Breathing systems and the anaesthetic machine Zsolt Molnár SZTE, AITI.

24
Breathing systems and the anaesthetic machine Zsolt Molnár SZTE, AITI

Transcript of Breathing systems and the anaesthetic machine Zsolt Molnár SZTE, AITI.

Page 1: Breathing systems and the anaesthetic machine Zsolt Molnár SZTE, AITI.

Breathing systems and the anaesthetic machine

Zsolt Molnár

SZTE, AITI

Page 2: Breathing systems and the anaesthetic machine Zsolt Molnár SZTE, AITI.

• Irish anaesthetist (1888-1986)– Endotracheal tube, anaesthesia (The Magill endotracheal tube, oral & nasal

designs)– An anaesthetic breathing system (Magill circuit and expiratory valve)– Intubating Forceps– Laryngoscope a straight bladed version– Catheter mount (endotracheal tube-to-circuit connector)– Endotracheal tube connectors, oral & nasal versions– Magill tracheal spray (glottis anaesthetic)– Single-lung anaesthesia– Endobronchial tubes and bronchial blockers– Bobbin flowmeters

Sir Ivan Whiteside Magill

Page 3: Breathing systems and the anaesthetic machine Zsolt Molnár SZTE, AITI.

• British anaesthetist

• Cardiff 1952– Mapleson WW. The elimination of rebreathing in various semi-closed

anaesthetic systems. Br J Anaesth 1954; 26: 323–32

– Mapleson WW, Mushin WW. Relaxant action in man: an experimental study. Anaesthesia 1955; 10: 265–78, 379–90

Mapleson WW. Br J Anaesth 2004; 93: 319-321.

WW Mapleson

Page 4: Breathing systems and the anaesthetic machine Zsolt Molnár SZTE, AITI.

• „Auld” definitions• Open-, semi open-• Semi-closed, closed-systems

• „New”: Mapleson systems 1954• Rebreathing• Non-rebreathing

Introduction

Page 5: Breathing systems and the anaesthetic machine Zsolt Molnár SZTE, AITI.

• VT• 4-7 ml/kg

• VD• 2-3 ml/kg

• Peak inspiratory flow (PIF)• At rest: 20-30 l/min

• Breathing cycle• Inspiration – expiration – expiratory pause

Physiolgy refreshment course

Page 6: Breathing systems and the anaesthetic machine Zsolt Molnár SZTE, AITI.

• Parts• FG inlet• Rezervoire balloon• Pop-off valve

• Efficacy• FGF – to prevent CO2 rebreathing

Mapleson systems

Page 7: Breathing systems and the anaesthetic machine Zsolt Molnár SZTE, AITI.

• Parts• Disc• Sharp edge • Spring

• Safety opening P• 60-70 cmH2O

Pressure relieving valve

Page 8: Breathing systems and the anaesthetic machine Zsolt Molnár SZTE, AITI.

• „Magill attachment”• >110 cm• FGF = 1 x VA (70ml/kg/m)• VA = ?• PIF = ?• Monitoring: capnograph

Mapleson A – spont. breathing

Page 9: Breathing systems and the anaesthetic machine Zsolt Molnár SZTE, AITI.

• A: normal

• B: rebreathing

• C: diaphragm movements

• D: cardiac oscillation

• E: sampling error

• F: bronchospasm

Molnár ‘99

A B

C D

E F

pCO2

Capnography

Page 10: Breathing systems and the anaesthetic machine Zsolt Molnár SZTE, AITI.

History

• Birth of: intensive therapy, blood gas machine, and the use of Mapleson C

• 1952 Coppenhagen: polio-epidemic

• 31 patients – 21 dead within 72 hrs

• Bjorn Ibsen (1915-2007)• Waters: To-and-fro

Lassen HCA. 1956. Management of poliomyelitis. In HCA Lassen Management of life threatening poliomyelitis. Copenhagen, 1952-1956; E.S. Livingstone Ltd, London

SZEGEDI INTENZÍVES TALÁLKOZÓ 2010

SZINT

Page 11: Breathing systems and the anaesthetic machine Zsolt Molnár SZTE, AITI.

Water’s circuit

Molnár ‘99

Mask, ET-tube

O2

Valve

Rezervoire baloon

Page 12: Breathing systems and the anaesthetic machine Zsolt Molnár SZTE, AITI.

• E: Ayre’s T-piece• Spont: FGF = 1.5 x VA

• IPPV: FGF = 3 x VA

• F: Jackson-Rees• Monitoring• PEEP

Mapleson E and F

Page 13: Breathing systems and the anaesthetic machine Zsolt Molnár SZTE, AITI.

T-piece

Molnár ‘99

T-piece

Inlet

Outlet

Page 14: Breathing systems and the anaesthetic machine Zsolt Molnár SZTE, AITI.

A happy chapy on T-piece

Molnár ‘99

Page 15: Breathing systems and the anaesthetic machine Zsolt Molnár SZTE, AITI.

The anaesthetic machine

• Cylinders (O2, N2O, Air)• Rotameters• Vaporizers• Breathing circuit• CO2 absorber• Safety:

• FiO2 + alarm• O2 bypass• Pressure relieving valve

Page 16: Breathing systems and the anaesthetic machine Zsolt Molnár SZTE, AITI.

Rebreathing system (circuit)

Page 17: Breathing systems and the anaesthetic machine Zsolt Molnár SZTE, AITI.

CO2 absorber

• CaOH2, NaOH, sylicate

• CO2 + 2NaOH Na2CO3 + H2O + heat• Na2CO3 + CaOH2 2NaOH + CaCO3

Page 18: Breathing systems and the anaesthetic machine Zsolt Molnár SZTE, AITI.

Rotameters

Page 19: Breathing systems and the anaesthetic machine Zsolt Molnár SZTE, AITI.

Rotameters

10

4

8

πpp

l

RI

• Gas specific• Safety features:

• Min FiO2 = 0.3• FiO2 alarm

Page 20: Breathing systems and the anaesthetic machine Zsolt Molnár SZTE, AITI.

Vaporisers ages ago…

• Inside the circuit• Small vaporizing surface• Quick cooling

Page 21: Breathing systems and the anaesthetic machine Zsolt Molnár SZTE, AITI.

…and today

• Outside the circiut• Well insulated containers

Page 22: Breathing systems and the anaesthetic machine Zsolt Molnár SZTE, AITI.

…and today

• Outside the circiut• Well insulated containers• Agent safe

Page 23: Breathing systems and the anaesthetic machine Zsolt Molnár SZTE, AITI.

…and today

• Outside the circiut• Well insulated containers• Agent safe• Heat compensation

Page 24: Breathing systems and the anaesthetic machine Zsolt Molnár SZTE, AITI.

• Mapleson systems: advantages• Simple, cheap, safe• Low resistance• Good for practice

• Mapleson systems: disadvantages• Non-economic• Not environment friendly• High heat and humid loss

Summary

Get to know your equipment!