AVR Assistant Manual - EMRI · EMRI standard CAN HS connectors are not capable of powering the AVR...

24
Version 1.7 Configure, analyse and diagnose AVR ASSISTANT User manual Product version V1.1.3.0

Transcript of AVR Assistant Manual - EMRI · EMRI standard CAN HS connectors are not capable of powering the AVR...

Page 1: AVR Assistant Manual - EMRI · EMRI standard CAN HS connectors are not capable of powering the AVR Assistant, unless otherwise specified. The AVR Assistant comes with a dedicated

Version 1.7

Configure, analyse and diagnose

AVR ASSISTANT

User manual Product version V1.1.3.0

Page 2: AVR Assistant Manual - EMRI · EMRI standard CAN HS connectors are not capable of powering the AVR Assistant, unless otherwise specified. The AVR Assistant comes with a dedicated

Generator rewinding & repair. Voltage regulator products

Version 1.7 Page 2 of 24

Electronics AVR Assistant Manual

Revision history

The table below provides a historical summary of the changes made to this document.

Revisions are listed in chronological order.

Version Date Change

1.6 2016-11-17 Manual V1.1.2.0

1.7 2017-11-09 Manual V1.1.3.0

Applicability

This manual is applicable to the following product versions.

Product Product description Applicable

Handset V1.1.2.0 AVR Assistant Yes, FW V1.2

Handset V1.1.3.0 AVR Assistant Yes

Page 3: AVR Assistant Manual - EMRI · EMRI standard CAN HS connectors are not capable of powering the AVR Assistant, unless otherwise specified. The AVR Assistant comes with a dedicated

Generator rewinding & repair. Voltage regulator products

Version 1.7 Page 3 of 24

Electronics AVR Assistant Manual

WARNINGS

The manual does not cover all technical details of the product. Specifications may be

modified by the manufacturer without notice. For further information, the manufacturer should

be contacted.

ELECTRICAL HAZARDOUS VOLTAGES

DANGEROUS DO NOT OPERATE WHEN

NOT FAMILIAR WITH GENERATORS

WARNING

Dangerous voltages are present at the voltage regulator

board. Accidental contact with live conductors could result

in serious electrical shock or electrocution.

Disconnect the power source before making repairs,

connecting test instruments, or removing or making

connections to the voltage regulator or generator.

WARNING

The system should not be installed, operated, serviced or

modified except by qualified personnel who understand

the danger of electric shock hazards and have read and

understood the user instructions

WARNING

Never work on a LIVE generator. Unless there is another

person present who can switch off the power supply or

stop the engine

Page 4: AVR Assistant Manual - EMRI · EMRI standard CAN HS connectors are not capable of powering the AVR Assistant, unless otherwise specified. The AVR Assistant comes with a dedicated

Generator rewinding & repair. Voltage regulator products

Version 1.7 Page 4 of 24

Electronics AVR Assistant Manual

1. TABLE OF CONTENTS

2. General description ........................................................................................................ 5

3. Device overview ............................................................................................................. 6

4. Specifications ................................................................................................................. 7

4.1. Connector ............................................................................................................... 7

4.2. Basic features ......................................................................................................... 8

4.2.1. Power ............................................................................................................... 8

4.2.2. Power saving.................................................................................................... 8

4.2.3. Status bar ......................................................................................................... 9

4.2.4. Reset ............................................................................................................... 9

4.2.5. Display brightness ............................................................................................ 9

4.2.6. Menu navigation ..............................................................................................10

5. Connectivity ...................................................................................................................11

6. Features ........................................................................................................................12

6.1. AVR Advanced settings ..........................................................................................12

6.1.1. Safety ..............................................................................................................13

6.1.2. Compatibility ...................................................................................................13

6.1.3. Reading advanced settings .............................................................................14

6.1.4. Browse advanced settings...............................................................................15

6.1.5. Editing advanced settings ...............................................................................16

6.1.6. Writing advanced settings ...............................................................................17

6.2. Process variables ...................................................................................................18

6.2.1. Process variables page 1: Measurements .......................................................19

6.2.2. Process variables page 2: Protections ............................................................19

6.2.3. Process variables page 3: Setpoints ...............................................................19

7. Settings .........................................................................................................................20

7.1.1. Access Level ...................................................................................................20

7.1.2. Firmware upgrade ...........................................................................................20

7.1.3. Keypad test .....................................................................................................20

8. Troubleshooting.............................................................................................................21

9. Known limitations ..........................................................................................................22

Page 5: AVR Assistant Manual - EMRI · EMRI standard CAN HS connectors are not capable of powering the AVR Assistant, unless otherwise specified. The AVR Assistant comes with a dedicated

Generator rewinding & repair. Voltage regulator products

Version 1.7 Page 5 of 24

Electronics AVR Assistant Manual

2. GENERAL DESCRIPTION

The AVR Assistant will assist the commissioner of EMRI Automatic Voltage Regulators with

configuration and analysis. The handheld device features:

o Live display of measurements.

o Live readout of protections.

o Reading of advanced settings.

o Modifying of advanced settings.

o Transfer of advanced settings.

Page 6: AVR Assistant Manual - EMRI · EMRI standard CAN HS connectors are not capable of powering the AVR Assistant, unless otherwise specified. The AVR Assistant comes with a dedicated

Generator rewinding & repair. Voltage regulator products

Version 1.7 Page 6 of 24

Electronics AVR Assistant Manual

3. DEVICE OVERVIEW

Figure 1 Front view

Figure 2 Back view

Figure 3 Bottom view

4x case screws

Battery cover screw

Battery cover screw

Including magnet

+

-

+

-

+

-

3x AAA battery cover

AVR AssistantFirmware: V0.x

Nr: 1234Hardware: V0.x

Prod: 0106 Ede NL

Display

Enter key

Power key

Communications and

power port

Directional keys

12

5 m

m

69 mm

Clear or Return

33

mm

Page 7: AVR Assistant Manual - EMRI · EMRI standard CAN HS connectors are not capable of powering the AVR Assistant, unless otherwise specified. The AVR Assistant comes with a dedicated

Generator rewinding & repair. Voltage regulator products

Version 1.7 Page 7 of 24

Electronics AVR Assistant Manual

4. SPECIFICATIONS

Symbol Parameter Condition Min. Typ Max. Unit

Vbat Batteries 3 x AAA Alkaline 3 4 5 Vdc

Vext External power Isolated 5 or 12 Vdc 5

12 13

VDC

Tamb Operating temperature 95 % RHD non-condensing -10 +50 °C

Tstg Storage temperature 95 % RHD non-condensing -40 +70 °C

CAN_H CAN_L

Can data lines Maximum values Nominal ISO 11898

-2 8 V

Rcan Can termination 120 0.2

Ω W

Width 69 mm

Height With and without cable 125 170 mm

Depth 34 mm

Weight With and without batteries 140 170 G

Cable lenght 1.8 M

IP-code Cable is attached. IP41 Table 1 Recommended operating conditions

CONSIDERATIONS o Avoid leaving the handset in direct sunlight.

o Avoid leaving the device unprotected from weather influences.

o Avoid exposing the device to physical shocks and vibrations.

4.1. CONNECTOR

The main connector of type Binder Series 719 (09-9766-30-04) has the following pinout.

Notch

GND Vext

CAN_L CAN_H Table 2 Connector pinout, see figure 3

Page 8: AVR Assistant Manual - EMRI · EMRI standard CAN HS connectors are not capable of powering the AVR Assistant, unless otherwise specified. The AVR Assistant comes with a dedicated

Generator rewinding & repair. Voltage regulator products

Version 1.7 Page 8 of 24

Electronics AVR Assistant Manual

4.2. BASIC FEATURES

The following chapters describes the basic features for operating the device.

4.2.1. POWER

To power the handset on, batteries must be inserted or an external power supply must be

connected.

Supported battery types are:

AAA Alkaline non-rechargeable.

AAA Rechargeable NiCd or NiMH.

AAA Lithium batteries are not supported.

The handset is turned on by pressing the Power button:

The handset is turned off by holding the Power button for more than 2 seconds in any menu.

4.2.2. POWER SAVING

Because the handset is battery powered, it has some basic power saving features.

The handset will dim backlight after 10 seconds of inactivity. The handset will eventually turn off

completely. During live view the timeout is extended when the communications are active.

A special condition exist when AVR advanced setting data was read. Instead of powering off, the

handset will enter a memory retain mode for 1 hour before turning off.

This is to prevent accidental loss of settings memory while moving to a new AVR.

Average battery life is estimated at 20 hours on-time.

Page 9: AVR Assistant Manual - EMRI · EMRI standard CAN HS connectors are not capable of powering the AVR Assistant, unless otherwise specified. The AVR Assistant comes with a dedicated

Generator rewinding & repair. Voltage regulator products

Version 1.7 Page 9 of 24

Electronics AVR Assistant Manual

4.2.3. STATUS BAR

Figure 4 Annonated statusbar

The status bar at the top of the screen shows general information, such as:

The battery level.

External power

3/3 capacity

2/3 capacity

1/3 capacity

Batteries near depletion.

The access level. With User, Service or OEM.

The backlight brightness, from 1 to 5. (default 3)

Current advanced setting memory status.

o Empty, there is no avr settings data. (5 min power-off timeout)

o Ready, there is valid avr data. (1 hour power off timeout)

o Modified, modified data in buffer. (1 hour power off timeout)

The connection status.

o Offline, there is no link.

o Idle, there is a link, without active communication. (eg: the AVR is off) or there

are devices active on the CAN bus that are not supported by the AVR Assistant.

o Online, there is active communication with the AVR.

o Error, there is a link error. Refer to Troubleshooting.

4.2.4. RESET

A hard reset can be generated by holding the and for more than 5 seconds.

4.2.5. DISPLAY BRIGHTNESS

In the main menus, the LCD brightness can be adjusted to one of the 5 levels.

Hold or to adjust the brightness level.

1 is the lowest level and 5 is the highest level. Using brightness level 5 will deplete the batteries at

a faster rate than the default level 3.

U Offline3 Empty

Battery status

Access Level

Backlight

Memory status

Link status

Page 10: AVR Assistant Manual - EMRI · EMRI standard CAN HS connectors are not capable of powering the AVR Assistant, unless otherwise specified. The AVR Assistant comes with a dedicated

Generator rewinding & repair. Voltage regulator products

Version 1.7 Page 10 of 24

Electronics AVR Assistant Manual

4.2.6. MENU NAVIGATION

When the device is turned on, the main menu is shown on the display.

Navigate the menu using the arrow keys: .

Pressing will navigate to or execute the selected item. Such as “Advanced Settings” in

screenshot 1.

Pressing the shift button will show the description texts. Which also appears after 3 seconds

of inactivity.

The menu rolls over from bottom to top, or vice versa, when scrolling.

U Online

Main Menu

Select device

Information

Settings

Settings

Advanced settings

->

->

->

->

->

Modify device settings

Screenshot 1 Main menu.

Page 11: AVR Assistant Manual - EMRI · EMRI standard CAN HS connectors are not capable of powering the AVR Assistant, unless otherwise specified. The AVR Assistant comes with a dedicated

Generator rewinding & repair. Voltage regulator products

Version 1.7 Page 11 of 24

Electronics AVR Assistant Manual

5. CONNECTIVITY

Most EMRI products are fitted with a 4 way green connector. This is an EMRI standard CAN High

Speed connector. The AVR Assistant interfaces with EMRI products using the CAN High Speed

interface.

Figure 5 CAN HS connector on LXCOS Figure 6 CAN HS connector on LX449

The CAN1-CAN2 jumper must be linked to ensure a reliable connection with the AVR Assistant.

Most EMRI standard CAN HS outputs are electrically isolated from all other terminals, except in the

AVR Assistant itself. The DC GND used to externally power the AVR Assistant unit is equal to CAN

GND.

An extern power supply (12V DC) must be electrically isolated for use with the AVR Assistant.

EMRI standard CAN HS connectors are not capable of powering the AVR Assistant, unless

otherwise specified.

The AVR Assistant comes with a dedicated 1.8 meter data cable to connect to an EMRI standard

CAN HS connector.

The connector is polarized using arcs on one side of the connector shrouding.

Figure 7 AVR Communication cable

Page 12: AVR Assistant Manual - EMRI · EMRI standard CAN HS connectors are not capable of powering the AVR Assistant, unless otherwise specified. The AVR Assistant comes with a dedicated

Generator rewinding & repair. Voltage regulator products

Version 1.7 Page 12 of 24

Electronics AVR Assistant Manual

6. FEATURES

This chapter describes the features of the AVR Assistant.

6.1. AVR ADVANCED SETTINGS

The AVR Assistant is capable of reading the advanced settings of the products listed in the “Select

Device” menu.

The supported version will be listed in the following menu.

If your version is not listed, please contact EMRI.

U Online

Main Menu

Select device

Information

Settings

3

4

->

->

->

->

->

Select a device

to view or modify

U Online

Select device

LX449

LXCOS

LX321

AFD

Advanced settings

->

->

->

->

->

Any EMRI LX449 AVR

U Online

Version?

V1.4.0.0

V1.3.9.0

V1.3.7.0

V1.3.6.0

V1.3.5.0

->

->

->

->

->

Any EMRI AVR

V1.3.4.0->

Page 13: AVR Assistant Manual - EMRI · EMRI standard CAN HS connectors are not capable of powering the AVR Assistant, unless otherwise specified. The AVR Assistant comes with a dedicated

Generator rewinding & repair. Voltage regulator products

Version 1.7 Page 13 of 24

Electronics AVR Assistant Manual

6.1.1. SAFETY

When reading or writing an AVR it is important that the following conditions are met:

The AVR is not regulating.

(eg: the AVR On link is removed or the exciter wires are disconnected)

The AVR is not hardware self-excited. (eg: the exciter wires are removed)

The AVR is powered by either remanence voltage or an external source.

It is possible to put the AVR in a mode not supported by the current application.

This might damage equipment or even cause injury to personnel.

Refer to the user manual of the AVR for details about applicability of AVR advanced settings.

6.1.2. COMPATIBILITY

The AVR Assistant is not capable of determining the AVR type, and AVR Version attached.

The user of the AVR Assistant must ensure the correct version of AVR is selected prior to reading

or writing settings from or to the AVR.

It is not possible to connect multiple AVR units to a single CAN-bus.

All EMRI AVR-units use CAN-bus at a speed of 250 Kilobit per second.

Contact EMRI for more details.

Page 14: AVR Assistant Manual - EMRI · EMRI standard CAN HS connectors are not capable of powering the AVR Assistant, unless otherwise specified. The AVR Assistant comes with a dedicated

Generator rewinding & repair. Voltage regulator products

Version 1.7 Page 14 of 24

Electronics AVR Assistant Manual

6.1.3. READING ADVANCED SETTINGS

To read settings follow these steps:

1. Choose Select Device.

2. Select the device and the version.

3. Choose Advanced settings.

Screenshot 2 Advanced settings menu

4. Select Read settings.

Depending on the product and version, extra requirements will show.

Such as:

a. AVR Idle.

The AVR On link must be removed.

This is sufficient for LXCOS from V1.2.0.0 and LX449 from V1.5.3.0.

b. MSelect 0.

In older versions, access to advanced settings is only possible by setting MSelect to 0.

This way, when linking opt, no changes are made by the build-in advanced settings

method.

c. OPT linked.

In older versions, access to advanced settings require a physical connection to be

made.

Screenshot 3 Reading data

5. The AVR will be read.

6. Now it is possible to View the settings, or write to clipboard or other product.

The clipboard can be used to temporarily store the settings of an AVR to read and browse

a different AVR. The clipboard is maintained as long as the AVR Assistant is powered on.

U Online

Settings

Read settings

Browse settings

Write settings

Copy to clipboard

Paste from clipboard

->

->

->

->

->

Retrieve settings from AVR

Restart AVR->

Empty

U Online

MSELECT 0 ?

AVR IDLE ?

OPT LINKED ?

Empty

yes

yes

no

Page 15: AVR Assistant Manual - EMRI · EMRI standard CAN HS connectors are not capable of powering the AVR Assistant, unless otherwise specified. The AVR Assistant comes with a dedicated

Generator rewinding & repair. Voltage regulator products

Version 1.7 Page 15 of 24

Electronics AVR Assistant Manual

6.1.4. BROWSE ADVANCED SETTINGS

When settings are read. It is possible to view them.

Select “Browse Settings” in the advanced settings menu, and several pages of settings shown on

the screen.

U Online1/3

Build-up gain

Loop gain

More info, press shift

[ Stability (1/1) ]

1x

1.000x

Ready

Screenshot 4 Browse AVR settings

Change the current selected item using or .

Move pages using and and return to the menu’s using .

Press to show more info about the selected item.

Press to edit the selected item.

While editing, press to save the value, or press to discard any changes.

Page 16: AVR Assistant Manual - EMRI · EMRI standard CAN HS connectors are not capable of powering the AVR Assistant, unless otherwise specified. The AVR Assistant comes with a dedicated

Generator rewinding & repair. Voltage regulator products

Version 1.7 Page 16 of 24

Electronics AVR Assistant Manual

6.1.5. EDITING ADVANCED SETTINGS

Note: the numerical settings using the multiplier is available since firmware V0.7.

Note: the numerical settings with a decimal point is available since firmware V1.2.

There are several types of settings to edit. They are described

below.

In the editing screen, the variable name, current value, type of

min/max of the variable are shown.

As well as the currently assigned keypad layout.

BOOLEAN A Boolean setting can be set to On or Off using Up or Down.

Press Ok to save and C to return without changes.

LIST A list is a set of predefined settings of which one can be selected

using Up or Down.

Press Ok to save and C to return without changes.

NUMERICAL A numerical setting can be set anywhere between min and max.

Use the Up and Down keys to add or subtract the current multiplier

shown after the ± sign.

Use Left or Right to increase or decrease the multiplier.

Press Ok to save and C to return without changes.

Some numerical settings work with a decimal point.

Operation is identical.

U Online1/3

Edit: Analog In

Off

Ok

list

U Online1/3

Edit: UnderFreq (+/-)

1.1 Hz

+

++

-

-- Ok

Min: -12.7

Max: 12.7

±0.1

U Online1/3

Edit: Excitation Loss

ON

OFF

Ok

ON

on/off

Table 3 Boolean setting

Table 5 List setting

Table 4 Numerical setting

Page 17: AVR Assistant Manual - EMRI · EMRI standard CAN HS connectors are not capable of powering the AVR Assistant, unless otherwise specified. The AVR Assistant comes with a dedicated

Generator rewinding & repair. Voltage regulator products

Version 1.7 Page 17 of 24

Electronics AVR Assistant Manual

6.1.6. WRITING ADVANCED SETTINGS

To write settings follow these steps:

1. The memory status in the status bar must show “ready” or “modified”.

When no settings were read since power-on, the current memory is

invalid and the text shows “Empty”.

2. The connected AVR is of the same type and version as the AVR that

was read before.

3. Select Write settings in the Advanced settings menu.

A warning will be shown.

4. Hold OK.

Depending on the product and version, extra requirements will show.

Such as:

AVR Idle.

The AVR On link must be removed.

This is sufficient for LXCOS from V1.2.0.0 and LX449 from

V1.5.3.0.

MSelect 0.

In older versions, access to advanced settings is only possible by

setting MSelect to 0. This way, when linking opt, no changes are

made.

OPT linked.

In older versions, access to advanced settings require a physical

connection to be made.

The settings will be written.

5. Press Any key to continue when the Power Cycle warning shows.

6. Select Restart AVR or temporarily remove power from the AVR to

activate the new settings.

If the writing cycle of the setting is aborted before finishing, the settings in the AVR will be

corrupted and the AVR must be reset to factory defaults using the procedure described in

the AVR manual.

It is not supported to write settings of a different version to the AVR.

e.g. Transferring from V1.2.2.0 to V1.4.2.0 might result in erroneous behaviour.

U Online

Power Cycle AVR

to apply settings

Modified

Press any key

U Online

Write settings

Are you sure

Previous settings

Will be lost

Power cycle

Or

Restart AVR

To apply settings

Modified

Hold OK to conitnue

Press C to return.

Screenshot 6 Warning after writing

Screenshot 5 Warning before writing

Page 18: AVR Assistant Manual - EMRI · EMRI standard CAN HS connectors are not capable of powering the AVR Assistant, unless otherwise specified. The AVR Assistant comes with a dedicated

Generator rewinding & repair. Voltage regulator products

Version 1.7 Page 18 of 24

Electronics AVR Assistant Manual

6.2. PROCESS VARIABLES

When the AVR is regulating, the AVR Assistant provides a way to display process variables.

Process variables are values and setpoints used in the regulator process. These also include any

active error conditions.

Enter the Process Variables menu, select the applicable device and version, and the following

values will show on screen.

Screenshot 7 Process variables

Move pages using and and return to the menu’s using .

In this mode, the screen will dim in 5 minutes, and the unit will power off, or enter memory retain

mode after 15 minutes.

Viewing process variables severely reduces battery life.

U Online

Voltage (V)

400

Current (%)

10

Power Factor

1.00

AVR (°C)

29

Mode

Frequency

50.0I Exc. (mA)

250

Status

S-T (ohm)

5294

A1-A2 (mV)

290

normal voltage

U Online

Voltage pot

50.7%

Droop pot

49.5%

I gain

47.8%

Exc. Limit

53.9%

I-Limit

P gain

50.5%Acc-range

50.7%

Cosphi pot

PF / SE

0 / 1

M-Select

5 (175%)

0 (CV)

46.0%

U Online

ProtectionsOver VoltageOver CurrentExcitation ErrorOvertemp AVROvertemp GENPhase ErrorLoss of ExcLoss of CT

|

Page 19: AVR Assistant Manual - EMRI · EMRI standard CAN HS connectors are not capable of powering the AVR Assistant, unless otherwise specified. The AVR Assistant comes with a dedicated

Generator rewinding & repair. Voltage regulator products

Version 1.7 Page 19 of 24

Electronics AVR Assistant Manual

6.2.1. PROCESS VARIABLES PAGE 1: MEASUREMENTS

Generator voltage

Generator current in percentage of input channel.

Generator frequency in steps of 0.5 Hertz.

Generator power factor.

Exciter current in milliamperes.

AVR Temperature

AVR status.

AVR control mode.

A1-A2 Accessories voltage

S-T External potentiometer voltage

6.2.2. PROCESS VARIABLES PAGE 2: PROTECTIONS

This page shows active errors in red. Errors need to be active for a few seconds before shown.

Overvoltage.

Overcurrent.

Overexcitation.

Loss of voltage sensing.

Overtemperature AVR.

Overtemperature Generator.

Loss of Excitation.

Loss of current sensing.

6.2.3. PROCESS VARIABLES PAGE 3: SETPOINTS

Voltage potentiometer. (excluding fine potentiometer)

Droop potentiometer

P-gain potentiometer

I-gain potentiometer

Accessories range potentiometer

Excitation Limit potentiometer

Cosphi setpoint potentiometer

Current limit selector

Mode selector

PF/SE contacts.

Page 20: AVR Assistant Manual - EMRI · EMRI standard CAN HS connectors are not capable of powering the AVR Assistant, unless otherwise specified. The AVR Assistant comes with a dedicated

Generator rewinding & repair. Voltage regulator products

Version 1.7 Page 20 of 24

Electronics AVR Assistant Manual

7. SETTINGS

The AVR Assistant has additional features in the settings menu. These settings apply to the AVR

Assistant itself.

7.1.1. ACCESS LEVEL

The AVR Assistant contains 3 authority levels. These are used to allow access to specific settings,

such as AVR calibration values and internal process variables.

The three access levels are:

Level Statusbar code Code

User U 0000

Service S ****

OEM (Manufacturer) O **** Table 6 Access levels

The access level can be changed using the Access Level option in the Settings menu.

Enter the number using the numerical keypad, use shift to change key modes.

Screenshot 8 Access Level

Only use access codes when instructed by EMRI personnel.

7.1.2. FIRMWARE UPGRADE

The AVR Assistant has a bootloader to upgrade firmware. The bootloader process cannot be

aborted without removing the batteries. Specialized tooling is required to utilize the bootloader.

And is therefore protected by OEM code.

7.1.3. KEYPAD TEST

A keypad test feature is added to quickly test the keypad functionality. Hold the C button to exit.

U Online

_ _ _ _

1 3402

(Use Shift)

Enter 0000 for User Level

U Online

_ _ _ _

(Use Shift)

Enter 0000 for User Level

5 7896

Page 21: AVR Assistant Manual - EMRI · EMRI standard CAN HS connectors are not capable of powering the AVR Assistant, unless otherwise specified. The AVR Assistant comes with a dedicated

Generator rewinding & repair. Voltage regulator products

Version 1.7 Page 21 of 24

Electronics AVR Assistant Manual

8. TROUBLESHOOTING

A troubleshooting table is provided to solve the most frequent problems with the device.

Symptoms Cause Solution

The device does not turn on. The device only turns on for a short blink.

Empty batteries Replace batteries.

The display blinks on every operation Low batteries It is recommended to replace the batteries.

The device does not show anything on the display but the backlight is blinking.

Hardware failure Contact EMRI.

Link Idle is shown but the cable is attached.

CAN bus connectivity failure

Make sure the AVR is powered on.

Link error is shown but the cable is attached and the AVR is powered on.

CAN bus connectivity failure

Make sure the connectors and cables are not damaged.

My AVR version is not listed Version is not supported.

Contact EMRI for details.

Page 22: AVR Assistant Manual - EMRI · EMRI standard CAN HS connectors are not capable of powering the AVR Assistant, unless otherwise specified. The AVR Assistant comes with a dedicated

Generator rewinding & repair. Voltage regulator products

Version 1.7 Page 22 of 24

Electronics AVR Assistant Manual

9. KNOWN LIMITATIONS

This chapter describes known limitations of the AVR Assistant.

Page 23: AVR Assistant Manual - EMRI · EMRI standard CAN HS connectors are not capable of powering the AVR Assistant, unless otherwise specified. The AVR Assistant comes with a dedicated

Generator rewinding & repair. Voltage regulator products

Version 1.7 Page 23 of 24

Electronics AVR Assistant Manual

A.1 Contacts

Manufacturer: EMRI Electronics B.V. Morsestraat 10 6716 AH, Ede, Netherlands Tel: (+31) 0318 620427 Fax: (+31) 0318 634615 Website: www.emri.nl E-mail: [email protected]

Lucio Vicencio y CIA.LTDA Santiago, Chile Tel: +1-281-334-2904 Fax::+1-832-221-5642 Website: www.luciovicencio.cl

Email: [email protected]

Power Solutions, India Tel: +911 29404 1409 Fax::+911 29404 1410 Website: www.psolindia.com

Page 24: AVR Assistant Manual - EMRI · EMRI standard CAN HS connectors are not capable of powering the AVR Assistant, unless otherwise specified. The AVR Assistant comes with a dedicated

Generator rewinding & repair. Voltage regulator products

Version 1.7 Page 24 of 24

Electronics AVR Assistant Manual

A.1 Contacts

Manufacturer: EMRI Electronics B.V. Morsestraat 10 6716 AH, Ede, Netherlands Tel: (+31) 0318 620427 Fax: (+31) 0318 634615 Website: www.emri.nl E-mail: [email protected]

Ramtec Marine Systems LLC Kemah, Texas Tel: +1-281-334-2904 Fax::+1-832-221-5642 Website: www.ramtec-marine.com Email: [email protected]

Elektrisk Drivteknik EDT AB Kungälv, Sweden Tel: +46-705-28 20 60 Tel: +46-709-50 47 90 Website: www.edtab.se Email:[email protected]

Rafeining ehf Hafnarfjordur - Iceland Tel: +354 565 3049 Fax: +354 565 3048 Website: www.rafeining.is E-mail: [email protected]

Myren & Co. AB Askim - Sweden Tel: +46 317481860 Fax: +46 317481870 Website: www.myren.com E-mail: [email protected]

Marel Serwis Szczezin-Mierzyn - Poland Tel: +48 91 48 58 388 Fax: +48 91 48 79 948 Website: www.marel.szczecin.pl E-mail: [email protected]

Frydenbø Electric A/S Bergen - Norway Tel: + 47 55 34 91 00 Fax: + 47 55 34 91 10 Website: www.frydenbo.no E-mail: [email protected]

KDU Technical Services Sharjah - United Arab Emirates Tel: +971-6-5575480 Fax: +971-6-5575490 Website: www.kdutech.ae E-mail: [email protected]

An-Elec Sp. z o.o. Gdynia - Poland Tel: +48 58 668 44 00 Fax: +48 58 668 44 66 Internet: http://an-elec.pl E-mail: [email protected]

Yneldo Electronics Roodepoort - South Africa Tel: +27(0)117637053 Fax: +27(0)117634212 Internet: www.yneldo.com E-mail: [email protected]

Cyclect Electrical Engineering Singapore Tel: +65 6868 6013 Fax: +65 6863 6260 Internet: www.cyclect.com.sg E-mail: [email protected]

MJR Controls Stockton on Tees - United Kingdom Tel: +44 1642 762 151 Fax: +44 1642 762 502 Internet: www.mjrcontrols.com Email: [email protected]

Stavros Kassidiaris S.A. Piraeus - Greece Tel: +30 210 4636000 Fax: +30 210 4624471 Internet: www.kassidiaris.gr E-mail: [email protected]

Electricidad Juan Betancor S.L. Las Palmas – Canary Islands Tel: +34 928327054 Fax: +34 928327055 Email:[email protected]

North Star USA inc. Fort Lauderdale, Florida Tel: +1 800 571-4453 Fax: +1 3052985727 Email:[email protected]

See our website www.emri.nl for local suppliers