ANAESTHESIA MACHINE AND EQUIPMENT
Q & A SESSION
Dr.R.SelvakumarProfessor Of AnaesthesiologyK.A.P.Viswanatham Govt medical college,TrichirapalliTamilnadu - India
PART -I
![Page 2: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/2.jpg)
1.NAME ANY TWO GASES THAT ARE MARKETED IN LIQUID FORM IN THE CYLINDERS TO THE O.T?
1.NITROUS OXIDE2.CO23.CYCLOPROPANE
Q-1
![Page 3: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/3.jpg)
![Page 4: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/4.jpg)
Q-2
2. MENTION ANY 4 TESTS PERFORMED ON THE FINISHED CYLINDERS:
1. TENSILE TEST2. FLATTENING TEST3. IMPACT TEST4. BEND TEST5. HYDRAULIC TEST
![Page 5: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/5.jpg)
Q-3
3. WHAT IS THE COLOUR CODING OF MEDICAL AIR CYLINDER?
Ans. BLACK BODY , BLACK AND WHITE SHOULDER
![Page 6: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/6.jpg)
![Page 7: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/7.jpg)
Q-4
4. WHAT IS THE PIN INDEX OF OXYGEN CYLINDER?
Ans: 2,5
![Page 8: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/8.jpg)
Q-5
5. WHAT IS THE SERVICE PRESSURE OF SMALL AND BULK OXYGEN CYLINDERS?
Ans: SAME … 1987 PSI / 137 bar
![Page 9: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/9.jpg)
Q-6
6.WHICH GAS HAS BEEN ALLOTTED TO THE PIN INDEXOF 3,4?
Ans: NONE
![Page 10: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/10.jpg)
![Page 11: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/11.jpg)
Q-7
7. WHAT IS THE SERVICE PRESSURE OF N2O CYLINDER WHEN IT IS HALF FULL?
Ans: 750 PSI / 51 bar
![Page 12: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/12.jpg)
Q-8
8. WHAT ARE THE PARTS OF A CYLINDER?
• BODY• VALVE• PORT• STEM• NECK• SHOULDER• PRESSURE RELIEF
DEVICE
![Page 13: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/13.jpg)
Q-9
9. WHAT IS THE MATERIAL USED FOR THE MANUFACTURE OF CYLINDERS?
Ans: MOLYBDENUM STEEL
![Page 14: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/14.jpg)
Q-10
10. WHICH MATERIAL IS USED TO CONSTRUCT THE CYLINDER OUTLET VALVES?
Ans: BRONZE / BRASS WITH NICKEL / CHROMIUM COATING
![Page 15: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/15.jpg)
Q-11
11.WHAT ARE THE TYPES OF CYLINDER VALVES?
1. FLUSH TYPE2. BULL NOSE
![Page 16: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/16.jpg)
FLUSH TYPE VALVEBULL NOSE TYPE
![Page 17: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/17.jpg)
Q-12
12.WHAT ARE THE TYPES OF CYLINDER VALVE AS PER THE INTERIOR DESIGN?
1. PACKED TYPE2. DIAPHRAGM TYPE
![Page 18: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/18.jpg)
PACKED TYPE DIAPHRAGM TYPE
![Page 19: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/19.jpg)
Q-13
13. WHERE IS WOODS METAL USED?
IT IS A FUSIBLE PLUG USED IN THE CYLINDERVALVE WHICH MELTS AT HIGH TEMPERATURETHEREBY PREVENTING EXPLOSION.
![Page 20: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/20.jpg)
FUSIBLE PLUG
![Page 21: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/21.jpg)
Q-14
14. WHAT IS BODOK SEAL ?
IT IS A NEOPRENE WASHER WITH ALUMINIUMEDGES WHICH IS USED TO PREVENT THE LEAKBETWEEN THE CYLINDER AND YOKE ASSEMBLY
![Page 22: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/22.jpg)
Q-15
15. WHAT IS THE COLOUR CODING OF ENTONOX CYLINDER?
BLUE BODYWHITE AND BLUE QUARTER ON THE SHOULDER
![Page 23: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/23.jpg)
ENTONOX CYLINDER
![Page 24: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/24.jpg)
Q-16
16. WHAT DO YOU MEAN BY CRACKING THE CYLINDER ?
THE CYLINDER VALVE IS OPENED TO ATMOSPHEREBEFORE IT IS BEING FITTED TO THE YOKE TO CLEAR OFF THE DUST.
![Page 25: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/25.jpg)
Q-17
17. WHAT DO YOU MEAN BY FILLING RATIO?
RATIO OF THE WEIGHT OF THE LIQUEFIABLE GAS INTRODUCED IN THE CYLINDER TO THE WEIGHT OF THE WATER THE CYLINDER WILL HOLD.
![Page 26: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/26.jpg)
Q-18
18. WHAT IS THE FILLING RATIO OF OXYGEN?
FILLING RATIO IS ONLY FOR THE LIQUEFIABLE GASES
![Page 27: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/27.jpg)
Q-19
19. WHAT IS THE DRIVING PRESSURE OF ANAESTHESIA VENTILATOR?
55 -60 PSI
![Page 28: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/28.jpg)
Q-20
20. AT WHICH TEMPERATURE, LIQUID OXYGEN SHOULD BE STORED?
BELOW - 184 c
![Page 29: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/29.jpg)
Q.21
21.WHAT DO YOU MEAN BY “ZEOLITE”?
HYDRATED ALUMINIUM SILICATE OF THE ALKALINE EARTH METALS. IT SELECTIVELYABSORBS NITROGEN,LEAVING O2 . ITS BEING USED IN OXYGEN CONCENTRATOR.
![Page 30: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/30.jpg)
Q.22
22. WHAT IS THE SPECIFIC NAME OF THE QUICK COUPLERS?
SHRADDER VALVE AND PROBE
![Page 31: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/31.jpg)
Q.23
23. WHAT DO YOU MEAN BY DISS ?
DIAMETER – INDEXED SAFETY SYSTEM : NON-INTERCHANGABLE,THREADED CONNECTION FOR MEDICALGAS PIPELINES.
![Page 32: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/32.jpg)
Q.24
24. WHAT DO YOU MEAN BY LINK – 25 SYSTEM ?
OXYGEN PROPORTION LIMITING CONTROL DEVICE.IT ENSURES A MINIMUM % OF O2 TO BE DELIVERED AT 25% .
![Page 33: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/33.jpg)
![Page 34: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/34.jpg)
Q.25
25. WHAT IS THE DISADVANTAGE OF LINK – 25 SYSTEM ?
• MECHANICAL DEVICE – SO WEAR & TEAR AFFECTS THE GEAR SYSTEM.• IF A 3RD GAS LIKE AIR IS GIVEN, THIS SYSTEM DOESN’T HAVE A CONTROL.
![Page 35: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/35.jpg)
Q.26
26.WHAT DO YOU MEAN BY SLAVE REGULATOR ?
THE OUTLET PRESSURE OF O2 PRESSURE REGULATOR CONTROLS THE OUTPUT FROM THE N2O REGULATOR. SINCE THE N2O REGULATORACTS AS A SLAVE TO THE PRIME O2 REGULATOR,IT IS CALLED THE SLAVE REGULATOR.
![Page 36: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/36.jpg)
Q.27
27. WHAT IS THE OUTLET PRESSURE FROM THE PRESSURE REGULATOR ?
50 PSI
![Page 37: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/37.jpg)
Q.28
28. WHAT IS THE PIPE LINE PRESSURE ? WHY IS IT KEPT SLIGHTLY HIGHER THAN THE CYLINDER OUTLET PRESSURE ?
60 PSI - TO MAKE IT TO BE USED PREFERENTIALLY OVER CYLINDERS.
![Page 38: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/38.jpg)
Q.29
29. WHAT DO YOU MEAN BY PUMPING EFFECT ?
THE EXCESSIVE BACK PRESSURE CONDUCTED FROM SQUEEZING THE BAG CAUSES THE VAPORISEROUTPUT TO BECOME MORE THAN THE SET DIAL CONCENTRATION.
![Page 39: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/39.jpg)
Q.3030. IF THE OXYGEN FLOW METER SHOWSA 200 ml FLOW EVEN AFTER CLOSUREOF THE FLOW CONTROL VALVE.. WHATCOULD BE THE REASON?
MINIMUM MANDATORY FLOW
![Page 40: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/40.jpg)
Q.31
31. WHAT IS THE OPENING PRESSURE OF SAFETY RELIEF VALVE SITUATED JUST PRIOR TO COMMON GAS OUTLET?
200 CM OF H2O
![Page 41: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/41.jpg)
Q.32
32. WHAT ARE THE TWO TYPES OF SCAVANGING SYSTEM ?
• ACTIVE• PASSIVE
![Page 42: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/42.jpg)
Q.33
33. WHAT IS THE PHYSICAL PRINCIPLE OF PRESSURE REGULATOR?
A LARGE FORCE ACTING ON A SMALL AREA ISBALANCED BY SMALL FORCE ACTING ON A LARGE AREA.
![Page 43: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/43.jpg)
Q.34
34. IN FLOW METERS, THE FLOW OF THE GAS AT LOW FLOW DEPENDS UPON WHICH FACTOR ?
VISCOSITY
![Page 44: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/44.jpg)
Q.35
35. WHICH LAW GOVERNS THE FLOW RATE THROUGH THE FLOW METER USING THE LAMINAR FLOW ?
HAGEN – POISEUITTE LAW
![Page 45: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/45.jpg)
Q.36
36. GIVE FOUR EXAMPLES OF VARIABLE ORIFICE FLOW METERS
• ROTAMETER• COXETER BOBBIN METER• HEIDBRINK METER• CONNEL METER
![Page 46: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/46.jpg)
Q.37
37. GIVE AN EXAMPLE OF FIXED ORIFICE FLOW METER
BOURDON PRESSURE GAUGE METER
![Page 47: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/47.jpg)
Q.38
38. WHAT IS THE SPECIALITY OF O2 FLOW CONTROL KNOB IN THE FLOW METER ASSEMBLY ?
OXYGEN IS LARGE, FLUTED, PROJECTILE KNOB, WHITE IN COLOR
![Page 48: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/48.jpg)
Q.39
39. WHAT IS THE NAME OF THE FLOW METER IN USE IN OUR BOYLES MACHINE ?
ROTAMETER
![Page 49: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/49.jpg)
Q.40
40. WHAT IS THE NAME OF THE TUBE BEING USED IN ROTAMETER ? WHAT IS ITS SPECIALITY ?
THORPES TUBE – IT IS CONICAL IN SHAPE, TAPERED VERTICALLY WITH THE SMALLEST DIAMETER AT THE BOTTOM.
![Page 50: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/50.jpg)
Q.41
41. WHICH FACTOR DECIDES THE FLOW IN THE ROTAMETER WITH BALL TYPE FLOAT ?
DENSITY
![Page 51: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/51.jpg)
Q.42
42. WHAT ARE THE SAFETY FEATURES IN THE FLOW METER BANK ?
• O2 IN THE DOWN STREAM• COLOR AND TOUCH CODED FLOW CONTROL KNOBS• TRANSILLUMINANT BACK SCREEN• ROTATING BOBBIN• STOPS AT THE END OF THORPES TUBE• ORM C• LINK 25• MINIMUM MANDATORY OXYGEN FLOW
![Page 52: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/52.jpg)
Q.43
43. WHAT IS THE S.V.P OF DESFLURANE AT 20◦ C ?
664 MM OF HG
![Page 53: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/53.jpg)
Q.44
44. WHICH METAL IS USED FOR THE CONSTRUCTION OF VAPORISER?
COPPER
![Page 54: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/54.jpg)
Q.45
45. WHY COPPER IS SELECTED ?
• REASONABLE SPECIFIC HEAT CAPACITY• HIGH THERMAL CONDUCTIVITY• MALLEABLE
![Page 55: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/55.jpg)
Q.46
46. WHAT DO YOU MEAN BY PLENUM VAPORISER ?
VAPORISER WITH BAFFLES WHICH REQUIRESPRESSURISED FRESH GAS FLOW .HIGH RESISTANCE VAPORISERS.
![Page 56: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/56.jpg)
Q.47
47. GIVE AN EXAMPLE FOR A MEASURED FLOW VAPORISER
COPPER KETTLE, DRAGER DIVA
![Page 57: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/57.jpg)
Q.48
48. GIVE AN EXAMPLE OF BUBBLE THROUGH VAPORISER
COPPER KETTLE AND VERNITROL
![Page 58: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/58.jpg)
Q.49
49. WHAT IS THE MAIN DRAWBACK OF VIC VAPORISER?
THE DELIVERED CONCENTRATION OF THE VAPORISER MAY BE MORE THAN THE SET CONCENTRATION.
![Page 59: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/59.jpg)
Q.50
50. IN WHICH FLOW RATE, PUMPING EFFECT IS PREDOMINANT ?
LOW FRESH GAS FLOWS
![Page 60: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/60.jpg)
Q.51
51. HOW WILL YOU MINIMISE THE PUMPING EFFECT ?
1. INCREASING THE SIZE OF BYEPASS CHAMBER2. INCREASING THE LENGTH OF INLET TUBE INTO THE VAPORISER3. AVOIDING WICKS NEAR INLET TUBE
![Page 61: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/61.jpg)
Q.52
52. WHAT IS THE IMPACT OF DECREASING ATMOSPHERIC PRESSURE ON THE TEC SERIES VAPORISERS ?
THE VAPORISER OUTPUT WILL INCREASE AS THEOUTPUT IS CALCULATED IN VOLUME % BASED ON THE ATMOSPHERIC PRESSURE. BUT THE PARTIAL PRESSUREREMAINS THE SAME.
![Page 62: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/62.jpg)
Q.53
53. WHAT DO YOU MEAN BY ANA CONDA ?
• ANAESTHETIC CONSERVING DEVICE• A HEAT AND MOISTURE EXCHANGER LIKE DEVICE WHERE LIQUID ANAESTHETIC IS DELIVERED BY A SYRINGE PUMP.
![Page 63: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/63.jpg)
Q.54
54. WHAT IS THE COMMON NAME FOR TRIMETHYLENE ?
CYCLOPROPANE
![Page 64: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/64.jpg)
Q.55
55. WHAT IS THE COLOR CODING OF CYCLOPROPANE CYLINDER ?
ORANGE
![Page 65: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/65.jpg)
Q.56
56. WHAT IS A “BELLAMY GARDENER BOTTLE” ?
ETHER DROPPER USED FOR SCHIMMELBUSCH MASH
![Page 66: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/66.jpg)
Q.57
57. EXPAND THE E.M.O VAPORISER ?
EPSTEIN – MCINTOSH – OXFORD VAPORISER
![Page 67: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/67.jpg)
Q.58
58. WHAT IS THE MATERIAL USED IN EMO VAPORISER FOR THERMO BUFFERING ?
WATER BATH
![Page 68: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/68.jpg)
Q.59
59. WHAT IS THE PURPOSE OF THE PLUNGER IN BOYLES ETHER VAPORISER ?
TO INCREASE THE OUTPUT CONCENTRATION, THE PLUNGER HAS TO BE DEPRESSED
![Page 69: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/69.jpg)
Q.60
60. FGF ENTERS NEAR THE PATIENT END IN MAPLESON C.
TRUE OR FALSE ?
TRUE
![Page 70: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/70.jpg)
Q.6161. FGF RUNS THROUGH THE OUTER TUBE AND EXPIRATION FLOWS THROUGH THE INNER TUBE.
TRUE OR FALSE ?
TRUE
![Page 71: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/71.jpg)
Q.62
62. WHAT IS FLOW RATE IN LOW FLOW ANAESTHESIA?
FGF 500 – 1000 ml/min ( up to 2 l/min )
![Page 72: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/72.jpg)
Q.63
63. WHAT IS THE CONCERN OF USING SEROFLURENE IN LOW FLOW ANAESTHESIA ?
ACCUMULATION OF COMPOUND-A & METHANOL IN THE CLOSED CIRCUIT.
![Page 73: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/73.jpg)
Q.64
64. WHICH FACE MASK IS RECOMMENDED FOR PAEDIATRIC PATIENTS ?
RENDELL-BAKER-SOUCEK MASKBECAUSE OF ITS LOW DEAD SPACE
![Page 74: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/74.jpg)
Q.65
65. WHAT ARE THE PARTS OF A LARYNGOSCOPE ?
HANDLE, BATTERY, HINGE, WEB, BASE, FLANGE, HEEL, SPATULA,BULB, TIP OR BEAK
![Page 75: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/75.jpg)
Q.66
66. NAME ANY 2 STRAIGHT BLADED LARYNGOSCOPE
• MILLER• WISCONSIN• FLAGG
![Page 76: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/76.jpg)
Q.6767. WHAT IS THE PURPOSE OF MURPHY’S EYE IN ENDOTRACHEAL TUBE ?
IT ACTS AS AN ALTERNATE CHANNEL ,IF BEVEL IS BLOCKED.
![Page 77: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/77.jpg)
Q.68
68. WHAT DO YOU MEAN BY “CE” MARKING ON THE ETT ?
CONFORMITE EUROPEENE
![Page 78: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/78.jpg)
Q.69
69. WHAT DO YOU MEAN BY Z79 – IT ?
IMPLANTATION TESTED AS SUGGESTED BY Z79 TOXICITY SUBCOMMITTEE OF THE AMERICAN NATIONAL STANDARDS INSTITUE SET UP IN 1968.
![Page 79: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/79.jpg)
Q.70
70. HOW IS THE FRENCH SIZE OF AN ETT IS CALCULATED?
EXTERNAL DIAMETER IN m.m X 3 WILL GIVE YOU THE FRENCH SIZING.
![Page 80: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/80.jpg)
Q.71
71. WHAT IS THE FORMULA USED TO CALCULATE THE SIZE OF ETT IN CHILDREN ?
AGE BELOW 6yrs – AGE IN ( YEARS/3 ) + 3.5ABOVE 6yrs – AGE IN ( YEARS/4 ) + 4.5
![Page 81: Anesthesia machine and equipment Q & A Part -I](https://reader035.fdocuments.in/reader035/viewer/2022062412/5882178a1a28ab3f4c8b5c0f/html5/thumbnails/81.jpg)
THANK YOU
dr.r.selvakumar
Continued in part –II presentation