Breathings Systems

download Breathings Systems

of 67

Transcript of Breathings Systems

  • 7/30/2019 Breathings Systems

    1/67

    1

    PENDAHULUAN

    The function of breathing is to maintain a supply of oxygen to the lungs for the blood to transport

    to the tissues and to remove carbon dioxide from the body. The delivery systems which conduct

    anaesthetic gases from an anaesthetic machine to the patient are known as the breathing systems

    or circuits. A breathing circuit must enable a patient to breathe satisfactorily without significantly

    increasing the work of breathing or the physiological deadspace. It must also conduct

    inhalational anaesthetic agents to the patient. The volume of gas inspired and expired with each

    breath is the tidal volume (normally 6-10mls/kg), the total volume breathed in a minute is the

    minute volume and the volume of gas in the lungs at the end of normal expiration is the

    Functional Residual Capacity (FRC).

    The concentration of carbon dioxide in an exhaled breath varies with time; the first portion

    contains no carbon dioxide and comes from the upper respiratory tract where no gas exchange

    takes place (the anatomical dead space - 2mls/kg). The concentration of carbon dioxide then rises

    rapidly to a plateau of about 5% as alveolar gas is breathed out. The volume of alveolar gas

    expired per minute is called the alveolar minute ventilation. The anatomical dead space is 25-

    35% of each tidal volume. Any areas of lung that are ventilated with gas but are not perfused by

    blood cannot take part in gas exchange and represent the alveolar dead space. The total dead

    space in the patient is the physiological dead space.

    The term rebreathing implies that expired alveolar gas containing 5% carbon dioxide (and less

    oxygen than normal) is inspired as part of the next tidal volume. Anaesthetic circuits are

    designed to minimise this occuring as it may lead to serious elevations in blood CO2 levels. The

    amount of rebreathing that occurs with any particular anaesthetic breathing system depends on

    four factors; the design of the individual breathing circuit, the mode of ventilation (spontaneous

    or controlled), the fresh gas flow rate and the patient's respiratory pattern. Circuits may eliminate

    rebreathing either by ensuring an adequate flow of fresh gas which flushes the circuit clear ofalveolar gas, or, in the case of a circle system by the use of sodalime which absorbs the CO2 so

    that low fresh gas flows may be used. For each of the circuits described below, fresh gas flow

    rates that will ensure minimal rebreathing will be suggested.

  • 7/30/2019 Breathings Systems

    2/67

    2

    They are designed to allow either spontaneous respiration or intermittent positive pressure

    ventilation (IPPV) and consist of a reservoir bag, anaesthetic tubing, and a pressure relief valve.

    A number of mechanical ventilators include a specific breathing system eg the Manley series.

    Other ventilators have been designed to operate with existing breathing systems e.g. the Penlon

    Nuffield 200.

    DEFINITION:

    A breathing system is defined as an assembly of components which connects the patients

    airway to the anaesthetic machine creating an artificial atmosphere, from and into which the

    patient breathes.

    It primarily consists of

    a) A fresh gas entry port/delivery tube through which the gases are delivered from the machine

    to the systems;

    b) A port to connect it to the patients airway;

    c) A reservoir for gas, in the form of a bag or a corrugated tube to meet the peak inspiratory flow

    requirements;

    d) An expiratory port/valve through which the expired gas is vented to the atmosphere;

    e) A carbon dioxide absorber if total rebreathing is to be allowed and

    f) Corrugated tubes for connecting these components.

    Flow directing valves may or may not be used.

    Properties of the ideal

    breathing system

    1. Simple and safe to use.

    2. Delivers the intended inspired gas

  • 7/30/2019 Breathings Systems

    3/67

    3

    mixture.

    3. Permits spontaneous, manual and

    controlled ventilation in all age

    groups.

    4. Efficient, requiring low fresh gas

    flow rates.

    5. Protects the patient from

    barotrauma.

    6. Sturdy, compact and lightweight

    in design.

    7. Permits the easy removal of waste

    exhaled gases.

    8. Easy to maintain with minimal

    running costs.

    REQUIREMENTS OF A BREATHING SYSTEM

    The components when assembled should satisfy certain requirements, some essential and

    others desirable.

    Essential:

    The breathing system must

    a) deliver the gases from the machine to the alveoli in the same concentration as set and in the

    shortest possible time;

    b) effectively eliminate carbon-dioxide;

    c) have minimal apparatus dead space; and

    d) have low resistance.

    Desirable:

  • 7/30/2019 Breathings Systems

    4/67

    4

    The desirable requirements are

    a) economy of fresh gas;

    b) conservation of heat;

    c) adequate humidification of inspired gas;

    d) Light weight;

    e) convenience during use;

    f) efficiency during spontaneous as well as controlled ventilation (Efficiency is determined in

    terms of CO2 elimination and fresh gas utilization);

    g) adaptability for adults, children and mechanical ventilators;

    h) provision to reduce theatre pollution.

    Components of the

    breathing systemsADJUSTABLE PRESSURE LIMITING

    (APL) VALVE

    A valve which allows the exhaled

    gases and excess fresh gas flow to

    leave the breathing system (Fig. 4.2).

    It does not allow room air to enter the

    breathing system. Synonymous terms

    for the APL valve are expiratory

    valve, spill valve and relief valve.

  • 7/30/2019 Breathings Systems

    5/67

    5

    Fig. 4.1 Mapleson classification of anaesthetic breathing systems. The arrow indicates

    entry of fresh gas to the system (reproduced with permission from Aitkenhead R and

    Smith G; Textbook of Anaesthesia 3rd edition, 1996, Churchill Livingstone).

  • 7/30/2019 Breathings Systems

    6/67

    6

    Fig. 4.2 Diagram of an adjustable

    pressure-limiting (APL) valve (reproducedwith permission from Aitkenhead R and

    Smith G; Textbook of Anaesthesia 3rd

    edition, 1996, Churchill Livingstone).

    Components

  • 7/30/2019 Breathings Systems

    7/67

    7

    1. Three ports: the inlet, the patient

    and the exhaust ports. The latter

    can be open to the atmosphere or

    connected to the scavenging

    system using a shroud.

    2. A lightweight disc rests on a knifeedge

    seating. The disc is held onto

    its seating by a spring. The tension

    in the spring, and therefore the

    valves opening pressure, are

    controlled by the valve dial.

    Mechanism of action

    1. This is a one-way, adjustable,

    springloaded valve. The spring is

    used to adjust the pressure

    required to open the valve. The

    disc rests on a knife-edge seating

    in order to minimize its area of

    contact.

    2. The valve allows gases to escape

    when the pressure in the breathing

    system exceeds the valves opening

    pressure.

    3. During spontaneous ventilation,

    the patient generates a positive

    pressure in the system during

    expiration, causing the valve to

    open. A pressure of less than

    1 cmH2O (0.1 kPa) is needed to

    actuate the valve when it is in the

    open position.

  • 7/30/2019 Breathings Systems

    8/67

    8

    4. During positive pressure

    ventilation, a controlled leak is

    produced by adjusting the valve

    dial during inspiration. This

    allows control of the patients

    airway pressure.

    Problems in practice and safety

    features

    1. Malfunction of the scavenging

    system may cause excessive

    negative pressure. This can lead to

    the APL valve remaining open

    throughout respiration. This leads

    to an unwanted enormous

    increase in the breathing systems

    dead space.

    2. The patient may be exposed to

    excessive positive pressure if the

    valve is closed during assisted

    ventilation. A pressure relief safety

    mechanism actuated at a pressure

    of about 60 cmH2O is present in

    some designs (Fig. 4.3).

    3. Water vapour in exhaled gas may

    condense on the valve. The surface

    tension of the condensed water

    may cause the valve to stick. The

    disc is usually made of a

    hydrophobic (water repelling)

    material, which prevents water

    condensing on the disc.

  • 7/30/2019 Breathings Systems

    9/67

    9

    Adjustable pressure limiting valve

    (APL)

    n One-way springloaded valve

    with three ports.

    n The spring adjusts the pressure

    required to open the valve.

    RESERVOIR BAG

    The reservoir bag is an important

    component of most breathing

    systems.

    Components

    1. It is made of antistatic rubber or

    plastic. Latex free versions also

    exist. Designs tend to be

    ellipsoidal in shape.

    2. The standard adult size is 2 litres.

    The smallest size for paediatric use

    is 0.5 litre. Volumes from 0.5 to 6

    litres exist. Bigger size reservoir

    bags are useful during inhalational

    induction, e.g. adult induction

    with sevoflurane.

    Mechanism of action

    1. It accommodates the fresh gas

    flow during expiration acting as a

    reservoir available for the

    following inspiration.

    Components of the breathing systems

    41

  • 7/30/2019 Breathings Systems

    10/67

    10

    Fig. 4.3 Intersurgical APL valve. In the open position (left) the valve is actuated by pressures of

    less than 0.1 kPa (1 cmH2O) with

    minimal resistance to flow. A 3/4 clockwise turn of the dial takes the valve through a range of

    pressure-limiting positions to the closed

    position (centre). In the closed position, the breathing system pressure, and therefore theintrapulmonary pressure, is protected by a

    pressure relief mechanism (right) actuated at 6 kPa (60 cmH2O). This safety relief mechanism

    cannot be overridden.

    High pressure

    spring

    Green

    valve

    Gas flow

    Light pressure

    spring

    Scavenged gas

    Gas flow Gas flow

    Scavenged gas

    2. It acts as a monitor of the patients

    ventilatory pattern during

    spontaneous breathing. It serves

    as a very inaccurate guide to the

    patients tidal volume.

    3. It can be used to assist or control

  • 7/30/2019 Breathings Systems

    11/67

    11

    ventilation.

    4. When employed in conjunction

    with the T-piece (Mapleson F), a

    0.5 litre double-ended bag is used.

    The distal hole acts as an

    expiratory port (Fig. 4.4).

    Problems in practice and safety

    features

    1. Because of its compliance, the

    reservoir bag can accommodate

    rises in pressure in the breathing

    system better than other parts.

    When grossly overinflated, the

    rubber reservoir bag can limit the

    pressure in the breathing system to

    about 40 cmH2O. This is due to

    the law of Laplace dictating that

    the pressure (P) will fall as the

    bags radius (r) increases

    {P = 2 tension/r}.

    2. The size of the bag depends on the

    breathing system and the patient.

    A small bag may not be large

    enough to provide a sufficient

    reservoir for a large tidal volume.

    3. Too large a reservoir bag makes it

    difficult for it to act as a

    respiratory monitor.

    Reservoir bag

    n Made of rubber or plastic.

    n 2 litre size commonly used for

  • 7/30/2019 Breathings Systems

    12/67

    12

    adults. Bigger sizes can be used

    for inhalational induction in

    adults.

    n Accommodates fresh gas flow.

    n Can assist or control

    ventilation.

    n Limits pressure build-up in the

    breathing system.

    TUBING

    Specific configurations are described

    below.

  • 7/30/2019 Breathings Systems

    13/67

    13

    Fig. 4.4 A 0.5 litre double-ended

    reservior.

    Classification of breathing systems

    One will realize the reason for the failure of the attempts at classification in the 50s to

    60s, if this definition and requirements are taken into account. There are numerous

    classifications of breathing systems according to the whims and fancy of the person classifying.

    Many of them are irrelevant as they do not define a breathing system. Different authors classified

    the same system under different headings, adding to confusion1. McMohan in 1951 classified

    them as open, semiclosed and closed taking the level of rebreathing into account. It as follows:

    Open no rebreathing

    Semiclosed partial rebreathing

    Closed total rebreathing

  • 7/30/2019 Breathings Systems

    14/67

    14

    Dripps et al have classified them as Insufflation, Open, Semiopen, Semiclosed and Closed taking

    into account the presence or absence of Reservoir, Rebreathing, CO2 absorption and Directional

    valves1. The ambiguity of the terminology used as open, semi open, semi closed and closed

    allowed inclusion of apparatus that are not breathing systems at all into the classification.

    To overcome this problem Conway2

    suggested that a functional classification be used and

    classified according to the method used for CO2 elimination as:

    1. Breathing systems with CO2 absorber and

    2. Breathing systems without CO2 absorber.

    Miller.D.M.3

    in 1988 widened the scope of this classification so as to include the enclosed

    afferent reservoir system.

    A new breathing system called The Maxima4

    has been designed by Miller in 1995 and to

    include it in the classification5, the enclosed afferent reservoir systems have been grouped under

    displacement afferent reservoir systems.

    BREATHING SYSTEMS WITHOUT

    CO2 ABSORPTION.

    I BREATHING SYSTEMS WITH CO2

    ABSORPTION.

    Unidirectional flow:

    a) Non rebreathing systems.

    B) Circle systems.

    Unidirectional flow

    Circle system with absorber.

    Bi-directional flow:

    a) Afferent reservoir systems.

    Mapleson A

    Mapleson B

    Bi-directional flow

    To and Fro system.

  • 7/30/2019 Breathings Systems

    15/67

    15

    Mapleson C

    Lacks system.

    B) Enclosed afferent reservoir systems

    Millers (1988)

    c) Efferent reservoir systems

    Mapleson D

    Mapleson E

    Mapleson F

    Bains system

    d) Combined systems

    Humphrey ADE

    This classification also has a personal bias as the Humphrey ADE system is not included in the

    classification, even though he preferred to compare his system with that of Humphreys6. The

    classification suggested in table.1. is a partial modification of Millers3

    classification.

    BREATHING SYSTEMS WITHOUT CO2 ABSORPTION

    Unidirectional F low

  • 7/30/2019 Breathings Systems

    16/67

    16

    a) NONREBREATHING SYSTEMS. They use non rebreathing valves and there is no mixing

    of fresh gas and the expired gas.

    Functional analysis:When the patient takes a breath, or if the reservoir bag is squeezed, the

    inspiratory unidirectional valve opens and the gases flow into the patients lungs(Fig.1). The

    expiratory unidirectional valve closes the expiratory port during spontaneous breathing. The

    inspiratory unidirectional valve itself closes the expiratory port during controlled ventilation. At

    the start of expiration, the inspiratory unidirectional valve returns back to position and expiration

    takes place through the expiratory port, opening the expiratory valve.

    The fresh gas flow (FGF) should be equal to the minute ventilation (MV) of the patient.

    These systems satisfy all four essential requirements, but are not very popular because of the

    following reasons:

    1) Fresh gas flow has to be constantly adjusted and is not economical.

    2) There is no humidification of inspired gas.

    3) There is no conservation of heat.

    4) They are not convenient as the bulk of the valve has to be positioned near the patient.

  • 7/30/2019 Breathings Systems

    17/67

    17

    5) The valves can malfunction due to condensation of moisture and lead to complications.

    B) CIRCLE SYSTEMS: These systems are designed with a CO2 absorber as an essential

    component of the system. To use it without absorber is uneconomical as it needs a FGF more

    than the alveolar ventilation.

    The effect of arrangements of various components in the effective elimination of CO2 and

    fresh gas economy when used with high flows were analysed by Egar and Ethans7. Detailed

    discussion on this is beyond the purview of this review. However, some aspects of this is

    discussed under the section, Circle system with absorber.

    Bi-Di rectional F low:

    Systems with bi-directional flow are extensively used. These systems depend on the FGF

    for effective elimination of CO2. Understanding these systems is most important as their

    functioning can be manipulated by changing parameters like Fresh gas flow, alveolar

    ventilation, apparatus dead space, etc. We will analyze these in detail.

    Fresh Gas Supply; Fresh gas flow (FGF) forms one of the essential requirements of a

    breathing system. If there is no FGF into the system, the patient will get suffocated. If the FGF

    is low, most systems do not eliminate carbon-dioxide effectively, and if there is an excess flow

    there is wastage of gas. So, it becomes imperative to specify optimum FGF for a breathing

    system for efficient functioning.

    If the system has to deliver a set concentration in the shortest possible time to the alveoli,

    the FGF should be delivered as near the patients airway as possible.

    Elimination Of Carbon-Dioxide: The following may be taken as an example for better

    understanding of CO2 elimination by the bi-directional flow systems. Normal production of

    carbon-dioxide in a 70 kg adult is 200 ml per minute and it is eliminated through the lungs.

    Normal end-tidal concentration of carbon-dioxide is 5%. Hence, for eliminating 200 ml ofcarbon-dioxide as a 5% gas mixture, the alveolar ventilation has to be:

    200 x 100 = 4,000 ml.

    5

  • 7/30/2019 Breathings Systems

    18/67

    18

    This 4000 ml or 4 litres is the normal alveolar ventilation. Any breathing system

    connected to an adults airway should provide a minimum of 4 litres per minute of carbon-

    dioxide free gas to the alveoli for eliminating carbon-dioxide. If the alveolar ventilation becomes

    less than 4 litres per minute, it would lead to hypercarbia. If the alveoli are ventilated with 5

    litres/minute of a gas containing 1% carbon-dioxide, or 8 litres/minute of a gas containing 2.5%

    carbon-dioxide, it could still eliminate 200 ml of carbon-dioxide per minute from the alveoli. It

    may be construed as 4 litres of CO2 free gas and 1 litre of gas with 5% CO2 in the first instant

    and as 4 litres of CO2 free gas and 4 litres of gas with 5% CO2 in the second instant. In effect, 4

    litres of alveolar ventilation with CO2 free gas is provided in both cases.

    Apparatus Dead Space: It is the volume of the breathing system from the patient-end to the

    point up to which, to and fro movement of expired gas takes place.

    In an afferent reservoir system with adequate FGF, the apparatus dead space extends up to the

    expiratory valve positioned near the patient (fig.2).

    If the FG enters the system near the patient-end as in an efferent reservoir system, the

    dead space extends upto the point of FG entry. In systems where inspiratory and expiratory

    limbs are separate, it extends upto the point of bifurcation. The dynamic dead space will depend

    on the FGF and the alveolar ventilation. The dead space is minimal with optimal FGF. If the

    FGF is reduced below the optimal level, the dead space increases and the whole system will act

  • 7/30/2019 Breathings Systems

    19/67

    19

    as dead space if there is no FGF. Increasing the FGF above the optimum level will only lead to

    wastage of FG.

    Sub-Classif ication Of Bi -Di rectional F low Systems:

    Mapleson8 did a theoretical analysis of the fresh gas requirements of the semiclosed

    systems available at that time. It is only proper to refer to it as Mapleson systems as he gave a

    nomenclature as A, B, C, D and E for easy identification as per their construction. For better

    understanding of the functional analyses, they have been classified as:

    1 Afferent reservoir system (ARS).

    2 Enclosed afferent reservoir systems (EARS).

    3 Efferent reservoir systems (ERS).

    4 Combined systems.

    The afferent limb is that part of the breathing system which delivers the fresh gas from the

    machine to the patient. If the reservoir is placed in this limb as in Mapleson A, B, C and Lacks

    systems, they are called afferent reservoir systems (ARS).

    The efferent limb is that part of the breathing system which carries expired gas from the patient

    and vents it to the atmosphere through the expiratory valve/port. If the reservoir is placed in this

    limb as in Mapleson D, E, F and Bain systems, they are called efferent reservoir systems (ERS).

    Enclosed afferent reservoir system has been described by Miller and Miller.

    AFFERENT RESERVOIR (AR) SYSTEMS

    The Mapleson A, B and C systems have the reservoir in the afferent limb, and do not have an

    efferent limb (Fig.3). Lack system has an afferent limb reservoir and an efferent limb through

    which the expired gas traverses before being vented into the atmosphere (Fig.4). This limb is

    coaxially placed inside the afferent limb.

  • 7/30/2019 Breathings Systems

    20/67

    20

    These AR systems work efficiently during spontaneous breathing provided the expiratory valve is

    separated from the reservoir bag and FGF by at least one tidal volume of the patient and

    apparatus dead space is minimal. They do not function efficiently during controlled ventilation.

    If the FGF is close to the expiratory valve as in Mapleson B & C, the system is inefficient both

    during spontaneous and controlled ventilation. The efficiency is determined in terms of carbon-

    dioxide elimination and FGF utilization.

    Mapleson8

    has analysed these bi-directional flow systems using mathematical calculations. He

    made a few basic assumptions while analyzing breathing systems. These are

    (1) Gases move enbloc. They maintain their identity as fresh gas, dead space gas and

    alveolar gas. There is no mixing of these gases.

    (2) The reservoir bag continues to fill up, without offering any resistance till it is full.

    (3) The expiratory valve opens as soon as the reservoir bag is full and the pressure inside

    the system goes above atmospheric pressure.

    (4) The valve remains open throughout the expiratory phase without offering any

    resistance to gas flow and closes at the start of the next inspiration.

    Mapleson A/Magills system:

    Functional analysis:

  • 7/30/2019 Breathings Systems

    21/67

    21

    Spontaneous breathing: The system is filled with fresh gas before connecting to the

    patient. When the patient inspires, the fresh gas from the machine and the reservoir bag flows to

    the patient, and as a result the reservoir bag collapses (Fig.5a). During expiration, the FG

    continues to flow into the system and fill the reservoir bag. The expired gas, initial part of which

    is the dead space gas, pushes the FG from the corrugated tube into the reservoir bag and collects

    inside the corrugated tube (Fig.5b).

    As soon as the reservoir bag is full, the expiratory valve opens and the alveolar gas is vented into

    the atmosphere (Fig.5c). During the expiratory pause, alveolar gas that had come into the

    corrugated tube is also pushed out through the valve, depending on the FGF. The system is filled

    with only fresh gas and dead space gas at the start of the next inspiration when FGF is equal to

    the alveolar ventilation (Fig.5d). The entire alveolar gas and dead space gas is vented through the

    valve and some FG also escapes, if the FGF is higher than the minute ventilation. Some amount

    of alveolar gas will remain in the system and lead to rebreathing with a FGF less than the

    alveolar ventilation. This has been confirmed theoretically and experimentally by many

    investigators8,9

    . The system functions at maximum efficiency, when the FGF equals the alveolar

    ventilation and the dead space gas (which has not taken part in gas exchange) is allowed to be

    rebreathed and utilized for alveolar ventilation.

  • 7/30/2019 Breathings Systems

    22/67

    22

    Controlled ventilation: To facilitate IPPV the expiratory valve has to be partly closed. During

    inspiration, the patient gets ventilated with FG and part of the FG is vented through the valve

    (Fig.6a) after sufficient pressure has developed to open the valve. During expiration, the FG

    from the machine flows into the reservoir bag and all the expired gas (i.e., dead space gas and

    alveolar gas) flows back into the corrugated tube till the system is full (Fig.6b). During the next

    inspiration the alveolar gas is pushed back into the alveoli followed by the FG. When sufficient

    pressure is developed, part of the expired gas and part of the FG escape through the valve

    (Fig.6c). This leads to considerable rebreathing, as well as excessive waste of fresh gas. Hence

    these systems are inefficient for controlled ventilation.

    Lacks system:

    This system functions like a Mapleson A system both during spontaneous and controlled

    ventilation. The only difference is that the expired gas instead of getting vented through the

    valve near the patient, is carried by an efferent tube placed coaxially and vented through the

    valve placed near the machine end (Fig.4). This facilitates easy scavenging of expired gas.

  • 7/30/2019 Breathings Systems

    23/67

    23

    Mapleson B & C systems:

    In order to reduce the rebreathing of alveolar gas and to improve the utilization of FGduring controlled ventilation, the FG entry was shifted near the patient(Fig.3). This allows a

    complete mixing of FG and expired gas. The end result is that these systems are neither efficient

    during spontaneous nor during controlled ventilation.

    ENCLOSED AFFERENT RESERVOIR (EAR) SYSTEMS

    This has been described by Miller & Miller10

    . The system consisted of a Mapleson A

    system enclosed within a non distensible structure (Fig.7a). It may also be constructed by

    enclosing the reservoir bag alone in a bottle and connecting the expiratory port to the bottle with

    a corrugated tube and a one way valve (Fig.7b). To the bottle is also attached a reservoir bag and

    a variable orifice for providing positive pressure ventilation.

  • 7/30/2019 Breathings Systems

    24/67

    24

    Functional analysis:During spontaneous ventilation, the gas is vented from the system

    in a manner which is identical to the Mapleson A system. In this mode the variable orifice is

    kept widely open to allow free communication to the atmosphere. In controlled ventilation the

    reservoir bag B is squeezed intermittently and the variable orifice is partly closed to allow

    building up of pressure in the bottle. The pressure thus developed (1) closes the expiratory

    valve, (2) squeezes the enclosed afferent reservoir and the patient gets ventilated. The expiration

    takes place in a manner similar to that described during spontaneous ventilation when the

    pressure is released in reservoir B,. Hence this system should function efficiently during

    spontaneous and controlled ventilation with a FGF equivalent to alveolar ventilation. The fresh

    gas requirement and the utilization of this system has been investigated by a group of

    investigators from Manchester11-13

    and a group from Wales14

    under the guidance of Mapleson.

    They have reported varying figures for utilization as 82%, 93% and 74% respectively11,12,14

    . The

    reasons for this lesser percentage of utilization have been quoted as faulty methodology for

    calculation15

    , resistance offered by the reservoir bag and tubing and early opening of the

    unidirectional valve during expiration14

    etc. Though the fresh gas requirement is higher than the

    alveolar ventilation in this system as shown by the above studies, it is still more efficient than the

    Bain system for controlled ventilation.

  • 7/30/2019 Breathings Systems

    25/67

    25

    EFFERENT RESERVOIR (ER) SYSTEMS:

    Bain system

    The Mapleson D, E, F and Bain systems have a 6 mm tube as the afferent limb that supplies the

    FG from the machine. The efferent limb is a wide-bore corrugated tube to which the reservoir

    bag is attached and the expiratory valve is positioned near the bag. In Mapleson E system, the

    corrugated tube itself acts as the reservoir (Fig.8). In Bain system, the afferent and efferent

    limbs are coaxially placed (Fig.9).

    All these ER systems are modifications of Ayres T-piece. This consists of a light metal tube 1

    cm in diameter, 5 cm in length with a side arm (Fig.10). Used as such, it functions as a non-

    rebreathing system. Fresh gas enters the system through the side arm and the expired gas is

    vented into the atmosphere and there is no rebreathing. The dead space is minimal as it is only

    up to the point of FG entry and elimination of CO2 is achieved by breathing into the atmosphere.

    FGF equal to peak inspiratory flow rate of the patient has to be used to prevent air dilution.

  • 7/30/2019 Breathings Systems

    26/67

    26

    In an attempt to reduce FGF requirements, ER systems are constructed with reservoirs in the

    efferent limb. The functioning of all these systems are similar. These systems work efficiently

    and economically for controlled ventilation as long as the FG entry and the expiratory valve are

    separated by a volume equivalent to atleast one tidal volume of the patient. They are not

    economical during spontaneous breathing.

    Spontaneous respiration: The breathing system should be filled with FG before connecting to

    the patient. When the patient takes an inspiration, the FG from the machine, the reservoir bag

    and the corrugated tube flow to the patient (Fig.11a). During expiration, there is a continuous

    FGF into the system at the patient end. The expired gas gets continuously mixed with the FG asit flows back into the corrugated tube and the reservoir bag (Fig.11b). Once the system is full the

    excess gas is vented to the atmosphere through the valve situated at the end of the corrugated

    tube near the reservoir bag. During the expiratory pause the FG continues to flow and fill the

    proximal portion of the corrugated tube while the mixed gas is vented through the valve

    (Fig.11c). During the next inspiration, the patient breaths FG as well as the mixed gas from the

    corrugated tube (Fig.11d). Many factors influence the composition of the inspired mixture.

    They are FGF, respiratory rate, expiratory pause, tidal volume and CO2 production in the body.

    Factors other than FGF cannot be manipulated in a spontaneously breathing patient. It has been

    mathematically calculated and clinically proved8,16

    that the FGF should be atleast 1.5 to 2 times

    the patients minute ventilation in order to minimise rebreathing to acceptable levels.

  • 7/30/2019 Breathings Systems

    27/67

    27

    Controlled ventilation: To facilitate intermittent positive pressure ventilation, the expiratory

    valve has to be partly closed so that it opens only after sufficient pressure has developed in the

    system. When the system is filled with fresh gas, the patient gets ventilated with the FGF from

    the machine, the corrugated tube and the reservoir bag (Fig.12a). During expiration, the expired

    gas continuously gets mixed with the fresh gas that is flowing into the system at the patient end.

    During the expiratory pause the FG continues to enter the system and pushes the mixed gas

    towards the reservoir (12B). When the next inspiration is initiated, the patient gets ventilated

    with the gas in the corrugated tube i.e., a mixture of FG, alveolar gas and dead space gas

    (Fig.12c). As the pressure in the system increases, the expiratory valve opens and the contents of

    the reservoir bag are discharged into the atmosphere.

  • 7/30/2019 Breathings Systems

    28/67

    28

    Factors that influence the composition of gas mixture in the corrugated tube with which the

    patient gets ventilated are the same as for spontaneous respiration namely FGF, respiratory rate,

    tidal volume and pattern of ventilation. The only difference is that these parameters can be

    totally controlled by the anaesthesiologist and do not depend on the patient. Using a low

    respiratory rate with a long expiratory pause and a high tidal volume, most of the FG could be

    utilized for alveolar ventilation without wastage.

    Analyzing the performance of these systems during controlled ventilation, two relationships have

    become evident. 1) When FGF is very high the PaCO2 becomes ventilation dependent (as during

    spontaneous respiration). 2) When the minute volume exceeds the FGF substantially, the PaCO2

    is dependent on the FGF17

    . Combining these influences a graph can be constructed as shown in

    Fig.13. An infinite number of combinations of FGF and minute ventilation can be chosen to

    achieve a desired PaCO2. One can use a high FGF and a normal minute volume of 70 ml/kg to

    achieve a normal PaCO2 of 40 mm Hg. This is uneconomical and leads to low humidity and heat

    loss. Alternately, a FGF equivalent to the predicted minute volume i.e., 70 ml/kg can be chosen

    and the patient ventilated with at least twice the predicted minute volume i.e. 140 ml/kg. Here a

    deliberate controlled rebreathing is allowed in order to maintain normal PaCO2 along with high

    humidity, less heat loss and greater economy of fresh gas. Combinations between these two

  • 7/30/2019 Breathings Systems

    29/67

    29

    extremes can also be used. It is important to remember that using a low FGF with normal minute

    ventilation, can lead to hypercarbia; a moderate FGF and hyperventilation, can lead to

    hypocarbia.

    COMBINED SYSTEMS

    To over come the difficulties of changing the breathing systems for different modes of

    ventilation, Humphrey designed a system called Humphrey ADE18

    , with two reservoirs, one in

    the afferent limb and the other in the efferent limb. While in use, only one reservoir will be in

    operation and the system can be changed from ARS to ERS by changing the position of a lever.

    It can be used for adults as well as children. The functional analysis is the same as Mapleson A

    in ARS mode and as Bain in ERS mode. It is not yet widely used.

    BREATHING SYSTEMS WITH CO2 ABSORPTION

    Systems so far described have relied on FGF for effective elimination of CO2. Any desire to

    economize on FGF by allowing a total rebreathing, should be accompanied by removal of the

    expired CO2 by chemical absorption using sodalime or baralyme. The systems designed for

    these purpose are again classified as:

    Unidirectional flow.

    -Circle system.

    Bi-directional flow.

    -To and fro system.

    The essential components of the circle system are, (1) a sodalime canister, (2) Two unidirectional

    valves, (3) Fresh gas entry, (4) Y-piece to connect to the patient, (5) Reservoir bag (6) a relief

    valve and (7) low resistance interconnecting tubing. The arrangement of the components is

    shown in fig.14. For efficient

  • 7/30/2019 Breathings Systems

    30/67

    30

    functioning of the system the following criteria should fulfilled. (1) There should be two

    unidirectional valves on either side of the reservoir bag, (2) Relief valve should be positioned in

    the expiratory limb only, (3) The FGF should enter the system proximal to the inspiratory

    unidirectional valve.

    Functional analysis: During inspiration the FG along with the CO2 free gas in the reservoir bag

    flow through the inspiratory limb and inspiratory unidirectional valve to the patient. No flow

    takes place in the expiratory limb as the expiratory unidirectional valve is closed by back

    pressure transmitted to the valve. During expiration the inspiratory unidirectional valve closes

    and the expired gas flows through the expiratory unidirectional valve in the expiratory limb to

    the sodalime canister and to the reservoir bag. The CO2 is absorbed in the canister. The FGF

    from the machine continues to fill the reservoir bag. When the reservoir is full the relief valve

    opens and the excess gas is vented to atmosphere. By selecting a suitable position for the relief

    valve, the expired gas can be selectively vented when the FGF is more than the alveolar

    ventilation. To facilitate controlled ventilation the relief valve has to be partly closed and the

    excess gas is vented during inspiration. The gas flow pattern is similar to that described above.

    The advantages and disadvantages of the various arrangements of the components were

    analyzed by Eger and Ethans7. The relative positions of the components of the circle system are

    of particular importance to the functioning of the system only when the FGF is high, the gas

    components of the system unmixed and CO2 absorber not used. When the FGF is reduced below

  • 7/30/2019 Breathings Systems

    31/67

    31

    the alveolar ventilation, the CO2 absorber is a must as the gas in the system become more

    uniformly mixed, and the relative position of the systems components become less important

    Total ly closed system:

    The systems with CO2 absorption can be used in a completely closed mode. After a

    period of approximately 10-20 minutes breathing with high inflow of fresh gas for

    denitrogenation, the expiratory valve is closed. The FGF is then adjusted to meet only the

    patients basal oxygen requirements together with anaesthetic. A number of advantages have

    been demonstrated for totally closed systems.

    A) Economy: The FGF could be reduced to as low as 250 - 500 ml of oxygen. The consumption

    of Halothane/Isoflurane has been found to be around 3.5 ml/hour19

    .

    b) Humidification: In the completely closed system, once the equilibrium has been established,

    the inspired gas will be fully saturated with water vapour20

    .

    C) Reduction of heat loss: In addition to conserving water the totally closed system will also

    conserve heat. The CO2 absorption is an exothermic reaction and the system may actively

    assist in maintaining body temperature.

    D) Reduction in atmospheric pollution: Once the expiratory valve has been closed, no

    anaesthetic escapes, except for the small percutaneous loss from the patient.

    E) Control of anaesthesia: It is possible to compute the time course of uptake of anaesthetic in a

    patient of known size and add the appropriate quantity of the anaesthetic to the circuit at a

    rate decreasing in a manner calculated to maintain a constant alveolar concentration21

    . In

    practice an alveolar concentration of about 1.3 x MAC is found to be suitable.

    The technique has several potential disadvantages.

    i) A greater knowledge of uptake and distribution is required to master closed circuit anaesthesia.

    ii) Inability to alter any concentration quickly.

  • 7/30/2019 Breathings Systems

    32/67

    32

    iii) Real danger of hypercapnia may result from, a) an inactive absorber, B) incompetent

    unidirectional valves and c) incorrect use of absorber bypass.

    BI -DIRECTIONAL FLOW SYSTEMS

    The Waters to and fro system is valveless and conveniently portable. It has been widely used in

    the past and now is only of historical importance. The reader may refer to any standard text

    book for furtherdetails.

    A number of classifications exist and the one introduced in 1954 by Professor W W Mapleson is

    most commonly used in the UK (Figure 1). It does not however, include systems with carbon

    dioxide absorption.

    The Mapleson A (Magill) system was designed by Sir Ivan Magill in the 1930's and remains an

    excellent system for spontaneous ventilation (Figure 2). Fresh gas enters the system at the fresh

    gas outlet of the anaesthesia machine. The expiratory valve (Heidbrink valve) is very close to the

    patient to reduce the dead space.

    http://www.nda.ox.ac.uk/wfsa/html/u07/u07b_p01.htm
  • 7/30/2019 Breathings Systems

    33/67

    33

    The respiratory cycle has three phases during spontaneous breathing; inspiration, expiration and

    the expiratory pause. During inspiration gas is inhaled from the 2 litre reservoir (breathing) bag

    which partially collapses giving a visual confirmation that breathing is occurring.

    This breathing system is popular and

    widely used in the UK.

    Components

    1. Corrugated rubber or plastic

    tubing (usually 110180 cm in

    length).

    2. A reservoir bag mounted at the

    machine end.

    3. An APL valve situated at the

    patient end.

    Mechanism of action

    1. As the patient exhales (Fig. 4.5C),

    initially the gases from the

    anatomical dead space are

    channelled through the tubing

    back towards the reservoir bag

    which is filled continuously with

    fresh gas flow.

    http://www.nda.ox.ac.uk/wfsa/html/u07/u07b_p02.htm
  • 7/30/2019 Breathings Systems

    34/67

    34

    2. Pressure builds up opening the

    APL valve and expelling the

    alveolar gases first (Fig. 4.5D). By

    that time the patient inspires again

    (Fig. 4.5B) getting a mixture of

    fresh gas flow and the rebreathed

    anatomical dead space gases.

    3. It is a very efficient system for

    spontaneous breathing. Because

    there is no gas exchange in the

    anatomical dead space, the fresh

    gas flow requirements to prevent

    rebreathing of alveolar gases is

    theoretically equal to the patients

    alveolar minute volume (about 70

    ml/kg/min).

    4. The Magill system is not an

    efficient system for controlled

    ventilation. A fresh gas flow rate

    of three times the alveolar minute

    volume is required to prevent

    rebreathing.

    Problems in practice and safety

    features

    1. It is not suitable for use with

    children of less than 25 to 30 kg

    body weight. This is because of

    the increased dead space caused

    by the systems geometry at the

    patient end. Dead space is further

    increased by the angle piece and

  • 7/30/2019 Breathings Systems

    35/67

    35

    face mask.

    2. One of its disadvantages is the

    heaviness of the APL valve at the

    patients end, especially if

    connected to a scavenging system.

    This places a lot of drag on the

    connections at the patient end.

  • 7/30/2019 Breathings Systems

    36/67

    36

    Fig. 4.5 Mechanism of action of the Magill breathing system during spontaneous

    ventilation; see text for details (FGF = fresh gas flow) (reproduced with permission from

    Aitkenhead R and Smith G; Textbook of Anaesthesia 3rd edition, 1996, Churchill

    Livingstone)

    Fig. 4.6 (A) The coaxial Lack breathing system. (B) The parallel Lack breathing system.

    During expiration the bag and tubing are initially refilled with a combination of exhaled dead

    space gas (containing no carbon dioxide) and fresh gas flowing from the anaesthetic machine.

    Once the bag is full the pressure within the breathing system rises and the expiratory valve near

    the patient opens allowing the alveolar gas (containing carbon dioxide) to be vented from the

    system. During the expiratory pause more fresh gas enters the system driving any remaining

    alveolar gas back along the corrugated tubing and out through the valve. If the fresh gas flow is

    sufficiently high all the alveolar gas is vented from the circuit before the next inspiration and no

    rebreathing will take place. With careful adjustment the fresh gas flow can be reduced until there

  • 7/30/2019 Breathings Systems

    37/67

    37

    is only fresh gas and dead space gas in the breathing system at the start of inspiration. When the

    system is functioning correctly, without any leaks, a fresh gas flow (FGF) equal to the patients

    alveolar minute ventilation is sufficient to prevent rebreathing. In practice however, a FGF closer

    to the patients total minute ventilation (including dead space) is usually selected to provide a

    margin of safety. An adult's minute volume is approximately 80mls/kg /min and thus for a 75kg

    man a FGF of 6 litres per minute will prevent rebreathing. This is an efficient system for

    spontaneously breathing patients if carbon dioxide absorption is not available.

    During controlled ventilation the Magill circuit works in a different way and becomes wasteful

    and inefficient, requiring high fresh gas flows to prevent rebreathing. The inspiratory force is

    provided by the anaesthetist squeezing the reservoir bag after partly closing the expiratory valve

    next to the patient. During lung inflation some of the gas is vented from the circuit and at the endof inspiration the reservoir bag is less than half full. During expiration, dead space and alveolar

    gas pass down the corrugated tubing and may reach the bag which will then contain some carbon

    dioxide. During the next inspiration when the bag is compressed alveolar gas re-enters the

    patients lungs followed by a mixture of fresh, dead space and alveolar gas. A FGF of two and a

    half times the patient's minute volume is required to vent enough alveolar gas to minimise

    rebreathing (FGF of about 12-15 litres /min) which is obviously very inefficient. In practice the

    Magill circuit should not be used for positive pressure ventilation except for short periods of a

    few minutes at a time.

    Modifications of the Mapleson A system

    A simple modification of the Mapleson A circuit is required to make it more efficient for

    controlled ventilation. This is achieved by substituting a non-rebreathing valve (such as an Ambu

    E valve) for the Heidbrink valve at the patient end of the circuit. Not only does this arrangement

    prevent rebreathing, but during manual ventilation the delivered minute volume will be the same

    as the desired FGF which should be set at the rotameters. It is, however, a dangerous

    arrangement for spontaneous respiration because the valve may jam if the fresh gas flow is

    greater that the patient's minute volume.

    The Lack circuit

  • 7/30/2019 Breathings Systems

    38/67

    38

    A disadvantage of the Magill system is that the expiratory valve is attached close to the patient

    making it awkward to use (particularly when a scavenging circuit is added). The Lack circuit

    (Figure 3) is a Mapleson A system in which the exhaled gases travel down a central tube located

    within an outer corrugated tube towards the expiratory valve (co-axial system).

    The inner tubing is wide enough to prevent an increase in the work of breathing and the

    expiratory valve is placed next to the reservoir bag, by the common gas outlet. The fresh gas

    flows required for both spontaneous and controlled ventilation are as described for the standard

    Mapleson A system.

    Components

    1. 1.8 m length coaxial tubing (tube

    inside a tube). The fresh gas flows

    through the outside tube, and the

    exhaled gases flow through the

    inside tube (Fig. 4.6A).

    2. The inside tube is wide in

    diameter(14 mm) to reduceresistance to expiration. The outer

    tubes diameter is 30 mm.

    3. The reservoir bag is mounted at

    the machine end.

    4. The APL valve is mounted at the

    http://www.nda.ox.ac.uk/wfsa/html/u07/u07_012.htmhttp://www.nda.ox.ac.uk/wfsa/html/u07/u07b_p03.htmhttp://www.nda.ox.ac.uk/wfsa/html/u07/u07_012.htmhttp://www.nda.ox.ac.uk/wfsa/html/u07/u07b_p03.htm
  • 7/30/2019 Breathings Systems

    39/67

    39

    machine end eliminating the drag

    on the connections at the patient

    end, which is a problem with the

    Magill system.

    Mechanism of action

    1. The mechanism is similar to that

    of the Magill system except the

    Lack system is a coaxial version.

    The fresh gas flows through the

    outside tube whereas the exhaled

    gases flow through the inside tube.

    2. A fresh gas flow rate of about 70

    ml/kg/min is required in order to

    prevent rebreathing. This makes it

    an efficient breathing system for

    spontaneous ventilation.

    3. Since it is based on the Magill

    system, it is not suitable for

    controlled ventilation.

    4. Instead of the coaxial design, a

    parallel tubing version of the

    system exists (Fig. 4.6B). This has

    separate inspiratory and

    expiratory tubing and retains the

    same flow characteristics as the

    coaxial version.

  • 7/30/2019 Breathings Systems

    40/67

    40

    Fig. 4.7 The Bain breathing system (reproduced with permission from Aitkenhead R

    and Smith G; Textbook of Anaesthesia 3rd edition, 1996, Churchill Livingstone).

    The Mapleson B and C breathing systems (Figure 1) are similar in construction, with the fresh

    gas flow entry and the expiratory valves located at the patient end of the circuit. They are not

    commonly used in anaesthetic practice, although the C system is used on intensive care units.

    High flows of gases are needed to prevent rebreathing of CO2 and this system was at one time

    combined with a canister of sodalime to absorb CO2 (Waters' "To and Fro" Circuit). However the

    cannister proved too bulky for practical use and there was a risk of the patient inhaling soda lime

    dust.

    Components

    1. A reservoir bag. In the B system

    corrugated tubing is attached to

    the bag, and this also acts as a

    reservoir.

    2. An APL valve at the patients end.

    3. Fresh gas flow is added just

    proximal to the APL.

    Mechanism of action

    1. Both systems are not efficient

    http://www.nda.ox.ac.uk/wfsa/html/u07/u07b_p01.htmhttp://www.nda.ox.ac.uk/wfsa/html/u07/u07_012.htmhttp://www.nda.ox.ac.uk/wfsa/html/u07/u07_012.htmhttp://www.nda.ox.ac.uk/wfsa/html/u07/u07b_p01.htm
  • 7/30/2019 Breathings Systems

    41/67

    41

    during spontaneous ventilation. A

    fresh gas flow of 1.5 to 2 times the

    minute volume is required to

    prevent rebreathing.

    2. During controlled ventilation, the

    B system is more efficient because

    of the corrugated tubing acting as

    a reservoir. A fresh gas flow of

    more than 50% of the minute

    ventilation is still required to

    prevent rebreathing.

    Bain system

    (Mapleson D)

    The Bain system is a coaxial version

    of the Mapleson D System (Fig. 4.7).

    It is lightweight and compact at the

    patient end. It is useful where access

    to the patient is limited, such as

    during head and neck surgery.

    A Manley ventilator which has

    been switched to spontaneous

    ventilation mode is an example of a

    non-coaxial Mapleson D system.

    Components

    1. A length of coaxial tubing (tube

    inside a tube). The usual length is

    180 cm but it can be supplied at

    270 cm (for dental or ophthalmic

    surgery) and 540 cm (for MRI

    scans where the anaesthetic

    machine needs to be kept outside

  • 7/30/2019 Breathings Systems

    42/67

    42

    the scanners magnetic field).

    Increasing the length of the tubing

    does not affect the physical

    properties of the breathing system.

    2. The fresh gas flows through the

    inner tube while the exhaled gases

    flow through the outside tube. The

    internal lumen has a swivel mount

    at the patients end. This ensures

    that the internal tube cannot kink,

    so ensuring delivery of fresh gas to

    the patient.

    3. The reservoir bag is mounted at

    the machine end.

    4. The APL valve is mounted at the

    machine end.

    Mechanism of action

    1. During spontaneous ventilation,

    the patients exhaled gases are

    channelled back to the reservoir

    bag and become mixed with fresh

    gas (Fig. 4.8B). Pressure build-up

    within the system will open the

    APL valve allowing the venting of

    the mixture of the exhaled gases

    and fresh gas (Fig. 4.8C).

    2. The fresh gas flow required to

    prevent rebreathing (as seen in

    Fig. 4.8D) during spontaneous

    ventilation is about 1.5 to 2 times

    the alveolar minute volume. A

  • 7/30/2019 Breathings Systems

    43/67

    43

    flow rate of 150 to 200 ml/kg/min

    is required. This makes it an

    inefficient and uneconomical

    system for use during spontaneous

    ventilation.

    3. It is a more efficient system for

    controlled ventilation. A flow of

    70 to 100 ml/kg/min will maintain

    normocapnia. A flow of 100

    ml/kg/min will cause moderate

    hypocapnia during controlled

    ventilation.

    4. Connection to a ventilator is

    possible (Fig. 4.9). By removing

    the reservoir bag, a ventilator such

    as the Penlon Nuffield 200 can be

    connected to the bag mount using

    a 1 metre length of corrugated

    tubing (the volume of tubing must

    exceed 500 ml if the driving gas

    from the ventilator is not to enter

    the breathing system). The APL

    valve must be fully closed. A

    newly designed modification

    incorporates a switch knob at the

    valve end to switch between

    spontaneous and controlled

    ventilation. This allows the

    ventilator to be connected to the

    breathing system all the time.

    5. A parallel version of the D system

  • 7/30/2019 Breathings Systems

    44/67

    44

    is also available.

    Problems in practice and safety

    features

    1. The internal tube can kink

    preventing fresh gas from being

    delivered to the patient.

    2. The internal tube can become

    disconnected at the machine end

    causing a large increase in the

    dead space, resulting in

    hypoxaemia and hypercapnia.

    Movement of the reservoir bag

    during spontaneous ventilation is

    not therefore an indication that

    the fresh gas is being delivered to

    the patient.

    Fig. 4.8 Mechanism of action of the Mapleson D breathing system during

    spontaneous ventilation (reproduced with permission from Textbook of Anaesthesia 2nd

  • 7/30/2019 Breathings Systems

    45/67

    45

    edn, Aitkenhead and Smith 1990, Churchill Livingstone).

    Fig. 4.9 The Bain breathing system connected to a ventilator (e.g. Penlon Nuffield

    200) via tubing connected to the bag mount (reproduced with permission from Textbook

    of Anaesthesia 2nd edn, Aitkenhead and Smith 1990, Churchill Livingstone).

    T-piece system

    (Mapleson E and F)

    This is a valveless breathing system

    used in anaesthesia for children up to

    25 to 30 kg body weight (Fig. 4.10).

    It is suitable for both spontaneous

    and controlled ventilation.Components

    1. A T-shaped tubing with three open

    ports (Fig. 4.11).

    2. Fresh gas from the anaesthetic

    machine is delivered via a tube to

  • 7/30/2019 Breathings Systems

    46/67

    46

    one port.

    3. The second port leads to the

    patients mask or tracheal tube.

    The connection should be as short

    as possible to reduce dead space.

    4. The third port leads to reservoir

    tubing. Jackson-Rees added a

    double-ended bag to the end of the

    reservoir tubing (making it

    Mapleson F).

    5. A recent modification exists where

    an APL valve is included before a

    closed-ended 500 ml reservoir

    bag. A pressure relief safety

    mechanism in the APL valve is

    actuated at a pressure of

    30 cmH2O (Fig. 4.12). This design

    allows effective scavenging.

    Mechanism of action

    1. The system requires a fresh gas

    flow of 2.5 to 3 times the minute

    volume to prevent rebreathing

    with a minimal flow of 4 litre/min.

    2. The double-ended bag acts as a

    visual monitor during

    spontaneous ventilation. In

    addition, the bag can be used for

    assisted or controlled ventilation.

    3. The bag can provide a degree of

    CPAP during spontaneous

    ventilation.

  • 7/30/2019 Breathings Systems

    47/67

    47

    4. Controlled ventilation is

    performed either by manual

    squeezing of the double-ended bag

    (intermittent occlusion of the

    reservoir tubing in the Mapleson

    E) or by removing the bag and

    connecting the reservoir tubing to

    a ventilator such as the Penlon

    Nuffield 200.

    5. The volume of the reservoir tubing

    determines the degree of

    rebreathing (too large a tube) or

    entrainment of ambient air (too

    small a tube). The volume of the

    reservoir tubing should

    approximate to the patients tidal

    volume.

    Problems in practice and safety

    features

    1. Since there is no APL valve used in

    the standard breathing system,

    scavenging can be a problem.

    2. Patients under 6 years of age have

    a low functional residual capacity

    (FRC). Mapleson E was designed

    before the advantages of CPAP

    were recognised for increasing the

    FRC. This problem can be

    partially overcome in the

    Mapleson F with the addition of

    the double-ended bag.

  • 7/30/2019 Breathings Systems

    48/67

    48

    Fig. 4.10 A T-piece breathing system.

    Fig. 4.11 Mechanism of action of the Tpiece

  • 7/30/2019 Breathings Systems

    49/67

    49

    breathing system.

    Fig. 4.12 Intersurgical T-piece incorporating an APL valve and closed reservoir bag to

    enable effective scavenging.

    The Mapleson D, E and F systems are all functionally similar(Figure 1).

    They act as T pieces with the FGF delivered to the patient end of the circuit and differ only in the

    presence of valves or breathing bags at the expiratory end of the circuit. These systems are all

    inefficient for spontaneous respiration (Figure 4).

    http://www.nda.ox.ac.uk/wfsa/html/u07/u07b_p01.htmhttp://www.nda.ox.ac.uk/wfsa/html/u07/u07b_p01.htm
  • 7/30/2019 Breathings Systems

    50/67

    50

    During expiration exhaled gas and fresh gas mix in the corrugated tubing and travel towards the

    reservoir bag. When the bag is full the pressure in the system rises and the expiratory valve

    opens venting to the atmosphere a mixture of fresh and exhaled gas. During the expiratory pause

    fresh gas continues to push exhaled alveolar gas down the tubing towards the valve. However,

    unless the FGF is at least twice the patient's minute volume, rebreathing of alveolar gas will

    occur. A FGF of at least 8-10 litres/min (150mls/kg/min) is required to prevent rebreathing in an

    adult.

    When used for controlled ventilation the Mapleson D system functions more efficiently. During

    expiration the corrugated tubing and reservoir bag fill with a mixture of fresh and exhaled gas.

    Fresh gas fills the distal part of the corrugated tube during the expiratory pause prior to

    inspiration. When the bag is compressed this fresh gas enters the lungs and when the expiratory

    valve opens a mixture of fresh and exhaled gas is vented. The degree of rebreathing that occurs

    depends on the FGF. A FGF of 70ml/kg/min is usually adequate for controlled ventilation;

    100mls/kg/min will result in a degree of hypocapnia (lowered CO2 level in the blood).

    Modifications of the Mapleson D system

    The Bain Circuit (Figure 3) is the most commonly used form of the Mapleson D system. It is a

    co -axial circuit which was introduced in 1972 by Bain and Spoerel. Unlike the Lack co-axial

    circuit described above, fresh gas flows down the central narrow bore tubing (7mm i.d.) to the

    patient and exhaled gases travel in the outer corrugated tubing (22mm i.d.). The reservoir bag

    may be removed and replaced by a ventilator such as the Nuffield Penlon 200 for mechanical

    ventilation. Before use the Bain circuit should be carefully checked by the anaesthetist. The outer

    http://www.nda.ox.ac.uk/wfsa/html/u07/u07b_p03.htmhttp://www.nda.ox.ac.uk/wfsa/html/u07/u07_012.htmhttp://www.nda.ox.ac.uk/wfsa/html/u07/u07b_p04.htmhttp://www.nda.ox.ac.uk/wfsa/html/u07/u07_012.htmhttp://www.nda.ox.ac.uk/wfsa/html/u07/u07b_p04.htmhttp://www.nda.ox.ac.uk/wfsa/html/u07/u07b_p03.htm
  • 7/30/2019 Breathings Systems

    51/67

    51

    tubing of a Bain circuit is made of clear plastic and the inner green or black. If a leak develops in

    the inner tubing or it becomes detached from the fresh gas port, a huge increase in apparatus

    dead space occurs. In order to check for this, the lumen of the green tubing should be occluded

    with a finger or the plunger of a 2ml syringe when a rise in gas pressure within the anaesthetic

    circuit should be observed.

    The degree of rebreathing that occurs suring IPPVwill depend on the FGF. In an adult, fresh gas

    flows of 70-80mls/kg/min (6-7litres/min) will maintain a normal arterial carbon dioxide tension

    (normocapnia) and a flow of 100mls/kg/min will result in mild hypocapnia.

    The Mapleson E system performs in a similar way to the Mapleson D, but because there are no

    valves and there is very little resistance to breathing it has proved very suitable for use with

    children. It was originally introduced in 1937 by P Ayre and is known as the Ayre's T-piece. The

    version most commonly used is the Jackson-Rees modification which has an open bag attached

    to the expiratory limb (classified as a Mapleson F system although it was not included in the

    original description by Professor Mapleson). Movement of the bag can be seen during

    spontaneous breathing, and the bag can be compressed to provide manual ventilation. As in the

    Bain circuit, the bag may be replaced by a mechanical ventilator designed for use with children.

    This system is suitable for children under 20kg. Fresh gas flows of 2 - 3 times minute volume

    should be used to prevent rebreathing during spontaneous ventilation, with a minimum flow of 3

    litres/minute, eg a 4 year old child weighing 20kg has a normal minute volume of 3 litres/min

    and would required a FGF of 6-9litres/min. During controlled ventilation in children

    normocapnia can be maintained with a fresh gas flow of 1000mls + 100mls/kg. e.g. a 4 year old

    weighing 20kg would need a total FGF of around 3litres/min. systems are all functionally

    similar.

    Combination of the Mapleson A, D and E Systems - The Humphrey A D E Circuit

    The Mapleson A circuit is inefficient for controlled ventilation as is the Mapleson D circuit for

    spontaneous ventilation. David Humphrey has designed a single circuit (Figure 5) that can be

    changed from a Mapleson A system to a Mapleson D by moving a lever on the metallic block

    which connects the circuit to the fresh gas outlet on the anaesthetic machine. The reservoir bag is

  • 7/30/2019 Breathings Systems

    52/67

    52

    situated at the fresh gas inlet end of the circuit, and gas is conducted to and from the patient

    down the inspiratory and expiratory limbs of the circuit.

    Depending on the position of the control lever at the Humphrey block, gases either pass through

    the expiratory valve or the ventilator port. When the lever is "up" the reservoir bag and the

    expiratory valve are used, creating a Mapleson A type circuit. When the lever is in the "down"

    position the bag and valve are by-passed and the ventilator port is opened creating a Mapleson D

    system for controlled ventilation. If no ventilator is attached and the port is left open the system

    will function like an Ayre's T piece (Mapleson E ).

    Like all pieces of equipment, it is essential that the anaesthetist fully understands the function ofa particular circuit. If the lever on the Humphrey block is moved from "up" to "down" whilst

    gases are flowing the breathing bag will remain full of gas but manual ventilation of the patient`s

    lungs by compressing the bag will be impossible and may resemble complete obstruction of the

    breathing circuit. This has led to anaesthetists ocasionally concluding that their endotracheal tube

    required changing.

    The Humphrey ADE

    breathing system

    This is a very versatile breathing

    system which combines the

    advantages of Mapleson A, D and E

    systems. It can therefore be used

    http://www.nda.ox.ac.uk/wfsa/html/u07/u07b_p05.htm
  • 7/30/2019 Breathings Systems

    53/67

    53

    efficiently for spontaneous and

    controlled ventilation in both adults

    and children. The mode of use is

    determined by the position of one

    lever which is mounted on the

    Humphrey block (Fig. 4.13). Both

    parallel and coaxial versions exist

    with similar efficiency. The parallel

    version will be considered here.

    Components

    1. Two lengths of 15 mm smoothbore

    tubing (corrugated tubing is

    not recommended). One delivers

    the fresh gas and the other carries

    away the exhaled gas. Distally

    they are connected to a Yconnection

    leading to the patient.

    Proximally they are connected to

    the Humphrey block.

    2. The Humphrey block is at the

    machine end and consists of

    a) an APL valve featuring a visible

    indicator of valve performance

    (Fig. 4.14)

    b) a 2 litre reservoir bag

    c) a lever to select either

    spontaneous or controlled

    ventilation

    d) a port to which a ventilator can

    be connected, e.g. Penlon

    Nuffield 200

  • 7/30/2019 Breathings Systems

    54/67

    54

    e) a safety pressure relief valve

    which opens at pressures in

    excess of 60 cmH2O

    f) a new design incorporating a

    soda lime canister is also

    available.

    Mechanism of action

    1. With the lever up (Fig. 4.15A) in

    the spontaneous mode, the

    reservoir bag and APL valve are

    connected to the breathing system

    as in the Magill system.

    2. With the lever down (Fig. 4.15B)

    in the ventilator mode, the

    reservoir bag and the APL valve

    are isolated from the breathing

    system as in the Mapleson E

    system. The expiratory tubing

    channels the exhaled gas via the

    ventilator port. Scavenging occurs

    at the ventilators expiratory valve.

    3. The system is suitable for

    paediatric and adult use. The

    tubing is rather narrow, with a low

    internal volume. Because of its

    smooth bore there is no significant

    increase in resistance to flow

    compared to 22 mm corrugated

    tubing used in other systems. Small

    tidal volumes are possible during

    controlled ventilation and less

  • 7/30/2019 Breathings Systems

    55/67

    55

    energy is needed to overcome the

    inertia of gases during spontaneous

    ventilation.

    4. The presence of an APL valve in

    the breathing system offers a

    physiological advantage during

    paediatric anaesthesia, since it is

    designed to offer a small amount

    of PEEP (1 cmH2O).

    5. During spontaneous ventilation

    a) a fresh gas flow of about 50 to

    60 ml/kg/min is needed in

    adults

    b) the recommended initial fresh

    gas flow for children weighing

    less than 25 kg body weight is 3

    litre/min, this offers a

    considerable margin for safety.

    6. During controlled ventilation

    a) a fresh gas flow of 70 ml/kg is

    needed in adults

    b) the recommended initial fresh

    gas flow for children weighing

    less 25 kg body weight is 3

    litres/min. However,

    adjustment may be necessary to

    maintain normocarbia.

  • 7/30/2019 Breathings Systems

    56/67

    56

    Fig. 4.13 The parallel Humphrey ADE

    breathing system.

    Fig. 4.14 The Humphrey ADE breathing

    systems APL valve (reproduced with

    permission from Dr D. Humphrey).

  • 7/30/2019 Breathings Systems

    57/67

    57

    Fig. 4.15 Mechanism of action of the parallel Humphrey ADE breathing system. With the lever

    up (A) the system functions in its

    Mapleson A mode for spontaneous ventilation. For mechanical ventilation the lever is down (B)

    and the system functions in its

    Mapleson E mode (reproduced with permission from Dr D. Humphrey).

    Circle Systems

    An alternative to using high flow circuits is to absorb CO2 from the expired gases which are then

    recirculated to the patient. These circuits are known as circle systems, were first devised by

    Brian Sword in 1926 and require smaller amounts of fresh gas each minute.

    Carbon dioxide is removed from the expired gas by passage through soda lime, a mixture of 94%

    calcium hydroxide and 5% sodium hydroxide, and 1% potassium hydroxide which reacts with

    CO2 to form calcium carbonate. Soda lime also contains small amounts of silica to make the

    granules less likely to disintegrate into powder and a chemical dye which changes colour with

    pH. As more carbon dioxide is absorbed the pH decreases and the colour of the dye changes

  • 7/30/2019 Breathings Systems

    58/67

    58

    from pink to yellow/white. When around 75% of the soda lime has changed colour it should be

    replaced . The soda lime canister should be mounted vertically on the anaesthetic machine to

    prevent the gases passing only through a part of the soda lime (streaming).

    Fresh soda lime contains 35% water by weight which is necessary for the reaction between

    carbon dioxide and soda lime to take place. This generates considerable heat. The soda lime may

    rise in temperature to 40 centigrade. There are therefore additional advantages of using circle

    systems in that the gases within the circle are warmed and humidified prior to inspiration.

    (Baralyme is a commercially available CO2 absorber which contains 5% barium hydroxide

    instead of sodium hydroxide.)

    Design of Circle Systems

    A circle system (Figure 6) is composed of two one way valves (one inspiratory and one

    expiratory), a reservoir bag, a fresh gas inlet, a canister of soda lime and an expiratory spill

    valve. Although there may be slight differences in the positioning of these components, all the

    systems function in the same way.

    Vaporiser Position. The vaporiser may be placed either outside the circle (VOC) on the

    anaesthetic machine in its conventional position, or rarely within the circle itself (VIC). Normal

    plenum vaporisers, with high internal resistance, cannot be used within the circle and a low

    internal resistance type vaporiser (such as the Goldman) is required. Drawover vaporisers such as

    http://www.nda.ox.ac.uk/wfsa/html/u07/u07b_p06.htm
  • 7/30/2019 Breathings Systems

    59/67

    59

    the OMV are not recommended for use within the circle because of the risk of over-dosage.

    Since the gases are recirculated, if the vaporiser is placed in the circle, gas already containing

    volatile anaesthetic agent will re-enter the vaporiser and the resulting output will exceed the

    vaporiser setting. This is a particular danger during controlled ventilation when dangerously high

    concentrations can build up. Vaporisers should only be placed inside the circle (VIC) when

    inspired volatile anaesthetic agent monitoring is available. It is safer to use conventional plenum

    vaporisers mounted on the anaesthetic machine outside the circle. In this case the maximum

    volatile anaesthetic agent concentration achievable within the circle cannot exceed that set on the

    vaporiser.

    Practical Use of Circle Systems. During the first 5 - 10 minutes of an inhalational anaesthetic

    using a volatile anaesthetic agent in oxygen and nitrous oxide, large amounts of the anaestheticagent and nitrous oxide will be taken up by the patient, and the nitrogen contained in the patient's

    lungs and dissolved in their body will be washed out. If low fresh gas flows are used

    immediately the patient is connected to the circuit the nitrogen will not be flushed out of the

    circle system and will dilute the anaesthetic agent concentration. This may be prevented by using

    conventional fresh gas flows of 6litres/min for the first 5-10 minutes of each anaesthetic before

    reducing the flow rates.

    Reducing the fresh gas flow rates. Inspired anaesthetic gases should contain no carbon dioxide

    and a minimum of 30% oxygen. Exhaled alveolar gas contains a lower concentration of oxygen

    and around 5% carbon dioxide which is removed from the exhaled gas on passage through the

    soda lime. A small amount of fresh gas is added before the next breath. At low fresh gas flow

    rates (

  • 7/30/2019 Breathings Systems

    60/67

    60

    These comments are less important if only oxygen and a volatile agent is being used in the circle.

    Under these circumstances there is no risk of oxygen dilution and the flows may be reduced to

    1000mls/min.

    With flows of >1500mls/min the inspired concentration of volatile agent will be similar to that

    set on the vaporisers. With flows

  • 7/30/2019 Breathings Systems

    61/67

    61

    gas flow to the patient and the

    other to receive exhaled gases

    from the patient, each port

    incorporates a unidirectional

    valve

    c) an APL valve connected to a 2

    litre reservoir bag.

    2. Inspiratory and expiratory tubings

    connected to the canister.

    3. A vaporizer mounted on the

    anaesthetic machine back bar

    (vaporizeroutside circleVOC)

    or a vaporizer positioned on the

    expiratory limb within system

    (vaporizerinside circleVIC).

    4. Soda lime consists of 94% calcium

    hydroxide and 5% sodium

    hydroxide with a small amount of

    potassium hydroxide. Silica is

    added to prevent disintegration of

    the granules into powder. A dye is

    added to change the granules

    colour when the soda lime is

    exhausted. Colour changes can be

    from white to violet or from pink

    to white.

    5. The size of soda lime granules is 4

    to 8 mesh. Strainers with 4 or 8

    mesh have four and eight openings

    per inch respectively. Therefore,

    the higher the mesh number, the

  • 7/30/2019 Breathings Systems

    62/67

    62

    smaller the particles that are

    retained. Soda lime can also be

    made to a uniform shape of 3 to 4

    mm spheres allowing a more even

    flow of gases and a reduction in

    channelling. This results in a

    longer life with lower dust content

    and lower resistance to flow. One

    kilogram of soda lime can absorb

    more than 120 litres of CO2.

    Mechanism of action

    1. Exhaled gases are circled back to

    the canister where carbon dioxide

    absorption takes place and water

    and heat are produced. The

    warmed and humidified gas joins

    the fresh gas flow to be delivered

    to the patient (Fig. 4.17).

    CO2 + 2NaOH Na2CO3 + H2O

    + heat

    Na2CO3 + Ca(OH)22NaOH +

    CaCO3

    2. The direction of gas flow is

    controlled via the unidirectional

    disc valves. These are mounted in

    see-through plastic domes so that

    they can be seen to be working

    satisfactorily.

    3. The canister is positioned

    vertically to prevent exhaled gas

    channelling through unfilled

  • 7/30/2019 Breathings Systems

    63/67

    63

    portions.

    4. The initial fresh gas flow

    required is several litres per

    minute for a period of 5 to 10

    minutes. This high flow washes

    any nitrogen from both the circle

    system and the patients

    functional residual capacity. This

    can later be reduced to 0.5 to 1

    litre/min.

    5. The circle system can be used for

    both spontaneous and controlled

    ventilation.

    6. Disposable circle breathing

    systems exist. They feature coaxial

    inspiratory tubing. The inner

    tubing delivers the fresh gas flow

    from the anaesthetic machine and

    the outer tubing delivers the

    recircled gas flow. Both gas flows

    mix distally. This allows a more

    rapid change in the inhalational

    gas and vapour concentration at

    the patient end.

    USE OF VAPORIZERS IN THE

    CIRCLE BREATHING SYSTEM

    VOC vaporizers (Fig. 4.18A) are

    positioned on the back bar of the

    anaesthetic machine. They are high

    efficiency vaporizers that can deliver

    high output concentrations at low

  • 7/30/2019 Breathings Systems

    64/67

    64

    flows. They have high internal

    resistance.

    1. The vaporizer should be able to

    deliver accurate concentrations of

    inhalational agent with both high

    and low fresh gas flows. This is

    easily achieved by most modern

    vaporizers (e.g. the Tec series).

    2. The volume of the circle system is

    large in relation to the low fresh

    gas flow used. Rapid changes in

    the concentration of the inspired

    vapour can be achieved by

    increasing the fresh gas flow to the

    circle system. Delivering the fresh

    gas flow distally, using a coaxial

    inspiratory tubing design, allows

    faster changes in inspired vapour

    concentration compared to

    conventional circle systems at low

    flows.

    VIC vaporizers (Fig. 4.18B) are

    designed to offer minimal resistance

    to gas flow and have no wicks on

    which water vapour might condense

    (e.g. Goldman vaporizer). The VIC is

    a low efficiency vaporizer adding

    only small amounts of vapour to the

    gas recirculating through it.

    1. Fresh gas flow will be vapour free

    and thus dilutes the inspired

  • 7/30/2019 Breathings Systems

    65/67

    65

    vapour concentration.

    2. During spontaneous ventilation,

    respiration is depressed with

    deepening of anaesthesia. Uptake

    of the anaesthetic agent is

    therefore reduced. This is an

    example of feedback safety

    mechanism. The safety

    mechanism is lost during

    controlled ventilation.

    Problems in practice and safety

    features

    1. Adequate monitoring of inspired

    oxygen, end-tidal carbon dioxide

    and inhalational agent

    concentrations are essential.

    2. The unidirectional valves may

    stick and fail to close because of

    water vapour condensation. This

    leads to an enormous increase in

    dead space.

    3. The resistance to breathing is

    increased especially during

    spontaneous ventilation.

    4. Compound A is produced when

    sevoflurane is used in conjunction

    with soda lime. Newer designs of

    soda lime claim lesser production

    of compound A.

    5. The circle system is bulkier, less

    portable and more difficult to

  • 7/30/2019 Breathings Systems

    66/67

    66

    clean.

    6. Soda lime is corrosive. Protective

    clothing, gloves and eye/face

    protection can be used.

    7. Because of the many connections,

    there is an increased potential for

    leaks and disconnection.

    Fig. 4.16 The circle breathing system.

  • 7/30/2019 Breathings Systems

    67/67

    Fig. 4.17 Mechanism of action of the circle breathing system.

    Fig. 4.18 Diagrammatic representation of the circle system with (A) vaporizer outsidethe circle (VOC) and (B) vaporizer inside the circle (VIC)