Rev # Date/Quote Description I have read the appropriate requirements from the Skytron Pre-Installation Guide for Ergon booms and consulted with the individual trades. I understand there are specific ceiling height, medical gas, electrical, video/communications and structural requirements that must be supplied by the appropriate trade representatives to support this project. Please be aware that modification to the site specific document could result in additional Change Order/Drawing Change fees. These fees take place after the second Submittal Revision or the first Fabrication Revision. Please speak with your Skytron Representative if you have any questions with regards to these fees. All final tie-ins of electrical connections, plumbing and media must be made by a qualified and licensed individual. Skytron does not provide final tie-in services due to local licensing regulations. Finish work (e.g., caulking and trim) is the responsibility of others. Installation of standard product moldings or trim is included in the pricing provided. PLEASE READ THE FOLLOWING CAREFULLY EACH DRAWING MUST BE INITIALED AND DATED Accepted by: (signature) Accepted by: (print) Requested Delivery Date Date: P.O. #: New Preliminary Package Drawing Package Revision Summary 0 11-14-17 Q-18272 Accepted - with changes Rejected Accepted - as is Equipment References Drawing Index # Note that not all drawings will be required for each mount / drawings package A (C-21554) B (C-21556) C (C-21558) D (C-21561) E (C-21562) F (C-21563) G (C-21564) H (C-21570) I (C-21572) 00a Room Layout - Contains overhead view of Skytron equipment Integration Elevation Details - Contains elevation drawings of the Integration system Carrier Details - Contains front, left and right side views of equipment carriers detailing accessories and outlet placement Medical Gas Details - Contains medical gas details required for the riser plate installation Electrical Wiring Detail - Contains detailed wiring diagram and circuit requirements for equipment booms and lights Communication Details - Contains a breakdown of required communications cabling and connections Light Fixture Details - Contains details of light wiring for fixture mounts and back box details S1a _2 _3 _4 _5 _6 Integration Mounting Details - Contains mounting details of the Integration system S1b Specification for the SKYTRON products found within this information package are specifically designated for your facility. Please ensure that all of your custom specifications have been included. Deviating from the depicted specific equipment placement could cause equipment conflicts in the room and must be approved by a Skytron Representative. 1 2 3 4 5 6 7 _1 _1.1 Elevation Details - Contains elevation view drawings on equipment Mounting Structure Details - Contains mounting structure details of equipment CHILDREN'S HOSPITAL NOLA New Submittal Package 11-16-17 Q-18272 Added VAC to C & D Updated Layouts 12-8-17 Q-18272 Updated equipment schedules Updated gasses on mounts C & D Updated VST lengths for mounts C & I 6-4-17 Q-18272 Updated Item IDs 6-18-17 Q-18272

Transcript of CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311...

Page 1: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

Rev # Date/Quote Description

I have read the appropriate requirements from the Skytron Pre-Installation Guide for Ergon booms and consulted with the individualtrades. I understand there are specific ceiling height, medical gas, electrical, video/communications and structural requirements that

must be supplied by the appropriate trade representatives to support this project. Please be aware that modification to the site specificdocument could result in additional Change Order/Drawing Change fees. These fees take place after the second Submittal Revision orthe first Fabrication Revision. Please speak with your Skytron Representative if you have any questions with regards to these fees. All

final tie-ins of electrical connections, plumbing and media must be made by a qualified and licensed individual. Skytron does not providefinal tie-in services due to local licensing regulations. Finish work (e.g., caulking and trim) is the responsibility of others. Installation of

standard product moldings or trim is included in the pricing provided.

PLEASE READ THE FOLLOWING CAREFULLY

EACH DRAWING MUST BE INITIALED AND DATED

Accepted by:(signature)

Accepted by:(print)

Requested Delivery Date

Date:

P.O. #:

New Preliminary Package

Drawing Package Revision Summary

0 11-14-17Q-18272

Accepted - with changes RejectedAccepted - as is

Equipment References DrawingIndex # Note that not all drawings will be required for each mount / drawings package

A (C-21554)B (C-21556)C (C-21558)D (C-21561)E (C-21562)F (C-21563)G (C-21564)H (C-21570)I (C-21572)

00a Room Layout - Contains overhead view of Skytron equipment

Integration Elevation Details - Contains elevation drawings of the Integration system

Carrier Details - Contains front, left and right side views of equipment carriers detailing accessories and outlet placement

Medical Gas Details - Contains medical gas details required for the riser plateinstallation

Electrical Wiring Detail - Contains detailed wiring diagram and circuit requirements for equipment booms and lights

Communication Details - Contains a breakdown of required communications cabling and connections

Light Fixture Details - Contains details of light wiring for fixture mounts and back box details

S1a

_2_3

_4

_5_6

Integration Mounting Details - Contains mounting details of the Integration systemS1b

Specification for the SKYTRON products found within this information package are specifically designated for your facility. Please ensure that all of your custom specifications have been included. Deviating from the depicted specific equipmentplacement could cause equipment conflicts in the room and must be approved by a Skytron Representative.

1

2

3

4

5

6

7

_1_1.1

Elevation Details - Contains elevation view drawings on equipment Mounting Structure Details - Contains mounting structure details of equipment

CHILDREN'S HOSPITAL NOLA

New Submittal Package11-16-17Q-18272

Added VAC to C & DUpdated Layouts

12-8-17Q-18272

Updated equipment schedulesUpdated gasses on mounts C & DUpdated VST lengths for mounts C & I

6-4-17Q-18272

Updated Item IDs6-18-17Q-18272

Page 2: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

D2302

O.R.3

D2

PERFWOR

51" [1300mm]

35.25"/895mm

43"/1

090m

m

35.25"/895mm

3'-6"

3'-6"

5' 5'

ID 11021

ID 1767

ID 12461

ID 10966

DESC

RIPT

ION:

ROO

M LA

YOUT

00aSHEETINITIAL:

DATE:

CHILD

REN'

S HO

SPIT

ALNO

LAPR

OJEC

T #:

17-3

28SU

BMIT

TAL

PLOT

DAT

E:

ROOM

TYP

E: H

EART

OR

REV.

#: 0

6/20/2

018

*SITE SPECIFIC LAYOUT*

* IF SKYVISION IS APPLICABLE REFER TO THE LAST PORTION OF BOOM PACKAGE*

DIMENSION PROVIDED FROM WALLS ARE BASED ON CURRENTINFORMATION PROVIDED TO SKYTRON AT TIME OF DOCUMENT CREATION.

IN THE EVENT THAT FIXTURE LOCATIONS NEED TO CHANGE PLEASE CHECK WITH SKYTRON ON THE REVISED LAYOUT LOCATION

FINAL CONSTRUCTION DOCUMENTS SHOULD BE PROVIDED BY THE ARCHITECT.

EQUIPMENT SCHEDULE

ID DESCRIPTION

A ID 12461

C ID 1767

D ID 10966

H ID 11021

Page 3: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

D2334

PERFUSIONWORKROOM

D2304

O.R.4

51" [1300mm]

35.25"/895mm 43"/1090mm

35.25"/895mm

3'-6"3'-6"

5'

5' ID 10966

ID 12461

ID 11021

ID 1767

DESC

RIPT

ION:

ROO

M LA

YOUT

01aSHEETINITIAL:

DATE:

CHILD

REN'

S HO

SPIT

ALNO

LARO

OM T

YPE:

GEN

OR

REV.

#: 0

6/20/2

018

*SITE SPECIFIC LAYOUT*

* IF SKYVISION IS APPLICABLE REFER TO THE LAST PORTION OF BOOM PACKAGE*

DIMENSION PROVIDED FROM WALLS ARE BASED ON CURRENTINFORMATION PROVIDED TO SKYTRON AT TIME OF DOCUMENT CREATION.

IN THE EVENT THAT FIXTURE LOCATIONS NEED TO CHANGE PLEASE CHECK WITH SKYTRON ON THE REVISED LAYOUT LOCATION

FINAL CONSTRUCTION DOCUMENTS SHOULD BE PROVIDED BY THE ARCHITECT.

PROJ

ECT

#: 17

-328

SUBM

ITTA

LPL

OT D

ATE:

EQUIPMENT SCHEDULE

ID DESCRIPTION

A ID 12461

C ID 1767

D ID 10966

H ID 11021

Page 4: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

D2305

O.R.5

D2306

O.R.6

4 1 1 /

2 " [10

50mm ]

35.25"/895mm

35.25

"/895

mm

3'-6"

3'-6"

5'

4 1 1 /

2 " [10

50mm ]

35.25"/895mm

35.25

"/895

mm

3'-6"

3'-6"

5'

ID 10966

ID 11793

ID 1767

ID 11793

ID 10966

ID 1767

DESC

RIPT

ION:

ROO

M LA

YOUT

02aSHEETINITIAL:

DATE:

CHILD

REN'

S HO

SPIT

ALNO

LARO

OM T

YPE:

GEN

OR

REV.

#: 0

6/20/2

018

EQUIPMENT SCHEDULE

ID DESCRIPTION

B ID 11793

C ID 1767

D ID 10966

*SITE SPECIFIC LAYOUT*

* IF SKYVISION IS APPLICABLE REFER TO THE LAST PORTION OF BOOM PACKAGE*

DIMENSION PROVIDED FROM WALLS ARE BASED ON CURRENTINFORMATION PROVIDED TO SKYTRON AT TIME OF DOCUMENT CREATION.

IN THE EVENT THAT FIXTURE LOCATIONS NEED TO CHANGE PLEASE CHECK WITH SKYTRON ON THE REVISED LAYOUT LOCATION

FINAL CONSTRUCTION DOCUMENTS SHOULD BE PROVIDED BY THE ARCHITECT.

PROJ

ECT

#: 17

-328

SUBM

ITTA

LPL

OT D

ATE:

Page 5: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

D2307

O.R.7

D2306

O.R.6

41 1/

2" [1

050m

m]

35.25"/895mm

35.25

"/895

mm

3'-6"

3'-6"

5'

ID 11793

ID 10966

ID 1767

DESC

RIPT

ION:

ROO

M LA

YOUT

03aSHEETINITIAL:

DATE:

CHILD

REN'

S HO

SPIT

ALNO

LARO

OM T

YPE:

GEN

OR

REV.

#: 0

6/20/2

018

*SITE SPECIFIC LAYOUT*

* IF SKYVISION IS APPLICABLE REFER TO THE LAST PORTION OF BOOM PACKAGE*

DIMENSION PROVIDED FROM WALLS ARE BASED ON CURRENTINFORMATION PROVIDED TO SKYTRON AT TIME OF DOCUMENT CREATION.

IN THE EVENT THAT FIXTURE LOCATIONS NEED TO CHANGE PLEASE CHECK WITH SKYTRON ON THE REVISED LAYOUT LOCATION

FINAL CONSTRUCTION DOCUMENTS SHOULD BE PROVIDED BY THE ARCHITECT.

EQUIPMENT SCHEDULE

ID DESCRIPTION

B ID 11793

C ID 1767

D ID 10966

PROJ

ECT

#: 17

-328

SUBM

ITTA

LPL

OT D

ATE:

Page 6: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

D2308

(SMALL)O.R.

8

41 1/2" [1050mm]

35.25"/895mm

35.25"/895mm

3'-6" 3'-6"

5'

ID 10966

ID 11793

ID 1767

DESC

RIPT

ION:

ROO

M LA

YOUT

04aSHEETINITIAL:

DATE:

CHILD

REN'

S HO

SPIT

ALNO

LARO

OM T

YPE:

GEN

OR

REV.

#: 0

6/20/2

018

*SITE SPECIFIC LAYOUT*

* IF SKYVISION IS APPLICABLE REFER TO THE LAST PORTION OF BOOM PACKAGE*

DIMENSION PROVIDED FROM WALLS ARE BASED ON CURRENTINFORMATION PROVIDED TO SKYTRON AT TIME OF DOCUMENT CREATION.

IN THE EVENT THAT FIXTURE LOCATIONS NEED TO CHANGE PLEASE CHECK WITH SKYTRON ON THE REVISED LAYOUT LOCATION

FINAL CONSTRUCTION DOCUMENTS SHOULD BE PROVIDED BY THE ARCHITECT.

EQUIPMENT SCHEDULE

ID DESCRIPTION

B ID 11793

C ID 1767

D ID 10966

PROJ

ECT

#: 17

-328

SUBM

ITTA

LPL

OT D

ATE:

Page 7: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

D2310

STORAGED2311

O.R.10

41 1/ 2" [1050mm]

35.25"/895mm

35.25"/895mm

3'-6" 3'-6"

5'

41 1/2" [1050mm]

35.25"/895mm

35.25"/895mm

3'-6"

3'-6"

5'

ID 10966

ID 11793

ID 11793

ID 1767

ID 10966ID 1767

DESC

RIPT

ION:

ROO

M LA

YOUT

05aSHEETINITIAL:

DATE:

CHILD

REN'

S HO

SPIT

ALNO

LARO

OM T

YPE:

GEN

OR

REV.

#: 0

6/20/2

018

*SITE SPECIFIC LAYOUT*

* IF SKYVISION IS APPLICABLE REFER TO THE LAST PORTION OF BOOM PACKAGE*

DIMENSION PROVIDED FROM WALLS ARE BASED ON CURRENTINFORMATION PROVIDED TO SKYTRON AT TIME OF DOCUMENT CREATION.

IN THE EVENT THAT FIXTURE LOCATIONS NEED TO CHANGE PLEASE CHECK WITH SKYTRON ON THE REVISED LAYOUT LOCATION

FINAL CONSTRUCTION DOCUMENTS SHOULD BE PROVIDED BY THE ARCHITECT.

EQUIPMENT SCHEDULE

ID DESCRIPTION

B ID 11793

C ID 1767

D ID 10966

PROJ

ECT

#: 17

-328

SUBM

ITTA

LPL

OT D

ATE:

Page 8: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

D2312

O.R.11

41 1/2" [1050mm]

35.25"/895mm

35.25"/895mm

3'-6"

3'-6"

5'

41 1/2" [1050mm]

35.25"/895mm

35.25"/895mm

3'-6"

3'-6"

5'

ID 11793ID 11793

ID 10966

ID 1767

ID 10966

ID 1767

DESC

RIPT

ION:

ROO

M LA

YOUT

06aSHEETINITIAL:

DATE:

CHILD

REN'

S HO

SPIT

ALNO

LARO

OM T

YPE:

GEN

OR

REV.

#: 0

6/20/2

018

*SITE SPECIFIC LAYOUT*

* IF SKYVISION IS APPLICABLE REFER TO THE LAST PORTION OF BOOM PACKAGE*

DIMENSION PROVIDED FROM WALLS ARE BASED ON CURRENTINFORMATION PROVIDED TO SKYTRON AT TIME OF DOCUMENT CREATION.

IN THE EVENT THAT FIXTURE LOCATIONS NEED TO CHANGE PLEASE CHECK WITH SKYTRON ON THE REVISED LAYOUT LOCATION

FINAL CONSTRUCTION DOCUMENTS SHOULD BE PROVIDED BY THE ARCHITECT.

EQUIPMENT SCHEDULE

ID DESCRIPTION

B ID 11793

C ID 1767

D ID 10966

PROJ

ECT

#: 17

-328

SUBM

ITTA

LPL

OT D

ATE:

Page 9: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

D2314

O.R.1

41 1/2" [1050mm]

35.25"/895mm

35.25"/895mm

3'-6"

3'-6"

5'

ID 11793

ID 10966

ID 1767

DESC

RIPT

ION:

ROO

M LA

YOUT

07aSHEETINITIAL:

DATE:

CHILD

REN'

S HO

SPIT

ALNO

LARO

OM T

YPE:

GEN

OR

REV.

#: 0

6/20/2

018

*SITE SPECIFIC LAYOUT*

* IF SKYVISION IS APPLICABLE REFER TO THE LAST PORTION OF BOOM PACKAGE*

DIMENSION PROVIDED FROM WALLS ARE BASED ON CURRENTINFORMATION PROVIDED TO SKYTRON AT TIME OF DOCUMENT CREATION.

IN THE EVENT THAT FIXTURE LOCATIONS NEED TO CHANGE PLEASE CHECK WITH SKYTRON ON THE REVISED LAYOUT LOCATION

FINAL CONSTRUCTION DOCUMENTS SHOULD BE PROVIDED BY THE ARCHITECT.

EQUIPMENT SCHEDULE

ID DESCRIPTION

B ID 11793

C ID 1767

D ID 10966

PROJ

ECT

#: 17

-328

SUBM

ITTA

LPL

OT D

ATE:

Page 10: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

OR

D2301

(SMALL)O.R.

2

41 1/2" [1050mm]

35.2

5"/89

5mm

35.25"/895mm

3'-6"3'-6"

5'

ID 10966

ID 11793

ID 1767

DESC

RIPT

ION:

ROO

M LA

YOUT

08aSHEETINITIAL:

DATE:

CHILD

REN'

S HO

SPIT

ALNO

LARO

OM T

YPE:

GEN

OR

REV.

#: 0

6/20/2

018

*SITE SPECIFIC LAYOUT*

* IF SKYVISION IS APPLICABLE REFER TO THE LAST PORTION OF BOOM PACKAGE*

DIMENSION PROVIDED FROM WALLS ARE BASED ON CURRENTINFORMATION PROVIDED TO SKYTRON AT TIME OF DOCUMENT CREATION.

IN THE EVENT THAT FIXTURE LOCATIONS NEED TO CHANGE PLEASE CHECK WITH SKYTRON ON THE REVISED LAYOUT LOCATION

FINAL CONSTRUCTION DOCUMENTS SHOULD BE PROVIDED BY THE ARCHITECT.

EQUIPMENT SCHEDULE

ID DESCRIPTION

B ID 11793

C ID 1767

D ID 10966

PROJ

ECT

#: 17

-328

SUBM

ITTA

LPL

OT D

ATE:

Page 11: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

B1144A

B1144

ORTHO

35.25

"/895

mm

27.5"/700mm

5'

CAUTIONTHIS BOOM HAS

ROTATIONALCONFLICTS

ID 4903 ID 12173

DESC

RIPT

ION:

ROO

M LA

YOUT

09aSHEETINITIAL:

DATE:

CHILD

REN'

S HO

SPIT

ALNO

LARO

OM T

YPE:

ORT

HORE

V. #:

06/2

0/201

8

EQUIPMENT SCHEDULE

ID DESCRIPTION

G ID 4903

I ID 12173

*SITE SPECIFIC LAYOUT*

* IF SKYVISION IS APPLICABLE REFER TO THE LAST PORTION OF BOOM PACKAGE*

DIMENSION PROVIDED FROM WALLS ARE BASED ON CURRENTINFORMATION PROVIDED TO SKYTRON AT TIME OF DOCUMENT CREATION.

IN THE EVENT THAT FIXTURE LOCATIONS NEED TO CHANGE PLEASE CHECK WITH SKYTRON ON THE REVISED LAYOUT LOCATION

FINAL CONSTRUCTION DOCUMENTS SHOULD BE PROVIDED BY THE ARCHITECT.

PROJ

ECT

#: 17

-328

SUBM

ITTA

LPL

OT D

ATE:

Page 12: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

1C5

B1145BB1C

B1145

RESUSCITATION 1

CAUTIONTHIS BOOM HAS

ROTATIONALCONFLICTS

35.25"/895mm ID 4903ID 1441

DESC

RIPT

ION:

ROO

M LA

YOUT

10aSHEETINITIAL:

DATE:

CHILD

REN'

S HO

SPIT

ALNO

LARO

OM T

YPE:

RES

USCI

TATI

ONRE

V. #:

06/2

0/201

8

EQUIPMENT SCHEDULE

ID DESCRIPTION

F ID 1441

G ID 4903

*SITE SPECIFIC LAYOUT*

* IF SKYVISION IS APPLICABLE REFER TO THE LAST PORTION OF BOOM PACKAGE*

DIMENSION PROVIDED FROM WALLS ARE BASED ON CURRENTINFORMATION PROVIDED TO SKYTRON AT TIME OF DOCUMENT CREATION.

IN THE EVENT THAT FIXTURE LOCATIONS NEED TO CHANGE PLEASE CHECK WITH SKYTRON ON THE REVISED LAYOUT LOCATION

FINAL CONSTRUCTION DOCUMENTS SHOULD BE PROVIDED BY THE ARCHITECT.

PROJ

ECT

#: 17

-328

SUBM

ITTA

LPL

OT D

ATE:

Page 13: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

1C5

B1145BB1C

B1146

RES 2

35.25"/895mm ID 4903ID 1441

DESC

RIPT

ION:

ROO

M LA

YOUT

11aSHEETINITIAL:

DATE:

CHILD

REN'

S HO

SPIT

ALNO

LA6/2

0/201

8

*SITE SPECIFIC LAYOUT*

* IF SKYVISION IS APPLICABLE REFER TO THE LAST PORTION OF BOOM PACKAGE*

DIMENSION PROVIDED FROM WALLS ARE BASED ON CURRENTINFORMATION PROVIDED TO SKYTRON AT TIME OF DOCUMENT CREATION.

IN THE EVENT THAT FIXTURE LOCATIONS NEED TO CHANGE PLEASE CHECK WITH SKYTRON ON THE REVISED LAYOUT LOCATION

FINAL CONSTRUCTION DOCUMENTS SHOULD BE PROVIDED BY THE ARCHITECT.

ROOM

TYP

E: R

ESUS

CITA

TION

REV.

#: 0

EQUIPMENT SCHEDULE

ID DESCRIPTION

F ID 1441

G ID 4903 PROJ

ECT

#: 17

-328

SUBM

ITTA

LPL

OT D

ATE:

Page 14: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

E2542AA E2542AB

E2543

ENDO E2542A

DECONTAM.

E2542

ENDO

E2542B

E2542AC

41 1/2" [1050mm]

41 1/2" [1050mm]

ID 1768ID 1768

DESC

RIPT

ION:

ROO

M LA

YOUT

12aSHEETINITIAL:

DATE:

CHILD

REN'

S HO

SPIT

ALNO

LA6/2

0/201

8

EQUIPMENT SCHEDULE

ID DESCRIPTION

E ID 1768

*SITE SPECIFIC LAYOUT*

* IF SKYVISION IS APPLICABLE REFER TO THE LAST PORTION OF BOOM PACKAGE*

DIMENSION PROVIDED FROM WALLS ARE BASED ON CURRENTINFORMATION PROVIDED TO SKYTRON AT TIME OF DOCUMENT CREATION.

IN THE EVENT THAT FIXTURE LOCATIONS NEED TO CHANGE PLEASE CHECK WITH SKYTRON ON THE REVISED LAYOUT LOCATION

FINAL CONSTRUCTION DOCUMENTS SHOULD BE PROVIDED BY THE ARCHITECT.

ROOM

TYP

E: E

NDO

REV.

#: 0

PROJ

ECT

#: 17

-328

SUBM

ITTA

LPL

OT D

ATE:

Page 15: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

E2547

DENTALPROCEDURE

B113

P

35.25"/895mm

CAUTIONTHIS BOOM HAS

ROTATIONALCONFLICTS

2'-1"

5'-11"

ID 10966

ID 1441

DESC

RIPT

ION:

ROO

M LA

YOUT

13aSHEETINITIAL:

DATE:

CHILD

REN'

S HO

SPIT

ALNO

LA6/2

0/201

8

EQUIPMENT SCHEDULE

ID DESCRIPTION

D ID 10966

F ID 1441

*SITE SPECIFIC LAYOUT*

* IF SKYVISION IS APPLICABLE REFER TO THE LAST PORTION OF BOOM PACKAGE*

DIMENSION PROVIDED FROM WALLS ARE BASED ON CURRENTINFORMATION PROVIDED TO SKYTRON AT TIME OF DOCUMENT CREATION.

IN THE EVENT THAT FIXTURE LOCATIONS NEED TO CHANGE PLEASE CHECK WITH SKYTRON ON THE REVISED LAYOUT LOCATION

FINAL CONSTRUCTION DOCUMENTS SHOULD BE PROVIDED BY THE ARCHITECT.

ROOM

TYP

E: D

ENTA

LRE

V. #:

0PR

OJEC

T #:

17-3

28SU

BMIT

TAL

PLOT

DAT

E:

Page 16: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

B1135A

B1180A B1185AE2548 E2549

MINORPROCEDURE MINOR

PROCEDURE

35.25"/895mm

35.25"/895mm

ID 1441 ID 1441

DESC

RIPT

ION:

ROO

M LA

YOUT

14aSHEETINITIAL:

DATE:

CHILD

REN'

S HO

SPIT

ALNO

LA6/2

0/201

8

EQUIPMENT SCHEDULE

ID DESCRIPTION

F ID 1441

*SITE SPECIFIC LAYOUT*

* IF SKYVISION IS APPLICABLE REFER TO THE LAST PORTION OF BOOM PACKAGE*

DIMENSION PROVIDED FROM WALLS ARE BASED ON CURRENTINFORMATION PROVIDED TO SKYTRON AT TIME OF DOCUMENT CREATION.

IN THE EVENT THAT FIXTURE LOCATIONS NEED TO CHANGE PLEASE CHECK WITH SKYTRON ON THE REVISED LAYOUT LOCATION

FINAL CONSTRUCTION DOCUMENTS SHOULD BE PROVIDED BY THE ARCHITECT.

ROOM

TYP

E: M

INOR

PRO

CEDU

RERE

V. #:

0PR

OJEC

T #:

17-3

28SU

BMIT

TAL

PLOT

DAT

E:

Page 17: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

6'-6"

6'-8" 2'10'-10"

41 1/2" [1050mm]

35.25

"/895

mm

35.25

"/895

mm

CAUTIONSURGICAL LIGHT PROVIDEDPER RFP092217 WILL NOT

REACH THE INTENDEDTARGET AREA

CAUTIONSURGICAL LIGHT PROVIDEDPER RFP092217 WILL NOT

REACH THE INTENDEDTARGET AREA

ID 1767

ID 10966ID 11793

DESC

RIPT

ION:

ROO

M LA

YOUT

15aSHEETINITIAL:

DATE:

CHILD

REN'

S HO

SPIT

ALNO

LA6/2

0/201

8

EQUIPMENT SCHEDULE

ID DESCRIPTION

B ID 11793

C ID 1767

D ID 10966

*SITE SPECIFIC LAYOUT*

* IF SKYVISION IS APPLICABLE REFER TO THE LAST PORTION OF BOOM PACKAGE*

DIMENSION PROVIDED FROM WALLS ARE BASED ON CURRENTINFORMATION PROVIDED TO SKYTRON AT TIME OF DOCUMENT CREATION.

IN THE EVENT THAT FIXTURE LOCATIONS NEED TO CHANGE PLEASE CHECK WITH SKYTRON ON THE REVISED LAYOUT LOCATION

FINAL CONSTRUCTION DOCUMENTS SHOULD BE PROVIDED BY THE ARCHITECT.

ROOM

TYP

E: H

YBRI

D OR

REV.

#: 0

PROJ

ECT

#: 17

-328

SUBM

ITTA

LPL

OT D

ATE:

Page 18: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

Nx6 3

T

3

T4

35.25"/895mm

35.25"

/895m

m

CAUTIONSURGICAL LIGHT PROVIDEDPER RFP092217 WILL NOT

REACH THE INTENDEDTARGET AREA

7'

3'-3"

9'-9"

6'

CAUTIONSURGICAL LIGHT PROVIDEDPER RFP092217 WILL NOT

REACH THE INTENDEDTARGET AREA

ID 10966

ID 1441

ID 1767

DESC

RIPT

ION:

ROO

M LA

YOUT

16aSHEETINITIAL:

DATE:

CHILD

REN'

S HO

SPIT

ALNO

LA6/2

0/201

8

*SITE SPECIFIC LAYOUT*

* IF SKYVISION IS APPLICABLE REFER TO THE LAST PORTION OF BOOM PACKAGE*

DIMENSION PROVIDED FROM WALLS ARE BASED ON CURRENTINFORMATION PROVIDED TO SKYTRON AT TIME OF DOCUMENT CREATION.

IN THE EVENT THAT FIXTURE LOCATIONS NEED TO CHANGE PLEASE CHECK WITH SKYTRON ON THE REVISED LAYOUT LOCATION

FINAL CONSTRUCTION DOCUMENTS SHOULD BE PROVIDED BY THE ARCHITECT.

ROOM

TYP

E: C

ATH

LAB

REV.

#: 0

EQUIPMENT SCHEDULE

ID DESCRIPTION

C ID 1767

D ID 10966

F ID 1441

PROJ

ECT

#: 17

-328

SUBM

ITTA

LPL

OT D

ATE:

Page 19: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

72"

120" CEILING HEIGHT

0" FLOOR

10"

712"

3912" 3Q1-L

3112" 32S

11"

11"

3UBM-W

82"

ITEM ID: 12461

MDL:

200 S

ERIE

SQT

Y.: 2

RE

V. #:

0

A1SHEETINITIAL:

DATE:

DESC

RIPT

ION:

ELE

VATI

ON D

ETAI

LS

CHILD

REN'

S HO

SPIT

ALNO

LA

*SITE SPECIFIC ELEVATION DETAILS*

PROJ

ECT

#: 17

-323

SUBM

ITTA

LPL

OT D

ATE:

6/20/2

018PLEASE REFER TO MOUNTING

STRUCTURE DETAIL PAGE FORMOUNTING STRUCTURE EXAMPLE

SCALE: 34" = 1'MAX. FIXTURE WEIGHT: MAX. MOMENT LOAD: EQUIPMENT CAPACITY:597 LBS. 2310 FT. LBS. 3UBMW - 108 LBS.

Page 20: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

STRUCTURAL REQUIREMENTS - Architect and Structural Engineer

Mounting Structure ComponentsThe fabrication of each mounting structure may be slightly different but they each require the same basic components to ensurestability.

Sway Bracing (by others)Sway bracing is designed to rigidly affix the mounting plate to the structural ceiling. The primary purpose of sway bracing is toeliminate sway, or lateral twisting and flexing of the mounting structure as it "reacts" to dynamic load changes caused by moving thefixture radial arms. The sway bracing should be welded to the mounting plate and extend away from the center of the mount. Aminimum of four sway braces placed 90° apart and positioned at a 45° and 60° angle is recommended.

Minimum recommended material for sway bracing is 3" x 3" x 1/4" angle iron. It is recommended that in all applications thesway bracing be fastened to the structural ceiling.

Structural Ceiling Plate (by others)The structural ceiling plate rigidly attaches the mount to the structural ceiling using structural anchors appropriate for the ceilingconstruction. The structural ceiling plate should be a minimum of, 1" ASTM A36 steel plate, 17" diameter with (6) 5/8" diameterholes for structural anchors and is fabricated by others.

Expansion Anchors (by others)Test 50% of the anchors at 2,000 pounds (907 kg) tension, or 50 ft. lb. (68 N●m) torque per CBC 1925A.3.5. Installed anchors must meet the following criteria:1. Hydraulic Ram Method: The anchor should have no observable movement at the applicable test load. For wedge and sleevetype anchors, a practical way to determine observable movement is that the washer under the nut becomes loose.2. Torque Wrench Method (Wedge or Sleeve Type): The applicable test torque must be reached within one-half (1/2) turn of the nut.Testing should occur no sooner than 24 hours after installation of anchors. If any anchor fails testing, test all anchors until 20 consecutive anchors pass, then resume the initial testing frequency. Test equipment is to be calibrated by an approved testinglaboratory in accordance with standard recognized procedures.

Support Tube (by others)The support tube required to attach the mounting plate to the structural ceiling plate should be ASTM 500 Grade B, 6" diameter tube.Support tube is to be welded to structural ceiling plate and mounting plate. A minimum of 6 gussets placed 60° apart should bewelded to support tube at the structural ceiling plate and the mounting plate.

Mounting Plate (SKYTRON supplied)The 17.5" x 17.5" x 1" ASTM A36 steel mounting plate is a SKYTRON supplied item. The Support tube and sway bracing are welded to the mounting plate. The mounting plate contains the corresponding bolt pattern for attaching the fixture and provides themounting areas for the junction box and gas riser plates.

Mounting Structure DesignSeismic structural applications differ. Please contact your local SKYTRON representative for specific calculations. Themounting structure must be designed and fabricated to position the bottom of the SKYTRON mounting plate as shown on sitespecific elevation page. This is a critical dimension in order to accommodate proper clearance required for ceiling cover function. The mounting plate must be perfectly level (+/- 0.1°) and allow no more than two-tenths of a degree (0.2°) of rotation atthe mounting plate when the specified load is applied. The mounting structure must be tested for strength and stiffness prior toinstallation of the fixture. Please contact your SKYTRON representative to schedule testing.

Please consult your SKYTRON representative during early stages of construction to facilitate this process. The testingprocess is a required, documented function prior to closing of the finished ceiling.

Ceiling RequirementsA 24" x 24" access door must be mounted adjacent to the mounting structure for entry by service personnel for serviceaccess.

SKYTRON provides a 24" ceiling cover designed to cover 20" diameter ceiling hole cutout.

Additional Items(SKYTRON supplied)In addition to the pre-installation kit. SKYTRON provides the following items:

(6) 1-1/4" x 10" threaded rods. (24) 1-1/4" hex nuts, pump enclosure (if applicable)

STRUCTURALCEILING

STRUCTURALCEILING

STRUCTURAL CONCRETE SLABNORMAL WEIGHTfc=3,000 PSI (20,684 kPa)

SWAY BRACING3" x 3" x 1/4" (4)

GUSSET 6" DIA.SUPPORT

TUBE

FINISHED CEILING

20"DIAMETER OPENING

REQUIRED(CENTERED UNDER MOUNT)

12" MIN. / 60" MAX.

REFER TO JOB SPECIFIC ELEVATION PAGE (-0/+1)

MOUNTING PLATE(SKYTRON SUPPLIED)

45°- 60°

FIXTURE HUBTYP. 1/4"

PJP

ACCESS DOOR24" x 24"

(REQUIRED)

TYP. 1/4"TOP & BOTTOM

STRUCTURALCEILING PLATE

17" DIA. x 1"

EXPANSION ANCHORS:HILTI KB-III (ICBO 4627)

CARBON STEEL

NOTES:• This illustration depicts recommended mounting structure design and its components. Always consult

specific structural criteria defined by a structural engineer.• Do not cover or block any holes with sway bracing, gussets, weld or weld slag.• Typical dimensions shown. Refer to specific structural drawings and/or Seismic drawings for each

application.• *Critical Dimension

12" HIGH CLEAR SPACEABOVE MOUNTING PLATE

REQUIRED ON 3 SIDES FORGAS AND ELECTRICAL

COMPONENT INSTALLATION

*TYPICAL BOOM MOUNTING STRUCTURE DETAILS**ALWAYS CONSULT SPECIFIC STRUCTURAL CRITERIA DEFINED BY A STRUCTURAL ENGINEER*

NOTES:• This illustration depicts a generic mounting structure design and its components. Always consult

specific structural criteria defined by a structural engineer.• 1 14" Mounting bolts and nuts shipped with fixture.

JUNCTION BOX(SKYTRON SUPPLIED)

BRACKET(SKYTRON SUPPLIED)

PLATE(SKYTRON SUPPLIED)

INLET TUBE(SKYTRON SUPPLIED)

RISER PLATE(SKYTRON SUPPLIED)

CHECK VALVEREMOVAL TOOL(SKYTRON SUPPLIED)

A1.1SHEETINITIAL:

DATE:

DESC

RIPT

ION:

ELE

VATI

ON D

ETAI

LS11

/16/20

17

CHILD

REN'

S HO

SPIT

ALNO

LAMD

L: 20

0 SER

IES

QTY.

: 2

REV.

#: 0

PROJ

ECT

#: 17

-328

SU

BMIT

TAL

PLOT

DAT

E:

Page 21: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

DIMENSIONS:INITIAL:

ELECTRICAL:GAS OUTLETS: A2

PLO

T DA

TE: 1

1/16

/201

7

CHIL

DREN

'S H

OSP

ITAL

NOLA

PRO

JECT

#: 1

7-32

8

DESC

RIPT

ION:

CAR

RIER

DET

AILS

SHEET

SUBM

ITTA

LQ

TY: 2

REV

#: 0

DATE:

MDL

: 200

SER

IES

CARRIER

*SITE SPECIFIC CARRIER DETAILS*

DISS (8) 125V, 20A DUPLEX - RED8 1/2"H x 26"W x 16 1/2"D

ACCESSORY LISTNONE

QTY GASES

1 CARBON DIOXIDE

0 HELIOX

2 MEDICAL AIR

0 NITROGEN

1 NITROUS OXIDE

2 OXYGEN

2 VACUUM

2 WAGD

ONLY PLACESCOMMUNICATIONS

CAN BE ADDED

VACUUM

VACUUM

NITROUS OXIDEMEDICAL AIR

CARBON DIOXIDE

WAGDOXYGEN

SEPARATOR PLATE

Page 22: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

A3SHEET

MEDICAL GAS REQUIREMENTS - Medical Gas / Piping EngineerNotes:

• Beacon Medaes manufactured gas outlets supplied unless otherwise specified. Pricing and lead time could vary if another Mfgr. is chosen.• Medical gas risers comply with NFPA 9999-2012 OR CAN-CSA Z 73961-12 Edition Cleanliness of gas outlets must be maintained through

installation.• Purge gas lines and test at least 24 hrs. prior to installation.• Test results must be made available to SKYTRON representative for verification & comparison.• Check valve must be removed during installation to prevent damage to o-rings, refer to Ergon pre-installation guide for removal

instructions.** Caution - Check valves have 2 different sizes, please keep Instrument Air and Oxygen separate from the others **

Each boom fixture is fabricated in accordance to the specifications required by the customer. The configuration drawings supplied bySKYTRON will indicate the type and quantity of gas supply lines required. D.I.S.S. connection medical grade hoses connect the fixture tothe riser plate connectors. The customer is responsible to deliver the appropriate medical gas from the facility supply to the riser plateconnectors and for the design of a medical gas system with adequate flow capacity capable of compensating for the accumulative flowrestrictions associated with conventional construction methods i.e. flex gas hoses. SKYTRON provides medical gas riser plate(s) and theappropriate connectors for attachment to the Mounting Plate. The riser plates attach to the mounting plate and will accommodate up to 12gas connectors. The connectors are D.I.S.S. type medical gas connectors with removable single check valves and provide a 1/2" coppertube for attachment to facility supply lines. Caps are provided with Vacuum & WAGD gas connectors to facilitate testing procedures.

All connection and testing of medical gas piping to be performed in accordance with NFPA 99-2012 OR CAN-CSA Z 73961-12requirements.

Recommended Gas Riser PlacementA 6" clearance must be maintained between the finished ceiling and the gas inlet tubes. If this clearance cannot be accomplished atthe mounting plate, then horizontal or remote mounting of the riser plate will be required.When mount is in center of room, riser plate should face head end of room, when mount is near walls, riser plate should be adjacentto wall. If more than one riser plate is required they should be placed next to each other.

GAS CONNECTOR ASSEMBLY (w/ SECONDARY CHECK)*SUPPLIED w/FIXTURE*

9"

DISS CONNECTOR(w/ REMOVABLE PRIMARY

CHECK. PASSIVE EVAC,VACUUM & WAGD DO NOT

HAVE PRIMARY CHECK.)*SUPPLIED w/PRE-INSTALL KIT*

D I S SCHEMETRON

CGA ColorStandard

Note #1 - Any room (Critical Care Area) designed for a permanently located respiratory ventilator or anesthesia machine shall have an outlet capable of a transient flow rate of:170 LPM (6 SCFM) for 3 seconds at the station outlet.

Note #2 - For testing and certification purposes, individual station inlets shall be capable of delivering a flow rate of: 3 SCFM, while maintaining a system pressure of not lessthan 12" (300mm) at the nearest adjacent vacuum inlet. Facility supply must be 115 LPM MINIMUM. (Vacuum D.I.S.S. connectors omit primary check valves for optimal flow). 12in/HG.

Note #3 - WAGD (Waste Anesthetic Gas Disposal) systems employing a design where the WAGD lines are "tied in" to MedVac lines must produce the same flow rates as theMedVac inlets.

Note #4 - Nitrogen system requires nitrogen supplied directly from facility supply line rated at 185psi MIN to 200psi MAX. Avoid designs which feature multiple-in-line Nitrogencontrol systems in order to avoid loss of flow capability.

Note #5 - Pipe sizing: piping systems shall be designed and sized to deliver the required flow rates at the utilization pressures while accounting for natural flow losses throughvarious components including flexible hoses and terminal outlets. The Skytron medical gas and vacuum outlets meet and exceed these standards at the time of manufacture. However piping source capacity, sizing, and restrictions may prevent outlets from attaining required valves.

Additional references: Health Care Facilities code 2012 edition, Section 5.1.12.3.10.2, 5.1.12.3.10.3, 5.1.12.3.10.4, 5.1.12.3.10.5

• TYPE K COPPERTUBING

• 1/2" O.D. EXCEPTPASSIVE EVAC

• 3/4" O.D. FOR PASSIVEEVAC

Medical Air(Yellow)

Test GasISO ColorStandardMedical Air

(Black)

AbbreviatedName

MedAir

StandardPressure

50 - 55 psig

MaximumPressure

55 psig

AllowablePressure Drop

5 psig

Minimum Flow Rates

3.5 SCFM per outlet (100NL/min)N2NF see note #1, 5Carbon Dioxide

(Gray)Carbon Dioxide

(Gray) CO2 50 - 55 psig 55 psig 5 psig 3.5 SCFM per outlet (100NL/min)N2NFHeliox

(Brown)Heliox

(Brown) Heliox 50 - 55 psig 55 psig 5 psig 3.5 SCFM per outlet (100NL/min)N2NFNitrogen(Black)

Nitrogen(Black) N2 or HPN2 160 - 185 psig 200 psig 5 psig

5 SCFM per outlet (140NL/min)free air per outletN2NF see note #4, 5

Nitrous Oxide(Blue)

Nitrous Oxide(Blue) N2O 50 - 55 psig 55 psig 5 psig 3.5 SCFM per outlet (100NL/min)N2NF

Oxygen(Green)

Oxygen(White) O2 50 - 55 psig 55 psig 5 psig 3.5 SCFM per outlet (100NL/min)N2NF see note #1

Vacuum(White)

Vacuum(Yellow) MedVac 12in/Hg (300mm) N/A 3 SCFM per outlet (85NL/min) see note #2

Waste AnestheticGas Disposal

(Purple) WAGDVaries withsystem type

At a minimum, each inlet must be able todraw a continuous 50 lpm (1.8 SCFM)

through the interface

Waste AnestheticGas Disposal

(Purple) see note #3

PURITANOHMEDA

DESC

RIPT

ION:

MED

ICAL

GAS

DET

AILS

*SITE SPECIFIC GAS DETAILS* 11/16

/2017

CHILD

REN'

S HO

SPIT

ALNO

LAPR

OJEC

T #:

17-3

28SU

BMIT

TAL

PLOT

DAT

E:

MDL:

200 S

ERIE

SQT

Y.: 2

REV.

#: 0

MEDI

CAL G

AS

MAX OF 12 INLET TUBES PER MOUNTING PLATE. IF MORE ARE REQUIREDTHEY WILL NEED REMOTE MOUNTING.

CHECK VALVE REPLACEMENTTOOL

VALVES

Use above for:Medical AirCarbon DioxideHelioxNitrous OxideNitrogen

Use above for:Instrument AirOxygen

1.18"1.10"

see note #5

see note #5

see note #5

CGA ColorStandard

Nitrogen(Black)

Test GasISO ColorStandard

Nitrogen(Black)

AbbreviatedName

AllowablePressure Drop Minimum Flow Rates

Vacuum(White)

Vacuum(Yellow)

N2 or HPN2

MedVac

(10 psi) 70kpa 400L/min (14.1scfm)@1,250kpa

30L/min (1.1scfm)@54kpa

N2NF

N2NF

INITIAL:DATE:

VERIFY AND INITIAL GAS FACEPLATE STYLE REQUESTED

(4 inhg) 13kpa

CAN-CSA section 7.3.2. Z 7396.1-12

N2NF Other Gas Various MedVac 120L/min (4.2scfm)@345kpa(4 psi) 28kpa

Page 23: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

INITIAL:DATE: INITIAL:

DESC

RIPT

ION:

ELE

CTRI

CAL W

IRIN

G DE

TAILS

A4SHEET

11/16

/2017

REQUIRED FOR FABRICATIONVERIFY AND INITIAL

POWER TYPE INITIAL:

4 INP

UT LE

ADS

120 V

AC, 2

0 AMP

120 V

AC, 2

0 AMP

120 V

AC, 2

0 AMP

CIRC

UIT

2

CIRC

UIT

3

CIRC

UIT

4

(8) 1

25V,

20A,

DUP

LEX

OUTL

ETS

- RED

*SITE SPECIFIC WIRING DETAILS* INTERNAL FIXTURE WIRING TYPICAL, EXTRA FLEX STEEL CONDUIT

TYPE: 12AWG, 600V, XHHW-2, 90°C - UNLESS NOTED

MOUN

TING

HUB

T B T B T B T B

ELECTRICAL REQUIREMENTS - Electrical Engineer

Each boom fixture is fabricated in accordance to the specifications required by the customer. The Configuration drawings supplied by SKYTRONwill indicate the type and quantity of circuits required. SKYTRON provides all wiring and electrical materials for connection from fixture to junctionbox or pump enclosure (if applicable). SKYTRON supplies an electrical junction box (8-5/8" x 4-5/8" x 1-3/4") to facilitate field wiring for up to sixcircuits that is mounted on the mounting plate in the correct position and if applicable, a hydraulic pump enclosure/junction box (18"L x 6"W x 12"H) that is to be remote mounted within 24" of the mounting structure (by contractor). The pump enclosure can be shipped with the installationkit upon request. Typical wire type is 12AWG, 600V, XHHW-2. Each circuit requires a separate, properly circuit protected, 120VAC, 60Hz powersupply line enclosed in rigid metal conduit. All electrical materials for connection to SKYTRON supplied junction box or pump enclosure and installation labor for such materials to be provided by customer. All wiring and materials to be in accordance with federal, state and local codes.It is the customer's responsibility to meet conformity to NFPA and NEC standards with respect to the number of receptacles provided in a patientcare area.

Specific conductor colors and/or wiring for isolated applications are available upon request.

ARM COVERARM COVER

ARM COVERARM COVER

GROUND(FOR ELECTRICAL)

PROTECTED EARTH(FOR CHASSIS)

NORMALDUPLEX

CRITICALDUPLEX

CRITICALDUPLEX

ULTRAFLEXTYPE USL CONDUIT

ULTRAFLEXTYPE USL CONDUIT

TB

TERMINALBLOCKDETAIL

120VAC, 20ASUPPLY

TB

ISOLATED POWERBROWN W/YELLOW STRIPE, ORANGE W/BLUE STRIPE,

GREEN W/YELLOW STRIPE

NON-ISOLATED POWERBLACK, WHITE, GREEN

Internally LocatedElectrical Terminal

Block Detail

120VAC, 20ASUPPLY

carrier contains aninternally located

Terminal Blockthat provides

interconnectivitybetween the

outlets and theinternal conductors

CARR

IER

UTILI

TIES

PAN

EL

TB

TB

CIRC

UIT

112

0 VAC

, 20 A

MP

GENERIC BOOM ELECTRICAL WIRING DIAGRAM FOR FIXED ARMS

CHILD

REN'

S HO

SPIT

ALNO

LAPR

OJEC

T #:

17-3

28SU

BMIT

TAL

PLOT

DAT

E:

MDL:

200 S

ERIE

SQT

Y.: 2

REV.

#: 0

Page 24: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

72"

120" CEILING HEIGHT

0" FLOOR

4112" 3T2-C

10"

36" 3AFC

18"

12"35"

REFER TO AUA3 DATA SHEETFOR SURGICAL LIGHT

SPECIFICATIONS

21"

3FLM-32

86 12"

80 12"

ITEM ID: 11793

B1SHEETINITIAL:

DATE:

DESC

RIPT

ION:

ELE

VATI

ON D

ETAI

LS

*SITE SPECIFIC ELEVATION DETAILS*

6/20/2

018PLEASE REFER TO MOUNTING

STRUCTURE DETAIL PAGE FORMOUNTING STRUCTURE EXAMPLE

MDL:

110 S

ERIE

SQT

Y.: 1

1RE

V. #:

0

SCALE: 34" = 1'MAX. FIXTURE WEIGHT: MAX. MOMENT LOAD: EQUIPMENT CAPACITY:484 LBS. 2207 FT. LBS. 3FLM1-32 - 64.5 LBS.

CHILD

REN'

S HO

SPIT

ALNO

LAPR

OJEC

T #:

17-3

23SU

BMIT

TAL

PLOT

DAT

E:

Page 25: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

STRUCTURAL REQUIREMENTS - Architect and Structural Engineer

Mounting Structure ComponentsThe fabrication of each mounting structure may be slightly different but they each require the same basic components to ensurestability.

Sway Bracing (by others)Sway bracing is designed to rigidly affix the mounting plate to the structural ceiling. The primary purpose of sway bracing is toeliminate sway, or lateral twisting and flexing of the mounting structure as it "reacts" to dynamic load changes caused by moving thefixture radial arms. The sway bracing should be welded to the mounting plate and extend away from the center of the mount. Aminimum of four sway braces placed 90° apart and positioned at a 45° and 60° angle is recommended.

Minimum recommended material for sway bracing is 3" x 3" x 1/4" angle iron. It is recommended that in all applications thesway bracing be fastened to the structural ceiling.

Structural Ceiling Plate (by others)The structural ceiling plate rigidly attaches the mount to the structural ceiling using structural anchors appropriate for the ceilingconstruction. The structural ceiling plate should be a minimum of, 1" ASTM A36 steel plate, 17" diameter with (6) 5/8" diameterholes for structural anchors and is fabricated by others.

Expansion Anchors (by others)Test 50% of the anchors at 2,000 pounds (907 kg) tension, or 50 ft. lb. (68 N●m) torque per CBC 1925A.3.5. Installed anchors must meet the following criteria:1. Hydraulic Ram Method: The anchor should have no observable movement at the applicable test load. For wedge and sleevetype anchors, a practical way to determine observable movement is that the washer under the nut becomes loose.2. Torque Wrench Method (Wedge or Sleeve Type): The applicable test torque must be reached within one-half (1/2) turn of the nut.Testing should occur no sooner than 24 hours after installation of anchors. If any anchor fails testing, test all anchors until 20 consecutive anchors pass, then resume the initial testing frequency. Test equipment is to be calibrated by an approved testinglaboratory in accordance with standard recognized procedures.

Support Tube (by others)The support tube required to attach the mounting plate to the structural ceiling plate should be ASTM 500 Grade B, 6" diameter tube.Support tube is to be welded to structural ceiling plate and mounting plate. A minimum of 6 gussets placed 60° apart should bewelded to support tube at the structural ceiling plate and the mounting plate.

Mounting Plate (SKYTRON supplied)The 17.5" x 17.5" x 1" ASTM A36 steel mounting plate is a SKYTRON supplied item. The Support tube and sway bracing are welded to the mounting plate. The mounting plate contains the corresponding bolt pattern for attaching the fixture and provides themounting areas for the junction box and gas riser plates.

Mounting Structure DesignSeismic structural applications differ. Please contact your local SKYTRON representative for specific calculations. Themounting structure must be designed and fabricated to position the bottom of the SKYTRON mounting plate as shown on sitespecific elevation page. This is a critical dimension in order to accommodate proper clearance required for ceiling cover function. The mounting plate must be perfectly level (+/- 0.1°) and allow no more than two-tenths of a degree (0.2°) of rotation atthe mounting plate when the specified load is applied. The mounting structure must be tested for strength and stiffness prior toinstallation of the fixture. Please contact your SKYTRON representative to schedule testing.

Please consult your SKYTRON representative during early stages of construction to facilitate this process. The testingprocess is a required, documented function prior to closing of the finished ceiling.

Ceiling RequirementsA 24" x 24" access door must be mounted adjacent to the mounting structure for entry by service personnel for serviceaccess.

SKYTRON provides a 24" ceiling cover designed to cover 20" diameter ceiling hole cutout.

Additional Items(SKYTRON supplied)In addition to the pre-installation kit. SKYTRON provides the following items:

(6) 1-1/4" x 10" threaded rods. (24) 1-1/4" hex nuts, pump enclosure (if applicable)

STRUCTURALCEILING

STRUCTURALCEILING

STRUCTURAL CONCRETE SLABNORMAL WEIGHTfc=3,000 PSI (20,684 kPa)

SWAY BRACING3" x 3" x 1/4" (4)

GUSSET 6" DIA.SUPPORT

TUBE

FINISHED CEILING

20"DIAMETER OPENING

REQUIRED(CENTERED UNDER MOUNT)

12" MIN. / 60" MAX.

REFER TO JOB SPECIFIC ELEVATION PAGE (-0/+1)

MOUNTING PLATE(SKYTRON SUPPLIED)

45°- 60°

FIXTURE HUBTYP. 1/4"

PJP

ACCESS DOOR24" x 24"

(REQUIRED)

TYP. 1/4"TOP & BOTTOM

STRUCTURALCEILING PLATE

17" DIA. x 1"

EXPANSION ANCHORS:HILTI KB-III (ICBO 4627)

CARBON STEEL

NOTES:• This illustration depicts recommended mounting structure design and its components. Always consult

specific structural criteria defined by a structural engineer.• Do not cover or block any holes with sway bracing, gussets, weld or weld slag.• Typical dimensions shown. Refer to specific structural drawings and/or Seismic drawings for each

application.• *Critical Dimension

12" HIGH CLEAR SPACEABOVE MOUNTING PLATE

REQUIRED ON 3 SIDES FORGAS AND ELECTRICAL

COMPONENT INSTALLATION

*TYPICAL BOOM MOUNTING STRUCTURE DETAILS**ALWAYS CONSULT SPECIFIC STRUCTURAL CRITERIA DEFINED BY A STRUCTURAL ENGINEER*

NOTES:• This illustration depicts a generic mounting structure design and its components. Always consult

specific structural criteria defined by a structural engineer.• 1 14" Mounting bolts and nuts shipped with fixture.

JUNCTION BOX(SKYTRON SUPPLIED)

BRACKET(SKYTRON SUPPLIED)

PLATE(SKYTRON SUPPLIED)

INLET TUBE(SKYTRON SUPPLIED)

RISER PLATE(SKYTRON SUPPLIED)

CHECK VALVEREMOVAL TOOL(SKYTRON SUPPLIED)

B1.1SHEETINITIAL:

DATE:

DESC

RIPT

ION:

ELE

VATI

ON D

ETAI

LS11

/16/20

17

CHILD

REN'

S HO

SPIT

ALNO

LAMD

L: 11

0 SER

IES

QTY.

: 11

REV.

#: 0

PROJ

ECT

#: 17

-328

SU

BMIT

TAL

PLOT

DAT

E:

Page 26: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

TYPE USL CONDUITBY SKYTRON

SPECIAL GROUNDING REQUIREMENTS - Electrical Engineer

1. Use of approved metal conduit shall be employed throughout the fixture's wiring circuit where applicable. Facility supplied, circuit breaker protected, 100-240VAC 50/60 Hz power source wiring.

2. Grounding - Proper performance and safety of this fixture can only achieved by an adequate grounding system. Fixture ground must be a dedicated ground point ultimately bonded to the facilities grounding system to prevent the migration of electricalinterference generated by other devices.

Protective Means - To avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective earth ground. This fixture requires a properly circuit protected, appropriately sized, dedicated circuit. An isolated powersupply circuit must be protected by an appropriately sized double pole, single throw circuit breaker

BY SKYTRON

BY OTHERS

See note 1 See note 2

THIS DIAGRAM IS FOR INFORMATIONAL PURPOSES ONLY. THIS WILL NOT MATCH YOUR EXACT MODEL.

JUNCTION BOX BY OTHERWITHIN 12" OF MOUNT 3/4" METAL CONDUIT

100-240VAC INPUT DEDICATED

3/4" METAL CONDUIT 12AWG, 3 WIRES PERLIGHTHEAD PLUS FIXTURE GROUNDRED, GREEN/YELLOW, BLUE 24VDC OUTPUT

POWER SUPPLY JUNCTION BOXWITHIN 3' OF MOUNT(SUPPLIED BY SKYTRON)

B6SHEETINITIAL:

DATE:

DESC

RIPT

ION:

LIGH

T FI

XTUR

E DE

TAILS

11/16

/2017

MDL:

110 S

ERIE

SQT

Y.: 1

1RE

V. #:

0CH

ILDRE

N'S

HOSP

ITAL

NOLA

PROJ

ECT

#: 17

-328

SUBM

ITTA

LPL

OT D

ATE:

Page 27: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

72"

120" CEILING HEIGHT

0" FLOOR

10"

312"

51" 3EC

3912" 32F-L

28"

3VBM48 67"

20"

35"

REFER TO AUA3 DATA SHEETFOR SURGICAL LIGHT

SPECIFICATIONS

21"

80 12"

ITEM ID: 1767

C1SHEETINITIAL:

DATE:

DESC

RIPT

ION:

ELE

VATI

ON D

ETAI

LS

*SITE SPECIFIC ELEVATION DETAILS*

6/20/2

018PLEASE REFER TO MOUNTING

STRUCTURE DETAIL PAGE FORMOUNTING STRUCTURE EXAMPLE

MDL:

310 S

ERIE

SQT

Y.: 1

4RE

V. #:

1

SCALE: 34" = 1'MAX. FIXTURE WEIGHT: MAX. MOMENT LOAD: EQUIPMENT CAPACITY:877 LBS. 4758 FT. LBS. 3VBM48 - 218 LBS.

CHILD

REN'

S HO

SPIT

ALNO

LAPR

OJEC

T #:

17-3

23SU

BMIT

TAL

PLOT

DAT

E:

Page 28: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

STRUCTURAL REQUIREMENTS - Architect and Structural Engineer

Mounting Structure ComponentsThe fabrication of each mounting structure may be slightly different but they each require the same basic components to ensurestability.

Sway Bracing (by others)Sway bracing is designed to rigidly affix the mounting plate to the structural ceiling. The primary purpose of sway bracing is toeliminate sway, or lateral twisting and flexing of the mounting structure as it "reacts" to dynamic load changes caused by moving thefixture radial arms. The sway bracing should be welded to the mounting plate and extend away from the center of the mount. Aminimum of four sway braces placed 90° apart and positioned at a 45° and 60° angle is recommended.

Minimum recommended material for sway bracing is 3" x 3" x 1/4" angle iron. It is recommended that in all applications thesway bracing be fastened to the structural ceiling.

Structural Ceiling Plate (by others)The structural ceiling plate rigidly attaches the mount to the structural ceiling using structural anchors appropriate for the ceilingconstruction. The structural ceiling plate should be a minimum of, 1" ASTM A36 steel plate, 17" diameter with (6) 5/8" diameterholes for structural anchors and is fabricated by others.

Expansion Anchors (by others)Test 50% of the anchors at 2,000 pounds (907 kg) tension, or 50 ft. lb. (68 N●m) torque per CBC 1925A.3.5. Installed anchors must meet the following criteria:1. Hydraulic Ram Method: The anchor should have no observable movement at the applicable test load. For wedge and sleevetype anchors, a practical way to determine observable movement is that the washer under the nut becomes loose.2. Torque Wrench Method (Wedge or Sleeve Type): The applicable test torque must be reached within one-half (1/2) turn of the nut.Testing should occur no sooner than 24 hours after installation of anchors. If any anchor fails testing, test all anchors until 20 consecutive anchors pass, then resume the initial testing frequency. Test equipment is to be calibrated by an approved testinglaboratory in accordance with standard recognized procedures.

Support Tube (by others)The support tube required to attach the mounting plate to the structural ceiling plate should be ASTM 500 Grade B, 6" diameter tube.Support tube is to be welded to structural ceiling plate and mounting plate. A minimum of 6 gussets placed 60° apart should bewelded to support tube at the structural ceiling plate and the mounting plate.

Mounting Plate (SKYTRON supplied)The 17.5" x 17.5" x 1" ASTM A36 steel mounting plate is a SKYTRON supplied item. The Support tube and sway bracing are welded to the mounting plate. The mounting plate contains the corresponding bolt pattern for attaching the fixture and provides themounting areas for the junction box and gas riser plates.

Mounting Structure DesignSeismic structural applications differ. Please contact your local SKYTRON representative for specific calculations. Themounting structure must be designed and fabricated to position the bottom of the SKYTRON mounting plate as shown on sitespecific elevation page. This is a critical dimension in order to accommodate proper clearance required for ceiling cover function. The mounting plate must be perfectly level (+/- 0.1°) and allow no more than two-tenths of a degree (0.2°) of rotation atthe mounting plate when the specified load is applied. The mounting structure must be tested for strength and stiffness prior toinstallation of the fixture. Please contact your SKYTRON representative to schedule testing.

Please consult your SKYTRON representative during early stages of construction to facilitate this process. The testingprocess is a required, documented function prior to closing of the finished ceiling.

Ceiling RequirementsA 24" x 24" access door must be mounted adjacent to the mounting structure for entry by service personnel for serviceaccess.

SKYTRON provides a 24" ceiling cover designed to cover 20" diameter ceiling hole cutout.

Additional Items(SKYTRON supplied)In addition to the pre-installation kit. SKYTRON provides the following items:

(6) 1-1/4" x 10" threaded rods. (24) 1-1/4" hex nuts, pump enclosure (if applicable)

STRUCTURALCEILING

STRUCTURALCEILING

STRUCTURAL CONCRETE SLABNORMAL WEIGHTfc=3,000 PSI (20,684 kPa)

SWAY BRACING3" x 3" x 1/4" (4)

GUSSET 6" DIA.SUPPORT

TUBE

FINISHED CEILING

20"DIAMETER OPENING

REQUIRED(CENTERED UNDER MOUNT)

12" MIN. / 60" MAX.

REFER TO JOB SPECIFIC ELEVATION PAGE (-0/+1)

MOUNTING PLATE(SKYTRON SUPPLIED)

45°- 60°

FIXTURE HUBTYP. 1/4"

PJP

ACCESS DOOR24" x 24"

(REQUIRED)

TYP. 1/4"TOP & BOTTOM

STRUCTURALCEILING PLATE

17" DIA. x 1"

EXPANSION ANCHORS:HILTI KB-III (ICBO 4627)

CARBON STEEL

NOTES:• This illustration depicts recommended mounting structure design and its components. Always consult

specific structural criteria defined by a structural engineer.• Do not cover or block any holes with sway bracing, gussets, weld or weld slag.• Typical dimensions shown. Refer to specific structural drawings and/or Seismic drawings for each

application.• *Critical Dimension

12" HIGH CLEAR SPACEABOVE MOUNTING PLATE

REQUIRED ON 3 SIDES FORGAS AND ELECTRICAL

COMPONENT INSTALLATION

*TYPICAL BOOM MOUNTING STRUCTURE DETAILS**ALWAYS CONSULT SPECIFIC STRUCTURAL CRITERIA DEFINED BY A STRUCTURAL ENGINEER*

NOTES:• This illustration depicts a generic mounting structure design and its components. Always consult

specific structural criteria defined by a structural engineer.• 1 14" Mounting bolts and nuts shipped with fixture.

JUNCTION BOX(SKYTRON SUPPLIED)

BRACKET(SKYTRON SUPPLIED)

PLATE(SKYTRON SUPPLIED)

INLET TUBE(SKYTRON SUPPLIED)

RISER PLATE(SKYTRON SUPPLIED)

CHECK VALVEREMOVAL TOOL(SKYTRON SUPPLIED)

C1.1SHEETINITIAL:

DATE:

DESC

RIPT

ION:

ELE

VATI

ON D

ETAI

LS11

/16/20

17

CHILD

REN'

S HO

SPIT

ALNO

LAMD

L: 31

0 SER

IES

QTY.

: 14

REV.

#: 0

PROJ

ECT

#: 17

-328

SU

BMIT

TAL

PLOT

DAT

E:

Page 29: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

DISS (8) 125V, 20A DUPLEX - RED48 1/2"H x 22"W x 30"D

ACCESSORY LIST(1) MOUNTING BLOCK FOR VST MONITOR MOUNT(1) FRONT EQUIPMENT RAIL(2) PMSH SHELF(1) PMSH-H SHELF(1) CORD WRAP - RAIL MOUNTED(1) 22" x 19.5" x 7" DRAWER(1) REAR EQUIPMENT RAIL(1) GLOVE BOX HOLDER - RAIL MOUNTED

QTY GASES

1 CARBON DIOXIDE

0 HELIOX

1 MEDICAL AIR

1 NITROGEN

0 NITROUS OXIDE

1 OXYGEN

2 VACUUM

0 WAGD

*COMMUNICATION CANONLY BE ADDED ABOVETHE SEPARATOR PLATE

GAS OUTLETS:DIMENSIONS:

NOLA

CHIL

DREN

'S H

OSP

ITAL

INITIAL:

PLO

T DA

TE: 6

/4/2

018

SUBM

ITTA

LPR

OJE

CT #

: 17-

328

DESC

RIPT

ION:

CAR

RIER

DET

AILS

SHEETC2

QTY

: 14

REV

#: 1

DATE: ELECTRICAL:

MDL

: 310

SER

IES

CARRIER

*SITE SPECIFIC CARRIER DETAILS*

SEPARATOR PLATE

NITROGEN

OXYGEN

MEDICAL AIR

VACUUM

CARBON DIOXIDE

Page 30: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

C3SHEET

MEDICAL GAS REQUIREMENTS - Medical Gas / Piping EngineerNotes:

• Beacon Medaes manufactured gas outlets supplied unless otherwise specified. Pricing and lead time could vary if another Mfgr. is chosen.• Medical gas risers comply with NFPA 9999-2012 OR CAN-CSA Z 73961-12 Edition Cleanliness of gas outlets must be maintained through

installation.• Purge gas lines and test at least 24 hrs. prior to installation.• Test results must be made available to SKYTRON representative for verification & comparison.• Check valve must be removed during installation to prevent damage to o-rings, refer to Ergon pre-installation guide for removal

instructions.** Caution - Check valves have 2 different sizes, please keep Instrument Air and Oxygen separate from the others **

Each boom fixture is fabricated in accordance to the specifications required by the customer. The configuration drawings supplied bySKYTRON will indicate the type and quantity of gas supply lines required. D.I.S.S. connection medical grade hoses connect the fixture tothe riser plate connectors. The customer is responsible to deliver the appropriate medical gas from the facility supply to the riser plateconnectors and for the design of a medical gas system with adequate flow capacity capable of compensating for the accumulative flowrestrictions associated with conventional construction methods i.e. flex gas hoses. SKYTRON provides medical gas riser plate(s) and theappropriate connectors for attachment to the Mounting Plate. The riser plates attach to the mounting plate and will accommodate up to 12gas connectors. The connectors are D.I.S.S. type medical gas connectors with removable single check valves and provide a 1/2" coppertube for attachment to facility supply lines. Caps are provided with Vacuum & WAGD gas connectors to facilitate testing procedures.

All connection and testing of medical gas piping to be performed in accordance with NFPA 99-2012 OR CAN-CSA Z 73961-12requirements.

Recommended Gas Riser PlacementA 6" clearance must be maintained between the finished ceiling and the gas inlet tubes. If this clearance cannot be accomplished atthe mounting plate, then horizontal or remote mounting of the riser plate will be required.When mount is in center of room, riser plate should face head end of room, when mount is near walls, riser plate should be adjacentto wall. If more than one riser plate is required they should be placed next to each other.

GAS CONNECTOR ASSEMBLY (w/ SECONDARY CHECK)*SUPPLIED w/FIXTURE*

9"

DISS CONNECTOR(w/ REMOVABLE PRIMARY

CHECK. PASSIVE EVAC,VACUUM & WAGD DO NOT

HAVE PRIMARY CHECK.)*SUPPLIED w/PRE-INSTALL KIT*

D I S SCHEMETRON

CGA ColorStandard

Note #1 - Any room (Critical Care Area) designed for a permanently located respiratory ventilator or anesthesia machine shall have an outlet capable of a transient flow rate of:170 LPM (6 SCFM) for 3 seconds at the station outlet.

Note #2 - For testing and certification purposes, individual station inlets shall be capable of delivering a flow rate of: 3 SCFM, while maintaining a system pressure of not lessthan 12" (300mm) at the nearest adjacent vacuum inlet. Facility supply must be 115 LPM MINIMUM. (Vacuum D.I.S.S. connectors omit primary check valves for optimal flow). 12in/HG.

Note #3 - WAGD (Waste Anesthetic Gas Disposal) systems employing a design where the WAGD lines are "tied in" to MedVac lines must produce the same flow rates as theMedVac inlets.

Note #4 - Nitrogen system requires nitrogen supplied directly from facility supply line rated at 185psi MIN to 200psi MAX. Avoid designs which feature multiple-in-line Nitrogencontrol systems in order to avoid loss of flow capability.

Note #5 - Pipe sizing: piping systems shall be designed and sized to deliver the required flow rates at the utilization pressures while accounting for natural flow losses throughvarious components including flexible hoses and terminal outlets. The Skytron medical gas and vacuum outlets meet and exceed these standards at the time of manufacture. However piping source capacity, sizing, and restrictions may prevent outlets from attaining required valves.

Additional references: Health Care Facilities code 2012 edition, Section 5.1.12.3.10.2, 5.1.12.3.10.3, 5.1.12.3.10.4, 5.1.12.3.10.5

• TYPE K COPPERTUBING

• 1/2" O.D. EXCEPTPASSIVE EVAC

• 3/4" O.D. FOR PASSIVEEVAC

Medical Air(Yellow)

Test GasISO ColorStandardMedical Air

(Black)

AbbreviatedName

MedAir

StandardPressure

50 - 55 psig

MaximumPressure

55 psig

AllowablePressure Drop

5 psig

Minimum Flow Rates

3.5 SCFM per outlet (100NL/min)N2NF see note #1, 5Carbon Dioxide

(Gray)Carbon Dioxide

(Gray) CO2 50 - 55 psig 55 psig 5 psig 3.5 SCFM per outlet (100NL/min)N2NFHeliox

(Brown)Heliox

(Brown) Heliox 50 - 55 psig 55 psig 5 psig 3.5 SCFM per outlet (100NL/min)N2NFNitrogen(Black)

Nitrogen(Black) N2 or HPN2 160 - 185 psig 200 psig 5 psig

5 SCFM per outlet (140NL/min)free air per outletN2NF see note #4, 5

Nitrous Oxide(Blue)

Nitrous Oxide(Blue) N2O 50 - 55 psig 55 psig 5 psig 3.5 SCFM per outlet (100NL/min)N2NF

Oxygen(Green)

Oxygen(White) O2 50 - 55 psig 55 psig 5 psig 3.5 SCFM per outlet (100NL/min)N2NF see note #1

Vacuum(White)

Vacuum(Yellow) MedVac 12in/Hg (300mm) N/A 3 SCFM per outlet (85NL/min) see note #2

Waste AnestheticGas Disposal

(Purple) WAGDVaries withsystem type

At a minimum, each inlet must be able todraw a continuous 50 lpm (1.8 SCFM)

through the interface

Waste AnestheticGas Disposal

(Purple) see note #3

PURITANOHMEDA

DESC

RIPT

ION:

MED

ICAL

GAS

DET

AILS

*SITE SPECIFIC GAS DETAILS* 11/16

/2017

CHILD

REN'

S HO

SPIT

ALNO

LAPR

OJEC

T #:

17-3

28SU

BMIT

TAL

PLOT

DAT

E:

MEDI

CAL G

AS

MAX OF 12 INLET TUBES PER MOUNTING PLATE. IF MORE ARE REQUIREDTHEY WILL NEED REMOTE MOUNTING.

CHECK VALVE REPLACEMENTTOOL

VALVES

Use above for:Medical AirCarbon DioxideHelioxNitrous OxideNitrogen

Use above for:Instrument AirOxygen

1.18"1.10"

see note #5

see note #5

see note #5

CGA ColorStandard

Nitrogen(Black)

Test GasISO ColorStandard

Nitrogen(Black)

AbbreviatedName

AllowablePressure Drop Minimum Flow Rates

Vacuum(White)

Vacuum(Yellow)

N2 or HPN2

MedVac

(10 psi) 70kpa 400L/min (14.1scfm)@1,250kpa

30L/min (1.1scfm)@54kpa

N2NF

N2NF

INITIAL:DATE:

VERIFY AND INITIAL GAS FACEPLATE STYLE REQUESTED

(4 inhg) 13kpa

CAN-CSA section 7.3.2. Z 7396.1-12

N2NF Other Gas Various MedVac 120L/min (4.2scfm)@345kpa(4 psi) 28kpa

MDL:

310 S

ERIE

SQT

Y.: 1

4RE

V. #:

0

Page 31: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

INITIAL:DATE: INITIAL:

DESC

RIPT

ION:

ELE

CTRI

CAL W

IRIN

G DE

TAILS

C4SHEET

11/16

/2017

REQUIRED FOR FABRICATIONVERIFY AND INITIAL

POWER TYPE INITIAL:

5 INP

UT LE

ADS

120 V

AC, 2

0 AMP

120 V

AC, 2

0 AMP

120 V

AC, 2

0 AMP

CIRC

UIT

2

CIRC

UIT

3

CIRC

UIT

4

(8) 1

25V,

20A,

DUP

LEX

OUTL

ETS

- RED

*SITE SPECIFIC WIRING DETAILS* INTERNAL FIXTURE WIRING TYPICAL, EXTRA FLEX STEEL CONDUIT

TYPE: 12AWG, 600V, XHHW-2, 90°C - UNLESS NOTED

MOUN

TING

HUB

T B T B T B T B

ELECTRICAL REQUIREMENTS - Electrical Engineer

Each boom fixture is fabricated in accordance to the specifications required by the customer. The Configuration drawings supplied by SKYTRONwill indicate the type and quantity of circuits required. SKYTRON provides all wiring and electrical materials for connection from fixture to junctionbox or pump enclosure (if applicable). SKYTRON supplies an electrical junction box (8-5/8" x 4-5/8" x 1-3/4") to facilitate field wiring for up to sixcircuits that is mounted on the mounting plate in the correct position and if applicable, a hydraulic pump enclosure/junction box (18"L x 6"W x 12"H) that is to be remote mounted within 24" of the mounting structure (by contractor). The pump enclosure can be shipped with the installationkit upon request. Typical wire type is 12AWG, 600V, XHHW-2. Each circuit requires a separate, properly circuit protected, 120VAC, 60Hz powersupply line enclosed in rigid metal conduit. All electrical materials for connection to SKYTRON supplied junction box or pump enclosure and installation labor for such materials to be provided by customer. All wiring and materials to be in accordance with federal, state and local codes.It is the customer's responsibility to meet conformity to NFPA and NEC standards with respect to the number of receptacles provided in a patientcare area.

Specific conductor colors and/or wiring for isolated applications are available upon request.

ARM COVERARM COVER

ARM COVERARM COVER

GROUND(FOR ELECTRICAL)

PROTECTED EARTH(FOR CHASSIS)

NORMALDUPLEX

CRITICALDUPLEX

CRITICALDUPLEX

ULTRAFLEXTYPE USL CONDUIT

ULTRAFLEXTYPE USL CONDUIT

TB

TERMINALBLOCKDETAIL

120VAC, 20ASUPPLY

TB

ISOLATED POWERBROWN W/YELLOW STRIPE, ORANGE W/BLUE STRIPE,

GREEN W/YELLOW STRIPE

NON-ISOLATED POWERBLACK, WHITE, GREEN

Internally LocatedElectrical Terminal

Block Detail

120VAC, 20ASUPPLY

carrier contains aninternally located

Terminal Blockthat provides

interconnectivitybetween the

outlets and theinternal conductors

CARR

IER

UTILI

TIES

PAN

EL

TB

TB

CIRC

UIT

112

0 VAC

, 20 A

MP

GENERIC BOOM ELECTRICAL WIRING DIAGRAM FOR FIXED ARMS

120 V

AC, 1

5 AMP

ASTR

O LI

GHT

LIGH

T CI

RCUI

T TO

POW

ER S

UPPL

Y BO

XSE

E PA

GE 6

CIRC

UIT

5

CHILD

REN'

S HO

SPIT

ALNO

LAPR

OJEC

T #:

17-3

28SU

BMIT

TAL

PLOT

DAT

E:

MDL:

310 S

ERIE

SQT

Y.: 1

4RE

V. #:

0

Page 32: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

TYPE USL CONDUITBY SKYTRON

SPECIAL GROUNDING REQUIREMENTS - Electrical Engineer

1. Use of approved metal conduit shall be employed throughout the fixture's wiring circuit where applicable. Facility supplied, circuit breaker protected, 100-240VAC 50/60 Hz power source wiring.

2. Grounding - Proper performance and safety of this fixture can only achieved by an adequate grounding system. Fixture ground must be a dedicated ground point ultimately bonded to the facilities grounding system to prevent the migration of electricalinterference generated by other devices.

Protective Means - To avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective earth ground. This fixture requires a properly circuit protected, appropriately sized, dedicated circuit. An isolated powersupply circuit must be protected by an appropriately sized double pole, single throw circuit breaker

BY SKYTRON

BY OTHERS

See note 1 See note 2

THIS DIAGRAM IS FOR INFORMATIONAL PURPOSES ONLY. THIS WILL NOT MATCH YOUR EXACT MODEL.

JUNCTION BOX BY OTHERWITHIN 12" OF MOUNT 3/4" METAL CONDUIT

100-240VAC INPUT DEDICATED

3/4" METAL CONDUIT 12AWG, 3 WIRES PERLIGHTHEAD PLUS FIXTURE GROUNDRED, GREEN/YELLOW, BLUE 24VDC OUTPUT

POWER SUPPLY JUNCTION BOXWITHIN 3' OF MOUNT(SUPPLIED BY SKYTRON)

C6SHEETINITIAL:

DATE:

DESC

RIPT

ION:

LIGH

T FI

XTUR

E DE

TAILS

11/16

/2017

MDL:

310 S

ERIE

SQT

Y.: 1

4RE

V. #:

0CH

ILDRE

N'S

HOSP

ITAL

NOLA

PROJ

ECT

#: 17

-328

SUBM

ITTA

LPL

OT D

ATE:

Page 33: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

72"

120" CEILING HEIGHT

0" FLOOR

10"

712"

3912" 3Q1-L

3112" 32F

25"

3VBM36 67 12"

30"

ITEM ID: 10966

D1SHEETINITIAL:

DATE:

DESC

RIPT

ION:

ELE

VATI

ON D

ETAI

LS

*SITE SPECIFIC ELEVATION DETAILS*

6/20/2

018PLEASE REFER TO MOUNTING

STRUCTURE DETAIL PAGE FORMOUNTING STRUCTURE EXAMPLE

MDL:

200 S

ERIE

SQT

Y.: 1

5RE

V. #:

0

SCALE: 34" = 1'MAX. FIXTURE WEIGHT: MAX. MOMENT LOAD: EQUIPMENT CAPACITY:597 LBS. 2310 FT. LBS. 3VBM36 - 219 LBS.

CHILD

REN'

S HO

SPIT

ALNO

LAPR

OJEC

T #:

17-3

23SU

BMIT

TAL

PLOT

DAT

E:

Page 34: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

STRUCTURAL REQUIREMENTS - Architect and Structural Engineer

Mounting Structure ComponentsThe fabrication of each mounting structure may be slightly different but they each require the same basic components to ensurestability.

Sway Bracing (by others)Sway bracing is designed to rigidly affix the mounting plate to the structural ceiling. The primary purpose of sway bracing is toeliminate sway, or lateral twisting and flexing of the mounting structure as it "reacts" to dynamic load changes caused by moving thefixture radial arms. The sway bracing should be welded to the mounting plate and extend away from the center of the mount. Aminimum of four sway braces placed 90° apart and positioned at a 45° and 60° angle is recommended.

Minimum recommended material for sway bracing is 3" x 3" x 1/4" angle iron. It is recommended that in all applications thesway bracing be fastened to the structural ceiling.

Structural Ceiling Plate (by others)The structural ceiling plate rigidly attaches the mount to the structural ceiling using structural anchors appropriate for the ceilingconstruction. The structural ceiling plate should be a minimum of, 1" ASTM A36 steel plate, 17" diameter with (6) 5/8" diameterholes for structural anchors and is fabricated by others.

Expansion Anchors (by others)Test 50% of the anchors at 2,000 pounds (907 kg) tension, or 50 ft. lb. (68 N●m) torque per CBC 1925A.3.5. Installed anchors must meet the following criteria:1. Hydraulic Ram Method: The anchor should have no observable movement at the applicable test load. For wedge and sleevetype anchors, a practical way to determine observable movement is that the washer under the nut becomes loose.2. Torque Wrench Method (Wedge or Sleeve Type): The applicable test torque must be reached within one-half (1/2) turn of the nut.Testing should occur no sooner than 24 hours after installation of anchors. If any anchor fails testing, test all anchors until 20 consecutive anchors pass, then resume the initial testing frequency. Test equipment is to be calibrated by an approved testinglaboratory in accordance with standard recognized procedures.

Support Tube (by others)The support tube required to attach the mounting plate to the structural ceiling plate should be ASTM 500 Grade B, 6" diameter tube.Support tube is to be welded to structural ceiling plate and mounting plate. A minimum of 6 gussets placed 60° apart should bewelded to support tube at the structural ceiling plate and the mounting plate.

Mounting Plate (SKYTRON supplied)The 17.5" x 17.5" x 1" ASTM A36 steel mounting plate is a SKYTRON supplied item. The Support tube and sway bracing are welded to the mounting plate. The mounting plate contains the corresponding bolt pattern for attaching the fixture and provides themounting areas for the junction box and gas riser plates.

Mounting Structure DesignSeismic structural applications differ. Please contact your local SKYTRON representative for specific calculations. Themounting structure must be designed and fabricated to position the bottom of the SKYTRON mounting plate as shown on sitespecific elevation page. This is a critical dimension in order to accommodate proper clearance required for ceiling cover function. The mounting plate must be perfectly level (+/- 0.1°) and allow no more than two-tenths of a degree (0.2°) of rotation atthe mounting plate when the specified load is applied. The mounting structure must be tested for strength and stiffness prior toinstallation of the fixture. Please contact your SKYTRON representative to schedule testing.

Please consult your SKYTRON representative during early stages of construction to facilitate this process. The testingprocess is a required, documented function prior to closing of the finished ceiling.

Ceiling RequirementsA 24" x 24" access door must be mounted adjacent to the mounting structure for entry by service personnel for serviceaccess.

SKYTRON provides a 24" ceiling cover designed to cover 20" diameter ceiling hole cutout.

Additional Items(SKYTRON supplied)In addition to the pre-installation kit. SKYTRON provides the following items:

(6) 1-1/4" x 10" threaded rods. (24) 1-1/4" hex nuts, pump enclosure (if applicable)

STRUCTURALCEILING

STRUCTURALCEILING

STRUCTURAL CONCRETE SLABNORMAL WEIGHTfc=3,000 PSI (20,684 kPa)

SWAY BRACING3" x 3" x 1/4" (4)

GUSSET 6" DIA.SUPPORT

TUBE

FINISHED CEILING

20"DIAMETER OPENING

REQUIRED(CENTERED UNDER MOUNT)

12" MIN. / 60" MAX.

REFER TO JOB SPECIFIC ELEVATION PAGE (-0/+1)

MOUNTING PLATE(SKYTRON SUPPLIED)

45°- 60°

FIXTURE HUBTYP. 1/4"

PJP

ACCESS DOOR24" x 24"

(REQUIRED)

TYP. 1/4"TOP & BOTTOM

STRUCTURALCEILING PLATE

17" DIA. x 1"

EXPANSION ANCHORS:HILTI KB-III (ICBO 4627)

CARBON STEEL

NOTES:• This illustration depicts recommended mounting structure design and its components. Always consult

specific structural criteria defined by a structural engineer.• Do not cover or block any holes with sway bracing, gussets, weld or weld slag.• Typical dimensions shown. Refer to specific structural drawings and/or Seismic drawings for each

application.• *Critical Dimension

12" HIGH CLEAR SPACEABOVE MOUNTING PLATE

REQUIRED ON 3 SIDES FORGAS AND ELECTRICAL

COMPONENT INSTALLATION

*TYPICAL BOOM MOUNTING STRUCTURE DETAILS**ALWAYS CONSULT SPECIFIC STRUCTURAL CRITERIA DEFINED BY A STRUCTURAL ENGINEER*

NOTES:• This illustration depicts a generic mounting structure design and its components. Always consult

specific structural criteria defined by a structural engineer.• 1 14" Mounting bolts and nuts shipped with fixture.

JUNCTION BOX(SKYTRON SUPPLIED)

BRACKET(SKYTRON SUPPLIED)

PLATE(SKYTRON SUPPLIED)

INLET TUBE(SKYTRON SUPPLIED)

RISER PLATE(SKYTRON SUPPLIED)

CHECK VALVEREMOVAL TOOL(SKYTRON SUPPLIED)

D1.1SHEETINITIAL:

DATE:

DESC

RIPT

ION:

ELE

VATI

ON D

ETAI

LS11

/16/20

17

CHILD

REN'

S HO

SPIT

ALNO

LAMD

L: 20

0 SER

IES

QTY.

: 15

REV.

#: 0

PROJ

ECT

#: 17

-328

SU

BMIT

TAL

PLOT

DAT

E:

Page 35: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

DISS (8) 125V, 20A DUPLEX - RED39"H x 11"W x 8 1/2"D

ACCESSORY LIST(1) REAR EQUIPMENT RAIL(1) WINGBAR W/ 4 HOOK IV TREE

QTY GASES

0 CARBON DIOXIDE

0 HELIOX

1 MEDICAL AIR

0 NITROGEN

1 NITROUS OXIDE

2 OXYGEN

2 VACUUM

1 WAGD

*COMMUNICATION CANONLY BE ADDED ABOVETHE SEPARATOR PLATE

GAS OUTLETS:DIMENSIONS:

NOLA

CHIL

DREN

'S H

OSP

ITAL

INITIAL:

PLO

T DA

TE: 6

/4/2

018

SUBM

ITTA

LPR

OJE

CT #

: 17-

328

DESC

RIPT

ION:

CAR

RIER

DET

AILS

SHEETD2

QTY

: 15

REV

#: 2

DATE: ELECTRICAL:

MDL

: 200

SER

IES

CARRIER

*SITE SPECIFIC CARRIER DETAILS*

SEPARATOR PLATE

VACUUM

WAGDMEDICAL AIRNITROUS OXIDEOXYGENOXYGEN

VACUUM

Page 36: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

D3SHEET

MEDICAL GAS REQUIREMENTS - Medical Gas / Piping EngineerNotes:

• Beacon Medaes manufactured gas outlets supplied unless otherwise specified. Pricing and lead time could vary if another Mfgr. is chosen.• Medical gas risers comply with NFPA 9999-2012 OR CAN-CSA Z 73961-12 Edition Cleanliness of gas outlets must be maintained through

installation.• Purge gas lines and test at least 24 hrs. prior to installation.• Test results must be made available to SKYTRON representative for verification & comparison.• Check valve must be removed during installation to prevent damage to o-rings, refer to Ergon pre-installation guide for removal

instructions.** Caution - Check valves have 2 different sizes, please keep Instrument Air and Oxygen separate from the others **

Each boom fixture is fabricated in accordance to the specifications required by the customer. The configuration drawings supplied bySKYTRON will indicate the type and quantity of gas supply lines required. D.I.S.S. connection medical grade hoses connect the fixture tothe riser plate connectors. The customer is responsible to deliver the appropriate medical gas from the facility supply to the riser plateconnectors and for the design of a medical gas system with adequate flow capacity capable of compensating for the accumulative flowrestrictions associated with conventional construction methods i.e. flex gas hoses. SKYTRON provides medical gas riser plate(s) and theappropriate connectors for attachment to the Mounting Plate. The riser plates attach to the mounting plate and will accommodate up to 12gas connectors. The connectors are D.I.S.S. type medical gas connectors with removable single check valves and provide a 1/2" coppertube for attachment to facility supply lines. Caps are provided with Vacuum & WAGD gas connectors to facilitate testing procedures.

All connection and testing of medical gas piping to be performed in accordance with NFPA 99-2012 OR CAN-CSA Z 73961-12requirements.

Recommended Gas Riser PlacementA 6" clearance must be maintained between the finished ceiling and the gas inlet tubes. If this clearance cannot be accomplished atthe mounting plate, then horizontal or remote mounting of the riser plate will be required.When mount is in center of room, riser plate should face head end of room, when mount is near walls, riser plate should be adjacentto wall. If more than one riser plate is required they should be placed next to each other.

GAS CONNECTOR ASSEMBLY (w/ SECONDARY CHECK)*SUPPLIED w/FIXTURE*

9"

DISS CONNECTOR(w/ REMOVABLE PRIMARY

CHECK. PASSIVE EVAC,VACUUM & WAGD DO NOT

HAVE PRIMARY CHECK.)*SUPPLIED w/PRE-INSTALL KIT*

D I S SCHEMETRON

CGA ColorStandard

Note #1 - Any room (Critical Care Area) designed for a permanently located respiratory ventilator or anesthesia machine shall have an outlet capable of a transient flow rate of:170 LPM (6 SCFM) for 3 seconds at the station outlet.

Note #2 - For testing and certification purposes, individual station inlets shall be capable of delivering a flow rate of: 3 SCFM, while maintaining a system pressure of not lessthan 12" (300mm) at the nearest adjacent vacuum inlet. Facility supply must be 115 LPM MINIMUM. (Vacuum D.I.S.S. connectors omit primary check valves for optimal flow). 12in/HG.

Note #3 - WAGD (Waste Anesthetic Gas Disposal) systems employing a design where the WAGD lines are "tied in" to MedVac lines must produce the same flow rates as theMedVac inlets.

Note #4 - Nitrogen system requires nitrogen supplied directly from facility supply line rated at 185psi MIN to 200psi MAX. Avoid designs which feature multiple-in-line Nitrogencontrol systems in order to avoid loss of flow capability.

Note #5 - Pipe sizing: piping systems shall be designed and sized to deliver the required flow rates at the utilization pressures while accounting for natural flow losses throughvarious components including flexible hoses and terminal outlets. The Skytron medical gas and vacuum outlets meet and exceed these standards at the time of manufacture. However piping source capacity, sizing, and restrictions may prevent outlets from attaining required valves.

Additional references: Health Care Facilities code 2012 edition, Section 5.1.12.3.10.2, 5.1.12.3.10.3, 5.1.12.3.10.4, 5.1.12.3.10.5

• TYPE K COPPERTUBING

• 1/2" O.D. EXCEPTPASSIVE EVAC

• 3/4" O.D. FOR PASSIVEEVAC

Medical Air(Yellow)

Test GasISO ColorStandardMedical Air

(Black)

AbbreviatedName

MedAir

StandardPressure

50 - 55 psig

MaximumPressure

55 psig

AllowablePressure Drop

5 psig

Minimum Flow Rates

3.5 SCFM per outlet (100NL/min)N2NF see note #1, 5Carbon Dioxide

(Gray)Carbon Dioxide

(Gray) CO2 50 - 55 psig 55 psig 5 psig 3.5 SCFM per outlet (100NL/min)N2NFHeliox

(Brown)Heliox

(Brown) Heliox 50 - 55 psig 55 psig 5 psig 3.5 SCFM per outlet (100NL/min)N2NFNitrogen(Black)

Nitrogen(Black) N2 or HPN2 160 - 185 psig 200 psig 5 psig

5 SCFM per outlet (140NL/min)free air per outletN2NF see note #4, 5

Nitrous Oxide(Blue)

Nitrous Oxide(Blue) N2O 50 - 55 psig 55 psig 5 psig 3.5 SCFM per outlet (100NL/min)N2NF

Oxygen(Green)

Oxygen(White) O2 50 - 55 psig 55 psig 5 psig 3.5 SCFM per outlet (100NL/min)N2NF see note #1

Vacuum(White)

Vacuum(Yellow) MedVac 12in/Hg (300mm) N/A 3 SCFM per outlet (85NL/min) see note #2

Waste AnestheticGas Disposal

(Purple) WAGDVaries withsystem type

At a minimum, each inlet must be able todraw a continuous 50 lpm (1.8 SCFM)

through the interface

Waste AnestheticGas Disposal

(Purple) see note #3

PURITANOHMEDA

DESC

RIPT

ION:

MED

ICAL

GAS

DET

AILS

*SITE SPECIFIC GAS DETAILS* 11/16

/2017

CHILD

REN'

S HO

SPIT

ALNO

LAPR

OJEC

T #:

17-3

28SU

BMIT

TAL

PLOT

DAT

E:

MEDI

CAL G

AS

MAX OF 12 INLET TUBES PER MOUNTING PLATE. IF MORE ARE REQUIREDTHEY WILL NEED REMOTE MOUNTING.

CHECK VALVE REPLACEMENTTOOL

VALVES

Use above for:Medical AirCarbon DioxideHelioxNitrous OxideNitrogen

Use above for:Instrument AirOxygen

1.18"1.10"

see note #5

see note #5

see note #5

CGA ColorStandard

Nitrogen(Black)

Test GasISO ColorStandard

Nitrogen(Black)

AbbreviatedName

AllowablePressure Drop Minimum Flow Rates

Vacuum(White)

Vacuum(Yellow)

N2 or HPN2

MedVac

(10 psi) 70kpa 400L/min (14.1scfm)@1,250kpa

30L/min (1.1scfm)@54kpa

N2NF

N2NF

INITIAL:DATE:

VERIFY AND INITIAL GAS FACEPLATE STYLE REQUESTED

(4 inhg) 13kpa

CAN-CSA section 7.3.2. Z 7396.1-12

N2NF Other Gas Various MedVac 120L/min (4.2scfm)@345kpa(4 psi) 28kpa

MDL:

200 S

ERIE

SQT

Y.: 1

5RE

V. #:

0

Page 37: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

INITIAL:DATE: INITIAL:

DESC

RIPT

ION:

ELE

CTRI

CAL W

IRIN

G DE

TAILS

D4SHEET

11/16

/2017

REQUIRED FOR FABRICATIONVERIFY AND INITIAL

POWER TYPE INITIAL:

4 INP

UT LE

ADS

120 V

AC, 2

0 AMP

120 V

AC, 2

0 AMP

120 V

AC, 2

0 AMP

CIRC

UIT

2

CIRC

UIT

3

CIRC

UIT

4

(8) 1

25V,

20A,

DUP

LEX

OUTL

ETS

- RED

*SITE SPECIFIC WIRING DETAILS* INTERNAL FIXTURE WIRING TYPICAL, EXTRA FLEX STEEL CONDUIT

TYPE: 12AWG, 600V, XHHW-2, 90°C - UNLESS NOTED

MOUN

TING

HUB

T B T B T B T B

ELECTRICAL REQUIREMENTS - Electrical Engineer

Each boom fixture is fabricated in accordance to the specifications required by the customer. The Configuration drawings supplied by SKYTRONwill indicate the type and quantity of circuits required. SKYTRON provides all wiring and electrical materials for connection from fixture to junctionbox or pump enclosure (if applicable). SKYTRON supplies an electrical junction box (8-5/8" x 4-5/8" x 1-3/4") to facilitate field wiring for up to sixcircuits that is mounted on the mounting plate in the correct position and if applicable, a hydraulic pump enclosure/junction box (18"L x 6"W x 12"H) that is to be remote mounted within 24" of the mounting structure (by contractor). The pump enclosure can be shipped with the installationkit upon request. Typical wire type is 12AWG, 600V, XHHW-2. Each circuit requires a separate, properly circuit protected, 120VAC, 60Hz powersupply line enclosed in rigid metal conduit. All electrical materials for connection to SKYTRON supplied junction box or pump enclosure and installation labor for such materials to be provided by customer. All wiring and materials to be in accordance with federal, state and local codes.It is the customer's responsibility to meet conformity to NFPA and NEC standards with respect to the number of receptacles provided in a patientcare area.

Specific conductor colors and/or wiring for isolated applications are available upon request.

ARM COVERARM COVER

ARM COVERARM COVER

GROUND(FOR ELECTRICAL)

PROTECTED EARTH(FOR CHASSIS)

NORMALDUPLEX

CRITICALDUPLEX

CRITICALDUPLEX

ULTRAFLEXTYPE USL CONDUIT

ULTRAFLEXTYPE USL CONDUIT

TB

TERMINALBLOCKDETAIL

120VAC, 20ASUPPLY

TB

ISOLATED POWERBROWN W/YELLOW STRIPE, ORANGE W/BLUE STRIPE,

GREEN W/YELLOW STRIPE

NON-ISOLATED POWERBLACK, WHITE, GREEN

Internally LocatedElectrical Terminal

Block Detail

120VAC, 20ASUPPLY

carrier contains aninternally located

Terminal Blockthat provides

interconnectivitybetween the

outlets and theinternal conductors

CARR

IER

UTILI

TIES

PAN

EL

TB

TB

CIRC

UIT

112

0 VAC

, 20 A

MP

GENERIC BOOM ELECTRICAL WIRING DIAGRAM FOR FIXED ARMS

CHILD

REN'

S HO

SPIT

ALNO

LAPR

OJEC

T #:

17-3

28SU

BMIT

TAL

PLOT

DAT

E:

MDL:

200 S

ERIE

SQT

Y.: 1

5RE

V. #:

0

Page 38: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

72"

120" CEILING HEIGHT

0" FLOOR

10"

312"

51" 3EC

4112" 3T2-C

36" 3AFC

12"

18"

3VBM48 67"

20"

3FLM-32

81"

3912" 32F-L

28"

ITEM ID: 1768

E1SHEETINITIAL:

DATE:

DESC

RIPT

ION:

ELE

VATI

ON D

ETAI

LS

*SITE SPECIFIC ELEVATION DETAILS*

6/20/2

018PLEASE REFER TO MOUNTING

STRUCTURE DETAIL PAGE FORMOUNTING STRUCTURE EXAMPLE

MDL:

320 S

ERIE

SQT

Y.: 2

REV.

#: 0

SCALE: 34" = 1'MAX. FIXTURE WEIGHT: MAX. MOMENT LOAD: EQUIPMENT CAPACITY:1029 LBS. 5754 FT. LBS.

3VBM48 - 218 LBS.3FLM1-32 - 64.5 LBS.

CHILD

REN'

S HO

SPIT

ALNO

LAPR

OJEC

T #:

17-3

23SU

BMIT

TAL

PLOT

DAT

E:

Page 39: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

STRUCTURAL REQUIREMENTS - Architect and Structural Engineer

Mounting Structure ComponentsThe fabrication of each mounting structure may be slightly different but they each require the same basic components to ensurestability.

Sway Bracing (by others)Sway bracing is designed to rigidly affix the mounting plate to the structural ceiling. The primary purpose of sway bracing is toeliminate sway, or lateral twisting and flexing of the mounting structure as it "reacts" to dynamic load changes caused by moving thefixture radial arms. The sway bracing should be welded to the mounting plate and extend away from the center of the mount. Aminimum of four sway braces placed 90° apart and positioned at a 45° and 60° angle is recommended.

Minimum recommended material for sway bracing is 3" x 3" x 1/4" angle iron. It is recommended that in all applications thesway bracing be fastened to the structural ceiling.

Structural Ceiling Plate (by others)The structural ceiling plate rigidly attaches the mount to the structural ceiling using structural anchors appropriate for the ceilingconstruction. The structural ceiling plate should be a minimum of, 1" ASTM A36 steel plate, 17" diameter with (6) 5/8" diameterholes for structural anchors and is fabricated by others.

Expansion Anchors (by others)Test 50% of the anchors at 2,000 pounds (907 kg) tension, or 50 ft. lb. (68 N●m) torque per CBC 1925A.3.5. Installed anchors must meet the following criteria:1. Hydraulic Ram Method: The anchor should have no observable movement at the applicable test load. For wedge and sleevetype anchors, a practical way to determine observable movement is that the washer under the nut becomes loose.2. Torque Wrench Method (Wedge or Sleeve Type): The applicable test torque must be reached within one-half (1/2) turn of the nut.Testing should occur no sooner than 24 hours after installation of anchors. If any anchor fails testing, test all anchors until 20 consecutive anchors pass, then resume the initial testing frequency. Test equipment is to be calibrated by an approved testinglaboratory in accordance with standard recognized procedures.

Support Tube (by others)The support tube required to attach the mounting plate to the structural ceiling plate should be ASTM 500 Grade B, 6" diameter tube.Support tube is to be welded to structural ceiling plate and mounting plate. A minimum of 6 gussets placed 60° apart should bewelded to support tube at the structural ceiling plate and the mounting plate.

Mounting Plate (SKYTRON supplied)The 17.5" x 17.5" x 1" ASTM A36 steel mounting plate is a SKYTRON supplied item. The Support tube and sway bracing are welded to the mounting plate. The mounting plate contains the corresponding bolt pattern for attaching the fixture and provides themounting areas for the junction box and gas riser plates.

Mounting Structure DesignSeismic structural applications differ. Please contact your local SKYTRON representative for specific calculations. Themounting structure must be designed and fabricated to position the bottom of the SKYTRON mounting plate as shown on sitespecific elevation page. This is a critical dimension in order to accommodate proper clearance required for ceiling cover function. The mounting plate must be perfectly level (+/- 0.1°) and allow no more than two-tenths of a degree (0.2°) of rotation atthe mounting plate when the specified load is applied. The mounting structure must be tested for strength and stiffness prior toinstallation of the fixture. Please contact your SKYTRON representative to schedule testing.

Please consult your SKYTRON representative during early stages of construction to facilitate this process. The testingprocess is a required, documented function prior to closing of the finished ceiling.

Ceiling RequirementsA 24" x 24" access door must be mounted adjacent to the mounting structure for entry by service personnel for serviceaccess.

SKYTRON provides a 24" ceiling cover designed to cover 20" diameter ceiling hole cutout.

Additional Items(SKYTRON supplied)In addition to the pre-installation kit. SKYTRON provides the following items:

(6) 1-1/4" x 10" threaded rods. (24) 1-1/4" hex nuts, pump enclosure (if applicable)

STRUCTURALCEILING

STRUCTURALCEILING

STRUCTURAL CONCRETE SLABNORMAL WEIGHTfc=3,000 PSI (20,684 kPa)

SWAY BRACING3" x 3" x 1/4" (4)

GUSSET 6" DIA.SUPPORT

TUBE

FINISHED CEILING

20"DIAMETER OPENING

REQUIRED(CENTERED UNDER MOUNT)

12" MIN. / 60" MAX.

REFER TO JOB SPECIFIC ELEVATION PAGE (-0/+1)

MOUNTING PLATE(SKYTRON SUPPLIED)

45°- 60°

FIXTURE HUBTYP. 1/4"

PJP

ACCESS DOOR24" x 24"

(REQUIRED)

TYP. 1/4"TOP & BOTTOM

STRUCTURALCEILING PLATE

17" DIA. x 1"

EXPANSION ANCHORS:HILTI KB-III (ICBO 4627)

CARBON STEEL

NOTES:• This illustration depicts recommended mounting structure design and its components. Always consult

specific structural criteria defined by a structural engineer.• Do not cover or block any holes with sway bracing, gussets, weld or weld slag.• Typical dimensions shown. Refer to specific structural drawings and/or Seismic drawings for each

application.• *Critical Dimension

12" HIGH CLEAR SPACEABOVE MOUNTING PLATE

REQUIRED ON 3 SIDES FORGAS AND ELECTRICAL

COMPONENT INSTALLATION

*TYPICAL BOOM MOUNTING STRUCTURE DETAILS**ALWAYS CONSULT SPECIFIC STRUCTURAL CRITERIA DEFINED BY A STRUCTURAL ENGINEER*

NOTES:• This illustration depicts a generic mounting structure design and its components. Always consult

specific structural criteria defined by a structural engineer.• 1 14" Mounting bolts and nuts shipped with fixture.

JUNCTION BOX(SKYTRON SUPPLIED)

BRACKET(SKYTRON SUPPLIED)

PLATE(SKYTRON SUPPLIED)

INLET TUBE(SKYTRON SUPPLIED)

RISER PLATE(SKYTRON SUPPLIED)

CHECK VALVEREMOVAL TOOL(SKYTRON SUPPLIED)

E1.1SHEETINITIAL:

DATE:

DESC

RIPT

ION:

ELE

VATI

ON D

ETAI

LS11

/16/20

17

CHILD

REN'

S HO

SPIT

ALNO

LAMD

L: 32

0 SER

IES

QTY.

: 2RE

V. #:

0PR

OJEC

T #:

17-3

28

SUBM

ITTA

LPL

OT D

ATE:

Page 40: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

VACUUM

OXYGENMEDICAL AIR

NITROGEN

VACUUM

SEPARATOR PLATE

DISS (8) 125V, 20A DUPLEX - RED48 1/2"H x 22"W x 30"D

ACCESSORY LIST(1) MOUNTING BLOCK FOR VST MONITOR MOUNT(1) FRONT EQUIPMENT RAIL(1) 3PMSH-H SHELF(2) 3PMSH SHELF(2) CORD WRAP - RAIL MOUNTED(1) GLOVE BOX HOLDER - RAIL MOUNTED(1) REAR EQUIPMENT RAIL(1) 22" x 19.5" x 7" PLASTIC DRAWER

QTY GASES

0 CARBON DIOXIDE

0 HELIOX

1 MEDICAL AIR

1 NITROGEN

0 NITROUS OXIDE

1 OXYGEN

2 VACUUM

0 WAGD*COMMUNICATION CANONLY BE ADDED ABOVETHE SEPARATOR PLATE

GAS OUTLETS:DIMENSIONS:

NOLA

CHIL

DREN

'S H

OSP

ITAL

INITIAL:

PLO

T DA

TE: 1

2/8/

2017

SUBM

ITTA

LPR

OJE

CT #

: 17-

328

DESC

RIPT

ION:

CAR

RIER

DET

AILS

SHEETE2

QTY

: 2

R

EV #

: 1

DATE: ELECTRICAL:

MDL

: 320

SER

IES

CARRIER

*SITE SPECIFIC CARRIER DETAILS*

Page 41: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

E3SHEET

MEDICAL GAS REQUIREMENTS - Medical Gas / Piping EngineerNotes:

• Beacon Medaes manufactured gas outlets supplied unless otherwise specified. Pricing and lead time could vary if another Mfgr. is chosen.• Medical gas risers comply with NFPA 9999-2012 OR CAN-CSA Z 73961-12 Edition Cleanliness of gas outlets must be maintained through

installation.• Purge gas lines and test at least 24 hrs. prior to installation.• Test results must be made available to SKYTRON representative for verification & comparison.• Check valve must be removed during installation to prevent damage to o-rings, refer to Ergon pre-installation guide for removal

instructions.** Caution - Check valves have 2 different sizes, please keep Instrument Air and Oxygen separate from the others **

Each boom fixture is fabricated in accordance to the specifications required by the customer. The configuration drawings supplied bySKYTRON will indicate the type and quantity of gas supply lines required. D.I.S.S. connection medical grade hoses connect the fixture tothe riser plate connectors. The customer is responsible to deliver the appropriate medical gas from the facility supply to the riser plateconnectors and for the design of a medical gas system with adequate flow capacity capable of compensating for the accumulative flowrestrictions associated with conventional construction methods i.e. flex gas hoses. SKYTRON provides medical gas riser plate(s) and theappropriate connectors for attachment to the Mounting Plate. The riser plates attach to the mounting plate and will accommodate up to 12gas connectors. The connectors are D.I.S.S. type medical gas connectors with removable single check valves and provide a 1/2" coppertube for attachment to facility supply lines. Caps are provided with Vacuum & WAGD gas connectors to facilitate testing procedures.

All connection and testing of medical gas piping to be performed in accordance with NFPA 99-2012 OR CAN-CSA Z 73961-12requirements.

Recommended Gas Riser PlacementA 6" clearance must be maintained between the finished ceiling and the gas inlet tubes. If this clearance cannot be accomplished atthe mounting plate, then horizontal or remote mounting of the riser plate will be required.When mount is in center of room, riser plate should face head end of room, when mount is near walls, riser plate should be adjacentto wall. If more than one riser plate is required they should be placed next to each other.

GAS CONNECTOR ASSEMBLY (w/ SECONDARY CHECK)*SUPPLIED w/FIXTURE*

9"

DISS CONNECTOR(w/ REMOVABLE PRIMARY

CHECK. PASSIVE EVAC,VACUUM & WAGD DO NOT

HAVE PRIMARY CHECK.)*SUPPLIED w/PRE-INSTALL KIT*

D I S SCHEMETRON

CGA ColorStandard

Note #1 - Any room (Critical Care Area) designed for a permanently located respiratory ventilator or anesthesia machine shall have an outlet capable of a transient flow rate of:170 LPM (6 SCFM) for 3 seconds at the station outlet.

Note #2 - For testing and certification purposes, individual station inlets shall be capable of delivering a flow rate of: 3 SCFM, while maintaining a system pressure of not lessthan 12" (300mm) at the nearest adjacent vacuum inlet. Facility supply must be 115 LPM MINIMUM. (Vacuum D.I.S.S. connectors omit primary check valves for optimal flow). 12in/HG.

Note #3 - WAGD (Waste Anesthetic Gas Disposal) systems employing a design where the WAGD lines are "tied in" to MedVac lines must produce the same flow rates as theMedVac inlets.

Note #4 - Nitrogen system requires nitrogen supplied directly from facility supply line rated at 185psi MIN to 200psi MAX. Avoid designs which feature multiple-in-line Nitrogencontrol systems in order to avoid loss of flow capability.

Note #5 - Pipe sizing: piping systems shall be designed and sized to deliver the required flow rates at the utilization pressures while accounting for natural flow losses throughvarious components including flexible hoses and terminal outlets. The Skytron medical gas and vacuum outlets meet and exceed these standards at the time of manufacture. However piping source capacity, sizing, and restrictions may prevent outlets from attaining required valves.

Additional references: Health Care Facilities code 2012 edition, Section 5.1.12.3.10.2, 5.1.12.3.10.3, 5.1.12.3.10.4, 5.1.12.3.10.5

• TYPE K COPPERTUBING

• 1/2" O.D. EXCEPTPASSIVE EVAC

• 3/4" O.D. FOR PASSIVEEVAC

Medical Air(Yellow)

Test GasISO ColorStandardMedical Air

(Black)

AbbreviatedName

MedAir

StandardPressure

50 - 55 psig

MaximumPressure

55 psig

AllowablePressure Drop

5 psig

Minimum Flow Rates

3.5 SCFM per outlet (100NL/min)N2NF see note #1, 5Carbon Dioxide

(Gray)Carbon Dioxide

(Gray) CO2 50 - 55 psig 55 psig 5 psig 3.5 SCFM per outlet (100NL/min)N2NFHeliox

(Brown)Heliox

(Brown) Heliox 50 - 55 psig 55 psig 5 psig 3.5 SCFM per outlet (100NL/min)N2NFNitrogen(Black)

Nitrogen(Black) N2 or HPN2 160 - 185 psig 200 psig 5 psig

5 SCFM per outlet (140NL/min)free air per outletN2NF see note #4, 5

Nitrous Oxide(Blue)

Nitrous Oxide(Blue) N2O 50 - 55 psig 55 psig 5 psig 3.5 SCFM per outlet (100NL/min)N2NF

Oxygen(Green)

Oxygen(White) O2 50 - 55 psig 55 psig 5 psig 3.5 SCFM per outlet (100NL/min)N2NF see note #1

Vacuum(White)

Vacuum(Yellow) MedVac 12in/Hg (300mm) N/A 3 SCFM per outlet (85NL/min) see note #2

Waste AnestheticGas Disposal

(Purple) WAGDVaries withsystem type

At a minimum, each inlet must be able todraw a continuous 50 lpm (1.8 SCFM)

through the interface

Waste AnestheticGas Disposal

(Purple) see note #3

PURITANOHMEDA

DESC

RIPT

ION:

MED

ICAL

GAS

DET

AILS

*SITE SPECIFIC GAS DETAILS* 11/16

/2017

CHILD

REN'

S HO

SPIT

ALNO

LAPR

OJEC

T #:

17-3

28SU

BMIT

TAL

PLOT

DAT

E:

MEDI

CAL G

AS

MAX OF 12 INLET TUBES PER MOUNTING PLATE. IF MORE ARE REQUIREDTHEY WILL NEED REMOTE MOUNTING.

CHECK VALVE REPLACEMENTTOOL

VALVES

Use above for:Medical AirCarbon DioxideHelioxNitrous OxideNitrogen

Use above for:Instrument AirOxygen

1.18"1.10"

see note #5

see note #5

see note #5

CGA ColorStandard

Nitrogen(Black)

Test GasISO ColorStandard

Nitrogen(Black)

AbbreviatedName

AllowablePressure Drop Minimum Flow Rates

Vacuum(White)

Vacuum(Yellow)

N2 or HPN2

MedVac

(10 psi) 70kpa 400L/min (14.1scfm)@1,250kpa

30L/min (1.1scfm)@54kpa

N2NF

N2NF

INITIAL:DATE:

VERIFY AND INITIAL GAS FACEPLATE STYLE REQUESTED

(4 inhg) 13kpa

CAN-CSA section 7.3.2. Z 7396.1-12

N2NF Other Gas Various MedVac 120L/min (4.2scfm)@345kpa(4 psi) 28kpa

MDL:

320 S

ERIE

SQT

Y.: 2

RE

V. #:

0

Page 42: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

INITIAL:DATE: INITIAL:

DESC

RIPT

ION:

ELE

CTRI

CAL W

IRIN

G DE

TAILS

E4SHEET

11/16

/2017

REQUIRED FOR FABRICATIONVERIFY AND INITIAL

POWER TYPE INITIAL:

4 INP

UT LE

ADS

120 V

AC, 2

0 AMP

120 V

AC, 2

0 AMP

120 V

AC, 2

0 AMP

CIRC

UIT

2

CIRC

UIT

3

CIRC

UIT

4

(8) 1

25V,

20A,

DUP

LEX

OUTL

ETS

- RED

*SITE SPECIFIC WIRING DETAILS* INTERNAL FIXTURE WIRING TYPICAL, EXTRA FLEX STEEL CONDUIT

TYPE: 12AWG, 600V, XHHW-2, 90°C - UNLESS NOTED

MOUN

TING

HUB

T B T B T B T B

ELECTRICAL REQUIREMENTS - Electrical Engineer

Each boom fixture is fabricated in accordance to the specifications required by the customer. The Configuration drawings supplied by SKYTRONwill indicate the type and quantity of circuits required. SKYTRON provides all wiring and electrical materials for connection from fixture to junctionbox or pump enclosure (if applicable). SKYTRON supplies an electrical junction box (8-5/8" x 4-5/8" x 1-3/4") to facilitate field wiring for up to sixcircuits that is mounted on the mounting plate in the correct position and if applicable, a hydraulic pump enclosure/junction box (18"L x 6"W x 12"H) that is to be remote mounted within 24" of the mounting structure (by contractor). The pump enclosure can be shipped with the installationkit upon request. Typical wire type is 12AWG, 600V, XHHW-2. Each circuit requires a separate, properly circuit protected, 120VAC, 60Hz powersupply line enclosed in rigid metal conduit. All electrical materials for connection to SKYTRON supplied junction box or pump enclosure and installation labor for such materials to be provided by customer. All wiring and materials to be in accordance with federal, state and local codes.It is the customer's responsibility to meet conformity to NFPA and NEC standards with respect to the number of receptacles provided in a patientcare area.

Specific conductor colors and/or wiring for isolated applications are available upon request.

ARM COVERARM COVER

ARM COVERARM COVER

GROUND(FOR ELECTRICAL)

PROTECTED EARTH(FOR CHASSIS)

NORMALDUPLEX

CRITICALDUPLEX

CRITICALDUPLEX

ULTRAFLEXTYPE USL CONDUIT

ULTRAFLEXTYPE USL CONDUIT

TB

TERMINALBLOCKDETAIL

120VAC, 20ASUPPLY

TB

ISOLATED POWERBROWN W/YELLOW STRIPE, ORANGE W/BLUE STRIPE,

GREEN W/YELLOW STRIPE

NON-ISOLATED POWERBLACK, WHITE, GREEN

Internally LocatedElectrical Terminal

Block Detail

120VAC, 20ASUPPLY

carrier contains aninternally located

Terminal Blockthat provides

interconnectivitybetween the

outlets and theinternal conductors

CARR

IER

UTILI

TIES

PAN

EL

TB

TB

CIRC

UIT

112

0 VAC

, 20 A

MP

GENERIC BOOM ELECTRICAL WIRING DIAGRAM FOR FIXED ARMS

CHILD

REN'

S HO

SPIT

ALNO

LAPR

OJEC

T #:

17-3

28SU

BMIT

TAL

PLOT

DAT

E:

MDL:

320 S

ERIE

SQT

Y.: 2

RE

V. #:

0

FLAT

SCRE

EN B

RACK

ETMO

UNTI

NG H

UB

Page 43: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

72"

120" CEILING HEIGHT

0" FLOOR

6"

35"

80"

REFER TO AUA3 DATA SHEETFOR SURGICAL LIGHT

SPECIFICATIONS

29"

ITEM ID: 1441

F1SHEETINITIAL:

DATE:

DESC

RIPT

ION:

ELE

VATI

ON D

ETAI

LS

*SITE SPECIFIC ELEVATION DETAILS*

6/20/2

018PLEASE REFER TO MOUNTING

STRUCTURE DETAIL PAGE FORMOUNTING STRUCTURE EXAMPLE

MDL:

AUA3

-895

QTY.

: 6RE

V. #:

0

SCALE: 34" = 1'MAX. FIXTURE WEIGHT: MAX. MOMENT LOAD: EQUIPMENT CAPACITY:93 LBS. 305 FT. LBS. N/A

CHILD

REN'

S HO

SPIT

ALNO

LAPR

OJEC

T #:

17-3

23SU

BMIT

TAL

PLOT

DAT

E:

Page 44: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

CEILING COVER(SKYTRON SUPPLIED)

FINISHED CEILING

STRUCTURAL CEILING

16-1/2" DIA. ACCESS HOLE20 1/2" DIA. CEILING COVER

15" DIA. MOUNTING. PLATE

2.2"1.1"

FIXTUREJUNCTION

BOX

5.9" MAX3.9" MIN.

3/4" METAL FLEXCONDUIT (see note 3)

45° 45°

STRUCTURAL CEILING

11.8" x 11.8" x .4"STIFFENER PLATE

3/4" ALL THREAD ROD(see note 1)

(4) 2" x 2" ANGLE SWAY BRACING(see note 2)

STRUCTURAL REQUIREMENTS - Architect and Structural Engineer

Notes1. 3/4” support rods located for total support of light, all labor and materials for fabrication are supplied by General Contractor. 3/4” nuts

and washers for support of SKYTRON mounting plate supplied by contractor (8 each required).

2. The mounting structure must be attached to structural ceiling and BRACED TO ALLOW NO TWISTING OR LATERAL MOTION andshall be designed not to provide a degree of deflection greater than two-tenths of a degree at the mounting plate.

3. 3/4” metal conduit and minimum 12AWG wire size (3 wires per lighthead plus fixture ground) required between fixture and the powersource. All metal conduit, wiring, and other electrical materials as well as installation labor for such materials associated with theinstallation of the SKYTRON surgical light to be provided by Electrical Contractor. All installations of SKYTRON procedural lightsshould be under the direct supervision of a SKYTRON representative. All wiring to be in accordance with local codes and by a certifiedelectrician.

This is a GENERAL GUIDELINE ONLY.

NOTES:• This illustration depicts a generic mounting structure design and its components. Always consult

specific structural criteria defined by a structural engineer.• Do not cover or block any holes with sway bracing, gussets, weld, weld slag or etcetera• Typical dimensions shown. Refer to specific structural drawings and/or Seismic drawings for each

application.

*TYPICAL LIGHT MOUNTING STRUCTURE DETAILS**ALWAYS CONSULT SPECIFIC STRUCTURAL CRITERIA DEFINED BY A STRUCTURAL ENGINEER*

Hole forAccess to

FixtureJunction

Box

TOP VIEWMOUNTING PLATE DETAIL

2" x 2"Angle Sway Bracing90° Apart - 4 Each(see note 2)

5.9"

Hole forAccess to

FixtureJunction

Box

TOP VIEWMOUNTING PLATE DETAIL

2" x 2"Angle Sway Bracing90° Apart - 4 Each(see note 2)

11.8"

9 - 1/2"

5.9"

11.8"9 - 1/2"

F1.1SHEETINITIAL:

DATE:

DESC

RIPT

ION:

ELE

VATI

ON D

ETAI

LS11

/17/20

17

CHILD

REN'

S HO

SPIT

ALNO

LAMD

L: AU

A3-8

95QT

Y.: 6

REV.

#: 0

PROJ

ECT

#: 17

-328

SU

BMIT

TAL

PLOT

DAT

E:

Page 45: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

SPECIAL GROUNDING REQUIREMENTS - Electrical Engineer

1. Use of approved metal conduit shall be employed throughout the fixture's wiring circuit where applicable. Facility supplied, circuit breaker protected, 100-240VAC 50/60 Hz power source wiring.

2. Grounding - Proper performance and safety of this fixture can only achieved by an adequate grounding system. Fixture ground must be a dedicated ground point ultimately bonded to the facilities grounding system to prevent the migration of electricalinterference generated by other devices.

Protective Means - To avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective earth ground. This fixture requires a properly circuit protected, appropriately sized, dedicated circuit. An isolated powersupply circuit must be protected by an appropriately sized double pole, single throw circuit breaker

3. Mounting and anchorage: please refer to the ASTRO Installation Manual for mounting requirements. Seismic applications will differ in construction requirements. Please request seismic calculations and mounting requirements from your SKYTRONrepresentative.

BY SKYTRON

BY OTHERS

See note 3

THIS DIAGRAM IS FOR INFORMATIONAL PURPOSES ONLY. THIS WILL NOT MATCH YOUR EXACT MODEL.

3/4" METAL CONDUIT 12AWG, 3 WIRES PERLIGHTHEAD PLUS FIXTURE GROUNDRED, GREEN/YELLOW, BLUE 24VDC OUTPUT

See note 1 See note 2

JUNCTION BOX BY SKYTRON3/4" METAL CONDUIT100-240VAC INPUT DEDICATED

POWER SUPPLY JUNCTION BOXWITHIN 3' OF MOUNT(SUPPLIED BY SKYTRON)

MDL:

AUA3

-895

QTY.

: 6RE

V. #:

0

F6SHEETINITIAL:

DATE:

DESC

RIPT

ION:

LIGH

T FI

XTUR

E DE

TAILS

11/16

/2017

CHILD

REN'

S HO

SPIT

ALNO

LAPR

OJEC

T #:

17-3

28SU

BMIT

TAL

PLOT

DAT

E:

Page 46: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

72"

108" CEILING HEIGHT

0" FLOOR

10"

712"

48" 3E1-L

3912" 32F-L

9"3VBM48 70"

23"

ITEM ID: 4903

G1SHEETINITIAL:

DATE:

DESC

RIPT

ION:

ELE

VATI

ON D

ETAI

LS

*SITE SPECIFIC ELEVATION DETAILS*

6/20/2

018PLEASE REFER TO MOUNTING

STRUCTURE DETAIL PAGE FORMOUNTING STRUCTURE EXAMPLE

MDL:

300 S

ERIE

SQT

Y.: 3

REV.

#: 0

SCALE: 34" = 1'MAX. FIXTURE WEIGHT: MAX. MOMENT LOAD: EQUIPMENT CAPACITY:657 LBS. 3968 FT. LBS. 3VBM48 - 235 LBS.

CHILD

REN'

S HO

SPIT

ALNO

LAPR

OJEC

T #:

17-3

23SU

BMIT

TAL

PLOT

DAT

E:

Page 47: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

STRUCTURAL REQUIREMENTS - Architect and Structural Engineer

Mounting Structure ComponentsThe fabrication of each mounting structure may be slightly different but they each require the same basic components to ensurestability.

Sway Bracing (by others)Sway bracing is designed to rigidly affix the mounting plate to the structural ceiling. The primary purpose of sway bracing is toeliminate sway, or lateral twisting and flexing of the mounting structure as it "reacts" to dynamic load changes caused by moving thefixture radial arms. The sway bracing should be welded to the mounting plate and extend away from the center of the mount. Aminimum of four sway braces placed 90° apart and positioned at a 45° and 60° angle is recommended.

Minimum recommended material for sway bracing is 3" x 3" x 1/4" angle iron. It is recommended that in all applications thesway bracing be fastened to the structural ceiling.

Structural Ceiling Plate (by others)The structural ceiling plate rigidly attaches the mount to the structural ceiling using structural anchors appropriate for the ceilingconstruction. The structural ceiling plate should be a minimum of, 1" ASTM A36 steel plate, 17" diameter with (6) 5/8" diameterholes for structural anchors and is fabricated by others.

Expansion Anchors (by others)Test 50% of the anchors at 2,000 pounds (907 kg) tension, or 50 ft. lb. (68 N●m) torque per CBC 1925A.3.5. Installed anchors must meet the following criteria:1. Hydraulic Ram Method: The anchor should have no observable movement at the applicable test load. For wedge and sleevetype anchors, a practical way to determine observable movement is that the washer under the nut becomes loose.2. Torque Wrench Method (Wedge or Sleeve Type): The applicable test torque must be reached within one-half (1/2) turn of the nut.Testing should occur no sooner than 24 hours after installation of anchors. If any anchor fails testing, test all anchors until 20 consecutive anchors pass, then resume the initial testing frequency. Test equipment is to be calibrated by an approved testinglaboratory in accordance with standard recognized procedures.

Support Tube (by others)The support tube required to attach the mounting plate to the structural ceiling plate should be ASTM 500 Grade B, 6" diameter tube.Support tube is to be welded to structural ceiling plate and mounting plate. A minimum of 6 gussets placed 60° apart should bewelded to support tube at the structural ceiling plate and the mounting plate.

Mounting Plate (SKYTRON supplied)The 17.5" x 17.5" x 1" ASTM A36 steel mounting plate is a SKYTRON supplied item. The Support tube and sway bracing are welded to the mounting plate. The mounting plate contains the corresponding bolt pattern for attaching the fixture and provides themounting areas for the junction box and gas riser plates.

Mounting Structure DesignSeismic structural applications differ. Please contact your local SKYTRON representative for specific calculations. Themounting structure must be designed and fabricated to position the bottom of the SKYTRON mounting plate as shown on sitespecific elevation page. This is a critical dimension in order to accommodate proper clearance required for ceiling cover function. The mounting plate must be perfectly level (+/- 0.1°) and allow no more than two-tenths of a degree (0.2°) of rotation atthe mounting plate when the specified load is applied. The mounting structure must be tested for strength and stiffness prior toinstallation of the fixture. Please contact your SKYTRON representative to schedule testing.

Please consult your SKYTRON representative during early stages of construction to facilitate this process. The testingprocess is a required, documented function prior to closing of the finished ceiling.

Ceiling RequirementsA 24" x 24" access door must be mounted adjacent to the mounting structure for entry by service personnel for serviceaccess.

SKYTRON provides a 24" ceiling cover designed to cover 20" diameter ceiling hole cutout.

Additional Items(SKYTRON supplied)In addition to the pre-installation kit. SKYTRON provides the following items:

(6) 1-1/4" x 10" threaded rods. (24) 1-1/4" hex nuts, pump enclosure (if applicable)

STRUCTURALCEILING

STRUCTURALCEILING

STRUCTURAL CONCRETE SLABNORMAL WEIGHTfc=3,000 PSI (20,684 kPa)

SWAY BRACING3" x 3" x 1/4" (4)

GUSSET 6" DIA.SUPPORT

TUBE

FINISHED CEILING

20"DIAMETER OPENING

REQUIRED(CENTERED UNDER MOUNT)

12" MIN. / 60" MAX.

REFER TO JOB SPECIFIC ELEVATION PAGE (-0/+1)

MOUNTING PLATE(SKYTRON SUPPLIED)

45°- 60°

FIXTURE HUBTYP. 1/4"

PJP

ACCESS DOOR24" x 24"

(REQUIRED)

TYP. 1/4"TOP & BOTTOM

STRUCTURALCEILING PLATE

17" DIA. x 1"

EXPANSION ANCHORS:HILTI KB-III (ICBO 4627)

CARBON STEEL

NOTES:• This illustration depicts recommended mounting structure design and its components. Always consult

specific structural criteria defined by a structural engineer.• Do not cover or block any holes with sway bracing, gussets, weld or weld slag.• Typical dimensions shown. Refer to specific structural drawings and/or Seismic drawings for each

application.• *Critical Dimension

12" HIGH CLEAR SPACEABOVE MOUNTING PLATE

REQUIRED ON 3 SIDES FORGAS AND ELECTRICAL

COMPONENT INSTALLATION

*TYPICAL BOOM MOUNTING STRUCTURE DETAILS**ALWAYS CONSULT SPECIFIC STRUCTURAL CRITERIA DEFINED BY A STRUCTURAL ENGINEER*

NOTES:• This illustration depicts a generic mounting structure design and its components. Always consult

specific structural criteria defined by a structural engineer.• 1 14" Mounting bolts and nuts shipped with fixture.

JUNCTION BOX(SKYTRON SUPPLIED)

BRACKET(SKYTRON SUPPLIED)

PLATE(SKYTRON SUPPLIED)

INLET TUBE(SKYTRON SUPPLIED)

RISER PLATE(SKYTRON SUPPLIED)

CHECK VALVEREMOVAL TOOL(SKYTRON SUPPLIED)

G1.1SHEETINITIAL:

DATE:

DESC

RIPT

ION:

ELE

VATI

ON D

ETAI

LS11

/16/20

17

CHILD

REN'

S HO

SPIT

ALNO

LAMD

L: 30

0 SER

IES

QTY.

: 3RE

V. #:

0PR

OJEC

T #:

17-3

28

SUBM

ITTA

LPL

OT D

ATE:

Page 48: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

DISS (8) 125V, 20A DUPLEX - RED48 1/2"H x 22"W x 30"D

ACCESSORY LIST(1) MOUNTING BLOCK FOR VST MONITOR MOUNT(1) FRONT EQUIPMENT RAIL(1) REAR EQUIPMENT RAIL(2) CORD WRAP - RAIL MOUNTED(1) PMSH SHELF(2) PREMIUM DRAWER

QTY GASES

0 CARBON DIOXIDE

0 HELIOX

1 MEDICAL AIR

0 NITROGEN

1 NITROUS OXIDE

2 OXYGEN

2 VACUUM

0 WAGD

ELECTRICAL:GAS OUTLETS:DIMENSIONS:INITIAL:

G2

PLO

T DA

TE: 1

1/16

/201

7

CHIL

DREN

'S H

OSP

ITAL

NOLA

PRO

JECT

#: 1

7-32

8

DESC

RIPT

ION:

CAR

RIER

DET

AILS

SHEET

SUBM

ITTA

LQ

TY: 3

REV

#: 0

DATE:

MDL

: 300

SER

IES

CARRIER

*SITE SPECIFIC CARRIER DETAILS*

MEDICAL AIR

NITROUS OXIDE

OXYGEN

OXYGEN

VACUUMVACUUM

SEPARATOR PLATE

*COMMUNICATION CANONLY BE ADDED ABOVETHE SEPARATOR PLATE

Page 49: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

G3SHEET

MEDICAL GAS REQUIREMENTS - Medical Gas / Piping EngineerNotes:

• Beacon Medaes manufactured gas outlets supplied unless otherwise specified. Pricing and lead time could vary if another Mfgr. is chosen.• Medical gas risers comply with NFPA 9999-2012 OR CAN-CSA Z 73961-12 Edition Cleanliness of gas outlets must be maintained through

installation.• Purge gas lines and test at least 24 hrs. prior to installation.• Test results must be made available to SKYTRON representative for verification & comparison.• Check valve must be removed during installation to prevent damage to o-rings, refer to Ergon pre-installation guide for removal

instructions.** Caution - Check valves have 2 different sizes, please keep Instrument Air and Oxygen separate from the others **

Each boom fixture is fabricated in accordance to the specifications required by the customer. The configuration drawings supplied bySKYTRON will indicate the type and quantity of gas supply lines required. D.I.S.S. connection medical grade hoses connect the fixture tothe riser plate connectors. The customer is responsible to deliver the appropriate medical gas from the facility supply to the riser plateconnectors and for the design of a medical gas system with adequate flow capacity capable of compensating for the accumulative flowrestrictions associated with conventional construction methods i.e. flex gas hoses. SKYTRON provides medical gas riser plate(s) and theappropriate connectors for attachment to the Mounting Plate. The riser plates attach to the mounting plate and will accommodate up to 12gas connectors. The connectors are D.I.S.S. type medical gas connectors with removable single check valves and provide a 1/2" coppertube for attachment to facility supply lines. Caps are provided with Vacuum & WAGD gas connectors to facilitate testing procedures.

All connection and testing of medical gas piping to be performed in accordance with NFPA 99-2012 OR CAN-CSA Z 73961-12requirements.

Recommended Gas Riser PlacementA 6" clearance must be maintained between the finished ceiling and the gas inlet tubes. If this clearance cannot be accomplished atthe mounting plate, then horizontal or remote mounting of the riser plate will be required.When mount is in center of room, riser plate should face head end of room, when mount is near walls, riser plate should be adjacentto wall. If more than one riser plate is required they should be placed next to each other.

GAS CONNECTOR ASSEMBLY (w/ SECONDARY CHECK)*SUPPLIED w/FIXTURE*

9"

DISS CONNECTOR(w/ REMOVABLE PRIMARY

CHECK. PASSIVE EVAC,VACUUM & WAGD DO NOT

HAVE PRIMARY CHECK.)*SUPPLIED w/PRE-INSTALL KIT*

D I S SCHEMETRON

CGA ColorStandard

Note #1 - Any room (Critical Care Area) designed for a permanently located respiratory ventilator or anesthesia machine shall have an outlet capable of a transient flow rate of:170 LPM (6 SCFM) for 3 seconds at the station outlet.

Note #2 - For testing and certification purposes, individual station inlets shall be capable of delivering a flow rate of: 3 SCFM, while maintaining a system pressure of not lessthan 12" (300mm) at the nearest adjacent vacuum inlet. Facility supply must be 115 LPM MINIMUM. (Vacuum D.I.S.S. connectors omit primary check valves for optimal flow). 12in/HG.

Note #3 - WAGD (Waste Anesthetic Gas Disposal) systems employing a design where the WAGD lines are "tied in" to MedVac lines must produce the same flow rates as theMedVac inlets.

Note #4 - Nitrogen system requires nitrogen supplied directly from facility supply line rated at 185psi MIN to 200psi MAX. Avoid designs which feature multiple-in-line Nitrogencontrol systems in order to avoid loss of flow capability.

Note #5 - Pipe sizing: piping systems shall be designed and sized to deliver the required flow rates at the utilization pressures while accounting for natural flow losses throughvarious components including flexible hoses and terminal outlets. The Skytron medical gas and vacuum outlets meet and exceed these standards at the time of manufacture. However piping source capacity, sizing, and restrictions may prevent outlets from attaining required valves.

Additional references: Health Care Facilities code 2012 edition, Section 5.1.12.3.10.2, 5.1.12.3.10.3, 5.1.12.3.10.4, 5.1.12.3.10.5

• TYPE K COPPERTUBING

• 1/2" O.D. EXCEPTPASSIVE EVAC

• 3/4" O.D. FOR PASSIVEEVAC

Medical Air(Yellow)

Test GasISO ColorStandardMedical Air

(Black)

AbbreviatedName

MedAir

StandardPressure

50 - 55 psig

MaximumPressure

55 psig

AllowablePressure Drop

5 psig

Minimum Flow Rates

3.5 SCFM per outlet (100NL/min)N2NF see note #1, 5Carbon Dioxide

(Gray)Carbon Dioxide

(Gray) CO2 50 - 55 psig 55 psig 5 psig 3.5 SCFM per outlet (100NL/min)N2NFHeliox

(Brown)Heliox

(Brown) Heliox 50 - 55 psig 55 psig 5 psig 3.5 SCFM per outlet (100NL/min)N2NFNitrogen(Black)

Nitrogen(Black) N2 or HPN2 160 - 185 psig 200 psig 5 psig

5 SCFM per outlet (140NL/min)free air per outletN2NF see note #4, 5

Nitrous Oxide(Blue)

Nitrous Oxide(Blue) N2O 50 - 55 psig 55 psig 5 psig 3.5 SCFM per outlet (100NL/min)N2NF

Oxygen(Green)

Oxygen(White) O2 50 - 55 psig 55 psig 5 psig 3.5 SCFM per outlet (100NL/min)N2NF see note #1

Vacuum(White)

Vacuum(Yellow) MedVac 12in/Hg (300mm) N/A 3 SCFM per outlet (85NL/min) see note #2

Waste AnestheticGas Disposal

(Purple) WAGDVaries withsystem type

At a minimum, each inlet must be able todraw a continuous 50 lpm (1.8 SCFM)

through the interface

Waste AnestheticGas Disposal

(Purple) see note #3

PURITANOHMEDA

DESC

RIPT

ION:

MED

ICAL

GAS

DET

AILS

*SITE SPECIFIC GAS DETAILS* 11/16

/2017

CHILD

REN'

S HO

SPIT

ALNO

LAPR

OJEC

T #:

17-3

28SU

BMIT

TAL

PLOT

DAT

E:

MDL:

300 S

ERIE

SQT

Y.: 3

REV.

#: 0

MEDI

CAL G

AS

MAX OF 12 INLET TUBES PER MOUNTING PLATE. IF MORE ARE REQUIREDTHEY WILL NEED REMOTE MOUNTING.

CHECK VALVE REPLACEMENTTOOL

VALVES

Use above for:Medical AirCarbon DioxideHelioxNitrous OxideNitrogen

Use above for:Instrument AirOxygen

1.18"1.10"

see note #5

see note #5

see note #5

CGA ColorStandard

Nitrogen(Black)

Test GasISO ColorStandard

Nitrogen(Black)

AbbreviatedName

AllowablePressure Drop Minimum Flow Rates

Vacuum(White)

Vacuum(Yellow)

N2 or HPN2

MedVac

(10 psi) 70kpa 400L/min (14.1scfm)@1,250kpa

30L/min (1.1scfm)@54kpa

N2NF

N2NF

INITIAL:DATE:

VERIFY AND INITIAL GAS FACEPLATE STYLE REQUESTED

(4 inhg) 13kpa

CAN-CSA section 7.3.2. Z 7396.1-12

N2NF Other Gas Various MedVac 120L/min (4.2scfm)@345kpa(4 psi) 28kpa

Page 50: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

INITIAL:DATE: INITIAL:

DESC

RIPT

ION:

ELE

CTRI

CAL W

IRIN

G DE

TAILS

G4SHEET

11/16

/2017

REQUIRED FOR FABRICATIONVERIFY AND INITIAL

POWER TYPE INITIAL:

4 INP

UT LE

ADS

120 V

AC, 2

0 AMP

120 V

AC, 2

0 AMP

120 V

AC, 2

0 AMP

CIRC

UIT

2

CIRC

UIT

3

CIRC

UIT

4

(8) 1

25V,

20A,

DUP

LEX

OUTL

ETS

- RED

*SITE SPECIFIC WIRING DETAILS* INTERNAL FIXTURE WIRING TYPICAL, EXTRA FLEX STEEL CONDUIT

TYPE: 12AWG, 600V, XHHW-2, 90°C - UNLESS NOTED

MOUN

TING

HUB

T B T B T B T B

ELECTRICAL REQUIREMENTS - Electrical Engineer

Each boom fixture is fabricated in accordance to the specifications required by the customer. The Configuration drawings supplied by SKYTRONwill indicate the type and quantity of circuits required. SKYTRON provides all wiring and electrical materials for connection from fixture to junctionbox or pump enclosure (if applicable). SKYTRON supplies an electrical junction box (8-5/8" x 4-5/8" x 1-3/4") to facilitate field wiring for up to sixcircuits that is mounted on the mounting plate in the correct position and if applicable, a hydraulic pump enclosure/junction box (18"L x 6"W x 12"H) that is to be remote mounted within 24" of the mounting structure (by contractor). The pump enclosure can be shipped with the installationkit upon request. Typical wire type is 12AWG, 600V, XHHW-2. Each circuit requires a separate, properly circuit protected, 120VAC, 60Hz powersupply line enclosed in rigid metal conduit. All electrical materials for connection to SKYTRON supplied junction box or pump enclosure and installation labor for such materials to be provided by customer. All wiring and materials to be in accordance with federal, state and local codes.It is the customer's responsibility to meet conformity to NFPA and NEC standards with respect to the number of receptacles provided in a patientcare area.

Specific conductor colors and/or wiring for isolated applications are available upon request.

ARM COVERARM COVER

ARM COVERARM COVER

GROUND(FOR ELECTRICAL)

PROTECTED EARTH(FOR CHASSIS)

NORMALDUPLEX

CRITICALDUPLEX

CRITICALDUPLEX

ULTRAFLEXTYPE USL CONDUIT

ULTRAFLEXTYPE USL CONDUIT

TB

TERMINALBLOCKDETAIL

120VAC, 20ASUPPLY

TB

ISOLATED POWERBROWN W/YELLOW STRIPE, ORANGE W/BLUE STRIPE,

GREEN W/YELLOW STRIPE

NON-ISOLATED POWERBLACK, WHITE, GREEN

Internally LocatedElectrical Terminal

Block Detail

120VAC, 20ASUPPLY

carrier contains aninternally located

Terminal Blockthat provides

interconnectivitybetween the

outlets and theinternal conductors

CARR

IER

UTILI

TIES

PAN

EL

TB

TB

CIRC

UIT

112

0 VAC

, 20 A

MP

GENERIC BOOM ELECTRICAL WIRING DIAGRAM FOR FIXED ARMS

CHILD

REN'

S HO

SPIT

ALNO

LAPR

OJEC

T #:

17-3

28SU

BMIT

TAL

PLOT

DAT

E:

MDL:

300 S

ERIE

SQT

Y.: 3

REV.

#: 0

Page 51: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

72"

0" FLOOR

120" CEILING HEIGHT10"

51" [1300mm]3T2

36" 3AFC

12"

18"

3FLM-32

86 12"

35"43"

80"

REFER TO AUA33 DATA SHEETFOR SURGICAL LIGHT

SPECIFICATIONS

81"

21" 15"

ITEM ID: 11021

H1SHEETINITIAL:

DATE:

DESC

RIPT

ION:

ELE

VATI

ON D

ETAI

LS

*SITE SPECIFIC ELEVATION DETAILS*

6/20/2

018PLEASE REFER TO MOUNTING

STRUCTURE DETAIL PAGE FORMOUNTING STRUCTURE EXAMPLE

SCALE: 34" = 1'MAX. FIXTURE WEIGHT: MAX. MOMENT LOAD: EQUIPMENT CAPACITY:

CHILD

REN'

S HO

SPIT

ALNO

LAMD

L: 11

0 SER

IES

QTY.

: 2RE

V. #:

0

484 LBS. 2207 FT. LBS. 3FLM1-32 - 64.5 LBS.

PROJ

ECT

#: 17

-323

SUBM

ITTA

LPL

OT D

ATE:

Page 52: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

STRUCTURAL REQUIREMENTS - Architect and Structural Engineer

Mounting Structure ComponentsThe fabrication of each mounting structure may be slightly different but they each require the same basic components to ensurestability.

Sway Bracing (by others)Sway bracing is designed to rigidly affix the mounting plate to the structural ceiling. The primary purpose of sway bracing is toeliminate sway, or lateral twisting and flexing of the mounting structure as it "reacts" to dynamic load changes caused by moving thefixture radial arms. The sway bracing should be welded to the mounting plate and extend away from the center of the mount. Aminimum of four sway braces placed 90° apart and positioned at a 45° and 60° angle is recommended.

Minimum recommended material for sway bracing is 3" x 3" x 1/4" angle iron. It is recommended that in all applications thesway bracing be fastened to the structural ceiling.

Structural Ceiling Plate (by others)The structural ceiling plate rigidly attaches the mount to the structural ceiling using structural anchors appropriate for the ceilingconstruction. The structural ceiling plate should be a minimum of, 1" ASTM A36 steel plate, 17" diameter with (6) 5/8" diameterholes for structural anchors and is fabricated by others.

Expansion Anchors (by others)Test 50% of the anchors at 2,000 pounds (907 kg) tension, or 50 ft. lb. (68 N●m) torque per CBC 1925A.3.5. Installed anchors must meet the following criteria:1. Hydraulic Ram Method: The anchor should have no observable movement at the applicable test load. For wedge and sleevetype anchors, a practical way to determine observable movement is that the washer under the nut becomes loose.2. Torque Wrench Method (Wedge or Sleeve Type): The applicable test torque must be reached within one-half (1/2) turn of the nut.Testing should occur no sooner than 24 hours after installation of anchors. If any anchor fails testing, test all anchors until 20 consecutive anchors pass, then resume the initial testing frequency. Test equipment is to be calibrated by an approved testinglaboratory in accordance with standard recognized procedures.

Support Tube (by others)The support tube required to attach the mounting plate to the structural ceiling plate should be ASTM 500 Grade B, 6" diameter tube.Support tube is to be welded to structural ceiling plate and mounting plate. A minimum of 6 gussets placed 60° apart should bewelded to support tube at the structural ceiling plate and the mounting plate.

Mounting Plate (SKYTRON supplied)The 17.5" x 17.5" x 1" ASTM A36 steel mounting plate is a SKYTRON supplied item. The Support tube and sway bracing are welded to the mounting plate. The mounting plate contains the corresponding bolt pattern for attaching the fixture and provides themounting areas for the junction box and gas riser plates.

Mounting Structure DesignSeismic structural applications differ. Please contact your local SKYTRON representative for specific calculations. Themounting structure must be designed and fabricated to position the bottom of the SKYTRON mounting plate as shown on sitespecific elevation page. This is a critical dimension in order to accommodate proper clearance required for ceiling cover function. The mounting plate must be perfectly level (+/- 0.1°) and allow no more than two-tenths of a degree (0.2°) of rotation atthe mounting plate when the specified load is applied. The mounting structure must be tested for strength and stiffness prior toinstallation of the fixture. Please contact your SKYTRON representative to schedule testing.

Please consult your SKYTRON representative during early stages of construction to facilitate this process. The testingprocess is a required, documented function prior to closing of the finished ceiling.

Ceiling RequirementsA 24" x 24" access door must be mounted adjacent to the mounting structure for entry by service personnel for serviceaccess.

SKYTRON provides a 24" ceiling cover designed to cover 20" diameter ceiling hole cutout.

Additional Items(SKYTRON supplied)In addition to the pre-installation kit. SKYTRON provides the following items:

(6) 1-1/4" x 10" threaded rods. (24) 1-1/4" hex nuts, pump enclosure (if applicable)

STRUCTURALCEILING

STRUCTURALCEILING

STRUCTURAL CONCRETE SLABNORMAL WEIGHTfc=3,000 PSI (20,684 kPa)

SWAY BRACING3" x 3" x 1/4" (4)

GUSSET 6" DIA.SUPPORT

TUBE

FINISHED CEILING

20"DIAMETER OPENING

REQUIRED(CENTERED UNDER MOUNT)

12" MIN. / 60" MAX.

REFER TO JOB SPECIFIC ELEVATION PAGE (-0/+1)

MOUNTING PLATE(SKYTRON SUPPLIED)

45°- 60°

FIXTURE HUBTYP. 1/4"

PJP

ACCESS DOOR24" x 24"

(REQUIRED)

TYP. 1/4"TOP & BOTTOM

STRUCTURALCEILING PLATE

17" DIA. x 1"

EXPANSION ANCHORS:HILTI KB-III (ICBO 4627)

CARBON STEEL

NOTES:• This illustration depicts recommended mounting structure design and its components. Always consult

specific structural criteria defined by a structural engineer.• Do not cover or block any holes with sway bracing, gussets, weld or weld slag.• Typical dimensions shown. Refer to specific structural drawings and/or Seismic drawings for each

application.• *Critical Dimension

12" HIGH CLEAR SPACEABOVE MOUNTING PLATE

REQUIRED ON 3 SIDES FORGAS AND ELECTRICAL

COMPONENT INSTALLATION

*TYPICAL BOOM MOUNTING STRUCTURE DETAILS**ALWAYS CONSULT SPECIFIC STRUCTURAL CRITERIA DEFINED BY A STRUCTURAL ENGINEER*

NOTES:• This illustration depicts a generic mounting structure design and its components. Always consult

specific structural criteria defined by a structural engineer.• 1 14" Mounting bolts and nuts shipped with fixture.

JUNCTION BOX(SKYTRON SUPPLIED)

BRACKET(SKYTRON SUPPLIED)

PLATE(SKYTRON SUPPLIED)

INLET TUBE(SKYTRON SUPPLIED)

RISER PLATE(SKYTRON SUPPLIED)

CHECK VALVEREMOVAL TOOL(SKYTRON SUPPLIED)

H1.1SHEETINITIAL:

DATE:

DESC

RIPT

ION:

ELE

VATI

ON D

ETAI

LS11

/16/20

17

CHILD

REN'

S HO

SPIT

ALNO

LAMD

L: 11

0 SER

IES

QTY.

: 2RE

V. #:

0PR

OJEC

T #:

17-3

28

SUBM

ITTA

LPL

OT D

ATE:

Page 53: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

TYPE USL CONDUITBY SKYTRON

SPECIAL GROUNDING REQUIREMENTS - Electrical Engineer

1. Use of approved metal conduit shall be employed throughout the fixture's wiring circuit where applicable. Facility supplied, circuit breaker protected, 100-240VAC 50/60 Hz power source wiring.

2. Grounding - Proper performance and safety of this fixture can only achieved by an adequate grounding system. Fixture ground must be a dedicated ground point ultimately bonded to the facilities grounding system to prevent the migration of electricalinterference generated by other devices.

Protective Means - To avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective earth ground. This fixture requires a properly circuit protected, appropriately sized, dedicated circuit. An isolated powersupply circuit must be protected by an appropriately sized double pole, single throw circuit breaker

BY SKYTRON

BY OTHERS

See note 1 See note 2

THIS DIAGRAM IS FOR INFORMATIONAL PURPOSES ONLY. THIS WILL NOT MATCH YOUR EXACT MODEL.

JUNCTION BOX BY OTHERWITHIN 12" OF MOUNT 3/4" METAL CONDUIT

100-240VAC INPUT DEDICATED

3/4" METAL CONDUIT 12AWG, 3 WIRES PERLIGHTHEAD PLUS FIXTURE GROUNDRED, GREEN/YELLOW, BLUE 24VDC OUTPUT

POWER SUPPLY JUNCTION BOXWITHIN 3' OF MOUNT(SUPPLIED BY SKYTRON)

MDL:

110 S

ERIE

SQT

Y.: 1

REV.

#: 0

H6SHEETINITIAL:

DATE:

DESC

RIPT

ION:

LIGH

T FI

XTUR

E DE

TAILS

11/16

/2017

CHILD

REN'

S HO

SPIT

ALNO

LAPR

OJEC

T #:

17-3

28SU

BMIT

TAL

PLOT

DAT

E:

Page 54: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

72"

0" FLOOR

108" CEILING HEIGHT6"

2712"

35"

80"

REFER TO AUA33 DATA SHEETFOR SURGICAL LIGHT

SPECIFICATIONS

81"

11"17"

ITEM ID: 12173

I1SHEETINITIAL:

DATE:

DESC

RIPT

ION:

ELE

VATI

ON D

ETAI

LS

*SITE SPECIFIC ELEVATION DETAILS*

6/20/2

018PLEASE REFER TO MOUNTING

STRUCTURE DETAIL PAGE FORMOUNTING STRUCTURE EXAMPLE

SCALE: 34" = 1'MAX. FIXTURE WEIGHT: MAX. MOMENT LOAD: EQUIPMENT CAPACITY:

CHILD

REN'

S HO

SPIT

ALNO

LAMD

L: AU

A33-

895/7

00QT

Y.: 1

REV.

#: 1

166 LBS. 539 FT. LBS. N/A

PROJ

ECT

#: 17

-323

SUBM

ITTA

LPL

OT D

ATE:

Page 55: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

CEILING COVER(SKYTRON SUPPLIED)

FINISHED CEILING

STRUCTURAL CEILING

16-1/2" DIA. ACCESS HOLE20 1/2" DIA. CEILING COVER

15" DIA. MOUNTING. PLATE

2.2"1.1"

FIXTUREJUNCTION

BOX

5.9" MAX3.9" MIN.

3/4" METAL FLEXCONDUIT (see note 3)

45° 45°

STRUCTURAL CEILING

11.8" x 11.8" x .4"STIFFENER PLATE

3/4" ALL THREAD ROD(see note 1)

(4) 2" x 2" ANGLE SWAY BRACING(see note 2)

STRUCTURAL REQUIREMENTS - Architect and Structural Engineer

Notes1. 3/4” support rods located for total support of light, all labor and materials for fabrication are supplied by General Contractor. 3/4” nuts

and washers for support of SKYTRON mounting plate supplied by contractor (8 each required).

2. The mounting structure must be attached to structural ceiling and BRACED TO ALLOW NO TWISTING OR LATERAL MOTION andshall be designed not to provide a degree of deflection greater than two-tenths of a degree at the mounting plate.

3. 3/4” metal conduit and minimum 12AWG wire size (3 wires per lighthead plus fixture ground) required between fixture and the powersource. All metal conduit, wiring, and other electrical materials as well as installation labor for such materials associated with theinstallation of the SKYTRON surgical light to be provided by Electrical Contractor. All installations of SKYTRON procedural lightsshould be under the direct supervision of a SKYTRON representative. All wiring to be in accordance with local codes and by a certifiedelectrician.

This is a GENERAL GUIDELINE ONLY.

NOTES:• This illustration depicts a generic mounting structure design and its components. Always consult

specific structural criteria defined by a structural engineer.• Do not cover or block any holes with sway bracing, gussets, weld, weld slag or etcetera• Typical dimensions shown. Refer to specific structural drawings and/or Seismic drawings for each

application.

*TYPICAL LIGHT MOUNTING STRUCTURE DETAILS**ALWAYS CONSULT SPECIFIC STRUCTURAL CRITERIA DEFINED BY A STRUCTURAL ENGINEER*

Hole forAccess to

FixtureJunction

Box

TOP VIEWMOUNTING PLATE DETAIL

2" x 2"Angle Sway Bracing90° Apart - 4 Each(see note 2)

5.9"

Hole forAccess to

FixtureJunction

Box

TOP VIEWMOUNTING PLATE DETAIL

2" x 2"Angle Sway Bracing90° Apart - 4 Each(see note 2)

11.8"

9 - 1/2"

5.9"

11.8"9 - 1/2"

I1.1SHEETINITIAL:

DATE:

DESC

RIPT

ION:

ELE

VATI

ON D

ETAI

LS11

/17/20

17

CHILD

REN'

S HO

SPIT

ALNO

LAMD

L: AU

A33-

895/7

00QT

Y.: 1

REV.

#: 0

PROJ

ECT

#: 17

-328

SU

BMIT

TAL

PLOT

DAT

E:

Page 56: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

SPECIAL GROUNDING REQUIREMENTS - Electrical Engineer

1. Use of approved metal conduit shall be employed throughout the fixture's wiring circuit where applicable. Facility supplied, circuit breaker protected, 100-240VAC 50/60 Hz power source wiring.

2. Grounding - Proper performance and safety of this fixture can only achieved by an adequate grounding system. Fixture ground must be a dedicated ground point ultimately bonded to the facilities grounding system to prevent the migration of electricalinterference generated by other devices.

Protective Means - To avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective earth ground. This fixture requires a properly circuit protected, appropriately sized, dedicated circuit. An isolated powersupply circuit must be protected by an appropriately sized double pole, single throw circuit breaker

3. Mounting and anchorage: please refer to the ASTRO Installation Manual for mounting requirements. Seismic applications will differ in construction requirements. Please request seismic calculations and mounting requirements from your SKYTRONrepresentative.

BY SKYTRON

BY OTHERS

See note 3

THIS DIAGRAM IS FOR INFORMATIONAL PURPOSES ONLY. THIS WILL NOT MATCH YOUR EXACT MODEL.

3/4" METAL CONDUIT 12AWG, 3 WIRES PERLIGHTHEAD PLUS FIXTURE GROUNDRED, GREEN/YELLOW, BLUE 24VDC OUTPUT

See note 1 See note 2

JUNCTION BOX BY SKYTRON3/4" METAL CONDUIT100-240VAC INPUT DEDICATED

POWER SUPPLY JUNCTION BOXWITHIN 3' OF MOUNT(SUPPLIED BY SKYTRON)

MDL:

AUA3

3-89

5/700

QTY.

: 1RE

V. #:

0

I6SHEETINITIAL:

DATE:

DESC

RIPT

ION:

LIGH

T FI

XTUR

E DE

TAILS

11/16

/2017

CHILD

REN'

S HO

SPIT

ALNO

LAPR

OJEC

T #:

17-3

28SU

BMIT

TAL

PLOT

DAT

E:

Page 57: CHILDREN'S HOSPITAL NOLAmplusmedtech.com/16001/pdf/1767.pdf · 2020. 5. 27. · d2310 storage d2311 o.r. 10 41 /2"[1050 m m] 3 5. 2 5 " / 8 95 mm 35 .2 5 "/ 895 mm 3'-6" 3'-6" 5'

INITIAL:DATE:

MOUNTING STRUCTURE TEST JIG

WEIGHT SUPPORT TUBE

1-1/4" NUTS

1-1/4" THREADED RODS (6)

MOUNTING HUBASSEMBLY

12mm NUT (6)

12mm x 30mm BOLT (4)

CHAIN

CHAIN FALL

LOWER RADIAL ARM

UPPER RADIAL ARM

5/8" NUT (2)

12mm NUT (2)

WEIGHTS (6)

TESTING - MOUNTING STRUCTURE

Each mounting structure must be tested by a Skytron representative to verify that there is no more than two-tenths of a degree of rotational movement at the mounting plate prior to installing a fixture. A mounting structure test jig is available from a Skytron representative to facilitate thisprocess. The mounting structure test jig is a fixture which simulates the weights and moment loads created by a Skytron fixture. It consists of ahub assembly, an upper and lower radial arm, a chain fall, weight support tube and six 100 lb. weights (refer to illustration).

The drawing package and the test jig instructions are required to perform an accurate test.

The test jig instructions include the load simulator weight chart, the test jig report form and complete instructions for performing the test.

fixture weight moment load * test weight

N/A200 lbs300 lbs

N/AN/A

N/AN/A

145 lbs 650 ft lbs

100110120200300310

330COLUMN (ALL)

(ALL) LFSLFS(ALL) EXTENDED LIGHT ARMS

SERIES NUMBER

200 lbs300 lbs

216 lbs 1011 ft lbs

400 lbs500 lbs

597 lbs 2310 ft lbs

1029 lbs6934 ft lbs

499 lbs657 lbs

657 lbs877 lbs

2207 ft lbs3213 ft lbs

3968 ft lbs4758 ft lbs

100 lbs

200 lbs

3201257 lbs

600 lbs5754 ft lbs600 lbs

WARNINGTesting the weight and moment load on Ergon 3can only be performed by a authorized Skytron

distributor ONLY.

WARNINGBecause testing the weight and moment load ofErgon 3 is hazardous work it is important to use

proper Personal Protection Equipment (PPE) and adhere toOSHA (Occupational Safety and Health Administration) for

materials handling and storage for the safe equipment used inhoists and with slings.

The Skytron weight and moment load simulator is a device which simulates the weight and moment load on an actual ceiling mounted articulating fixture in orderto test the weight bearing capacity and performance of the Ergon 3 mounting structure.

Skytron recommends using a 600 lb (272 kg) maximum load simulator to test all mounting structure systems for maximum load and weight capacity.

Load simulation is an essential part of readiness assessment and provides reassurance the mounting structure can accept larger more complex articulatingsystems if the need arises for equipment changes or future expansion.

In the event that the mounting structure system cannot deliver performance for the maximum load(600lbs), the weight matrix provided above provides lesserfixture weight and moment load as a means to ensure that the minimum criteria is met.

Refer to project specific drawings for exact weight capacity