:

    1. Click on the Block-Tag using the right mouse button if you want to change it.

    2. Set MODE_BLK_TARGET to "Out-of-Service".

    3. If necessary, assign a new tag description (TAG_DESC).

    4. Configure the device and the sensor.

    5. Set MODE_BLK_TARGET to "AUTO".

    6.5.7 Analog Input Blocks

    Click on the Analog Input Blocks one after the other and:1. Click on the Block-Tag using the right mouse button if you want to change it.

    2. Set MODE_BLK_TARGET to "Out-of-Service".

    3. If necessary, assign a new tag description (TAG_DESC).

    4. Set the channel to the measured value or distance.

    5. Set L_TYPE to "DIRECT" if you want to have the OUT value specified in technical units.

    Set L_TYPE to "INDIRECT" if you want to scale the OUT value.

    6. Set the desired output damping in PV_TIME.

    7. If necessary, set the indicating and critical alarms.

    8. Set MODE_BLK_TARGET to "AUTO".

    6.6 Quick Setup (local operation)

    1. Connect the supply voltage.

    2. Wait for the initialization to complete.

    3. Completely work through the "Quick Setup" submenu in "SETUP".

  • 7/30/2019 CM42 pH ORP Operating Instruction

    30/44

    Commissioning CM42-MAA301EAE00

    30 Endress+Hauser

    6.7 Measuring display (MEAS)

    There are various display modes:

    (Press the navigator button to change the mode)

    1. Main value

    2. Primary and secondary value

    3. All measured values

    Path:SETUP / Quick setup

    Function Options Info

    Language Options

    Second language English

    Factory settingSecond language

    You selected the "second language" in the CM42 order code.

    If the language is changed, the settings configured for thedevice remain intact.You can load another second language from the SystemDAT.--> SETUP/DAT menu

    Date format Options DDMMYYYY MMDDYYYY

    Factory settingDDMMYYYY

    Editing mode:DD (day): 1 to 31MM (month): 1 to 12

    YYYY (year): 2005 to 2100

    Set date Depends on the format DDMMYYYY

    Time format Options hhmmss (24h) hhmmss (am/pm)

    Factory settinghhmmss (24h)

    24-hour display or 12-hour display

    Editing mode:hh (hour): 0 to 23 / 0 am to 12 pmmm (minutes): 0 to 59ss (seconds): 0 to 59

    Set time Depends on the format hh:mm:ss

    Measured value Options(depending on sensor type) pH ORP mV ORP % rH

    Factory settingpHorORP mV

    Some of the subsequent menus and their optionsdepend on the option selected here.

    Measuring range rH: 0 to 42

    Temperature unit Options C F

    Factory settingC

    Decide whether you want to use the ISO or the Americantemperature unit.

    All the values and settings are automatically converted tomatch the option selected.

  • 7/30/2019 CM42 pH ORP Operating Instruction

    31/44

    CM42-MAA301EAE00 Calibration and adjustment

    Endress+Hauser 31

    When you exit the measuring mode, the current measured value is displayed in the status bar in

    the top left-hand corner:

    7 Calibration and adjustment

    The calibration provides important information on the condition of your sensor and the quality of

    the pH measurement.

    pH glass electrodes

    In addition to the slope that has an ideal value of approx. -59 mV/pH at 25 C, the change in the zero

    point also provides important information to the user. It is an indicator of the condition of the gel-like

    layer of the pH glass and could point to a blockage in the reference system. Suitable maintenance would

    involve cleaning or regenerating the sensor.

    To perform a quick test, simply immerse the sensor in a buffer solution with the same pH value as the

    internal buffer (e.g. pH 7). 0 mV would be an ideal value. The greater the deviation from the ideal value,

    the poorer the condition of the sensor (20 mV is still acceptable in most situations).

    7.1 DefinitionsCalibration (as per DIN 1319):

    A calibration is defined as a set of operations that establish the relationship between the measured value

    or expected value of the output variable and the related true or correct value of the measured variable

    (input variable) for a measuring system under specified conditions.

    A calibration does not alter the performance of the measuring device.

    Adjustment

    An adjustment corrects the value displayed by a measuring device, in other words the

    measured/displayed value (the actual value) is corrected so that the reading agrees with the correct, set

    value.

    Sensor type Main value Main value / secondary value

    All values

    pH, glass pH value pH value, temperature Main value, raw value, temperature,glass impedance, reference impedance

    pH and ORPcombined sensor

    pH value or ORP or rH value pH value or ORP or rH value,temperature

    Main values pH, ORP and rH(depending on measured value), raw

    value, temperature, glass impedance,reference impedance

    ORP ORP ORP, temperature Main value, raw value, temperature

  • 7/30/2019 CM42 pH ORP Operating Instruction

    32/44

    Calibration and adjustment CM42-MAA301EAE00

    32 Endress+Hauser

    The value determined during calibration is used to calculate the correct measured value and saved in

    the sensor.

    7.2 Zero point and slope

    Using a mathematical function, the transmitter converts the input signal of the sensor y (raw measuredvalue) to the measured value x. In many cases, this function is a simple linear of the form y = a + b . x.

    The linear element "a" is usually equivalent to the zero point and the factor "b" is the slope of the line.

    The Nernst equation, which is used to calculate the pH value, is a typical linear relationship:

    pH = -lg(aH+), aH+ ... activity of the hydrogen ions

    Ui ... raw measured value in mV

    U0 ... zero point (=voltage at pH 7)

    R ... relative gas constant (8.3143 J/molK)

    T ... temperature [K]F ... Faraday constant (26.803 Ah)

    The slope of the Nernst equation (-2.303RT/F) is known as the Nernst factor and has the value

    -59.16 mV/pH at 25 C.

    7.3 Notes on the calibration

    The following rules apply for all parameters:

    Calibrate in a way that reflects conditions in the process.

    If the process medium is constantly moving, also move the calibration solution accordingly (e.g.

    use a magnetic stirrer if calibrating in the laboratory).

    a0013593

    Fig. 33: Linear function

    a Zero point b Slope

    U = Ui 0 -2.303 RT

    FpH

  • 7/30/2019 CM42 pH ORP Operating Instruction

    33/44

    CM42-MAA301EAE00 Calibration and adjustment

    Endress+Hauser 33

    If your medium is relatively stationary, calibrate in solutions that are also stationary.

    Make sure that the samples are homogeneous for reference measurements, sample calibration etc.

    Use the same menu settings as those in the process to perform the calibration.

    Example: If you automatically compensate for the temperature effect during pH measurement, switch

    on automatic temperature compensation for the calibration also.

    It is advisable to perform the laboratory calibration using the "Memobase" database software (-->

    "Accessories"). This improves the availability of your measuring points and all the calibration and

    sensor data records are stored securely in the database.

    7.4 Calibration intervals

    The service life of a pH glass electrode is limited. This is due, in part, to the deterioration and aging of

    the pH-sensitive membrane glass. This aging causes the gel-like layer to change and become thicker

    over time.

    Symptoms of aging include:

    Higher membrane resistance

    Slow response

    Decrease in the slope

    A change in the reference system (e.g. due to contamination, i.e. unwanted redox reactions at the

    reference electrode) or electrolyte solution dissolving away in the reference half cell can change the

    reference potential, which, in turn, causes a zero point shift in the measuring electrode.

    To ensure a high level of accuracy, it is important to readjust the pH sensors at set intervals.

    The calibration interval depends heavily on the area of application of the sensor, as well as the requiredlevel of accuracy and reproducibility. The calibration interval can vary between daily and once every

    few months.

    Defining the calibration interval for the process

    1. Check the sensor with a buffer solution, e.g. pH 7.

    Proceed as specified in Step 2 only if the value deviates from the set point. No

    calibration/adjustment is necessary if the value is within the defined deviation tolerance range

    (see the Technical Information for the sensor).

    2. Calibrate and adjust the sensor.

    3. After 24 hours, check again with the buffer solution.a. If the deviation is within the permitted tolerance range, increase the checking interval by

    doubling it for example.

    b. If the deviation is larger, you must shorten the interval.

    4. Continue to proceed as defined in Steps 2 and 3 until you have identified the suitable interval.

  • 7/30/2019 CM42 pH ORP Operating Instruction

    34/44

    Calibration and adjustment CM42-MAA301EAE00

    34 Endress+Hauser

    7.5 Types of calibration

    7.5.1 pH measurement

    The following types of calibration are possible:

    Two-point calibration

    With calibration buffers

    Single-point calibration

    Entry of a reference value

    Sample calibration with laboratory comparative value

    Temperature adjustment by entering a reference value

    If using a combined sensor (CPS16D/CPS76D/CPS96D) you must calibrate both the pH electrode

    and the ORP electrode to get reliable rH values, for example.

    7.5.2 Measuring the ORP

    The following types of calibration are possible:

    Two-point calibration with medium samples (only ORP %)

    Single-point calibration with calibration buffer (only ORP mV)

    Data entry of an offset (only ORP mV)

    Temperature adjustment via reference value

    7.6 Two-point calibration

    7.6.1 General information

    Two-point calibration is the preferred method for pH sensors, particularly in the following

    applications:

    Municipal and industrial wastewater

    Natural waters and drinking water

    Boiler feedwater and condensates

    Beverages

    Calibrating with buffers with pH 7.0 and 4.0 is recommended for most applications.

    Alkaline buffer solutions have the disadvantage that carbon dioxide from the air can alter the pH value

    of the buffer on the long term. If calibrating with alkaline buffers it is best to do so in closed systems,such as flow assemblies or retractable assemblies with a rinse chamber, to minimize the effect of air.

    7.6.2 With calibration buffers

    pH measurement

    You use calibration buffers to perform two-point calibration. The quality buffers supplied by

    Endress+Hauser are certified and measured in an accredited laboratory. The accreditation (DAR

    registration number "DKD-K-52701") confirms that the actual values and the maximum deviations are

    correct and traceable.

  • 7/30/2019 CM42 pH ORP Operating Instruction

    35/44

    CM42-MAA301EAE00 Calibration and adjustment

    Endress+Hauser 35

    To calibrate the sensor, remove it from the medium and calibrate it in the laboratory. Since Memosens

    sensors save the data, you can always work with "precalibrated" sensors and do not have to stop

    monitoring the process to perform a calibration.

    Only use calibration buffers once.

    Measurement of the ORP in % (2-point cal.)

    To obtain useful ORP % values, you must adapt the sensor to your process. This is achieved through

    two-point calibration. The two calibration points are characteristic of the most important states your

    medium can assume in the process.

    You require two different compositions of your medium that represent the characteristic limits of your

    process (e.g. 20% and 80% value).

    The absolute value in mV is not relevant for the ORP % measurement.

    Measuring the ORP in % (data entry)

    As an alternative to two-point calibration with medium samples, you can also define the two calibration

    points via data entry. To do so, use data pairs in mV and % for the two calibration points that represent

    the most important states of your process medium. Ideally, you will have determined the mV values

    with a reference measurement.

    7.6.3 Entering data for the zero point, slope and temperature (only pH)

    You enter the slope, zero point and temperature manually. The function for determining the pH value

    is calculated from these values. Thus, the data entry returns the same result as two-point calibration.

    You must determine the slope, zero point and temperature alternatively.

    7.7 Single-point calibration

    7.7.1 Numeric input of a reference value (1-point cal.)

    pH measurement

    Single-point calibration is particularly useful if the deviation of the pH value from a reference value,

    and not the absolute pH value itself, is of interest to the user. Applications for single-point calibration

    include:

    Process control

    Quality assurance

    Fluctuations in the process value should not exceed 0.5 pH and the process temperature must remain

    relatively constant. As the measuring range is limited as a result, it is possible to set the slope to -59

    mV/pH (at 25 C).

    Alternatively, you can also use the "sample calibration". Here, you take a sample from the process and

    determine the pH value in the laboratory. In the case of the laboratory sample, you must make sure that

    the pH value is determined at the process temperature.

  • 7/30/2019 CM42 pH ORP Operating Instruction

    36/44

    Calibration and adjustment CM42-MAA301EAE00

    36 Endress+Hauser

    Measuring the ORP in mV

    You enter the offset directly with this type of calibration. Use the measured value of a reference

    measurement, for example, to determine the offset.

    7.7.2 Sample calibration (grab sample cal., only pH)

    With this type of calibration, you take a sample of the medium and determine its pH value (at process

    temperature) in the laboratory. You then use this laboratory value to adjust the sensor. This does not

    change the slope of the calibration function.

    7.7.3 With ORP buffers (only measurement of the ORP)

    With this type of calibration, you work with calibration buffers, e.g. ORP buffers from Endress+Hauser.

    For this purpose, you remove the sensor from the medium and calibrate it in the laboratory. Since

    Memosens sensors save the data, you can always work with "precalibrated" sensors and do not have to

    stop monitoring the process for extended periods to perform a calibration.Only use calibration buffers once.

    7.8 Temperature adjustment

    The temperature sensor has to be calibrated at regular intervals to ensure the measured value is not

    falsified by incorrect temperature measurement.

    1. Determine the temperature with a reference measurement and, depending on the result, decide

    whether you want to adjust the temperature sensor.

    2. If you do, go to the "CAL/Temp. adjustment" menu.The current offset is displayed.

    3. Start the adjustment and enter the new offset which you determined beforehand with the

    reference measurement.

    You can cancel the calibration any time by pressing the "ESC" key. No new data are then used to adjust

    the sensor.

    7.9 Calibrate

    1. Press the soft key for "CAL".

    2. Select the type of calibration.

    3. Enter the necessary parameters.

    (The parameters that have to be configured depend on the type of calibration selected.)

    4. Start the calibration.

    5. Follow the instructions in the menu.

    6. Decide whether to use the calibration data captured, or to abort or repeat the calibration.

    After calibration, the transmitter automatically switches back to the measuring mode and your

    measuring point is now ready for operation.

    If calibration is aborted using ESC, or if the calibration is faulty, the system continues to use theoriginal calibration data.

  • 7/30/2019 CM42 pH ORP Operating Instruction

    37/44

    CM42-MAA301EAE00 Calibration and adjustment

    Endress+Hauser 37

    A calibration error is shown as plain text on the display.

    Any offset set is automatically deleted after accepting the calibration.

  • 7/30/2019 CM42 pH ORP Operating Instruction

    38/44

    Software history CM42-MAA301EAE00

    38 Endress+Hauser

    8 Software history

    Date Version Changes in the software Documentation: edition

    04/2012 Package 07

    to 10

    Extension

    Support for pH/ORP combined sensors

    Improvement Enhanced user administration Damping can be set for up to 600 s Choice of measured variables modified Modified configuration of diagnostics messages Modified diagnostics messages

    BA00381/07/xx/14.12

    BA00382/07/xx/14.12

    11/2010 Package 06 Extension Support for Korean

    Improvement Sensor state display

    Sensor check

    BA381/07/xx/13.10BA382/07/xx/13.10

    11/2009 Package 05 Extension Manual hold also in the "Basic" software version

    Improvement Two-point calibration for special buffers corrected Improved table entry

    BA381/07/xx/10.03.01BA382/07/xx/10.03.01

    10/2008 10.05.00 Extension User administration upgraded Monitoring functions DAT functionalities: Upgrade, update, saving user settings by DAT Redesign of measuring menus

    BA381/07/xx/08.08.01BA382/07/xx/08.08.01

    02/2008 10.04.05 Improvement 3.6 mA functionality

    BA381/07/xx/08.02.01BA382/07/xx/08.02.01

    05/2007 10.04.05 Extension Memobase functions

    BA381/07/xx/07.05.01BA382/07/xx/07.05.01

    06/2006 10.04.00 Extension Advanced functionality:

    Medium compensation pH Memosens limit contactor

    ORP %

    BA381/07/xx/06.10.01BA382/07/xx/06.10.01

    04/2006 10.03.00 Extension Advanced functionality:

    Pfaudler electrodes Predictive maintenance: PCS, SCC, calibration timer

    Software update via DAT modules Date and time format selectable Language extension Diagnosis table, advanced diagnosis codes

    BA381/07/xx/06.05.01BA382/07/xx/06.05.01

    09/2005 10.02.00 Extension PROFIBUS Profile 3.0 Upload / Download mode Extension of the Memosens data

    BA381/07/xx/05.11.01BA382/07/xx/05.11.01

  • 7/30/2019 CM42 pH ORP Operating Instruction

    39/44

    CM42-MAA301EAE00 Software history

    Endress+Hauser 39

    07/2005 10.01.00 Improvement Correction of the Memosens functionality Extended error codes Additional types of calibration

    Improvement of the update behavior

    BA381/07/xx/05.08.01BA382/07/xx/05.08.01

    06/2005 10.00.00 Original software BA381/07/xx/05.05.24BA382/07/xx/05.05.24

    Date Version Changes in the software Documentation: edition

  • 7/30/2019 CM42 pH ORP Operating Instruction

    40/44

    Technical data CM42-MAA301EAE00

    40 Endress+Hauser

    9 Technical data

    9.1 Input

    9.1.1 Measured variables

    --> Documentation of the connected sensor

    9.1.2 Measuring range

    --> Documentation of the connected sensor

    9.1.3 Cable specification

    9.2 Output

    9.2.1 Output signal

    FOUNDATION Fieldbus

    9.2.2 Signal on alarm

    Digital via fieldbus

    9.3 Electrical connection

    9.3.1 Supply voltage and signal voltage

    9.3.2 Cable cross-section

    Max. cable cross-section: 2.5 mm2 (i14 AWG), GND 4 mm2 (i12 AWG)

    9.4 Performance characteristics

    9.4.1 Measured value resolution

    --> Documentation of the connected sensor

    9.4.2 Measured error--> Documentation of the connected sensor

    CYK10 with Memosens Max. cable length 100 m (330 ft)

    FOUNDATION Fieldbus: 9 to 32 V DC

    Bus current consumption: 22 mA

  • 7/30/2019 CM42 pH ORP Operating Instruction

    41/44

    CM42-MAA301EAE00 Technical data

    Endress+Hauser 41

    9.4.3 Repeatability

    --> Documentation of the connected sensor

    9.5 Environment

    9.5.1 Ambient temperature range

    30 to 70 C (20 to 160 F)

    9.5.2 Storage temperature

    40 to 80 C (40 to 175 F)

    9.5.3 Electromagnetic compatibility

    Interference emission and interference immunity to EN 61326-1: 2004, Category B (residential envi-

    ronments)

    9.5.4 Degree of protection

    IP66 / IP 67 / NEMA 4X

    9.5.5 Relative humidity

    10 to 95%, not condensing

    9.5.6 Pollution degree

    The product is suitable for pollution degree 3.

    9.6 Mechanical construction

    9.6.1 Weight

    1.5 kg (3.3 lbs)

    9.6.2 Material

    Housing

    Housing seals

    Module housing

    Soft keys,

    Cable mounting rail

    Display panel

    PC-FR (polycarbonate, flame-retardant)

    Foamed silicone, EPDM

    PC (polycarbonate)

    TPE

    Stainless steel 1.4301 (AISI 304)

    PC-FR (polycarbonate, flame-retardant)

  • 7/30/2019 CM42 pH ORP Operating Instruction

    42/44

    CM42-MAA301EAE00

    42

    Index

    AAdjustment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31

    Analog Input Block . . . . . . . . . . . . . . . . . . . . . . . 29

    Approvals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

    BBus configuration . . . . . . . . . . . . . . . . . . . . . . . . 28

    Bus configuration (hardware) . . . . . . . . . . . . . . . 26

    CCable cross-section . . . . . . . . . . . . . . . . . . . . . . . 40

    Cable grounding . . . . . . . . . . . . . . . . . . . . . . . . . 18

    Cable specification . . . . . . . . . . . . . . . . . . . . . . . 40Cable terminals. . . . . . . . . . . . . . . . . . . . . . . . . . 19

    CAL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31

    Calibrate. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36

    Calibration

    Calibration intervals . . . . . . . . . . . . . . . . . . . . 33

    Numeric input . . . . . . . . . . . . . . . . . . . . . . . . 35

    Sample calibration . . . . . . . . . . . . . . . . . . . . . 36

    Single-point calibration . . . . . . . . . . . . . . . . . . 35

    Two-point calibration . . . . . . . . . . . . . . . . . . . 34

    Types of calibration. . . . . . . . . . . . . . . . . . . . . 34Check

    Electrical connection. . . . . . . . . . . . . . . . . . . . 23

    Function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

    Post-connection check . . . . . . . . . . . . . . . . . . 23

    Post-installation . . . . . . . . . . . . . . . . . . . . . . . 16

    Commissioning. . . . . . . . . . . . . . . . . . . . . . . . . . 24

    Commissioning with bus communication . . . . . . 27

    Communication . . . . . . . . . . . . . . . . . . . . . . . . . 27

    Connecting ground. . . . . . . . . . . . . . . . . . . . . . . 18

    Connection data . . . . . . . . . . . . . . . . . . . . . . . . . 40

    DDDs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28

    Degree of protection . . . . . . . . . . . . . . . . . . . . . . 41

    Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

    Device configuration. . . . . . . . . . . . . . . . . . . . . . 28

    Device description . . . . . . . . . . . . . . . . . . . . . . . 10

    Device description files . . . . . . . . . . . . . . . . . . . . 28

    Device designation . . . . . . . . . . . . . . . . . . . . . . . . 7

    Device identification . . . . . . . . . . . . . . . . . . . . . . 27

    DIL switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26Dimensions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

    Display

    Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

    Status display. . . . . . . . . . . . . . . . . . . . . . . . . 25

    EElectrical connection . . . . . . . . . . . . . . . . . . 23, 40

    Cable cross-section . . . . . . . . . . . . . . . . . . . . 40

    Power supply. . . . . . . . . . . . . . . . . . . . . . . . . 19

    Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

    Signal output. . . . . . . . . . . . . . . . . . . . . . 19, 40

    Supply voltage . . . . . . . . . . . . . . . . . . . . . . . . 40

    Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . 17

    Electrical technician. . . . . . . . . . . . . . . . . . . . . . 17Electromagnetic compatibility . . . . . . . . . . . . . . 41

    Environment

    Ambient temperature range . . . . . . . . . . . . . . 41

    Degree of protection . . . . . . . . . . . . . . . . . . . 41

    Electromagnetic compatibility . . . . . . . . . . . . 41

    Pollution degree . . . . . . . . . . . . . . . . . . . . . . 41

    Relative humidity . . . . . . . . . . . . . . . . . . . . . 41

    Storage temperature . . . . . . . . . . . . . . . . . . . 41

    FFieldbus

    Analog Input Block . . . . . . . . . . . . . . . . . . . . 29

    Resource Block . . . . . . . . . . . . . . . . . . . . . . . 29

    Transducer Block. . . . . . . . . . . . . . . . . . . . . . 29

    Function group

    CAL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31

    GGrounding concept

    Capacitive connection to ground . . . . . . . . . . 20

    Hard grounding . . . . . . . . . . . . . . . . . . . . . . . 19

    HHousing

    Closed. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

    Open. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

    Housing grounding . . . . . . . . . . . . . . . . . . . . . . 18

    Housing opening . . . . . . . . . . . . . . . . . . . . . . . . 17

    IIdentification

    Nameplate. . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

    http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/
  • 7/30/2019 CM42 pH ORP Operating Instruction

    43/44

    CM42-MAA301EAE00

    Endress+Hauser

    Serial number. . . . . . . . . . . . . . . . . . . . . . . . . . 7

    Input

    Cable specification . . . . . . . . . . . . . . . . . . . . . 40

    Measured variables. . . . . . . . . . . . . . . . . . . . . 40

    Measuring range. . . . . . . . . . . . . . . . . . . . . . . 40

    Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

    Installation instructions

    Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . 12

    Panel mounting . . . . . . . . . . . . . . . . . . . . . . . 16

    Wall or field mounting . . . . . . . . . . . . . . . . . . 14

    Interpreting the order code . . . . . . . . . . . . . . . . . . 8

    MMaterial . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41

    Maximum measured error . . . . . . . . . . . . . . . . . 40

    Measured value resolution . . . . . . . . . . . . . . . . . 40Measured variables. . . . . . . . . . . . . . . . . . . . . . . 40

    Measuring range. . . . . . . . . . . . . . . . . . . . . . . . . 40

    Mechanical construction

    Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41

    Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41

    Mounting plate. . . . . . . . . . . . . . . . . . . . . . . . . . 13

    NNameplate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

    Nernst equation . . . . . . . . . . . . . . . . . . . . . . . . . 32

    Numeric input . . . . . . . . . . . . . . . . . . . . . . . . . . 35

    OOnline Configurator . . . . . . . . . . . . . . . . . . . . . . . 8

    Opening the housing . . . . . . . . . . . . . . . . . . . . . 17

    Operation concept . . . . . . . . . . . . . . . . . . . . . . . 26

    Ordering information

    Documentation . . . . . . . . . . . . . . . . . . . . . . . . 8

    Output

    Output signal . . . . . . . . . . . . . . . . . . . . . . . . . 40

    Signal on alarm. . . . . . . . . . . . . . . . . . . . . . . . 40

    PPerformance characteristics

    Maximum measured error . . . . . . . . . . . . . . . 40

    Measured value resolution . . . . . . . . . . . . . . . 40

    Repeatability. . . . . . . . . . . . . . . . . . . . . . . . . . 41

    PLC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

    Pollution degree . . . . . . . . . . . . . . . . . . . . . . . . . 41

    Post-installation

    Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

    Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

    QQuick Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29

    R

    Relative humidity . . . . . . . . . . . . . . . . . . . . . . . . 41Reordering documentation . . . . . . . . . . . . . . . . . . 8

    Repeatability . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41

    Resource Block . . . . . . . . . . . . . . . . . . . . . . . . . . 29

    SSample calibration. . . . . . . . . . . . . . . . . . . . . . . . 36

    Scope of delivery. . . . . . . . . . . . . . . . . . . . . . . . . . 8

    Sensor

    Connection. . . . . . . . . . . . . . . . . . . . . . . . . . . 22

    Serial number . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

    Signal output. . . . . . . . . . . . . . . . . . . . . . . . . 19, 40Single-point calibration . . . . . . . . . . . . . . . . . . . . 35

    Slope. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32

    Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28

    Software history . . . . . . . . . . . . . . . . . . . . . . . . . 38

    Status display . . . . . . . . . . . . . . . . . . . . . . . . . . . 25

    Storage temperature . . . . . . . . . . . . . . . . . . . . . . 41

    Supply voltage. . . . . . . . . . . . . . . . . . . . . . . . . . . 40

    System setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

    TTechnical data. . . . . . . . . . . . . . . . . . . . . . . . . . . 40Environment. . . . . . . . . . . . . . . . . . . . . . . . . . 41

    Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40

    Mechanical construction . . . . . . . . . . . . . . . . . 41

    Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40

    Performance characteristics . . . . . . . . . . . . . . . 40

    Transducer Block . . . . . . . . . . . . . . . . . . . . . . . . 29

    Two-point calibration . . . . . . . . . . . . . . . . . . . . . 34

    UUse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

    WWeight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41

    Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17

    ZZero point . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32

    http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/http://cm42-maa301eae00%2171162304%21e%21120419.pdf/
  • 7/30/2019 CM42 pH ORP Operating Instruction

    44/44