    ASCLD/LAB

    Electronic Criteria File2005 Accreditation Manual

    Insert Name of Laboratory

    Insert Name of Laboratory Director

    Insert Name of Individual Responsible for Preparation of Criteria File

    Date of Preparation

    1

  • 8/2/2019 ASCLD Electronic Criteria File 2005 Accreditation Manual - Copy

    2/40

    ASCLD/LAB-Legacy

    Electronic Criteria File

    Instructions:

    Each criterion must be graded as YES, NO orN/A. .

    In addition to grading each criterion please provide, where applicable, the three categories of

    information requested in the table following each criterion. The information will enable the

    Inspection Team to determine compliance with accreditation requirements in accordance with thelaboratorys written quality system. The three categories of information are:

    Quality System Requirement(s) Location(s) of Compliance Record(s)

    Objective Evidence Observed

    Definitions/information category descriptions:

    A laboratorys quality system includes all written policies, procedures and directives

    regardless of title which contain requirements for the management of the laboratory.

    A quality system requirement(s) is any specific component(s) of the laboratorys qualitysystem describing the laboratory's manner of compliance with the clause (e.g., a section of

    the Quality manual, a lab procedure, or a work instruction).

    The location(s) of compliance record(s) is the location(s) where quality system compliance

    records are maintained and available for review (e.g., proficiency testing records, court

    monitoring records, audits and case records).

    Objective Evidence Observed is a description of the evidence reviewed during the

    laboratorys internal inspection/audit conducted in preparation of the application for

    accreditation. Examples of evidence observed include a quality system document (i.e. policy

    on proficiency testing), compliance records, case records, security logs, direct observation,

    and employee or customer interview responses.

    Mandatory responses identified: . If the category title is printed in bold font, a response is

    required to successfully complete the criteria file. For some criteria it may not be necessary toprovide one or more of the three categories of information requested in each table. When the

    category of information title is printed in normal font, a response is not mandatory but may be

    appropriate for a particular laboratorys quality system. If you believe a category response is notapplicable, simply indicate N/A in the appropriate space.

    An information category response is not required for any criterion you deem N/A or grade

    as a NO, unless you believe an explanation may clarify an issue for the Inspection Team.

    2

  • 8/2/2019 ASCLD Electronic Criteria File 2005 Accreditation Manual - Copy

    3/40

    ASCLD/LAB inspectors will have the option to use the same process identified by the laboratory

    to establish compliance, or may elect an alternative process.

    Providing a fully electronic criteria file (with hyperlinks to policies/procedures) is preferred,

    but not required for accreditation. A laboratory may create a hard copy Criteria File for on-site

    review by the Inspection Team, or may provide a combination of electronic and hard copy data.In any case, this document must be completed and submitted to the Team Captain prior to the on-

    site inspection.

    When submitting a criteria file document that is not accompanied by fully electronic

    attachments, please contact the assigned Team Captain so there can be some agreement as

    to what hard copy attachments to include.

    When hyperlinks are provided, you may hyperlink to any type of file provided the software

    needed to open the file is in common use (i.e., Acrobat Reader, Microsoft PowerPoint, etc.). If

    you are unable to set an appropriate bookmark to locate a specific topic within a multi-page

    document, please specify the page(s) and section number(s) to be reviewed. For example:

    Quality Manual, Section V. page 12 14 / Proficiency testing.

    The purpose of the criteria file is to provide a systematic method for advising the InspectionTeam of what the laboratory believes to be evidence of compliance with each applicable criterion

    in accordance with its quality system. Please feel free to include as much information as you

    believe is necessary to explain and/or direct inspectors to evidence of compliance.

    Please feel free to contact the ASCLD/LAB Executive Director or Legacy ProgramManager at 919-773-2600 should you have any questions about completing the

    Criteria File document.

    Important Document Navigation Instructions:

    After selecting Yes, No, or N/A, please insert appropriate responses in the table provided

    for each accreditation requirement. PLEASE DO NOT INSERT RESPONSES IN ANYSHADED ROW WITHIN A TABLE. You may use the tab key to move from row to row

    within the table. However, to prevent adding additional rows to the table, please do not use the

    tab key to exit the last row of the table. Rather, please use the mouse or the down arrow key to

    move to the next Yes, No, or N/A response.

    All blank, non-shaded rows within each table will expand as you type.

    This document is the property of ASCLD/LAB. Laboratories engaged in the accreditation

    process, including those preparing to apply for accreditation are encouraged to use this

    guide. Its use for fee or profit is prohibited by copyright law.

    3

  • 8/2/2019 ASCLD Electronic Criteria File 2005 Accreditation Manual - Copy

    4/40

    All of the following criteria must be scored YES, NO or N/A:

    1.1.1.1 (I) 1DOES THE LABORATORY HAVE A WRITTEN STATEMENT OF ITSOBJECTIVES?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.1.1.2 (I) 1DO THE OBJECTIVES APPEAR TO BE RELEVANT TO THE NEEDS OF

    THE COMMUNITY SERVICED BY THE LABORATORY?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.1.1.3 (D) 1DOES THE LABORATORY STAFF UNDERSTAND AND SUPPORT THE

    OBJECTIVES?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.1.2.1 (I) DOES THE LABORATORY OR ITS PARENT AGENCY HAVE A FORMAL

    WRITTEN BUDGET?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    4

    STANDARDS AND CRITERIA

  • 8/2/2019 ASCLD Electronic Criteria File 2005 Accreditation Manual - Copy

    5/40

    1.1.2.2 (I) IS THE BUDGET ADEQUATE TO MEET THE WRITTEN OBJECTIVES?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.1.2.3 (E) DOES CLEARLY WRITTEN AND WELL UNDERSTOOD PROCEDURE

    EXIST FOR HANDLING AND PRESERVING THE INTEGRITY OF

    EVIDENCE?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.1.2.4 (E) DOES CLEARLY WRITTEN AND WELL UNDERSTOOD PROCEDURE

    EXIST FOR LABORATORY SECURITY?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.1.2.5 (E) DOES CLEARLY WRITTEN AND WELL UNDERSTOOD PROCEDURE

    EXIST FOR THE PREPARATION, STORAGE, SECURITY AND

    DISPOSITION OF CASE RECORDS AND REPORTS?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.1.2.6 (E) DOES CLEARLY WRITTEN AND WELL UNDERSTOOD

    DOCUMENTATION OR PROCEDURE EXIST FOR THE CONTROL OF

    MATERIALS AND SUPPLIES?

    5

  • 8/2/2019 ASCLD Electronic Criteria File 2005 Accreditation Manual - Copy

    6/40

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.1.2.7 (E) DOES CLEARLY WRITTEN AND WELL UNDERSTOOD PROCEDUREEXIST FOR THE MAINTENANCE ANDCALIBRATION OF EQUIPMENT

    AND INSTRUMENTS?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)

    Objective Evidence Observed1.1.2.8 (E) DOES CLEARLY WRITTEN AND WELL UNDERSTOOD PROCEDURE

    EXIST FOR THE OPERATION OF INDIVIDUAL CHARACTERISTIC

    DATABASES?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)

    Objective Evidence Observed

    1.1.2.9 (D) DOES CLEARLY WRITTEN AND WELL UNDERSTOOD

    DOCUMENTATION EXIST FOR JOB REQUIREMENTS AND

    DESCRIPTIONS?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)

    Objective Evidence Observed

    1.1.2.10 (D) DOES CLEARLY WRITTEN AND WELL UNDERSTOOD

    DOCUMENTATION OR PROCEDURE EXIST FOR PERSONNEL

    EVALUATIONS AND OBJECTIVES?

    6

  • 8/2/2019 ASCLD Electronic Criteria File 2005 Accreditation Manual - Copy

    7/40

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.1.2.11 (D) DOES CLEARLY WRITTEN AND WELL UNDERSTOOD

    DOCUMENTATION OR PROCEDURE EXIST FOR EMPLOYEECOMPLIANTS CONCERNING THE QUALITY SYSTEM?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.1.2.12 (I) DOES THE LABORATORY HAVE AND USE A MANAGEMENT

    INFORMATION SYSTEM?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.2.1.1 (D) DOES THE ORGANIZATIONAL STRUCTURE GROUP THE WORK AND

    PERSONNEL IN A MANNER THAT ALLOWS FOR EFFICIENCY OF

    OPERATION, TAKING INTO ACCOUNT THE INTERRELATION OF

    VARIOUS FORENSIC DISCIPLINES?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.2.1.2 (D) HAS THE LABORATORY DIRECTOR CONSIDERED AND TAKEN

    APPROPRIATE ACTION TO CORRECT ANY DISCREPANCIES WITH

    REGARD TO NUMBERS OF PERSONNEL WHEN GROUPING WORK AND

    RESOURCES?

    7

  • 8/2/2019 ASCLD Electronic Criteria File 2005 Accreditation Manual - Copy

    8/40

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.2.2.1 (I) IS THE LABORATORY DIRECTORS AUTHORITY WELL DEFINED?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.2.2.2 (I) DOES THE LABORATORY DIRECTOR HAVE AUTHORITYCOMMENSURATE WITH RESPONSIBILITIES?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.2.2.3 (I) IS THERE SUFFICIENT DELEGATION OF AUTHORITY?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.2.2.4 (I) IS AUTHORITY OF SUPERVISORS COMMENSURATE WITH THEIR

    RESPONSIBILITIES?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    8

  • 8/2/2019 ASCLD Electronic Criteria File 2005 Accreditation Manual - Copy

    9/40

    1.2.2.5 (I) IS EACH SUBORDINATE ACCOUNTABLE TO ONE AND ONLY ONE

    IMMEDIATE SUPERVISOR PER FUNCTION?

    Yes No Not Applicable

    Quality System Requirement(s):

    Location(s) of Compliance Record(s)Objective Evidence Observed

    1.2.2.6 (I) ARE PERFORMANCE EXPECTATIONS ESTABLISHED AND ARE THEY

    UNDERSTOOD BY LABORATORY PERSONNEL?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.3.1.1 (D) IS THERE CONSTRUCTIVE DISCUSSION BETWEEN SUPERVISORS AND

    SUBORDINATES?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.3.1.2 (I) DO SUPERVISORS CAREFULLY AND OBJECTIVELY REVIEW

    LABORATORY ACTIVITIES AND PERSONNEL?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)

    Objective Evidence Observed

    1.3.1.3 (D) DO THE SUPERVISORY TECHNIQUES ENCOURAGE CREATIVE,

    OBJECTIVE THINKING AND RECOGNIZE MERITORIOUS

    PERFORMANCE?

    9

  • 8/2/2019 ASCLD Electronic Criteria File 2005 Accreditation Manual - Copy

    10/40

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.3.2.1 (D) DOES AN EFFECTIVE MEANS OF COMMUNICATION EXIST WITHIN

    THE LABORATORY?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.3.3.1 (E) DOES THE LABORATORY HAVE AND USE A DOCUMENTED

    TRAINING PROGRAM IN EACH DISCIPLINE AND SUBDISCIPLINE FOR

    EMPLOYEES WHO ARE NEW, UNTRAINED OR IN NEED OF REMEDIAL

    TRAINING?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.3.3.2 (I) DOES THE LABORATORY HAVE AN EMPLOYEE DEVELOPMENT

    PROGRAM?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.3.3.3 (I) DOES THE FORENSIC LIBRARY CONTAIN CURRENT BOOKS,

    JOURNALS, AND OTHER LITERATURE DEALING WITH EACH

    FUNCTIONAL AREA?

    Yes No Not Applicable

    Quality System Requirement(s):

    10

  • 8/2/2019 ASCLD Electronic Criteria File 2005 Accreditation Manual - Copy

    11/40

    Location(s) of Compliance Record(s)Objective Evidence Observed

    1.3.3.4 (I) DOES A SYSTEM EXIST TO ENCOURAGE EACH EXAMINER TO

    REVIEW APPROPRIATE NEW LITERATURE?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.4.1.1 (E) DOES THE LABORATORY HAVE A WRITTEN OR SECUREELECTRONIC CHAIN OF CUSTODY RECORD WITH ALL NECESSARY

    DATA WHICH PROVIDES FOR COMPLETE TRACKING OF ALL

    EVIDENCE?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.4.1.2 (E) IS ALL EVIDENCE MARKED FOR IDENTIFICATION?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.4.1.3 (E) IS EVIDENCE STORED UNDER PROPER SEAL?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    11

  • 8/2/2019 ASCLD Electronic Criteria File 2005 Accreditation Manual - Copy

    12/40

    1.4.1.4 (E) IS EVIDENCE PROTECTED FROM LOSS, CROSS TRANSFER,

    CONTAMINATION AND/OR DELETERIOUS CHANGE?

    Yes No Not Applicable

    Quality System Requirement(s):

    Location(s) of Compliance Record(s)Objective Evidence Observed

    1.4.1.5 (E) IS THERE A SECURE AREA FOR OVERNIGHT AND/OR LONG-TERM

    STORAGE OF EVIDENCE?

    Yes No Not Applicable

    Quality System Requirement(s):

    Location(s) of Compliance Record(s)Objective Evidence Observed

    1.4.1.6 (E) HAS THE LABORATORY ESTABLISHED WHETHER INDIVIDUAL

    CHARACTERISTIC DATABASE SAMPLES ARE TREATED AS

    EVIDENCE, REFERENCE MATERIALS, OR EXAMINATIONDOCUMENTATION?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.4.1.7 (E) 1IS EACH INDIVIDUAL CHARACTERISTIC DATABASE SAMPLE

    UNDER THE CONTROL OF THE LABORATORY UNIQUELY

    IDENTIFIED?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    12

  • 8/2/2019 ASCLD Electronic Criteria File 2005 Accreditation Manual - Copy

    13/40

    1.4.1.8 (E) 11ARE INDIVIDUAL CHARACTERISTIC DATABASE SAMPLES

    PROTECTED FROM LOSS, CROSS TRANSFER, CONTAMINATION

    AND /OR DELETERIOUS CHANGE?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.4.1.9 (E) 111IS ACCESS TO INDIVIDUAL CHARACTERISTIC DATABASE

    SAMPLES RESTRICTED TO THOSE PERSONS AUTHORIZED BY THE

    LABORATORY DIRECTOR?

    Yes No Not ApplicableQuality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.4.2.1 (E) DOES THE LABORATORY HAVE A COMPREHENSIVE QUALITYMANUAL?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.4.2.2 (E) IS AN INDIVIDUAL DESIGNATED AS THE QUALITY MANAGER?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.4.2.3 (E) DID THE ACCREDITED LABORATORY CONDUCT AND DOCUMENTAN ANNUAL AUDIT OF ITS OPERATIONS AND SUBMIT AN ANNUAL

    13

  • 8/2/2019 ASCLD Electronic Criteria File 2005 Accreditation Manual - Copy

    14/40

    ACCREDITATION AUDIT REPORT TO ASCLD/LAB BY THE REQUIRED

    DEADLINE?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.4.2.4 (E) DOES THE LABORATORY CONDUCT AND DOCUMENT AN ANNUAL

    REVIEW OF ITS QUALITY SYSTEM?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.4.2.5 (E) ARE THE PROCEDURES USED GENERALLY ACCEPTED IN THE FIELD

    OR SUPPORTED BY DATA GATHERED AND RECORDED IN A

    SCIENTIFIC MANNER?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.4.2.6 (E) ARE NEW TECHNICAL PROCEDURES SCIENTIFICALLY VALIDATED

    BEFORE BEING USED IN CASEWORK AND IS THE VALIDATION

    DOCUMENTATION AVAILABLE FOR REVIEW?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.4.2.7 (E) ARE THE TECHNICAL PROCEDURES USED BY THE LABORATORY

    DOCUMENTED AND ARE THE DOCUMENTS AVAILABLE TO

    LABORATORY PERSONNEL FOR REVIEW?

    14

  • 8/2/2019 ASCLD Electronic Criteria File 2005 Accreditation Manual - Copy

    15/40

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.4.2.8 (E) ARE APPROPRIATE CONTROLS AND STANDARDS SPECIFIED IN THEPROCEDURES AND ARE THEY USED AND DOCUMENTED IN THE

    CASE RECORD TO ENSURE THE VALIDITY OF EXAMINATION

    RESULTS?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.4.2.9 (E) IS THE QUALITY OF THE STANDARD SAMPLES AND REAGENTS

    ADEQUATE FOR THE PROCEDURE USED?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)

    Objective Evidence Observed

    1.4.2.10 (E) DOES THE LABORATORY ROUTINELY CHECK THE RELIABILITY OF

    ITS REAGENTS?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.4.2.11 (I) ARE THE INSTRUMENTS/EQUIPMENT ADEQUATE FOR THE

    PROCEDURES USED?

    Yes No Not Applicable

    Quality System Requirement(s):

    15

  • 8/2/2019 ASCLD Electronic Criteria File 2005 Accreditation Manual - Copy

    16/40

    Location(s) of Compliance Record(s)Objective Evidence Observed

    1.4.2.12 (I) ARE THE INSTRUMENTS/EQUIPMENT IN PROPER WORKING ORDER?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.4.2.13 (E) ARE THE INSTRUMENTS/EQUIPMENT PROPERLY CALIBRATED?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.4.2.14 (E) DOES THE LABORATORY CREATE AND MAINTAIN, A UNIQUELY

    IDENTIFIED CASE RECORD FOR ALL EXAMINATION AND

    ADMINISTRATIVE DOCUMENTATION GENERATED AND/OR

    RECEIVED BY THE LABORATORY FOR EACH CASE INVOLVING THEANALYSIS OF EVIDENCE?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.4.2.15 (E) 1DOES THE LABORATORYS UNIQUE CASE IDENTIFIER APPEAR ON

    EACH PAGE OF EXAMINATION DOCUMENTATION, AND DOES THE

    HANDWRITTEN INITIALS (OR SECURE ELECTRONIC EQUIVALENT)

    OF THE PERSON GENERATING THE EXAMINATION

    DOCUMENTATION APPEAR ON EACH PAGE GENERATED BY THAT

    PERSON?

    Yes No Not Applicable

    Quality System Requirement(s):

    16

  • 8/2/2019 ASCLD Electronic Criteria File 2005 Accreditation Manual - Copy

    17/40

    Location(s) of Compliance Record(s)Objective Evidence Observed

    1.4.2.16 (E) 1ARE CONCLUSIONS AND OPINIONS IN REPORTS SUPPORTED BY

    DATA AVAILABLE IN THE CASE RECORD, AND ARE THE

    EXAMINATION DOCUMENTS SUFFICIENTLY DETAILED SUCH THAT,IN THE ABSENCE OF THE EXAMINER(S), ANOTHER COMPETENT

    EXAMINER OR SUPERVISOR COULD EVALUATE WHAT WAS DONE

    AND INTERPRET THE DATA?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.4.2.17 (E) 1IS EXAMINATION DOCUMENTATION OF A PERMANENT NATURE

    AND IS IT FREE OF OBLITERATIONS AND ERASURES?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.4.2.18 (E) 1HAS EACH PERSON(S) IN THE LABORATORY WHO ISSUED FINDINGS

    BASED ON EXAMINATION DOCUMENTATION GENERATED BY

    ANOTHER PERSON, COMPLETED A REVIEW OF ALL RELEVANT

    PAGES OF EXAMINATION DOCUMENTATION AND DOCUMENTED

    THE REVIEW IN THE CASE RECORD?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.4.2.19 (E) 1DOES THE LABORATORY GENERATE WRITTEN REPORTS FOR ALL

    ANALYTICAL WORK PERFORMED ON EVIDENCE, AND DO THE

    REPORTS CONTAIN THE CONCLUSIONS AND OPINIONS THAT

    17

  • 8/2/2019 ASCLD Electronic Criteria File 2005 Accreditation Manual - Copy

    18/40

    ADDRESS THE PURPOSE FOR WHICH THE ANALYTICAL WORK WAS

    UNDERTAKEN?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.4.2.20 (E) 1WHERE ASSOCIATIONS ARE MADE, IS THE SIGNIFICANCE OF THE

    ASSOCIATION COMMUNICATED CLEARLY AND QUALIFIED

    PROPERLY IN THE REPORT?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.4.2.21 (E) 11DOES THE NAME OF THE AUTHOR(S) APPEAR IN THE REPORT?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)

    Method(s) of Compliance Verification Used by Laboratory:

    1.4.2.22 (E) DOES THE LABORATORY HAVE, USE AND DOCUMENT A SYSTEM OF

    TECHNICAL REVIEW OF THE REPORTS TO ENSURE THAT THE

    CONCLUSIONS OF ITS EXAMINERS ARE REASONABLE AND WITHIN

    THE CONSTRAINTS OF SCIENTIFIC KNOWLEDGE?

    1

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.4.2.23 (E) DOES THE LABORATORY CONDUCT AND DOCUMENT

    ADMINISTRATIVE REVIEWS OF ALL REPORTS ISSUED?

    18

  • 8/2/2019 ASCLD Electronic Criteria File 2005 Accreditation Manual - Copy

    19/40

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.4.2.24 (E) DOES THE LABORATORY MONITOR THE TESTIMONY OF EACH

    EXAMINER AT LEAST ANNUALLY AND IS THE EXAMINER GIVENFEEDBACK FROM THE EVALUATION1?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.4.2.25 (E) IF THE LABORATORY HAS AN INDICATION OF A SIGNIFICANT

    TECHNICAL PROBLEM, IS THERE A PROCEDURE IN WRITING AND IN

    USE WHEREBY THE LABORATORY INITIATES A REVIEW AND TAKES

    ANY CORRECTIVE ACTION REQUIRED1?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)

    Method(s) of Compliance Verification Used by Laboratory:

    1.4.3.1 (E) DOES THE LABORATORY HAVE A DOCUMENTED PROGRAM OF

    PROFICIENCY TESTING?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.4.3.2 (E) DOES THE LABORATORY PARTICIPATE IN PROFICIENCY TESTING

    PROGRAMS CONDUCTED BY APPROVED TEST PROVIDERS, OR BY

    OTHER EXTERNAL PROVIDERS(S) WHEN NO APPROVED PROVIDER

    IS AVAILABLE?

    19

  • 8/2/2019 ASCLD Electronic Criteria File 2005 Accreditation Manual - Copy

    20/40

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.4.3.3 (I) WAS EACH EXAMINER PROFICIENCY TESTED ANNUALLY IN EACH

    SUBDISCIPLINE IN WHICH CASEWORK WAS PERFORMED?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    1.4.3.4 (I) DOES THE LABORATORY CONDUCT PROFICIENCY TESTING USING

    RE-EXAMINATION OR BLIND TECHNIQUES?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    2.1.1 (I) DOES THE LABORATORY DIRECTOR POSSESS A DEGREE IN A

    NATURAL SCIENCE, CRIMINALISTICS OR IN A CLOSELY RELATED

    FIELD, OR IS THE LABORATORY DIRECTOR SUPPORTED BY

    SCIENTIFIC PERSONNEL OF SUFFICIENT MANAGERIAL RANK AND

    AUTHORITY?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    2.1.2 (D) DOES THE LABORATORY DIRECTOR HAVE AT LEAST FIVE YEARS OF

    FORENSIC SCIENCE EXPERIENCE?

    Yes No Not Applicable

    Quality System Requirement(s):

    20

  • 8/2/2019 ASCLD Electronic Criteria File 2005 Accreditation Manual - Copy

    21/40

    Location(s) of Compliance Record(s)Objective Evidence Observed

    2.1.3 (D) DOES THE LABORATORY DIRECTOR HAVE SOME FORMAL TRAINING

    IN MANAGEMENT?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    2.1.4 (D) DOES THE LABORATORY DIRECTOR HAVE AT LEAST TWO YEARS OFMANAGERIAL EXPERIENCE?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    2.2.1 (E) DOES EACH CONTROLLED SUBSTANCE EXAMINER POSSESS A

    BACCALAUREATE OR ADVANCED DEGREE IN A NATURAL SCIENCE,CRIMINALISTICS OR IN A CLOSELY RELATED FIELD AND DOES

    EACH HAVE EXPERIENCE/TRAINING COMMENSURATE WITH THE

    EXAMINATIONS AND TESTIMONY PROVIDED?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    2.2.2 (E) DOES EACH CONTROLLED SUBSTANCE EXAMINER UNDERSTAND

    THE INSTRUMENTS, AND THE METHODS AND PROCEDURES USED?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)

    21

  • 8/2/2019 ASCLD Electronic Criteria File 2005 Accreditation Manual - Copy

    22/40

    Objective Evidence Observed

    2.2.3 (E) DID EACH CONTROLLED SUBSTANCE EXAMINER SUCCESSFULLY

    COMPLETE A COMPETENCY TEST PRIOR TO ASSUMING CASEWORK

    RESPONSIBILITY?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    2.2.4 (E) DID EACH CONTROLLED SUBSTANCE EXAMINER SUCCESSFULLY

    COMPLETE AN ANNUAL PROFICIENCY TEST?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    2.3.1 (E) DOES EACH TOXICOLOGY EXAMINER POSSESS A BACCALAUREATE

    OR ADVANCED DEGREE IN A NATURAL SCIENCE, TOXICOLOGY,

    CRIMINALISTICS OR IN A CLOSELY RELATED FIELD AND DOESEACH HAVE EXPERIENCE/TRAINING COMMENSURATE WITH THE

    EXAMINATIONS AND TESTIMONY PROVIDED?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    2.3.2 (E) DOES EACH TOXICOLOGY EXAMINER UNDERSTAND THEINSTRUMENTS, AND THE METHODS AND PROCEDURES USED?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)

    22

  • 8/2/2019 ASCLD Electronic Criteria File 2005 Accreditation Manual - Copy

    23/40

    Objective Evidence Observed2.3.3 (E) DID EACH TOXICOLOGY EXAMINER SUCCESSFULLY COMPLETE A

    COMPETENCY TEST PRIOR TO ASSUMING CASEWORKRESPONSIBILITY?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    2.3.4 (E) DID EACH TOXICOLOGY EXAMINER SUCCESSFULLY COMPLETE AN

    ANNUAL PROFICIENCY TEST?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    2.4.1 (E) DOES EACH TRACE EVIDENCE EXAMINER POSSESS A

    BACCALAUREATE OR ADVANCED DEGREE IN A NATURAL SCIENCE,

    CRIMINALISTICS OR IN A CLOSELY RELATED FIELD AND DOES

    EACH HAVE EXPERIENCE/TRAINING COMMENSURATE WITH THEEXAMINATIONS AND TESTIMONY PROVIDED?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    2.4.2 (E) DOES EACH TRACE EVIDENCE EXAMINER UNDERSTAND THE

    INSTRUMENTS, AND THE METHODS AND PROCEDURES USED?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    23

  • 8/2/2019 ASCLD Electronic Criteria File 2005 Accreditation Manual - Copy

    24/40

    2.4.3 (E) DID EACH TRACE EVIDENCE EXAMINER SUCCESSFULLY COMPLETE

    A COMPETENCY TEST IN EACH OF THE SUBDISCIPLINES PROCESSED

    PRIOR TO ASSUMING CASEWORK RESPONSIBILITY?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Method(s) of Compliance Verification Used by Laboratory:

    2.4.4 (E) DID EACH TRACE EVIDENCE EXAMINER SUCCESSFULLY COMPLETE

    AN ANNUAL PROFICIENCY TEST?

    Yes No Not ApplicableQuality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    2.5.1 (E) DOES EACH BIOLOGY EXAMINER POSSESS A BACCALAUREATE OR

    ADVANCED DEGREE IN A NATURAL SCIENCE, CRIMINALISTICS OR

    IN A CLOSELY RELATED FIELD AND DOES EACH HAVE EXPERIENCE/

    TRAINING COMMENSURATE WITH THE EXAMINATIONS AND

    TESTIMONY PROVIDED?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    2.5.2 (E) DOES EACH BIOLOGY EXAMINER PERFORMING DNA ANALYSISHAVE EDUCATION, TRAINING AND EXPERIENCE CONSISTENT WITH

    THOSE REQUIRED BY THE QUALITY ASSURANCE AUDITDOCUMENT?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)

    24

  • 8/2/2019 ASCLD Electronic Criteria File 2005 Accreditation Manual - Copy

    25/40

    Objective Evidence Observed2.5.3 (E) DOES EACH BIOLOGY EXAMINER UNDERSTAND THE INSTRUMENTS,

    AND THE METHODS AND PROCEDURES USED?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    2.5.4 (E) DID EACH BIOLOGY EXAMINER SUCCESSFULLY COMPLETE A

    COMPETENCY TEST PRIOR TO ASSUMING CASEWORK

    RESPONSIBILITY?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    2.5.5 (E) DID EACH BIOLOGY EXAMINER SUCCESSFULLY COMPLETE AN

    ANNUAL PROFICIENCY TEST?

    Yes No Not ApplicableQuality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    2.5.6 (E) DID EACH BIOLOGY EXAMINER PERFORMING DNA ANALYSIS

    SUCCESSFULLY COMPLETE TWO ANNUAL PROFICIENCY TESTS

    FROM AN APPROVED TEST PROVIDER?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    25

  • 8/2/2019 ASCLD Electronic Criteria File 2005 Accreditation Manual - Copy

    26/40

    2.6.1 (I) DOES EACH FIREARMS/TOOLMARK EXAMINER POSSESS A

    BACCALAUREATE DEGREE WITH SCIENCE COURSES?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    2.6.2 (E) DOES EACH FIREARMS/TOOLMARK EXAMINER UNDERSTAND THE

    INSTRUMENTS, AND THE METHODS AND PROCEDURES USED?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    2.6.3 (E) DID EACH FIREARMS/TOOLMARK EXAMINER HAVE EXTENSIVE

    TRAINING FROM A QUALIFIED EXAMINER AND DOES EACH HAVE

    EXPERIENCE COMMENSURATE WITH THE EXAMINATIONS AND

    TESTIMONY PROVIDED?

    Yes No Not Applicable

    Quality System Requirement(s):

    Location(s) of Compliance Record(s)Objective Evidence Observed

    2.6.4 (E) DID EACH FIREARMS/TOOLMARK EXAMINER SUCCESSFULLY

    COMPLETE A COMPETENCY TEST PRIOR TO ASSUMING CASEWORK

    RESPONSIBILITY?

    Yes No Not Applicable

    Quality System Requirement(s):

    Location(s) of Compliance Record(s)Objective Evidence Observed

    2.6.5 (E) DID EACH FIREARM/TOOLMARK EXAMINER SUCCESSFULLY

    COMPLETE AN ANNUAL PROFICIENCY TEST?

    26

  • 8/2/2019 ASCLD Electronic Criteria File 2005 Accreditation Manual - Copy

    27/40

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    2.7.1 (I) DOES EACH QUESTIONED DOCUMENT EXAMINER POSSESS ABACCALAUREATE DEGREE WITH SCIENCE COURSES?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    2.7.2 (E) DOES EACH QUESTIONED DOCUMENT EXAMINER UNDERSTAND

    THE INSTRUMENTS, AND THE METHODS AND PROCEDURES USED?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    2.7.3 (E) DID EACH QUESTIONED DOCUMENT EXAMINER HAVE EXTENSIVE

    TRAINING FROM A QUALIFIED EXAMINER AND DOES EACH HAVE

    EXPERIENCE COMMENSURATE WITH THE EXAMINATIONS AND

    TESTIMONY PROVIDED?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    2.7.4 (E) DID EACH QUESTIONED DOCUMENT EXAMINER SUCCESSFULLY

    COMPLETE A COMPETENCY TEST PRIOR TO ASSUMING CASEWORK

    RESPONSIBILITY?

    Yes No Not Applicable

    27

  • 8/2/2019 ASCLD Electronic Criteria File 2005 Accreditation Manual - Copy

    28/40

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    2.7.5 (E) DID EACH QUESTIONED DOCUMENT EXAMINER SUCCESSFULLY

    COMPLETE AN ANNUAL PROFICIENCY TEST?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    2.8.1 (I) DOES EACH LATENT PRINT EXAMINER POSSESS A BACCALAUREATE

    DEGREE WITH SCIENCE COURSES?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    2.8.2 (E) DOES EACH LATENT PRINT EXAMINER UNDERSTAND THEINSTRUMENTS, AND THE METHODS AND PROCEDURES USED?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    2.8.3 (E) DID EACH LATENT PRINT EXAMINER HAVE EXTENSIVE TRAINING

    FROM A QUALIFIED EXAMINER AND DOES EACH HAVE EXPERIENCE

    COMMENSURATE WITH THE EXAMINATIONS AND TESTIMONY

    PROVIDED?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)

    28

  • 8/2/2019 ASCLD Electronic Criteria File 2005 Accreditation Manual - Copy

    29/40

    Objective Evidence Observed

    2.8.4 (E) DID EACH LATENT PRINT EXAMINER SUCCESSFULLY COMPLETE A

    COMPETENCY TEST PRIOR TO ASSUMING CASEWORK

    RESPONSIBILITY?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    2.8.5 (E) DID EACH LATENT PRINT EXAMINER SUCCESSFULLY COMPLETE AN

    ANNUAL PROFICIENCY TEST?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    2.9.1 (E) DO TECHNICAL SUPPORT PERSONNEL MEET THE REQUIREMENTS OF

    THEIR JOB DESCRIPTIONS?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    2.9.2 (E) ARE THE TECHNICAL SUPPORT PERSONNEL JOB DESCRIPTIONS AND

    THE DUTIES PERFORMED IN AGREEMENT?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    29

  • 8/2/2019 ASCLD Electronic Criteria File 2005 Accreditation Manual - Copy

    30/40

    2.9.3 (E) DID EACH MEMBER OF THE TECHNICAL SUPPORT STAFF

    SUCCESSFULLY COMPLETE AN APPROPRIATE COMPETENCY TEST

    PRIOR TO ASSUMING CASEWORK RESPONSIBILITY?

    Yes No Not Applicable

    Quality System Requirement(s):

    Location(s) of Compliance Record(s)Objective Evidence Observed

    2.9.4 (E) DID ALL TECHNICAL SUPPORT PERSONNEL SUCCESSFULLY

    COMPLETE AN APPROPRIATE PROFICIENCY TEST, ANNUALLY?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    2.9.5 (E) DID ALL TECHNICAL SUPPORT PERSONNEL PERFORMING DNA

    ANALYSIS SUCCESSFULLY COMPLETE TWO ANNUAL PROFICIENCY

    TESTS FROM AN APPROVED TEST PROVIDER?

    Yes No Not Applicable

    Quality System Requirement(s):

    Location(s) of Compliance Record(s)Objective Evidence Observed

    2.10.1 (E) DO CRIME SCENE EXAMINERS MEET THE REQUIREMENTS OF THEIR

    JOB DESCRIPTIONS?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    2.10.2 (E) DOES EACH CRIME SCENE EXAMINER UNDERSTAND THE

    EQUIPMENT, METHODS AND PROCEDURES USED?

    30

  • 8/2/2019 ASCLD Electronic Criteria File 2005 Accreditation Manual - Copy

    31/40

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    2.10.3 (E) DID EACH CRIME SCENE EXAMINER HAVE EXTENSIVE TRAINING

    FROM A QUALIFIED EXAMINER AND DOES EACH HAVE EXPERIENCECOMMENSURATE WITH THE EXAMINATIONS/DOCUMENTATION

    AND TESTIMONY PROVIDED?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)

    Objective Evidence Observed2.10.4 (E) DID EACH CRIME SCENE EXAMINER SUCCESSFULLY COMPLETE A

    COMPETENCY TEST(S) PRIOR TO PRIMARY RESPONSIBILITY FOR

    THE EXAMINATION, DOCUMENTATION AND PROCESSING OF A

    CRIME SCENE?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    2.10.5 (E) DID EACH CRIME SCENE EXAMINER SUCCESSFULLY COMPLETE AN

    ANNUAL PROFICIENCY TEST?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)

    Objective Evidence Observed

    2.11.1 (I) DOES EACH EXAMINER POSSESS A BACCALAUREATE DEGREE WITH

    SCIENCE COURSES?

    Yes No Not Applicable

    31

  • 8/2/2019 ASCLD Electronic Criteria File 2005 Accreditation Manual - Copy

    32/40

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    2.11.2 (E) DOES EACH EXAMINER UNDERSTAND THE EQUIPMENT, PROGRAMS,

    METHODS AND PROCEDURES USED?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    2.11.3 (E) DOES EACH EXAMINER HAVE EXPERIENCE COMMENSURATE WITH

    THE EXAMINATIONS/DOCUMENTATION AND TESTIMONYPROVIDED?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    2.11.4 (E) DID EACH EXAMINER SUCCESSFULLY COMPLETE A COMPETENCY

    TEST IN EACH SUBDISCIPLINE PRIOR TO ASSUMING CASEWORK

    RESPONSIBILITY?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    2.11.5 (E) DID EACH EXAMINER SUCCESSFULLY COMPLETE AN ANNUAL

    PROFICIENCY TEST?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)

    32

  • 8/2/2019 ASCLD Electronic Criteria File 2005 Accreditation Manual - Copy

    33/40

    Objective Evidence Observed

    3.1.1 (I) DOES EACH EMPLOYEE HAVE ADEQUATE WORK SPACE TO

    ACCOMPLISH ASSIGNED TASKS?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    3.1.2 (D) IS THERE SUFFICIENT SPACE PROVIDED FOR STORAGE OF SUPPLIES,

    EQUIPMENT AND TOOLS?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    3.1.3 (I) IS THERE ADEQUATE SPACE AVAILABLE FOR EXAMINERS FOR

    WRITING REPORTS AND OTHER OFFICIAL COMMUNICATIONS?

    Yes No Not ApplicableQuality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    3.1.4 (I) IS THERE ADEQUATE AND APPROPRIATE SPACE AVAILABLE FOR

    RECORDS, REFERENCE WORKS AND OTHER NECESSARY

    DOCUMENTS?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    33

  • 8/2/2019 ASCLD Electronic Criteria File 2005 Accreditation Manual - Copy

    34/40

    3.1.5 (I) IS ADEQUATE SPACE AVAILABLE FOR EACH INSTRUMENT TO

    FACILITATE ITS OPERATION?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    3.1.6 (D) ARE ACCESSORIES STORED NEAR EACH INSTRUMENT TO

    FACILITATE ITS USE AND OPERATION?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    3.2.1 (I) DOES THE PHYSICAL DESIGN PERMIT THE EFFICIENT FLOW OF

    EVIDENCE FROM THE TIME OF ITS ACCEPTANCE UNTIL ITS PROPER

    DISPOSAL?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    3.2.2 (D) DO THE RELATIVE LOCATIONS OF FUNCTIONAL AREAS FACILITATE

    THE USE OF EQUIPMENT AND INSTRUMENTS?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    3.2.3 (I) IS THERE ADEQUATE AND PROPER LIGHTING AVAILABLE FOR

    PERSONNEL TO CARRY OUT ASSIGNED TASKS?

    Yes No Not Applicable

    34

  • 8/2/2019 ASCLD Electronic Criteria File 2005 Accreditation Manual - Copy

    35/40

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    3.2.4 (I) IS THERE ADEQUATE AND PROPER PLUMBING AND WIRING

    AVAILABLE AND ACCESSIBLE FOR PERSONNEL TO CARRY OUT

    ASSIGNED TASKS?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    3.2.5 (I) DOES THE LABORATORY HAVE PROPER GENERAL VENTILATION?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    3.2.6 (I) IS THE HEATING, COOLING AND HUMIDITY CONTROL IN THELABORATORY ADEQUATE?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    3.3.1 (E) IS ACCESS TO THE OPERATIONAL AREA OF THE LABORATORYCONTROLLABLE AND LIMITED?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    35

  • 8/2/2019 ASCLD Electronic Criteria File 2005 Accreditation Manual - Copy

    36/40

    3.3.2 (E) DO ALL EXTERIOR ENTRANCE/EXIT POINTS HAVE ADEQUATE

    SECURITY CONTROL?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    3.3.3 (E) DO ALL INTERNAL AREAS REQUIRING LIMITED/CONTROLLED

    ACCESS HAVE A LOCK SYSTEM?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    3.3.4 (E) IS DISTRIBUTION OF ALL KEYS, MAGNETIC CARDS, ETC.,

    DOCUMENTED AND IS DISTRIBUTION LIMITED TO THOSE

    INDIVIDUALS DESIGNATED BY THE LABORATORY DIRECTOR TO

    HAVE ACCESS?

    Yes No Not ApplicableQuality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    3.3.5 (E) IS THE LABORATORY SECURED DURING VACANT HOURS BY MEANS

    OF AN INTRUSION ALARM OR BY SECURITY PERSONNEL?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    3.3.6 (I) DOES THE LABORATORY HAVE A FIRE DETECTION SYSTEM?

    36

  • 8/2/2019 ASCLD Electronic Criteria File 2005 Accreditation Manual - Copy

    37/40

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    3.4.1 (I) DOES THE LABORATORY HAVE AN EFFECTIVE HEALTH ANDSAFETY PROGRAM DOCUMENTED IN A MANUAL?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    3.4.2 (I) IS AN INDIVIDUAL DESIGNATED AS THE HEALTH AND SAFETY

    MANAGER?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    3.4.3 (I) IS THE HEALTH AND SAFETY PROGRAM MONITORED REGULARLY

    AND REVIEWED ANNUALLY TO ENSURE THAT ITS REQUIREMENTS

    ARE BEING MET?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    3.4.4 (I) DOES THE LABORATORY HAVE AVAILABLE AND ENCOURAGE THE

    USE OF SAFETY DEVICES, PARTICULARLY THOSE REQUIRED BY ITS

    HEALTH AND SAFETY MANUAL?

    Yes No Not Applicable

    Quality System Requirement(s):

    37

  • 8/2/2019 ASCLD Electronic Criteria File 2005 Accreditation Manual - Copy

    38/40

    Location(s) of Compliance Record(s)Objective Evidence Observed

    3.4.5 (I) DOES THE LABORATORY HAVE PROPER EQUIPMENT AND

    MATERIAL AVAILABLE FOR THE HANDLING OF CARCINOGENIC,

    TOXIC AND/OR OTHER DANGEROUS MATERIAL SPILLS?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    3.4.6 (I) DOES THE LABORATORY HAVE SAFETY SHOWER AND EYE WASH

    EQUIPMENT IN APPROPRIATE LOCATIONS AND IN GOOD WORKING

    CONDITION?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    3.4.7 (I) ARE SUFFICIENT EXHAUST HOODS AVAILABLE TO MAINTAIN A

    SAFE WORK ENVIRONMENT?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    3.4.8 (I) ARE SUFFICIENT FIRST-AID KITS AVAILABLE AND STRATEGICALLY

    LOCATED?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)

    38

  • 8/2/2019 ASCLD Electronic Criteria File 2005 Accreditation Manual - Copy

    39/40

    Objective Evidence Observed3.4.9 (I) DOES THE LABORATORY HAVE AN ADEQUATE NUMBER OF

    PERSONNEL HOLDING CURRENT CERTIFICATION IN FIRST-AID?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    3.4.10 (I) IS APPROPRIATE SPACE PROVIDED FOR SAFE STORAGE OF

    VOLATILE, FLAMMABLE, EXPLOSIVE AND OTHER HAZARDOUS

    MATERIALS?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    3.4.11 (I) ARE THE EMERGENCY EXITS FROM THE LABORATORY ADEQUATE

    FOR SAFE EXIT IN AN EMERGENCY?

    Yes No Not ApplicableQuality System Requirement(s):Location(s) of Compliance Documentation:Objective Evidence Observed

    3.4.12 (D) IS THERE GENERAL CLEANLINESS AND APPARENT GOOD-

    HOUSEKEEPING IN THE LABORATORY?

    Yes No Not Applicable

    Quality System Requirement(s):Location(s) of Compliance Record(s)Objective Evidence Observed

    39

  • 8/2/2019 ASCLD Electronic Criteria File 2005 Accreditation Manual - Copy

    40/40

    SUMMATION OF CRITERIA RATINGS

    Total Total Total

    Possible Yes No

    Essential Important Desirable

    Areas sought for accreditation are as follows:

    Controlled Substances Biology

    Toxicology serology only

    drug identification screening only

    blood alcohol only nuclear DNA

    Trace Evidence mitochondrial DNA

    fire debris platform explosives other fibers offender database analysis

    GSR Latent Prints

    glass processing only

    hair AFIS databasepaint Crime Scene (optional)

    specify others: Firearms/Toolmarks Digital & Multimedia Evidence

    firearms computer forensics

    toolmarks forensic audio

    individual characteristic database video analysis

    Questioned Documents image analysis

    Impression evidence, specify under which disciplines Serial number restoration, specify under which disciplines

    Prepared by:

    L b t Di t