BACKGROUND Biomechanical properties of cartilage vary throughout joint surface Manual positioning &...
-
Upload
bryce-harrington -
Category
Documents
-
view
216 -
download
0
Transcript of BACKGROUND Biomechanical properties of cartilage vary throughout joint surface Manual positioning &...
BACKGROUND• Biomechanical properties of cartilage vary
throughout joint surface• Manual positioning & testing of each site is time-
consuming• Proposed Robotic Mechanical Testing System
(RMTS) incorporates a 6-DOF robotic platform to automate testing
• Indentation testing non-destructively characterizes mechanical properties of cartilage on joints
OBJECTIVES• Design an automated positioning system to allow
high throughput mechanical testing• Outfit robotic platform for mechanical testing• Incorporate modular design that can be used for
future applications
HARDWARE DESIGN
ROBOTIC MECHANICAL TESTING SYSTEM (RMTS)Meena Siddiqui, Sahil D Patel, Bradley C Hansen, Johnny C Tam, Cynzia X Fan
Won C Bae, PhD, Robert L Sah, MD, ScD, Graeme M. Bydder, MDDepartment of Bioengineering, University of California-San Diego
ACKNOWLEDGMENTS Calit2 Undergraduate Research Scholarship
Figure 1. RMTS concept design showing the major sub-systems.
• Specifications:• Weight: 800 lbs• Dimensions: Ø 36”, Height 24”• Platform capacity: 220 lbs• Accuracy: ± .002”• Repeatability: .001”• Working volume
• X,Y,Z: 4”• A,B axis: 15o
• C axis: ± 720o
2. Computer • Coordinate movement of robot, indentor, and laser• Acquire and analyze laser and indentor data• Various algorithms were designed to coordinate
movement of robot with the indentor, using data from the laser system
September 21, 2007
3. HYDRATION CONTAINMENT
4. LASER 5. INDENTOR
7. FRAME
2. COMPUTER
6. HYDRATION
1. ROBOT
7. Frame• House robot, laser, and hydration system• Provide attachments for modular subsystems• Support forces generated by the robotic platform with minimal deflection
6. Hydration SystemMaintain joint in Phosphate Buffered Saline during testingUse spray nozzle to minimize fluid volume & leakage of fluid
5. Indenting System• Apply displacement and measure force to determine mechanical stiffness• Develop active indentation system with actuator and force transducer to allow use of various probe tips • Align sample via robotic platform
4. Laser Scanning System• Obtain 3D geometric data of the cartilage surface on the joint• Measure displacements as robotic platform moves sample
• Robot Calibration Procedure Using a Laser:1. Find distance vs. volt ratio, and range of laser2. Measure translation (X, Y, Z), set up laser along
axis of motion, record at 5mm increments
3. Measure rotation (A, B, C), calculate angle of motion by:i. Representing motion with isosceles triangleii. Using law of cosines: to determine input angle α
LASER
Design Goals
Protect Platform
Fit many clamps
Fit on robot
Not affect accuracy
Resist corrosion
3. Hydration Containment System • Adapter 1: ¾” thick aluminum, ½-13 counterbores, 3/8-16 taps, self-sealing hex bolts• Basin: Anodized aluminum 16” baking pan• Adapter 2: ¾” stainless steel, 3/8-16 counterbores, ¼-20 taps, T-slots, O-ring grooves
Figure 5. Sectional view of adapters and basin assembled w/ bolts
INDENTOR
PATELLA
1.R1000 Robot• Assist laser scanning system by providing translation of sample• Align joint surface perpendicular to indentation
probe• Robot calibration necessary to examine 6DOF precision, accuracy and backlash of R1000
Figure 2. R1000 robot
PLATFORM
• Test plans:• Insert cobalt chloride moisture strips inside O-ring grooves• Blue to pink with water contact
Figure 4.18” Adapter 1 (left)10” Adapter 2 (right)
PLATFORM
Figure 3. Calibration set-up
XY
Figure 7. Indentation probe testing human patella
• NIH
• HHMI