Beyond Conventional Stereotaxic Targeting: Using Multiplanar Computed Tomography, Optical Coherence Tomography and Intracranial/Extravascular Ultrasound of the Subarachnoid Space to Locate the Rhesus Vestibular Nerve for Single Unit Recording By Shiyao Dong A thesis submitted to Johns Hopkins University in conformity with the requirements for the degree of Master of Science in Engineering Baltimore, Maryland August, 2016 © 2016 Shiyao Dong All Rights Reserv

Transcript of Beyond Conventional Stereotaxic Targeting: Using ...

Page 1: Beyond Conventional Stereotaxic Targeting: Using ...

Beyond Conventional Stereotaxic Targeting:

Using Multiplanar Computed Tomography,

Optical Coherence Tomography and

Intracranial/Extravascular Ultrasound of the

Subarachnoid Space to Locate the Rhesus

Vestibular Nerve for Single Unit Recording

By

Shiyao Dong

A thesis submitted to Johns Hopkins University in conformity with the

requirements for the degree of Master of Science in Engineering

Baltimore, Maryland

August, 2016

© 2016 Shiyao Dong

All Rights Reserv

Page 2: Beyond Conventional Stereotaxic Targeting: Using ...

ii

Abstract

Single-neuron electrophysiologic recording of activity in the vestibular nerve and

vestibular nuclei of alert rhesus monkeys is a uniquely informative technique in the arsenal

of neurophysiologists who seek to elucidate adaptive neuronal signal processing that

mediate learning within vestibulo-ocular and vestibulo-spinal reflex pathways. The

traditional approach to targeting these structures with a recording microelectrode -

stereotactic guidance based on atlas coordinates - has changed little over the past half

century and is made difficult by the vestibular nerve’s mobility, mechanical compliance,

small caliber and relatively long distance from where a microelectrode enters the cranium

(typically at the base of a recording chamber surgically affixed to the parietal aspect of the

skull). The goal of the present study was to determine whether imaging techniques that

have recently gained prominence in clinical care might augment or replace the traditional

neurophysiologic method. Toward that end, we evaluated 3D multiplanar computed

tomography (3DCT), optical coherence tomography (OCT) and an intracranial/cisternal

adaptation of intravascular ultrasound (IVUS) as adjuncts to facilitate targeting the rhesus

vestibular nerve. Image guidance using CT scans acquired after recording chamber

implantation proved to be a simple and useful complement to traditional atlas-based

stereotaxis; however, atlas-based stereotaxis, 3DCT, OCT and IVUS offer complementary

advantages and disadvantages for targeting cranial nerves. Although OCT and IVUS

ultimately proved needlessly complex for our application, adaptation of those techniques

for intracranial/cisternal or intracranial imaging of cranial nerves, spinal nerves and other

structures adjacent to cerebrospinal fluid spaces may hold promise for intraoperative

guidance during minimal-access rhizotomy, biopsy and other neurosurgical procedures.

Page 3: Beyond Conventional Stereotaxic Targeting: Using ...

iii

Keywords: CT-guidance, vestibular nerve, stereotactic surgery, OCT, ultrasound

Thesis Review Committee:

Advisor:

Charles C. Della Santina, Ph.D. M.D.

Thesis Reviewer:

Chenkai Dai, Ph.D. M.D.

Kathleen Cullen, Ph.D.

Page 4: Beyond Conventional Stereotaxic Targeting: Using ...

iv

Acknowledgement

I really appreciate those people who help me a lot throughout my Master research,

especially members of Vestibular NeuroEngineering Laboratory. Without their help, this

thesis project wouldn’t be possible. I would like to thank Dr. Paul Bottomley, Dr. Jon Resar,

Dr. Caroline Garret, Dr. Yu Chen, Cardiac Cath lab at Johns Hopkins Hospital, Shashank

Sathyanarayana Hegde, Kelly Lane, and Nicole McIntosh for the help with equipment and

animals. I would like to thank Pengyu Ren for the help with animal surgery. I would like

to thank Dr. Chenkai Dai for the valuable advice and helpful guidance on my research

project. I would like thank Dr. Kathleen Cullen for reviewing my thesis.

Most of all, I’m really thankful to Dr. Charley Della Santina, my friendly, helpful

and responsible advisor, mentor and friend. He shared his experience and vision with me,

and provided me with essential skills and tools for research. His guidance helped me get

through lots of difficulties during my master research.

I would like to thank my family who taught me, loved me and supported me

throughout my life endeavors.

The project described was supported by NIH/NIDCD (R01-DC2390, R01DC9255)

and donors to the Johns Hopkins Vestibular NeuroEngineering Laboratory.

Page 5: Beyond Conventional Stereotaxic Targeting: Using ...

v

Table of contents

Abstract .............................................................................................................................. ii

Acknowledgement ............................................................................................................ iv

Table of contents ............................................................................................................... v

List of tables..................................................................................................................... vii

List of Figures ................................................................................................................. viii

Chapter 1 Introduction ................................................................................................. 1

1.1 Challenge in Single-Unit Recording in Rhesus Vestibular Nerve ................... 3

1.2 Current approaches to targeting a cranial nerve .............................................. 7

1.2.1 Conventional Stereotaxic Procedure .................................................... 8

1.2.2 Image-guided Stereotaxis ..................................................................... 9

1.3 Potential imaging techniques for nerve targeting .......................................... 12

1.3.1 Intravascular Ultrasound .................................................................... 13

1.3.2 Optical Coherence Tomography ........................................................ 15

1.4 Project Aims .................................................................................................. 20

Chapter 2 Post-operative CT Guidance for Single-unit Recording in Rhesus

Vestibular Nerve ............................................................................................................. 22

2.1 Experimental Subjects ................................................................................... 22

2.2 Materials and Methods .................................................................................. 22

2.2.1 Recording Chamber Implantation Surgical Procedures ..................... 23

2.2.2 CT Imaging and Analysis .................................................................. 26

2.3 Results............................................................................................................ 31

2.3.1 Measurement of Vestibular Nerve Location ...................................... 31

2.3.2 Experimental Validation and Errors Calibration ............................... 32

2.4 Discussion ...................................................................................................... 37

Chapter 3 Intracranial/Extravascular Application of OCT and IVUS for Imaging

the Subarachnoid Space ................................................................................................. 41

3.1 Experimental Subjects ................................................................................... 41

3.1.1 Phantom ............................................................................................. 41

3.1.2 Chinchilla Subject .............................................................................. 41

3.1.3 Rhesus Monkey Subjects ................................................................... 42

3.2 Surgical Procedures ....................................................................................... 42

3.2.1 Chinchilla Surgery ............................................................................. 42

Page 6: Beyond Conventional Stereotaxic Targeting: Using ...

vi

3.3 Intracranial/Extravascular Optical Coherence Tomography ......................... 43

3.3.1 Method ............................................................................................... 43

3.3.2 Results ................................................................................................ 46

3.4 Intracranial/Extravascular Ultrasound Imaging of the Subarachnoid Space . 52

3.4.1 Method ............................................................................................... 52

3.4.2 Results ................................................................................................ 52

3.5 Discussion ...................................................................................................... 55

Chapter 4 Conclusion ................................................................................................. 58

4.1 Post-Operative CT Guidance ......................................................................... 58

4.2 Intracranial Imaging of the Subarachnoid Space Using OCT and IVUS ...... 58

4.3 Future Directions ........................................................................................... 59

Reference List .................................................................................................................. 61

Curriculum Vitae ............................................................................................................ 68

Page 7: Beyond Conventional Stereotaxic Targeting: Using ...

vii

List of tables

Table 1-1 Comparison of OCT and IVUS ........................................................................ 17

Table 2-1. General information of experimental subjects ................................................ 22

Table 2-2. Location of porus acusticus in CT stereotaxic coordinate............................... 36

Table 2-3. Location of porus acusticus in CT chamber coordinate .................................. 36

Table 3-1.Comparison of OCT and IC/EVUS .................................................................. 55

Page 8: Beyond Conventional Stereotaxic Targeting: Using ...

viii

List of Figures

Figure 1-1. Rhesus Monkey Brain Atlas (Interaural -6.00 mm) ......................................... 5

Figure 1-2. Rhesus Monkey Brain Atlas (Interaural -5.55 mm) ......................................... 6

Figure 1-3.MRI-guided stereotaxic framework overview ................................................ 10

Figure 1-4. IVUS images of aortic dissection ................................................................... 14

Figure 1-5 Basic schematic of OCT working principle .................................................... 16

Figure 1-6. A camera image of the OCT probe on top of the human basal ganglia ......... 18

Figure 1-7. OCT images of a tympanic membrane (TM) ................................................. 19

Figure 2-1. Stereotaxic frame parameters ......................................................................... 25

Figure 2-2. CT 3D reconstruction in stereotaxic head coordinates .................................. 28

Figure 2-3. CT 3D reconstruction in recording chamber coordinate (left ear) ................. 29

Figure 2-4. XY Micromanipulator for single-unit recording ............................................ 30

Figure 2-5. Region of interest in CT 3D reconstruction ................................................... 33

Figure 2-6. Measurement of inner acoustic opening location in CT stereotaxic coordinate

(Animal ID: RhF234D) ..................................................................................................... 34

Figure 2-7. Measurement of inner acoustic opening location in CT chamber coordinate

(Animal ID: RhF234D) ..................................................................................................... 35

Figure 2-8.Location of porus acusticus in transverse plane of chamber coordinate ......... 38

Figure 3-1. Placement of OCT catheter inside penne embedded in gelatin ...................... 44

Figure 3-2.Simulation of placing OCT catheter through the floor of bulla in chinchilla

right ear under camera view .............................................................................................. 45

Figure 3-3. OCT image of a penne in the gelatin phantom .............................................. 47

Figure 3-4.OCT image of inner acoustic canal in chinchilla ............................................ 48

Page 9: Beyond Conventional Stereotaxic Targeting: Using ...

ix

Figure 3-5.Location of OCT catheter placement into the internal auditory canal and

cerebellopontine angle of a chinchilla placed just after euthanasia and then dissected after

OCT image acquisition. .................................................................................................... 49

Figure 3-6.Rhesus monkey inner acoustic opening on left side ....................................... 50

Figure 3-7.A slice of OCT side-view scan when catheter is placed in CSF space in the

post-mortem rhesus monkey on left side .......................................................................... 51

Figure 3-8.IC/EVUS side-view image of penne embedded in gelatin phantom ............... 53

Figure 3-9. Images of subarcuate parafloccular recess obtained using

intracranial/extravascular ultrasound in the cerebellopontine angle of a rhesus monkey

specimen. .......................................................................................................................... 54

Figure 3-10.Comparison of OCT and IC/EVUS in gelatin phantom ................................ 57

Page 10: Beyond Conventional Stereotaxic Targeting: Using ...

1

Chapter 1 Introduction

In normal individuals, the two vestibular labyrinths provide the central nervous

system (CNS) with sensation of head rotation and linear accelerations due to both gravity

and translational motion. These sensory inputs drive compensatory reflexes that maintain

posture and stabilize gaze so as to maximize visual acuity during head movement. Bilateral

loss of vestibular sensation due to ototoxic injury or other insults to both labyrinths is

disabling. Patients suffer from chronic disequilibrium, increased frequency of falls and

unstable vision during head movements typical of common daily activities like walking

and driving.[1-3] Most individuals with mild or moderate loss eventually compensate

through rehabilitative exercises that augment residual function, but those with profound

loss who fail to compensate have no good treatments options. There are approximately

64000 U.S adults who experience chronic, symptomatic profound loss of bilateral sensation

with no solution for relief.[4, 5] No clinical-approved treatment beyond rehabilitation

exists for these patients. Vestibular implants, which are neuroelectronic vestibular

prostheses designed to restore motion-modulated vestibular nerve activity, should provide

a solution to people with bilateral vestibular deficiency who experience chronic dizziness

even after rehabilitation, by restoring sensation of head motion and gravitational

orientation that normally drive vestibular reflexes.[6-10]

Normally, the eyes rotate opposite the direction of head rotation in order to stabilize

images on retinae. The angular vestibulo-ocular reflex (aVOR) drives the compensatory

eye rotation. Sensory input to the aVOR is provided by three mutually orthogonal

semicircular canals (SCC) in each inner ear’s vestibular labyrinth. In each SCC’s ampullary

Page 11: Beyond Conventional Stereotaxic Targeting: Using ...

2

nerve, the firing rates of vestibular afferents are modulated by the component of head

angular velocity about that SCC’s axis.[11-14] By measuring the resulting eye movement

in 3D, relative excitation of the ampullary nerves driving the reflex can be estimated.[15-

19] Suzuki, Cohen et al. demonstrated that delivering pulses of electrical stimulation to the

ampullary nerve via wire electrodes can elicit an aVOR about axes of rotation similar to

those elicited by the hydrodynamic excitation of the individual canals.[13, 20-26] The

vestibular prosthesis modulates electrical stimulation of surviving vestibular afferents

based on motion sensor input. A vestibular prosthesis directly stimulates vestibular afferent

fibers, which synapse on regions of the CNS that generate vestibular-mediated reflexes.

Thus, the vestibular prosthesis can partly replace function of the lost labyrinth with an

electronic solution.

The Vestibular NeuroEngineering Laboratory (VNEL) at Johns Hopkins has done

significant work with the Johns Hopkins Multichannel Vestibular Prosthesis (MVP) from

its initial development,[6, 27, 28] including reduction in size and power consumption,[29]

reduction in channel cross-talk,[30] developing an understanding of the effect of stimulus

parameters,[31] and conducting chronic stimulation experiments.[32, 33] After initial

development in small mammals, VNEL translated MVP technology to rhesus monkeys,

demonstrating similarly favorable outcomes. VNEL recently initiated a first-in-human

clinical trial of a vestibular implant based on a modified cochlear implant.[34]

One of the key remaining challenges to vestibular prosthesis development is

vestibulo-ocular reflex (VOR) misalignment due to current spread to vestibular afferent

fibers beyond those in the targeted nerve branch.[30] When current spread incurs spurious

excitation of adjacent vestibular nerve branches, the prosthesis cannot selectively stimulate

Page 12: Beyond Conventional Stereotaxic Targeting: Using ...

3

each ampullary nerve. Higher current stimuli can increase the aVOR magnitude, but

meanwhile the eye rotation axis increasingly deviates from the ideal because of spurious

stimulation of other vestibular nerve branches.[30]

The 3D VOR axis is a good estimate of relative activity in the 3 ampullary nerves of

the implanted ear, but that 3-value vector is a greatly simplified estimate of neuron activity.

Single-unit recording, a more direct assay of measuring single-unit neuronal activity, is

needed to guide further optimization of designs that reduce VOR misalignment. By

combining single-unit recording in rhesus vestibular nerve during MVP stimulation with

3D VOR data, we can better understand current spread and refine a rigorous biophysical

model of neuronal activation by current spread.

1.1 Challenge in Single-Unit Recording in Rhesus Vestibular Nerve

Accurate targeting is required for single-unit recording in the rhesus vestibular nerve;

however, targeting the vestibular nerve for electrophysiology is a challenging task. Also

known as the eighth cranial nerve, the vestibulocochlear nerve splits into two large

divisions: the cochlear nerve and the vestibular nerve. The peripheral parts of the eighth

nerve travel a short distance medially from their respective neurosensory epithelia to

ganglia (i.e., clusters of nerve cell bodies near the vestibular labyrinth [Scarpa’s ganglion]

and within the central spiraling modiolus of the cochlea [the spiral ganglion]). From there,

the central part of the nerve travels through the internal auditory meatus (also called the

internal auditory canal, or IAC) alongside the facial nerve. Departing the IAC at the porus

acusticus (also called the internal acoustic meatus), the eighth nerve traverses a

cerebrospinal fluid filled space in the cerebellopontine angle before entering the brainstem

Page 13: Beyond Conventional Stereotaxic Targeting: Using ...

4

at the junction of the pons and medulla lateral to the facial nerve. The diameter of

vestibulocochlear nerve is about 0.9 mm in rhesus and 1 mm in human. [35-38] The length

of the vestibulocochlear nerve, from the glial-Schwann junction, where cochlear nerve and

vestibular nerve join together within the IAC, to the brainstem, is 10-13 mm in human.

Page 14: Beyond Conventional Stereotaxic Targeting: Using ...

5

Figure 1-1. Rhesus Monkey Brain Atlas (Interaural -6.00 mm)

Before entering brainstem, the rhesus vestibulocochlear nerve is about 45 mm inferior to bregma in

stereotactic coordinates. Adapted from: Paxinos, G., X.-F. Huang, and A.W. Toga, The rhesus monkey

brain in stereotaxic coordinates. 2000.

Page 15: Beyond Conventional Stereotaxic Targeting: Using ...

6

Figure 1-2. Rhesus Monkey Brain Atlas (Interaural -5.55 mm)

After entering brainstem, the rhesus vestibular nerve is about 44 mm inferior to bregma in stereotactic

coordinates. Adapted from: Paxinos, G., X.-F. Huang, and A.W. Toga, The rhesus monkey brain in

stereotaxic coordinates. 2000.

Page 16: Beyond Conventional Stereotaxic Targeting: Using ...

7

and 7-10 mm in females. Figure 1-1 shows a section of rhesus monkey brain from a

stereotaxic atlas by Paxinos, Huang and Toga,[39] illustrating the position of the

vestibulocochlear nerve before entering the brainstem. Figure 1-2 shows another section,

illustrating the position of vestibulocochlear nerve after entering brainstem. Rhesus

monkeys have similar size of vestibulocochlear nerve, the average diameter of which is

about 0.9 mm.[38] Only the portion of vestibulocochlear nerve that exits the IAC before

entering brainstem is available for placing recording electrode in, which is shorter than 10

mm. The superior-inferior distance from vestibulocochlear nerve to bregma (confluence of

sutures of frontal and parietal bones) according to the Paxinos et al. stereotaxic atlas is

about 44-45 mm.[39] Thus, the small size of the vestibular nerve, long distance to the nerve

from parietal cranium surface (where a recording electrode must enter the cranium in our

preparation), and inter-subject difference among rhesus monkeys combine to make

targeting the vestibular nerve difficult.

1.2 Current approaches to targeting a cranial nerve

Conventional stereotaxic guidance based on atlas coordinates has long been

successfully employed by investigators experienced in that technique to locate small targets

in brain and to perform action such as injection, stimulation and implantation, etc. The

development of computed tomography (CT) and more recently magnetic resonance

imaging (MRI) has greatly facilitated the conventional stereotactic procedure.

Page 17: Beyond Conventional Stereotaxic Targeting: Using ...

8

1.2.1 Conventional Stereotaxic Procedure

Conventional stereotaxic surgery includes three main components: (1) a stereotaxic

planning system including atlas and coordinate calculator, (2) a stereotaxic apparatus and

(3) a method for stereotaxic localization. Each available rhesus monkey brain atlas is based

one single animal,[39-41] which precludes correction for inter-subject variability, apart

from simply scaling all dimensions according to the relative length of the distance between

the left and right bony ear canal entrances (as determined by ear bar settings on a

stereotactic frame). In a typical stereotactic atlas, the three dimensions are: lateral-medial,

anterior-posterior and superior-inferior. Rhesus monkey brain atlases often use the bony

external auditory meatus, the inferior orbital ridges, and the bregma as landmarks. The

simple orthogonal stereotaxic apparatus uses a set of three Cartesian coordinates (x, y and

z) in an orthogonal frame of reference. Reid’s horizontal (Z) plane passes through the

interaural axis and the lowest point of the cephalic edge of each infraorbital ridge. Reid’s

coronal (X) plane is perpendicular to the Z plane and contains the interaural axis. Reid’s

midsagittal (Y) plane is perpendicular to the other two planes and lies along the head’s

plane of symmetry. Positive X, Y and Z are anterior, left and superior. Guide-bars in x, y

and z directions, fitted with high precision Vernier scales allow a neurophysiologist or

neurosurgeon to position the point of probe or electrode inside the brain, at the calculated

coordinates for the target structure, through a small craniotomy.

Single-unit recording of electrophysiological activity from individual afferent

neurons within the rhesus vestibular nerve has long been traditionally guided by atlas-based

stereotaxic planning.[42-44] To determine how deep the electrode should be inserted to get

the vestibular nerve, experienced researchers in field of vestibular study listen to the

Page 18: Beyond Conventional Stereotaxic Targeting: Using ...

9

characteristic sounds of neural activity when fed into an audio monitor.[45-47] They

approach the vestibular nerve through the floccular complex, which is identified by its eye

movement-related neural activity. As the recording microelectrode is advanced inferiorly

via a trajectory starting at posterosuperolateral parietal bone and passing through the

flocculus, entry into the vestibular nerve is typically preceded by a silence, indicating that

the electrode has left the cerebellum and is passing through cerebrospinal fluid in the

subarachnoid space of the cerebellopontine angle. After pushing all the way through and

then exiting the nerve, the microelectrode may travel a short distance (typically less than a

few mm) before hitting mater over the petrous part of the temporal bone, and event that

can blunt or otherwise damage the electrode tip and which is sometimes signified by abrupt

onset of 60 Hz noise in the output of the microelectrode preamplifier.

1.2.2 Image-guided Stereotaxis

In 1978, Russel Brown, an American physician, invented a device known as the N-

localizer.[48-55] The N-localizer enables guidance of stereotaxic surgery or radiosurgery

using tomographic images that are obtained via computed tomography (CT),[56-59]

magnetic resonance imaging (MRI),[60-62] or positron emission tomography (PET).[63,

64] Surgical precision is significantly improved because CT, MRI and PET permit accurate

visualization of intracranial anatomic detail. These imaging techniques can display

intracranial anatomy in 3-D images, allowing surgeons to register a patient’s cranial image

space to stereotaxic physical space with the help of external markers before surgery.

Surgeons now commonly use intracranial images preoperatively for stereotaxic planning

or during surgical interventions as guidance, and several commercial systems for image-

guided navigation during sinus and intracranial surgery are available.

Page 19: Beyond Conventional Stereotaxic Targeting: Using ...

10

Functional magnetic resonance imaging (fMRI) allows localization of brain regions

specialized for different perceptual and higher cognitive functions; however, the spatial

resolution with which fMRI can identify different levels of physiologic activity is orders

of magnitude more coarse than is possible with single-unit (i.e., individual neuron) electro-

physiological recording of neural activity using tungsten needle microelectrodes or glass

micropipettes. Ohayon et al. designed a novel framework for MRI-stereotaxic registration

and chamber placement for precise electrode guidance to recording site defined in MRI

space.[65] Figure 1-3 shows overview of the frame. The framework allows positioning a

recording chamber, according to pre-surgery planning, and is not limited to

Figure 1-3.MRI-guided stereotaxic framework overview

A region of interest is selected for targeting and a virtual chamber is placed. MRI scan is performed after

markers are attached. The software calculates the parameters to align the manipulators. Adapted

from[65].

Page 20: Beyond Conventional Stereotaxic Targeting: Using ...

11

vertical penetrations. It also permits implantation of chambers while the animal is simply

head fixed in the primate chair, detached from the stereotaxic frame.

CT-guided stereotaxic targeting has been used in humans for implantation of deep

brain stimulation electrodes and for craniofacial surgery.[66, 67] Eggers et al. proposed

intraoperative CT-guidance for craniofacial surgery with a fully automated registration,

which allows surgeons to navigate the operation without delay of patient-to-image

registration.[67] Neurosurgeons and neurotologists have increasingly embraced

preoperative virtual planning and intraoperative navigation to reduce the risk of

complications and to increase the efficiency and confidence with which they can work near

critical anatomic structures.[68] Similarly, neuroscience researchers should benefit from

CT-aided stereotaxis to refine their approaches to recording neurophysiologic activity of

small, hard-to-reach intracranial structures like the vestibular nerve.

Although experienced surgeons can complete procedures without image guidance,

the ability to precisely target and/or identify anatomic points of interest is cost-effective in

scenarios that require the surgeon to target a hard-to-find structure while avoiding large

blood vessels and other vital or fragile structures.[69] Similarly, although experienced

neurophysiologists can find their electrophysiologic targets through a combination of

frame-based stereotaxis, systematic sampling of a 3D grid volume encompassing the

expected location of the nerve of interest, directed search relying on the location of neurons

with recognizable firing patterns, patience and luck, image-guidance should provide a

useful adjunct to traditional stereotactic techniques.

Page 21: Beyond Conventional Stereotaxic Targeting: Using ...

12

Realizing this, Watanabe et al. developed a new system which introduces the idea

of CT-guided stereotaxis into conventional open neurosurgery[70]. Using metal markers,

they coregistered a preoperative CT imaging data set (for high resolution bone anatomy)

with patient’s head. They reported that this system makes it possible for a surgeon to

identify the operating site by obtaining real-time positional information in surgery

translated back onto CT images. However, for targeting the rhesus vestibular nerve, that

approach might suffer from errors arising during recording chamber implantation surgery.

CT scans obtained before chamber implantation may not be sufficient to accurately target

the nerve in each recording experiment after surgery. A more efficient solution involves

first surgically attaching a rigid recording chamber to the skull (using traditional

stereotactic targeting), then performing a high-resolution CT and 3D reconstructions along

mutually orthogonal planes that either contain or are perpendicular to the recording

chamber’s axis. We evaluated that approach in the present study.

1.3 Potential imaging techniques for nerve targeting

In addition to CT and MRI guidance, biomedical imaging techniques such as

intravascular ultrasound (IVUS) and optical coherence tomography (OCT) can image soft

tissue with better resolution, which could possibly help visualize vestibular nerve

intraoperatively or even directly observe the microelectrode as it is advanced into the nerve.

Stereotaxic procedures that require insertion of needle-type instrument into the brain serve

crucial roles in neurosurgery. The procedures are employed as aids to diagnosis in

management of various medical conditions including biopsy of tumors, assessment of

vessel wall plaques, and placement of electrodes that are used for signal recording and deep

brain stimulation (DBS).[71-73] Catheter-based IVUS and OCT have both established

Page 22: Beyond Conventional Stereotaxic Targeting: Using ...

13

roles in interventional cardiology as visualization aids during stent implantation. The small

size of IVUS and OCT catheters and the high resolution of these imaging modalities gives

these two intravascular imaging technologies potential to be applied in the subarachnoid

space to provide guidance for targeting cranial and spinal nerves.

1.3.1 Intravascular Ultrasound

The concept of intravascular ultrasound was first introduced by Born et al. in

1972.[74] IVUS is a catheter based system that allows physicians to acquire images of

diseased vessels from inside the artery. In the late 1980s, commercially available IVUS

catheter-based probes were used in both animals and humans.[75] Intravascular ultrasound

imaging, working in the range of 20-45 MHz, is the most commonly used intravascular

imaging modality (after dye-contrasted cine or flat panel CT fluoroscopic angiography) for

diagnosing coronary artery diseases or coronary heart diseases.[76] Imaging through blood

with large penetration (>5 mm) makes IVUS able to facilitate vessel modeling and plaque

morphology.[77, 78] The use of IVUS is currently being incorporated into several

modalities that will offer more real-time information in both the aorta and the treatment of

peripheral vascular disease.[76, 79-83] Figure 1-4 illustrates IVUS images along the

imaging catheter inside the subclavian artery.[81]

IVUS can provide a cross-sectional image of the vessel or other tissue within the

range, with resolution of 100-200 um and the imaging depth around 10 mm [81, 84]. These

features endow IVUS with potential to be applied outside the vasculature, deployed via a

recording microelectrode’s guide tube trocar through brain tissue and into cerebrospinal

fluid spaces, where it can be used for image guidance as an aid for nerve targeting in single-

Page 23: Beyond Conventional Stereotaxic Targeting: Using ...

14

unit recording or intracranial neurosurgery. An analogous approach might be used in the

cranium, in which IVUS might be renamed intracranial/extravascular ultrasound

(IC/EVUS).

Figure 1-4. IVUS images of aortic dissection

In aortic dissection, entry tear is the initial intimal tear that causes the blood to flow between the layers of

the wall of the aorta. These images, adapted from [81], show that IVUS can images a vessel wall with

sufficient resolution to identify an entry tear.

Page 24: Beyond Conventional Stereotaxic Targeting: Using ...

15

1.3.2 Optical Coherence Tomography

Optical coherence tomography (OCT) is a recently developed biomedical imaging

that has accelerated the research in the fields of medicine and biology. The capability of

OCT to perform high resolution, cross-sectional tomographic imaging of microstructure in

biological tissues has led this technology to be applied to surgical guidance and medical

diagnostics. Compared with IVUS, OCT provides higher resolution of around 10-20

um.[84-86] Like IVUS, needles and probes have been developed to integrate OCT into

intravascular catheters.

In essence, OCT probes detect and report the occurrence of reflection backward

along a beam path extending like a laser beam from and back to the OCT sensor.

Interferometry allows detection of these reflections and determination of how far each

reflection-generating boundary is from the sensor (Figure 1-5). The light source has a

bandwidth in the near-infrared spectrum with central wavelengths ranging from 1,250 to

1,350 nm. Tissue penetration, and therefore the depth over which OCT images tissue, is

limited to 1-3 mm.[87] Rapid, repetitive and systematic reorientation of the OCT sensor

allows a system to scan either a slice of tissue perpendicular to the catheter (side-viewing

OCT) or a conic section of tissue concentric with the catheter (forward-viewing OCT).

Side-view imaging OCT probes generate a 2D image of a disk of tissue

perpendicular to the axis of the OCT probe’s introducing catheter. A side-viewing system

consists of an optical fiber, a focusing lens and a prism or mirror for lateral deflection of

light. The entire apparatus spins within a surrounding catheter, allowing the OCT sensor’s

single line of sight to be scanned like a submarine periscope or an air traffic control radar

Page 25: Beyond Conventional Stereotaxic Targeting: Using ...

16

Figure 1-5 Basic schematic of OCT working principle

system. A constant-velocity “pull-back” of the rotating sensor along the axis of the catheter

allows one to sample many disc slices and then reconstruct them to image a 3D cylindrical

volume.[88, 89] In addition to intravascular imaging of vessel walls, side-view imaging

probes are used to image gland ducts and other hollow organs.[87]

Compared with side imaging probes, forward imaging probes are more challenging

to construct, because 3D image construction cannot easily be done by back-forth “pull-

back” translation. To acquire the 3D image of object, for example, the scanning is realized

by a piezoelectric transducer, galvanometer scanners or a microelectromechanical system.

[90-92] The scanning device in a forward imaging endoscope is usually placed at the distal

end of the system. That makes it very difficult to maintain the diameter small and have a

large working distance (WD) and broad field of view (FOV). If these technical challenges

can be overcome, Forward-viewing OCT seems especially well-suited to providing useful

image guidance during intracranial stereotaxic procedures, because the guide tube through

Page 26: Beyond Conventional Stereotaxic Targeting: Using ...

17

which the OCT imaging system is introduced is nominally pointing at the target one wants

to see.

Forward-scanning OCT (also called endoscopic OCT, or EOCT) has been applied

to detect diseases of the tympanic membrane and the middle ear, such as otitis media or

cholesteatoma.[94-96] Burkhardt et al. designed a 3D forward imaging EOCT at a large

WD with a large FOV and acquired images of tympanic membrane (Figure 1-7).[97]

Although a forward-scanning needle probe can be used to provide real-time feedback along

the line of sight as a microelectrode guide tube is inserted, the forward-viewing OCT

systems described so far do not provide as much field of view as side-view imaging systems.

Chen et al. at University of Maryland, College Park designed a forward-imaging needle-

type OCT probe for stereotaxic procedures intended to overcome this problem by using a

gradient-index (GRIN) rod lens, which has a gradual variation of refractive index and

provides variable length focus without the weight, size and fragility of a traditional lens,

(Figure 1-6).[93].

Table 1-1 summarizes performance parameters for IVUS and endoscopic OCT,

which are promising techniques for real-time guidance for intracranial surgery.[87]

Modality Axial Resolution Lateral Resolution Penetration depth Smallest

Catheter size

OCT 12-18 um 20-90 um 1.5-3 mm ~ 1 mm

IVUS 150-200 um 150-300 um 4-8 mm ~ 1 mm

Table 1-1 Comparison of OCT and IVUS

Page 27: Beyond Conventional Stereotaxic Targeting: Using ...

18

Figure 1-6. A camera image of the OCT probe on top of the human basal ganglia

The top yellow bar shows the full track of OCT reconstruction along the probe insertion direction. Labels:

extreme capsule (ex), claustrum (Cl), external capsule (ec), putamen (PUT), lateral medullary lamina (lml),

globus pallidus externa (GPe) and globus pallidus interna (GPi).Adapted from [93].

Page 28: Beyond Conventional Stereotaxic Targeting: Using ...

19

Figure 1-7. OCT images of a tympanic membrane (TM)

(A) The red arrow points at the umbo, the deepest point of the TM. (B) Manubrium of malleus (MM). (C)

Top view OCT image of the TM. (D) Top view camera picture, the umbo (*) and the MM (**).(E) 3-D OCT

image of the TM. Scale bars:1 mm. Adapted from [97].

Page 29: Beyond Conventional Stereotaxic Targeting: Using ...

20

1.4 Project Aims

Although experienced neuroscience researchers can successfully target the rhesus

vestibular nerve for single-unit recording using traditional atlas-based stereotaxis, finding

the nerve can be a challenging and laborious task that requires weeks of months of hunting

with a microelectrode, systematically making passes through a grid of tracks surrounding

the trajectory estimated using skull landmarks and a stereotactic atlas from a different

animal. Differences between atlas and experimental subjects, small size of the nerve and

long distance to the nerve from craniotomy combine to make it take long time targeting the

nerve before recording electrophysiological activity. Thus, there is a need for a guidance

method to help locate the nerve more efficiently.

The aim of this project was to evaluate a method of image guidance based on post-

operative CT scans (obtained after surgical placement of a recording chamber on the

parietal calvarium) in searching for vestibular afferents that:

1. individualizes targeting for the specific subject under study, rather than relying

on an atlas created from a different specimen;

2. provides direct guidance to help calibrate the placement of recording

microelectrode; and is

3. accessible to researchers with less experienced in stereotaxic procedures.

Although our focus was on image guidance in the form of CT 3D reconstructions,

which relies on technology now widely available in most academic medical centers, we

also explored catheter-based imaging techniques including intravascular ultrasound (IVUS)

Page 30: Beyond Conventional Stereotaxic Targeting: Using ...

21

and optical coherence tomography (OCT) to facilitate targeting the vestibular nerve more

precisely following the CT guidance.

Page 31: Beyond Conventional Stereotaxic Targeting: Using ...

22

Chapter 2 Post-operative CT Guidance for Single-unit

Recording in Rhesus Vestibular Nerve

We performed multiplanar CT scans on rhesus monkeys after recording chamber

implantation. Locations of the internal acoustic opening for both sides, where the vestibular

nerve bundle leaves the internal auditory canal to traverse the cerebellopontine angle on its

way from the inner ear to the brainstem, were measured with respect to recording chamber

coordinates for recording navigation and in stereotaxic coordinate for comparison. Details

of the experiment procedure are described in respective sections.

2.1 Experimental Subjects

Five rhesus monkeys were used for experiments, which were performed in

accordance with a protocol approved by the Johns Hopkins Animal Care and Use

Committee. Details about the five rhesus monkeys are shown in Table 2-1.

No. Animal ID Weight (kg) Sex Implanted Ears IVUS test OCT test

1 RhF234D 5.2 F Both No No

2 RhF247E 5.3 F Both No No

3 RhM46xH 4.5 M Both No No

4 RhF4617J 6.0 F Both Yes(post-

mortem)

Yes(post-

mortem)

5 RhM54AM 6.0 M Both No No

Table 2-1. General information of experimental subjects

2.2 Materials and Methods

All experiments were performed in accordance with protocols approved by the

Johns Hopkins Animal Care and Use Committee, which is accredited by the Association

Page 32: Beyond Conventional Stereotaxic Targeting: Using ...

23

for the Assessment and Accreditation of Laboratory Animal Care (AAALAC) International

and consistent with European Community Directive 86/609/EEC.

2.2.1 Recording Chamber Implantation Surgical Procedures

Five rhesus monkeys weighting 4-7kg were used for evaluating CT guidance to

facilitating single unit recording from vestibular nerve afferent neurons. Surgical

procedures were carried out under strict sterile conditions. Animals were sedated with

ketamine (0.2 mg/kg), treated with prophylactic IV cefazolin, endotracheally intubated and

maintained under deep anesthesia via inhalation of 1.5-5 % isoflurane in oxygen. Surgical

preparation was done in stages so as to limit the operative time and shorten the recovery

period for any single anesthetic exposure. In the first surgery, a head-restraining cap was

permanently affixed on the calvarium using dental acrylic affixed to multiple titanium bone

screws.

Before chamber implantation, we positioned a skull from a monkey of body size

similar to the animal on which we planned to operate in the stereotaxic frame, securing it

in place using ear bars, an incisor bar and a nose clamp. We then positioned a chamber

holder so that its axis passed from ipsilateral parietal cortex through the cerebellum to the

porus acusticus (i.e., the medial opening of the internal auditory canal) while avoiding

major blood vessels (Figure 2-1). Stereotaxic manipulator parameters were recorded for

later use during chamber implantation surgery.

We then positioned the animal to be studied on the stereotactic frame. Using the

previously determined settings, we positioned a 1.9 cm diameter titanium recording

chamber over the parietal calvarium and marked its location on the scalp and, after skin

Page 33: Beyond Conventional Stereotaxic Targeting: Using ...

24

incision and elevation, on the skull. A ~1 cm diameter craniotomy was drilled within the

region of calvarium encircled by the mark, and the chamber was cemented in place using

dental cement (3M ESPE ProTemp®II or OrthoJet®). A hairless, split-thickness skin graft

was harvested, meshed and applied to the exposed dura, then held in place using antibiotic-

impregnated GelFoam® and sterile packing.[98] The packing was removed and replaced

every other day starting 7 days post-op and continuing until the dura was completely

covered by skin. This technique can yield a clean, dry, easy-to-maintain recording chamber

floor that is almost entirely free of granulation tissue, significantly reducing the risk of

infection, blood loss, CSF leakage and animal distress that might otherwise occur with

daily debridement. The skin grafts act as a biologic dressing, sealing itself closed after

guide tube removal.

A second recording chamber was placed using the same techniques, typically on

the same day and in a position approximately at the mirror image across the midsagittal

plane from the first chamber.

Page 34: Beyond Conventional Stereotaxic Targeting: Using ...

25

Figure 2-1. Stereotaxic frame parameters

Stereotaxic setting parameters: (a) XY-slide for right and left, Medial-Lateral slide for right and left. (b)

Vertical azimuth for right, horizontal azimuth for right. (c) Vertical azimuth for left, horizontal azimuth for

left and upper slide.

Page 35: Beyond Conventional Stereotaxic Targeting: Using ...

26

2.2.2 CT Imaging and Analysis

Monkeys maintained under propofol IV infusion anesthesia and endotracheally

intubated were imaged in a high-resolution CT scanner (Toshiba Aquilion One). A guide

tube containing a tungsten recording microelectrode was inserted along the axis of each

recording chamber, advanced through skin graft, dura, parietal/occipital cortex, tentorium,

and cerebellum toward the presumed location of the porus acusticus, then left in place

temporarily while a CT scan was performed using a bone algorithm, step scan mode, a 120

kV source collimated to 0.5-mm slice thickness at full-width half maximum, and a (120

mm)2 region of interest imaged on a 512 x 512 voxel/slice matrix, creating voxels of 0.4 x

0.4 x 0.3 mm.

To define the location of the porus acusticus, we used two different coordinate

systems (Figure 2-2 and Figure 2-3). We used stereotactic coordinates using Reid’s planes.

Reid’s horizontal (Z) plane passes through the interaural axis and the lowest point of the

cephalic edge of each infraorbital ridge. Reid’s coronal (X) plane is perpendicular to the Z

plane and contains the interaural axis. Reid’s midsagittal (Y) plane is perpendicular to the

other two planes and lies along the head’s plane of symmetry. Positive X, Y and Z are

anterior, left and superior. To define locations with respect to each of the two cylindrical

recording chambers, we defined a chamber coordinate system resulting in which the Zc

axis for a given chamber is the axis of that chamber, the Yc axis is perpendicular to Zc of

the same chamber and intersects the Zc axis of the other chamber, and Xc is perpendicular

Page 36: Beyond Conventional Stereotaxic Targeting: Using ...

27

to both Zc and Yc. Positive Xc, Yc and Zc are approximately anterior, left and superior

with respect to the skull and the stereotactic coordinate system.

In single-unit recording experiments, we placed an X-Y micromanipulator (Figure

2-4) on the chamber with its X, Y-axis in alignment with the chamber coordinate. Based

on the location of target measured in CT chamber coordinate, we changed

micromanipulator settings and passed tungsten electrodes through a guide tube to target

the ipsilateral porus acusticus.

Page 37: Beyond Conventional Stereotaxic Targeting: Using ...

28

Figure 2-2. CT 3D reconstruction in stereotaxic head coordinates

Page 38: Beyond Conventional Stereotaxic Targeting: Using ...

29

Figure 2-3. CT 3D reconstruction in recording chamber coordinate (left ear)

Page 39: Beyond Conventional Stereotaxic Targeting: Using ...

30

Figure 2-4. XY Micromanipulator for single-unit recording

Page 40: Beyond Conventional Stereotaxic Targeting: Using ...

31

2.3 Results

2.3.1 Measurement of Vestibular Nerve Location

As described previously, the vestibular nerve the petrous temporal bone’s

posteromedial face via the porus acusticus (inner acoustic opening). Therefore, we used

CT reconstructions to measure the location of porus acusticus as the landmark for targeting

the nerve.

In the rhesus monkey skull anatomy, another constant feature is a subarcuate

parafloccular recess, which is close to the porus acusticus on the petrous part of the

temporal bone and similar in size (Figure 2-5). The recess, within which a portion of the

cerebellar paraflocculus sits encircled by the three semicircular canals of the ipsilateral

labyrinth, is posterior and superior to the porus acusticus. For different individuals, the

distance between subarcuate parafloccular recess and porus acusticus may be different, but

the relative position is almost the same. This relation served as our criterion to distinguish

these two similar structures and find the location of porus acusticus. Once the porus

acusticus is identified in CT transverse view, we choose one point as our target location for

measurement, which would be about 0.5 mm medial of the porus acusticus and far enough

from nearby bone surfaces to reduce the risk that the recording electrode may hit the bone

in later experiment. Figure 2-6 and Figure 2-7 show the example of measurement in CT

scan. Same measurement method has been applied to all five experimental subjects. Table

2-2 and table 2-3 shows the measurement results. The internal auditory canals are roughly

coaxial with the external bony auditory canals (where the stereotaxic ear bars go) and the

porus acusticus in rhesus is about 1-1.1 cm off midline. Since internal acoustic opening is

Page 41: Beyond Conventional Stereotaxic Targeting: Using ...

32

a hole instead of a single point, one cannot uniquely define or measure its precise location

in CT scan. We measured the location estimated as the center of the nerve exiting the porus

acusticus twice per ear and took the average as the final result. The distances from porus

acusticus to the chamber axis in X-Y plane on both sides for five monkeys were all smaller

than the chamber radius, indicating that the traditional method of stereotactic targeting we

used for chamber placement was adequate to at least ensure that the nerve would be

accessible via each chamber, if only one knows where to look.

2.3.2 Experimental Validation and Errors Calibration

As described previously, based on measurements of vestibular nerve location in

chamber coordinates, we inserted recording electrode through guide tube navigated by the

XY micromanipulator to target ipsilateral vestibular nerve. If we didn’t successfully target

the nerve at the first time, we would pull out and insert the electrode several times or change

XY micromanipulator setting by 0.1 mm per time, then reinsert electrode through guide

tube. We found the vestibular nerve and successfully recorded afferent units in a single

afternoon of electrophysiologic recording during the first experiment after chamber

implantation for monkey 54A and for monkey 4617. Previously, without using CT

guidance, it usually took us 3-4 months to find the vestibular nerve through a systematic

electrophysiologic survey of activity in a volume about the location expected from

traditional stereotactic targeting.

Page 42: Beyond Conventional Stereotaxic Targeting: Using ...

33

Figure 2-5. Region of interest in CT 3D reconstruction

In CT 3D reconstruction :(a) skull base with guide tube in left side chamber. (b) view of left lateral skull with

guide tube from medial side. (c) view through craniotomy from left chamber floor.

Page 43: Beyond Conventional Stereotaxic Targeting: Using ...

34

Figure 2-6. Measurement of inner acoustic opening location in CT stereotaxic coordinate (Animal ID: RhF234D)

Page 44: Beyond Conventional Stereotaxic Targeting: Using ...

35

Figure 2-7. Measurement of inner acoustic opening location in CT chamber coordinate (Animal ID: RhF234D)

Page 45: Beyond Conventional Stereotaxic Targeting: Using ...

36

Table 2-2. Location of porus acusticus in CT stereotaxic coordinate

Location of internal acoustic opening on left and right side in five monkeys measured from 3D reconstruction

in stereotaxic coordinate (cm). Lat+ = lateral, Med- = medial, Ant+ = anterior, Pos- = Posterior, Sup+ =

superior, Inf- = inferior

Table 2-3. Location of porus acusticus in CT chamber coordinate

Location of internal acoustic opening on left and right side in five monkeys measured from 3D reconstruction

in chamber coordinate (cm). X+/X- = X-axis, Y+/Y- = Y-axis, Z+/Z- = Z-axis.

Page 46: Beyond Conventional Stereotaxic Targeting: Using ...

37

2.4 Discussion

Guidance using 3D reconstructions of a CT scan obtained after initial placement of

a recording chamber using traditional stereotactic targeting is an efficient and effective

method that helps target the rhesus vestibular nerve for single-unit recording. We

accomplish this by registering the location of the porus acusticus relative to a chamber’s

axis identified in CT scans and determining parameters for a micromanipulator XY stage

mounted on the chamber that navigates and holds a guide tube and recording electrode.

Although a low sample number precludes a statistically rigorous comparison between the

traditional stereotaxic method alone and traditional stereotaxis augmented by post-

operative CT guidance, the latter was more efficient in our experience, With the help of the

post-operative CT guidance, vestibular afferent neurons were precisely located in the first

experiment after chamber implantation in the two rhesus monkeys for which we used this

technique,.

CT guidance using 3D reconstructions obtained after surgical attachment of a

recording chamber to the calvarium provides a safety net in case of inaccuracies of

stereotactic targeting (e.g., due to differences between the study subject and atlas), and it

gives the surgeon flexibility. For example, if a large meningeal vessel is in the path that

had been planned for a guide tube, the surgeon can reorient the chamber on the fly during

surgery without fear of later failing to find the nerve.

In the rhesus brain stereotaxic atlas we use,[39] the position the portion of

vestibular nerve that comes out of porus acusticus is about: posterior 0.06 cm, inferior 0.10

Page 47: Beyond Conventional Stereotaxic Targeting: Using ...

38

cm and lateral 0.85 cm. Compared with the atlas, in our results, the position is more anterior

and superior, which shows quite obvious inter-subject difference.

Although we still used a traditional stereotactic approach to initially place the

recording chambers, our method does not require that a chamber be very accurately

oriented, as long as the target lies along a path accessible by a guide tube passed parallel

to the chamber’s axis via a craniotomy. As Figure 2-8 shows, in all five monkeys, the

distances from porus acusticus to the chamber axis are all smaller than the chamber radius,

which means the nerve can be reached within the range of chamber. Navigation of electrode

is all based on the CT scan measurements.

Figure 2-8.Location of porus acusticus in transverse plane of chamber coordinate

Part (a) shows the location of porus acusticus in chamber coordinate on left side for five rhesus monkeys,

part (b) shows the right side. Red dots represent targets. The black dot represents chamber axis, and the

black ring shows the location of the ~3.5 cm diameter chamber.

Page 48: Beyond Conventional Stereotaxic Targeting: Using ...

39

The CT method also provides accurate and precise information regarding the

distance from parietal dura to the vestibular nerve, which is another parameter needed to

identify the location of the nerve efficiently. This saves time and reduces the risk of damage

to the animal’s brainstem and to the recording microelectrodes, because it permits one to

quickly and reliably pass the guide tube until it stops just before the nerve while not injuring

the nerve or adjacent vessels. Inserting electrodes with image guidance can therefore

reduce complications compared to standard stereotaxic surgery and therefore reduce the

risk to an animal subject. Moreover, by helping to maximize data yields per animal, CT

guidance can reduce the number of animals needed to obtain data for a given experiment.

Despite the advantages of CT guidance, several modes of targeting failure remain.

First, the guide tube may bend or tilt subtly when penetrating scar tissue near the

craniotomy skull surface, which causes the end of guide tube to move away from planned

position. The tendency for this to occur can be reduced by ensuring the guide tube is rigid,

held to the XY stage in a way that ensures it remains perpendicular to the stage, and filed

so that its tip is concentrically beveled. (Typical slant-beveled spinal needles tend to bend

as they are advanced through tissue.) Second, the outer diameter of tungsten

microelectrodes (120 um) is typically much smaller than the inner diameter of guide tube

(1mm), so a microelectrode advanced beyond the tip of a guide tube may bend when

penetrating hard scar tissue or being squeezed by the core of tissue that may enter the guide

tube lumen as the guide tube is inserted. Third, one can accurately target the

vestibulocochlear nerve with a guide tube and yet not selectively impale individual axons

in a way that yields high signal-to-noise recordings. Because the cell bodies of vestibular

nerve afferents are in the fundus of the internal auditory canal, they are shielded by bone

Page 49: Beyond Conventional Stereotaxic Targeting: Using ...

40

and inaccessible to microelectrodes inserted via a parietal craniotomy. Tungsten

microelectrodes with tips of ~1-3 um may be too large to isolate vestibular primary afferent

axons with diameter less than ~10 um. Fourth, one may accurately target the 7th/8th cranial

nerve bundle but miss the vestibular divisions by being ~200 um too anterior, instead

hitting only facial nerve motor neurons and cochlear afferents. Finally, changes in head

orientation can cause the brainstem to shift position within the skull. The effect of such

movement should be minimized by targeting the nerve where it leaves the bone at the porus

acusticus. However, clinical magnetic resonance imaging scans acquired with patients

lying supine commonly show the 7th/8th nerve complex resting against the posterior lip of

the porus (C Della Santina, personal communication), suggesting that a point coaxial with

the internal auditory canal may not necessarily lying within the center of the nerve.

Page 50: Beyond Conventional Stereotaxic Targeting: Using ...

41

Chapter 3 Intracranial/Extravascular Application of OCT and

IVUS for Imaging the Subarachnoid Space

We explored two real-time imaging techniques that are currently applied to

cardiology intervention surgery: intravascular optical coherence and intravascular

ultrasound. We sought to determine whether these highly refined intravascular

technologies, which are already approved by the United States Food & Drug

Administration for intravascular use in interventional cardiology and interventional

neuroradiology, might be adapted for intracranial/extravascular imaging of structures

in the subarachnoid space.

3.1 Experimental Subjects

3.1.1 Phantom

We made a phantom for image testing by embedding spaghetti and penne pasta in

a bowl full of red gelatin to simulate the CSF space. The inner diameter of penne pasta is

about 5mm and the thickness of penne wall is about 1.5 mm.

3.1.2 Chinchilla Subject

An immediately post-mortem adult wild-type 450g female chinchilla (Chinchilla

lanigera) was used for experiments, which were performed just after euthanasia following

a separate, terminal single unit recording experiment performed under deep general

anesthesia in accordance with a protocol approved by the Johns Hopkins Animal Care and

Use Committee.

Page 51: Beyond Conventional Stereotaxic Targeting: Using ...

42

3.1.3 Rhesus Monkey Subjects

A post-mortem rhesus monkey (RhF4617J) was used for IVUS and OCT

experiments, which was performed in accordance with a protocol (Protocol#: PR12M318)

approved by the Johns Hopkin Animal Care and Use Committee.

3.2 Surgical Procedures

3.2.1 Chinchilla Surgery

A chinchilla was used to study OCT immediately after euthanasia following a

terminal single-unit recording experiment performed as part of a different study. The

chinchilla had been anesthetized with 3-5% isoflurane anesthesia and sterilely prepped.

Skin was locally anesthetized (0.5% bupivacaine with 1:100K epinephrine). Enrofloxacin

antibiotic prophylaxis was given. A patch of skin was excised between the mastoid bullae,

and holes were drilled in the roof of each bulla using otologic drill. GelFilm® and/or

GelFoam® were used to shield the middle ear and ossicles from the mastoid bulla. Dental

cement (3M ESPE ProTemp®II or OrthoJet®) was used to conformally fill the

superomedial recess of each bulla, with the cement joining across the midline adjacent to

the skull to form a rigid cast around the central portion of the skull. A phenolic post was

embedded in the cement and oriented 10o pitched back a plane tangent to the central portion

of the visible skull (so that when animals was restrained with the post in the 45o pitch-nose-

down slot of a Plexiglass restraint, the chinchilla’s mean horizontal semicircular canal axis

was Earth vertical).

Page 52: Beyond Conventional Stereotaxic Targeting: Using ...

43

Immediately following the single unit experiment, the animal was euthanized under

deep inhalational anesthesia. A hole was then opened on the floor of bulla 1mm anterior

and medial to the anterior semicircular canal ampulla using an otologic drill, to create a

port into the internal auditory canal for inserting an OCT imaging catheter.

3.3 Intracranial/Extravascular Optical Coherence Tomography

3.3.1 Method

For the intracranial OCT imaging experiment, we used a St. Jude Medical® C7XR

unit with a C7 Dragonfly IV imaging catheter with OCT flush, 100 frames/sec, pullback

rate 2cm/s. the outer diameter of catheter is 0.9mm. The imaging resolution is about 20um

in greatest dimension. The system is available to perform pullback side-viewing image

scan with length of 55mm.

First, we performed an OCT imaging test in the penne-gelatin phantom to simulate

the situation of cerebrospinal fluid (CSF) space around porus acusticus. Figure 3-1 shows

the OCT images obtained by inserting a 30 mm-length of the catheter roughly along the

axis of penne inside the gelatin. We triggered the pullback scanning and the transducer

moved backward while scanning.

For the chinchilla experiment, we performed an OCT imaging scan to examine the

inner acoustic canal in an animal immediately after euthanasia by passing the catheter

through a hole opened on the floor of bulla via which the 7th/8th nerve complex had been

accessed for single unit recording as part of a separate research protocol (Figure 3-2).

(Whereas we approach the rhesus vestibular nerve for single unit recording via the parietal

cranium, we typically approach the chinchilla internal auditory canal via the bulla, so the

Page 53: Beyond Conventional Stereotaxic Targeting: Using ...

44

never is directly visible in the latter case.) The chinchilla head was then dissected to

determine where the OCT probe had been.

Next, we placed the catheter near the left porus acusticus in a rhesus monkey skull

and performed the image scanning to visualize its bone structure in OCT.

Finally, we passed an OCT catheter inside CSF through the guide tube navigated

by the CT measurement of target in a post-mortem monkey head (RhF4617J).

Figure 3-1. Placement of OCT catheter inside penne embedded in gelatin

Page 54: Beyond Conventional Stereotaxic Targeting: Using ...

45

Figure 3-2.Simulation of placing OCT catheter through the floor of bulla in chinchilla right ear

under camera view

Page 55: Beyond Conventional Stereotaxic Targeting: Using ...

46

3.3.2 Results

Stacks of OCT side-view images were acquired with imaging diameter of 5mm,

which nicely constructed a 3D image of the penne pasta (Figure 3-3). The tiny circle in the

center of image is the catheter around the OCT probe. The larger, oval orange curve with

fading shade outside it is the imaging of interface between gelatin and wall of penne. The

infrared light hardly penetrates the penne’s inner wall and mostly reflects back or is

absorbed by the pasta, so the outer wall/gelatin boundary is not detected by OCT.

An OCT 3D reconstruction was acquired via pullback scanning after the catheter

was passed through the floor of the bulla and into the internal auditory canal and

cerebellopontine angle of a chinchilla immediately post-mortem (Figure 3-4). We can

clearly identify two black holes near each other on the surface of some dense matter which

is likely to be the bone. The landmark in 3D reconstructed image is identical to the real

bone structure, which was exposed after dissecting the chinchilla’s head (Figure 3-5).

OCT 3D reconstruction (Figure 3-6 b) was also done after pullback scanning near

porus acusticus in a rhesus monkey skull (Figure 3-6 a), which clearly shows the porus

acusticus and subarcuate parafloccular recess compared with CT 3D reconstruction (Figure

3-6 c).

Unfortunately, in the OCT image acquired from scanning in the post-mortem rhesus

monkey assisted by post-operative CT guidance (Figure 3-7), we can hardly identify any

structures. In one of the image scans, a dark blurry area close to the shields of catheters

could be CSF since no reflection occurred there. The areas that have bright boundary and

fade in radial gradient could be either soft tissue or bone. We could not confidently relate

Page 56: Beyond Conventional Stereotaxic Targeting: Using ...

47

them to any anatomic landmarks that would help find the porus acusticus or the vestibular

nerve.

Figure 3-3. OCT image of a penne in the gelatin phantom

Top left section shows the 3D reconstruction of part of the penne’s inner wall within the pullback scanning

range. Top right section shows the slice of side-view image, location of which is indicated by the white circle

in the left section. Lower section shows a cross-sectional image along catheter axis, orientation of which is

indicated by the yellow in the top left section.

Page 57: Beyond Conventional Stereotaxic Targeting: Using ...

48

Figure 3-4.OCT image of inner acoustic canal in chinchilla

Two black holes can be identified as porus on the surface of dense matter, facing toward a small space free

of reflection.

Page 58: Beyond Conventional Stereotaxic Targeting: Using ...

49

Figure 3-5.Location of OCT catheter placement into the internal auditory canal and cerebellopontine

angle of a chinchilla placed just after euthanasia and then dissected after OCT image acquisition.

Red circle is the area of internal auditory canal, where vestibular afferent nerves of different semicircular

canals join together. The view is from the contralateral side, after removal of the brain and brainstem.

Page 59: Beyond Conventional Stereotaxic Targeting: Using ...

50

Figure 3-6.Rhesus monkey inner acoustic opening on left side

Rhesus monkey inner acoustic canal: (a) endoscopic photograph of an OCT catheter placed near internal

acoustic opening on bony skull specimen, (b) Optical coherence tomography shows bone/CSF interfaces, (c)

CT 3D reconstruction.

Page 60: Beyond Conventional Stereotaxic Targeting: Using ...

51

Figure 3-7.A slice of OCT side-view scan when catheter is placed in CSF space in the post-mortem

rhesus monkey on left side

Page 61: Beyond Conventional Stereotaxic Targeting: Using ...

52

3.4 Intracranial/Extravascular Ultrasound Imaging of the

Subarachnoid Space

3.4.1 Method

For the intracranial/extravascular ultrasound (IC/EVUS) experiment, we used a

Volcano CORE® mobile unit, operating at 10 frames/sec with the Volcano Eagle-Eye®

Platinum IVUS imaging catheter. The outer diameter of catheter is 1.1mm. The system is

available to perform side-viewing scans and construct a cross-sectional image centered on

the transducer. The imaging resolution is about 200 um in greatest dimension. The

ultrasound unit can provide a real-time side-view image with imaging diameter of 10mm

but it doesn’t have the automatic pullback function we used with our OCT imaging unit.

First, we performed IC/EVUS imaging test in the pasta-gelatin phantom to simulate

the situation of cerebrospinal fluid (CSF) space around porus acusticus by inserting the

catheter roughly along the axis of penne inside the gelatin, which is similar to the approach

in the OCT experiment. We then used post-operative CT guidance on a rhesus monkey

head specimen with implanted chambers and passed the IC/EVUS catheter through brain

tissue through the guide tube navigated by the CT measurement of target. We manually

pulled back the catheter. Due to constraints on availability of OCT and IVUS equipment

for use with animal specimens, we could not perform IC/EVUS in a live animal.

3.4.2 Results

As Figure 3-8 shows, the ultrasound penetrates the penne and a cross-sectional

image of penne is readily detected. The small bright white circle is identified as the shield

Page 62: Beyond Conventional Stereotaxic Targeting: Using ...

53

of catheter and the larger bright circle-like shape is identified as the transverse view of

penne. In one of the ultrasound images scanned in CSF space on the right side in the rhesus

monkey head specimen (Figure 3-9 a), we found an angular structure formed by two short

bright edges.

Figure 3-8.IC/EVUS side-view image of penne embedded in gelatin phantom

The boundary between the outer wall of the penna and the gelatin can be clearly identified, which cannot be

seen in OCT due to light absorption.

Page 63: Beyond Conventional Stereotaxic Targeting: Using ...

54

Figure 3-9. Images of subarcuate parafloccular recess obtained using intracranial/extravascular

ultrasound in the cerebellopontine angle of a rhesus monkey specimen.

Subarcuate parafloccular recess in (a) IC/EVUS side-view scan (b) CT 3D reconstruction (c) CT cross-

sectional image (d) CT cross-sectional image with measurement of the angular structure

Page 64: Beyond Conventional Stereotaxic Targeting: Using ...

55

The angle is approximately 62o, and lengths of two edges are about 4mm and 5mm after

scaling to real space. We retrieved 3D CT reconstruction of the specimen (Figure 3-9 b),

and found that the subarcuate parafloccular recess had a sharp-angle opening. A cross-

sectional CT image (Figure 3-9 c) shows a similar structure as seen in the IC/EVUS image.

As measured in CT scan, the angle is approximately 60° and the lengths of bone edges are

about 4.9mm and 5.3mm (Figure 3-9 d). Measurements from the IC/EVUS image and the

CT image are basically the same, which suggests that IC/EVUS can identify bone/CSF

boundaries like the porus acusticus and subarcuate parafloccular recess, if they have

already been seen on a CT 3D reconstruction. However, even with advance knowledge of

the anatomy from the CT dataset, we could not identify any structure that appeared to be

the vestibular nerve on IC/EVUS.

3.5 Discussion

Comparing the OCT and IC/EVUS images of the penne-gelatin phantom (Figure

3-10), we can see that OCT has better spatial resolution than ultrasound, but IC/EVUS is

able to penetrate the penne and provides both a larger imaging depth and definition of the

outer pasta/gelatin boundary. Summary of comparison is shown in Table 3-1.

Techniques Catheter Model Imaging Depth Resolution Catheter

Diameter

Automatic

Pullback

Function

Images

through

optical

opaque

material

OCT C7 Dragonfly

(St. Jude

Medical)

5 mm in

diameter

~20 um 2.7F

(0.9mm)

Yes No

IC/EVUS Eagle-Eye

Platinum

10 mm in

diameter

~200 um 3.5F

(1.2 mm)

No Yes

Table 3-1.Comparison of OCT and IC/EVUS

Page 65: Beyond Conventional Stereotaxic Targeting: Using ...

56

In the immediately postmortem chinchilla specimen, OCT 3D reconstruction

apparently showed the porus acusticus and subarcuate parafloccular fossa, which although

unnecessary in that experimental preparation (we can already see the chinchilla 7th/8th nerve

directly), served as a useful proof of concept. Unfortunately, our attempts at performing

OCT imaging inside CSF space in the post-mortem rhesus monkey did not result in images

from which we could identify the vestibular nerve, porus acusticus, or other anatomic

landmarks. There are several possible causes for the failure. First, soft tissue and possibly

blood in CSF make the OCT light reflect back and generate blurring artifacts, which limits

the field of view and penetration depth. Regardless of any other source of failure, the major

limitation of OCT for our application was the shallow penetration depth. Second, the guide

tube might not have been pointed at the right target, so the porus acusticus or subarcuate

parafloccular recess might have been completely out of the imaging range. Third, inability

to image along the guide tube axis with our side-viewing OCT system precluded seeing

where the guide tube was headed. Finally, the lack of OCT catheters designed and intended

for our intended use required that we instead use a clinical model of OCT catheter intended

for coronary artery imaging that would not allow us to advance the OCT imaging unit to

the end of the unit’s protective catheter sheath. To overcome this, we had to cut the end of

the sheath, allowing bodily fluids to back fill into the catheter and perhaps cloud the image.

As for IC/EVUS, although its resolution is not as high as OCT, it has larger

penetration through soft tissue. The IC/EVUS images we obtained (Figure 3-9) clearly

show the bone boundaries of the subarcuate parafloccular recess, as confirmed by CT scan.

IC/EVUS provided clear contrast between bone and other tissue, which can help identify

bone landmarks that can help one target the vestibular nerve if the catheter is within 10 mm

Page 66: Beyond Conventional Stereotaxic Targeting: Using ...

57

of the porus acusticus. However, the IVUS probe is slightly larger than the 1.0 mm inner

diameter guide tube we normally use.

Figure 3-10.Comparison of OCT and IC/EVUS in gelatin phantom

Images of the penne in gelatin: (a) camera view of the penne in gelatin (b) OCT side-view image of the penne

(c) IC/EVUS side-view image of the penne

Page 67: Beyond Conventional Stereotaxic Targeting: Using ...

58

Chapter 4 Conclusion

4.1 Post-Operative CT Guidance

Compared with the conventional stereotaxic method, augmenting traditional

stereotaxis with post-operative CT guidance as presented in this thesis was more efficient

and more effective in helping researchers find the rhesus vestibular nerve. This method

doesn’t completely depend on an atlas or require the monkey to be physically attached to

the stereotaxic frame in recording stage, which allows for more flexibility in surgical

technique and allows a neurophysiologist to find the vestibular nerve despite inter-subject

anatomic differences. Moreover, this method provides intuitive image guidance on where

and how deep the guide tube and electrode should be inserted to reach the target. Having

accurate measurements from a CT scan greatly reduces the risk of overinsertion, so one

can avoid inadvertently cutting the nerve during guide tube insertion or damaging the

microelectrode by driving it into a bone wall. CT imaging units with adequate resolution

are widely available in academic medical centers, and 3D reconstruction software is readily

available to research labs.

4.2 Intracranial Imaging of the Subarachnoid Space Using OCT and

IVUS

OCT and ultrasound within CSF spaces like the cerebellopontine angle offer

complementary advantages for targeting cranial nerves. When passed through the guide

tube of a single-unit recording set-up, optical coherence tomography probes can provide

direct view of intracranial tissue and bone structures through a small craniotomy.

Page 68: Beyond Conventional Stereotaxic Targeting: Using ...

59

Unfortunately, in our experience, optical blurring and a shallow penetration depth limited

the ability to visualize the porus acusticus or the vestibular nerve in post-mortem and live

rhesus monkeys using OCT.

Our intracranial adaptation of IVUS offered a larger imaging depth and ability to

see through optically opaque or translucent material (probably including nerve, dura and

brain), but its resolution is not as high as optical coherence tomography. The large

penetration depth enables it to detect bone structures beyond soft tissues within the range.

Visualization of the bone landmarks in US images has the potential to serve as real-time

guidance for targeting cranial nerves in electrophysiologic experiments.

Adaptation of intravascular OCT and IVUS systems for intracranial/extravascular

imaging of cranial and spinal nerves may hold promise for intraoperative guidance in

minimal-access rhizotomy and other clinical procedures that require precise spatial

targeting of neural structures through narrow channels. Further development of OCT is

needed in imaging penetration. It would be great if OCT and IVUS are combined in one

modality for targeting, which will have large imaging penetration depth for detecting more

structures and high resolution for detecting soft tissues.

4.3 Future Directions

Although the post-operative CT guidance can already give us direct and precise

guidance to reach the porus acusticus, imprecision in our estimate of the exact direction

and size of vestibular nerve coming out of the porus acusticus still make targeting a

challenge. To visualize the vestibular nerve, coregistering 3D reconstructions of both CT

and MRI imaging could be considered. To perform MRI after chamber implantation, the

Page 69: Beyond Conventional Stereotaxic Targeting: Using ...

60

chambers and guide tube would need to be replaced by non-ferrous, nonmagnetic material.

Recent progress in intravascular MRI [99-101] also holds promise as another means by

which a high resolution imaging unit can be introduced via a guide tube into the

subarachnoid space, whether for targeting the vestibular nerve in rhesus monkeys or for

analogous neurosurgical procedures in humans.

Page 70: Beyond Conventional Stereotaxic Targeting: Using ...

61

Reference List

1. Blatt, P.J., et al., The reliability of the Vestibular Autorotation Test (VAT) in

patients with dizziness. J Neurol Phys Ther, 2008. 32(2): p. 70-9.

2. Minor, L.B., Gentamicin-induced bilateral vestibular hypofunction. JAMA, 1998.

279(7): p. 541-4.

3. Crawford, J., Living without a Balancing Mechanism. New England Journal of

Medicine, 1952. 246(12): p. 458-460.

4. Ward, B.K., et al., Prevalence and impact of bilateral vestibular hypofunction:

results from the 2008 US National Health Interview Survey. JAMA

Otolaryngology–Head & Neck Surgery, 2013. 139(8): p. 803-810.

5. Della Santina, C.C., et al., Current and future management of bilateral loss of

vestibular sensation - an update on the Johns Hopkins Multichannel Vestibular

Prosthesis Project. Cochlear Implants Int, 2010. 11 Suppl 2: p. 2-11.

6. Della Santina, C.C., A.A. Migliaccio, and A.H. Patel, A multichannel semicircular

canal neural prosthesis using electrical stimulation to restore 3-D vestibular

sensation. Ieee Transactions on Biomedical Engineering, 2007. 54(6): p. 1016-

1030.

7. Gong, W.S. and D.M. Merfeld, Prototype neural semicircular canal prosthesis

using patterned electrical stimulation. Annals of Biomedical Engineering, 2000.

28(5): p. 572-581.

8. Gong, W.S. and D.M. Merfeld, System design and performance of a unilateral

horizontal semicircular canal prosthesis. Ieee Transactions on Biomedical

Engineering, 2002. 49(2): p. 175-181.

9. Chiang, B., G.Y. Fridman, and C.C. Della Santina, Enhancements to the Johns

Hopkins Multi-Channel Vestibular Prosthesis Yield Reduced Size, Extended

Battery Life, Current Steering and Wireless Control. Association for Research in

Otolaryngology. 2009. Baltimore, MD.

10. Della Santina, C.C., A.A. Migliaccio, and A.H. Patel, Electrical stimulation to

restore vestibular function - development of a 3-D vestibular prosthesis. 27th

Annual IEEE Engineering in Medicine and Biology. 2005. Shanghai,China.

11. Della Santina, C.C., et al., Orientation of human semicircular canals measured by

three-dimensional multiplanar CT reconstruction. J Assoc Res Otolaryngol, 2005.

6(3): p. 191-206.

12. Carey, J.P. and C.C. Della Santina, Principles of applied vestibular physiology.

Cummings: Otolaryngology: Head & Neck Surgery. 4th ed. Philadelphia, PA:

Elsevier Mosby, 2005. 3: p. 3115-3159.

13. Ewald, J., Physiologische untersuchugen uber das Endorgans des Nervus Octavus:

Bergmann. 1892.

14. Wilson, V. and G. Jones, Mammalian vestibular physiology Plenum New York.

1979.

15. Robinson, F.P. and P.G. Murphy, The Validity of Measuring Eye Movements by

Direct Observation. Science, 1932. 76(1964): p. 171-2.

Page 71: Beyond Conventional Stereotaxic Targeting: Using ...

62

16. Robinson, D.A., A Method of Measuring Eye Movement Using a Scleral Search

Coil in a Magnetic Field. Ieee Transactions on Biomedical Engineering, 1963.

Bm10(4): p. 137-&.

17. Aw, S.T., et al., Three-dimensional vector analysis of the human vestibuloocular

reflex in response to high-acceleration head rotations .2. Responses in subjects

with unilateral vestibular loss and selective semicircular canal occlusion. Journal

of Neurophysiology, 1996. 76(6): p. 4021-4030.

18. Aw, S.T., et al., Three-dimensional vector analysis of the human vestibuloocular

reflex in response to high-acceleration head rotations .1. Responses in normal

subjects. Journal of Neurophysiology, 1996. 76(6): p. 4009-4020.

19. Migliaccio, A.A., et al., Inexpensive system for real-time 3-dimensional video-

oculography using a fluorescent marker array. J Neurosci Methods, 2005. 143(2):

p. 141-50.

20. Cohen, B., S. Shanzer, and J. Suzuki, Nystagmus Induced by Stimulation of Single

Nerves from Semicircular Canals. Federation Proceedings, 1963. 22(2): p. 338-&.

21. Cohen, B., J. Suzuki, and M.B. Bender, Eye Movements from Semicircular Canal

Nerve Stimulation in Cat. Annals of Otology Rhinology and Laryngology, 1964.

73(1): p. 153-&.

22. Cohen, B. and J. Suzuki, Eye Movements Produced by Vestibular Nerve

Stimulation. Electroencephalography and Clinical Neurophysiology, 1963. 15(1):

p. 152-&.

23. Cohen, B. and J.I. Suzuki, Eye Movements Induced by Ampullary Nerve Stimulation.

American Journal of Physiology, 1963. 204(2): p. 347-&.

24. Cohen, B. and J.I. Suzuki, Phases of Central Excitability Evoked by Ampullary

Nerve Stimulation. Federation Proceedings, 1964. 23(2p1): p. 414-&.

25. Suzuki, J., B. Cohen, and M.B. Bender, Compensatory Eye Movements Induced by

Vertical Semicircular Canal Stimulation. Experimental Neurology, 1964. 9(2): p.

137-&.

26. Suzuki, J.I., et al., Implantation of electrodes near individual vestibular nerve

branches in mammals. Ann Otol Rhinol Laryngol, 1969. 78(4): p. 815-26.

27. Della Santina, C., A. Migliaccio, and A. Patel, Electrical stimulation to restore

vestibular function development of a 3-d vestibular prosthesis. Conf Proc IEEE Eng

Med Biol Soc, 2005. 7: p. 7380-5.

28. Rahman, M.A., et al., Restoring the 3D vestibulo-ocular reflex via electrical

stimulation: the Johns Hopkins multichannel vestibular prosthesis project. Conf

Proc IEEE Eng Med Biol Soc, 2011. 2011: p. 3142-5.

29. Chiang, B., et al., Design and Performance of a Multichannel Vestibular Prosthesis

That Restores Semicircular Canal Sensation in Rhesus Monkey. Ieee Transactions

on Neural Systems and Rehabilitation Engineering, 2011. 19(5): p. 588-598.

30. Fridman, G.Y., et al., Vestibulo-ocular reflex responses to a multichannel

vestibular prosthesis incorporating a 3D coordinate transformation for correction

of misalignment. J Assoc Res Otolaryngol, 2010. 11(3): p. 367-81.

31. Davidovics, N.S., et al., Effects of biphasic current pulse frequency, amplitude,

duration, and interphase gap on eye movement responses to prosthetic electrical

stimulation of the vestibular nerve. IEEE Trans Neural Syst Rehabil Eng, 2011.

19(1): p. 84-94.

Page 72: Beyond Conventional Stereotaxic Targeting: Using ...

63

32. C. Dai, G.Y.F., N.S. Davidovics, B. Chiang, J.H. Ahn, and C.C. Della Santina,

Restoration of 3D vestibular sensation in rhesus monkeys using a multichannel

vestibular prosthesis. Hear Res, Aug 2011.

33. C. Dai, G.Y.F., B. Chiang, N.S. Davidovics, T.A. Melvin, K.E. Cullen, and C.C.

Della Santina, Cross-axis adaptation improves 3D vestibulo-ocular reflex

alignment during chronic stimulation via a head-mounted multichannel vestibular

prosthesis. Exp Brain Res, Mar 4 2011.

34. Carey, J., Multichannel Vestibular Implant Early Feasibility Study. In:

ClinicalTrials.gov[Internet].Bethesda(MD): National Library of Medicine (US).

2000-[cited 2016 Aug 9]. Available from:

https://clinicaltrials.gov/ct2/show/NCT02725463 NLM Identifier: NCT02725463.

35. Sildiroglu, O., et al., Evaluation of cochlear nerve size by magnetic resonance

imaging in elderly patients with sensorineural hearing loss. Radiol Med, 2010.

115(3): p. 483-7.

36. Jaryszak, E.M., et al., Cochlear Nerve Diameter in Normal Hearing Ears Using

High-Resolution Magnetic Resonance Imaging. Laryngoscope, 2009. 119(10): p.

2042-2045.

37. Nakamichi, R., et al., Establishing Normal Diameter Range of the Cochlear and

Facial Nerves with 3D-CISS at 3T. Magnetic Resonance in Medical Sciences, 2013.

12(4): p. 241-247.

38. Gacek, R.R. and G.L. Rasmussen, Fiber analysis of the statoacoustic nerve of

guinea pig, cat, and monkey. Anat Rec, 1961. 139: p. 455-63.

39. Paxinos, G., X.-F. Huang, and A.W. Toga, The rhesus monkey brain in stereotaxic

coordinates. 2000.

40. Saleem, K.S. and N.K. Logothetis, Atlas of the rhesus monkey brain in stereotaxic

coordinates: a combined mri and histology. 2006: Academic Press.

41. Saleem, K.S. and N.K. Logothetis, A combined MRI and histology atlas of the

rhesus monkey brain in stereotaxic coordinates. 2012: Academic Press.

42. Sadeghi, S.G., et al., Effects of canal plugging on the vestibuloocular reflex and

vestibular nerve discharge during passive and active head rotations. Journal of

neurophysiology, 2009. 102(5): p. 2693-2703.

43. Haque, A., D.E. Angelaki, and J.D. Dickman, Spatial tuning and dynamics of

vestibular semicircular canal afferents in rhesus monkeys. Exp Brain Res, 2004.

155(1): p. 81-90.

44. Angelaki, D.E. and J.D. Dickman, Spatiotemporal processing of linear

acceleration: primary afferent and central vestibular neuron responses. J

Neurophysiol, 2000. 84(4): p. 2113-32.

45. Sylvestre, P.A. and K.E. Cullen, Quantitative analysis of abducens neuron

discharge dynamics during saccadic and slow eye movements. Journal of

Neurophysiology, 1999. 82(5): p. 2612-2632.

46. Sadeghi, S.G., L.B. Minor, and K.E. Cullen, Response of vestibular-nerve afferents

to active and passive rotations under normal conditions and after unilateral

labyrinthectomy. Journal of neurophysiology, 2007. 97(2): p. 1503-1514.

47. Cullen, K.E. and L.B. Minor, Semicircular canal afferents similarly encode active

and passive head-on-body rotations: implications for the role of vestibular

efference. J Neurosci, 2002. 22(RC226): p. 1-7.

Page 73: Beyond Conventional Stereotaxic Targeting: Using ...

64

48. Khan, F.R. and J.M. Henderson, Deep brain stimulation surgical techniques.

Handb Clin Neurol, 2013. 116: p. 27-37.

49. Brown, R.A. and J.A. Nelson, Invention of the N-Localizer for Stereotactic

Neurosurgery and Its Use in the Brown-Roberts-Wells Stereotactic Frame.

Neurosurgery, 2012. 70(6): p. 173-176.

50. Brown, R.A. and J.A. Nelson, The Origin and History of the N-Localizer for

Stereotactic Neurosurgery. Cureus, 2015. 7(9): p. e323.

51. Tse, V.C., M.Y.S. Kalani, and J.R. Adler, Techniques of Stereotactic Localization.

Principles and Practice of Stereotactic Radiosurgery, 2015: p. 25.

52. Saleh, H. and B. Kassas, Developing stereotactic frames for cranial treatment.

Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy, 2014: p. 153.

53. Arle, J., Development of a classic: The Todd-Wells apparatus, the BRW, and the

CRW stereotactic frames, in Textbook of Stereotactic and Functional Neurosurgery.

2009, Springer. p. 453-467.

54. Sharan, A.D. and D.W. Andrews, Stereotactic frames: technical considerations.

NEUROLOGICAL DISEASE AND THERAPY, 2003. 58: p. 11-20.

55. Apuzzo, M.L. and C.A. Fredericks, The Brown-Roberts-Wells system, in Modern

stereotactic neurosurgery. 1988, Springer. p. 63-77.

56. Heilbrun, M.P., et al., Preliminary experience with Brown-Roberts-Wells (BRW)

computerized tomography stereotaxic guidance system. J Neurosurg, 1983. 59(2):

p. 217-22.

57. Thomas, D.G., R.E. Anderson, and G.H. du Boulay, CT-guided stereotactic

neurosurgery: experience in 24 cases with a new stereotactic system. J Neurol

Neurosurg Psychiatry, 1984. 47(1): p. 9-16.

58. Goerss, S., et al., A Computed Tomographic Stereotactic Adaptation System.

Neurosurgery, 1982. 10(3): p. 375-379.

59. Brown, R.A., A computerized tomography-computer graphics approach to

stereotaxic localization. J Neurosurg, 1979. 50(6): p. 715-20.

60. Heilbrun, M.P., et al., Brown-Roberts-Wells stereotactic frame modifications to

accomplish magnetic resonance imaging guidance in three planes. Appl

Neurophysiol, 1987. 50(1-6): p. 143-52.

61. Leksell, L., D. Leksell, and J. Schwebel, Stereotaxis and nuclear magnetic

resonance. J Neurol Neurosurg Psychiatry, 1985. 48(1): p. 14-8.

62. Thomas, D.G., et al., Stereotaxic biopsy of the brain under MR imaging control.

AJNR Am J Neuroradiol, 1986. 7(1): p. 161-3.

63. Maciunas, R.J., et al., Positron emission tomography imaging-directed stereotactic

neurosurgery. Stereotact Funct Neurosurg, 1992. 58(1-4): p. 134-40.

64. Levivier, M., et al., Use of stereotactic PET images in dosimetry planning of

radiosurgery for brain tumors: Clinical experience and proposed classification.

Journal of Nuclear Medicine, 2004. 45(7): p. 1146-1154.

65. Ohayon, S. and D.Y. Tsao, MR-guided stereotactic navigation. Journal of

Neuroscience Methods, 2012. 204(2): p. 389-397.

66. Vega, R.A., K.L. Holloway, and P.S. Larson, Image-guided deep brain stimulation.

Neurosurg Clin N Am, 2014. 25(1): p. 159-72.

Page 74: Beyond Conventional Stereotaxic Targeting: Using ...

65

67. Eggers, G., B. Kress, and J. Muhling, Fully automated registration of

intraoperative computed tomography image data for image-guided craniofacial

surgery. J Oral Maxillofac Surg, 2008. 66(8): p. 1754-60.

68. Ferroli, P., et al., Advanced 3-Dimensional Planning in Neurosurgery.

Neurosurgery, 2013. 72: p. A54-A62.

69. Ferroli, P., et al., Advanced 3-dimensional planning in neurosurgery. Neurosurgery,

2013. 72 Suppl 1: p. 54-62.

70. Watanabe, E., et al., Three-dimensional digitizer (neuronavigator): new equipment

for computed tomography-guided stereotaxic surgery. Surgical neurology, 1987.

27(6): p. 543-547.

71. Colbassani, H.J., et al., CT-assisted stereotactic brain biopsy: value of

intraoperative frozen section diagnosis. J Neurol Neurosurg Psychiatry, 1988.

51(3): p. 332-41.

72. Venkataramana, N.K., et al., Open-labeled study of unilateral autologous bone-

marrow-derived mesenchymal stem cell transplantation in Parkinson's disease.

Transl Res, 2010. 155(2): p. 62-70.

73. Lang, A.E., et al., Randomized controlled trial of intraputamenal glial cell line-

derived neurotrophic factor infusion in Parkinson disease. Ann Neurol, 2006.

59(3): p. 459-66.

74. Bom, N., C. Lancee, and F. Van Egmond, An ultrasonic intracardiac scanner.

Ultrasonics, 1972. 10(2): p. 72-76.

75. Liu, J.B. and B.B. Goldberg, Catheter-based intraluminal sonography. J

Ultrasound Med, 2004. 23(2): p. 145-60.

76. Li, X., et al., Integrated IVUS-OCT imaging for atherosclerotic plaque

characterization. Selected Topics in Quantum Electronics, IEEE Journal of, 2014.

20(2): p. 196-203.

77. Nissen, S.E., et al., Intravascular Ultrasound Assessment of Lumen Size and Wall

Morphology in Normal Subjects and Patients with Coronary-Artery Disease.

Circulation, 1991. 84(3): p. 1087-1099.

78. Nicholls, S.J., et al., Intravascular ultrasound in cardiovascular medicine.

Circulation, 2006. 114(4): p. E55-E59.

79. Lee, J.T. and R.A. White. Basics of intravascular ultrasound: an essential tool for

the endovascular surgeon. in Seminars in vascular surgery. 2004. Elsevier.

80. Lee, J.T., T.D. Fang, and R.A. White. Applications of intravascular ultrasound in

the treatment of peripheral occlusive disease. in Seminars in vascular surgery.

2006. Elsevier.

81. Marrocco, C.J., et al., Intravascular Ultrasound. Seminars in Vascular Surgery,

2012. 25(3): p. 144-152.

82. Rogacka, R., A. Latib, and A. Colombo, IVUS-guided stent implantation to improve

outcome: a promise waiting to be fulfilled. Current cardiology reviews, 2009. 5(2):

p. 78-86.

83. Jiang, J., et al., The application of intravascular ultrasound imaging in the

diagnosis of aortic dissection. Zhonghua wai ke za zhi [Chinese journal of surgery],

2003. 41(7): p. 491-494.

Page 75: Beyond Conventional Stereotaxic Targeting: Using ...

66

84. Brezinski, M.E., et al., Assessing atherosclerotic plaque morphology: comparison

of optical coherence tomography and high frequency intravascular ultrasound.

Heart, 1997. 77(5): p. 397-403.

85. Brezinski, M.E., et al., Optical coherence tomography for optical biopsy.

Properties and demonstration of vascular pathology. Circulation, 1996. 93(6): p.

1206-13.

86. Fujimoto, J.G., et al., High resolution in vivo intra-arterial imaging with optical

coherence tomography. Heart, 1999. 82(2): p. 128-33.

87. Bezerra, H.G., et al., Intracoronary optical coherence tomography: a

comprehensive review clinical and research applications. JACC Cardiovasc Interv,

2009. 2(11): p. 1035-46.

88. Yaqoob, Z., et al., Methods and application areas of endoscopic optical coherence

tomography. J Biomed Opt, 2006. 11(6): p. 063001.

89. Kang, W., et al., Endoscopically guided spectral-domain OCT with double-balloon

catheters. Opt Express, 2010. 18(16): p. 17364-72.

90. Moon, S., et al., Semi-resonant operation of a fiber-cantilever piezotube scanner

for stable optical coherence tomography endoscope imaging. Opt Express, 2010.

18(20): p. 21183-97.

91. Sun, J. and H. Xie, MEMS-Based Endoscopic Optical Coherence Tomography.

International Journal of Optics, 2011. 2011.

92. Pan, Y., et al., Hand-held arthroscopic optical coherence tomography for in vivo

high-resolution imaging of articular cartilage. J Biomed Opt, 2003. 8(4): p. 648-

54.

93. Liang, C.P., et al., A forward-imaging needle-type OCT probe for image guided

stereotactic procedures. Opt Express, 2011. 19(27): p. 26283-94.

94. Nguyen, C.T., et al., Non-invasive optical interferometry for the assessment of

biofilm growth in the middle ear. Biomed Opt Express, 2010. 1(4): p. 1104-1116.

95. Djalilian, H.R., et al., Optical coherence tomography of cholesteatoma. Otol

Neurotol, 2010. 31(6): p. 932-5.

96. Djalilian, H.R., et al., Imaging the human tympanic membrane using optical

coherence tomography in vivo. Otol Neurotol, 2008. 29(8): p. 1091-4.

97. Burkhardt, A., et al., Endoscopic optical coherence tomography device for forward

imaging with broad field of view. J Biomed Opt, 2012. 17(7): p. 071302.

98. Ahn, J.H., C. Dai, and C.C. Della Santina, Skin grafting facilitates the maintenance

of head recording chambers for neurophysiological recording. Journal of

neuroscience methods, 2013. 215(2): p. 161-163.

99. Hegde, S.S., Y. Zhang, and P.A. Bottomley. Accelerated, motion-corrected high-

resolution intravascular MRI at 3T. in Proceedings of the International Society for

Magnetic Resonance in Medicine... Scientific Meeting and Exhibition.

International Society for Magnetic Resonance in Medicine. Scientific Meeting and

Exhibition. 2013. NIH Public Access.

100. Hegde, S.S., Y. Zhang, and P.A. Bottomley, Acceleration and motion‐correction

techniques for high‐resolution intravascular MRI. Magnetic resonance in medicine,

2015. 74(2): p. 452-461.

Page 76: Beyond Conventional Stereotaxic Targeting: Using ...

67

101. Bottomley, P., Y. Zhang, and S.S. Hegde, Methods and apparatus for accelerated,

motion-corrected high-resolution mri employing internal detectors or mri

endoscopy. 2013, Google Patents.

Page 77: Beyond Conventional Stereotaxic Targeting: Using ...

68

Curriculum Vitae

Shiyao Dong was born in Xi’an, Shaanxi, China on February 1st, 1992. After

graduating from the Gaoxin No.1 High School, he attended the Zhejiang University. In

2010, he graduated from the Zhejiang University with a B.S in Biomedical Engineering.

His research interests include: biomedical engineering and data mining.