Chapter 1: Introduction
description
Transcript of Chapter 1: Introduction
Chapter 1: IntroductionChapter 1: Introduction
Analytical ChemistryAnalytical Chemistry Qualitative analysisQualitative analysis Quantitative analysisQuantitative analysis
Classification of Analytical Classification of Analytical MethodsMethods Classical methods (wet chemical Classical methods (wet chemical
methods)methods) Instrumental methodsInstrumental methods
Comparisons of Wet Comparisons of Wet Chemical and Instrumental Chemical and Instrumental
MethodsMethodsWet chemical Instrumental
Separation methods
PrecipitationExtractionDistillation
Chromatographiceletrophoretic
Detection methods ColorsBoiling or melting pointsSolubilitiesOdors
ConductivityElectrode potential\light absorption, emissionMass/charge ratio (M/Z)fluorescence
Types of Instrumental MethodsTypes of Instrumental Methods Separation techniquesSeparation techniques
Chromatographic Chromatographic GasGas LiquidLiquid …………..
ElectrophoresisElectrophoresis Many typesMany types
Detection techniquesDetection techniques Optical spectroscopyOptical spectroscopy
AbsorptionAbsorption EmissionEmission FluorescenceFluorescence
Mass spectroscopyMass spectroscopy AtomicAtomic MolecularMolecular
Instruments for AnalysisInstruments for Analysis
SignalGenerator
SampleAnalytical
Signal
InputTransducer(Detector)
Signal Processor
OutputTransducer(Readout)
Instruments for AnalysisInstruments for Analysis
Selection of An Instrumental Selection of An Instrumental MethodMethod
Defining the problemDefining the problem Performance characteristicsPerformance characteristics
Figure of MeritsFigure of Merits: : Quantitative performance Quantitative performance criteria of instruments that are expressed criteria of instruments that are expressed in numerical termsin numerical terms
PrecisionPrecisionA measure of the random or indeterminate A measure of the random or indeterminate
errorerror
Relative Standard Deviation
BiasBiasA measure of the systematic or determinate A measure of the systematic or determinate
errorerror
Bias = µ - XBias = µ - Xtt
µ - the population mean for the µ - the population mean for the concentration of an analyte concentration of an analyte
XXt t – true concentration– true concentration
Standard reference materials (SRM)Standard reference materials (SRM)
SensitivitySensitivityA measure of the ability of an analytical A measure of the ability of an analytical method to discriminate between small method to discriminate between small differences in analyte concentrationdifferences in analyte concentration Calibration sensitivity (S = Calibration sensitivity (S = mmc + Sc + Sblbl))
SSblbl
Signal
Concentration
xx
x
x
Concentration
SSblbl
Signal
x x
x x
SensitivitySensitivity Analytical sensitivity (Analytical sensitivity ( = m/s = m/sSS))
m – slope; sm – slope; sSS – standard deviation of the – standard deviation of the measurementmeasurement
Relative insensitive to amplification factors
Signal
Concentration
Increase the gain of the instrument by a Factor of two
Detection limit (Limit of detection, LOD)Detection limit (Limit of detection, LOD)
The analyte concentration giving a signal equal The analyte concentration giving a signal equal the blank signal, Sthe blank signal, Sblbl, plus three time standards , plus three time standards deviation of the blank, sdeviation of the blank, sblbl
Blank signal
Standard deviation ofthe Blank signal
SSmm = = SSblbl + ks + ksblbl
Analytical Signal
Convert signal response, SSmm,, to analyte concentration
Dynamic RangeDynamic RangeLowest concentration (LOQ) to the Lowest concentration (LOQ) to the
concentration where the calibration curve concentration where the calibration curve departure from linearity (limit of departure from linearity (limit of linearity, LOL)linearity, LOL)
Blank signal
Standard deviation ofthe Blank signal
SSmm = = SSblbl + ks + ksblbl
Analytical Signal
Convert signal response, SSmm,, to analyte concentration
Dynamic RangeDynamic Range
Guideline for Reporting Data Guideline for Reporting Data (Recommended by ACS Committee of (Recommended by ACS Committee of Environmental Improvement)Environmental Improvement)
AnalyteAnalyteConcentrationConcentration Region of reliabilityRegion of reliability< 3< 3σσ Region of Questionable detection Region of Questionable detection
(unacceptable)(unacceptable)
33σσ Detection limitDetection limit
33σσ-10-10σσ Region of less certain Region of less certain quantitationquantitation
1010σσ Limit of quantitationLimit of quantitation
> 10> 10σσ Region of quantitationRegion of quantitation
99.7%
Question: When the value of 10 sbl
(standard deviation of the blank) is used as LOQ, what is the relative standard deviation of the measurement at this point at the 99.7% confidential level, and why?
SelectivitySelectivityThe degree to which the method is The degree to which the method is
free from interference by other species free from interference by other species contained in the sample matrixcontained in the sample matrix
S = mS = mAAccAA + m + mBBccBB + m + mCCccC C + S+ Sblbl
The selectivityThe selectivity coefficient for B with respect to coefficient for B with respect to AA
kkB,AB,A = m = mBB/m/mAA
…………………………………………………………………………Note: this is the selectivity of an analytical Note: this is the selectivity of an analytical detectiondetection technique technique
Calibration MethodsCalibration Methods
Analytical response Analytical response
Comparison with StandardComparison with Standard Direct comparisonDirect comparison
ColorimetricColorimetric
TitrationTitration
Analyte Concentration
Calibration MethodsCalibration Methods
External External Calibration CurveCalibration Curve
Signal
Concentration
XX
X
X
Normally use the method of least squares
Calibration MethodsCalibration Methods
External External Calibration CurveCalibration Curve
Two Assumptions:Two Assumptions:
1. Linear relationship1. Linear relationship
2. deviation of the individual point 2. deviation of the individual point from the straight line arises from from the straight line arises from the error in the the error in the measurementmeasurement
Calibration MethodsCalibration Methods Standard AdditionStandard Addition
Signal
Concentration
Without sample matrix effect
With sample matrix effect (signal suppressed)
Standard AdditionStandard Addition
Signal
Concentration
C0 C1 C2 C3
xx
xx
Conc. of analyte C0
Sample signalSbl
C1, C2, and C3 are the concentrations of analyte after spiked with stnard, with counting the amount of analyte in the original sample solution.C0 is the concentrations of analyte in sample without spiking
S = mC + Sbl
C0 =Sbl/m
Internal Standard Internal Standard Add a substance in a constant amount to Add a substance in a constant amount to all samples and, blanks, and calibration all samples and, blanks, and calibration standard in an analysisstandard in an analysis
Concentration
Sanalyte/Sinternal