Ankersmid peliter cooler datasheet 0

2
Peltier Cooler Gas Sample Cooler Through lowering the dew point of humid gas, the Ankersmid Peltier Cooler is able to reduce the risk of condensation occurring in the analyser The Peltier Cooler is equipped with an adjustable dew point stabiliser. The stabiliser is set at 4 o C as standard with the option to adjust the value to between 1 and 15 o C at any time. Digitally Controlled The digital controlled cooler has many control and warning features like programmable alarms, mA-output, digital inputs and Modbus or RS485 communication. Using smart technology the alarm status of the cooler changes when the temperature deviates ± 3 °C from a set point Humidifies Calibration Gases The peltier cooler is equipped with a silicon coated (glass) heat exchanger, as standard with the added availability of PFA coating as an alternative. Using a special liquid inlet, the Peltier Cooler humidifies calibration gases to avoid volumetric errors. KEYBENEFITS Removable heat-exchanger with unique design Humidified heat-exchanger for calibration cross-interference compensation Digital controlled high stable outlet dew point ± 0,1°C Compact design Digitally controlled Affordable & compact, the peltier cooler cools below ambient temperature and eliminates the risk of condensation

Transcript of Ankersmid peliter cooler datasheet 0

Peltier Cooler Gas Sample Cooler

Through lowering the dew point of humid gas, the Ankersmid Peltier Cooler is able to reduce the risk of condensation occurring in the analyser

The Peltier Cooler is equipped with an adjustable dew point stabiliser.

The stabiliser is set at 4oC as standard with the option to adjust the

value to between 1 and 15oC at any time.

Digitally Controlled The digital controlled cooler has many control and warning features

like programmable alarms, mA-output, digital inputs and Modbus or

RS485 communication. Using smart technology the alarm status of the

cooler changes when the temperature deviates ± 3 °C from a set point

Humidifies Calibration Gases The peltier cooler is equipped with a silicon coated (glass) heat exchanger,

as standard with the added availability of PFA coating as an alternative.

Using a special liquid inlet, the Peltier Cooler humidifies calibration gases

to avoid volumetric errors.

KEYBENEFITS• Removable heat-exchanger with

unique design

• Humidified heat-exchanger for

calibration cross-interference

compensation

• Digital controlled high stable outlet

dew point ± 0,1°C

• Compact design

• Digitally controlled

Affordable & compact, the peltier cooler cools below ambient temperature and eliminates

the risk of condensation

“ “

TECHNICALSPECIFICATION

GENERAL • Lowers the dew point of humidity

• Ambient temperature up to +50°C

• Interchangeable heat exchangers

OPERATION DATA • Gas inlet dew point: Max. 65°C

• Gas inlet temperature: Max. 190°C

• Gas outlet temperature: +1°C to +15°C,

factory setting: +4°C

• Ambient temperature: +5°C to 45°C

• Total cooling capacity: Max. 245kJ/h

• Stability: 0.1°C at ambient temperature 20°C

POWER

• Mains connection: Electrical terminals 2.5mm² / Cable

gland 2 x PG13

• Alarm contact: Free programmable contact 1NO / 1NC,

rating: 250V, 16A AC

• Alarm set points: < +2°C / > +10°C

• Protection class: IP20 EN 60529 / EN 61010

• Electrical protection: Fuse 1,6A

• Power consumption: 2 Peltier elements á 34W

PHYSICAL• Dimensions: 200 x 280 x 190mm (H x L x D)

• Weight: Approx 4.5kg

• Housing: Stainless steel/Aluminium anodised,

Wall-mounting

11 Ark Royal Way, Lairdside Technology Park, Tranmere, Wirral, CH41 9HTT: +44(0)151 666 8300 F: +44(0)151 666 8329 E: [email protected] W: www.a1-cbiss.com Buy Online: shop.a1-cbiss.com

Cert No. 996ISO 9001

MODELOPTIONS

Model Number 14xx 15xx 16xx

Heat exchangers 1 2 1

Gas flow rate 1x 150Nl/h max 2x 150Nl/h max 1x 350Nl/h max

Sealing Fluoroelastomer Fluoroelastomer Fluoroelastomer

Max. pressure 10 bar a 10 bar a 10 bar a

Pressure drop 2mbar at 150Nl/h 2mbar at 150Nl/h 5mbar at 350Nl/h

Dead volume 35cm³ 35cm³ 100 cm³

Sample gas inlet 1x 1/4”f NPT 1x 1/4”f NPT 1x 1/4”f NPT

Sample gas outlet 1x 1/4”f NPT 1x 1/4”f NPT 1x 1/4”f NPT

Condesate outlet 1x 3/8”f NPT 1x 3/8”f NPT 1x 3/8”f NPT