Influence of Pacing Site Characteristics on Response to Cardiac ...
Cardiac Pacing for the Surgeons
-
Upload
rezwanul-hoque-bulbul -
Category
Documents
-
view
245 -
download
1
Transcript of Cardiac Pacing for the Surgeons
-
8/2/2019 Cardiac Pacing for the Surgeons
1/46
Cardiac Pacing for the Surgeons
Dr. Rezwanul HoqueMBBS, MS, FCPS, FRCSG, FRCSEd
Associate ProfessorDepartment of Cardiac surgery
BSMMU, Dhaka, Bangladesh
-
8/2/2019 Cardiac Pacing for the Surgeons
2/46
A pacemaker or artificial pacemaker is a medical device designed to regulate
the beating of the Heart.
The purpose of an artificial pacemaker is to stimulate the heart when either the
heart's native pacemaker is not fast enough or if there are blocks in the heart's
electrical conduction system preventing the propagation of electrical impulses
from the native pacemaker to the lower chambers of the heart, known as theventricles.
Pacemaker
http://en.wikipedia.org/wiki/Image:Pacemaker_GuidantMeridianSR.jpg -
8/2/2019 Cardiac Pacing for the Surgeons
3/46
An ICD(Implantable cardioverter & defibrillator) is a specialized device designed to
directly treat a cardiac tachydysrhythmia.
If a patient has a ventricular ICD and the device senses a ventricular rate that exceeds
the programmed cut-off rate of the ICD, the device performs
cardioversion/defibrillation.
Alternatively, the device, if so programmed, may attempt to pace rapidly for a number
of pulses to attempt pace-termination of the ventricular tachycardia.
The newer devices are a combination of ICD and pacemaker in one unit. These
combination ICD/pacemakers are implanted in patients who require both devices.
ICD
-
8/2/2019 Cardiac Pacing for the Surgeons
4/46
Transvenous
Transcutaneous
EpicardialTransesophageal
-
8/2/2019 Cardiac Pacing for the Surgeons
5/46
Acute myocardial infarction with:
CHB, Mobitz type 2 AV block, medically refractory symptomatic bradycardia,
alternating BBB, new bifascicular block, new BBB with anterior MI
In absence of acute MI : SSS, CHB, Mobitz type 2 AV block
Treatment of tachyarrhythmias : VT
-
8/2/2019 Cardiac Pacing for the Surgeons
6/46
Transcutaneous pacing
Transcutaneous pacing (TCP), also called external pacing, is recommended for theinitial stabilization of hemodynamically significant bradycardia of all types.
The procedure is performed by placing two pacing pads on the patient's chest, either
in the anterior/lateral position or the anterior/posterior position.
The rescuer selects the pacing rate, and gradually increases the pacing current
(measured in ma) until electrical capture (characterized by a wide QRS complex with a
tall, broad T wave on the ECG) is achieved, with a corresponding pulse.
Pacing artifact on the ECG and severe muscle twitching may make this determination
difficult.
External pacing should not be relied upon for an extended period of time. It is an
emergency procedure that acts as a bridge until transvenous pacing or other therapies
can be applied.
-
8/2/2019 Cardiac Pacing for the Surgeons
7/46
Transvenous pacing
Transvenous pacing, or temporary internal pacing, is an alternative to transcutaneous
pacing.
A wire is placed under sterile conditions via a central venous catheter. The distal tip of
the wire is placed into either the right atrium or right ventricle. The proximal tip of thewire is attached to the pacemaker generator, outside of the body.
Transvenous pacing is often used as a bridge to permanent pacemaker placement.
Under certain conditions, a person may require temporary pacing but would not
require permanent pacing. In this case, a temporary pacing wire may be the optimal
treatment option.
-
8/2/2019 Cardiac Pacing for the Surgeons
8/46
Epicardial pacing
Epicardial pacing wires are routine leads used for temporary cardiac pacing after open
heart surgery.
During surgery the heart is subjected to stress which can lead to myocardial ischemia
and cardiac depression.
Temporary pacing may be necessary to re-establish electrical conduction.
Atrial wires (a wires) are positioned on the right atrial surface and exit the chest wall to
the right of the sternum. Atrial wires are attached via a connecting cable to A marked
terminals on the pacemaker box.
Ventricular wires (v wires) are positioned on the right ventricular surface and exit thechest wall to the left of the sternum. Ventricular wires are attached via connecting cable
to V marked terminals on the pacemaker box.
If a skin wire is used it is attached to positive pole and the ventricular wire is attached to
negative pole of the generator.
Epicardial pacing after open heart surgery allows for the treatment of dysrhythmias, the
improvement of hemodynamic functioning, and to maintain backup rate with maze
procedure.
The patient should remain on a cardiac monitor 2-hours after the removal of epicardial
wires. The patient cannot have an MRI while pacing wires are in place.
-
8/2/2019 Cardiac Pacing for the Surgeons
9/46
FAILURE TO CAPTURE/PACE:
MEASURES:
Secure connections
Replace battery/pacemaker as needed
Increase ma
Reduce sensitivity (turn mv dial counter clockwise or to a higher numericalsetting).
Reverse polarity
Check stimulation threshold
Monitor electrolytes and ABG
Obtain EKG to check for ischemia (ST depression)
-
8/2/2019 Cardiac Pacing for the Surgeons
10/46
Eliminate any electrical interference.Note: check to ensure that all electrical equipment attached to the patient isgrounded
Replace battery/pacemaker as needed
If undersensing [not recognizing patients' own rhythm]increase sensitivity(turn mv dial clockwise to lower mv numerical value)
Note: this will make the pacemaker more sensitive to patients own rate
If oversensing [when A-V pacing, the pacemaker may sense the atrial pacingspike as the qrs complex and not fire (cross talk) or the patient is havingbreakthrough rate and pacemaker does not fire] decrease sensitivity (turn mvdial counter clockwise to raise the numerical mv value)
Note: this will make the pacemaker less sensitive to the patient's rate.
If patient has an adequate underlying rhythm, turn off the pacemaker
-
8/2/2019 Cardiac Pacing for the Surgeons
11/46
When to consider PPM
Patients with symptoms directly attributable tobradycardia (even SB)
Disease within AV node as manifest by extreme PRprolongation, normal QRS
Disease belowAV node as manifest by normal or mildlyprolonged PR and wide QRS
Disease in His purkinje system (less stable thanabove)
-
8/2/2019 Cardiac Pacing for the Surgeons
12/46
Degree Pacemaker necessary Pacemakerprobablynecessary
Pacemaker notnecessary
Third Symptomatic congenitalcomplete heart block
Aquired symptomatic completeheart block
Atrial fibrillation with completeheart block
Acquired asymptomaticcomplete heart block
Second Symptomatic type ISymptomatic type II Asymptomatictype IIAsymptomatictype I at intra-Hisor infra-His levels
Asymptomatic typeI at supra-His (AVnodal) block
First Asymptomatic orsymptomatic
-
8/2/2019 Cardiac Pacing for the Surgeons
13/46
Pacemaker Pacemaker probably
necessary
Pacemaker not
necessary
Symptomatic bradycardia Symptomatic patients withsinus node dysfunction
with documented rates of
-
8/2/2019 Cardiac Pacing for the Surgeons
14/46
Pacing for HemodynamicImprovement
Hypertrophic Obstructive
Cardiomyopathy
Cardiac ResynchronizationTherapy
Evolving Indications for Pacing
Long QT syndromes
Sleep apneaNeurally mediated syncope
-
8/2/2019 Cardiac Pacing for the Surgeons
15/46
Patient
Lead
Pacemaker
Programmer
Lead
Pacemaker
-
8/2/2019 Cardiac Pacing for the Surgeons
16/46
Basic pacemaker function
Modern pacemakers usually have multiple functions.
The most basic form listens to the heart's native electrical rhythm, and if the device
doesn't sense any electrical activity within a certain time period, the device will stimulatethe vetricles of heart with a set amount of energy, measured in joules.
The more complex forms include the ability to sense and/or stimulate both the atrial
and ventricular chambers.
-
8/2/2019 Cardiac Pacing for the Surgeons
17/46
Permanent pacing
Permanent pacing with an implantable pacemaker involves placement of one or more
pacing wires within the chambers of the heart. One end of each wire is attached to the
muscle of the heart. The other end is screwed into the pacemaker generator.
The pacemaker generator is a hermetically sealed device containing a power sourceand the computer logic for the pacemaker.
Most commonly, the generator is placed below the subcutaneous fat of the chest wall,
above the muscles and bones of the chest.
The outer casing of pacemakers is so designed that it will rarely be rejected by the
body's immune system. It is usually made of titanium, which is very inert in the body.
-
8/2/2019 Cardiac Pacing for the Surgeons
18/46
Codes are combined to describe:
Mode of pacing
Mode of sensing
How the pacemaker will respond to the
presence or absence of intrinsic beats
-
8/2/2019 Cardiac Pacing for the Surgeons
19/46
Responseto Sensing
III IV
NASPE / BPEG Generic (NBG) Code
Bernstein, A.D., et al. PACE 2002; 25:260-264
Letters Used:
O
R
None
Atrium
Ventricle
Dual A+V
O O None O None None O None
A A Atrium T TriggeredRate
Modulation
A Atrium
V V Ventricle I Inhibited V Ventricle
D D Dual A+V D Dual T+I D Dual A+V
Chamber(s)Paced
Chamber(s)Sensed
RateModulation
MultisitePacing
Category:
I II VPosition:
SingleA or V
S S SingleA or V
Manufacturers Designation Only:
-
8/2/2019 Cardiac Pacing for the Surgeons
20/46
Pacemaker energy output is dependent upon the signal amplitude and pulsewidth. Signal amplitude is measured in electrical units of volts ormilliamperes. Pulse width is a measure of output duration and is measured inmilliseconds. For proper permanent pacer operation, signal amplitude andwidth are set high enough to reliably achieve capture of the myocardium, yetlow enough to prolong battery life.
Pulse generators can be set to a fixed-rate (asynchronous) or demand(synchronous) mode. In the fixed-rate mode, an impulse is produced at a setrate and has no relationship to the patient's intrinsic cardiac activity. Thismode carries a small but inherent danger of producing lethal dysrhythmiasshould the impulse coincide with the vulnerable period of the T wave. In thedemand mode, the sensing circuit searches for an intrinsic depolarization
potential. If this is absent, a pacing response is generated. This mode canclosely mimic the intrinsic electric activity pattern of the heart.
-
8/2/2019 Cardiac Pacing for the Surgeons
21/46
Ventricular pacing
No sensing
Ventricular asynchronous pacing at lower programmedpacing rate
VOO
Ventricular lead
-
8/2/2019 Cardiac Pacing for the Surgeons
22/46
VVI
Ventricular Lead (I)
Ventricular pacing
Ventricular sensing
Sensed intrinsic QRS inhibits ventricular pacing
-
8/2/2019 Cardiac Pacing for the Surgeons
23/46
AOO
Atrial pacing
No sensing
Atrial asynchronous pacing at lower programmed pacingrate
Atrial Lead
-
8/2/2019 Cardiac Pacing for the Surgeons
24/46
AAI
Atrial pacing
Atrial sensing
Intrinsic P-wave inhibits atrial pacing
Atrial Lead (I)
-
8/2/2019 Cardiac Pacing for the Surgeons
25/46
Pacing in atrium and ventricle
Sensing in atrium and ventricle
Intrinsic P-wave and intrinsic QRS can inhibit pacing
Intrinsic P-wave can trigger a paced QRS
VentricularLead (I)
Atrial Lead(T/I)
DDD Pacing
-
8/2/2019 Cardiac Pacing for the Surgeons
26/46
Dual-chamber pacing capable of pacing and sensing in both the atrial and ventricularchambers of the heart
Four distinct patterns can be observed with DDD pacing
Sensing in the atrium and sensing in the ventricle (AsVs)
Sensing in the atrium and pacing in the ventricle(P wave tracking) (AsVp)
Pacing in the atrium and sensing in the ventricle (ApVs)
Pacing in the atrium and pacing in the ventricle (ApVp) Adapts to changes post-implant
May resemble AAI, VAT, VDD, DVI modes
Will strive to maintain AV synchrony with variable atrial rates and AV conduction
DDD Pacing
-
8/2/2019 Cardiac Pacing for the Surgeons
27/46
In rate-responsive pacing (modes ending with R), sensor(s) inthe pacemaker are used to detect changes in physiologic needsand increase the pacing rate accordingly
Sensors
Detect changes in metabolic demand
Sense motion or physiologic indicator
accelerometer or minute ventilationDefinition: Chronotropic Incompetence
inability to increase and maintain heart rateappropriately with exercise*
*Ellenbogen, Kenneth A. Cardiac Pacing and ICDs. Malden, Mass.: Blackwell Pub.,2005.
-
8/2/2019 Cardiac Pacing for the Surgeons
28/46
Example of dual-chamber, rate-responsive pacing
-
8/2/2019 Cardiac Pacing for the Surgeons
29/46
Mode Selection Considerations
Status of Atrial Rhythm- Intrinsic vs. Paced- Presence of Atrial Tachyarrhythmias:
Acute/Chronic
Status of AV ConductionNormal -Slowed-Blocked
Presence of ChronotropicIncompetence
-
8/2/2019 Cardiac Pacing for the Surgeons
30/46
Mode Representative Indications
VVI Rare Need for Pacing Simple; System Desirable (Old, Debilitated Patient; Very Young)
VVIR Chronic AF With Symptomatic Pauses; Inadequate Ventricular Response
After AV Nodal Ablation for Rate Control in AF
AAI Symptomatic Sinus Pauses With Otherwise Normal Sinus Node Function and Intact AV
Conduction
AAIR Sinus Node Dysfunction With Intact AV Conduction and No Atrial Arrhythmias
VDD AV Block With Normal Sinus Node Function; Uses Single Lead; No Atrial Pacing,
Therefore Not Appropriate for Patients With Sinus Bradycardia
DDD AV Block With Normal Sinus Node Function
DDDR AV Block With Sinus Node Dysfunction; Mode Switching Feature if AnyParoxysmal Atrial Arrhythmias
DDIR An Alternate to DDDR for Patients Who Require Dual-Chamber Pacing but Have
Paroxysmal Atrial Arrhythmias Does Not Permit Atrial Tracking in Sinus Rhythm
-
8/2/2019 Cardiac Pacing for the Surgeons
31/46
Biventricular Pacing (BVP)
A biventricular pacemaker, also known as CRT (cardiac resynchronisation therapy) is a
type of pacemaker that can pace both ventricles (right and left) of the heart.
By pacing both sides of the heart, the pacemaker can resynchronize a heart that doesnot beat in synchrony, which is common in heart failure patients.
CRT devices have three leads, one in the atrium, one in the right ventricle, and a final
one is inserted through the coronary sinus to pace the left ventricle.
CRT devices are shown to reduce mortality and improve quality of life in groups ofheart failure patients.
-
8/2/2019 Cardiac Pacing for the Surgeons
32/46
Coronary sinus lead
Right atrial lead
Right ventricular lead
N Engl J Med 2003
-
8/2/2019 Cardiac Pacing for the Surgeons
33/46
LV lead
RV coil
SVC coil
RA lead
-
8/2/2019 Cardiac Pacing for the Surgeons
34/46
Venous accessPneumothorax, hemothoraxAir embolismPerforation of central veinInadvertent arterial entry
Lead placementBrady tachyarrhythmiaPerforation of heart, veinDamage to heart valve
GeneratorPocket hematomaImproper or inadequate connection of lead
-
8/2/2019 Cardiac Pacing for the Surgeons
35/46
Lead-relatedThrombosis/embolizationSVC obstructionLead dislodgementInfectionLead failure
Perforation, pericarditis
Generator-relatedPainErosion, infectionMigration
Damage from radiation, electric shock
Patient-relatedTwiddler syndrome
-
8/2/2019 Cardiac Pacing for the Surgeons
36/46
Fatigue, dizziness, hypotension
Caused by pacing the ventricle asynchronously, resulting in AV dissociation orVA conduction
Mechanism: atrial contraction against a closed AV valve and release of atrialnatriuretic peptide
Worsened by increasing the ventricular pacing rate, relieved by lowering the
pacing rate or upgrading to dual chamber system
Therapy with fludrocortisone/volume expansion NOT helpful
-
8/2/2019 Cardiac Pacing for the Surgeons
37/46
Failure to output
Failure to output occurs when no pacing spike is present despite an indication to
pace. This may be due to battery failure, lead fracture, a break in lead insulation,
oversensing (inhibiting pacer output), poor lead connection at the takeoff from the
pacer, and "cross-talk" (i.e., a phenomenon seen when atrial output is sensed by a
ventricular lead in a dual-chamber pacer).
Management of pacer output complications includes medications to increase the
intrinsic heart rate and placement of a temporary pacer. A chest radiograph is
warranted to check pacer leads, with close scrutiny to evaluate for possible lead
fracture, which occurs most commonly at the clavicle/first rib location. The
patient's pacer identification card should be obtained and his/her
electrophysiologist/cardiologist consulted.
-
8/2/2019 Cardiac Pacing for the Surgeons
38/46
Failure to capture
Failure to capture occurs when a pacing spike is not followed by either an
atrial or a ventricular complex.
This may be due to lead fracture, lead dislodgement, a break in lead
insulation, an elevated pacing threshold, myocardial infarction at the lead
tip, certain drugs (eg, flecainide), metabolic abnormalities (eg, hyperkalemia,acidosis, alkalosis), cardiac perforation, poor lead connection at the takeoff
from the generator, and improper amplitude or pulse width settings.
Management of pacer capture complications is the same as for output
complications, with extra consideration given to treating metabolic
abnormalities and potential myocardial infarction.
-
8/2/2019 Cardiac Pacing for the Surgeons
39/46
Oversensing
Oversensing occurs when a pacer incorrectly senses noncardiac electrical activity and
is inhibited from correctly pacing.
This may be due to muscular activity, particularly oversensing of the diaphragm or
pectoralis muscles, electromagnetic interference such as MRIs, or lead insulation
breakage. More recently, cellular phones held within 10 cm of the pulse generator
may elicit this response.
Undersensing
Undersensing occurs when a pacer incorrectly misses intrinsic depolarization and
paces despite intrinsic activity.
This may be due to poor lead positioning, lead dislodgment, magnet application, lowbattery states, or myocardial infarction. Management is similar to that for other types
of failures.
-
8/2/2019 Cardiac Pacing for the Surgeons
40/46
Operative failures
A final category of pacer failures is termed operative. This includes malfunction
due to mechanical factors, such as pneumothorax, pericarditis, infection, skin
erosion, hematoma, lead dislodgment, and venous thrombosis.
Treatment depends on the etiology.
Pneumothoraces, being dependent on size, may require chest thoracostomy.
Erosion of the pacer through the skin, while rare, requires pacer replacement
and systemic antibiotics. Hematomas may require drainage. Lead dislodgmentusually occurs within 2 days following implantation of a permanent pacer and
may be seen on chest radiography. If the lead is floating freely in the ventricle,
malignant arrhythmias may develop. Thrombosis is rare and usually presents as
unilateral arm edema. Treatment includes arm elevation and anticoagulation.
Advanced life support protocols, including defibrillation, may safely be executed
in patients with pacemakers in place. Sternal paddles are placed at a safe
distance (10 cm) from the pulse generator. Temporary transcutaneous pacing
may become necessary in cases of myocardial infarction.
-
8/2/2019 Cardiac Pacing for the Surgeons
41/46
MedicalMRI
Lithotripsy
Electrocautery/cryosurgery
External defibrillators
Therapeutic radiation
NonmedicalArc welding equipment
Automobile engines
Radar Transmitters
-
8/2/2019 Cardiac Pacing for the Surgeons
42/46
Devices with pacemaker function
Sometimes devices resembling pacemakers, called ICDs (implantable cardioverter-
defibrillator) are implanted.
These devices are often used in the treatment of patients at risk from sudden cardiac
death.
An ICD has the ability to treat many types of heart rhythm disturbances by means of
pacing, cardioversion, or defibrillation.
-
8/2/2019 Cardiac Pacing for the Surgeons
43/46
Secondary prevention of SCD:VF arrest, sustained VT not secondary to reversible
cause
Primary prevention of SCD:
LVEF < 36%, class II-III symptoms of CHF
CAD, h/o MI, LVEF 40%, inducible sustained VT
Familial or inherited conditions with high risk for SCD:HCM, long QT syndrome, Brugada syndrome
-
8/2/2019 Cardiac Pacing for the Surgeons
44/46
Pacemaker programming can be performed noninvasively by an electrophysiologist or
cardiologist. Because of the myriad of pacemaker types, patients should carry a card
with them providing information about their particular model.
This information is crucial when communicating with the cardiologist about a pacer
problem.
Most pacemaker generators, however, have an x-ray code that can be seen on a
standard chest x-ray. The markings, along with the shape of the generator, may assist
with deciphering the manufacturer of the generator and pacemaker battery.
This may be helpful in the event a patient neither recalls the company nor has thepermanent pacemaker card.
-
8/2/2019 Cardiac Pacing for the Surgeons
45/46
Placing a magnet over a permanent pacemaker causes sensing to be inhibited by
closing an internal reed switch. This only temporarily "reprograms" the pacer into the
asynchronous mode, where pacing is initiated at a set rate. It does not turn the
pacemaker off.
Each pacemaker type has a unique asynchronous rate for beginning-of-life (BOL),
elective replacement indicator (ERI), and end-of-life (EOL).
Therefore, application of a magnet can determine if the pacer's battery needs to be
replaced.
Further interrogation, or manipulating of the device, should be performed by an
individual skilled in the technique.Patients should carry a card that contains information about their particular
pacemaker, since these rates are dependent on the manufacturer and the model.
-
8/2/2019 Cardiac Pacing for the Surgeons
46/46
+8801711560305
mailto:[email protected]:[email protected]