Blood Pressure and Sound (2) Dept. of Biomedical Engineering 2003200449 YOUNHO HONG

Transcript of Blood Pressure and Sound (2) Dept. of Biomedical Engineering 2003200449 YOUNHO HONG.

Page 1: Blood Pressure and Sound (2) Dept. of Biomedical Engineering 2003200449 YOUNHO HONG.

Blood Pressure and Sound (2)

Dept. of Biomedical Engineering2003200449

YOUNHO HONG

Page 2: Blood Pressure and Sound (2) Dept. of Biomedical Engineering 2003200449 YOUNHO HONG.

IBP ( Invasive BP ) measurement

t

Pi

catheterBloodvessel

Pi

pill up withsome liquid(ex.saline)

diaphragm

strain gages

cablePo

t

Po

If we choose sticky and dense liquid, We can’t get the signal of Pi just like the graph.To get the signal, We should concern with distance, diameter of a catheter and liquid, air-bubble inside a catheter.

Page 3: Blood Pressure and Sound (2) Dept. of Biomedical Engineering 2003200449 YOUNHO HONG.

IBP ( Invasive BP ) measurement

# Equivalent Circuit Model Of Catheter-Sensor System

(1) Resistance

AL

A

LVRRiV

, : resistance Electrical

resistivity

P1

P2

)21( PPF R

FRop : resistance Liquid

A

LR

viscosity

Page 4: Blood Pressure and Sound (2) Dept. of Biomedical Engineering 2003200449 YOUNHO HONG.

IBP ( Invasive BP ) measurement

(2) Capacitance or Compliance

x

AC

dt

dvCi ,

dt

dpCf modulus sYoung'

YC

(3) Inductance or Inertance

dt

diLV

2 ,

A

mL

dt

dfLP

Page 5: Blood Pressure and Sound (2) Dept. of Biomedical Engineering 2003200449 YOUNHO HONG.

IBP ( Invasive BP ) measurement

multiphysics Electric Circuit Fluid Mechanics

voltagecurrentcharge

pressure flow volume

dtdVI

C

dtdIV

L

A

L

I

VR

)( ][

modulus sYoung'

)(

)8

( ][

2

43

C

r

L

dtdFP

L

r

Lm

sPaF

R

Page 6: Blood Pressure and Sound (2) Dept. of Biomedical Engineering 2003200449 YOUNHO HONG.

Equivalent Circuit Model of IBP

catheterPi

liquid

diaphragm

strain gages

cablePo

L c

C d

R c

Vi

Compliance of diaphragm

+Vo -

i

ODEorder 2nd :

20

20

0

00

dt

VdCL

dt

dVCRVV

dt

dVCiV

dt

diLiRV

dcdci

dcci

)()(]12

[

operator :

2

2

tKVtVW

D

W

D

dt

dD

ionn

dc

n

c

dc

CLW

L

CRK

1

2 ,1

dampingratio

naturalfrequency

Page 7: Blood Pressure and Sound (2) Dept. of Biomedical Engineering 2003200449 YOUNHO HONG.

Equivalent Circuit Model of IBP

(1) Frequency Transfer Function

nnnn

i

o

Ww

jWw

Wjw

WjwjwV

jwVjwH

2)(1

1

12)(

1

)(

)()(

22

)2

(tan

)(4])(1[

1

))(1

2

(tan

)(4])(1[

1)(

1

2222

2

1

2222

w

W

Ww

Ww

Ww

WwWw

Ww

Ww

jwH

n

nnn

n

n

nn

Page 8: Blood Pressure and Sound (2) Dept. of Biomedical Engineering 2003200449 YOUNHO HONG.

Equivalent Circuit Model of IBP

w

|H|

Wn

)dunderdampe( 5.0

)damped critically( 1

)overdamped( 2

w

-π/2

∠H)dunderdampe( 5.0

)damped critically( 1

)overdamped( 2

Page 9: Blood Pressure and Sound (2) Dept. of Biomedical Engineering 2003200449 YOUNHO HONG.

Equivalent Circuit Model of IBP

Methods to solve 2nd order ODE io VV

dt

dVRC

dt

VdLC 0

20

2

function transfer loperationa : 1

1)(

dt

d )1(

)1(

2

2

2

RCDLCDV

VDH

DVVRCDLCD

VVRCDVVLCD

i

o

io

iooo

jwD

i

o

iooo

DH

jwRCjwLCV

VjwH

VVRCjwVVjwLC

)(

1)()(

1)(

)( )2(

2

2

Page 10: Blood Pressure and Sound (2) Dept. of Biomedical Engineering 2003200449 YOUNHO HONG.

Steady State Freq. Response

)4

2sin()(

)2sin()(

1

1

tfKAtV

tfAtV

o

i

|H|

ff1 f2

K

0.5

)8.12sin(5.0)(

)2sin()(

2

2

tfAtV

tfAtV

o

i

f2f1

-4/π

-1.8π

∠H

Page 11: Blood Pressure and Sound (2) Dept. of Biomedical Engineering 2003200449 YOUNHO HONG.

Unit Step ResponseIn reality, We need a unit step function for a starting point.

For example, should be )()2sin()( 1 tutfAtVi )2sin()( 1tfAtVi

input signal

underdamping

critical damping

overdamping

Page 12: Blood Pressure and Sound (2) Dept. of Biomedical Engineering 2003200449 YOUNHO HONG.

Transient Step Response

Pbulb

balloon

saline

Po

underdamping

critical damping

overdamping

Page 13: Blood Pressure and Sound (2) Dept. of Biomedical Engineering 2003200449 YOUNHO HONG.

Example(7.1) A 5mm-long air bubble has formed in the rigid-walled catheter to a Statham P23Dd sensor. The catheter is 1m long, 6 French diameter, and filled with water at 20 ℃. Plot the frequency-response curve of the system with and without the bubble.

033.0)(4

91)1

(2

21

3

21

P

VL

r

HzL

P

L

rfn

137.0

22,

bub

bubn Hzf

log f1.34 1.95

Page 14: Blood Pressure and Sound (2) Dept. of Biomedical Engineering 2003200449 YOUNHO HONG.

Example(7.2) By changing only the radius of the catheter, redesign the (no-bubble) catheter of Figure 7.9 to achieve the damping ratio ζ=1. Calculate the resulting natiral frequency fn.

Hzf

rrff

r

r

rr

n

nn

29

147.0

0032.0

00

3

03

03

log f1.46

Page 15: Blood Pressure and Sound (2) Dept. of Biomedical Engineering 2003200449 YOUNHO HONG.

Thank you.