Biological indicators role in CSSD equipment qualification and monitoring Paulo Roberto Laranjeira...

download Biological indicators role in CSSD equipment qualification and monitoring Paulo Roberto Laranjeira 1.

If you can't read please download the document

description

Orion C.E. - Brazil Founded employees 12 equipment qualification teams Certified Kaye System maintenance and calibration services 3

Transcript of Biological indicators role in CSSD equipment qualification and monitoring Paulo Roberto Laranjeira...

Biological indicators role in CSSD equipment qualification and monitoring Paulo Roberto Laranjeira 1 Brazil Population (2010) Brazil So Paulo State Greater So Paulo City Quantity Hospitals9.000 Pharmeutical Industries823 2 Orion C.E. - Brazil Founded employees 12 equipment qualification teams Certified Kaye System maintenance and calibration services 3 Regulations and Standards RegulationsStandards By country; Establishes minimum qualification requirements: o Frequency; o Load; o Responsibilities. Requirements for load monitoring International and local; Qualification and re- qualification steps; Critical normative criteria; Informative criteria; Load monitoring. 4 Manufacturer IFUs Standards o ISO o ISO Standard 134C 4 minutes Cycle 5 First Step Protocols must be written and approved prior to equipment challenge o Qualification protocols o Indicators (biological, chemical and physical) protocols o CHANGE CONTROL 6 Thermal Destruction of Microorganisms Log Log Log Log Log Log Log ,1Log ,01Log ,001Log ,0001Log ,00001Log ,000001Log Theory: Fo o Tref = 121C o D = 1 minute o Z = 10C OverKill = 12 logs 7 Thermal Destruction of Microorganisms Log Log Log Log Log Log Log ,1Log ,01Log ,001Log ,0001Log ,00001Log ,000001Log Biological Indicator Real evidence Fo = 1,8 * (Log 1,2 x Log 1,0 x ) Fo C 9 Cycle 121C/15 min (exposure only) Calculated Lethality during Exposure TempoTmn.Let. Inst.Let. Acum.No 16:03:30100,00,00 1,20E+06 16:04:30105,00,00 1,20E+06 16:05:30110,00,00 1,20E+06 16:06:30115,00,00 1,20E+06 16:07:30121,01,00 3,34E+05 16:08:30121,01,002,009,29E+04 16:09:30121,01,003,002,59E+04 16:10:30121,01,004,007,19E+03 16:11:30121,01,005,002,00E+03 16:11:30121,01,006,005,57E+02 16:12:30121,01,007,001,55E+02 16:13:30121,01,008,004,31E+01 16:14:30121,01,009,001,20E+01 16:15:30121,01,0010,003,34E+00 16:16:30121,01,0011,009,29E-01 16:17:30121,01,0012,002,59E-01 16:18:30121,01,0013,007,19E-02 16:19:30121,01,0014,002,00E-02 16:20:30121,01,0015,005,57E-03 16:29:30115,00,0015,005,57E-03 16:30:30110,00,0015,005,57E-03 16:31:30105,00,0015,005,57E-03 16:32:30100,00,0015,005,57E-03 Fo (required) = 21 minutes 10 Kill x Overkill 11 12 Cycle: 121C/15 minutes 121C 100C 50C Start Lethality Calculated for Tmin. > 100C TimeTmin.Let. Inst. Let. Acum.No END BIs in the Qualification 13 134C/4 min Qualification Theory Based o D 121 = 1 minute o Z = 10C o Tref = 121C 14 134C/4 min Qualification Based on BI 121C o D 121 = 1,9 minutes o Z = 20C o Tref = 121C 15 134C/4 min Qualification Based on calculated D 134 value o D 134 = 0,43 minutes o Z = 20C o Tref = 134C 16 Integrating 134C/4 min cycles Theory Based o D 121 = 1 minute o Z = 10C o Tref = 121C Based on 121C o D 121 = 1,9 minutes o Z = 20C o Tref = 121C Based on calculated D 134 value o D 134 = 0,43 minutes o Z = 20C o Tref = 134C Lethality = 159 minutes Lethality = 25 minutes Lethality = 4,96 minutes 17 Monitoring 134C/4 min cycles 134C 125C 50C 121C Sal = 1 x Recent occurrence 19 20 134C/4 min 134C 125C 50C 121C Start Lethality Calculated for Tmin. > 100C TimeTmin.Let. Inst. Let. Acum.No End Sterilization Cycle 1 21 Sterilization Cycle 2 22 Equipment Control 134C 125C 3 minutes 23 Equipment Control joules 2727 joules Water Water + Steam Saturated Steam Super Heated Steam Sensor error Deviation Temp CLethality 1,026% 0,512% 0,37% 25 Monitoring best practices Biological Indicators o Monitor half cycles o Results depends on incubation period o Real evidence of sterilization o When placed inside a PCD will adequately monitor: NCG Superheated steam Chemical Indicators o Full cycle monitor o Immediate results o Requires: EN 285 ISO ISO o When placed inside a PCD will adequately monitor: NCG 26 Thank You! Eng. Paulo Roberto Laranjeira Celular e Whatsapp: Facebook: Pictures and data: Google e OrionCE 27