Cardiac Pacing for the Surgeons

download Cardiac Pacing for the Surgeons

of 46

Transcript of Cardiac Pacing for the Surgeons

  • 8/2/2019 Cardiac Pacing for the Surgeons

    1/46

    Cardiac Pacing for the Surgeons

    Dr. Rezwanul HoqueMBBS, MS, FCPS, FRCSG, FRCSEd

    Associate ProfessorDepartment of Cardiac surgery

    BSMMU, Dhaka, Bangladesh

  • 8/2/2019 Cardiac Pacing for the Surgeons

    2/46

    A pacemaker or artificial pacemaker is a medical device designed to regulate

    the beating of the Heart.

    The purpose of an artificial pacemaker is to stimulate the heart when either the

    heart's native pacemaker is not fast enough or if there are blocks in the heart's

    electrical conduction system preventing the propagation of electrical impulses

    from the native pacemaker to the lower chambers of the heart, known as theventricles.

    Pacemaker

    http://en.wikipedia.org/wiki/Image:Pacemaker_GuidantMeridianSR.jpg
  • 8/2/2019 Cardiac Pacing for the Surgeons

    3/46

    An ICD(Implantable cardioverter & defibrillator) is a specialized device designed to

    directly treat a cardiac tachydysrhythmia.

    If a patient has a ventricular ICD and the device senses a ventricular rate that exceeds

    the programmed cut-off rate of the ICD, the device performs

    cardioversion/defibrillation.

    Alternatively, the device, if so programmed, may attempt to pace rapidly for a number

    of pulses to attempt pace-termination of the ventricular tachycardia.

    The newer devices are a combination of ICD and pacemaker in one unit. These

    combination ICD/pacemakers are implanted in patients who require both devices.

    ICD

  • 8/2/2019 Cardiac Pacing for the Surgeons

    4/46

    Transvenous

    Transcutaneous

    EpicardialTransesophageal

  • 8/2/2019 Cardiac Pacing for the Surgeons

    5/46

    Acute myocardial infarction with:

    CHB, Mobitz type 2 AV block, medically refractory symptomatic bradycardia,

    alternating BBB, new bifascicular block, new BBB with anterior MI

    In absence of acute MI : SSS, CHB, Mobitz type 2 AV block

    Treatment of tachyarrhythmias : VT

  • 8/2/2019 Cardiac Pacing for the Surgeons

    6/46

    Transcutaneous pacing

    Transcutaneous pacing (TCP), also called external pacing, is recommended for theinitial stabilization of hemodynamically significant bradycardia of all types.

    The procedure is performed by placing two pacing pads on the patient's chest, either

    in the anterior/lateral position or the anterior/posterior position.

    The rescuer selects the pacing rate, and gradually increases the pacing current

    (measured in ma) until electrical capture (characterized by a wide QRS complex with a

    tall, broad T wave on the ECG) is achieved, with a corresponding pulse.

    Pacing artifact on the ECG and severe muscle twitching may make this determination

    difficult.

    External pacing should not be relied upon for an extended period of time. It is an

    emergency procedure that acts as a bridge until transvenous pacing or other therapies

    can be applied.

  • 8/2/2019 Cardiac Pacing for the Surgeons

    7/46

    Transvenous pacing

    Transvenous pacing, or temporary internal pacing, is an alternative to transcutaneous

    pacing.

    A wire is placed under sterile conditions via a central venous catheter. The distal tip of

    the wire is placed into either the right atrium or right ventricle. The proximal tip of thewire is attached to the pacemaker generator, outside of the body.

    Transvenous pacing is often used as a bridge to permanent pacemaker placement.

    Under certain conditions, a person may require temporary pacing but would not

    require permanent pacing. In this case, a temporary pacing wire may be the optimal

    treatment option.

  • 8/2/2019 Cardiac Pacing for the Surgeons

    8/46

    Epicardial pacing

    Epicardial pacing wires are routine leads used for temporary cardiac pacing after open

    heart surgery.

    During surgery the heart is subjected to stress which can lead to myocardial ischemia

    and cardiac depression.

    Temporary pacing may be necessary to re-establish electrical conduction.

    Atrial wires (a wires) are positioned on the right atrial surface and exit the chest wall to

    the right of the sternum. Atrial wires are attached via a connecting cable to A marked

    terminals on the pacemaker box.

    Ventricular wires (v wires) are positioned on the right ventricular surface and exit thechest wall to the left of the sternum. Ventricular wires are attached via connecting cable

    to V marked terminals on the pacemaker box.

    If a skin wire is used it is attached to positive pole and the ventricular wire is attached to

    negative pole of the generator.

    Epicardial pacing after open heart surgery allows for the treatment of dysrhythmias, the

    improvement of hemodynamic functioning, and to maintain backup rate with maze

    procedure.

    The patient should remain on a cardiac monitor 2-hours after the removal of epicardial

    wires. The patient cannot have an MRI while pacing wires are in place.

  • 8/2/2019 Cardiac Pacing for the Surgeons

    9/46

    FAILURE TO CAPTURE/PACE:

    MEASURES:

    Secure connections

    Replace battery/pacemaker as needed

    Increase ma

    Reduce sensitivity (turn mv dial counter clockwise or to a higher numericalsetting).

    Reverse polarity

    Check stimulation threshold

    Monitor electrolytes and ABG

    Obtain EKG to check for ischemia (ST depression)

  • 8/2/2019 Cardiac Pacing for the Surgeons

    10/46

    Eliminate any electrical interference.Note: check to ensure that all electrical equipment attached to the patient isgrounded

    Replace battery/pacemaker as needed

    If undersensing [not recognizing patients' own rhythm]increase sensitivity(turn mv dial clockwise to lower mv numerical value)

    Note: this will make the pacemaker more sensitive to patients own rate

    If oversensing [when A-V pacing, the pacemaker may sense the atrial pacingspike as the qrs complex and not fire (cross talk) or the patient is havingbreakthrough rate and pacemaker does not fire] decrease sensitivity (turn mvdial counter clockwise to raise the numerical mv value)

    Note: this will make the pacemaker less sensitive to the patient's rate.

    If patient has an adequate underlying rhythm, turn off the pacemaker

  • 8/2/2019 Cardiac Pacing for the Surgeons

    11/46

    When to consider PPM

    Patients with symptoms directly attributable tobradycardia (even SB)

    Disease within AV node as manifest by extreme PRprolongation, normal QRS

    Disease belowAV node as manifest by normal or mildlyprolonged PR and wide QRS

    Disease in His purkinje system (less stable thanabove)

  • 8/2/2019 Cardiac Pacing for the Surgeons

    12/46

    Degree Pacemaker necessary Pacemakerprobablynecessary

    Pacemaker notnecessary

    Third Symptomatic congenitalcomplete heart block

    Aquired symptomatic completeheart block

    Atrial fibrillation with completeheart block

    Acquired asymptomaticcomplete heart block

    Second Symptomatic type ISymptomatic type II Asymptomatictype IIAsymptomatictype I at intra-Hisor infra-His levels

    Asymptomatic typeI at supra-His (AVnodal) block

    First Asymptomatic orsymptomatic

  • 8/2/2019 Cardiac Pacing for the Surgeons

    13/46

    Pacemaker Pacemaker probably

    necessary

    Pacemaker not

    necessary

    Symptomatic bradycardia Symptomatic patients withsinus node dysfunction

    with documented rates of

  • 8/2/2019 Cardiac Pacing for the Surgeons

    14/46

    Pacing for HemodynamicImprovement

    Hypertrophic Obstructive

    Cardiomyopathy

    Cardiac ResynchronizationTherapy

    Evolving Indications for Pacing

    Long QT syndromes

    Sleep apneaNeurally mediated syncope

  • 8/2/2019 Cardiac Pacing for the Surgeons

    15/46

    Patient

    Lead

    Pacemaker

    Programmer

    Lead

    Pacemaker

  • 8/2/2019 Cardiac Pacing for the Surgeons

    16/46

    Basic pacemaker function

    Modern pacemakers usually have multiple functions.

    The most basic form listens to the heart's native electrical rhythm, and if the device

    doesn't sense any electrical activity within a certain time period, the device will stimulatethe vetricles of heart with a set amount of energy, measured in joules.

    The more complex forms include the ability to sense and/or stimulate both the atrial

    and ventricular chambers.

  • 8/2/2019 Cardiac Pacing for the Surgeons

    17/46

    Permanent pacing

    Permanent pacing with an implantable pacemaker involves placement of one or more

    pacing wires within the chambers of the heart. One end of each wire is attached to the

    muscle of the heart. The other end is screwed into the pacemaker generator.

    The pacemaker generator is a hermetically sealed device containing a power sourceand the computer logic for the pacemaker.

    Most commonly, the generator is placed below the subcutaneous fat of the chest wall,

    above the muscles and bones of the chest.

    The outer casing of pacemakers is so designed that it will rarely be rejected by the

    body's immune system. It is usually made of titanium, which is very inert in the body.

  • 8/2/2019 Cardiac Pacing for the Surgeons

    18/46

    Codes are combined to describe:

    Mode of pacing

    Mode of sensing

    How the pacemaker will respond to the

    presence or absence of intrinsic beats

  • 8/2/2019 Cardiac Pacing for the Surgeons

    19/46

    Responseto Sensing

    III IV

    NASPE / BPEG Generic (NBG) Code

    Bernstein, A.D., et al. PACE 2002; 25:260-264

    Letters Used:

    O

    R

    None

    Atrium

    Ventricle

    Dual A+V

    O O None O None None O None

    A A Atrium T TriggeredRate

    Modulation

    A Atrium

    V V Ventricle I Inhibited V Ventricle

    D D Dual A+V D Dual T+I D Dual A+V

    Chamber(s)Paced

    Chamber(s)Sensed

    RateModulation

    MultisitePacing

    Category:

    I II VPosition:

    SingleA or V

    S S SingleA or V

    Manufacturers Designation Only:

  • 8/2/2019 Cardiac Pacing for the Surgeons

    20/46

    Pacemaker energy output is dependent upon the signal amplitude and pulsewidth. Signal amplitude is measured in electrical units of volts ormilliamperes. Pulse width is a measure of output duration and is measured inmilliseconds. For proper permanent pacer operation, signal amplitude andwidth are set high enough to reliably achieve capture of the myocardium, yetlow enough to prolong battery life.

    Pulse generators can be set to a fixed-rate (asynchronous) or demand(synchronous) mode. In the fixed-rate mode, an impulse is produced at a setrate and has no relationship to the patient's intrinsic cardiac activity. Thismode carries a small but inherent danger of producing lethal dysrhythmiasshould the impulse coincide with the vulnerable period of the T wave. In thedemand mode, the sensing circuit searches for an intrinsic depolarization

    potential. If this is absent, a pacing response is generated. This mode canclosely mimic the intrinsic electric activity pattern of the heart.

  • 8/2/2019 Cardiac Pacing for the Surgeons

    21/46

    Ventricular pacing

    No sensing

    Ventricular asynchronous pacing at lower programmedpacing rate

    VOO

    Ventricular lead

  • 8/2/2019 Cardiac Pacing for the Surgeons

    22/46

    VVI

    Ventricular Lead (I)

    Ventricular pacing

    Ventricular sensing

    Sensed intrinsic QRS inhibits ventricular pacing

  • 8/2/2019 Cardiac Pacing for the Surgeons

    23/46

    AOO

    Atrial pacing

    No sensing

    Atrial asynchronous pacing at lower programmed pacingrate

    Atrial Lead

  • 8/2/2019 Cardiac Pacing for the Surgeons

    24/46

    AAI

    Atrial pacing

    Atrial sensing

    Intrinsic P-wave inhibits atrial pacing

    Atrial Lead (I)

  • 8/2/2019 Cardiac Pacing for the Surgeons

    25/46

    Pacing in atrium and ventricle

    Sensing in atrium and ventricle

    Intrinsic P-wave and intrinsic QRS can inhibit pacing

    Intrinsic P-wave can trigger a paced QRS

    VentricularLead (I)

    Atrial Lead(T/I)

    DDD Pacing

  • 8/2/2019 Cardiac Pacing for the Surgeons

    26/46

    Dual-chamber pacing capable of pacing and sensing in both the atrial and ventricularchambers of the heart

    Four distinct patterns can be observed with DDD pacing

    Sensing in the atrium and sensing in the ventricle (AsVs)

    Sensing in the atrium and pacing in the ventricle(P wave tracking) (AsVp)

    Pacing in the atrium and sensing in the ventricle (ApVs)

    Pacing in the atrium and pacing in the ventricle (ApVp) Adapts to changes post-implant

    May resemble AAI, VAT, VDD, DVI modes

    Will strive to maintain AV synchrony with variable atrial rates and AV conduction

    DDD Pacing

  • 8/2/2019 Cardiac Pacing for the Surgeons

    27/46

    In rate-responsive pacing (modes ending with R), sensor(s) inthe pacemaker are used to detect changes in physiologic needsand increase the pacing rate accordingly

    Sensors

    Detect changes in metabolic demand

    Sense motion or physiologic indicator

    accelerometer or minute ventilationDefinition: Chronotropic Incompetence

    inability to increase and maintain heart rateappropriately with exercise*

    *Ellenbogen, Kenneth A. Cardiac Pacing and ICDs. Malden, Mass.: Blackwell Pub.,2005.

  • 8/2/2019 Cardiac Pacing for the Surgeons

    28/46

    Example of dual-chamber, rate-responsive pacing

  • 8/2/2019 Cardiac Pacing for the Surgeons

    29/46

    Mode Selection Considerations

    Status of Atrial Rhythm- Intrinsic vs. Paced- Presence of Atrial Tachyarrhythmias:

    Acute/Chronic

    Status of AV ConductionNormal -Slowed-Blocked

    Presence of ChronotropicIncompetence

  • 8/2/2019 Cardiac Pacing for the Surgeons

    30/46

    Mode Representative Indications

    VVI Rare Need for Pacing Simple; System Desirable (Old, Debilitated Patient; Very Young)

    VVIR Chronic AF With Symptomatic Pauses; Inadequate Ventricular Response

    After AV Nodal Ablation for Rate Control in AF

    AAI Symptomatic Sinus Pauses With Otherwise Normal Sinus Node Function and Intact AV

    Conduction

    AAIR Sinus Node Dysfunction With Intact AV Conduction and No Atrial Arrhythmias

    VDD AV Block With Normal Sinus Node Function; Uses Single Lead; No Atrial Pacing,

    Therefore Not Appropriate for Patients With Sinus Bradycardia

    DDD AV Block With Normal Sinus Node Function

    DDDR AV Block With Sinus Node Dysfunction; Mode Switching Feature if AnyParoxysmal Atrial Arrhythmias

    DDIR An Alternate to DDDR for Patients Who Require Dual-Chamber Pacing but Have

    Paroxysmal Atrial Arrhythmias Does Not Permit Atrial Tracking in Sinus Rhythm

  • 8/2/2019 Cardiac Pacing for the Surgeons

    31/46

    Biventricular Pacing (BVP)

    A biventricular pacemaker, also known as CRT (cardiac resynchronisation therapy) is a

    type of pacemaker that can pace both ventricles (right and left) of the heart.

    By pacing both sides of the heart, the pacemaker can resynchronize a heart that doesnot beat in synchrony, which is common in heart failure patients.

    CRT devices have three leads, one in the atrium, one in the right ventricle, and a final

    one is inserted through the coronary sinus to pace the left ventricle.

    CRT devices are shown to reduce mortality and improve quality of life in groups ofheart failure patients.

  • 8/2/2019 Cardiac Pacing for the Surgeons

    32/46

    Coronary sinus lead

    Right atrial lead

    Right ventricular lead

    N Engl J Med 2003

  • 8/2/2019 Cardiac Pacing for the Surgeons

    33/46

    LV lead

    RV coil

    SVC coil

    RA lead

  • 8/2/2019 Cardiac Pacing for the Surgeons

    34/46

    Venous accessPneumothorax, hemothoraxAir embolismPerforation of central veinInadvertent arterial entry

    Lead placementBrady tachyarrhythmiaPerforation of heart, veinDamage to heart valve

    GeneratorPocket hematomaImproper or inadequate connection of lead

  • 8/2/2019 Cardiac Pacing for the Surgeons

    35/46

    Lead-relatedThrombosis/embolizationSVC obstructionLead dislodgementInfectionLead failure

    Perforation, pericarditis

    Generator-relatedPainErosion, infectionMigration

    Damage from radiation, electric shock

    Patient-relatedTwiddler syndrome

  • 8/2/2019 Cardiac Pacing for the Surgeons

    36/46

    Fatigue, dizziness, hypotension

    Caused by pacing the ventricle asynchronously, resulting in AV dissociation orVA conduction

    Mechanism: atrial contraction against a closed AV valve and release of atrialnatriuretic peptide

    Worsened by increasing the ventricular pacing rate, relieved by lowering the

    pacing rate or upgrading to dual chamber system

    Therapy with fludrocortisone/volume expansion NOT helpful

  • 8/2/2019 Cardiac Pacing for the Surgeons

    37/46

    Failure to output

    Failure to output occurs when no pacing spike is present despite an indication to

    pace. This may be due to battery failure, lead fracture, a break in lead insulation,

    oversensing (inhibiting pacer output), poor lead connection at the takeoff from the

    pacer, and "cross-talk" (i.e., a phenomenon seen when atrial output is sensed by a

    ventricular lead in a dual-chamber pacer).

    Management of pacer output complications includes medications to increase the

    intrinsic heart rate and placement of a temporary pacer. A chest radiograph is

    warranted to check pacer leads, with close scrutiny to evaluate for possible lead

    fracture, which occurs most commonly at the clavicle/first rib location. The

    patient's pacer identification card should be obtained and his/her

    electrophysiologist/cardiologist consulted.

  • 8/2/2019 Cardiac Pacing for the Surgeons

    38/46

    Failure to capture

    Failure to capture occurs when a pacing spike is not followed by either an

    atrial or a ventricular complex.

    This may be due to lead fracture, lead dislodgement, a break in lead

    insulation, an elevated pacing threshold, myocardial infarction at the lead

    tip, certain drugs (eg, flecainide), metabolic abnormalities (eg, hyperkalemia,acidosis, alkalosis), cardiac perforation, poor lead connection at the takeoff

    from the generator, and improper amplitude or pulse width settings.

    Management of pacer capture complications is the same as for output

    complications, with extra consideration given to treating metabolic

    abnormalities and potential myocardial infarction.

  • 8/2/2019 Cardiac Pacing for the Surgeons

    39/46

    Oversensing

    Oversensing occurs when a pacer incorrectly senses noncardiac electrical activity and

    is inhibited from correctly pacing.

    This may be due to muscular activity, particularly oversensing of the diaphragm or

    pectoralis muscles, electromagnetic interference such as MRIs, or lead insulation

    breakage. More recently, cellular phones held within 10 cm of the pulse generator

    may elicit this response.

    Undersensing

    Undersensing occurs when a pacer incorrectly misses intrinsic depolarization and

    paces despite intrinsic activity.

    This may be due to poor lead positioning, lead dislodgment, magnet application, lowbattery states, or myocardial infarction. Management is similar to that for other types

    of failures.

  • 8/2/2019 Cardiac Pacing for the Surgeons

    40/46

    Operative failures

    A final category of pacer failures is termed operative. This includes malfunction

    due to mechanical factors, such as pneumothorax, pericarditis, infection, skin

    erosion, hematoma, lead dislodgment, and venous thrombosis.

    Treatment depends on the etiology.

    Pneumothoraces, being dependent on size, may require chest thoracostomy.

    Erosion of the pacer through the skin, while rare, requires pacer replacement

    and systemic antibiotics. Hematomas may require drainage. Lead dislodgmentusually occurs within 2 days following implantation of a permanent pacer and

    may be seen on chest radiography. If the lead is floating freely in the ventricle,

    malignant arrhythmias may develop. Thrombosis is rare and usually presents as

    unilateral arm edema. Treatment includes arm elevation and anticoagulation.

    Advanced life support protocols, including defibrillation, may safely be executed

    in patients with pacemakers in place. Sternal paddles are placed at a safe

    distance (10 cm) from the pulse generator. Temporary transcutaneous pacing

    may become necessary in cases of myocardial infarction.

  • 8/2/2019 Cardiac Pacing for the Surgeons

    41/46

    MedicalMRI

    Lithotripsy

    Electrocautery/cryosurgery

    External defibrillators

    Therapeutic radiation

    NonmedicalArc welding equipment

    Automobile engines

    Radar Transmitters

  • 8/2/2019 Cardiac Pacing for the Surgeons

    42/46

    Devices with pacemaker function

    Sometimes devices resembling pacemakers, called ICDs (implantable cardioverter-

    defibrillator) are implanted.

    These devices are often used in the treatment of patients at risk from sudden cardiac

    death.

    An ICD has the ability to treat many types of heart rhythm disturbances by means of

    pacing, cardioversion, or defibrillation.

  • 8/2/2019 Cardiac Pacing for the Surgeons

    43/46

    Secondary prevention of SCD:VF arrest, sustained VT not secondary to reversible

    cause

    Primary prevention of SCD:

    LVEF < 36%, class II-III symptoms of CHF

    CAD, h/o MI, LVEF 40%, inducible sustained VT

    Familial or inherited conditions with high risk for SCD:HCM, long QT syndrome, Brugada syndrome

  • 8/2/2019 Cardiac Pacing for the Surgeons

    44/46

    Pacemaker programming can be performed noninvasively by an electrophysiologist or

    cardiologist. Because of the myriad of pacemaker types, patients should carry a card

    with them providing information about their particular model.

    This information is crucial when communicating with the cardiologist about a pacer

    problem.

    Most pacemaker generators, however, have an x-ray code that can be seen on a

    standard chest x-ray. The markings, along with the shape of the generator, may assist

    with deciphering the manufacturer of the generator and pacemaker battery.

    This may be helpful in the event a patient neither recalls the company nor has thepermanent pacemaker card.

  • 8/2/2019 Cardiac Pacing for the Surgeons

    45/46

    Placing a magnet over a permanent pacemaker causes sensing to be inhibited by

    closing an internal reed switch. This only temporarily "reprograms" the pacer into the

    asynchronous mode, where pacing is initiated at a set rate. It does not turn the

    pacemaker off.

    Each pacemaker type has a unique asynchronous rate for beginning-of-life (BOL),

    elective replacement indicator (ERI), and end-of-life (EOL).

    Therefore, application of a magnet can determine if the pacer's battery needs to be

    replaced.

    Further interrogation, or manipulating of the device, should be performed by an

    individual skilled in the technique.Patients should carry a card that contains information about their particular

    pacemaker, since these rates are dependent on the manufacturer and the model.

  • 8/2/2019 Cardiac Pacing for the Surgeons

    46/46

    [email protected]

    +8801711560305

    mailto:[email protected]:[email protected]