Automatic Programmable Melting Point Apparatus
Model : µThermoCal50 | Make : ANALAB | HSN Code : 90278090
Supplied with Melting Capillary One End Sealed (Pack of 100 Nos.), Boiling Tube,
Sample Tapping Rod, Two Melting Standards 49ºC and 238 ºC, Forceps, Power Cord,
Operational & Instruction Manual with SOP, Calibration Test Certicate (without
Traceability Document) and Warranty Card.
www.analab.co.in
Technical Specication:
www.analab.co.in
Temperature Controller : Microcontroller Based
Detection Principal of Melting Point : Automatic by Silicon Photodiode
Heating Media : By Aluminum Block
Temperature Range : +5°C above ambient to 350°C
Readability : 0.1°C
Accuracy of Temperature : A) +5°C above ambient to 200°C………… ± 0.2°C
B) 200°C to 350°C …………................…..± 0.5°C
Overheating Protection : In-built - Set by user in method parameter
Heating Before EMP* : Automatic heating depending upon temperature difference
Heating Rate : 0.2, 0.5, 1.0, 2.0, 3.0 & 5.0 ºC/min -Selectable
Display : 20 x 4 Line Alphanumeric Backlit LCD
Detection of Melting : A) Melting Point B) Melting Range
Melting Point Average Reading : Display of average reading for three same sample
Type of Detection : Auto / Manual
Detection of Boiling Point : Manually by viewing in TFT Display – One Boiling Tube
Magnication : 5X
Keypad : Membrane waterproof polycarbonate soft touch keypad compliance
to UL/ROHS
Viewing of Melting Capillary : Camera with TFT Display for viewing Melting Capillary and Boiling
Tube in CAPillaryVIEW
Method : 40 method for Melting & 10 for Boiling sample with View, Print &
Delete facility
Calibration : Seven Calibration Standards
(As per Pharmacopeia AND / OR Customer's known standards)
Melting Capillary Ports : Three - one sample at a time
Glass Melting Capillary : OD: 1.4 to 1.6mm | ID: 1.0mm| Length: 100mm | One End Sealed
Sample Filling Height : 3mm
Password Protection : Date & Time, Calibration Mode, Delete Stored Analysis Data and
Restore Factory Set Calibration Data
Maximum Heating Time : Around 6 minutes from 50°C to 350°C
Maximum Cooling Time : Around 6 minutes from 350°C to 50°C
Temperature Sensor : Duplex PT-100, One for internal used and One for calibration purpose
by external source
www.analab.co.in
Calibration Data Storage : Non Volatile Memory which can store, print & view calibration date &
time, company name, standard name with temperature, actual
reading of standard temperature, calibration done by, instrument
serial number, instrument ID, model number etc.
Searching of data : With record number, batch number, Date and by scrolling
Report Formation : As per GLP Compliance
Analysis Data Storage : More than 750 analysis data can be stored with corresponding
calibration data with batch number/ID, users name, sample name,
instrument serial number, instrument ID, model number, calibration
point, calibration date & time, analysis date & time, upper & lower
calibration standard based on sample temperature, heating rate,
melting range (MP1# & MP2#), signature etc. Secondly, analysis
data can print directly from analysis mode without saving analysis
data in instrument.
Printer Interface : Centronics Parallel Port / Serial (RS-232 C) Port
Printer : 40 / 80 Column Dot Matrix Printer
(Printer have to procure optionally)
Power Supply : 230 VAC ±10%, 50 Hz
Speci�cation & design are subject to change without prior notice for our continuous development.
www.analab.co.in
Plot No. 361, Sayajimarket,
N. H. No. 8 By-Pass,
At & Post : Sayajipura,
Vadodara - 390019, Gujarat, India.
+91 265 2514404, 9714503393
www.analab.co.in, www.analabscienti�c.com
CIN: U33110GJ2005PTC046555
AN ISO 9001 : 2015 CERTIFIED CO.
Optional
1. IQ / OQ / PQ / DQ Documents
Note:
a) Please provide Customer's Name & Address prior to dispatch.
b) To perform onsite documentation, visit charge will be extra depending upon place of Installation.
c) Traceability documents along with installation report.
d) Standards / Chemicals / Buffers required to perform documents are in customer's scope.
e) We don't send documents by email.
2. Traceability documents
3. Calibration Standard Kit for Melting Point Apparatus - (ANALAB Certied)
Consist of 07 Melting Standards as specied bellow, Melting Capillary One End Sealed 100 mm Long
(Pack of 100) – 02 Pack, Boiling Tube – 02 Nos., Sample Tapping Rod, Forceps, Glass for Heating Block
– 02 Nos., Glass for Camera – 02 Nos.
Melting Standards :
a) Benzophenone - Melting Range : 47 - 49 ºC, Capacity : 5gm
b) Vanillin - Melting Range : 81 - 83 ºC, Capacity : 5gm
c) Benzoic Acid - Melting Range : 121.5 - 123.5 ºC, Capacity : 5gm
d) Sulfanilamide - Melting Range : 164 - 167 ºC, Capacity : 5gm
e) Caffeine - Melting Range : 235 - 238 ºC, Capacity : 5gm
f) Phenolphthalein - Melting Range : 258 - 262 ºC, Capacity : 5gm
g) Potassium Nitrate - Melting Range : 334 – 336 ºC, Capacity : 5gm
Note: Melting Range / Melting Point specied on standard bottle is re-pack based on the values of batch
number specied by manufacturer and their COA.
4. GMP Enclosure made up of Stainless Steel
5. Keypad Screen Guard