Alcolyzer Plus Beer

download Alcolyzer Plus Beer

of 129

Transcript of Alcolyzer Plus Beer

  • 7/22/2019 Alcolyzer Plus Beer

    1/129

    Reference Manual

    Alcolyzer Plus Beer

    Beer Analyzing SystemFirmware Version: V5.02.09

  • 7/22/2019 Alcolyzer Plus Beer

    2/129

  • 7/22/2019 Alcolyzer Plus Beer

    3/129

    Reference Manual

    Alcolyzer Plus Beer

    Beer Analyzing SystemFirmware Version: V5.02.09

  • 7/22/2019 Alcolyzer Plus Beer

    4/129

    Anton Paar GmbH assumes no liability for technical or printing errors or omissions in this document.

    Nor is any liability assumed for damages resulting from information contained in the document.

    Anton Paar GmbH reserves the right to content changes. This also extends to changes to delivery

    volumes or any features of delivered parts.

    All rights reserved (including translation). This document, or any part of it, may not be reproduced,

    changed, copied, or distributed by means of electronic systems in any form (print, photocopy, microfilm

    or any other process) without prior written permission by Anton Paar GmbH.

    Trademarks, registered trademarks, trade names, etc. may be used in this manual without being marked

    as such. They are the property of their respective owner.

    Published by Anton Paar

    Printed: Anton Paar, Austria

    Copyright 2009 Anton Paar GmbH, Graz, Austria

    Address: Anton Paar GmbH

    Anton-Paar-Str. 20

    A-8054 Graz / Austria Europe

    Tel: +43 316 257-0

    Fax: +43 316 257-257

    E-Mail: [email protected]

    Web: www.anton-paar.com

    Date: 06.03.2009

    Document number C27IB02M.fm

  • 7/22/2019 Alcolyzer Plus Beer

    5/129

    C27IB02 5

    Contents

    1 About the Reference Manual .......................................................................................................... 9

    2 Safety Instructions........................................................................................................................... 9

    2.1 General Safety Instructions ..................................................................................................... 9

    3 About Alcolyzer Plus Beer............................................................................................................ 11

    4 Testing the Supplied Parts............................................................................................................ 11

    5 Putting into Operation................................................................................................................... 14

    5.1 Installing the Instruments....................................................................................................... 14

    5.2 Mounting the pH Electrode .................................................................................................... 16

    5.2.1 Preparing and Mounting the pH Electrode "Hamilton Polilyte Lab"........................ 16

    5.2.2 Preparing and Mounting the pH Electrode "Orion model 91-63"............................ 175.3 Mounting the Hoses without Option pH/Color........................................................................ 19

    5.4 Mounting the Hoses with Option pH/Color............................................................................. 20

    5.5 Completing the Installation .................................................................................................... 21

    5.6 Connecting a Printer or PC.................................................................................................... 24

    5.6.1 Connecting Alcolyzer Plus Beer to a Printer .......................................................... 24

    5.6.2 Connecting Alcolyzer Plus Beer to a PC................................................................ 24

    6 Functional Components................................................................................................................ 25

    6.1 Display and Keypad............................................................................................................... 25

    6.1.1 Display in the Measuring Mode.............................................................................. 25

    6.1.2 Soft Keys in the Measuring Mode .......................................................................... 27

    6.2 Rear View .............................................................................................................................. 28

    7 General Instrument Settings......................................................................................................... 29

    7.1 Setting the Display Contrast .................................................................................................. 29

    7.2 Setting Date and Time ........................................................................................................... 29

    7.3 Configuring the Measuring System........................................................................................ 29

    7.3.1 Setting the General Parameters............................................................................. 29

    7.3.2 Defining the Peripheral Devices ............................................................................. 30

    7.3.3 Defining the Interface Settings ............................................................................... 30

    7.4 Setting a Custom Function .................................................................................................... 30

    7.4.1 Setting a User-Defined Ethanol Table (in %V/V).................................................... 30

    7.4.2 Setting a User-Defined Color Unit .......................................................................... 31

    7.5 Selecting the Output Data...................................................................................................... 32

    7.5.1 Selecting the Output Data for the Display .............................................................. 33

    7.5.2 Selecting the Output Data for the Printer ............................................................... 33

    7.5.3 Selecting the Output Data for the Data Memory..................................................... 33

    7.6 Settings for the Sample Changer........................................................................................... 33

    8 Setting the Measuring Methods.................................................................................................... 34

    8.1 Description of the Measuring Methods .................................................................................. 348.2 Editing a Measuring Method.................................................................................................. 35

  • 7/22/2019 Alcolyzer Plus Beer

    6/129

    6 C27IB02

    8.2.1 Editing a Predefined Method.................................................................................. 35

    8.2.2 Editing a New Method ............................................................................................ 35

    8.3 Selecting the Measuring Method........................................................................................... 37

    9 SOP (Standard Operating Procedure).......................................................................................... 38

    10 Checking the Adjustment Validity................................................................................................ 41

    10.1 Checking Procedure without Option pH/Color....................................................................... 41

    10.1.1 Performing a Checking Procedure with Water (0%V/V)......................................... 41

    10.1.2 Performing a Checking Procedure with a Water/Ethanol Solution of Known

    Concentration ........................................................................................................................ 43

    10.2 Checking Procedure with Option pH/Color............................................................................ 44

    10.2.1 Checking Procedure for the Measuring Parameter "alcohol" ................................. 44

    10.2.2 Checking Procedure for the Measuring Parameter "pH"........................................ 44

    10.2.3 Checking Procedure for the Measuring Parameter "color"..................................... 45

    11 Adjusting the Instrument .............................................................................................................. 4511.1 Adjustment without Option pH/Color...................................................................................... 45

    11.1.1 Adjustment with Water (Zero Point)........................................................................ 46

    11.1.2 Adjustment with a Water/Ethanol Solution of Known Alcohol Concentration ......... 48

    11.2 Adjustment with Option pH/Color........................................................................................... 49

    11.2.1 Adjustment Alcohol................................................................................................. 49

    11.2.2 Adjustment Color.................................................................................................... 49

    11.2.2.1 Adjustment with Water (Zero) ................................................................. 49

    11.2.2.2 Adjustment with a Color Standard (Tartrazine) of Known Concentration 51

    11.2.3 Adjustment pH........................................................................................................ 5211.2.3.1 Automatic ................................................................................................ 53

    11.2.3.2 Manual .................................................................................................... 54

    11.2.3.3 Edit Buffer Data....................................................................................... 55

    11.2.3.4 Sort Buffer Data ...................................................................................... 56

    11.3 Density Adjustment................................................................................................................ 56

    12 Measurement.................................................................................................................................. 58

    12.1 Identifying the Sample and Assigning the Method................................................................. 59

    12.1.1 Types of Sample Identification ............................................................................... 59

    12.1.2 Sample Identification for "text + number" and "text"............................................... 60

    12.1.3 Sample Identification for "list" and "# + list"............................................................ 60

    12.1.3.1 Entering a Sample List ............................................................................ 60

    12.1.3.2 Editing the Sample List during Operation................................................ 62

    12.1.3.3 Deleting a Sample List ............................................................................ 62

    12.2 Performing a Measurement without Option pH/Color ............................................................ 63

    12.3 Performing a Measurement with Option pH/Color................................................................. 64

    12.4 Correcting Calculated Analysis Results................................................................................. 64

    13 Data Memory .................................................................................................................................. 66

    14 Cleaning Routines ......................................................................................................................... 67

    14.1 Cleaning the Measuring Cells without Option pH/Color......................................................... 67

  • 7/22/2019 Alcolyzer Plus Beer

    7/129

    C27IB02 7

    14.1.1 Between Series of Measurements on the Same Day............................................. 67

    14.1.2 At the End of the Day ............................................................................................. 67

    14.1.3 Once a Week.......................................................................................................... 68

    14.2 Cleaning the Measuring Cells with Option pH/Color.............................................................. 69

    14.2.1 Storing the pH Electrode ........................................................................................ 69

    14.2.2 Between Series of Measurements on the Same Day............................................. 70

    14.2.3 At the End of the Day ............................................................................................. 70

    14.2.4 Once a Week.......................................................................................................... 71

    14.3 Cleaning after Wort Measurements....................................................................................... 72

    14.4 Maintaining the pH Electrode................................................................................................. 72

    14.4.1 Maintaining the Hamilton pH Electrode.................................................................. 72

    14.4.2 Maintaining the Orion pH Electrode ....................................................................... 73

    15 Menu Operation and Description ................................................................................................. 74

    15.1 Menu Operation..................................................................................................................... 7415.1.1 Operating the Soft Keys on the Instrument ............................................................ 74

    15.1.2 Using an External PC Keyboard............................................................................. 75

    15.1.3 Using a Bar Code Reader ...................................................................................... 75

    15.2 Menu Description and Function............................................................................................. 76

    15.2.1 "adjustment" ........................................................................................................... 76

    15.2.1.1 "adjust alcohol"........................................................................................ 76

    15.2.1.2 "adjust pH" .............................................................................................. 77

    15.2.1.3 "adjust color" ........................................................................................... 77

    15.2.2 "instrument settings"............................................................................................... 7815.2.3 "method settings" ................................................................................................... 82

    15.2.4 "user functions" ...................................................................................................... 83

    15.2.5 "data memory"........................................................................................................ 83

    15.2.6 "utilities".................................................................................................................. 84

    15.2.7 "parameter calculator" ............................................................................................ 90

    15.2.8 "service mode" ....................................................................................................... 91

    15.2.9 "auto sampler settings"........................................................................................... 91

    15.3 Entering and Changing a Password ...................................................................................... 91

    15.4 Installing a New Firmware Version ........................................................................................ 92

    16 Calculations.................................................................................................................................... 93

    17 Filling Samples with the SF-1 Sample Filling Unit...................................................................... 97

    Appendix A: Degassing Beer Samples.......................................................................................... 98

    Appendix B: Available Parameters ................................................................................................ 99

    B.1 Method Parameters............................................................................................. 99

    B.2 Printer Protocol Parameters.............................................................................. 101

    Appendix C: Status and Error Messages .................................................................................... 102

    C.1 Status/Error Messages for the Alcohol Measurement....................................... 102

    C.2 Status/Error Messages for the Color Measurement .......................................... 103

    C.3 Status/Error Messages for the pH Measurement .............................................. 103

  • 7/22/2019 Alcolyzer Plus Beer

    8/129

    8 C27IB02

    C.4 System Error Messages.................................................................................... 104

    C.5 Other Messages................................................................................................ 104

    Appendix D: Technical Data ......................................................................................................... 105

    Appendix E: Interface Commands for Communication PC and Alcolyzer Plus Beer ............. 106

    Appendix F: Configuration for Communication between DMA and Alcolyzer Plus Beer ...... 108

    Appendix G: Firmware Versions .................................................................................................. 110

    Appendix H: Wetted Parts............................................................................................................. 112

    Appendix I: Material Safety Data Sheet for Tartrazine.............................................................. 113

    Appendix J: Density/Concentration Table for Binary Aqueous Ethanol Solutions ................ 115

    Appendix K: Menu Tree................................................................................................................. 126

    Index................................................................................................................................................... 128

  • 7/22/2019 Alcolyzer Plus Beer

    9/129

    1 About the Reference Manual

    C27IB02 9

    1 About the Reference Manual

    This reference manual informs you about the installation and the safe handling

    and use of the instrument. Make sure that the reference manual is easily

    accessible to all personnel involved with the instrument. Pay special attention to

    the safety instructions and warnings in the manual and on the instrument.

    Symbols in this reference manual

    This reference manual features the following symbols:

    2 Safety Instructions

    Read the reference manual at hand before using Alcolyzer Plus Beer.

    Follow all hints and instructions contained in this reference manual to ensure

    the correct use and safe functioning of Alcolyzer Plus Beer.

    2.1 General Safety Instructions

    Liability

    The reference manual at hand does not claim to address all safety issues

    associated with the use of the instrument and samples. It is your

    responsibility to establish health and safety practices and determine the

    applicability of regulatory limitations.

    Anton Paar GmbH only warrants the proper functioning of Alcolyzer Plus

    Beer if no adjustments have been made to the mechanics, electronics, and

    firmware.

    Only use Alcolyzer Plus Beer for the purpose described in this reference

    manual. Anton Paar GmbH is not liable for damages caused by incorrect use

    of Alcolyzer Plus Beer.

    Caution:

    This symbol calls attention to the risk of instrument damage or

    measurement errors. Do not proceed until the indicated conditions for

    averting this threat are fully understood and met.

    Information:

    This symbol calls attention to any additional information of use to the

    operator.

  • 7/22/2019 Alcolyzer Plus Beer

    10/129

    2 Safety Instructions

    10 C27IB02

    Installation and use

    Alcolyzer Plus Beer is not an explosion-proof instrument and therefore must

    not be operated in areas with risk of explosion.

    The installation procedure shall only be carried out by authorized personnel

    who are familiar with the installation instructions.

    Do not use any accessories or wearing parts other than those supplied or

    approved by Anton Paar GmbH.

    Make sure all operators are trained to use the instrument safely and correctly

    before starting any applicable operations.

    In case of damage or malfunction, do not continue operating Alcolyzer Plus

    Beer. Do not operate the instrument under conditions which could result in

    damage to goods and/or injuries and loss of life.

    Check Alcolyzer Plus Beer for chemical resistance to the samples and

    cleaning agents.

    Maintenance and service

    The results delivered by Alcolyzer Plus Beer not only depend on the correct

    function of the instrument, but also on various other factors. We therefore

    recommend to have the results checked (e.g. plausibility tested) by skilled

    personnel before consequential actions are taken based on the results.

    Service and repair procedures may only be carried out by authorizedpersonnel or by Anton Paar GmbH.

    Disposal

    Concerning the disposal of Alcolyzer Plus Beer, observe the legal

    requirements in your country.

    Precautions for highly flammable samples and cleaning agents

    Observe and adhere to your national safety regulations for handling the

    measured samples (e.g. use of safety goggles, gloves, respiratory protection

    etc.).

    Only store the minimum required amount of sample, cleaning agents and

    other inflammable materials near the Alcolyzer Plus Beer.

    Do not spill sample/cleaning agents or leave their containers uncovered.

    Immediately remove spilled sample/cleaning agents.

    Make sure that the setup location is sufficiently ventilated. The environment

    of Alcolyzer Plus Beer must be kept free of inflammable gases and vapors.

  • 7/22/2019 Alcolyzer Plus Beer

    11/129

    3 About Alcolyzer Plus Beer

    C27IB02 11

    Connect Alcolyzer Plus Beer to the mains via a safety switch located at a

    safe distance from the instrument. In an emergency, turn off the power using

    this switch instead of the power switch on Alcolyzer Plus Beer.

    Supply a fire extinction unit.

    Ensure the sufficient supervision of Alcolyzer Plus Beer during operation.

    3 About Alcolyzer Plus Beer

    Alcolyzer Plus Beer is based on a highly accurate and patented NIR

    spectrometer, which measures the alcohol content of beer in the concentration

    range of 0 - 12%V/V.

    Optionally, Alcolyzer Plus Beer can also be equipped with a color measurement

    device and a pH measurement device. The color value (EBC or ASBC value) is

    determined by the measurement of light absorption. For the pH measurement, an

    electrode Hamilton (Polilyte Lab) or a measuring cell Orion (model 91-63) is used.

    In combination with an Anton Paar density meter DMA 4500/5000 and a sample

    changer SP-1m, the automatic determination of relevant beer parameters can be

    achieved.

    4 Testing the Supplied PartsAlcolyzer Plus Beer was tested and packed carefully before shipment. However,

    damage may occur during transport.

    1. Keep the packaging material (box, foam piece, transport protection) for

    possible returns and further questions from the transport and insurance

    company.

    2. Check the delivery for completion by comparing the supplied parts to those

    noted in Table 4.1.

    3. If a part is missing, contact your Anton Paar representative.

    4. If a part is damaged, contact the transport company and your Anton Paar

    representative.

    http://-/?-http://-/?-
  • 7/22/2019 Alcolyzer Plus Beer

    12/129

    4 Testing the Supplied Parts

    12 C27IB02

    Table 4.1: Supplied Parts

    Symbol Pcs. Item/Spare Part Mat. No.

    1 Alcolyzer Plus Beer (without option pH/color)

    or

    Alcolyzer Plus Beer

    with option color + pH Hamilton or

    with option color + pH Orion

    8921

    8921

    9918

    12167

    1 Instruction manualEnglish or

    German

    including a CD with the reference manual

    12665

    12666

    1 Power cable

    USA:

    UK:

    Europe:

    52656

    61865

    65146

    1 Accessory kit 9860

    2m Hose 3 x 5 silicone / 1m

    Hose 1.6 x 3.2 Tygon / 1m

    50814

    67540

    5 Syringe 5 mL / 1pc. 63130

    2 Filling nozzle Luer Alcolyzer Plus / 1pc. 7046

    2 Adapter Luer " UNF / 1pc. 64792

    2 Male Luer plug / 1pc. 63865

  • 7/22/2019 Alcolyzer Plus Beer

    13/129

    4 Testing the Supplied Parts

    C27IB02 13

    3 Adapter Luer cone / 1pc. 63863

    2 Filling nozzle fitting Alcolyzer Plus / 1pc. 10759

    2 Filling nozzle UNF " Alcolyzer Plus / 1pc. 10758

    2 Tube d=2.1, D=2.4, L=20 1.4571 / 1pc. 67982

    1 Needle (only delivered with SF-1) / 1pc. 67394

    3m Hose 2 x 4 silicone /1m

    (only delivered with SF-1)

    51273

    1 Cable male 9 Pin / male 9 Pin / 1pc. 2359

    1

    (option)

    Backup Card DMA 4500/5000 Alcolyzer for

    firmware versions DMA V4.xxx or higher;

    Alcolyzer V5.xx / 1pc.

    23913

    1 Waste vessel 500 mL 6210

    1 Option color + pH Hamilton 9918

    1 Electrode Hamilton pH = 0-14 0-80 9466

    1 Option color + pH Orion 12167

    1 Electrode Orion model 91-63 12555

    Table 4.1: Supplied Parts

    Symbol Pcs. Item/Spare Part Mat. No.

  • 7/22/2019 Alcolyzer Plus Beer

    14/129

    5 Putting into Operation

    14 C27IB02

    5 Putting into Operation

    5.1 Installing the Instruments

    1. Put SP-1m into operation as described in the SP-1m instruction manual.

    2. Connect the power cord of SP-1m to a power terminal with ground

    connection.

    3. Place the density meter on SP-1m.

    4. Connect the connecting power cable (SP-1m accessory, Mat. No. 66178) to

    the power outlet of SP-1m and the power outlet of the DMA.

    5. Mount the filling nozzle fittings as described in the SP-1m instruction manual.

    6. Place Alcolyzer Plus Beer on top of the density meter.

    7. Connect the power cord of Alcolyzer Plus Beer to a power terminal with

    ground connection.

    Delivered with both

    options color + pH

    1 Connection cable type K1 BNC 1m 62521

    1 Buffer solution pH=4.01 Hamilton (500 mL) 125541 Buffer solution pH=7.00 Hamilton (500 mL) 12556

    1 pH storage solution 500 mL Hamilton 12632

    1 Color adjustment set

    Color standard approx. 50 EBC 2x50 mL

    Ultra pure water 0.0 EBC 3x50 mL

    12627

    Table 4.1: Supplied Parts

    Symbol Pcs. Item/Spare Part Mat. No.

    Information:

    Alcolyzer Plus Beer does not require any special installation conditions.

    Conditions in the installation area should be those in a typical laboratory.

    To guarantee temperature stability, do not place Alcolyzer Plus Beer

    - near a heater

    - near an air conditioner

    - in direct sunlight.

    For air humidity restrictions seeAppendix D.

    Caution:

    To ensure the correct functioning of the built-in ventilator, make sure the

    airflow at the bottom and the rear of the instrument is not blocked.

  • 7/22/2019 Alcolyzer Plus Beer

    15/129

    5 Putting into Operation

    C27IB02 15

    8. Screw the filling nozzle fittings Alcolyzer Plus (accessory box) into the

    openings of the measuring cell.

    9. If the option pH/color is integrated, check that 2# Filling nozzle Luer

    Alcolyzer Plus adapters are screwed into the openings of the pH measuring

    cell (see Fig. 5 - 1).

    Fig. 5 - 1 Mounting the adapters onto the pH measuring cell

    10. Connect COM2 at the rear of Alcolyzer Plus Beer to COM1 at the rear of

    DMA 4500/5000 using the connection cable (Mat. No. 2359).

    11. If a printer is used, connect the printer interface cable to COM1 at the rear of

    Alcolyzer Plus Beer.

    12. Connect the interface cable of SP-1m to the S-Bus outlet of Alcolyzer Plus

    Beer.

    Caution:

    Screw the adapters into the openings of the measuring cell hand-tight(without using any tools). Otherwise, the connections between the adapters

    and the measuring cell are not leak-tight and liquid may enter the electronics

    and affect operation.

    Caution:

    Before switching on Alcolyzer Plus Beer, check that the given voltage and

    frequency (AC100 to 240 V, 50 to 60 Hz) corresponds with the local mains

    voltage and mains frequency. If large voltage fluctuations are to be

    expected, the use of a constant voltage source (UPS) is recommended.

    The non-fused earth conductor of the power cord must be connected to

    earth.

  • 7/22/2019 Alcolyzer Plus Beer

    16/129

    5 Putting into Operation

    16 C27IB02

    5.2 Mounting the pH Electrode

    5.2.1 Preparing and Mounting the pH Electrode "Hamilton

    Polilyte Lab"

    Preparing

    1. Leave the pH electrode in an upright position for at least one hour.

    2. Carefully remove the watering cap.

    3. Flush the electrode with distilled water.

    4. Check the inside of the electrode for unwanted air bubbles. Shaking the

    electrode like a clinical thermometer will cause any bubbles to rise to the top.

    Mounting

    1. Loosen the protective tube by twisting it counterclockwise (see Fig. 5 - 2).

    Fig. 5 - 2 Loosening the protective tube

    2. Screw the connecting cable (Mat. No. 62521) to the rear end of the pH

    electrode.

    3. Connect the other end of the cable to the rear of Alcolyzer Plus Beer (see

    Chapter 6.2).

    4. Insert the pH electrode into the protective tube carefully until the end of the

    electrode touches the bottom of the pH measuring cell.

    5. Tighten the protective tube by turning it clockwise.

    The pH electrode is fixed.

  • 7/22/2019 Alcolyzer Plus Beer

    17/129

    5 Putting into Operation

    C27IB02 17

    Fig. 5 - 3 pH electrode Hamilton

    6. Check that both Filling nozzle Luer Alcolyzer Plus are screwed into the

    openings of the pH measuring cell (see Fig. 5 - 1).

    5.2.2 Preparing and Mounting the pH Electrode "Orion

    model 91-63"

    Preparing

    1. Remove the protective shipping cap from the sensing element and save for

    storage.

    2. Clean any salt deposits from the exterior by rinsing with distilled water.

    3. Uncover the filling hole and add filling solution to the electrode.

    4. Place the pH electrode in the electrode holder and suspend in air for 15

    minutes to thoroughly wet the reference junction.

    5. Shake the electrode (like a clinical thermometer) to remove air bubbles.

    6. Soak the electrode in pH electrode storage solution for one hour.

    Information:

    To maintain an adequate flow rate, the level of filling solution must always

    be above the reference junction.

    The filling hole must be open whenever the electrode is in use.

    Caution:

    Once the junction is wet, do not allow the electrode to dry out.

  • 7/22/2019 Alcolyzer Plus Beer

    18/129

    5 Putting into Operation

    18 C27IB02

    Mounting

    1. Screw the connecting cable (Mat. No. 62521) to the rear end of the pH

    electrode.

    2. Connect the other end of the cable to the rear of Alcolyzer Plus Beer (see

    Chapter 6.2).

    3. Insert the pH electrode into the pH measuring cell carefully until the end of

    the electrode touches the bottom of the pH measuring cell.

    Fig. 5 - 4 pH electrode Orion

    4. Tighten the knurled nut (see Fig. 5 - 4). The pH electrode is fixed.

    5. Check that both Filling nozzle Luer Alcolyzer Plus are screwed into the

    openings of the pH measuring cell (see Fig. 5 - 1).

    Knurled nut

  • 7/22/2019 Alcolyzer Plus Beer

    19/129

    5 Putting into Operation

    C27IB02 19

    5.3 Mounting the Hoses without Option pH/Color

    Fig. 5 - 5 Hose connections without option pH/color

    1. Cut a piece of the silicone hose 3 x 5 mm (1) approx. 24 cm in length.

    2. Connect one end of the hose to the air pump outlet of the DMA.

    3. Connect the Adapter Luer cone to the other end of the hose.

    4. Connect a Tygon hose (2) approx. 26 cm in length to the upper end of the

    SP-1m needle and to the front adapter of the density meter.

    5. Connect a Tygon hose (3) approx. 22 cm in length to the rear adapter of the

    density meter and to the lower adapter of Alcolyzer Plus Beer.

    6. Connect a Tygon hose (4) approx. 22 cm in length to the upper adapter of

    Alcolyzer Plus Beer.

    7. Connect the tube, Mat. No. 67982, (5) to the other end of the hose.

    8. Connect a silicone hose 2x4 mm (6) to the tube.

    9. Lead the silicone hose via the peristaltic pump of SP-1m to a waste vessel.

    10. The length of this hose can be selected according to the distance of the

    waste vessel.

    4 (T)

    3 (T)

    2 (T)

    5 (Tube)

    6 (S) to waste

    6

    1 (S)

    (T) .... Tygon

    (S) .... Silicone

  • 7/22/2019 Alcolyzer Plus Beer

    20/129

    5 Putting into Operation

    20 C27IB02

    5.4 Mounting the Hoses with Option pH/Color

    Fig. 5 - 6 Hose connections with option pH/color

    1. Connect an adapter Luer cone (1) to each opening of the pH measuring cell.

    2. Cut a piece of the silicone hose 3 x 5 mm (2) approx. 24 cm in length.

    3. Connect one end of the hose to the air pump outlet of the DMA.

    4. Connect the adapter Luer cone to the other end of the hose.

    5. Connect a Tygon hose (3) approx. 26 cm in length to the upper end of the

    SP-1m needle and to the front adapter of the density meter.

    6. Connect a Tygon hose (4) approx. 22 cm in length to the rear adapter of the

    density meter and to the lower adapter of Alcolyzer Plus Beer.

    7. Connect a Tygon hose (5) approx. 22 cm in length to the upper adapter of

    Alcolyzer Plus Beer.

    8. Connect the tube, Mat. No. 67982, (6) to the other end of the hose.

    9. Connect a silicone hose 2x4 mm (7) approx. 50 cm in length to the tube.

    10. Lead the silicone hose via the peristaltic pump of SP-1m to the front adapterof the pH measuring cell.

    5 (T)

    8 (S) to waste

    3 (T)

    6 Tube

    7 (S)

    4 (T)

    2 (S)

    1

    (T) .... Tygon

    (S) .... Silicone

    to airpump

  • 7/22/2019 Alcolyzer Plus Beer

    21/129

    5 Putting into Operation

    C27IB02 21

    11. Connect a 2x4 mm silicone hose (8) to the rear adapter of the pH measuring

    cell.

    12. Lead this hose to a waste vessel.

    13. Select the length of this hose according to the distance of the waste vessel.

    5.5 Completing the Installation

    1. Turn on the SP-1m and the density meter and wait for the initialization of

    both instruments to be completed.

    2. Configure the DMA 4500/5000 according toAppendix F.

    3. Turn on Alcolyzer Plus Beer using the power switch at the rear of the

    instrument.

    4. Select the menu instrument settings > device listand configure the device

    list as follows (for details about operating the menu, see Chapter 15.1):

    Caution:

    Before switching on Alcolyzer Plus Beer, check that the given voltage and

    frequency (AC 100 to 240 V, 50 to 60 Hz) corresponds with the local

    mains voltage and mains frequency. If large voltage fluctuations are to be

    expected, the use of a constant voltage source (UPS) is recommended.

    The non-fused earth conductor of the power cord must be connected to

    earth.

    Information:

    Once switched on, Alcolyzer Plus Beer performs a system check and anattemperating procedure:

    After transport, full stabilization of Alcolyzer Plus Beer might take up to

    one day.

    Usually Alcolyzer Plus Beer is ready for measurement after a start-up

    period of one hour.

    To guarantee an optimal temperature stability over a long time, always

    leave the system switched on.

  • 7/22/2019 Alcolyzer Plus Beer

    22/129

    5 Putting into Operation

    22 C27IB02

    Fig. 5 - 7 Menu: instrument settings/device list

    Now Alcolyzer Plus Beer initializes the communication between all

    connected peripherals.

    The COM2 interface of Alcolyzer Plus Beer is automatically configured

    (connected to: PIB, baud rate: 9600, format: 8 bit No) and cannot be

    changed unless densityis set to not used.

    If all devices are detected properly, the following message is displayed:

    Fig. 5 - 8 Device check

    5. If for a certain peripheral the message "offline" is displayed, check the

    interface settings and/or connection of the corresponding connection cable

    (seeAppendix F).

    INSTRUMENT SETTINGS

    filling devicedensityCO2O2

    SP-1m

    device list

    ::::

    DMA xxxx series

    not usednot used

    DEVICE CHECK

    filling unitdensity :: SP-1mpassed

  • 7/22/2019 Alcolyzer Plus Beer

    23/129

    5 Putting into Operation

    C27IB02 23

    Fig. 5 - 9 Device check failed

    To repeat the device check, select device check.

    To edit the device list, select change device list.

    6. Set the control parameters for SP-1m in the menu auto sampler settings

    (see Chapter 15.2.9).

    To make sure that the measured sample is fully replaced with the next

    sample, we recommend using the control parameter 1100 (100 seconds

    filling time, approx. 30 mL sample). For further information, see the SP-1m

    instruction manual.

    Fig. 5 - 10 Measuring system: Alcolyzer Plus Beer Analyzing System

    BUS ERROR

    filling unit : offline

    device check 20 secchange device list

  • 7/22/2019 Alcolyzer Plus Beer

    24/129

    5 Putting into Operation

    24 C27IB02

    5.6 Connecting a Printer or PC

    5.6.1 Connecting Alcolyzer Plus Beer to a Printer

    Connect the printer cable (Mat. No. 64433) to the COM1 on the rear side of

    Alcolyzer Plus Beer.

    Depending on the printer that you use, you can choose between 4 different printer

    types for optimal communication between the connected printer and Alcolyzer

    Plus.

    Set the COM1 interface to the appropriate printer in the menu instrument

    settings > COM1.

    5.6.2 Connecting Alcolyzer Plus Beer to a PC

    To transfer the measured results to a PC, connect the PC to Alcolyzer Plus Beer.

    For further details, seeAppendix F.

    1. Connect the interface cable DMA-PC (Mat. No. 72979) to the PC and to

    COM1 on the rear side of Alcolyzer Plus Beer.

    2. If you use the Anton Paar software VisioLab for Alcolyzer, set the COM1

    interface to VisioLabin the menu instrument settings > COM1.

    3. If you use the Anton Paar software AP-SoftPrint, set the COM1 interface to

    AP-SoftPrintin the menu instrument settings > COM1.

    Printer type Holding time

    after 1 line

    End-of-line

    character

    Remarks

    printer_1 800 msec LF CR after 40

    characters

    printer_2 1500 msec LF

    printer_3 800 msec LF

    printer_4 10 msec LF+CR

    AP-SoftPrint 10 msec LF

  • 7/22/2019 Alcolyzer Plus Beer

    25/129

    6 Functional Components

    C27IB02 25

    6 Functional Components

    6.1 Display and Keypad

    After switching on, Alcolyzer Plus Beer performs a self-test and the initialization.

    6.1.1 Display in the Measuring Mode

    Fig. 6 - 1 Display after the end of the measurement

    A ... MethodThe selected method is displayed.

    B ... Measuring status of the alcohol measurementFor details, see the reference manual, Appendix B.1.

    C ... Time

    D ... Sample nameE ... Measuring status of the color measurement

    For details, see the reference manual, Appendix B.2.

    F ... Measuring status of the pH measurementFor details, see the reference manual, Appendix B.3.

    G ... Measuring status of the density measurementThe current status of the density measurement is displayed.

    H ... Display area for measured values

    I ... Information about the sample changer

    J ... Soft keys

    Information:

    It may be necessary to adjust the contrast of the display with the black knob

    "Display contrast" at the rear of the instrument.

    BEERvalid09:44:21sample #validvalidvalid

    SP-1m

    %V/V 20 CBeer3c 7.4 EBCpH 4.43

    1.00521 g/cm3

    : stopped

    A

    B

    C

    D

    E

    F

    G

    H

    I

    J

  • 7/22/2019 Alcolyzer Plus Beer

    26/129

    6 Functional Components

    26 C27IB02

    Display of measured values:

    The alcohol content is displayed in %V/V (2 trailing digits) at 20C (percent

    volume per volume at the indicated temperature). If the measurement is not

    finished, "....." is displayed. As long as the measurement is still in progress

    ("predetermined"), only one trailing digit is displayed. If the measuring resultis lower than -0.50%V/V or higher than 30.00%V/V, "low" or "high" is

    displayed.

    The color measurement is displayed in EBC or ASBC with one trailing digit

    at 20C. If the measurement is not finished, "....." is displayed. As long as the

    measurement is still in progress ("predetermined"), no trailing digit is

    displayed. If the measuring result is lower than 0.5 EBC (-0.25 ASBC) or

    higher than 130.0 EBC (66 ASBC), "low" or "high" is displayed.

    The pH value is displayed in pH units (2 trailing digits) at 25C. If the

    measurement is not finished, "....." is displayed.

    Fig. 6 - 2 Display of further parameters

    K ... Display area for calculated valuesSelectable, calculated parameters are displayed in this area after pressing

    the key (see Chapter 7.5).

    BEERvalid09:44:21sample #validvalidalcoholpErEaRDF

    SP-1m

    %V/V 20 CBeer3c 7.4 EBCpH 4.43 3.86 %m/m

    11.12 %m/m 3.59 %m/m 1.80 %m/m 68.95 %

    : stopped

    KK

  • 7/22/2019 Alcolyzer Plus Beer

    27/129

    6 Functional Components

    C27IB02 27

    6.1.2 Soft Keys in the Measuring Mode

    Toggle key:

    To toggle between displays of measured andcalculated values

    To enter the menu (the key is

    available)

    Selects the measuring method.

    Selects the main menu for various settings and

    configurations. Access to the main menu can be

    protected by a password.

    Starts printing.

    This soft key is only displayed when interface

    COM1 or COM2 is configured for a printer in the

    menu instrument settings.

    Switches the air pump on.

    Switches the air pump off.

    Starts an automatic measuring process.

    If a sample changer SP-1m is connected to

    Alcolyzer Plus Beer and activated, the key

    on Alcolyzer Plus Beer is not displayed. Start the

    measurement with the key on the sample

    changer.

    Starts the next measurement.

    To enter the sample identification.

    Selects the list for the sample identification.

    Confirms the changes.

    Enters the SOP menu.

  • 7/22/2019 Alcolyzer Plus Beer

    28/129

    6 Functional Components

    28 C27IB02

    6.2 Rear View

    Fig. 6 - 3 Rear view

    A ... Cable of the pH electrode

    B ... pH electrode

    C ... "air pump inlet": Connection for a desiccator

    D ... "pH / color Module": Connection for the pH electrode

    E ... "COM2": Serial interface RS 232 for connecting a density meter or a

    printer

    F ... "Keyboard" Connection for a keyboard and/or bar code reader

    G ... Power inlet

    H ... Power switch

    I ... "COM1": Serial interface RS 232 for connecting a computer or printer

    J ... "S-Bus": Connection for a sample changer

    K ... "Display contrast"L ... Display contrast setting

    M ... Cooling fins

    A

    B

    C D E F G

    I

    J

    KL

    H

  • 7/22/2019 Alcolyzer Plus Beer

    29/129

    7 General Instrument Settings

    C27IB02 29

    7 General Instrument Settings

    7.1 Setting the Display Contrast

    If necessary, adjust the display contrast with the "Display contrast" knob at the

    rear of the instrument (see Chapter 6.2).

    7.2 Setting Date and Time

    Set date and time in the menu instrument settings > timeor dateusing the soft

    keys , or . Press to confirm the settings.

    7.3 Configuring the Measuring System

    1. Press and to enter the main menu.

    2. Select the menu instruments settings using the soft keys and .

    Confirm with .

    Fig. 7 - 4 Menu: instrument settings

    7.3.1 Setting the General Parameters

    Select the menu configurationto set general instrument settings.

    Use the soft keys , and the key to select the parameter.

    Press to confirm the settings.

    Information:

    For a detailed description of all setting options, see Chapter 15.2.2.

    INSTRUMENT SETTINGS

    device listCOM1COM2timedate

    configuration

  • 7/22/2019 Alcolyzer Plus Beer

    30/129

    7 General Instrument Settings

    30 C27IB02

    7.3.2 Defining the Peripheral Devices

    Change to the menu instrument settings > device list to select the

    connected peripherals like density meter, sample changer SP-1m or filling

    device SF-1.

    7.3.3 Defining the Interface Settings

    Configure the two interfaces COM1or COM2for computer or printer in the

    menu instrument settings.

    7.4 Setting a Custom Function

    You can define customized ethanol tables and color units.

    7.4.1 Setting a User-Defined Ethanol Table (in %V/V)

    If you want to use a different density/ethanol table than the OIML %V/V table, you

    can edit a polynomial function (4th order) that calculates an ethanol value based

    on your specific density/ethanol table.

    Calculating the coefficients

    The input value for the polynomial function is the ethanol %V/V value (c)

    measured by Alcolyzer Plus Beer. The result of the calculation depends on the

    entered coefficients according to the following formula.

    Polynomial formula: Output (%V/V) = A0 + A1 x (c [%V/V]) + A2 x (c [%V/V]) 2+

    A3 x (c [%V/V])3+ A4 x (c [%V/V])4

    The default values are the polynomial coefficients valid for the OIML %V/V table(see Fig. 7 - 5).

    Caution:

    You can only use density/ethanol tables based on %V/V.

    http://-/?-http://-/?-
  • 7/22/2019 Alcolyzer Plus Beer

    31/129

    7 General Instrument Settings

    C27IB02 31

    Fig. 7 - 5 Default setting of the polynomial coefficients

    1. To calculate the coefficients, take the appropriate table (%V/V) that converts

    the measured density into a corresponding ethanol value.

    2. By using an appropriate mathematical program, you can calculate the

    polynomial coefficients which are based on the table data.

    Editing the user-defined ethanol table

    1. Select the menu user functions > user defined ethanol table.

    2. Enter the 5 polynomial coefficients (A0 to A4) as calculated.

    Activating the user-defined ethanol table

    1. At Alcolyzer Plus Beer, select the menu instrument settings >

    configuration > calc. parameter.

    2. Select CU V/Vand press the key.

    7.4.2 Setting a User-Defined Color Unit

    If the color measurement according to EBC or ASBC is not desired, you can

    define a user-defined color unit. This is done by correcting the EBC value c [EBC]

    according to the formula shown in Fig. 7 - 6.

    Editing the user-defined color unit

    1. To enter the correction factor for the slope of the color measurement and the

    offset, select the menu user functions > user defined color unit.

    USER FUNCTIONS

    user defined ethanol table

    A0 : 9.982000e-01A1 : - 1.520000e-03A2 : + 1.924620e-05A3 : - 1.461290e-07A4 : - 4.574060e-09

    +

    Caution:

    To obtain a correct alcohol adjustment of Alcolyzer Plus Beer, use the same

    ethanol table in the DMA that you used for calculating the polynomial

    coefficients (seeAppendix F).

    http://-/?-http://-/?-
  • 7/22/2019 Alcolyzer Plus Beer

    32/129

    7 General Instrument Settings

    32 C27IB02

    The following window is displayed:

    Fig. 7 - 6 Menu: User functions/user defined color unit

    2. Enter a slope value between 0 and 7.

    Use the key to move the cursor. Edit the highlighted number using the and keys. Press to confirm the changes.

    3. Enter an offset value between -10 and +10.

    If not allowed values are entered, the slope is automatically set to 1.0 and

    the offset to 00.0 after pressing .

    Activating the user-defined color unit

    1. Select the menu instrument settings > configuration > color unit.

    2. Select CUand press the key.

    7.5 Selecting the Output Data

    When you select the parameters for display, printer and memory, you get a list

    containing all parameters that are available for the predefined methods BEER,

    CIDER, SOFT DRINK, WASH and MOLASSES (seeAppendix B.1).

    If you then select your appropriate measuring method, you get only the

    parameters that are of importance for this very method.

    Example: If you select the parameter p (original extract) [%Plato], this parameter

    is only available in the methods BEER, WASH and MOLASSES.

    USER FUNCTIONS

    user defined color unit

    c [CU] = .000 * c [EBC] + 00.001

  • 7/22/2019 Alcolyzer Plus Beer

    33/129

    7 General Instrument Settings

    C27IB02 33

    7.5.1 Selecting the Output Data for the Display

    Change to the menu utilities > display configuration.

    Select up to 5 parameters using the soft keys , and thekey.

    Press to confirm the settings.

    7.5.2 Selecting the Output Data for the Printer

    You can select output data for the printer and save this selection in a printer

    protocol. A printer protocol consists of a head and parameters. There are 4

    factory-set printer protocols and 6 free printer protocols available.

    Factory-set printer protocols:

    - Printer protocol No. 6, used with BEER method

    - Printer protocol No. 7, used with CIDER method

    - Printer protocol No. 8, used with WASH and MOLASSES method

    - Printer protocol No. 9, used with SOFT DRINK method.

    For details of the settings, seeAppendix B.2.

    If required, you can change these pre-settings according to your demands.

    To define or change a printer protocol, change to the menu utilities > output

    configuration > printer protocol:

    To activate/deactivate the parameters, use the key.

    Press to confirm the settings.

    7.5.3 Selecting the Output Data for the Data Memory

    Change to the menu utilities > output configuration > data memory.

    You can activate up to 10 parameters to be stored in the data memory:

    To activate/deactivate the parameters, use the key.

    Press to confirm the settings.

    7.6 Settings for the Sample Changer

    Enter the control parameter for SP-1m in the menu auto sampler settings

    as described in the SP-1m instruction manual.

    Information:

    If data should be sent to a LIMS system, activate additionally the parameters

    alcohol cond.and color cond.

  • 7/22/2019 Alcolyzer Plus Beer

    34/129

    8 Setting the Measuring Methods

    34 C27IB02

    This menu is only displayed if a sample changer is connected and configured

    in the menu instrument settings > device list > filling device.

    8 Setting the Measuring Methods

    8.1 Description of the Measuring Methods

    Alcolyzer Plus Beer provides 10 different methods:

    Fig. 8 - 1 Menu: method settings

    Method 1: BEER

    Standard method for beers. This method can be used for almost all samples.

    Method 2: CIDER

    Method for ciders.

    Method 3: SOFT DRINK

    Method for alcohol-free beverages.

    Method 4: WASH

    Method for wash. Wash is an intermediate product of whisky production. It is an

    alcohol containing liquid produced in the fermentation of worts by yeasts. The

    alcohol content in wash varies between 7 and 10%vol.

    Information:

    We recommend setting the following filling times: Without option pH/color: 80 seconds (sample changer parameter "1080").

    With option pH/color: 100 seconds (sample changer parameter "1100").

    METHOD SETTINGS

    no. 1no. 2no. 3no. 4no. 5no. 6no. 7no. 8

    no. 9no. 10

    ::::::::

    ::

    BEERCIDERSOFT DRINKWASH

    new method

    MOLASSES

  • 7/22/2019 Alcolyzer Plus Beer

    35/129

    8 Setting the Measuring Methods

    C27IB02 35

    Method 5: MOLASSES

    Method for molasses. Molasses is a honey-like, dark-brown sugar syrup which is

    used as a fermentation raw material in the production of bioethanol. With this

    method, you can determine the ethanol content of fermenting or ready-fermented

    molasses.

    Methods 6 to 10:

    These methods are sample-specific and can be edited as described in Chapter

    8.2.

    8.2 Editing a Measuring Method

    8.2.1 Editing a Predefined Method

    You can edit the predefined methods "BEER", "CIDER", "SOFT DRINK", "WASH"

    or "MOLASSES" to assign another printer protocol or enter a residual CO2content.

    1. In the menu method settings, select one of the predefined methods and

    confirm with .

    2. In protocol no.:, you can select one of 10 printer protocols to assign it to the

    selected method (For editing a printer protocol, see Chapter 7.5.2).

    8.2.2 Editing a New Method

    You can edit a new method to assign a special printer protocol (see Chapter 7.5.2)

    or to enter a sample-specific alcohol correction factor (see Chapter 12.4). For the

    types "BEER", "CIDER" and "SOFT DRINK", the residual CO2can be entered

    (see Chapter 8.2.1).

    1. Select a row between 6 and 10 in the menu method settings.

    "new method" is displayed.

    2. Confirm with .

    3. Enter a name for the method and confirm the method name with .

    Information:

    If you measure fermenting molasses, we recommend to filtrate the sample

    prior to measurement.

    Information:The menu item residual co2 is only used with PBA-B and the methods

    "BEER", "CIDER" and "SOFT DRINK".

  • 7/22/2019 Alcolyzer Plus Beer

    36/129

    8 Setting the Measuring Methods

    36 C27IB02

    4. Select a type"BEER", "CIDER", "SOFT DRINK", "WASH" or "MOLASSES"

    for the new method.

    The type has to correspond with the sample to be measured.

    5. If necessary, enter an offset between 0.99%vol and +0.99%vol (seeChapter 12.4).

    6. If necessary, enter a pH-offset between -0.99 and +0.99 in pH-deg. offset.

    7. In protocol no.:,you can assign a printer protocol from 1 to 10 to the new

    edited method.

    For details about editing a printer protocol, see Chapter 7.5.2.

    8. Press to confirm the settings.

    Fig. 8 - 2 Menu: method settings/new method/name

    EDIT METHOD no. 6

    typeoffsetpH-deg.offset

    BEER:::

    +0.50 %V/V

    name special beer:

    protocol no.residual co2

    ::

    +0.50050.00 g/l

    Information:

    The menu item residual co2 is only used with PBA-B and the methods

    "BEER", "CIDER" and "SOFT DRINK".

  • 7/22/2019 Alcolyzer Plus Beer

    37/129

    8 Setting the Measuring Methods

    C27IB02 37

    8.3 Selecting the Measuring Method

    1. Press the key in the measuring window.

    The following window is displayed:

    Fig. 8 - 3 Selecting a method

    2. Use the key to toggle between the different methods.

    3. Select the appropriate method (e.g. "BEER") with the key.

    The selected method is displayed in the left upper corner.

    BEER

    BEERvalid09:44:21

    validvalid

    method :

    %V/V 20 C

    c 7.4 EBCpH 4.43

    Information:

    If you have assigned another method to a certain sample in the sample list,

    this method instead of the selected method is used to measure the sample

    (see Chapter 12.1.3.1).

  • 7/22/2019 Alcolyzer Plus Beer

    38/129

    9 SOP (Standard Operating Procedure)

    38 C27IB02

    9 SOP (Standard Operating Procedure)

    To ensure that Alcolyzer Plus Beer gives correct results, check the adjustments

    of the measured parameters regularly.

    The integrated SOP (Standard Operating Procedure) allows you to perform an

    automated control measurement of all measuring parameters as well as an

    adjustment afterwards.

    Carrying out the SOP

    1. Select the menu utilities > SOP configuration and choose SOP #1 or

    SOP #2.

    "SOP #1" is defined as the "routine" SOP and contains the parameters water,

    alcohol, pH1 and pH2.

    2. Place the appropriate control solutions in the correct order into the sample

    changer magazine (for preparing the control solutions, see Chapter 10).

    The following control solutions are used for the SOP:

    - Bi-distilled water to check the zero point of the alcohol and the color

    adjustment

    - A binary ethanol/water solution with an alcohol content of 8 to 12%V/V to

    check the "known concentration" of the alcohol adjustment

    - pH buffer pH 7.00 and

    - pH buffer pH 4.01 to check the pH adjustment

    3. Press the key in the measuring window of Alcolyzer Plus Beer to

    access the following menu:

    Information:

    The SOP can only be used in combination with a sample changer.

    Information:

    Carry out the "routine" SOP once a day. If desired, an additional SOP

    (SOP #2) can be defined as an "extended" SOP with additional parameters

    (e.g. color standard).

  • 7/22/2019 Alcolyzer Plus Beer

    39/129

    9 SOP (Standard Operating Procedure)

    C27IB02 39

    Fig. 9 - 1

    Use the key to toggle between "SOP #1" and "SOP #2".

    4. Press the key to start the SOP.

    5. Press the key on the sample changer.

    The control solutions are measured automatically.

    After the last measurement, the following window is displayed:

    Fig. 9 - 2

    6. If a measured value deviates significantly from the desired value, this value

    can be readjusted now.

    STANDARD OPERATING PROCEDURE #1

    #1:2:

    3:4:

    #

    ##

    wateralcohol

    pH 1pH 2

    adjust ?adjust ?

    adjust ?

    STANDARD OPERATING PROCEDURE #1

    water

    alcoholpH 1pH 2

    SP-1m :

    -0.01 %V/V 0.0710.01

    EBC %V/V

    4.02 pH 6.99 pH

    stopped

    adjust ?

    10: 32: 54

  • 7/22/2019 Alcolyzer Plus Beer

    40/129

    9 SOP (Standard Operating Procedure)

    40 C27IB02

    7. Highlight the respective row "adjust?" and press the key (see Fig. 9 - 2).

    8. Enter the desired value and press .

    Fig. 9 - 3

    The following window is displayed (in this example, the alcohol value has

    been readjusted):

    Fig. 9 - 4

    Information:

    The adjustment of the known concentration for the alcohol can only be

    performed successfully if an accurate density meter and an appropriate

    density/alcohol concentration table are available.

    If the known concentration for the alcohol shall be adjusted, the alcohol

    value measured by the DMA is displayed in a line below the SOP

    parameters after pressing the key. This value can be taken directlyfor the readjustment.

    STANDARD OPERATING PROCEDURE #1

    water

    alcoholpH 1pH 2

    adjust ?adjust ?

    adjust ?

    10: 34: 27

    -0.01 %V/V 0.0710.01

    EBC %V/V

    4.02 pH 6.99 pH

    1concentration: 0.05 %V/V

    adjust ?adjustedadjust ?

    STANDARD OPERATING PROCEDURE #1

    water

    alcoholpH 1pH 2

    -0.01 %V/V 0.0710.05

    EBC %V/V

    4.02 pH 6.99 pH

    adjust ?

    10: 35: 16

  • 7/22/2019 Alcolyzer Plus Beer

    41/129

    10 Checking the Adjustment Validity

    C27IB02 41

    9. Finish the SOP by pressing the key or the key when all

    parameters are readjusted.

    10 Checking the Adjustment Validity

    If no SOP (see Chapter 9) is carried out, check the adjustments regularly as

    described below.

    10.1 Checking Procedure without Option pH/Color

    Before each series of measurements, check the zero point using degassed,

    bi-distilled water.

    To check the validity of the slope of the adjustment, use a water/ethanol

    solution of known concentration. Perform a slope check at least once everyweek.

    10.1.1 Performing a Checking Procedure with Water (0%V/V)

    1. Boil fresh, bi-distilled water for several minutes to remove dissolved gas.

    2. Fill a clean glass flask full with the boiled water and cover it.

    3. Wait until the water has cooled down to 20 2C.

    4. Clean the measuring cell as described in Chapter 14.

    Information:

    If one (or both) pH buffer deviates too much from the correct value, the

    measured pH value and the measured temperature will be displayed in a

    line below the SOP parameters after confirming the question "adjust ?".

    The pH value can then be corrected according to the pH/temperature

    table of the corresponding buffer.

    The readjustment for the pH measurement is carried out for both

    adjustment points simultaneously. Therefore, only one "adjust? is

    displayed for both buffers.

    Information:

    For the color adjustment, it is only necessary to adjust the zero point with

    bi-distilled water regularly. If it is desired to check also the known concentration for the color

    measurement (e.g. with a standard beer or color standard) this parameter

    can be added to the SOP. If the known concentration for the color

    measurement shall be adjusted, the color value of the color standard can

    be entered in a line below the SOP parameters after pressing the key. Color standards are available at Anton Paar. Contact your local distributor.

  • 7/22/2019 Alcolyzer Plus Beer

    42/129

    10 Checking the Adjustment Validity

    42 C27IB02

    5. Fill a sample vial with degassed, bi-distilled water and put it into the

    magazine of the sample changer.

    6. Press the key on the sample changer.

    7. Wait until the measuring status displays "valid".

    8. If the displayed value is 0 0.03%V/V and is stable for a few minutes,

    measurements can be performed.

    9. If the displayed value is outside this concentration range, clean the

    measuring cell and perform another control measurement.

    10. If the measuring result is still out of range (0 0.03%V/V) perform a zero

    point adjustment (see Chapter 11.1.1).

    11. If "no sample" is displayed, clean the measuring cells according to Chapter

    14.1.3or Chapter 14.2.4(with option pH/color).

    12. Then perform an adjustment for the zero point and for the known

    concentration.

    Checking the validity of the density adjustment

    1. As soon as the DMA gives a valid result (flashing display), check the density

    reading.

    2. If the result deviates more than 0.0001 g/cm3(DMA 4500) or 0.00005 g/cm3

    (DMA 5000) from the reference value (0.998203 g/cm3), perform a cleaning

    procedure according to Chapter 14.

    3. Repeat the checking procedure.

    4. If the density result still exceeds the max. allowed deviation, perform an air/

    water adjustment for the density meter (see Chapter 11.3).

    Information:

    If a density meter (DMA 4500, DMA 5000) is connected to Alcolyzer Plus

    Beer, check also the validity of the density adjustment as described below.

  • 7/22/2019 Alcolyzer Plus Beer

    43/129

    10 Checking the Adjustment Validity

    C27IB02 43

    10.1.2 Performing a Checking Procedure with a Water/

    Ethanol Solution of Known Concentration

    1. Prepare a water/ethanol solution with a concentration between 8 and

    12%V/V.

    2. Cover the flask and leave the solution for several hours at a temperature of

    20 3C.

    Before filling the measuring cell, the sample must be free of air bubbles.

    3. Clean the measuring cell, if necessary (see Chapter 14).

    4. Fill a sample vial with the water/ethanol solution and put it into the magazine

    of the sample changer.

    5. Press the key on the sample changer.

    6. Wait until Alcolyzer Plus Beer and the DMA show "valid".

    7. If the difference between the measuring values from Alcolyzer Plus Beer and

    the DMA is less than 0.03%V/V and the results remain stable within this

    range for a few minutes, the adjustment of Alcolyzer Plus Beer is correct.

    8. If the displayed value is out of this range, clean the measuring cell andperform another control measurement.

    9. If the measuring result is still out of range, perform an adjustment using a

    water/ethanol solution (see Chapter 11.1.2).

    Information:

    This measurement can only be performed successfully if an accurate density

    meter and an appropriate density / alcohol concentration table is available.

    Caution:

    Only use high-proof ethanol without denaturant!

    Information:

    If Alcolyzer Plus Beer is used in combination with a DMA 4500/5000, the

    alcohol concentration is directly displayed on the density meter.

  • 7/22/2019 Alcolyzer Plus Beer

    44/129

    10 Checking the Adjustment Validity

    44 C27IB02

    10.2 Checking Procedure with Option pH/Color

    10.2.1 Checking Procedure for the Measuring Parameter

    "alcohol"

    The checking procedure for the measuring parameter "alcohol" is performed as

    described in Chapter 10.1.

    10.2.2 Checking Procedure for the Measuring Parameter "pH"

    Check the validity of the pH adjustment with buffers pH 7 and pH 4 once a day

    (before starting the routine measurements).

    Control measurement for buffer pH 7

    1. Fill a sample vial with buffer pH 7 and place it in the sample changer

    magazine.

    2. Press the key on the sample changer.

    After filling is completed, the measurement starts automatically.

    3. Wait until the pH value is displayed in the measuring window ("valid").

    This value is valid for 25C.

    4. If the difference between the measured pH value and the set point is less

    than 0.05 pH units, the pH adjustment is correct.

    5. If the displayed value is out of this range, fill the pH measuring cell with the

    same buffer and repeat the control measurement.

    Caution:

    Always use fresh pH buffers.

    Information:

    Instead of using the sample changer, one of the following filling procedures

    can be used:

    Disconnect the silicone hose that is plugged to the front adapter of the pH

    measuring cell. Fill a 20 mL syringe with the buffer. Connect the syringe to

    the front adapter of the pH measuring cell and fill in the buffer.

    Or:

    Remove the pH electrode from the pH measuring cell and immerse the

    electrode into a vessel filled with pH buffer. Ensure that the buffer

    temperature does not deviate significantly from the temperature in the pH

    measuring cell as the temperature sensor for the pH measurement is

    located in the pH measuring cell.

  • 7/22/2019 Alcolyzer Plus Beer

    45/129

    11 Adjusting the Instrument

    C27IB02 45

    6. If the displayed value is still out of range, perform an adjustment (see

    Chapter 11.2.3).

    Control measurement for buffer pH 4

    The control measurement is performed in the same way as for buffer pH 7.

    10.2.3 Checking Procedure for the Measuring Parameter

    "color"

    Checking procedure with water (0.0 EBC)

    This procedure should preferably be performed together with the checking

    procedure for 0% alcohol (see Chapter 10.1.1)

    1. If the displayed value is 0 0.3 EBC (0 0.15 ASBC), measurements can beperformed.

    2. If the displayed value is out of this range, clean the measuring cells and

    repeat the control measurement.

    3. If the measuring result is still out of range, perform a "zero" adjustment (seeChapter 11.2.2).

    11 Adjusting the Instrument

    11.1 Adjustment without Option pH/Color

    Regular adjustments are performed with degassed bi-distilled water (0%V/V

    alcohol, zero point).

    Extended adjustments are performed with degassed bi-distilled water and a

    water/ethanol solution of exactly known concentration in the range of 2 to

    30 %V/V, preferably 8 to 12%V/V.

    An extended adjustment is necessary if the checking procedure with a water/

    ethanol solution of known concentration shows deviations of more than

    0.03 %V/V from the reference value, even after correct zero point

    adjustment.

    Information:

    During the first months, usually a drift of the color value of approx. 0.1 EBC

    per day can be observed. This drift (in the positive direction) is a result of a

    decreasing LED intensity.

  • 7/22/2019 Alcolyzer Plus Beer

    46/129

    11 Adjusting the Instrument

    46 C27IB02

    If the adjustment of Alcolyzer Plus Beer is properly performed, it is sufficient

    to check only the zero point before each series of measurements.

    11.1.1 Adjustment with Water (Zero Point)

    1. Boil fresh, bi-distilled water for several minutes to remove dissolved gas.

    2. Fill a clean glass flask full with the boiled water and cover it.

    3. Wait until the water has cooled down to approx. 20 2C.

    4. Select Alcolyzer Plus Beer menu adjustment > adjust alcohol > zero.

    5. Answer the question "Do you really want to adjust?" with .

    If there is no bi-distilled water in the measuring system, press on

    Alcolyzer Plus Beer. Then press the key on the sample changer.

    The bi-distilled water is now filled into the measuring system by the sample

    changer. After the filling procedure is completed, the adjustment procedure

    starts automatically.

    If bi-distilled water has already been filled into the measuring system, press

    the key instead of the key to start the adjustment

    procedure immediately.

    6. Wait until the adjustment has been completed.

    Information:

    Before the adjustment, Alcolyzer Plus Beer has to be turned on for at leastone hour. However, after transport full stabilization of Alcolyzer Plus Beer

    might take up to one day.

    The ambient temperature should be stable to < 3C.

    Caution:

    Before starting the adjustment, ensure that the measuring cells are properly

    cleaned (see Chapter 14).

  • 7/22/2019 Alcolyzer Plus Beer

    47/129

    11 Adjusting the Instrument

    C27IB02 47

    Fig. 11 - 1 Successful adjustment

    7. If the measurement value is stable to 0.02%V/V, the adjustment is

    successful and "adjustment saved" is displayed (see Fig. 11 - 1).

    Press to confirm the adjustment.

    Press the key three times to return to the measuring window.

    8. If "no stability - restart" is displayed after the adjustment, the required stability

    of the measuring result could not be achieved. Problems during sample

    filling (for example, air bubbles in the cell) could be the reason.

    Clean the cells again and repeat the adjustment with bi-distilled water.

    ADJUSTMENT ALCOHOL

    S1 = 87.30 % of factory adjustment

    adjustment 0.00 %V/V

    adjustment saved

    Information:

    Additionally, the value S1 gives information about the measured signal

    intensity of the current adjustment compared to the signal intensity at the

    factory adjustment. This value decreases by-and-by due to ageing of the LED

    and/or coating of the measuring cell.

    If the S1 value is below 80%, clean the measuring cell thoroughly (see

    Chapter 14) and repeat the zero adjustment. If the S1 value is still too low,

    contact your Anton Paar representative.

    Information:

    After this adjustment we recommend performing the zero adjustment for the

    optional color measurement because the bi-distilled water which is inside the

    cells can be used for the zero adjustment of the color measurement.

    http://-/?-http://-/?-
  • 7/22/2019 Alcolyzer Plus Beer

    48/129

    11 Adjusting the Instrument

    48 C27IB02

    11.1.2 Adjustment with a Water/Ethanol Solution of Known

    Alcohol Concentration

    1. Prepare a water/ethanol solution.

    2. Cover the flask and leave the solution for several hours at 20 3C. Before

    filling the measuring cell, the sample must be free of gas bubbles.

    3. Select the Alcolyzer Plus Beer menu adjustment > adjust alcohol > known

    concentration.

    4. Answer the question "Do you really want to adjust?" with .

    5. If there is no water/ethanol solution in the measuring system, press

    on Alcolyzer Plus Beer.

    6. Press the key on the sample changer.

    The water/ethanol solution is now filled into the measuring system by the

    sample changer. After the filling procedure is completed, the adjustment

    procedure starts automatically.

    If a DMA 4500/5000 is connected and configured properly (seeAppendix F),

    "DMA: measuring" is displayed on Alcolyzer Plus Beer. As soon as "DMA:

    valid" is displayed, the alcohol concentration that comes from the DMA is

    displayed in the first line of Alcolyzer Plus Beer display and can be used

    directly as the "known concentration" value for the adjustment.

    7. Press the key to continue the adjustment.

    8. Wait until the adjustment is finished.

    Caution:

    Only use high-proof ethanol without denaturant.

    Information:

    The concentration of the water/ethanol solution for the adjustment must be

    between 2 and 30%V/V, preferably between 8 and 12%V/V alcohol.

    Caution:

    Before starting the adjustment, ensure that the measuring cells are properly

    cleaned (see Chapter 14).

    Information:

    If a water/ethanol solution has already been filled into the measuring system,

    press the key instead of the key.

  • 7/22/2019 Alcolyzer Plus Beer

    49/129

    11 Adjusting the Instrument

    C27IB02 49

    9. If the measurement is stable to 0.02%V/V, the adjustment was successful

    and "adjustment saved" is displayed.

    Press to confirm the adjustment.

    Press the key three times to return to the measuring window.

    10. If "adjustment failed" is displayed after the adjustment, the required stability

    of the measuring result could not be achieved. Problems during sample

    filling (for example, air bubbles in the cell) could be the reason.

    Clean the cells again and repeat the adjustment with a water/ethanol

    solution.

    11.2 Adjustment with Option pH/Color

    11.2.1 Adjustment Alcohol

    The adjustment for the measuring parameter "alcohol" is performed as described

    in Chapter 11.1.

    11.2.2 Adjustment Color

    For the color adjustment, use the delivered color adjustment set.

    11.2.2.1 Adjustment with Water (Zero)

    1. Fill a clean sample vial with the delivered water standard (ultra pure water)

    and place the vial in the sample magazine.

    2. Select the menu adjustment > adjustment color > zero.

    3. Answer the question "Do you really want to adjust?" with .

    4. Press the key on Alcolyzer Plus Beer.

    5. Press the key on the sample changer.

    The ultra pure water is filled into the measuring system.

    Information:

    Alcohol solutions have a tendency to form gas bubbles. Handle the sample

    with care.

    Caution:

    Before starting the adjustment, ensure that the measuring cells are properly

    cleaned (see Chapter 14).

  • 7/22/2019 Alcolyzer Plus Beer

    50/129

    11 Adjusting the Instrument

    50 C27IB02

    After the filling procedure is completed, the adjustment procedure starts

    automatically.

    6. Wait until the adjustment is finished.

    Fig. 11 - 2 Successful adjustment

    If the measurement value is stable, the adjustment is successful,

    "adjustment saved" is displayed (see Fig. 11 - 2).

    Press to confirm the adjustment.

    Press the key three times to return to the measuring window.

    7. If "no stability - restart" is displayed after the adjustment, the required stability

    of the measuring result could not be achieved. Problems during sample

    filling (for example, air bubbles in the cell) could be the reason.

    Clean the cells again and repeat the adjustment with bi-distilled water.

    ADJUSTMENT COLOR

    S = 98.30 % of factory adjustment

    adjustment 0.00 EBC adjustment saved

    Information:

    Additionally, the value S gives information about the measured signal intensity

    of the current adjustment compared to the signal intensity at the factory

    adjustment. This value decreases by-and-by due to ageing of the LED and/orcoating of the measuring cell.

    If the S value is below 80%, clean the measuring cell thoroughly (see Chapter

    14) and repeat the zero adjustment. If the S value is still too low, contact your

    Anton Paar representative.

    Information:

    After the color adjustment has been done properly, it is only necessary to

    check the zero point adjustment with bi-distilled water regularly.

    http://-/?-http://-/?-
  • 7/22/2019 Alcolyzer Plus Beer

    51/129

    11 Adjusting the Instrument

    C27IB02 51

    11.2.2.2 Adjustment with a Color Standard (Tartrazine) of Known

    Concentration

    1. Fill a sample vial with the delivered color standard (tartrazine) and place it in

    the sample changer magazine.

    2. Select the menu adjustment > adjust color > known concentration.

    3. Answer the question "Do you really want to adjust?" with .

    4. Press the key on Alcolyzer Plus Beer.

    5. Press the key on the sample changer.

    The standard is filled into the measuring cells.

    6. Enter the EBC value of the delivered standard in the line "color".

    7. Press the key to continue with the adjustment.

    8. Wait until the adjustment is finished.

    9. If the measurement is stable, the adjustment was successful and

    "adjustment saved" is displayed.

    Press to confirm the adjustment.

    Press the key three times to return to the measuring window.

    10. If "adjustment failed" is displayed after the adjustment, the required stability

    of the measuring result could not be achieved. Problems during sample

    filling (for example, air bubbles in the cell) could be the reason.

    Clean the cells again and repeat the adjustment with the color standard.

    Information:

    The concentration of the color standard for the adjustment must be between1.00 and 60.00 EBC (0.50 to 30.48 ASBC), preferably at approx. 50.00 EBC

    (25.40 ASBC).

    Information:

    After the color adjustment has been done properly, it is only necessary to

    check the zero point adjustment with bi-distilled water regularly.

  • 7/22/2019 Alcolyzer Plus Beer

    52/129

    11 Adjusting the Instrument

    52 C27IB02

    11.2.3 Adjustment pH

    The adjustment of the pH electrode is performed with two buffers (pH 7 and pH 4).

    Check the validity of the pH adjustment with buffers pH 7 and pH 4 once a day

    (before starting the routine measurements). For details, see Chapters 9 and

    10.2.2.

    For the pH buffers "Hamilton Duracal" the buffer data (pH value depending on the

    temperature) are integrated in Alcolyzer Plus Beer.

    If the pH electrode is adjusted with these buffers, the correct buffer data are

    selected automatically.

    If you use a different buffer to "Hamilton Duracal", the buffer data (pH value

    depending on the temperature) can be entered in the menu edit buffer data. If

    the pH electrode is then adjusted with these buffers, the correct buffer data areselected automatically.

    Caution:

    Always use fresh pH buffers.

    Temperature (C) Buffer pH 7.00 Buffer pH 4.01

    5 7.09 4.01

    10 7.06 4.00

    15 7.04 4.00

    18 7.03 4.00

    20 7.02 4.00

    22 7.01 4.00

    25 7.00 4.01

    30 6.99 4.01

    35 6.98 4.02

    40 6.97 4.03

    45 6.97 4.04

    50 6.97 4.05

    Information:

    For the adjustment, the first buffer should be pH 7, the second buffer should

    be pH 4.

  • 7/22/2019 Alcolyzer Plus Beer

    53/129

    11 Adjusting the Instrument

    C27IB02 53

    11.2.3.1 Automatic

    The adjustment is usually performed with "Hamilton Duracal" buffers. However,

    the adjustment can be performed with other buffers. The data of these buffers

    have to be entered in the menu edit buffer data. The edited buffers have to be

    selected in the menu sort buffer data.

    1. Select menu adjustment > adjust pH > automatic.

    "clean pH sensor/expose to buffer 1" is displayed.

    2. To fill buffer pH 7 in the measuring cells using the sample changer SP-1m,

    press first the key on the Alcolyzer and then the key on

    the sample changer.

    3. After the filling procedure is completed, wait until the voltage and the pH

    value is stable (also check if the pH value is plausible) before pressing the

    key to start the adjustment.

    "Standard 7.00" (or the name of your buffer) as well as a number (in row"adjust") is displayed.

    4. Press the key.

    The adjustment with the first buffer is finished.

    "clean pH sensor/expose to buffer 2" is displayed.

    5. Fill buffer pH 4 in the measuring cells and wait until the voltage and the pH

    value is stable (also check if the pH value is plausible).

    6. Then press the key to start the adjustment.

    Information:

    Instead of using the sample changer, one of the following filling procedurescan be used:

    Disconnect the silicone hose that is plugged to the front adapter of the pH

    measuring cell. Fill a 20 mL syringe with the buffer. Connect the syringe to

    the front adapter of the pH measuring cell and fill in the buffer. Use at least

    15 mL buffer to make sure that the previous liquid in the pH measuring

    cell is fully replaced.

    Or:Remove the pH electrode from the pH measuring cell and immerse the

    electrode into a vessel filled with pH buffer. Ensure that the buffer

    temperature does not deviate significantly from the temperature in the pH

    measuring cell as the temperature sensor for the pH measurement is

    located in the pH measuring cell.

    Information:

    If "no buffer saved" is displayed, the measured voltage deviates from the

    stored tables (Hamilton or your own tables).

  • 7/22/2019 Alcolyzer Plus Beer

    54/129

    11 Adjusting the Instrument

    54 C27IB02

    "Standard 4.01" (or the name of your buffer) is displayed. The adjustment is

    finished. The voltages of the adjustment data (for the offset and the slope)

    are displayed.

    7. Press the key to finish the adjustment.

    11.2.3.2 Manual

    The pH adjustment can also be performed manually. To do this, any buffer can be

    used.

    1. Select menu adjustment > adjust pH > manual.

    "clean pH sensor/expose to buffer 1" is displayed.

    2. To fill buffer pH 7 in the measuring cells using the sample changer SP-1m,

    press first the key on the Alcolyzer and then the key onthe sample changer.

    3. After the filling procedure is completed, wait until the voltage and the pH

    value is stable (also check if the pH value is plausible) before pressing the

    key to start the adjustment.

    4. Enter the theoretical pH value of the buffer (pH value depending on the

    current temperature) using either the , , keys or a keyboard.

    5. Then press .

    6. Wait until the adjustment with the first buffer is finished.

    7. Then press the key.

    "clean pH sensor/expose to buffer 2" is displayed.

    8. Fill buffer pH 4 in the measuring cells and wait until the voltage and the pH

    value is stable (also check if the pH value is plausible).

    9. Then press the key to start the adjustment.

    Information:

    Instead of using the sample changer, one of the following filling procedures

    can be used:

    Disconnect the silicone hose that is plugged to the front adapter of the pH

    measuring cell. Fill a 20 mL syringe with the buffer. Connect the syringe to

    the front adapter of the pH measuring cell and fill in the buffer. Use at least

    15 mL buffer to make sure that the previous liquid in the pH measuring

    cell is fully replaced.

    Or:Remove the pH electrode from the pH measuring cell and immerse theelectrode into a vessel filled with pH buffer. Ensure that the buffer

    temperature does not deviate significantly from the temperature in the pH

    measuring cell as the temperature sensor for the pH measurement is

    located in the pH measuring cell.

  • 7/22/2019 Alcolyzer Plus Beer

    55/129

    11 Adjusting the Instrument

    C27IB02 55

    10. Enter the theoretical pH value of the buffer (pH value depending on the

    current temperature) using either the , , keys or a keyboard.

    11. Then press .

    12. Wait until the adjustment with the second buffer is finished.

    The voltages of the adjustment data (offset and slope) are displayed.

    13. Press the key to finish the adjustment.

    11.2.3.3 Edit Buffer Data

    If you use a different buffer to "Hamilton Duracal" for the adjustment, the buffer

    data (pH value depending on the temperature) can be edited in this menu.

    In the first two rows of the list the buffers "Hamilton Duracal" are predefined. Thedata of these buffers are already programmed and cannot be edited.

    The rows 3 10 are individually editable. For any buffer the buffer data can be

    entered and stored using a certain name for the new edited buffer.

    Entering buffer data

    1. Select a row between 3 and 10 using the keys and .

    "new buffer" appears on the display.

    2. Press to enter the designation (e.g. "buffer pH 4") of the new bufferusing the , , keys or a keyboard.

    3. Press the key.

    Now the list can be entered (temperature and corresponding pH value).

    4. Use the key to enter the data. Use the key to move the cursor.Use the key to go to the next row. Press to save all entered data.

    5. Answer the question "save new pH buffer?" with the key.

    Deleting buffer data

    1. To delete a newly edited buffer, press the key.

    2. Answer the question "really clear buffer data?" with and press the

    key.

    Caution:At least two pairs of temperature and corresponding pH value that enclose the

    temperature, at which the pH is later measured, have to be entered.

  • 7/22/2019 Alcolyzer Plus Beer

    56/129

    11 Adjusting the Instrument

    56 C27IB02