BLD 414 Notes

45
BLD 414 Notes 1/9/12 8:54 PM Osmolality Osm/Liter Osmole = species in solution (after disassociation) Full Disassociation 1mol NaCl 1mol Na + + 1mol Cl - 1 mole in 2 species in solution 1 mole NaCl = 2 Osm/L 1mol CaCl 2 1mol Ca 2+ + 2mol Cl - 1 mole in 3 species in solution 1 mole CaCl 2 = 3 Osm/L 1 mole Glucose: does not disassociate o Therefore 1 mole = 1 Osm/L Partial Disassociation 1 mole MgSO 4 disassociates 60% .4mol MgSO 4 ---partial--- .6mol Mg 2+ + .6mol SO 4 2- 1 mole MgSO 4 = 1.6 Osm/L

Transcript of BLD 414 Notes

Page 1: BLD 414 Notes

BLD 414 Notes 1/9/12 8:54 PM

Osmolality

Osm/Liter

Osmole = species in solution (after disassociation)

Full Disassociation

1mol NaCl 1mol Na+ + 1mol Cl-

1 mole in 2 species in solution

1 mole NaCl = 2 Osm/L

1mol CaCl2 1mol Ca2+ + 2mol Cl-

1 mole in 3 species in solution

1 mole CaCl2 = 3 Osm/L

1 mole Glucose: does not disassociate

o Therefore 1 mole = 1 Osm/L

Partial Disassociation

1 mole MgSO4 disassociates 60%

.4mol MgSO4 ---partial--- .6mol Mg2+ + .6mol SO42-

1 mole MgSO4 = 1.6 Osm/L

Page 2: BLD 414 Notes

Density

Mass/Volume

Grams/cm3

cm3 = mL

Usually for liquids

H2O at 4 degrees Celsius to standardize

o 1 gram per 1 cm3

Specific Gravity

o For liquids only

o Unitless

o g/mL (implied)

Make 1 liter of 2M MeOH (methanol)

o MeOH = 32 g/mol

o SG = .87

o 2mol/L * 32g/mol = 64 g/L needed

o .87g/mL = 64g/X

Page 3: BLD 414 Notes

X = 73mL QS to 1 liter

Percent Purity

Focus on liquids

Done by weight

Make 1 liter of 1M HCl

o HCl = 36.5g/mol

o SG = 1.029

o Bought 75% pure

o 1mol/L * 36.5g/mol = 36.5g/L

o 1.029g/mL * .75 = .772g/mL

o .772g/mL = 36.5g/X

X = 47.3 mL QS to 1 liter

Dilution

Ratio dilution: not common except in microbiology

o Part solute : part solvent

Page 4: BLD 414 Notes

o 1mL serum : 9mL water

1mL to 9mL

o Use colon

True dilution: used more often

o Part solute : total volume

o 1mL serum and 9mL water

1/10 = true dilution

1mL in 10mL total volume

o Use slash

Volumes of Dilution

o Make 40mL of 1/5 dilution of serum in saline

1/5 = x/40

x = 8mL, add 32mL of water

Dilution Factor (DF)

o Final concentration = original concentration * DF

o DF is the true dilution fraction

o Want .5M HCl starting with 2M HCl

.5 = 2*DF

.5 = 2*(1/x)

Page 5: BLD 414 Notes

1/x = .5/2

x = 4

DF = ¼

Independent Dilutions

o Non-equivalent dilutions

o Have 5M HCl and need 2M, .5M, & .2M

Method 1: use this

Add same amount of solute to each tube

Figure out DF for each

Calculate how much solvent to add to each

Method 2

Add same amount of solvent to each

Ex) add 4mL of water to each

x/(4+x)

x = volume of solute

Serial Dilutions: common

o Equivalent dilution factors

o n = tube number

o 1/x = DF

o To calculate concentration of any given tube

Page 6: BLD 414 Notes

1/x ^ n

pH

Relative concentration of two species (H+ and OH-)

Kd = limited by the disassociation of water

o H2O [H+] * [OH-]

o Kd = [H+][OH-] = 10-14

H+ and OH- are always in 1:1

o -14 = logH+ + logOH-

o 14 = -logH+ - logOH-

-logH+ = pH

[H+] and [OH-] always expressed in M (mole/Liter)

o 14 = pH + pOH

ex) 3mM HCl, what is pH?

o Complete disassociation

o HCl H+ + OH-

o 3mM 3mM + 3mM

o pH = -log[3*10-3]

pH = 2.5

ex) pH = 8.5, what is [H+]

o Take negative antilog

Page 7: BLD 414 Notes

o 10-8.5 = H+

o [H+] = 3.16*10-9 M

Weak Acids

Does not completely disassociate

Acetic acid

o HAC H+ + AC-

o Some HAC remains

o Kd = ([H+][AC-])/HAC

Want [H+]

o [H+] and [AC-] are equal (what if not?)

o Kd = x2/HAC – x

Use quadratic equation

Get rid of x in denominator if small enough

pK = -log[Kd]

If pK > 4, disregard x in denominator

Ex) Have .5M HAC, what is the pH?

o pK = -logKd = 4.76

o Kd = x2/HAC

o 10-4.76 = Kd = 1.74*10-5

o 1.74*10-5 = x2/.5

o x = 2.95*10-3

Page 8: BLD 414 Notes

o pH = -log[2.95*10-3]

pH = 2.5

Henderson-Heiselbach

HAC H+ + AC-

Kd = ([H+][AC-])/HAC

Kd/[H+] = [AC-]/HAC

o logKd – logH+ = log(AC-/HAC)

o –logH+ = -logKd + log(AC-/HAC)

o pH = pK + log(AC-/HAC)

Buffers

Solution contain weak acid resists change in pH in presence of

added base

Buffering Capacity (BC)

o From [base] = 0 to end of flat line region

o How much base to add to use all available protons

Protons from free H+ and from non-disassociated acid

o When acid molarity doubles, need 2x more base to reach BC

o Middle of flat line: AC- = HAC and pH = pK

pK = constant

ex) Prepare a 1 liter buffer solution of .1M acetate with pH = 5.2

and pK = 4.76

o pH = pK + log(AC-/HAC)

Page 9: BLD 414 Notes

o 5.2 = 4.76 + log(AC-/HAC)

o .44 = log(AC-/HAC)

o 2.75 = AC-/HAC maintain this unit-less ratio

o BC = .1M = [AC-] + [HAC]

o [AC-] = .1 – [HAC]

plug into ratio

o 2.75 = (.1 – HAC)/HAC

[HAC] = .027M

o AC- = .1 - .027

[AC-] = .073M

o Combine [HAC] and [AC-] and QS to 1 liter

Quality Assurance

Predictive Value Theory

Relevance of testing: epidemiology

Diagnostic Sensitivity (%)

o Performance of lab test on a known diseased population

o Correct lab result: true positive (TP)

o Incorrect lab result: false negative (FN)

o TP/(TP+FN)

Diagnostic Specificity (%)

Page 10: BLD 414 Notes

o Performance of lab test on a known non-diseased population

o Correct lab result: true negative (TN)

o Incorrect lab result: false positive (FP)

o TN/(TN+FP)

Predictive Value (PV)

o Index of degree of confidence associated with a positive or

negative result

Needed because of the effect of prevalence due to lab

performance done in a general population

o Positive PV

TP/(TP+FP)

o Negative PV

TN/(TN+FN)

o Diagnostic Efficiency

(TN+TP)/(TN+TP+FN+FP)

Diagnostic Indicators to Make Decisions

o High sensitivity needed

High PPV

Low NPV

o High specificity needed

Low PPV

High NPV

o High sensitivity and high specificity needed

Page 11: BLD 414 Notes

High PPV

High NPV

Normal Reference Range

Normal RR = 95% confidence interval on Gaussian curve

o Measure of variance: Z-test

o %CV = SD/mean * 100%

Misrepresentation of normal distribution results in compromising

diagnostic sensitivity and specificity

o Normal RR too wide

Increased specificity

Increased false negative

Decreased sensitivity

Decreased false positive

o Normal RR too narrow

Increased sensitivity

Increased false positive

Decreased specificity

Decreased false negative

Use exclusion criteria to identify normal RR

o Disease, drugs, etc.

Use partitioning factors to make subclasses of population

Page 12: BLD 414 Notes

o Age, gender, diet, etc.

o Priori: separate data using known information

o Posteriori: analyze collected data then partition

o Z-test

Requirements

o Ensure accuracy and precision, avoid transference

o 95% CI of normal population in parametric (Gaussian)

distribution

Chi Square Goodness of Fit Test

Pass = parametric

Pass: median is +/- 3% of mean

Can transform non-G to G if Chi2 test has 95% CI

o Reference data array

n = total frequency

Upper limit: (n+1)*.975

Lower limit: (n+1)*.025

Receive Operated Characteristic Curves (ROC Curves)

Cutoffs: don’t use 95% confidence interval

(x) = 1-specificity, (y) = sensitivity

Want to find inflection point

o Part of curve that has most change

o Highest sensitivity

Page 13: BLD 414 Notes

o Lowest false positive

o Best diagnostic performance

Quality Control

Observed error < allowable error

Observed error = accuracy error + precision error

o Accuracy: difference between true value and mean (x bar)

o Precision: 2 standard deviations out from mean

Allowable error

o Consensus: government, many labs

Analytic error : use mean as true value

2 SD/mean * 100%

o Medical Allowable Error: insurance $, professional societies,

few labs

Weighted average

2 segments/mean * 100%

o Use lower error of the two when publishing

o Expressed as %, [], or SD from mean

Internal and external QC

Establish temporary QC values

o T-test

o Mean +/- allowable error = floating mean

Page 14: BLD 414 Notes

o Mean = (temporary) Target

o Once n = 200, final target mean established

Change in mean is small as n increases

o Outdate: due to time with specimen degrading, etc.

Concentration of QC Material

o One for normal and for one for diseased populations

Monitoring Controls

o Usually once per day

o More frequent with drift: analytic system change, needs

recalibration

Detect QC Problems

o Using Westgard Rules on Levy-Jennings graph: internal

2 graphs: normal and abnormal, ran simultaneously

12s

13s

22s

R4s

41s

10x

o External QC: Proficiency Testing

Law

Certifying agencies

Page 15: BLD 414 Notes

Blind specimen, use values determined by allowable

error (MAE or consensus)

3 strikes in a row is bad

Method Evaluation

Compare observed to allowable error

Verify analytic machine technical specifications

Analytic Range: Quantitative Test

o Standard curve and working range

o Want working range as wide as possible

May need dilution to achieve this

o Upper Limit: Calculate R value

Best: R = 1

Look for point of inflection on graph of R values on Y-

axis and # of points on X-axis, then use that many data

points

o Lower Limit

Analytic

Zone near origin where value is always 0

Functional: sig figs

Regression line with 95% CI around it

n = same for each x column

Analytic Specificity: Accuracy

o Interfering substances

Page 16: BLD 414 Notes

o Percent Recovery

Make one tube the control and find the [base mean]

Measure the [observed] of the sample tubes

Report the % shy or over: is it within allowable error?

o Paired-Difference : want maximum physiologic concentrations

Make one tube the control and find the [base mean]

Make a 95% CI (2SD on each side of x) around the [base

mean]

Measure the [observed] of the sample tubes with

different possible interfering substances

If [observed] is within the 95% CI around [x], then

the substance is not interfering

Find point of interference

Monitored response on Y-axis

[interfering substance] on X-axis

Make regression line using the same n for each x

Make 95% CI around regression line

Point of no overlap

Error

o Observed compared to allowable

Observed: accuracy and precision/random

o MDL: concentrations of substances to diagnose

Within run: 2SD / mean = %CV

Between run: change everything about assay except

specimen

Page 17: BLD 414 Notes

Random/Precision Error (RE)

Between run between day

2 SD of the observed mean

Systematic/Accuracy Error (SE)

Compare using reference method vs. candidate

Plot reference = x, candidate = y

Draw regression: y = mx + b

Ideal: y = x

Constant error: same slope, different y-int

Proportional error: different slope

o Can be both (usually is)

Total Error = RE + SE

Take TE / [MDL] * 100%

Page 18: BLD 414 Notes

BLD 414 Notes 1/9/12 8:54 PM

LIS: Laboratory Information Systems

HIS: Hospital information system

o Allows for communication via LAN

In-house request

o Admission: gather patient info, wristband = barcode

o Panel: test most pertinent for situation/condition

Pre-prepared/standardized

CPOE: Computerized Physician Order Entry

o Accession number: _ _ _ _ _ _ _ _ / _ _ (identification)

o Compile/sort based on priority

Routine: at leisure, 24 hours; time and location

Timed: prior to timed test, location; certain analytes at

precise times

STAT: critical conditions; goes straight to lab without

sorting

o Compile to batch list

List of info per patient, printed after positive

identification

o Collect specimen, take to lab, sort/ID

Conveyor belt with forks

o Order any tests necessary via LIS

Upload: transfer info from instrument to mainframe

Download: transfer info from mainframe to instrument

o Data verification

Page 19: BLD 414 Notes

Auto: computer verifies normal RR results

Also does delta check for intraindividual variation

Manual: if failed delta check

o Output data

Auxiliary features: new/future

o Patient and specimen auditing

o QC: Westgard rules

o Balance department work load

Automation

Many accomplishments

Computing Relationships

o Integrated: all parts from single vendor; seamless system;

limited to what vendor can offer – non-modifiable; expensive

o Interfaced: multiple vendors; compatibility issues – software;

flexible – modifiable; cheaper; more common

Total Lab Auto

o No humans involved

o Processing/pre-analytical (similar to conveyor belt)

o Analytical, post analytical (storage)

Partially Auto Labs

o Most common

Instruments

Page 20: BLD 414 Notes

o Analytical

Batch analyzer

Dedicated system: for one task

Multichannel analyzer

Group of tests chosen by need

Continuous flow uses this

Random access

Chosen from very large menu of tests

Upload and download

o Configuration

Discrete

Each test separated into cuvettes in a chain

Tray or wheel

Continuous flow

Buffer flows continuously

Bubble-specimen-bubble

Multichannel series and parallel

Series: non-destructive – no addition of

reagent; direct detection

Continuous reagent: specimen introduced to

reagent

Terms/concepts

o Dwell time

Page 21: BLD 414 Notes

Time per test

Want to verify salesmen claims

Varies per analyte

o Throughput

Test in one hour

Relates to dwell time

Varies per analyte

Varies per pipetting rate

Wash cycle: more = lower throughput

Tests/specimen: more = higher throughput

o Turn Around Time

Lab TAT

Specimen arrival to lab until result output to LIS

Gathering specimen rate limiting step

Total TAT

Test ordered by doctor until result output

o STAT

Highest priority/life-threatening

Not all tests offered

1 hour

Acquiring specimen rate limiting step

Page 22: BLD 414 Notes

Basic Measurement

Express data

o Qualitative

o Semi-quantitative: relative scale, visual, correlate to values

o Quantitative: most data, instruments, standard curve, units

Signal = response

o Change in monitored value related to change in []: scale

o Light-based = lumens

o Electronic based = flowing electrons

o Physical: partial pressure, osmometry, etc.

o Change in signal

Amplitude: magnitude

Rate of generation

Per unit or kinetic

Frequency

Duration

o Detection

Direct: limited

Indirect: convert signal to different unit; processor

Measurement Scale

Page 23: BLD 414 Notes

o Standard curve

Interpolation: have signal, find []

Direct (positive) or Indirect (negative)

Linear or nonlinear

Transform nonlinear to linear by taking log

Want to go through origin: blanking

o Made from standards

Primary: government, NBS, 99% pure, expensive,

bought by manufacturers

Secondary: used by lab after [] established by

manufacturers

Factors Effecting Relationship

o Drift: stability of standard curve; redo curve when drift seen

Change in response over time

Upward drift: signal increase

Downward drift: signal drop

Monitor with QC

o Noise

Property of electronic analog systems

More apparent at high signal

Less apparent at low signal

Signal : Noise ratio

High signal for low detection is good

Page 24: BLD 414 Notes

Low noise: good

Big signal: high slope on standard curve

o Artifactual output

Point error

Analytic Systems

Review

Spectrophotometry 2/10/12

Electromagnetic radiation

o Wavelength: 1 complete cycle; nanometers

o E = hc/wavelength

More energy = smaller wavelength

o Pertinent regions (increasing in wavelength)

Gamma: ionize

X-ray

UV: far = 10-180; near: 180-380

Visible: purple = 380; red = 750

Infrared: generate heat

Radiowaves: RFID

o Intensity of Light: dim or bright; lumens

Bright = higher amplitude

Page 25: BLD 414 Notes

Power = ability to do work

o Interference

Constructive: in phase, positive

Same wavelength

Both maximums hit together

Increase amplitude: brighter

Destructive: 180 deg out of phase, negative

Max of one hits min of other

Negate each other: no light

o Diffraction

Light bends around edge

More bending = less intensity

Laser: monochromatic = single wavelength

Banding (constructive and destructive)

Blaze: closest degree band to original light

Highest intensity

Shorter wavelength: larger degree of banding

Sunset: red glow at start, purple at end

o Refraction: light bends moving between states of matter

More bend moving from gas to solid

More density change = more bend

Page 26: BLD 414 Notes

Longer wavelength = bent less

Shorter wavelength = bent more

o Polarized light

Natural light is omnidirectional

Dichroic crystal: layers of glass in single direction

Absorb all but one plane of light

Omniplanar in uniplanar out

Absorption Spectrophotometry: Part 1

o Each molecule has unique absorption

o Light hits electron and is absorbed

o Electrons close to nucleus has shorter wavelength

Higher energy

o If photon hitting electron is same wavelength: light absorbed

Electron thrown into high energy state

o Each peak: electron math at given wavelength

o Seen wavelength = no absorbed

o White light: total spectrum

o Colorimetry

Visual

Serial dilution of high purity standard used to

compare unknown

Page 27: BLD 414 Notes

Duboscq

Depends on volume (depth) of test tube and

concentration of color

Pre-light bulb

Eyepiece: try to adjust to produce 2 identical

(standard & unknown) colors – by changing

volume (depth)

Cs * hs = Cu * hu; find Cu

o Beer/Lambert

-log(transmittance) = absorbance

transmittance: out/in

need to linearize data by taking the log

used filter

absorbance is unit less

Molar absorptivity

Abs = epsilon * b * concentration

Epsilon = liters/mole*cm

b = 1cm; abs = epsilon * concentration

want biggest slope of abs vs. conc

large signal/noise ratio

epsilon at specific wavelength maximum

Absorptivity

Used when molecular weight unknown

a = liters/gram * cm

Page 28: BLD 414 Notes

o Quantitative Analysis

Direct: compounds that naturally absorb

Chromogenic reaction: absorbing substance made

through chemical reaction

Derivitization: substance binds to analyte, separate free

and bound

o Determining the [unknown]

Standard curve: graph

y = mx + b

m = epsilon = molar absorptivity

b = zero

x = concentration: solve for this

y = absorbance, measure this

Direct: no graph

Concentration = absorbance/epsilon

Conc = mole/liter

Assume 1cm test tube

Ratio: common for large medically allowable error

Assays with drift

Cu = (Absu/Abss) * Cs

Limited by upper limit of linearity

Big allowable error: 2 point curve

o Origin and one other point

Page 29: BLD 414 Notes

Small MAE: multipoint curve

o Interference

Reagent blank

Proteins

Find absorbance in absence of analyte

Specimen blank

Calcium

Find absorbance before adding color reagent

Chemical separation

2/15/12: Alan Correction

No discrete peak of interferent

Abs(a) = Abs(i) – [(Abs2 + Abs1)/2]

Abs(a) at wavelength max

Bichromic/Simultaneous Analysis

When interfering peak in same region as analyte

Total absorbance proportion by interferent

calculated by equations on handout

Absorption Photometry Instrumentation

Light source

o Intensity of emissions from sources vary of range of useful

wavelengths: non-constant emission

Wavelength selectors

Page 30: BLD 414 Notes

o Optical filter: glass/sandwich

o Interference filter: one-way mirrors

Mono-chromatic

Diffraction grating: can dial in desired wavelength

o Long wavelengths bend less, and vice-versa

o Linear separation of wavelengths

o Transmission: many edges; increase light intensity

Less pivot: more red

More pivot: more purple

o Reflection

Reflection at different angles gives different wavelength

Not monochromatic, range instead

Like a CD in the sun

Refraction: pass between states of matter

o Greater change in density: greater bend

o Select wavelength by pivot

Red: little pivot

Purple: more pivot

o Nonlinear distribution of wavelengths

Long wavelengths close together: reds

Page 31: BLD 414 Notes

Short wavelengths far apart: purple

Bandpass

o Measure of monochromicity

o Bandwidth: wavelength range with zero light intensity

o Bandpass: wavelengths associated with ½ max intensity, easy

to measure

o Narrow bandpass is better (for specificity)

o Narrow bandpass gives steeper slope in standard curve of

Absorbance vs. Concentration

Narrow bandpass: good for lower limit of detection

(sensitivity)

Higher signal to noise ratio

o Wavelength selectors

Prism: non-linear = more separated wavelengths

better bandpass

Grating: linear; best; slit width dependent

o Cuvet

Material and diameter

Absorbing UV is bad

Detectors

o Want to extend upper limit of linearity, limit noise

o Photo Electric Cell

Selenium: shift to conductor when light hits

Create current: more light = more Amps

Not good for dim light for detection

Page 32: BLD 414 Notes

Slow to react

o Phototube/Vacuum Photo Diode

Fast

Linear response

o PMT

Multiplier: 1 e- hits, 6 3- released

o Silicone Photo Diode

Standardizing

o Zeroing

o Use occlude: maintain intensity over wavelength range

o Going from dim to bright: apparent absorption is negative

Errors in Spectroscopy

Incident light is not monochromatic: wind bandpass

o Lets many wavelengths through: low sensitivity

o Solution: decrease bandpass by decreasing slit width

This also decreases light intensity to a limited point

Spectral stray light

o Outside of absorbed wavelengths

Stray Light Instrument

o Stray light is constant as concentration of analyte changes

Page 33: BLD 414 Notes

o Inappropriate light reaching detector: cuvette or leak

Above ideal behavior

o Insoluble particles: precipitate does not absorb

Fluorescence

Molecule in high energy electronic state

If light wavelength hits an electron of same wavelength

o Transfers energy

o Moves electron to different energy orbital

Wants to return to ground state

Fluorescence

o Rapid return to ground state

o Longer wavelength emitted than wavelength absorbed

o High intensity light

Phosphorescence

o Slower return to ground state

o Low intensity light

Internal Conversion

o Infrared: heat release

o Step-wise return to ground state

o No light emitted

Photochemical reaction

Page 34: BLD 414 Notes

o Unstable

o Ion pair created

Fluorescent Spectroscopy

o Use absorbed wavelength that hits electron:

excitation/primary wavelength

o Secondary wavelength emitted

Monitored response is intensity

o F = IC F is secondary wavelength emitted intensity

I is primary wavelength absorbed intensity

C is concentration, solve for

b is assumed 1cm for Beer’s law

is molar absorptivity

is efficiency constant

Direct correlation of intensity of light

Fluorescent Spectrophotometer (fluorometer)

o Want brightest light source

Requires less molecules in cuvette to cause emission

Highly sensitive

o Selecting for I: primary wavelength

o Primary filter: excitation wavelength

o Secondary filter: emission wavelength

Page 35: BLD 414 Notes

o Sources

Mercury: non-continuous intensity

Xenon: flicker; better

Compensated fluorometer: 2 detectors

Calculate mean of I: primary wavelength

intensity

Produces constant F intensity

Quantitative Fluorescent Assay

o Want to find concentration

o Lower limit of detection and upper limit of linearity

o RFU vs concentration

Relative fluorescent u_

Direct Assay

o Rare

o Interfering substances in the way

o Hit cuvette with primary, observe secondary

Fluorogenic Assay

o Reaction makes fluorescent compound

o A+R=F

o A-F bound

Fluorescent polarization

o Common

Page 36: BLD 414 Notes

o Basis

Di_crystal

Absorb light in specific plane of field

o

o Application: 2/22/12

Competitive-binding assay

Bind of antibody increases molecular weight, and slows

random Brownian motion

Increases chance of emission in same plane as

primary wavelength = only those detected

Antigen = variable, want to determine

Ag-F: created HMW, monitored

More Ag, less Ag-F (vice-versa)

Time-Delayed

o Extend duration of fluorescence with Europium (Eu) chelates

o Phosphorescing

Light Scattering Photometry

Basic theory

o Particulate must be greater than .2 micron to use light scatter

o Small particle: back-scatter

o Intermediate particle: side-scatter

o Large particle: forward-scatter

Turbidity

Page 37: BLD 414 Notes

o Only non-scattered light detected

o Compare light in to light detected

More particulates is less light detected

o Follows Beer’s Law

o Clinical application for CFU counter and PT times

Nephelometer

o Detects scattered light

o Do not use wavelength of light that can be absorbed by

particulate

o Minimize blank reading to improve signal:noise

Non-divergent light source

Angle of detection

Use angle that gives max intensity

o Clinical application for immunoassay and flow cytometer

PIN detector

Tiny barrier cell model with a circuit

Hydrodynamic focusing: physical property

Page 38: BLD 414 Notes

BLD 414 Notes 1/9/12 8:54 PM

2/24/12: Reflectance Photometry

Quantitate intensity of reflected light

Types

o Specular

Glare: changes reflection = bad

Not subject to change in intensity

o Diffuse

From matte surface

Instrumentation: reflectometer

o Blank pure porcelain as pure white: max intensity

o First blank with black AlO3 to zero light out

Reflectance: Beer’s Law

o R value = Rout/Rmax

o Linear relationship of density (Dr) vs. concentration

Luminescence

Assays: monitored response is generation of light (h)

o Chemiluminescence

A + Reag Product + h

More h = more A

Not all reactions produce light

Correct with H2O2

Page 39: BLD 414 Notes

A + R P + H2O2

Acridinium ester + H2O2 h

o Bioluminescence

Enzyme catalyzed

Fireflies

A + enzyme P + h

Enzyme: lucerferase

Depends on ATP and reduced luciferin =

Analyte: ATP

ATP + reduced luciferin (enzyme) ADP + oxidized

luciferin + h

Derivitization

o Immunoassay: ELISA

o Cannot use area under curve

Instrumentation: luminometer

o Has photo multiplier tube

o Monitor generation of light

2/27/12: Enzyme Kinetics

Correlate enzyme activity to concentration of substrate

Michaelis-Menton

o Hold enzyme concentration constant

Page 40: BLD 414 Notes

o Vary substrate concentration

Low substrate concentration: 1st order

Intermediate substrate concentration: pseudo 1st order

High substrate concentration: zero order

Reaches saturation point

Vmax

o If enzyme concentration changes, rate changes proportionally

o Use zero order, high substrate concentration

Gives linear relationship

o Km = substrate concentration at ½ Vmax

Affinity of substrate to enzyme

Low Km = high affinity; takes less substrate to saturate

enzyme

Determining enzyme concentration

o Kinetic assay

Zero order

Linear relationship: ideal

Product formed per unit time

Initial lag phase: substrate starts occupying enzyme

active site

Intermediate zero order/linear range phase: only use

this

Final bend over phase: substrate depletion

o Continuous monitor: lab instrument

Page 41: BLD 414 Notes

o Fixed period: more practical than continuous

Use 2 point slope: assumes constant reaction rate

T1 after lag: lower limit

T2 found by running up enzyme concentration until non-

linear relationship of product vs. time, then find

inflection of slightly less (20%) enzyme concentration

Units of enzyme activity

o IU: micromole/liter/minute

o Product formed: monitored by spectrophotometry – color

change

o EXAMPLE

1mL serum: specimen

Plus 4mL substrate containing buffer

DF – Dilution factor: 1/5 correlate to enzyme activity

T1 = 30sec, Abs = .1

T2 = 2min 30sec, Abs = .4

Change in absorbance: .3 in 2 minutes .15/min

Beers Law: .15/molar absorptivity (epsilon: 7500)

Gives mole/liter

Convert to micromole

Multiply by 5 due to DF

Answer: 100 IU

3/2/12: Osmometry

Page 42: BLD 414 Notes

Units

o Osmolarity: osmoles/liter

Depends on temperature

o Osmolality: osmoles/Kg solution

Temperature independent

o Osmoles: number of species in solution after disassociation

Colligative Properties

o Osmolarity increase, freezing point decrease

Add salt on icy roads

o Osmolarity increase, vapor pressure decrease

Add salt to boiling water

Vapor Pressure Osmometry

o Air-tight chamber

o Components

o Low osmolality: high vapor pressure, high humidity

o Underestimate osmolality when sample is organic

Freezing Point Depression Osmometry

o Freezing point of a solution decreases 1.86 deg C for

each osmole/Kg solution

o Components

Ethylene glycol absorbs heat of test tube sample

Cool sample below freezing point but still liquid

Page 43: BLD 414 Notes

Rapid freeze seeded by agitation with iron rod

Crystallization starts at seed then expands out

Heat of fusion warms solution to freezing point

Measure Osmolality

o COP: automated method

Quantitation

o Series of standards with CPU

3/12/12: Electrophoresis

Basis

o Isoelectric point: neutral charge protein; no migration

o Quaternary protein: positive charge

o Tertiary protein: negative charge

o Low pH: positive charged protein

o High pH: negative charged protein

o Anion (negative charged protein) pulled to anode

Rate of migration

o Charge on molecule: more absolute charge = faster

Charge to mass ratio

Lower: slower

Larger: faster

o Friction

Page 44: BLD 414 Notes

Viscosity

Weight: heavier moves slower

Physical barriers: restricting system/sieving/pores

Endo-osmosis

o At pH 8.6 all proteins have negative charge

o Barbital is positively charged at pH 8.6

o Protein migrates in opposite of expected direction due to

Barbital mobile-phase/friction

Heat: not good

o Associated with current = Amps (in constant voltage system)

o Increase heat, increase conductance

o Increase conductance, increase migration rate

o May denature proteins

o More heat more evaporation = decreased resistance, high

salt concentration, increased Amps, increased heat

o Wick effect

o Minimize heat

Constant current source: decrease voltage, decease

rate of migration

Refrigerate

Cut down evaporation: glass sandwich

Electrophoretic Apparatus

o Horizontal or vertical

o Tube: capillary

3/14/12: Support Media

o Paper: positive charge

Non-restricting: huge pores (cellulose channel)

May interact with proteins

High endo-osmosis

Trialing edge: bad

o Cellulose Acetate: neutral charge

Non-restricting, charge to mass ratio takes effect

Low endo-osmosis

No trailing edge